OPEN Validation of a method to assess


night myopia in a clinical setting
Gené‑Sampedro Andrés 1,2*, Basulto Marset Mercedes 1, Monsálvez Romin Daniel 1,
Montecelo Salvado Susana 3 & Bueno‑Gimeno Inmaculada 1

A study was conducted with 115 subjects who regularly drove at night to validate a refraction protocol
for detecting refractive visual changes from daytime to nighttime conditions. Objective and subjective
refractions were performed in both photopic and mesopic conditions, with a dark adaptation period
before the mesopic subjective refraction. The results showed that in mesopic conditions, visual
acuity decreased by 0.2 logMAR units on average (p < 0.01), and there was a myopic refractive shift of
− 0.36 ± 0.20 D (p < 0.01). Most subjects (92.2%) exhibited a myopic refractive shift of at least 0.12 D.
Compensation of refractive shift improved mesopic visual acuity by 0.06 logMAR on average (p < 0.01)
and higher refractive shifts showed higher improvement. Night Rx was preferred by 82.1% of subjects
with myopic refractive shift. Gender and age did not significantly affect the refractive shift, although
myopes showed a higher shift compared to emmetropes (p < 0.01). The refractive shift remained
stable over time when the time slot of the day did not change (p < 0.01). Night Rx protocol proved to
be a robust and accurate method for identifying drivers with refractive changes when transitioning
from photopic to mesopic conditions. The high prevalence and inter-individual variability of Rx shift
highlight the need of customized refraction.

The human eye can detect different levels of illumination, from bright to very dim light, and the visual system
can adapt its sensitivity to these c­ onditions1. Interestingly, it has been observed that some subjects, under low-
lighting conditions, become more myopic. The so-called “night myopia” has already been documented for more
than a century and several studies have been carried out to try to give an explanation, but even today it is not
fully understood and there is no single consensus about its ­causes2. To date, experimental works have tried
to isolate different factors separately (Purkinje effect, eye aberrations, neural factors, adaptation to darkness),
being accommodation role the one that has been suggested as the main ­component3. Furthermore, most of the
mentioned experimental studies have been carried out in monocular conditions, while it has been seen that
under binocular conditions this effect could be lower, probably due to fusional convergence that comes into
play, stabilizing a­ ccommodation4. There is probably not a single cause, but a conjunction of variable contribution
factors between subjects can lead to this phenomenon.
Despite all the doubts that still need to be clarified about night myopia, the manifestation of this phenomenon
has recently gained clinical interest due to its impact on every-day tasks for a significant number of people, such
as nighttime driving. Driving itself is a complex task that requires specific visual skills and optimal vision. In fact,
we know that the visual route is what provides us with more information and the one that mainly contributes
when maintaining a safe driving. Therefore, good vision is essential for safe and efficient d ­ riving5. Nighttime driv-
ing entails more considerable risks compared with daytime, and it involves a challenging visual environment with
dim street lighting, glare from oncoming headlights, and the need to quickly adapt across a wide range of lighting
levels. Certainly, fatality rates at night are up to three times higher than in the day when adjusted for driving
­exposure6. The most likely reason seems to be the reduced visibility of the nighttime road ­environment2,7–9. Night
myopia significantly impairs driving performance, particularly in very bad-lit road ­environments2. Wearing cor-
rective lenses can improve visual performance and reduce the impact of night myopia on driving ­performance10.
Thus, it would be ideal for visual processing capabilities to remain in optimal conditions to perform the task over
time, providing a prescription to correct night myopia.
The differences in night myopia results among studies sometimes are related to the measuring methods. Sub-
jective refraction is still considered the gold standard for determining the optimal prescription for the correction
of refractive errors, using objective methods as a baseline. However, standard subjective refractions are normally
performed in a photopic viewing context which does not represent the nighttime driving environment. In this

1
Department of Optics, and Optometry and Vision Science, University of Valencia, Facultad Física, c/ Dr. Moliner
50, 46100 Burjassot, Spain. 2INTRAS (Research Institute on Traffic and Road Safety), University of Valencia,
46022 Valencia, Spain. 3Department of Light and Life Vision Sciences, Research and Development EssilorLuxottica,
75012 Paris, France. *email: [email protected]

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 1

Vol.:(0123456789)
www.nature.com/scientificreports/

scenario, it would be more appropriate to assess the vision function under mesopic or low light ­conditions11,12.
Up to the authors’ best knowledge, there is no established standard clinical procedure in such conditions.
Using standardized protocols in visual examinations can lead to more accurate diagnoses, more reliable
treatment plans, and improved outcomes for ­patients13–15. A standardized protocol might help professionals to
assess mesopic visual function and thus, to better address night vision needs of drivers and to provide them with
custom correction. For all these reasons, this research aims to validate a protocol to perform subjective visual
refractions under low lighting levels (mesopic) to identify those subjects who manifest a shift towards more
myopic values with respect to their standard refraction. Also, it is desirable to estimate feasible time limits for its
assessment in the clinical setting, considering the dark-adaptation time and battery of mesopic tests. Moreover,
this investigation might help to understand the distribution of night myopia and the possible related individual
factors contributing to its manifestation.
The purpose of this study was to validate a measurement protocol for accurate and reliable refractive night-
time prescription that can be easily implemented in practice.

