CALIBRATION CERTIFICATE
Model No: INTERACOUSTICS AD226 Serial No. 1010598 Date: 07/10/2021 Technician: Nurul Location: Poliklinik An Nur
Due Date: 07/10/2022 Bangi Selangor
Transducer Information
Output Type Channel Serial No. Information Confirmed Test Equipment
Earphone DD45 Left WT445674 Equipment Details Y Bruel & Kjaer 1/2'' Sound Level Meter Type 2250
Earphone DD45 Right WT446262 Data Entered Y G.R.A.S 40EN 1” Ext Polarized Pressure Microphone
Insert 5A Left NA Bruel & Kjaer Artificial Mastoid
Insert 5A Right NA
BC B-71 Left/Right 163055032 Reference Standard
HF HDA200 Left/Right NA ISO 389 - 1 : Pure Tone Audiometer
FF1 VRA Left NA ISO 389 - 2 : Insert Phone Pure Tone
FF2 VRA Right NA ISO 389 - 3 : Pure Tone Bone Conductor
ISO 389 - 4 : Narrow Band Masking
Headphones DD45
Left Right
Frequency Att. Target Original Original
New Value Adjustment Variance New Value Adjustment Variance
Value Value
Hz dB HL dB SPL dB SPL dB SPL dB SPL dB SPL dB SPL dB SPL dB SPL dB SPL
125 40 87.5 87.3 87.3 0.0 -0.2 87.3 87.3 0.0 -0.2
250 60 87.0 86.7 87.2 0.5 0.2 86.7 87.2 0.5 0.2
500 75 88.0 87.4 87.9 0.5 -0.1 87.9 87.9 0.0 -0.1
750 80 86.5 86.0 86.5 0.5 0.0 86.1 86.3 0.2 -0.2
1000 80 86.0 85.7 86.2 0.5 0.2 85.8 85.8 0.0 -0.2
1500 80 88.0 87.9 87.9 0.0 -0.1 88.1 88.1 0.0 0.1
2000 80 88.0 88.0 88.0 0.0 0.0 88.0 88.0 0.0 0.0
3000 80 88.0 87.9 87.9 0.0 -0.1 87.9 87.9 0.0 -0.1
4000 80 89.0 89.1 89.1 0.0 0.1 89.2 89.2 0.0 0.2
6000 65 85.5 85.1 85.6 0.5 0.1 84.5 85.6 1.1 0.1
8000 75 87.0 85.8 87.3 1.6 0.3 86.1 87.1 1.0 0.1
Noise Calibrated
Narrow Band Y
Speech Y
White Noise Y
Page 1/2
Bone Conductor B71
Left/Right
Frequency Att. Target Original
New Value Adjustment Variance
Value
Hz dB HL dB VFL dB VFL dB VFL dB VFL dB VFL
250 10 58.5 55.0 58.5 3.5 0.0
500 20 59.1 60.7 59.2 -1.5 0.1
750 30 60.5 60.2 60.6 0.4 0.1
1000 35 59.4 59.8 59.8 0.0 0.4
1500 40 59.5 58.3 59.8 1.5 0.3
2000 45 59.5 63.2 59.2 -4.0 -0.3
3000 50 64.0 65.3 64.3 -1.0 0.3
4000 40 54.5 57.4 54.4 -3.0 -0.1
6000 30 43.0 43.7 43.2 -0.5 0.2
8000 20 34.0 38.6 34.2 -4.4 0.2
Noise Calibrated
Narrow Band NA
Speech NA
White Noise Y
White Noise Y
Prepared By:
Nurul Syazwani
Technical Support
Page 2/2
* The audiometer has been calibrated according to Peraturan-peraturan keselamatan dan kesihatan pekerjaan (Pendedahan Bising) 2019
P.U (A) 60/2019 dan Tataamalan Industri (ICOP) bagi Pengurusan Pendedahan Bising Pekerjaan Dan Pemeliharaan Pendengaran 2019