VegaLux Manual A4 Preview
VegaLux Manual A4 Preview
VegaLux Manual A4 Preview
USER
MANUAL
2 Reveal Lasers - User Manual
Reveal Lasers Ltd
reveallasers.com
All rights reserved. Contents of this publication may not be reproduced in any form
without the written permission of Reveal Ltd.
1.2 Maintenance 6
System Overview
1.1 Use of this Manual
The VEGA LUX™ system is designed to meet international safety and performance standards.
Personnel operating the system must have a thorough understanding of the proper operation of
the system. This manual has been prepared to aid medical and technical personnel to understand
and operate the system.
Do not operate the system before reading this manual and gaining a clear understanding of system
operation. If any part of this manual is not clear, please contact your representative for clarification.
The information provided in this manual is not intended to replace professional training on the
clinical use of the system. Please contact your representative for current information on available
training. For clinical information, refer to the clinical guides in this manual, which includes set
up guidelines for each application. This manual should always accompany the system and all
operating personnel must know its location.
Warning:
For further information about REVEAL™v, isit the company website at: www.reveallasers.com
Warning:
A Ampere(s) mm Millimeter(s)
Safety
2.1 Introduction
This chapter describes general safety issues regarding the use of the VEGA LUX™ system, with
special emphasis on optical and electrical safety.
With proper operation and maintenance, trained and qualified medical practitioners can use
the VEGA LUX™ system safely. The supervising physician and all other personnel operating or
maintaining the VEGA LUX™ must be familiar with the safety information provided in this chapter.
The primary consideration should be for the safety of the patient, the physician and other
personnel. Patient safety is mainly assured with a well-trained staff and a well laid out treatment
room. Patient education is also important, including information about the nature of the treatment.
1. semi-automatic circuit breaker, located on the service panel, protects the system by
A
tripping when power overload occurs. To resume normal operation, lift the circuit breaker
handle and restart the system.
2. here are fuses in the machine to protect the system from any electrical spikes.
T
A
watchdog cycle continuously monitors operation of the system during treatment.
If an error occurs, the system displays a warning message to the operator and disables
further operation. A self-test of the attached module is performed when the system is
turned on. The test includes module identification.
A
self-test of the electrical circuitry takes places after the system is turned on. The test
circuits continuously monitor system operation during treatment.
1. Closed light-guide geometry is used to transmit Light/Laser energy to the treatment site.
Light/ Laser energy is emitted only through the front plane of the light-guide.
2. An emergency shutoff knob expedites shutdown when necessary. When pressed, it
immediately shuts down system operation.
3. The system can be turned on only when the key is inserted in the key-switch. When not in
operation, the key should be removed to prevent unauthorized use of the system.
4. password on the service screen prevents unauthorized changes to the system's basic
A
operating parameters
5. The system features one emission indicators: a red emission indicator lamp located on the
handpiece and a buzzer.
6. ight emission is enabled only when the operator switches to Ready mode and presses
L
the footswitch (minimum risk).
7. Laser emission is enabled only if both the foot-switch and the module's trigger are pressed
at the same time, reducing the risk of unintentional Light/Laser emission.
8. ater is circulated through the module as soon as the system is turned on in order to cool
W
the light source.
9. The flow and temperature of the water are monitored in order to eliminate the risk of
module overheating. Light/Laser emission is not permitted if the water flow stops or if the
water temperature is equal to or higher than 40°C (104°F).
10. The system is equipped with an electric foot-switch for the ease of use.
Warning:
A
ny Light/Laser emitting device can cause injury if used improperly. High voltages are
present inside the VEGA LUX™ system. Personnel who work with lasers must always be
aware of the possible dangers and must take the proper safeguards as described in this
manual.
U
se carefully may cause serious burns. Do not use over sensitive skin areas or in the
presence of poor circulation. The unattended use of the VEGA LUX™ system by children
or incapacitated persons may be dangerous.
Caution:
Laser fume and/or plume may contain viable tissue particulates.
