ADVIA 120/2120/2120i

Configuration
Data Manager
Configuration
2 Document Version / Disclaimer / Copyright

Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.

Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens Healthcare GmbH or one of its affiliates or who are otherwise authorized by Sie‐
mens Healthcare GmbH or its affiliates to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens Healthcare GmbH or one of its affiliates are not entitled to
use this documentation without prior written authority.

Copyright
“© Siemens Healthineers, 2019” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 2 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Table of Contents 3

1 Data Manager 4

1.1 Using the FSE Parameters window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


1.2 Editable WST.INI parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Example of a WST.INI file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Setting up the metering manager. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.4.1 To Set Up the Metering Manager . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.5 Resetting the metering manager Cumulative report . . . . . . . . . . . . . . . . . . 22
1.5.1 To Reset the Cumulative Report . . . . . . . . . . . . . . . . . . . . . . . . . . 22
1.6 Changing the decimal separator transmitted to the Host Computer . . . . . . 23
1.7 Configuring formula (ratio) in TDC test dictionary . . . . . . . . . . . . . . . . . . . . 24
1.7.1 Report Lymphs (Lymph + LUC) . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

2 Changes to Previous Version 26

2.1 Version 02 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.2 Version 01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

3 List of Hazard IDs 28

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 3 of 29 ADVIA 120/2120/2120i


Restricted 08.19
4 1 Data Manager

1.1 Using the FSE Parameters window


1. At the Customize menu, select System Setup, then select Tools Modify.
2. After a moderate delay, the Modify Tools list appears.
3. Double-select FSE Parameters.
4. Type the password BØØMFSEA (letters must be uppercase), then select OK or press
Enter.
5. At the File menu, select Load. Now, select WST.INI, and then select OK or press En‐
ter.
6. While all the parameters appear in the window, you can only change those parame‐
ters preceded by a hyphen (-).
7. To change an editable parameter:
8. Find the parameter you want to edit. (Press ± to get the next editable parameter, or
press to get the previous one.)
9. In the parameter value text box, type the new value, then select OK or press Enter.
10. Continue for each parameter you want to edit.
11. At the File menu, select Save.
12. At the File menu, select Exit.
13. When the Modify Tools list appears, select Exit or select another tool.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 4 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 5

1.2 Editable WST.INI parameters


You can edit the following parameters
(Parameters appear with default values.)

 Eod_Open=Y is the End of Day window available in the Tools tab? Y for Yes or N for No.

Uiw_TenSix=M
Uiw_TenMinusSix=µ
Uiw_TenMinusThree=m
TotalTransmission=Y
Uiw_Custom_Color_Palette=N
Uiw_Insert=N
Uiw_Mode=REJ
Uiw_FormFeed=N
Uiw_RatioFlag=
Uiw_Result_Report_Header=<
Uiw_SystemNumber=1
Uiw_Warn_All_Matching=Y
Uiw_Keep_RequestedTestOnly=Y
Uiw_Enable_Westgard=N
XPosWnd=70
YPosWnd=150
Uiw_DotKeyIsDecimalSep=N
Uiw_Update_WoR&P#=Y
Transmit_Retry_Host_Delay=60
PRINTER=Y
Host_Hnxt_Images_Zoomed=N
Host_Hnxt_Images_WhiteBackground=Y
Ctrlp_Open=Y
Ctrlp_UnmatchedLikeInProcess=N
Tpa_BatchExpansion=Y
Tpa_Open=N
DBS_InProcess=N
Valid_EditControl=N
Valid_See_All_Test_To_Valid=Y
Oen_Open=N
Oen_Original_And_Diluted=Y

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 5 of 29 ADVIA 120/2120/2120i


Restricted 08.19
6 1 Data Manager

MeteringManager_Open=N
Eod_Open=Y
Qc_Open=Y
Calculs_DeleteFlaggedControlResult=N
Calculs_PatientQc_Formula=1
Calculs_OptimizedDeltaCheck=N
Calculs_Keep_WoR&P#=N
Fcu_NbSample=1
Metering_Management=N
Metering_Management_NewResultOnly=N

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 6 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 7

