Review Article Impedance Imaging of Cells and Tissues: Design and Applications
Review Article Impedance Imaging of Cells and Tissues: Design and Applications
BME Frontiers
Volume 2022, Article ID 9857485, 21 pages
https://doi.org/10.34133/2022/9857485
Review Article
Impedance Imaging of Cells and Tissues: Design and Applications
Due to their label-free and noninvasive nature, impedance measurements have attracted increasing interest in biological research.
Advances in microfabrication and integrated-circuit technology have opened a route to using large-scale microelectrode arrays for
real-time, high-spatiotemporal-resolution impedance measurements of biological samples. In this review, we discuss different
methods and applications of measuring impedance for cell and tissue analysis with a focus on impedance imaging with
microelectrode arrays in in vitro applications. We first introduce how electrode configurations and the frequency range of the
impedance analysis determine the information that can be extracted. We then delve into relevant circuit topologies that can be
used to implement impedance measurements and their characteristic features, such as resolution and data-acquisition time.
Afterwards, we detail design considerations for the implementation of new impedance-imaging devices. We conclude by
discussing future fields of application of impedance imaging in biomedical research, in particular applications where optical
imaging is not possible, such as monitoring of ex vivo tissue slices or microelectrode-based brain implants.
1. Introduction trode arrays (MEAs) [9] and their use for impedance analy-
sis enabled high-sensitivity and spatially highly resolved
Recent progress in in vitro cellular and molecular analysis measurements at subcellular resolution [10–13]. The label-
has greatly advanced our understanding of human physiol- free and high-resolution nature of impedance detection
ogy in healthy and diseased states. Owing to their label- offered by high-density MEAs has enabled to produce 2D
free and noninvasive nature, impedance-based detection impedance images of biological samples to, e.g., monitor
methods have been used for quantitative long-term charac- how cell attachment and mobility of adherent cells are
terization of live cells and tissues. In comparison to other affected by drug treatment [11] or to recognize the spatial
routine analysis methods, such as optical imaging, imped- organization of cells in live, ex vivo brain slices [12].
ance measurements typically require less costly equipment, In this review article, we first present different examples of
the respective setups can be miniaturized, and experiments how impedance measurements and impedance imaging is
can be parallelized [1, 2]. applied in neuroscience and biomedical research. We discuss
First reports on the use of impedance-based detection different methods that have been introduced for the character-
methods in biology date back to the 1920s, when impedance ization of cell and tissue models in vitro, with a particular focus
measurements were used to characterize membrane capaci- on methods and approaches for two- and three-dimensional
tance and resistance of blood cells [3, 4]. Technical improve- (2D and 3D) impedance-based imaging, and we present
ments and the establishment of theoretical models have then examples of impedance imaging in biomedical research. We
enabled the use of impedance methods for the analysis of cell then discuss the different detection methods that are used
volumes in suspension [5], for discerning different cell pop- for impedance analysis as well as related circuitry implementa-
ulations [5], and for monitoring cellular adhesion and tions and discuss the advantages and limitations of the differ-
growth [6–8]. The development of high-density microelec- ent approaches. Afterwards, we summarize the design options
2 BME Frontiers
that need to be considered during the conceptualization of mittivity, which is associated with its ability to store energy.
impedance-based MEA sensors for different applications. As the energy of an applied electric field can either be dissi-
Finally, we end the review with an outlook on trends in devel- pated or stored, the impedance of a dielectric material is
oping MEAs for impedance-based imaging of biological directly related to its permittivity and can be expressed as
entities. [17]
Electrode
Zel
108
Cell 𝛼
Permitivity
106
Cc Rc
104 𝛽
Rsol
102 𝛿
Cm Rm
𝛾
1
Zel 1 102 104 106 108 1010
Electrode Frequency (Hz)
(a) (b)
Figure 1: (a) Electrical equivalent-circuit model of a cell in suspension between a pair of electrodes. The overall measured impedance
includes three contributions: the cell impedance, which consists of the cytoplasm and membrane capacitive and resistive contributions
(C c , Rc and C m , Rm ), the electrode impedances (Z el ), and the medium resistance (Rsol ). (b) Idealized spectrum of the relative permittivity
of cells and tissues, showing four main dispersion regions [18, 19]; for details, see text.
|Z|
|Z|
f
fLF fHF
f Cell type 1
Cell type 2
𝜃Z Cell type 3
Electrical opacity
(|ZHF|/|ZLF|)
f
fm Electrical volume |ZLF|
Cell type 1
Cell type 2
Cell type 3
(a) (b)
positions influence the electric-field distribution and how reflects the voltage drop across the sample, while the influ-
electrical currents flow across the sample. For coplanar elec- ence of electrode impedance is eliminated. The location of
trodes, as shown in Figure 3(a), the electric field decreases the voltage-sensing electrodes needs to be carefully chosen,
with increasing distance to the electrode plane, and most in particular for large samples, as the measured voltage drop
of the current tend to flow along the surface of the electrode may otherwise not represent the average voltage drop across
plane. Therefore, the measured impedance is significantly the sample.
altered when a sample is placed on the plane in between Electric impedance tomography (EIT) is an impedance-
the electrodes. In contrast, if the same sample is located far based imaging technique, which relies on measurements of
from the electrode plane, it only marginally affects the sample impedance at different positions to investigate the
impedance measured between the electrodes. This internal properties of a sample [38, 39] (Figure 3(c)). A
position-dependence can be avoided by using facing elec- reconstruction algorithm, frequently based on a finite-
trodes, i.e., electrodes placed on opposing planes, which fea- element model, is used to infer the internal impedance dis-
ture uniform electric-field distribution, as in parallel plate tribution that corresponds to the experimentally determined
capacitors. However, the realization of facing electrodes is measurement values. EIT reconstruction is often an ill-posed
more difficult due to a more complex fabrication, and the problem, and algorithms have limitations concerning the
required alignment of the two electrodes. complexity of the reconstructed image. Reconstruction typi-
Regardless of the electrode configuration, the electrode- cally requires a priori information about the expected prop-
electrolyte-interface impedances add in series to the sample erties of the sample [40].
