INSTALLATION MANUAL

TABLE OF CONTENTS
1. Before
---------------------------------------------------------- 3
installation
1.1 Safety
------------------------------------------------------- 3
indication
1.2 Additional
------------------------------------------------------ 4
indication
1.3 Safety indication for
----------------------------------------------- 4
installation
1.4 Safety indication for loading and
-------------------------------------- 5
access
1.5 Specifications of fuse and screw ------------------------------------------ 5
1.6 Fitting method and precautions ------------------------------- 6
1.6.1 Configurations of fitting ------------------------------------------- 6
1.6.2 Fitting method -------------------------------------- 6

1.6.3 Precautions of fitting ---------------------------- 8

1.6.4 Specifications of fitting and tube of TAURUS G2 ------------------------- 8

2. Instruments
required for ---------------------------------------------------- 9

installation
3.
---------------------------------------------------------------- 10
Installation
3.1 Product
------------------------------------------------------- 10
dimensions
3.2
------------------------------------------------------------- 11
Packaging
3.2.1 Transport and
--------------------------------------------------- 11
storage
3.3 Product installation --------------------------------------------------- 12
3.3.1 Moving chair ------------------------------------------------------ 12
3.3.2 Moving unit ----------------------------------------------------- 12
3.3.3 Chair/unit base fixation --------------------------------------- 12
3.3.4Secondary base --------------------------------------------- 12

1
installation
3.3.5 Chair and unit cover disassembling ----------------------------------- 14
3.3.6 Connecting frame and cover
----------------------------------- 14
assembling
3.3.7 Transport handle removal -------------------------------------------- 15
3.3.8 Chair and unit wire connection ----------------------------------- 15
3.3.9 Power connection and chair position move ------------------------------- 16

3.3.10 Spittoon assembly installation ------------------------------ 16

3.3.11 Assistant table installation ---------------------------------------- 17


3.3.12 Table arm
----------------------------------------------- 18
installation
3.3.13 Junction box
------------------------------------------------ 20
installation
3.3.14 Table installatioin --------------------------------------------------- 21
3.3.15 Table options
--------------------------------------------- 22
installation
3.3.16 Monitor arm/ light installatioin ------------------------------------- 23
3.3.17 Armrest installation ----------------------------------------------- 24
3.3.18 Seat
------------------------------------------------------ 24
installation
3.3.19 Healdrest installation ----------------------------------------------- 25
3.3.20 Backrest
------------------------------------------------ 25
installation
3.3.21 Chair and unit leveling -------------------------------- 25
3.3.22 Light arm/ table arm adjustment -------------------------------- 26
3.3.23 Plumbling
----------------------------------------------- 28
connection in the clinic
3.3.24
--------------------------------------------------------- 28
Inspection
3.3.25 Cover
------------------------------------------------------ 29
Assembly

4. Product specification ---------------------------------------------------


30
---------------

1. before installation

2
Make sure that the electricity outlet has no problem with grounding and
electricity leakage before connecting product to it.

※ grounding specification
Resistance: under 100 Ω/ for general low voltage electrical device

1.1 Safety indication


This installation manual includes icons in below list to guide users to use
the product appropriately preventing product damage and loss.

(1) Figure out the meaning of each icon prior to reading this manual.
(2) After reading keep this manual where users find easily.
(3) Be sure to follow all safety instructions.

indicates a maximum hazardous situation that can directly cause serious injury or
death.

indicates a hazardous situation that can lead to property damage or minor to


moderate injury.

indicates a situation related to safety and caution.

indicates users and service technician rules or instructions to follow.

indicates prohibited acts

1.2 Additional indication


Follow additional indications that displays important points of installation

3
work as follows.

indicates ‘advice’ for installation work

indicates ‘confirmation’ for installation work

1.3 Safety indication for installation


(1) Make sure both functional and protective grounding is clearly
connected before power connection after the installation work.
(2) Do the inspection for the installation in the presence of safety manager
of the clinic or hospital and keep the report at both sides.
(3) Consult the manufacturer or official supplier in case of moving the uint
after installation.

Install the unit only under the given frequency[㎐], voltage[V], allowable
currency [A] (or consumption)
A dedicated power circuit must be connected to each unit. Combination with
other devices, branches should be avoided.
▶ Otherwise it may cause accidents such as fire or shock or malfunction.

