SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017,
ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006
TRESCAL, INC.
554 NW 77th St
Boca Raton, FL 33487
Nathan Thrasher Phone: 561 999 1896
CALIBRATION
Valid To: November 30, 2024 Certificate Number: 1877.02
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to
this laboratory to perform the following calibrations1, 7:
I. Dimensional
Parameter/Equipment Range CMC2 (±) Comments
Micrometers3
Spindle Linearity Up to 48 in 16 µin + 5.5 µin/in Gage blocks
Length Indicators3 Up to 1 in 74 µin Starrett 716 dial
(Drop, Test, LVDTs) indicator calibrator
Height Gages3 Up to 24 in 19 µin + 5.6 µin/in Gage blocks, surface
plate
Calipers3
Outside Diameter Up to 48 in 30 µin + 5.2 µin/in Gage blocks
Step & Depth 31 µin + 5.2 µin/in
(A2LA Cert. No. 1877.02) 05/31/2023 Page 1 of 34
II. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 4, 5, 6 (±) Comments
DC Voltage – Generate3 (0 to 330) mV 0.8 µV + 16 nV/mV Fluke 5520A
330 mV to 3.3 V 17 µV + 5 µV/V
(3.3 to 33) V 88 µV + 7 µV/V
(33 to 330) V 0.56 mV + 14 µV/V
(330 to 1000) V 2.7 mV + 14 µV/V
DC Voltage – Measure Up to 100 mV 0.33 µV + 17 µV/V Agilent 3458A
(0.1 to 1) V 0.52 µV + 8 µV/V
(1 to 10) V 8.8 µV + 8.1 µV/V
(10 to 100) V 35 µV + 10 µV/V
(100 to 1000) V 1.2 mV + 11 µV/V
DC High Voltage – (1 to 20) kV 2.3 % Fluke 80K-40, Agilent
Measure (20 to 35) kV 1.2 % 34401A
(35 to 40) kV 2.3 %
DC Current – Generate3 (0 to 330) µA 16 nA + 0.12 nA/µA Fluke 5520A
330 µA to 3.3 mA 41 nA + 80 nA/mA
(3.3 to 33) mA 0.21 µA + 80 nA/mA
(33 to 330) mA 2.1 µA + 80 nA/mA
330 mA to 1.1 A 33 µA + 0.16 mA/A
(1.1 to 3.0) A 0.37 mA + 0.13 mA/A
(3 to 11) A 0.49 mA + 0.39 mA/A
(11 to 20.5) A 9.4 mA + 0.8 mA/A
DC Clamp Meters (10 to 16.5) A 50 mA + 6.5 mA/A Fluke 5520A w/ 5500
(16.5 to 150) A 0.18 A + 3.4 mA/A coil
(150 to 1025) A 0.83 A + 3.3 mA/A
DC Current – Measure (0 to 100) nA 41 pA + 89 µA/A Agilent 3458A
(0.1 to 1) µA 70 pA + 21 µA/A
(1 to 10) µA 0.7 nA + 18 µA/A
(10 to 100) µA 5.8 nA + 9.2 µA/A
(0.1 to 1) mA 58 nA + 9.7 µA/A
(1 to 10) mA 0.58 µA + 8 µA/A
(10 to 100) mA 5.8 µA + 17 µA/A
(0.1 to 1) A 59 µA + 81 µA/A
(1 to 2) A 30 µA + 0.17 mA/A Fluke 8508A
(2 to 20) A 3.9 mA + 0.26 mA/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 2 of 34
Parameter/Equipment Range CMC2, 4, 6 (±) Comments
DC Resistance – (0 to 10.9999) Ω 1.2 mΩ + 24 µΩ/Ω Fluke 5520A
Generate (11 to 32.9999) Ω 1.5 mΩ + 24 µΩ/Ω
(33 to 109.9999) Ω 1.9 mΩ + 22 µΩ/Ω
(110 to 329.9999) Ω 4.1 mΩ + 22 µΩ/Ω
(330 to 1099.999) Ω 9.1 mΩ + 22 µΩ/Ω
(1.1 to 3.299 999) kΩ 41 mΩ + 22 µΩ/Ω
(3.3 to 10.999 99) kΩ 92 mΩ + 22 µΩ/Ω
(11 to 32.999 99) kΩ 0.41 Ω + 22 µΩ/Ω
(33 to 109.9999) kΩ 0.9 Ω + 22 µΩ/Ω
(110 to 329.9999) kΩ 8.4 Ω + 26 µΩ/Ω
(330 to 1099.999) kΩ 14 Ω + 26 µΩ/Ω
(1.1 to 3.299 999) MΩ 93 Ω + 48 µΩ/Ω
(3.3 to 10.999 99) MΩ 0.4 kΩ + 0.1 mΩ/Ω
(11 to 32.999 99) MΩ 4.4 kΩ + 0.2 mΩ/Ω
(33 to 109.9999) MΩ 16 kΩ + 0.4 mΩ/Ω
(110 to 329.9999) MΩ 0.35 MΩ + 2.4 mΩ/Ω
(330 to 1100) MΩ 4.4 MΩ + 12 mΩ/Ω
DC Resistance – (0 to 10) Ω 55 µΩ + 15 µΩ/Ω Agilent 3458A
Measure (10 to 100) Ω 0.52 mΩ + 13 µΩ/Ω
(0.1 to 1) kΩ 0.53 mΩ + 10 µΩ/Ω
(1 to 10) kΩ 5.3 mΩ + 10 µΩ/Ω
(10 to 100) kΩ 53 mΩ + 11 µΩ/Ω
(0.1 to 1) MΩ 2.3 Ω + 17 µΩ/Ω
(1 to 10) MΩ 0.1 kΩ + 55 µΩ/Ω
(10 to 100) MΩ 1 kΩ + 0.52 mΩ/Ω
(0.1 to 1) GΩ 10 kΩ + 5.1 mΩ/Ω
(1 to 2) GΩ 1.2 MΩ + 1.4 mΩ/Ω Fluke 8508A
(2 to 20) GΩ 12 MΩ + 1.4 mΩ/Ω
(A2LA Cert. No. 1877.02) 05/31/2023 Page 3 of 34
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Calibration of
Thermocouple Generate
& Measure –
Type E (-200 to -100) ºC 0.39 °C Fluke 5520A
(-100 to -25) ºC 0.13 °C
(-25 to 350) ºC 0.12 °C
(350 to 650) ºC 0.13 °C
(650 to 1000) ºC 0.17 °C
Type J (-210 to -100) ºC 0.21 °C
(-100 to -30) ºC 0.13 °C
(-30 to 150) ºC 0.12 °C
(150 to 760) ºC 0.14 °C
(760 to 1200) ºC 0.18 °C
Type K (-200 to -100) ºC 0.26 °C
(-100 to -25) ºC 0.15 °C
(-25 to 120) ºC 0.13 °C
(120 to 1000) ºC 0.21 °C
(1000 to 1372) ºC 0.31 °C
Type N (-200 to -100) ºC 0.31 °C
(-100 to -25) ºC 0.18 °C
(-25 to 120) ºC 0.15 °C
(120 to 410) ºC 0.15 °C
(410 to 1300) ºC 0.21 °C
Type R (0 to 250) ºC 0.44 °C
(250 to 400) ºC 0.28 °C
(400 to 1000) ºC 0.26 °C
(1000 to 1767) ºC 0.31 °C
Type S (0 to 250) ºC 0.37 °C
(250 to 1000) ºC 0.28 °C
(1000 to 1400) ºC 0.29 °C
(1400 to 1767) ºC 0.36 °C
Type T (-250 to -150) ºC 0.49 °C
(-150 to 0) ºC 0.21 °C
(0 to 120) ºC 0.15 °C
(120 to 400) ºC 0.14 °C
(A2LA Cert. No. 1877.02) 05/31/2023 Page 4 of 34
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Simulation of
RTD Indicators3–
Pt 385, 100 Ω (-200 to -80) °C 0.057 °C Fluke 5520A
(-80 to 0) °C 0.056 °C
(0 to 100) °C 0.068 °C
(100 to 300) °C 0.081 °C
