@@download
@@download
Name(s) of person
performing the work Users (Lab manager & Lab Technician & Tenants)
Specifications:
Measuring range: 0.00…14.00 pH and -2000…2000 mV.
Calibration: max. 3 point and linear.
Ph electrode:
Is a 3-in-1 plastic Mettler Toledo LE438 electrode with integrated temperature sensor measures pH, no tris
buffer, BNC + RCA (Cinch) plug connection, 1m cable length, suitable for general purpose applications and
field samples. When not in use the Ph electrode is stored in Ph electrode solution (KCL). The electrode
should never be left dry and uncovered otherwise it will be damaged. Always keep ph electrode in storage
solution between measurements and when not in use.
Operator must be trained in Ph Meter to guarantee safe daily use. Untrained Personnel are not be allowed to
operate the Ph Meter. Users should operate the Ph Meter according to instructions in the manual.
Note:
1. With the 1-point calibration only the offset is adjusted. If the sensor was previously calibrated with
multipoint calibration the previously stored slope will remain. Otherwise the theoretical slope (100
%) will be used.
2. With the 2-point calibration, both slope and offset are updated and shown on the right side of the
display.
3. With the 3-point calibration, both slope and offset are updated and shown on the right side of the
display. The slope and offset values are calculated using least square method through the three
calibration points (linear calibration). The FP20 offers the option of segmented calibration, where
slope and offset are calculated individually for each pair of adjacent buffers. Segmented calibration
is only meaningful for calibration with 3 or more points.
Safety precautions:
It is compulsory to wear protective clothing in the laboratory when working with hazardous or toxic
substances:
A lab coat should be worn.
Suitable eye protection such as goggles should be worn.
Use appropriate gloves when handling chemicals or hazardous substances, checking their
integrity before use.
All relevant safety measures are to be observed when working with chemicals.
Any spills should be wiped off immediately.
Page 2 of 7
Potential hazards
Substance or item handled Associated Existing Control Measures Risk Further Action required Risk
Hazard (s)
(L/M/H) (L/M/H)
Use of Ph Meter Electrical Only switch on the device if the device L No further action required if L
hazard - and power cable are undamaged. The the existing control measures
Electrical device has been properly installed and are adhere to.
shock – danger there is a preventative maintenance in
of death. place. Only trained personal are allowed
to use the machine. Incubator is earthed,
protective earth connection for the
machine is provided using 13A plug
fitted to the machine (RCD protected).
Make sure it has been PAT tested.
Measuring ph of chemicals Hazard PPE must be worn all the time (lab coat, L No further action required if L
associated with lab gloves and safety glasses). Spillage the existing control measures
chemical – must be cleared up immediately and are adhere to.
refer to decontaminated in accordance to COSHH
COSHH assessment. Instrument only to be used
assessment by trained personal
Ph electrode storage solution Irritant PPE must be worn all the time (lab coat, L No further action required if L
Harmful - lab gloves and safety glasses). Refer to the existing control measures
KCL COSHH assessment 0004. are adhere to.
Page 3 of 7
Persons potentially at risk:
Page 4 of 7
Arrangements for monitoring effectiveness of control:
Review of the Risk Assessment:
Yes No
Have there been any changes in the procedure or in the information available which
affect the estimated level of risk from the listed substances
Yes No
Page 5 of 7
Declaration by Tenant/Licensee/Technician:
I confirm that I have read this Risk Assessment and that I understand the hazards and
risks involved and will follow all of the safety procedures stated. Where PPE has been
identified as a control measure, I will ensure that it is worn.
Page 6 of 7
Name (Please print) Signature LM Date
Countersignature
Page 7 of 7