1.3 Example of a WST.INI file


;*******************************************************************
*****;
;;
; * WARNING * ;
;;
; THIS FILE IS TO BE MODIFIED BY SERVICE PERSONNEL ONLY ! ;
;;
; IMPROPER SETTING OF PARAMETERS CAN ADVERSELY AFFECT SYSTEM OPERATION ! ;
;;
;*******************************************************************
*****;

[Btrieve]
Options=/m:32 /p:2048 /b:16 /f:80 /l:1 /n:1 /t:btrieve.trn

[Uiw]
-Uiw_TenSix=M
-Uiw_TenMinusSix=µ
-Uiw_TenMinusThree=m
STATION=1
Deltacheck=DCK
Weeklimit=4
Monthlimit=2
Yearlimit=2
-TotalTransmission=Y
Uiw_CheckStackCounter=0
Uiw_CommentWarning=DOR
Uiw_Custom_Color_Palette=N
Uiw_Erase_Image_Background_Color=192,192,192
Uiw_Format_AllComplete=list,F
Uiw_Format_Complete=list,F
Uiw_Format_CurrentComplete=list,F
Uiw_Format_Hold=list,R
Uiw_Format_Incomplete=list,R
Uiw_Format_Pending=list,F
Uiw_Format_Rerun=list,F

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 7 of 29 ADVIA 120/2120/2120i


Restricted 08.19
8 1 Data Manager

Uiw_Format_Unmatched=list,R
Uiw_HoldStatus=N
Uiw_HR_Palette_1_Color_1=128,128,128
Uiw_HR_Palette_1_Color_2=255,0,0
Uiw_HR_Palette_1_Color_3=255,255,0
Uiw_HR_Palette_1_Color_4=0,255,0
Uiw_HR_Palette_1_Color_5=0,255,255
Uiw_HR_Palette_1_Color_6=0,0,255
Uiw_HR_Palette_1_Color_7=255,0,255
Uiw_HR_Palette_1_Color_8=255,255,255
Uiw_HR_Palette_1_Color_9=192,192,192
Uiw_HR_Palette_1_Color_10=128,0,0
Uiw_HR_Palette_1_Color_11=0,128,0
Uiw_HR_Palette_1_Color_12=128,128,0
Uiw_HR_Palette_1_Color_13=0,0,128
Uiw_HR_Palette_1_Color_14=128,0,128
Uiw_HR_Palette_1_Color_15=0,128,128
Uiw_HR_Palette_1_Color_16=128,128,128
Uiw_IdReader=Y
Uiw_Image=RD
-Uiw_Insert=N
Uiw_Manual_Rerun=H
Uiw_MaxTubeLimit=5000
Uiw_MisreadToMatch=Y
Uiw_UnmatchedToMatch=Y
-Uiw_Mode=REJ

Uiw_Seq_Sep=-
Uiw_Modulo1=999
Uiw_CarNumber1=3
Uiw_Modulo2=10
Uiw_CarNumber2=2
Uiw_Modulo3=0
Uiw_CarNumber3=0

Uiw_NbStation=1
Uiw_PageLength=60
-Uiw_FormFeed=N

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 8 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 9

Uiw_PosMode=Y
-Uiw_RatioFlag=
Uiw_ReceivedResultsFiles=P/H/I
Uiw_AnsweredQueryFiles=P/H/I
-Uiw_Result_Report_Header=<
Uiw_RoutineSampleType=WB
Uiw_SuspendTest=N
-Uiw_SystemNumber=1
Uiw_Treat_IO_Error=Y
Uiw_WarningNbTubeLimit=4800

Uiw_Dispatching=N
Uiw_SidOnly=N
Uiw_InstDownloading=N
Uiw_OtherReportsHeader=
Uiw_SortTest=Y
-Uiw_Warn_All_Matching=Y
Uiw_CumulQc_Formula=1
-Uiw_Keep_RequestedTestOnly=Y
Uiw_Enable_Westgard=N
-XPosWnd=70
-YPosWnd=150
LanguageType=ST
Normal End=Y
Downloading Order Send=
Result Send=
Review_None_For_Control=N
Uiw_Hnxt_Errors=Y
Uiw_TabbedView=Y
Uiw_Hnxt_UserCode=Y
Uiw_Hnext_Adjust_Seconds=N
Uiw_MovAvg_QCCloseOutDelay=30
-Uiw_DotKeyIsDecimalSep=N

