DataLog x2 & Terminal x2

Installation and User Guide

Document Ref: 04008373b


Date: December 2020

This document was prepared by the staff of Valeport Limited, the Company, and is the property of the
Company, which also owns the copyright therein. All rights conferred by the law of the copyright and by virtue
of international copyright conventions are reserved to the Company. This document must not be copied,
reprinted or reproduced in any material form, either wholly or in part, and the contents of this document, and
any method or technique available therefrom, must not be disclosed to any other person whatsoever without
the prior written consent of the Company.

Valeport Limited
St Peters Quay
Totnes Tel: +44 1803 869292
Devon, TQ9 5EW e mail: [email protected] | [email protected]
United Kingdom Web: www.valeport.co.uk

As part of our policy of continuous development, we reserve the right to alter, without prior notice, all
specifications, designs, prices and conditions of supply for all our equipment.
Table of Contents

Table of Contents
1. Introduction
............................................................................................................... 3
2. Installation
............................................................................................................... 4
3. Terminal
...............................................................................................................
X2 7
3.1. Establish
....................................................................................................................................
a Connection 9
3.2. Configure
....................................................................................................................................
an Instrument 11
3.3. Download
....................................................................................................................................
a file 15
3.4. Record
....................................................................................................................................
a File 17
4. Datalog
...............................................................................................................
X2 19
4.1. Home
....................................................................................................................................
ToolBar 19
4.1.1 File Manager
.................................................................................................................................... 19
4.1.2 Time Series Chart
.................................................................................................................................... 20
4.1.3 ....................................................................................................................................
Profile Chart 20
4.1.4 Scatter Chart
.................................................................................................................................... 21
4.1.5 ....................................................................................................................................
Polar Chart 22
4.1.6 Terminal Window
.................................................................................................................................... 22
4.1.7 ....................................................................................................................................
Table View 23
4.2. Instrument
....................................................................................................................................
ToolBar 24
4.2.1 Establish a connection
.................................................................................................................................... 24
4.2.2 Configure an Instrument
.................................................................................................................................... 27
4.2.3 ....................................................................................................................................
Download a File 31
4.2.4 Record a File
.................................................................................................................................... 33
4.2.5 ....................................................................................................................................
Instrument Tools 35
4.3. Chart
....................................................................................................................................
ToolBar 36
4.4. Recording
....................................................................................................................................
Toolbar 36
5. Tutorials
............................................................................................................... 37
5.1. Configure
....................................................................................................................................
a Model 106 Current Meter 37
5.2. Calibrate
....................................................................................................................................
a 740 Tide Gauge 41
5.3. Calculated
....................................................................................................................................
Density and Salinity 43

© 2020 Valeport Ltd Page 2


Introduction

1. Introduction
DataLog X2 and Terminal X2 are software packages to interface, configure and download data from
Valeport Instrumentation. DataLog x2, in addition to interfacing and download capability, has
extensive graphing and data analysis tools built in.
Terminal X2 will work with any Valeport Instrument that communicates via serial RS232.
Functionality may be limited with some older instruments.
At present DataLog X2 is not fully compatible with the following instruments and the listed packages
should be used instead:

Equipment Software to be Used

TideMaster TideMaster Express


MIDAS DWR WaveLog Express
MIDAS WTR WaveLog Express
MIDAS Surveyor SurveyLog
MIDAS BathyPack BathyLog
MIDAS CTD+ DataLog Pro
Model 803 ROVLog

Page 3 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

2. Installation
System Requirements:
DataLog X2 and Terminal X2 have been tested on Windows 8, 8.1 and 10.
Basic Requirements: 800MHz processor; 512MB of RAM; 1024 x 768 screen resolution; Free Disk-
space 20Mb (excluding data); IE6+ installed.
Recommended Requirements: 1 gigahertz (GHz) or faster 32-bit (x86) or 64-bit (x64) processor. 2
gigabyte (GB) RAM (32-bit) or 4 GB RAM (64-bit); 1280 x 1024 screen resolution; Free Disk-space
20Mb (excluding data of course!); IE9+ installed.

