0% found this document useful (0 votes)
58 views37 pages

Testrano 600 Getting Started ENU

Uploaded by

Alan Septian
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
58 views37 pages

Testrano 600 Getting Started ENU

Uploaded by

Alan Septian
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 37

TESTRANO 600

Getting Started
TESTRANO 600 Getting Started

Manual version: ENU 1161 03 02


© OMICRON electronics GmbH 2017. All rights reserved.
This manual is a publication of OMICRON electronics GmbH.
All rights including translation reserved. Reproduction of any kind, for example, photocopying,
microfilming, optical character recognition and/or storage in electronic data processing systems,
requires the explicit consent of OMICRON. Reprinting, wholly or in part, is not permitted.
The product information, specifications, and technical data embodied in this manual represent the
technical status at the time of writing and are subject to change without prior notice.
We have done our best to ensure that the information given in this manual is useful, accurate and entirely
reliable. However, OMICRON does not assume responsibility for any inaccuracies which may be
present.
The user is responsible for every application that makes use of an OMICRON product.
OMICRON translates this manual from the source language English into a number of other languages.
Any translation of this manual is done for local requirements, and in the event of a dispute between the
English and a non-English version, the English version of this manual shall govern.

2 OMICRON
Safety instructions

1 Safety instructions
1.1 Operator qualifications
Working on high-voltage assets can be extremely dangerous. Consequently, only personnel qualified,
authorized and skilled in electrical engineering and trained by OMICRON are allowed to operate
TESTRANO 600 and its accessories. Before starting to work, clearly establish the responsibilities.
Personnel receiving training, instructions, directions, or education on TESTRANO 600 must be under
constant supervision of an experienced operator while working with the equipment. The operator is
responsible for the safety requirements during the whole test.

1.2 Safety standards and rules


1.2.1 Safety standards
Testing with TESTRANO 600 must comply with the internal safety instructions and additional safety-
relevant documents.
In addition, observe the following safety standards, if applicable:
• EN 50191 (VDE 0104) "Erection and Operation of Electrical Test Equipment"
• EN 50110-1 (VDE 0105 Part 100) "Operation of Electrical Installations"
• IEEE 510 "IEEE Recommended Practices for Safety in High-Voltage and High-Power Testing"
Moreover, observe all applicable regulations for accident prevention in the country and at the site of
operation.
Before operating TESTRANO 600 and its accessories, read the safety instructions in this Getting Started
carefully.
Do not turn on or operate TESTRANO 600 if you do not understand the safety information in this manual.
If you have questions or do not understand some safety instructions, contact OMICRON before
proceeding.
Maintenance and repair of TESTRANO 600 and its accessories is only permitted by qualified experts at
OMICRON service centers (see "Support" on page 36).

1.2.2 Safety rules


Always observe the five safety rules:
► Disconnect completely.
► Secure against re-connection.
► Verify that the installation is dead.
► Carry out grounding and short-circuiting.
► Provide protection against adjacent live parts.

OMICRON 3
TESTRANO 600 Getting Started

1.3 Operating the measurement setup


► Before connecting or disconnecting test objects and/or cables, make sure that TESTRANO 600 is
turned off. Either use the power switch or press the Emergency Stop button.
► Do not connect or disconnect a test object while the outputs are active.
► After switching off TESTRANO 600, wait until the red warning light on the front panel is fully
extinguished (see 3.1.1 "TESTRANO 600 front panel" on page 8). As long as this warning light is lit,
there is still voltage and/or current potential on one or more of the outputs.
► Make sure that the test object’s terminals to be connected to TESTRANO 600 do not carry any
voltage potential.
► Make sure that during a test, TESTRANO 600 is the only power source for a test object.
► Switch off the high voltage. Always obey the five safety rules and follow the detailed safety
instructions.
► Before switching on the high voltage, leave the high-voltage test area.
Before performing tests using high voltage, observe the following instructions:
► Before operating TESTRANO 600, ground it as described in sections 1.2.2 "Safety rules" on page 3
and 4 "Application" on page 14.
► Make sure that the ground terminal of the test object is in good condition, clean, and free of oxidation.
► Do not connect any cable to the test object without a visible grounding of the test object.
► Do not remove any cables from TESTRANO 600 or the test object during a test.
► Do not use inadequately rated supply cords.
► Before connecting cables to TESTRANO 600’s high-voltage or current outputs, or other conducting
parts that are not protected against accidental contact, press the Emergency Stop button. Do not
release it unless an output signal is absolutely necessary for the test.
► Do not stand right next to or directly underneath a connection point. The clamps may fall off and hit
you.
The red warning light on the TESTRANO 600 front panel indicates hazardous voltage and/or current
levels on the outputs. The green light indicates that the TESTRANO 600 outputs are not active.
Note: If none or both lights on the front panel are lit, TESTRANO 600 is either not supplied by mains or
it is defective. In this case do not use it anymore.
► Always lock connectors properly.
► The counterpart of the sockets are locking connectors. To lock these connectors safely, insert them
carefully and turn clockwise until you feel them click into place. Check if they are locked by trying to
turn counterclockwise without pulling the silver latch.
► To remove the locking connectors, unlock them by pulling the silver latch.
► Do not insert any objects into any input/output socket.
► Do not operate TESTRANO 600 under ambient conditions that exceed the temperature and humidity
limits listed in chapter 7 "Technical data" on page 28.
► Check any additional equipment (for example, your computer) for environmental conditions before
use.
► Use dry and clean cables. In dusty regions, use protective caps. To avoid leakage current, make sure
that the cables have ground contact.
► Only use cables supplied by OMICRON.

