AN3358 OneWireViewer UsersGuide
AN3358 OneWireViewer UsersGuide
Keywords: onewireviewer, 1-wire,ibutton, onewire, viewer, software, evaluation, eval kit, java, web start, webstart, user's guide, manual, howto, debugging
Abstract: This User's Guide explains the OneWireViewer software program and how it can be used to evaluate the unique features of 1-Wire® and iButton®
devices.
Introduction
The OneWireViewer is a Java™-based software package to explore Maxim's 1-Wire and iButton devices with a personal computer. The 1-Wire and iButton devices
communicate over a single data line plus ground reference, using the 1-Wire protocol. Several 1-Wire adapters are available for USB and serial ports. The
OneWireViewer provides a user interface to evaluate the unique features of 1-Wire and iButton products, including Thermochron® (temperature logging),
Hygrochron™ (real-time temperature and humidity), A to D (analog to digital), clock, memory, and file operations.
Installation
Refer to application note 4373, "OneWireViewer and iButton® Quick Start Guide" for the installation instructions for the 1-Wire Drivers, which also install the
OneWireViewer. For additional/expanded help, see application note 5057, "OneWireViewer Tips and Tricks," starting with the section Identifying the System.
Uninstalling
Refer to the application note 5057, "OneWireViewer Tips and Tricks," section How to Install a Newer OneWireViewer Version.
Viewer Menus
File
View
Page 1 of 21
Show Tab in New Window (to view multiple functions or devices simultaneously)
To use this function, first select a device, then select one of the available tabs. Then use Show Tab in New Window, which moves the selected tab into a window of
its own. The main window will then revert back to the Description tab.
Show Device Alternate Names (to see alternate names in the device list)
Example: Thermochron instead of DS1921G-F5 to the right of the ROM ID
Tools
Pick Adapter (to change the communication port and 1-Wire port adapter)
Opens a new window for selecting one of several 1-Wire port adapters, port type, and port number. For more information see application note 5057,
"OneWireViewer Tips and Tricks," section How to Change 1-Wire Adapters and Ports.
XML Tagging (to load device tags that can be displayed in the device list)
Opens a new window to select the file that holds the XML tags of the devices in use. See application note 158, "1-Wire® Tagging with XML," for tag specification
and how to create tag files that are compatible with the OneWireViewer.
Device Poll Rate (to set the frequency at which the 1-Wire net is searched for devices)
The rates are 1s, 5s, 10s, 30s, 1min, 5min, and immediate poll. The typical value is 1s, which yields the fastest response.
Help
About (displays the version numbers of the software components that comprise the OneWireViewer)
Opens a window that displays: the site from which the source code of the 1-Wire API and OneWireViewer can be downloaded; the API version; the version numbers
of the various viewers; and the location of the onewireviewer.properties file on the local hard drive. The OneWireViewer is provided as a source code example in the
1-Wire API for Java Kit.
Below the device list, the total number of devices on the network and the type of port adapter are displayed. In this example, the adapter is a DS9097U. The { }
brackets around the adapter name indicate that it uses a native TMEX driver.
Pause All Searching (to stop/restart searching the network for arriving/departing devices)
When checked, the 1-Wire network is no longer searched. This minimizes the communication traffic on the network (e.g., to facilitate catching events with an
oscilloscope), and freezes the content of the device list. An error message will be generated if one selects a device that has departed from the network. With the
searching paused, one cannot access devices that have arrived because they are not included in the device list. To end the pause, check Show Normal Devices.
Tab Area
Once a device is selected, the Description tab appears with a short description of the device. Depending on the device, additional tabs may appear that provide
access to the applicable device function viewers. See the Supported Devices table below for a list of devices and applicable device viewers.
Page 2 of 21
The 1-Wire and iButton products support a broad set of features. The following table, sorted by family code, shows the applicable device function viewers.