Results
The mean age of the sample was 34.51 ± 13.09 years, ranging from 18 to 63 (56% men, 44% women). Age dis-
tribution was not homogeneous, the burden of the sample consisted of 71% of non-presbyopes (18–39 years),
whereas 29% were presbyopes (40–63 years). There were no significant differences in mean age according to
gender (p > 0.05). The mean photopic and mesopic pupil size under natural conditions were 2.75 ± 0.45 mm and
5.82 ± 0.99 mm respectively for non-presbyopes and 2.39 ± 0.29 mm and 4.81 ± 0.69 mm for presbyopes. The
mean spherical aberration (5-mm pupil) was − 0.00 ± 0.07 microns and 0.04 ± 0.05 microns for non-presbyopes
and presbyopes respectively.
The mean (spherical equivalent) M based on objective “day” refraction was − 1.94 ± 2.30 D and on “night”
refraction − 1.95 ± 2.29 D (mean change of − 0.01 ± 0.18D p > 0.05). In terms of ametropia distribution, 8.7%
of subjects were hyperopes (M > 0.875 D), 18.3% emmetropes (− 0.625 D < M < 0.875 D), and 73.0% myopes
(M < − 0.625 D). The mean manifest M based on standard (photopic) subjective refraction was − 1.93 ± 2.25 D,
with a mean cylinder of − 0.76 ± 0.57 D and on night (mesopic) subjective refraction − 2.29 ± 2.28 D (mean change
in sphere of − 0.36 ± 0.19 D. p < 0.001). The mean values of refraction (right eye) by light condition and age cat-
egory are shown in Table 1. The ametropia was balanced across gender (p > 0.05). According to age, the presbyopic
group included a higher number of hyperopic subjects, whereas non-presbyopes were more myopic (p < 0.001).
With the PA for the calculation of the Rx shift (Rx shift w/PA), the 92.2% of the subjects (n = 106) showed a
myopic shift equal or superior to − 0.12 D, 83.5% to − 0.25 D, 59.2% to − 0.37 D, 33.1% to − 0.50 D whereas the
7.8% had not myopic shift (n = 9). PA had a significant effect on Rx shift (p < 0.001), which was independent of
the degree of gain in Night visual acuity (VA) with Night Rx (p < 0.001). The results showed that the Rx shift is
not dependent on age or gender, neither for any age-gender category. When comparing ametropia groups, an
effect on Rx shift was found. Specifically, emmetropes showed lower Rx shift than myopes (p < 0.05), but there
was no effect of myopia degree on Rx shift (p > 0.05) (Fig. 1). There was no effect of astigmatism degree on Rx
shift either. There was no effect of other optical factors on subjective Rx shift as mesopic pupil size (Spearman’s
R = − 0.162 p = 0.083) or 5-mm pupil spherical aberration (valid n = 53; Spearman’s R = − 0.094 p > 0.05). However,
spherical aberration influenced the objective Rx shift (R = − 0.957 p < 0.001) which showed to be less myopic than
subjective Rx shift by 0.35 D in mean (p < 0.001).
VA decreases by 0.2 logMAR on average from photopic to mesopic conditions (p < 0.001). In mesopic condi-
tions, when Rx shift is compensated VA increases by − 0.06 logMAR on average (p < 0.001). VA shifts vary among
individuals, Standard VA shift ranges up to 0.40 logMAR and mesopic VA shift up to − 0.24 logMAR. The mean
values of VA by light conditions and age category are shown in Table 1. With Night Rx, non-presbyopes show
better mesopic VA (p = 0.006) and mesopic VA shift (p = 0.01) than presbyopic subjects. Values of VA shift by
light conditions and age category are shown in Table 2. 100% of both non- presbyopes and presbyopes show VA
loss from photopic to mesopic conditions (with Std Rx) and 89.2% of non-presbyopes and 78.1% of presbyopes
had VA improvement with the Night Rx.
According to quality of vision preferences, for those subjects who present myopic Rx shift, Night Rx was pre-
ferred by 82.1% of subjects, 13.2% had no preference and 4.7% preferred Std Rx. Higher myopic Rx shifts led to
more important subjective difference between Night Rx and Std Rx (p < 0.01) (Fig. 2a). However, in some cases,
low Rx shifts as − 0.25 D or − 0.37 D also led to perception of important differences, whereas moderate to high
Rx shifts, − 0.50 D to − 0.87 D, also yielded slight differences. On the other hand, subjects who had Rx shifts of
− 0.12 D did not perceive any difference. In terms of VA the higher the gain when Rx shift is compensated, the
more important the difference perceived between Night and Std Rx (Fig. 2b).
The assessment of the reproducibility of the proposed method (n = 54, mean age 27.1 ± 6.2 years) showed no
significant differences in either M or Rx shift between visit 1 (V1) and visit 2 (V2) without PA (wo/PA). However,
taking the PA into account, the Rx shift was more myopic in V2 (p < 0.01) (Fig. 3), although this difference was
still close to zero (− 0.07 D).
Differences of refractive and VA parameters as well as LoA between visits are shown in Table 3. The repro-
ducibility of Rx shift w/PA was near 60% within ± 0.12 D and 94% within ± 0.25 D. The number of days elapsed
between V1 and V2 [< 60 days (n = 28) and > 60 days (n = 26)] had no effect on Rx shift wo/PA. The Rx shift w/
PA was more myopic when elapsed time was greater than 60 days (mean difference − 0.08 D) (p = 0.047). There
was found an examiner effect. The intra-examiner effect (same examiner between visits) yielded more myopic
Rx shift values (w/PA) in V2 for examiner 2 (− 0.12 D in average, p < 0.01), but not for examiner 1 (p = 0.60)
(Fig. 4a). The inter-examiner (different examiners between visits) did not significantly change this Rx shift. Lastly,
the reproducibility of the Rx shift was influenced by the differences in the time of the day at which the protocol

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 2

Vol:.(1234567890)
www.nature.com/scientificreports/

Rx shift ≤ -0.12 D
All subjects NP = 82 P = 33 NP = 74 P = 32
Mean ± SD
Conditions Age p Mean ± SD Median (IQR) Median (IQR)
− 2.48 ± 1.98 − 2.37 ± 1.92
NP
− 2.25 (2.08) − 2.13 (2.03)
Day < 0.01
− 0.60 ± 2.52 − 0.68 ± 2.51
P
− 0.20 (3.57) − 0.22 (3.62)
Objective Rx
− 2.47 ± 1.99 − 2.36 ± 1.93
NP
− 2.20 (2.09) − 2.05 (2.00)
Night < 0.01
− 0.67 ± 2.53 − 0.75 ± 2.53
P
− 0.30 (3.58) − 0.33 (3.54)
− 2.43 ± 1.93 − 2.32 ± 1.87
NP
− 2.25 (2.10) − 2.20 (2.09)
Subjective Std Rx Photopic < 0.01
− 0.68 ± 2.52 − 0.75 ± 2.52
P
− 0.01 (3.60) − 0.19 (3.66)
− 2.74 ± 1.95 − 2.67 ± 1.92
NP
− 2.63 (2.03) − 2.54 (2.04)
Mesopic < 0.01
− 0.96 ± 2.58 − 1.04 ± 2.58
P
− 0.25 (3.89) − 0.45 (3.93)
Subjective Night Rx
− 2.80 ± 1.95 − 2.73 ± 1.92
NP
− 2.69 (1.98) − 2.67 (2.00)
Mesopic w/PA < 0.01
− 1.02 ± 2.58 − 1.10 ± 2.58
P
− 0.38 (3.77) − 0.51 (3.92)
− 0.31 ± 0.19 − 0.34 ± 0.15
NP
− 0.25 (0.25) − 0.25 (0.25)
Mesopic-Photopic wo/PA > 0.05
− 0.28 ± 0.21 − 0.29 ± 0.21
P
− 0.25 (0.50) − 0.25 (0.38)
Subjective Rx shift
− 0.37 ± 0.20 − 0.41 ± 0.16
NP
− 0.38 (0.25) − 0.38 (0.25)
Mesopic-Photopic w/PA > 0.05
− 0.34 ± 0.19 − 0.35 ± 0.18
P
− 0.38 (0.38) − 0.38 (0.31)
− 0.14 ± 0.06 − 0.15 ± 0.06
NP
− 0.14 (0.08) − 0.16 (0.08)
VA Std Rx Photopic > 0.05
− 0.15 ± 0.06 − 0.15 ± 0.06
P
− 0.16 (0.08) − 0.16 (0.08)
0.06 ± 0.07 0.06 ± 0.07
NP
0.06 (0.08) 0.06 (0.08)
VA Std Rx Mesopic > 0.05
0.06 ± 0.07 0.06 ± 0.07
P
0.04 (0.10) 0.04 (0.12)
− 0.01 ± 0.06 − 0.02 ± 0.06
NP
− 0.02 (0.10) − 0.02 (0.08)
VA Night Rx Mesopic < 0.05
0.02 ± 0.06 0.02 ± 0.06
P
0.02 (0.10) 0.02 (0.10)

Table 1.  Descriptive values of refraction of the right eye (M coefficient) and binocular visual acuity (logMAR)
by light condition and age category. NP non-presbyopes; P presbyopes; PA perceptual appreciation; Objective
Rx (WAM700); Subjective Rx, subjective Rx shift and VA (VR800); p: p value. Significant p-values are shown
in bold italics.

