DNA Technology - DTprime Operation Manual I
DNA Technology - DTprime Operation Manual I
DTprime
Real-Time PCR instrument
PART I
Operating Procedure
TS 9443-004-96301278-2010
Protvino
2
Safety Information
IMPORTANT: READ THIS MANUAL CAREFULLY BEFORE USING THE INSTRUMENT
General Safety
Electrical Safety
Prior to connecting the instrument to an electric grid, check the presence of protective
grounding in the socket to which the instrument will be connected as well as integrity of the
power cable. Connection to the socket without grounding is prohibited. Please use the
power cable provided in the set of delivery. If liquid is spilt inside the instrument, disconnect
it from the power supply immediately and contact local service center or distributor.
Please connect the instrument to an electric grid with the corresponding voltage, which is
mentioned on the identification plate of it.
During Operation
Do not expose the instrument to heat and sunlight or bright light from other intensive light
sources.
3
Maintenance
Do not open the cabinet of the instrument! It does not contain user serviceable
components.
Note! Settings of motor controllers, calibration parameters of optical and temperature
modules of the instrument cannot be changed by the end user. Calibration of the instrument
has to be performed by the manufacturer in accordance with the internal regulations of quality
control. Calibration data, if necessary, can be provided by the service department of DNA-
Technology Research& production
Operational Safety
indoors;
altitude up to 2 000 m;
wear suitable protective clothing, gloves and secure eyes and face;
do not pipette solutions by mouth;
do not eat, drink, smoke or use cosmetic products in the operational area;
after handling samples and chemical reagents wash your hands carefully;
dispose of leftover / used reagents and waste in compliance with the regulations in
force;
read the instruction for use provided with the PCR kit before running the assay;
while running the assay, follow the instructions for use;
do not use a PCR kit after its expiration date;
use only the chemical reagents provided along with the kit and those, which are
recommended by the manufacturer;
do not mix chemical reagents from different PCR kits or lots.
Environmental Effect
Deinstallation of the instrument for maintenance or recycling purposes: if properly operated,
the device doesn’t bear biologically hazardous.
Recycling of the instrument should be carried out in accordance with the regulations in force
at the area. The device does not contain any substances, which may represent a direct threat to
the environment.
4
Notice
This instruction for use contains information protected by copyright. No part of it may be
reproduced without preliminary written consent of “DNA-Technology, Research &
Production”, LLC. Any software products, equipment and designations mentioned in the text
may represent trademarks of their holders.
5
Table of Contents
1. General Information ............................................................................................................................ 8
1.1. Intended Use and modifications of the instrument ....................................................................... 8
1.2. Delivery Set .................................................................................................................................. 9
1.3 Technical specification .................................................................................................................. 9
1.4. Operating Conditions.................................................................................................................. 11
1.5. PC Requirements ........................................................................................................................ 11
2. Instrument description and principle of operation ............................................................................. 12
2.1. Fast Thermal Control System ..................................................................................................... 14
2.2. Optical system ............................................................................................................................ 14
2.3. Control and Display System ....................................................................................................... 15
2.4. Marking. Symbols used for marking. ......................................................................................... 16
3. Setting up the instrument ................................................................................................................... 18
3.1. Instrument Unpack ..................................................................................................................... 18
3.1.1. Instrument Carrying............................................................................................................. 18
3.2. Instrument installation ................................................................................................................ 18
3.3. Software Installation ................................................................................................................... 19
3.3.1. Instrument Driver Installation.............................................................................................. 19
3.3.2. Installation of RealTime_PCR software .............................................................................. 25
4. Operating Procedure .......................................................................................................................... 29
4.1. Instrument Switching-on and RealTime_PCR Launch ......................................................... 29
4.2. Preparing the Instrument for PCR .............................................................................................. 30
4.2.1. Checkup of Geometrical Settings ........................................................................................ 30
4.2.2. Optical Measurement Exposure Setting .............................................................................. 31
4.2.3. Checking the Cleanness of Thermal Plate Wells ................................................................. 32
