SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

DGI METROLOGY/DIMENSIONAL GAUGE COMPANY


2144 Hwy 292
Inman, SC 29349
Jim Bible Phone: 864 473 1188

CALIBRATION

Valid To: May 31, 2022 Certificate Number: 4870.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to
this laboratory to perform the following calibrations1, 7:

I. Acoustics & Vibration

Parameter/Equipment Range CMC2, 5 () Comments

Microphones (-50 to -20) dB 0.21 dB Precision microphone


re 1V/Pa;
(3 to 100) mV/Pa
250 Hz

Pistonphones 74 to 135 dB 0.09 dB Precision microphone


(Acoustic Calibrators) – re 1Vpa; DMM
250 Hz

Frequency 250 Hz; 0.002 Hz


1 kHz

Total Harmonic <1 % of reading 0.08 % NI PCI-4461


Distortion (%)

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 1 of 14


Parameter/Equipment Frequency CMC2, 5 () Comments

Vibration (5 to 9) Hz 2.1 % Precision accelerometer


(10 to 99) Hz 1.6 %
(0.1 to 1.999) kHz 1.1 %
(2 to 10) kHz 2.8 %

II. Dimensional

Parameter/Equipment Range CMC2, 4 () Comments

Surface Plates –

Mean Plain Flatness3 Up to 271'' D 0.83 μin/in Differential levels

Repeatability3 – 12'' D to 271'' D 0.09 D μin/in Repeat reading gage and


Local Flatness electronic amplifier

III. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 6 () Comments

DC Voltage – Measure (0 to 100) mV 9.0 μV/V + 0.3 μV Agilent 3458A


100 mV to 1 V 8.0 μV/V + 0.3 μV
(1 to 10) V 8.0 μV/V + 0.6 μV
(10 to 100) V 10 μV/V + 31 μV
(100 to 1000) V 10 μV/V + 0.1 mV

DC Voltage – Generate Up to 330 mV 18 μV/V + 670 nV Fluke 5522A


330 mV to 3.3 V 9.1 μV/V + 1.3 μV
(3.3 to 33) V 9.5 μV/V + 13 μV
(30 to 330) V 13 μV/V + 100 μV
330 V to 1.02 kV 12 μV/V + 1 mV

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 2 of 14


Parameter/Equipment Range CMC2, 6 () Comments

DC Current – Measure Up to 100 nA 30 μA/A + 0.04 nA Agilent 3458A


1 μA 20 μA/A + 0.04 nA
10 μA 20 μA/A + 0.1 nA
100 μA 20 μA/A + 0.81 nA
1 mA 20 μA/A + 5.2 nA
10 mA 20 μA/A + 52 nA
100 mA 35 μA/A + 0.52 uA
1A 110 μA/A + 10 μA

DC Current – Generate Up to 30 μA 100 μA/A + 13 nA Fluke 5522A


330 μA to 3.3 mA 68 μA/A + 33 nA
(3.3 to 33) mA 67 μA/A + 170 nA
(33 to 330) mA 67 μA/A + 1.7 μA
330 mA to 1.1 A 130 μA/A + 27 μA
(1.1 to 3) A 250 μA/A + 27 μA
(3 to 11) A 330 μA/A + 330 μA
(11 to 20.5) A 670 μA/A + 500 μA

Parameter/Range Frequency CMC2, 6 () Comments

AC Voltage – Generate

(1 to 33) mV (10 to 45) Hz 530 μV/V + 4 μV Fluke 5522A


45 Hz to 10 kHz 100 μV/V + 4 μV
(10 to 20) kHz 140 μV/V + 4 μV
(20 to 50) kHz 670 μV/V + 4 μV
(50 to 100) kHz 2.3 mV/V + 8 μV
(100 to 500) kHz 5.3 mV/V + 33 μV

(33 to 330) mV (10 to 45) Hz 200 μV/V + 5.3 μV


45 Hz to 10 kHz 98 μV/V + 5.3 μV
(10 to 20) kHz 110 μV/V + 5.3 μV
(20 to 50) kHz 230 μV/V + 5.3 μV
(50 to 100) kHz 530 μV/V + 21 μV
(100 to 500) kHz 1.3 mV/V + 47 μV

