Manual GeoTest
Manual GeoTest
GeoTest is a powerful software tool for data acquisition and controlling of geoelectric equipment using
many electrodes, so called multi-electrode-geoelectrics. Profiling and mapping measurements are
possible, as well as custom made measurements. Moving arrays with GPS position measurements and
completely automatic monitoring measurements are supported also.
Additional inversion software (not part of GeoTest) is used for Geoelectric Tomography.
Safety requirements
Be careful in using geoelectric equipment. Geoelectric devices can produce high
electrical voltages of 200 Volts or more.
Be always sure to switch the power off, when you are touching electrodes or
power outlets to avoid injuries.
GeoTest is a software for controlling geoelectric equipment in combination with many electrodes. This
type of geoelectric measurement is called "Multielectrode Geoelectrics". Using two-dimensional or
three-dimensional inversion techniques additionally for evaluation, this method is known for "Geoelectric
Tomography" or ″Electrical Resistivity Tomography″ (ERT).
Geoelectric tomography is a modern type of 4-point geoelectrics. "4 point" means, for one measurement
always four electrodes are used. Two electrodes serve as current emitting electrodes. These are usually
named A and B (or C1 and C2). Two other electrodes measure the potential difference (= voltage). These
are the M and N electrodes (or P1 and P2).
The measurements can be done in two principal different orientations. GeoTest supports both:
Profiling: this means measurements, where the electrodes are set along a linear profile on the
earth surface. The measurement profile is 1-dimensional (e.g. in x direction) and the result will be
a 2-dimensional depth section, which is a plane in e.g. x- and z-direction.
Mapping: the electrodes are set along a 2-dimensional grid or along several parallel profiles on
the earth surface. The measurements are done in a 2-dimensional plane (e.g. in x- and y-
directions). The result will be a 3 dimensional block containing x-, y- and z- coordinates.
The following pictures show some explanation of the technique to measure one geoelectric profile
("profiling").
For a further explanation of ERT describing the standard comfigurations like Wenner or Dipole-Dipole see
Berktold et al. (1997) or Friedel (1997). The investigation depth definition used in GeoTest is described in
Roy and Apparao (1971) and Barker (1989). Edwards (1977) introduced another definition of the
investigation depth. Newer configurations like the Multiple Gradient Array or the Full Range Gradient
Array″ are described in Dahlin & Zhou (2006) and in Zhou et al. (2020). See references.
2.1 Computer
GeoTest runs on desktop or notebook PCs with Windows 95 (TM) or later operating systems. GeoTest
runs on 64 bit systems like Win 10 as well. Communication with the measurement device is done by a
serial (RS232) interface. This RS232 interface may be an original COM1 port or a driver emulated COM
port using an USB-to-serial adapter. In this case, please be sure to install the device driver software
before using the adapter. The computer hardware specifications are not critical. We recommend:
Windows XP (TM) 32 bit version or later
at least 512 MB RAM
1 RS232 serial port or 1 USB port with an adapter USB-to-serial
in case of GPS measurements: another RS232 serial port or USB-to-serial adapter.
► ► ►
►
This example shows 3 installed COM ports: COM1 and COM2 are the standard COM ports of the desktop
computer. COM3 is an additionally installed USB-to-serial converter.
Geotest offers an automatic search for the suitable COM port with connected device (see following
chapter)
"4-Point-Light hp" or "4-Point-Light 10W", a high power, low cost induced polarization (IP) earth
resistivity meter: for Schlumberger soundings to L/2 = 1 km, for mapping, monitoring, geoelectric
tomography, lab work, induced polarization measurements, archeology.
"ActEle", active electrodes for electrical resistivity tomography with direct interface for 4point light.
Can be used to upgrade existing earth resistivity meters to state-of-the-art tomography systems.
Up to 255 electrodes are possible.
LGM
Erich Lippmann
Kornacker 4
94571 Schaufling
Germany
Tel: +49 (0) 9904 84076
Fax: +49 (0) 9904 8119802
email: [email protected]
http://www.l-gm.de/
This device uses a ground electrode to reference the potential measurements. This ground electrode is
important and absolutely necessary. The ground electrode is not part of the measurement profile or grid.
It has to be placed somewhere nearby the instrument but not near besides one of the profile electrodes.
Be sure to realize low contact resistance of the ground electrode. Connect the ground electrode to the
"GND" input terminal of the instrument. In a laboratory environment without access to the earth surface,
you may connect the GND Terminal to the electrode A terminal.
Computer interface
This equipment needs one serial interface to the computer (RS232). This can be the standard COM1 port
(if the computer has one) or some driver emulated COM port, e.g. by using an USB-to-serial converter.
Due to our experience, the using of the standard COM1 port is without any problems. Not every USB-to-
serial converter works properly. If you get some communication problem, it is most likely a result of not
proper functioning of the USB-to-serial converter. You could fix this problem by either (1) updating the
USB-to-serial converter driver software or by (2) using another converter.
