Aficheur Newport Idp

Download as pdf or txt
Download as pdf or txt
You are on page 1of 35

IDP

INFINITY® D Process Meter

Operator’s Manual

®
NEWPORT Electronics, Inc.
http://www.newportUS.com/manuals
Additional products from

®
NEWPORT Electronics, Inc.
Counters Rate Meters
Frequency Meters Timers
PID Controllers Totalizers
Clock/Timers Strain Gauge
Printers Meters
Process Meters Voltmeters
On/Off Multimeters
Controllers Soldering Iron
Recorders Testers
Relative pH pens
Humidity pH Controllers
Transmitters pH Electrodes
Thermocouples RTDs
Thermistors Thermowells
Wire Flow Sensors
For Immediate Assistance
®
In the U.S.A. and Canada: 1-800-NEWPORT
SM
In Mexico: (95) 800-NEWPORT
Or call your local NEWPORT Office.
NEWPORTnet On-Line Service
SM
Internet e-mail
http://www.newportUS.com [email protected]

It is the policy of NEWPORT to comply with all worldwide safety and EMC/EMI regulations that apply.
NEWPORT is constantly pursuing certification of its products to the European New Approach Directives.
NEWPORT will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but NEWPORT Electronics, Inc. accepts no liability for
any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
PATENT NOTICE: The “Meter Case Bezel Design” is a trademark of NEWPORT Electronics, Inc., registered in the U.S..
This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577 /
CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 /
UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. OTHER INTERNATIONAL PATENTS PENDING.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or
commissioning this device as it contains important information relating to safety and EMC.
PREFACE

MANUAL OBJECTIVES

This manual shows you how to set up and use the Programmable Process meter.

NOTES, WARNINGS and CAUTIONS

Information that is especially important to note is identified by three labels:

• NOTE
• WARNING
• CAUTION
• IMPORTANT

NOTE: provides you with information that is important to successfully


setup and use of the Process meter.

CAUTION or WARNING: tells you about the risk of electric shock.

CAUTION, WARNING or IMPORTANT: tells you of circumstances


or practices that can effect the meter's functionality and must refer to
accompanying documents.

i
Table of Contents
Section Page

SEC 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

SEC 2 ABOUT THE METER. . . . . . . . . . . . . . . . . . . . . . . . . . 3


2.1 Front of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Back of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

SEC 3 GETTING STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . 7


3.1 Changing Configuration Jumpers . . . . . . . . . . . . . . . . . . . 7
3.2 Mounting the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Connecting Sensor Input . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Connecting Main Power. . . . . . . . . . . . . . . . . . . . . . . . . 12

SEC 4 CONFIGURING THE METER . . . . . . . . . . . . . . . . . 13


4.1 Selecting the Input Range . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Selecting a Decimal Point Position. . . . . . . . . . . . . . . . . 14
4.3 Scaling the Meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.3.1 Internal Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3.2 Live Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4 Changing the Meter’s Calibration. . . . . . . . . . . . . . . . . . 17
APPENDICES
A Checking and Changing Main Board Power Jumpers . . . 19
B Reference Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
C Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
D CE Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

ii
List of Figures
Figures Page

2-1 Front-Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2-2 Connectors (ac-Powered and dc-Powered) . . . . . . . . . . . . . . .5
2-3 dc-Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
3-1 S1 - S4 Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
3-2 Meter - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
3-3 Panel Cut-Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3-4 3-Wire dc Voltage Input Connections
with Internal Excitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3-5 2-Wire dc Voltage Input Connections
with External Excitation . . . . . . . . . . . . . . . . . . . . . . . . . . .10
3-6 4-20 mA Input Connections with External Excitation . . . . . .10
3-7 4-20 mA Transmitter Hook-Up with Internal Excitation . . . .11
3-8 dc Current Input Connections with Current Source . . . . . . .11
3-9 Main Power Connections ac . . . . . . . . . . . . . . . . . . . . . . . . 12
3-10 Main Power Connections dc . . . . . . . . . . . . . . . . . . . . . . . .12
A-1 115Vac Jumper Wiring (default) . . . . . . . . . . . . . . . . . . . . . .19
A-2 230 Vac Jumper Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
C-1 Meter Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