Keep the unit away from water. Do not install the unit in watery space
▶ Otherwise it may cause accidents such as fire from shortage or heating,
electricity shock or malfunction.
Do not install the unit in a space keeping chemicals or where there is gas.
▶ Otherwise it may cause accidents such as fire or electricity shock or
malfunction.

Do not install the unit in a space where its air pressure, temperature, humidity,
ventilation, sunlight, dust, or salinity may defective the unit.
▶ Otherwise it may cause defectiveness or malfunction.

Do not install the unit in a place where safety is not secured by slope,
vibration, or physical impact.
▶ Otherwise it may cause defectiveness or malfunction.

1.4 Safety indication for equipping and connection

4
Equipping and connection of all parts need to be made after the main power
plug is off except for the cases operation is required.

All plumbing and pipes should be firmly connected and arranged not to be
bent.
▶ Not enough connection may be a reason for air and water leakage.
Too strong bending of plumbing and pipes may block the air and water flow
or cause air or water leakage by fracture or deviation of joints.

1.5 Fuse and screw specifications.

1.5.1 Fuse specification


▶ MAIN : 250V/5A (diameter : Φ5 mm, length : 30mm)
▶ TRANS : 250V/5A (diameter : Φ5 mm, length : 30mm

Fuse location

1.5.2 clamping bolt specification

Applicabl Applicable
Classification Morphology
e hole material

Standard Aluminum die


TAPTITE SCREW
hole casting

Tapped Tapped
METER SCREW
hole material

Standard metals
TAPPING SCREW,TAP2
hole plastics

※ As the thread pitches of taptite screw and meter screw are the same, mter
screw can replace taptite screw once it’s removed from its hole.

1.6 FITTING working methods and precautions

5
1.6.1 FITTING configuration
Body
Cap

O-Ring

Collet

1.6.2 FITTING work procedure

▶ tube insert
• Cut the end of a tube perpendicularly before insert work.

• Outer surface of the inserted tube should be clean without any scratch.

• When scissors, instead of tube cutter, are used for the cutting, make the
buckled hole round after cutting. Otherwise it may cause leakages

• Use screws for fitting after removing any debris on it. Otherwise the debris may
cause scratches on the screws and eventually leakages and breakage.
• For the fastening of fittings use o-rings if it’s ready with product. If not, use
Teflon tape and proper instruments.

O-ring

6
• Insert the tube to the stop position. Ignore slight jams in the middle and go the
end position.
.
• Non-completely inserted tubes may break away under high pressure.

Stop position

• In order to ensure full fastening, try pulling out the tube with proper strength.

• The teeth of collets prevent falling out of the tubes. O-rings prevent leakage.

• After tube insert, insert locking clip in the gap between the collet and the body.

• Locking clip prevents tubes from falling out and so leakage problem.

• Make sure no leakage is found after tube connection.

▶ Disconnection of tube
• Pull out the tube pushing the collet toward body as much as possible.
• Reuse of fitting is prohibited.

• Tubes can be reused under the condition that the end of the tubes are round as
circle and have no scratch.

7
1.6.3 Precautions of fitting work
(1) Do not make severe bumps on, rotation, pulling, twisting, and impact on the u
body. Otherwise it may cause the breakage of body and falling out of collets.

(2) Use of high-temperature fluids can harm the fitting body creating the
hydrolysis.

(3) Use only FDA certified tubes for water line.

(4) Check and make sure that the cross section of the tube end is cut
perpendicular, that no damage is found on the tube surface, and that the tube is
thrust to the very end.

(5) Make sure the pressure is ‘0’ when the tube is removed from the tubing.

Removing of tubes with excessive force under pressure may damage the fitting
body and tube.

(6) Tightening the fittings, use proper tools for the screws.

Excessive torque may damage the screw parts, and less than proper torque will
cause loosening and leakage

(7) Use band for the connection part when barb (or nipple) is used.

(8) Do not mix water and air line tubes, an activity creating leakage.

1.6.4 TAURUS G2 fitting and tube specification.

Classification color manufacturer Measures Material Others

POM Tube for


Fitting Gray DMT
(acetal) warmer
Water Inch Poly connection
Tube Blue DMT Ethylene is made of
nylon.
Fitting Black Sang-A PBT
Yellow Poly
Air meter
Tube Green Nittamoor Urethane
Translucence Poly Olefin

※ Do not mix water and air line tubes, an activity creating leakage.