4.2.4. Tubes Height Measurement ................................................................................................. 32
4.3 Creation of Plate protocol and Amplification Program ............................................................... 33
4.3.1. Filling in of Plate protocol ................................................................................................... 33
4.3.2. Creation/editing of Amplification Program ......................................................................... 34
4.4. Placement of Tubes with Samples ............................................................................................. 35
4.5. Startup of the amplification program.......................................................................................... 37
4.6 Amplification program running ................................................................................................... 38
5. Maintenance ...................................................................................................................................... 40
5.1 General Guidelines ...................................................................................................................... 40
5.2 Maintenance Procedure ............................................................................................................... 41
5.3 Disinfection requirements of the instrument prior to maintenance and repair procedures. ......... 41
5.4 Troubleshooting........................................................................................................................... 42
6. Storage and shipment ........................................................................................................................ 42
7. Recycling ........................................................................................................................................... 43
8. EMS declaration ................................................................................................................................ 43
9. Warranties.......................................................................................................................................... 44
10. Packing Certificate .......................................................................................................................... 45
11. Acceptance certificate...................................................................................................................... 45
12. Cards of warranty repair .................................................................................................................. 47
13. List of claims and instrument disinfection procedure...................................................................... 49
6
Introduction
The “DTprime” Detecting Thermocycler (hereafter – the “instrument” or “device”) is
intended to conduct qualitative and quantitative PCR analysis dropping out electrophoresis
stage of PCR products in agarose gel when using PCR kits based on the fluorescent detection
principle. Measurement of the quantity of accumulated product of PCR amplification takes
place right in the course of reaction temperature cycles (real-time PCR). The qualitative
analysis is based on evaluation of kinetics of PCR visible part, carried out using relevant
mathematical tool.
The use of fluorescent detection of PCR products has a list of substantial advantages:
high detection specificity (if oligonucleotide probes are applied, which allow detecting
only of certain amplicons);
high efficiency;
reduced assay time;
carrying out detection in a closed tube, which practically eliminates contamination of
further experiments;
availability of quantitative assessment of initial DNA-matrix;
registering and accounting of the data in electronic format.
7
1. General Information
1.1. Intended Use and modifications of the instrument
The “DTprime” real time PCR instrument is intended for carrying out an assay using
polymerase chain reaction (PCR) method and accumulated product detection in a real time
(real-time PCR). Detection of the accumulated PCR amplification product occurs directly
during the execution of the amplification program, which increases the informative value of
the study.
The field of application - clinical laboratory diagnostics in vitro;
Purpose of use – diagnostics;
Sample type - samples of human DNA and microorganisms extracted from samples of
biomaterial;
The instrument is manufacturing in the following modifications: 2М1, 2М3, 2М6, 4M1, 4М6,
5M1, 5M3, 5M6, 6М6, 2Х1 4X1, 5X1, 6Х1.
Modification of the instrument is defined by the following structure – XXX, where:
1) 1st X is figure from 4 to 5 and it indicates the number of optical detection channels;
2) 2nd X is the symbol M or X and it determines the thermal block format: M - 96 wells of
200 l, X - 384 wells of 45 l.
3) 3rd X is figure 1, 3 or 6 and it means a thermal block version such as 1 – monoblock; 3 or
6 - the number of thermal block sections.
The instrument is designed for simultaneous analysis of up to 96 or up to 384 samples
depending on the modification (M or X).
Preparation and initialization of experiments on the instrument is realized with an IBM-
compatible personal computer, and further implementation of a preset program may be
executed autonomously. On completing the amplification program, the instrument goes to the
storage mode, and if the option “Turn off device after the run” is activated on the window
“Running”, the instrument goes to the power standby mode.
The optical measurements results may preliminarily be evaluated during the execution of the
amplification program, whereas the final evaluation – at any time after completing the
program. All data of the last amplification are saved in the internal memory of the instrument
and may be read out even in the case of power failure during amplification. As soon as the
mains supply is restored, the instrument continues to function with full recovery of the
amplification program status. Neither failure of the operating system, nor disconnection of the
external computer will halt the experiment. Thus, the instrument is highly protected against
major factors effect, which may result in a failure during PCR implementation.