2.3.1 Precautions
Observe the following precautions when using the VEGA LUX™ system.
1. Physicians and clinicians should read this manual thoroughly before attempting to operate
the VEGA LUX™ system.
2. The module's light-guide must be kept clean at all times. Remember to clean the cooling
gel from the light-guide after each patient.
3. The system weighs approximately 154 Lbs (70 Kg) and may cause injury if proper care is
not used when moving it. The system is well balanced and is designed to be moved, but
should always be moved carefully and slowly. Never pull the system by the module or
its umbilical cable.
2.3.2 Cautions
Observe the following cautions when using the VEGA LUX™ system.
1. Only REVEAL™a uthorized personnel may service the VEGA LUX™s ystem. This includes
making internal adjustments to the power supply, cooling system, optics, modules, etc.
2. Verify that the VEGA LUX™ is wired for the appropriate electrical voltage of your country
(230V AC).
3. Maintenance performed by the operator must only take place when the system is
shut down and disconnected from the electrical power source. Performing maintenance
procedures with the system powered-up can be hazardous to the operator and/or
destructive to the system.
8. he module and tips must always be visually inspected before treating the patient. If any
T
wear or damage is apparent, do not use the module or tip.
9. Never leave the system turned on, open or unattended during system maintenance.
The treatment room must be clearly labelled with a sign indicating that high intensity light/laser
energy is in use. The treatment room sign that is supplied with the VEGA LUX™ system.
The treatment room should not include any light reflecting objects such as a mirror.
Access to the VEGA LUX™ treatment room should be allowed only to personnel essential to the
procedure and who are well trained in the required safety procedures.
Assure that all of the treatment room personnel are familiar with the VEGA LUX™ controls and
know how to shut down the system instantly.
Energy Emission
2. It is good practice to instruct the patient to close their eyes during treatment even when
wearing protective eyeglasses.
3. If the patient cannot wear the protective eyewear, fit the patient with opaque eye
protection that completely blocks light from the eyes.
4. If the treatment area is very close to the eyes (e.g. eyelids), protect the eyes with corneal
shields.
2. The safety eyewear and opaque eye protectors supplied with the system offers adequate
protection, and more can be ordered from your representative.
2. The system is not suitable for use in the presence of flammable mixtures with air or
oxygen.
3. o not operate in the presence of volatile solvents such as alcohol, gasoline or other
D
solvents.
4. Do not use any flammable substances such as alcohol or acetone in the preparation of
the skin for treatment. If necessary, use soap and water to clean before treatment.
5. If alcohol is used to clean and disinfect any part of the VEGA LUX™ system, allow it to dry
thoroughly before operating the system.
The VEGA LUX™ system produces very high voltages in various components. Some components
may retain a charge after the power supply has been turned off, so no part of the exterior housing
should be removed, except REVEAL™a uthorized personnel.
Whenever system maintenance is performed, never leave the VEGA LUX™ system turned on,
open or unattended.
Warning:
To avoid the risk of electric shock, this equipment must only be connected to a supply main with
protective earth.
To release the emergency shut-off knob, turn it clockwise. Otherwise, the system will remain off.
Caution:
Use the emergency shut-off knob only in the event of an emergency.
2. Intermittently during light/ laser emission, at the same rate as the pulse repetition rate.
2.6.5 Foot-switch
The system is equipped with electric foot-switch for the ease of use.
1. ince the Light/Laser energy is generated in the module itself and not in the console (as in
S
conventional lasers), there is no need for an articulated arm or other beam delivery system with
inherent beam quality and alignment concerns. Instead, the VEGA LUX™ modules incorporate
a onepiece light mixer that combines the emission from thousands of emitters to produce a
uniform square beam. Since light/laser emission is confined to the module, there is no hazardous
optical radiation in the console or the umbilical cable.
2. he light-guide is placed against the patient's skin during the system use, reducing stray light/
T
laser energy while increasing the therapeutic effect.
