Check-list de maintenance préventive

Générateur de dialyse NIPRO

N° Description Fréquence Durée Vérification

1 Extemals, connection tubes, and cables


Conditions of connection tubes (bending, etc.)
Connection to the drain pipe
Presence of damage and connection of cables 3 mois 5 min
Conditions of the grounding conductor
Damage/deformation that interfere with functions
Conditions of the light indicator
2 Water supply, concentrate feeder, and drainage areas
Conditions of tubes (bending, etc.)
Movement of the solenoid valve
Movement of the reducing valve 10
Check of the degassing part
3 mois
min
Movement of the degassing tank
Conditions of rinse ports
Stain, leakage, or blockage in filters
3 Closed line area
Movement of the solenoid valve
Flow meter display
Movement and leakage in concentrates/UF pumps
3 mois 5 min

Movement during gas purge


Movement and adjustment during BYPASS
4 Disinfectant/acid solution (option) feeder areas
Movement of the solenoid valve
Residual disinfectant/acid solution (option)
3 mois 5 min

Conditions of the tube (bending, etc.)


5 Blood pump area
Loosening and abnomal sound in the rotating part 3 mois 5 min
Consistency between displayed and actual flow volumes
6 Heparin pump area
Movement of the heparin pump 3 mois 5 min
Loosening of the heparin pump pulley
Consistency between displayed and infused volumes
7 Movements of monitoring and indication alarmns
15
Temperature-related alarms (calibration of temperature indication) 3 mois
min
Dialysate concentration alarm (calibration of concentration indication)
Venous pressure alarm (calibration of the venous pressure gauge)
Arterial pressure alarm (calibration of the arterial pressure gauge
Dialysate pressure alarm (calibration of the dialysate pressure gauge
Blood leak alarm (movement of the blood leak detector)
Tube error alarm (movement of the air bubble detector)
Insufficient water supply
Alam buzzer
Blood leak alam (calibration for blood-Heak detecion)
7.Movements of monitoring and indication alarmns
Temperature-related alarms (calibration of temperature indication)
Dialysate concentration alarm (calibration of concentration indication)
8 Variable control of dialysate flow Assy and hi Flow Assy (Option)
Filling flow rate (880 to 950 mL/min)
Filling complete pressure 0.113 to 0.118 MPa (1.15 to 1.2 kgficm)
Dialysate flow rate 10
Variable control of dialysate flow Assy and hi Flow Assy (Option)
3 mois
min
Filling flow rate (880 to 950 mL/min)
Filling complete pressure 0.113 to 0.118 MPa (1.15 to 1.2 kgficm)
Dialysate flow rate
1
Durée totale 1 H 30min