Contents

Chapter1 Safety Precautions ......................................................................................................... 1


Chapter2 Main Unit ...................................................................................................................... 6
Chapter4 External Interfaces ....................................................................................................... 10
Chapter5 Dry Battery/AC Adapter Installation ........................................................................... 13
5.1 Dry Battery Installation .............................................................................................. 14
5.2 Using the AC Adapter ................................................................................................. 14
Chapter6 Setting the Date and Time ........................................................................................... 16
Chapter7 Unit ............................................................................................................................. 19
Chapter8 Alarm Function ............................................................................................................ 20
Chapter9 Applying the Limb Cuff .............................................................................................. 22
Chapter10 BP Measurement ....................................................................................................... 26
10.1 Accurate Measurement Way ..................................................................................... 26
10.2 BP Measurement....................................................................................................... 28
Chapter11 Memory Function ...................................................................................................... 30
11.1 Review the Memory Value........................................................................................ 30
11.2 Delete Memory Values .............................................................................................. 32
Chapter13 Error Message ........................................................................................................... 37
Chapter14 Troubleshooting......................................................................................................... 39
Chapter15 Clean and Maintenance ............................................................................................. 41
Chapter16 Specification .............................................................................................................. 46
I
Chapter1 Safety Precautions
 Before use, carefully read "Safety Precautions" for a correct use.
 To prevent users suffered hurt or damnification due to improper use, see "Safety
Precautions", and use this product properly.
For safety reasons, be sure to comply with safety precautions.

Note

If not to use correctly, it exists that a potentially hazardous situation which may result in

injury to the user or patient or damage to the equipment or other property.

1
Note
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician.
Contact your physician for specific information about your animal’s blood pressure.
Please hand measurement results to the doctor who know your animal’s health to
accept diagnosis.
For severe blood circulation disorder or arrhythmia patients, please use the device
under the guidance of a doctor.
Otherwise it may lead to acute hemorrhage, or measurement error as a result of
squeezed arm.
This device is intended for using in measuring blood pressure and pulse rate.
Do not use for any other purpose.
Otherwise it may cause accident or holdback.
Please use special cuff.
Otherwise it is possible that measurement result is incorrect.
Do not disassemble or attempt to repair the unit or components without permission.
Otherwise it can not measure correctly.

2
Operation for AC Adapter (Separate Sale)

Note
Please use sold separately dedicated AC adapter.
Otherwise it may cause trouble.
The dedicated AC adapter must use AC 100V-240V.
Otherwise it may cause fire or electric shock.
When there is breakage of sold separately dedicated AC adapter plug or wire ,
please immediately pulled the plug from the socket.
Otherwise it may cause electric shock or injury.
Do not plug or unplug the adapter power cord with wet hands.
Otherwise it may cause electric shock or injury.

3
Operation for battery
Note
Please use 4 "AA" size manganese or alkaline batteries, do not use batteries of other types.
Otherwise it may cause fire.
New and old batteries, different kinds batteries can not be confusion.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
+ and - polarities of the batteries must match the polarities of the battery compartment as
indicated. When the batteries power exhausts, replace with four new batteries at the same time.
Please take out the batteries when you do not use the device for a long time.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water.
Contact a physician immediately.
Otherwise it will cause blindness or other hazards.
If electrolyte of the batteries immodestly glues on the skin or the clothes, immediately rinse with
plenty of clean water.
Otherwise it may hurt the skin.

4
Advice
Do not subject the device to strong shocks, such as dropping the unit on the floor.
Do not inflate before the cuff wrapps around the arm.
Do not inflect the cuff and the air tube forcibly.
Warning
You must not perform NIBP measurements on animals with sickle-cell disease or under any
condition which the skin is damaged or expected to be damaged.
For a thrombasthemia animal, it is important to determine whether measurement of the
blood pressure shall be done automatically. The determination should be based on the
clinical evaluation.
Ensure that the correct setting and blood pressure cuff are selected when performing
measurements on animal. It may be dangerous for the animal to use an over pressure level.
The device has normal useful life for five years since the first electrified use. After its
normal life, dispose the device and its accessories according to applicable local regulations
about environmental.