Figure 1.  Distribution of Rx shift by ametropia category **p < 0.01 *p < 0.05.

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 3

Vol.:(0123456789)
www.nature.com/scientificreports/

Standard VA s­ hift2 Standard VA ­shift2 Standard VA ­shift2 P Mesopic VA ­shift3 Mesopic VA ­shift3 Mesopic VA ­shift3 P
Category Range in l­ etters1 (n = 106) NP (n = 74) (n = 32) (n = 106) NP (n = 74) (n = 32)
Very high gain 18 or higher – – – – – –
High gain 13–17 – – – – – –
Moderate gain 8–12 – – – 13% (14) 18% (13) 3% (1)
Low gain 3–7 – – – 36% (38) 39% (29) 28% (9)
Slight gain 1–2 – – – 37% (39) 32% (24) 47% (15)
No change 0 – – – 8% (9) 7% (5) 13% (4)
Slight loss 1–2 1% (1) 1% (1) – 6% (6) 4% (3) 9% (3)
Low loss 3–7 14% (15) 15% (11) 13% (4) – – –
Moderate loss 8–12 61% (65) 62% (46) 59% (19) – – –
High loss 13–17 21% (22) 19% (14) 25% (8) – – –
Very high loss 18 or higher 3% (3) 3% (2) 3% (1) – – –

Table 2.  Distribution of binocular VA shift (in letters) by light condition and age category; Subjects with
Rx shift equal or superior to − 0.12 D NP non-presbyopes, P Presbyopes. 1 Each letter correspond to 0.02
logMAR units, 2VA change wearing Std Rx between mesopic and photopic conditions. 3VA change in mesopic
conditions between Night and Std Rx.

Figure 2.  Preference difference between NIGHT and STD Rx (a) by Rx shift and (b) mesopic VA gain (in
letters) when the Rx shift is compensated. Subjects with myopic shift (n = 101).

Figure 3.  Inter-session reproducibility of the refractive shift without (Rx shift wo/PA) and with perceptual
appreciation (Rx shift w/PA). V1 (first visit) and V2 (second visit). **p < 0.01.

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 4

Vol:.(1234567890)
www.nature.com/scientificreports/

Parameter (n = 54) Light Condition Visit 1 Mean ± SD Visit 2 Mean ± SD LoA p


Subjective Std Rx—M Photopic − 2.12 ± 1.62 − 2.12 ± 1.67 0.46 0.951
Subjective Std Rx – J0 Photopic 0.05 ± 0.38 0.04 ± 0.36 0.20 0.668
Subjective Std Rx – J45 Photopic 0.00 ± 0.17 − 0.02 ± 0.18 0.18 0.106
Subjective Night Rx -M Mesopic − 2.49 ± 1.67 − 2.52 ± 1.66 0.43 0.278
Subjective Rx shift wo/PA Mesopic-Photopic − 0.37 ± 0.15 − 0.40 ± 0.15 0.31 0.135
Subjective Rx shift w/PA Mesopic-Photopic − 0.43 ± 0.15 − 0.50 ± 0.15 0.31 0.006
VA Std Rx Photopic − 0.15 ± 0.06 − 0.15 ± 0.05 0.11 0.779
VA Std Rx Mesopic 0.06 ± 0.06 0.07 ± 0.07 0.15 0.331
VA Night Rx Mesopic − 0.02 ± 0.05 − 0.02 ± 0.06 0.10 0.238
Std VA shift Mesopic-Photopic(Std Rx) 0.21 ± 0.06 0.22 ± 0.06 0.13 0.297
Mesopic VA shift Mesopic (Night-Std Rx) − 0.08 ± 0.06 − 0.09 ± 0.05 0.14 0.839

Table 3.  Inter-session reproducibility of refractive coefficients M, J0 and J45 of the right eye (diopters)
and binocular visual acuity (logMAR) by light condition. Std Rx (Standard subjective refraction); Night Rx
(Mesopic subjective refraction); Rx shift (Refractive shift); woPA (without perceptual appreciation); wPA
(with perceptual appreciation); VA Std Rx (binocular visual acuity wearing Std Rx); VA Night Rx (binocular
visual acuity wearing Night Rx); Std VA shift (visual acuity shift from mesopic to photopic conditions, wearing
standard refraction); Mesopic VA shift (VA shift in mesopic conditions between Night and Std Rx); LoA: 95%
Limits of Agreement. p: p-value. Significant values are in bold italics.

Figure 4.  Reproducibility of refractive shift with perceptual appreciation (Rx shift w/PA) (a) Intra-examiner
effect (b) Effect of time slot of the day. V1 (first visit) and V2 (second visit) **p < 0.01, *p < 0.05.

was performed. When time of day did not change between visits (within ± 3 h), there was found no difference
in Rx shift (w/PA) independently of the number of days between visits or whether the examiner changes or not
(p > 0.05). When time of day changed between visits (more than ± 3 h), the Rx shift (w/PA) was more myopic in
V2 (mean difference − 0.09 D, p < 0.05) (Fig. 4b).

Discussion
In this study, we developed and tested a new protocol to measure mesopic subjective refraction (Night Rx
protocol). The use of protocols in the visual examination of refractive errors is important to ensure accurate
measurement and correction of refractive e­ rrors13,16,17. By following established protocols, clinicians can ensure
that all aspects of the examination are performed consistently and thoroughly, reducing the risk of errors or
oversights, leading to improved visual outcomes for patients. Validating protocols is essential for optimizing
refractive prescription ­accuracy16,18,19.
Subjects in this research were dark-adapted to a unique short period (5 min). In mesopic vision, both cones
and rods photoreceptors contribute to the visual response and both foveal and peripheral vision are involved. It
is a complex intermediate vision state as the degree of cones and rods interaction is constantly changing due to
the eye’s adaptation to a dynamic range of luminance levels and different visual stimuli. In mesopic vision, when
the light level decreases, cones increase their sensitivity by reduction of the spatial resolution whereas rods are
highly sensitive but slower in terms of adaptation and r­ esponse20. In the dark, cones adapt fast by increasing its
sensitivity until a maximum threshold after 5–10 min and rods (which are saturated) slowly start adapting over
time. Other studies have also shown a maximum improvement in mesopic VA after 6 min of dark ­adaptation21.

