PAD M3046A Software Revision E Data Sheet

M3/M4 Compact Portable


Patient Monitor
for the M3001A Multi-Measurement Server

Measurement Features
• Patient demographics, measurement data and
trends, acquired and stored in the MMS, can be
transferred to the M-Series (M3/M4)3 and Philips
IntelliVue (MP60/MP70/MP90) patient
monitors.

• EASI™-derived 12-lead ECG4 using a 5-electrode


cable set.

• Multi-lead Arrhythmia.

• Up to 3 channels of ECG with basic Arrhythmia


and ST Analysis capability.

• FAST-SpO2 measurement, using Fourier Artifact


Suppression Technology, for accurate performance
even with low perfusion.

• Choice of Philips, traditional Nellcor5 and Nellcor


The M3/M4 Compact Portable Patient Monitor
OxiMax compatible pulse oximetry.
(M3046A)1is a small, light patient monitor offering
a flexible monitoring solution to suit a broad • Respiration, Non-invasive and Invasive Blood
spectrum of monitoring needs. Pressure, and Temperature measurements.
The monitor works together with the M3001A2 • Sidestream CO2, a second Invasive Blood Pressure
Multi-Measurement Server (MMS) and the
and Temperature measurements, and enhanced
M3012A and M3015A Measurement Server
Arrhythmia capability, available optionally.
Extensions to give you standard-of-care
measurements.

3.Software Revision E only.


4.EASI™-derived 12-lead ECGs and their measurements are
approximations to conventional 12-lead ECGs. As the 12-lead
1.This data sheet describes monitors with Software Revision E. ECG derived with EASI™ is not exactly identical to the 12-lead
The M3046A is also available with Software Revision D conventional ECG obtained from a electrocardiograph, it
(M3046A option #C90) and there is a separate data sheet for should not be used for diagnostic interpretations.
this (publication number: 4522 981 83061). 5.Nellcor and OxiMax are trademarks of Nellcor Puritan Bennett
2.Not compatible with M3000A. Inc., a part of Tyco Healthcare.

PAD
Usability Features Display
The color display has a wide viewing angle, and
• Choice of serial interface for a dedicated external
strip chart recorder (as default), or IrDA interface excellent visibility from a distance, so you can easily
for connecting to a local printer. recognize and see the data you need for decision
making.
• 3-wave or 4-wave displays are available. For applications where a larger display is required, an
• Trend data, with manual and automatic event additional display can be connected to the monitor
storage, together with a range of report styles help via the standard VGA output (consult your Philips
you track and document your patient’s progress. representative for further information).

• Color display with a wide viewing angle. Networked Systems


• Optional battery informs you of remaining
In larger installations, several of these monitors can
capacity, and allows longer transport times. It can
be networked and connected to a Philips
be easily exchanged during transport with no loss
Information Center — with or without wires, for
of settings.
ultimate flexibility. Such a networked system not
• TouchBar human interface — quick to learn and only allows central monitoring, but also bed-to-bed
easy to use. overview on any connected monitor. A permanent
• Rubber bezel protector, available in four colors, indication at the top of the screen conveniently
gives additional protection in transport scenarios. shows the alarm status for beds assigned to the
caregiver’s group at all times.
Patient Transfers
Patient demographics, measurement data and trends Intended Use
(for up to the last eight hours at one minute
The M3 and M4 patient monitors are designed for
resolution for all numerics), are stored in the MMS,
use by clinicians for adult, pediatric and neonatal
and can be uploaded to other Philips patient
patients in Critical and Intermediate Care areas, in
monitors - up to the last four hours to the M-Series
Operating Rooms and General Wards, and for
(M3/M4)6, and the full eight hours to the IntelliVue transport inside and outside the hospital.
family.
U.S. Federal Law restricts this device to sale by or on
the order of a physician.

M3001A/M3046A Monitor and


Measurement Server Safety
Specifications
The M3046A monitor together with the M3001A
Patients can be transferred by disconnecting the Measurement Server and the M3015A Measurement
MMS from the monitor, and then reconnecting it at Server Extension comply with the Medical Device
a new monitor. Directive 93/42/EEC (CE0366).
In addition, the product complies with
Local Recorder
• IEC 601-1:1988 + A1:1991 + A2:1995
You can connect a dedicated, standalone strip chart EN60601-1:1990 + A1:1993 + A2:1995
recorder (GSI Lumonics XE-50p, order number • UL 2601-1
M3080A #H20) to your monitor via the optional • CAN/CSA C22.2#601.1-M90
serial recorder interface (see Ordering Information). • JIS T 1001-1992
The following recordings
are supported: Monitor Physical Specifications
• Real-time local strips
Size (W×D×H)
• Delayed local strips
Without Handle:
• Alarms
• Vital signs 210mm × 135mm × 210mm
• Trends and stored (8.27" × 5.31" × 8.27")
events
Weight
Without battery:
6.Software Revision E only. 3200 g (7.05 lb.)

