CTC - CW - Drewgard Xta-R201
CTC - CW - Drewgard Xta-R201
CTC - CW - Drewgard Xta-R201
turning the eyepiece (rubber hood) either clockwise or counter-clockwise. Be careful not to overturn the focusing mechanism lock. YES NO NO
4. Look at the scale through the eyepiece. The upper field of the view will be seen as blue and the lower field will be seen as white (Fig. 3).
5. Confirm that the boundary line crosses the scale at “0” (Fig. 4). If not, gently loosen the set screw on the calibration ring and turn the it 4 5 6
clockwise or counter-clockwise until the boundary line is at zero. Gently tighten the set screw. Do not over-tighten as the boundary line might 3 3 10
9
10
9
7 7
shift slightly. When the calibration is complete, gently wipe the prism dry using soft tissue paper. 2 2
8
7
6
8
7
6
6 6
GENERAL USE: The sample must be at ambient temperature prior to testing. 1 1 Brix % 5
4
5
4
20°C 5 5
3 3
1. Repeat steps 1 through 3 from above using sample solution instead of distilled water. 0 0 2
1
2
1
4 4
2. Look at the scale through the eyepiece, the upper field of the view will be seen as blue while the lower field will be seen as white (Fig. 5). The
0 0
3 3
reading is taken at the point where the boundary line of the blue and white fields crosses the scale (Fig. 6).
3. The DREWGARD XTA concentration in percentage (%) equals: THE SCALE READING * 2. 6
4. After each measurement, the sample must be cleaned from the prism using soft tissue paper and water.
MAKEUP WATER CHLORIDE TEST PROCEDURE WEEKLY TEST – MAKEUP WATER CHLORIDE
Note: If the sample is colored or turbid, filter before running the test. If the sample remains cloudy after the first filtering, the sample should again be filtered through the same paper since the filter becomes more retentive on the second filtration.
1a. For samples < 100 ppm chlorides: Rinse the Plastic Vial and fill to the mark (10 ml) with the sample to be tested (Fig. 1).
1b. For samples > 100 ppm chloride: Rinse the tall Glass Vial (test tube) and fill to the mark (2 ml) with the sample to be tested (Fig. 1). 1 2
2. Neutralize the sample by adding 3 drops of Phenolphthalein Indicator to the sample. Swirl to mix. If the sample turns PINK add Sulfuric Acid
OR &
N/10 drop by drop while swirling until the sample turns CLEAR. Afterwards, add one more drop (Fig. 2). OR
3. Adjust the color of the sample by adding 6 drops of Potassium Chromate. The sample will be YELLOW (Fig. 3).
10 ml
4. Counting the drops, add Silver Nitrate N/10 drop by drop. Swirl between drops until the sample turns ORANGE (Fig. 4). 2
ml
10 #
5. Calculate the chloride concentration: 10 ml 10 ml 10 ml
If plastic vial (< 100 ppm) was used in Step 4: - Number of drops of Silver Nitrate N/10 * 10 = ppm chloride
If glass test tube (> 100 ppm) was used in Step 4: - Number of drops of Silver Nitrate N/10 * 50 = ppm chloride
3 4
TEST RESULTS
Chloride 100 ppm (See Note 1) Maximum Limit #
Note 1: If manufacturer’s recommendations are more stringent, take those limits into account. 10 ml 10 ml
MAKEUP WATER HARDNESS TEST PROCEDURE WEEKLY TEST – MAKEUP WATER CHLORIDE
Note: If the sample is colored or turbid, filter before running the test. If the sample remains cloudy after the first filtering, the sample should again be filtered through the same paper since the filter becomes more retentive on the second filtration.
1. Rinse and fill the Sample Cup to the 25 ml mark with makeup water sample (Fig. 1).
1 2 Slide Over 3 Snap
2. Slide the flexible rubber end of the Valve Assembly over the tapered tip of the Titret to the white Reference Line for a snug fit (Fig. 2). Valve
Valve
3. Snap the tip of the Titret (ampoule) at the Score Mark (Fig. 3). ml
Score Score Assembly
25 Assembly
4. With the tip of the valve assembly immersed in the sample, squeeze the bead valve briefly to add a small amount of sample to the Titret 20 Mark Mark
(Fig. 4). The red indicator in the Valve Assembly will also be added to the Titret. 15
10
NOTE: Do not squeeze the bead value unless the tip of the Valve Assembly is immersed below the surface of the liquid. 5
Reference
Line
5. Rock the Titret to mix the contents. The contents of the Titret will turn a BLUE color. Titret Titret
6. Continue to add small amounts of sample until the liquid in the Titret turns from BLUE to PINK. Be sure to rock the Titret to mix the
contents after each addition of sample. When the color of the liquid in the Titret changes to PINK, the end point has been reached. Stop the
test, hold the Titret in a vertical position and carefully read the depth scale on the Titret that corresponds to the liquid level at the bottom 4 55
of the concave meniscus (Fig. 5). Results are in ppm total hardness as calcium carbonate CaCO3. Squeeze 26
Bead Value
28
Valve Read = 28
Here 30
TEST RESULTS
Hardness 170 ppm Maximum Limit
DREWGARD XTA: OPTISCOPE-10 (PCN 1AB2940) Beaker 100 ml Plastic: Beaker 100 ml - Clear Plastic (PCN 0247016)
Chlorides: Chloride LMP Test Kit (PCN 0373019) Filter Paper: Filter Paper - 100/BX (PCN 0225012)
Hardness: Total Hardness Titrets (PCN 0378019) Funnel: Funnel - Plastic (PCN 0221010)
Stirring Rod: Stirring Rod - 150 mm Plastic (PCN 0417015)