impedance, which may obstruct sample characterization at Electrode arrays allow for both, multiple two-terminal
specific frequencies, where electrode contributions domi- and multielectrode impedance measurements (Figure 3(d)).
nate. To circumvent this limitation, a multiple-electrode Two-terminal measurements can be taken between two elec-
arrangement can be used to reduce or even eliminate the trodes in the array [41, 42], or several electrodes can be
effects of the electrode impedance [29, 37]. In four- grouped to obtain larger pixels to enable multielectrode or
electrode measurements, the two outer electrodes (one per differential measurements [43]. Alternatively, one of the ter-
terminal) provide the stimulation current, while two inner minals can be used as a common reference electrode to per-
electrodes sense the voltage drop across the sample form parallel measurements [10, 44].
(Figure 3(b)). Voltage measurements require that the sens-
ing electrodes do not carry any current so that no voltage 2.2.3. Absolute and Relative Measurements. Absolute imped-
drop occurs at the electrodes. The measured signal at the ance measurements enable a quantitative characterization of
inner electrodes in an arrangement shown in Figure 3(b) the sample features of interest. However, absolute
BME Frontiers 5
I + V – I I I
V1 V2 V3 V4 V5 V6
(a) (c)
+ V – + V1 – + V2 – + V3 –
I I I1 I1 I2 I2 I3 I3
(b) (d)
Figure 3: Illustration of impedance measurement methods using a variable number of coplanar electrodes. (a) Two-electrode setup, in
which a stimulation current (I) is applied through the same electrodes that are used for voltage measurement (V). Note that voltage
stimulation and current measurements are also possible. (b) Four-electrode setup, using one outer pair of electrodes for current
stimulation and a second inner pair of electrodes for voltage measurements. (c) EIT uses two (outermost) electrodes of a microelectrode
array for current stimulation, while the other electrodes are used to simultaneously measure the voltage at different positions of the
sample. (d) Microelectrode array configured for several two-electrode measurements in parallel.
To increase the signal-to-noise ratio, impedance cytom- MEAs for real-time imaging of adherent cells is the work of
etry is often carried out using a differential detection scheme Laborde et al., where the authors presented a CMOS high-
by employing, e.g., a three-electrode setup [53]. Differential density microelectrode array (HD-MEA) with 65,536 elec-
measurements enable to directly measure the dielectric trodes of 90 nm radius on a 0:6 μm × 0:89 μm grid [11]. To
properties of the sample against the suspension medium increase the detection range to a few hundreds of microns
and to remove any effect caused by drifts in electrode perfor- from the sensor surface, the authors used a 50 MHz detection
mance, medium evaporation, or temperature variations. frequency to overcome the screening effect of the electrical
Owing to the relatively high throughput of impedance double layer (EDL) at the electrode interface (Figure 4(a)).
cytometry (~100-1000 cells per minute), this technique can While this approach enables to extend the sensing region of
be used to provide a snapshot of the sample condition at a the electrodes while reducing the impedance of the double-
defined time point and has been used to, e.g., identify cells layer capacitance at the sensing frequency, measuring only
in mixed cell populations [21, 58], recognize differentiated the capacitive contribution makes it impossible to perform
mesenchymal stem cells [59], measure the proportion of impedance spectroscopy for cell characterization and differen-
activated platelets in whole blood [60], and discriminate tiation. Nevertheless, the developed system enabled to follow,
activated T-cells [61]. Conversely, sample confinement in in real-time and at subcellular resolution, the adhesion,
the active area is a prerequisite to continuously monitor spreading, and dynamic attachment of cancer cells, which
dynamic processes on the same sample or specimen. As an were cultured on the sensor surface (Figure 4(b)).
example, hydrodynamic trapping was employed to capture A primary use of MEAs includes localized voltage
single yeast cells to follow yeast growth and budding by recordings, for which the systems have been used to perform
monitoring impedance variations over a large range of fre- long-term measurements of intra- and extracellular electrical
quencies [54]. Impedance characterization has also been potentials of electrogenic cells, in particular of neurons and car-
x
0.32
(i) (ii)
Potential (a.u.)
0.30
Cexp (fF)
0.5 0.28 (iii)
10 𝜇m (iv)
50 MHz
0.26
0.24
0 10 15
t (min)
0.0
(a) (b)
Figure 4: Impedance-imaging with HD-MEAs for the detection of cellular adhesion and micromotion. (a) Simulated spatial distribution of
the electrical potential in 150 mM salt solution at 10 kHz and 50 MHz stimulation frequencies. The fields are simulated for a 90 nm diameter
Au electrode and the presence of a 4.4 μm polystyrene bead (dashed line) positioned on the array. At low frequencies, the electric field only
interacts with the bead when the bead is located within ~10 nm distance from the electrode, due to the screening effect of the electrode-
electrolyte double layer. By increasing the stimulation frequency to 50 MHz, the electric field extends further into the solution. (b)
growth of EBs on the array for five days. After five days of 1 MHz. Impedance recordings at 1 kHz stimulation fre-
spreading and differentiation on chip, extracellular voltage quency were used to evaluate the adhesion and growth of
recordings were used to measure the spontaneous beating of primary hippocampal neurons on the chip (Figure 5(b))
cells that had differentiated into cardiomyocytes [12]. The [13]. The combination of voltage and impedance recordings
HD-MEA system was also used to acquire an impedance provided information on cell distribution to assess whether
image of an acute mouse cerebellar slice and enabled to iden- the absence of electrical activity in specific regions on the
tify four different regions in the slice, namely, (i) white matter, sensor surface corresponded to areas devoid of cells or
(ii) the granular cell layer, (iii) the Purkinje cell layer with large whether the regions were populated by cells without electri-
electrophysiological activity, and (iv) the molecular layer that cal activity. The impedance recording of the whole array at a
contained the dendritic trees of the Purkinje cells single frequency could be acquired in ~2 minutes, while
(Figure 5(a)). Impedance imaging enabled a label-free detec- imaging of the whole sensor surface with a confocal micro-
tion of different cell layers in the cerebellar slice, which could scope required ~30 minutes. The system enabled high spatial
be recognized by their different impedance signatures (i.e., resolution within the 480 × 480 μm2 sensing areas owing to
magnitude and phase). To perform a complete impedance the small electrode pitch of 15 μm. However, the spatial
scan of the whole sensing area, multiple sequential measure- arrangement of the electrode clusters, which featured a
ments were required due to the large number of active elec- 1020 μm intercluster distance, was optimized for multiwell
trodes and the limited number of impedance-detection packaging, which obviated to perform high-resolution elec-
circuitry modules. Although recording at medium/high fre- trical detection over large areas, e.g., for large tissue slices.
