(CR-IR 391V CL)   For Veterinary

Reference Guide (Options)


3rd Edition September 2019

This Reference Guide provides instructions and tips on options to FVS-1000 (CR-IR 391V
CL). Please read this Reference Guide before using the product. After you have finished
reading the Reference Guide, keep it in a safe place that is readily accessible for quick
reference. You are also encouraged to read "FVS-1000 (CR-IR 391V CL) Operation
Manual", which contains basic operating information and operation summaries, and
"FVS-1000 (CR-IR 391V CL) Reference Guide", which covers detailed product operating
instructions and precautions, in conjunction with this Reference Guide.
897N200794B
FVS-1000 (CR-IR 391V CL) OPERATION MANUALS

FVS-1000 (CR-IR 391V CL) Operation Manual

FVS-1000 (CR-IR 391V CL) Reference Guide

(Optional)

FVS-1000 (CR-IR 391V CL) Reference Guide (Options)

Trademarks
FCR is a trademark or a registered trademark of FUJIFILM Corporation.

Other holders’ trademarks


Microsoft, Windows, Windows 7, Windows 8 and Windows 10 are the registered trademarks or
trademarks of US Microsoft Corporation in the U.S.A. and other countries.
Windows is the registered trademark of US Microsoft Corporation in the U. S. A. and other countries.
All other company names and product names described in this manual are the trademarks or
registered trademarks of their respective holders.

Copyright © 2016-2019 FUJIFILM Corporation. All rights reserved.

FVS-1000 Reference Guide (Options) 897N200794B


ii
Contents at a Glance

1 Importing Endoscopic Images

2 Importing Ultrasonic Images

3 Importing General-Purpose Images

FVS-1000 Reference Guide (Options) 897N200794B


iii
How to Read This Reference Guide

Basic Page Organization


The pages of this Reference Guide are basically organized into the following elements:

Lead
Chapter title Describes information that is about
A brief explanation of
to be explained in this chapter
the operation that is
toge ther with supplementar y
covered in the chapter.
information.

1. Importing General-Purpose Images


Section Title You can import images (images shot with a digital camera or any other imaging device) stored on your personal computer
(PC) to the FVS-1000. Images imported are stored on the FVS-1000 in association with patient and/or study information, so
A brief explanation of the you can select studies from the “Main window” and edit the associated study information or view general-purpose images
operation that is covered in the “Image Viewer window”.

under the section. 1.1 Displaying the “General-Purpose Image 1.2 Importing General-Purpose Images
Import Window” To import general-purpose images stored on the PC to the
To display the “Examination data importing window” from the FVS-1000, follow these steps.

Explanatory Text “Main window” in the FVS-1000, follow these steps.


1 Click [Select Folder] and select the folder

A summary introduction 1 Select the [JPEG] on the "Import window" at


the lower of "Main window".
containing the image file of interest.
When the “Browse For Folder dialog box” appears, select
Operation procedure
to the topic described the folder and click [OK]. A sequence of numbered
The “Examination data importing window” appears.
under the section, with operational steps to
precautions that should guide the user through an
deserve special notice, if operation.
any.

Thumbnails of the image files contained in the folder


appear.

2 Check the thumbnail of the image you want


to import.
Display Windows
“Examination data importing window” Folder selection field
Leave all other images unchecked.
A window that appears The thumbnailed image can be rotated as needed.

during an operation. 3 Check patient and study information about


the study selected.
You can edit the patient and study information displayed
in the window. You may also invoke patient information
registered in the FVS-1000 database and enter it.
For instructions on how to edit patient information, see
"(4) Check or edit patient information", in "2.
Working with the “Import General-Purpose Images
window”".
For instructions on how to edit study information, see
"(5) Check or edit study information", in "2.
Working with the “Import General-Purpose Images
Thumbnailed image Study information Patient window”".
information
4 Click [Start Import].
The image you selected is imported to the FVS-1000.
A message may appear when you clicked [Start Import].
To see what actions to take when message is
displayed, see "3. Taking Actions Requested by
Messages".

