Testrano 600 Getting Started ENU
Testrano 600 Getting Started ENU
Getting Started
TESTRANO 600 Getting Started
2 OMICRON
Safety instructions
1 Safety instructions
1.1 Operator qualifications
Working on high-voltage assets can be extremely dangerous. Consequently, only personnel qualified,
authorized and skilled in electrical engineering and trained by OMICRON are allowed to operate
TESTRANO 600 and its accessories. Before starting to work, clearly establish the responsibilities.
Personnel receiving training, instructions, directions, or education on TESTRANO 600 must be under
constant supervision of an experienced operator while working with the equipment. The operator is
responsible for the safety requirements during the whole test.
OMICRON 3
TESTRANO 600 Getting Started
4 OMICRON
Safety instructions
► Position the measurement setup in a way that you can easily disconnect TESTRANO 600 from
mains. If permanently connected, make sure that the measurement setup is positioned in a way that
the switch or circuit breaker can be easily reached.
► Do not operate TESTRANO 600 and its accessories in the presence of explosives, gas or vapors.
► If TESTRANO 600 or its accessories do not seem to function properly (for example, in case of cable
damages, abnormal warming or overheating of components), stop using them and contact
OMICRON support (see "Support" on page 36).
► Observe the high-voltage areas.
► Always obey the internal safety instructions for working in areas with high voltage to avoid injury.
Distance:
min. 1.5 m/5 ft
Device under test
TESTRANO 600
CP TD1
Barrier height:
1 - 1.4 m / 3.3 - 4.6 ft
Figure 1-1: Illustration of safe area and high-voltage area established for working with TESTRANO 600
and CP TD1
OMICRON 5
TESTRANO 600 Getting Started
1.5 Disclaimer
Using TESTRANO 600 in any way differing from the one mentioned above is considered improper use,
and will invalidate all customer warranty claims and exempt the manufacturer from any liability to
recourse.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
1.7 Recycling
This test set (including all accessories) is not intended for household use. At
the end of its service life, do not dispose of the test set with household waste!
For customers in EU countries (incl. European Economic Area)
OMICRON test sets are subject to the EU Waste Electrical and Electronic
Equipment Directive 2012/19/EU (WEEE directive). As part of our legal obligations
under this legislation, OMICRON offers to take back the test set and ensure that it is
disposed of by authorized recycling agents.
For customers outside the European Economic Area
Please contact the authorities in charge for the relevant environmental regulations in
your country and dispose the OMICRON test set only in accordance with your local
legal requirements.
6 OMICRON
Introduction
2 Introduction
2.1 Designated use
In combination with the CP TD1 or as a stand-alone unit, TESTRANO 600 is a multi-purpose power
transformer test system for routine and diagnostic testing of power transformers during manufacturing,
commissioning and maintenance.
The various partly automated tests are defined and parameterized via the front panel control of a built-
in embedded PC or via an externally connected laptop.
OMICRON 7
TESTRANO 600 Getting Started
3 Hardware overview
3.1 TESTRANO 600
► Refer to user manual chapter 7 "Technical data" on page 28 for detailed hardware information.
8 OMICRON
Hardware overview
Power switch
16A
Resetable mains overcurrent SERIAL Mains power socket (one-phase)
protection Connection to CP TD1
NETWORK
RJ-45 socket for connecting TESTRANO 600 to an external computer
OMICRON 9
TESTRANO 600 Getting Started
► Always observe the warning lights while working with TESTRANO 600.
► Do not cover the warning lights (for example, with a computer) – they indicate possible hazards.
If no or both warning lights are on, the unit is defective or not supplied by mains.
► If TESTRANO 600 appears to be defective, do not use it anymore. Contact OMICRON support
(see "Support" on page 36).
10 OMICRON
Hardware overview
For disconnection:
► Hold the connector and pull back the
silver latch.
► Turn the connector to the left and gently
pull it out.
NOTICE
Equipment damage possible
► Do not pull the cable when disconnecting.
► Hold, turn and gently pull the connector for disconnection.
OMICRON 11
TESTRANO 600 Getting Started
High-voltage cable
marked red
• Polarity protection: only suitable for
HIGH VOLTAGE and LOW VOLTAGE
sockets
• 15 m length
• 8 poles
• cross-section:
4 × 4 mm2 for output
4 × 1 mm2 for measurement
• Neutrik® plug
Low-voltage cable
marked yellow
12 OMICRON
Hardware overview
3.2 CP TD1
► Refer to the CP TD1 User Manual for detailed information and safety instructions.
Equipotential ground
terminal
BOOSTER IN
Booster input
SERIAL
Serial interface
connector
IN_B
measuring input
IN_A
measuring input
Grounding terminal
High-voltage output
OMICRON 13
TESTRANO 600 Getting Started
4 Application
4.1 Safety precautions in the substation
Before setting TESTRANO 600 into operation and carrying out a test, it is essential that you have read
and understood chapter 1 "Safety instructions" on page 3.