quencies could be performed within a few minutes, low- Finally, MEAs can be used to perform localized electro-
frequency measurements required an acquisition time of ~1 chemical measurements, such as amperometry and voltamm-
hour, which is comparable to the time required for confocal etry, to provide real-time two-dimensional electrochemical
imaging. Impedance imaging still offers the advantage to imaging [44, 73–75]. As an example, a HD-MEA with
record in real time from live cell and tissue models, as there 17:5 × 15 μm2 interdigitated electrodes was used to detect
is no risk of potential damages by phototoxic effects during the spatiotemporal characteristics of neurotransmitter release
long-term optical and fluorescence imaging and as impedance (norepinephrine, epinephrine, and dopamine) in acute
imaging is label-free. murine adrenal-tissue slices upon stimulation with caffeine
Lopez et al. recently presented a multifunctional HD- [75]. A microfluidic channel was used to achieve temporally
MEA featuring 16,384 electrodes, which were arranged in defined chemical stimulation of the tissue slices so that the
16 active areas of 1024 electrodes each at an electrode pitch neurotransmitter release could be measured at high spatial
of 15 μm [72]. The system featured current and voltage stim- and temporal resolution. Electrochemical measurements
ulation, intracellular and extracellular voltage recordings, were also used to image the secretion of metabolites in bacte-
and two modes for impedance measurement: fixed fre- rial films of P. aeruginosa [74]. Interestingly, the use of
quency (1 and 10 kHz) measurements for fast impedance potential-sweep methods (square-wave voltammetry), as
monitoring (0.1-1 ms temporal resolution), and impedance opposed to fixed-potential detection, enabled to detect the
spectroscopy in a frequency range between 10 Hz and presence and concentration of multiple phenazine
8 BME Frontiers
Array y (mm)
GCL
WM
GCL
Electrophysiology recording units (1024 ch.) PCL
1.6
5
ML ML
0.8
Impedance (4) Impedance (4) 1 mm
0
Waveform gen.
Impedance (4)
Electrode array Impedance (4) 0 1 2 3 4
Digital ctrl
Impedance (4) Impedance (4)
Array x (mm)
50 𝜇m
1 mm
Impedance (4) Impedance (4)
Electrophysiological activity [Hz] Impedance phase (°)
Ref. el. 2.4 100 GCL
40
PCL
Array y (mm)
1.6
Electrophysiology recording units (1024 ch.) 50 0
0.8 ML
0 –30
0 1 2 3 4 0 1 2 3 4
Array x (mm) Array x (mm)
(a)
1 2 5 6
512 channels 512 channels
32 MUXs 3 4 7 8 32 MUXs
32 ADCs 32 ADCs
digital control 9 10 13 14 digital control Voltage (𝜇V)
11 12 15 16
0 100 200 300 400 500 600
32 stim. units
16 pixels
1 pixel
E1 E2
16 pixels
E3 E4
|Z| variation
15 𝜇m
0 0.2 0.4 0.6 0.8
(b)
Figure 5: Multifunctional HD-MEAs. (a) A micrograph of the HD-MEA chip with the main functional blocks highlighted in the picture
(left). The array featured 59,760 platinum electrodes at a 13.5 μm pitch. The HD-MEA was used for impedance imaging and
electrophysiological recordings of an acute brain slice (right). The electrical recordings are compared to a micrograph of the slice;
adapted with permission from Viswam et al., IEEE Trans. Biomed. Circuits Syst. 2018 (12). (b) HD-MEA with 16 active areas, each of
which includes 1024 TiN electrodes. The main functional blocks are indicated in the chip schematic, including 64 stimulation units and
1024 voltage-recording units, 64 multiplexers (MUX), and 64 analog-to-digital converters (ADC). Impedance recordings were used to
reconstruct the spatial distribution of primary neurons that were cultured on chip and to correlate their positions with their
electrophysiological activity (right); adapted with permission from Miccoli et al., Front. Neurosci. 2019 (13)–Creative Commons
Attribution License (CC BY).
metabolites released by the bacterial film. This MEA allowed uses a set of electrodes, placed at specific locations of the
for monitoring the gradient distribution of metabolite secre- sample, to inject small alternating currents and to measure
tion over a large sensing area of 8 × 8 mm2 at 225 μm spatial the resulting voltages [76]. Currently, EIT is used in medi-
resolution, which corresponds to the electrode pitch. Up to cine and industry, for example, for label-free and continuous
38 electrodes could be read out in parallel, and each potential monitoring of patient respiratory parameters during
sweep was carried out in 0.2 seconds. Impedance imaging can mechanical ventilation [77, 78], while its use for in vitro bio-
be used to provide information that is complementary to logical applications has also been explored [76, 79–81].
