Indir
Indir
Indir
IQ
Panametrics Oxygen Transmitter
Applications Features
An oxygen transmitter for use in: • Two-wire, loop-powered 4 to 20 mA transmitter
• Glove box purge and leak detection • Display with keypad
• Natural gas • Intrinsically safe option for Class 1 Div 1 and Div 2 locations,
and ATEX/IECEX Zone 0
• Semiconductor wafer machines
• Non Incendive for Div 2 locations
• Coating process machines
• Ex flameproof option
• Membrane air separators
• Proven galvanic fuel cell O2 sensor technology
• Inert welding gases
• User selectable ranges for ppm and percent oxygen
• Pure gaseous hydrocarbon streams
• User-friendly and intuitive user interface with diagnostics
• Process monitoring of gaseous monomers
• Microprocessor-based, all-digital technology for reliable
• Heat treating and bright annealing
operation
• Low maintenance, economical and compact
• Sensor failure output error
• Sensor lifetime indication
• NAMUR error indication
Panametrics oxygen transmitter Control at the tip of a finger
The oxy.IQ is a highly reliable and cost-effective two-wire, The rugged oxy.IQ is controlled by a microprocessor that
loop-powered transmitter with a linearized 4 to 20 mA enables the user to select the range, trim outputs and
output. It measures oxygen in ten ppm ranges and seven perform calibration. Improved user interface includes user-
percentage ranges. All ranges are user-selectable. This selectable and programmable ranges, sensor diagnostics
compact transmitter uses proven sensor technology including a sensor lifetime feature and sensor raw signal,
to accurately measure O2 in a variety of gases, even in low reading/sensor failure signal and temperature
hazardous environments. compensation. Programming is easily done using the
keypad and display. These keypad buttons allow complete
Proven sensor technology functionality in hazardous areas when equipped with a zener
barrier or a galvanic isolator per drawing 752-347.
The oxy.IQ’s oxygen sensor is an advanced galvanic fuel cell
that provides superior performance, accuracy, stability and
long life. The cell’s innovative design eliminates the potential
Installation flexibility
for negative signal output, reduces sources of contamination. The compact oxy.IQ, with its built-in microprocessor, is
designed to fit easily into any installation site. The oxy.IQ
The cell is unaffected by other background gases or
can be installed right at the sampling point, whereas other
hydrocarbons and is compatible with acid gases(OX-2 and
transmitters must be rack or panel mounted.
OX-4 cells). Recovery from air at low ppm levels takes just
a few minutes. Because the cell is self-contained, minimal
maintenance is required. There is no electrolyte to change or Sample systems
electrodes to clean. In addition to standard features and options, Panametrics
offers a full line of sample handling systems for a variety of
Intrinsically safe applications. If needed, Panametrics can design and build
a sample conditioning system to meet unique application
When equipped with optional zener barriers or a galvanic
requirements. Please contact Panametrics for details.
isolator, the oxy.IQ can be mounted in a hazardous
(classified) location.
Oxygen sensor interference gases
OX-1 and
OX-2,
Gas OX-5, OX-3, % OX-4, %
ppm.
ppm
SS process unit: 316 stainless steel, Viton® O-ring, gold plated Weight
sensor electrical contacts and glass (Chemraz O-Ring Option) 1.35 lb (612 grams)
User-selectable measurement ranges Sample flow rate
• PPM sensors:
1.0 SCFH (500 cc/min) recommended for process units
– 0 to 10 ppmv O2 (OX-1 or OX-2 only)
– 0 to 20 ppmv O2 (OX-1 or OX-2 only) Electrical classification/certification
– 0 to 50 ppmv O2 (OX-1 or OX-2 only) Intrinsically safe
– 0 to 100 ppmv O2 USA/Canada
– 0 to 200 ppmv O2 • IS for Class I, Div 1, Groups ABCD, T4
– 0 to 500 ppmv O2
ATEX and IECEx
– 0 to 1000 ppmv O2
• Ex ia IIC Ga T4 Tamb -20 to 60°C
– 0 to 2000 ppmv O2
– 0 to 5000 ppmv O2 IS package, non-incendive without use of Zener barrier or
– 0 to 10,000 ppmv O2 galvanic isolator:
• USA/Canada: Class 1, Div 2, Groups A, B, C, D; T4
• Percent sensors:
Flame proof package
– 0% to 1% O2
• ATEX/IECEX Ex d IIC T4 IP66
– 0% to 2% O2
– 0% to 5% O2 European compliance
– 0% to 10% O2 Complies with EMC directive 2014/30/EU
– 0% to 25% O2
– 0% to 50% O2
Dimensions
A - Model only
oxy.IQ Oxygen Transmitter; 4 to 20 mA output
B - Sensor
0 No sensor
1 Standard ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
2 Acid ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
3 Standard percent sensor
4 Acid percent sensor
5 Standard ppm, 0 to 100, 200, 500 and 1000 ppm
C - Package
1 Standard
3 Intrinsically safe for ATEX/IECEX and US/CAN Class 1 Div1 or Non-incendive for US/CAN Class 1 Div 2
4 EX&IP66 ATEX/IECEx 6mm fittings in aluminium enclosure
5 EX&IP66 ATEX/IECEx 1/4” fittings in aluminium enclosure
6 EX&IP66 ATEX/IECEx 6mm fittings in 316SS enclosure
7 EX&IP66 ATEX/IECEx 316SS 1/4”fittings in 316SS enclosure
D - Cable length
0 No cable
1 2 meter cable
2 10 meter cable
3 Demo cable
E - Zener barrier
0 None
1 Zener barrier
2 Galvanic isolator
F-Base adapter
0 Standard
1 O2X1 Adapter
2 TD Adapter
3 ISO KF/NW40 Adapter
4 Ambient Air Adapter
S- Special
0 No special selected
1 CHEMRAZ o-ring
2 Special consult factory
Note : For Option C=3 intrinsic safety, either zener barrier or galvanic isolator
must be selected. For Class 1 Div 2, no barriers needed. Please refer to dwg
752-347 installation guidelines.
__ - __ ___ ___ ___ ___ ___ [Use this number to order product]
Panametrics, a Baker Hughes Business, provides solutions in
the toughest applications and environments for moisture,
oxygen, liquid and gas flow measurement. Experts in flare
management, Panametrics technology also reduces flare
emissions and optimizes performance.
With a reach that extends across the globe, Panametrics’
critical measurement solutions and flare emissions
management are enabling customers to drive efficiency
and achieve carbon reduction targets across critical
industries including: Oil & Gas; Energy; Healthcare; Water and
Wastewater; Chemical Processing; Food & Beverage and
many others.
Join the conversation and follow us on LinkedIN
linkedin.com/company/panametricscompany