Page 3 of 21
Supported Devices
A
Supported Family
Thermochron Mission Temperature Humidity Switch to Clock Memory File Password
Devices Code
D
DS1990A,
01
DS1990R
DS1993L 06
DS1992L 08
DS1982,
09
DS2502
DS1995L 0A
DS1985,
0B
DS2505
DS1996L 0C
DS1920,
10
DS1820
DS2406 12
DS1971,¹
14 ( )
DS2430A¹
DS28E04 1C
DS2450 20
DS1921 21
DS1822 22
DS1973,
23
DS2433
DS1904,
24
DS2415
DS2438 26
DS2417 27
DS18B20 28
DS2408 29
DS2431 2D
DS2760,
30
DS2762
DS1977 37
DS2413 3A
DS1922,² (
41
DS2422² )
DS1923 41
DS28EA00³ 42
DS28EC20 43
¹It is possible to format this device; however, there is no memory left to store a file.
²A-to-D support applies to the DS2422 only.
³This device supports Chain Mode; see data sheet for more information.
The tab structure of the Thermochron viewer consists of two areas: Command (top), and device data tabs (bottom). The Command tab includes three buttons to
operate the device's logging function and to select the temperature scale (°C or °F). Depending on the size of the Thermochron tab, there can be scroll bars for both
areas.
Page 4 of 21
Sample Thermochron Viewer Tab
Start New Mission (to allow the user to enter all data necessary to start a new mission)
This command opens a new window in which the user enters the parameters for the next mission.
Automatically reads the device status, temperature, histogram, and alarm log.
Automatically displays the device status on the Status tab.
Displays the temperature log as a graphic (if you open the Temperatures tab).
Displays the temperature histogram in text form (if you open the Histogram tab).
Displays the alarm log in text form (if you open the Alarm Log tab).
Allows the temperature scale to be changed between °C (default) and °F; the selected scale is memorized.
Allows activation of automatic rescaling of the temperature log graph.
Allows the graph data to be exported by using the clipboard or saving as a CSV (Excel®) file.
Page 5 of 21
Sample Temperature Log Graphic
Sample Histogram
Page 6 of 21
Start New Mission Window (default)
It is recommended that you synchronize the device's real-time clock (RTC) to the PC. Enter the mission start delay in minutes. Sampling rate, mission start delay,
temperature alarms, and rollover depend on the technical requirements of the mission. See the respective DS1921-series data sheets for guidelines. The clock
alarm functions are not relevant for a device on a mission. The default settings, therefore, can be accepted.
Notes:
The time reference in the alarm log (e.g., "Low alarm started at: 21") indicates minutes after the start of the mission. To find the day and time when the alarm
occurred, manually add the indicated time to the mission start date and time.
The Thermochron viewer does not enable the temperature-high and temperature-low alarm search. As a result, if a temperature alarm occurs during the
mission, the device will not be identified as an alarming device unless there is also a timer alarm.
For the alarm-frequency and alarm-time values to be updated in the device, Enable Clock Alarm? must be checked. Otherwise, the current (i.e., previous
mission's) settings remain in effect for the new mission.
When a mission extends across time zones, the time shown in the exported data (clipboard or csv file) refers to the time at the location where the mission was
started.
The tab structure of the Mission viewer consists of two areas: Command (top), and device data tabs (bottom). The Command tab includes three buttons to operate
the logging function of the device, and to select the temperature scale (°C or °F). Depending on the size of the Mission tab, there may be scroll bars for both areas.
Page 7 of 21
Sample Mission Viewer Tab
Start New Mission (to allow the user to enter all data necessary to start a new mission)
This command opens a new window in which the user enters the parameters for the next mission. If passwords are enabled for the device, the user must first use
the Password viewer command Set Software Passwords, before access is granted to device data.
Automatically reads the device status, temperature log, and data log.
Automatically displays the device status on the Status tab.
Displays the temperature log as a graphic (if you open the Temperature Data Log tab).
Displays the humidity or data log as a graphic (if you open the Humidity Data Log or Data tab).
Allows the temperature scale to be changed between °C (default) and °F; the selected scale is memorized.
Allows activation of automatic rescaling of the log graphs.
Allows log data to be exported by using the clipboard or saving as a CSV (Excel) file.
