DSC New Manual
DSC New Manual
GT-DSC-300
Operation
Instructions
1
Directory
1. Instrument Introduction ____________________________________ 1
2
1. Instrument Introduction
The Differential scanning calorimetry (DSC) technique has been widely
used. The differential scanning calorimeter is both a routine quality test
tool and a research tool. Measmeasures the temperature, heat flow
associated with the thermal transition inside the material. Our
instrument is a heat flow type differential scanning calorimeter, with
good repetition and high accuracy, which is especially suitable for
accurate measurement of specific heat. The equipment is easy to calibrate,
uses a low melting point, is fast and reliable, and has a very wide range
of applications, especially in material research and development,
performance testing and quality control. The properties of materials,
such as glass transition temperature, cold crystallization, phase
transition, melting, crystallization, product stability, curing /
crosslinking, oxidation induction period, etc., are the research fields
of differential scanning calorimetry, and different models are selected
according to the experimental parameters and experimental requirements.
The sample and parameters were placed into the crucible separately
and placed in the furnace for programmed heating to change the temperature
of the sample and parameters. If the heat capacity of the reference and
the sample are the same, and the sample has no heat effect, the temperature
difference is almost "zero", and a smooth curve is obtained. As the
temperature increases, the sample has a thermal effect, and the parameter
does not produce a thermal effect, the temperature difference between the
two, in the DSC curve, the larger the temperature difference, the greater
the temperature difference changes, the more the number of peaks. Peak
1
up are called exothermic and peaks down are called heat absorption.
2. Instrument principle
Material is often accompanied by thermal effects during physical and
chemical changes, and heat release and heat absorption reflect the thermal
enthalpy of matter. A differential heat analyzer is a function of
measuring the temperature or time between the test sample and the
reference under the same thermal condition.
After the sample and the reference are put into the crucible, they
heat up at a certain rate. If the thermal capacity of the reference and
the sample is roughly the same, the ideal scanning calorimetric analysis
2
map can be obtained.
3
3. Instrument characteristics
1.New furnace structure, better resolution, resolution and baseline
stability.
4
4. Instrument interface
4.1.1 The Initial Status key is used to view the ambient temperature,
sample temperature, etc.
5
4.1.3 The Device Information key displays the device information. The
administrator channel is used for the internal personnel calibration
temperature.
4.1.4 The Start Run key to display the current data information after
the operation on the computer software.
6
5. Software description
5.1 Open the software, click [New] or [New] shortcut under the menu bar
5.2 After clicking "new", you will transfer to a new window, in the new
and you will hear a beep.Note two experiments, the sample name can not
be the same, otherwise the last data, resulting in the loss of the last data.as
7
The experimental parameters are set as follows:
5.2.1 "Parameter setting of oxidation induction period experiment" as
shown in the figure: (Stage 1 can choose constant temperature time 5-10
minutes, scanning rate 20, the end temperature is 190-210℃, commonly used
as 200℃.Phase 2 scanning rate 0, the end temperature is the same as phase
1, the time should be greater than the sample OIT time of more than 10
minutes.When the sample time is unknown, it can be set to 150 or
200min.Test Type Selection OIT)
8
After selecting the automatic mode, the instrument
9
5.2.3 "Parameter setting of two or more section lifting temperature"
(select Not OIT according to the estimated parameter setting of the
sample.) as illustrated in following figure:
10
11
5.3 After the software setting is complete, click [Connect
instrument] and start key " " in the upper left corner of the
to the set program, and the software will record the data in
5.4 First, save the map first to prevent loss. You can also use
12
followed by the following analysis.as illustrated in following
figure:
13
5.4.2 Oxidation induction analysis process: click the map to make it green,
that is, the selected map, click [] - [analysis in the oxidation induction
period] -about two black lines, drag the left analysis line in the front,
the right analysis line in the back end, after selection, click
[application], [sure], then click the curve, make it rose red,
analysis.The analyzed maps are shown in the figure below
14
15
Software OIT automatic analysis function, the instrument operation is finished,
