SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

& ANSI/NCSL Z540-1-1994

DYTRAN INSTRUMENTS INC.


21592 Marilla St
Chatsworth, CA 91311
Denis Varak Phone: 818 700 7818

CALIBRATION

Valid To: May 31, 2022 Certificate Number: 2672.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 7:

I. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 3, 4 (±) Comments

DC Voltage – Measure (10 to 100) mV 0.10 % Fluke 45


(100 to 1000) mV 0.09 %
(1 to 10) V 0.09 %
(10 to 100) V 0.09 %

Capacitance – Measure (150 to 330) pF 0.20 % Capacitance meter,


(330 to 15) µF 0.19 % 1 kHz

Resistance – Measure (0 to 300) kΩ 0.29 % Fluke 45


300 kΩ to 1 MΩ 0.13 %

Parameter/Range Frequency CMC2, 3, 4 (±) Comments

AC Voltage – Measure

(0 to 5) V (20 to 50) Hz 0.3 % Data acquisition card


50 Hz to 10 kHz 0.3 %

(A2LA Cert. No. 2672.01) 12/10/2020 Page 1 of 3


II. Mechanical

Parameter/Equipment Range5 CMC2, 4, 6 (±) Comments

Acceleration 1 Hz @ 0.25 G 2.1 % Data acquisition card


Sensitivity/Frequency (2 to 20) Hz @ 1 G 2.1 %
Response – Measure (20 to 100) Hz @ 1 G 1.9 %
(100 to 2500) Hz @ 1 G 1.5 %
(2500 to 10 000) Hz @ 1 G 2.8 %

Acceleration Sensitivity (-55 to 270) °C @ 100 Hz, 1 G 4.3 % Mini-shaker and oven
Deviation Due to
Temperature – Measure

Transverse Sensitivity – 328 Hz @ 1 G 3.6 % Back-to-back


Measure comparison using
electrodynamic shaker

Acoustic Pressure – 130 dB @ 1000 Hz 7.6 % Comparison to


Measure 140 dB @ 1000 Hz 2.7 % standard microphone
150 dB @ 1000 Hz 1.5 %
160 dB @ 1000 Hz 1.3 %

Static Force – Measure (10 to 500) lbf 0.95 % Ring dynamometer,


(500 to 1000) lbf 1.3 % strain gage, load cell,

(1000 10 000) lbf 1.1 % Ring dynamometer,


(10 000 to 100 000) lbf 1.2 % load cell

Fixed Points (1.876, 2.840, 3.290, 4.170) lbf 0.67 % Dead weights

Shock – Measure Up to 5000 G 3.0 % Load cell

Static Pressure – (5 to 100) psig 2.4 % Back-to-back


Measure (100 to 1000) psig 0.99 % comparison in
(1000 to 10 000) psig 1.0 % pressure pump

Impulse Force – Up to 1000 lbf 3.8 % Load cell


Measure (1000 to 5000) lbf 1.0 %

(A2LA Cert. No. 2672.01) 12/10/2020 Page 2 of 3


Parameter/Range5 Frequency CMC2, 4, 6 (±) Comments

Acceleration Amplitude
Linearity Vibration –
Measure

(1 to 80) G 100 Hz 1.9 % Back-to-back


300 Hz, 500 Hz 1.8 % comparison using
1000 Hz 2.9 % electrodynamic shaker

____________________________________________
1
This laboratory offers commercial calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured
values in the ranges indicated. CMCs are expressed as either a specific value that covers the full range
or as a percent or fraction of the reading plus a fixed floor specification.
4
In the statement of CMC, % denotes percentage of reading.
5
In the statement of CMC, G represents the gravitational acceleration.
6
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
7
This scope meets A2LA’s P112 Flexible Scope Policy.

(A2LA Cert. No. 2672.01) 12/10/2020 Page 3 of 3


Accredited Laboratory
A2LA has accredited

DYTRAN INSTRUMENTS INC.


Chatsworth, CA
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSL
Z540-1-1994 and R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates
technical competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 10th day of December 2020.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 2672.01
Valid to May 31, 2022