Page 8 of 21
Sample Temperature Data Log Graphic (Humidity Data Log looks similar)
It is recommended that you synchronize the device's RTC to the PC. Enter the sampling rate in seconds (between samples). Enter the mission's Start Delay in
minutes. SUTA means Start mission Upon Temperature Alarm. Sampling rate, mission start delay, resolution, alarms, and rollover depend on the technical
requirements of the mission. See the respective device data sheets for details and guidelines.
Notes:
Mission Samples Count on the Status tab indicates the number of valid samples in the datalog memory. This number stops incrementing when the memory is
full.
Total Mission Samples indicates the number of samples taken during a mission. This number continues incrementing when a rollover occurs.
The Use 1-Second Mission Test? is not needed for B1 revision parts (date code 2007 and newer; serial numbers 005A900h to 005C09Fh, 0064200h and
higher).
Although the part is found as an alarming device, a Temperature High Alarm is not flagged on the Status tab. If there is a Temperature Low Alarm, both the
high and low alarm are flagged on the Status tab.
When a mission extends across time zones, the time shown in the exported data (clipboard or csv file) refers to the time at the location where the mission was
started.
Page 9 of 21
Password
The Password viewer allows the user to set, enable, and disable passwords on a 1-Wire device, and to input the device-specific read and write passwords into the
viewer.
The tab structure of the Password viewer consists of two areas: Info (top) and Config (bottom). The Config tab has three buttons to set the software passwords, set
the device passwords, and to enable/disable the password protection on the 1-Wire device. Depending on the size of the Password tab, there may be scroll bars for
the Info area.
Set Device Passwords (to write a read password and a full-access password to the device)
If the user wants to activate passwords with the device, the passwords must first be written to the device and then enabled to take effect. This command prompts
the user to specify/enter a read password in a separate small window. Once the read password is entered, the user is asked to enter the full-access password. Next
the passwords are written to the respective device registers. The viewer does not memorize the passwords after the session with a particular device is closed.
Passwords entered with this command are automatically taken as software passwords; there is no need to reenter these passwords as software passwords.
Passwords need to be entered as 8-byte hexadecimal strings.
Enable/Disable Passwords on Device (to enable or disable the use of passwords with the device)
This command allows the user to activate (enable) passwords that were written to the device previously. If passwords are enabled, this command allows the user to
disable them. This command does not erase passwords in the device.
Real-Time Temperature
The Real-Time Temperature viewer gives a real-time temperature reading of the selected 1-Wire temperature device connected to the OneWireViewer. The real-
time temperature log on the screen is created from temperature data stored in the PC's memory. This is different from Thermochron and Mission viewers, where the
temperature logging is performed inside an iButton, whether or not the device is connected to the OneWireViewer.
The tab structure of the Real-Time Temperature viewer consists of three areas: Info (top), Graph (bottom left), and Thermometer (bottom right). Depending on the
size of the Real-Time Temperature tab, there may be scroll bars for the Thermometer, Graph, and Info area. The Real-Time Temperature viewer applies to devices
that contain a digital thermometer. See the Supported Devices table above for the list of devices.
Page 10 of 21
Sample Real-Time Temperature Tab
When the Real-Time Temperature tab is active, the Real-Time Temperature viewer:
Continuously issues the device-specific command to perform a temperature conversion.
After each conversion reads the temperature registers of the selected device.
Converts the reading from the device-specific format into a conventional format.
Displays the result as a number, graph, and thermometer scale.
Compresses the horizontal scale of the graph, as needed, to display additional readings.
Allows the temperature scale to be changed between °C (default) and °F; the selected scale is memorized.
Allows the resolution of the temperature display (some devices only) to be changed.
Allows activation of automatic rescaling of the graph.
Allows graph data to be exported by using the clipboard or saving as a CSV (Excel) file.
Notes:
The Real-Time Temperature viewer closes whenever the selected device is found to be missing from the network or when another device is selected.
To keep the Real-Time Temperature viewer running while evaluating other devices, first activate the Real-Time Temperature viewer. Then on the View menu
(see Program Main Window), select Show Tab in New Window which will restart the Real-Time Temperature viewer in a new window. This separate window
will remain open, even if the device disappears from the network.
With the DS1922L, DS1922T, and DS1923, the temperature readings displayed are software corrected. For the DS1922E, software correction does not apply.