(300 to 400) °C 0.088 °C
(400 to 630) °C 0.10 °C
(630 to 800) °C 0.19 °C
Pt 385, 200 Ω (-200 to -80) °C 0.051 °C
(-80 to 0) °C 0.051 °C
(0 to 100) °C 0.051 °C
(100 to 260) °C 0.057 °C
(260 to 300) °C 0.10 °C
(300 to 400) °C 0.11 °C
(400 to 600) °C 0.12 °C
(600 to 630) °C 0.13 °C
Pt 385, 500 Ω (-200 to -80) °C 0.051 °C
(-80 to 0) °C 0.056 °C
(0 to 100) °C 0.056 °C
(100 to 260) °C 0.062 °C
(260 to 300) °C 0.074 °C
(300 to 400) °C 0.074 °C
(400 to 600) °C 0.081 °C
(600 to 630) °C 0.095 °C
Pt 385, 1000 Ω (-200 to -80) °C 0.047 °C
(-80 to 0) °C 0.047 °C
(0 to 100) °C 0.051 °C
(100 to 260) °C 0.057 °C
(260 to 300) °C 0.062 °C
(300 to 400) °C 0.068 °C
(400 to 600) °C 0.068 °C
(600 to 630) °C 0.18 °C
(-200 to -190) °C 0.20 °C
Pt 3916, 100 Ω (-190 to -80) °C 0.051 °C
(-80 to 0) °C 0.056 °C
(0 to 100) °C 0.062 °C
(100 to 260) °C 0.068 °C
(260 to 300) °C 0.074 °C
(300 to 400) °C 0.081 °C
(400 to 600) °C 0.041 °C
(600 to 630) °C 0.18 °C
(A2LA Cert. No. 1877.02) 05/31/2023 Page 5 of 34
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Simulation of
RTD Indicators3 – (cont)
Pt 3926, 100 Ω
(-200 to -80) °C 0.056 °C Fluke 5520A
(-80 to 0) °C 0.056 °C
(0 to 100) °C 0.068 °C
(100 to 300) °C 0.081 °C
(300 to 400) °C 0.088 °C
(400 to 630) °C 0.10 °C
PtNi 385, 120 Ω
(-80 to 0) °C 0.062 °C
(0 to 100) °C 0.062 °C
(100 to 260) °C 0.11 °C
Cu 427, 10 Ω
(-100 to 260) °C 0.23 °C
(A2LA Cert. No. 1877.02) 05/31/2023 Page 6 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Voltage – Generate3
(1 to 33) mV (10 to 45) Hz 5.5 µV + 0.64 µV/mV Fluke 5520A
45 Hz to 10 kHz 5 µV + 0.12 µV/mV
(10 to 20) kHz 5 µV + 0.16 µV/mV
(20 to 50) kHz 5.6 µV + 0.8 µV/mV
(50 to 100) kHz 12 µV + 2.8 µV/mV
(100 to 200) kHz 47 µV + 6.4 µV/mV
(200 to 500) kHz 47 µV + 6.4 µV/mV
(33 to 330) mV (10 to 45) Hz 14 µV + 0.24 µV/mV
45 Hz to 10 kHz 24 µV + 87 nV/mV
(10 to 20) kHz 11 µV + 0.13 µV/mV
(20 to 50) kHz 35 µV + 0.28 µV/mV
(50 to 100) kHz 47 µV + 0.64 µV/mV
(100 to 200) kHz 0.11 mV + 1.6 µV/mV
(200 to 500) kHz 0.11 mV + 1.6 µV/mV
330 mV to 3.3 V (10 to 45) Hz 0.12 mV + 0.24 mV/V
45 Hz to 10 kHz 88 µV + 0.12 mV/V
(10 to 20) kHz 99 µV + 0.15 mV/V
(20 to 50) kHz 0.12 mV + 0.24 mV/V
(50 to 100) kHz 0.29 mV + 0.56 mV/V
(100 to 200) kHz 1.1 mV + 1.9 mV/V
(200 to 500) kHz 1.1 mV + 1.9 mV/V
(3.3 to 33) V (10 to 45) Hz 1.3 mV + 0.24 mV/V
45 Hz to 10 kHz 0.94 mV + 0.12 mV/V
(10 to 20) kHz 1.6 mV + 0.18 mV/V
(20 to 50) kHz 1.4 mV + 0.28 mV/V
(50 to 100) kHz 3.7 mV + 0.72 mV/V
(33 to 330) V (10 to 45) Hz 6.8 mV + 0.15 mV/V
45 Hz to 10 kHz 10 mV + 0.16 mV/V
(10 to 20) kHz 12 mV + 0.2 mV/V
(20 to 50) kHz 13 mV + 0.24 mV/V
(50 to 100) kHz 93 mV + 1.6 mV/V
(330 to 1020) V 45 Hz to 1 kHz 87 mV + 0.24 mV/V
(1 to 5) kHz 74 mV + 0.2 mV/V
(5 to 10) kHz 88 mV + 0.24 mV/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 7 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Voltage3 – Measure
Up to 100 mV (1 to 40) Hz 6.1 µV + 0.0003 mV/mV Agilent 3458A
40 Hz to 1 kHz 4.1 µV + 0.0002 mV/mV
(1 to 20) kHz 4.2 µV + 0.0003 mV/mV
(20 to 50) kHz 4.9 µV + 0.001 mV/mV
(50 to 100) kHz 9 µV + 0.005 mV/mV
(100 to 300) kHz 45 µV + 0.04 mV/mV
(1 to 40) Hz 12 µV + 0.000 07 mV/mV
(0.1 to 1) V 40 Hz to 1 kHz 2.7 µV + 0.000 07 mV/mV
(1 to 20) kHz 11 µV + 0.000 14 mV/mV
(20 to 50) kHz 012 µV + 0.0003 mV/mV
(50 to 100) kHz 17 µV + 0.0008 mV/mV
(100 to 300) kHz 47 µV mV + 0.003 mV/mV
(0.3 to 1) MHz 0.12 mV + 0.01 mV/mV
(1 to 2) MHz 3.8 mV + 0.01 mV/mV
(1 to 10) V
(1 to 40) Hz 0.12 mV + 0.000 07 V/V
40 Hz to 1 kHz 0.1 mV + 0.000 07 V/V
(1 to 20) kHz 0.11 mV + 0.000 14 V/V
(20 to 50) kHz 0.12 mV + 0.0003 V/V
(50 to 100) kHz 0.17 mV + 0.0008 V/V
(100 to 300) kHz 0.47 mV + 0.003 V/V
(0.3 to 1) MHz 1.2 mV + 0.010 V/V
(1 to 2) MHz 38 mV + 0.015 V/V
(10 to 100) V
(1 to 40) Hz 1.2 mV + 0.000 07 V/V
40 Hz to 1 kHz 1 mV + 0.000 07 V/V
(1 to 20) kHz 1 mV + 0.000 14 V/V
(20 to 50) kHz 1.2 mV + 0.0003 V/V
(50 to 100) kHz 1.7 mV + 0.0008 V/V
(100 to 300) kHz 4.7 mV + 0.003 V/V
(0.3 to 1) MHz 12 mV + 0.01 V/V
(1 to 2) MHz 0.38 V + 0.015 V/V
(100 to 1000) V (1 to 40) Hz 13 mV + 0.0002 V/V
40 Hz to 1 kHz 11 mV + 0.0002 V/V
(1 to 20) kHz 11 mV + 0.0002 V/V
(20 to 50) kHz 13 mV + 0.000 35 V/V
(50 to 100) kHz 21 mV + 0.0012 V/V
(100 to 300) kHz 57 mV + 0.004 V/V
(0.3 to 1) MHz 0.17 V + 0.015 V/V
(1 to 40) Hz 98 mV + 0.0004 V/V
40 Hz to 1 kHz 84 mV + 0.0004 V/V
(1 to 20) kHz 98 mV V + 0.0006 V/V
(20 to 50) kHz 0.14 V + 0.0012 V/V
(50 to 100) kHz 0.27 V + 0.003 V/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 8 of 34
Parameter/Range Frequency CMC2, 4, 5, 6 (±) Comments
AC High Voltage – Measure (1 to 40) kV 5.8 % Fluke 80K-40,
Agilent 34401A