One computer may control operation of several instruments, including ones of different types
(DTlite, DTprime) connected to it via USB ports.
To operate the instrument, the RealTime_PCR software is used, which allows performing the
following operations:
Loading and unloading of analyzed samples may be automatic in case if the instrument is
built in a robotic system.
Quantity
No. Description Reference Producer
pcs.
DNA-
1 DTPrime real time PCR instrument Technology 1
RT96-01IN
R&P
DNA-
2 Instruction for use (Part I and Part II) RT96-04IFU Technology 1
R&P
3
Computer communication cable USB 2.0 — China
1
High Speed А-В
DNA-
7 Packing unit RT96-03PG Technology 1
R&P
The instrument may be delivered along with portable PC with the pre installed software in
accordance with customer’s request. Note: PC should be purchased along with the order of
the instrument.
9
Thermal block format (modification X) 384-well microplate
10
Number of optical channels 4 (versions 2M1, 2M3,2M6,2Х1) 4 (versions
4M1, 4M3,4M6) 5 (versions 5M1, 5M3,5M6,
4X1,5X1) 4 (versions 6M1, 6M3,6M6,6Х1) *
Excitation/Detection wavelengths for each channel, 470/515, 530/560, 580/620, 630/660,687/731(*)
nm
Weight 27 kg
* spectral characteristics of optical channels can be changed within the specified limits 450 ÷
750 nm in accordance to customer request.
1.5. PC Requirements
The instrument operates under the control of IBM-compatible personal computer with the
following minimum requirements:
512 Mb RAM;
Realtime PCR software to control the instrument and perform data analysis.
11
The software may be downloaded for free from the company’s official website at
http://www.dna-technology.ru/eng/support/
The software is compatible with the following operating systems: Windows 98, Windows
ME, Windows 2k/XP, Windows Vista, Windows 7, Windows 8, Windows 10.
LCD panel;
12
Fig.2. Front view of DTprime instrument
The instrument rear panel accommodates the following (fig 3):
power switch;
two fuses;
power input;
USB connection (type B).
in M modification – 96 wells for standard 200 mcl PCR tubes with distance between
the rows as in standard 96-well plates. The well shape provides optimal heat transfer to the
samples with volume up to 50 mcl.
14
Optical path (light beams transmission module) ensures the transmission of the luminous
flux from light sources of the optical module to the tubes, and back transmission of the
fluorescent luminous flux from the tubes to CCD camera.
The module for light beams transmission is a rectangular duct. Rectangular mirrors located in
the duct corners provide the optical pathway in a required direction.
The system of light locks provides an absolute isolation of the optical channel from external
light.
opening and closing of the thermal block for placing the tubes into the thermal plate
before PCR running;
precise positioning of the thermal block under optical elements in their operating
position.
When the instrument operates under computer control, one can open or close the thermal
operator’s name;
Real-time PCR
The following warning signs “Fragile”, “Top”, “Keep dry”, “Stacking is forbidden” are
placed on the shipping box as well as:
16
Symbols used for marking:
« Fragile »
«Top»
« Keep dry »
« Stacking is forbidden »
Date of production
Manufacturer details
High voltage
17
3. Setting up the instrument
3.1. Instrument Unpack
DTprime real time PCR instrument is delivered packaged in a cardboard or wooden box with
inserts of polyethylene foam; the package is intended to protect the instrument from
mechanical damage when shipped or stored.
Before being packaged in the box, the instrument is placed to a special bag of durable and
waterproof material with zip fastener and transportation straps.
Caution! Heavy! The instrument has to be carried to the installation’s place in the
manufacturer’s package box by two persons.
Before taking the instrument out of the package box, perform the following procedures:
During operation, the instrument may be carried over a short distance within the facility in
unpacked condition by two persons or using movable table (trolley) to other preliminary
prepared working place in accordance with corresponding precautions.
Note! In case the instrument is transported over long distances or shipped by a carrier, it has
to be placed into the original manufacturer’s package.
Caution! When instrument is moved the thermal block has to be in closed position.