5
Chapter2 Main Unit
The production is in the package. Open the package and confirm whether the production is
whole.

6
Accessories:
Cuff:

The sphygmomanometer supports cuffs in a variety of sizes, please choose suited cuff according
to the size of the animal.
User Manual

Optional Accessories:
AC adapter

7
Input: voltage: AC 100V~240V
frequency: 50Hz/60HZ
Rated current: AC 150mA
Output: DC 6.0V±0.2V 1.0A
or DC5.0V±0.2V 1.0A

Cuff:
There are several suitable cuffs(range of limb circumference,middle of upper limb)
the range of limb circumference is 6-11cm
the range of limb circumference is 10-19cm
the range of limb circumference is 18-26cm
the range of limb circumference is 22-32cm
the range of limb circumference is 32-43cm
Note
◎The cuff is a consumable. Calculate by measuring 6 times a day(3 times each morning and
evening), the service life of the cuff is about 1 year.(using our experimental conditions)
◎In order to correctly measure blood pressure, please replace the cuff in time.
◎If the cuff leaks, please contact our company to buy a new one. The cuff purchased separately
does not include the airway tube plug. When replacing, please do not throw the airway tube plug
away, install it on the new cuff.
8
Note
When the product and accessories described in this manual are about to exceed the period of use,
they must be disposed according to relevant product handling specification. If you want to know
more information, please contact our company or representative organization.

9
Chapter3 Button Functions
All the operations to the Electronic Sphygmomanometer are through the buttons. The names of
the buttons are above them. They are:

【ON/OFF】 Hold the button to start or close the device.

【START/STOP】 Press to inflate the cuff and start a blood pressure


measurement. When measuring, press to cancel the measurement and deflate the cuff.
The three buttons correspond with the hint in the LCD screen
downside, pressing any button will carry on corresponding function, eg:【MENU】
【ENTER】 【LIST】 etc.

Up and down buttons respectively carry on the functions of moving the cursor

up and down, changing the parameters and switching the status.

10
Chapter4 External Interfaces

Note
Please hold the air plug to remove the NIBP cuff.

11
12
Chapter5 Dry Battery/AC Adapter Installation
The production can use dry battery or AC adapter as power source.

13
5.1 Dry Battery Installation

① Demount the battery cover in the direction of the arrow.


② Install "AA" dry batteries according to polarities.
③ Slide to close the battery cover.
Icon “ ”: the batteries power will exhaust.
Replace with four new batteries (the same sort) at the same time.
Turn the unit off before replacing the batteries.
Note
Dispose of the batteries according to applicable local regulations about environmental.

5.2 Using the AC Adapter

① Connect device and the AC Adapter. Insert the AC Adapter Plug into the AC Adapter Jack
on the right side of the device.
② Please insert the power plug of AC adapter into the 100V- 240V socket.
Note
①Hold and pull the Housing to remove the AC Adapter from the electrical outlet. Do not
remove by pulling on the cord.
14
②Remove the AC Adapter plug from the unit.
Please be sure to use dedicated AC adapter.
Note
You'd better take off batteries when use the AC adapter as power source. If there is any
damage to the AC adapter, you should use batteries to run the device.
When adapter and batteries are both used at the same time, the battery power will not be
consumed. Switch adapter and battery as power supply when the device is off, otherwise,
the device may shutdown due to power failure.

15
Chapter6 Setting the Date and Time

is necessary to set date and time after turning on the device.


The Electronic Sphygmomanometer can automatically stores measurement results, date
and time.
If dry battery power exhausts or removed, the time to stop.
At the moment, please reset date and time.

The Electronic Sphygmomanometer stores 100 groups of measurement results at most. If the
date and time are set correctly, the date and time when measuring will be correct in the memory,
otherwise it may not be correct.The results can be uploaded to PC via USB and processed with
the PC software. If the date and time are set correctly, the date and time when measuring will be
correct in the memory, otherwise it may not be correct.
Note
Correctly use data upload function:
1. First, make sure the time is right, there are three modes of time setting:
(1)When using the Sphygmomanometer for the first time or after the Sphygmomanometer has
been placed without power supply for a certain time(more than 5 minutes), after power on, there
is a prompt of time error on the main interface, set date and time with 【UP】, 【DOWN】 and

16
【ENTER】 button.

(2)Press 【MENU】 button on the main interface to enter system menu, then enter 【SYSTEM
TIME】 item, the current time will be displayed on the screen. Set date and time with 【UP】,
【DOWN】 and 【ENTER】 button

17
2. After setting, select 【CONFIRM】 option and press 【ENTER】 button to confirm the setting
value. If you do not want to change the time, select 【EXIT】 option and press 【ENTER】
button to return to the previous menu.
Note
The year can be set between 2010 and 2099. When the display reaches 2099, it will return to
2010 when pressing【UP】button.
Note
Please choose the computer which should be ensured compliance with the requirements of
IEC60950, or else it may damage the device.

