MAGNETOM
MR
Software
System
© Siemens 2007
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Final checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Introduction 0
Prerequisites 0
Abbreviations 0
The following table provides a number of important milestones for the software installation.
MRC (save set available) MRC (no valid save set) MRSC
Install software Install software Install software
Install software at the Imager n.a.
Restore site-specific data Initial configuration Restore site-specific data /
or import clone package
from the MRC
Check system language Configure the system Check system language
and weight and weight
Check if service packs are Perform system tune-up, Check if service packs are
available for this software quality assurance, and the available for this software
version installation protocol version
Check the image orienta- Check system language and n.a.
tion weight
Check if Update Instructions with service packs are available for this software version
NOTICE Verify the software baseline label with the Help, Info function.
The correct software label for the released syngo MR A30 is:
¹ COEM VD20F
¹ syngo VE31G
¹ N4_VA30A_LATEST_20070315
The BIOS update optimizes the setting of NUMARIS/4 for your host type, sets the
NUMARIS/4 BIOS password and boot device order
The following table describes how to install software on hardware without an operating sys-
tem.
• Perform the “BIOS update“
The following table describes how to install software on hardware that is bootable.
• Switch on the Host
Prerequisites 0
Installation procedure 0
NOTE The Imager will reboot twice following the installation. The
initialization of each raw-data disk takes approximately 2
minutes!
Checking the connectivity from Host to Imager and Imager raw drives
• To check the measurement network connectivity from the Host to the Imager, enter the
command ping -t mrir ↵ in a cmd tool at the MRC
• You can use the Remote Desktop Connection at the MRC to access the Imager:
(Advanced user must be enabled) Start, Programs, Accessories, Communications,
Remote Desktop Connection. The computer name of the Imager is mrir ↵), User:
MRSystem ↵, and enter the Password ...
• Open a cmd tool, check that the “MattauchBuffer“ and “rawexchange“ devices are avail-
able:
NOTE For Imager M450, only one raw disk (F:) is displayed.
C:\>dir F: ↵
...
09/20/2006 09:01 PM 4,697,620,480 MattauchBuffer1.dat
09/20/2006 08:52 PM 4,294,967,296 rawexchange.dat
...
C:\>dir G: ↵
...
09/20/2006 09:03 PM 4,697,620,480 MattauchBuffer2.dat
09/20/2006 08:54 PM 4,294,967,296 rawexchange.dat
...
C:\>dir H: ↵
...
09/20/2006 08:57 PM 4,294,967,296 rawexchange.dat
...
C:\>dir I: ↵
...
09/20/2006 08:59 PM 4,294,967,296 rawexchange.dat
...
Final checks 0
NOTE Check if hotfixes and Service Packs are available for this soft-
ware version according to various Speedinfos and Update
Instructions.
• Perform the “Quality Assurance“ procedure: “Image Orientation“ in the service platform.
• Make sure that the system is generating images.
NOTE Use only CD-R 's with medical grade, e.g. Siemens Medical
CD-R 80 or TDK Medical CD-R.
Save the site-specific data each time you make changes to the site-specific configuration,
tune-up, or sequences and protocols. In case of a hard disk crash, you will be able to
restore a valid set of site-specific data.
• (You are logged in as meduser / login as meduser.)
• Start the service platform, select Backup & Restore.
• Insert a recordable CD-disk in the CD/DVD-burner.
• Select Backup in the “Command“ pull-down menu.
• In the menu Drives, select the target drive CD/DVD-burner (drive letter S:) used to save
the data.
• Select Master-Backup in the “Packages“ menu.
• Click Go to start the “Backup“, check for errors in the log-window after the backup has
been completed.
Restoring site-specific data may be helpful in many situations. Please read the following
information prior to beginning the work.
After a software installation without “Silent-Restore“, a set of site-specific data will save you
plenty of work/time. The site-specific data e.g. Configuration, Tune-up, and QA are included
in the master backup.
• Login as meduser ↵.
• Start the “Internet Explorer“
• Click the link “Click here to start the Local Service Configuration“ in the Internet
Explorer (with the window title Start Up Check).
• Login to the service platform, select Backup & Restore.
• Insert the backup disk in the CD/DVD-ROM.
• Select Restore in the “Command“ pull-down menu.
• Select the source drive CD/DVD-ROM (drive letter R:) in the “Drives“ menu.
• Select your xxx_Master-Backup_xxxx_xxxx_ok.ar to be restored in the “Archive“
menu.
• In the menu “Groups“, select the necessary packages to be restored.
• Click Go to start the “Backup“, check for errors in the log window after the restore has
been completed. (Estimated time ~10-15 min.)
• Click Home.