► Be aware that all output sockets of TESTRANO 600 can carry life-hazardous voltage and current.
► Only use TESTRANO 600 with a solid connection to ground.
► Separate your working area – see Figure 1-1: "Illustration of safe area and high-voltage area
established for working with TESTRANO 600 and CP TD1" on page 5.
► Tests with high voltages and currents must only be carried out by authorized and qualified personnel.
► Personnel receiving training, instructions, directions, or education on high- voltage/current tests
should remain under the constant supervision of an experienced operator while working with the
equipment.
► The instructions have to be renewed at least once per year.
► The instructions must be available in written form and signed by each person assigned to do high-
voltage/current tests.
Prior to connecting a test object to TESTRANO 600, the following steps need to be carried out by an
authorized employee of the utility:
► Protect yourself and your working environment against an accidental re-connection of high voltage
by other persons and circumstances.
► Verify that the test object is safely isolated.
► Earth-connect and shorten out the test object’s terminals using a grounding set.
► Protect yourself and your working environment with a suitable protection against other (possibly live)
circuits.
► Protect others from accessing the high-voltage area and accidentally touching live parts by setting up
a suitable barrier and, if applicable, warning lights.
► If there is a longer distance between the location of TESTRANO 600 and the area of danger, a
second person with an additional Emergency Stop button is required.
14 OMICRON
Application
1. Make sure that the power switch on the TESTRANO 600 side panel is turned off.
2. Press the Emergency Stop button.
3. Connect TESTRANO 600 to:
• equipotential ground: Ground TESTRANO 600 with a cable of at least 6 mm2 cross-section as
close as possible to the operator.
• the computer (optional if TESTRANO 600 is used with TouchControl)
• the mains power supply
Optional
Connect the CP TD1 to TESTRANO 600
a) Connect to ground.
Without trolley:
► Properly connect the TESTRANO 600 and CP TD1 grounding terminals to substation ground.
With trolley (optional):
► Properly connect the TESTRANO 600 and CP TD1 grounding terminals to the trolley’s ground
bar.
► Connect the trolley’s ground bar to earth.
b) Connect the CP TD1 BOOSTER IN to the TESTRANO 600 EXTERNAL BOOSTER using the
booster cable.
c) Connect the CP TD1 SERIAL to TESTRANO 600 SERIAL with the data cable.
Note: The serial cable also provides the power supply for the CP TD1.
OMICRON 15
TESTRANO 600 Getting Started
16 OMICRON
Application
1. Connect the test leads to TESTRANO 600 as shown in the wiring diagram and in the order given
below:
TESTRANO 600
00
I. Connect the high-voltage (red), low-voltage (yellow) and tap changer cables to TESTRANO 600.
II. Connect the high-voltage (red) and low-voltage (yellow) cables to the transformer’s main
terminals.
III. Connect the tap changer cable to the appropriate terminals in the control cabinet of the
transformer (see Figure 4-2 below).
IV. Re-connect and turn on the voltage of the tap changer.
Note: If the tap changer control voltage exceeds 42 V, the orange warning light 2 on the side panel will
indicate a hazardous voltage on the TESTRANO 600 inputs (see 3.1.3 "Safety and warning indicators"
on page 10).
OMICRON 17
TESTRANO 600 Getting Started
PE Motor
Up
Down
Tap control
Optional
Connect the CP TD1 to the transformer
If you connected the CP TD1 to TESTRANO 600 (step 4 in section 4.2), connect the IN_A, IN_B and
high-voltage output of the CP TD1 to the transformer.
► Refer to the CP TD1 User Manual for more information on safely connecting the CP TD1 to a device
under test.
2. Erect a barrier separating the safe area from the high-voltage test area (see page 5).
3. Remove the grounding set from the test object.
4. Release the Emergency Stop button.
18 OMICRON
Application
4.4 Measurement
TouchControl Primary Test Manager
TouchControl displays the results in the Primary Test Manager displays the results in the
Measurement view of the test. Measurements section of the test.
OMICRON 19
TESTRANO 600 Getting Started
4.5 Disconnection
1. Wait until the green light on the TESTRANO 600 front panel is on and the warning lights on the front
and side panel are off.
Note: If the tap changer control voltage exceeds 42 V, the orange warning light 2 on the side panel
will indicate a hazardous voltage on the TESTRANO 600 inputs (see 3.1.3 "Safety and warning
indicators" on page 10).
► Disconnect the tap changer cable to extinguish warning light 2.
2. Press the Emergency Stop button on the TESTRANO 600 front panel.
DANGER
Death or severe injury caused by high voltage or current
► Never unplug any cables while the measurement is running.