Uiw_Update_WoR&P#=Y
Uiw_MovAvg_LastQCCloseOut=26/08/1997
Uiw_DateLastDisplaySplash=27/8/1997
Downloading Order Send=N

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 9 of 29 ADVIA 120/2120/2120i


Restricted 08.19
10 1 Data Manager

Result Send=N

[Transmit]
Transmit_Y_To_Inst=N
Transmit_YToInst_atStartUp=N
Transmit_Ctrl_To_Host=N
Transmit_Ctrl_To_Printer=N
Transmit_Patient_To_Host=N
Transmit_Patient_To_Printer=N
Transmit_PreValidationPrint=N
-Transmit_Retry_Host_Delay=60

[Com]
Com_SrchNxtSple=0
HOST=Y
Inst1=Y
Inst2=Y
Inst3=Y
Inst4=Y
Inst1_Tubeid_Field_Length=7
-PRINTER=Y
Trame_Spy=N
Host_Work_Order_Transfer_Mode=D
Query_To_Host_Timeout=7
Host_Images_Transfer=Y
Patient_To_Host_Image_Configuration=
Ctrl_To_Host_Image_Configuration=
Host_Hnxt_Images_Zoomed=N
Host_Hnxt_Images_WhiteBackground=Y
Host_Send_Asterisk=Y
Inst5=N
Inst6=N

[Ctrlp]
Ctrlp_Delay=10
-Ctrlp_Open=Y
-Ctrlp_UnmatchedLikeInProcess=N

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 10 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 11

[Tpa]
-Tpa_BatchExpansion=Y
Tpa_BatchSize=20
Tpa_Delay=35
-Tpa_Open=N

[Dbs]
DBS_Deviation=0
DBS_FileName=DBS.CPT
-DBS_InProcess=N
Dbs_WriteDelay=30

[Valid]
Valid_Create=Y
Valid_DefaultAccess=IU
Valid_Delay=300
Valid_Display_Asterisk=N
Valid_Display_Image=Y
-Valid_EditControl=N
Valid_Group1=
Valid_Group2=
Valid_Group3=
Valid_Group4=
Valid_Group5=
Valid_Group6=
Valid_Group_Image_CBC/DIFF=0
Valid_Group_Image_RETIC=0
Valid_Kind=ReviewAll
Valid_MorphoFlagsInFlagWindow=N
Valid_PrintButtonFormat=REPORT.PAR
Valid_PrintButtonDelay=3
Valid_Open=Y
Valid_Result_Entry_Mode=I
-Valid_See_All_Test_To_Valid=Y
Conflict_Error=Y
RerunValidation=H
strDilutionString1=NNN0000000000
Valid_Image_Configuration=

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 11 of 29 ADVIA 120/2120/2120i


Restricted 08.19
12 1 Data Manager

[Oen]
-Oen_Open=N
Oen_Options=S,P,N
Oen_OptionalFieldName=
Oen_OptionalFieldValue=
Oen_SampleType=WB
-Oen_Original_And_Diluted=Y

[Tools]
Alarms_Open=Y
Controls_Open=Y
CusEdit_Open=Y
Dic_Update=N
File_Management_Open=Y
ParEdit_Open=Y
RoutineParam_Open=Y
Tables_Open=Y
Tools_Open=Y
Tools_Delay=30000
Tools_Lock=Y
Tools_Screen_Configuration_Open=N
Tests_Open=Y
Texts_Open=Y
Ports_Open=Y
Tools_Spy_View_Open=Y
Print_Dictionaries_Open=Y
FseParameters_Open=N
-MeteringManager_Open=N
-Eod_Open=Y
Image_Configuration_Open=Y

[Qc]
-Qc_Open=Y
Qc_Delay=400

[Calculs]

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 12 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 13