Go to the GeoTest main menu → Set → Options → Getting started → Set measurement device.
GeoTest supports position measurements done by a GPS device connected to one of the serial data
ports of the computer. The GPS device must be able to send one of the "NMEA-0183" protocols GGA,
GLL, RMC or LLQ. Example of a NMEA-0183 GGA sentence data line:
char position:
1 2 3 4 5 6 7 8
1234567890123456789012345678901234567890123456789012345678901234567890123456789012
string:
$GPGGA,121357.00,5226.316328,N,01300.878187,E,2,07,1.1,35.59,M,45.19,M,5.2,0100*75
pos. contents
8 UTC of position fix in HHMMSS.SS format
18 Latitude in DD MM,MMMM format(0-7decimal places)
30 Direction of latitude N or S
32 Longitude in DDD MM,MMMM format(0-7decimal places)
45 Direction of longitude E or W
47 GPS Quality indicator; O= fix not valid; 1=GPS fix; 2=DGPS fix
49 Number of SVs in use, 00-12
52 HDOP
56 Antenna hight, MSL reference
62 M indicates that the altitude is in meters
64 Geoidal separation
70 M indicates that the geoidal separation is in meters
72 Age of differential GPS data record, Type 1. Null when DGPS not used
76 Base station ID, 0000-1023
Suggestion for a communication test: Connect the GPS device with the computer. Switch on the
device and set the correct COM and NMEA values. Let the GPS device send data continuously to the
serial data port. If the correct COM parameters are set, you should see the continously updated positions
in the box "Reading".
3.1 Installation
Complete installation: The user should install the program and all its components using the default
settings. The default settings are:
Updating: For updating the already installed program, it is sufficient to copy the files to be updated into
the <Windows ProgramFilesDir>\GeoTest folder. The already installed files will be overwritten.
The last settings will automatically be saved in the "GeoTest.ini" file, which is located in the folder:
<Windows user application data folder\GeoTest>.
3.2 Registration
After the first installation of the program, it will run in the limited modus. The
limitation refers to e.g. to the limited maximum amount of electrodes. This limited
modus is thought as a demo modus for interested people. Due to this limited
modus, a free copy and distribution of the program is prevented.
Locate the "Computer ID". This is your computer fingerprint. You should tell us this number. You could
use the "-> eMail" button to open your eMail client and to send your computer ID by eMail directly.
I will calculate the "Key" and create the "User name" and will sent these to the user.
The user must type in the "Key" (or copy-paste) and the "User name". After pressing the "OK" button
the key will be saved in the INI file (if the key matches the computer ID). There is no need to type in again
the key after a later starting of the program, nor after a later updating of the program. Nevertheless, I
recommend the user to save this key in a hardcopy paper form for an own reference.
If the user wants to install GeoTest on another computer, again another registration procedure is
necessary.
The program GeoTest can be used in a normal way by double-left-clicking on the GeoTest
icon on the computer desktop. The main form, described in the following chapter 4.1, will
appear. Alternatively, the user may start GeoTest using the command line options
described in chapter 4.3.
Frequency
settings
Transmitter
Receiver
Measurement
status
Battery
Data power
list
By clicking on the "mV" label,
the user can switch between
displaying of single voltage or
voltage list.
4.1.1.1 File
"Save": the current data file will be saved using the GeoTest
data format (ascii file with the extension *.tx0). All electrode
and cable parameters, settings, measured contact resistances,
error messages and measured data are saved. The saved
data format is explained in the "Data list" chapter.
"Save in 'Res2DInv' format": data is saved using the data format of the 'Res2DInv' geoelectric inversion
program (by 'GeoTomo software', M. H. Loke, Malaysia, www.geoelectrical.com). This is a 2 D inversion
software for evaluation of profile measurements of resistivities (done along 1 profile direction).
"Save in 'Res3DInv format": data is saved using the data format of 'Res3DInv' geoelectric inversion
program. (by 'GeoTomo software', M. H. Loke, Malaysia, www.geoelectrical.com). This is a 3 D inversion
software for geoelectrioc measurements done using 2 directional x-y-grids.
"Save in 'SensInv2D' format": This format is used by the 'SensInv2D' program of Geotomography, Dr.
Th. Fechner, Germany (www.geotomographie.de). This is a 2 D inversion software for evaluation of
profile measurements of resistivities (done along 1 profile direction).
"Save in 'DC2dInvRes' format": This format is used by the 2 D inversion software 'DC2dInvRes' of
Thomas Günther (http://www.resistivity.net/dc2dinvres/).
"Save in 'DC2dTree' format": This format is used by the 2 D inversion software 'DC2dTree' of Thomas
Günther and Carsten Rücker (http://www.resistivity.net/dc2dtree/). This software enables the
"Save in 'DC3dInvRes' format": This format is used by the 3 D inversion software 'DC2dInvRes' of
Thomas Günther (http://www.resistivity.net/dc3dinvres/).