Installation Process:
To Install: Run Setup.exe from the installation directory.

Note: You may need to accept a Windows security warning to proceed based upon your
system configuration.

Confirm additional settings as required

© 2020 Valeport Ltd Page 4


Installation

Confirm the installation tasks and click "Install" to finalise the installation.

Installation can be terminated at any time by pressing the "Cancel" button.

Page 5 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Uninstall
To uninstall DataLog X2, navigate to the Windows "Control Panel" and select "Remove Program",
DataLog X2 can be selected from the program list and uninstalled.
Starting DataLog X2
To start DataLog X2 use the following methods:
Navigate to the "All Programs" link under the Windows "START" menu
Navigate to the folder "Valeport Software"
Click the icon "DataLog X2"
Alternatively if during setup you chose to create a desktop icon, locate the icon on the desktop and
double-click to open DataLog X2.

© 2020 Valeport Ltd Page 6


Terminal X2

3. Terminal X2
Terminal x2 is a lightweight terminal program for serial communication, configuration and control of
Valeport Instruments. It is intended as a direct replacement for Hyper Terminal which is no longer
available for Windows PC's.
It has a number of additional features over and above most terminal packages including data
logging, wizard based instrument configuration, file downloads, multiple port connections and TCP-
IP serial connection for use over internet connections.

Terminal Toolbar

This allows the port to be configured, serial com ports or TCP/IP ports can be selected. Baud rates
can be changed whilst still connected. Default port settings are 8 bit, No stop bit, No 1 Parity Bit.
These settings can be changed but as standard will work with all Valeport instruments.
Other options available from this toolbar include echo on/off, wrapping, hexadecimal split screen,
hide/show special characters and time stamping.

Page 7 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Connection Status

Shows the current port, baud rate and protocol and connection status.

Connection Window

The connection window shows any communications between the PC and the connected
instrument. In the example above the time stamping is enabled. Once the window is full, it will scroll
with any received data. Data can be highlighted, selected, copied and pasted.
Commands can be typed directly into the connection window or entered into the command line
window below

Command Line and Dropdown

The command line can be used to send commands to a connected instrument. A history of sent
commands can be accessed with the drop-down functionality

Connection information

Shows volume of data received during currently connected session, number of lines of data
received and rate at which data is being received.

© 2020 Valeport Ltd Page 8


Terminal X2

3.1. Establish a Connection


There are two methods of establishing a connection. When the instrument type is known and when
a generic serial connection is required.

Clicking on the connect icon will


bring up the connection wizard.
The instrument type can be
selected from the drop down list.
The connection settings will
default to the factory settings for
that instrument. If these have
been changed by the user it may
be necessary to change these.

Alternatively, set the connection


up using the settings toolbar.
Once the appropriate port and
baud rate settings are selected,
press the connect icon to
establish a connection
To establish TCP-IP serial
connection, select TCP-IP port
from the drop down menu,
Enter the IP address and port in
the Host settings

Page 9 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Once the connection is established, the status bar will show the connection details and any
communications will be shown in the window.
Commands can be issued via typing directly into the command window or via the command
toolbar.
The command toolbar stores a history of issued commands.
The status toolbar shows the amount of data received, sent and the received rate of data.

© 2020 Valeport Ltd Page 10


Terminal X2

3.2. Configure an Instrument


The configuration wizard is instrument specific and the options given will depend on the instrument
connected, the example given here is for a MiniSVP.

Clicking on the configure tool will bring up the


configuration wizard.

Select the instrument type and the connection


settings will be set to the factory default for this
instrument. They can be altered if required.

Select the comm port the instrument is


connected to and click next.

At the next window, set the timeout setting for


interrupting the instrument.
If attempting to connect to an instrument
operating in Burst Mode, the timeout should be
set to greater than the burst interval.
When Next is clicked the configuration wizard will
attempt to interrupt the instrument and read out
the current configuration.

If the wizard fails to interrupt the instrument, then


check the connection settings and retry.
If the instrument is successfully interrupted,
commands and responses should be visible in
the command window behind the configuration
wizard.