Otherwise, in case of some problem with instrument please contact service department of
local distributor or “DNA-Technology” directly.
18
Prior to connecting the instrument to the supply voltage, check the presence of protective
grounding in the socket to which the instrument will be hooked up as well as integrity of the
connecting cable.
Caution! Hooking up to the socket without grounding is prohibited.
It is important that the instrument be correctly grounded. The instrument is supplied with
power cord that provides the instrument grounding, when connected to an appropriate power
outlet.
Also it is important that the instrument and PC must be connected to the same grounded
power outlet. Otherwise, there is a risk of the instrument damage.
Caution! After transport or storage under humid and cold conditions, dry out the instrument
(2-3 hrs) at indoor temperature (+18-25°С) before connecting it to the supply voltage. During
drying out the intrinsic protection may be impaired.
Driver installation for the instrument is identical with one for any standard device and should
be performed as follows:
Insert the USB flash drive with the drivers into the USB port of PC or download
drivers from DNA-technology official website at http://www.dna-technology.ru/eng/support/
Turn on the instrument using the power switch on the back panel.
The operating system indentifies a new device in 1-2 minutes (wait for message
In the next window Driver Updating select Manual driver search and installation,
click Browse, select Drivers_DT96x_W32_2014 folder (for 32-bit version) or
Drivers_DT96x_W64_2014 folder (for 64-bit version).
20
Fig.6. Selecting a type of installation
Select a path to the driver folder by ticking Add this location to the search and click
(Fig.7).
21
Fig. 8. Selecting the Drivers_DT96x folder on the USB flash drive
In the pop-up window Browse For Folder select Drivers_DT96x folder on the USB
flash drive and click (Fig.8).
Making sure the driver folder path is correct click (Fig.9),
22
Fig.10. Further Driver_DT96x installation
In the next window Locate file find this file in the Drivers DT96x folder on the CD,
select it and click (Fig.12)
23
Fig.12. Confirmation of Dt964R.sys file selection on the USB flash drive
Making sure that the path to the file in the window (Fig.11) is correct, click
.
After message that the installation of the driver has been completed (Fig.13), click
.
24
Note! Installation of the instrument’s driver is required upon the first connection only.
Note! The software must be installed on the computer by a user with administrative
privileges.
Make sure you are logged in with administrative privileges.
Installation of RealTime_PCR software should be performed as follows:
Launch Setup manager from the USB flash drive included in the delivery set (it can be
also downloaded official website http://www.dna-technology.ru/eng/support/).
Select Russian/English language to use during the installation (Fig. 14) and click
25
Select a folder for software installation (Fig. 16). RealTime_PCR software can be
installed into a default folder or any folder on the hard disk by clicking
(installation into a default folder is preferable).
- Select the name of the folder, where RealTime_PCR shortcuts will be created, and click
. (Fig. 17).
26
Fig. 17. Selecting of folder to create shortcuts
Confirmation window will appear, informing that software is ready for setup (Fig. 18).
Check the chosen options and click .
After the installation is completed you will see following window (Fig. 19). Click
to quit the Setup Wizard.
.
27
Fig.19. Completed operation of the RealTime_PCR installation wizard
Choosing the Show List_Device shows the list of instruments connected to the PC.
28
4. Operating Procedure
4.1. Instrument Switching-on and RealTime_PCR Launch
Check whether the instrument is connected to the PC via USB cable, and then switch the
instrument and the PC on.
Start the RealTime_PCR software using icon on the Desktop or via the Start menu of
the Windows. In the window of selecting operation modes (Fig. 20) Guest operator is a
default (not recommended to use). Select another operator by using drop-down menu or add a
new one (drop-down menu last line). Please type in the name of Operator and select the
Directory where run protocol files and assay results be stored. Click Ok and select Device
operation.
29
The instrument will be connected to the PC and the indication in the status line of the
operation procedure window will be changed to .
After instrument self testing, indication of its status in the status line will be changed to
and the instrument will be ready for running the amplification program.
A detailed procedure of starting and operating RealTime_PCR software is described in the
second part of the operation manual named Software guidance.
