CM-BASEα 3713 E05 アラーム機能追加版
CM-BASEα 3713 E05 アラーム機能追加版
CM-BASEα 3713 E05 アラーム機能追加版
CM-BASE
CM-BASE Special Scale Series
Instruction Manual
CONTENTS
1. Precautions for Use ................................................................................................................................................... 2
2. Refractive Index, Brix, and Special Scale ...................................................................................................... 7
(1) What is refractometer? .................................................................................................................................... 7
(2) What is the Refractive Index? ..................................................................................................................... 7
(3) Brix(%), Special scale ...................................................................................................................................... 7
(4) Temperature correction ................................................................................................................................... 7
3. Unpacking and Installation..................................................................................................................................... 8
(1) Unpacking .............................................................................................................................................................. 8
(2) Installation .............................................................................................................................................................. 8
4. Names and Functions of Components ............................................................................................................ 9
5. Mounting the Sample Inlet Unit ......................................................................................................................... 10
6. External Output .......................................................................................................................................................... 12
(1) Recorder Output ............................................................................................................................................... 12
(2) RS-232C Output ............................................................................................................................................... 12
(3) Alarm Output (high- and low-limiter output) ....................................................................................... 16
7. Power Supply .............................................................................................................................................................. 17
8. Measuring ..................................................................................................................................................................... 18
9. Zero-setting .................................................................................................................................................................. 19
10. Offset Function ........................................................................................................................................................... 20
11. Bias Adjustment ......................................................................................................................................................... 22
12. Setting Alarms .............................................................................................................................................................. 23
13. Error and Display Lamp......................................................................................................................................... 25
(1) During measurement (including when the power is turned on) ............................................... 25
(2) During zero-setting .......................................................................................................................................... 25
(3) During measurement or zero-setting ..................................................................................................... 25
14. Cleaning the Prism ................................................................................................................................................... 26
15. Consumable Parts and Optional Items ......................................................................................................... 27
(1) Consumable Parts ........................................................................................................................................... 27
(2) Optional Items .................................................................................................................................................... 27
16. Relationships between Brix(%) Values and Refractive Index (nD) Values ................................ 28
17. Specifications .............................................................................................................................................................. 29
18. Repair Service and Warranty Period .............................................................................................................. 35
19. ATAGO CO.,LTD. Service Centers .................................................................................................................. 36
1
1. Precautions for Use
Introduction
Thank you for purchasing the CM-BASEα series. Before using your CM-BASEα series, read
this instruction manual carefully, and keep it on hand for future reference.
Explanation of indications
WARNING If this indication is neglected and the instrument is handled
incorrectly, serious injury or death may result.
Explanation of symbols
This symbol denotes an item of which to be warned or cautioned.
The contents of warning are described in detail in or near the △.
This symbol denotes an action that must not be performed (a prohibited item).
The contents of prohibition are described in detail in or near the ○.
2
Handling of this instrument
WARNING
◇ When measuring a substance harmful to ◇ Immediately turn off the power (DC24V) if
the human body, be aware of its the unit begins to overheat, smoke or emit
properties and wear protective gloves, an abnormal smell. Fire or malfunction
mask, etc. may result if the instrument continues to
be used. Contact your ATAGO Distributor
for an inspection.
◇ Do not attempt to repair, modify, or ◇ If the instrument is dropped or is
disassemble the instrument yourself. subjected to a strong shock, have it
Improper servicing may result in fire, inspected by an Authorized ATAGO
electrical shock, or burns. Service Center.
Fire or malfunction may result if the
instrument is used.
CAUTION
◇ Do not apply water or sample to any part ◇ Do not measure very hot or highly acidic
of the instrument other than the surface of samples. This may damage the prism,
the prism. which would result in inaccurate
Applying water to any other part of the measurements.
instrument may result in a malfunction.
3
Handling of this instrument (Continued)
CAUTION
◇ Do not measure a sample that can ◇ If the sample could possibly stain the
damage the prism or sample inlet unit. prism, immediately clean the prism
When the power is turned on, do not according to the procedure described on
measure a sample that is below 10˚C or page 26.
above 50˚C.
◇ The CM-BASEα power (DC24V) should ◇ When transporting the instrument, pack
be turned off when it is not to be used for carefully.
a long period of time.
◇ Carefully read this instruction manual and fully understand the function and operation of
each part of the instrument before use.
◇ Check that each part of the instrument operates normally before use.
◇ Perform the necessary inspections and reference adjustments according to the instruction
manual.