Page 11 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

The first wizard window shows information about


the instrument. Boxes are greyed out as
information cannot be changed by the user.

The next screen allows the pressure sensor to


be configured.
Output units can be set and the Tare setting can
be configured.
To set a manual Tare, enter the required figure in
the box.
To take a measured Tare, press the Set Tare
Button.
The following dialog will pop-up.

The Latitude value is used to convert pressure to


depth. An approximate value can be used for the
operational area.

© 2020 Valeport Ltd Page 12


Terminal X2

If the instrument being configured has a Real


Time Clock, the option to sync it to PC time is
available.

Header is a free text field that will be inserted into


the header of every recorded file.
Sampling modes will depend on the instrument
connected.

The final screen will show a summary of the


commands that will be sent to the instrument.

Page 13 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

If settings have not been changed, then no


commands will be shown and the finish button is
greyed out.
Simply type a # into the command window and
the Finish button will be available.

Once the configuration is completed, the


instrument will be in a paused (‘interrupted’) state
and, therefore, will require prompting from the
user to set it running. This can be achieve in one
of 3 ways:
1. Select the RUN button in the instrument tab of
DataLog X2 (See image).
2. Enter the command #028 in the Terminal
window.
3. Disconnect, then reconnect the
Power/Comms cable to activate the
instrument.

© 2020 Valeport Ltd Page 14


Terminal X2

3.3. Download a file

Clicking on the download icon invokes the


Download Wizard. As with all wizards the
first stage is to identify the connected
insturment and comms settings.

Ensure the instrument is switched on and


connected. Extend the interrupt timeout if
the unit is operating in burst mode (or
cycle the power to the instrument)

Page 15 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Location to download files to can be


defined here. Sub folders by instrument
and date will automatically created.
For instruments that support individual file
downloads (mini and 400 series), a file
list with check boxes will be displayed.
For older instruments such as the model
106/108/308 and the model 740, the
entire memory has to be downloaded in
one
A status window will show the progress
of the download and estimated
completion time.
.

Files can be deleted by clicking on the bin


icon.

There is no recycle bin. If the file is


deleted from the instrument it is gone
forever

© 2020 Valeport Ltd Page 16


Terminal X2

3.4. Record a File

Real time data can be logged to


file with the Recording wizard.
As with the other wizards, set
up the instrument type and
communications settings.
If the instrument is not listed,
then a 'Generic Instrument' can
be selected from the dropdown
and suitable comms settings
entered.

The next screen sets the


interuppt settings, but also adds
the ability to filter the incoming
string.
For example, if an instrument
such as the TideMaster is
outputting multiple NMEA
strings such as $PVTM1 for
met, and $PVTMA for Tide, the
start filter can be set to $PVTMA
to only log the tide data.

Page 17 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

For data recorded on


instruments without a Real
Time Clock, there is the option
to append a timestamp to each
record.

Files will be saved to the


selected location. Files can be
automatically named,
incremented and numbered.
Data will continue to be
recorded until the record button
is clicked again, or Terminal is
closed.

© 2020 Valeport Ltd Page 18


Datalog X2

4. Datalog X2
DataLog X2 is a package for communication, configuration and control of Valeport Instruments. It
also contains basic graphing and data display functions for internally recorded and real time data
from Valeport Instruments.
It has all the features of Valeport Terminal including data logging, wizard based instrument
configuration, file downloads, multiple port connections and TCP-IP serial connection for use over
internet connections.

4.1. Home ToolBar

The Home toolbar gives access to graphing and file management tools. Data from recorded files
and downloaded files can be loaded and viewed as tables and graphs.
4.1.1. File Manager
In DataLog X2 a file is either a recorded data file from real-time connection, or a downloaded file
from an instrument. Downloaded files and recorded files are automatically loaded into the file
manager window.

Previously downloaded or recorded files can be added by clicking on the button


Files will be sorted in the order that they are added. The select button activates or deactivates the
file.
4.1.1.1. Workspaces
A workspace is a saved configuration of files, graphs and tables. When the workspace is loaded, it
will automatically load the files and create the graphs and tables saved in the workspace.