The Real-Time Temperature viewer cannot read the temperature of a Thermochron (DS1921, DS1922, or DS1923) that is currently running a mission.
Real-Time Humidity
The Real-Time Humidity viewer gives a real-time humidity reading of the selected 1-Wire humidity device connected to the OneWireViewer. The real-time humidity
log on the screen is created from humidity data stored in the PC's memory. This is different from Mission viewer, where the humidity logging is performed inside an
iButton whether or not it is connected to the OneWireViewer.
The tab of the Real-Time Humidity viewer consists of three areas: Info (top), Graph (bottom left), and Humidity (bottom right). Depending on the size of the
Humidity tab, there may be scroll bars for the graph and/or humidity area. The Real-Time Humidity viewer applies primarily to the DS1923. The viewer may also
support other devices that can be set up to measure humidity. See the Supported Devices table above for the list of devices.
Page 11 of 21
Sample Real-Time Humidity Tab
When the Real-Time Humidity tab is active, the Real-Time Humidity viewer:
Continuously issues the device-specific command to perform a humidity conversion.
After each conversion reads the humidity registers of the selected device.
Converts the reading from the device-specific format into a conventional format.
Displays the result as a number, graph, and humidity scale.
Compresses the horizontal scale of the graph, as needed, to display additional readings.
Allows activation of automatic rescaling of the graph.
Allows graph data to be exported by using the clipboard or saving as a CSV (Excel) file.
Notes:
The Real-Time Humidity viewer closes whenever the selected device is found to be missing from the network or when another device is selected.
To keep the Real-Time Humidity viewer running while evaluating other devices, first activate the Real-Time Humidity viewer. Then on the View menu (see
Program Main Window), select Show Tab in New Window which will restart the Real-Time Humidity viewer in a new window. This separate window will remain
open, even if the device disappears from the network.
To watch humidity and temperature simultaneously, open the Real-Time Humidity and Real-Time Temperature viewers in separate windows.
With the DS1923, the humidity readings displayed are software corrected; they are not compensated for temperature or saturation drift.
The Real-Time Humidity viewer cannot read the humidity of a DS1923 that is currently running a mission.
Switch
The Switch viewer gives real-time information on the I/O channels of the selected 1-Wire addressable switch device connected to the OneWireViewer. This
information includes the channel's state, the logic level at its output, and the state of the associated activity latch. The user can change (toggle) the state of the I/O
channel and clear its activity latch. Each I/O channel of a multichannel addressable switch device can be operated individually.
The tab of the Switch viewer consists of two areas: Features (top), and Channels (bottom). The Features area lists device-specific characteristics. The number of
switches (i.e., number of channels) for the selected device is indicated by the number of channel boxes in the channels area. A digital switch can be closed (State =
true) or open (State = false). Level shows the logic equivalent of the voltage at the switch output. Each channel has two command buttons associated with it.
Depending on the size of the Switch tab and/or the device, there may be scroll bars for both areas. The Switch viewer applies to all devices that contain one or
more electronic switches. See the Supported Devices table above for the list of devices.
Page 12 of 21
Sample Switch Tab
Notes:
To use a switch channel as a remote digital sensor, the switch must be off (State = false).
See the respective device data sheets for more details on how the individual channels operate.
A to D (Analog to Digital)
The A to D viewer gives a real-time voltage reading on the analog inputs of the selected 1-Wire ADC connected to the OneWireViewer. The user can select which
channels of a multichannel ADC are included in the polling.
The tab of the A to D viewer (for an analog-to-digital converter, ADC) consists of two areas: Features (top), and Channels (bottom). The Features area lists device-
specific characteristics. For each channel, there is a channel box in the Channels area. A to D Voltage shows the voltage reading of the analog input. Depending
on the size of the A to D tab and/or the number of channels, there may be scroll bars for both areas. The A to D viewer applies to devices that contain an ADC with
one or more input channels. See the Supported Devices table above for the list of devices.
Page 13 of 21
Sample A to D Tab
Notes:
If a channel is not included in the poll, the display for this channel will "freeze", and not show the actual input voltage.