◇ The manufacturer shall not be held responsible for any damage that results from using the
instrument for other than its intended purposes (measurement of Brix, sugar content, liquid
concentration).
◇ The prism in contact with the sample is a consumable item.
◇ ATAGO shall not be held responsible if the use of the instrument has an undesired effect
on the measured materials.
4
Handling of plug
WARNING
◇ Supply the CM-BASEα with DC24V only ◇ Do not use a power cable that is
(allowable fluctuation is ±10 %). damaged, cut, broken, or altered. Use
Short-circuit, smoke, or fire may occur if may result in fire, electrical shock, or
other voltages are used. burn.
For repair service of the power cable,
contact an Authorized ATAGO Service
Center.
CAUTION
◇ Do not insert or pull out the connector and
plug with wet hands.
5
Conditions to be observed when using
Environmental conditions
◇ Use the instrument at an altitude below 5,000m (above sea level).
◇ Use the instrument indoors.
◇ Use the instrument where the temperature is between 5 to 40˚C.
◇ Do not leave the instrument in a location exposed to direct sunlight or near a heating unit
where the temperature may rise.
◇ Do not change the environmental temperature of the instrument suddenly.
◇ Do not place the instrument in a place where it may be subject to strong vibrations.
◇ Do not use the instrument where there is much dust.
◇ Do not leave the instrument where the temperature is extremely low.
◇ Do not leave the instrument in a damp place.
◇ Do not place or drop heavy objects on the instrument.
◇ Use this instrument under the condition where humidity is 80% at 31˚C or lower, falling
linearly to 50% at 40˚C.
◇ Main supply voltage fluctuation should not to exceed ±10% the nominal voltage.
◇ Installation categories (Overvoltage Categories): ll
◇ The pollution degree is 2 (according to IEC60664).
Handling
◇ Do not drop the instrument or subject it to any strong shock.
◇ The power cable may be damaged by:
Bending the cable.
Pulling the cable.
Twisting the cable.
Placing the cable under heavy objects.
Catching the cable between objects.
Daily maintenance
◇ If the instrument becomes dirty, wipe it with a soft cloth.
◇ Do not use benzine, paint thinner, etc. to clean the instrument.
6
2. Refractive Index, Brix, and Special Scale
The CM-BASEα is a refractometer that detects the Refractive Index of a sample and outputs
the Brix(%) value on the display.
The "CM-BASEα special scale series" is a refractometer series that detects the refractive
index of a sample and outputs the concentration, specific density, and Baume degree.
(1) What is refractometer?
When a straw is placed into a glass filled with water, the straw appears to bend. When a
straw is placed into a glass filled with sugar water, the straw appears to bend much more
sharply than in the case of water alone. This phenomenon is known as the refraction of
light. The refractometer is an instrument that measures the Refractive Index by utilizing
this principle (the Refractive Index increases in proportion to the concentration of the
solution), and was developed by Dr. Ernst Abbe at the end of the 19th century.
(2) What is the Refractive Index?
If the Refractive Index of air under atmospheric
pressure is 1, then when light enters medium x, the
ratio of the sine of the incident angle <alpha>
measured against the phase boundary to the sine of
the refracting angle <beta> is called the Refractive
Index of the medium x.
The Refractive Index varies with the wavelength of
light and temperature and is represented as follows:
n : Represents the Refractive Index
t : Temperature (˚C)
D : D-line of natrium (589nm)
For example, Refractive Index of water at 20˚C under the D-line is:
= 1.33299 (Generally expressed as nD = 1.33299.)
m The Refractive Index is based upon the supposition that the Refractive Index in a
vacuum is 1 and is called the absolute Refractive Index. Generally, however, this index
is seldom used.
(3) Brix(%), Special scale
The CM-BASEα is programmed with the Brix(%) scale, based on the Refractive Index of
water (nD = 1.33299) as the reference (0%). Models from the CM-BASEα Special Scale
Series are equipped with various dedicated scales that were created based on the
correlation between specific sample's concentration(%) and refractive index values.
The Brix(%) scale represents the weight of sucrose expressed by percentage (sucrose
weight in grams contained in 100 grams of sucrose solution). Therefore, this scale
corresponds with the sucrose concentration. For samples such as water-soluble cutting
oil, it is possible to create a conversion table in advance to show the relation between the
sample’s concentration % and Brix %, and convert the Brix measurement value into its
concentration. However, it is very common to create a concentration standard using Brix
values, and monitor with Brix values in everyday operations.