Page 19 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

4.1.2. Time Series Chart


A time series chart can be for single parameter or multiple parameters. Time is always on the X
axis. If multiple parameters are selected then multiple Y axes will be displayed.

4.1.3. Profile Chart

A profile chart can show a single or multiple


parameters plotted against depth/pressure.
Depth/Pressure is always on a reversed Y axis.

© 2020 Valeport Ltd Page 20


Datalog X2

4.1.4. Scatter Chart


A scatter chart allows one parameter to be plotted against another. A good example of this is T-S
plot traditionally used in oceanography to identify water masses.

Temperature is show here on the Y axis with Salinity plotted on the X axis.

Page 21 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

4.1.5. Polar Chart


Polar Charts can be used to display vector type data with a magnitude + direction. eg. water current
direction and speed or wind speed and direction.

4.1.6. Terminal Window


The terminal window shows any comms between the insturment and DataLog X2. It can be used to
manually interact with the instrument. It gives access to all the functionality of Terminal X2

© 2020 Valeport Ltd Page 22


Datalog X2

4.1.7. Table View


The Table view shows the data contained in the file currently highlighted in the File Manager.

Greyed out data is data that has been filtered using the Filtering Tools available in the Chart Toolbar

Page 23 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

.
4.2. Instrument ToolBar

The instrument toolbar gives access to a range of functionality for connecting, configuring and
downloading data from a Valeport Instrument

4.2.1. Establish a connection


The connection wizard is shard with DataLog X2

There are two methods of establishing a connection. When the instrument type is known and when
a generic serial connection is required.

Clicking on the connect icon will


bring up the connection wizard.
The instrument type can be
selected from the drop down list.
The connection settings will
default to the factory settings for
that instrument. If these have
been changed by the user it may
be necessary to change these.

Alternatively, set the connection


up using the settings toolbar.
Once the appropriate port and
baud rate settings are selected,

© 2020 Valeport Ltd Page 24


Datalog X2

press the connect icon to


establish a connection
To establish TCP-IP serial
connection, select TCP-IP port
from the drop down menu,
Enter the IP address and port in
the Host settings

Page 25 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Once the connection is established, the status bar will show the connection details and any
communications will be shown in the window.
Commands can be issued via typing directly into the command window or via the command
toolbar.
The command toolbar stores a history of issued commands.
The status toolbar shows the amount of data received, sent and the received rate of data.

© 2020 Valeport Ltd Page 26


Datalog X2

4.2.2. Configure an Instrument


The configuration wizard is shared with DataLog X2.

The configuration wizard is instrument specific and the options given will depend on the instrument
connected, the example given here is for a MiniSVP.

Clicking on the configure tool will bring up the


configuration wizard.

Select the instrument type and the connection


settings will be set to the factory default for this
instrument. They can be altered if required.

Select the comm port the instrument is


connected to and click next.

At the next window, set the timeout setting for


interrupting the instrument.
If attempting to connect to an instrument
operating in Burst Mode, the timeout should be
set to greater than the burst interval.
When Next is clicked the configuration wizard will
attempt to interrupt the instrument and read out
the current configuration.

If the wizard fails to interrupt the instrument, then


check the connection settings and retry.
If the instrument is successfully interrupted,
commands and responses should be visible in
the command window behind the configuration
wizard.

Page 27 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

The first wizard window shows information about


the instrument. Boxes are greyed out as
information cannot be changed by the user.

The next screen allows the pressure sensor to


be configured.
Output units can be set and the Tare setting can
be configured.
To set a manual Tare, enter the required figure in
the box.
To take a measured Tare, press the Set Tare
Button.
The following dialog will pop-up.

The Latitude value is used to convert pressure to


depth. An approximate value can be used for the
operational area.

© 2020 Valeport Ltd Page 28


Datalog X2

If the instrument being configured has a Real


Time Clock, the option to sync it to PC time is
available.

Header is a free text field that will be inserted into


the header of every recorded file.
Sampling modes will depend on the instrument
connected.

The final screen will show a summary of the


commands that will be sent to the instrument.