functional analysis through voltage recordings or electro- A first example of EIT application in pharmaceutical
chemical imaging for an in-depth analysis of cell and tissue research included its use to monitor the dissolution of phar-
models in vitro. maceutical tablets [76]. 80 electrodes were integrated within
Electrode impedance tomography (EIT) is a technique to a test vessel, and the dissolution of sodium chloride tablets in
generate 2D or 3D impedance-based images of a sample. EIT water was monitored in real time by detecting local
BME Frontiers 9
0.53
0.318
0.106
–0.106
GND
–0.318
Chamber –0.53
Common ground
(A) (B) (C) (D)
+ Imaging
software
1
10 mm
15 mm 0.9
Correlation coefficient
0.8
0.7
0.6
0.5
15 mm ID Socket to EIT
0.4
system 0.3
2 6
1 5 0.2
0.1
0
0.07 0.16 0.28 0.44 0.64 0.87 1.14 1.44 1.78 2.15 2.56 3.00
Area ratio between the sensor and the spheroid (%)
(a) (b)
Figure 6: EIT sensor chip. (a) Schematics and picture of a planar EIT sensor at the bottom of a 15 mm diameter tissue-culture well. (b) The
sensor enabled to reconstruct the 3D shape of microtissues of different size positioned within the well (top). The correlation between the
EIT-reconstructed image and microscopy image for each microtissue size was reported (bottom); adapted with permission from Wu
et al., Analyst 2018-Published by The Royal Society of Chemistry [79].
VDD IS IS-IR
Rs VOut
+ Is ADC
IS VDD IR
– + Vs –
CE Фc
Vin IR
WE
Фc
RE WE
RE Фd CR Фd Cint Фd
(a) (d)
RE R
LPF
VI
Vin WE Vc VOut
RS
WE
TIA
90° LPF
VQ RE
Ccell
(b)
BPF
RE VOut1
ADC t
(c) (e)
Figure 7: Simplified schematics of the circuit topologies commonly used for impedance sensing. (a) Potentiostat-based impedance sensing.
The voltage difference between reference (RE) and working electrode (WE) is controlled by an operational amplifier, which injects a current
through the counter electrode (CE). The injected current that is required to keep the sample voltage equal to the stimulation voltage (V in )
depends on the sample impedance. The injected current can be measured by using a known resistance (Rs ) in series with the WE and by
measuring the potential drop (V s ) across it. (b) Lock-in detection. A stimulation voltage is applied to the sample, and the resulting
current is converted to a voltage using a TIA. The TIA output is multiplied with in-phase and quadrature-phase reference signals (RS) to
demodulate the signal at the frequency of interest. The demodulated signals are then low-pass filtered to eliminate out-of-target-
frequency noise and to calculate the signal amplitude and phase. (c) Simultaneous multiple-frequency detection. The applied stimulation
signal can be a square pulse or a linear combination of multiple sinusoidal waveforms. The response of the sample is measured with a
wide-range amplifier, which is then followed by multiple band-pass filters (BPFs), and digitized by a set of analog-to-digital converters
(ADCs) in parallel to simultaneously measure the sample response at multiple frequencies. (d) Charge-based capacitive sensing. The
sample capacitance and a reference capacitor (C R ) are charged to a constant voltage (V DD ) using two switches (Φc ), while measuring the
current flowing through each capacitor (I S and I R , respectively). The difference between these two currents is injected into a third
capacitor, which acts as an integrator (C int ). The resulting voltage is proportional to the difference between sample and reference
capacitances. After digitizing the output voltage, all capacitors are discharged (Φd ). (e) Capacitance-to-frequency conversion based on a
relaxation oscillator. The sample acts as a capacitor, which is repeatedly charged and discharged between two threshold voltages (V Lo
and V Hi ) by a comparator with hysteresis and a feedback resistor (R). The oscillation period is proportional to the time constant of the
resulting RC circuit, which depends on the sample capacitance.
voltage and the measured current, i.e., by using a known high dynamic range, and suitability for electrochemical mea-
resistor in series to the sample. surements. This detection method is not common in
Goikoetxea et al. designed an impedance measurement CMOS integrated circuits for impedance detection, as it
system, which featured a 20-channel, passive MEA with typically requires high circuit complexity and large silicon
TiN electrodes [87], where the electrode impedance was area and potentially entails high power consumption, all
monitored with a benchtop potentiostat by sequentially mul- features that ultimately limit parallelizability and large-
tiplexing between the electrodes. Zhang et al. presented a scale integration.