Indication of instrument status in the status line means:
The status line represents current date and time , instrument serial
number and a version of optical controller firmware .
Attention! At the first switch on of the instrument after its transportation or any other change
of its position, checking the optic module’s geometrical settings is strongly recommended.
A detailed procedure of the geometrical settings checkup is described in the second part of the
operation manual named Software guidance.
The image of thermal plate wells outlining the boundaries of a measuring region (red circles)
has to completely encompass the light spots. If the circles are displaced relative to the light
spots, correction of the geometry of an optical image must be carried out (Fig. 22).
30
a)
b)
Preset working exposure values for each channel are the following:
Fam – 1000; Hex – 1000; Rox – 1000; Cy5 – 1000; Cy5.5 – 1000
Note! Preset parameters of optical measurements for each channel are optimal to operate with
DNA-Technology PCR kits and could be suitable for the most of third party kits.
Exposure change is required only when levels of different channels’ optical signals
substantially differ, as it may result in incorrect color compensation in the process of analysis
of optical measurements. For exposure setting window see Fig. 23.
Caution! In case of using third party kits the optimal exposition values for channels must be
provided by manufacturer.
31
Fig.23. Exposure setting window
A detailed procedure of exposure value selecting is described in the second part of the
operation manual named Software guidance.
While checking the optical unit geometrical settings, one has to increase the value of
exposure in 2 times in comparison with the operational values. There must be no bright spots
on the image of empty thermal plate wells.
It is recommended to check the purity of the wells before placing tubes, strips or plates with
PCR mixes, if there was a risk of contamination before.
Note! Routine check of well cleanness must be carried out by the operator in compliance with
the instrument maintenance procedure (see item 5).
Attention! The height of the test tubes must be measured every time new type or brand of
plastic consumables is used, to ensure proper pressure of the hot lid.
32
Save measured value by clicking .
Caution! To measure tubes height, use at least 32 tubes or 4 strips putting them evenly
throughout thermal plate.
The window for plate protocol filling in (Fig. 24) is displayed immediately after starting
RealTime_PCR and comprises: data input table, control buttons for plate protocol filling in
and a field of graphic representation of tubes arrangement in the thermal block.
by using Test procedure with saved standard assay parameters (recommended for
standard assays in clinical laboratories);
Note! Do not fill in the plate protocol without using Test procedure, as window Analysis
Parameters won’t be active, and the optical calibration coefficients of the instrument
will not be used.
In the course of the filling in the plate protocol, enter the following data:
fluorophore is used (fluorescent label) and its purpose (specific target, IC or not
detected);
operator’s name.
The procedure of creating (editing) the plate protocol is described in details in the second part
of the operation manual Software guidance.
33
Fig.24. Window of plate protocol filling-in
After filling in the plate protocol proceed to the Start run tab. The amplification program,
specified in the Test procedure used for plate protocol filling in, will be displayed as the
default. If the plate protocol was filled in without using the Test procedure, the amplification
program of the last run will be displayed.
34
Fig.25. Creation of amplification program using template
a)
b)
Fig.27. Exterior appearance of the thermal block in open position (a – M modification, b – X
modification)
1. Click in the Start run tab or push the button of thermal block manual
control and wait until the thermal block is opened (Fig.27).
2. If it is the first start of the instrument, rub the wells of the thermal block with a cloth
moistened in alcohol (96-% ethanol or 100-% isopropyl alcohol may be used) as described in
section 8 of instrument maintenance. In the course of operation, wipe the wells as set forth in
the Maintenance section.
3. Place the tubes with samples into the thermal block wells in compliance with filled in plate
protocol.
The same message will appear at the instrument’s display as well. When warming up is
completed, the instrument will automatically proceed to the RUN execution mode.
37
In case of necessity of instrument’s autonomous functioning after the RUN completion and its
automatic turning to sleep mode perform the following:
1. In Running window activate the option - turn off device after the RUN
2. Close the Real Time PCR software by clicking in the upper right corner of its
window.
3. When RUN is completed the instrument will switch to the sleep mode, what is confirmed
by blinking diode on the front panel of the instrument.