List of Tables
Table Page

2-1 Connector Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5


3-1 S1 - S4 DIP Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3-2 ac Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4-1 Meter Calibration Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
B-1 Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
B-2 Menu Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . 21
B-3 Run Mode Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
B-4 Factory Preset Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

iii
NOTES

iv
SECTION 1. INTRODUCTION

1.1 UNPACKING

Remove the Packing List and verify that you have received all equipment. If
there are any questions about the shipment, use the phone numbers listed on the
back cover to contact the Customer Service Department nearest you.

Upon receipt of shipment, inspect the container and equipment for any signs of
damage. Note any evidence of rough handling in transit. Immediately report any
damage to the shipping agent.

The carrier will not honor any claims unless all shipping material is
saved for their examination. After examining and removing contents,
save packing material and carton in the event reshipment is necessary.

Verify that you receive the following items in the shipping box:

QTY DESCRIPTION

1 Programmable indicator/controller with all applicable connectors


attached.

1 Owner’s Manual

1 Quick Start Manual

If you ordered any of the available options (except the “BL” blank Lens
option), they will be shipped in a separate container to avoid any
damage to your indicator/controler.

1
1.2 SAFETY CONSIDERATIONS

This device is marked with the international caution symbol. It is important to


read this manual before installing or commissioning this device as it contains
important information relating to Safety and EMC (Electromagnetic Compatibility).

This instrument is a panel mount device protected in accordance with EN 61010-


1:2001, electrical safety requirements for electrical equipment for measurement, con-
trol and laboratory. Installation of this instrument should be done by qualified person-
nel. In order to ensure safe operation, the following instructions should be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall


be included in the building installation as a disconnecting device. It shall be marked to
indicate this function, and it shall be in close proximity to the equipment within easy
reach of the operator. The switch or circuit-breaker shall not interrupt the Protective
Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and
IEC 947-3 (International Electrotechnical Commission). The switch shall not be incor-
porated in the main supply cord.

Furthermore, to provide protection against excessive energy being drawn from the
main supply in case of a fault in the equipment, an overcurrent protection device shall
be installed.

• Do not exceed voltage rating on the label located on the top of the instrument
Note ☞ housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
• Unit mounting should allow for adequate ventilation to ensure instrument does not
exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power
requirements. Install without exposing bare wire outside the connector to minimize
electrical shock hazards.

EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems
persist.

Failure to follow all instructions and warnings may result in injury!

2
SECTION 2. ABOUT THE METER

2.1 FRONT OF THE METER

Figure 2-1 shows each part of the front of the meter.

Figure 2-1. Front Panel

METER DISPLAY:

Digital LED display: ¡.99. 9


. . or 9.99
. 9
. . 4-digit 7–segment, 0.56" high LED
display with programmable decimal point.

METER MODES:

Run Mode The meter is in the run mode when the display is actively
showing a process.

Configuration Mode The meter is in the configuration mode when you press the
MENU button to enable meter configurations.

3
METER BUTTONS

T-RST button - Clears the tare value.

£/MAX button - In the run mode, press the £/MAX button to show the
maximum value if jumper S3 is installed: the meter momentarily shows
“####”, then flashes the maximum value encountered since the last peak
reset. Press the £/MAX button to show the VALLEY value if jumper S3 is
removed: the meter momentarily shows “™™™™”, then flashes the minimum
value encountered since the last valley reset. In the configuration mode, press
the £/MAX button to change the flashing digit’s value (increments by one).

£/TARE button - In the run mode, press the £/TARE button to tare the
display value to zero. In the configuration mode, press the £/TARE button to
scroll through available choices, or to move one digit to the right.

MENU button - In the run mode, press the MENU button to enable meter
configuration changes. Refer to specific procedure for more details.

RESET button - In the run mode, press the RESET button to reset the Peak
and Valley registers: the meter momentarily shows “PrSt” or “VrSt” , then
resets both values. In the configuration mode, press the RESET button twice
quickly to escape from the configuration mode.