2.6.8 Compliance
In compliance with these standards, the system is equipped with:
Proper labeling
3.1 Introduction 17
Installation
3.1 Introduction
The VEGA LUX™ system is designed for installation in an office or a clinic and requires minimal site
preparation. When the VEGA LUX™ system is purchased, complete on-site installation, including
initial system testing and calibration, is included.
System transportation and installation is carried out by authorized technical personnel, who will
do the following:
Caution:
In order to guarantee safe disconnection of the system in any hazardous situation, system shall
be allocated in such a way that appliance coupler always accessible.
Caution:
The system should be connected to a separate power line with separate circuit breakers. REVEAL™
cannot guarantee adequate performance unless the system is connected to a dedicated circuit.
Caution:
In order to guarantee safe disconnection of the system in any hazardous situation, system shall
be allocated in such a way that appliance coupler always accessible.
Caution:
The system should be connected to a separate power line with separate circuit breakers. REVEAL™
cannot guarantee adequate performance unless the system is connected to a dedicated circuit.
Water Quality:
the system should be operated using de-ionized water only. Regular tap water contains sediments
that may damage the cooling system.
Temperature:
to ensure that the system performs optimally, maintain room temperature between 17°C and 25°C
(68°F - 77°F) and relative humidity of less than 80%.
Note: When the system is used intensively, it emits heat. Therefore, it is recommended to install
air conditioning in the room in which the system will be used
and Transportation
Transportation of the VEGA LUX™ system should be under normal environmental conditions
within the following range:
Note: If the system is to be stored in a cold area where the temperature may fall below 0°C (32°F),
the deionized water must be drained from the cooling system. Only authorized service personnel
may perform this procedure.
The VEGA LUX™ cooling system's water reservoir must be filled (or refilled) with de-ionized water
in the following instances:
1. water filling tube, with a funnel connected to one side and a quick-connector
A
attached to the other side.
1. Connect the small connector to the left port marked VENT on the lower right side of the
service panel
2. Connect the clear tube with the funnel to the port marked FILL on the upper side of the
service Panel
3. se the supplied 5 liter (1.3 gal.) container of de-ionized water and pour it slowly into
U
the funnel. Align the clear tubing with the service panel. Stop filling as soon as the water
starts to spill from the VENT port.
1. Insert the module connector into the connection port (Fig. 22), as far as it will go in.
3. After you turn the latch (Fig. 24), the connector will continue to go into the port until you
can turn it no more - now the module is properly connected to the system.
Caution:
Never use tools to tighten the module connector - finger-tight pressure is sufficient.
Caution:
Never use the module or the umbilical cable to move the system.
Do not use the module mast for moving or lifting the system.
4.1 Introduction 25
System Description
4.1 Introduction
This chapter provides a detailed description of the VEGA LUX™ system. The description covers
the system's main components, controls and functional sub- systems and system & module
specifications.
Diode laser technology: 2D (755 nm + 810 nm combined in 2 cm2spot size module) + Radiofrequency
(2Mhz). The 2D+RF, 755nm, 810nm Diode Laser and Radiofrequency energy handpieces are
intended for the removal of unwanted hair and to effect stable, long-term hair reduction. The
complete system consists of its console, the various available modules and a footswitch. The
module is pressed against the patient's skin and a light pulse is delivered when the module's
trigger and the foot-switch are activated. The module's tip is cooled by the VEGA LUX™ cooling
system. Output parameters and other system features are controlled from the touch-screen
control panel on the console, which provides an interface to the system's micro-controller
through & LCD touch-screen
2. Modules (with umbilical cable and connector) that incorporates the optical head, tissue
cooling system (the cold plate) and the trigger.
3. oot-switch
F
1. Control panel
2. Power supply modules
3. System controller
4. ooling system
C
5. Isolation transformer
1. Control Panel - this touch-controlled screen provides information on the status and
settings of the VEGA LUX™ system. The various screens of the display are described
in detail
2. Emergency Shut-off Knob - this is a red, mushroom-like knob designed for emergency
shutdown of the system.
