MAGNETOM
MR

Software
System

Software Installation syngo MR A30


(NUMARIS/4VA30A)

© Siemens 2007
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: MR-000.816.26.01.02 English


Replaces: n.a. Doc. Gen. Date: 04.07
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

MAGNETOM MR-000.816.26.01.02 Page 2 of 22 Siemens


04.07 CS SD MR
Table of Contents 3
0Table of Contents

1 _______ General ________________________________________________________ 4

syngo MR system software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 _______ Installing MRC / MRC+MRSC ______________________________________ 5

Overview of installing software MRC/MRC+MRSC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5


BIOS update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Loading BIOS defaults for Celsius R630 / R640 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
NUMARIS/4 installation (host without operating system) . . . . . . . . . . . . . . . . . . . . . . . . . 7
NUMARIS/4 reinstallation (on booting HW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3 _______ Installation of Imager____________________________________________ 10

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Final checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

4 _______ Site-specific Data_______________________________________________ 12

Backup site-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12


Restore site-specific data (manually after SW installation) . . . . . . . . . . . . . . . . . . . . . . 13
Restore site-specific data (on running system) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

5 _______ Configuration __________________________________________________ 15

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

6 _______ Appendix _____________________________________________________ 16

Installing patch CD VA30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16


Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Preparation for Service Pack installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Service pack installation at the MRIR VA30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Service Pack installation at the MRC/MRSC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Siemens MR-000.816.26.01.02 Page 3 of 22 MAGNETOM


04.07 CS SD MR
4 General
1-
syngo MR system software
1General

Introduction 0

The installation and service of equipment described herein is to be performed by qualified


personnel employed or authorized by Siemens Medical Solutions or one of its affiliates.
Assemblers and other personnel not employed by nor directly affiliated with nor authorized
by Siemens Medical Solutions or one of its affiliates` technical services are directed to con-
tact the local office of Siemens Medical Solutions or its affiliates before attempting installa-
tion or service procedures.

Prerequisites 0

To configure the syngo-based MR system, the following training is required in combination


with this software manual:
• syngo training / syngo MR training
• Network training (TCP/IP)
• DICOM training

NOTE All passwords required to install and configure the system


are available in the “CS Knowledge Base“
http://www-tdkb.med.siemens.de under “Password List“
(except for the service key (variable password), which is han-
dled by the districts).

Abbreviations 0

EDB Event Data Base


n.a. Not applicable
SRS Siemens Remote Service
QDB Quality assurance Data Base

MAGNETOM MR-000.816.26.01.02 Page 4 of 22 Siemens


04.07 CS SD MR
Installing MRC / MRC+MRSC 5
2-
Overview of installing software MRC/MRC+MRSC
2Installing MRC / MRC+MRSC

The following table provides a number of important milestones for the software installation.

MRC (save set available) MRC (no valid save set) MRSC
Install software Install software Install software
Install software at the Imager n.a.
Restore site-specific data Initial configuration Restore site-specific data /
or import clone package
from the MRC
Check system language Configure the system Check system language
and weight and weight
Check if service packs are Perform system tune-up, Check if service packs are
available for this software quality assurance, and the available for this software
version installation protocol version
Check the image orienta- Check system language and n.a.
tion weight
Check if Update Instructions with service packs are available for this software version

NOTICE Verify the software baseline label with the Help, Info function.
The correct software label for the released syngo MR A30 is:
¹ COEM VD20F
¹ syngo VE31G
¹ N4_VA30A_LATEST_20070315

Siemens MR-000.816.26.01.02 Page 5 of 22 MAGNETOM


04.07 CS SD MR
6 Installing MRC / MRC+MRSC

BIOS update 2.1

The BIOS update optimizes the setting of NUMARIS/4 for your host type, sets the
NUMARIS/4 BIOS password and boot device order

Loading BIOS defaults for Celsius R630 / R640 0

• Reboot the system (power on the Host).