FVS-1000 Reference Guide (Options) 897N200794B


Ge 1

Page number
An indication of the
order of a particular
page in a chapter.

FVS-1000 Reference Guide (Options) 897N200794B


iv
Marks
Marks are used in the descriptions to mark supplementary information and information that must be
observed while using the FVS-1000. For your safety and correct use, always observe a Warning or
Caution.

Indicates a dangerous situation which, if you ignored or could not


follow for some reason, could result in death or serious injury.

Indicates a hazardous situation which, if you ignored or could not follow for
some reason, could result in a slight or moderate degree of personal injury or
material damage.

Indicates information that should be noted.

Indicates information that may be helpful.

Indicates an item that provides details of the procedure or related information.

« Indicates a procedure that is dependent on the situation.

Describes the functions that can be used only on the server PC.

Notation
In this Reference Guide, button names, menu actions, window names and messages appearing in the
operational windows are designated in the following manners:

● Buttons (example)
[Complete]
------------------------------ Names of buttons in the operational windows are enclosed
in brackets [ ], regardless of their shape.

● Menus (example)
[Annotation] – [Rectangle]
--------
Menu names are enclosed in brackets [ ].
Sequences of menu choices are designated in the format [ ]
– [ ].
In this example, users are supposed to choose [Annotation]
first, then [Rectangle].

● Windows (example)

"Read image" window


-----
Window names are enclosed in quotation marks " ".

● Messages (example)

The selected study will be updated


------ to have been verified.
+Are you sure?

Part of a message appearing in the message dialog is


designated.

FVS-1000 Reference Guide (Options) 897N200794B


v
Importing Endoscopic Images Contents

1. Importing Endoscopic Images.................................................. ES-1


1.1 Displaying the “Examination Data Importing Window”........ ES-1
1.2 Importing Endoscopic Images............................................. ES-1
1.3 Confirming Endoscopic Images in the “Viewer Window”.... ES-2
2. Working with the “Examination data importing window”.... ES-3
3. Taking Actions Requested by Messages................................ ES-4
4. Handling of Imported Endoscopic Images............................. ES-5
4.1 Handling in the “Viewer Window” and Limitations.................. ES-5
4.2 Limitations on Storing Endoscopic Images............................ ES-5
4.3 Limitation on Printing Endoscopic Images............................. ES-5
5. Precautions................................................................................... ES-5
5.1 Precautions in Using the Endoscopic Image Importing Function.... ES-5
5.2 Precautions in Using Images of Endoscope Together with
Endoscopic Network Processor Equipment.......................... ES-5

FVS-1000 Reference Guide (Options) 897N200794B


vi
Importing Ultrasonic Images Contents

1. Overview of Ultrasonic Image Import Software..................... US-1


2. Confirming Imported Images..................................................... US-2
3. Handling Ultrasonic Images....................................................... US-2
3.1 Handling in the “Viewer Window” and Limitations.................. US-2
3.2 Limitations on Storing Ultrasonic Images............................... US-2
3.3 Limitations on Printing Ultrasonic Images.............................. US-2
4. Precautions................................................................................... US-2

FVS-1000 Reference Guide (Options) 897N200794B


vii
Importing General-Purpose Images Contents

1. Importing General-Purpose Images......................................... Ge-1


1.1 Displaying the “General-Purpose Image Import Window”...... Ge-1
1.2 Importing General-Purpose Images....................................... Ge-1
1.3 Confirming General-Purpose Images in the
“Image Viewer Window”........................................................ Ge-2
2. Working with the “Import General-Purpose Images window”... Ge-3
3. Taking Actions Requested by Messages................................ Ge-4
4. Handling of Imported General-Purpose Images.................... Ge-4
4.1 Handling in the “Image Viewer Window” and Limitations....... Ge-4
4.2 Limitations on Storing General-Purpose Images.................... Ge-4
4.3 Limitation on Printing General-Purpose Images..................... Ge-4
5. Precautions................................................................................... Ge-4

