1.0 PURPOSE :
To establish and maintain an instruction and assign responsibilities to control, calibrate and maintain
the Monitoring, Measuring and Testing Devices (collectively, “devices”) used to demonstrate the
product conformity to the specified requirements.
2.0 SCOPE :
This Instruction is applicable to PT. Fujii Ironworks Teknologi and all associated operating units.
4.0 DEFINITION :
-
Table 1, CalibrationStandard
Monitoring and Measuring DevicesCalibration RequirementsTable
All new/unused measuring devices taken out of the original box (when supplied with a
manufacturer’s calibrationcertificate) shall be considered calibrated from thedateitis put in service
for the frequency assigned below.
Calibration
Device Tolerance/Method/Requirements & Restrictions
Frequency
Master Gauge 5 Years Method Traceable to international standard
Blocks Use: Used to calibrate micrometres, OD dial and digital callipers.
Master gauge blocks will be GradeA+ (2) or better.
Tolerance:
Plus Minus
NominalSize (In)
(Micro In) (Micro In)
1"or less 4 2
2" 8 4
3" 10 5
4" 12 6
5" 14 7
6" 16 8
7" 18 9
8" 20 10
Dial Gauge 1 Year Tolerance: see Clause 6 below
(Gage Maker) Method Traceable to international standard
Mic-Trac 1 Year Tolerance: ± 0.00033"
Precision linear- Method Traceable to international standard
Measuring Use: Used to calibrate Dial gauges, micrometres, Vernier and digital
Machine calipers if needed
Table 2,WorkingMonitoring,MeasuringDevices
All new/unused measuring devices taken out of the original box (when supplied with a manufacturer’s
calibrationcertificate) shall be considered calibrated from the date it is put inservice for the frequency
assigned below.
Calibration
Device Tolerance/Method/Requirements & Restrictions
Frequency
Dial Indicator 1 Year Range (mm) : 10
Accuracy (mm) : 0.01
Tolerance: ± 0.015 mm
Method Traceable to international standard.
NC 38 (3 1/2 IF) Tolerance : based on the requirements of API Specifications 7-2 2nd
Work Ring & 500 Edition, Clause 8.2.3, table B.13 & Fig. 20
Plug Usage Acceptance : based on the requirements of API Specifications 7-2 2nd
Edition, Clause 9.2
Calibration method: Calibration was performed at 20 ˚C ± 1 ˚C ( 68 ˚F ±
1.8 ˚F ). With standards which are certified accurate and traceable to
the National Institute of Standards and Technology and / or A2LA
calibration labs.
NC 50 (4 1/2 IF) Tolerance : based on the requirements of API Specifications 7-2 2nd
Work Ring & 500 Edition, Clause 8.2.3, table B.13 & Fig. 20
Plug Usage Acceptance : based on the requirements of API Specifications 7-2 2nd
Edition, Clause 9.2
Calibration method: Calibration was performed at 20 ˚C ± 1 ˚C ( 68 ˚F ±
1.8 ˚F ). With standards which are certified accurate and traceable to
the National Institute of Standards and Technology and / or A2LA
calibration labs.
Profile Tolerance: based on the requirements of API Specifications 5B 16th
Template & 1 Year Edition, Clause 5.4.3 and API Specifications 7-2 2nd Edition, Clause 7.4.1.
Lead Standard Acceptance : based on the requirements of API Specifications 7-2 2nd
Edition, Clause 7.4.2
Calibration method: Calibration was performed at 20 ˚C ± 1 ˚C ( 68 ˚F ±
1.8 ˚F ). With standards which are certified accurate and traceable to
the National Institute of Standards and Technology and / or A2LA
calibration labs.
4. CALIBRATION PROCEDURE
4.1 Initial Examination
A complete visual examination shall be made to ensure that all moving parts are intact and
free of any excessive wear or damage. Adjust, repair or replace gauges when the measured
readings do not meet the accuracy acceptance criteria defined by this procedure.
5. Calibration Frequency
The normal calibration frequency for Dial gauges, at frequent intervals no less than once per year,
however, if the dial gauges is not used in the 1 year period, calibration is not required until
subsequent future usage.
7.0 DETAIL INSTRUCTION FOR PRESSURE CHART RECORDERS CALIBRATION OF PRESSURE CHART
RECORDER:
1. REQUIRED EQUIPMENT OR STANDARDS
© 2019 PT. Fujii Ironworks Teknologi
This document is intended for the sole use of PT. Fujii Ironworks Teknologi and no part of this book may be reproduced in any form or by
any means without permission in writing. Printed copies and electronic files saved outside FUJITEK permit are not controlled. Latest
revision can be found at FUJITEK Quality Management System
PT. SAKA
TEHNIK UTAMA TECHNICAL SPECIFICATION PROCEDURE
REV: PAGE: ORIGINAL ISSUE DATE: REVISION DATE:
DOCUMENT NUMBER:
TP.QCT-07 2 9 of 14 OCT 7, 2019 -
TITLE : IMTE STANDARD REQUIREMENT & CALIBRATION ACCEPTED
2. CALIBRATION PROCEDURE
2.1. VISUAL EXAMINATION
2.1.1. A thorough examination shall be made to ensure that all moving parts are intact,
functional and free from any excessive wear or damage.
2.1.2. In the event excessive wear and/or damage are detected, then the recorder shall be
taken out of service and repaired or replaced.
3. Acceptance Criteria
The error in indicated measurement shall be within +/- 2% of the full recording scale value.