4 OMICRON
Safety instructions

► Position the measurement setup in a way that you can easily disconnect TESTRANO 600 from
mains. If permanently connected, make sure that the measurement setup is positioned in a way that
the switch or circuit breaker can be easily reached.
► Do not operate TESTRANO 600 and its accessories in the presence of explosives, gas or vapors.
► If TESTRANO 600 or its accessories do not seem to function properly (for example, in case of cable
damages, abnormal warming or overheating of components), stop using them and contact
OMICRON support (see "Support" on page 36).
► Observe the high-voltage areas.
► Always obey the internal safety instructions for working in areas with high voltage to avoid injury.

Distance:
min. 1.5 m/5 ft
Device under test

TESTRANO 600

CP TD1

Barrier height:
1 - 1.4 m / 3.3 - 4.6 ft

Safe area High-voltage test area

Figure 1-1: Illustration of safe area and high-voltage area established for working with TESTRANO 600
and CP TD1

1.4 Orderly measures


The TESTRANO 600 Getting Started or alternatively the e-book always has to be available on the site
where TESTRANO 600 is operated.
The users of TESTRANO 600 must read this manual before operating TESTRANO 600 and observe the
safety, installation, and operation instructions therein.
TESTRANO 600 and its accessories may be used only as described in this Getting Started. Any other
use is not in accordance with the regulations. The manufacturer and the distributor are not liable for
damage resulting from improper use. The user alone assumes all responsibility and risk.
Following the instructions provided in this Getting Started is also considered part of being in accordance
with the regulations.
Opening TESTRANO 600 or its accessories invalidates all warranty claims.

OMICRON 5
TESTRANO 600 Getting Started

1.5 Disclaimer
Using TESTRANO 600 in any way differing from the one mentioned above is considered improper use,
and will invalidate all customer warranty claims and exempt the manufacturer from any liability to
recourse.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.

1.6 Compliance statement


Declaration of conformity (EU)
The equipment adheres to the guidelines of the council of the European Community for meeting the
requirements of the member states regarding the electromagnetic compatibility (EMC) directive, the low
voltage directive (LVD) and the RoHS directive.

FCC compliance (USA)


This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be
required to correct the interference at his own expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.

Declaration of compliance (Canada)


This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.

1.7 Recycling
This test set (including all accessories) is not intended for household use. At
the end of its service life, do not dispose of the test set with household waste!
For customers in EU countries (incl. European Economic Area)
OMICRON test sets are subject to the EU Waste Electrical and Electronic
Equipment Directive 2012/19/EU (WEEE directive). As part of our legal obligations
under this legislation, OMICRON offers to take back the test set and ensure that it is
disposed of by authorized recycling agents.
For customers outside the European Economic Area
Please contact the authorities in charge for the relevant environmental regulations in
your country and dispose the OMICRON test set only in accordance with your local
legal requirements.

6 OMICRON
Introduction

2 Introduction
2.1 Designated use
In combination with the CP TD1 or as a stand-alone unit, TESTRANO 600 is a multi-purpose power
transformer test system for routine and diagnostic testing of power transformers during manufacturing,
commissioning and maintenance.
The various partly automated tests are defined and parameterized via the front panel control of a built-
in embedded PC or via an externally connected laptop.

2.2 Device variants


TESTRANO 600 is available with two interface variants.
With multi-touch screen and USB port
Controlled via the embedded PC using TouchControl or via a connected laptop using the Primary Test
Manager software
Without touch screen and embedded PC
Controlled only via laptop using the Primary Test Manager software

OMICRON 7
TESTRANO 600 Getting Started

3 Hardware overview
3.1 TESTRANO 600
► Refer to user manual chapter 7 "Technical data" on page 28 for detailed hardware information.

3.1.1 TESTRANO 600 front panel


Multi-touch screen* USB* Emergency Stop button
Port for USB drive

Red warning light Green light


Indicates possibly hazardous voltage Safe operation
and/or current levels on the Start/stop button
TESTRANO 600 outputs
*Display version only

Figure 3-1: TESTRANO 600 front panel with display

8 OMICRON
Hardware overview

3.1.2 TESTRANO 600 side panel


TAP CHANGER: Connection to tap changer

EXTERNAL BOOSTER*: Connection to CP TD1

Equipotential ground terminal

Warning light 2 (steady-on orange):


voltage > 42 V

Warning light 1 (flashing red):


current > 3 mA

HIGH VOLTAGE terminal*


for connection to primary side

LOW VOLTAGE terminal*


for connection to secondary side
EtherCAT®
RJ-45 socket for connecting an
external EtherCAT® slave to
TESTRANO 600

LEDs indicating the EtherCAT®


communication state

SAFETY 1 (IN for remote control)


Master connector for OMICRON
safety accessories

SAFETY 2 (OUT for remote control)


Slave connector for OMICRON safety
accessories

Power switch

16A
Resetable mains overcurrent SERIAL Mains power socket (one-phase)
protection Connection to CP TD1

NETWORK
RJ-45 socket for connecting TESTRANO 600 to an external computer

* Max. output voltage: 300 VRMS

Figure 3-2: TESTRANO 600 side panel

OMICRON 9
TESTRANO 600 Getting Started

3.1.3 Safety and warning indicators


TESTRANO 600 provides the following warning lights to indicate safe and dangerous operating
conditions.