AC Current – Generate3
(29 to 329.99) µA (10 to 20) Hz 0.13 µA + 1.6 nA/µA Fluke 5520A
(20 to 45) Hz 0.12 µA + 1.2 nA/µA
45 Hz to 1 kHz 0.11 µA + 1 nA/µA
(1 to 5) kHz 0.19 µA + 2.4 nA/µA
(5 to 10) kHz 0.35 µA + 6.4 nA/µA
(10 to 30) kHz 0.70 µA + 13 nA/µA
330 µA to 3.2999 mA (10 to 20) Hz 0.65 µA + 1.6 µA/mA
(20 to 45) Hz 0.45 µA + 1 µA/mA
45 Hz to 1 kHz 0.39 µA + 0.8 µA/mA
(1 to 5) kHz 0.69 µA + 1.6 µA/mA
(5 to 10) kHz 2.9 µA + 1.6 µA/mA
(10 to 30) kHz 3.4 µA + 8 µA/mA
(3.3 to 32.999) mA (10 to 20) Hz 6.4 µA + 1.4 µA/mA
(20 to 45) Hz 4 µA + 0.72 µA/mA
45 Hz to 1 kHz 2.7 µA + 0.32 µA/mA
(1 to 5) kHz 3.7 µA + 0.64 µA/mA
(5 to 10) kHz 7.7 + 1.6 µA/mA
(10 to 30) kHz 14 µA + 3.2 µA/mA
(33 to 329.99) mA (10 to 20) Hz 65 µA + 1.4 µA/mA
(20 to 45) Hz 42 µA + 0.72 µA/mA
45 Hz to 1 kHz 29 µA + 0.31 µA/mA
(1 to 5) kHz 68 µA + 0.8 µA/mA
(5 to 10) kHz 0.13 mA + 1.6 µA/mA
(10 to 30) kHz 0.27 mA + 3.2 µA/mA
330 mA to 1.099 99 A (10 to 45) Hz 0.56 mA + 1.4 mA/A
45 Hz to 1 kHz 0.22 mA + 0.4 mA/A
(1 to 5) kHz 2.4 mA + 4.8 mA/A
(5 to 10) kHz 11 mA + 20 mA/A
(1.1 to 2.999 99) A (10 to 45) Hz 1.7 mA + 1.4 mA/A
45 Hz to 1 kHz 0.61 mA + 0.48 mA/A
(1 to 5) kHz 6.1 mA + 4.8 mA/A
(5 to 10) kHz 26 mA + 20 mA/A
(3 to 10.9999) A (45 to 100) Hz 3.1 mA + 0.48 mA/A
100 Hz to 1 kHz 4 mA + 0.8 mA/A
(1 to 5) kHz 74 mA + 24 mA/A
(11 to 20.5) A (45 to 100) Hz 15 mA + 0.96 mA/A
100 Hz to 1 kHz 17 mA + 1.2 mA/A
(1 to 5) kHz 0.27 A + 24 mA/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 9 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Current – Generate3
(cont)
Toroidal
(10 to 16.5) A (45 to 65) Hz 26 mA + 2.1 mA/A Fluke 5520A w/
(16.5 to 150) A 50 mA + 1.9 mA/A 5500coil
(150 to 1025) A 0.34 A + 1.9 mA/A
(10 to 16.5) A (65 to 440) Hz 60 mA + 6 mA/A
(16.5 to 150) A 0.11 A + 5.3 mA/A
(150 to 1025) A 0.86 A + 5.3 mA/A
Non-Toroidal
(10 to 16.5) A (45 to 65) Hz 60 mA + 3.8 mA/A
(16.5 to 150) A 0.23 A + 3.7 mA/A
(150 to 1025) A 1.2 A + 3.7 mA/A
(10 to 16.5) A (65 to 440) Hz 90 mA + 7.2 mA/A
(16.5 to 150) A 0.28 A + 6.7 mA/A
(150 to 1025) A 1.6 A + 6.7 mA/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 10 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Current3 – Measure
(5 to 100) µA (10 to 20) Hz 57 nA + 4 mA/A Agilent 3458A
(20 to 45) Hz 45 nA + 1.5 mA/A
45 Hz to 5 kHz 40 nA + 0.64 mA/A
(0.1 to 1) mA (10 to 20) Hz 0.43 µA + 0.0041 mA/mA
(20 to 45) Hz 0.31 µA + 0.0016 mA/mA
(45 to 100) Hz 0.26 µA + 0.000 67 mA/mA
(0.1 to 5) kHz 0.25 µA + 0.000 37 mA/mA
(5 to 20) kHz 0.26 µA + 0.000 67 mA/mA
(20 to 50) kHz 0.63 µA + 0.0042 mA/mA
(50 to 100) kHz 1.8 µA + 0.0056 mA/mA
(1 to 10) mA (10 to 20) Hz 2.4 µA + 0.0041 mA/mA
(20 to 45) Hz 2.3 µA + 0.0016 mA/mA
(45 to 100) Hz 2.3 µA + 0.000 67 mA/mA
(0.1 to 5) kHz 2.3 µA + 0.000 37 mA/mA
(5 to 20) kHz 2.3 µA + 0.000 67 mA/mA
(20 to 50) kHz 4.5 µA + 0.0041 mA/mA
(50 to 100) kHz 16 µA + 0.0056 mA/mA
(10 to 100) mA (10 to 20) Hz 42 µA + 0.004 mA/mA
(20 to 45) Hz 30 µA + 0.0015 mA/mA
(45 to 100) Hz 25 µA + 0.0006 mA/mA
(0.1 to 5) kHz 24 µA + 0.0003 mA/mA
(5 to 20) kHz 25 µA + 0.0006 mA/mA
(20 to 50) kHz 62 µA + 0.004 mA/mA
(50 to 100) kHz 0.18 mA + 0.0055 mA/mA
(0.1 to 1) A (10 to 20) Hz 0.42 mA + 0.0041 A/A
(20 to 45) Hz 0.3 mA + 0.0017 A/A
(45 to 100) Hz 0.26 mA + 0.0009 A/A
(0.1 to 5) kHz 0.27 mA + 0.0011 A/A
(5 to 20) kHz 0.37 mA + 0.0031 A/A
(20 to 50) kHz 0.92 mA + 0.01 A/A
(1 to 2) A 10 Hz to 2 kHz 0.42 mA + 0.5 mA/A Fluke 8508A
(2 to 10) kHz 0.3 mA + 0.66 mA/A
(10 to 30) kHz 0.96 mA + 2.1 mA/A
(2 to 20) A 10 Hz to 2 kHz 2.4 mA + 0.71 mA/A
(2 to 10) kHz 2.7 mA + 1.9 mA/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 11 of 34
Parameter/Equipment Range CMC2, 4, 5, 6 (±) Comments
AC Power3 – Generate Up to 1020 V
(45 to 65) Hz, PF=1 Up to 33 W 0.1 % Fluke 5520A
(33 to 330) W 0.062 %
330 W to 1.1 kW 0.072 %
(1.1 to 3) kW 0.075 %
(3 to 11) kW 0.087 %
(11 to 20.9) kW 0.14 %
DC Power3 – Generate Up to 1020 V
Up to 3 W 0.017 % Fluke 5520A
(3 to 30) W 0.014 %
(30 to 300) W 0.014 %
300 W to 3 kW 0.033 %
(3 to 20.9) kW 0.085 %
Parameter/Range Frequency CMC2, 4, 5, 6 (±) Comments
Phase Generate3 – (10 to 65) Hz 0.1° Fluke 5520A
(0 to 360)° (65 to 500) Hz 0.26°
500 Hz to 1 kHz 0.51°
(1 to 5) kHz 2.6°
(5 to 10) kHz 5.1°
(10 to 30) kHz 10°
Capacitance3 – Generate
(0.19 to 0.3999) nF 10 Hz to 10 kHz 8.8 pF + 4 pF/nF Fluke 5520A
(0.4 to 1.0999) nF 10 Hz to 10 kHz 9.7 pF + 4 pF/nF
(1.1 to 3.2999) nF 10 Hz to 3 kHz 13 pF + 4 pF/nF
(3.3 to 10.9999) nF 10 Hz to 1 kHz 15 pF + 2 pF/nF
(11 to 32.9999) nF 10 Hz to 1 kHz 0.1 nF + 2 pF/nF
(33 to 109.999) nF 10 Hz to 1 kHz 0.15 nF + 2 pF/nF
(110 to 329.999) nF 10 Hz to 1 kHz 0.46 nF + 2 pF/nF