8
2. Tools required for installation

+, - drivers Spanners

Monkey spanner Long nose

Nipper Wrenches

Box wrench Box

TUBE CUTTER

9
3. Installation
3.1 Product dimensions

10
3.2 Packaging
Classificatioin Main box Ⅰ Main box Ⅱ Option box
□ UNIT ASS’Y □ CHAIR ASS’Y □ FORM SEAT
□ LIGHT ASS’Y □ DOCTOR TABLE □ FORM BACKREST
□ TABLE ARM ASS’Y ASS’Y □ HEADREST ASS’Y
□ ASSISTANT TABLE □ G-PEDAL ASS’Y □ BACKREST COVER
ASS’Y □ ARMREST (left- □ FOOTREST COVER
Component □ SPITTON ASS’Y fixed) □ STOOL BOX
□ JUNCTION BOX □ STOOL BASE BOX
ASS’Y □ ARMREST (Right-
revolving)
□ LCD ARM
□ ACCESSORIES
Size 570×1,495×1,135 570×1,495×1,1135 570×940×800
(mm) (L×D×H) (L×D×H) (L×D×H))
Weight 129 kgf 105 kgf 59 kgf

3.2.1 Transport and storage

NOTE Transport and storage condition must be maintained during repair and
service works.
Below labeling instructions are attached on the exterior boxes.
Obey following labeling instructions.
The direction of the arrow facing upward, placed upright

Fragile products. Protection against external shocks

Stay dry

The maximum allowable loading capacity

n Maximum loading singular

Allowable temperature range

Range of air humidity

Allowable atmospheric pressure

▶ storage and transport conditions


1) Ambient temperature : -40℃ ~ 70℃
2) Relative humidity: 10-95%: 10% ~ 95%
3) Atmospheric pressure : 50㎪ ~ 106㎪

11
3.3 Product installation
3.3.1 Moving chair
▶ Disconnect chair Ass’y from the pallet
▶ Move holding both bottom ends of chair base(2 people required) .
3.3.2 Moving unit
▶ Disconnedt unit ass’y from the pallet.
▶ Fix transport handles(4pcs) using hex head screw (M16×L40, 8pcs)
▶ Move two people holding transport handles.

3.3.3 Chair/unit base fixation


▶ Using Φ12 drill, make 4 holes on the floor for chair base and
Unit and fix them with anchor bolts.

The chair base of this product must be fixed onto the floor with proper
method.
▶ After the completion of the connection of all parts, the chair base must be
fixed to the floor.

3.3.4 Secondary base installation (option)


The installation of secondary base is recommended for poor floor
condition and for a situation when light trembling is severe
▶ Install a secondary base-A to positon.
▶ After the assembly of enclosed bolts(M14xL20, 8pcs) to their location, adjust
the level horizontally. .

Secondary base-A

Horizontal adjustment is even more difficult after the unit installation is


completed..

12
▶ Unit and Chair Ass’y fixation precedure

◎ Fix the unit with screws(hex head M14xL25, 4pcs)


◎ Insert connecting frame between unit and chair and fix it in the hole inside
of the chair base with screw bolts(hex head M14xL25, 2pcs).
Adjust the space between the unit and chair (divided by 20mm with 4 holes)

[ unit fixatioin ] [ connecting frame assembling]

◎ Fix the connecting frame to chair frame(refer to 3.3.6 14th page)


◎ Adjust the horizontal level using 4 screws on the chair base.

[connecting frame fixation]


[horizontal adjustment ]

▶ Put secondary base-B behind the unit, adjust the horizontal level with(M14xL20,
3pcs) and install junction box ass’y.

Junction box ass’y


Level adjustment
(3)
Secondary base-B

[Junction box ass’y installation]


[secondary base-B installation ]

13
3.3.5 Chair and unit cover disassembling
▶ Move base cover from chair front to its left side pushing both sides of it.
※ As main power switch is connected to the base cover, do not strongly pull it
back.
▶ Dismantle ‘Chair base bracket’ (M4×L10, 4pcs)
▶ Open the unit cover turning the locker on upper part of it.