Table 3-1: Warning lights


Light Description Device sState Operating condition
Front panel
Green light on the front TESTRANO 600 is up and running Safe operating condition as
panel is on in the stand-by mode. long as no voltage is
applied from outside – as
Blue ring on the long as the warning symbol
A test is ready to start.
Start/Stop button is lit on the side panel is off.
The Start/Stop button has just
Blue ring on the been pressed. There may be Dangerous
Start/Stop button is hazardous voltage and/or current operating
flashing levels on the TESTRANO 600 condition
outputs.
A test is running. There probably
Red warning light on Dangerous
are hazardous voltage and/or
the front panel is operating
current levels on the
flashing condition
TESTRANO 600 outputs.
Side panel
There are hazardous current
Warning light 1 on the levels (>3 mA) on the Dangerous
side panel is flashing TESTRANO 600 inputs/outputs operating
(red) independent of the measurement condition
state.
There are hazardous voltage
Warning light 2 on the levels (>42 V) on the Dangerous
side panel is on TESTRANO 600 inputs/outputs operating
(orange) independent of the measurement condition
state.

► Always observe the warning lights while working with TESTRANO 600.
► Do not cover the warning lights (for example, with a computer) – they indicate possible hazards.
If no or both warning lights are on, the unit is defective or not supplied by mains.
► If TESTRANO 600 appears to be defective, do not use it anymore. Contact OMICRON support
(see "Support" on page 36).

10 OMICRON
Hardware overview

3.1.4 Emergency Stop button


Pressing the Emergency Stop button immediately shuts off all
TESTRANO 600 outputs and stops the running measurement. When the
Emergency Stop button is pressed, you cannot start any measurements.
► Only use the Emergency Stop button in an emergency or to guaranty the
safe connection/disconnection of cables.
► During regular operation, stop tests via the Start/Stop button or the
software.

3.1.5 TESTRANO 600 measuring cables


► To connect a measuring cable to TESTRANO 600, insert the connector and turn it to the right until it
locks with a "click".

For disconnection:
► Hold the connector and pull back the
silver latch.
► Turn the connector to the left and gently
pull it out.

NOTICE
Equipment damage possible
► Do not pull the cable when disconnecting.
► Hold, turn and gently pull the connector for disconnection.

OMICRON 11
TESTRANO 600 Getting Started

Table 3-2: TESTRANO 600 measuring cables

Item Picture Description

High-voltage cable
marked red
• Polarity protection: only suitable for
HIGH VOLTAGE and LOW VOLTAGE
sockets
• 15 m length
• 8 poles
• cross-section:
4 × 4 mm2 for output
4 × 1 mm2 for measurement
• Neutrik® plug
Low-voltage cable
marked yellow

• Polarity protection: only suitable for


TAP CHANGER socket
• 15 m length
Tap changer cable
• 8 poles
• 8 × 2.5 mm2 cross-section
• Neutrik® plug

12 OMICRON
Hardware overview

3.2 CP TD1
► Refer to the CP TD1 User Manual for detailed information and safety instructions.

3.2.1 Grounding terminal and booster input

Equipotential ground
terminal
BOOSTER IN
Booster input

Figure 3-3: Left side view of the CP TD1

3.2.2 Serial interface connector and measuring inputs

SERIAL
Serial interface
connector
IN_B
measuring input
IN_A
measuring input

Figure 3-4: Right side view of the CP TD1

3.2.3 High-voltage connector

Grounding terminal

High-voltage output

Figure 3-5: Back side view of the CP TD1

OMICRON 13
TESTRANO 600 Getting Started

4 Application
4.1 Safety precautions in the substation
Before setting TESTRANO 600 into operation and carrying out a test, it is essential that you have read
and understood chapter 1 "Safety instructions" on page 3.
► Be aware that all output sockets of TESTRANO 600 can carry life-hazardous voltage and current.
► Only use TESTRANO 600 with a solid connection to ground.
► Separate your working area – see Figure 1-1: "Illustration of safe area and high-voltage area
established for working with TESTRANO 600 and CP TD1" on page 5.
► Tests with high voltages and currents must only be carried out by authorized and qualified personnel.
► Personnel receiving training, instructions, directions, or education on high- voltage/current tests
should remain under the constant supervision of an experienced operator while working with the
equipment.
► The instructions have to be renewed at least once per year.
► The instructions must be available in written form and signed by each person assigned to do high-
voltage/current tests.
Prior to connecting a test object to TESTRANO 600, the following steps need to be carried out by an
authorized employee of the utility:
► Protect yourself and your working environment against an accidental re-connection of high voltage
by other persons and circumstances.
► Verify that the test object is safely isolated.
► Earth-connect and shorten out the test object’s terminals using a grounding set.
► Protect yourself and your working environment with a suitable protection against other (possibly live)
circuits.
► Protect others from accessing the high-voltage area and accidentally touching live parts by setting up
a suitable barrier and, if applicable, warning lights.
► If there is a longer distance between the location of TESTRANO 600 and the area of danger, a
second person with an additional Emergency Stop button is required.

14 OMICRON
Application

4.2 Preparing the test setup


DANGER
Death or severe injury caused by high voltage or current
The output sockets of TESTRANO 600 can carry life-hazardous voltage potential and
life-hazardous currents.
► Do not use TESTRANO 600 without a solid connection to ground.
► Before switching on TESTRANO 600, make sure it is completely dry.
► Before connecting any cables, check them for damage. Make sure that the connectors
are clean and dry and that the insulation is intact.