C The relationship between Brix(%) and the Refractive Index (nD) is outlined on
page 28.
(4) Temperature correction
The Refractive Index of a substance varies with temperature. Thus, when using a refractometer
to measure the Refractive Index of a liquid, the measurement value will vary with the sample
temperature. The CM-BASEα series always detects the prism temperature. The value of the
measurement is automatically corrected for temperature by a built-in processor, so that the
displayed value is equal to the value measured at 20˚C (provided that the sample temperature is
within the range of the temperature compensation).
7
3. Unpacking and Installation
(1) Unpacking
① Unpack the CM-BASEα and confirm that there is no external damage.
② Check that the following items are included.
(2) Installation
8
4. Names and Functions of Components
① Liquid Crystal Display (LCD) ⑥ Screw mounting holes
Displays the measured values. Used to mount the main unit to the
stand (optional), etc.
② ZERO key
Press for zero-setting and offset / bias ⑦ Connector
function setting. Terminal to connect the power
(DC24V) and output cable.
③ Set key
Press when setting the alarm output ⑧ Prism
Corrosion resistant optical glass, with
④ Indicator Lamp (Green) a polished surface to reflect light.
The light flashes during zero-setting,
and stays lit during measurement. ⑨ Prism stage
Connected to the sample inlet and
⑤ Indicator Lamp (Red) fastened by a clamp band.
The light comes on/flashes when an
error occurs.
9
5. Mounting the Sample Inlet Unit
◇ Turn off the power (DC24V) before mounting. Electrical shock
may occur if the unit is mounted with the power connected.
WARNING
◇ Ensure that the prism is clean and free of any damage and/or
scratches. Completely remove any sample before installing the
CAUTION CM-BASEα to piping or equipment.
Sample inlet units are available as pipe fittings: straight type and L type.
The straight type should be installed on straight sections of piping. The L type inlet
should serve as the joining piece for piping intended to align in an L shape.
There are three ways to connect a sample inlet to a pipe: IDF/ISO clamp union (ferrule),
IDF/ISO screw union (screw), and JIS Flange.
Other sample inlets available include small diameter series compression fitting (oute r
diameter 10mmφ) and hose connector (outer diameter 12mmφ).
10
Mounting procedure (Fig. 5-1, Fig. 5-2, and Fig. 5-3)
① Install the CM-BASEα so that the prism surface is at a right angle to the ground.
② Attach the sample inlet unit to the CM-BASEα with O-ring inserted between them, and
fasten them together with the clamp band.
③ Install the inlet unit so that the sample solution runs from the lower nozzle to the upper
nozzle to prevent air bubbles from forming.
④ When connecting the tubes to the hose connector, clamp them with a tie band. The tie
bands provided with the hose connector are made of plastic. If chemicals corrosive to
plastics are used, substitute with tie bands made of other material.
⑤ The prism surface may become contaminated with solids, dirt and/or grease. If this
happens, the prism surface must be cleaned by hand. (See page 26.) The sample
inlet unit should be installed in such a manner that it can be easily removed to allow
access to the prism for cleaning.
Prism stage
O-ring
(silicon or EPDM)
Flow of solution
Sample inlet Prism stage
Clamp band
unit
(optional) Top
ex.hose connector
Bottom
O-ring (silicon
or EPDM) Flow of solution
Dimensions
CM-BASEα
11
6. External Output
◇ Turn off the power (DC24V) when connecting the cable from the
terminal.
WARNING
From the bottom of the main unit`s body, there is an external data output for RS-232C
and Recorder Output (4 to 20mA).
The provided cable connects to the base of the CM-BASEα series, and it provides power
(DC24V) to the instrument and transmits Brix and temperature data via RS-232C, and
transmits Brix data via DC 4-20mA. Brix values ranging from -2.0 to 53.5% are
communicated as DC ranging from 4 to 20mA respectively.
Connector pin
Code color Signal name
number
1 Black DC24V
DC24V
2 Black / White
GND
3 Red DC4 to 20mA
4 Yellow Low-limiter output
5 Brown High-limiter output
6 Yellow / White Limiter Com
7 Red / White DC4 to 20mA GND
8 - -
9 Green / White RS-232C GND
10 Green RS-232C
Please refer to the specification page for the special scale series.
12
Computer Output
① Computer - Data Setting
13
(3) The serial settings are conformed to the instrument settings.
Enter the port number selected in step (1).
Click OK.
14
② Data Output From Main Unit to PC
Every time a measurement is taken, a new row of data appears in the Tera Term
window.
