Operator’s Manual
AGH6200/AGH6200F
Version:3.3.1901
1
AIYI Technologies
Dear customer:
Yours sincerely.
1
AIYI Technologies
Copyright Statement
Disclaimer
This manual, as well as the software described in it, is furnished under license and may be used or
copied only in accordance with the terms of such license. The content of this manual is furnished
for informational use only, is subject to change without notice, and should not be construed as a
commitment by AIYI Technologies. (AIYI Technologies ) AIYI Technologies assumes no responsibility
or liability for any errors or inaccuracies that may appear in the manual.
Except as permitted by such license, no part of this publication may be reproduced, stored in a
retrieval system, or transmitted, in any form or by any means, electronic, mechanical, recording, or
otherwise, without the prior written permission of AIYI Technologies.
2
AIYI Technologies
Statement
The manual is applicable to the use and maintenance of AGH6200 & AGH6200F
series gas detector.
3
AIYI Technologies
CONTENT
1.4 Specifications………………………………………………………………09
Chapter 2: Operation
Chapter 3: Maintenance
4
AIYI Technologies
▎Various alarm modes of sound and light and vibration, effectively alert.
▎Standard Micro USB interface for easy charging and data export.
▎GB 50493-2009 Specification for design of combustible gas and toxic gas de-
tection and alarm for petrochemical industry
▎GB 12358-2006 Gas monitors and alarms for workplace General technical
requirements
▎GB 16838-2005 Envionmental test methods and severities for fire electronic
products
▎GBZ 2.1-2007 Occupational exposure limits for hazardous agents in the work-
place Part1:Chemical hazardous agents
6
AIYI Technologies
7
AIYI Technologies
1 Task bar
Zero calibration
Span calibration
Self-test
USB connection
Power
High alarm
%LEL lower explosive limit
5 Time bar
% Percentage
8
AIYI Technologies
1.4 Specifications
9
AIYI Technologies
Chapter 2 Operation
2.1 Packing list
Please confirm if the package is complete before use. If there is any loss, please con-
tact us as soon as possible. Normally, the gas detector contains the following products
and accessories:
Quan Quan Quan
Item name Item name Item name
tity tity tity
Calibration certificate &
Detector 1 1 Calibration adapter 1
QC certificate
Charger(including USB
Manual 1 1
Cable)
10
AIYI Technologies
▎Power off: Press and hold the power button to enter the shutdown interface.
All the indicator lights are turned off, the detector enters the shutdown countdown,
and the display is turned off from OFF 004 to OFF 000. After the countdown ends, the
display is turned off, and the Green indicator keep flashing for about 10 seconds,
then the instrument is turned off.
Shutdown interface
▎Factory reset: After the boot process is displayed, the detector enters the facto-
ry reset interface. If you need to restore the factory settings, press the function
button twice in rapid succession, otherwise the detector will automatically enter the
main interface.
▎Fault indication: When the gas sensor is shorted or broken, the corresponding
detection channel has a fault indication, as shown in the figure below. At this point,
the green indicator is off and the yellow fault indicator is keeps on.
11
AIYI Technologies
2.4 Operation
▎Menu:
1.Press and hold the power button to enter the shutdown interface,as shown in
the figure below.
2.During the shutdown countdown process, press the power and function button
at the same time to enter the menu. Menu is shown in the figure below.
3.Short press the function button to select the zero calibration, span calibration
or self-test. The selected icon will flash.
4.Press and hold the function button to exit the menu interface and return to the
main interface. Main interface is shown in the figure below.
12
AIYI Technologies
▎Zero calibration:
1.After entering the menu, short press the function button to select the zero
calibration, the zero calibration indicator flashes, then short press the power
button to enter the zero calibration interface. At this time, the zero calibration
indicator keeps display, and the span calibration indicator and self-test indicator
are not displayed. Zero calibration interface is shown in the figure below.
2.Short press the function button to select the specific channel, and the gas
name keep flashes.
3. Short press the power button to do zero calibration. If the calibration is com-
pleted, the “End” prompt will be displayed. If the calibration fails, the “Err”
prompt will be displayed, as shown in the figure.