Page 29 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

If settings have not been changed, then no


commands will be shown and the finish button is
greyed out.
Simply type a # into the command window and
the Finish button will be available.

Once the configuration is completed, the


instrument will be in a paused (‘interrupted’) state
and, therefore, will require prompting from the
user to set it running. This can be achieve in one
of 3 ways:
1. Select the RUN button in the instrument tab of
DataLog X2 (See image).
2. Enter the command #028 in the Terminal
window.
3. Disconnect, then reconnect the
Power/Comms cable to activate the
instrument.

© 2020 Valeport Ltd Page 30


Datalog X2

4.2.3. Download a File


The download wizard is shared with DataLog X2, it operates in the same way as in Terminal with
the addition that any downloaded files are automatically loaded into the File Manager.

Clicking on the download icon invokes the


Download Wizard. As with all wizards the
first stage is to identify the connected
insturment and comms settings.

Ensure the instrument is switched on and


connected. Extend the interrupt timeout if
the unit is operating in burst mode (or
cycle the power to the instrument)

Page 31 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

Location to download files to can be


defined here. Sub folders by instrument
and date will automatically created.
For instruments that support individual file
downloads (mini and 400 series), a file
list with check boxes will be displayed.
For older instruments such as the model
106/108/308 and the model 740, the
entire memory has to be downloaded in
one
A status window will show the progress
of the download and estimated
completion time.
.

Files can be deleted by clicking on the bin


icon.

There is no recycle bin. If the file is


deleted from the instrument it is gone
forever

© 2020 Valeport Ltd Page 32


Datalog X2

4.2.4. Record a File


The recording wizard in DataLog X2 differs from the recording wizard in Terminal X2. Rather than
just recording any data incoming on the serial port, DataLog X2 will interpret and parse the data into
the current workspace to allow real-time data to be displayed and potted using the charting tools
available on the Home Toolbar

Real time data can be logged to


file with the Recording wizard.
As with the other wizards, set
up the instrument type and
communications settings.

If the instrument is not listed,


then a 'Generic Instrument' can
be selected from the dropdown
and suitable comms settings
entered.

The next screen sets the


interrupt settings, selects the
incoming Data format and adds
the ability to filter the incoming
string.

If the data format is not known,


select generic instrument, the
next screen will allow the format
to be defined.

Filtering is also available. For


example, if an instrument such
as the TideMaster is outputting
multiple NMEA strings such as
$PVTM1 for met, and $PVTMA
for Tide, the start filter can be
set to $PVTMA to only log the
tide data.

Page 33 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

If the data format outputted by


the instrument is recognised,
the columns of data should
automatically populate.

If Generic format has been


selected in the previous screen,
the delimters, columns names
and units can be defined.

For data recorded on


instruments without a Real
Time Clock, there is the option
to append a timestamp to each
record.

© 2020 Valeport Ltd Page 34


Datalog X2

Files will be saved to the


selected location. Files can be
automatically named,
incremented and numbered.

Data will continue to be


recorded until the record button
is clicked again.

4.2.5. Instrument Tools

The instrument tools can be used a


shortcut to carry out task on a connected
instrument such as interupting a running
intrument, set a pressure Tare, clearing
the memory and setting the instrument
into Run mode.
For certain instruments such as the
model 740 that require site calibrations,
the calibrate button will be available.

Page 35 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

4.3. Chart ToolBar

The Chart Toolbar allows for some basic filtering to be applied to the datasets in the workspace.
Date and Depth filters can be applied using the toolbar buttons, One additional filter can be added
using the Add button. This brings up the filter dialogue where the parameter to filter with can be
selected and upper/lower limits set.

To clear the filter, click the clear button.

4.4. Recording Toolbar


The Recording toolbar becomes active when the Record button on the Home Toolbar is clicked.
Clicking the record button will activate the Recording Wizard.

The recording toolbar can be used to manage incoming data. The toolbar informs on the location
that files are being saved to, the file increment and the port settings of the attached instrument.
When DataLog X2 is actively recording data, a new file can be started by pressing the next file
button.

This is will also have the effect of resetting any graphs or charts currently active.