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 5

Vol.:(0123456789)
www.nature.com/scientificreports/

Most of the studies assessed mesopic visual function under a high range of dark-adaptation times from 5 to
20 ­min21,22. Since we aim to implement this protocol in a clinical setting, it is desirable to optimize the duration
of the measurements to avoid fatigue. For this reason and based on previous findings, a dark-adaptation of 5 min
was conducted prior to mesopic ­refraction21,22.
We corroborated that VA measures deteriorated in mesopic conditions compared to photopic conditions.
While driving at night, visibility of targets decreases, which increases visual processing time and the risk of acci-
dents, even with small ­reductions9,23,24. Myopic Rx shifts measured by the night Rx protocol in this study varied
from − 0.12 D to − 0.87 D (− 0.36 D on average). Several researchers have suggested that optical correction of
night myopia would only have a marginal benefit for V ­ A25,26. However, Hessler et al.27 reported that the optical
compensation of this refractive shift improves the visual function under mesopic conditions. Thus, optimizing
night driving performances might improve road safety. The influence of undercorrected refractive error as low
as − 0.50 D impacts night-driving performances as reading road ­signs28 or pedestrian ­recognition29. Quality of
vision and comfort also improves with optimized correction for mesopic l­evels27,30. In our study, whereas Rx
shift of − 0.12 D showed no benefit on perceived quality of vision on simulated real-life scene, changes as low
as − 0.25 D led to significant improvement in individual’s perception (Fig. 2a). Further research is necessary to
assess the clinical relevance of Rx shift magnitude in night driving activity.
In this study, myopic subjects had more night myopia compared to emmetropic and hyperopic ones. Atchison
et al.31 found that people with myopia had reduced visual performance under low-light conditions, including
reduced VA and contrast sensitivity. Other studies showed that night myopia values of − 0.75 D or higher affect
night-driving performance and increase number of a­ ccidents8. After reviewing our findings, myopic subjective
Rx shift (measured with Night Rx protocol and ETDRS chart) is not reflected by objective Rx measurements
(WAM700) which is significant less myopic by − 0.35 D, most likely due to spatial configuration of visual stimuli
and other instrument related factors. Furthermore, optical factors as pupil size and 5 mm-pupil spherical aber-
ration do not have influence in subjective Rx shift under low-light conditions, at least in healthy phakic patients.
Influence of pupil-related factors as spherical aberration on night vision has been reported ­previously2,4. As light
environment decreases pupil size increases and effect of excess of positive spherical aberration on visual per-
formance is greatest. Subjective refractive error becomes more myopic for some stimulus configuration typical
of night ­driving32. Previous research shown that spherical and chromatic aberrations are unlikely to affect night
­myopia3. Hessler et al.27 did not found relationship between subjective luminance-dependent refraction and pupil
size-dependent objective ­refraction24. In our study, spherical aberration does influence the objective Rx shift
but not the subjective change. Thus, subjective Night Rx and thus, the mesopic Rx shift should be influenced by
other factors than optical characteristics of the eye such as neural factors—receptoral (retina) and post-receptoral
complex visual processing-linked to the change in light e­ nvironment25,26.
The participants in our study were healthy. However, it is crucial to understand the importance of various opti-
cal considerations, especially when there are disturbances in the ocular media. The level of disturbances tends to
rise as the pupils dilate, which can lead to a heightened intraocular scattering effect Moreover, straylight increases
with age even in normal ­eyes33. This effect can potentially impact various visual functions, including contrast
sensitivity and disability glare. It is noteworthy that this impact is not limited to individuals with compromised
ocular health but can also affect those with seemingly healthy eyes and excellent visual a­ cuity34.
As in previous ­studies35, in our research, presbyopes and non-presbyopes presented different physiological
characteristics; presbyopes showed to be less myopic or more hyperopic, the pupil size decreases, and positive
spherical aberration (5-mm pupil) increases compared to non-presbyopes. However, these physiological differ-
ences are not reflected in Rx shift compared to non presbyopes. Change in refraction at low-light levels has been
related to accommodation errors more than to spherical aberration or pupil ­size3,36. In our study, physiological
decrease of accommodation with age does not impact measured subjective Rx shift when comparing non-presby-
opic and presbyopic populations. One of the key aspects of vision in driving is VA, which refers to the sharpness
of vision. Good VA allows drivers to see clearly and quickly identify important information, such as road signs
and other vehicles. Sivak and ­Schoettle37 showed that VA was a significant predictor of night driving performance,
and that older drivers with poorer VA had more difficulty driving at night. Poor VA, on the other hand, can make
it difficult for drivers to see important details, which can increase the risk of accidents. Owsley et al.38 showed that
drivers with poorer VA were more likely to have a crash, and that the risk of a crash increased as VA decreased.
The risk of accidents increased significantly as VA d ­ ecreased39. Besides VA, other visual functions playing an
important role on night driving such as contrast ­sensitivity11, disability ­glare12 and halo ­perception12,40,41 should
be evaluated in order to characterize the role of vision in night driving performances.
For our research, we have chosen mesopic VA as a benchmark. High-contrast, photopic VA has been shown
to be a poor predictor of closed-road night-time d ­ riving11, with mesopic VA being a better predictor of night-
11,12
time driving than standard photopic ­measures . However, there are limitations to measurements of mesopic
VA, which can be unreliable due to variations in chart luminance at these low light l­evels42,43. Previous studies
demonstrated repeatability of mesopic VA at 0.75 cd/m244,45 or about 2.0 log units lower than the photopic t­ est44.
However, measures at 0.75 cd/m2 are clinically more significant than those at higher mesopic luminance levels as
3 cd/m244. Our results showed similar reproducibility limits for 1.1 cd/m2 (0.10 95% LoA for Night VA). Moreover,
nighttime road lighting standards are between 0.3 and 2.0 cd/m246. For all these reasons, mesopic luminance of
1.1 cd/m2 as an intermediate value of lighting standards and inferior by 2.0 log units than photopic level (181 cd/
m2), seems to be an adequate first choice for mesopic VA assessment for both age categories. Photopic VA chart
luminance levels in this research follows standards for subjective ­refraction47 (ISO NORM 10938:2016) usually
conducted on clinical practice. Nevertheless, higher levels of VA chart for determining Std Rx which are not
evaluated in this study would give more insight on shift in refraction from photopic to mesopic conditions.
Our results agree with those of the study by J­ ohnson25, who measured VA under several different low lumi-
nance levels using sinusoidal gratings. Even with night myopia correction, VA was significantly decreased in