"Save in 'EarthImager' format: This format is used by the EarthImager 2D and 3D inversion software of
Advanced Geosciences Inc., USA (www.agiusa.com). Two files will be saved: a *.URF file contains the
measured data and a *.TRN file ("terrain file") contains the topography information.
"Save in 'ZOND' format: This is the format is used by the inversion softwares 'ZondRes2D' and
'ZondRes3D', written by Alex Kaminsky (http://zond-geo.com). Files will be saved using the extension
'z2d' and 'z3d' respectively.
"Save in 'CSV TXT format': This is a simple CSV = "comma separated values" ascii text file. This file
could be used for user made correlations e.g. by Excel.
"Read measured data from device": data measured in the autonomous mode (MON, TOM) of the
geoelectrical device (without controlling by a computer) will be read from the device, transfered into the
GeoTest data format and displayed.
"Save parameter set to device": A list of currently set parameters (electrode configurations and
measurement settings) will be transferred into the geoelectrical device. Afterwards, the device can
measure tomography data in an autonomous mode, using this parameter set , without any computer for
controlling the measurement.
These 2 device options are enabled only when using the "4 Point Light 10W" geoelectrical device. See
the respective hardware manual for more details.
"Exit program": the program GeoTest will be closed. If there is unsaved data, the user will be asked to
save before closing the program.
4.1.1.2 Set
Electrode chains (electrode address #) means chains (or sequences) of active, addressed electrodes
controlled by a bus system. Each electrode address is unique and fixed physically to one switch box and
one electrode. The addresses may be in increasing or in decreasing order as well. The user can choose
electrode chains by checking the respective check boxes. The actual electrode numbers of each used
cable are shown and can be edited.
Electrodes constants: The editable yellow boxes characterize the very 1st electrode in the profile or the
1st electrode in a mapping grid. These values remain constant even during a "Roll-on" measurement.
By clicking the Roll on button, the cable sequence will be changed in
that way, that the first cable will be placed after the last cable.
GeoTest will re-calculate the profile electrode numbers. See the
chapter ″4.2 Roll-on procedure″ for an explanation.
Defaults: The defaults values are: the electrode chains (active or not, adresses) and the electrode
constants (yellow fields).
Set new user defaults will ask for new settings, which will be stored as the new default values.
Reset will use the default values as the new actual settings.
Set will set update actual values.
Example: In the above example, 3 chains are used (chain 1,3 and 5) with the corresponding addresses:
1..20 + 41..60 + 81..100. These are 60 adresses. GeoTest will use 60 profile electrodes. The number of
the 1st electrode is given to 100, therefore the profile electrodes will cover the range from 100 to 159. The
1st profile electrode 100 corresponds to electrode address 1. The last profile electrode 159 corresponds to
electrode address 100. The electrode separation is set to 5 m. The start electrode is the number 1 at
profile position 1 * 5 m = 5 m. The 1 st profile electrode is located at 100 * 5 m = 500 m. The last profile
electrode is located at 159 * 5 m = 795 m.
Check electrode chains: The program
checks the cable connection to the
electrodes and the additionally the mean
cable resistance. The cable test will find e.g.
broken cables. During this test, all electrodes
may have contact to the earth as usually (=
grounded) or not, that means only with the
multi core cable connected to the instrument.
In the latter case the terminal ″A″ has to be
connected with the terminal ″GND″.
"Data header": The data name will be used as default for the data file name. Some additional comment
can be given to describe the measurement.
"Data folder": The default values for the disk drive and path (data folder), where data files will be stored.
The user may give another folder or file name during saving.
4.1.1.2.2.2 Profiling
This tab sheet gives access to all settings which are necessary for profiling measurements. A profile is a
linear sequence of electrodes situated along the Earth surface. All electrodes have to be equidistant.
Remark: Gradient measurements with remote A and B electrodes give no resistivity values, if the
remote electrodes are assumed to be infinitely far away. In this case, no geometry factor K is
possible. GeoTest is able to calculate resistivities when using the exact positions of the remote
electrodes.
"Spreads to be used": This box is active only if the Dipole-Dipole or
Schlumberger configuration is chosen. The spread is the length of the
electrode array from the first to the last electrode along the profile. The
spread depends on the actual settings of the dipole separations and the
dipole lengths. The program increases the spread by increasing dipole
separation and dipole length simultanously in different ways:
only increasing spreads are used (=spreads which are longer than the previous used spreads).
This is the standard option.
all possible spreads are used (= all dipole separation und length
combinations). This option may result in doubled data point.
all spreads which differ by at least 1 electrode distance.
all spreads which differ by at least 2 electrode distances.