1. Make sure that the power switch on the TESTRANO 600 side panel is turned off.
2. Press the Emergency Stop button.
3. Connect TESTRANO 600 to:
• equipotential ground: Ground TESTRANO 600 with a cable of at least 6 mm2 cross-section as
close as possible to the operator.
• the computer (optional if TESTRANO 600 is used with TouchControl)
• the mains power supply
Optional
Connect the CP TD1 to TESTRANO 600
a) Connect to ground.
Without trolley:
► Properly connect the TESTRANO 600 and CP TD1 grounding terminals to substation ground.
With trolley (optional):
► Properly connect the TESTRANO 600 and CP TD1 grounding terminals to the trolley’s ground
bar.
► Connect the trolley’s ground bar to earth.
b) Connect the CP TD1 BOOSTER IN to the TESTRANO 600 EXTERNAL BOOSTER using the
booster cable.
c) Connect the CP TD1 SERIAL to TESTRANO 600 SERIAL with the data cable.
Note: The serial cable also provides the power supply for the CP TD1.

4. Turn on the power switch on the TESTRANO 600 side panel.


5. The green light and the blue ring of the Start/Stop button are switched on, showing that
TESTRANO 600 does not output dangerous voltage or current.
6. If the PE connection is defective or if the power supply has no galvanic connection to ground, a
warning message appears.
Note: If TESTRANO 600 is supplied by mains and switched on, and no or both warning lights are on,
the unit might be defective. Contact OMICRON support (see "Support" on page 36).

OMICRON 15
TESTRANO 600 Getting Started

4.3 Connecting to the transformer


Preparing the software
TouchControl Primary Test Manager

► Select a test. ► Create a job with tests or select a manual test.


► After defining the asset’s vector group, tap
► View the wiring diagram in the General tab of
Wiring to display the wiring diagram for the
the test.
test.
► Lock TESTRANO 600 using the Software
► Lock your computer.
lock.
► Connect the test leads to TESTRANO 600 (or the CP TD1, if applicable).

16 OMICRON
Application

Connection to the transformer


DANGER
Death or severe injury caused by high voltage or current possible
► Before connecting any test leads to the transformer, turn off and disconnect any
voltage to and from the transformer (e. g. high voltage on the main terminals, control
voltage of the tap changer).
► Ground and short-circuit its terminals using a grounding set.

1. Connect the test leads to TESTRANO 600 as shown in the wiring diagram and in the order given
below:

TESTRANO 600
00

Figure 4-1: Connection sequence TESTRANO 600 to transformer

I. Connect the high-voltage (red), low-voltage (yellow) and tap changer cables to TESTRANO 600.
II. Connect the high-voltage (red) and low-voltage (yellow) cables to the transformer’s main
terminals.
III. Connect the tap changer cable to the appropriate terminals in the control cabinet of the
transformer (see Figure 4-2 below).
IV. Re-connect and turn on the voltage of the tap changer.

Note: If the tap changer control voltage exceeds 42 V, the orange warning light 2 on the side panel will
indicate a hazardous voltage on the TESTRANO 600 inputs (see 3.1.3 "Safety and warning indicators"
on page 10).

OMICRON 17
TESTRANO 600 Getting Started

Tap changer connection


► For the measurement of motor current and voltage, connect VIn+ and VIn- (green) as illustrated
below. On a three-phase motor VIn+ can be connected to either L1, L2 or L3.
Connect the current clamp to CurrentIn+ and CurrentIn-.
► For tap changer control connect TapUp+ and TapUp- (blue) to the connectors controlling the upward
switching of the tap changer.
Connect TapDown+ and TapDown- (purple) to the connectors controlling the downward switching
of the tap changer.

PE Motor

Measuring motor current and voltage

Up

Down

Tap control

Figure 4-2: Connection scheme tap changer cable to tap changer

Optional
Connect the CP TD1 to the transformer
If you connected the CP TD1 to TESTRANO 600 (step 4 in section 4.2), connect the IN_A, IN_B and
high-voltage output of the CP TD1 to the transformer.
► Refer to the CP TD1 User Manual for more information on safely connecting the CP TD1 to a device
under test.

2. Erect a barrier separating the safe area from the high-voltage test area (see page 5).
3. Remove the grounding set from the test object.
4. Release the Emergency Stop button.

18 OMICRON
Application

4.4 Measurement
TouchControl Primary Test Manager

► Establish the connection between PTM and


► Disable the Software lock.
TESTRANO.
► Enter/adjust the test settings in the Settings
► Enter/adjust the test settings in the Settings and conditions area.
view. ► Select a standard in the Assessment area (if
applicable).
► Tap ► Press
The blue ring on the Start/stop button lights up.
► Press the Start/stop button to start the test.
The blue ring and the red warning light are now flashing for approx. 3 seconds.
► To suspend the test, press the Start/stop button on the TESTRANO 600 front panel.
► In an emergency, press the Emergency Stop button to stop the test.
After the measurement is completed or stopped, the green warning light switches on.

TouchControl displays the results in the Primary Test Manager displays the results in the
Measurement view of the test. Measurements section of the test.

► To perform additional tests, repeat the steps in chapters 4.3 to 4.4.