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 6

Vol:.(1234567890)
www.nature.com/scientificreports/

mesopic conditions compared to photopic conditions. This result strongly suggests that factors other than refrac-
tion influence VA under low luminance conditions. It has been suggested that decreases in VA occur at lower
light levels because of neural factors associated with decreasing retinal illuminance, more than optical blur from
myopic ­shifts22,25,26. Our results further support this idea, changing light environment degrades binocular VA by
2 lines on average independently of the subject’s Rx shift. Moreover, Rx shift compensation leads to a VA gain
of 1 line on average which increases with the Rx shift, myopic Rx shifts higher than − 0.50 D improve by twice
the VA in mesopic conditions compared to lower myopic Rx shifts (− 0.25 to − 0.37 D). Population shows a high
inter-individual variability in VA shifts, a loss up to 4 VA lines due to change in light environment and a gain
with the Night Rx up to 2 VA lines. Both age groups show VA improvement with Night Rx. However, difference
in mesopic VA with Night Rx is of 2 letters lower for presbyopes. Other factors as gender do not influence VA
gain when Rx shift is compensated.
Quality of vision at nighttime depends not only on VA, but PA needs to be considered. In line with previ-
ous research in photopic conditions on dioptric power s­ ensitivity16, our results show that PA adjustment does
influence subjective refraction by improving visual perception and comfort compared to the MPMVA endpoint
in mesopic conditions. Subjects were able to benefit of PA adjustment which is reflected by a Rx shift refining
in more precise steps (0.12 D) than the clinical standards (0.25 D). PA impacts Rx shift and improves quality of
vision independently of VA gain produced by the Rx shift compensation and other physiological factors as age,
gender, or ametropia. Benefit of Night Rx is appreciated subjectively. For those subjects who present myopic Rx
shift, Night Rx is preferred over Std Rx (82.1%), and for half of the subjects, the difference was seen as moderate to
important. Subjective benefit increases for higher myopic Rx shifts or higher mesopic VA gains. However, the high
inter-individual variability in the perceived benefit evidences a variability in inter-individual mesopic sensitivity.
Low myopic Rx shift and low mesopic VA gain due to Rx shift compensation may also lead to a high subjec-
tive benefit and vice versa, high changes in visual performances may be perceived as slighter benefit. The high
inter-individual variability of the Rx shift compensation evidence differences in subject’s sensitivity in mesopic
conditions. Thus, it is desirable to custom mesopic subjective refraction according to individual’s specific needs.
From our data, we have observed that refraction coefficients in photopic conditions are stable over time; Mean
differences between visits were close to zero and 95% LoA are within the expected limits based on previous stud-
ies assessing repeatability of conventional subjective Rx, from ± 0.48 D and ± 0.63 D for the M (SE) c­ oefficient48
and from ± 0.16 D to ± 0.50 D in astigmatic coefficients. Therefore, we can conclude that Std Rx is reproducible
and no significant biological evolution of subject ametropia occurred.
In our understanding there is no previous research on reproducibility of subjective refraction in mesopic
conditions. The results of this study are in line with those in the literature for photopic ­conditions48. Rosenfield
and ­Chiu49 who assessed repeatability of subjective refraction under ideal conditions such as a single masked
examiner, same timeslot of day and 5 repeated times, showed 95% LoA of ± 0.29 D of M suggesting that a change
of ± 0.50 D should be considered as clinically significant. Results suggest that, when using a guided standardized
protocol in a controlled environment, the reproducibility of mesopic subjective refraction is good and comparable
to previous studies of Rx in photopic conditions, and differences are clinically and statistically not relevant in
optimized experimental conditions as same examiner and same moment of the day.
Although all the procedures were carefully designed and conducted, there are several limitations and future
research lines to mention. First, it would have been more informative to have additional age categories for
validating any distinct results or age-related concerns that may arise among this demographic. However, due
to the insufficient sample size the subjects were analyzed in two main groups based on their presbyopic status,
which were representative for statistical analysis. Second, reproducibility has been assessed in non-presbyopic
population due to potential physiological variations in younger eyes are lower and presbyopic sample was more
limited varying in age categories. Assessment of reproducibility in older subjects should be considered to gain
a more comprehensive understanding of effect of night myopia in this specific population. Third, majority of
the population in this investigation is myope. It would have been of interest to extend night myopia assessment
to higher number of hyperopes and emmetropes for more insight of ametropia effect. Fourth, according to
our research, measuring and comparing VA and perceived visual quality with on real-life night driving driver
scene can help evaluate night myopia. Since there may be multiple causes for night myopia, it is recommended
to include additional visual performance metrics like contrast sensitivity, stereo acuity, and glare halos for a
thorough assessment of how it affects night driving. Fifth, moreover, small variation of chart background lumi-
nance and dark adaptation undertaken period should be investigated due to the potential influence on mesopic
refraction and associated impact on VA and perceived quality of vision.
We used specific refraction instrument to perform 0.01 D precise refraction and adapted screen to attain and
control mesopic luminance level of visual stimuli. However, for clinicians who lack access to specific instrumen-
tation, the protocol can still be implemented using traditional optometric setups such as a manual phoropter or
trial frame with 0.25/0.12 D steps. While it may not result in an exact night refraction, it is possible to perform.
Mesopic VA can be measured without the need for mesopic-vision dedicated screens. However, it is crucial to
regulate the background luminance of the VA chart screen as well as the ambient l­ighting21. If a VA chart pro-
jector is being used, neutral density ­filters21,44 or two crossed linear polarizing sheets can be utilized to decrease
the luminance to the desired mesopic conditions. In this latter case, in some parts of the refractive or sensory
examination, it is not recommended to perform tests that require additional polarized lenses to avoid affecting
the test results. For instance, when adjusting biocular balance refraction, instead of using polarized lenses to
dissociate the vision of each eye, it is recommended to use vertical prisms. This approach ensures accurate results
without any interference from the lenses.
92.2% of subjects in this study showed different correction between photopic and mesopic vision. When
considering the use of different visual corrections for night and day vision, it’s important to carefully weigh
the advantages and practical challenges. One key benefit is the potential for improved night vision. In low light

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 7

Vol.:(0123456789)
www.nature.com/scientificreports/

conditions and with increased glare, visual issues like refractive errors can be ­worsened12,27,28. Specialized correc-
tions for night vision can help people see more clearly in these situations and reduce visual disturbances caused
by headlights and streetlights, making driving at night safer. Additionally, wearing separate corrections for night-
time activities can enhance overall comfort, especially for those with specific visual needs as nighttime driving.
One of the challenges of using separate glasses for day and night vision is the added cost and inconvenience, it
can be difficult to remember to bring the appropriate pair or keep track of both. Moreover, not everyone may
need this approach, as it primarily benefits individuals with specific night vision difficulties or requirements. A
case-by-case decision should be made based on individual visual needs and preferences, with consultation from
an eye care professional to determine suitability. For instance, those subjects with a night myopia of − 0.12 D did
not report any difference between using it or not. Thus, it is suggested that individuals with night myopia ≤ − 0.25
D should correct their distance vision for activities in low-light conditions (see Fig. 2).
Finally, although the visual task analysis in this manuscript is considered from the point of view of night-time
driving, visual tasks are relevant also in other mesopic applications such as marine, aviation, security lighting,
and the safe movement and visual orientation of pedestrians in urban areas as well as bad weather conditions.