4
2.2 BACK OF THE METER

Figure 2-2 shows the connectors on the back of the meter. Table 2-1 gives a
brief description of each connector at the back of the meter. Refer to Figure 2-3
for dc-power connection information.

Figure 2-2. Connectors for ac-Power (Top) and dc-Power (Bottom)

Table 2-1. Connector Description


Connector Description

J1-1 ac line connection (No Connection on dc-powered units).


J1-2 ac neutral connection (+ input on dc-powered units).
J1-3 ac earth ground (dc-power return on dc-powered units).
J2-1 Remote tare connection with momentary switch to J2-2.
J2-2 Digital Return (RTN).
J2-3 Tare reset with momentary switch to J2-2.
J3-1 Hold input (active low) - Connect to J2-2 to hold meter reading
(does not update Min/Max).
J3-2 +EXC: Positive excitation connection from meter (10, 24 Vdc).
J3-3 -EXC: Negative excitation connection from meter (10, 24 Vdc).
J4-1 Analog Input return (RTN).
J4-2 Analog Current input (4-20 mA).
J4-3 Analog Voltage input (0-5 V, 0-10 V etc.).

5
2.2 BACK OF THE METER (Continued)

Figure 2-3. dc-Power Connections

2.3 DISASSEMBLY

You may need to open up the meter for one of the following reasons:

* To check or change the 115 or 230 Vac power jumpers.


* To install or remove calibration jumper on the main board.

Disconnect the power supply before proceedings.

To remove and access the main board, follow these steps:

1. Disconnect the main power from the meter.


2. Remove the front lens.
3. Pull the meter forward, out of its case.

6
SECTION 3. GETTING STARTED

CAUTION: The meter has no power-on switch, so it will be in


operation as soon you apply power.

If you power off/on the meter momentarily shows the following: “idP” for
model type, “codE” , and “rxx” for the micro controller revision code.

Keep track of revision code for future reference.

3.1 CHANGING CONFIGURATION JUMPERS

The S1, S2 and S3 DIP switches are at the front of the meter near the digital
display. The S4 DIP switch is on side of the main board (refer to
Figure 3–3). Remove the lens cover to access these jumpers.

Table 3-1. S1 - S4 DIP Switches (Defaults in bold and italics)

Jumper Description
S1 Installed: 10 volt excitation
Removed: 24 volt excitation
S2 Installed: Front-panel buttons locked out
Removed: All buttons operable
S3 Installed: PEAK shows when £/MAX button is pushed
Removed: VALLEY shows when £/MAX button is
pushed
S4 Installed: Calibration enabled
Removed: Calibration disabled

7
3.1 CHANGING CONFIGURATION JUMPERS (Continued)

Front of the meter

Figure 3-1. S1 - S4 Jumpers

3.2 MOUNTING THE METER

PANEL CUTOUT

Figure 3-2. Meter - Exploded View

8
3.2 MOUNTING THE METER (Continued)
PANEL THICKNESS
1. Cut a hole in your panel according to 6,4 (.25) MAX
0,8 (.03) MIN
the dimensions specified in Figure 3-3.
1,5
R(.06) 45,00 + 0,61/-0,00
2. Insert the meter into the hole. Be sure 4 PLCS (1.772 + .024/–.000)

the front bezel is flush to the panel. 92,00 + 0,81/–0,00


(3.622 + .032/–.000)

NOTE: Dimensions in Millimeters (Inches)

Figure 3-3. Panel Cut-Out

3. Proceed to Section 3.4 to connect your sensor input.

3.3 CONNECTING SENSOR INPUT

Figures 3-4 through 3-8 describe how to connect common sensor types.