Wenner
Schlumberger = Sounding
A-----M---N---B x = (xN+xM)/2
←n∙a→ ←a→ ←a→ z = min[(xB-x),(x-xA)]∙fac
← (n+2)∙a → fac = 0.5 to 1
Dahlin & Zhou (2006)
A---M---N-----B x = (xM+xN)/2
←a→ ←a→ ←n∙a→ z = min[(xB-x),(x-xA)]∙fac
← (n+2)∙a → fac = 0.5 to 1
A-----M---N---B x = (xM+xN)/2
←n∙a→ ←a→ ←a→ z = min[(xB-x),(x-xA)]∙fac
← (n+2)∙a → fac = 0.5 to 1
A-------------B-----N---M x = (xM+xN)/2
← (n+2)∙a → z = (x-xB)∙fac
←n∙a→ ←a→ fac = 0.5 to 1
Zhou et al. (2020)
K = 2∙Π / (1/AM - 1/AN - 1/BM + 1/BN) with AM=│xA-xM│, AN=│xA-xN│, BM=│xB-xM│, BN=│xB-xN│
Berkold et al. (1997)
"Sorting of data"
By default, the data is sorted by data number. The user can change the type of sorting and also the
direction of sorting.
"Reduce pseudodepths"
The user may confine the pseudodepth values of the data. By checking of
this option, the data will be reduced inside or outside the given depth
interval. Example: Data between 10 m and 20 m pseudodepths will be
reduced to 40 % of the original amount.
This tab sheet controls mapping measurements. The electrodes are situated along a 2-dimensional grid
on the Earth surface. The grid can be a regular and rectangular grid with given X- and Y- spacings and X-
and Y- grid sizes. The grid can also be a part of a regular grid (e.g. U-shaped) or can be a circular grid.
This box controls the setup of the measurement grid. The yellow fields within the box " Start electrode
[1,1]" control the coordinates of the starting electrode. The grid sizes and also the spacings are given for
X- and Y- direction separately.
The box "Electrode positions" gives access to the type of the measurement grid.
"Regular loops" and "Regular zig-zag" describe a regular and rectangular grid of electrode positions.
The grid sizes and spacings for the X- and for the Y- direction are given. The examples show a grid of the
dimensions 7 x 7 electrodes, X grid index between 1 and 7, Y grid index between 1 and 7. Starting
electrode[1,1] is located at [1m, 1m]. Last electrode[7,7] is located at [7m, 7 m].
"L-shaped" "Circle"
"From electrodes table": This option reads the electrodes positions from the list in the "electrodes
position" tab.This option may be useful for the setup of a custom made or not regular electrode grid.
Remark: The 3D inversion program Res3DInv does not support all electrode configurations,
which are possible in GeoTest.
Configuration
"Type" selects the electrode configuration type. Pole-Pole uses 2 field electrodes (A and
M) and additionally 2 "remote" electrodes (N and B). The actual positions of the "remote"
electrodes can be given in the tab sheet "Remote electrode positions". If the coordinates are
known, the exact geometric factors can be calculated (see "Program options"). Wenner
(alpha) and Wenner (beta) are linear configurations which are used in profiling
measurements also. Square uses 4 field electrodes in square configuration. Gradient uses
2 field potential electrodes (M and N) and 2 remote current electrodes (A and B). This
setting allows the computation of resistivities only for the case that the exact geometric K-
factors can be calculated from the known positions of the remote A and B electrodes. SP
Self Potential uses only 1 field electrode (M) for measuring the intrinsic voltage, without
applying any current to the ground. [Custom made] gives access to self made electrode
lists (see the respective chapter following).
"Spreads to be used": The direction of the field electrode array is given. The array
can spread in X-, Y- or diagonal direction as well. The
option "all" calculates all available spreads.
This tabsheet controls measurements with a moving array of electrodes. This option is only available in
the profiling mode.
The typical example is an array of 20 electrodes mounted on a vehicle. This car transports the electrodes
from position to position and Schlumberger soundings with 9 measurements at each position are
performed. The positions of the soundings are measured by a GPS device. The following explanations
refer to this example.
The user can force the GeoTest program to measure GPS positions continu-
ously or to measure one position per geoelectric cycle (default option).
The button "GPS setup" gives access to the GPS settings like
COM port, baudrate or NMEA sentence. The antenna height is
the vertical distance between the GPS antenna and the ground.
This tab sheet displays the 3-dimensional coordinates of all used field electrodes. In case of profiling
measurements, these are the equidistant profile electrodes, in case of mapping measurements, these are
the electrodes situated in the grid nodes. The remote electrodes, if there are any, are displayed in the
next tab sheet named "Remote electrode positions".
The coordinate system is the same for profile electrodes, grid electrodes and remote electrodes.
The units are meters (can be e.g. Gauss-Krüger coordinates). By default, the X-direction is the
direction of the profile. The Y-direction will run perpendicular to the "left" of the profile and the z-
direction is positive upwards. That means depths below the earth surface will get negative values.
The table gives the electrode number and the respective X, Y and Z coordinates. The above shown
example displays the electrode positions of a profile line. The line expands in positive x direction and
reaches from 1 m (1st electrode) to 49 m (last electrode). The y coordinates are perpendicular to the
profile direction and are all zero. The z coordinates display the elevation of each electrode and are all
zero too.