Conclusion
Nighttime driving poses unique challenges to drivers, and vision also plays a crucial role in ensuring safety on
the road. A standardized protocol to conduct mesopic refraction may help eye care professionals to address
night vision complaints related to VA loss during night driving. Results prove the accuracy and feasibility of the
Night Rx protocol to identify drivers who presented a different Rx between photopic and mesopic conditions.
Night Rx and Rx shift are reproducible over time and compensation of Rx shift leads to a significant VA gain
and subjective benefit in terms of quality of vision. Conducting a 5-min dark-adaptation, the Night Rx protocol
is optimized in time and feasible for clinical setting. There is a high prevalence of the Rx shift and a high inter-
individual variability, thus the need of customized measurement. In conclusion, drivers with night myopia
should consider wearing corrective lenses to improve their visual performance in nighttime driving and reduce
the risk of accidents on the road.

Methods
Sample
This study included 115 subjects (82 non-presbyopes and 33 presbyopes). Inclusion criteria were age (between 18
and 65 years old), best-corrected distance visual acuity (BCDVA) of + 0.10 logMAR or better, distance spherical
equivalent between − 8.00 and + 6.00 diopters (D), cylindrical error up to − 3.00 D, anisometropia up to 1.50 D,
and driving at night regularly (at least 4 days per week). Exclusion criteria were any previous ocular surgery,
pathology, untreated and/or uncontrolled systemic condition, any medical treatment or medication which might
have an influence on vision or interfere with study assessments. Subjects with migraine, epilepsy, binocular vision
problems, permanent wearing of contact lenses, or tinted lenses and with knowledge in optometry, ophthalmic
lenses or optics were also excluded from the study.

Experimental design
This was a monocentric, exploratory, comparative, and controlled study performed at the University of Valen-
cia and approved by the Ethics Committee of Research in Humans of the Ethics Commission in Experimental
Research of University of Valencia and adhered to the tenets of the Declaration of Helsinki. All subjects were
informed of the study characteristics and protocols, and informed consent form was obtained.

Procedure
Subjects were given a general health questionnaire for the inclusion criteria. The same experienced examiner/s
(A.G-S. and M.B-M) performed all the measurements. A preliminary ocular examination including BCDVA,
binocular screening, stereo acuity, and ocular motility test were carried out to assess inclusion requirements.
Objective refraction, ocular aberrometry and pupillometry were performed with Wave Analyzer Medica 700
(WAM700), (Essilor, France). This multifunctional eye diagnostic d ­ evice16 provides in one single measurement
two monocular refraction values for photopic (or “Day”) and mesopic (or “Night”) conditions, giving both low
and high order aberrations at 3-mm and 5-mm pupil apertures. It uses Hartmann-Shack aberrometer. The stimu-
lus luminance was 4 cd/m2 (Konica Minolta LS-110 luminance meter). For each subject, a single measurement
was conducted in a dark room to ensure sufficient pupil width. Objective refractions were recorded in both eyes
and pupil size and spherical aberration were recorded only in the right eye because data from both eyes is similar
­ opulation50. The subjective refraction was evaluated with the Vision-RTM800 (VR-800) automated
in a healthy p
phoropter (Essilor, France) which allows resolution of 0.01 D steps by continuous lens power v­ ariation16. Two
subjective refractions (photopic and mesopic) were performed at 6.0 m under controlled light environment fol-
lowing 3 steps. Firstly, under photopic conditions (lightroom 45 lx), secondly, a dark adaption under scotopic
conditions (dark room < 0.01 lx), and thirdly, under mesopic conditions (lightroom 0.01 lx). Spectral irradiance
measurements were done at eye level using a Konica Minolta CL-200 Chroma Meter illuminance spectropho-
tometer. Randomized ETDRS charts were used to avoid the learning effect in the VA measurement. These charts
were presented on the Vision-C 600 (VC600) screen (Essilor, France), at two different background luminance:
photopic level 181 cd/m2 (standard refraction) and mesopic level 1.1 cd/m2 (night refraction) and letter contrast
(Weber fraction) higher than 90%. In order to attain desired mesopic luminance level, the screen backlight
was decreased and ETDRS charts were adapted with a grey background. Luminance measurements were done
using a ColorCAL MKII Colorimeter. In Supplementary Fig. S1 online shows a flowchart with the four steps of
the refraction protocol. In photopic conditions, objective refraction was used as starting point for the standard

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 8

Vol:.(1234567890)
www.nature.com/scientificreports/

subjective refraction. Monocular fogging/defogging and cross-cylinder technique, biocular and binocular bal-
ance were ­performed16. The most positive refraction that provided the best visual acuity (MPMVA) on ETDRS
chart was considered for both monocular and binocular endpoints. Subject was then dark-adapted for 5 min
(VR800 refractor head’s occluders on and room light off < 0.01 lx). In mesopic conditions, a fogging of + 0.50 D
was applied while subject was looking at the ETDRS chart (mesopic). Then, the MPMVA was obtained by adjust-
ing the binocular sphere from the photopic MPMVA value maintaining the astigmatic component unchanged.
The endpoint of the mesopic refraction was the perceptual appreciation (PA) which is a method to “refine” the
refraction value to give visual comfort to the subject. An algorithm based on the PA test adjusts the mesopic
binocular MPMVA in 0.12 D steps (see Supplementary Fig. S2 online). Subjects compared the binocular MPMVA
with addition of − 0.25 D sphere and graded the difference in slight/moderate or important. Finally, subjects were
able to compare photopic and mesopic refractions on a simulated real-life night-driving scene displayed on the
VC600 under mesopic conditions and graded their quality of vision preference as slight, moderate, or important.
To validate the protocol, reproducibility was tested on a group of subjects from the study (n = 54). After an
average of 60 days following the initial visit, refractive tests were repeated under the same experimental condi-
tions. The results of refraction, the spherical equivalent (M coefficient) and refraction shift (Rx shift), and VA
were compared between the first and second visits.
The metrics derived from each examination are defined as follows:

• Photopic or standard refraction (Std Rx): standard binocular refraction in photopic conditions.
• Mesopic or night refraction (Night Rx): binocular refraction in mesopic conditions.
• Refractive shift (Rx shift): difference in binocular sphere between the night (mesopic) refraction and the
standard (photopic) refraction. It is usually myopic.
• Photopic visual acuity (in logMAR units): binocular VA in photopic conditions wearing the Std Rx.
• Mesopic visual acuity (in logMAR units): binocular VA in mesopic conditions wearing the Night Rx.
• Standard visual acuity shift (in logMAR units): change in binocular VA between mesopic and photopic
lighting conditions wearing the Std Rx.
• Mesopic visual acuity shift (in logMAR units): change in binocular VA in mesopic conditions between the
Night Rx and Std Rx. It is the change in VA due to compensation of the Rx shift.