• During the boot, press F2 to access the BIOS setup
• Enter the BIOS password.
• Check date / time, correct if necessary. Installations in Asia should change the date
going back 3 days for the installation to avoid problems with the licenses.
• Check in Security that the entry Start from IDE CD-ROM is set to Enabled.
• Select the Boot tab.
• Highlight the Boot Device Priority and press Enter.
• Move the CD-ROM entry to the top of the list.
• Press F9 and accept (Yes) by pressing Enter.
• If you see an ! next to the entry, press SHIFT+ Insert BIOS-CD Vxxx:
xxxxxxx-MR005BIOS-CD in to the DVD/CD-ROM drive.
• Press F10 and accept "Yes" by pressing Enter.
• The BIOS update will run, wait until the “Ready for Software Installation“ message is
displayed.
• Remove the BIOS CD after the automatic eject.
• Restart the system by pressing Ctrl + Alt + Del simultaneously.

MAGNETOM MR-000.816.26.01.02 Page 6 of 22 Siemens


04.07 CS SD MR
Installing MRC / MRC+MRSC 7

NUMARIS/4 installation (host without operating system) 2.2

The following table describes how to install software on hardware without an operating sys-
tem.
• Perform the “BIOS update“

Action MRC MRSC


Insert the “Numaris/4 .... DVD 1/2“ in to the DVD/CD-ROM (R:) drive.
Insert the “Master Backup-CD“ into the CD/DVD burner drive DVD/CD
burner (S:)
... automatic boot will follow
A pseudo-graphic user interface is displayed with the warning that all
data will be lost after pressing a key for confirmation. At this point, press
Enter to accept the warning that all data will be destroyed.
... automatic reboot will follow
A pseudo-graphic menu with the following windows will be displayed:
“Selection Menu“, “Hardware Detection Results“, and “Information Win-
dow“
Press Enter twice (2 times) to continue the installation.
... several reboots will follow
... Symantec ghost window will be displayed with a “progress indicator“
that this will take ~25 min.
Remove the "syngoMR xxx Customer Image" DVD from the
CD/DVD-ROM drive (R:\). When the following message is displayed:
Plug and Play is running, please wait! Depending on
the hardware, this may take 2-5 minutes
Insert the Service Pack CD into the CD/DVD-ROM drive (R:\)
The automatic installation will continue, this will take ~25 min. At the end
of the installation, the “Windows XP login window“ will be displayed.
Log in as “meduser“
Click the link Click here to start the Local Service Configuration in the
Internet Explorer (with the window title Start Up Check).
Enter the valid Service Key and Password, click OK.
Select Configuration, navigate through the configuration, check to
ensure that the network settings and host name are correct.

Siemens MR-000.816.26.01.02 Page 7 of 22 MAGNETOM


04.07 CS SD MR
8 Installing MRC / MRC+MRSC

NUMARIS/4 reinstallation (on booting HW) 2.3

The following table describes how to install software on hardware that is bootable.
• Switch on the Host

Action MRC MRSC


Login as administrator
Insert the “Numaris/4 ... DVD 1/2“ into the DVD/CD-ROM drive.
Insert the “Master Backup-CD“ into the CD/DVD burner drive (DVD/CD
burner)
Navigate to “R:\start.bat“ (use File Explorer)
Double-click “start.bat“
A DOS shell pops up with the following message: Start to irreversible pre-
pare the system for installation [Y/N]?
Press Y to continue
... automatic reboot will follow
A pseudo-graphic user interface is displayed with the warning that all
data will be lost after pressing a key for confirmation. At this point, press
Enter to accept the warning that all data will be destroyed.
... automatic reboot will follow
A pseudo-graphic menu with the following windows will be displayed:
“Selection Menu“, “Hardware Detection Results“, and “Information Win-
dow“
Press Enter twice (2 times) to continue the installation.
... several reboots will follow
... Symantec ghost window will be displayed with a “progress indicator“
that this will take ~25 min.
Remove the "syngoMR xxx Customer Image" DVD from the
CD/DVD-ROM drive (R:\). When the following message is displayed:
Plug and Play is running, please wait! Depending on
the hardware, this may take 2-5 minutes
Insert the Service Pack CD into the CD/DVD-ROM drive (R:\)
The automatic installation will continue, this will take ~25 min. At the end
of the installation, the “Windows XP login window“ will be displayed.
Log in as “meduser“
Click the link Click here to start the Local Service Configuration in the
Internet Explorer (with the window title Start Up Check).