OMICRON 19
TESTRANO 600 Getting Started

4.5 Disconnection
1. Wait until the green light on the TESTRANO 600 front panel is on and the warning lights on the front
and side panel are off.
Note: If the tap changer control voltage exceeds 42 V, the orange warning light 2 on the side panel
will indicate a hazardous voltage on the TESTRANO 600 inputs (see 3.1.3 "Safety and warning
indicators" on page 10).
► Disconnect the tap changer cable to extinguish warning light 2.
2. Press the Emergency Stop button on the TESTRANO 600 front panel.
DANGER
Death or severe injury caused by high voltage or current
► Never unplug any cables while the measurement is running.
► Only disconnect cables when all of the following apply:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.
3. To prevent anyone from starting a test, use the Software lock in the TouchControl software (see
chapter 5.5 "Software lock" on page 32) and/or lock your computer.
4. Remove the barrier between the high-voltage area and the safe area.
DANGER
Death or severe injury caused by high voltage or current possible
► Before touching any part of the transformer, ground and short-circuit its terminals using
a grounding set.
5. Disconnect all cables from the transformer.
6. Disconnect all cables from TESTRANO 600 and, if applicable, from the CP TD1.
7. Switch TESTRANO 600 off by pressing the mains power switch on the side panel.
8. Disconnect the mains power cord.
9. Remove the equipotential ground as the last connection that is removed first from TESTRANO 600
and then on the substation side.

20 OMICRON
Testing with TouchControl

5 Testing with TouchControl


5.1 Getting started
The following table lists the basic steps necessary to complete a measurement using TouchControl.
► For more information on each step refer to the user manual chapters listed on the right.

Step User manual chapter


Safety instructions
Hardware overview
1. SAFETY Safety and warning indicators
Emergency Stop button
Application

2. Start TESTRANO 600 TESTRANO 600 side panel

3. Enter asset info Asset info

4. Add tests TouchControl tests

Safety instructions
TESTRANO 600 measuring cables
5. Connect to transformer
TESTRANO 600
Application
Wiring diagrams

6. Prepare test Settings view

7. Measurement Measurement

OMICRON 21
TESTRANO 600 Getting Started

5.2 Test view


The TESTRANO 600 tests are divided into the Settings and Measurement views.
► Swipe right or left or tap the arrow on the side of the screen to change between the views.
In the Settings view, you can enter asset data and the values necessary for your test. The
Measurement view displays test results in a live, table or plot view, depending on the test.

5.2.1 Settings view

Wiring diagram for this test

TESTRANO 600 cables:


HV = red, LV = yellow

OLTC/DETC position
(HV or LV side)
Tap or swipe left to switch to
Measurement view

Tap to expand and collapse the submenu

Tap to start the test

Figure 5-1: Turns ratio test - Settings view

Entering values
► Tap a box, then use the numpad to enter or correct a value.
► If needed, tap the metric prefixes below after entering a value:
– k for kilo-
– M for mega-
– m for milli-
► Use the slider to increase or decrease the displayed value. Release the slider to stop at the desired
value.
Note: The slider will stop at the minimum/maximum value.

22 OMICRON
Testing with TouchControl

Defining a tap changer


► To define a tap changer press the tap changer icon in the test’s Settings view.
► Tap Copy to all to copy the settings to all tests that have not yet been executed.

Table 5-1: Steps in the Define Tap Changer view

Tap changer settings – Define Tap Changer


► Mark the type of tap changer on the left and tap Yes to confirm and display
Available
the settings.
Position ► Choose the tap changer’s transformer side: HV or LV.
Tap scheme ► Select the notation scheme for tap identification from the drop-down box.
No. of taps ► Enter the number of taps.
Voltage table The Voltage table displays the voltage for each tap. You can either enter each
values manually or have them calculated.
► Enter at least the first two values and press Calculate .
► Compare the calculated values with the nominal values on the nameplate
before proceeding.
Add more taps at the end of the voltage table.

Delete a tap from the voltage table.

Delete all taps from the voltage table.

Insert a tap below the marked tap.


Middle ► Enter the voltage for the middle tap (rated voltage) and the deviation value for
the calculation, then press Calculate.
First/middle/last ► Enter the voltages for the first, middle and last tap, then press Calculate.

OMICRON 23
TESTRANO 600 Getting Started

5.2.2 Measurement view


Switch between
Enter full screen view
views

Test assessment
(see section 5.4)
Tap arrows to sort list in ascending or
descending order

Delete one or all results

Tap or swipe right to switch to Settings view

Tap to expand and collapse submenus and drop-down lists

Tap to start the test

Figure 5-2: Turns ratio test – Measurement view

5.2.3 Wiring diagrams


► Tap Wiring to display the wiring diagram for the test and vector group.
The colors in the wiring diagram represent the cable ends (see 3.1.5 "TESTRANO 600 measuring
cables" on page 11).

Figure 5-3: Wiring diagram for a TTR on a transformer with vector group YNd5

24 OMICRON
Testing with TouchControl

5.3 Measurement
► Refer to chapter 1 "Safety instructions" on page 3 for detailed information about safe testing.
► If in doubt, contact OMICRON support (see "Support" on page 36).

DANGER
Death or severe injury caused by high voltage or current
► Do not unplug any cables while the measurement is running.
► Only remove cables when all of the following apply to TESTRANO 600:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.

WARNING
Death or severe injury caused by high voltage or current possible
► Do not enter the high-voltage area during the test.
► Do not touch any part of the transformer before grounding and short-circuiting its
terminals.