single-cell impedance measurement system with 128 passive
electrodes that used dielectric forces to locate and drive the 3.2. Lock-In Amplifiers. A lock-in amplifier enables to extract
cells on top of the measurement electrodes [88]. Moreover, low-amplitude signals with known periodicity from a noisy
the authors used an external potentiostat-based impedance background and to measure single-frequency signals with
analyzer to carry out the impedance recordings. remarkable accuracy. Lock-in detection is carried out by
Potentiostat-based detection is commonly used in com- multiplying the readout signal with in-phase and quadrature
mercial equipment owing to its wide frequency spectrum, (i.e., 90°-shifted) reference periodic signals, the periodicity of
BME Frontiers 11
which equals that of the signal of interest. The multiplication range. A similar sequential approach was also used in the
results in a demodulation of the signal information at low integrated HD-MEA, developed by Lopez et al. [72]. How-
frequencies, typically referred to as in-phase/quadrature (I/ ever, here, the authors used square-wave current-stimulation
Q) demodulation, where undesired components and out- signals, while the cell impedance was measured by detecting
of-target-frequency noise can be easily eliminated using a voltages at the sensing electrodes. This impedance measure-
low-pass filter. By multiplying with both, the in-phase and ment structure takes advantage of current generators and
quadrature reference signals, lock-in detection enables to amplifiers that were already implemented in the HD-MEA
measure the real and imaginary component of the signal for cellular stimulation and voltage recording, which resulted
[89, 90]. The same periodic signal can be used for stimula- in a compact and power-efficient design. Electrode voltage
tion and for generation of the reference signals for lock-in recording was performed by using an amplifier, equipped
demodulation, which enables to measure the absolute mag- with choppers, which allowed to demodulate the impedance
nitude and phase delay of the detected signal (Figure 7(b)) signal to low frequencies (<10 kHz) and to subsequently dig-
[10]. Typically, a sinusoidal voltage stimulation is generated itize those on-chip. The system featured 64 channels, which
by the lock-in amplifier. The current flowing through the test could be configured to provide either the in-phase or
system is converted into a voltage using a TIA before feeding it quadrature-phase signals so as to calculate the real and imag-
back to the lock-in detector for multiplication with the refer- inary parts of the sample impedance off-chip. The spatial res-
ence signals and subsequent low-pass filtering. Impedance olution of the system was limited by the multiplexed structure
values can then be extracted by calculating the ratio of the of the array, on which electrodes were arranged in pixels with
applied stimulation voltage and the measured current. four electrodes per pixel. While the electrode pitch was
Lock-in detection is widely used in active CMOS-MEAs 15 μm, only one of the four electrodes in each pixel could
for obtaining impedance recordings of cells on electrode be recorded at a given time.
could be obtained from a single measurement, postproces- circuit used the CBCM technique to observe the capacitance
sing of the signals to extract the impedance values compli- change on the electrodes and to obtain real-time capacitance
cated data acquisition. Furthermore, since the stimulus images of cells. Finally, another example of capacitive sensing
signal contained power at multiple frequencies, the power for impedance analysis was described by Prakash et al. [100].
spectral density at each frequency inevitably was low, which The authors presented a charge-sharing-based capacitive sen-
rendered the output signal more sensitive to noise. The accu- sor for tracking cell adhesion and cell health. The authors
racy of the measurement may be even more decreased dur- could differentiate healthy and diseased cells by detecting the
ing postprocessing due to the crosstalk between signals at differences in cell adherence to the electrode surface. The
different frequencies. stronger adherence of healthy cells led to a higher capacitive
Hamilton et al. proposed a cell impedance sensor based coupling at the electrode, which could be detected by charging
on a “silicon cochlea” to provide simultaneous impedance and discharging the electrode capacitance.
sensing over a wide frequency range (from Hz to MHz) These examples show how capacitive coupling can be
[95]. The simulated circuitry included both, a multiple- used to obtain high-resolution electrical imaging of cells on
frequency signal generator (current stimulator) and an ana- electrode arrays and information on cell adherence. How-
lyzer part. In this implementation, the voltage of a sensing ever, the use of high-frequency detection limits the amount
electrode was fed to a voltage-to-current converter, and the of information that can be extracted with this technique.
corresponding current was split up into components at dif-
ferent frequencies with cascaded low-pass filters. However, 3.5. Impedance-to-Frequency Conversion. With the previous
the multiple analog outputs of the circuit require digitaliza- methods, impedance is quantified by measuring the current
tion for each frequency output of the electrodes, which mas- through (or voltage across) a sample to which a stimulus is
sively increases the complexity of the detection system. applied. However, impedance can also be estimated by inte-
Table 1: Applications, circuitry architectures, and design options of some representative MEA-based impedance measurement systems.
Number Electrode
Frequency Electrode Electrode Circuitry MEA
Application of pitch
range (Hz) size material architecture type
electrodes (μm)
Potentiostat-
Liu et al. 2009 Detection of cell adhesion, spreading,
1-1 M 100 80 μm∗ Pt 200 based Passive
[120] and proliferation
sensing
Impedance-
Mucha et al. Detection of cell reactions and 55 × 55
— 64 Au — to-frequency Active
2011 [121] adhesion μm2
converter
Potentiostat-
Mamouni et al. Detection of cell adhesion, spreading, 15 × 15
10-1 M 50 Au — based Passive
2011 [122] and proliferation μm2†
sensing
Impedance imaging of cells, detection
Widdershoven Capacitive
of dynamic attachment, and cellular 1 M–70 M 65,536 180 nm∗ Au 0:6 × 0:89 Active
et al. 2015 [99] sensing
micromotion
Detection of cell adhesion,
Dragas et al. 3 × 7:5 Lock-in
differentiation, and spreading, 1-1 M 59,760 Pt black 13.5 Active
2018 [44] μm2 detection
impedance imaging of tissue slices
2:5 × 3:5
MEA system for impedance detection. These factors include increase of the electrode impedance as a consequence of
the geometry and material of the electrodes, the MEA topol- the smaller surface area and the smaller electrical double-
ogy, signal digitalization, and multiplexing approaches, as layer capacitance at the electrode-electrolyte interface [50].
well as sensitivity and noise characteristics. The different The effects of a high initial impedance of small electrodes
design options will be discussed in the following sections. may superimpose to or even obscure cell-dependent varia-
We then summarize a variety of representative solutions that tions at low frequencies, in particular, when the electrode
have been reported in literature in a table which compares diameter is below ~50 μm. For such small electrodes, the
different approaches to designing MEA or HD-MEA sys- electrode impedance in physiological solutions can be in
tems (Table 1). the MΩ range for frequencies up to ~10 kHz, if no surface
modification is applied to increase the effective electrode
4.1. Electrode Geometries. Electrode size, the surface-area area [104]. The selection of a suitable electrode size strongly
ratio between reference and sensing electrodes, and the elec- depends on the application of interest. For example, elec-
trode pitch are key parameters that directly affect the signal trodes of hundreds of microns in diameter can be employed
quality, detection sensitivity, and lateral resolution. for monitoring the formation of a confluent cell layer [7, 50];
Cells sedimenting on or attachment to an electrode such large electrodes feature a large sensing area and allow
causes an increase of the electrode impedance, as they block for using simplified readout circuitry schemes, however
the electric currents through fractions of the electrode sur- strongly affecting the spatial resolution and signal sensitivity
face. The measured impedance increases as the ratio between attainable.