► Only disconnect cables when all of the following apply:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.
3. To prevent anyone from starting a test, use the Software lock in the TouchControl software (see
chapter 5.5 "Software lock" on page 32) and/or lock your computer.
4. Remove the barrier between the high-voltage area and the safe area.
DANGER
Death or severe injury caused by high voltage or current possible
► Before touching any part of the transformer, ground and short-circuit its terminals using
a grounding set.
5. Disconnect all cables from the transformer.
6. Disconnect all cables from TESTRANO 600 and, if applicable, from the CP TD1.
7. Switch TESTRANO 600 off by pressing the mains power switch on the side panel.
8. Disconnect the mains power cord.
9. Remove the equipotential ground as the last connection that is removed first from TESTRANO 600
and then on the substation side.
20 OMICRON
Testing with TouchControl
Safety instructions
TESTRANO 600 measuring cables
5. Connect to transformer
TESTRANO 600
Application
Wiring diagrams
7. Measurement Measurement
OMICRON 21
TESTRANO 600 Getting Started
OLTC/DETC position
(HV or LV side)
Tap or swipe left to switch to
Measurement view
Entering values
► Tap a box, then use the numpad to enter or correct a value.
► If needed, tap the metric prefixes below after entering a value:
– k for kilo-
– M for mega-
– m for milli-
► Use the slider to increase or decrease the displayed value. Release the slider to stop at the desired
value.
Note: The slider will stop at the minimum/maximum value.
22 OMICRON
Testing with TouchControl
OMICRON 23
TESTRANO 600 Getting Started
Test assessment
(see section 5.4)
Tap arrows to sort list in ascending or
descending order
Figure 5-3: Wiring diagram for a TTR on a transformer with vector group YNd5
24 OMICRON
Testing with TouchControl
5.3 Measurement
► Refer to chapter 1 "Safety instructions" on page 3 for detailed information about safe testing.
► If in doubt, contact OMICRON support (see "Support" on page 36).
DANGER
Death or severe injury caused by high voltage or current
► Do not unplug any cables while the measurement is running.
► Only remove cables when all of the following apply to TESTRANO 600:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.
WARNING
Death or severe injury caused by high voltage or current possible
► Do not enter the high-voltage area during the test.
► Do not touch any part of the transformer before grounding and short-circuiting its
terminals.
1. Tap Start.
5.4 Assessment
► Tap Assess to expand the available assessment states:
OMICRON 25
TESTRANO 600 Getting Started
Location view
4. Location and asset
Asset view
5. Jobs Jobs
Test view
PTM Transformer tests
6. Tests
PTM Bushing tests
Device-independent PTM tests
Safety instructions
TESTRANO 600 measuring cables
7. Connection to device under test
TESTRANO 600
Application
Connecting to the transformer
8. Test settings Performing tests
26 OMICRON
Testing with Primary Test Manager
6.2 Measurement
► Refer to chapter 1 "Safety instructions" on page 3 for detailed information about safe testing.
► If in doubt, contact OMICRON support (see "Support" on page 36).
DANGER
Death or severe injury caused by high voltage or current
► Do not unplug any cables while the measurement is running.
► Only remove cables when all of the following apply to TESTRANO 600:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.
WARNING
Death or severe injury caused by high voltage or current possible
► Do not enter the high-voltage area during the test.
► Do not touch any part of the transformer before grounding and short-circuiting its
terminals.
OMICRON 27
TESTRANO 600 Getting Started
7 Technical data
At the time of factory adjustment all units are within the typical accuracy values specified in this
document.
Typical accuracy means that 98 % of all units meet the specified values at 23 °C ± 5 °C/73 ºF ± 10 ºF,
after a warm-up time of more than 25 min., and in a frequency range of 45 Hz to 65 Hz or DC.
The typical accuracy values multiplied by 3 are guaranteed at an ambient temperature of
23 ºC ± 5 ºC/73 ºF ± 10 ºF, after a warm-up time more than 25 min., and in a frequency range of
45 Hz to 65 Hz or DC.
Accuracy values indicate that the error is smaller than:
± (value read × reading error [rd] + full scale of range × range error [rg]).
For mains voltages below 190 V AC the system is subject to power restrictions.
OMICRON suggests that you send in your unit for calibration at least once a year.
Technical data are subject to change without notice.
CAT level
The CAT level required depends on the TESTRANO 600 application. All CAT ratings are defined for sea
levels below 2000 m. There are some limitations between 2000 m and 5000 m sea level (see
section 7.4 "Environmental conditions" on page 34).
CAT I is required when the measured voltage is generated by the test set itself. No voltages from other
sources are measured.
CAT II is required when measuring within electrical devices or between mains supply and devices.