Base cover

Chair base bracket


Chair base

3.3.6 Connecting frame and cover assembling


▶ Place the chair on the installation site, push connecting frame into chair base,
and insert wires from unit into a hole of chair base.
▶ Assemble connecting frame cover
▶ Check the space between the chair and unit. Tighten three screws(M10×L20) on
it with (M10×L20)에 spring washer(M10), flat washer 2pcs (M10, 1.2mm) each.

Connecting frame cover

Connecting frame

20mm

▶ The distance between the chair and unit is adjustable(max. 20mm) to the
request of users.

▶ Adjusting the distance between the chair and unit, make sure the backrest
form and assistant table(rotating by 45degrees) do not meet each other. Then
firmly fix connecting frame
▶ If the backrest form and assistant table meet each other, lengthen the
distance. Otherwise it may damage the backrest form.

14
3.3.7 Transport handle removal
▶ Remove transport handles attached to the unit loosening hex screws(M16×L40,
8pcs)

▶ Fasten wrench bolts(M16×L14) in the place where transport handles are

removed.

3.3.8 Chair and unit wire connection


▶ Communication cable, homing connector connection
① Connect CN10 connect of the unit and CN10 of the MPU PCB of the chair.
② Connect homing connector to the MPU PCB of the chair.
Communication cable
(CN10)

HOMMING connector (CN12, Blue)

▶ Power cable connection


Connect pins to the terminal block on chair base.
Refer to the attached circuit diagram.

▶ Ground wire connection


Connect the ground wire of the unit to the ground block located at the front of
the chair base.

15
3.3.9 Power connection and chair position move
▶ Plug the power cord into the outlet
▶ Turn on the main power switch on the chair base cover
▶ Lift chair to its highest end and lower the backrest to its lowest.
▶ Turn off the main power switch and pull out the plug from the outlet.

Make sure the unit is unplugged. Otherwise danger of short circuit remains
during assembly

3.3.10 Spittoon assembly installation


▶ Place the cover shaft on the unit.

▶ Pass wires and tubing from the spittoon through unit top cover. .

◎ wires: hole on the right ◎ water tube: hole on the left.


[Spittoon Ass’y insert angle]
▶ Insert Spittoon assembly into the axis in the unit with 45
degree angle, and push it down till the connector shaft meets the opposite end.

▶ Connect the Auto-cup connector attached to the spittoon assembly with the
connector attached to the unit.

▶ Connect the spittoon rinsing tube(short) attached to spittoon and Auto-cup


tubing(long) to its solenoid valves respectively and insert locking clips.

▶ After the assembling check the direction of rotation and whether the micro
switch works well.

Spittoon Ass’y

Autocup tubing
Spittoon tubing
Connector Shaft

Left hole
Unit top cover

Right hole

16
3.3.11 Assistant table installation
▶ Insert the data cable and ground wire connected to the assistant table into the
connector AT hole.
▶ Connect the table arm of assistant table with the connector AT.

▶ Tighten two screws(M4×L8) on the arm.

connector
AT

▶ Communication cable connection


① Connect the communication cable CN11(blue-5PIN) and CN12(blue-3PIN) of
assistant table to the UNIT BASE BOARD PCB CN11 & CN12 of unit.

▶ Ground wire connection


Connect the ground wire of assistant table to ground hole with tapping
screw(PH, Φ4×L8, TAP2).

① CN11(5PIN)
Ground wire
① CN12(3PIN)

UNIT BASE BOARD PCB

▶ Connect suction tubing and 3-WAY syringe tubing to the unit and mount them
on the assistant table.

17
3.3.12 Table arm installation

▶ Table arm installation must be done after unit is fully fixed.


Otherwise the unit may tremble.

▶ Secure with tape the end of wires and tubes of table arm so that it is easily
inserted into the pipe.
▶ Apply grease to the outer diameter of the table arm pole fixed on the unit.
▶ Disassemble the stopper behind the table 1st arm.
▶ Insert arm cap on the pole.
▶ Insert table 1st arm while pushing wires and tubes into the pole.