• Click OK to reboot the system.
Restoring the site-specific data may be helpful in many situations. Please read the following
information before beginning with the work.
In case of corrupted site-specific data, we recommend saving the actual status of the
site-specific data to another medium before restoring an old save set. This will save time if
the problem is resolved by reinstalling the software. The actual tune-up data may save a
great deal of work, especially if the customer optimized his protocols recently.
• Login as administrator ↵.
• Start the “Internet Explorer“
• Login to the service platform, select Backup & Restore.
• Insert the backup disk in the CD/DVD-ROM.
• Select Restore in the “Command“ pull-down menu.
• Select the source drive CD/DVD-ROM (drive letter R:) in the “Drives“ menu.
• Select your xxx_Master-Backup_xxxx_xxxx_ok.ar to be restored in the “Archive“
menu.
• In the menu “Groups“, select the necessary packages to be restored.
• Click Go to start the “Backup“, check for errors in the log window after the restore has
been completed. (Estimated time ~10-15 min.)
For detailed information regarding the configuration of syngo MR systems, refer to Help in
the Service Software Platform
If you have changed the configuration within the Service Software Platform and the mes-
sage “reboot system pending“ is displayed in the “status bar“, reboot the system as follows:
Click Home, confirm the reboot message window with OK.
Remarks 0
• On some systems, the following or a similar error message may occasionally be dis-
played:
Fig. 1:
• Accept the message by clicking ’OK’: installation will continue.
ACTION MRSC
• Log off Meduser and log in as Administrator (in the mainUI, select "System"
-> "End Session", select "Log Off", and press Yes in the next pop-up. Con-
firm all further pop-ups displayed).
Note: Software patch can be installed only under user administrator.
ACTION MRC
• Log off Meduser and log in as Administrator (in the main UI, select "System"
-> "End Session", select "Log Off", and press Yes in the next pop-up. Con-
firm all further pop-ups displayed).
Note: Software patch can be installed only under user administrator.
ACTION MRIR
• Insert the CD Service Pack CD/DVD
ACTION MRC/MRI
• Start the remote desktop connection at the MRC by selecting START -->
RUN and entering ShowMeMRIR.bat <Enter>.
If the task bar is not visible, press Ctrl + ESC on the keyboard to access it
(if a monitor is not hooked up to the imager).
• Log into the MRIR with user MRSystem.
If the MRIR does not accept the password, check the keyboard language
setting. (Y <-> Z)
Username: MRSystem
Password: MRSystempwd
Note:
The following steps are performed using the remote desktop connection on the MRIR
(imager).
• Start Windows NT Explorer:
- Click START -> Programs -> Accessories -> Windows Explorer
• Navigate to the CD/DVD-ROM drive E:
ACTION MRIR
• Remove the CD Service Pack CD/DVD.
ACTION MRC
• Insert CD Service Pack CD/DVD
ACTION MRC
• Reboot the entire system (MRC and MRIR [imager]).
- Reboot the MRIR (imager)
- Select the CMD tool window, if one is available, or open a new one.
- Enter the following command
shutdown -r -m \\mrir and press <ENTER>
(Use the same syntax for the command as shown above !)
The MRIR reboots in 30 sec.
- Reboot the MRC from the start menu.
• The system automatically logs in under the "meduser" account, or log in as
"meduser" if auto login was not activated.
• After the login, the progress window from the license update is displayed
automatically.
Fig. 2:
The progress window automatically closes when the license update finishes
successfully
• Select Options --> Service -> Local Service.
• Select Configuration.
• Select Next.
ACTION MRC
• Continue with Finish.
Fig. 3:
- Click "OK".
The reboot starts (takes about 5 min. until the reboot starts).
Note:
This reboot has to be performed as specified above. Do not perform this reboot
manually. Otherwise, Numaris will not boot and the start-up check for Numaris
will fail.
ACTION MRSC
• Insert the CD Service Pack CD/DVD
ACTION MRSC
• Run program MR_Update.bat by double-clicking it.
NOTE:
The program will copy the software update package to the system first. After
copying is complete, installation of the package will start in a new CMD Window
(this will take approx. 10 min!)
The original CMD window displays ‘Press any key to continue ...’ This can be
safely done at any time. The CD will be ejected AUTOMATICALLY. Logfile entries
such as:
”Installation finished with ERRORLEVEL: 42” are indicating no errors!!
The installation has finished when the following message is displayed in the CMD
window:
....
....
- all packages have been installed successfully.
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
--> Please reboot the system now. <--
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
-----------------------------------------------------
Finished installation of update packages. (actual date
actual time)
*****************************************************
• Remove/Insert the CD Service Pack CD/DVD