18
Chapter7 Unit
There are two units: "mmHg" and "kPa".
The default is: "mmHg".
To switch "mmHg" and "kPa" units, enter the 【SYSTEM SETUP】 submenu in 【SYSTEM
MENU】, and complete switching in 【UNIT】 item.

19
Chapter8 Alarm Function
Alarms are classified into two categories: technical alarm and physiological alarm.
Physiological Alarm
User can press 【MENU】 button to enter system menu, select 【ALARM SETUP】 item to enter
its interface, and then set alarm on-off and the high and low alarm limits, when blood pressure is
higher than the high limit or lower than the low limit, the physiological alarm will occur.

Press any button to cancel the alarm in physiological alarm state and this method brings no
affection to the next alarm. To close alarm in 【ALARM SETUP】 will disable the alarm
function until switch alarm on.

20
Technical Alarm
When power is about to exhaust and alarm is on , then the alarm will occur. This alarm can not
be cancelled unless being closed or the power replaced.

21
Chapter9 Applying the Limb Cuff

The measurement can be carried out by applying the cuff on left or right forelimb.
Carry out the operation in a room with comfortable temperature.

In order to measure accurately, pay attention to applying the cuff properly. Select a proper cuff
according to different animals, then select proper cuff type(large, middle and small).
There are two ways to select the cuff mode, in the time interface, press CUFF button to switch
cuff mode; or enter the 【SYSTEM SETUP】 submenu in 【SYSTEM MENU】, and complete
switching in 【CUFF TYPE】 item.

22
Wear the cuff accurately following these steps:

① Make sure the air plug is securely inserted in the main unit.

② Apply the blood pressure cuff to the animal's leg following the instructions below .
 Ensure that the cuff is completely deflated.
 Apply the appropriate size cuff to the animal. Ensure that the cuff is not wrapped too
tightly around the limb. Excessive tightness may cause discoloration and eventual
ischemia of the extremities.

23
Note
Every CUFF TYPE corresponds to different cuff sizes, which are identified on the cuff. As
follows:

24
Animal Type Limb perimeter

Small <13

Middle 8 ~ 26 cm

Large >25 cm
 Make sure that the cuff edge falls within the range of mark.If it does not, use a larger or
smaller cuff that fits better.
Warning
Before starting a measurement, verify that you have selected a appropriate setting of
animal type (large,middle,or small)for your animal according to its limb perimeter.
Do not apply the cuff to a limb that has an intravenous infusion or catheter in place.This
could cause tissue damage around the catheter when infusion is slowed or blocked during
cuff inflation.

25
Chapter10 BP Measurement
10.1 Accurate Measurement Way
The limb chosen for taking the measurement should be placed at the same level as the animal's
heart. If this is not possible you should apply the following corrections to the measured values:If
the cuff is placed higher than the heart level, add 0.75 mmHg for each inch of difference.If it is
placed lower than the heart level, deduct 0.75 mmHg for each inch of difference.

26
Try to measure your animal’s blood pressure at the same time each day with the
same arm
and the same pose for consistency.
Do not touch Electronic Sphygmomanometer, cuff and windpipe during measurement.
Measurements should be taken in a quiet place and the body relax.
Remain still 4~5 minutes before measurement.
Relax the body, do not let the muscle activity.
Do not bark or move during the measurement.
Wait 4~5 minutes between measurements.
Do not use precision instruments near the Sphygmomanometer.
Please use the sphygmomanometer at an environment of suitable temperature and
humidity(refer to Chapter 16), otherwise it will cause measurement error.

27
10.2 BP Measurement
①Press 【START/STOP】 button to take a measurement.

Start measurement measurement result


During measurement, please keep correct pose and quiet state, do not move.
Stop Measurement
During measurement, if you wish to stop measuring, press 【START/STOP】 button to stop and
deflate.
②Confirm Measurement Value
Measurement value can be automatically stored.([memory function] refer to Chapter 11)
*Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions
28
of your physician.

Note
 Wait at least 4-5 minutes between measurements.
When repeated measurements, because, the limb appears congestion, it may not get correct
blood pressure measurement. After the blood flow, take a measurement once again.
 When some factors affect the measurement results in measurement process, error
messages hints will appear on the screen, you can obviate the malfunction and restart a
measurement.

