PART II
RealTime_PCR
Software guidance for
DTprime and DTlite
3
6.4 Color compensation in DTPrime and DTLite ........................................................... 71
6.4.1 Enabling signal compensation ................................................................................. 71
6.5 Diagnostics of the device.............................................................................................. 72
7 User settings ......................................................................................................................... 73
8 Parameters of the analysis settings .................................................................................... 74
9 Emergency situations .......................................................................................................... 76
Appendix 1. Glossary of Basic Terms ....................................................................................... 78
Appendix 2. Types of analysis (only the basic ones are represented) .................................... 80
Appendix 3 DTReport library description .............................................................................. 86
4
1 Introduction
The RealTime_PCR software is designed to control the detecting thermocyclers DTPrime and DTLite.
The Real_Time PCR software allows creating a RUN and to analyze the results of optical
measurements.
The RealTime_PCR software is designed for clinical laboratories usage, that is why the instrument
management functions and the results analysis settings are simplified as much as possible. At the same
time, if desired, the user can change the parameters of the instrument operation and data analysis with
the help of additional settings of the RealTime_PCR software.
3. Ability to operate several devices at the same time: RealTime_PCR software is launched in a
separate window to control each of the connected devices "multi-windowing mode". This
mode allows monitoring simultaneously the operation of several devices.
5
5. The possibility of introducing a vertical or horizontal temperature gradient across the matrix for
device with a solid matrix.
6. The possibility of independently setting temperature for each of six sections for the device with a
6-sectional matrix.
8. Two alternative methods for analyzing the DNA accumulation curve during PCR: Threshold
(Ct) and Geometric (Cp), which are based on different approaches to determining the indicator
cycle of amplification. By default, when analyzing optical measurements, RealTime_PCR
software uses the Geometric method (Cp).
2.3 PC requirements
PC requirements are described in section 1.5 of the Part I of operation manual.
The RealTime_PCR software is constantly updating by DNA-Technology. The latest version of the
RealTime_PCR software could be downloaded on the website:
http://www.dna-technology.ru/eng/support/
It is recommended to follow the updates and use the latest version of The RealTime_PCR software.
Before the first run of the device, the geometrical settings of the optical unit have to be checked as well
as the purity of the wells, the height of the tubes shall be measured, and the exposure of the optical
measurements has to be set.
Optimal exposure values have to be provided by the PCR kits manufacturer. When using the
DNA-Technology PCR kits, the factory settings are optimal and preset in the device. When
the PCR kits of different manufacturers are used, the exposure values should be checked
before each start of the amplification program.
Measurement of the tubes height has to be made in case of using plastic consumables of
different producers.
Checking the purity of the thermal block wells has to be carried out periodically to exclude
the possibility of obtaining unreliable results as a cause of background fluorescence
increase.
The geometric settings of the device's optical unit and exposure should be checked once during the
installation procedure. It might be necessary to change these settings when changing a manufacturer of
6
commercial kits and also in case it is stated in the instructions for kits or in case of any doubt the device
settings are correct.
The main menu of the RealTime_PCR software (Fig.1) consists of four sections:
Mode
Test (available in the "Device operation" mode only)
Preference
Help
1. Mode section is designed to control the operation of the software and includes the following
commands:
Operating mode the command is active in the Archive viewer mode and is intended for
switching to the mode Device operation.
Archive viewer the command is active in the mode Device operation and is intended
for switching to the mode Archive viewer.
Amplification program editor opens the Amplification program editor window.
Viewing CCD shots archive opens the window Viewing CCD shots archive, which
allows viewing the video of a matrix with tubes for each measurement, provided that after
starting the amplification program a field View the current video image has been ticked in the
Current image window.
Exit exit from the software with the closing of the corresponding window.
2. Section Test is intended for creation and editing tests and includes the following commands
(available in the mode "Device operation" only):
7
Create/Edit test displays Create/Edit test window to create a new test or edit an
existing one.
Copy group of tests displays the window to copy the tests of the operator specified in
the left part of the window, or from ini files with test parameters for Real_Time PCR software
(ini-files, which are provided by the kits manufacturer) to the operator selected in the right part
of the window (or to all operators).
3. Section Preferences is intended to display and edit the device settings. The "Instrument
operation" mode includes the following commands:
View image displays View image window in the process of setting up and diagnosing
the device (detailed description, on p.65, section 6 Settings and diagnostics of the device).
Set exposure displays Set exposure window in the process of setting up and diagnosing
the device (detailed description, on p 65, section 6 Settings and diagnostics of the device).
Last Run in device reading the results of optical measurements from the last run on the
device.
List of devices displays List of devices window with the function of connecting or
disconnecting them.
Device diagnostics sub-menu of the Preference, intended for service maintenance and
diagnostics of operation of the device’s various systems:
Command line displays the window where the diagnostics of the optical, temperature, motor
and display modules of the device are performed;
Monitoring opens the window for continuous display of service information necessary for
diagnostics of the device operation;
8
Measure height of the tube starts the procedure for measuring the height of the tubes while
maintaining the obtained value for further work with the device;
Read SD card displays window for diagnosing the operation of the SD card;
View CCD shots archive opens the window Viewing CCD shots archive, which allows viewing
the video of a matrix with tubes for each measurement, provided that after starting the
amplification program a field View image has been ticked in the Current image window;
Service engineer displays the window for optical calibration of the device (used by service
engineer only);
Options displays Options window to set user and analysis parameter settings.
Others displays Others window to change the parameters of the analysis of optical
measurements and adjust advanced settings for the data processing and contain the following
options:
Attention! Settings for operating with optical measurement data set by default are optimal for working
with commercial kits. These settings are stored in hidden system files and do not change when the
RealTime_PCR software is updated.
Type _AnalyseExpo: Selecting the type of "stitching" of optical data when the saturation level is
reached: linking the two exposures or selecting one of the two (Fig.2).
Expo_coef: a factor that determines how many times the second exposure is less than the first (Fig.4).
9
Fig.4 Tab «Expo_coef»
Signal_compenstaion: determine the existence of algorithms for "color" compensation between optical
channels, correction of the optical signal along the thermal block and between channels (Fig.5).
Filter_RawData: the presence of a digital filter in the analysis of the original data and the use of
additional analysis on the final "buildup". The borders of validity in describing the PCR curve with the
sigmoid (Fig.6).
10
Fig.7 Tab «Save to SD»
PCR_settings: define the left boundary of the search for the "Take_of point" in the case of PCR
analysis. Coefficient of reduce Cp relative to the location of the maximum of the second derivative in
PCR sigmoid (Fig.8).
Export results: the general settings for the export of results are defined, as well as the features of the
report output in the LIS (Fig.10).
11
Fig.10 Tab «Export results»
Template NameResult: setting the name of the protocol (RUN) to be saved according to a predefined
pattern (Fig.11).
4. The Help section displays information about the RealTime_PCR software and describing its
current version:
2. Use the icon on the desktop (when the software is installed, this icon is placed on the
Desktop automatically).
12
After the software is launched, a window will appear on the screen for selecting the operating mode of
the software (Fig. 12).
Fig.12 The window for selecting the operating mode of the software
Select an existing operator from the Operator list. To add a new operator select Add operator item
from the Operator list. When selecting Add operator item, the Add operator window appears on the
screen (Fig. 13).
In this window, the fields Operator and Directory shall be filled in. Enter the operator name in the
Operator field.
Path to the operator's working directory on the hard disk of the computer in the Directory field shall be
specified. To select a folder use the Browse button. While working under this operator, all files (plate
protocols, amplification programs, RUNs) will be stored in the specified folder, unless otherwise
indicated.
Press OK button to save the specified data and return to the window for selecting the operating mode of
the software.
Note: It is not recommended to create a working folder of the operator on the system hard disk, instead
of it use any other hard disks of a PC for files’ storage. While working with the device under the Guest
operator, the RUNs will be stored by default in the RealTime_PCR software folder. It is not
recommended to operate under Guest operator option.
13
2. Choose one of two possible modes for operation with the software:
Note: It is possible to work simultaneously in two modes (for example, analyzing the data obtained
earlier during the execution of amplification programs by devices, or preliminary analysis of data
obtained during the current run).
4 Device operation
4.1 Device operation mode
4.1.1 Mode assignment
Device operation mode of the RealTime_PCR software is intended to create the plate protocols,
create/copy tests, create amplification programs, conduct the amplification using the devices and
configure device technical parameters. (Fig.14).