MissingResultReview=Y
NbMaxAutoRerun=1
Calculs_DelayDCK=14D
-Calculs_DeleteFlaggedControlResult=N
Calculs_PatientQc_BatchSize=20
Calculs_PatientQc_ExclBySample=Y
-Calculs_PatientQc_Formula=1
Calculs_PatientQc_MaxWait=0
Controls=7,9,4
Calcul_Open=Y
Calculs_MaxAutoDispRerun=1
-Calculs_OptimizedDeltaCheck=N
Calculs_ResultsOneByOne=N
-Calculs_Keep_WoR&P#=N

[Fcu]
-Fcu_NbSample=1
Fcu_Delay=30000
Fcu_EraseCompleteWorkOrder=N
Fcu_Fast_Delay=3000
Fcu_Open=Y

[Purge]
Purge_Complete=N
Purge_CurrentComplete=N
Purge_EndOfDay_Trigger=4800
Purge_EndOfDay_DeleTableTubes=500
Purge_Hold=N
Purge_Incomplete=N
Purge_Pending=N
Purge_UnMatched=N

[Eod]
Eod_DateBackupBDD=
Eod_TimeBackupBDD=
Eod_MovAvg_RetentionDelay=15
Eod_NbDiskBackupDICT=1

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 13 of 29 ADVIA 120/2120/2120i


Restricted 08.19
14 1 Data Manager

[DebugUiw]
DebugUiw=
ExtraSize=0
Display_Requests=N
Trace_Requests=N
Tracepile=N
Set_Colors=Y

[DataBaseCheck]
DataBaseCheck_Link=Y
DataBaseCheck_TubeCheck=Y
DataBaseCheck_CannotCheck=Y

[AccessPath]
Error_Spy=C:\wst\
FilePath=C:\wst\
ImageFilePath=C:\wst\ima\
Tools=C:\wst\
AccessPath_ReportFile=C:\wst\
HelpFilePath=C:\WST\

[Parametres UIW]
Version=V07.51
Bdd_Dll=B00MBD72.DLL

[Inst1]
-Metering_Management=N
-Metering_Management_NewResultOnly=N

[Printer]
Result_Report_Open=Y
Result_Device=
Result_File=
Report_Image_Configuration=
QCReport_Image_Configuration=
PreValid_Image_Configuration=

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 14 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 15

PrintDedicated_Image_Configuration=

[Crt]
Crt_Delay=50
Crt_Help=N
Crt_NbTimes=10

Parameter Description
Fcu_NbSample=1 determines the maximum number (0 to
9) of sample records for this patient to
be searched if no previous test result is
provided by the host computer for delta
check purposes.
Host_Hnxt_Images_WhiteBackground=Y This parameter determines the back‐
ground color of the images transmitted
to the host computer.
Y System transmits images with a white
background.
N System transmits images with a black
background.
Host_Hnxt_Images_Zoomed=N This parameter determines the size of
the images transmitted to the host com‐
puter.
Y System transmits large images
(420KB).
N System transmits standard images
(120KB).
MeteringManager_Open=N Is the Metering Management window
available in the Tools tab?
Y Yes.
N No.
Do not change this parameter value un‐
less specifically authorized.
Metering_Management=N Is the Metering Management available
for the instrument # (1 to 3) according
to the protection key connected?
Y Yes.
N No.
Do not change this parameter value un‐
less specifically authorized.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 15 of 29 ADVIA 120/2120/2120i


Restricted 08.19
16 1 Data Manager

Parameter Description
Metering_Management_NewResultOn‐ If all test results are retransmitted after
ly=N only one test is inserted or rerun,
should the Metering Management func‐
tion count only the new result?
Y Yes. Only the new test result is coun‐
ted.
N No. All test results are counted.
Do not change this parameter value un‐
less specifically authorized.
Oen_Open=N Is the Order Entry tab available?
Y Yes.
N No.
Oen_Original_And_Diluted=Y When managing the manual dilution,
should the test(s) be run on the diluted
and original samples?
Y Yes. Tests are run on the diluted and
original samples.
N No. Tests are run only on the diluted
samples.
In either case, operator can select or de‐
select the original tests to run for the
sample.
PRINTER=Y Is a printer connected to the system?
Y Yes.
N No.
Qc_Open=Y Is the Quality Control tab available?
Y Yes.
N No.
TotalTransmission=Y Y Yes. The sample results are transmit‐
ted to the Host when the results of all
requested tests are received from the
instrument.
N No. Allows transmission of results
when record returns to the Pending file
for rerun (partial transmission feature).
Tpa_BatchExpansion=Y Are samples identified individually by
SID# after their batch group is released
via the Sample List in the Sample Con‐
trol Panel tab?
Y Yes.
N No.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 16 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 17