1. System Software - The VEGA LUX™ system incorporates embedded software. The
software supports the graphic user interface, controls the system operating parameters
and runs routine tests to ensure proper operation.
2. Error Detection - This VEGA LUX™ system is equipped with self-testing software that
continuously monitors system operation, using the watchdog feature. The software
continuously checks the hardware status, and if an error condition is detected:
A
n error message is displayed
T
he audible alarm signal is activated
2. Integrated contact tissue cooling system - The module light-guides are cooled by
a thermoelectric cooling method to reduce patient discomfort during treatment & to
reduce post-procedure side effects, such as local skin redness & swelling. The tissue is
cooled through a metallic ring & a cold sapphire window. The light-guide's temperature
is reduced to 4°C (39°F) during system operation. Thermistor monitors & keeps the
temperature at this level to prevent overcooling of the treated area. The tissue cooling
system is designed to cease operation in case of a pause in operation lasting for more
than 5 minutes. This feature is included to prevent undesired condensation of water
drops on the tip.
3. Fans - Five fans are located inside the system. Four fans cool the water system and the
other fan cools the interior space of the console.
1. Stamping Mode
2. Smooth Mode™
Laser energy is delivered through a 2cm2 tip built into the module. The modules provide a
preselected pulse duration as a function of the selected pulse type and fluence. The VEGA
LUX™ diode laser modules are intended for hair removal and long-term hair reduction. It is
indicated for use on all skin types (Fitzpatrick Skin Types I-VI).
The optical bench and the tissue cooling system are located inside the ergonomically designed
module. The diode laser module is divided into 3 parts: an umbilical cable, a connector and the
module itself. Power and control signals are exchanged with the console via the umbilical cable.
The length of the cable is 170 cm (67"), which enables free access to any body part. The diode
laser module is shown It incorporates the laser diode array with the light-guide, the protective
sapphire window, the cold plate, and the module's trigger.
Caution:
T
he module contains delicate optical components which may suffer severe damage if
dropped. Except during treatment, the module should be always kept in its cradle.
W
hen moving the system, the module should be disconnected from the system and
stored in its carrying case.
RF Power 200W
UV Power 5W
Stamping: Up to 60
Energy Density
Smooth Mode: Up to 20
Stamping: 0.5-3Hz
Frequency
Smooth Mode: 3, 5, 10Hz
Language English
Net Weight
5.1 Introduction 31
5.2.5 Key-switch 32
5.3.5 Foot-switch 33
Controls and
Indicators
5.1 Introduction
This chapter details the controls, indicators and connection ports of the VEGA LUX™ system.
operating menu after the system is activated. To switch the system into operational mode,
Warning:
To avoid misuse of the system, do not leave the key in the key-switch while the system is
unattended.
Warning:
If the system is in Ready mode, laser beam emission occurs when the module trigger and
the foot-switch are activated. This requirement for simultaneous activation of both triggers
minimizes the risk of unintentional laser beam emission.
5.3.5 Foot-switch
The foot-switch is enabled only when the system is in Ready mode. Activation of the footswitch
in Standby mode does not result in light/laser emission.
The foot-switch is an electric pedal that connects to the system through the foot-switch
connection port on the service panel.
Warning:
If the system is in Ready mode, light/laser energy emission from the modules occurs when the
footswitch or foot-switch and module's trigger are activated.
6.1 Introduction 35
6.4.1 Key-switch 38
Controls and
Indicators
6.1 Introduction
This chapter describes in detail the operating instructions for the VEGA LUX™ system.
Caution:
T
his system may only be operated by a licensed practitioner, according to the local laws
in every country.
A
light/laser radiation danger sign, supplied with this system, should be placed at the
entrance to the treatment room whenever the system is in use.
Improper use or adjustment of this system may invalidate the VEGA LUX™ service warranty
agreement. Please contact your authorized distributor before attempting to use the system
in any manner other than those specified in this manual.