③In no physiological alarm state, press any button to carry on the corresponding button function;
in audio alarm state, press any button (except 【ON/OFF】 button) to clear up the audio alarm.
④Take off the cuff, hold 【ON/OFF】 button to turn the device off.
*The device will automatically turn off after two minutes in which there is no operation to the
device, even if you forget to turn the power off.

29
Chapter11 Memory Function
The device is designed to store and display the blood pressure, pulse rate values and the date and
time when measured, which are up to 100 items. If there have been 100 items stored, when the
101 measurement have been taken, the earliest results are deleted.

11.1 Review the Memory Value

1.In the main interface (interface when boot-strap), press 【MEMORY】 button to review the
most recent measurement values in large-print with the serial number from 1 to 100.

30
2.Press 【UP】/【DOWN】 button to circularly switch the former measurement values.

*The right figure shows that there is no Measurement result.

3.Press 【LIST】 button to switch to data table interface.

31
4.Press 【TREND】 button to display trend interface.

5.Finish displaying the measurement values


Press 【EXIT】 to return the main interface or hold 【ON/OFF】 button to turn the power off.
*The device will automatically turn off after two minutes in which there is no operation to the
device, even if you forget to turn the power off.
11.2 Delete Memory Values
Users can delete all values stored in the memory of each user separately instead of deleting a
special item.
1.Press 【MENU】 button to enter the system menu, select 【DELETE DATA】 to enter the
confirm interface, after confirming again, all measurement results will be deleted.

32
2.Finish Operation
Select 【CONFIRM】 or 【EXIT】 to return the previous menu, or hold 【ON/OFF】 button
to turn the power off.
*The device will automatically turn off after two minutes in which there is no operation to the
device, even if you forget to turn the power off.

33
Chapter12 Keys and Symbols

Signal Description

Warning – See User Manual

SYS Systolic pressure

MAP MAP pressure

DIA Diastolic pressure

PR Pulse rate (bpm)

LARGE Large animal

MIDDLE Middle animal

SMALL Small animal

INFO Information

34
Open the alarm sound indication

Close the alarm sound indication

Low-power

Full-power

1.no NIBP data to review


2.An indicator of signal inadequacy

ClassⅡequipment

WEEE (2002/96/EC)

Type BF Applied Part

35
SN Serial number

This item is compliant with Medical Device Directive 93/42/EEC


of June 14, 1993, a directive of the European Economic Community.

European Representative

USB socket connect SpO2 probe(This item only fit for European
Union market)

36
Chapter13 Error Message
Error message will be displayed in the screen if there is something wrong when measuring. The
causes and solutions are shown as follows:

Error Message Causes Solutions


Self-test failure
Function abnormal Please contact us
System failure
Cuff is not connected
Loose cuff Correctly connect cuff (refer to Chapter 9)
correctly.
Make sure the cuff plug is securely inserted
Air leakage Cuff plug fall off
in the windpipe (refer to Chapter 9)
Air pressure error Air pressure error Refer to the troubleshooting
The pulse signal is too
Weak signal Correctly connect cuff (refer to Chapter 9)
weak or the cuff is loose.
Cuff is blocked or
overpressure Correctly connect cuff (refer to Chapter 9)
squeezed

37
Excessive The signal extent is too
movement big owing to the arm or
Over range body moving or other Keep limb, body still, measure again
Saturated signal reasons when measuring
Time out It takes too much time

38
Chapter14 Troubleshooting
Abnormal
Causes Solutions
Phenomenons
Correctly connect cuff (refer to
BP measurement Cuff is not connected correctly.
Chapter 9)
values too high or
Keep quiet and restart a
too low. Move limbs when measuring
measurement
Cuff leakage Buy a new cuff
The cuff windpipe is not
Correctly connect
No pressure correctly connected with cuff
Stop using the device and contact
Cuff is not inflated
us
Cuff deflates in
Loose cuff Correctly apply cuff
short time
It can not carry on measurement when press the Switch on the power once again
measurement button and restart a measurement
Power off No use for a long time, the
Replace all four batteries with new
suddenly when power of batteries can be
ones.
inflating exhausted owing to the changed
39
temperature
Power of batteries can be Replace all four batteries with new
Hold the on/off exhausted ones.
button but can not Check the battery installation for
start the device The battery polarities is reversed proper placement of the battery
polarities.
Cuff inflation start before press the measurement Pull out the cuff to deflate. Stop
button or never stop inflating when measuring using the device and contact us.
Pull out the cuff to deflate. Stop
Cuff never deflation
using the device and contact us.
No deflation or deflation error or Pull out the cuff to deflate. Stop
inflation without stop using the device and contact us.
Air pressure error
Keep arm, body still, measure
Others
again.
No press value displayed or the value unchanged or Pull out the cuff to deflate. Stop
change erratically when cuff inflated using the device and contact us.
Switch on the power once again
and restart an operation.
Other phenomenon
Replace the batteries.
If no, please contact us.