330 mV to 3.3 V (10 to 45) Hz 200 μV/V + 33 μV


45 Hz to 10 kHz 100 μV/V + 40 μV
(10 to 20) kHz 130 μV/V + 40 μV
(20 to 50) kHz 200 μV/V + 33 μV
(50 to 100) kHz 470 μV/V + 83 μV
(100 to 500) kHz 1.6 mV/V + 400 μV

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 3 of 14


Parameter/Range Frequency CMC2, 6 () Comments

AC Voltage – Generate
(cont)

(3.3 to 33) V (10 to 45) Hz 200 μV/V + 530 μV Fluke 5522A


45 Hz to 10 kHz 100 μV/V + 400 μV
(10 to 20) kHz 160 μV/V + 400 μV
(20 to 50) kHz 230 μV/V + 400 μV
(50 to 100) kHz 600 μV/V + 1.1 mV

(33 to 330) V 45 Hz to 1 kHz 130 μV/V + 1.3 mV


(1 to 10) kHz 140 μV/V + 4 mV
(10 to 20) kHz 170 μV/V + 4 mV
(20 to 50) kHz 200 μV/V + 4 mV
(50 to 100) kHz 1.3 mV/V + 33 mV

330 V to 1.02 kV 45 Hz to 1 kHz 200 μV/V + 6.7 mV


(1 to 5) kHz 200 μV/V + 6.7 mV
(5 to 10) kHz 200 μV/V + 6.7 mV

AC Voltage – Measure

(0 to 10) mV (1 to 40) Hz 0.30 mV/V + 3.0 μV Agilent 3458A


40 Hz to 1 kHz 0.20 mV/V + 1.1 μV
(1 to 20) kHz 0.30 mV/V + 1.1 μV
(20 to 50) kHz 1.0 mV/V + 1.1 μV
(50 to 100) kHz 5.0 mV/V + 1.1 μV
(100 to 300) kHz 40 mV/V + 20 μV
300 kHz to 1 MHz 12 mV/V + 50 μV
(1 to 4) MHz 70 mV/V + 70 μV
(4 to 8) MHz 200 mV/V + 80 μV

100 mV (1 to 40) Hz 0.08 mV/V + 4 μV


40 Hz to 1 kHz 0.08 mV/V + 2 μV
(1 to 20) kHz 0.14 mV/V + 2 μV
(20 to 50) kHz 0.30 mV/V + 2 μV
(50 to 100) kHz 0.8 mV/V + 2 μV
(100 to 300) kHz 3 mV/V + 10 μV
300 kHz to 1 MHz 10 mV/V + 10 μV
(1 to 2) MHz 15 mV/V + 10 μV
(2 to 4) MHz 40 mV/V + 70 μV
(4 to 8) MHz 40 mV/V + 80 μV
(8 to 10) MHz 150 mV/V + 100 μV

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 4 of 14


Parameter/Range Frequency CMC2, 6 () Comments

AC Voltage – Measure
(cont)

1V (1 to 40) Hz 0.08 mV/V + 0.04 mV Agilent 3458A


40 Hz to 1 kHz 0.08 mV/V + 0.02 mV
(1 to 20) kHz 0.14 mV/V + 0.02 mV
(20 to 50) kHz 0.30 mV/V + 0.02 mV
(50 to 100) kHz 0.80 mV/V + 0.02 mV
(100 to 300) kHz 3.0 mV/V + 0.10 mV
300 kHz to 1 MHz 10 mV/V + 0.10 mV
(1 to 2) MHz 15 mV/V + 0.10 mV
(2 to 4) MHz 40 mV/V + 0.70 mV
(4 to 8) MHz 40 mV/V + 0.80 mV
(8 to 10) MHz 150 mV/V + 1.0 mV