1st arm stopper

Arm cap

③table stopper

▶ Stopper fixation
① Fix the table 1st arm stopper(truss wrench bolt, M6×L12) behind the arm.
② Remove the 1st arm cap and fix a stopper(wrench bolt, M5×L10).
③ Remove table 2nd arm cover and fix a stopper(wrench bolt, M4×L8)
④ Assemble table 2nd arm cover.

▶ The stopper for table 1st arm is only for preventing the table from hitting
the wall. So when this function is not required, this stopper is not necessary to
be fixed.

▶ Power cable connection


Connect the connecter of table to the connector at the bottom of unit.
Tie the wires with a clamp.
▶ Ground wire connection
Connect table ground wire fastening a tapping screw(PH, Φ4×L8, TAP2) to the
ground hole.

① connecter
② ground wire

18
▶ Communication cable connection
Connect the CN13(노랑) connecter of table to BASE BOARD PCB CN13.

` CN13(yellow)

UNIT BASE BOARD PCB

▶ AIR / WATER LINE Connection


① Connect signal air(green) to Y-fitting.
② Connect drive air(yellow) to servo motor L-fitting after the installation of
junction box.
③ Connect water line(blue) to T-fitting at the assistant table syringe tube and
insert locking clip.

▶ Once all connection is done, lift arm cap and adjust the horizontal level of table
arm fastening properly the four wrench bolts around the pole.

19
3.3.13 Junction box installation

▶ Dismantle junction box cover removing tapping screw(TH ø4×L8, 4pcs) at the
sides.
▶ Install the junction box frame to the unit base with two tapping screws(PH
ø4×L8).

Tapping screws

▶ Connect the water line(blue) and signal air(green) from unit and drive
air(yellow) from the table to the Junction box as instructed in below diagram.

▶ Fix the servomotor connector-A and Solenoid connecter-A from the unit.

Servo motor connecter-A

Solenoid valve connecter-A

20
3.3.14 Table installation
▶ Withdraw three electric wires, three tubes through the hole on table base.
▶ Place the table on the table arm.
▶ Fasten wrench bolts(M8×L10-Short, 2pcs) at the rear, wrench bolts(M8×L35-
long, 2pcs) at the front of the table.
▶ When the table is not parallel to the ground, adjust it handling adjustment
bolts(4pcs) and fix it with fixation bolts.

Adjustment bolt (4pcs) Fixation bolt (M8×L10-short)

Fixation bolt (M8×L35-long)

▶ Power cable connection


Connect the block pin on the table terminal strip

▶ Connecting the table terminal block pins, match the their colors
▶ Refer to the wiring diagram attached to the bracket assembled inside the
table.
▶ Communication cable connection
Connect the data communication cable connecter to the PCB
▶ Ground wire connection: Fix the table ground wire to the ground hole.
▶ Insert USB terminal to the back of the table, cover with a bracket and fasten
screws. (Loosen the readily fastened screws)
▶ Connect the drive air(yellow), signal air(green) and water line(blue) from table
arm to the relevant fittings. Ground wire

Communication cable
USB bracket

21
3.3.15 Table options installation
▶ Mouse tray
Remove the tray of dentist table. Attach a mouse tray with tapping
screws(Ф4×L12 2pcs top, 1pc down, total 3pcs) at the left or right side of the
table as per customers’ request.
▶ Assistant tray
Insert the nipple of assistant table holder into the hole at the left side of dentist
table. Fasten taptite screws(M4×L12, 4pcs).

Hole for nipple

Assistant tray holder

Assistant tray

Mouse tray

Using screws longer than 12mm for the installation of assistant tray holder to
the bottom of dentist table may damage PC boards in the table

▶ Waste bottle installation


Loosen wrench bolt (M4×L10) for valve block at the right bottom of the dentist
table. Assemble the waste bottle bracket with the valve block tightening the
loosened bolt.

Waste bottle

22
3.3.16 Monitor arm/ light installation
Ground Power cable
▶ Tape wires of vertical light arm.
wire
connecter
▶ Apply grease on the bore of vertical light arm. connecter

▶ Put arm cap on the vertical arm, insert wires,


and insert vertical arm pole
▶ Fix with screws (M5×L8, 2pcs) with washers.
▶ Connect the power cable and ground wire of
light first arm to the connecters of vertical arm.
Assemble the light first arm pulling out the wires
from the unit slowly while making sure the wires
are not caught by the shaft.
▶ Uncover the first arm cap and fix a stopper (wrench bolt M5×L10).