(0.33 to 1.099 99) μF (10 to 600) Hz 1.5 nF + 2 nF/µF
(1.1 to 3.299 99) μF (10 to 300) Hz 4.6 nF + 2 nF/µF
(3.3 to 10.9999) μF (10 to 150) Hz 15 nF + 2 nF/µF
(11 to 32.9999) μF (10 to 120) Hz 59 nF + 3.2 nF/µF
(33 to 109.999) μF (10 to 80) Hz 0.2 µF + 3.6 nF/µF
(110 to 329.999) μF (0 to 50) Hz 0.64 µF + 3.6 nF/µF
(0.33 to 1.099 99) mF (0 to 20) Hz 2 µF + 3.6 µF/mF
(1.1 to 3.299 99) mF (0 to 6) Hz 7 µF + 3.5 µF/mF
(3.3 to 10.9999) mF (0 to 2) Hz 38 µF + 2.6 µF/mF
(11 to 32.9999) mF (0 to 0.6) Hz 90 µF + 6 µF/mF
(33 to 110) mF (0 to 0.2) Hz 0.37 mF + 8.8 µF/mF
(A2LA Cert. No. 1877.02) 05/31/2023 Page 12 of 34
Parameter/Equipment Range CMC2, 4, 5, 6 (±) Comments
Oscilloscopes3 –
Volts – DC Signal
Into 50 Ω ± 0 mV to 6.6 V 33 µV + 2 mV/V Fluke 5520A
Into 1 MΩ ± 0 mV to 130 V 33 µV + 0.39 mV/V
Volts – Squarewave
Into 50 Ω 1 mVpp to 6.6 Vpp; 91 µV + 4.3 mV/V
Into 1 MΩ 1 mVpp to 130 Vpp;
10 Hz to 1 kHz 0.2 mV + 0.8 mV/V
(1 to 100) kHz 0.24 mV + 2 mV/V
Edge Into 50 Ω ≤ 300 ps (Up to 2 MHz) 82 ps
≤ 350 ps (Above 2 MHz) 82 ps
Leveled Sine Wave 50 kHz Reference 0.33 mV + 17 mV/V
50 kHz to 100 MHz 0.4 mV + 31 mV/V
(100 to 300) MHz 0.43 mV + 36 mV/V
(300 to 600) MHz 0.53 mV + 57 mV/V
(600 to 1100) MHz 0.58 mV + 67 mV/V
Time Marker 5 s to 50 ms 0.000 63 % + T x 0.1 % T = time in seconds
50 ms to 1 ns 0.000 52 %
IV. Fluid Quantities
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Liquid Flow – (0.7 to 180) GPM 1.2 % Blancett 110-500
Measure3 (lin)/B2800 turbine
flow meter
(A2LA Cert. No. 1877.02) 05/31/2023 Page 13 of 34
V. Mechanical
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Mass – Measure, 1 mg 2.9 µg Substitution method
Fixed Points 2 mg 2.9 µg using mass standards
3 mg 3.4 µg with mass comparators
5 mg 2.9 µg
10 mg 2.9 µg
20 mg 2.9 µg
30 mg 3.4 µg
50 mg 2.9 µg
100 mg 2.9 µg
200 mg 2.9 µg
300 mg 3.4 µg
500 mg 2.9 µg
1g 6.5 µg
2g 6.5 µg
3g 9 µg
5g 6.5 µg
10 g 9.4 µg
20 g 14 µg
30 g 17 µg
50 g 24 µg
100 g 46 µg
200 g 96 µg
300 g 0.15 mg
500 g 0.32 mg
1 kg 0.48 mg
2 kg 0.66 mg
3 kg 1.3 mg
4 kg 1.3 mg
5 kg 3.9 mg
10 kg 7.3 mg
20 kg 0.11 g
Mass – Measure, Up to 5.1 g 0.43 µg + 0.000 28 % UMX5 comparator
Direct Weigh (> 5.1 to 520) g 0.33 mg XP505 comparator
(> 0.52 to 2.3) kg 1.7 mg XP2004S comparator
(> 2.3 to 5.1) kg 5.4 mg XP5003S comparator
(> 5.1 to 32) kg 66 mg XPE32003LC
comparator
(A2LA Cert. No. 1877.02) 05/31/2023 Page 14 of 34
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Balances Up to 10 kg 0.87 x R Class Ultra, 0 & 4
(10 to 32) kg 1.1 x R Weights
R = Unit under test
resolution
Scales Up to 3000 lbs 0.82 x R Class F weights
R = Unit under test
resolution
Scales – Substitution (3000 to 6000) lbs 1.2 x R EURAMET CG-18
Testing
R = Unit under test
resolution
Pressure – Measuring
Equipment3
Pneumatic Gage Up to 10 000 psi 0.024 % DH instruments
Pressure RPM4/HPMS
A70MS/A20MS-AF
reference pressure
monitor
(-14.7 to 15) psi 0.012 % Mensor CPC6000
(15 to 300) psi 0.012 %
(300 to 750) psi 0.012 %
(750 to 1500) psi 0.024 %
Hydraulic Gage Up to 36 000 psi 43 psi Additel ADT681-10-
Pressure GP2500-BAR pressure
gage
Torque Measure – Torque 5 lbf·in to 600 lbf·ft 0.31 % CDI torque transducers
Tools
(A2LA Cert. No. 1877.02) 05/31/2023 Page 15 of 34
VI. Thermodynamic
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Infrared Temperature – (-15 to 12) °C 1.2 °C Fluke 4180
Measuring Equipment (-12 to -9) °C 1.1 °C
(-9 to -6) °C 1 °C
(-6 to -2) °C 0.94 °C
(-2 to 0) °C 0.82 °C
(0 to 120) °C 0.87 °C + 0.01 °C/°C
(35 to 500) °C 0.62 °C + ΔTChange from 35 Fluke 4181
°C x 0.017
Temperature – Measure (-197 to 660) °C 0.027 °C Fluke 2560, 5628 PRT
Temperature Measuring (-80 to 100) °C 0.018 °C + ΔTChange from - Fluke 5628
Equipment 40 °C x 0.00001 w/reference
baths/blocks
(100 to 300) °C 0.020 °C + ΔTChange from
100 °C x 0.000079
(300 to 400) °C 0.032 °C + ΔTChange from
300 °C x 0.0001
(400 to 660) °C 0.078 °C + ΔTChange from
400 °C x 0.000 034
Relative Humidity – (10 to 80) % RH 0.63 % RH Thunder Scientific
Measuring Equipment 2500LT
Relative Humidity (10 to 70) % RH 1.9 % RH Fluke DewK 1620
Measuring Equipment – (> 70 to 100) % RH 3.5 % RH
Calibration of Humidity
Probes3
(A2LA Cert. No. 1877.02) 05/31/2023 Page 16 of 34
VII. Time & Frequency
Parameter/Equipment Range CMC2, 5, 8 (±) Comments
Frequency – Generate 100 µHz to 15 MHz 0.0024 % HP 33120A function
generator
Frequency – Measure 1 Hz to 225 MHz 0.000 59 % HP 53132A counter
Stopwatches & Timers3 0 s to 24 hr 0.047 s Witschi QT6000
Optical Tachometers3 (1 to 100 000) rpm 0.00035 rpm + 0.000 038 Fluke 5520A, LED
rpm/rpm
(A2LA Cert. No. 1877.02) 05/31/2023 Page 17 of 34
SATELLITE LOCATION
TRESCAL, INC.