the electrode area and cell dimension decreases [103]. Gen- The electrode size and pitch directly define the spatial
erally, a decrease in electrode size or area results in an resolution that can be attained in impedance imaging. The
14 BME Frontiers
development of integrated MEAs towards HD-MEAs featur- measurements of different sensors. Finally, porous TiN coat-
ing large numbers of small electrodes with dimensions com- ings have also been used for HD-MEAs in in vitro and
parable to the size of cells at low pitch enabled them to attain in vivo applications, as this material has shown electrical
more detailed information including tissue morphology and performance comparable to that of Au and Pt and high sta-
spreading [11]. bility under physiological conditions [13, 109]. A recent
Impedance imaging is usually carried out in conjunction trend towards the use of 3D in vitro models has also fueled
with other types of measurements, such as voltage measure- the development of 3D-MEA structures. Metal-based elec-
ments, so that an electrode size needs to be found that yields trodes (typically Au and Pt) can be fabricated on different
sufficient sensitivity and acceptable noise characteristics for micro- and nanostructures, such as flexible polyimide pillars
all implemented detection methods. Simply decreasing elec- [110] or silicon-based shanks [111], to realize 3D electrode
trode size and increasing the electrode number to improve arrays, which then can be interfaced with standard, planar
spatial resolution results in complex readout architectures readout circuitry.
to handle the large number of electrodes or increased mea- Metal materials, however, are nontransparent and pre-
surement time in case of a limited number of readout units, vent optical access to the sample with inverted microscopes,
as well as higher microfabrication costs. which represents a major limitation for standard in vitro
Moreover, the counter electrode in three-electrode investigations of biological samples. Therefore, transparent,
setups or the reference electrode in two-electrode setups conductive materials, such as indium tin oxide (ITO) or
has to be much larger than the sensing or working elec- functionalized iridium oxide (IrO2), have been explored as
trodes to provide low-impedance paths for current injec- alternative electrode materials for passive MEAs on glass
tion [50]. In MEA architectures that use alternating array wafers to enable simultaneous optical and electrical charac-
electrodes as counter/reference electrodes for each sensing terization of cells in vitro [112–115].
CMOS technologies [11–13]. Therefore, the availability of of high-performance ADCs, both in terms of resolution and
CMOS HD-MEA technology is key for the realization of speed, to avoid losing information during signal digitaliza-
impedance-imaging systems, as the spatiotemporal resolution tion and multiplexing. However, complex signal processing,
of the imaging is determined by electrode numbers and size, such as signal filtering and multiplication, can be imple-
electrode pitch, and the achievable data acquisition rate. mented in the analog domain and the digitalization of the
processed signal can be done afterwards [12]. In such an
4.4. Signal Conditioning, Digitalization, and Multiplexing. As approach, the design challenges mostly concern the design
mentioned above, signal conditioning, multiplexing, and of analog processing circuits, however, with the benefit of
digitalization strategies form part of fundamental design relaxed requirements for the design of the ADCs.
considerations for CMOS HD-MEA-based impedance sen-
sors. Signal conditioning typically includes amplification 4.5. Sensitivity and Noise. The detection limit of a sensor is
and filtering to transform weak and noisy signals, detected determined by its signal-to-noise ratio (SNR), which defines
at the electrodes, into robust signals for processing in subse- the minimum signal that can be reliably detected given the
quent stages. The design of the amplifiers has to be done measurement noise. Here, we refer to measurement noise
with due regard to the expected input signals and the as any fluctuation in the recorded signal that is not consid-
dynamic range and resolution in subsequent stages. The gain ered a signal, i.e., is caused by a change in the sample. High
must be large enough to ensure that the smallest features of SNR requires high sensitivity, i.e., large variations of the out-
interest remain detectable during further processing. How- put signal upon small variations in the sample, and low noise
ever, extremely large gains may cause system saturation in levels, i.e., small signal variations that are not related to
the presence of large input signals. To be able to record input changes in the sample impedance.
signals with different amplitude ranges, amplifiers can fea- Maximizing the sensitivity of impedance measurements
downmodulated signal after multiplication can be carried analysis units, especially in case of multifrequency measure-
out using a dedicated low-pass filter [12]. Finally, differential ments. Therefore, digitalization and signal preprocessing,
schemes, whether applied during signal acquisition or post- such as Fast Fourier Transformations (FFTs), directly on-
processing, are an effective means to greatly reduce the effect chip will likely play a key role in the development of highly
of power-supply noise or environmental noise. Acquisition parallelized HD-MEAs as they offer the possibility to rapidly
of differential signals, however, requires more circuitry, acquire impedance recordings over a large spectral range at
therefore larger areas for implementation, which could high temporal resolution. However, the development of
impact electrode density or increase fabrication costs. While CMOS-MEAs requires—owing to design complexity and
differential values can often be calculated during data post- fabrication costs of integrated circuits—much larger time
processing, this solution may yield lower accuracy and less and financial investments in comparison to the fabrication
noise suppression than differential signal acquisition. of passive MEAs. Therefore, the development of custom-
designed CMOS-MEAs for specific applications can only
5. Outlook be justified by either unique functionality or performance
that cannot be achieved otherwise or by large numbers of
Impedance imaging is a real-time, label-free, and noninva- devices that can be produced at comparably low costs.
sive measurement method, which can be used for long- Finally, technological and theoretical improvements in
term characterization of cells and tissue models. The small electrical impedance tomography (EIT) may also increase
form factors of the sensor chips and the electronic nature the relevance of impedance techniques for studying biologi-
of the detection render this technique particularly suitable cal samples. Multifrequency impedance tomography (also
for parallelization and automation. Real-time monitoring known as electrical impedance tomography spectroscopy;
of cell and tissue dynamics can be performed very efficiently EITS) is a promising noninvasive technique to analyze
by the Swiss National Science Foundation under Contracts models of epithelial tissues,” Frontiers in Bioengineering and
205320_188910 and CR32I2_166329 and a Marie Heim- Biotechnology, vol. 7, 2020.
