SM300 Manual Version 1.3
SM300 Manual Version 1.3
SM300
Soil Moisture Sensor
SM300-UM-1.3
Notices
Copyright
All parts of the SM300 design and documentation are the exclusive
right of Delta-T Devices and covered under copyright law.
Copyright © 2014 Delta-T Devices Ltd.
Patents
The SM300 is protected under international law by the following
patents:-
USA: Patent US7944220
Europe: Patent EP1836483
Australia: Patent AU2005315407
China: Patent CN101080631
EMC Compliance
See page 33.
Design changes
Delta-T Devices Ltd reserves the right to change the designs and
specifications of its products at any time without prior notice.
Operation 10
Cable Connections 10
Installation 11
Logger connections and configuration 12
GP1 Logger 12
GP2 Logger Controller 13
DL6 Logger 14
DL2e Logger 15
Other data loggers 16
Logging Advice 16
Logger Grounding 17
HH2 Meter 20
Calibration 21
Soil calibration 21
Sensor calibration 24
Soil moisture reading 25
Troubleshooting 27
Technical Reference 29
Specifications 29
Volumetric water content 29
Temperature 29
Definitions 34
References 36
Technical Support 37
Appendix 1 39
SM300 User Manual 1.3 3
Soil-specific Calibration 39
Laboratory calibration for non-clay soils 40
Laboratory calibration for clay soils 43
Appendix 2: 46
The SM300 Temperature Sensor 46
SM300 Temperature Measurement 47
Effect of Temperature on Water Permittivity 49
Resistance to Temperature Lookup Table 50
Index 51
Description
The SM300 measures soil moisture content and temperature.
Its sealed plastic body is attached to two sensing rods which
insert directly into the soil for taking readings.
A waterproof plug connects to a choice of signal cables.
Both extension cables and extension tubes can be used.
The soil moisture output signal is a differential analogue DC
voltage. This is converted to soil moisture by a data logger or
meter using the supplied general soil calibrations.
It can also be calibrated for specific soils.
Features
Soil moisture accurate to ± 2.5%
Soil temperature to ± 0.5°C over 0-40°C
Low salinity sensitivity
Excellent stability
Minimal soil disturbance
Easy installation at depth in augured holes
Waterproof connector to IP68
Rugged, weatherproof and can be buried.
Good electrical immunity
Choice of cabling system options
Cable connector, cylindrical profile and extension tube
design simplifies removal for servicing.
See also Specifications on page 29
Cable connector
sealed to IP68
M12, 5 pin, male
12
Thread ¾inch BSP
for connecting to
92 mm
Extension Tube(s)
67 mm
51mm
22mm
40 mm
5m cable with
SMSC/lw-05 200mm flying leads
to connect to logger
Take care when attaching cables to ensure that the connectors are
clean, undamaged and properly aligned before pushing the parts
together.
If you feel strong resistance when inserting the SM300 into soil, it is
likely you have encountered a stone. Stop pushing and re-insert at a
new location.
Cable Connections
brown Power 0V
white Power V+
blue Signal HI
black Signal LO
grey Temperature +
green Cable shield
5m 5m with
0.9m 10m 200 mm bare
25m leads
Take care when attaching cables to ensure that the connectors are
clean, undamaged and properly aligned before pushing the parts
together.
Installing at depth
Make a 45mm diameter hole, preferably at
about 10° to the vertical using the
SM-AUG-100 auger.
Connect an extension tube e.g. ML/EX50
Push the SM300 into the soil until rods are
fully inserted. Ensure good soil contact.
Alternatively
Dig a trench, and install horizontally.
1
You need the PC logger software DeltaLINK version 2.4 or later. A free upgrade can
be obtained from www.delta-t.co.uk or from the Software and Manuals DVD.
For configuration details see the DeltaLINK2 software sensor Info Panel and
Help or the GP2 User Manual.
2
You need the PC logger software DeltaLINK version 3.0 or later. A free upgrade can
be obtained from www.delta-t.co.uk or from the Software and Manuals DVD.
3You need the PC logger software DeltaLINK version 2.4 or later. A free upgrade can
be obtained from www.delta-t.co.uk or from the Software and manuals DVD.
4
You need a PC running Ls2Win version 1.0 SR6 or later. A free upgrade can be
obtained from www.delta-t.co.uk or from the Software and manuals DVD.
Logging Advice
Allow 20mins for the temperature readings to stabilise after installation
Do not log faster than 1 minute to avoid SM300 self-heating, which
could affect the accuracy of temperature readings.