Select the Operating mode from Mode menu of the Archive viewer window.
14
1. Click on the button of the Archive viewer window.
A dialog window List of devices appears on the screen with a list of devices connected to the computer
(Fig.15).
To work with a specific device, select it in the list and then press the button . The
RealTime_PCR software is ready to work with the selected device.
Multi-windowing mode allows working with each device, provided that the RealTime_PCR software is
launched in a separate window by one of the methods described in section 2.2 item 3. This mode allows
simultaneously monitoring the operation of several devices.
Note: In case several devices cannot be connected to the computer at the same time, it is possible to
work with these devices through a single-window mode. In single-window mode, after starting the
amplification program in the device, disconnect one using the button and connect
another device.
When working with devices frequently, it is necessary to create plate protocols. It shall include the
following information:
15
9. If necessary, additional settings.
After completion of the amplification program, the data analysis contains optical measurements
information in addition to the above mentioned settings and stored as a file located in the folder of the
operator, unless otherwise indicated.
To view the data analysis after the amplification is completed, open the saved RUN file.
When commercial kits are used and similar studies are conducted, the parameters of the plate protocols
are standard and differ from each other, usually in the number of samples and their placement in the
thermal block of the device.
To simplify the plate protocol creation procedure, it is recommended to save the standard plate protocol
parameters in the form of a Test.
4.1.4 Test
The set of all RUN parameters, that is necessary for conducting PCR and subsequent analysis of the
results is described as a Test.
The RUN can consist of one or more Tests having one amplification program and one type of analysis.
The Test can be created once and remain in the list of available Tests for a certain Operator.
To add a Test to the list of operator Tests, the procedure of creating Test using the option Create/Edit
test has to be performed, or by using the procedure Copy group of tests.
Copying a group of Tests involves the presence of ini files with all parameters for the RUN (ini-files,
provided by the DNA-Technology company). When creating a Test using an ini-file, the final stage of
process after the completed RUN is the formation of a result form using the Form of the answer .
When preparing to a RUN, it is sufficient to indicate Test in the Add test window (all the necessary
parameters will be loaded automatically), the number of samples to be examined and their placement in
the thermal block of the device should be indicate by Operator.
The RUN can be created without using the option Test, in this case, all necessary information shall be
entered in the plate protocol fields manually. When operating in this mode, Operator cannot select the
type of analysis. Note: The default type of analysis for such RUN creation is Qualitative.
16
Attention! It is not recommended to create a protocol without using the Test procedure since the
Analysis parameters window will not be active, and the optical calibration coefficients of the device
will not be used (see Section 6.4).
1. To create a new test, click the menu Test of the RealTime_PCR window and choose the
command Create/edit test. In the appeared window (Fig.16) click Create a new test DNA-
Technology, Research & Production LLC
2. Then enter the name of the new Test in the next window (Fig.17), if necessary activate the
Multi-test mode option (see section Multi-test mode) and click on OK button.
3. The Test window appears. Enter Test parameters in the appropriate fields in the Test window
(see Test Parameters section).
4. Press OK button to save the specified parameters and exit the Test window.
1. To edit the Test in the window Create/Edit test select the Test and click Edit current test.
The Test window appears (Fig.18).
2. Change the required fields in the Test window (see section Test Parameters).
3. Press OK button to save the specified parameters and exit the Test window.
The description and Test parameters are indicated in the Test window (Fig.18).
17
Drop-down list of Test is intended for selecting a Test. When selecting an item from the list in the Test
window the parameters of the selected test are displayed.
In the Description field, it is recommended to specify the purpose and features of the test.
Then specify Test parameters necessary for conducting PCR and analyzing the results.
Note: If commercial kits are used, the required parameters are specified in the kits’ instruction for use.
Parameters description:
Analysis
In Analysis section, select the Analysis type and Analysis method options.
Type
Method
Geometric (Cp) determination of the indicator cycle on the basis of mathematical analysis of
the shape of the amplification curve;
18
Threshold (Ct) determination of the Ct (threshold cycle). This method is based on conducting
a threshold line and determining the cycle number for each sample, at which the amplification
curve crosses the threshold line.
Note: To change the values of the threshold cycle, press the button Additional and change it in the
opened window Parameters of the analysis "point 2.Threshold value". The value of the threshold line
can also be changed using the mouse in the graphical representation window.
For more information about advanced options, see «Parameters of the analysis».
Tubes
The types of samples that are used in the analysis should be marked.
Standards
When choosing the Quantitative with standards type of analysis it is necessary to describe the used
calibration samples ("standards").
fill in the Quantity field indicate the number of different concentrations of calibration samples
("standards");
fill in the Copy field indicate the number of repetitions of each concentration of calibration
samples ("standards");
fill in the table of "standards" values (specify the name and concentration of each "standard").
The number of rows in the table corresponds to the number of variants of "standards". For each
"standard" you need to specify a name (by default this is "Standard_1", "Standard_2", etc.) and specify
the value (concentration of the "standard"). Units (copies, picograms, IU) can be selected from the drop-
down list in the title bar of the right column of the table.
The table of "standards" values can be filled manually or automatically. For automatic filling, press the
button Serial dilution, in the appeared window fill in the fields Initial value and Dilution step and
click the button Apply.
Data in the table of "standards" values can be presented in two formats: standard and scientific. Select
the presentation format using the switch Format.
Example. Indication of values in the standard and scientific format. Note: the degree of 10 (order) of the
value is indicated after the letter "E" in scientific format).
19
Standard Scientific
5 5,00E+00
15 1.50E+01
250 2.50E+02
9,500 9.50E+03
18,500,000 1.85E+07
Controls
In Controls section, it is necessary to indicate the number of tubes used in the Test with control
samples.
Note: The number of tubes with calibration and control samples, if necessary, can be changed during the
compilation of the plate protocol (see Section 4.2 Creating a plate protocol).
In PCR-mix volume field, it is necessary to indicate the volume of the reaction mixture without taking
into account the volume of oil or paraffin (if used).
For the most effective and accurate performance of temperature regimes of amplification, an "active"
method for controlling the temperature of the reaction mixture in a tube is used in the device. The
method is based on the application of a mathematical model of the reaction tube and requires an
accurate indication of the volume of the reaction mixture.
Fluorophores
In the Fluorofors table the fluorophore and its purpose has to be selected and indicated from the drop-
down lists for each channel when analyzing the results:
If some channel is not used, select in the Fluorofors table the item Not detected.
Note: If R6G dye is used in PCR kit, then while creating (editing) a Test in the HEX channel choose
When choosing the Analysis of polymorphisms (2 probes) or Analysis of polymorphisms (MC) auto
or Analysis of polymorphisms (MC) type of analysis, it is necessary to specify the allele for
fluorophores signal (by default this is "a" and "b").
20
When analyzing Analysis of polymorphisms (2 probes), the results are evaluated by the presence and
magnitude of relative displacement amplification curves recorded in one tube for two fluorophores (dCp
or dCt).
In the Analysis of polymorphisms (MC), melting type, the interpretation of the results is based on the
correspondence of the melting temperatures of the signal probes to the temperatures indicated in the
Criteria for polymorphisms analysis field.
In the Analysis of polymorphisms (MC) auto type, the interpretation of the results is based on
comparing the characteristics of reference melting curves contained in the library file, and melting
curves obtained in the present experiment.
Amplification program
In section Amplification program an amplification program has to be specified in one of the following
ways:
The selected amplification program will be displayed in this section. For more information about
creating an amplification program, see section 4.1.6 "Creating and editing amplification programs".
Additional
When you press the Additional button the Parameters of the analysis window opens.
See the description of additional analysis parameters in section 8 «Parameters of the analysis».
Attention! This section is intended for advanced users only. In most cases, there is no need to change
the default values of the analysis parameters. Changing the analysis parameters stored in the software,
may lead to incorrect interpretation of the obtained data analysis.
4.1.5.4 Create a new Test using the Copy group of tests option
Creating a Test is also possible using ini files with test parameters for the Real_Time PCR software, if
they are provided by a kits manufacturer (ini-files are provided by DNA technology). Files with test
parameters already contain all necessary information that is entered in the Test window. When copying
tests, the information is saved in the list of operator tests.
