101 - Basics of Irradiance in Weathering Testing 2018
101 - Basics of Irradiance in Weathering Testing 2018
101 - Basics of Irradiance in Weathering Testing 2018
Weathering Testing
Introduction
In exposure tests it is important to know both the amplitude of the radiation impinging on a test specimen’s surface
as well as the accumulated energy dose over the exposure period. Following Newton’s inverse square law, with a
constant point source output, that amplitude at the test specimen surface will decrease with the square of the
distanceNote 1. Therefore, describing the
intensity only in terms of the emission
power of the source is not useful. We
need to know the amplitude at the test
specimen surface, from which we can
calculate the accumulated exposure dose
over time.
Dose
The accumulated radiation dose deposited on a sample surface is irradiance integrated over time and is the Radiant
Atlas_TG101_Basics of Irradiance in Weathering Testing 2018-02-16_AZ.docx www.atlas-mts.com 1
2 -1 -1
exposure (H), usually expressed in units of J/m nm where J is Joules and nm is the same wavelength or range of
the spectral irradiance value. Therefore, the general shorthand formula used is:
H (J/m2 nm-1) = E (W/m2 nm-1) X t (seconds) Equation 1.
While not shown, the units convert correctly since 1 Joule = 1 Watt-second. Frequently the radiant exposure will be
stated in terms of kilojoules (J X 103) or Megajoules (J X 106) and, although frequently omitted, is always at the same
wavelength or wavelength range as the spectral irradiance
value. Note that spectral ranges cannot be mixed in
Equation 1. Also, time is more commonly expressed in
hours (3,600 seconds). Therefore, factors such as hours X
3.6 (to convert to kiloJoules/m2) or hours X 0.0036 (to
convert to Megajoules/m2) are often substituted for time (t)
in seconds in this equation. Note that this is the radiation
exposure time only, as some test cycles have dark periods.
Normalization
As the amplitude and shape of the spectral curve will vary, for example between solar radiation and different
laboratory light source simulations, it is often desirable to compare SPD’s based on a common reference point. The
process of normalization adjusts data sets by commonizing the curves to a specific irradiance value at a particular
wavelength. This can only be performed when the curve shape changes linearly with irradiance, as in filtered xenon
arc solar simulators such as Atlas xenon weathering instruments. An example of normalization to an irradiance of
0.55 W/m2 • 340nm is shown in Figure 6. Terrestrial solar radiation data should not be adjusted, since the SPD from
the sun will constantly change due to the sun’s position in the sky and atmospheric conditions. Therefore, the shape
of solar SPD does not remain constant at different amplitudes during the day. It is recommended that only the artificial
simulation spectra be normalized relative to the natural solar radiation, and not to adjust natural values relative to an
artificial one.
Irradiance will also change as the source to test specimen distance is altered, but will not affect the spectral balance.
Exposure “equivalences”
A very common question is “how long do I need to run laboratory test method “X” to equal “Y” years exposure
outdoors in location “Z”? The issues of exposure correlation and specific material degradation are very complex and
there are no simple answers to this question. Also, the answer is different for different test condition/test material
combinations. Actual correlation can only be established by comparing the outdoor and laboratory test
Figure 7. Atlas Miami test site solar radiation data summary for calendar 2017.
results for a specific material formulation and a specific property change. However, this presents a “which came first,
the chicken or the egg?” problem in that often the very reason for laboratory testing is because there is no real-time
outdoor test data for the material, and/or it will take too long to produce.
Note 1. Newton’s inverse square law only directly applies to point light sources. Laboratory exposure devices
typically use source geometries that are not point sources, or use multiple sources, so there is some deviation from
the mathematical formula and the actual irradiance value should be measured rather than calculated. However, the
concept of the relationship between source-to-specimen distance and irradiance still applies.
Author: Allen Zielnik, Atlas Material Testing Technology LLC, 16 February 2018