FVS-1000 Reference Guide (Options) 897N200794B


viii
(CR-IR 391V CL)   For Veterinary

Reference Guide
(Importing Endoscopic Images)
FVS-1000 Reference Guide (Options) 897N200794B
ii
1. Importing Endoscopic Images
You can import still pictures captured on Endoscopic Study device (Endoscopic image) to the FVS-1000 through the
network. Endoscopic images imported are stored on the FVS-1000 in association with patient and/or study information, so
you can select studies from the “Main window” and edit the associated study information or view endoscopic images in
the “Viewer window”.

1.1 Displaying the Examination Data 1.2 Importing Endoscopic Images


Importing Window To import endoscopic images selected in the study
To display the Examination data importing window importing list to the FVS-1000, follow these steps.
from the Main window in the FVS-1000, follow these
steps.
1 Select one study to import an image from
the study importing list.
1 Select the [ES] on the Import window at
Select the study which displays “ES” (Endoscopic Study)
the lower of Main window .
for [Modality]. Thumbnail images in the study displays.
The “Examination data importing window” appears, and Yet, select [Update] to update the dislay of study importing
the list of study which can import displays. list.

2 Check the thumbnails of the images you


want to import.
Leave all other images unchecked.
Thumbnailed images can be rotated as needed.

3 Check patient and study information about


the study selected.
You can edit the patient and study information displayed
in the window. You may also invoke patient information
registered in the FVS-1000 database and enter it.
For instructions on how to edit patient information, see
“(5) Check or edit patient information”, in “2. Working
“Examination data importing window” Study importing list with the “Examination data importing window””.

For instructions on how to edit study information, see


“(6) Check or edit study information”, in “2. Working
with the “Examination data importing window””.

4 Click [Start Import].


The endoscopic image of study selected is imported to
the FVS-1000.
A message may appear when you clicked [Start Import]
depends on the state of study selected.
To see what actions to take when message is
displayed, see “3. Taking Actions Requested by
Messages”.

Thumbnailed image Study information Patient


information

FVS-1000 Reference Guide (Options) 897N200794B


ES 1
1.3 Confirming Endoscopic Images in
the Viewer Window
To d i s p l a y e n d o s c o p i c i m a g e s i n t h e V i e w e r
window , follow these steps:
1 Select the study for the endoscopic images that
have been imported to the FVS-1000 in the “Main
window” and double-click.
[Modality] of endoscopic study is ES .

For details about the “Main window” and the “Viewer


window , see "FVS-1000 (CR-IR 391V CL) Reference
Guide".

FVS-1000 Reference Guide (Options) 897N200794B


2 ES
2. Working with the “Examination data importing window”
Instructions about handling the components of the
“Examination data importing window” are given below. If the PC is linked to an optional electronic medical chart system
(1) (2) or receipt computer, patient information is retrieved from the
destination database, in which case the order of search
depends on the installation setting.

(4)
(5) Check or edit patient information
Patient information relevant to the selected study is displayed. The
(3) information can be edited.
(5)

When the information cannot be received from the endoscopic


devices, the “Species” and “Breed” column will be blank even
for a registered patient.
Patient ID : Required. If an ID number is not displayed, enter
(6)
one without using the following characters:
(9) (8) (7) “, #, %, &, ’, +, ?, \, /, *, @, <
Patient Name : Up to 64 characters can be entered.
Responsible : Enter a responsible person for patient.
The information displayed in the study list is the information set with
person
the endoscopic devices.
Responsible : Enter a name of responsible organization or
When the patient ID set with the endoscopic devices includes different
organization association for patient.
patient information in the FVS-1000 patient information database, the
sections (5) and (6) of the above screen show the information registered in Sex : Required. Selectable from the pulldown menu.
the FVS-1000 patient information database, and the contents of these Neutered state : Select the Neutered state from the pull-down menu.
sections (5) and (6) may be different from the study list. Date of Birth : Enter the date of birth with characters as
directed by the sample entry.
(1) Change the display order of study list
Lists can be sorted by the item in the ascending order by clicking Height : Enter the length of body. Use a unit set in the
the item on the header part of the study importing list. Customize window.