10 V (1 to 40) Hz 0.08 mV/V + 0.40 mV


40 Hz to 1 kHz 0.08 mV/V + 0.20 mV
(1 to 20) kHz 0.14 mV/V + 0.20 mV
(20 to 50) kHz 0.30 mV/V + 0.20 mV
(50 to 100) kHz 0.80 mV/V + 0.20 mV
(100 to 300) kHz 3.0 mV/V + 1.0 mV
300 kHz to 1 MHz 10 mV/V + 1.0 mV
(1 to 2) MHz 15 mV/V + 1.0 mV
(2 to 4) MHz 40 mV/V + 7.0 mV
(4 to 8) MHz 40 mV/V + 8.0 mV
(8 to 10) MHz 150 mV/V + 10 mV

100 V (1 to 40) Hz 0.20 mV/V + 4.0 mV


40 Hz to 1 kHz 0.20 mV/V + 2.0 mV
(1 to 20) kHz 0.20 mV/V + 2.0 mV
(20 to 50) kHz 0.35 mV/V + 2.0 mV
(50 to 100) kHz 1.2 mV/V + 2.0 mV
(100 to 300) kHz 4.0 mV/V + 30 mV
300 kHz to 1 MHz 15 mV/V + 100 mV

1000 V (1 to 40) Hz 0.40 mV/V + 40 mV


40 Hz to 1 kHz 0.40 mV/V + 20 mV
(1 to 20) kHz 0.60 mV/V + 20 mV
(20 to 50) kHz 1.2 mV/V + 20 mV
(50 to 100) kHz 3.0 mV/V + 20 mV

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 5 of 14


Parameter/Range Frequency CMC2, 6 () Comments

AC Current – Generate

(29 to 330) μA (10 to 20) Hz 1.4 mA/A + 67 nA Fluke 5522A


(20 to 45) Hz 1 mA/A + 67 nA
45 Hz to 1 kHz 830 μA/A + 67 nA
(1 to 5) kHz 2 mA/A + 100 nA
(5 to 10) kHz 5.3 mA/A + 130 nA
(10 to 30) kHz 11 mA/A + 270 nA

330 μA to 3.3 mA (10 to 20) Hz 1.4 mA/A + 100 nA


(20 to 45) Hz 850 μA/A + 100 nA
45 Hz to 1 kHz 670 μA/A + 100 nA
(1 to 5) kHz 1.3 mA/A + 130 nA
(5 to 10) kHz 3.3 mA/A + 200 nA
(10 to 30) kHz 6.7 mA/A + 400 nA

(3.3 to 33) mA (10 to 20) Hz 1.2 mA/A + 1.3 μA


(20 to 45) Hz 600 μA/A + 1.3 μA
45 Hz to 1 kHz 270 μA/A + 1.3 μA
(1 to 5) kHz 530 μA/A + 1.3 μA
(5 to 10) kHz 1.3 mA/A + 2 μA
(10 to 30) kHz 2.7 mA/A + 2.7 μA

(33 to 330) mA (10 to 20) Hz 1.2 mA/A + 13 μA


(20 to 45) Hz 600 μA/A + 13 μA
45 Hz to 1 kHz 270 μA/A + 13 μA
(1 to 5) kHz 670 μA/A + 33 μA
(5 to 10) kHz 1.3 mA/A + 67 μA
(10 to 30) kHz 2.7 mA/A + 130 μA

