Integration of Vieworks DR With Shimadzu
Integration of Vieworks DR With Shimadzu
SMEA_IVDRS_2020 (Rev:A)
1 Purpose ................................................................................................................. 3
4 DR System .............................................................................................................. 5
6 DAP Setup............................................................................................................... 13
1. Purpose
This document is written for instructing how to use generator service functions to integrate VXvue with the Shimadzu
generator.
2. Basic Information
GS.SHIMADZU is a component for integrating with Shimadzu generator and communicating with the generator using RS-232c. It
will be installed on the same PC of Vieworks.
VXvue is the software of Vieworks and can transmit the dose setting value and others to Shimadzu generator through this
component in one-way only. However, the actual irradiated dose value (Post Dose Value) can be delivered from the Generator.
The component supports 1P (AEC) and 3P (mA/ms) modes, so refer to ‘Restrictions and Features’ in this document for the
detailed information. The dose table can be different depending on the generator models.
3. How to Setup
3.1 Terms
Terms Description
It is developed / provided as a component of VXvue and provides integrating with Shimadzu generator.
GS.SHIMADZU GS stands for ‘Generator Service’.
VXSetup It is a component of VXvue that can set up integration with Shimadzu generator.
The following conditions should be met to use the Shimadzu Generator Service:
To set and use the Shimadzu generator with VXvue normally, be sure to check the following restrictions:
1. 1P : Controls kVp
Supports 1P (AEC) and 3P
Selecting mode
(mA/ms) modes 2. 3P : Controls kVp, mA, ms
2. User can change the default dose value and dose table.
Setting GS (Dose Setting and changing dose table Setting and
value) changing default dose values 3. mAs table is not supported.
4. Do not change the values other than the default dose value and
dose table.
Post Dose Displaying dose value that actually exposed Supporting dose value that actually exposed while shooting X-ray.
Setting AEC Available for setting AEC Density (-3 ~ 3) can be set.
The center field cannot be set simultaneously with left or right field.
Selecting Bucky Available for selecting Bucky Available for selecting NO-BUCKY, BUCKY 1, BUCKY 2
❖ We must start with installing the DR System and ensure that it’s working fine and fully functional before
proceeding in the integration with the Generator
4.1 DR Installation
Starting from Prerequisites, Installation to Calibration of the FPD’s. Please refer to the “VIVIX Setup Operation Manual” for
more information.
As for the below port Setup, It’s mentioned in details as it’s important step to establish the communication through the RS 232
Cable.
window.
2. Select Ports (COM & LPT) from the menu on the Device Manager screen to display the Communications Port menu.
5. Generator Communications
There are 2 types of Generators are connected with Vieworks and tested as mentioned earlier:
❖ The Next Scheme will describe the systems we have integrated with Vieworks and the options needed to have the
generator communication established:
5.2 SHIMADZU Generator Installation
❖ Refer to SHIMADZU Installation Manual (M501-E362R) for setting the SHIMADZU generator.
1. Execute VXSetup.
3. Specify the Generator Service Path as [The folder where VXvue is installed]\GENERATOR\Protocol\GS.SHIMADZU
\GS.SHIMADZU.exe.
4. Click the Echo button to confirm that communication with SHIMADZU generator is normal.
5.4 Generator Integration:
Please refer to the below description that is related to the connection between the SCU “System Control Unit” of Viewroks and
the Terminal PCB inside the generator:
❖ The below Figure 1 will show the connection and the Technique Code that should be used for Stand and table:
- Figure 1-
❖ The below Picture is the actual Connection of the Terminal PCB and description:
- Figure 2-
❖ Communication Setup:
From the Generator Console side, we must adjust the Communication Settings as Initial Setup by accessing the Communication
from the Setup screen and adjust the UD-COM I/F as below:
❖ Figure 4 will show the actual connection of the Handshaking Cable from the “SCU” to the Generator:
-Figure 4-
❖ Communication Setup:
In Rad FIT system we only need to adjust SW2-7 on NEX-SH PCB to make it DR.
DAP (Dose Area Product) is the medical equipment for measuring the exact X-ray dose exposed per unit area. You can connect
PC and DAP to read values from VXvue.
• The DAP meter will be connected to the generator and the Generator will be connected to VXvue (PC).
• The Indirect Integration with Shimadzu Generator is still not integrated yet.
• It’s Direct connection between the DAP meter and the PC as below:
• The DAP models that support direct integration with VXvue are as follows.
• We have used VacuDAP Standard (with Display Unit) and it’s working fine.