Parameter Description
Tpa_Open=N Is the Test Panel displayed in the Sam‐
ple Control Panel tab?
Y Yes. The Ctrlp_Open parameter must
also be Y (yes) to display the Test Panel.
N No. Not opened in the Modify mode.
Do not change this parameter value un‐
less specifically authorized.
Transmit_Retry_Host_Delay=60 Specifies the waiting time after 4 con‐
secutive NACKs or a Z1 message (tube
protected on the Host) before the sam‐
ple results ("R" message) are retransmit‐
ted to the host computer. Value is in
seconds (30 s minimum).
During this time, the sample record
moves to the Current Complete status.
Uiw_Custom_Color_Palette=N Do not change this parameter value un‐
less specifically authorized.
Uiw_DotKeyIsDecimalSep=N Determines which character appears
when you press the numeric keypad dot
key (.) on the Review / Edit, Format,
Control Dictionary, and Test Diction‐
ary windows.
Y Yes. The character is determined by
the Decimal symbol box on the Number
tab of the Windows NT Regional Set‐
tings Control Panel.
N No. The period character (.) appears.
Uiw_Enable_Westgard=N This parameter enables the Westgard
Rules functions on the Sample Control
Panel tab.
Y Yes. Westgard Rules enabled.
N No. Westgard Rules disabled.
Do not change this parameter.
Uiw_FormFeed=N Y Yes. Insert a blank page between the
different sections in the printed reports
for File Management, QC Summary, and
Dictionary Printout.
N No. Do not insert a blank page.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 17 of 29 ADVIA 120/2120/2120i


Restricted 08.19
18 1 Data Manager

Parameter Description
Uiw_Insert=N Y Yes. A test can be inserted using "In‐
sert" option in the Disposition dialog
box of the Review / Edit tab. The "I" dis‐
position flag is sent to the host comput‐
er for this result.
N No. A test cannot be inserted.
Applicable only if Uiw_Keep_Requested‐
TestOnly parameter is set to N.
Uiw_Keep_RequestedTestOnly=Y Y Yes. Only show requested tests on the
Review / Edit tab.
N No. Show all tests run for sample.
Uiw_Mode=REJ Select the desired response if the instru‐
ment retransmits the sample results
with the same SID without requesting a
rerun.
n REJ
Reject. The results are rejected and a
low error is triggered. After the first
aspiration, any additional aspirations
with the same SID are ignored.
IMPORTANTSystems that will oper‐
ate in Position Mode must be set to
Uiw_Mode=REJ.
n OVR
Overwrite. The previous results are
replaced.
n HIS
History. The previous results move
to the Prev column on the Review /
Edit tab.

Uiw_RatioFlag= Select the character that appears in the


Flag column of the Review / Edit tab for
a ratio if one of its component tests is
flagged. The operator can set this pa‐
rameter using the FLAG CHARACTER
FOR "RATIO" TEST field on the Customer
Parameters window.
Uiw_Result_Report_Header=< Select the character that indicates the
end of the subheader in the result re‐
port parameter files. Any text preceding
this character appears in the heading on
each page of the printed report.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 18 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 19