1. Turn on the system. If you want to sterilize the Hand piece tip, click sterilize on the home
screen. The UV light turns ON for 30 seconds and the tip gets sterilized
4. Select the Gender "Male / Female" and the area to be treated. Toggle for the area to
be treated in Front/Back/ Face. The selected area will be highlighted on the screen.
5. Select the Skin Type in the "Manual" segment between "I to VI". (Auto mode is disabled
in this version) and click continue.
T
he VEGA LUX™ system emits intense light/laser pulses through the module. Make
sure that all personnel are protected against accidental exposure to these pulses, either
directly from the module or indirectly from a reflecting surface.
T
o protect against eye damage and discomfort, make sure that everyone present in the
room is wearing recommended protective eyewear.
N
ever look directly at the pulse coming from the module, even when wearing appropriate
protective eyewear.
N
ever point the module so that it discharges into free space. Make sure that the module
is pointed at the treatment site during actual treatment.
T
he VEGA LUX™ system emits intense light/laser pulses through the module. Make
sure that all personnel are protected against accidental exposure to these pulses, either
directly from the module or indirectly from a reflecting surface.
T
o protect against eye damage and discomfort, make sure that everyone present in the
room is wearing recommended protective eyewear.
N
ever look directly at the pulse coming from the module, even when wearing appropriate
protective eyewear.
N
ever point the module so that it discharges into free space. Make sure that the module
is pointed at the treatment site during actual treatment.
Energy: Set the total invested energy parameters, click on the energy
window. To increase or decrease the parameter use + or - arrow keys
on the right of the energy window. The exhibited value expresses the
total energy applied per session counted in Kilo Joule(KJ)
Total Shots: The field counts the total number of laser pulses emitted by the Hand piece and it
cannot be reset. The field will reset to zero only when a new Handpiece is connected.
Grids: To increase or decrease the no of grids use + or - arrow keys on the screen. The total
number of grids depend upon the area to be treated.
For example, if you are treating lower hands and you select the 100cm2 grid, there would be
around 5 to 6 grids to be treated.
Grid Size: The Grid needs to be marked on the treatment area with the stencil. Select the Grid
size 50 / 100 / 150 / 300 cm2 on the screen with respect to the treatment area marked on the
skin.
RF percentage: The percentage of RF Frequency can be adjusted from this setting, use the +
or - arrow to adjust the Frequency from 0 to 100%. The RF can be disabled by clicking on the
RED button and enabled by clicking the ON button.
Switch mode: Click on Switch mode at the bottom corner of the screen for the
Frequency: Frequency can be adjusted from 0.5 to 3 Hz, by clicking on + and -. Shot count: This
field displays the total number of laser pulses emitted by the module and cannot be reset.
Pulse duration: The pulse duration is to be adjusted between short, Medium and Long.
Standby/Ready mode:
This soft key toggles the VEGA Lux between standby and ready modes. Press the Ready button
when you are ready to start the treatment. Once the treatment is done toggle the machine back
to standby mode.
5. o stop light/ laser emission, release the foot-switch or the foot-switch and the module's
T
trigger.
Warning:
To prevent unauthorized use of the system, do not leave the key in the key-switch unattended.
Treatment Guidelines
7.1 Safety checks
1. Check All Equipment and Laser.
6. A test patch should be performed on each of the areas the patient is interested in.
7. he patient should come with some hair for test patch, so the hair can be assessed
T
for color and thickness
10. Use three different fluence levels starting with the lowest.
11. Choose the highest fluence that gives the best efficacy with no adverse side effects.
Ask the patient to hide lips/ protect vermilion border. You can use a wooden spatula to
block off each nostril as you work close to it. If using Air cooler to cool skin lower the level
and supplement with ice packs. You can ask the patient to hold the nose or take deep
breaths after each shot.
2. Ears
Place dry cotton wool down ear canal. Place hand piece so that the light fires away from
the ear canal.
• If the patient experiences excessive pain, stop treatment, cool the skin
Excessive Pain and moisturize with Aloe Vera gel.
• Review and then if all is OK restart treatment.
