GSS2-Plus Manual
GSS2-Plus Manual
This document has been prepared for the titled project or named part thereof and should not be relied upon or used for any other project
without an independent check being carried out as to its suitability and prior written authority of CAPCIS Ltd being obtained. CAPCIS Ltd
accepts no responsibility or liability for the consequences of this document being used for a purpose other than the purposes for which it was
commissioned. Any person using or relying on the document for such other purposes agrees, and will by such use or reliance be taken to
confirm his agreement to indemnify CAPCIS Ltd for all loss or damage resulting therefrom. CAPCIS Ltd accepts no responsibility or liability for
this document to any party other than the person by whom it was commissioned.
Registered in England No. 2291712. Registered office 25 Savile Row, London, W1S 2ES
CAPCIS Ltd. is an Intertek Group plc company
CAPCIS Ltd
www.capcis.co.uk
1. INTRODUCTION .................................................................................................... 1
1.1 Applications................................................................................................. 1
2. DELIVERABLES ..................................................................................................... 2
2.1 Parts ............................................................................................................ 2
2.2 Operation .................................................................................................... 2
2.2.1 Power Supply and Charger ............................................................ 2
2.2.2 Battery Low Indicator ...................................................................... 2
2.2.3 Connections .................................................................................... 2
3. FUNCTION TEST PROCEDURE FOR PM205 USING GSS2 PLUS ................... 3
3.1 Test 1: 100 Ohm Test ................................................................................. 3
3.2 Test 2: 25 Ohm Test ................................................................................... 3
3.3 Test 3: 10 Ohm Test ................................................................................... 4
3.4 Test 4: Voltage CAL Test............................................................................ 4
3.5 Test 5: Voltage Test.................................................................................... 4
3.6 Test 6: Voltage No Trip Test....................................................................... 5
3.7 Test 7: Channel Separation Test................................................................ 5
4. PM205 CALIBRATION USING GSS2 PLUS ......................................................... 6
4.1 Calibration ................................................................................................... 6
4.2 Test 8: RF Test ........................................................................................... 7
5. BATTERY CHARGING........................................................................................... 8
6. WARRANTY............................................................................................................ 9
7. SPECIFICATIONS ................................................................................................ 10
Pictures of the GSS2-Plus Kit
From Guy Tipton with Intertek
October 29, 2010
Note these pictures were added by Tom Vanderveen (Weatherford) to the Capcis/Intertek
Document for the purposes of clarity to what is included in the GSS2-Plus kit.
Page 1
1. INTRODUCTION
The GSS2 plus is a functional tool for testing the full operation of the PM205
product range. The GSS2 plus is portable, lightweight and rugged and has
been designed for use within a factory or service bay environment.
1.1 APPLICATIONS
The Hanson PM205 should be checked for calibration and functionality
regularly, the manufacturer recommends an annual check. The GSS2 plus is
part of that process and should be used when checking the electronic
functionality of any of the PM205 range.
Page 2
2. DELIVERABLES
2.1 PARTS
The GSS2 plus is made up of the following parts:
Rugged plastic case
GSS2 plus
Test lead for testing RF connection
Test lead for testing Wellhead connection and Alarm Relay connection
Mains battery charger
2mm plugs for access to Wellhead output
Instruction Manual
2.2 OPERATION
2.2.1 Power supply and charger
The GSS2 plus is supplied with a rechargeable battery that allows up to 10
hours continuous operation between charges.
The GSS2 plus’s battery charger will charge the on board batteries in approx 4
hours.
NOTE: The GSS2 plus should not be used when charging.
The charger conforms to the relevant European and North American standards.
The CHARGER should NOT be used outside.
2.2.2 Battery low indicator
There is a battery low light that glows on and off when the on board battery
requires charging. The light is labelled “BATT”.