The voltage range, resolution, voltage alarm thresholds, and alternate channel functions can be set on the byte level through the Memory tab. See the
respective device data sheet for more details.
Clock
The Clock viewer gives a real-time reading of the RTC in the selected 1-Wire device connected to the OneWireViewer. This viewer also tells the time difference
between the RTC in the 1-Wire device and in the PC. The user can synchronize the RTC in the 1-Wire device to the one in the PC or halt the RTC oscillator in the
1-Wire device.
The tab of the Clock viewer consists of two areas: Info (top), and Config (bottom). Depending on the size of the tab area, there may be scroll bars (horizontal,
vertical) for the Info area. The current time is always displayed in the 24-hour format with time zone. The Clock viewer applies to all devices that contain an RTC.
See the Supported Devices table above for the list of devices.
Page 14 of 21
When the Clock tab is active, the Clock viewer:
Continuously reads the time registers of the selected device.
Converts the reading from the device-specific format into a conventional format.
Displays the time read from the device (automatically updated with every reading).
Allows the RTC oscillator to be stopped and restarted.
Allows the time/date of the computer (PC or workstation) to be copied to the time registers in the device (i.e., click on Synchronize Clock to PC Time); this
function also restarts a halted RTC oscillator.
Notes:
If the 1-Wire device uses a binary counter as RTC (e. g., the DS1904, DS2415, DS2417), the Clock viewer sets the device clock to UTC (also known as
Greenwich Mean Time, GMT). If the RTC of the 1-Wire device counts seconds, hours, days, months, and years in separate registers (e.g., the DS1921,
DS1922), the Clock viewer sets the device clock to the local time.
Memory
The Memory viewer displays in hexadecimal format the data of the selected memory bank of the selected 1-Wire memory device connected to the OneWireViewer.
The user can change the data (see Hex Editor below) and write it back to the 1-Wire device.
The tab of the Memory viewer consists of three areas: Banks (top), Info (bottom left), and Contents (center right). Two buttons below the Contents area allow the
data of a selected memory bank to be reread (Refresh) and data that was manually entered in the contents area to be written to the 1-Wire device (Commit
Changes). Depending on the size of the tab area, there may be scroll bars for the Info and Contents area. The Memory viewer applies to all devices that contain
general-purpose user memory and/or special function/register pages. See the Supported Devices table above for the list of devices. A memory bank must first be
selected before its data (Contents) and characteristics (Info) will be displayed. The Contents area is automatically refreshed whenever the bank selection changes.
Page 15 of 21
Allows the contents of the selected bank to be changed.
Allows the new data to be copied to the selected bank.
Notes:
The number of the memory banks and their respective names depend on the type of the device viewed.
The display format in the content field varies with the device and memory bank selected. For most devices there are 2 x 16 bytes per page. With EPROM
devices, there are n x 8 bytes (special function registers only). With the DS1977 the format is 2 x 32 bytes per page.
If the selected memory bank is a scratchpad, the target address TA1, TA2, and the E/S byte (offset) are displayed above the scratchpad contents. With other
memory banks and depending on the device, one will see device-specific information associated with a memory page.
To write to the device, select the bytes to be changed in the Contents field, type new data, and click on Commit Changes. This method applies to all memory
blocks shown in the Memory tab. The viewer will perform the necessary steps (e.g., write scratchpad, copy scratchpad) to update the device memory or
register. See the section Auxiliary Functions text editor (hex/ASCII editor) for more details.
Writing to an EPROM device requires a port adapter that supports 12V programming pulses.
File
The File viewer displays the file names and directory structure of the selected 1-Wire memory device connected to the OneWireViewer. The user can format the
device, create/delete directories and files, edit files (see Hex Editor and ASCII Editor), resize files, and write them back to the 1-Wire device.
The File viewer supports the 1-Wire File Structure. (See application note 114, "1-Wire File Structure.") The tab of the File viewer consists of two areas: Directory
(center left), and File Contents (center right). In addition, there are eight command buttons. The vertical bar between Directory and File Contents can be moved
horizontally to change the size of the areas. Depending on the size of the file tab, there may be scroll bars for the Directory and File Contents area. The File viewer
applies to all devices that contain general-purpose user memory. See the Supported Devices table above for the list of devices.