MAGNETOM MR-000.816.26.01.02 Page 8 of 22 Siemens


04.07 CS SD MR
Installing MRC / MRC+MRSC 9

Action MRC MRSC


Enter the valid Service Key and Password, click OK.
Select Configuration, navigate through the configuration, check to
ensure that the network settings and host name are correct.

Siemens MR-000.816.26.01.02 Page 9 of 22 MAGNETOM


04.07 CS SD MR
10 Installation of Imager
3-
General
3Installation of Imager

This chapter describes how to install the Imager software.

NOTE The MAGNETOM syngo-MR Imager is delivered with fac-


tory-installed software. The following installation procedure
is required only when replacing the host.

Prerequisites 0

• Customer-DVD 2/2 MR B30


• The measurement network is connected

Checking / Importing the Imager license to the MRC


• Login as administrator (MRC)
• Use the Windows file Explorer to copy/paste the license file from the floppy disk to the
following target on the MRC: C:\MedCom\tftp\ ↵
• Rename the license file for the Imager to license.dat

Installation procedure 0

• Insert the Customer-DVD 2/2 in the DVD-ROM drive


• Switch off / on the power to force a re-boot from the CD-ROM drive
• Please wait ~10 minutes, the DVD-ROM drive opens the carriage automatically.
Remove the DVD.
• The silent installation will need ~45 minutes additionally
• Boot the MRC (Host).
• Start the MR application.
• Power on the Imager

NOTE The Imager will reboot twice following the installation. The
initialization of each raw-data disk takes approximately 2
minutes!

Checking the connectivity from Host to Imager and Imager raw drives
• To check the measurement network connectivity from the Host to the Imager, enter the
command ping -t mrir ↵ in a cmd tool at the MRC
• You can use the Remote Desktop Connection at the MRC to access the Imager:
(Advanced user must be enabled) Start, Programs, Accessories, Communications,
Remote Desktop Connection. The computer name of the Imager is mrir ↵), User:
MRSystem ↵, and enter the Password ...

MAGNETOM MR-000.816.26.01.02 Page 10 of 22 Siemens


04.07 CS SD MR
Installation of Imager 11

• Open a cmd tool, check that the “MattauchBuffer“ and “rawexchange“ devices are avail-
able:

NOTE For Imager M450, only one raw disk (F:) is displayed.

C:\>dir F: ↵
...
09/20/2006 09:01 PM 4,697,620,480 MattauchBuffer1.dat
09/20/2006 08:52 PM 4,294,967,296 rawexchange.dat
...
C:\>dir G: ↵
...
09/20/2006 09:03 PM 4,697,620,480 MattauchBuffer2.dat
09/20/2006 08:54 PM 4,294,967,296 rawexchange.dat
...
C:\>dir H: ↵
...
09/20/2006 08:57 PM 4,294,967,296 rawexchange.dat
...
C:\>dir I: ↵
...
09/20/2006 08:59 PM 4,294,967,296 rawexchange.dat
...

Final checks 0

NOTE Check if hotfixes and Service Packs are available for this soft-
ware version according to various Speedinfos and Update
Instructions.

• Perform the “Quality Assurance“ procedure: “Image Orientation“ in the service platform.
• Make sure that the system is generating images.

Siemens MR-000.816.26.01.02 Page 11 of 22 MAGNETOM


04.07 CS SD MR
12 Site-specific Data
4-
Backup site-specific data
4Site-specific Data

Required time: 10-15 min.