Figure 2 This diagrams shows that if the logger is not well grounded then a
potential ground loop can exist between the sensor and the power supply and/or
mains power, and also the risk of galvanic corrosion to buried M12 connectors on
the SM300 and cable.
Figure 3 This diagram shows the logger, power supply and sensor all grounded.
This minimises the possibility of galvanic corrosion of the SM300 connector. It also
minimises the possibility of nearby AC-powered machinery injecting noise into the
readings. The potential ground loop between logger and sensor is mitigated by the
differential wiring of the SM300
Logger grounded
Figure 4 This diagram shows a battery or high quality power supply with no
leakage current to ground. The logger is grounded. The potential ground loop
between logger and sensor is mitigated by the differential wiring of the SM300.
This is the best arrangement.
Make sure the HH2 is correctly configured for your soil type:
► At the start-up screen, press Set and scroll down to the Soil Type option.
► Press Set again and scroll down to the
appropriate soil type (use Mineral for sand,
Soil Type:
silt or clay soils or Organic for peaty soils) Mineral
► Press Set to confirm this choice.
6.0 6.0
4.0 4.0
√ε √ε
2.0 2.0
0.0 0.0
0% 20% 40% 60% 80% 0.0 0.2 0.4 0.6 0.8 1.0 1.2
θ (%vol) SM300 output (V)
Organic
60% Mineral
40%
θ
20%
0%
0.0 0.2 0.4 0.6 0.8 1.0 1.2
SM300 output (V)
a0 a1
where the coefficients a0 and a1 conveniently parameterise the
dielectric properties of soils.
Soil calibrations
8.0
Soil refractive index (√ε)
6.0
4.0
Slope (a1)
2.0
Offset (a0)
0.0
0.0 0.2 0.4 0.6 0.8
Soil moisture (m3.m-3)
Note that:
a0 dry _ soil is usually between 1.3 to 2.3
a1 corresponds approximately to water 1 and usually takes a
value about 8.0. Real soil values for a0 and a1 can vary significantly
from these guidelines when they are affected by other factors – in
particular, bound water in clay may result in higher values of a1.
a0 a1
Mineral soils 1.6 8.4
Organic soils 1.3 7.7
Mineral
6.0
Organic
4.0
2.0
0.0
0.0 0.2 0.4 0.6 0.8
3 -3
Soil Moisture (m .m )
These values have been used to generate the polynomial
conversions and linearisation tables in the Soil moisture reading
section.
Soil-specific calibration
Instead of adopting these general calibrations, you may wish to
determine specific calibration values of a0 and a1 for your soil. This
procedure is fairly straightforward if you can get access to standard
laboratory equipment and is described in detail in Appendix 1 on
page 38.
Soil specific calibration can significantly improve the accuracy of
individual readings - but make less of an improvement to readings
where installation and sampling errors are high.
6.0
4.0
2.0
0.0
0.0 0.2 0.4 0.6 0.8 1.0 1.2
SM300 output (Volts)
This response can be approximated either by a polynomial (below)
or by a linearisation table (see next page):
Polynomial (for use over the full range of SM300 readings)
√𝜖 = 1.0 + 14.868𝑉 − 33.56𝑉 2 + 51.223𝑉 3 − 36.283𝑉 4 + 9.715𝑉 5
where V is the SM300 output in Volts
V V V V V
0.000 1.000 0.300 3.556 0.600 4.956 0.900 6.471 1.200 7.961
0.075 1.963 0.375 3.913 0.675 5.327 0.975 6.841 1.275 8.411
0.150 2.641 0.450 4.254 0.750 5.708 1.050 7.205 1.350 8.971
0.225 3.149 0.525 4.598 0.825 6.092 1.125 7.571 1.425 9.724
Polynomial conversion
Combining the Soil calibrations and Sensor calibration
steps, the conversion equation becomes:
The following table lists the values used for the DL2e data logger:
Soil Mineral Organic Soil Mineral Organic
moisture soil soil moisture soil soil
%vol Volts Volts %vol Volts Volts
-4 -2.090 -2.090 52 0.801 0.670
0 0.044 0.021 56 0.867 0.731
4 0.074 0.045 60 0.934 0.791
8 0.108 0.072 64 1.003 0.852
12 0.147 0.102 68 1.072 0.913
16 0.193 0.137 72 1.140 0.976
20 0.248 0.178 76 1.204 1.039
24 0.313 0.226 80 1.260 1.102
28 0.384 0.281 84 1.309 1.163
32 0.457 0.343 88 1.351 1.220
36 0.530 0.410 92 1.388 1.270
40 0.600 0.477 96 1.419 1.314
44 0.669 0.544 100 1.447 1.352
48 0.735 0.608 104 2.090 2.090
Check Versions
Check you have the correct versions:
HH2 Meter: Firmware version 2.5 and PC software HH2read
version 5 or later are recommended.