directly appears in the following figure:
5.4.3 Glass analysis operation: click the map to make it green, that is,
select the map, click [analysis] - [glass transformation] in the
taskbar-about two black lines, drag the left analysis line in the front,
the right analysis line in the back end, select, click [application],
[sure], then click the curve, make it blue, analysis.The analyzed maps
are shown in the figure below
16
17
5.4.4 Initial melting point, final melting point analysis: click the map
to make it green, that is, selected map, click the task bar [analysis]
- [initial melting point] or [final melting point] -appear about two black
lines, drag the left analysis line in the front, right analysis line in
the back end, select, click [application], [sure], then click the curve,
make it blue, analysis.The analyzed map is shown below:
18
19
5.4.5 Specific heat capacity analysis: open the three maps of the test,
click the map to make it green, that is, select the map, right click, select
the corresponding items (baseline, standard sample, sample), click
[Analysis] - - [specific heat-specific heat method] in the task bar-
-enter the temperature to be compared-click calculation. The analyzed map
is shown as follows below:
20
21
5.5 For all analyzed maps, click [File] - [Save as status T]
below:
22
Operation procedure for temperature calibration:
1、 Enter "456" in the administrator channel, click input, and enter the temperature correction
interface. The following below:
23
2, at the correction point 2, you can input the corresponding melting point value of indium; the
correction point 3, input the corresponding melting point value of tin, click Save. Note that
correction point 3 needs to be higher than correction point 2. The following below:
The following table shows the melting points and theoretical thermal
enthalpy values of the commonly used calibration.
24
Theoretical
Standard Theoretical
enthalpy of melting
substances melting point, ℃
heat is J/g
Inum In 156.6 28.6
Tin Xi 231.9 60.5
Zinc-based Zn 419.5 107.5
7. Instrument application
25
Thermal enthalpy is a state function of the energy of a matter system
that is numerically equal to the product of its internal energy U plus
pressure P and volume V, namely H=U + PV.Under certain conditions, the
internal energy and enthalpy changes can be measured from the heat
transfer between the system and the environment.In the absence of other
work, all the heat absorbed by the system in the isovolume process is used
to increase the internal energy, and all the heat absorbed by the system
in the isotropic process is used to increase the enthalpy. Because the
general chemical reactions are mostly carried out under the isovoltage,
the enthalpy is of more practical value.In the DSC curve, we can obtain
the melting heat enthalpy of the sample by calculating the peak area,
namely the BCD in Fig.
26
in the real-time data column. If the measured phase change temperature
is close to the actual temperature of the sample, the thermal enthalpy
value can be recorded and the instrument coefficient can be calculated
as the coefficient of the instrument.Set the following figure:
27
conditions. Therefore, when marking the glass transition temperature of
a certain polymer, the measured method and conditions shall be indicated.
28
dish, the maximum temperature of the test shall not exceed 550 ℃.
6. When preparing DSC samples, do not sprinkle the samples on the edge
of the crucible to avoid polluting the sensor and damaging the instrument.
The bottom and all external surfaces of the crucible shall be free of
samples and impurities to avoid affecting the experimental results.
29
11. Do not use hard objects to clean the sample holder and test area,
so as to avoid irreversible damage to the instrument.
14. do not adjust the flow of purified gas in the process of collecting
data, because the slight change of gas flow will affect the DSC curve.
15. After the experiment, be careful of the furnace cover of DSC. When
the temperature drops below 100 ℃, handle it gently with tweezers to
avoid scalding or damage to the furnace cover.
17. Disconnect the data cable and close the software before closing
the instrument. To prevent online and communication errors( This problem
can be found in XP and SP3 systems (other systems have not been tested).
Solution:
30
9.Bating list
Host 1
The U disk 1
Data line 2
Power cord 1
The Aluminum
200
Crucible
Metal cover 3
10A Fuses 5
Eear ball 1
Thracheal 2
Instructions 1
Warranty form 1
Certificate of
1
qualification
31
产品保修卡
Product Warranty Card
______________________________
维修服务记录
Repair Service Record
产品名
产品型号 产品编号
称
Model Product No.
Product
发票号 购买日期 免费保修期
12 个月
Invoice Date of Period of
12Months
No. purchase Warranty
日期 记录 维修人
Date Description Serviced by
32
User Instructions
In order to ensure the user interests, I have established a quality
Card.
33