Note! The instrument will proceed to the sleep mode only if “Turn off device after the run”
option is activated and the executed amplification program finishes as Holding mode.
pause running;
stop running;
add any number of cycles in a current block or skip execution of the remaining cycles
and proceed to the execution of the next temperature block;
to check optical measurement data of the selected tubes in the form of 3D histogram in
the Distribution of optical measurements over the plate window by clicking
(Fig. 28).
38
Fig.28. Distribution of optical measurements over the plate window
If option “Turn off device after the run” was not activated, the instrument passes to the
Holding mode after completion of the RUN (Fig. 29).
In this case, a blinking message Holding… and a information about storage temperature will
appear on the chart.
When the temperature reaches the value close to holding temperature information message
appears on the screen notifying that the RUN has completed and instrument passes to the
holding mode (Fig. 30).
39
Fig.30. Storage mode indication
5. Maintenance
5.1 General Guidelines
The maintenance of the instrument is necessary to keep it at optimum working condition,
which minimises the risk of having unscheduled downtime and providing for its maximum
operational life.
The instrument belongs to IVD medical equipment. Maintenance should be carried out only
by competent personnel, who have studied the present operation manual.
The instrument is required for minimal routine maintenance during normal laboratory
operation.
40
their replacement. De-energizing of the instrument is achieved by removing the power cord
connector from it or from the power outlet and disconnection of USB cable as well.
The instrument should be protected from mechanical shocks and any liquids getting on its
case.
To keep thermal block matrix and elements of the optical system clean, the instrument
thermal block has to be always closed (except operations for placement / removal of PCR
tubes with samples).
Caution! To avoid failure of the instrument’s optical system it is strictly forbidden to use any
substances (heat-resistant pastes, oils, etc.) to improve contact between PCR tubes and
thermal block wells.
Caution! Never use any metal objects (paper-clips, wire, etc.) for wells cleaning!
Attention ! When cleaning the wells and thermo block matrix do not allow a liquid
flowing inside of the thermal block and the instrument!
Cleaning frequency: after 20 executed RUNs image of thermal block matrix should be
checked for all channels (as it is described in point 4.2 of the present manual). In case of some
bright spots will be detected for any ranges of fluorescence registration, cleaning of the wells
have to done.
Prior sending the instrument for repair or maintenance to local servise office or manufacturer
it is necessary to perform the disinfection operations with accordance to local rules and
regulations and fill out the disinfection certificate (see Appendix 1).
41
5.4 Troubleshooting
Trouble description Probable cause Corrective measures
42
During transportation, the instrument should be protected from dust and precipitation. It is
forbidden to tilt and turn over the instrument! Prior to the shipment, the instrument should
be secured to ensure its stable position and exclude any displacement and shocks.
Instrument can be shipped by any kinds of transport and it has to be carried in the
manufacturer’s transport package and treated with accordance to manipulation signs applied
on the outer surface of transport package. Acceptable environmental conditions for
transportation are the following: temperature from - 50°C to + 50°C and relative air humidity
up to 80 % at + 6°C.
Note! The instrument contains precision mechanical parts. To avoid their damage during
handling and shipping, the requirements of manipulation signs applied on package should be
strictly followed.
Attention! The instrument has to be shipped with the closed thermal block only.
7. Recycling
7.1 Recycling of medical devices has to be carried out in accordance with the classification,
rules for the collection, use, disinfection, storage, transportation, accounting and disposal of
medical waste established by the authorized federal executive authority.
7.2 Recycling of the instrument has to be carried out by special companies, which have the
appropriate license, using specially equipped places in accordance with the requirements
stipulated by existing Federal laws and in compliance with the mandatory requirements for
environmental protection in accordance with the local regulations.
8. EMS declaration
The instrument complies with the immunity and electromagnetic emission requirements of
EN 61326-2-6: 2013.
The instrument is intended for use in the electromagnetic environment described below:
- The device was designed and tested in accordance with the requirements of
CISPR 11 and it is suitable for use in all rooms, including ones, which are
directly connected to the public low-voltage mains network.