GP1 & DL6 Loggers: DeltaLINK version 2.4 or later is required.
GP2 Logger: DeltaLink 3 or later is required
DL2e Logger: Ls2Win 1.0 SR6 or later is required
Consult the user manuals or the on-line help for these devices and
their related software.
Try alternative types of equipment if you have them available.
Check that you are using an appropriate soil calibration and the
correct conversion method – see Calibration section.
The SM300 itself
Try to isolate the problem into one of the following areas
The SM300 or the connecting cable
Then try to narrow down the area further
Mechanical problems faults, or damage
Electrical or electronic problems or faults
Air reading
Hold the SM300 away from other objects and take a reading using
an HH2 meter, or voltmeter or logger.
The reading should be 0 ±4mV when used with a 5m cable.
Specifications
Volumetric water content
Accuracy 2.5% vol over 0 to 50% vol and 0-60°C
Measurement range 0 to 100% vol with reduced accuracy5
Salinity error (see p.30) ≤3.5%vol over 50 to 1000 mS.m-1 and 0-40% vol
Output Signal 0-1V differential ≈ 0 to 60% vol nominal
Output compatible with GP1, GP2, DL6, DL2e, HH2
SM300 must be fully buried to
Temperature
accurately measure soil temperature
Sensor accuracy 0.5°C over 0-40°C
not including logger or cabling error
5 In water (no soil present) the reading may not be 100% vol. It depends on a0 and a1 but can still be
used as a quick check that the unit is working. See page 24.
6 See Appendix 2 on page 42.
7 Note: The DL6 has only one temperature channel. The DL6 error contribution to SM300 temperature
measurement is negligible compared to the accuracy of the SM300 temperature sensor itself. The two
only become comparable below -15C.
8 DL2e logger users can apply a correction in the Ls2Win logging software (for cable lengths >5m)
1.2 ~20%
1.0
0.8
0.6
0.4
0.2
0.0
0 100 200 300 400 500 600
Conductivity ECp (mS.m-1)
0 2 4 6 8 10 12 14 16 dS.m-1
Classification of salinity
θ ~ 0.4
Apparent soil moisture content (m3.m-3)
θ ~ 0.3
0.4
θ ~ 0.2
θ ~ 0.1
bound water
0.3
0.2
0.1
0
0 10 20 30 40 50
Note: ice has a quite different refractive index from water, so SM300 soil
moisture readings cannot be interpreted reliably when inserted into soil
below 0°C.
For best results keep a gap of at least 25mm (1 inch) between the
body of the sensor and the wall of the plant pot.
0%
-1%
Error (% of reading)
-2%
-3%
-4%
-5%
-6%
0 5 10 15 20 25 30 35 40
Gap between SM300 body and wall of pot (mm)
General information
SM300 is a Class A product, intended for operation in non-
residential environments.
Only use cables and accessories authorised by Delta-T (sensor
cables from other sources for example may adversely affect
product performance and affect quality of results).
If possible route cables along the soil surface or bury them – this
also reduces possible trip hazard and animal damage.
Do not modify the product or its supplied accessories.
See also SM300 EMC Guidance on the Software and Manuals CD
Issue 3.
Regulatory information
Europe
This device conforms to the essential requirements of the EMC
directive 2004/108/EC, based on the following test standards:
EN61326-1:2006 Electrical requirement for measurement, control
and laboratory use. EMC requirements: Group 1, Class A
equipment – (emissions section only).
EN61326-1:2006 Electrical requirement for measurement, control
and laboratory use. EMC requirements: Basic Immunity (immunity
section only).
Soil Moisture Content varies from approx. 0.02 m3.m-3 for sandy
soils at the permanent wilting point, through approx. 0.4 m 3.m-3 for
clay soils at their field capacity, up to values as high as 0.85 m 3.m-3
in saturated peat soils.