330 mA to 1.1 A (10 to 45) Hz 1.2 mA/A + 67 μA


45 Hz to 1 kHz 330 μA/A + 67 μA
(1 to 5) kHz 4 mA/A + 670 μA
(5 to 10) kHz 17 mA/A + 3.3 mA

(1.1 to 3) A (10 to 45) Hz 1.2 mA/A + 67 μA


45 Hz to 1 kHz 400 μA/A + 67 μA
(1 to 5) kHz 4.0 mA/A + 670 μA
(5 to 10) kHz 17 mA/A + 3.3 mA

(3 to 11) A (45 to 100) Hz 400 μA/A + 1.3 mA


100 Hz to 1 kHz 690 μA/A + 1.3 mA
(1 to 5) kHz 20 mA/A + 1.3 mA

(11 to 20.5) A (45 to 100) Hz 800 μA/A + 3.3 mA


100 Hz to 1 kHz 1 mA/A + 3.3 mA
(1 to 5) kHz 20 mA/A + 3.3 mA

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 6 of 14


Parameter/Range Frequency CMC2, 6 () Comments

AC Current – Measure

Up to 100 μA (10 to 20) Hz 4.0 mA/A + 30 nA Agilent 3458A


(20 to 45) Hz 1.5 mA/A + 30 nA
(45 to 100) Hz 0.6 mA/A + 30 nA
100 Hz to 5 kHz 0.6 mA/A + 30 nA

100 μA to 1 mA (10 to 20) Hz 4.0 mA/A + 200 nA


(20 to 45) Hz 1.5 mA/A + 200 nA
(45 to 100) Hz 0.6 mA/A + 500 nA
100 Hz to 5 kHz 0.3 mA/A + 0.2 μA
(5 to 20) kHz 0.6 mA/A + 0.2 μA
(20 to 50) kHz 4.0 mA/A + 0.4 μA
(50 to 100) kHz 5.5 mA/A + 1.5 μA

(1 to 10) mA (10 to 20) Hz 4.0 mA/A + 2.0 μA


(20 to 45) Hz 1.5 mA/A + 2.0 μA
(45 to 100) Hz 0.6 mA/A + 2.0 μA
100 Hz to 5 kHz 0.3 mA/A + 2.0 μA
(5 to 20) kHz 4.0 mA/A + 4.0 μA
(20 to 50) kHz 5.5 mA/A + 15 μA
(50 to 100) kHz 4.0 mA/A + 20 μA

(10 to 100) mA (10 to 20) Hz 1.5 mA/A + 20 μA


(20 to 45) Hz 0.6 mA/A + 20 μA
(45 to 100) Hz 0.3 mA/A + 20 μA
100 Hz to 5 kHz 0.6 mA/A + 20 μA
(5 to 20) kHz 4.0 mA/A + 40 μA
(20 to 50) kHz 5.5 mA/A + 150 μA
(50 to 100) kHz 4.0 mA/A + 200 μA

100 mA to 1 A (10 to 20) Hz 4.0 mA/A + 200 μA


(20 to 45) Hz 1.6 mA/A + 200 μA
(45 to 100) Hz 0.8 mA/A + 200 μA
(5 to 20) kHz 3.0 mA/A + 0.2 mA
(20 to 50) kHz 10 mA/A + 0.4 mA

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 7 of 14


Parameter/Equipment Range CMC2, 6 () Comments

Resistance – Generate Up to 11 Ω 30 μΩ/Ω + 670 μΩ Fluke 5522A


(11 to 33) Ω 24 μΩ/Ω + 1 mΩ
(33 to 110) Ω 21 μΩ/Ω + 930 μΩ
(110 to 330) Ω 20 μΩ/Ω + 1.3 mΩ
330 Ω to 1.1 kΩ 19 μΩ/Ω + 1.3 mΩ
(1.1 to 3.3) kΩ 19 μΩ/Ω + 13 mΩ
(3.3 to 11) kΩ 19 μΩ/Ω + 13 mΩ
(11 to 33) kΩ 19 μΩ/Ω + 130 mΩ
(33 to 110) kΩ 19 μΩ/Ω + 130 mΩ
(110 to 330) kΩ 26 μΩ/Ω + 1.3 Ω
330 kΩ to 1.1 MΩ 22 μΩ/Ω + 1.3 Ω
(1.1 to 3.3) MΩ 41 μΩ/Ω + 20 Ω
(3.3 to 11) MΩ 87 μΩ/Ω + 33 Ω
(11 to 33) MΩ 170 μΩ/Ω + 1.7 kΩ
(33 to 110) MΩ 330 μΩ/Ω + 2 kΩ
(110 to 330) MΩ 3.8 mΩ/Ω + 67 kΩ
330 MΩ to 1.1 GΩ 10 mΩ/Ω + 330 kΩ

