1395 03
1395 03
1395 03
SIMCO ELECTRONICS
783 N. Grove Road, Suite 106
Richardson, TX 75081
Erwin Kanins Phone: 972 994 0120
CALIBRATION
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 6:
I. Dimensional
Thread Plugs –
UN Major Diameter:
Up to 1 in 10 µin Pratt & Whitney
(1 to 12) in (4.8 + 5L) µin LabMaster™ Universal
and gage blocks
Steel Rulers and Tape Up to 40 in 0.0014 in + 43L µin Linear scale calibrator
Measures
10 Ω 24 µΩ/Ω
19 Ω 25 µΩ/Ω
1 kΩ 8.6 µΩ/Ω
1.9 kΩ 8.0 µΩ/Ω
100 kΩ 10 µΩ/Ω
190 kΩ 10 µΩ/Ω
Resistance3 – Generate
(cont.)
Electrical Simulation
of Thermocouples3 –
AC Voltage3–
Generate
AC Voltage3 –
Generate (cont)
AC Voltage3 –
Measure
AC Voltage3 –
Measure (cont)
AC Voltage3 –
Measure (cont)
AC Voltage3 –
Measure (cont)
AC Flatness3 –
Measure
AC Flatness3 –
Measure (cont)
AC Current3 –
Generate
AC Current – Clamp
Meters3–
AC Current3 –
Measure
Oscilloscope3 –
Oscilloscope3 – (cont)
Inductance3 – Measure
(100 to 200) μH 1 kHz, 120 Hz 0.58 % Gen Rad 1657 RLC bridge
200 µH to 200 H 0.25 %
Capacitance3 – Generate
Capacitance3 –
Generate (cont)
Fixed Points
10 pF 1 MHz 0.12 %
2 MHz 0.12 %
3 MHz 0.12 %
4 MHz 0.12 %
5 MHz 0.12 %
10 MHz 0.12 %
13 MHz 0.12 %
Capacitance3 –
Measure
(1.0 to 200) µF 120 Hz, 1 kHz 0.25 % GenRad 1657 RLC bridge
200 µF to 1 mF 1.2 %
RF Power3 – Generate
(+20 to -100) dBm 2.5 MHz to 26.5 GHz 0.6 dB HP 8340B synthesizer
w/ HP 8902 audio
analyzer; HP 11722A,
HP 11792A sensors, HP
1793A converter
RF Power3 – Measure
(-20 to + 30) dBm 100 kHz to 2.6 GHz 0.063 dB + 1 digit HP 8902A measuring
50 MHz to 26.5 GHz 0.13 dB + 1 digit receiver w/ HP 11792A
power sensor
Tuned RF Power –
Measure (Relative)
(+10 to 0) dB (0.15 to 1300) MHz 0.035 dB + 1 digit HP 8902A measuring
(0 to -10) dB 0.035 dB + 1 digit receiver w/ HP 11722A
(-10 to -20) dB 0.040 dB + 1 digit power sensor
(-20 to -30) dB 0.045 dB + 1 digit
(-30 to -40) dB 0.051 dB + 1 digit
(-40 to -50) dB 0.058 dB + 1 digit
(-50 to -60) dB 0.058 dB + 1 digit
(-60 to -70) dB 0.064 dB + 1 digit
(-70 to -80) dB 0.070 dB + 1 digit
(-80 to -90) dB 0.076 dB + 1 digit
(-90 to -100) dB 0.083 dB + 1 digit
(-100 to -110) dB 0.14 dB + 1 digit
(-110 to -120) dB 0.19 dB + 1 digit
Amplitude Modulation3 –
Measure
Frequency Modulation3 –
Measure
Phase Modulation –
Measure
Rate: 200 Hz to 10 kHz 150 kHz to 10 MHz 4 % + 0.015 rad HP 8902A measuring
receiver w/
Rate: 20 Hz to 200 kHz 10 MHz to 26.5 GHz 5 % + 0.019 rad HP 11792A power
sensor
Harmonic Distortion –
Measure
Pressure3 –
Measuring Equipment
Nitrogen Gas (0 to 90) psia 0.0021 % + 0.009 psia + 0.58R Additel ADT780
(>90 to 300) psia 0.0054 % + 0.015 psia + 0.6R pressure calibrator
Hydraulic (-14.5 to 300) psig 0.0054 % + 0.015 psi + 0.6R Crystal reference
(300 to 1000) psig 0.01 % + 0.01 psi + 0.6R pressure gauge
V. Thermodynamics
Frequency Reference 10 MHz Reference 1.8 × 10-12 Hz/Hz Fluke 910 GPS receiver
Signal
Frequency3 – 1 µHz to 26.5 GHz 1.8 × 10-12 Hz/Hz +0.6R GPS receiver w/
Measuring Equipment synthesized generator
Frequency3 – Measure DC to 225 MHz 1.5 × 10-9 Hz/Hz + 5 µHz GPS receiver w/
53132 counter
225 MHz to 26.5 GHz 1.8 × 10-12 Hz/Hz + 1.3 5343A counter
Hz
1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMC represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSL
Z540-1-1994 and the requirements of ANSI/NCSL Z540.3-2006 and R205 – Specific Requirements: Calibration Laboratory
Accreditation Program. This accreditation demonstrates technical competence for a defined scope and the operation of a
laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 1395.03
Valid to September 30, 2023
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.