4. Press and hold the function button to exit the zero calibration operation and
return to the main interface.
13
AIYI Technologies
▎Span calibration:
1.After entering the menu, short press the function button to select the span
calibration, the span calibration indicator flashes, the zero calibration indicator
and the self-test indicator keep displayed, then short press the power button to
enter the span calibration interface. At this time, the span calibration indicator
keeps displayed, and the zero calibration and self-test indicator are not displayed.
Span calibration interface is shown in the figure below.
2.Short press the function button to select the specific channel, and the gas
name in the selected channel flashes.
3.Short press the power button to do span calibration. If the calibration is com-
pleted, the “End” prompt will be displayed. If the calibration fails, the “Err”
prompt will be displayed, as shown in the figure.
4.Press and hold the function button to exit the span calibration and return to
the main interface.
▎Self-test:
1. After entering the menu, short press the function button to select the self-test,
the self-test indicator flashes, the zero calibration indicator and span calibration
indicator keeps displayed, then short press the power button, the detector enters
self-test.
2. The detector will display all the contents in turn, then alarm with sound, light,
vibration once. as shown in the figure.
14
AIYI Technologies
▎Bump test:
15
AIYI Technologies
2.In the main interface, press and hold the function button to enter the bump
test preparation interface. At this time, the self-test indicator flashes, short press
the power button to enter the bump test.
3.Short press the function button to select the quick detection zone and the gas
name indicator flashes.
4.After a stable readings, press the power button to enter bump test, and the gas
name flashes. About 15 seconds later, if the test is normal, the concentration
value will flash for 3 seconds, and then the "End" prompt will be displayed. If the
test is not normal, the “Err” prompt will flash for 3 seconds, and then the gas
concentration will be calibrated for processing. If the calibration is successful, the
correct concentration value will be displayed for 3 seconds, and then the “End”
prompt will be displayed. If the calibration fails, the “Err” prompt will flash for 3
seconds, then the “End” prompt will be displayed, followed by the sensor fault as
“---”.
5.The bump test is end and the detector automatically back to the main interface.
16
AIYI Technologies
2.5 Charging
There is a power indicator in the upper right corner of the display. Please keep
charge in time with original charger and cable when low power. The charging port
is a standard USB charging port. When charging, the charging indicator of the
detector keeps red, and there is also a charging prompt in the upper right corner
of the display. The charge indicator turns green when fully charged.
In the power on state, connect the detector to the computer with the Micro USB
cable, and the computer will automatically recognize and display the correspond-
ing drive letter. After double-clicking on it, you can implement data export and
other operations.
Chapter 3 Maintenance
3.1 Cautions
▎It is strictly forbidden to expose the detector to a gas with high concentration,
corrosive gas, etc. for a long time, otherwise it will cause a serious damage to the
sensor.
▎In order to ensure the performance of the detector, calibration should be carried
out regularly. The calibration frequency depends on the specific working conditions.
It is recommended to calibrate the detector every 3 months. The calibration record
sheet should be filled in each calibration. Considering if factors such as high concen-
tration impact and corrosive gases happen, it is necessary to shorten the calibration
period. In either case, the calibration period must not exceed one year.
▎Keep the detector clean, and if the sensor vent is blocked, it may affects the
detection sensitivity and even damage it.
17
AIYI Technologies
▎Replace sensor in time once the sensor life is over. The replacement of the sen-
sor should be carried out under the guidance of the manufacturer or returned to
the factory for replacement.
▎Keep use the origin parts provided by manufacturer while the maintenance and
the replacement of the parts. Otherwise it may affect the performance and safety
of the instrument itself.
▎The operation of the detector, calibration and other maintenance work should
be carried out by qualified professionals.
▎For the faults encountered in daily maintenance, please contact the manufactur-
er in time.
3.2 Warranty
AIYI Technologies promises that the product will be guaranteed for one year from
sale and will be serviced for life. If the product fails or any problem during the use,
please feel free to contact us, we will be happy to provide you with timely and
quality service!
3.3. Trouble-Shooting
For the faults encountered during use, you can refer to the following ways to solve
the problem, or you can contact the manufacturer directly for consultation.