Statistical analysis
It was performed using the SPSS software for macOS (version 28.0, IBM Corp., USA) and the Statistica soft-
ware for PC (version 14.0.1.25 TIBCO Inc USA) and it was analysed in descriptive terms. Parametric or non-
parametric tests were used according to sample size and/or normality data. The normality was checked using the
Kolmogorov–Smirnov test. For those paired variables that did not follow a normal distribution, the Wilcoxon
sign test was used. For the repeated measures, the non-parametric Friedman test was used, equivalent to the
analysis of variance for variables that do not meet the normality condition. For bi-variants independent samples
the Mann–Whitney test was used and the Kruskal–Wallis for more than two variables.
This study analyzed both objective and subjective refractive data for each eye individually. Pairwise Wilcoxon
tests were conducted to check for differences between the eyes. Since no significant differences were found, this
manuscript reports only the analysis of refractive data from the right eye.
The level of significance was defined as p < 0.05. Reproducibility analysis was used to quantify the contribution
of the number of days elapsed between visits, examiner, and time of the day on refraction components. Limits
of agreement (LoA) were estimated by ± 1.96 standard deviation.

Data availability
The datasets used and/or analysed during the current study will are available from the corresponding author
on reasonable request.

Received: 10 August 2023; Accepted: 29 December 2023

References
1. Perlman, I. & Normann, R. A. Light adaptation and sensitivity controlling mechanisms in vertebrate photoreceptors. Prog. Retin.
Eye Res. 17, 523–563 (1998).
2. Charman, W. N. Night myopia and driving. Ophthalmic Physiol. Opt. 16, 474–485 (1996).
3. Artal, P., Schwarz, C., Cánovas, C. & Mira-Agudelo, A. Night myopia studied with an adaptive optics visual analyzer. PLoS One.
7, e40239 (2012).
4. Chirre, E., Prieto, P. M., Schwarz, C. & Artal, P. Night myopia is reduced in binocular vision. J. Vis. 16, 1–10. https://​doi.​org/​10.​
1167/​16.8.​16 (2016).
5. Lijarcio, I., Useche, S. A., Llamazares, J. & Montoro, L. Are your eyes “on the road”? Findings from the 2019 national study on
vision and driving safety in Spain. Int. J. Environ. Res. Public Health. 17, 3195 (2020).
6. NHTSA. Traffic safety facts passenger vehicle occupant fatalities by day and night: A contrast (DOT HS 810 637) in National center
for statistics and analysis, NHTSA (ed. DOT HS 809923 ed. (National Highway Traffic Safety Administration, US Department of
Transportation, 2007).
7. Wood, J. M. Nighttime driving: Visual, lighting and visibility challenges. Ophthalmic Physiol. Opt. 40, 187–201 (2020).
8. Cohen, Y. et al. Relationship between night myopia and night-time motor vehicle accidents. Acta Ophthalmol. Scand. 85, 367–370
(2007).
9. Plainis, S., Murray, I. J. & Pallikaris, I. G. Road traffic casualties: understanding the night-time death toll. Inj. Prev. 12, 125–128
(2006).
10. Chang, A., Koozekanani, D., Wang, L. & Lee, J. Effect of spherical aberration and night myopia on visual acuity and contrast
sensitivity in a simulated night-driving scenario. Clin. Exp. Optom. 100, 650–657 (2017).