Salinity
0 2 4 6 8 10 12 14 16 dS.m-1
Classification of salinity
4. White, I., J.H. Knight, S.J. Zegelin, and Topp, G.C. 1994
Comments on ‘Considerations on the use of time-domain
reflectometry (TDR) for measuring soil water content’ by W R
Whalley
Journal of Soil Sci. 45, 503-508
Technical Support
Users in countries that have a Delta-T distributor or technical
representative should contact them in the first instance.
Technical Support is available on Delta-T products and systems. Your
initial enquiry will be acknowledged immediately with a reference number.
Make sure to quote the reference number subsequently so that we can
easily trace any earlier correspondence.
In your enquiry, always quote instrument serial numbers, software version
numbers, and the approximate date and source of purchase where these
are relevant.
Contact details:
Soil-specific Calibration
This note provides details of 2 techniques for generating soil-specific
calibrations:
Laboratory calibration for substrates* and non-clay soils
Laboratory calibration for clay soils
* We use the term substrate to refer to any artificial growing medium.
Underlying principle
Soil moisture content () is proportional to the refractive index of the soil
() as measured by the SM300 (see Calibration section).
The goal of calibration is to generate two coefficients (a0, a1) which can be
used in a linear equation to convert probe readings into soil moisture:
a0 a1
This is the easiest technique, but it’s not suitable for soils that shrink or
become very hard when dry.
Equipment you will need:
SM300 and meter
Soil corer (if doing a calibration for a cohesive soil rather than sand
or a substrate)
Heat-resistant beaker ( 0.5 litre)
Weighing balance (accurate to < 1g)
Temperature controlled oven (for mineral soils or substrates)
Re-insert the SM300 into the dry sample and record this
reading.
V0 = 0.051V
gives 0 1.68
Since 0 = 0, this is the value needed for a0
a0 = 1.68
Calculate w The water content of the wet soil, w, can be calculated from
the weight of water lost during drying, (Ww – W0) and its
w = 0.25
Calculate a1 In the wet soil V = Vw = 0.350 Volts and substituting gives
w 3.79
Finally
a1 w 0 w 0 3.79 1.68 0.25 0 8.44
a1 = 8.44
Result a0 = 1.68
a1 = 8.44
a0 a1
See also page Underlying principle on page 39
Dry the sample until still moist, ~15% water content. Gentle
warming can be used to accelerate the process, but take care
not to over-dry in places, and allow time for the water content
to equilibrate throughout the sample before taking a reading.
Reweigh.
Wm = 693.2g
a0 a1
See also page Underlying principle on page 39
DL2e Logger
The linearization table for the ‘S3T’ sensor code (‘SM300
Temperature’) provides supply current correction for the
SMSC/lw-05 5m logger cable ONLY.
Other loggers
If your logger can be programmed so that the soil moisture and
temperature readings can be taken sequentially (i.e. the sensor is
not powered during the temperature reading), then the temperature
can be obtained directly from the response curve on page 50.
Otherwise, correct the resistance reading before applying the
response curve.
You need to know the resistance of the Power 0V wire in the
SM300 cable (Rc) and establish whether your logger uses voltage
or current excitation for resistance measurement.
9
Note: This equation only applies to Delta-T SM300 cables
SM300 User Manual 1.3 Appendix 2: 48
Voltage Excited
You need to know the excitation voltage (Vref), reference resistance (Rref).
The correct resistance is given by the equation:
R = a0 + a1 * Rmeas
Where:
a0 = – Ic.Rc.Rref / Vref
a1 = 1 – Ic.Rc / Vref
Ic = 18 mA (SM300 sensor supply current)
Current Excited
You need to know the excitation current (Iex).
The corrected resistance is given by the equation (using terms defined
above):
R = Rmeas – Ic.Rc/Iex
Temperature Resistance
degrees C Kohms
-25 90.538
SM300 Resistance to Temperature
-22 77.683 Conversion Chart
-19 66.854
-16 57.713
100
-13 49.968
-10 43.379
Resistance
-7 37.759 90
(kOhms)
-4 32.957
-1 28.844 80
2 25.299
5 22.244
70
8 19.608
11 17.321
14 15.334 60
17 13.606
20 12.098 50
23 10.780
26 9.623
40
29 8.611
32 7.720
35 6.935 30
38 6.241
41 5.627 20
44 5.080
47 4.595
10
50 4.162
53 3.775
56 3.430 0
Note: This table has been optimised for use as a look-up table.
To minimise linear interpolation errors the data points fall either side of the
manufacturers’ specified sensor response curve. This helps optimise the
overall accuracy of readings.