Required tools / parts: CD-R

NOTE Use only CD-R 's with medical grade, e.g. Siemens Medical
CD-R 80 or TDK Medical CD-R.

Save the site-specific data each time you make changes to the site-specific configuration,
tune-up, or sequences and protocols. In case of a hard disk crash, you will be able to
restore a valid set of site-specific data.
• (You are logged in as meduser / login as meduser.)
• Start the service platform, select Backup & Restore.
• Insert a recordable CD-disk in the CD/DVD-burner.
• Select Backup in the “Command“ pull-down menu.
• In the menu Drives, select the target drive CD/DVD-burner (drive letter S:) used to save
the data.
• Select Master-Backup in the “Packages“ menu.
• Click Go to start the “Backup“, check for errors in the log-window after the backup has
been completed.

Setting that are not part of the site-specific data:


Remote access mode (No Access, Limited Access, etc in the customer UI).
Auto-transfer flag (“Configuration“, “Service“, “Auto Transfers“)
Postscript Printer Configuration (Customer paper printer)
ImageText Settings (ImageText Settings from ImageText Dialog)
Time zone
Transfer Rules

NOTE During backup at the MRSC, a “CorRea.bru“ error may be


displayed, please ignore this error.

NOTE If there are errors regarding “ITO“, check if CA Unicenter has


been installed previously. In case CA Unicenter is installed.
this is an error. Manually write down the CA Unicenter config-
uration settings. If CA Unicenter has not been installed,
please ignore this error.

MAGNETOM MR-000.816.26.01.02 Page 12 of 22 Siemens


04.07 CS SD MR
Site-specific Data 13

Restore site-specific data (manually after SW installation) 4.1

Restoring site-specific data may be helpful in many situations. Please read the following
information prior to beginning the work.
After a software installation without “Silent-Restore“, a set of site-specific data will save you
plenty of work/time. The site-specific data e.g. Configuration, Tune-up, and QA are included
in the master backup.
• Login as meduser ↵.
• Start the “Internet Explorer“
• Click the link “Click here to start the Local Service Configuration“ in the Internet
Explorer (with the window title Start Up Check).
• Login to the service platform, select Backup & Restore.
• Insert the backup disk in the CD/DVD-ROM.
• Select Restore in the “Command“ pull-down menu.
• Select the source drive CD/DVD-ROM (drive letter R:) in the “Drives“ menu.
• Select your xxx_Master-Backup_xxxx_xxxx_ok.ar to be restored in the “Archive“
menu.
• In the menu “Groups“, select the necessary packages to be restored.
• Click Go to start the “Backup“, check for errors in the log window after the restore has
been completed. (Estimated time ~10-15 min.)
• Click Home.
• Click OK to reboot the system.

Siemens MR-000.816.26.01.02 Page 13 of 22 MAGNETOM


04.07 CS SD MR
14 Site-specific Data

Restore site-specific data (on running system) 4.2

Restoring the site-specific data may be helpful in many situations. Please read the following
information before beginning with the work.
In case of corrupted site-specific data, we recommend saving the actual status of the
site-specific data to another medium before restoring an old save set. This will save time if
the problem is resolved by reinstalling the software. The actual tune-up data may save a
great deal of work, especially if the customer optimized his protocols recently.
• Login as administrator ↵.
• Start the “Internet Explorer“
• Login to the service platform, select Backup & Restore.
• Insert the backup disk in the CD/DVD-ROM.
• Select Restore in the “Command“ pull-down menu.
• Select the source drive CD/DVD-ROM (drive letter R:) in the “Drives“ menu.
• Select your xxx_Master-Backup_xxxx_xxxx_ok.ar to be restored in the “Archive“
menu.
• In the menu “Groups“, select the necessary packages to be restored.
• Click Go to start the “Backup“, check for errors in the log window after the restore has
been completed. (Estimated time ~10-15 min.)