Resistance – Measure Up to 10 Ω 15 μΩ/Ω + 0.1 mΩ Agilent 3458A


100 Ω 12 μΩ/Ω + 0.5 mΩ
1 kΩ 10 μΩ/Ω + 0.3 mΩ
10 kΩ 10 μΩ/Ω + 5.0 mΩ
100 kΩ 17 μΩ/Ω + 0.1 Ω
1 MΩ 15 μΩ/Ω + 2.0 Ω
10 MΩ 50 μΩ/Ω + 0.1 kΩ
100 MΩ 0.5 mΩ/Ω + 1.0 kΩ
1GΩ 5.0 mΩ/Ω + 10 kΩ

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 8 of 14


Parameter/Equipment Range CMC2, 6 () Comments

Capacitance – Generate

10 Hz to 10 kHz (220 to 400) pF 4.6 mF/F + 6.7 pF Fluke 5522A


10 Hz to 10 kHz 400 pF to 1.1 nF 3.7 mF/F + 6.7 pF
10 Hz to 3 kHz (1.1 to 3.3) nF 3.5 mF/F + 6.7 pF
10 Hz to 1 kHz (3.3 to 11) nF 2.5 mF/F + 6.7 pF
10 Hz to 1 kHz (11 to 33) nF 2.1 mF/F + 6.7 pF
10 Hz to 1 kHz (33 to 110) nF 2.5 mF/F + 6.7 pF
10 Hz to 1 kHz (110 to 330) nF 1.8 mF/F + 20 pF
(100 to 600) Hz 330 nF to 1.1 μF 1.7 mF/F + 670 pF
(10 to 300) Hz (1.1 to 3.3) μF 1.8 mF/F + 2 nF
(10 to 150) Hz (3.3 to 11) μF 1.8 mF/F + 6.7 nF
(10 to 120) Hz (11 to 33) μF 2.8 mF/F + 20 nF
(10 to 80) Hz (33 to 110) μF 3 mF/F + 67 nF
(0 to 50) Hz (110 to 330) μF 3 mF/F + 200 nF
(0 to 20) Hz 330 μF to 1.1 mF 3 mF/F + 670 nF
(0 to 6) Hz (1.1 to 3.3) mF 3.3 mF/F + 2 μF
(0 to 2) Hz (3.3 to 11) mF 3.6 mF/F + 6.7 μF
(0 to 0.6) Hz (11 to 33) mF 5.1 mF/F + 20 μF
(0 to 0.2) Hz (33 to 110) mF 7.4 mF/F + 67 μF

Capacitance – Measure Up to 2 nF 13 mF/F + 67 pF RIGOL DM3068


(Constant Current) (2 to 20) nF 6.7 mF/F + 80 pF
(20 to 200) nF 6.7 mF/F + 400 pF
200 nF to 2 μF 6.7 mF/F + 4 nF
(2 to 20) μF 6.7 mF/F + 40 pF
(20 to 200) μF 6.7 mF/F + 400 pF
200 μF to 2 mF 6.7 mF/F + 4 μF
(2 to 20) mF 6.7 mF/F + 40 μF
(20 to 100) mF 20 mF/F + 130 μF

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 9 of 14


Parameter/Equipment Range CMC2, 5 () Comments

Electrical Simulation of
Thermocouple
Indicators7 –

Type B (600 to 800) °C 0.29 °C + 0.58 % Fluke 5522A


(800 to 1000) °C 0.23 °C + 0.58 %
(1000 to 1550) °C 0.2 °C + 0.58 %
(1550 to 1820) °C 0.22 °C + 0.58 %

Type C (0 to 150) °C 0.2 °C + 1.2 %


(150 to 650) °C 0.17 °C + 1.2 %
(650 to 1000) °C 0.21 °C + 1.2 %
(1000 to 1800) °C 0.33 °C + 1.2 %
(1800 to 2316) °C 0.56 °C + 1.2 %