△△.△ : Temperature
Example
Output Data Temperature(˚C)
9.5 9.5˚C
23.5 23.5˚C
Please refer to the specification page for the special scale series.
15
(3) Alarm Output (high- and low-limiter output)
User can decide own acceptable Brix range, and set it as upper and lower limit of alarm
output.When the value exceeds the acceptable range, alarm output will be operated.
N During alarm output is operated, the value on display will be blinking.
User can also set whether alarm output can be continued (delay time setting) after alarm
output is operated and a few seconds have passed.
Minimum width
Factory settings Setting range of setting range
Lower Limit Brix -2.0% Brix -2.0 to 52.5%
Brix 0.1%
Upper Limit Brix 53.5% Brix -1.0 to 53.5%
Delay Time 0 second 0 to 99 seconds 1 second
Please refer to the specification page for the special scale series.
C For upper and lower limit setting and delay time setting, please refer “12.Setting
Alarms” on page 23.
V+
Photocoupler R
Sequencer
COM
※R (resistance) and V+ are not necessary if the input to a sequencer or the like is pulled up.
16
7. Power Supply
◇ NEVER connect the CM-BASEα to power other than DC24V
(allowable fluctuation is ±10%).
WARNING ◇ NEVER use a power cable that is damaged, cut, broken, or
altered. Fire, electrical shock or burns may occur.
To purchase replacement power cables, contact an Authorized
ATAGO Distributor.
② Once the power (DC24V) is turned on to the instrument, either the green light or red
light will come on.
m The CM-BASEα has no power switch. When power (DC24V) is supplied to the
CM-BASEα, the measurement value display illuminates and the Brix(%)
measurement commences.
The green light comes on when a sufficient amount of liquid with a Brix value of -2.0 to
53.5% is passing by the prism. When there is not enough liquid passing by, the red light
turns on and error message "LLL" is displayed. When the Brix of the liquid is above
53.6% or more, the red light turns on and "HHH" is displayed.
17
8. Measuring
◇ DO NOT measure any sample that can damage the prism or the
sample inlet unit. Sample temperature should be kept between
CAUTION 10˚C and 50˚C when the power (DC24V) is turned on.
◇ If the sample solution could potentially stain the prism,
immediately clean the prism after measurement (following the
instructions on page 26 of this instruction manual).
① When power (DC24V) is supplied to the CM-BASEα, the Brix(%) measurement starts.
(*) The CM-BASEα has no power switch.
② The Brix(%) value will be displayed when sample flows onto the prism surface.
③ If there is no sample on the prism surface (only air), the red light is on and the error
message LLL will be displayed.
External output relationship is shown in the table below.
m See page 25 if the red light is on even though the prism surface is sufficiently
covered with liquid.
④ The CM-BASEα constantly detects the temperature of the prism. When the sample
temperature is within the range of 10 to 50˚C, the Brix(%) values are automatically
compensated for temperature.
m A left arrow will appear on the screen when the sample temperature is 9.4˚C or
less, or 50.6˚C or more. The Brix values displayed while the arrow is displayed
are not properly temperature-compensated.
Please refer to the specification page for the special scale series.
18
9. Zero-setting
m Before zero-setting, confirm that the prism surface is clean.
① Confirm that the sample inlet unit is properly connected to the piping.
⑥ The green light and “000” on the LCD will flash three times.
Once zero-setting is complete, the green light will remain lit and it will start taking the
measurements again.
m If zero-setting is not successfully completed, the red light will flash and “AAA”
will display on the LCD. Pressing the ZERO key once more will turn off the red
light and will restart taking measurements. Turn off the power (DC24V), clean
the prism, and try zero-setting again with an ample amount of tap water.
19
10. Offset Function
There is an offset function that allows to enter and display the coefficient (a) and constants
(b) of the measured value. (CM-BASEα, CM-BASEα-101S only)
For cutting oil, the actual concentration and the displayed value on the instrument differs
depending on the type of oil being measured. For those case, the actual concentration can be
displayed by caluculating the converstion factor and the coefficient (a).
① Hold the ZERO key for about 3 seconds (while the power is on).
④ Confirm by pressing SET key and ZERO key at the same time.
Next, move to setting of coefficient (a).
⑦ Confirm by pressing SET key and ZERO key at the same time.
⑧ Setting completed.
It automatically returns to measurement.
20
*To remove the Offset function.
To remove the offset function, set the default values (a: 1.0, b: 0.0).
It can also be removed automatically by holding the ZERO key during setting.