CAT III is required when measuring in electrical installations such as control cubicles that are still
connected to the station battery or mains. The electrical installations are protected by a fuse.
28 OMICRON
Technical data
Characteristic Rating
Frequency DC or 15 Hz … 599 Hz
Vmains P30 s Pcontinuous
Power >100 VRMS 1500 W 1000 W
>190 VRMS 4000 W 2400 W
3 × 0 … ±56 VDC
DC low range 33 ADC
1 × 0 … ±170 VDC
AC high range
3 × 0 … 230 VRMS (LN) 100 mARMS
low current
3 × 0 … 80 VRMS (LN)
AC high range 16 ARMS
1 × 0 … 240 VRMS
3 × 0 … 40 VRMS (LN)
AC low range 33 ARMS
1 × 0 … 120 VRMS
Characteristic Accuracy1
Voltage accuracy DC 0.033 % rd ± 0.017 % range
Voltage accuracy AC (50 Hz) at burden open
0.33 % rd ± 0.17 % range
load
Phase accuracy AC (50 Hz) burden open load,
± 0.36°
V>20 VRMS
1. Typical accuracy at 23 °C ±5 K
OMICRON 29
TESTRANO 600 Getting Started
Characteristic Accuracy1
Current accuracy DC 0.033 % rd ± 0.017 % range
Current accuracy AC
0.33 % rd ± 0.17 % range
50/60 Hz at burden 0.1 Ω
1. Typical accuracy at 23 °C ±5 K
30 OMICRON
Technical data
OMICRON 31
TESTRANO 600 Getting Started
Characteristic Rating
Voltage 300 VRMS
32 OMICRON
Technical data
Range name
Voltage at HV Range1 Accuracy2
(LV voltage range)
1 0.03 % rd +
300 VRMS ------------------
1 …10 0.043 % range
1 0.027 % rd +
30 VRMS -------------------------
10 …100 0.043 % range
230 VRMS HV (LN)
1 0.027 % rd +
3 VRMS -------------------------------
100 …1000 0.043 % range
1 0.027 % rd +
300 mVRMS --------------------------------------
1000 …10000 0.043 % range
LV
1. Range = --------
HV
2. Typical accuracy at 23 °C ±5 K
OMICRON 33
TESTRANO 600 Getting Started
Characteristic Rating
Nominal 100 V … 240 VAC
Voltage
Permitted 85 V … 264 VAC
Current Nominal 16 A
Nominal 50 Hz/60 Hz
Frequency
Permitted 45 Hz … 65 Hz
Automatic circuit breaker with magnetic
Power fuse
overcurrent tripping at I >16 A
Continuous <3.6 kW
Power consumption
Peak <5.0 kW
Current consumption, continuous <16 AAC
Connector type IEC320/C20, 1 phase
Characteristic Rating
Operating -10 ºC … +55 ºC/+14 ºF…+131 ºF
Temperature
Storage -30 ºC … +70 ºC/-22 ºF…+158 ºF
2000 m/6550 ft, up to 5000 m/16400 ft with
Operating
Max. altitude limited specifications1
Storage 12 000 m/40 000 ft
1. Output TAP CHANGER (CAT III / 300 V): from 2000 m/6550 ft to 5000 m/16400 ft altitude only CAT II compliance or CAT III
compliance with half voltage
Characteristic Rating
464 × 386 × 229 mm
With cover, without handles
Dimensions 18.3 × 15.2 × 9 in
(w × h× d) 580 × 386 × 229 mm
With cover, with handles
22.8 × 15.2 × 9 in
Device with display 20.6 kg/45.5 lb
Weight
Device without display 19.5 kg/43 lb
34 OMICRON
Technical data
7.6 Standards
Table 7-17: Standards conformity
EMC, safety
EMC IEC/EN 61326-1
(industrial electromagnetic environment)
FCC subpart B of part 15, class A
Safety IEC/EN/UL 61010-1,
IEC/EN/UL 61010-2-30
Other
IEC/EN 60068-2-27
Shock
(15 g/11 ms, half-sinusoid, 3 shocks in each axis)
IEC/EN 60068-2-6
Vibration (frequency range 10 Hz…150 Hz, acceleration 2 g continuous
(20 m/s2/65 ft/s2), 20 cycles per axis)
IEC/EN 60068-2-78
Humidity (5 % … 95 % relative humidity, no condensation),
tested at 40 ºC/104 ºF for 48 hours
OMICRON 35
TESTRANO 600 Getting Started
Support
When you are working with our products we want to provide you with the greatest
possible benefits. If you need any support, we are here to assist you!
OMICRON electronics GmbH, Oberes Ried 1, 6833 Klaus, Austria, +43 59495
36 OMICRON
ENU 1161 03 02