(2) Delete or edit the selected studies Weight : Enter the weight. Use a unit set in the Customize
For the studies selected from the study list, the following functions window.
can be used. Species : Enter a species.
[Delete Study] : Deleting the selected study. Breed : Enter a breed.
- If the confirmation message is displayed - Patient : Enter the comment for patient up to 1024
Studies selected will be deleted. comment words.
* Do you wish to proceed? (6) Check or edit study information
 [OK] : Delete the study. The name of the login user (study doctor) in charge of the selected
 [Cancel] : Cancel the deletion. study is displayed, along with the study start date and time.
Items other than [Study explanation] are required.
[Complete] : Change the recieving status to “Complete”.

When the information cannot be received from the endoscopic


If the checkmark is placed to “Display the message for images devices, the “Study Description” column will be blank even for a
already being imported”, the studies including the study which registered patient.
has finished importing will be displayed in the study list.
Study : Enter a comment on the in up to 64 characters.
(3) Select the images to be imported Description
Thumbnail images of study selected in the study importing list appear. Endoscopy : Selectable from the pulldown menu.
Check the thumbnails of the images you want to import to the
Technician
FVS-1000.
Study start : Select a date from the pulldown menu and
: Check all images date and time enter the time with characters.
: Uncheck all images (7) Import the endoscopic images to FVS-1000
Import the images of checked thumbnails to the FVS-1000, along with
the relevant patient information and study information.
The study data which completes importing images will be The study which completed the importing deletes automatically
deleted, and also images will be deleted. when updating the study importing list.
Images that are unchecked will be deleted without importing, so
check it carefully and then start importing. (8) Updating the study importing list to the latest information
Delete the study data which importing completed from the list, and
(4) Retrieve patient information from the database
display the new study data which can import in the list.
Enter [ID Number] and click when the ID number has not
been retrieved from the Endoscopic Study device. (9) Close the “Examination data importing window”
Then, relevant patient information is retrieved from the FVS-1000 Close the “Examination data importing window” and exit this
database and is displayed. function.

FVS-1000 Reference Guide (Options) 897N200794B


ES 3
3. Taking Actions Requested by Messages In addition, messages may be displayed in the
Meanings of the messages that are displayed when following situations to call for certain actions.
importing images are given below, along with • If the patient or study information’s incomplete, the image
suggested responses. import process is canceled with a message appearing.
Correct the problem and retry the image import process.
• If an image file failing to meet the specifications is selected, the
- If the following message is displayed -
image import process is canceled with a message appearing.
Patient ID specified does not exist on Patient Database. Correct the problem and retry the image import process.
Verify the image file specifications.
This message is displayed if the patient ID you entered does not • If the image import process fails due to a corrupt image file,
exist in the database when searching for patient information with network failure, or any other reason, an error message is
. displayed. Take action suggested in the error message.
« Click [OK].
You may have entered the wrong ID number. Enter the correct ID
number and then click again.

- If the following message is displayed -

Do you wish to reflect the patient information for studies that


are import target upon database?

This message is displayed when the image has been imported


after editing the patient information.
« Click [Yes] to import the study data by updating the database with
the edited version of patient information.
« Click [No] to import the study data without updating the database
with the edited version of patient information.

- If the following message is displayed -


Images that have not been selected exist.
*Unselected images will be deleted after importing has been
completed. Do you wish to continue the processing?
This message is displayed when importing images has been
started with having the image has not been checked on the
thumbnail image.
« Click [OK] to automatically delete unchecked images after importing.
« Click [Cancel] to return to the window before importing, so check
whether or not the check mark is on and import the image again.