Locked electrodes
This button will open an additional form displaying all electrode coordinates.
This form can be used for "drag and drop" information from e.g. an Excel
spreadsheet.
This button saves the electrode coordinates displayed in the data list. The user
may edit the electrode position table by himself (e.g. in case of custom made
electrode grids or by giving elevations to each electrode position). After doing this,
the user has to press this button to make the changes active.
These yellow boxes display the start electrode coordinates. During profiling, the start electrode is the 1 st
electrode in the profile, during mapping, this is the electrode situated at the 1 st grid node [1,1].
This box is enabled in the profiling mode only. The box "End electrode" displays the coordinates of the
last electrode used in this electrode setup.
The electrode positions table will automatically be updated after clicking the "Set" button after choosing
profiling or mapping parameters.
This tab sheet displays information about the remote electrodes. Remote electrodes are situated far away
from the profile or the measurement grid. Some of the electrode configurations, like Pole-Pole, use
remote electrodes, some others, like Wenner, do not.
"A", "B", "M" and "N" display the positions of the remote electrodes and indicate if these electrodes are
used as remote electrodes. Usually, these remote electrodes are connected directly to the geoelectric
hardware device using e.g. the terminal ″B″. Alternatively, the remote electrode may be realized as an
addressed electrode too.
The above example shows B as the remote electrode of a Pole-Dipole configuration. B is situated at
profile position x = 10 m, with a distance of y = 1000 m perpendicular to the profile. This remote electrode
is connected not by the terminal ″B″ to the device, but as electrode number 201. A direct connection to
the respective terminal is indicated by an address of -1.
"Calculate geometric factors": The user can choose between exact and approximate calculations.
Approximate means that the remote electrodes are considered to be infinitely far away.
Exact means, that the real positions of the remote electrodes are used to calculate the geometric
factor. In case of "Gradient" measurements (M and N within the grid, A and B remote), a
resistivity value can only be calculated using the "exact geometric factor" option.
If remote electrodes are used without setting of the positions, only approximate calculations are possible.
If no remote electrodes are used, exact and approximate calculations will give the same results. This can
also be set in the "Program options" menu, see chapter 4.1.1.2.4.
"Save": This button allows to edit and save the remote electrode data. The program will re-calculate
geometry factors und resistivities. Other profile or mapping parameters will not be updated.
"Save" will save the electrode configurations into a measurement file (option: program data) or into a
data list (option: A B M N list).
"Set" will accept all changes. Please be sure to press this button to accept all
changes made.
"Current (mA)":
this box controls the measurement current. The "min" and "max" boxes
limit the minimum and maximum currents. Normally, these boxes display the really available minimum
and maximum current settings. The "now" box displays the start setting. "No current: measure self
potential (SP)": If this checkbox is active, no current will be applied during the measurement. Only the self
potential is measured. Please note: proper self potential measurements require the use of unpolarizable
electrodes (no steel electrodes). This option does not work with all hardware versions.
"Frequency (Hz)":
This box can choose between standard measurements with using of one single frequency or frequency
soundings. The one frequency or the multiple frequencies are set.
Measurement device "4 point light hp". The user should take care of the correct frequency setting. Users
in Europe: We recommend to use 1.04 Hz, 2.08 Hz, 4.16 Hz or 8.33 Hz (these settings give maximum
suppression of the 50 Hz mains frequency as well as the 16.7 Hz railway frequency). Users in the United
States: We recommend to use 1.25 Hz, 2.5 Hz, 5.0 Hz or 10.0 Hz (due to 60 Hz mains frequency).
Measurement device "4 point light 10W". This device is pre-set to one of two frequency sets. One set is
suitable for measurements in Europe, the other set is suitable for measurements in USA. Please notice: The
device frequency presetting (within the device menu) must be the same as the frequencies displayed in the
GeoTest software. See chapter 2.2.1. for details.
The "Set" button will accept all settings. It is necessary to press this button,
otherwise the settings will not be accepted.
"Getting started"
"ResXDInv"
This menu gives access to saving options for users of Res2DInv and
Res3DInv inversion software.
"Include topography": if this option is set, GeoTest will export
topography information into the ResXDinv data file. The topography
results from the settings in the "Electrode positions" section (elevation
= Z coordinate).
"General array data format": This setting exports ResXDinv data in
the so called "general array" data format. This is a useful option for
mixed arrays, that means measurements with different array types like
Wenner and Dipole-Dipole in one single measurement.
If "Save remote electrode positions" is clicked, these positions are saved too. This is useful only if
exact geometric factors are used additionally.
"Data format": The user can choose between rho = resistivities only or rho+phi = resistivities and phase
angles to be saved into the ResXDinv data file.
