Securespan
Securespan
Laminoplasty System
Aesculap Spine
Aesculap® SecureSpan®
Content
A Foreword 3
B Features 4
C Surgical Technique 6
D Ordering Information
D1 Implants 16
D2 Instruments 17
2
Foreword A
Cervical Laminoplasty is a surgical procedure originating in Japan
that has gained acceptance in the United States and other parts
of the world. Originally targeted toward the Asian population
susceptible to Ossification of the Posterior Longitudinal Ligament
(OPLL), it has expanded into treatment for spinal stenosis.
3
Aesculap® SecureSpan®
Features
4
Implants
5
Aesculap® SecureSpan®
Surgical Technique C
6
Fig. 1 Fig. 2
Approach
7
Aesculap® SecureSpan®
Surgical Technique
Fig. 3 Fig. 4
Fig. 3a Fig. 4a
Preparation Laminotomy
❚ Prepare the open side trough along the junction of the lamina ❚ Remove soft tissue at the caudal and cranial endpoint
and lateral mass. Use a high speed burr to transect the lamina as needed.
1 cm lateral to the midline. Take care to not contact the under- ❚ Utilizing the lamina lifter, lift the lamina away from the
lying dura. spinal canal (Fig. 4).
Decorticate the lamina on the contralateral side by scoring.
Using a high speed drill, create a hinge. The hinge trough
should be slightly more lateral than the lamina-facet interval
with a depth to the dorsal cortex only (Fig. 3).
Note: Note:
Cutting too far lateral will lead to extensive burring of The laminae are opened slowly, creating greenstick
the facet without detachment of the lamina, whereas fractures of the thinned laminae on the hinge side.
cutting too medial will leave a lateral bony edge that
may impinge the spinal cord as it drifts backwards.
8
Fig. 5 Fig. 6
Fig. 5a Fig. 6a
❚ Using the plate holder, pick plate out of the caddy (Fig. 5). Plate sizes are in terms of laminar gap.
❚ 4 mm brown
❚ 6 mm turquoise
❚ 8 mm yellow
❚ 10 mm light blue
❚ 12 mm grey / silver
Using the plate benders LX185R, to bend plate to the desired
angles (Fig. 6).
Caution:
SecureSpan® plates should always be bent in one direction
only. SecureSpan® plates should never be bent back.
9
Aesculap® SecureSpan®
Surgical Technique
Fig. 7 Fig. 8
Fig. 7a
❚ To load screw onto the screwdriver, place screwdriver ❚ Place the stabilization block, e.g. autograft, in the preparation
over screw head and turn screwdriver’s thumbwheel device ME739. Attach the plate to the stabilization block
until threaded internal shaft is fully engaged into the by drilling a hole and inserting the self-tapping 5 mm center
threaded receptacle in the screw head (Fig. 7). screw FG923T (Fig. 8).
10
Fig. 9
Fig. 9a
Note:
SecureSpan® should always be used with a stabilization
block.
11
Aesculap® SecureSpan®
Surgical Technique
Fig. 10 Fig. 11
Drill
❚ Drill holes into the proper location on the lateral mass. ❚ An alternative drill and drill guide with an integrated depth
Next drill holes on the lamina (Fig. 10). stop is provided (Fig. 11).
Caution:
Ensure safe drill trajectory.
12
Fig. 12 Fig. 13
Secure Plate
❚ To load plate screw onto the screwdriver, place screwdriver ❚ Screw plate onto lamina at the predetermined hole location
over screw head and turn screwdriver’s thumbwheel until (Fig. 13).
threaded internal shaft is fully engaged into the threaded
receptacle in the screw head (Fig. 12).
13
Aesculap® SecureSpan®
Surgical Technique
Fig. 14 Fig. 15
Fig. 14a
Explantation
❚ Revision screws are provided if needed (Fig. 14). ❚ The Aesculap® SecureSpan® laminoplasty screwdriver has the
capability to explant the laminoplasty screw incorporated into
❚ Repeat previous steps at adjacent levels as its design.
necessary (Fig. 15). The locking feature of the internal threaded screwdriver shaft
allows for explantation of the screws even if the screw spins
freely in the bone.
Explantation is the reverse of the implantation procedure.
14
15
Aesculap® SecureSpan®
Screws
FG915T Ø 2.0 mm, Length 4 mm Self Drilling / Tapping
FG916T Ø 2.0 mm, Length 5 mm Self Drilling / Tapping
FG917T Ø 2.0 mm, Length 6 mm Self Drilling / Tapping
FG918T Ø 2.0 mm, Length 8 mm Self Drilling / Tapping
FG919T Ø 2.0 mm, Length 10 mm Self Drilling / Tapping
FG920T Ø 2.0 mm, Length 12 mm Self Drilling / Tapping
Screws
FG925T Ø 2.4 mm, Length 5 mm Revision Screw
FG926T Ø 2.4 mm, Length 6 mm Revision Screw
FG927T Ø 2.4 mm, Length 7 mm Revision Screw
FG928T Ø 2.4 mm, Length 9 mm Revision Screw
FG929T Ø 2.4 mm, Length 11 mm Revision Screw
FG930T Ø 2.4 mm, Length 13 mm Revision Screw
Screw
Allograft Screw for stabilization block
FG923T
(to fix spacer on plate)
16
Ordering Information – Instruments D2
Instruments
Art. no. Description
FG847R Lamina Lifter
17
Aesculap® SecureSpan®
Instruments
Art. no. Description
FG856SU Drill, Length 4 mm
FG857SU Drill, Length 6 mm
FG858SU Drill, Length 8 mm
FG859SU Drill, Length 10 mm
FG860SU Drill, Length 12 mm
FG848R Screwdriver
18
Instruments
Art. no. Description
FG849R Graft Holder
19
The main product trademark ‘Aesculap’ and
the product trademark ‘SecureSpan’ are
registered trademarks of Aesculap AG.