1
2 J2
3

1 HOLD
2 EXC J3
3 EXC
EXC

COM
1 RTN
SIG
0 10Vdc 2 I J4
3 V

Figure 3-4. 3-Wire dc Voltage Input Connection


with Internal Excitation

9
3.3 CONNECTING SENSOR INPUT (Continued)

1
2 J2
3

1
2 J3
3

1 RTN
2 I J4
0 10Vdc
3 V
SIG SIG
LOW HI
SOURCE

Figure 3-5. 2-Wire dc Voltage Input Connections


with External Excitation

1
2 J2
3

1
2 J3
3
EXTERNAL SUPPLY
1 RTN
2 I J4
3 V

Figure 3-6. 4-20 mA Input Connections


with External Excitation
10
3.3 CONNECTING SENSOR INPUT (Continued)

1
2 J2
3

1 HOLD
2 EXC J3
3 EXC

1 RTN
2 I J4
3 V

Figure 3-7. 4-20 mA Transmitter Hook-Up


with Internal Excitation

1
2 J2
3

1
2 J3
3

1 RTN
2 I J4
3 V

Figure 3-8. dc Current Input Connections with Current Source

11
3.4 CONNECTING MAIN POWER

Connect the ac main power connections as shown in Figure 3-9.

WARNING: Do not connect ac power to your meter until you have complet-
ed all input and output connections. Failure to do so may result in injury!

J1

1 2 3

AC Power Earth
Line Neutral Ground

FUSE

Figure 3-9. Main Power Connections - ac

Table 3-2 shows the wire color and respective terminal connections for both
USA and Europe.
Table 3-2. ac-Power Connections
WIRE COLORS
J1 AC POWER EUROPE USA
1 ~ ac Line Brown Black
2 ~ ac Neutral Blue White
3 ~ ac Earth Ground Green/Yellow Green
Connect the dc main power connections as shown in Figure 3-10.

J1

1 2 3

+ DC - DC
DC POWER

Figure 3-10. Main Power Connections - DC


12
SECTION 4. CONFIGURING THE METER

4.1 SELECTING THE INPUT RANGE (INPT)

Refer to Table B-2 for a summary list of menu configuration.

To select your appropriate input range, follow these steps:

1. Press the MENU button. The meter flashes “InP”.

2. Press the ¢/TARE button. The meter flashes one of the following:

* 0 - 10 (for 0-10 volt)


* 4 - 20 (for 4-20 mA)
* 20 - 4 (for 20-4 mA)
* 0 - 5 (for 0-5 volt)
* 1 - 5 (for 1-5 volt)

3. Press the ¢/TARE button to scroll through available choices.

4. Press the MENU button to store your choice. The meter momentarily
shows “StOr”, followed by “dEc.P” (Decimal point).

13
4.2 SELECTING A DECIMAL POINT POSITION (DEC.P)

Refer to Table B-2 for a summary list of menu configuration.

To select a decimal point display position, follow these steps:

1. Press the MENU button until the meter shows “dEc.P”.

2. Press the ¢/TARE button. The meter shows one of the following:

* FFF.F
* FFFF.
* FFFF
* F.FFF
* FF.FF

3. Press the ¢/TARE button to move the decimal point to the right one
position.

4. Press the MENU button to store your choice. The meter momentarily
shows “Stor”, followed by “ScAL” (Scaling).

4.3 SCALING THE METER

There are two scaling methods. One method is internal scaling, “int” (scaling
without known loads): you decide upon low and high reading values and
manually enter them through the keyboard. The meter then operates with those
new parameters. Another method is live scaling, “LivE” : you apply low and
high input signals and change the meter reading accordingly. The meter
calculates a new scale factor and then operates with those new parameters.

14
4.3.1 Internal Scaling
Use internal scaling if you do not have an actual input signal to the meter. With
internal scaling the input values are assumed to be the low and high signal input
based on the selection input (e.g. if you selected 4-20 mA input, the “rd1” input
value will be 0400 for 4 mA and “rd2” will be 2000 for 20 mA). All you need to
do is change what the meter should read at these two points. For internal scaling,
follow these steps:

Refer to Table B-2 for a summary list of menu configuration.