④ "a" flashes.
21
11. Bias Adjustment
The Bias Adjustment feature allows the instrument to display readings that have been
adjusted by a fixed value added/substracted from the actual measured value.
(CM-BASEα and CM-BASEα-101S have an offset function (see P.21 for details)).
Hold the ZERO key for approximately 5 seconds to activate the TEST mode, and the
displayed value will start flashing.
The display value increases by 0.1 each time the ZERO key is pressed. The value can only
increase, not decrease.
Continue to press the ZERO key until the maximum value +2.0 is dsplayed, and then -2.0 will
appear. From then on, the value increases from -2.0 by 0.1. (Example: 2.0 → 1.9)
Example:
Displaying 10.0% as 10.5%
② Flashing
④ Confirm by pressing SET key and ZERO key at the same time.
⑤ Setting completed.
The bias adjustment will reset when the ZERO setting is done.
Writing down thr biaz setting is highly recommended.
22
12. Setting Alarms
For lower limit (ALL)
① Hold the SET key for about 3 seconds (while the power is on).
④ Confirm by pressing SET key and ZERO key at the same time.
Next, move to setting of upper limit (ALH).
⑦ Confirm by pressing SET key and ZERO key at the same time.
Next, move to setting of delay time (ALd).
23
For delay time (ALd)
⑩ Confirm by pressing SET key and ZERO key at the same time.
⑪ Setting completed.
It automatically returns to measurement.
24
13. Error and Display Lamp
(1) During measurement (including when the power is turned on)
●error sign: The red light is on and "LLL" is displayed.
<situation> No or an insufficient amount of sample is on the prism surface.
<situation> When a sufficient amount of sample within the indication range is on the prism
surface.
solution: Turn the power (DC24V) off and ensure that the prism is clean. If the error persists
even when there is a sufficient amount of sample to cover the prism, air bubbles or
undissolved solids may be the cause.
<situation> When a sufficient amount of sample within the indication range is poured.
solution: Turn the power (DC24V) off and ensure that the prism is clean. If the error persists
even when there is a sufficient amount of sample to cover the prism, air bubbles or
undissolved solids may be the cause.
<situation> When pressing the ZERO key, with an ample amount of tap water on the prism
surface.
solution: Turn the power (DC24V) off and ensure that the prism is clean. Try zero-setting
again with an ample amount of tap water.
25
14. Cleaning the Prism
◇ Before running hazardous substance(s) through any system,
necessary precautions should be taken to ensure the safe
WARNING handling of the hazardous substance(s). If using a sample inlet
unit, use caution when disconnecting the CM-BASEα unit.
◇ Before cleaning the prism, be sure to turn off the power
(DC24V).
① Detach the clamp band that connects the main unit to the sample inlet unit, piping or
tank.
② Clean the prism surface carefully with a soft tissue soaked with warm water or ethyl
alcohol. If the sample solution contains oil or grease, use ethyl alcohol to ensure the
prism surface does not develop a film. Development of a film on the prism could
cause erroneous measurements.
③ NEVER clean the prism with an abrasive material. Cleaning the prism with an
abrasive material could cause scratches on the prism which could lead to erroneous
measurements.
④ After cleaning is complete, re-attach the CM-BASEα unit to the sample inlet unit,
piping or tank.
26
15. Consumable Parts and Optional Items
(1) Consumable Parts
The following consumable parts are available for the instrument.
These items should be monitored and replaced as necessary. To place an order, please
contact an Authorized ATAGO Distributor.
Part name Part number Description
Tie band RE-8507 Maximum temperature: Max. 150˚C
Quantity: 10 pieces
For water bath connection
O-ring (Silicon) RE-68100
O-ring used to connect the sample inlet unit to the
O-ring (EPDM) RE-68115
prism stage unit.
O-ring (Viton) RE-68002
27
16. Relationships between Brix(%) Values and
Refractive Index (nD) Values
The relationships between Brix(%) values and Refractive Index (nD) values are listed in
this table for your reference.
m Refractive Index values correlating to Brix 0 to 85% in the table above have been
officially determined by ICUMSA (International Committee of Uniform Method of
Sugar Analysis held in 1974).
28
17. Specifications
Cat.No 3713 CM-BASEα
Measurement scale Brix(%) (Automatic Temperature Compensation)
Measurement range Brix 0.0 to 53.0% (indication -2.0 to 53.5)
Minimum indication Brix 0.1%
Measurement accuracy Brix ±0.5% (for measurement value of Brix 0.0 to 53.0%)
Measurement temperatures 10 to 50˚C
Temperature accuracy ±2.0˚C
Measurement Interval
Approx. 2 seconds per measurement .