© 2020 Valeport Ltd Page 36


Tutorials

5. Tutorials
5.1. Configure a Model 106 Current Meter
Configuration of a Model 106 current meter is via the configuration wizard, this can be accessed via
the Instrument ToolBar.

Clicking on the Configure button will invoke the configuration wizard.

Select Model 106 Current Meter


from the dropdown menu.
This will populate the
connection settings with the
default setting for a Model 106.
Select the COM port that the
instrument is connected to.
Click Next.

Page 37 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

This will set the interrupt


timeouts for the instrument. As
default DataLog X2 will attempt
to interrupt an instrument for
60s.
If the instrument is not currently
connected to the PC, or the
seawater switch is not
connected if using an older
model, do so now and click
next.
If the interrupt is successful,
then commands should be
displayed bottom left being sent
to the instrument.
If no connection is established,
check connections and the
correct com port /baud rate is
being used and try again.

DataLog X2 will read the current


setting of the instrument and
display on screen.
Settings that are greyed out, are
factory set and cannot be
changed by the user

© 2020 Valeport Ltd Page 38


Tutorials

The instruments internal clock


can be manually set, or synced
to the PC time. This will be in
the time zone that the PC is
currently set to.

Sample Period and Averaging


Period can be set.
The combination of these two
will determine how often the
Model 106 will log and output
data.
Sample Period is how long the
106 counts impeller revolutions
for. This is set to a minimum of
3 seconds.
Average Period is how many
Sample Periods are averaged
before a data point is recorded
or output.
I.E
Sample Period = 3s
Average Period = 1
Data recorded once every 3s
Sample Period = 3s
Average Period = 20
Data recorded once every 60s.

Page 39 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

The final screen shows the


commands to be sent to the
instrument. If no changes have
been made to the instrument
setup, then no commands will
be shown here.
Manual commands can be
added here, for example if you
want to put the instrument
straight into run mode, then add
#028 to the list.

The final screen shows the


commands to be sent to the
instrument. If no changes have
been made to the instrument
setup, then no commands will
be shown here.
Manual commands can be
added here, for example if you
want to put the instrument
straight into run mode, then add
#028 to the list.
Alternatively, click finish to apply
setup commands, and then
start the model 106 running by
using the Instrument Tools run
button

© 2020 Valeport Ltd Page 40


Tutorials

5.2. Calibrate a 740 Tide Gauge


There are two routes to calibrate a 740 tide gauge in DataLog X2, it will appear as part of the
configuration wizard while configuring a model 740, or can be directly accessed via the Instrument
Tools Toolbar. To access via the Instrument Tools section, a connection will first have to be
established with the Model 740.

There are three options for the


calibration of a Model 740.
1: No calibration
2: Site Calibration - A site
specific calibration is calculated
by measured the pressure at
known high and low tide heights.
See Model 740 manual or
TideMaster manual for more
detail on the operation procedure
used to carry this out. Height can
be output in metres or feet.
3: User Calibration - a datum
offset and gain factor is manually
applied.

Selecting the Site Calibration


method activates the window
shown.
Select the required units, enter
the known tide heights and click
the << button to measure the
pressure at that height. 60s of
pressure data will be averaged.

Page 41 © 2020 Valeport Ltd


04008373b - DataLog x2 & Terminal x2 Installation and User Guide

If a User Calibration is selected,


the option to choose output units
and enter and offset and gain
factor are given.
Offset is the height of the
pressure transducer above or
below a known datum point IE
Chart Datum or Lowest
Astronomical Tide.
Gain is the factor applied to
compensate for the density of the
water in the pressure conversion
to depth.

Once the calibration method has been completed, click on next to commit settings to the model
740.

© 2020 Valeport Ltd Page 42


Tutorials

5.3. Calculated Density and Salinity


By uploading a data set that contains Sound Velocity, Pressure and Temperature DataLog x2 will
automatically calculate the Practical Salinity and from that Density at each point in the profile using
Valeport's proprietary computational techniques.
Salinity is calculated to an accuracy of ±0.05 PSU and Density to an accuracy of ±0.05kg/m 3