"Display"
Pseudo section
The color scale of the boxes
within the pseudosections can
be set. The same color scale will
be used for the resistivity (rho)
and phase angle (phi) display as
well. The colors may be
interpolated for a more detailed
view.
Show ...
The user may choose the displays on the right side of the main form. For example,
the checkbox "Date and time" will display the current date and time at the lower right
corner on the right side of the main form.
"more ..."
This form controls the measurement of contact resistances. The contact resistance is the resistance (=
impedance, measured in Ohm = Ω) between one electrode and ground. Although the measurements are
done by pairs of electrodes, the contact resistance of each measured electrode is calculated and shown
against the respective electrode number. For comparison, the respective electrode address is shown too.
The lower the contact resistance, the more current can be applied to the ground. High quality geoelectric
measurements need low contact resistances. Optimal contact resistances are < 600 Ω. Nevertheless,
even much higher contact resistances (> 50'000 Ω = 50 kΩ) may give useable geoelectric measurements.
The measured contact resistances are shown as a data list and in a graphics display. The graphics show
also the maximum allowed contact resistance as a horizontal red line (in the example 10 kΩ).
Electrodes: The numbers of 1st and last electrodes to be measured can be set. By default, these are the
1st and last available electrodes, formerly set in the "Electrodes and data header" (chapter 4.1.1.2.2.1)
menu. The user may choose only a part of the available electrodes.
Calibration: Three electrodes are used for the calibration measurement. By
default, the calibration electrodes are the first three electrodes. The user
may choose onother electrodes for calibration when one of the three
electrodes has bad ground contact or is locked. The 1 st of the calibration
electrode (Ref) is the reference electrode for all following contact resistance
measurements.
Current: The program uses the test current for the measurement cycle. If ″auto″ is checked, the test
current will automatically be adjusted for proper voltage readings.
Frequency: The test frequency for the mesasurement cycle.
Num: The number of measurements per one averaging.
Maximum contact resistance: The user gives the maximum allowed contact resistance (in the example
10 kΩ).
Measurements
The active A and B electrodes are displayed, as well as the active transmitter voltage
Buttons
"Clear" will clear all displayed values. The table and graphics are
cleared and all measured contact resistances are resetted.
"Show all" will show all measured contact resistance data. This may
be useful after the measurement of only few electrodes.
"min + max" will display a short statistics of measured contact
resistances like to be seen at the right side.
"Reset" will reset all settings to defaults and will reset all measured
contact resistances as well.
"Start ..." will start the measurements (alternativeley pressing the
<RET> key). The button caption will change to "... Stop". Pressing the button now (alternatively pressing
the <RET> or the <ESC> key) will stop the actual measurement.
"Close" will close the form..
Calibration: Three electrodes are used for a calibration of the measurement procedure. By default, these
are the first 3 electrodes. One of these 3 electrodes will serve further on as the reference electrode. The
calibration procedure results in the known contact resistance of this reference electrode.
Measurement: After the auto calibration, resistances are measured between the reference electrode and a
2nd electrode. The 2nd electrode moves fro the beginning to the end of all available electrodes. The reference
electrode and the 2nd electrode are choosen as current emitting electrodes A and B. The test current is
applied. The transmitter voltage is measured. The contact resistance of the 2 nd electrode computes from the
transmitter voltage devided by the transmitter current minus the known contact resistance of the reference
electrode.
Display: The measured contact resistances are displayed in the data list and in the graphics display. If the
measured contact resistance is less then the maximum allowed value, "OK" will be displayed in the data list.
... the form "Start and stop #" will appear. This window controls the
minimum and maximum number of measurements, which will be
performed. By default, the minimum and maximum available
numbers are displayed. The user is able to change these numbers,
for example to limit the range of numbers to be measured.
Active GPS option: In case of an active GPS option the user will be
asked to start the GPS position measurement immediately without
asking for start and stop numbers. The GPS measurements will be
stopped by pressing "Stop".
The "Show details" option will display all available information of the actual
data set and all measurerd data sets.
The "Show overview" option will display an overview of the measured data.
Not all available data is displayed. The displayed overview is the same as in
the data list window in the main program form (see chapter "Data list").
"Show error messages" will display a window containing error messages. Errors could be e.g. "voltage
too low" or "overload" or "open".
The displayed example allows the user to delete all data, which contain the electrode number 11 as
active potential electrode M or N. After pressing the ″Re-measure″ button, the formerly deleted data
points will be measured again.
4.1.1.5 Device
4.1.1.6.1 Monitoring
This option allows time dependent measurements. More than one configuration may
be measured. All current settings (cable, profile and measurement parameters) will
be used or, alternatively, the settings from different template files. The
measurements will be triggered automatically, depending on the system time, and
saving will be done automatically as well. The following forms give access to all
settings.
4.1.1.6.3 Registration
4.1.1.6.4 Language
The program supports the languages "Deutsch" and "English". Most of the
program captions (boxes, buttons, windows) will immediately switch to the new
language.