1. Press the MENU button until the meter flashes “ScAL”.


2. Press the ¢/TARE button until the meter shows “int”.
3. Press the £/MAX button. The meter momentarily flashes “rd1” (read
1), then shows the low input value based on the selected input range (e.g.
if the input range is 0 - 10 V, the meter shows “0000”, or if input range is
4-20 mA, the meter shows “0400” ).
4. Enter the desired reading for this low input value using the £/MAX and
¢/TARE buttons.
5. Press the MENU button for the meter to accept the new value. The meter
then momentarily flashes “rd2” , then shows the high input value based on
the selected input range (e.g. if the input range is 0 - 10 V, the meter shows
“9999” or if the input range is 4-20 mA, the meter shows “2000” ).
6. Enter the desired reading for this high input value using the £/MAX and
¢/TARE buttons.
7. Press the MENU button for the meter to accept the new value, save
changes and return to the run mode. The meter shows “Stor” and
calculates the new scale factor based on the new high and low values.
8. If the display is not zero with no load on your sensor, press the ¢/TARE
button. Scaling is now complete.

15
4.3.2 Live Scaling

Use live scaling when a stable input source is available (e.g. 4-20 mA or
voltage calibrator). Also use live scaling when the actual input from your
sensor can be set to output values close to the low and high ends of your
input. For live scaling, follow these steps:

Refer to Table B-2 for a summary list of menu configuration.


1. Press the MENU button until the meter flashes “ScAL” .
2. Press the ¢/TARE button until the meter shows “LivE”.
3. Press the £/MAX button. The meter momentarily flashes “rd1” .
4. Apply the low input signal to the input. The meter then shows the live
low input source.
5. Change the meter reading as required for the applied low input using the
£/MAX and ¢/TARE buttons.
6. Press the MENU button for the meter to accept the new value. The
meter then momentarily flashes “rd2” .
7. Apply the high input signal to the input. The meter then shows the live
high input source.
8. Change the meter reading as required for the applied high input using the
£/MAX and ¢/TARE buttons.
9. Press the MENU button to save changes and return to the run mode.
The meter shows “Stor” and calculates the new scale factor based on
the new high and low values.

16
4.4 CHANGING THE METER S CALIBRATION

CAUTION: It is not necessary to calibrate a brand new meter, it


arrives completely calibrated. The following procedure modifies
the calibration of the meter.

This procedure should only be performed by qualified personnel with


accurate test equipment.

To change the meter’s calibration, follow these steps:

Disconnect the main power from the meter.

1. Install jumper S4 to enable the calibration procedure (refer to Figure 3-3).


2. Slide the main board back into the case and replace the cover.
3. Power on the meter.
4. Select the input range you want to calibrate (refer to Section 4.1,
Selecting Input Range)
5. Press the MENU button until the meter flashes “LINE”. The meter will
indicate the line frequency it is currently using. Line frequency default is
60H for 60 Hz and 50H for 50 Hz.
6. Press the ¢/TARE button until the correct line frequency shows.
7. Press the MENU button to store this line frequency selection. Once you
store your selection, the meter flashes “cAL”.
8. Press the ¢/TARE button to initiate the calibration procedure. The
meter momentarily flashes the selected input range that you are about to
calibrate, then flashes “rd1” .
9. Apply 0 Vdc or 0 mA and change the display to “0000” . Use the
¢/TARE and £/MAX buttons to change the digits.

17
4.4 CHANGING THE METER S CALIBRATION (Continued)

10. Press the MENU button for the meter to accept the new value. The
meter then flashes “rd2” .

11. Apply the high input voltage or current for the selected input range (e.g.
apply 9.900 V if you are calibrating 0-10 V range).

To calibrate the entire meter, you only need to calibrate the


following three ranges: 0-10 V, 0-5 V and 4-20 mA.

Table 4-1. Meter Calibration Table


Range Low Input rd1* Input High rd2*
0-10V 0 Vdc 0000 9.9 Vdc 9900
0-5, 1-5V 0 Vdc 0000 5.0 Vdc 5000
4-20, 20-4 mA 0 mA 0000 20.0 mA 2000
* Disregard decimal point

12. Change the display to equal the input source (e.g. if you apply
9.900 V for the high input voltage, change the meter reading to “9900”).
Press the £/MAX button to change the flashing digit’s value (increments
by one) and press the ¢/TARE button to move from one digit to the
next.