▶ Make sure to assemble light first arm pulling out the wires from the unit.
Otherwise it may damage the wire.

▶ Assemble light power cable and ground wire


to the unit.
Horizontal control

▶ Lift arm cap and adjust the horizontal


position of the vertical light pole cont
Ground
rolling wrench bolts(4 pcs),

Power cable connecter


▶Place wires for monitor{HDMI terminal cable,
, power cable, video cable(yellow jack), audio
cable(black jack), antenna} to the rear of unit
and connect HDMI terminal to the junction box.

▶ In case the wires for monitor are not used,


cover the wires with a wrapper enclosed in the
package and place them at the rear of the unit.

▶ In case the wires for monitor are required to be connecter

connected to a computer, push them through the


duct hole of junction box.
▶ In case DVI terminal is used, get a DVI-HDMI gender
LIGHT
from the market.
BOARD

23
3.3.17 Armrest installation
▶ Place an armrest at the top left side of seat frame.
▶ Fix with wrench bolts(M5ⅹL16, 3pcs).

armrest (movable,)
armrest (fixed)

Nut

Seat frame part cut


Arm cap

▶ Adjustment of the right armrest tightness.


① Uncover the arm cap using a pin.

② Adjust the tightness turning the nut inside.

- Note that if the nut is too loose, the right arm may shake severely and
sounds loud while moving.

▶ To retrofit a right armrest, cut the seat frame part with a nipper and grind
the cut part with a file or knife.

3.3.18 Seat installation


▶ Turn the main power switch on, lift the seat to the highest, lay the backrest
down to the lowest, and turn the main power switch off.
▶ Hang the hooks of seat foam on the grooves of chair frame.
▶ Assemble footrest using tapping screws(Ф4 × L16, 5pcs).

▶ Assemble footrest cover using tapping screws(Ф4 × L16, 4pcs).

(Do not use big-head screws)

Seat foam
Foot rest

Foot rest cover

24
3.3.19 Headrest installation
Sliding brake
▶ Disassemble slide stopper from the backrest
And insert headrest assembly. Slide stopper
▶ Fix slide stopper (wrench bolt M4×5)..
▶ Adjust the height of sliding brake for the
headrest can move easily.

3.3.20 Backrest installation


▶ Assemble backrest cover with tapping screws(TH, Ф4×L12, 6pcs)
▶ Place the backrest foam on the cover.
▶ Assemble with tapping screws(stainless Ф4×L10, 4pcs).
▶ Finish with rubber plugs(cap backrest-A/B, 2pcs respectively)..

▶ Note that it’s important not to drop or damage with tools the backrest foam.
Care for especially its edges.

Backrest cover Tapping screw Backrest foam

Rubber plug

3.3.21 Chair and unit leveling

If the position of chair and unit is not horizontally balanced,


- Light, dentist table may tremble..
- Unnecessary noise and vibration during operation occur.
- Product may be damaged and the lifespan of the product may be also
damaged.
▶ Loosen the wrench bolts with L-wrench as long it doesn’t meet the floor.
▶ Control the balance of the base of chair and unit fastening the bolts.
If you fasten only one side, it may make a unnecessary gaps.
▶ Finish with rubber plugs at the rear of chair base.

▶ Check again the horizontal balance of table arm and light arm.
(refer to “3.2.13 table arm installation”, “3.2.15 table installation”, and “3.2.17
Monitor arm / light arm installation”)

25
3.3.22 Light arm/ table arm adjustment

▶ In case the dentist table moves by itself to the left or right


① Remove the table 2nd arm cover sliding it out.

② Loosen nut(M4), but do not remove it completely.

③ See the table movement fastening wrench bolts(3pcs).


④ Fasten the nut completely and assemble the 2nd arm cover.

Table 2nd arm

▶ In case the table 2nd arm moves by itself to the left or right

① Remove the cap of table 2nd arm.

② Unblock the wrench bolt(M4X6), loosening a nut(M4) a little.


③ See the movement fastening wrench bolts(3pcs).

④ When the table 2nd stops self-movement, fasten the nut and reassemble the

cap.