Parameter Description
Uiw_SystemNumber= Enter the workstation number (1 to
999) that is sent to the host computer.
Default is a space.
Uiw_TenMinusSix=µ Specifies the character that expresses .
Uiw_TenMinusThree=m Specifies the character that expresses .
Uiw_TenSix=M Specifies the character that expresses .
Uiw_Update_WoR&P#=Y When enabled, this parameter sets the
system to update the workorder in the
instrument database with the Rack and
Position # received from the analyzer.
The system uses this setting only when
Id_Reader=Y and Sid_Only=Y.
Y Yes. System updates the workorder
with the R&P#.
N No. System does not update the
workorder
Uiw_Warn_All_Matching=Y Do you want to be warned each time
you correct an unmatched sample?
Y Yes. A warning message and a low er‐
ror occur each time you match a sam‐
ple.
N No. A warning message and a low er‐
ror occur only if you match a sample us‐
ing a different SID#.
Valid_EditControl=N Should the Edit disposition be made
available for a control result in the Re‐
view / Edit tab?.
Y Yes.
N No.
Valid_See_All_Test_To_Valid=Y A maximum of 21 test results are visible
at one time on the Review / Edit tab. If
more than 21 tests are ordered, must
the operator view all the results before
the sample can be validated?
Y Yes. Operator must view all results be‐
fore the OK button is activated.
N No. It is not mandatory to view all re‐
sult to activate the OK button.
XPosWnd=70 Specifies the horizontal position of the
window from the left edge.
Do not change this parameter value un‐
less specifically authorized.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 19 of 29 ADVIA 120/2120/2120i


Restricted 08.19
20 1 Data Manager

Parameter Description
Calculs_DeleteFlaggedControlResult=N Should the Delete disposition be auto‐
matically assigned to a flagged control
result in the Review / Edit tab?.
Y Yes. Each flagged control result is au‐
tomatically assigned a Delete disposi‐
tion.
N No. The operator decides the disposi‐
tion.
D Dictionary. A flagged control result is
assigned a Delete disposition only if
"Delete" is selected for the flag in the
Alarm Dictionary.
Calculs_OptimizedDeltaCheck=N Y Yes. If the current and previous results
are outside the rerun range but the Del‐
ta Check threshold has not been
reached, the current result is automati‐
cally validated, and is not displayed in
the Review / Edit tab.
N No. If these conditions are not satis‐
fied, the current result must be valida‐
ted.
Calculs_PatientQc_Formula=1 Determines the formula that calculates
the Patient QC statistics.
1 Bull formula (moving average)
2 arithmetic mean formula
other Patient QC calculations are not
performed.
Ctrlp_Open=Y Is the Sample Control Panel tab availa‐
ble?
Y Yes.
N No.
Ctrlp_UnmatchedLikeInProcess=N This parameter is not used.
DBS_InProcess=N Do not change this parameter value un‐
less specifically authorized.
YPosWnd=150 Specifies the vertical position of the
window from the top edge.
Do not change this parameter value un‐
less specifically authorized.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 20 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 21

1.4 Setting up the metering manager

1.4.1 To Set Up the Metering Manager


1. Use FSE Parameters in Tools Modify to edit the WST.INI file parameters as follows:
MeteringManager_Open = Y
Metering_Management = Y
Metering_Management_NewResultOnly = Y
2. Set the system to back up the Metering Manager data during the End of Day backup
procedure.
a) Exit the ADVIA software.
b) Log in to Windows using a field service password.
c) When the ADVIA Shell opens, select Explorer.
d) In Explorer, double-select the icon for the WST file.
e) Double-select the icon labeled Dic.par.
The Open with dialog box opens.
f) Scroll down the list of programs and select NOTEPAD.
g) Select OK.
h) When the document Dic.par opens, type Histo.dat on a new line.
i) At the Notepad window File menu, select Save.
j) Close Notepad and restart the ADVIA system.
3. Log on and begin normal system operation.
The Tools list will include Metering Manager in both View and Modify.
C IMPORTANT
You cannot use the Metering Manager window until you have run at least one sample on
the system. If you try to open the window before you have run a sample, the system will
display an error message.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 21 of 29 ADVIA 120/2120/2120i


Restricted 08.19
22 1 Data Manager

1.5 Resetting the metering manager Cumulative report

1.5.1 To Reset the Cumulative Report


1. View the Cumulative Report.
2. (Optional) Select Print on the File menu to print a copy of the Cumulative Report.
3. Select the parameters you want to reset. Click the first entry you want to delete and
do one of the following:
- Hold down the Shift key while you click the last consecutive entry you want to de‐
lete.
- Hold down the Ctrl key while you click each additional entry you want to delete.
4. At the File menu, select Reset.
A password box appears.
5. Enter the password RESET (use all capital letters) and select OK.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 22 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 23