(Same as output interval)
Temperature correction
Based on the temperature correction table for sucrose.
values
Brix -2.0 to 53.5% = DC4 to 20mA
Output
RS-232C output: Brix and temperature
Source LED (D-line approximation)
Factory Minimum width
Setting range of setting range
settings
Lower Limit Brix Brix
(ALL) -2.0% -2.0 to 92.5% Brix
Alarm Output 0.1%
Upper Limit Brix Brix
(ALH) 93.5% -1.0 to 93.5%
Delay Time 0 second 0 to 99 seconds 1 second
(ALd)
Temperature sensor Thin film platinum sensor
Materials in contact with the Prism: Sapphire
solution Prism stage : SUS304
Resistible pressure on the
0.98MPa(10kgf/cm 2)
prism unit
International Protection
IP67
Class
Power supply DC24V (Allowable fluctuation is ±10%)
Power consumption 0.6VA
・Use the instrument at an altitude below 5,000m (above
sea level).
・Use the instrument indoors.
・Use the instrument where the temperature is between 5
to 40˚C.
Environmental conditions ・Use the instrument under the condition where humidity
is 80% at 31˚C or lower, falling linearly to 50% at 40˚C.
・Main supply voltage fluctuation should not to exceed
±10% the nominal voltage.
・Installation categories (Overvoltage Categories):ll
・The pollution degree is 2 (according to IEC60664).
29
CM-BASEα Special Scale Series
Model CM-BASEα-03S CM-BASEα-04S CM-BASEα-05S CM-BASEα-06S
Cat.No. 3783 3784 3785 3786
Measurement Salinity of
Salinity Salinity (S.G.) Salinity (Baume)
items Seawater
Measurement
0.0 to 28.0% 1.000 to 1.217 0.0 to 25.7° 0 to 100‰
range
Minimum
0.1% 0.001 0.1° 1‰
indication
Measurement
±0.5% ±0.005 ±0.5° ±5‰
accuracy
Temperature
10.0 to 50.0˚C 10.0 to 40.0˚C 10.0 to 40.0˚C 10.0 to 40.0˚C
compensation
Indication range -1.5 to 28.5% 1.000 to 1.230 -1.5 to 26.2° -15 to 100‰
Indication
temperature 9.5 to 50.5˚C 9.5to 40.5˚C 9.5 to 40.5˚C 9.5 to 40.5˚C
range
Bias Adjustment ±2.0% ±0.020 ±2.0° ±20‰
Factory settings Factory settings Factory settings Factory settings
ALL:-2.0% ALL:1.000 ALL:-1.5% ALL:-15‰
ALH:28.5% ALH:1.230 ALH:26.2% ALH:100‰
ALd:0 second ALd:0 second ALd:0 second ALd:0 second
Setting range Setting range Setting range Setting range
Alarm Output
ALL:-2.0 to 27.5% ALL:1.000 to 1.220 ALL:-1.5 to 25.2% ALL:-15 to 90‰
ALH:-1.0 to 28.5% ALH:1.010 to 1.230 ALH:-0.5 to 26.2% ALH:-5 to 100‰
Minimum width: Minimum width: Minimum width: Minimum width:
0.1% 0.001 0.1% 1‰
ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds
aaa,bb.b
aa.a,bb.b a.aaa,bb.b aa.a,bb.b
serial aaa :
aa.a : a.aaa : aa.a :
communication Salinity of
Salinity(%: W/W) Salinity (S.G.) Salinity (Baume)
Transmit data Seawater(‰)
bb.b : bb.b : bb.b :
(Fixed range) bb.b :
Temperature Temperature Temperature
Temperature
* Other specifications are the same as CM-BASEα (Cat.No.3713).