100 N. Babcock St
Melborne, FL 33935
Nathan Thrasher Phone: 321 622 1553
I. Device Specific Parameters
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Ionizing Fans, Blowers &
Guns
Ion Balance Up to 50 V 0.28 V Ion System 210 charge
plate monitor, HP 34401A
Discharge Time Up to 30 sec 0.039 s
Electrostatic Discharge
Mats – Resistance Up to 1 GΩ 12 % Desco 19787 surface
resistance meter
Soldering Systems –
Measure
Leak Voltage Up to 40 mV 0.7 mV Hakko FG-101 solder
tester
Tip-to-Ground Ω Up to 40 Ω 0.7 Ω
Temperature Stability (32 to 1300) °F 7 °F
(A2LA Cert. No. 1877.02) 05/31/2023 Page 18 of 34
II. Dimensional
Parameter/Equipment Range CMC2 (±) Comments
Micrometers3
Spindle Linearity Up to 20 in 16 µin + 5.9 µin/in Gage blocks
Anvil Flatness 50 µin 6.2 µin Optical flat
Parallelism 50 µin 11 µin Optical flat
Depth Micrometer Base Up to 200 µin 31 µin Gage blocks
Flatness
Calipers3
Outside Diameter Up to 20 in 30 µin + 5.2 µin/in Gage blocks
Inside Diameter Up to 20 in 34 µin + 5.1 µin/in
Step & Depth Up to 20 in 31 µin + 5.2 µin/in
Height Gages3 Up to 24 in 18 µin + 5.7 µin/in Gage blocks
Gage Amp w/ Probe3 Up to 0.005 in 30 µin Gage blocks, Tesa TT-
80 electronic length
measuring instrument
Length & Plain Diameter3 Trimos TULM210
(Length Standards, Cylindrical Up to 4 in length measuring
Plugs & Pins, Feeler Gages) instrument
Steel 34 µin + 1.2 µin/in
Aluminum 34 µin + 1.4 µin/in
Length 3 Up to 20 in 33 µin + 4.2 µin/in Gage blocks, Tesa TT-
80 electronic length
measuring instrument
Dial & Digital Indicators3 Up to 1 in 49 µin Itall 700 dial indicator
calibrator
(1 to 2) in 18 µin + 5.2 µin/in Gage blocks
(A2LA Cert. No. 1877.02) 05/31/2023 Page 19 of 34
Parameter/Equipment Range CMC2 (±) Comments
Threaded Plug Gages,
Go & NoGo3
60° Pitch Diameter Up to 4 in 82 µin Trimos TULM210
(4.5 to 80 TPI) length measuring
instrument, Vermont
Major Diameter Up to 4 in 34 µin + 1.2 µin/in gage thread wire set
Protractors3 (0 to 180)° 0.0077° Sine bar, gage
blocks, surface plate
III. Dimensional Testing/Calibration
Parameter/Equipment Range CMC2 (±) Comments
Dimensional
Measurements3–
Checking Angle Flatness: Up to 2000 µin 33 µin + 3 µin/in Tesa TT-80
Plates, V-Blocks, electronic length
Fixtures & Parallelism: Up to 2000 µin 33 µin + 3 µin/in measuring
Workpieces instrument
IV. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 4, 6 (±) Comments
DC Voltage – Generate3 (0 to 330) mV 0.8 µV + 16 nV/mV Fluke 5520A
330 mV to 3.3 V 17 µV + 5 µV/V
(3.3 to 33) V 88 µV + 7 µV/V
(33 to 330) V 0.56 mV + 14 µV/V
(330 to 1000) V 2.7 mV + 14 µV/V
DC Voltage – Measure Up to 100 mV 0.33 µV + 17 µV/V Agilent 3458A
(0 to 1) V 0.52 µV + 8 µV/V
(0 to 10) V 8.8 µV + 8.1 µV/V
(0 to 100) V 35 µV + 10 µV/V
(0 to 1000) V 1.2 mV + 11 µV/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 20 of 34
Parameter/Equipment Range CMC2, 4, 5, 6 (±) Comments
DC Current – Generate (0 to 330) µA 16 nA + 0.12 nA/µA Fluke 5520A
330 µA to 3.3 mA 41 nA + 80 nA/mA
(3.3 to 33) mA 0.21 µA + 80 nA/mA
(33 to 330) mA 2.1 µA + 80 nA/mA
330 mA to 1.1 A 33 µA + 0.16 mA/A
(1.1 to 3.0) A 0.37 mA + 0.13 mA/A
(3 to 11) A 0.49 mA + 0.39 mA/A
(11 to 20.5) A 9.4 mA + 0.8 mA/A
DC Current – Measure (0 to 100) nA 41 pA + 89 µA/A HP 3458A
(0.1 to 1) µA 70 pA + 21 µA/A
(1 to 10) µA 0.7 nA + 18 µA/A
(10 to 100) µA 5.8 nA + 9.2 µA/A
(0.1 to 1) mA 58 nA + 9.7 µA/A
(1 to 10) mA 0.58 µA + 8 µA/A
(10 to 100) mA 5.8 µA + 17 µA/A
(0.1 to 1) A 59 µA + 81 µA/A
(1 to 200) A 0.0046 % Ohm-Lab CS-200
shunt, HP 34401A
(200 to 1000) A 0.18 % Empro B-1000-100
shunt, HP 34401A
Resistance – Generate (0 to 10.9999) Ω 1.2 mΩ + 24 µΩ/Ω Fluke 5520A
(11 to 32.9999) Ω 1.5 mΩ + 24 µΩ/Ω
(33 to 109.9999) Ω 1.9 mΩ + 22 µΩ/Ω
(110 to 329.9999) Ω 4.1 mΩ + 22 µΩ/Ω
(330 to 1099.999) Ω 9.1 mΩ + 22 µΩ/Ω
(1.1 to 3.299 999) kΩ 41 mΩ + 22 µΩ/Ω
(3.3 to 10.999 99) kΩ 92 mΩ + 22 µΩ/Ω
(11 to 32.999 99) kΩ 0.41 Ω + 22 µΩ/Ω
(33 to 109.9999) kΩ 0.9 Ω + 22 µΩ/Ω
(110 to 329.9999) kΩ 8.4 Ω + 26 µΩ/Ω
(330 to 1099.999) kΩ 14 Ω + 26 µΩ/Ω
(1.1 to 3.299 999) MΩ 93 Ω + 48 µΩ/Ω
(3.3 to 10.999 99) MΩ 0.4 kΩ + 0.1 mΩ/Ω
(11 to 32.999 99) MΩ 4.4 kΩ + 0.2 mΩ/Ω
(33 to 109.9999) MΩ 16 kΩ + 0.4 mΩ/Ω
(110 to 329.9999) MΩ 0.35 MΩ + 2.4 mΩ/Ω
(330 to 1100) MΩ 4.4 MΩ + 12 mΩ/Ω
(A2LA Cert. No. 1877.02) 05/31/2023 Page 21 of 34
Parameter/Equipment Range CMC2, 4, 6 (±) Comments
Resistance – Measure (0 to 10) Ω 55 µΩ + 15 µΩ/Ω HP 3458A
(10 to 100) Ω 0.52 mΩ + 13 µΩ/Ω
(1.1 to 1) kΩ 0.53 mΩ + 10 µΩ/Ω
(1 to 10) kΩ 5.3 mΩ + 10 µΩ/Ω
(10 to 100) kΩ 53 mΩ + 11 µΩ/Ω
(1.1 to 1) MΩ 2.3 Ω + 17 µΩ/Ω
(1 to 10) MΩ 0.1 kΩ + 55 µΩ/Ω
(10 to 100) MΩ 1 kΩ + 0.52 mΩ/Ω
(1.1 to 1) GΩ 10 kΩ + 5.1 mΩ/Ω
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Voltage –
Generate3
(1 to 33) mV (10 to 45) Hz 5.5 µV + 0.64 µV/mV Fluke 5520A
45 Hz to 10 kHz 5 µV + 0.12 µV/mV
(10 to 20) kHz 5 µV + 0.16 µV/mV
(20 to 50) kHz 5.6 µV + 0.8 µV/mV
(50 to 100) kHz 12 µV + 2.8 µV/mV
(100 to 200) kHz 47 µV + 6.4 µV/mV