1. Tap Start.

2. The blue ring on the Start/stop button lights up.

3. Press the Start/stop button to start the test.


4. The blue ring and the red warning light are now flashing for approx. 3 seconds.
► To suspend the test, press the Start/stop button on the TESTRANO 600 front panel.
► In an emergency, press the Emergency Stop button to stop the test.
5. After the measurement is completed or stopped, the green warning light switches on and
TouchControl displays the results in the Measurement view.

5.4 Assessment
► Tap Assess to expand the available assessment states:

Pass: results accepted; measurement OK

Fail: results rejected; measurement failed

Investigate: marked for further investigation


Note: Each measurement of an assessed test will individually be tagged with the selected assessment
when displayed in Primary Test Manager or a report.

OMICRON 25
TESTRANO 600 Getting Started

6 Testing with Primary Test Manager


6.1 Getting started
The following table lists the basic steps necessary to complete a measurement using TESTRANO 600
and the Primary Test Manager guided workflow.
► For more information refer to the user manual chapters listed on the right.
Step User manual chapter
Safety instructions
Hardware overview
1. SAFETY Safety and warning indicators
Emergency Stop button
Application

2. Connection to TESTRANO 600 Preparing the test setup


TESTRANO 600

TESTRANO 600 side panel


3. Start device and software
Software start and device update

Location view
4. Location and asset
Asset view

5. Jobs Jobs

Test view
PTM Transformer tests
6. Tests
PTM Bushing tests
Device-independent PTM tests
Safety instructions
TESTRANO 600 measuring cables
7. Connection to device under test
TESTRANO 600
Application
Connecting to the transformer
8. Test settings Performing tests

9. Test assessment Assessing measurement results

10. Measurement Measurement

26 OMICRON
Testing with Primary Test Manager

6.2 Measurement
► Refer to chapter 1 "Safety instructions" on page 3 for detailed information about safe testing.
► If in doubt, contact OMICRON support (see "Support" on page 36).

DANGER
Death or severe injury caused by high voltage or current
► Do not unplug any cables while the measurement is running.
► Only remove cables when all of the following apply to TESTRANO 600:
• The red warning light on the front panel is off.
• The warning lights on the side panel are off.
• The green light on the front panel is on.
If all lights on TESTRANO 600 are off, the device is defective or not supplied by mains.

WARNING
Death or severe injury caused by high voltage or current possible
► Do not enter the high-voltage area during the test.
► Do not touch any part of the transformer before grounding and short-circuiting its
terminals.

1. Press Start in Primary Test Manager.

2. The blue ring on the Start/stop button lights up.


3. Press the Start/stop button to start the test.
4. The blue ring and the red warning light are now flashing for approx. 3 seconds.
► To suspend the test, press the Start/stop button on the TESTRANO 600 front panel.
► In an emergency, press the Emergency Stop button to stop the test.
5. After the measurement is completed or stopped, the green warning light switches on and
Primary Test Manager displays the results in the Measurements view.

OMICRON 27
TESTRANO 600 Getting Started

7 Technical data
At the time of factory adjustment all units are within the typical accuracy values specified in this
document.
Typical accuracy means that 98 % of all units meet the specified values at 23 °C ± 5 °C/73 ºF ± 10 ºF,
after a warm-up time of more than 25 min., and in a frequency range of 45 Hz to 65 Hz or DC.
The typical accuracy values multiplied by 3 are guaranteed at an ambient temperature of
23 ºC ± 5 ºC/73 ºF ± 10 ºF, after a warm-up time more than 25 min., and in a frequency range of
45 Hz to 65 Hz or DC.
Accuracy values indicate that the error is smaller than:
± (value read × reading error [rd] + full scale of range × range error [rg]).
For mains voltages below 190 V AC the system is subject to power restrictions.
OMICRON suggests that you send in your unit for calibration at least once a year.
Technical data are subject to change without notice.

CAT level
The CAT level required depends on the TESTRANO 600 application. All CAT ratings are defined for sea
levels below 2000 m. There are some limitations between 2000 m and 5000 m sea level (see
section 7.4 "Environmental conditions" on page 34).
CAT I is required when the measured voltage is generated by the test set itself. No voltages from other
sources are measured.
CAT II is required when measuring within electrical devices or between mains supply and devices.
CAT III is required when measuring in electrical installations such as control cubicles that are still
connected to the station battery or mains. The electrical installations are protected by a fuse.

28 OMICRON
Technical data

7.1 TESTRANO 600 specifications


7.1.1 Output specifications
Table 7-1: General output specifications

Characteristic Rating
Frequency DC or 15 Hz … 599 Hz
Vmains P30 s Pcontinuous
Power >100 VRMS 1500 W 1000 W
>190 VRMS 4000 W 2400 W

Table 7-2: Voltage source (HV and LV connectors)

Source Range Imax, continuous


3 × 0 … ±113 VDC
DC high range 16 ADC
1 × 0 … ±340 VDC

3 × 0 … ±56 VDC
DC low range 33 ADC
1 × 0 … ±170 VDC
AC high range
3 × 0 … 230 VRMS (LN) 100 mARMS
low current
3 × 0 … 80 VRMS (LN)
AC high range 16 ARMS
1 × 0 … 240 VRMS

3 × 0 … 40 VRMS (LN)
AC low range 33 ARMS
1 × 0 … 120 VRMS

Table 7-3: Voltage source accuracy

Characteristic Accuracy1
Voltage accuracy DC 0.033 % rd ± 0.017 % range
Voltage accuracy AC (50 Hz) at burden open
0.33 % rd ± 0.17 % range
load
Phase accuracy AC (50 Hz) burden open load,
± 0.36°
V>20 VRMS
1. Typical accuracy at 23 °C ±5 K