In the main menu of the RealTime_PCR software select the Test, See (Fig.19):
Click Copy group of tests in the drop-down list. The Copy group of tests window opens (Fig.20):
21
Fig.20 Window « Copy group of tests »
In the left part of the opened window, select the item "from *.ini file", see (Fig.21):
Click to choose the necessary test from the list on the left or select all tests using the button
and click . After that, the test should move to the right list and the following message will be
displayed, see Fig.23.
22
Fig.23 The message about the successful completion of the test copy
Multi-test mode is the option that allows combining several tests in one RUN with the same
amplification program and analysis type, creating batch studies.
Multi-test mode can be used for the following analysis types: Analysis of polymorphisms (MC),
Relative and Qualitative.
While creating batch analysis (see section 4.1.5 Creating a new test using the Create/Edit test option),
the option multi-test mode has to be activated, See (Fig 24).
In the opened window Test (Fig.25) the operator can select all previously created tests, selecting from
which the operator forms a list of new batch analysis. Test list of this batch analysis in the required
order is displayed in the Location tests window, see (Fig.26).
23
Fig.25 Multi-test test window
The selected tests should have the same amplification program. If a test with another amplification
program is added, a warning message will appear, see (Fig.27).
24
Fig.27 Message about the presence of tests with different amplification programs
After creating a new batch analysis list, click OK. The created Test will be saved in the list of the
operator tests and will be marked with an icon .
When preparing to RUN for batch studies, Operator marks the number of samples in a window Add
test, there is an indication of the number of tubes for each sample in brackets (Fig.28).
When using the option Copy group of tests, the amplification program is loaded into the software
automatically.
When creating a test with the option Create/Edit test, the amplification program is entered by the
operator.
When preparing to RUN without the help of an Add test option, the amplification program is created by
Operator or added from the list of previously created programs just before each start of the RUN.
25
4.1.6.2 Creating a new amplification program
There is an option to create a new amplification program at various stages of the RealTime_PCR
software:
then select the Menu bar item Create a new program in the Amplification programs editor
window.
When creating a new amplification program by any of these methods, an Amplification
program template window appears on the screen (Fig.29). Using the template, specify the
structure of the amplification program.
A new amplification program can be created with or without the Amplification program template.
To create a new amplification program using a template, the following steps shall be proceeded:
1. Choose a Create new program command. After that, an Amplification program editor
window and an Amplification program template window will be displayed. The Amplification
26
program template window (Fig.29) schematically shows the parts of the amplification
program: two- and three-stage blocks of temperature cycles, preheating and storage modes.
2. Select the template that best matches the structure.
3. If necessary, add Preheating and Standby modes (enable the corresponding options).
4. After selecting the most suitable template, press the button Apply.
The information specified by the template will be displayed in the Amplification program
editor window (Fig. 30).
enter the name of the program in the Program title field (Note! strictly Latin letters and
numbers without spaces and symbols);
fill in the Description field if necessary;
change the required data in the amplification program table.
Each row in the table corresponds to one temperature range of the amplification program. Temperature
shelves can be combined into blocks. The number of rows in the block determines the number of
temperature shelves at this stage of the program. In the table, the blocks are separated from each other
by empty lines.
Adding the temperature shelf is active after completing the editing of other lines
only.
In each line for each temperature shelf, the following information shall be edited:
shelf temperature ;
27
shelf length and ;
the presence of optical measurements .
Attention! Only one optical measurement is allowed within one block. Empty fields in the program are
not allowed when entering temperature and time values, except for blocks Pause and Standby.
1. Cycle (specified by default when creating a block) sets the mode for repeating the program at
this stage until a specified number of cycles is reached.
2. Pause pauses in program execution mode.
3. Standby sets storage temperature after program execution.
Attention! The «Standby» mode can be set at the end of the program only.
4. Melting curve allows obtaining melting profiles for estimating the lengths and composition of
the amplified fragments.
Selecting the Melting curve allows creating the default temperature block of one temperature
shelf with an initial temperature of 90 °C, the number of cycles equal to 100 and incremented on
each temperature cycle of -0.5 °C. The initial settings of the "Melting curve" block can be edited
using the Temperature increment.
To display the specified parameters in the table, enable the corresponding options (Fig.33).
28
Fig.33 Advanced setting
The temperature shelf can be changed by editing the graphic image of the amplification program. To do
this, bring the mouse cursor to the temperature shelf (the mouse cursor becomes ) and, while holding
down the left mouse button, move (in the vertical direction) the temperature shelf to the necessary
temperature.
The graphic representation of the amplification program can be changed using the context menu, which
can be opened by the right mouse button, and contains the following items:
To create a new amplification program without using a template, procced the following steps:
1. Choose a Create new program command. After that, an Amplification program editor
window and an Amplification program template window will be displayed.
2. Press the button Cancel in the Amplification program template window.
3. Enter the name of the amplification program in the Program title field.
4. Fill in the Description field if necessary.
5. Fill in the table with the amplification program, similar to creating an amplification program
using a template.
29
Note: The table with the amplification program can be filled in either using the edit buttons to the right
of the table or by the keyboard.
To fill in the table of the amplification program using the keyboard, it is necessary to fill in the fields of
temperature and time (minutes, seconds). Entering each new value completed by pressing the <Enter>
key (Fig.35).
After introduction of all parameters (temperature, minutes and seconds) of the current line, a new line of
the same block with empty fields for input appears automatically (Fig.36).
To complete the block, press <Enter> without entering the temperature and time in the new line. After
that, the editing of the value of the number of cycles in the block becomes active. Pressing the <Enter>
key entering a new value for "Number of cycles" finishes and jumps to enter a new block (Fig.37).
30
Next, the type of each block shall be specified, enable optical measurements and, if necessary, indicate
the temperature change on a matrix, increment on time and temperature (see related topics).
Using temperature gradient allows entering horizontal or vertical temperature gradient on the matrix if
the device is equipped with a solid matrix, or independently to adjust the temperature of each of the six
blocks of the matrix (4 x 4), if the device is equipped with pseudo blocks. The maximum value of the
gradient on the matrix or the temperature difference between the different zones is ± 10 °C. Thus, using
a temperature difference (when setting different temperatures in blocks), six different temperature
profiles can be used with one RUN, which allows performing simultaneous reactions with different
temperature parameters. The temperature gradient (smooth horizontal or vertical temperature change in
the block) is usually used in scientific applications to optimize temperature profiles.
31
Fig.39 Horizontal gradient with the option "strip"
If you select horizontal gradient, the option can be set, using in this case the field of the 8x8
matrix with improved linearity of the gradient (the colored area on Fig.39).
Fig.40 The temperature drop across the blocks for DTprime (only for devices with pseudo blocks)
32
select the symbol of the difference and enter its value with respect to the shelf temperature for
each pseudo block of the matrix (Fig.40).
Increment by time
Using the increment by time allows changing a duration of the given temperature shelf for each
subsequent cycle by the value of a given increment.
2. Using the button open the window in the corresponding row of the table with the
amplification program Use increment (Fig. 41).
3. Enable option Use time increment (the input field of the increment in time becomes active
).
4. Enter the increment by time.
5. Click OK to complete the entry, otherwise click Cancel.
In the course of editing the increment value, the duration of temperature shelf is automatically
recalculated for each subsequent cycle and it is displayed graphically as a dependency from the cycle
number.
Use the temperature increment to create a "Cycle" block, that allows changing the value of this matrix
temperature to perform the temperature shelves for each subsequent cycle by the value of the specified
increment.
33
To indicate the temperature increment:
2. Using the button open the window in the corresponding row of the table with the
amplification program Use increment (Fig.42) to edit the temperature increment.
3. Enable option Use temperature increment.
4. Enter the value of the increment by temperature or the temperature shelf for
the final block cycle (the increment will be calculated automatically) .
5. Click OK to complete the entry, otherwise click Cancel.
In the course of editing the increment value, the temperature of the shelf is automatically recalculated
for each subsequent cycle and it is displayed graphically as a dependency from the cycle number.
Using the temperature increment when creating the "Melting curve" block allows editing the initial
parameters of this block.
2. Using the button open the window in the corresponding row of the table with the
amplification program Melting curve (Fig.43) to edit the initial parameters of this block.
3. Set three of the four parameters in the window Melting curve (for example, initial temperature
, temperature increment , and the number of cycles
). The fourth parameter will be calculated automatically.