30
CM-BASEα Special Scale Series
Model CM-BASEα-07S CM-BASEα-08S CM-BASEα-27S CM-BASEα-29S
Cat.No. 3787 3788 3791 3792
Measurement Sea Water Concentration of
Sea Water (S.G.) Citric Acid
items (Baume) Soy milk
Measurement
1.000 to 1.070 0.0 to 10.0° 0.0 to 20.0% 0.0 to 10.0%
range
Minimum
0.001 0.1° 0.1% 0.1%
indication
Measurement
±0.005 ±0.5° ±0.5% ±0.5%
accuracy
Temperature
10.0 to 40.0˚C 10.0 to 40.0˚C 10.0 to 50.0˚C 10.0 to 40.0˚C
compensation
Indication range 1.000 to 1.075 -1.5 to 10.5° -1.5 to 20.5% -2.0 to 10.5%
Indication
temperature 9.5 to 40.5˚C 9.5 to 40.5˚C 9.5 to 50.5˚C 9.5 to 40.5˚C
range
Bias Adjustment ±0.020 ±2.0° ±2.0% ±2.0%
Factory settings Factory settings Factory settings Factory settings
ALL:1.000 ALL:-1.5° ALL:-1.5% ALL:-2.0%
ALH:1.075 ALH:10.5° ALH:20.5% ALH:10.5%
ALd:0 second ALd:0 second ALd:0 second ALd:0 second
Setting range Setting range Setting range Setting range
Alarm Output
ALL:1.000 to 1.065 ALL:-1.5 to 9.5° ALL:-1.5 to 19.5% ALL:-2.0 to 9.5%
ALH:1.010 to 1.075 ALH:-0.5 to 10.5° ALH:-0.5 to 20.5% ALH:-1.0 to 10.5%
Minimum width: Minimum width: Minimum width: Minimum width:
0.001 0.1° 0.1% 0.1%
ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds
aa.a,bb.b aa.a,bb.b aa.a,bb.b
a.aaa,bb.b
serial aa.a : aa.a : aa.a :
a.aaa :
communication Sea Water Concentration of Citric Acid
Sea Water (S.G.)
Transmit data (Baume) Soy milk (%: W/W)
bb.b :
(Fixed range) bb.b : bb.b : bb.b :
Temperature
Temperature Temperature Temperature
* Other specifications are the same as CM-BASEα (Cat.No.3713).
31
CM-BASEα Special Scale Series
Model CM-BASEα-30S CM-BASEα-38S CM-BASEα-39S CM-BASEα-40S
Cat.No. 3793 3794 3795 3796
Hydrogen
Measurement Dimethylformamid Sodium Hydroxide
Acetic acid Peroxide Solution
items e Concentration Concentration
Concentration
Measurement
0.0 to 20.0% 0.0 to 45.0% 0.0 to 50.0% 0.0 to 38.0%
range
Minimum
0.2% 0.2% 0.2% 0.1%
indication
Measurement
±1.0% ±1.0% ±1.4% ±0.5%
accuracy
Temperature
10.0 to 40.0˚C 10.0 to 40.0˚C 10.0 to 35.0˚C 10.0 to 40.0˚C
compensation
Indication range -2.0 to 20.6% -2.0 to 45.6% -3.0 to 51.0% -1.0 to 38.5%
Indication
temperature 9.5 to 40.5˚C 9.5 to 40.5˚C 9.5 to 35.5˚C 9.5 to 40.5˚C
range
Bias Adjustment ±2.0% ±2.0% ±2.0% ±2.0%
Factory settings Factory settings Factory settings Factory settings
ALL:-2.0% ALL:-2.0% ALL:-3.0% ALL:-1.0%
ALH:20.6% ALH:45.6% ALH:51.0% ALH:38.5%
ALd:0 second ALd:0 second ALd:0 second ALd:0 second
Setting range Setting range Setting range Setting range
Alarm Output
ALL:-2.0 to 19.6% ALL:-2.0 to 44.6% ALL:-3.0 to 50.0% ALL:-1.0 to 37.5%
ALH:-1.0 to 20.6% ALH:-1.0 to 45.6% ALH:-2.0 to 51.0% ALH:0.0 to 38.5%
Minimum width: Minimum width: Minimum width: Minimum width:
0.1% 0.1% 0.1% 0.1%
ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds
aa.a,bb.b
aa.a,bb.b aa.a,bb.b aa.a,bb.b
serial aa.a :
aa.a : aa.a : aa.a :
communication Acetic Acid
DMF (%: W/W) H2O2 (%: W/W) NaOH (%: W/W)
Transmit data (%: W/W)
bb.b : bb.b : bb.b :
(Fixed range) bb.b :
Temperature Temperature Temperature
Temperature
* Other specifications are the same as CM-BASEα (Cat.No.3713).