2.2.3 Connections
There are various test connections on the GSS2 plus, their names and
functions are listed below:
TEST/I-O Used for connection to the Wellhead and Alarm relay
connections on the back of the PM205
RF OUT Used for connection to the RF connector on the back of the
PM205
Test sockets There are 2 test sockets on the top edge of the GSS2 plus,
these sockets enable the connection of a digital meter that
should be used in conjunction with the GSS2 plus when
checking the calibration of the PM205.
Charger Used to connect the mains powered charger.
NOTE: Only charge GSS2 plus using charger supplied.
NOTE: Never use GSS2 plus when connected to charger.
Page 3
Page 4
3.3 TEST 3: 10 OHM TEST
Optional: Set digital meter to volts
Set the “OFF-AC-RF” switch to AC
Set the “NOTRIP-CAL-TEST-VARI” switch to VARI
Turn the “VOLTAGE” knob to 0 volts
Turn the “RESISTANCE” knob to 10 ohms (Check with Digital
Meter in calibration)
Press “TEST” button
Ground integrity display will show 10 ohms: F+/-5ohms
C+/-0.3ohms
Page 5
3.6 TEST 6: VOLTAGE NO TRIP TEST
Optional: Set digital meter to AC volts 1volt range
Set the “OFF-AC-RF” switch to AC
Set the “NOTRIP-CAL-TEST-VARI” switch to NO TRIP
Turn the “VOLTAGE” knob to 0 volts
Turn the “RESISTANCE” knob to 0 ohms (Check with Digital
Meter in calibration)
Wellhead Voltage display will show 0.200 volts: F+/-0.05volts
C+/-0.010volts
There may be approx 5ohms on the Ground integrity display.
Alarm activity (this is the trip threshold) Alarm will reset
Alarm light will go off
3.7 TEST 7: CHANNEL SEPARATION TEST
Set the “OFF-AC-RF” switch to AC
Set the “NOTRIP-CAL-TEST-VARI” switch to NO TRIP
Page 6
Page 7
17. Wellhead Voltage 0.350 ± 0.004
18. Ground Integrity 35.0 ± 00.4
4.2 TEST 8: RF TEST
Set the “OFF-AC-RF” switch to RF
Set the “NOTRIP-CAL-TEST-VARI” switch to NO TRIP
RF button out Alarm will reset
Alarm light will go off
RF button in Alarm will sound
Alarm light will glow
Page 8
5. BATTERY CHARGING
ONLY USE NiMH (Nickel metal hydride) BATTERIES
Turn Gss2 plus OFF, plug the charger connector into “Charger” socket and
switch charger ON.
The following Lights on the charger should come on as follows
Red “Power” shows it is powered up and charging.
Green ON then flashing it is charging steady ON the battery is fully charged.
IF Green is flashing and Red is OFF then either the battery is not connected or
reversed.
NB Should the tip of the charger lead become detached; ensure that the 2 +
symbols line up when re-fitting.
Page 9
6. WARRANTY
CAPCIS Limited warrants this instrument to be free from defective material or
workmanship for a period of ninety days from date of original purchase.
CAPCIS Limited under this warranty is limited to repairing the defective device
when returned to the factory, shipping charges prepaid, within ninety days of
original purchase.
Units returned to CAPCIS Limited that have been subject to abuse, misuse,
damage or accident; have been connected, installed or adjusted contrary to the
instructions furnished by CAPCIS Ltd, or repaired by unauthorised persons will
not be covered by this warranty.
CAPCIS Limited reserves the right to discontinue models; change specification,
price or design of this device at any time without notice and without incurring
any obligation whatsoever.
CAPCIS Limited shall not be liable in any way for consequential damages
resulting from use of this device.
This warranty is in lieu of all other representations or warranties expressed or
implied and no agent or representative of CAPCIS Limited is authorised to
assume any other obligation in connection with the sale and purchase of this
device.
Page 10
7. SPECIFICATIONS
Charger unit
Input range 100-240 Vac
47-63 Hz
17 VA
Output 18 VDC
0.84A
Enclosure rating IP 60
No protection against
water