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 9

Vol.:(0123456789)
www.nature.com/scientificreports/

11. Wood, J. M. & Owens, D. A. Standard measures of visual acuity do not predict drivers’ recognition performance under day or
night conditions. Optom. Vis. Sci. 82, 698–705 (2005).
12. Kimlin, J. A., Black, A. A. & Wood, J. M. Nighttime driving in older adults: Effects of glare and association with mesopic visual
function. Investig. Ophthalmol. Vis. Sci. 58, 2796–2803 (2017).
13. Mutti, D. O. et al. Refractive error, axial length, and relative peripheral refractive error before and after the onset of myopia. Investig.
Ophthalmol. Vis. Sci. 48, 2510–2519 (2007).
14. Atowa, U. C., Wajuihian, S. O. & Hansraj, R. A review of paediatric vision screening protocols and guidelines. Int. J. Ophthalmol.
12, 1194–1201 (2019).
15. Kumar, M., Mahaseth, A., Parveen, S., Rafeeq, U. & Chauhan, L. Refractive errors among commercial drivers. Indian J. Ophthalmol.
70, 2112–2116 (2022).
16. Venkataraman, A. P., Sirak, D., Brautaset, R. & Dominguez-Vicent, A. Evaluation of the performance of algorithm-based methods
for subjective refraction. J. Clin. Med. 9, 3144 (2020).
17. Rodríguez-Vallejo, M., Burguera, N., Rocha-de-Lossada, C., Aramberri, J. & Fernández, J. Refraction and defocus curves in eyes
with monofocal and multifocal intraocular lenses. J. Optom. 16, 236–243 (2023).
18. Rodriguez-Lopez, V. & Dorronsoro, C. Beyond traditional subjective refraction. Curr. Opin. Ophthalmol. 33, 228–234 (2022).
19. Kobashi, H. et al. Comparison of subjective refraction under binocular and monocular conditions in myopic subjects. Sci. Rep. 5,
12606. https://​doi.​org/​10.​1038/​srep1​2606 (2015).
20. Barbur, J. L. & Stockman, A. Photopic, mesopic and scotopic vision and changes in visual performance. Encycl. Eye 3, 323–331
(2010).
21. Wood, L. J., Jolly, J. K., Buckley, T. M., Josan, A. S. & MacLaren, R. E. Low luminance visual acuity as a clinical measure and clinical
trial outcome measure: A scoping review. Ophthalmic Physiol. Opt. 41, 213–223 (2021).
22. Hiraoka, T., Hoshi, S., Okamoto, Y., Okamoto, F. & Oshika, T. Mesopic functional visual acuity in normal subjects. PLoS One 10,
e0134505 (2015).
23. Rice, T. M., Peek-Asa, C. & Kraus, J. F. Nighttime driving, passenger transport, and injury crash rates of young drivers. Inj. Prev.
9, 245–250 (2003).
24. Owens, D. A., Wood, J. M. & Owens, J. M. Effects of age and illumination on night driving: a road test. Hum. Factors. 49, 1115–1131
(2007).
25. Johnson, C. A. Effects of luminance and stimulus distance on accommodation and visual resolution. J. Opt. Soc. Am. 66, 138–142
(1976).
26. Arumi, P., Chauhan, K. & Charman, W. N. Accommodation and acuity under night-driving illumination levels. Ophthalmic Physiol.
Opt. 17, 291–299 (1997).
27. Hessler, P., Stinn, M., Dolata, J. & Degle, S. Evaluation of a night spectacle correction concerning an improvement of mesopic
vision quality. Investig. Ophthalmol. Vis. Sci. 58, 1127–1127 (2017).
28. Wood, J. M. et al. Differential effects of refractive blur on day and nighttime driving performance. Investig. Ophthalmol. Vis. Sci.
55, 2284–2289 (2014).
29. Wood, J. M. et al. Refractive blur affects judgement of pedestrian walking direction at night. Ophthalmic. Physiol. Opt. 41, 582–590
(2021).
30. Chamorro, E., Cleva, J. M., Alvarez, M., Subero, M. S. & Garcia, M. Mesopic visual function attained by lenses specifically designed
for mesopic vision. Investig. Ophthalmol. Vis. Sci. 60, 5935–5935 (2019).
31. Atchison, D. A., Guo, H. & Charman, W. N. Subjective blur limits for cylinder and sinusoidal gratings and the effect of myopia.
Optom. Vis. Sci. 84, 904–911 (2007).
32. López-Gil, N., Peixoto-de-Matos, S. C., Thibos, L. N. & Gonzalez-Meijome, J. M. Shedding light on night myopia. J. Vis. 12, 1–9
(2012).
33. van den Berg, T. J., Franssen, L., Kruijt, B. & Coppens, J. E. History of ocular straylight measurement: A review. Z. Med. Phys. 23,
6–20 (2013).
34. Mueller-Schotte, S., van der Schouw, Y. T. & Schuurmans, M. J. Ocular straylight: a determinant of quality of life in the elderly?.
Gerontol. Geriatr. Med. 1, 2333721415610193 (2015).
35. Fernández, J., Rodríguez-Vallejo, M., Martínez, J., Tauste, A. & Piñero, D. P. From presbyopia to cataracts: A critical review on
dysfunctional lens syndrome. J. Ophthalmol. 10, e4318405 (2018).
36. Owens, D. A. & Leibowitz, H. W. Night myopia: Cause and a possible basis for amelioration. Am. J. Optom. Physiol. Opt. 53, 709–717
(1976).
37. Sivak, M. & Schoettle, B. Visual acuity and night driving performance of older drivers. Accid. Anal. Prev. 41, 758–766 (2009).
38. Owsley, C., McGwin, G. Jr. & Ball, K. Visual acuity and crash risk in older adults. J. Am. Geriatr. Soc. 49, 1318–1323 (2001).
39. Owsley, C., McGwin, G. Jr., Ball, K. & Sloane, M. E. Vision impairment, eye disease, and injurious motor vehicle crashes in the
elderly. Ophthalmic Epidemiol. 23, 84–92 (2016).
40. Puell, M. C. & Palomo-Álvarez, C. Effects of light scatter and blur on low-contrast vision and disk halo size. Optom. Vis. Sci. 94,
505–510 (2017).
41. Mainster, M. A. & Turner, P. L. Glare’s causes, consequences, and clinical challenges after a century of ophthalmic study. Am. J.
Ophthalmol. 153, 587–593 (2017).
42. Sturgis, S. P. & Osgood, D. J. Effects of glare and background luminance on visual acuity and contrast sensitivity: Implications for
driver night vision testing. Hum. Factors. 24, 347–360 (1982).
43. Johnson, C. A. & Casson, E. J. Effects of luminance, contrast and blur on visual acuity. Optom. Vis. Sci. 72, 864–869 (1995).
44. Lin, R. J., Ng, J. S. & Nguyen, A. L. Determinants and standardization of mesopic visual acuity. Optom. Vis. Sci. 92, 559–565 (2015).
45. Barrio, A., Antona, B. & Puell, M. C. Repeatability of mesopic visual acuity measurements using high- and low-contrast ETDRS
letter charts. Graefes Arch. Clin. Exp. Ophthalmol. 253, 791–795 (2015).
46. Fryc, I., Czyżewski, D., Fan, J. & Gălăţanu, C. D. The drive towards optimization of road lighting energy consumption based on
mesopic vision-A suburban street case study. Energies 14, 1175 (2021).
47. Livingstone, I. A. et al. Photometric compliance of tablet screens and retro-illuminated acuity charts as visual acuity measurement
devices. PLoS One. 11, e0150676 (2016).
48. MacKenzie, G. E. Reproducibility of sphero-cylindrical prescriptions. Ophthalmic Physiol. Opt. 28, 143–150 (2008).
49. Rosenfield, M. & Chiu, N. N. Repeatability of subjective and objective refraction. Optom. Vis. Sci. 72, 577–579 (1995).
50. Telek, H. H., Erdol, H. & Turk, A. The effects of age on pupil diameter at different light amplitudes. Beyoglu Eye J. 3, 80–85 (2018).

Acknowledgements
The authors would like to thank Juliette Papin and Adèle Longo from EssilorLuxottica R&D Instruments for
their work on instrumentation support.

Scientific Reports | (2024) 14:293 | https://doi.org/10.1038/s41598-023-51062-8 10

Vol:.(1234567890)
www.nature.com/scientificreports/

Author contributions
The authors A.G.S, M.B.M and S.M.S have contributed to the study protocol definition. The authors A.G-S and
M.B.M. have contributed to the acquisition of clinical measurements. The authors A.G.S., M.B.M, D.M.R, and
S.M.S have contributed to the definition of the concept and writing of the first draft of the manuscript. The
authors A.G-S., S.M.S. and I.B-G. have contributed to the reviewing of the manuscript, statistical analysis, and
discussion of the data. The authors A.G-S. and S.M.S. have contributed to supervision. All authors reviewed the
manuscript.

Competing interests
This research was funded by Essilorluxottica.

Additional information
Supplementary Information The online version contains supplementary material available at https://​doi.​org/​
10.​1038/​s41598-​023-​51062-8.
Correspondence and requests for materials should be addressed to G.-S.A.
Reprints and permissions information is available at www.nature.com/reprints.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and
institutional affiliations.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International
License, which permits use, sharing, adaptation, distribution and reproduction in any medium or
format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the
Creative Commons licence, and indicate if changes were made. The images or other third party material in this
article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the
material. If material is not included in the article’s Creative Commons licence and your intended use is not
permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from
the copyright holder. To view a copy of this licence, visit http://​creat​iveco​mmons.​org/​licen​ses/​by/4.​0/.

© The Author(s) 2024