4.1.1.7 Help
In case of the profiling mode, a depth section will appear. This is a x- z- slice.
In case of the mapping mode, horizontal x- y- slices are displayed. The (pseudo) depth is
encoded in boxes with different sizes (the smaller, the deeper).
The pseudo section displays the measured data as colored boxes. The boxes are situated in some table
type refering to the horizontal and vertical positions of the respective measurement. These sections are
called "pseudo", because the vertical positions of the boxes are - of course - not the real vertical positions
of the respective data point. For getting quasi real vertical positions, some evaluation of data is necessary
(computer inversion modelling).
Depending on the type of hardware and type of performed measurement, the user can switch between
different displays.
"rho (Ohmm)" displays the measured resistivities in [Ohm*m] = [m]. These result from the in-phase
component of the measured voltages.
"phi (mrad)" displays the information from the out-of-phase component of the measured voltages. Phi in
mrad is defined as = U90 [mV} / U [mV] * 1000. This option is not availble for all hardware types.
"U (mV)" displays the measured voltages (U and U90 or self potentials). This option is only available for
some measurement configurations (SP or gradient) and not available for all hardware types.
If frequency dependent measurements are performed, 2 additional options are available:
"rho (f)" displays information about the frequency effect. The
displayed values are calculated by = rho(f1) / rho(f2).
"phi (f)" displays the out-of-phase frequency effect. The
definition is = phi(f1) - phi(f2).
The user can handle the pseudosections by mouse actions like this:
<left-click>: The data line in the data list (chapter 4.1.3) corresponding to the
respective box is selected. The data is searched from the beginning (num = 1). If
there are overlapping boxes, the 1st (lowermost) box is used.
<double-left-click> or <shift>+<left-click>: The data line in the data list (chapter
4.1.3) corresponding to the respective box is selected. The data list is searched in reverse direction. If there are
overlapping boxes, the last (uppermost) box is used.
<right-click>: the measurement can be re-measured. Search of data list from the beginning.
<shift>+<right-click>: the measurement can be re-measured. Search of data list from the end.
The section displays the measured data as colored boxes. The boxes are situated in some table type
refering to the horizontal positions (X and Y coordinates) of the respective measurement. The size of the
boxes depends on the respective pseudo depth. "Pseudo", because the pseudo depths are not the real
depths.
This example displays the blank section of a grid of Pole-Pole- measurements. Blank means, there is not
yet data which fills the boxes.
As in the pseudo section display of the profiling measurements, the user can choose between drawing of
resistivities [rho (Ohmm)] or phase angles [phi (mrad)]. In case of frequency dependent measurements,
also the frequency functions rho(f) and phi(f) are available (see former chapter).
The data list contains the summary of measured data. This is the same data as available by pressing the
"Data" → Show overview" option of the main menu and the same data to be stored into a file. The
displayed data list displays the results of 1 frequency setting. If there exist measurements of more than
one frequency, the user can switch to another frequency by cklicking the "Frequency settings" option.
Explanation of data:
Please note: there are some different definitions of depths obtained in tomography measurements.
Edwards (1977) introduced the "effective depth" This depth definition is different to the "investigation
depth" definition of Roy and Apparao (1971). ResXDinv inversion software uses the "effective depth"
definition of Edwards (1977).
Clicking the left mouse button, the respective data box in the
pseudosection will blink. This makes it easier to find the corresponding
box in the pseudosection.
Remark: the "Edit data" menu gives access to the data (see chapter 4.1.1.4). Within this menu,
the user can see some statistics for each parameter and the user is able to edit each parameter.
The currently active electrode numbers are displayed. The white fields display the
electrode numbers calculated by the program. The blue "Address" fields give the real
electrode addresses (in case of active electrodes). The actual number and the actual
address need not to be the same. The slider at the bottom of this box gives a sketch
of the electrode postions. A and B are the current emitting electrodes. M and N are
the potential electrodes.
In case of profiling and suitable data sorting: during the
measurement procedure a message will apear just below the
electrode switch box as soon as the first or last used cable (or
electrode chain) could be un-plugged (see chapter 4.1.1.2.2.2.).
4.1.6 Transmitter
The current settings of the transmitter are displayed. The needle indicator shows the
current step (or transmitter voltage, depending on the settings, see chapter
4.1.1.2.3), the text box shows the current in mA. The green symbol shows the
proper function of the transmitter. If there is a problem (e.g. open A-B connection),
the green symbol will disappear and an "open" symbol will appear. If the "show
transmitter voltage" option in the measurement parameter section is checked
(chapter 4.1.1.2.3), the needle indicator will display the transmitter voltage (in V), not
the transmitter current step. In this case, one additional measurement (namely the
transmitter voltage) is necessary.
4.1.7 Receiver
This box shows the current voltage measurement. Only the in-
phase part of the voltage is shown. The green symbol in the middle
shows correct voltage readings. Too low or too high voltages will
result in red symbols.