Vögtlin grant to R.B., and by an ETH Postdoctoral Fellow- [16] W. Franks, I. Schenker, P. Schmutz, and A. Hierlemann,
ship to F. C. “Impedance characterization and modeling of electrodes for
biomedical applications,” IEEE Transactions on Biomedical
Engineering, vol. 52, no. 7, pp. 1295–1302, 2005.
References [17] K. R. Foster and H. P. Schwan, “Dielectric properties of tis-
sues and biological materials: a critical review,” Critical
[1] A. Yúfera, G. Huertas, and A. Olmo, “A microscopy tech- Reviews in Biomedical Engineering, vol. 17, no. 1, pp. 25–
nique based on bio-impedance sensors,” Procedia Engineer- 104, 1989.
ing, vol. 47, pp. 1049–1052, 2012. [18] G. H. Markx and C. L. Davey, “The dielectric properties of
[2] Y. Xu, X. Xie, Y. Duan, L. Wang, Z. Cheng, and J. Cheng, “A biological cells at radiofrequencies: applications in biotech-
review of impedance measurements of whole cells,” Biosen- nology,” Enzyme and Microbial Technology, vol. 25, no. 3-5,
sors & Bioelectronics, vol. 77, pp. 824–836, 2016. pp. 161–171, 1999.
[3] H. Fricke, “The electrical capacity of suspensions with special [19] C. Gabriel, S. Gabriel, and E. Corthout, “The dielectric prop-
reference to blood,” The Journal of General Physiology, vol. 9, erties of biological tissues: I. Literature survey,” Physics in
no. 2, pp. 137–152, 1925. medicine & biology, vol. 41, no. 11, pp. 2231–2249, 1996.
[4] H. Fricke and S. Morse, “The electric resistance and capacity of [20] L. Y. Zhang, C. B. M. Du Puch, C. Dalmay et al., “Discrimina-
blood for frequencies between 800 and 4½ million cycles,” The tion of colorectal cancer cell lines using microwave biosen-
Journal of General Physiology, vol. 9, no. 2, pp. 153–167, 1925. sors,” Sensors and Actuators A: Physical, vol. 216, pp. 405–
[5] W. H. Coulter, Means for Counting Particles Suspended in a 416, 2014.
Spectroscopya,” Annals of the New York Academy of Sciences, [48] R. T. Kurulugama, D. O. Wipf, S. A. Takacs, S. Pongmayteegul,
vol. 873, pp. 51–58, 1999. P. A. Garris, and J. E. Baur, “Scanning electrochemical micros-
[32] M. Gheorghiu, E. Gersing, and E. Gheorghiu, “Quantitative copy of model neurons: constant distance imaging,” Analytical
analysis of impedance spectra of organs during ischemia,” Chemistry, vol. 77, no. 4, pp. 1111–1117, 2005.
Annals of the New York Academy of Sciences, vol. 873, no. 1 [49] C. R. Keese and I. Giaever, “Use of electric fields to monitor
ELECTRICAL BI, pp. 65–71, 1999. the dynamical aspect of cell behavior in tissue culture,” IEEE
[33] K. Darowicki and P. Ślepski, “Determination of electrode Transactions on Biomedical Engineering, vol. BME-33, no. 2,
impedance by means of exponential chirp signal,” Electro- pp. 242–247, 1986.
chemistry Communications, vol. 6, no. 9, pp. 898–902, 2004. [50] X. Zhang, W. Wang, A. N. Nordin, F. Li, S. Jang, and
[34] A. Mejía-Aguilar and R. Pallàs-Areny, “Electrical impedance I. Voiculescu, “The influence of the electrode dimension on
measurement using voltage/current pulse excitation,” in Pro- the detection sensitivity of electric cell-substrate impedance
ceedings XIX IMEKO World Congress: Fundamental and sensing (ECIS) and its mathematical modeling,” Sensors and
Applied Metrology, pp. 662–667, Lisbon, Portugal, 2009. Actuators B: Chemical, vol. 247, pp. 780–790, 2017.
[35] S. Gawad, T. Sun, N. G. Green, and H. Morgan, “Impedance [51] P. O. Bagnaninchi and N. Drummond, “Real-time label-free
spectroscopy using maximum length sequences: application monitoring of adipose-derived stem cell differentiation with
to single cell analysis,” The Review of Scientific Instruments, electric cell-substrate impedance sensing,” Proceedings of
vol. 78, no. 5, article 054301, 2007. the National Academy of Sciences of the United States of
[36] D. Rairigh, A. Mason, and C. Yang, “Analysis of on-chip America, vol. 108, no. 16, pp. 6462–6467, 2011.
impedance spectroscopy methodologies for sensor arrays,” [52] C. Petchakup, H. Li, and H. W. Hou, “Advances in single cell
Sensor Letters, vol. 4, no. 4, pp. 398–402, 2006. impedance cytometry for biomedical applications,” Micro-
[37] J. J. Ackmann, “Complex bioelectric impedance measure- machines, vol. 8, no. 3, p. 87, 2017.