Page 16 of 21
Sample File Tab (DS1922L)
Read File (to read and display the contents of the selected file)
The file contents can be viewed as data bytes or as text. Click on the Hex or Ascii tab to change the view.
Page 17 of 21
Sample File Tab with Directory and File Data Displayed
Notes:
The 1-Wire File Structure allows multiple devices to be formatted as one cluster. For this reason when formatting a device, the device needs to be selected
again, even if the File viewer already accesses it.
Before data can be written to a file, the file must first be created as a new file.
The file size is a multiple of 28 bytes (59 bytes with the DS1977). Changing the file size through Resize File automatically updates the file size in the display's
directory area. Changing the file size to 0 bytes is equivalent to first deleting and then creating the same file again.
A new file is filled with 00h bytes and has a size of 28 bytes (59 bytes with the DS1977).
To create a file inside a directory, first click on the directory and then use the Create New File function.
To change the contents of a file (hex as well as ASCII format), select the section (bytes/characters) to be changed, type the new contents, and click on Write
File. If the file is edited on the ASCII tab, there will be an automatic resize when the file is written back to the device. See Auxiliary Functions, and sections
Hex Editor with File Viewer and ASCII Editor for more details.
Writing to an EPROM device requires a port adapter that supports 12V programming pulses.
When accessing another device, the file contents area still shows data from the most recently accessed file.
To copy data from a disk file to the File viewer and vice versa, use the editing, copy, and paste functions of the operating system.
Auxiliary Functions
Hex Editor
The Memory viewer allows the user to enter data in hex format. The File viewer expects hex input when the Hex tab is selected. There are three ways to use the
hex editor:
1. Change: select one or more bytes and replace them with the same number of new bytes.
2. Delete: select one or more bytes and delete them.
3. Insert: position the cursor between bytes in the contents field and enter one or more bytes.
With special function registers or EPROM devices one should only change (overwrite) bytes, but not insert or delete bytes.
Changing Using the cursor, select a range of adjacent bytes in the editing area and type as many new bytes as are selected. Entering data overwrites the
data selected range and changes the background color of the field to yellow. When finished, click on Commit Changes.
Using the cursor, select one or more adjacent bytes in the editing area and press the "delete" key on the keyboard. This removes as many
bytes as are selected in the editing area and changes the background color of the affected field to yellow. When finished, click on Commit
Deleting
Changes. The viewer will ask for permission to pad the affected field(s) in the editing area with 00h bytes. This padding, however, does not
data
happen and data at the end of the affected field(s) remains unchanged. All other data in the field(s) is shifted to the left (lower address)
according to the number of bytes deleted.
Inserting Position the cursor in a field of the editing area and type the data to be inserted. This changes the background color of the field to yellow and
data shifts existing data to the right of the cursor out of the field, where it is lost. When finished, click on Commit Changes.
Verifying
Click the Refresh button and scroll to the updated field in the editing area.
data
Error Numbers from 0 to 9 and characters (upper/lower case) from A to F are accepted. Other characters can be entered in the editing area, but are
handling not written to the device. Spaces are optional between bytes; they are taken as delimiters.
Page 18 of 21
Hex Editor with File Viewer
Multiple rows of 16 bytes each in the Hex tab of the viewer; the last row can be shorter (partially editable). Each row is treated as an
Editing independent entity. More than one row can be updated before committing the changes. Use the Resize button to set the size of the editing
Area area, as needed. The description below assumes that the editing area contains data, which was obtained by selecting a file in the directory area
and then clicking the Read File button.
Changing Using the cursor, select a range of adjacent bytes in the editing area and type as many new bytes as are selected. Entering data overwrites the
data selected range and changes the background color of the row to yellow. When finished, click on Write File.
Using the cursor, select one or more adjacent bytes in the editing area and press the "delete" key on the keyboard. This removes as many
Deleting bytes as are selected in the editing area and changes the background color of the affected field to yellow. When finished, click on Write File.
data There is no padding at the end, and data deleted at the end remains unchanged. All other data in the row(s) is shifted to the left (lower
address) according to the number of bytes deleted.