10. Not cleaning laser consumables, glass pieces, distant gauges, inserts etc.
7. Massage areas during / after treatment. This can really help with pain on areas such as
upper lip.
2. A topical soothing skin care product such as Aloe Vera gel may be applied following
treatment. If the area still feels hot you can apply a cold flannel or cool pack at home.
3. If your skin is particularly sensitive & prone to histamine reactions, you may take an
antihistamine medication.
4. Makeup can be used after the treatment if the skin is not irritated.
6. DO NOT depilate between laser treatments with waxing, plucking, threading or creams
that will disturb the hair follicle. Shaving is the preferred method.
7. Anytime from 10-21 days after the treatment, shedding of the treated hair may occur &
this appears as new hair growth.
8. This is NOT new hair growth. You can clean & remove the hair by washing or wiping the
area with a wet cloth or loofah sponge.
10. There are no restrictions on bathing except to treat the skin gently, as if you had
sunburn, for the first 24 hours.
12. It is extremely important to avoid sun exposure before, during and after treatment.
13. After Laser your skin is more sensitive, and you are more likely to burn and therefore
get hyper pigmentation or hypo pigmentation issues.
14. Laser treatment increases light sensitivity; all the areas treated by a laser must be
covered with plenty of sunscreen because the skin is sensitive to darkening.
15. Recently tanned skin cannot be treated with a laser because it increases the risk of
developing hypo pigmentation.
16. We recommend using sunscreen protection of at least SPF 30 all day
17. Irritated skin may be more susceptible to injury. Recently tanned skin cannot be
treated as you may develop hypo- pigmentation (loss of skin color).
18. You may be asked to stop any topical medications 1-2 weeks before treatment.
19. You MUST avoid waxing, threading, plucking, epilating or bleaching the hair 4 weeks
prior to treatment.
20. The melanin containing hair must be present in the follicle as it is the 'target' for the
laser light.
21. You will need to prepare the area to be treated in advance of treatment.
22. The area will need to be shaved & kept clean & free from any cosmetic products on
the day of the treatment.
24. A medical history is taken at the consultation, but we rely on you to inform us of any
changes to this.
25. We need to know of any medication you are taking including those bought from a
chemist or herbalist/homeopath (inter sessions as well).
2. Beauty products containing Glycolic Acid, Benzoyl Peroxide, Bleach, Salicylic Acid &
bleach: These need to be stopped before treatment, they contain strong chemical &
make the skin more sensitive.
3. Glycolic Acid Peel: Wait 1-2 weeks before Laser hair removal. If 70 % wait longer.
8. Cut, broken, irritated or infected skin: The appointment will need to be rescheduled as
you cannot treat over these areas.
Communicate with the client and gain an understanding of what they want from the
treatment.
Good questions to ask would be: What is it that concerns you most? How do you remove
hair at present? How often do you need to? How quickly does your hair grow back?
The hair needs to be dark and there needs to be enough contrast between hair and
skin color for the Laser to work.
You need to ensure you have the correct equipment to treat the patient's skin type.
Explain to the client in simple words how the laser works - e.g. -the light from laser
heats the hair follicle and destroys it completely.
The treatment can be uncomfortable and should be described as such to the client.
Client may feel mildly uncomfortable to extremely uncomfortable depending on their
pain threshold, the thickness of the hair and the area being treated. Cooling before and
after can reduce the pain.
2. Results are not immediately visible. After laser treatment some hair will be vaporized
immediately, and some will take 3-7 days to shed.
4. Results to expect: Differs for different hair and skin types and body areas being treated.
100% clearance cannot be guaranteed.
All of these conditions are temporary however in some cases they can be permanent.
2. It is a good idea to take picture of the area before session to catalogue the progress
3. As treatments progress clients may forget how bad the problem was.
2. Patient consent form must be signed prior to the treatment, and every session if any
changes to body parts or extending areas.
3. Which explains any possible side effects, and confirmation that the patient has been
given an explanation of the procedure and expected outcome.
USER MANUAL