2
Monitor Environmental Specifications Interfaces
- Server link between monitor and measurement
Temperature Range (without wireless server
network interface)
- Network (RJ45)
Operating: - Infrared Interface to local printer
0 to 45°C (32 to 113°F). - Serial Interface to local recorder (supported
Storage:
recorder: GSI Lumonics XE-50p)
- Nurse Call relay (1/8" phone jack, ≤60W, ≤2A,
-20 to 60°C (-4 to 140°F).
≤36VDC, ≤25VAC, active=closed)
- ECG Output/Marker Input (1/4" stereo phone
Temperature Range (with wireless network
interface)
jack with tip, ring, sleeve)
ECG Output:
Operating:
Signal Gain: 320 to 3200 (depending upon ECG
0 to 35°C (32 to 95°F). display gain)
Storage:
Full Scale on Display: 3.2Vpp
-20 to 60°C (-4 to 140°F).
Gain Error: <20%
Humidity Range
Baseline Offset: <150mV
Operating:
Noise:<20mVRMS at signal gain 1000
95% RH max. @ 40°C (104°F)
Storage: Bandwidth: 1 to 80Hz in ECG Diagnostic mode
85% RH max. @ 50°C (122°F) Output Impedance:
<2.2KΩ ±20% ECG Output (ring)
Altitude Range <2.5kΩ ±20% ECG Output/Marker Input (tip)
Operating: Signal delay: <30ms
-500m to 4,600m
Marker Input Requirements:
(-1,600' to 15,000')
Signal Type: 0 to -12V, negative edge pulse.
Storage:
-500m to 13,100m Pulse Source Impedance: <7kΩ
(-1,600' to 43,000') Pulse Fall Time: <100µs

Electrical Specifications Pulse Duration: >4ms


VGA Interface
100 to 240VAC, 50/60Hz,
0.4 to 0.7A Refresh rate: 60 Hz.
Power Fail Protection. Horizontal frequency: 31.5 kHz
Resolution: 640 x 480 pixels
Monitor Performance Specifications
Video Signal: 0.7 Vpp @ 75 Ohm
Display HSYNC/VSYNC signal type: TTL
6.5" (diagonal) active color LCD (TFT) Connector: 15-pin D-SUB
Resolution: Wireless Network Interface (optional)
640 × 480 pixels Antenna: integrated into handle.
Technology: Frequency Hopping Spread
Indicators
Spectrum (FHSS)
- Up to 4 waves (depending on monitor option)
Frequency band: 2.4 GHz Industrial Scientific
- Alarms Off (red crossed-bell LED).
Medical (ISM) band
- Alarms (red/yellow LED, 3 different alarm
tones). Output Power: 100mW (max.)
- On/Standby (green LED).
Weight: < 200g
- AC Power (green LED)
- Battery Status (green/yellow/red LED).
- QRS tone or tick, or SpO2 modulation tone.