13. Press the MENU button once more to save changes. The meter shows
“Stor” and is now calibrated for the selected range.

14. Power off the meter. Remove the S4 jumper (to ensure no accidental
changes). Slide main board back into the case.

15. Power on the meter.

18
APPENDIX A

CHECKING AND CHANGING MAIN BOARD POWER JUMPERS

IMPORTANT: If you want to change the Factory preset jumpers,


do the following steps:

Disconnect the power from the unit before proceeding.

1. Remove the main board from the case. Refer to Section 2.6.
2. Locate the jumpers W1, W2, W3 and W4 (located near the edge of the
main board alongside the transformer).
3. If your power requirement is 115 Vac, install jumpers W1 and W2. If
your power requirement is 230 Vac, install jumpers W4 and W3. Refer
to Figures A-1 and A-2.

Refer to Section 2.5, Back of the Meter, for more information on dc-
power connections.

JUMPER
FOR
115Vac

W4
W2
W3
W1

Figure A-1. 115Vac Jumper Wiring (Default)


19
APPENDIX A -
CHECKING AND CHANGING MAIN BOARD POWER JUMPERS
(Continued)

JUMPER
FOR
230Vac

W4
W2
W3
W1

Figure A-2. 230 Vac Jumper Wiring

20
APPENDIX B

REFERENCE TABLES

Table B-1. Display Messages


MESSAGE DESCRIPTION
Peak value to follow
Valley value to follow
123* Peak value reached overload
123* Valley value reached overload
ER 1 Scaling format error
PrSt Peak reset
VrST Valley reset
T-RST Tare reset. Clears tare value.

* Represents any value

Table B-2. Menu Configuration Displays


MENU ¢/TARE £/MAX
InP 0-10
4-20
20-4
0-5
1-5
DEC.P FFF.F
FFFF.
FFFF
F.FFF
FF.FF
ScAL int
LivE rd 1*
XXXX
rd 2*
XXXX

* Not displayed unless you press the £/MAX key.

21
APPENDIX B - REFERENCE TABLES (Continued)
Table B-3. Run Mode Displays
Display ¢/TARE £/MAX RESET T-RST Jumpers

PrST Press to S3
Peak Reset activate installed

VrST Press to S3
Valley activate removed
Reset

Press to S3
Peak Value activate* installed
to follow

Press to S3
Valley activate* removed
value to
follow

T-RST Press to
Clears tare activate
value

Tare Press to
Display activate

* Press ¢/MAX again to remove peak or valley reading and display process value.
Table B-4. Factory Preset Values
Menu Item Description
INPT Input Range: 4-20 mA = 0 - 100.0
DEC.P Decimal Point Position: FFF.F
Sensor Excitation: 24 Vdc
Line Frequency: 60 Hz

22
APPENDIX C
SPECIFICATIONS
Analog Input Ranges: 4-20 mA, 0-5 Vdc, 1-5 Vdc, 0-10 Vdc
Input Impedance: Voltage: 1.0 Meg
Current: 100 Ω
Isolation: Dielectric strength to 2500V transient per 3mm
spacing base on EN61010 for 260Vrms or dc
working voltage
Accuracy: 0.05%R +/- LSB
Tempco: +/- 50 PPM/°C
Excitation Voltage: 24 Vdc @ 25mA
10 Vdc @ 25mA
Display: LED 7-segment, 14.2 mm (0.56”)
Range: +9999 to -1999
Decimal Point: 4 positions
Input Power Information:
Voltage ac: 115/230 Vrms ±10%
dc: 9-32Vdc (Isolated up to 300 Vp)
Power: 1.5 watt Excitation Unloaded
2.5 Watts Maximum (with 20mA excitation load)
Frequency: 50/60 Hz
External Fuse Protection Recommended:
ac 115V UL slow blow, 40mA, 250V
ac 230V UL slow blow, 20mA, 250V
Environment:
Operating Temperature: 0° to 60°C (32° to 140° F)
Storage Temperature: -40° to 85°C (-40° to 185° F)
Relative Humidity: 95% at 40°C (non-condensing)
Mechanical
Panel Cutout: DIN 4896-100 / 1.77 x 3.66” (45 x 92 mm)
Depth Behind Bezel: 3.94 in (100 mm)
Weight: 11 oz (312 g)
Case Material: Polycarbonate, 94 V-O UL rated
External Features: Tare, Display Hold, Tare Reset