34
4. Click to complete the entry, otherwise click .
In the course of editing the parameters of the Melting curve window, the temperature of the shelf is
automatically recalculated for each subsequent cycle and it is displayed graphically as a dependency
from the cycle number.
select item of the Menu of Amplification programs editor window: - open the program
from the hard disk or a data storage device.
1. Choose the Edit program command, then Amplification program editor window will be
displayed.
2. Edit the amplification program, similar to creating an amplification program using a template.
35
1. Press the button (to save the amplification program and exit) the Amplification
program editor.
2. In the dialog box that appears (Fig.44), enter the name of the created amplification program in
the field File name, using the Latin letters only.
3. Using a Folder field, select a folder on the hard disk or a data storage device to save it (by
default the software will offer the operator's working directory).
4. Press the button Save. The amplification program will be saved to a file with *.rta extension.
Before the amplification, it is necessary to fill in the plate protocol (Tab Protocol in the window
RealTime_PCR).
36
2. Additional information:
The plate protocol data is displayed in a table, each line of which contains information about a single
tube (Fig.45).
The location of the tubes in the thermal unit of the device is schematically displayed in the lower right
part of the "RealTime_PCR" window (Fig.45).
Editing the plate protocol is performed using the buttons located to the right of the table (Fig.46).
1. Using the option Test with the saved standard parameters of the analysis.
Tests (using options Create/edit or Copy group of tests) are available to the Operator who
created them and may be accessed by the button Add test.
2. By filling the plate protocol manually line by line.
3. Use the previously saved plate protocols.
37
4.2.2 Import protocol data from the LIS (Import button ).Create a plate protocol using
the Test option
To create a plate protocol using the Test option, perform the following actions:
1. Log into the RealTime_PCR software under one of the operators, using the Device operation
mode and select the device to connect.
2. If necessary, specify the plate protocol number in the Protocol number field
.
3. Click on the button Add test of the Protocol tab. The window Add test is
opened (Fig.47).
4. From the drop-down list of Test (located to the right) select the test type: Single (for single-tube
tests), Complex (for multi-tube tests) or All tests.
5. The drop-down list of Test located to the left will display the tests of selected type.
6. Select a test name from the drop-down list of Test. The parameters of the selected test are
displayed in the window.
7. Specify the number of test samples and the number of replicates of each sample in the section
Samples.
8. Specify the number of tubes with control samples in the section Controls.
9. When carrying out the quantitative analysis, specify the number of calibration samples and the
number of repetitions of each sample in the section Standards.
Note: During the analysis, it is possible to use "standards" from another RUN (for more details,
see section Additional standards), to enable it proceed the following:
38
using the Browse button, open the file selection window with additional standards;
select the corresponded RUN.
The path to the selected file will be displayed in the Standards option.
Attention! The obligatory condition for the use of additional standards is the identity of the plate
protocols for the amplification program, as well as the absence of changes in the list of tests of
this Operator.
10. To save data and return to the RealTime_PCR window click the button OK, to return to the
RealTime_PCR window without saving the data, press the button Cancel.
11. When using the procedure Add test, the fields of the plate protocol are filled in automatically.
12. Only tests of the same type (matching by type of analysis and amplification program) can be
added in one plate protocol. Additional editing of the plate protocol is possible (change of
identificators names, concentration, standards, etc.) manually.
4.2.3 The manual line-by-line plate protocol filling
In this option all information in the plate protocol fields is manually entered.
Note: With a line-by-line plate protocol creation, it is impossible to change the type of data analysis, a
qualitative analysis type will automatically be used.
1. Log into the RealTime_PCR software under one of the operators, using the Device operation
mode and select the device to connect.
2. If necessary, specify the plate protocol number in the Protocol number field.
3. To add a new line (description of the new tube), press the button Add line .
4. The following data in the line can be edited:
No. (tube number). Filled automatically when adding a new line, if necessary, the data can be
edited;
Identificator (short designation of the sample) - an arbitrary description of the tube such as
patient's name, sample number, etc;
R (adding duplicates). Each time pressing a button in the corresponding row, a linked line is
created in the R field with an identical description of the tube (duplicate). Creating duplicates is
possible after entering all the information about the tube. To change information about the block
of duplicates, the "head line" of the block of duplicates has to be edited;
39
Tube type. The following types have to be selected from the drop-down list: Sample, Standard,
Control + or Control - ;
Concentration. Filled with the Standard type of tube. It is necessary to indicate the
concentration of the calibration sample;
Fam, Hex, Rox, Cy5 and Cy5.5 (for a 5-channel device) - choose from the drop-down list the
purpose of the fluorophore (specific target, IC, not detected).
Note: When preparing to RUN in this way, the type of analysis is Qualitative.
Once the amplification program is completed, the following changes can be made to the plate protocol:
the type of analysis, the values of the calibrators, the type of tubes, duplicates, identificators.
Attention! Do not fill a plate protocol without using the Test procedure (using options Create/edit or
Copy group of tests) since the Analysis parameters window (see section 8 of this manual) will not be
active, and the optical calibration coefficients of the device will not be used.
After filling in the plate protocol fields with the tubes’ data, it is necessary to specify the location of the
tubes in the thermal block of the device. Tubes can be placed in the matrix automatically or manually.
When filling the plate protocol by any of the above methods, the tubes are automatically placed in the
thermal block matrix.
Using the button choose the order of filling: by rows, or by columns (Fig.48).
40
(For DTprime) (For DTlite)
Fig.48 Schematic arrangement of test tubes in the thermal unit matrix
Tubes can be placed manually in a free order. To do this, disable the mode, tick the
mode, clear the matrix field using the button , place all the
tubes in the matrix using the mouse cursor.
Button is intended for displaying a graphic image of a matrix window. This image can be printed or
saved to a file.
For different types of device, the schematic arrangement of the tubes will be different. This depends on
matrix capacity of the device.
To edit and start the amplification program, press the button Apply.
After clicking the button Apply the software automatically switches to the Start run.
Start the amplification program on the tab Start run of the RealTime_PCR window after creating a
plate protocol (see "Creating a plate protocol").
By default, the tab displays the program entered by the operator when creating/editing test. When
creating a Test using ini-files, the necessary program is loaded automatically (Fig 49).
When creating a plate protocol using the option Add test, it is not necessary to edit parameters in this
window.
If the plate protocol was created manually line-by-line, in this window select an amplification program
from a previously created amplification, edit one of the previously created, or create a new one and
specify the volume of working mixture in tubes.
41
Fig.49 "Start RUN " tab in the "RealTime_PCR" window
After the thermal block is in working position a dialog box appears on the screen of the display that
prompts to specify the name of the RUN file in which the results of optical measurements will be stored
and a place to save it (by default it is a working directory of operator) (Fig.50).
6. Select the directory to save the RUN file (in case it differs from the default directory).
7. File name: enter the name (agree with the unique file name proposed by software or create
another file name).
42
8. File type: the data saving format will be specified with the following extensions, which are
corresponded to type of the device: r48,*.r96, *.384.
After clicking on the Save button, the RealTime_PCR software processes into the mode of executing the
amplification program while saving the results of optical measurements in the RUN file.
In the status line of the software, one of the following states represents the device status:
the device is switched on, but not ready for operation (warming up the device for at
least 5 minutes).
Note! The activation of the amplification program is possible with the status of the device "Device is
ON" (green or yellow indicator in the status line of the software). In the case of a status of the device
"Device is OFF" (red indicator), a message will be displayed that the device is not connected (switched
off).
Attention! If you pressed the Start RUN button during warm-up (yellow indicator is in the status line
), in the Temperature profile window a flashing message will be displayed on the
continuation of the warm-up process of the device (Fig.51):
43
A similar message will be displayed on the device’s display. After the end of warm-up, the device will
automatically switch to the mode of the amplification program execution.
4.2.6 Use the option Switch off the device after the RUN
If there is a need for autonomous operation of the device with its automatic shutdown (transition to sleep
mode) after the completion of the amplification program, perform the following:
2. Close the software by pressing in the upper right corner of its window.
3. Upon completion of the amplification program, the device processes into a sleep mode, and
indicates by the blinking LED on the OSD menu button (only for DTprime device).
Attention! The device will switch into a sleep mode only if the option Turn off device after the run is
activated and the amplification program performed therein ends in a Holding mode.