3
Battery (optional) Ordering Information
Moltech NJ1020HP Smart Battery, 3500mAh
(typical), Nickel-Metal-Hydride (removable). Also
Monitor Options
possible: Toshiba TR36, Duracell DR36 M3046A #A03
M3 with 3 wave channels.
Discharge Time: Supports C02 from M3015A Measurement Server
2.5h minimum [new and fully charged battery, Extension.
with NBP every 15 minutes, ECG, SpO2, and M3046A #A04
automatic reduction of display brightness.] M4 with 4 wave channels.
Supports C02 from M3015A Measurement Server
With a CO2 measurement in use, typically 15 Extension, and the Pressure/Temperature channel from
minutes shorter. M3012A Measurement Server Extension.
M3046A #C01
Charge Time (after connection to AC power):
Add enhanced arrhythmia capability
- typically 24h with monitor on and functioning.
M3046A #C15
- typically 4.5h with monitor off. Add basic network capability
Weight: M3046A #C16
600 g (1.3 lb.) Add full network capability
(requires #C15)
Real-time Clock M3046A #C40
Add battery.
Accuracy: M3046A #C90
<2 seconds per day. Software Revision D (works with M3000A, not with
M3001A). No local recorder support.
Operating Time:
M3046A #C91
6h without battery or AC power, otherwise Software Revision E (works with M3001A, not with
unlimited. M3000A). Support for local recorder M3080A #H20.
M3046A #J14
Active Settings and Stored Data Add IrDA Interface to connect to local printer
(you cannot also order M3046A #J16)
Operating Time:
M3046A #J16
6h without battery or AC power, otherwise Add Serial Interface to connect to local recorder
unlimited. (you cannot also order M3046A #J14; not compatible
with M3046A #C90)
Trends M3046A #J20
Add Wireless Network Interface
Short trend:
(requires #C15).
10 hours with 1 minute resolution. Complies with the requirements of the
Council Directive 1999/5/EEC of 9 0560
Long trend:
March 1999 (Radio and Telecommunications Terminal
48 hours with 5 minutes resolution. Equipment Directive)
Both trends are stored.
Monitor Mounting Options
M3080A #A10
Table Top Mount
M3080A #A11
Universal Hanger (including Table Top Mount)
M3080A #A13
Wall Rail
M3080A #A14
Tilt/Swivel Wall Mount
(includes wall rail)
M3080A #A15
GCX Wall Channel
M3080A #A21
Transport Bed Hanger
M3080A #A22
Roller Top Stand
M3080A #C05
Universal Pole Mount Clamp
M3080A #C06
Mounting Clamp for transport of Measurement Server
and Measurement Server Extension
M3080A #C10

4
Monitor Rail Mount Local Printer Options
M3080A #H05
Measurement Server Mounting Wireless Infrared Printer Connector
Options
M3080A #A01 Local Recorder Options
Measurement Server Mounting Plate For GSI Lumonics XE-50p:
(pack of five plates)
M3080A #H20
M3080A #A02
GSI Lumonics XE-50p Strip Chart Recorder
Rotating Clamp Mount (Software Revision E and Serial Interface connector
required)
Monitor Accessory Options M3080A #A30
Roller Stand Mount
M3080A #C12
Monitor carrying case M3080A #A31
Wall Mount
M3080A #C32
12V Adapter for use with a vehicle 12V power supply M4816A
(for CE countries only) Paper, 10 Rolls
[Original Equipment Manufacturer Product Number: M4817A
NotePower 75/NotePower 75i] Paper, 80 Rolls
M3080A #C40 M3081A #A70
Spare battery 1.0m serial recorder cable

Upgrade Options Battery Charger/Conditioner


M3060A #A01 M8043A
Upgrade software to the latest revision (E) Monitor battery charger and conditioner. Supplied
M3060A #A03 without power cord, orderable separately.
Upgrade M2 to M3
M3060A #A04
Upgrade M3 to M4
M3060A #C01
Upgrade to full Arrhythmia
M3060A #C15
Upgrade to basic network
M3060A #C16
Upgrade to full network
M3060A #J14
Upgrade to IrDA Interface for connection to external
printer (exchanging the Serial Interface for the external
recorder)
M3060A #J16
Upgrade to Serial Interface for connection to external
recorder (exchanging the IrDA Interface for the
external printer). Revision E software is required.
M3060A #K01
Rubber Bezel Protector - Gray
M3060A #K02
Rubber Bezel Protector - Blue
M3060A #K03
Rubber Bezel Protector - Aqua
M3060A #K04
Rubber Bezel Protector - Yellow

5
Philips Medical Systems is part of Asia
Tel: +852 2821 5888
Royal Philips Electronics
Europe, Middle East, Africa
Interested? Tel: +31 40 27 63005
Would you like to know more about our
imaginative products? Please do not hesitate to Latin America
contact us. We would be glad to hear from you. Tel: +55 11 2125 0764

On the web North America


www.medical.philips.com Tel: +1 800 229 6417

Via e-mail
© 1995-2005 Koninklijke Philips Electronics N.V.
[email protected] All Rights Reserved.

By fax Printed in The Netherlands.


+31 40 27 64 887 4522 962 05851 JULY 2005

By postal service
Philips Medical Systems
Global Information Center
P.O Box 1168
5602 BD Eindhoven
The Netherlands

0366

This device complies with the requirements of the


Council Directives 93/42/EEC (Medical Device
Directive) of 14 June 1993.