23
APPENDIX C - SPECIFICATIONS (Continued)

48.00(1.890)
96.00 (3.780)

0.73 (18.67)

3.22 (81.7)

3.74 (95.0)
MAX

SIDE VIEW
TOP VIEW
PANEL THICKNESS:
6,4 (.25) MAX
0,8 (.03) MIN

1,5
R (.06) 45,01 +0,61/-0,00
4 PLCS (1.772 +.024/-.000)

92,00 +0,81/-0,00
(3.622 +.032/-.000)

PANEL CUTOUT

Figure C-1. Meter Dimensions

24
APPENDIX D - CE INFORMATION

This product conforms to the EMC directive 89/336/EEC amended by


93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.

Electrical Safety EN61010-1:2001


Safety requirements for electrical equipment for measurement, control
and laboratory.

• Double Insulation: Primary to Secondaries


• Pollution Degree 2
• Measurement *Category I
Note 1: *Measurements performed on circuits not directly connected
to the Mains Supply (power). Maximum Line-to-Neutral working
voltage 50Vac/dc.
Note 2: This unit should not be used in Measurement Categories II, III, IV.
• Dielectric withstand Test per 1 min (50/60Hz)
Power to Input/Output 2500Vac
• Transients Overvoltage
Input Power Transients Overvoltage 2500V
Input/Output Signals Transients Overvoltage 500V

EMC EN61326:1997 + and A1: 1998 + A2: 2001


Immunity and Emissions requirements for electrical equipment for
measurement, control and laboratory.

• EMC Emissions Class B, EN61326


• EMC Immunity** Table 1, EN61326

** I/O Signal and Control Lines require shielded cables and these
cables must be located on conductive cable trays or in conduits.
Furthermore, the length of these cables should not exceed 30 meters.

Refer to the EMC and Safety installation considerations


(Guidelines) of this manual for additional information.

25
INDEX
Calibration
and the S4 jumper 7
changing 17, 18
installing/removing calibration jumper 6
Configuration jumpers
changing 7, 8
Configuration mode 3
escaping from 4
RESET button 4
ßMAX button 4
∂TARE button 4
Connecting sensor input 9-11
Connector label 5
dc-power 6
2-wire voltage input connections 10
3-wire voltage input connection 9
connectors 5, 6
current input connections 11
Decimal point
disregarding in meter calibration 18
factory default 20
selecting 14
Disassembly 6
Front-panel buttons
MENU 4
RESET 4
T-RST 4
ßMAX 4
∂TARE 4
Input range
and internal scaling 15
calibrating 17
factory default 20
selecting 13
Installing the meter 9
Internal scaling 15

26
INDEX
Live scaling 16
Main board power jumpers 23, 24
MENU button 4
Meter modes
configuration mode 3
run mode 3
micro controller revision code
accessing 7
Mounting the meter 9
RESET button 4
Resetting peak and valley registers 4, 20
Run mode 3
MENU button 4
RESET button 4
ßMAX button 4
∂TARE button 4
S1, S2 & S3 DIP switches
changing 7
Scaling
error message 19
internal 14, 15
live 14, 16
Scaling without known loads 14, 15
Sensor excitation
factory default 20
Sensor input 9-11
connecting 9-11
T-RST button 4
Voltage calibrator
and live scaling 16
ßMAX button 4
∂TARE button 4