1.6 Changing the decimal separator transmitted to the Host


Computer
The decimal separator is normally specified in Windows Regional Settings. Sites can over‐
ride that setting and transmit a different character if they use Host Spec. 79 protocol for
data transmission to the LIS

1. Shut down the ADVIA software.


2. Log on to Windows using a field service password.
3. At the ADVIA Shell, select Explorer.
4. At the View menu, select Details.
5. At the View menu, select Options.
6. At the View tab, uncheck Hide file extensions for known file types. Select OK.
7. In the directory C:\Wst, double-click the Lis.ini icon The file will open in Notepad.
8. Place the cursor under the last line, Host_Delay_Before_Retry=0.
9. Type Host_Decimal=<character> and press the Enter key. This is the line that speci‐
fies the new decimal separator.
10. Type Host_Thousand=<character>
C IMPORTANT
The characters you enter for Host_Decimal and Host_Thousand must not be the
same.
Example: If you want the decimal separator to be a comma, the last lines of the Lis.ini
file should look like this:
Host_Delay_Before_Retry=0
Host_Decimal=,
Host_Thousand=.
11. At the File menu, select Exit. At the message Do you want to save the changes, se‐
lect Yes.
12. Close Explorer and restart the ADVIA software.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 23 of 29 ADVIA 120/2120/2120i


Restricted 08.19
24 1 Data Manager

1.7 Configuring formula (ratio) in TDC test dictionary

1.7.1 Report Lymphs (Lymph + LUC)

1.7.1.1 Configure Test Dictionary


The following is an example of adding a calculated ratio test:

1. At the Customize menu select System setup and then select Tools Modify.
 The Modify Tools list appears.
2. Select Test Dictionary.
3. Type the name of the sum (LY+LU) in the Code field and then press Enter.
4. Select WB in the Type field.
5. Leave the Test Number field blank.

 A test number is not added when performing a ratio test.

6. Type the next available host number in the Host Number field.

 The host number should be set to an available host number that is being used by the
site. Host numbers 21 and 15 are not available if the defaults are being used.

7. Select sum of lymph and luc in the name field.


8. Select the Auto Request check box.
9. Type [%LYMPH]+[%LUC] in the Formula field.
10. Select Ok to save settings.

1.7.1.2 Configure Table Dictionaries

1. At the Customize menu select Tools Modify and then select Table Dictionary.
 The Modify Tools list appears.
2. Select Other and then select ?.
3. Select WB.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 24 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Data Manager 1 25

4. Add %TLYMP and #TLYMP; delete LUC and then select OK.
Fig. 1:

5. Select Profiles and then select ?.


6. Select C/D.
7. Add #TLYMP and %TLYMP; delete LUC. and then select OK.

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 25 of 29 ADVIA 120/2120/2120i


Restricted 08.19
26 2 Changes to Previous Version

2.1 Version 02
Change No. Changes
CHG0052921 Added section 1.7, Configuring formula (ratio) in
TDC test dictionary, as per SKB 29200.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 26 of 29 © Siemens Healthineers, 2019


08.19 Restricted
Changes to Previous Version 2 27

2.2 Version 01
Initial creation of document

© Siemens Healthineers, 2019 DCHT-A04.843.06.02.02 Page 27 of 29 ADVIA 120/2120/2120i


Restricted 08.19
28 3 List of Hazard IDs

There are no Hazard IDs in this document.

ADVIA 120/2120/2120i DCHT-A04.843.06.02.02 Page 28 of 29 © Siemens Healthineers, 2019


08.19 Restricted
- Restricted - All documents may
only be used by authorized person‐
nel for rendering services on Sie‐
mens Healthcare Products. Any
document in electronic form may
be printed once. Copy and distribu‐
tion of electronic documents and
hardcopies is prohibited. Offenders
will be liable for damages. All other
rights are reserved.

siemens-healthineers.com

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens-healthineers.com

Print No.: DCHT-A04.843.06.02.02 | Replaces: DCHT-A04.843.06.01.02


Doc. Gen. Date: 08.19 | Language: English
© Siemens Healthineers, 2019

You might also like