Type E (-250 to -100) °C 0.33 °C+ 1.2 %


(-100 to -25) °C 0.11 °C + 1.2 %
(-25 to 0) °C 0.09 °C + 1.2 %
(0 to 350) °C 0.09 °C + 0.58 %
(350 to 650) °C 0.11 °C + 0.58 %
(650 to 1000) °C 0.14 °C + 0.58 %

Type J (-210 to -100) °C 0.18 °C + 0.87 %


(-100 to -30) °C 0.11 °C + 0.87 %
(-30 to 150) °C 0.09 °C + 0.87 %
(150 to 760) °C 0.11 °C + 0.87 %
(760 to 1200) °C 0.15 °C + 0.87 %

Type K (-200 to -100) °C 0.22 °C + 2.3 %


(-100 to -25) °C 0.12 °C + 2.3 %
(-25 to 0) °C 0.11 °C + 2.3 %
(0 to 120) °C 0.11 °C + 0.87 %
(120 to 1000) °C 0.17 °C + 0.87 %
(1000 to 1372) °C 0.27 °C + 0.87 %

Type L (-200 to -100) °C 0.25 °C + 0.87 %


(-100 to 800) °C 0.17 °C + 0.87 %
(800 to 900) °C 0.11 °C + 0.87 %

Type N (-200 to -100) °C 0.27 °C + 2.3 %


(-100 to -25) °C 0.15 °C + 2.3 %
(-25 to 0) °C 0.13 °C + 2.3 %
(0 to 120) °C 0.13 °C + 0.87 %
(120 to 410) °C 0.12 °C + 0.87 %
(410 to 1300) °C 0.18 °C + 0.87 %

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 10 of 14


Parameter/Equipment Range CMC2, 5 () Comments

Electrical Simulation of
Thermocouple
Indicators7 – (cont)

Type R (0 to 250) °C 0.38 °C + 0.29 % Fluke 5522A


(250 to 400) °C 0.23 °C + 0.29 %
(400 to 1000) °C 0.22 °C + 0.29 %
(1000 to 1767) °C 0.27 °C + 0.29 %

Type S (0 to 250) °C 0.31 °C + 0.29 %


(250 to 1000) °C 0.24 °C + 0.29 %
(1000 to 1400) °C 0.25 °C + 0.29 %
(1400 to 1767) °C 0.31 °C + 0.29 %

Type T (-250 to -150) °C 0.42 °C + 1.7 %


(-150 to 0) °C 0.16 °C + 1.7 %
(0 to 120) °C 0.11 °C + 0.87 %
(120 to 400) °C 0.09 °C + 0.09 %

Type U (-200 to 0) °C 0.37 °C + 1.2 %


(0 to 600) °C 0.18 °C + 0.87 %

Electrical Simulation of
RTDs7 –

Pt 385 (100 Ω) (-200 to 0) °C 0.35 °C + 0.005 [t] Fluke 5522A


(0 to 100) °C 0.35 °C + 0.005 [t] [t] = temperature
(100 to 300) °C 0.35 °C + 0.005 [t]
(300 to 400) °C 0.35 °C + 0.005 [t]
(400 to 630) °C 0.36 °C + 0.005 [t]
(630 to 800) °C 0.38 °C + 0.005 [t]

Pt 3926 (100 Ω) (-200 to 0) °C 0.35 °C + 0.005 [t]


(0 to 100) °C 0.35 °C + 0.005 [t]
(100 to 300) °C 0.35 °C + 0.005 [t]
(300 to 400) °C 0.35 °C + 0.005 [t]
(400 to 630) °C 0.36 °C + 0.005 [t]

Pt 3916 (100 Ω) (-200 to -190) °C 0.38 °C + 0.005 [t]


(-190 to -80) °C 0.35 °C + 0.005 [t]
(-80 to 0) °C 0.35 °C + 0.005 [t]
(0 to 100) °C 0.35 °C + 0.005 [t]
(100 to 260) °C 0.35 °C + 0.005 [t]
(260 to 300) °C 0.35 °C + 0.005 [t]
(300 to 400) °C 0.35 °C + 0.005 [t]
(400 to 600) °C 0.35 °C + 0.005 [t]
(600 to 630) °C 0.38 °C + 0.005 [t]