counting,” Sensors and Actuators B: Chemical, vol. 181, [78] C. Putensen, B. Hentze, S. Muenster, and T. Muders, “Electri-
pp. 842–849, 2013. cal impedance tomography for cardio-pulmonary monitor-
[64] Y. Chen, C. C. Wong, T. S. Pui et al., “CMOS high density ing,” Journal of Clinical Medicine, vol. 8, no. 8, p. 1176, 2019.
electrical impedance biosensor array for tumor cell detec- [79] H. Wu, Y. Yang, P. O. Bagnaninchi, and J. Jia, “Electrical
tion,” Sensors and Actuators B: Chemical, vol. 173, pp. 903– impedance tomography for real-time and label-free cellular
907, 2012. viability assays of 3D tumour spheroids,” The Analyst,
[65] C. Begly, D. Ackart, J. Mylius, R. Basaraba, A. J. Chicco, and vol. 143, no. 17, pp. 4189–4198, 2018.
T. W. Chen, “Study of real-time spatial and temporal behav- [80] E. Lee, H. Wi, A. McEwan et al., “Design of a microscopic
ior of bacterial biofilms using 2-D impedance spectroscopy,” electrical impedance tomography system for 3D continuous
IEEE Transactions on Biomedical Circuits and Systems, non-destructive monitoring of tissue culture,” Biomedical
vol. 14, no. 5, pp. 1051–1064, 2020. Engineering Online, vol. 13, no. 1, p. 142, 2014.
[66] C. Stagni, C. Guiducci, L. Benini et al., “CMOS DNA sensor [81] Y. Yang, J. Jia, S. Smith, N. Jamil, W. Gamal, and P.-O. O.
array with integrated A/D conversion based on label-free Bagnaninchi, “A miniature electrical impedance tomogra-
capacitance measurement,” IEEE Journal of Solid-State Cir- phy sensor and 3-D image reconstruction for cell imag-
cuits, vol. 41, no. 12, pp. 2956–2964, 2006. ing,” IEEE Sensors Journal, vol. 17, no. 2, pp. 514–523,
[67] H. M. Jafari and R. Genov, “CMOS impedance spectrum ana- 2017.
lyzer with dual-slope multiplying ADC,” in 2011 IEEE Bio- [82] Z. Chen, Y. Yang, and P.-O. Bagnaninchi, “Hybrid learning-
medical Circuits and Systems Conference (BioCAS), pp. 361– based cell aggregate imaging with miniature electrical imped-
364, San Diego, CA, USA, 2011. ance tomography,” IEEE Transactions on Instrumentation
[68] F. Heer, W. Franks, A. Blau et al., “CMOS microelectrode and Measurement, vol. 70, pp. 1–10, 2021.
array for the monitoring of electrogenic cells,” Biosensors & [83] M. E. J. Obien, K. Deligkaris, T. Bullmann, D. J. Bakkum, and
[92] D. Jung, J. S. Park, G. V. Junek et al., “A 21952-pixel multi- ance sensing,” Sensors and Actuators B: Chemical, vol. 288,
modal CMOS cellular sensor array with 1568-pixel parallel pp. 707–715, 2019.
recording and 4-point impedance sensing,” in 2019 Sympo- [106] R. Bounik, M. Gusmaroli, P. M. Misun, V. Viswam,
sium on VLSI Circuits, pp. C62–C63, Kyoto, Japan, June A. Hierlemann, and M. M. Modena, “Integration of discrete
2019. sensors and microelectrode arrays into open microfluidic
[93] Y. Liu, E. W. Abel, J. J. F. Belch, and S. Chen, “An impulse hanging-drop networks,” in 2019 IEEE 32nd International
response flow cytometric technique for blood cell Conference on Micro Electro Mechanical Systems (MEMS),
characterisation-instrumentation and preliminary evalua- pp. 441–444, Seoul, Korea (South), 2019.
tion,” in Proceedings of the 20th Annual International Confer- [107] K. Cheung, S. Gawad, and P. Renaud, “Impedance spectros-
ence of the IEEE Engineering in Medicine and Biology Society. copy flow cytometry: on-chip label-free cell differentiation,”
Vol.20 Biomedical Engineering Towards the Year 2000 and Cytometry Part A, vol. 65, pp. 124–132, 2005.
Beyond (Cat. No.98CH36286), pp. 1881–1884, Hong Kong,
[108] V. Srinivasaraghavan, J. Strobl, D. Wang, J. R. Heflin, and
China, 1998.
M. Agah, “A comparative study of nano-scale coatings on gold
[94] R. Bragós, R. Blanco-Enrich, O. Casas, and J. Rosell, “Charac- electrodes for bioimpedance studies of breast cancer cells,” Bio-
terisation of dynamic biologic systems using multisine based medical Microdevices, vol. 16, no. 5, pp. 689–696, 2014.
impedance spectroscopy,” in Conference Record - IEEE
[109] J. J. Jun, N. A. Steinmetz, J. H. Siegle et al., “Fully integrated
Instrumentation and Measurement Technology Conference,
silicon probes for high-density recording of neural activity,”
vol. 1, pp. 44–47, Budapest, Hungary, 2001.
Nature, vol. 551, no. 7679, pp. 232–236, 2017.
[95] T. J. Hamilton, N. M. Nelson, D. Sander, and P. Abshire, “A
[110] D. A. Soscia, D. Lam, A. C. Tooker et al., “A flexible 3-
cell impedance sensor based on a silicon cochlea,” in 2009
dimensional microelectrode array for in vitro brain models,”
IEEE Biomedical Circuits and Systems Conference, pp. 117–
Lab on a Chip, vol. 20, no. 5, pp. 901–911, 2020.
120, Beijing, China, 2009.