- If the following message is displayed -


Import of endoscopic study whose reception status is not
“complete” will start.
*Do you wish to proceed?
>Images received after the completion of import will be
deleted without being imported. Make sure all images have
been received before starting import.
This message is displayed when importing images has been
started for the study in which receiving the image from the
endoscopic device.
« Click [OK] to start importing images, but received images will be
deleted after importing completed.
« Click [Cancel] to return to the window before importing, so click
[Update] to update the study importing list to the latest information,
and then import the image again.

FVS-1000 Reference Guide (Options) 897N200794B


4 ES
4. Handling of Imported Endoscopic Images 5. Precautions
Similar to CR study images in the Viewer window , The following are the precautions in importing
you can view endoscopic images that have been endoscopic images.
added to the Study List in the Main window .
The following describes the characteristics of the 5.1 Precautions in Using the Endoscopic
process of handling endoscopic images in the Viewer Image Importing Function
window and limitations imposed on it.
• The endoscopic images that appear in the “Viewer window”
4.1 Handling in the Viewer Window are reference images and should not be used for diagnostic
and Limitations purposes.
Among the toolbar button functions and the functions of • When deleting images, make sure that it is no more needed,
the right-clicked pulldown menu, the following functions are and use maximum discretion to avoid mishandling.
not available: • Before importing any images, make sure that the images to be
•Reverse •Gradation imported and the associated patient information are correctly
•FCR Image Processing •Menu Change configured to prevent study data from being mixed up.
•Preset •Additional Exposures Because the original file of an image remains undeleted after
•%-Size Image •Exchange Imgs the image is imported, be careful not to import the same
•Film Print image in duplicate.

4.2 Limitations on Storing Endoscopic 5.2 Precautions in Using Images of


Images Endoscope Together with Endoscopic
• Endoscopic images cannot be directed to a DICOM server. Network Processor Equipment
• Density and FCR Image Processing cannot be specified for
Images may not be imported correctly by setting
endoscopic image work disk output.
inproperly or operating incorrectly when using images
• Endoscopic images can be saved only in the JPEG format of endoscope together with the endoscopic network
when they are saved automatically or from the “Main window”. processor equipment.
Those items that are not found in the patient information of
To prevent this, follow these steps:
endoscopic images appear blank.
• Import the endoscopic image on the same day the image is
transferred.
4.3 Limitation on Printing Endoscopic
• Do not change the software of the server settings which use in
Images network transferring.
• Endoscopic images can not be printed on film (optional). • To use the endoscopic image importing function and
authorized processor, perform the endoscopic study in order
to the operation of instruction manual of the processor.
Also, to perform the endoscopic study, you must input the
patient ID in the processor side for using the patient
information and images correctly.
Yet, backup the data if you can backup in the processor side.
• Images can be imported from the inserted media into the
image recorder if using this device together with our authorized
image recorder when the data cannot be transferred due to
the network difficulty.
In such a case, use the recommended drive and media.

FVS-1000 Reference Guide (Options) 897N200794B


ES 5
FVS-1000 Reference Guide (Options) 897N200794B
6 ES
(CR-IR 391V CL)   For Veterinary

Reference Guide
(Importing Ultrasonic Images)
FVS-1000 Reference Guide (Options) 897N200794B
ii
1. Overview of Ultrasonic Image Import Software

The ultrasonic image import software is launched on a sever PC, together with the FVS-1000, to import images from the
ultrasonic diagnostic imaging equipment and register relevant study data automatically. You can manage images imported
as study data the same way as you handle X-ray images, so you can view, save them to disk or other media from both the
server and client PCs.

The workflow from acquiring ultrasonic images on the ultrasonic diagnostic imaging equipment to saving them is shown
below.