OMICRON 29
TESTRANO 600 Getting Started

Table 7-4: Current source (HV and LV)


Source Range Vmax, continuous
3 × 0 ... ±33 ADC or
56 VDC
DC source high range 1 × 0 ... ±100 ADC (3 × 33.33 ADC)
1 × 0 ... ±33 ADC 170 VDC

3 × 0 ... ±16 ADC


113 VDC
DC source low range 1 × 0 ... ±50 ADC (3 × 16.66 ADC)

1 × 0 ... ±16 ADC 340 VDC


AC source high range 3 × 0 ... 33 ARMS (LN) or 40 VRMS
1 × 0 ... 100 ARMS (3 × 33.33 ARMS)
1 × 0 ... 33 ARMS 120 VRMS
AC source low range 3 × 0 ... 16 ARMS (LN) or 80 VRMS

1 × 0 ... 50 ARMS (3 × 16.66 ARMS)

1 × 0 ... 16 ARMS 240 VRMS

Table 7-5: Current source accuracy

Characteristic Accuracy1
Current accuracy DC 0.033 % rd ± 0.017 % range
Current accuracy AC
0.33 % rd ± 0.17 % range
50/60 Hz at burden 0.1 Ω
1. Typical accuracy at 23 °C ±5 K

Table 7-6: Voltage source (Booster)

Source Range Imax, cont.1 Imax, 30 s1


Power – 3 kVA 4.4 kVA
AC high voltage 1 × 0 ... 240 VRMS 16 ARMS 20 ARMS
Characteristic Rating
Channels 1
Voltage accuracy2 AC (50/60 Hz) 0.33 % rd ± 0.16 % range
at burden open load
1. Within the above specified power limit
2. Typical accuracy at 23 °C ±5 K

30 OMICRON
Technical data

7.1.2 Input specifications


Table 7-7: Voltage inputs (HV and LV) 3 phase

Range name Range value Accuracy1


AC
300 mVRMS 0 ... 300 mVRMS 0.01 % rd + 0.003 % range
3 VRMS 0 ... 3 VRMS 0.01 % rd + 0.003 % range
30 VRMS 0 ... 30 VRMS 0.01 % rd + 0.003 % range
300 VRMS 0 ... 300 VRMS 0.012 % rd + 0.003 % range
DC
42.4 mVDC 0 ... 42.4 mVDC 0.022 % rd + 0.032 % range
424 mVDC 0 ... 424 mVDC 0.01 % rd + 0.017 % range
4.24 VDC 0 ... 4.24 VDC 0.007 % rd + 0.012 % range
42.4 VDC 0 ... 42.4 VDC 0.01 % rd + 0.017 % range
424 VDC 0 ... 424 VDC 0.007 % rd + 0.012 % range
1. Typical accuracy at 23 °C ±5 K

Typical phase accuracy at 50/60 Hz, V>30 % of used range: 0.017°

Table 7-8: Voltage input (Booster)

Range name Range value Accuracy1


280 VRMS 0 ... 280 VRMS 0.012 % rd + 0.003 % range
1. Typical accuracy at 23 °C ±5 K

Typical phase accuracy at 50/60 Hz, V>30 % of used range: 0.017°

Table 7-9: Current inputs (internal)

Range name Range value Accuracy1


AC
4 ARMS 0 ... 4 ARMS 0.036 % rd + 0.0033 % range
40 ARMS 0 ... 40 ARMS 0.023 % rd + 0.013 % range
DC
0.56 ADC 0 ... 0.56 ADC 0.01 % rd + 0.023 % range
5.6 ADC 0 ... 5.6 ADC 0.037 % rd + 0.026 % range
56 ADC 0 ... 56 ADC 0.008 % rd + 0.01 % range
1. Typical accuracy at 23 °C ±5 K

Typical phase accuracy at 50/60 Hz, I>30 % of used range: 0.017°

OMICRON 31
TESTRANO 600 Getting Started

Table 7-10: On-load tap changer measurement (tap changer connector)

Characteristic Rating
Voltage 300 VRMS

Accuracy1 AC (50/60 Hz)/DC 0.07 % rd + 0.07 % range


Current clamp input 3 VRMS
Tap up/down switch current 300 mA continuous, 9 A for 0.7 s
(AC permitted only)
Tap up/down switch voltage 300 VRMS
(AC permitted only)
1. Typical accuracy at 23 °C ±5 K

7.2 Combined values


Table 7-11: Resistance measurement AC

Range name Current Range Accuracy1


1 Ω ... 10 Ω 0.053 % rd + 0.033 % range
0.1 Ω ... 1 Ω 0.053 % rd + 0.033 % range
40 ARMS 30 ARMS 0.01 Ω ... 0.1 Ω 0.053 % rd + 0.033 % range
0.001 Ω ... 0.01 Ω 0.053 % rd + 0.033 % range
0.0001 Ω ... 0.001 Ω 0.063 % rd + 0.033 % range
10 Ω ... 100 Ω 0.053 % rd + 0.037 % range
1 Ω ... 10 Ω 0.053 % rd + 0.037 % range
4 ARMS 3 ARMS 0.1 Ω ... 1 Ω 0.053 % rd + 0.037 % range
0.01 Ω ... 0.1 Ω 0.053 % rd + 0.037 % range
0.001 Ω ... 0.01 Ω 0.067 % rd + 0.037 % range
1. Typical accuracy at 23 °C ±5 K