(200 to 500) kHz 47 µV + 6.4 µV/mV
(33 to 330) mV (10 to 45) Hz 14 µV + 0.24 µV/mV
45 Hz to 10 kHz 24 µV + 87 nV/mV
(10 to 20) kHz 11 µV + 0.13 µV/mV
(20 to 50) kHz 35 µV + 0.28 µV/mV
(50 to 100) kHz 47 µV + 0.64 µV/mV
(100 to 200) kHz 0.11 mV + 1.6 µV/mV
(200 to 500) kHz 0.11 mV + 1.6 µV/mV
330 mV to 3.3 V (10 to 45) Hz 0.12 mV + 0.24 mV/V
45 Hz to 10 kHz 88 µV + 0.12 mV/V
(10 to 20) kHz 99 µV + 0.15 mV/V
(20 to 50) kHz 0.12 mV + 0.24 mV/V
(50 to 100) kHz 0.29 mV + 0.56 mV/V
(100 to 200) kHz 1.1 mV + 1.9 mV/V
(200 to 500) kHz 1.1 mV + 1.9 mV/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 22 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Voltage –
Generate3 (cont)
(3.3 to 33) V (10 to 45) Hz 1.3 mV + 0.24 mV/V Fluke 5520A
45 Hz to 10 kHz 0.94 mV + 0.12 mV/V
(10 to 20) kHz 1.6 mV + 0.18 mV/V
(20 to 50) kHz 1.4 mV + 0.28 mV/V
(50 to 100) kHz 3.7 mV + 0.72 mV/V
(33 to 330) V (10 to 45) Hz 6.8 mV + 0.15 mV/V
45 Hz to 10 kHz 10 mV + 0.16 mV/V
(10 to 20) kHz 12 mV + 0.2 mV/V
(20 to 50) kHz 13 mV + 0.24 mV/V
(50 to 100) kHz 93 mV + 1.6 mV/V
(330 to 1020) V 45 Hz to 1 kHz 87 mV + 0.24 mV/V
(1 to 5) kHz 74 mV + 0.2 mV/V
(5 to 10) kHz 88 mV + 0.24 mV/V
AC Voltage3 –
Measure
Up to 100 mV (1 to 40) Hz 6.1 µV + 0.0003 mV/mV Agilent 3458A
40 Hz to 1 kHz 4.1 µV + 0.0002 mV/mV
(1 to 20) kHz 4.2 µV + 0.0003 mV/mV
(20 to 50) kHz 4.9 µV + 0.001 mV/mV
(50 to 100) kHz 9 µV + 0.005 mV/mV
(100 to 300) kHz 45 µV + 0.04 mV/mV
(0.1 to 1) V (1 to 40) Hz 12 µV + 0.000 07 mV/mV
40 Hz to 1 kHz 2.7 µV + 0.000 07 mV/mV
(1 to 20) kHz 11 µV + 0.000 14 mV/mV
(20 to 50) kHz 012 µV + 0.0003 mV/mV
(50 to 100) kHz 17 µV + 0.0008 mV/mV
(100 to 300) kHz 47 µV mV + 0.003 mV/mV
(0.3 to 1) MHz 0.12 mV + 0.01 mV/mV
(1 to 2) MHz 3.8 mV + 0.01 mV/mV
(1 to 10) V (1 to 40) Hz 0.12 mV + 0.000 07 V/V
40 Hz to 1 kHz 0.1 mV + 0.000 07 V/V
(1 to 20) kHz 0.11 mV + 0.000 14 V/V
(20 to 50) kHz 0.12 mV + 0.0003 V/V
(50 to 100) kHz 0.17 mV + 0.0008 V/V
(100 to 300) kHz 0.47 mV + 0.003 V/V
(0.3 to 1) MHz 1.2 mV + 0.01 V/V
(1 to 2) MHz 38 mV + 0.015 V/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 23 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Voltage3 –
Measure (cont)
(10 to 100) V (1 to 40) Hz 1.2 mV + 0.000 07 V/V Agilent 3458A
40 Hz to 1 kHz 1 mV + 0.000 07 V/V
(1 to 20) kHz 1 mV + 0.000 14 V/V
(20 to 50) kHz 1.2 mV + 0.0003 V/V
(50 to 100) kHz 1.7 mV + 0.0008 V/V
(100 to 300) kHz 4.7 mV + 0.003 V/V
(0.3 to 1) MHz 12 mV + 0.01 V/V
(1 to 2) MHz 0.38 V + 0.015 V/V
(100 to 1000) V (1 to 40) Hz 13 mV + 0.0002 V/V
40 Hz to 1 kHz 11 mV + 0.0002 V/V
(1 to 20) kHz 11 mV + 0.0002 V/V
(20 to 50) kHz 13 mV + 0.000 35 V/V
(50 to 100) kHz 21 mV + 0.0012 V/V
(100 to 300) kHz 57 mV + 0.004 V/V
(0.3 to 1) MHz 0.17 V + 0.015 V/V
(1 to 40) Hz 98 mV + 0.0004 V/V
40 Hz to 1 kHz 84 mV + 0.0004 V/V
(1 to 20) kHz 98 mV V + 0.0006 V/V
(20 to 50) kHz 0.14 V + 0.0012 V/V
(50 to 100) kHz 0.27 V + 0.003 V/V
(A2LA Cert. No. 1877.02) 05/31/2023 Page 24 of 34
Parameter/Range Frequency CMC2, 4, 6 (±) Comments
AC Current – Generate3
(29 to 329.99) µA (10 to 20) Hz 0.13 µA + 1.6 nA/µA Fluke 5520A
(20 to 45) Hz 0.12 µA + 1.2 nA/µA
45 Hz to 1 kHz 0.11 µA + 1 nA/µA
(1 to 5) kHz 0.19 µA + 2.4 nA/µA
(5 to 10) kHz 0.35 µA + 6.4 nA/µA
(10 to 30) kHz 0.70 µA + 13 nA/µA
330 µA to 3.2999 mA (10 to 20) Hz 0.65 µA + 1.6 µA/mA
(20 to 45) Hz 0.45 µA + 1 µA/mA
45 Hz to 1 kHz 0.39 µA + 0.8 µA/mA
(1 to 5) kHz 0.69 µA + 1.6 µA/mA
(5 to 10) kHz 2.9 µA + 1.6 µA/mA
(10 to 30) kHz 3.4 µA + 8 µA/mA
(3.3 to 32.999) mA (10 to 20) Hz 6.4 µA + 1.4 µA/mA
(20 to 45) Hz 4 µA + 0.72 µA/mA
45 Hz to 1 kHz 2.7 µA + 0.32 µA/mA
(1 to 5) kHz 3.7 µA + 0.64 µA/mA
(5 to 10) kHz 7.7 + 1.6 µA/mA
(10 to 30) kHz 14 µA + 3.2 µA/mA
(33 to 329.99) mA (10 to 20) Hz 65 µA + 1.4 µA/mA
(20 to 45) Hz 42 µA + 0.72 µA/mA
45 Hz to 1 kHz 29 µA + 0.31 µA/mA
(1 to 5) kHz 68 µA + 0.8 µA/mA
(5 to 10) kHz 0.13 mA + 1.6 µA/mA
(10 to 30) kHz 0.27 mA + 3.2 µA/mA
330 mA to 1.099 99 A (10 to 45) Hz 0.56 mA + 1.4 mA/A
45 Hz to 1 kHz 0.22 mA + 0.4 mA/A
(1 to 5) kHz 2.4 mA + 4.8 mA/A
(5 to 10) kHz 11 mA + 20 mA/A
(1.1 to 2.999 99) A (10 to 45) Hz 1.7 mA + 1.4 mA/A
45 Hz to 1 kHz 0.61 mA + 0.48 mA/A
(1 to 5) kHz 6.1 mA + 4.8 mA/A
(5 to 10) kHz 26 mA + 20 mA/A
(3 to 10.9999) A (45 to 100) Hz 3.1 mA + 0.48 mA/A
100 Hz to 1 kHz 4 mA + 0.8 mA/A
(1 to 5) kHz 74 mA + 24 mA/A
(45 to 100) Hz 15 mA + 0.96 mA/A
(11 to 20.5) A 100 Hz to 1 kHz 17 mA + 1.2 mA/A
(1 to 5) kHz 0.27 A + 24 mA/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 25 of 34
Parameter/Range Frequency CMC2, 5, 6 (±) Comments
AC Current3 – Measure
(5 to 100) µA (10 to 20) Hz 57 nA + 4 mA/A Agilent 3458A
(20 to 45) Hz 45 nA + 1.5 mA/A
45 Hz to 5 kHz 40 nA + 0.64 mA/A
(0.1 to 1) mA (10 to 20) Hz 0.43 µA + 0.0041 mA/mA
(20 to 45) Hz 0.31 µA + 0.0016 mA/mA
(45 to 100) Hz 0.26 µA + 0.000 67 mA/mA
(0.1 to 5) kHz 0.25 µA + 0.000 37 mA/mA
(5 to 20) kHz 0.26 µA + 0.000 67 mA/mA
(20 to 50) kHz 0.63 µA + 0.0042 mA/mA