In case of closing the RealTime_PCR software (during the autonomous work of the device), the
computer remained turned on, the file with the results of optical measurements will be saved in the
chosen directory and the operator will be available to view it after the amplification program is
completed.
Note! In case of closing the RealTime_PCR software (during the autonomous work of the device) the
computer was turned off, the file with the results of optical measurements will not be saved in the
chosen directory. The RUN file can be downloaded to the computer via the option Last run in device in
the menu Preferences of the RealTime_PCR software window.
If the amplification program starts successfully, the tab Start run of RealTime_PCR window is
replaced by a tab Running (Fig.53). After starting the amplification program, the device does not
require user intervention. If necessary, during the program execution, the operator can change the
display of the current processes and also change the amplification program (only for advanced users).
After the amplification program is completed, the software automatically switches to the data analysis
viewing mode with the results of optical measurements.
44
Fig.53 The Running tab of the RealTime_PCR window
The top part of the tab displays general information about the current RUN:
protocol number;
operator name;
date;
comments on the RUN, which, if necessary, can be supplemented in the process of program
execution;
two fields in the right corner of the window, where the following time parameters for displaying
can be selected from the drop-down list:
In the central part of the tab, a graph of fluorescence values is displayed: the DNA accumulation curves
in each tube in the coordinates "Fluorescence level amplification cycle number".
The choice of the fluorescence channel for displaying fluorescence values on the graph is performed
using the following drop-down list:
45
To control the display mode of the graph and detailed analysis of the data, the following elements can be
used:
button to display the dialog box for changing the scale of the axes (Fig.54):
Location and color of the tubes button, by this option the color of the tubes, as well as the
view of their type and numbering can be changed.
In the Distribution of optical measurements over the plate window you can see the value of the
selected fluorescence channel for each cycle .
Using the arrow buttons located at the bottom of the window, zoom in and out in a horizontal or vertical
plane.
The window can be resized by fixing the mouse cursor on the border of the window until the double
arrow appears, then moving the border using the mouse.
46
(For DTprime) (For DTlite)
Current image button is intended to display the window with the image of the matrix field
with tubes on this cycle for a given measurement (Fig.56).
button is intended for switching on/off of the calculation mode of the basic (background)
component for each tube by the "first" cycles of the amplification program. All measurements are output
after deduction of this base component.
Note (for advanced users): To specify the boundaries of the first cycles, select item Options... in the
menu Preferences of the RealTime_PCR window. In the appeared Options window select a tab
Optical data acquisition, enable the option Consider background input starting from and indicate
the boundaries of the cycles in the corresponding windows (Fig.57).
47
Fig.57 Specifying the boundaries of the first cycles
Change the representation of optical data button is intended to display a multi-window view
of the curves (each kinetic curve is displayed in its own window See (Fig.58).
At the bottom of the Running tab, the Temperature profile is shown (diagram of the matrix
temperature versus the time of the amplification program). Temperature shelves with optical
measurements are highlighted in red, the moments of optical measurements are displayed by flashing
the icon on the diagram.
On the right above the field with the diagram, there is the button for displaying the dialog
box for changing the scale of the axes.
The area indicated by the symbol shows the progress of the amplification program and contains
the following information (Fig.59) :
48
Fig.59 Information on the progress of the amplification program
In the area indicated by the symbol , there are buttons for controlling the progress of the program:
enable/disable Pause mode (pause the RUN execution). The pause occurs on
the first temperature shelf after switching on this mode. When the pause is switched off, a transition to
the next temperature shelf immediately occurs, without waiting for the end of the current shelf, at which
stage a pause of the amplification process was occurred.
skip the cycles remaining until the end of the block execution.
Attention! These operations can be performed only with the enabled option
To automatically turn off the device after execution of the amplification program, enable the option
.
During the execution of the amplification program, the progress bar of the running process is displayed
in the status line .
During the optical measurements (until the end of the amplification program), if necessary it is possible
to come back to the tab Protocol and edit the plate protocol for the following parametrs: the number,
type, location of the tubes (except for the amplification program). When going back to the tab Running
after editing the plate protocol, it is requested to save the plate protocol or discard changes.
Note (for advanced users): During the optical measurements, it is possible to change the settings of the
measurement and data representation modes. For this, it is necessary to select item Options... in the
menu Preferences of the RealTime_PCR window. In the appeared Options window select a tab
Optical data acquisition.
49
Show calibrated data show the data of the optical measurements taking into account the
compensation of the optical uniformity of the matrix;
Use DT normalization method apply the DT normalization method to calibrate optical
measurements;
Consider background input starting from the background component for each tube is
calculated according to the first (indicated) cycles. All measurements are output after deduction
of this background component;
Auto exposition adjustment automatic selection of the exposure value of optical measurements
(temporarily unavailable);
Number of exposures for one measurement the number of measurements (1-4) within the
same optical channel: Fam, Hex.
After the amplification program has finished, a window appears on display with a message that the
program enters to the Holding mode (Fig.60).
In this case, a flashing inscription Holding... and a marker with a storage temperature appear on the
temperature’s graph of the matrix dependence against the time of the amplification program.
When the storage temperature is reached, a window appears on the screen indicating that the
amplification program has completed its execution and switches to the storage mode (Fig.61).
When clicking OK button, the RealTime_PCR software switches to the mode Archive viewer (tab
Data analysis).
1. Select mode Archive viewer when you run the RealTime_PCR software.
50
Fig.62 «Program RealTime_PCR start» window
2. Select item Archive viewer in the menu Mode of the RealTime_PCR window.
51
The RUN file with the optical measurement data is stored in the operator folder and can be opened for
viewing in the Archive window as well.
- double click to open the RUN file, that is stored on the computer or any other source, with the
results of optical measurements (file type *.r48, *.r96, *.384);
- in the window "Archive" click the button in the Data analysis tab and select the
necessary RUN file in the "Open run file" window.
Note: If for some reason the computer communication with the device was interrupted during the
amplification program execution, there is the possibility to download data from the device, that was last
Run. To do this, select the item Last Run in device in the menu Preferences of the RealTime_PCR
window (option is available only in the "Device operation" mode).
In the Archive window the following information about the executed RUN is provided:
1. RUN data (RUN number, operator name, date and time of optical data obtaining, comments to
the RUN). The RUN data is displayed at the top of the window.
2. Additional RUN data (the name of the RUN file, the name of the amplification program,
parameters of the amplification program, the volume of the reaction mixture in the tube).
52
For more information about the protocol, click the button Information about amplification
program, which opens the Information window. The window contains two tabs:
- Amplification program tab contains the name of amplification program, its parameters and
volume of the reaction mixture in the tube.
- General information tab contains the path to the RUN file, its name, serial number of the
device and plate barcode (Fig.65):
Basic toolbar:
, where
Panel of device:
1. - opening and saving RUN measurements:
- open Run
- save RUN
- export results in xml format (old version)
- export results in txt format
- Export of results in xml format to LIS
- sending RUN via email
2. - general information on the results of analysis and measurement data:
53
- location and color of the tubes (Fig.66).
- RUN information: amplification program, RUN file name, serial number of the device,
etc (Fig.67).
54
Fig.69 The parameters of fitting of PCR- sigmoid
55
- display of tubes in the Test/Sample mode (Fig.72).
Additional panels required for work can be dragged to any place in the working window of the software
using the mouse, and afterwards the toolbar can be removed.
The following elements can be used to change the way of results displaying in Archive window:
the window for selecting the displayed fluorescence channel (using the drop-down
list).
the button displays an additional window Scale of axes for manual scale adjustment on
the X and Y axes.
automatic scale adjustment button.
56
the option to display the graph on a semi-log scale (it is recommended to turn off the
fitting option that is located on the right corner above the graph).
If necessary, select the color and number of lines in the fluorescence graph, press the button (in the
window RealTime_PCR tab Running or in the window Archive in the tab Data analysis). In the
opened window Location and color of the tubes, the matrix of the thermal block displayed. The color
of the tube corresponds to the color of the line in the graph. The number of tubes corresponds to the
number of lines displayed in the graph.
To display the numbering of tubes according to the RUN, press the button , to display the type of
tubes press the button .
To change the color of the tube, click right mouse button on the selected tube, in the opened Color
window select the desired color and press OK (Fig.74).