32 OMICRON
Technical data

Table 7-12: Resistance measurement DC

Range name Current Range Accuracy1


1 Ω ... 10 Ω 0.037 % rd + 0.017 % range
0.1 Ω ... 1 Ω 0.04 % rd + 0.027 % range
40 ARMS 30 ADC 0.01 Ω ... 0.1 Ω 0.033 % rd + 0.017 % range
0.001 Ω ... 0.01 Ω 0.037 % rd + 0.027 % range
0.0001 Ω ... 0.001 Ω 0.05 % rd + 0.043 % range
10 Ω ... 100 Ω 0.1 % rd + 0.18 % range
1 Ω ... 10 Ω 0.1 % rd + 0.267 % range
4 ARMS 3 ADC 0.1 Ω ... 1 Ω 0.1 % rd + 0.18 % range
0.01 Ω ... 0.1 Ω 0.1 % rd + 0.267 % range
0.001 Ω ... 0.01 Ω 0.113 % rd + 0.433 % range
1. Typical accuracy at 23 °C ±5 K

Table 7-13: Ratio measurement

Range name
Voltage at HV Range1 Accuracy2
(LV voltage range)
1 0.03 % rd +
300 VRMS ------------------
1 …10 0.043 % range
1 0.027 % rd +
30 VRMS -------------------------
10 …100 0.043 % range
230 VRMS HV (LN)
1 0.027 % rd +
3 VRMS -------------------------------
100 …1000 0.043 % range
1 0.027 % rd +
300 mVRMS --------------------------------------
1000 …10000 0.043 % range
LV
1. Range = --------
HV

2. Typical accuracy at 23 °C ±5 K

OMICRON 33
TESTRANO 600 Getting Started

7.3 Power supply specifications

Table 7-14: Power supply specifications

Characteristic Rating
Nominal 100 V … 240 VAC
Voltage
Permitted 85 V … 264 VAC
Current Nominal 16 A
Nominal 50 Hz/60 Hz
Frequency
Permitted 45 Hz … 65 Hz
Automatic circuit breaker with magnetic
Power fuse
overcurrent tripping at I >16 A
Continuous <3.6 kW
Power consumption
Peak <5.0 kW
Current consumption, continuous <16 AAC
Connector type IEC320/C20, 1 phase

7.4 Environmental conditions


Table 7-15: Climate

Characteristic Rating
Operating -10 ºC … +55 ºC/+14 ºF…+131 ºF
Temperature
Storage -30 ºC … +70 ºC/-22 ºF…+158 ºF
2000 m/6550 ft, up to 5000 m/16400 ft with
Operating
Max. altitude limited specifications1
Storage 12 000 m/40 000 ft
1. Output TAP CHANGER (CAT III / 300 V): from 2000 m/6550 ft to 5000 m/16400 ft altitude only CAT II compliance or CAT III
compliance with half voltage

7.5 Mechanical data


Table 7-16: Mechanical data

Characteristic Rating
464 × 386 × 229 mm
With cover, without handles
Dimensions 18.3 × 15.2 × 9 in
(w × h× d) 580 × 386 × 229 mm
With cover, with handles
22.8 × 15.2 × 9 in
Device with display 20.6 kg/45.5 lb
Weight
Device without display 19.5 kg/43 lb

34 OMICRON
Technical data

7.6 Standards
Table 7-17: Standards conformity

EMC, safety
EMC IEC/EN 61326-1
(industrial electromagnetic environment)
FCC subpart B of part 15, class A
Safety IEC/EN/UL 61010-1,
IEC/EN/UL 61010-2-30

Other
IEC/EN 60068-2-27
Shock
(15 g/11 ms, half-sinusoid, 3 shocks in each axis)
IEC/EN 60068-2-6
Vibration (frequency range 10 Hz…150 Hz, acceleration 2 g continuous
(20 m/s2/65 ft/s2), 20 cycles per axis)
IEC/EN 60068-2-78
Humidity (5 % … 95 % relative humidity, no condensation),
tested at 40 ºC/104 ºF for 48 hours

OMICRON 35
TESTRANO 600 Getting Started

Support
When you are working with our products we want to provide you with the greatest
possible benefits. If you need any support, we are here to assist you!

24/7 Technical support – get support


www.omicronenergy.com/support
At our technical support hotline, you can reach well-educated technicians for all
of your questions. Around the clock – competent and free of charge.
Make use of our 24/7 technical support hotlines:
Americas: +1 713 830-4660 or +1 800-OMICRON
Asia-Pacific: +852 3767 5500
Europe / Middle East / Africa: +43 59495 4444
Additionally, you can find our Service Center or Sales Partner closest to you at
www.omicronenergy.com.

Customer area – stay informed


www.omicronenergy.com/customer
The customer area on our website is an international knowledge exchange
platform. Download the latest software updates for all products and share your
own experiences in our user forum.
Browse through the knowledge library and find application notes, conference
papers, articles about daily working experiences, user manuals and much more.

OMICRON Academy – learn more


www.omicronenergy.com/academy
Learn more about your product in one of the training courses offered by the
OMICRON Academy.

OMICRON electronics GmbH, Oberes Ried 1, 6833 Klaus, Austria, +43 59495

36 OMICRON
ENU 1161 03 02

You might also like