Ivium-n-Stat Multichannel potentiostat

More for less:


The first affordable multi-channel
high performance instrument!
• Batteries

• Fuel cells

• Electrode
development

• Biotechnology

• Corrosion

Up to 8 channels per frame


Stackable up to 8 frames with 64 channels

Each channel with integrated impedance analyser!


analyser!

Ivium-n-Stat:
• 1-8 channels per frame, stackable up to 64 channels
• Independent floating operation
• Modules encapsulated for easy handling
• 2-mode software:
IviumSoft: for independent operation per channel
IviumNcontrol: synchronous control of multiple channels
s-Module:
• ± 2.5 A @ ±10 V (expandable to ± 5 A @ ±10 V)
• Each channel with integrated impedance analyser: 10µHz to 250 kHz
• Data acquisition rate: 100 kHz
• Current ranges 10 nA – 10 A; min. resolution: 1 pA
• Peripheral analog/digital channels

Ivium Technologies B.V. www.ivium.nl


De Zaale 11 E-mail: [email protected]
5612 AJ Eindhoven Tel. +31 40 2390600
The Netherlands Fax. +31 40 2390601
Ivium-n-Stat s-Module

System Performance
4 Electrodes WE, CE, RE and S
Current compliance ±2.5 A
Maximum output Voltage ± 10 V below 1A and ±8 V up to 2.5 A
Potentiostat Bandwidth 250 kHz
Stability settings High Speed, Standard, and High Stability
Programmable response filter 1 MHz , 100 kHz , 10 kHz , 1 kHz , 10 Hz
Signal acquisition dual channel 16 bit ADC, 100.000 samples/sec

Potentiostat
Applied potential range ±10 V, with 0.33 mV resolution
Applied potential accuracy 0.2%, or 2 mV
Current ranges ±10 nA to ±10 A in 10 steps
Measured current resolution 0.015% of current range, minimum 1 pA

Galvanostat
Applied current resolution 0.33% of applied current range
Applied current accuracy 0.2%
Potential ranges ±10 mV, ±100 mV, ±1 V, ±10 V
Measured potential resolution 0.038% of potential range, minimum 3 uV

Impedance Analyser
Frequency range 10µHz to 250 kHz
Amplitude 0.015mV to 1.0V, or 0.03% to 100% of current range
DC offset 16 bit dc offset subtraction, and 2 dc-decoupling filters

Electrometer
Input impedance >1000 Gohm //<8pF
Input bias current <10 pA
Bandwidth >1 MHz

Peripheral connections
2 Analog in 0 to +5.000 V, 16 bits resolution, bandwidth 40 kHz
1 analog out 0 to +4.096 V, 16 bits resolution
1 Digital input, 3 Digital outputs 0 to +5 V
I-out and E-out analog monitor for cell current and potential
AC-out ±0.5 V sinewave 10µHz-250kHz with variable attenuation
Channel-X, Channel-Y inputs ±4 V: to record impedance from peripheral devices

Environment
Size w x d x h = 3 x 35 x 13 cm
Weight 0.7 kg
PC requirements Windows XP or Vista, with free USB port

Specifications: IviumNstat main frame


Slot positions 8 (can mount up to 8 S- or B-modules)
Common connectors GND and combined emergency off control monitor
Power requirements 100-240 V, 47-63 Hz, 300 W
Interfacing USB 1.1 or 2.0 compliant
Size w x d x h = 47 x 36 x 14 cm
Weight 6.2 kg (no modules)
11.8 kg (with 8 modules)
Power booster
B-module: power booster ±5 A ±10 V power booster for S-module
Available Q2 of 2010

Specifications subject to change, Ivium Technologies BV ©2009