Nut & wrench bolt

cap

26
▶ In case the light 2nd arm goes up or down by itself.
① Remove the light 2nd arm cover, loosening 2 screws on top of it.
② If the 2nd arm goes up, turn the adjustment nut clockwise, and if it goes
down, counterclockwise.
③ Reassemble the cover and fasten the screws.
Screws

Cover
Adjustment
nut

▶ In case the light 2nd arm moves by itself to the left or right
① Remove the cap of the light 1st arm.
② Loosen nut(M4), but do not remove it completely.

③ See the movement fastening wrench bolts(3pcs).


④ When the light 2nd stops self-movement, fasten the nut and reassemble the

Nut & wrench bolt

cap

▶ In case the light 1st arm moves by itself to the left or right.
① Fasten the wrench bolt on the light pole.

Light 1st arm

Wrench bolt

27
3.3.23 Plumbing connection in the clinic
▶ Connect the air, water, and suction pipes installed in the clinic to the junction
box.
▶ Finish with enclosed duct hose.

▶ Make sure to check the leakage of air, water, and suction pipes.

3.3.24 Inspection

▶ Plug the power code into the outlet and turn on the power switch at chair base.

▶ Open the main air and water valves in the junction box.

▶ Check the air and water pressure and adjust with air and water regulators.
(Recommended pressure: water – 0.18~0.2 ㎫, air – 0.4 ㎫ )
※ In case a distilled water tank or sanitation bottle is installed, make sure the
regulator of the unit is set at 0.15 ㎫.

▶ Check whether the inbound water is tap water(City) or distilled water and set
the toggle valve at the front of the unit accordingly.

▶ Check the condition of the handpieces, suction, scaler, and syringes.

▶ Check the water supply to auto-cup and spittoon rinsing tube.


In the first inspection air in the pipes makes water splash. Watch it out.

▶ Check the air or water leakage in junction box, unit and table and connection
of each fitting with the naked eye.

▶ Check the functions of all touch keys on both dentist and assistant panels.

▶ Make sure the assistant table do not hit the wall and the backrest.

▶ Make sure the light and table arms do not tremble and check the balance of
the dentist table.

▶ Using option mode 1 and 2, check the settings of equipment. Set the duration
of auto-cup and spittoon rinsing.
Adjust the chair movement speed, if necessary. Note that the faster the chair
moves the bigger its noise is.

28
3.3.25 Cover Assembly
▶ Assemble the chair base bracket with taptite screws(M4ⅹL8).
▶ Assemble the chair base cover.

Chair base bracket Chair base cover

▶ Assemble left and right unit service covers.


▶ Assemble unit front door. (Make sure the hinge is not bent too much.)
If the door and unit base meets each other, loosen the screws a little of the
bracket that holds the hinge and lift up the bracket, then fasten the screws.
▶ Assemble the cover of junction box with tapping screws(stainless steel TH
ø4ⅹL8)

Service cover

Junction box cover

Front door

▶ Assemble the protection cover with taptite screws(M4ⅹL8, 2pcs).


▶ Assemble the table cover with screws(M4ⅹL8, 4pcs).
▶ Cover with the silicon pad.

Silicon pad

Table cover

Protection bracket

29
4. Product Specification

1) Product specification
Input voltage and frequency AC 220V, 50/60HZ
Power consumption 1,000VA
Power supply classification Class 1
Electric protection BF type
Ground Class 3
Inlet water pressure 0.2 ~ 0.59MPa
Water inflow 5 ℓ/min or more.
Inlet air pressure 0.39 ~ 0.78MPa
Slop of drainage pipe minimum 10mm/M
Diameter of drainage pipe 40mm
Weight (Net) 200±5 kgf
Chair elevation range Min 380±20mm , Max 780±20mm
Spacing range between chair and unit 20mm
Chair operation cycle ON : 25sec , OFF : 400sec
EMC standard IEC60601-1-2:2007,
EN61000-3-2:2006+A2:2009
EN61000-3-3:2008

2) Working environment
Slop of installation surface 0˚ ~ 5˚
Ambient temperature 10 ~ 40℃
Relative humidity 30 ~ 75%
Air pressure 70 ~ 106㎪
Water temperature under 25℃
Allowable weight on the chair 150 kgf

3) Required installation space


Width 1,800mm and more
Length 2,400mm and more
height 2,000mm and more

30