57
Fig.74 Choice of color
To select or deselect the tube, click left mouse button on the selected tube.
It is possible to select a group of column or row of tubes, for that procced the following:
In order to clear the entire field of the thermal block matrix, use the button (Fig.76).
58
To access additional color control buttons and the location of the tubes, press the button Briefly and
then in Detail.
In the expanded window, additional buttons appear, with the help of which it is possible:
to return to the initial state (initial location and color palette of the tubes);
to designate an arbitrary color of tubes;
to designate the same color for duplicates;
to select the color of the tube from the preset color set;
undo the last change in color.
View the results of the settings either in All tests mode or by selecting one test from the drop-down list:
The following elements can be used to change the way results are processed:
By default, the analysis method specified when preparing to RUN using the Test procedure. If the RUN
is prepared without using the Test procedure, the Geometric (Cp) method will be specified in the
corresponded field.
button is intended to open the window Parameters of the analysis (see "Parameters of the
analysis "). The default values are preset and optimal. When conducting scientific research and for a
more detailed analysis of the obtained data, change the preset parameters for processing the results of
optical measurements and data analysis (Fig.77).
button is intended for switching on/off fitting (smoothing) of the PCR curves.
59
The Fitting option is applicable only after subtracting the "base" line for samples in which fragments of
the desired DNA were found (for "positive" tubes). The button is not active when selecting the analysis
method Initial area and when choosing the type of analysis Melting curve.
1. Analysis information:
To create a report based on the analysis results, perform the following actions:
1. Press the button on the tab Data analysis in the window Archive. The Report window is
opened (Fig. 78).
With the help of the appropriate functions, the following options are available:
60
add fluorescence curves for the corresponding fluorescence channels;
61
Fig.79 Format selection
In the opened window Save the analysis results as... specify a file name and select a folder
on the hard disk or any other device to which it will be saved (Fig.80). After clicking the
button Save the file will be saved to the specified directory.
Export the data for transfer to the LIS using the button Exporting results as XML. Report in XML
format containing the calculated results in numerical form will be generated.
Also, the measurement results can be exported as an Excel table or copied to the clipboard. For this, in
the "Archive" window choose the "Data analysis" tab and click right button of the mouse on the
"Results" table (Fig.81).
62
Fig.81 Copy the contents of the table
Settings of technical parameters of the device and its diagnostics are carried out in Device operation
mode.
Using the RealTime_PCR software the following parameters can be checked and set:
6.1 Checking the geometric settings of the optical unit and the purity of the wells.
When switching on the device for the first time after its transporation or any movement, it is
recommended to check the geometric settings of the optical unit. For this, proceed the following steps:
1. Select item View image in the menu Preferences of the RealTime_PCR window.
2. In the opened window View image (Fig.82) using the drop-down list select
the measurement channel Fam.
3. Set the required exposure value in the window Exposure to get enough of optical signal levels in
the range of 1000 ... 3000 (the scale value when obtaining the digital values of the fluorescence
signals in the window Digitization of the signal).
63
(For DTprime)
(For DTlite)
Fig.82 Window for checking and correcting geometric settings of the device
In the resulting image, the contours denoting the boundaries of the measured area (red circles)
should not go beyond the light spot.
If the center of the pixel constituting the light spot is inside the red circle, it is taken into account
during the measurement. Pixels whose centers are outside the boundary of the red circle will not
be taken into account (Fig.83).
64
If the circles are offset relative to the light spots, it is necessary to correct the geometry of the
optical image.
7. To correct the geometry of the optical image in the Image preview window, select the point to be
corrected with the left mouse button. For more accurate correction, press the button Zoom. A
window will appear that zooms the image of the selected point.
8. Correct the coordinates of the corner points and the size of the light spot by changing the values
in the corresponding windows.
Note! To cancel the changes that are made press the button Return to initial values.
(For DTprime)
65
(For DTlite)
Fig.84 The window for adjusting the geometric settings of the device, adjusting the displacement of the
coordinates of the central points.
To reset the offset coordinates to zero position, press the button , to return to the previous values
press the button .
13. To obtain the digital values of the fluorescence signals, select the option Digitization of the
signal for displaying an additional window Digitization of the signal (Fig. 85).
The option Marker allows displaying numerical values of measurements for each tube.
Selection of the option Statistics opens the window Statistics of optical measurement (Fig.86)
with an indication of the time duration of the optical measurements and the average value of the
video signal of the selected channel.
66
Fig.86 Statistics of optical measurement window
14. Repeat step 5-8 for measuring channels Hex, Rox and Cy5 to make sure the settings are
correct.
The cleanliness of the thermal block wells is indicated by the absence of the bright spots on the image
of the empty wells when checking the geometric settings of the optical unit (fig 87).
To check the cleanliness of wells, and the lack of stray reflected optical signals perform the
following:
1. Close the thermal block, make sure there are no tubes in the wells.
2. Open the window View Image (Preference/View image), choose Digitalization of signal and
open this window, set the exposition.
3. Press the button , start the process of optical measurement in FAM channel
(selected by default).
4. Make sure that the optical signal levels for empty wells in the Digitization of video signal window
(see fig.88) does not exceed 1500 conventional units, otherwise clean luminous wells and repeat the
measurement.
67
a)
b)
Fig.88 The view of empty wells DTprime (a) and DTlite (b) devices
5. Successively choose the rest channels of the device and set for the exposition values (twice
bigger than the working exposition values), repeat these measurements for all channels (by
pressing the button and controlling the levels of optical signals in the Digitization of
video signal window).
68
Note! It is recommended to check the purity of the wells immediately before placing tubes with PCR
samples if there is any possibility of contamination.
The regular check of the purity of wells should be carried out by the operator in accordance with the
device maintenance recommendations.
click the button and place the tubes or strips evenly over the thermal block
matrix;
select the Preferences/Device diagnostics/Measure height of the tubes (Fig 89), and wait for the
message (Fig.90):
69
Fig.90 Successful change of test tube height
Attention! When measuring the height of the tube, place at least 32 test tubes or 4 strips evenly over the
thermal block matrix.
Preset exposure parameters of optical measurements are optimal for DNA-Technology PCR kits. The
change of exposure parameters could be required if using PCR kits from other producers.
If necessary, set the exposure value for each channel. It is obligatory to set the exposure of optical
measurements for all channels by selecting them sequentially in the upper right corner of the window
Exposure.
70
The buttons with arrows under the diagram allow rotating the image in the horizontal and vertical
planes, the option Marker allows displaying numerical values for each tube.
Note! The red color of the column warns that the fluorescence value in this thermal block well is above
the linear measuring range.
In order to expand the optical range, each measurement can contain from 1 to 2 exposures for DTprime
and DTlite (the latter is five times shorter than the first). In this case, the maximum exposure time is set
in the window Exposure. The number of exposures can be adjusted in the menu
Preferences/Options/Optical data acquisition (Fig. 92).
The cross-channel effect, arising from a partial overlapping of the absorption and emission spectra of
fluorescent dyes, is eliminated by the software method. For this purpose, the matrices of the optical and
spectral coefficients obtained by calibrating of the device with pure fluorescent dyes are stored in the
device memory.
Attention! The color compensation works only if the plate protocol is filled using the procedure
, that is, each tube corresponds to a specific test, which indicates the
composition of the dyes included in the PCR kit used. Otherwise, the compensation will not be
performed correctly.
Color compensation is enabled by default, as can be verified by selecting Others… from the menu
Preferences (Fig. 93).
71
Fig.93 Enable signal compensation in the Others window
In the opened window Others on the tab Signal compensation the Use_ColorCompensation must be
enabled. If this option is disabled, the color compensation will not be applied.
Attention! It is not recommended to use this diagnostic service options by the end user.
To diagnose the operation of the device, select Device diagnostics from menu Preferences of the
RealTime_PCR software window:
to enter the diagnostic commands for the device, select item Command line. The Command
string window will be opened (Fig.94). Buttons are intended for the selection of
the controller, the diagnostics of which has to be performed;
72
to display the diagnostic information, select item Monitoring. The Monitoring window will be
opened (Fig.95).
Fig.95 Monitoring
7 User settings
In a RealTime_PCR software it is possible to save individual settings for each user (operator) of the
software: operator name, a folder for saving files, font style, window presentation style (only for
Windows XP operating system).
To change the user settings, open the window Options... (select item Options... in the menu
Preferences).