Inserting Position the cursor in a row of the editing area and type the data to be inserted. This changes the background color of the field to yellow and
data shifts existing data to the right of the cursor out of the row, where it is lost. When finished, click on Write File.
Verifying
After writing, the file is automatically read back and its data is displayed.
data
Error Numbers from 0 to 9 and characters (upper/lower case) from A to F are accepted. Other characters can be entered in the editing area, but an
handling error message appears if there is an attempt to write the file to the device. Spaces are optional between bytes; they are taken as delimiters.
Changing Using the cursor, select a range of adjacent characters in the editing area and type as many new characters as are selected. Entering data
data overwrites the selected range. When finished, click Write File.
Deleting Using the cursor, select one or more adjacent characters in the editing area and press the "delete" key on the keyboard. This removes as many
data characters as are selected in the editing area. When finished, click on Write File. Instead of padding at the end, the file is automatically resized.
Inserting Position the cursor in the editing area and type the text to be inserted. No text is deleted. When finished, click on Write File. The file is
data automatically resized.
Verifying
After writing, the file is automatically read back and its data is displayed.
data
Error
All characters are accepted, including those that are not found on the English keyboard.
handling
DS9097U-009#
DS9097U-S09#
COM DS9097U Overdrive, power delivery, built-in serial number (009-suffix only)
DS1411-009#
DS1411-S09#
DS9097#
COM DS9097E* None
DS1413#
DS9490B#
USB DS9490 Overdrive, power delivery, built-in serial number
DS9490R#
USB DS9481R DS9481R-3C7+ Overdrive, power delivery, EPROM programming (12V, 7V)
*Not recommended for new designs.
Extended Features
Overdrive
The standard 1-Wire data rate is 15.3kbps. All iButtons and 1-Wire devices support this communication speed. Most 1-Wire devices also support overdrive speed,
which is typically 125kbps.
Power Delivery
Most iButtons and 1-Wire devices are parasite powered. This means that they derive their operating energy from the 1-Wire net during communication. Some
devices, however, require more energy than can be accumulated in this fashion. For these devices, choose an adapter with the power-delivery feature.
Page 19 of 21
EPROM Programming
iButton and 1-Wire devices that are one-time-programmable (OTP) incorporate technology called electrically programmable read only memory (EPROM). EPROM
programming requires 12V programming for writing. Although all adapters can read an EPROM iButton, only adapters with EPROM programming capability can write
to them.
NetAdapter
The NetAdapter is a virtualized 1-Wire adapter. Using TCP/IP protocol, it is possible for the OneWireViewer to communicate through a 1-Wire adapter attached to
another PC, as long as both PCs are connected through a TCP/IP network. Another requirement is that the other PC must be running a small host program, such
as the "StartNetAdapterHost" demo program found in the 1-Wire API for Java Software Development Kit. It is recommended that suitable software drivers for the 1-
Wire adapter be installed on the host PC, such as the 1-Wire Drivers. For a more in-depth discussion on this, please see application note 193, "Extending 1-Wire
Range with Network Proxies."
On the host PC where the 1-Wire adapter of interest resides, the StartNetAdapterHost Java program should be run from a command line. This is done by simply
copying the program and the OneWireAPI.jar file to a directory and typing the following example command-line parameters:
java -cp .;OneWireAPI.jar StartNetAdapterHost -adapterName {DS9490} -adapterPort USB1 -listenPort 6161 -secret secret
This sets up the host program to communicate to a DS9490 on USB1 and allows NetAdapter connections to it through the TCP/IP port of 6161 with the secret set to
"secret". Then, on the client OneWireViewer, simply set up the NetAdapter as shown in the screen shot below. In the "Adapter Name" field, type in "NetAdapter". In
the "Adapter Port" field, type in the IP address of the host computer, the TCP/IP socket port number, and the secret separated by colons.
Related Parts
More Information
For Technical Support: http://www.maximintegrated.com/support
For Samples: http://www.maximintegrated.com/samples
Other Questions and Comments: http://www.maximintegrated.com/contact