SOP For PH Meter
SOP For PH Meter
TABLE OF CONTENTS
1. PURPOSE
2. SCOPE
3. RESPONSIBILITIES & AUTHORITIES
4. OPERATING INSTRUCTION
5. PRECAUTION
6. CLEANING AND MAINTAINANCE
AMENDMENT STATUS
1. Purpose:
It is established for the proper operation of pH meter.
2. Scope:
It is applicable in the testing in quality control department.
3. Responsibilities and Authorities:
Quality Control Manager
Quality Control Officer
Asst. Quality Control
4. Operating Instructions:
Wash pH electrode with distilled water carefully and clean its external surface with tissue paper
Dip this electrode in Buffer pH 4.00 and press CAL
Allow the meter to read buffer and then again press CAL
The device enter in calibration mode and show CAL 1
The screen shows END to confirm the calibration process completed
Again wash pH electrode with distilled water carefully and clean its external surface with tissue
paper
Dip this electrode in Buffer pH 7.00 and press CAL
The device enter in calibration mode and show CAL 2
The screen shows END to confirm the calibration process completed
Dip this electrode in Buffer pH 10.00 and press CAL
The device enter in calibration mode and show CAL 3
The screen shows END to confirm the calibration process completed
Press ENTER to end two point calibration
Again wash pH electrode with distilled water carefully and dip in the sample whose pH is to be
measured
Stir gently with electrode and allow to stable
Note the reading
Wash pH electrode with distilled water carefully and place in 3M KCl solution.
5. Precautions:
Avoid buffer contamination
Protect electrodes from damage
6. Cleaning and maintenance:
Clean the external surface with dry clean dust.