© 2019 PT. Fujii Ironworks Teknologi
This document is intended for the sole use of PT. Fujii Ironworks Teknologi and no part of this book may be reproduced in any form or by
any means without permission in writing. Printed copies and electronic files saved outside FUJITEK permit are not controlled. Latest
revision can be found at FUJITEK Quality Management System
PT. SAKA
TEHNIK UTAMA TECHNICAL SPECIFICATION PROCEDURE
REV: PAGE: ORIGINAL ISSUE DATE: REVISION DATE:
DOCUMENT NUMBER:
TP.QCT-07 2 10 of 14 OCT 7, 2019 -
TITLE : IMTE STANDARD REQUIREMENT & CALIBRATION ACCEPTED
4. Calibration Frequency
The normal calibration frequency for Pressure Chart Recorders shall be once every three (3)
months.
3. CALIBRATION PROCEDURE
4.1. Initial Examination
A complete visual examination shall be made to ensure that all moving parts are intact and
free of any excessive wear or damage.
Using a deadweight tester or master pressure-measuring device, measure and record the
pressure gauge readings at least three equidistant points of full scale (excluding zero and full
scale as required points of calibration). The preferred points of calibration are at 25%, 50%
and 75% of full scale. Adjust, repair or replace gauges when the measured readings do not
meet the accuracy acceptance criteria defined by this procedure.
4.2. Calibration Adjustment(s)
When excessive wear, damage or a need for adjustment is detected, the gauge shall be
pulled from service and repaired, adjusted or replaced.
Remove the gauge socket assembly (body) from the case and connect to a deadweight tester
or a pressure comparator with a master gauge.
Reset the pointer (needle) at the low (0) position and verify that the positive spring (hair
spring) is in the fully wound position.
Verify and adjust full span accuracy as follows:
Pressurize the gauge to 100% of full scale and check the high range reading.
If the pointer does not read correctly, adjust the full span as needed.
Depressurize the gauge and check the low (0) reading and adjust as needed until both
the high and low reading is correct.
NOTE: It is extremely important to have approximately four sector teeth below the pinion.
The gauge angularity adjustment (mid-range reading) shall be checked as follows:
© 2019 PT. Fujii Ironworks Teknologi
This document is intended for the sole use of PT. Fujii Ironworks Teknologi and no part of this book may be reproduced in any form or by
any means without permission in writing. Printed copies and electronic files saved outside FUJITEK permit are not controlled. Latest
revision can be found at FUJITEK Quality Management System
PT. SAKA
TEHNIK UTAMA TECHNICAL SPECIFICATION PROCEDURE
REV: PAGE: ORIGINAL ISSUE DATE: REVISION DATE:
DOCUMENT NUMBER:
TP.QCT-07 2 11 of 14 OCT 7, 2019 -
TITLE : IMTE STANDARD REQUIREMENT & CALIBRATION ACCEPTED
Pressurize the gauge to approximately 50% of the full scale and compare the reading
with the requirements of Section 6.0 of the procedure.
If the reading is too high, bend the Bourdon tube tip outward to change the angle of
pull.
If the reading is too low, the Bourdon tube tip should be bent inward.
Depressurize the gauge and check the low reading. If needed, reset the pointer to zero and
repeat the adjustments noted above until both the mid-range and low readings are correct.
After all adjustments are completed, the gauge accuracy shall be verified to at least three
equidistant points of full scale (excluding zero and full scale as points of calibration). The
preferred points of calibration are at 25%, 50% and 75% of full scale.
4. Calibration Frequency
The normal calibration frequency for Pressure Gauges shall be a maximum of three (3) months
until recorded calibration history can be established to justify a longer frequency. Longer
intervals in maximum increment of three month can be established. When indicated by the
calibration history, the frequency shall be shortened accordingly to prevent the use of out of
tolerance pressure gauges.
5. Acceptance Criteria
The tolerance is +/- 2% of the full-scale range.
For example : Using the preferred test ranges, pressure gauges certified for pressure tests
would be tested in accordance with the following :
Gauge Rating 25% of Range 50% of Range 75% of Range Permissible Error
( psi ) ( psi ) ( psi ) ( psi ) ( psi )
100 25 50 75 2
300 75 150 225 6
1000 250 500 750 20
3000 750 1500 2250 60
5000 1250 2500 3750 100
6000 1500 3000 4500 120
10000 2500 5000 7500 200
15000 3750 7500 11250 300
20000 5000 10000 15000 400
30000 7500 15000 22500 600
50000 12500 25000 37500 1000
© 2019 PT. Fujii Ironworks Teknologi
This document is intended for the sole use of PT. Fujii Ironworks Teknologi and no part of this book may be reproduced in any form or by
any means without permission in writing. Printed copies and electronic files saved outside FUJITEK permit are not controlled. Latest
revision can be found at FUJITEK Quality Management System
PT. SAKA
TEHNIK UTAMA TECHNICAL SPECIFICATION PROCEDURE
REV: PAGE: ORIGINAL ISSUE DATE: REVISION DATE:
DOCUMENT NUMBER:
TP.QCT-07 2 12 of 14 OCT 7, 2019 -
TITLE : IMTE STANDARD REQUIREMENT & CALIBRATION ACCEPTED
NOTE: These ranges are shown for the gauges normally used for these applications. Values
must be calculated by the calibrator for other gauge pressure ranges or use of other ranges for
the three equidistant points of calibration.
9.0 DETAIL PROCEDURE FOR DRIFT MANDRELS FOR API APPLICATIONS
1. Calibrations of Drift Mandrel
1.1. Required Equipment or Standards
Outside Diameter Micrometers and/ or Calipers
2. Calibration Frequency
The normal calibration frequency for drift Mandrels shall be once every thirty-six (36) months. And
Verification OD dimension Before usage.
3. Acceptance Criteria