Ultrasonic diagnostic imaging equipment FVS-1000

1 Acquire ultrasonic images. 2 Import images


automatically (server PC). FVS-1000

3 Confirm images in
the "Viewer window"
2
(server PC/client PC).

Storage Disk
Work Disk
Miscellaneous Disk

4 Save to media.

FVS-1000 Reference Guide (Options) 897N200794B


US 1
2. Confirming Imported Images 3.2 Limitations on Storing Ultrasonic
Images
Ultrasonic images that have been imported to the FVS-1000
Ÿ Ultrasonic images cannot be directed to a DICOM server.
are added to the Study List in the “Main window”, with “US”
Ÿ Density and FCR Image Processing cannot be specified for
in [Modality].
ultrasonic image work disk output.

If imported images are not shown on the Study List, click Ÿ Ultrasonic study images can be saved only in the JPEG
format when they are saved automatically or from the "Main
[Update] to refresh the list.
window".
Those items that are not found in the patient information of
To display ultrasonic images in the “Image Viewer window”,
ultrasonic images appear blank.
follow these steps:

1 Select the study for the ultrasonic images that 3.3 Limitations on Printing Ultrasonic
have been imported to the FVS-1000 in the “Main Images
window” and double-click.
Ÿ Ultrasonic images cannot be printed on film.

4. Precautions
The following are the precautions in using the ultrasonic
image import function.

Ÿ Do not display ultrasonic images (study data) before their


import completes, with [Study Status] not yet changed to
[Study Completed], in the “Image Viewer window”. Otherwise,
an error would be indicated in the ultrasonic diagnostic
imaging equipment.
To view ultrasonic images in the “Image Viewer window”, click
For details about the “Main window” and the “Image [Update the list] to make sure that [Study Status] has changed
Viewer window”, see "FVS-1000 (CR-IR 391V CL) to [Completed] or the [Number of Images] has reached the
Reference Guide". actual number of images acquired.
Ÿ The ultrasonic images that appear in the “Image Viewer
3. Handling Ultrasonic Images window” are reference images and should not be used for
diagnostic purposes.
Similar to CR study images, you can view ultrasonic Ÿ When deleting any ultrasonic image, make sure that it is no
images that have been added to the Study List in the “Main more needed, and use maximum discretion to avoid
window” in the “Image Viewer window”. mishandling.
The following describes the characteristics of handling Ÿ In handling images, check the nature of the images and
ultrasonic images in the “Image Viewer window” and relevant patient information.
limitations imposed on it.

3.1 Handling in the “Viewer Window” and


Limitations

Among the toolbar button functions and the functions of the


right-clicked pulldown menu, the following functions are not
available:

ŸReverse ŸGradation
ŸFCR Image Processing ŸMenu Change
ŸPreset ŸAdditional Exposures
Ÿ%-Size Image ŸExchange Imgs
ŸFilm Print

FVS-1000 Reference Guide (Options) 897N200794B


2 US
(CR-IR 391V CL)   For Veterinary

Reference Guide
(Importing General-Purpose Images)
FVS-1000 Reference Guide (Options) 897N200794B
ii
1. Importing General-Purpose Images
You can import images (images shot with a digital camera or any other imaging device) stored on your personal computer
(PC) to the FVS-1000. Images imported are stored on the FVS-1000 in association with patient and/or study information, so
you can select studies from the “Main window” and edit the associated study information or view general-purpose images
in the “Image Viewer window”.

1.1 Displaying the “General-Purpose Image 1.2 Importing General-Purpose Images


Import Window” To import general-purpose images stored on the PC to the
To display the “Examination data importing window” from the FVS-1000, follow these steps.
“Main window” in the FVS-1000, follow these steps.
1 Click [Select Folder] and select the folder
containing the image file of interest.
1 Select the [JPEG] on the "Import window" at
the lower of "Main window". When the “Browse For Folder dialog box” appears, select
the folder and click [OK].
The “Examination data importing window” appears.

Thumbnails of the image files contained in the folder


appear.