MAGNETOM MR-000.816.26.01.02 Page 14 of 22 Siemens


04.07 CS SD MR
Configuration 15
5-
General
5Configuration

For detailed information regarding the configuration of syngo MR systems, refer to Help in
the Service Software Platform
If you have changed the configuration within the Service Software Platform and the mes-
sage “reboot system pending“ is displayed in the “status bar“, reboot the system as follows:
Click Home, confirm the reboot message window with OK.

NOTE How to configure Siemens Remote Service is described sep-


arately in the document “Installation and Startup for SRS“.

NOTE The use of non-released combinations of hardcopy cameras


and BU products is prohibited and does not meet the require-
ments defined. For detailed information regarding the CAM-
ERA configuration, refer to technical documentation
“Camera Info ...“ valid for your NUMARIS/4 software version.

NOTE The “CCC“ Connectivity Competence Center provides a pow-


erful configuration management tool see:
http://cs.med.siemens.de/cms/en/home/
For Service, DICOM, in the DICOM Homepage (CCC) navigate
to
“Interoperability DB“, Download, Acquisition Tool).

NOTE Remember to create a backup if the configuration has


changed.

Siemens MR-000.816.26.01.02 Page 15 of 22 MAGNETOM


04.07 CS SD MR
16 Appendix
6-
Installing patch CD VA30
6Appendix

Remarks 0

• On some systems, the following or a similar error message may occasionally be dis-
played:

Fig. 1:
• Accept the message by clicking ’OK’: installation will continue.

Preparation for Service Pack installation 0

• Steps at the MRSC, if available

ACTION MRSC
• Log off Meduser and log in as Administrator (in the mainUI, select "System"
-> "End Session", select "Log Off", and press Yes in the next pop-up. Con-
firm all further pop-ups displayed).
Note: Software patch can be installed only under user administrator.

• Steps at the MRC

ACTION MRC
• Log off Meduser and log in as Administrator (in the main UI, select "System"
-> "End Session", select "Log Off", and press Yes in the next pop-up. Con-
firm all further pop-ups displayed).
Note: Software patch can be installed only under user administrator.

• Steps at the MRIR

ACTION MRIR
• Insert the CD Service Pack CD/DVD

Service pack installation at the MRIR VA30 0

• Steps at the MRC

MAGNETOM MR-000.816.26.01.02 Page 16 of 22 Siemens


04.07 CS SD MR
Appendix 17

ACTION MRC/MRI
• Start the remote desktop connection at the MRC by selecting START -->
RUN and entering ShowMeMRIR.bat <Enter>.
If the task bar is not visible, press Ctrl + ESC on the keyboard to access it
(if a monitor is not hooked up to the imager).
• Log into the MRIR with user MRSystem.
If the MRIR does not accept the password, check the keyboard language
setting. (Y <-> Z)
Username: MRSystem
Password: MRSystempwd
Note:
The following steps are performed using the remote desktop connection on the MRIR
(imager).
• Start Windows NT Explorer:
- Click START -> Programs -> Accessories -> Windows Explorer
• Navigate to the CD/DVD-ROM drive E:

• Run program MR_Update.bat by double-clicking it.


NOTE:
The program will copy the software update package to the system first. After
copying is complete, installation of the package will start in a new CMD Win-
dow (this will take approx. 10 min!)
The original CMD window displays ‘Press any key to continue ...’ This can be
safely done at any time. The CD will be ejected AUTOMATICALLY. Logfile
entries such as:
”Installation finished with ERRORLEVEL: 42” are indicating no errors!! The
installation has finished when the following message is displayed in the CMD
window:
....
....
- all packages have been installed successfully.
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
--> Please reboot the system now. <--
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
-----------------------------------------------------
Finished installation of update packages. (actual date
actual time)
*****************************************************
• Close the remote desktop connection to the MRIR (imager) by clicking the
X on the remote desktop bar.

Siemens MR-000.816.26.01.02 Page 17 of 22 MAGNETOM


04.07 CS SD MR
18 Appendix

• Steps at the MRIR

ACTION MRIR
• Remove the CD Service Pack CD/DVD.