(50 to 100) kHz 1.8 µA + 0.0056 mA/mA
(1 to 10) mA (10 to 20) Hz 2.4 µA + 0.0041 mA/mA
(20 to 45) Hz 2.3 µA + 0.0016 mA/mA
(45 to 100) Hz 2.3 µA + 0.000 67 mA/mA
(0.1 to 5) kHz 2.3 µA + 0.000 37 mA/mA
(5 to 20) kHz 2.3 µA + 0.000 67 mA/mA
(20 to 50) kHz 4.5 µA + 0.0041 mA/mA
(50 to 100) kHz 16 µA + 0.0056 mA/mA
(10 to 100) mA (10 to 20) Hz 42 µA + 0.004 mA/mA
(20 to 45) Hz 30 µA + 0.0015 mA/mA
(45 to 100) Hz 25 µA + 0.0006 mA/mA
(0.1 to 5) kHz 24 µA + 0.0003 mA/mA
(5 to 20) kHz 25 µA + 0.0006 mA/mA
(20 to 50) kHz 62 µA + 0.004 mA/mA
(50 to 100) kHz 0.18 mA + 0.0055 mA/mA
(0.1 to 1) A (10 to 20) Hz 0.42 mA + 0.0041 A/A
(20 to 45) Hz 0.3 mA + 0.0017 A/A
(45 to 100) Hz 0.26 mA + 0.0009 A/A
(0.1 to 5) kHz 0.27 mA + 0.0011 A/A
(5 to 20) kHz 0.37 mA + 0.0031 A/A
(20 to 50) kHz 0.92 mA + 0.01 A/A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 26 of 34
Parameter/Equipment Range CMC2, 4, 6 (±) Comments
Capacitance3 –
Generate
(0.19 to 0.3999) nF 10 Hz to 10 kHz 8.8 pF + 4 pF/nF Fluke 5520A
(0.4 to 1.0999) nF 10 Hz to 10 kHz 9.7 pF + 4 pF/nF
(1.1 to 3.2999) nF 10 Hz to 3 kHz 13 pF + 4 pF/nF
(3.3 to 10.9999) nF 10 Hz to 1 kHz 15 pF + 2 pF/nF
(11 to 32.9999) nF 10 Hz to 1 kHz 0.1 nF + 2 pF/nF
(33 to 109.999) nF 10 Hz to 1 kHz 0.15 nF + 2 pF/nF
(110 to 329.999) nF 10 Hz to 1 kHz 0.46 nF + 2 pF/nF
(0.33 to 1.099 99) μF (10 to 600) Hz 1.5 nF + 2 nF/µF
(1.1 to 3.299 99) μF (10 to 300) Hz 4.6 nF + 2 nF/µF
(3.3 to 10.9999) μF (10 to 150) Hz 15 nF + 2 nF/µF
(11 to 32.9999) μF (10 to 120) Hz 59 nF + 3.2 nF/µF
(33 to 109.999) μF (10 to 80) Hz 0.2 µF + 3.6 nF/µF
(110 to 329.999) μF (0 to 50) Hz 0.64 µF + 3.6 nF/µF
(0.33 to 1.099 99) mF (0 to 20) Hz 2 µF + 3.6 µF/mF
(1.1 to 3.299 99) mF (0 to 6) Hz 7 µF + 3.5 µF/mF
(3.3 to 10.9999) mF (0 to 2) Hz 38 µF + 2.6 µF/mF
(11 to 32.9999) mF (0 to 0.6) Hz 90 µF + 6 µF/mF
(33 to 110) mF (0 to 0.2) Hz 0.37 mF + 8.8 µF/mF
(A2LA Cert. No. 1877.02) 05/31/2023 Page 27 of 34
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Simulation of
Thermocouple–
Type J (-210 to -100) ºC 0.21 °C Fluke 5520A
(-100 to -30) ºC 0.13 °C
(-30 to 150) ºC 0.12 °C
(150 to 760) ºC 0.14 °C
(760 to 1200) ºC 0.18 °C
Type K
(-200 to -100) ºC 0.26 °C
(-100 to -25) ºC 0.15 °C
(-25 to 120) ºC 0.13 °C
(120 to 1000) ºC 0.21 °C
(1000 to 1372) ºC 0.31 °C
Type R
(0 to 250) ºC 0.44 °C
(250 to 400) ºC 0.28 °C
(400 to 1000) ºC 0.26 °C
(1000 to 1767) ºC 0.31 °C
Type S
(0 to 250) ºC 0.37 °C
(250 to 1000) ºC 0.28 °C
(1000 to 1400) ºC 0.29 °C
Type T (1400 to 1767) ºC 0.36 °C
(-250 to -150) ºC 0.49 °C
(-150 to 0) ºC 0.21 °C
(0 to 120) ºC 0.15 °C
(120 to 400) ºC 0.14 °C
(A2LA Cert. No. 1877.02) 05/31/2023 Page 28 of 34
Parameter/Equipment Range CMC2, 4, 5, 6 (±) Comments
Oscilloscopes3 –
Volts – DC Signal
Into 50 Ω ± 0 mV to 6.6 V 33 µV + 2 mV/V Fluke 5520A
Into 1 MΩ ± 0 mV to 130 V 33 µV + 0.39 mV/V
Volts – Squarewave
Into 50 Ω 1 mVpp to 6.6 Vpp; 91 µV + 4.3 mV/V
Into 1 MΩ 1 mVpp to 130 Vpp;
10 Hz to 1 kHz 0.2 mV + 0.8 mV/V
(1 to 100) kHz 0.24 mV + 2 mV/V
Edge into 50 Ω ≤ 300 ps (Up to 2 MHz) 82 ps
≤ 350 ps (Above 2 MHz) 82 ps
Leveled Sine Wave 50 kHz Reference 0.33 mV + 17 mV/V
50 kHz to 100 MHz 0.4 mV + 31 mV/V
(100 to 300) MHz 0.43 mV + 36 mV/V
(300 to 600) MHz 0.53 mV + 57 mV/V
(600 to 1100) MHz 0.58 mV + 67 mV/V
Time Marker 5 s to 50 ms 0.000 63 % + T x 0.1 % T = time in seconds
50 ms to 1 ns 0.000 52 %
V. Electrical – RF/Microwave
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
RF Microwave, Absolute 0.05 to 10V pk-pk
Power – Generate Up to 100 kHz 0.4 % HP 33120A
100 kHz to 1 MHz 0.42 %
(1 to 15) MHz 0.55 %
(-100 to 15) dBm 1.7 dB HP 8648C
100 kHz to 3.2 GHz
RF Microwave, Absolute (-30 to 20) dBm 3.7 % Agilent E4419B w/
Power – Measure3 100 kHz to 4.2 GHz HP 8482A
(A2LA Cert. No. 1877.02) 05/31/2023 Page 29 of 34
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Distortion – Measure 20 Hz to 50 kHz 1.2 dB HP 8903B
(50 to 100) kHz 2.3 dB
100 kHz to 2.9 GHz 2.7 dB HP 8594E
VI. Mechanical
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Accelerometers
(1 to 10) g (0.5 to 1) Hz 1.6 % VR9500, PCB 301A11
(> 1 to 10) Hz 1.3 % reference sensor
(11 to 99) Hz 1.6 %
100 Hz 1.3 %
(101 to 920) Hz 1.7 %
921 Hz to 5 kHz 2%
(5 to 10) kHz 2.4 %
Force– Measuring
Equipment
Crimpers/Banding (0 to 120) lbf 0.62 lb DMC PT-100A pull
Tools tester
Force Gages (0 to 100) lbf 0.0023 %
Class F weights
Pneumatic Pressure – (> 0 to 5) psiv 0.06 % Ashcroft AQS & HQS
Measuring Equipment (> 0 to 60) psia 0.06 % pressure modules
(> 0 to 30) psig 0.061 %
(> 0 to 100) psig 0.061 %
(> 0 to 500) psig 0.061 %
(> 0 to 250) inH20 0.06 %
(> 0 to 500) mmHg 0.033 %
Scales Up to 200 lbs 0.82 x R Class F weights
R = Unit under test
resolution
(A2LA Cert. No. 1877.02) 05/31/2023 Page 30 of 34
Parameter/Equipment Range CMC2, 5, 6, 8 (±) Comments
Balances
Resolution 0.0001 g ≤ 500 mg 1.2 x R Class S-1 weights
> 500 mg to 1 g 1.4 x R
(> 1 to 2) g 1.7 x R R = Unit under test