Tab List of operators of the Options window (Fig.96) allows performing the following actions: adding,
editing and deleting operators, assigning a folder to save files.
After clicking the buttons Add or Change, the Add operator window will be opened, in which
specify/edit Operator name and Directory (folder) for saving files (for more details, see "Starting the
RUN ").
73
Fig.96 List of operators tab of the Options window
Tab Interface Parameters (Fig.97) is intended for selecting the font of the software and the language.
Analysis parameters, set by default, are optimal for daily operations. When conducting scientific
research and for more detailed analysis of the obtained data, change the preset parameters for
processing the results of optical measurements and data analysis.
Attention! When working with commercial kits, the settings of analysis and device settings have to
be requested from the kit manufacturer.
74
Parameters change is available at various stages of the software:
In this case, a window Parameters of the analysis is displayed (Fig.97) and the following settings are
available:
Criterion of the PCR positive result (in the range from 60 to 100%) allows changing the
sensitivity to the rate of growth of PCR products. By decreasing the value of this parameter,
visualization and calculation of more gently sloping curves can be achieved;
Threshold value allows establishing a "threshold" level of fluorescence when determining the
indicator cycle (it is considered in the threshold method of curve analysis);
Criteria of validity of results are the parameters that allow adjusting the conditions for
calculating the curve under the geometric method of analysis:
the threshold of the positive result is the lower limit;
the threshold of data normalization is the upper limit.
Normalization: the use of the "amplitude normalization" method.
Method of analysis is selected during the Test creation stage or directly in the Analysis of optical
measurements window.
Threshold (Ct) and Geometric (Cp) methods for analyzing the DNA accumulation curve during PCR are
based on different approaches to determining the indicator cycle of amplification.
The Threshold (Ct) method is based on conducting a threshold line that is parallel to X line and
determining the cycle number (threshold cycle), at which the amplification curve for this tube
crosses the threshold line. Threshold cycle is one of the variants of indicator cycles;
Geometric method (Cp) is based on a mathematical analysis of the shape of the DNA
accumulation curve during PCR.
75
By default, when analyzing optical measurements, RealTime_PCR uses the Geometric method (Cp). In
case of non-standard situations, when the amplification curve differs significantly from the classical
DNA accumulation curve in the course of PCR (sigmoid), it is possible to analyze the results using the
Threshold method (Ct), which, in such cases, gives a better estimate.
Along with the processed results (deduction of the baseline, fitting), when analyzing the results of
optical measurements, it is possible to view raw data, selecting the appropriate position in the drop-
down list of analysis methods.
9 Emergency situations
Attention! There was a malfunction when the device was The device identification failed when
initialized! connecting to the PC. It is necessary to repeat
Try to repeat the selected operation... the procedure of connecting to the device...
Attention! An error occurred while reading the optical mask of An error occurred while reading the setting
the device... parameters from the device.
You need to restart the application! It is necessary to repeat the procedure of
Attention! An error occurred while reading the parameter connecting to the device...
block from the device...
You need to restart the application!
Attention! An incorrect exposure value when reading the The validation of the parameter block from the
parameter block from the device is detected... device revealed the presence of incorrect values
It is necessary to check the exposure values and correct if of the individual parameters of the tuning
necessary! factors: exposure, coordinates of the optical
Attention! An incorrect coordinate value of the geometrical geometry, etc.
setting of the optical image was detected when reading the It is necessary to correct these parameters and
parameter block from the device... save them again in the device. Contact the
It is necessary to check the geometric setting values and technical support of the manufacturer.
correct if necessary!
Attention! An incorrect value of the geometric adjustment
coordinate offset is detected...
It is necessary to contact the service engineer of the
manufacturer!
Attention! An error occurred while reading the protocol from When starting a new protocol or connecting to
the device... device in which the protocol is already running,
It is necessary to repeat the reading procedure! the protocol reading/writing failed, respectively.
Attention! An error occurred while writing the protocol to the It is necessary to repeat the read/write
device... procedure.
It is necessary to repeat the writing procedure!
Attention! An incorrect value of the coefficients of optical The validation of the optical coefficients of the
nonuniformity is detected... device revealed the presence of incorrect values.
It is necessary to contact the service engineer of the Please contact the technical support of the
manufacturer! manufacturer.
Attention! An incorrect value of the spectral nonuniformity
coefficients is detected...
It is necessary to contact the service engineer of the
manufacturer!
Attention! An incorrect value of Unequal Nonuniformity
Coefficients has been detected...
It is necessary to contact the service engineer of the
manufacturer!
Attention! An error occurred while writing the protocol data to A glitch occurred when writing data to the
the device! idevice at the SD card level. Please repeat the
writing to the device procedure.
Attention! The temperature gradient/difference does not The temperature gradient/difference in the
correspond to the temperature block of the device! amplification program loaded into the device
It is necessary to make appropriate changes in the does not correspond to the temperature unit of
amplification program... the device, or MAY NOT correspond (in the
Attention! The temperature gradient/difference may not case of old device, the information about the
correspond to the temperature block of the device! thermal block is missing). It is necessary to
Do you want to continue the program? make appropriate changes in the amplification
program...
76
Attention! Incorrect values of geometric settings are The validation of geometric settings in the
detected... "Preview Image" procedure revealed the
Continue the saving procedure? presence of incorrect values.
Attention! Problems when choosing colour-filters (MBUSY)... During the acquisition of the optical picture in
the "Preview Image" procedure, there was a
malfunction in the device when the color-filter
was selected. Repeat the procedure for reading
the optical image.
Attention! Incorrect values of exposure values are detected... The validation of exposure values in the
Continue the saving procedure? procedure "Exposure of measurements"
revealed the presence of incorrect values.
Attention! During the execution of the program, errors were During the execution of the protocol, errors in
found while storing the measurement values in the results file. the procedure for recording optical
It is necessary to consider the last run in the device (the measurement data in a file with results were
corresponding item in the menu) immediately after the detected. It is necessary to read the last start in
program is executed in the device! the device immediately after the program is
Attention! There were troubles reading the part of executed in the device.
optical measurements data from the device!
Upon completion of the amplification program, it is necessary
to
read the last start from the device...
Attention! An emergency stop of the program has occurred! During the execution of the amplification
program in the device, an emergency stop of the
program occurred. Please contact the technical
support of the manufacturer and provide log
files (if possible).
Attention! An emergency has been detected in the device! At the end of the amplification program in the
device, an emergency situation was detected.
Please contact the technical support of the
manufacturer and provide log files (if possible).
Can not download the protocol! A discrepancy between the When creating a protocol and using an already
test list and the test in the protocol was found! existing protocol as a template, the list of tests
for the operator and the list of tests in the
template protocol is validated.
Attention! An inconsistency between the current list of tests When reading the last run, the device found a
and the list of tests in the last run was detected! mismatch in the tests in the read protocol and
Check the name of the operator under which the last run was with the list of tests of the current operator,
read... under whom the procedure for reading the last
run is started.
Attention! The block is not in working order! It is necessary to perform the procedure of
opening/closing the thermal block.
Attention! In case of contacting the customer support service, provide the following:
77
Appendix 1. Glossary of Basic Terms
Thermocycler, Instrument, Device the device, intended for polymerase chain reaction (PCR); it is a
programmable thermostat capable of heating and cooling a tube with a reaction mixture according to a
certain program.
The internal control sample (internal control, IC) DNA or RNA added to the analyzed sample in
order to monitor the correct performance of all operations during the analysis. If there is no DNA or
RNA in the sample, the internal control should be amplified (otherwise the result of the analysis is
considered as invalid).
Duplicate (duplicates) a parallel set of two or several absolutely identical reactions. It is used to
average the effect of random factors on the result.
Intercalating dyes substances that intensively fluoresce when bound to double-stranded DNA or
RNA (eg, ethidium bromide, SYBR Green I).
Qualitative analysis determination of the presence or absence of nucleic acids of the organism to be
detected in the test sample.
Quantitative analysis determination of the concentration of nucleic acids of the detected organism in
the test sample.
Melting curve a plot of fluorescence versus temperature upon heating (or cooling) the reaction
mixture after the PCR termination.
Multi-test mode the option that allows combining several tests in one analysis with the same
amplification program and analysis type.
Amplification program a program that specifies the mode of carrying out the amplification reaction.