- The quality of the electricity supply has to comply with typical conditions of
use in commercial institutions or hospitals.
- Power frequency of magnetic fields has to be at a level corresponded with
typical conditions of use in commercial institutions or hospitals.
- The instrument uses radiofrequency energy exclusively for its internal
function. The level of radiofrequency emission is very low and does not lead to
disturbances in the operation of electronic equipment located close to it.
- It is forbidden to use the instrument near sources of strong electromagnetic
radiation that could interfere its normal functioning.
Notes:
1. The manufacturer is responsible for providing the consumer information about the
electromagnetic compatibility of the instrument.
2. The customer is responsible for maintaining electromagnetic environment, which
ensures instrument normal operation in accordance with its intended purpose.
43
9. Warranties
The manufacturer guarantees proper operation of the instrument and its conformity to TS
9443-004-96301278-2010 while meeting the requirements of this Operations Manual.
The warranty period for the instrument is 24 months from the date of its sale. Warranty repair
is acceptable upon providing of warranty repair card along with the filled deficiency report.
Average life of the instrument is not less than 5 years from the start of its operation.
Guaranteed storage life while meeting storage conditions (indoor under temperature 5 ° C - 40
° C) of the instrument is 12 months from the date of manufacture.
During the warranty period, the manufacturer bears responsibility to eliminate the
instrument's defects free of charge by repairing or replacing it on a similar one, provided that
the occurred defects are fault of the manufacturer.
Execution the warranty repair of the failed instrument by the manufacturer entails
prolongation of the warranty period for the time of such repair.
The manufacturer is not responsible for compatibility of the provided software with any
hardware or software supplied by other manufacturers, unless otherwise stipulated.
In no event shall the manufacturer and seller be liable for any losses, including loss of data,
loss of profits and other incidental, consequential or indirect losses arising from improper
installation, maintenance and operation or related to failure / temporary no operability of the
instrument.
The manufacturer is not responsible for defects and malfunctions of the instrument that
arising from the following:
- disregard of rules transportation, storage conditions, operation or improper
installation;
- improper actions connected to improper usage of the instrument as well as
disregard of the requirements set out in the operational manual;
- non authorized repair or changes in the design of instrument as well as cases of
violation of warranty seals on the body of the instrument;
- force majeure (fire, flood, earthquake, etc.) or influence of some random
external factors (power surges, etc.);
- Entering foreign objects (substances, liquids, insects, etc.) inside the
instrument.
The warranty does not cover instruments that have external defects (obvious mechanical
damage, cracks, cleavage on the body and inside, broken contacts of connectors and so on).
44
10. Packing Certificate
L.S.
Note. The form is to be filled in at the factory that packed the product.
L.S.
45
46
12. Cards of warranty repair
(QCD stamp)
Note of sale______________________________
(manufacturer’s name)
“……….”………………………., 20.…. manufacturer’s stamp
(date) (personal signature)
Owner and his address …………………………………………………………………
…………………………………………………………………………………….……
(personal signature)
…………………………………………………………………………………………
(QCD stamp)
Note of sale______________________________
(manufacturer’s name)
“……….”………………………., 20.…. manufacturer’s stamp
(date) (personal signature)
Owner and his address …………………………………………………………………
…………………………………………………………………………………….……
(personal signature)
…………………………………………………………………………………………
47
Reverse of CARD No. 1 to be filled in by a service center
48
13. List of claims and instrument disinfection procedure
49
Manufacturer: DNA-Technology, Research & Production, LLC
20 Zheleznodorozhnaya Street, Protvino,
Moscow Region, Russia, 142281
Phone/fax: +7(4967) 31-06-70
E-mail: [email protected]
http://www.dna-technology.ru
Customer support:
Tel.: 8 800 200-75-15 (free for Russia)
E-mail: [email protected]
Feedback form see on DNA-Technology’s website
http://www.dna-technology.ru/customer_support/
Service department:
Tel.: +7(4967) 31-14-67, +7(4967) 31-06-71 (ex. 3126)
E-mail: [email protected]
Version 14-2021