CoA TDS1000 - LRAD7678

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Document 20550009 Version 2.

Certificate of Analysis - Certified Reference Material


Total Dissolved Solids 1000 mg/L Calibration Standard
Product no.: TDS1000-100ML
Lot no.: LRAD7678
Expiry Date: January 2027
Manufacturing Date: January 2024
Storage: ROOM TEMPERATURE (2 °C to 30 °C)
Solvent/Matrix: DI WATER
Certificate version: LRAD7678.01 (Note: Certificates may be updated due to
the availability of new data. Check our website at:
www.sigma-aldrich.com for the most current version.)

Certified Values:

Certified Raw Material Raw Material


Analyte Units Value Purity,% Lot

TOTAL DISSOLVED SOLIDS mg/L 1000 ± 3 99.99 20K305207


Additional Information: SAMPLE PREPARATION
The sample is ready to use.
No sample preparation is required.
Metrological traceability: Traceable to the SI and higher order standards from NIST through an unbroken
chain of comparisons. The balance used to weigh raw materials is accurate to
+/-0.0001 g and calibrated regularly using mass standards traceable to NIST. All
dilutions were performed gravimetrically. Additionally, individual analytes are
traceable to NIST SRMs where available and specified above.
Measurement method: Where applicable, the assigned value is based on a purity determination by mass
balance and gravimetrically prepared value.
Intended use: Intended for R&D and Analytical Use only. Not for drug, household or other uses.

Packaging: 100 ML IN HDPE BOTTLE


Instructions for handling Use on the as is basis. The internal pressure of the container may be slightly
and correct use: different from the atmospheric pressure at the user`s location. Open slowly and
carefully to avoid dispersion of the material.
Health and safety All chemical reference materials should be considered potentially hazardous and
information: should be used only by qualified laboratory personnel. Please refer to the Safety
Data Sheet for detailed information about the nature of any hazard and
appropriate precautions to be taken.
Accreditation: Sigma-Aldrich RTC is accredited by the US accreditation authority ANAB as a
registered reference material producer AR-1470 in accordance with ISO 17034.
Certificate issue date: 03 JAN 2024

Andy Ommen - QC Manager Scott Stetler - QA Manager

Sigma-Aldrich RTC, 2931 Soldier Springs Rd. Laramie, WY 82070, USA;


Tel. 1 307-742-5452; Fax 1 855-831-9211; www.sigmaaldrich.com
Sigma-Aldrich RTC is a subsidiary of Merck KGaA, Darmstadt, Germany.

Certificate Page 1 of 2 Certificate version 01


Details on metrological This standard has been gravimetrically prepared using balances that have been
traceability: fully qualified and calibrated to ISO 17025 requirements. All calibrations utilize
NIST traceable weights which are calibrated externally by a qualified ISO 17025
accredited calibration laboratory to NIST standards. Qualification of each balance
includes the assignment of a minimum weighing by a qualified and ISO 17025
accredited calibration vendor taking into consideration the balance and installed
environmental conditions to ensure compliance with USP tolerances of NMT
0.10% relative error. Fill volume to predetermined specifications is
gravimetrically verified throughout the dispensing process using qualified and
calibrated balances. Further traceability to a corresponding Primary Standard may
be achieved through a direct comparison assay. Where a Primary Standard is
available, the assay value will be included in the specified section of the COA.

Associated uncertainty: Ucrm - Uncertainty values in this document are expressed as Expanded
Uncertainty (Ucrm) corresponding to the 95% confidence interval. Ucrm is
derived from the combined standard uncertainty multiplied by the coverage factor
k, which is obtained from a t-distribution and degrees of freedom. If k is not
provided, assume a value of 2.0. The components of combined standard
uncertainty include the uncertainties due to characterization, homogeneity, long
term stability, and short term stability (transport). The components due to
stability are generally considered to be negligible unless otherwise indicated by
stability studies. The mathematical representation of the Ucrm calculation is as
follows:
2 2 2
𝑈𝑐𝑟𝑚 = 𝑢𝑐ℎ𝑎𝑟𝑎𝑐𝑡𝑒𝑟𝑖𝑧𝑎𝑡𝑖𝑜𝑛 + 𝑢ℎ𝑜𝑚𝑜𝑔𝑒𝑛𝑒𝑖𝑡𝑦 + 𝑢𝑠𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦 × 𝑘

Homogeneity assessment: Homogeneity was assessed in accordance with ISO Guide 35. Completed units
were sampled using a random stratified sampling protocol. The results of
chemical analysis were then compared by Single Factor Analysis of Variance
(ANOVA). The uncertainty due to homogeneity was derived from the ANOVA.
Heterogeneity was not detected under the conditions of the ANOVA.
Stability assessment: Significance of the stability assessment will be demonstrated if the analytical
result of the study and the range of values represented by the Expanded
Uncertainty do not overlap the result of the original assay and the range of its
values represented by the Expanded Uncertainty. The method employed will
usually be the same method used to characterize the assay value in the initial
evaluation.

Certificate of analysis revision history:

Certificate version Date Reason for version


LRAD7678.01 03 JAN 2024 Original release date
Disclaimer: The purchaser is required to determine the suitability of this product for
any particular application. Sigma-Aldrich RTC makes no warranty of any kind, express
or implied, other than its products meet all quality control standards set by
Sigma-Aldrich RTC. We do not guarantee that the product can be used for any
particular application.
The vibrant M, Supelco, TraceCERT and Sigma-Aldrich are trademarks of Merck KGaA, Darmstadt, Germany or its affiliates.
All other trademarks are the property of their respective owners. Detailed information on trademarks is available via
publicly accessible resources. © 2018 Merck KGaA, Darmstadt, Germany and/or its affiliates. All Rights Reserved.

Certificate Page 2 of 2 Certificate version 01 LRAD7678