(> 2 to 5) g 2.2 x R resolution
(> 5 to 10) g 3.0 x R
> 10 g to 2 kg 4.1 x R
Resolution 0.001 g ≤ 20 g 0.91 x R
(> 20 to 50) g 1.1 x R
(> 50 to 100) g 1.4 x R
(> 100 to 200) g 2.5 x R
> 200 g to 2 kg 3.6 x R
Resolution 0.01 g ≤ 100 g 0.82 x R
(> 100 to 200) g 0.85 x R
(> 200 to 500) g 1.0 x R
> 500 g to 2 kg 1.4 x R
Resolution 0.1 g, 1 g, ≤ 2 kg 0.82 x R
& 10 g
Torque Transducers Up to 250 lbf-ft 0.012 % Class F weights,
calibration arms
Torque – Measuring (1 to 10) ozf-in 1.2 % AWS torque transducers
Equipment (10 to 50) ozf-in 0.58 % & CDI torque calibration
system
Snap-Action & 50 ozf-in to 250 lbf-ft 0.31 %
Breakover Wrenches,
Screwdrivers,
Electronic, Digital &
Dial Indicating
VII. Optical Quantities
Parameter/Equipment Range CMC2, 5 (±) Comments
Luminance Meters (0.5 to 1700) fL 2.6 % Hoffman Engineering
LS-65-8C
(A2LA Cert. No. 1877.02) 05/31/2023 Page 31 of 34
VIII. Thermodynamics
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Temperature – (-80 to 350) °C 0.13 °C Fluke 1523, 5626 PRT,
Measuring Equipment calibration bath
(-80 to 350) °C 0.14 °C + ΔTChange from Fluke 1523, 5618B PRT,
-80 °C x 0.0002 calibration bath
Temperature – Measure (-196 to 661) °C 0.018 °C + ΔTChange Fluke 1523, 5626 PRT
from -196 °C x 0.000 033
(-196 to 500) °C 0.029 °C + ΔTChange Fluke 1523, 5618B PRT
from -196 °C x 0.000 28
(662 to 1000) °C 3.2 °C + ΔTChange from TC probe
662 °C x 0.011
Relative Humidity – (20 to 85) % RH 1.5 % RH Vaisala M170 w/ HMP-
Measure & Generate (85 to 90) % RH 2.2 % RH 77(B) probe
Temperature – (-50 to 100) °C 0.86 °C Fluke 2635A Hydrabucket
Environmental w/Type T SLE
Chambers, Temperature thermocouple
Uniformity Survey
(A2LA Cert. No. 1877.02) 05/31/2023 Page 32 of 34
Radiation Thermometry Source using CI Systems & SBIR calibration standards & contact
thermometry method
CI Systems CI System Santa Barbara IR Sources
SR800N, SR800R SR800 Source
Sources
Temperature CMC2, 6, 8 (±) CMC2, 6, 8 (±) CMC2, 6, 8 (±)
-40 °C 0.0086 °C 0.32 °C
-30 °C 0.0087 °C 0.32 °C
-20 °C 0.0088 °C 0.32 °C
-10 °C 0.0089 °C 0.32 °C
0 °C 0.009 °C 0.33 °C 0.015 °C
10 °C 0.009 °C 0.35 °C 0.017 °C
20 °C 0.0091 °C 0.36 °C 0.018 °C
30 °C 0.0092 °C 0.37 °C 0.020 °C
40 °C 0.0093 °C 0.39 °C 0.021 °C
50 °C 0.0094 °C 0.40 °C 0.023 °C
60 °C 0.0095 °C 0.41 °C 0.025 °C
70 °C 0.0096 °C 0.42 °C 0.026 °C
80 °C 0.0097 °C 0.44 °C 0.028 °C
90 °C 0.0098 °C 0.44 °C 0.029 °C
100 °C 0.0098 °C 0.44 °C 0.031 °C
110 °C 0.0099 °C 0.44 °C
120 °C 0.01 °C 0.44 °C
130 °C 0.01 °C 0.44 °C
140 °C 0.01 °C 0.44 °C
150 °C 0.01 °C 0.44 °C
IX. Time & Frequency
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Frequency – Measuring
Equipment
Time Base 10 MHz 0.59 mHz 58305A GPS receiver, HP
53132A frequency counter
Frequency – Measure 0.1 Hz to 225 MHz 74 nHz + 0.1 µHz/Hz HP 53132A frequency
counter
(A2LA Cert. No. 1877.02) 05/31/2023 Page 33 of 34
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Frequency – Generate Up to 15 MHz 44 µHz + 0.12 µHz/Hz HP 33120A function
generator
100 kHz to 3.2 GHz 5.8 Hz + 0.1 µHz/Hz HP 8648C signal generator
Stopwatch Up to 24 Hr 0.039 s 58305A GPS receiver, HP
53132A frequency counter
___________________________
1
This laboratory offers commercial calibration services and field calibration services.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC uncertainty found
on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of
calibration and for other possible adverse effects such as those caused by transportation of the calibration
equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g.
resolution) must also be considered and this, on its own, could result in the actual measurement
uncertainty achievable on a customer’s site being larger than the CMC uncertainty.
4
The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
5
In the statement of CMC, all percentages are percent of reading unless otherwise indicated. In the statement
of CMC, R represent the resolution of the unit under test.
6
Uncertainty components that can be reasonably attributed to the Unit Under Test have not been utilized in
the calculation of the CMC value for this measurement parameter.
7
This scope meets A2LA’s P112 Flexible Scope Policy.
8
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
(A2LA Cert. No. 1877.02) 05/31/2023 Page 34 of 34
Accredited Laboratory
A2LA has accredited
TRESCAL, INC.
Boca Raton, FL
for technical competence in the field of
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of
ANSI/NCSL Z540-1-1994, ANSI/NCSL Z540.3-2006 and R205 – Specific Requirements: Calibration Laboratory Accreditation Program.
This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality
management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
Presented 31st day of May 2023.
_______________________
Mr. Trace McInturff, Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 1877.02
Valid to November 30, 2024
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.