2 Check the thumbnail of the image you want


“Examination data importing window” Folder selection field to import.
Leave all other images unchecked.
The thumbnailed image can be rotated as needed.

3 Check patient and study information about


the study selected.
You can edit the patient and study information displayed
in the window. You may also invoke patient information
registered in the FVS-1000 database and enter it.
For instructions on how to edit patient information, see
"(4) Check or edit patient information", in "2.
Working with the “Import General-Purpose Images
window”".
For instructions on how to edit study information, see
"(5) Check or edit study information", in "2.
Working with the “Import General-Purpose Images
Thumbnailed image Study information Patient window”".
information
4 Click [Start Import].
The image you selected is imported to the FVS-1000.
A message may appear when you clicked [Start Import].
To see what actions to take when message is
displayed, see "3. Taking Actions Requested by
Messages".

FVS-1000 Reference Guide (Options) 897N200794B


Ge 1
1.3 Confirming General-Purpose Images in
the “Image Viewer Window”
To display general-purpose images in the “Image Viewer
window”, follow these steps.:

1 Select the study for the general-purpose images


that have been imported to the FVS-1000 in the
“Main window” and double-click.

For details about the “Main window” and the “Image


Viewer window”, see "FVS-1000 (CR-IR 391V CL)
Reference Guide".

FVS-1000 Reference Guide (Options) 897N200794B


2 Ge
2. Working with the “Import General-Purpose Images window”

Instructions about handling the components of the “Import


General-Purpose Images window” are given below. Image files compliant with the following specifications can be
imported to the FVS-1000:
Format : JPEG (JFIF)
Color : 8-bit grayscale or 24-bit color
(1) Size : Minimum is 200 x 200 pixels or more
(2)
Maximum is 4000 x 4000 pixels
(4) Check or edit patient information.
(3) Patient information relevant to the selected study is displayed. The
(4)
information can be edited.
Patient ID : Required. If an ID number is not displayed, enter
one without using the following characters:
“, #, %, &, ’, +, ?, \, /, *, @, <
Patient Name : Up to 64 characters can be entered.
Responsible : Enter a responsible person for patient.
(5) person
Responsible : Enter a name of responsible organization or
(7) (6)
organization association for patient.
Sex : Required. Selectable from the pulldown menu.
Neutered state : Select the Neutered state from the pull-down
(1) Select the folder containing the image files of interest
menu.
Select the target folder from the “Folder selection field”.
Date of Birth : Enter the date of birth with characters as
(2) Retrieve patient information from the database
directed by the sample entry.
Enter [ID Number] and click . Then, relevant patient Height : Enter the length of body. Use a unit set in the
information is retrieved from the FVS-1000 database and is
Customize window.
displayed.
Weight : Enter the weight. Use a unit set in the Customize
window.
If the PC is linked to an optional electronic medical chart system Species : Enter a species.
or receipt computer, patient information is retrieved from the Breed : Enter a breed.
destination database, in which case the order of search
Patient : Enter the comment for patient up to 1024
depends on the installation setting.
comment words.
(3) Select the images to be imported
(5) Check or edit study information
Thumbnails of the image files contained in the selected folder
The name of the login user (study doctor) in charge of the selected
appear thumbnailed.
study is displayed, along with the study start date and time.
Check the thumbnails of the images you want to import to the
Items other than [Study explanation] are required.
FVS-1000.
Modality : The modality of the study can be selected from
: Check all images
the pulldown menu. Available choices are XC,
: Uncheck all images US, OT, CR, CT, MR, NM, ES, PT, RG, XA, RF
and DX.
: Rotate a selected image 90 degrees left Study : Enter a comment on the in up to 64 characters.
: Rotate a selected image 90 degrees right Description
Study Doctor : Selectable from the pulldown menu.
If a thumbnail is rotated, the corresponding image is imported
Study Start : Select a date from the pulldown menu and enter
rotated likewise.
Date and Time the time with characters.
Sequence of images : Sort the image from the oldest or
newest. When importing the images, the (6) Import the general-purpose images to FVS-1000
number will be issued according to the Import the images of checked thumbnails to the FVS-1000, along
order of selected images. with the relevant patient information and study information.
Delete the original : If the checkmark is placed, the original Images are not deleted from the PC.
image being imported image being imported will be deleted (7) Close the “Import General-Purpose Images window”
automatically. Close the “Import General-Purpose Images window” and exit this
Display the : If the checkmark is placed, the message function.
confirmation message will be displayed to confirm to delete the
when images are to image when importing the images.
be deleted