40
Chapter15 Clean and Maintenance
*Please follow the instructions in the user manual. If you do not comply, our company will not
assume responsibilities of the quality.
Warning
Must remove batteries and cut off the alternating power supply before cleaning the
sphygmomanometer, accessories must be clean separately.
Caution
 Do not high pressure disinfect the sphygmomanometer and accessories.
 Do not immerse the sphygmomanometer and accessories into any liquid.
 Do not use any sphygmomanometer and accessories that may be damaged or deteriorated.
Cleaning:
 If there is something dirty on the sphygmomanometer and accessories, please wipe it with
a dry and soft cloth.
 If the sphygmomanometer and accessories are very dirty, you can clean them with a soft
cloth dampened with water or neutral detergent after the cloth is full twisted.
Maintenance:
*Please follow the instructions in the user manual. If you do not comply, our company will not
assume responsibilities of the quality.
 Frequently clean the Electronic Sphygmomanometer.

41
 The device should be inspected regularly(or according to the inspection procedure of
hospital). The device can be inspected by inspection organizations which designated by the state,
or inspected by professionals, also, users can contact with our company.

Warning
Do not submerge the device in water.
Advice
 Do not use any naphtha, thinner or gas to clean.

42
 Do not attempt to clean or wash the cuff.

Keeping
Advice
Do not place the machine in the following areas:
 Easy to splash water areas.
 Direct sunlight, extreme hot, humidity, dust, causticity gas areas.

43
 Lean or the area which can cause vibration, impact.

 Chemicals or corrosive gas storage areas.


 Remove the batteries if the unit will not be used for long time.
Warning
It is recommended that you check if there is any damage on the Sphygmomanometer or the
accessories regularly, if you find any damage, stop using it, and contact our Customer
Service immediately.
Warning
The disposal of scrap instrument and its accessories and packing (including batteries,
plastic bags, foams and paper boxes) should follow the local laws and regulations.
EMC declaration:
 When this device is installed or putted into service, EMC should be paid more attention, as
44
the portable and mobile RF communications equipment with higher EM interference can affect
this device.
 The internal components and cables should not be changed,as this may decreased
IMMUNITY of the device.
 The Electronic Sphygmomanometer should not be used adjacent to or stacked with other
equipment.
Classification:
EMC: Group I Class B.
According to the MDD 93/42, the classification of this medical device: Ⅱa.
The type of protection against electroshock: ClassⅡequipment.
The degree of protection against electroshock: type BF applied part.
The degree of protection against ingress of water: IPX1.
According to the mode of operation: Continuous.
According to the degree of safely of application in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence
of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.

45
Chapter16 Specification
Name Electronic Sphygmomanometer
Display 2.8'' color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working modes Automatic
Operation mode Continuous operation
Pressure:0~290mmHg(0~38.6kPa)
Measurement Range
Pulse: 40~240/min
Resolution Pressure:1mmHg
Accuracy Static pressure: ±3mmHg(±0.4kPa)
The BP Value of the device is equivalence with the measurement
Error value of Stethoscopy. The error meets all the conditions in the
ANSI/AAMI SP-10:2002+A1:2003 +A2:2006.
Operating +5ºC~40 ºC . 15%RH~80%RH
Temperature/ Humidity
Transport by general vehicle or according to the order contract,
Transport
avoid pounded, shake and splash by rain and snow in transportation.

46
Temperature: -20ºC~+55ºC; Relative humidity: ≤95%; No corrosive
Storage
gas and drafty.
Atmospheric pressure 700hPa~1060hPa
Power supply 4 "AA" alkaline batteries, AC Adapter(AC, 100V-240V, optional)
When the temperature is 23ºC, limb circumference is 270mm, the
Battery life measured blood pressure is normal, 4 "AA" alkaline batteries cab be
used about 300 times.
Main Unit Dimensions 130(L)*110(W)*80mm(H)
Main Unit Weight 300gram
Safety classification class Ⅱ, type BF applied equipment
The service life of the device is five years or 10000 times of BP
Service life
measurement.

47