32
CM-BASEα Special Scale Series
CM-BASEα-
Model CM-BASEα-43S CM-BASEα-Urea CM-BASEα-101S
Dampening Water
Cat.No. 3797 3789 3798 3782
Measurement Magnesium Urea
Cutting Oil Dampening Water
items Chloride Concentration
Measurement
0.0 to 30.0% 0.0 to 50.0% 0.0 to 50.0% 0.0 to 50.0%
range
Minimum
0.1% 0.1% 0.1% 0.1%
indication
Measurement
±0.5% ±0.5% ±0.5% ±0.5%
accuracy
Temperature
10.0 to 40.0˚C 10.0 to 40.0˚C 10.0 to 50.0˚C 10.0 to 50.0˚C
compensation
Indication range -1.0 to 30.5% -2.0 to 50.5% -2.0 to 50.5% -2.0 to 50.5%
Indication
temperature 9.5 to 40.5˚C 9.5 to 40.5˚C 9.5 to 50.5˚C 9.5 to 50.5˚C
range
Bias Adjustment ±2.0% ±2.0% - ±2.0%
Factory settings Factory settings Factory settings Factory settings
ALL:-1.0% ALL:-2.0% ALL:-2.0% ALL:-2.0%
ALH:30.5% ALH:50.5% ALH:50.5% ALH:50.5%
ALd:0 second ALd:0 second ALd:0 second ALd:0 second
Setting range Setting range Setting range Setting range
Alarm Output
ALL:-1.0 to 29.5% ALL:-2.0 to 49.5% ALL:-2.0 to 49.5% ALL:-2.0 to 49.5%
ALH:0.0 to 30.5% ALH:-1.0 to 50.5% ALH:-1.0 to 50.5% ALH:-1.0 to 50.5%
Minimum width: Minimum width: Minimum width: Minimum width:
0.1% 0.1% 0.1% 0.1%
ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds ALd:0 to 99seconds
aa.a,bb.b
aa.a,bb.b aa.a,bb.b aa.a,bb.b
serial aa.a :
aa.a : aa.a : aa.a :
communication Magnesium
Urea Cutting Oil Dampening Water
Transmit data Chloride (%: W/W)
bb.b : bb.b : bb.b :
(Fixed range) bb.b :
Temperature Temperature Temperature
Temperature
* Other specifications are the same as CM-BASEα (Cat.No.3713).
33
RS-232C
Measurement Display Indicator Lamp DC
(Depends on the product)
Below the Red
LLL 4mA LLL/LL.L/L.LLL
indication range
within the Green
Measured value 4 to 20mA Measured value
indication range
Above the Red
HHH 20mA HHH/HH.H/H.HHH
indication range
34
18. Repair Service and Warranty Period
The CM-BASEα series is a precise electronic instrument which incorporates both optical
and electrical components. Due to the complex interaction of these components, repairs
and/or adjustments of the CM-BASEα series unit must be performed by an ATAGO
engineer or a properly trained service technician at an ATAGO Authorized Service Center.
Authorized Service Center technicians have completed maintenance courses and have a
vast knowledge of ATAGO instruments. Any simple inspection or replacement of parts
described in this manual can be performed by the end-user. Only ATAGO engineers and
properly trained service technicians are allowed to perform repairs or disassemble the
CM-BASEα series. Any attempt to make repairs or disassemble the unit will void the
warranty.
The warranty period of the CM-BASEα series is one year from the date of purchase. This
warranty covers manufacturer's defects. If any manufacturer's defect is found during the
warranty period, the CM-BASEα series will be repaired under warranty. The prism of the
CM-BASEα series is considered a consumable item and is not covered under the
warranty.
All instruments received for repair are subject to a possible inspection fee. If the unit is
inspected and found to be either in good working order or is not covered by the warranty,
the customer will be responsible for any inspection fees, repair costs including labor,
parts and materials use, and shipping charges.
35
19. ATAGO CO.,LTD. Service Centers
ATAGO has Authorized Service Centers around the world. Below is the list of countries where
you can find an ATAGO Authorized Service Center. If your ATAGO instrument requires
servicing please contact ATAGO at the following e-mail address.
[email protected]
Please provide your company name, address and telephone number so that we can direct
your inquiry to the Authorized Service Center nearest you. The Authorized Service Center in
your area will contact you within 1 to 2 business days.
36
Headquarters: The Front Tower Shiba Koen, 23rd Floor
2-6-3 Shiba-koen, Minato-ku, Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 FAX: 81-3-3431-1945
[email protected] https://www.atago.net/
TEL: 55 16 3913-8400
[email protected]
TEL: 39 02 36557267
[email protected]
TEL: 86-20-38108256
[email protected]
TEL: 7-812-777-96-96
[email protected]
TEL: 7-727-257-08-95
[email protected]