FVS-1000 Reference Guide (Options) 897N200794B


Ge 3
3. Taking Actions Requested by Messages 4. Handling of Imported General-Purpose Images
Meanings of the messages that are displayed when importing Similar to CR study images, you can view general-purpose images
images are given below, along with suggested responses. that have been added to the Study List in the “Main window” in the
“Image Viewer window”.
- If the following message is displayed -
The following describes the characteristics of the process of
No patient ID has been input. handling general-purpose images in the “Image Viewer window”
and limitations imposed on it.
This message is displayed when no [Patient ID] has been entered.
« Click [OK].
4.1 Handling in the “Image Viewer window”
Enter [Patient ID] and then click [Start Import] again.
and Limitations
- If the following message is displayed - Among the toolbar button functions and the functions of the
right-clicked pulldown menu, the following functions are not
Modality has not been selected.
available:
This message is displayed when no modality has been entered. ŸReverse ŸGradation
« Click [OK]. ŸFCR Image Processing ŸMenu Change
Select [Modality] and then click [Start Import] again. ŸPreset ŸAdditional Exposures
- If the following message is displayed - Ÿ%-Size Image ŸExchange Imgs
ŸFilm Print
Patient ID specified does not exist on Patient Database.
4.2 Limitations on Storing General-Purpose
This message is displayed if the patient ID you entered does not
Images
exist in the database when searching for patient information with
. Ÿ General-purpose images cannot be directed to a DICOM server.
« Click [OK]. Ÿ Density and FCR Image Processing cannot be specified for
You may have entered the wrong ID. Enter the correct ID and then general-purpose image work disk output.
click again.
Ÿ General-purpose images can be saved only in the JPEG
- If the following message is displayed - format when they are saved automatically or from the "Main
window".
Do you wish to reflect the patient information for studies that Those items that are not found in the patient information of
are import target upon database? general-purpose images appear blank.

This message is displayed when the image has been imported


4.3 Limitation on Printing General-Purpose
after editing the patient information.
« Click [Yes] to import the study data by updating the database with
Images
the edited version of patient information. Ÿ General-purpose images cannot be printed on film (optional).
« Click [No] to import the study data without updating the database
with the edited version of patient information.
5. Precautions
In addition, messages may be displayed in the
following situations to call for certain actions. The following are the precautions in importing general-
purpose images.
Ÿ If the patient information or study information is incomplete,
Ÿ The general-purpose images that appear in the “Image Viewer
the image import process is canceled with a message
window” are reference images and should not be used for
appearing.
diagnostic purposes.
Correct the problem and retry the image import process.
Ÿ When deleting any ultrasonic image, make sure that it is no
Ÿ If an image file failing to meet the specifications is selected, the
more needed, and use maximum discretion to avoid
image import process is canceled with a message appearing.
mishandling.
Correct the problem and retry the image import process.
Verify the image file specifications. Ÿ Before importing any images, make sure that the images to be
imported and the associated patient information are correctly
Ÿ If the image import process fails due to a corrupt image file,
configured to prevent study data from being mixed up.
network failure, or any other reason, an error message is
Because the original file of an image remains undeleted after
displayed. Take action suggested in the error message.
the image is imported, be careful not to import the same
image in duplicate.

FVS-1000 Reference Guide (Options) 897N200794B