This box displays the current status of the measurements to be performed. The
extrapolated remaining measurement time is displayed too.
The battery power of the geoelectric device is shown. Depending on the hardware
used, also the external 12 V power and notebook battery capacity may be shown.
This display is updated automatically every 30 seconds.
″Roll-on″ or ″Roll-along″ is the name of the procedure to perform measurements with a total profile length
which is greater than the available cable lengths. Roll-on works using mapping configurations as well. In
the following, the ″Roll-on″ procedure is explained using the profiling configuration as an example:
The following example explains the procedure in more detail using a profiling configuration. Supposed:
The user wants to perform a 695 m long measurement consisting of 1 standard and 2 roll-on procedures
using the Wenner configuration and 5 m electrode separation using 5 electrode chains with 20 electrodes
each. The user starts by choosing main menu > set > cables / electrodes chains (see chapter ″4.1.1.2.1.
Setup cables..″):
This setup shows in total 100 electrodes. The 1 st electrode in use (address 1) is at profile start = 0 m. The
last electrode number 100 (address 100) is situated at the profile end = 495 m. The user fixes this setup
by pressing ″Set″, sets all profile settings (main menu > set > electrodes and data header) and starts the
measurement (main menu > measure > start ...). After finishing the 1 st measurement, the respective
pseudosection of the example looks like this:
The 1st electrode now in use is the electrode number 21 (address 21) at position = 100 m. The last
electrode in use is the number 120 (address 20) at position = 595 m. This setup is used for the next
measurement. After the 2nd measurement, the resulting pseudosection looks like this:
The data is saved. The example procedure is repeated again: roll-on + measure + save. After 2 nd roll-on
procedure (in total 3rd measurement), the total profile reaches now up to electrode number 140 = 695 m.
GeoTest can be run using command line parameters. GeoTest will read the command line parameters
immediately after starting the program. The following command line parameters are possible (see the
following example):
1st parameter (example = "d:\_\test.tx0"): name and folder of the data file to be read.
Automatically read the data file which includes all settings, including electrodes and data header,
measurements parameters, monitoring settings, all series of ABMN settings. The data file may
contain already measured data or may be blank.
2nd parameter (= "monitoring"): automatically switch into the monitoring mode
3rd parameter (= "start"): automatically run the measurement in monitoring mode.
If the user copies the GeoTest shortcut with the respective parameter strings into the Windows "Auto-
Start" folder, the program GeoTest will automatically be started after starting (or re-starting) of Windows.
This may be used for completely automatic and autonomous monitoring measurements without the need
for handling by a user.
Licence agreement
The user of the software accepts the following conditions.
The user is allowed:
- to install and use the software on one single computer.
- to make one copy of the program as a safety copy.
- to give the software to a third person only after a written consent of the producer.
The user is not allowed:
- to give under-licences or to sale or to lend the software or parts of the software to third persons.
- to try to get access to the source code of the program by reverse engineering, decompiling or
other methods.
- to use a former version of the software, in the case of getting an upgrade of the software.
Extended licence agreement
The user has the duty to comply with additional licence agreements, which are valid for the delivered additional
software products, if there are...
Guarantee terms
The Author guarantees, that the software is developped carefully. The software is very complex, therefore the author
cannot guarantee the absolute absence of software bugs or errors.
Liability
There is no liability of the author for any possible consequences of this software.
This manual (version 17.3.2021) refers to the software program "GeoTest" in version 3.00 (march 2021).
6 References
Barker R.D. (1989): Depth of investigation of collinear symmetrical four-electrode arrays. Geophysics, vol. 54, No. 8 (august 1989),
1031-1037.
Berktold & 21 Coautoren (1997): Geoelektrik. In Knödel K., Krummel H. & Lange G. (Herausg.): Geophysik. Handbuch zur
Erkundung des Untergrundes von Deponien und Altlasten, Band 3, 65-367, Springer-Verlag.
Dahlin & Zhou (2006): multiple gradient array measurements for multichannel 2D resistivity imaging. Near Surface Geophysics,
2006, 113-123.
Edwards (1977): A modified pseudosection for resistivity and IP. Geophysics, Vol. 42, No 5, 1020-1036.
Friedel S. (1997): Hochauflösende Geoelektrik - Geoelektrische Tomographie. In: M. Beblo (Herausg.): Umweltgeophysik, 131-151,
Berlin.
Hennig Th., Weller A., Canh T. (2005): The effect of dike geometry on different resistivity configurations. Journal of Applied
Geophysics 57, 278-292.
Roy and Apparao (1971): Depth of investigation in direct current methods. Geophysics, Vol. 36, No. 5, P. 943-959
Trimble (1999): AgGPS(TM) 124/132 operation manual. Trimble Navigations Ltd. www.trimble.com.
Zhou et al. (2020): A full-range gradient survey for 2D electrical resistivity tomography. Near Surface Geophysics, 2020, 18, 609-
626.