Service Pack installation at the MRC/MRSC 0

• Steps at the MRC

ACTION MRC
• Insert CD Service Pack CD/DVD

• Start Windows NT Explorer:


- Click START -> Programs -> Accessories -> Windows Explorer
• Navigate to the CD/DVD-ROM drive R:

• Run program MR_Update.bat by double-clicking it.


NOTE:
The program will copy the software update package to the system first. After
copying is complete, installation of the package will start in a new CMD Window
(this will take approx. 10 min!)
The original CMD window displays ‘Press any key to continue ...’ This can be
safely done at any time. The CD will be ejected AUTOMATICALLY. Logfile entries
such as:
”Installation finished with ERRORLEVEL: 42” are indicating no errors!! The
installation has finished when the following message is displayed in the CMD
window:
....
....
- all packages have been installed successfully.
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
--> Please reboot the system now. <--
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
-----------------------------------------------------
Finished installation of update packages. (actual date
actual time)
*****************************************************
• Remove the CD Service Pack CD/DVD

MAGNETOM MR-000.816.26.01.02 Page 18 of 22 Siemens


04.07 CS SD MR
Appendix 19

ACTION MRC
• Reboot the entire system (MRC and MRIR [imager]).
- Reboot the MRIR (imager)
- Select the CMD tool window, if one is available, or open a new one.
- Enter the following command
shutdown -r -m \\mrir and press <ENTER>
(Use the same syntax for the command as shown above !)
The MRIR reboots in 30 sec.
- Reboot the MRC from the start menu.
• The system automatically logs in under the "meduser" account, or log in as
"meduser" if auto login was not activated.
• After the login, the progress window from the license update is displayed
automatically.

Fig. 2:
The progress window automatically closes when the license update finishes
successfully
• Select Options --> Service -> Local Service.

• Enter the valid service key.

• Select Configuration.

• Select Next.

• Select the page "Measurement" Meas. Settings


Accept the settings as they are. Do not change any!
• Click Save and confirm the pop-up window with OK (configuration data are
saved).

Siemens MR-000.816.26.01.02 Page 19 of 22 MAGNETOM


04.07 CS SD MR
20 Appendix

ACTION MRC
• Continue with Finish.

• Restart the system as follows:


- Click Home in the top right corner.
- A window prompting you to reboot the system is displayed.

Fig. 3:
- Click "OK".
The reboot starts (takes about 5 min. until the reboot starts).
Note:
This reboot has to be performed as specified above. Do not perform this reboot
manually. Otherwise, Numaris will not boot and the start-up check for Numaris
will fail.

• Steps at the MRSC, if available

ACTION MRSC
• Insert the CD Service Pack CD/DVD

• Start Windows NT Explorer:


- Click START -> Programs -> Accessories -> Windows Explorer
• Navigate to the CD/DVD-ROM drive R:

MAGNETOM MR-000.816.26.01.02 Page 20 of 22 Siemens


04.07 CS SD MR
Appendix 21

ACTION MRSC
• Run program MR_Update.bat by double-clicking it.
NOTE:
The program will copy the software update package to the system first. After
copying is complete, installation of the package will start in a new CMD Window
(this will take approx. 10 min!)
The original CMD window displays ‘Press any key to continue ...’ This can be
safely done at any time. The CD will be ejected AUTOMATICALLY. Logfile entries
such as:
”Installation finished with ERRORLEVEL: 42” are indicating no errors!!
The installation has finished when the following message is displayed in the CMD
window:
....
....
- all packages have been installed successfully.
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
--> Please reboot the system now. <--
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
-----------------------------------------------------
Finished installation of update packages. (actual date
actual time)
*****************************************************
• Remove/Insert the CD Service Pack CD/DVD

• Reboot the MRSC from the start menu.

Siemens MR-000.816.26.01.02 Page 21 of 22 MAGNETOM


04.07 CS SD MR
22 Appendix

MAGNETOM MR-000.816.26.01.02 Page 22 of 22 Siemens