78
Fluorescent dye a chemical compound capable of effectively fluorescing. Can absorb light energy of
a certain wavelength (excitation wavelength) and re-emit it at another wavelength (wavelength of
radiation).
Exposure of measurements – the parameter that regulates the sensitivity of the optical system of the
device.
79
Appendix 2. Types of analysis (only the basic ones are represented)
When creating tests and working with different types of analysis, the settings of the analysis and
device settings are requested from the kits manufacturer.
1. Qualitative analysis
Qualitative analysis allows determining the presence or absence of the target DNA fragment in the
sample.
To perform a qualitative analysis (Fig.99) it is necessary to select a Qualitative analysis from the Type
of analysis drop-down list when creating a Test. In case of creation a plate protocol manually without
using the Test option, this type of analysis is used by default.
In the first and second columns of the Results tab, the tube numbers and their identificators are
indicated. The choice of the numbering type (by location in the thermal block or by the number in the
plate protocol), as well as the display of the "color" of the tube, is made using the button View data
representation modes below the table:
The results of a qualitative analysis are displayed in a column Qualitative analysis in the form of the
following values:
there are fragments of the desired DNA in the sample (the exponential growth of the
signal for the Specifics channel is recorded).
there are no fragments of the desired DNA in the sample (there is no exponential
signal growth for the Specific channel, in the case of using an internal control sample (IC), an
exponential growth of the signal for the channel IC).
the unreliable result (when using IC - there is no exponential growth of the signal,
both for the Specifics channel, and for the channel IC).
80
In columns : the values of the output cycles for the fluorophores used for each sample (Ct or
Cp depending on the selected analysis method in the drop-down list Method of analysis).
Quantitative analysis using calibration samples ("standards") allows determining the amount of the
desired DNA fragment in the sample. For the quantitative analysis, it is necessary to include the samples
with a known amount of the desired DNA in the plate protocol (calibration samples, "standards").
There is option in the software that allows using calibration samples from other RUN file.
For quantitative analysis (Fig.100) it is necessary to select a Quantitative analysis with standards
from the Type of analysis drop-down list when creating a Test.
After the analysis, if there are calibration samples, the software will automatically build a calibration
plot (you can click the button to view it Show the "Standard curve" plot in the basic toolbar
"Panel of device ") and will determine the concentration of the desired DNA in unknown samples.
When using option of adding the calibration samples from other RUN file, it is necessary to perform the
following actions:
proceed to the tab Additional standards in the Archive mode window;
click the button , select the file with the corresponding RUN (file type *.r48, *.r96, *.384),
the "standards" of which will be used as additional standards. For detailed information see
Additional standards.
The results of the quantitative analysis are displayed in a column Absolute concentration.
81
In columns : the values of the "characteristic" cycles for the fluorophores used for each
sample, (Ct or Cp depending on the selected analysis method in the drop-down list Method of
analysis).
Interpretation of the results of the analysis is carried out in accordance with the instruction for use to the
PCR kit used.
Additional standards
For analysis, it is possible to use "standards" (samples with the known content of initial concentration)
from another RUN file or Additional Standards (Fig.101).
In this case, there is option to conduct analysis in one run up to 96 tubes with unknown samples.
Fig.101 Display of additional standards when selecting the type of analysis "Quantitative with
standards"
Attention! An obligatory condition for the use of additional standards is the identity of the RUN
(amplification program, test type).
at the stage of creating a plate protocol using the procedure Add test;
at the stage of the results analysis in the tab Additional standards.
Standard curve
The following data is displayed in the Standard curve window (see Fig.102):
82
coordinate system (X-axis is the logarithm of the sample concentration Log10 (Concentration),
the Y-axis is the value of Ct or Cp);
calibration samples (indicated by );
studied samples (indicated by );
"standard curve";
the efficiency of PCR, the standard deviation, the reliability of the approximation.
Fig.102 The window Standard curve when selecting the type of analysis "Quantitative with standards"
Note! The software constructs straight line of the form Y = A*x+B from the points of the calibration
samples , for which the concentration and the value of Ct (or Cp) are known. The equation of the
resulting straight line is indicated in the name of the Standard curve window.
On the constructed straight line the investigated samples are displayed in the form of icons , in
accordance with the obtained value of Ct (Cp). Based on the Log10 concentration found, the software
calculates the concentration of the analyzed samples.
The quality of the approximation of the "Standard curve" is represented by the Confidence level of
approximation and the Standard deviation values.
3. Relative analysis
The Relative analysis makes it possible to estimate the amount of the desired DNA fragment in the test
samples relative to the "base" tube. The value of the relative concentration of the "base" tube is one.
When conducting a relative analysis, it is assumed that the efficiency of PCR for all samples is the
same. Any of the samples can be selected as a "base".
To carry out a relative analysis (Fig.103), select a Relative analysis from the Type of analysis drop-
down list when creating a Test.
open the window Relative concentration, located in the upper left corner of the fluorescence
graph (click the icon );
in the opened Relative concentration window choose one of the following values:
83
a sample with a minimum value of Ct (or Cp),
an arbitrary sample.
By default, the relative concentration is calculated assuming a 100% efficiency of the PCR reaction.
If the titration indicators of the individual samples are known, the result of the values of the relative
concentration can be corrected by changing the value of the efficiency coefficient using the field
The results of the relative analysis are displayed in a column Relative concentration as the relative
concentration of the amplified fragments (amplicons) in each of the tubes relative to the tube chosen as
the base one.
In columns : the values of the "characteristic" cycles for the fluorophores used for each
sample (Ct or Cp depending on the selected analysis method in the drop-down list Method of
analysis).
4. Multiplex detection
Multiplex detection in this section means the types of data representation. For all RUNs in which
classical PCR is present, it is possible to select the type of presentation Multiplex detection from the
84
types of analysis (Fig.104). Information on all channels in each tube shows information of a qualitative,
quantitative or relative analysis, respectively. This type of presentation is designed to work in multi-
channel mode, and it allows detecting up to five specific DNA fragments simultaneously.
Note! To conduct a relative analysis, select a Multiplex detection from the Type of analysis drop-
down list when creating a Test.
5. Melting curve
For all RUNs in which the procedure of "melting" of probes is present, select the Melting curve type of
data representation from the corresponding item of analysis types that allows estimating the values of
the temperature melting peaks for future analysis of genetic polymorphisms (Fig.105).
Note! To conduct a Melting curve (MC) analysis, select the related Analysis of polymorphism (MC)
from the Type of analysis drop-down list when creating a Test.
85
Appendix 3 DTReport library description
The DTReport library is intended for implementation of algorithms’ analysis and results' calculation
based on data obtained from RealTime_PCR software.
The library includes the following components:
- DTReport core;
- a file with code on the DTReport computation language ("dtr file");
- a file with PCR test parameters («ini-file»);
- report generator.
The DTReport library provides wide opportunities for assay calculations and for visualizing the
calculated results in the form of tables and reports.
1. Computing modules
Any type of analysis is implemented by means of a calculating module (hereinafter referred to as a
module) developed in the DTReport Studio environment (application is not included in the library).
The module contains the following components:
- calculation of results;
- generating reports;
- visualization of results;
- settings of assay parameters.
The module is built into the application using the DTReport core. Such architecture allows using of the
same compute modules with RealTime 7 software.
Fig. 106 Architecture of interaction between RealTime_PCR software, DTReport core and
computational modules
2. Data structure
DTReport operates with data organized in a hierarchical structure of the following structure:
Document (protocol)
Request
Sample
Test
Result
3. Localization
DTReport Studio includes localization tools that allow to localize both the system core and the
computing modules without rebuilding the software, including:
- report form;
- string constants used in calculations;
- GUI elements (settings of the results' presentation).
86
87
Manufacturer: DNA-Technology, Research & Production, LLC
20 Zheleznodorozhnaya Street, Protvino,
Moscow Region, Russia, 142281
Phone/fax: +7(4967) 31-06-70
E-mail: [email protected]
http://www.dna-technology.ru
Customer support:
Tel.: 8 800 200-75-15 (free for Russia)
E-mail: [email protected]
Feedback form is avalible on DNA-Technology’s website
http://www.dna-technology.ru/customer_support/
Service department:
Tel.: +7(4967) 31-14-67, +7(4967) 31-06-71 (ex. 3126)
E-mail: [email protected]
Version 07-2021
88