27
Warranty/Disclaimer
NEWPORT Electronics, Inc. warrants this unit to be free of defects in materials and workmanship for a period of one
(1) year from the date of purchase. In addition to NEWPORT’s standard warranty period, NEWPORT Electronics will
extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned
to NEWPORT.
If the unit should malfunction, it must be returned to the factory for evaluation. NEWPORT’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon
examination by NEWPORT, if the unit is found to be defective it will be repaired or replaced at no charge. NEWPORT’s
WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to
mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification.
This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being
damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of NEWPORT’s control. Components which wear are not
warranted, including but not limited to contact points, fuses, and triacs.
NEWPORT is pleased to offer suggestions on the use of its various products. However, NEWPORT neither
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the
use of its products in accordance with information provided by NEWPORT, either verbal or written. NEWPORT
warrants only that the parts manufactured by it will be as specified and free of defects. NEWPORT MAKES NO
OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED,
EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION
OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of NEWPORT
with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or
otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event
shall NEWPORT be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications
or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application,
or used on humans, or misused in any way, NEWPORT assumes no responsibility as set forth in our basic
WARRANTY / DISCLAIMER language, and additionally purchaser will indemnify NEWPORT and hold NEWPORT
harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to the NEWPORT Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO NEWPORT, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM NEWPORT’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the outside of the return package and on any
correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in
transit.

FOR WARRANTY RETURNS, please have the FOR NON-WARRANTY REPAIRS, consult
following information available BEFORE NEWPORT for current repair charges. Have the
contacting NEWPORT: following information available BEFORE
1. P.O. number under which the product was contacting NEWPORT:
PURCHASED, 1. P.O. number to cover the COST of
2. Model and serial number of the product under the repair,
warranty, and 2. Model and serial number of product, and
3. Repair instructions and/or specific problems 3. Repair instructions and/or specific problems
relative to the product. relative to the product.

NEWPORT’s policy is to make running changes, not model changes, whenever an improvement is possible. This
affords our customers the latest in technology and engineering.
NEWPORT is a registered trademark of NEWPORT Electronics, Inc.
© Copyright 2007 NEWPORT Electronics, Inc. All rights reserved. This document may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in
part, without prior written consent of NEWPORT Electronics, Inc.
For immediate technical or application assistance please call:
®

Newport Electronics, Inc.


2229 South Yale Street • Santa Ana, CA • 92704 • U.S.A.
TEL: (714) 540-4914 • FAX: (203) 968-7311
Toll Free: 1-800-639-7678 • www.newportUS.com • e-mail:[email protected]
ISO 9001 Certified

Newport Technologies, Inc.


976 Bergar • Laval (Quebec) • H7L 5A1 • Canada
TEL: (514) 335-3183 • FAX: (514) 856-6886
Toll Free: 1-800-639-7678 • www.newport.ca • e-mail:[email protected]

Newport Electronics, Ltd.


One Omega Drive • River Bend Technology Centre
Northbank, Irlam • Manchester M44 5BD • United Kingdom
Tel: +44 161 777 6611 • FAX: +44 161 777 6622
Toll Free: 0800 488 488 • www.newportuk.co.uk • e-mail:[email protected]

Newport Electronics B.V. - Benelux


TEL: +31 20 3472121 • FAX: +31 20 6434643
Toll Free: 0800 0993344 • www.newport.nl • e-mail: [email protected]

Newport Electronics spol s.r.o.


Frystatska 184, 733 01 Karviná • Czech Republic
TEL: +420 59 6311899 • FAX: +420 59 6311114
Toll Free: 0800-1-66342 • www.newport.cz • e-mail: [email protected]

Newport Electronics GmbH


Daimlerstrasse 26 • D-75392 Deckenpfronn • Germany
TEL: 49 7056 9398-0 • FAX: 49 7056 9398-29
Toll Free: 0800 / 6397678 • www.newport.de • e-mail: [email protected]

Newport Electronique S.A.R.L. - France


TEL: +33 1 61 37 29 00 • FAX: +33 1 30 57 54 27
Toll Free: 0800 466 342 • www.newport.fr • e-mail: [email protected]

Mexico and Latin America


FAX: 001 (203) 359-7807
TEL En Español: 001 (203) 359-7803

M1866/0307 12160ML-01 Rev. K

You might also like