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 11 of 14


Parameter/Equipment Range CMC2, 5 () Comments

Electrical Simulation of
RTDs7 – (cont)

Pt 385 (200 Ω) (-200 to 100) °C 0.35 °C + 0.005 [t] Fluke 5522A


(100 to 260) °C 0.35 °C + 0.005 [t] [t] = temperature
(260 to 300) °C 0.36 °C + 0.005 [t]
(300 to 400) °C 0.36 °C + 0.005 [t]
(400 to 600) °C 0.36 °C + 0.005 [t]
(600 to 630) °C 0.11 °C + 0.005 [t]

Pt 385 (500 Ω) (-200 to -80) °C 0.35 °C + 0.005 [t]


(-80 to 100) °C 0.35 °C + 0.005 [t]
(100 to 260) °C 0.35 °C + 0.005 [t]
(260 to 400) °C 0.35 °C + 0.005 [t]
(400 to 600) °C 0.35 °C + 0.005 [t]
(600 to 630) °C 0.35 °C + 0.005 [t]

Pt 385 (1000 Ω) (-200 to 0) °C 0.42 °C + 0.005 [t]


(0 to 100) °C 0.35 °C + 0.005 [t]
(100 to 260) °C 0.35 °C + 0.005 [t]
(260 to 300) °C 0.35 °C + 0.005 [t]
(300 to 600) °C 0.35 °C + 0.005 [t]
(600 to 630) °C 0.38 °C + 0.005 [t]

PtNi 385 (120 Ω) (-80 to 100) °C 0.35 °C + 0.005 [t]


(100 to 260) °C 0.36 °C + 0.005 [t]

Cu 427 (10 Ω) (-100 to 260) °C 0.40 °C + 0.005 [t]

IV. Mechanical

Parameter/Equipment Range CMC2, 5 () Comments

Force –

Impulse Hammers and Up to 5000 lbf 2.2 % Calibrated mass and


Dynamic Force Cells accelerometers

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 12 of 14


Parameter/Equipment Range CMC2, 4, 5 () Comments

Pressure Up to 2000 psi 0.16 % Precision pressure gage


(Analog and
Transducers) Up to 1 in H2O 0.0053 % FS + 0.0076 % GE Ruska 7250LP
(1 to 3) in H2O 0.0045 % FS + 0.0076 %
(3 to 10) in H2O 0.0041 % FS + 0.0076 %
(10 to 30) in H2O 0.0035 % FS + 0.0076 %

V. Time & Frequency

Parameter/Equipment Range CMC2, 8 () Comments

Frequency – Measuring 0.01 Hz to 2.000 MHz 1.7 μHz/Hz + 3.3 μHz Fluke 5522A-SC600
Equipment

________________________________
1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs
represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually
using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device
and to influences from the circumstances of the specific calibration.
3
Field calibration service is available for this calibration. Please note the actual measurement
uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found
on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of
calibration and for other possible adverse effects such as those caused by transportation of the
calibration equipment. The usual allowance for the actual uncertainty introduced by the item being
calibrated, (e.g., resolution) must also be considered and this, on its own, could result in the actual
measurement uncertainty achievable on a customer’s site being larger than the CMC.
4
In the statement of CMC, D is the numerical value of the diagonal length of the device measured in
inches; FS represents “Full Scale”.
5
In the statement of CMC, percentages are percentages of reading, unless otherwise indicated.
6
Field The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured
value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range
or as a fraction/percentage of the reading plus a fixed floor specification
7
This scope meets A2LA’s P112 Flexible Scope Policy.

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 13 of 14


8
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges
indicated for the listed measurement parameter.

(A2LA Cert. No. 4870.01) Revised 04/22/2022 Page 14 of 14


Accredited Laboratory
A2LA has accredited

DGI METROLOGY/DIMENSIONAL GAUGE CO.


Inman, SC
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017
General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 –
Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical
competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 16th day of April 2020.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 4870.01
Valid to May 31, 2022
Revised April 22, 2022