DHO900 Performance Verification Guide
DHO900 Performance Verification Guide
Copyright
Trademark Information
Notices
• RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
• RIGOL reserves the right to modify or change parts of or all the specifications and pricing
• RIGOL shall not be liable for either incidental or consequential losses in connection with the
• Any part of this document is forbidden to be copied, photocopied, or rearranged without prior
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as
well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard
conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact
RIGOL.
E-mail: [email protected]
Website: http://www.rigol.com
Section Description Page
List of Figures................................................................................................................................III
List of Tables..................................................................................................................................IV
1 Safety Requirement .........................................................................................................1
1.1 General Safety Summary .................................................................................................................. 1
1.2 Safety Notices and Symbols ............................................................................................................ 3
2 Specifications .....................................................................................................................4
3 Document Overview ....................................................................................................... 6
4 Overview ............................................................................................................................. 8
4.1 Test Preparations ................................................................................................................................. 8
4.2 Test Result Record ............................................................................................................................... 9
4.3 Specifications ......................................................................................................................................10
5 Performance Verification Test ....................................................................................11
5.1 Impedance Test ..................................................................................................................................11
5.1.1 Specification ...........................................................................................................................11
5.1.2 Test Connection Diagram ..................................................................................................12
5.1.3 Test Procedures .....................................................................................................................12
5.1.4 Test Record From ..................................................................................................................12
5.2 Input Capacitance Test .................................................................................................................... 13
5.2.1 Specification ...........................................................................................................................13
5.2.2 Test Connection Diagram ..................................................................................................13
5.2.3 Test Procedures .....................................................................................................................13
5.2.4 Test Record From ..................................................................................................................14
5.3 DC Gain Accuracy Test .....................................................................................................................14
5.3.1 Specification ...........................................................................................................................14
5.3.2 Test Connection Diagram ..................................................................................................15
5.3.3 Test Procedures .....................................................................................................................15
5.3.4 Test Record From ..................................................................................................................16
5.4 DC Offset Accuracy Test ..................................................................................................................18
5.4.1 Specification ...........................................................................................................................18
5.4.2 Test Connection Diagram ..................................................................................................18
5.4.3 Test Procedures .....................................................................................................................18
5.4.4 Test Record From ..................................................................................................................19
5.5 Bandwidth Test ...................................................................................................................................20
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
I
5.5.1 Specification ...........................................................................................................................20
5.5.2 Test Connection Diagram ..................................................................................................21
5.5.3 Test Procedures .....................................................................................................................21
5.5.4 Test Record From ..................................................................................................................22
5.6 Time Base Accuracy Test .................................................................................................................23
5.6.1 Specification ...........................................................................................................................23
5.6.2 Test Connection Diagram ..................................................................................................24
5.6.3 Test Procedures .....................................................................................................................24
5.6.4 Test Record From ..................................................................................................................25
6 Appendix: Test Record Form ......................................................................................26
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
II
List of Figures
Figure 4.1 Self-calibration Menu ...........................................................................................9
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
III
List of Tables
Table 2.1 Specifications ............................................................................................................ 4
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
IV
Safety Requirement
1 Safety Requirement
Only the power cord designed for the instrument and authorized for use within
the local country could be used.
The instrument is grounded through the Protective Earth lead of the power cord.
To avoid electric shock, connect the earth terminal of the power cord to the
Protective Earth terminal before connecting any input or output terminals.
If a probe is used, the probe ground lead must be connected to earth ground.
Do not connect the ground lead to high voltage. Improper way of connection
could result in dangerous voltages being present on the connectors, controls or
other surfaces of the oscilloscope and probes, which will cause potential hazards
for operators.
To avoid fire or shock hazard, observe all ratings and markers on the instrument
and check your manual for more information about ratings before connecting
the instrument.
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach
the product. Otherwise, the operator might be exposed to the danger of an
electric shock.
Do not insert objects into the air outlet, as doing so may cause damage to the
instrument.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
1
Safety Requirement
Do not touch exposed junctions and components when the unit is powered on.
If you suspect that any damage may occur to the instrument, have it inspected
by RIGOL authorized personnel before further operations. Any maintenance,
adjustment or replacement especially to circuits or accessories must be
performed by RIGOL authorized personnel.
To avoid short circuit inside the instrument or electric shock, never operate the
instrument in a humid environment.
Do not expose the battery (if available) to high temperature or fire. Keep it out of
the reach of children. Improper change of battery (note: lithium battery) may
cause explosion. Use the RIGOL specified battery only.
Please handle with care during transportation to avoid damage to keys, knob
interfaces and other parts on the panels.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
2
Safety Requirement
WARNING
Indicates a potentially hazardous situation or practice which, if not avoided, will result in
serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not avoided, could result
in damage to the product or loss of important data.
• DANGER
• WARNING
• CAUTION
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
3
Specifications
2 Specifications
This chapter lists the technical specifications of the DHO900 series oscilloscope.
DHO900 series digital oscilloscope achieves 1.25 GSa/s sample rate, a fast waveform
capture rate of 1,000,000 wfms/s with the UltraAcquire mode, up to 50 Mpts memory
depth, 12-bit vertical resolution, all combined with excellent noise floor performance
and vertical accuracy to meet your requirements for more accurate measurements.
The following table shows some of the technical specifications of DHO914, DHO914S,
DHO924, and DHO924S. For other technical specifications of DHO900, please refer to
DHO900 Data Sheet.
Analog Bandwidth 125 MHz 125 MHz 250 MHz 250 MHz
No. of Input
4 analog channel inputs, 16 digital channel inputs
Channels
Input Impedance 1 MΩ ± 1%
Bandwidth Limit
20 MHz, FULL; selectable for each channel
(Typical)
Max. Analog
1.25 GSa/s (single channel[1]), 625 MSa/s (half channels[2]),
Channel Sample
312.5 MSa/s (all channels[3])
Rate
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
4
Specifications
Trigger Source Analog channels (CH1 to CH4), digital channels (D0 to D15)
NOTE
[1]: If any one of the channels is enabled, it is called single channel mode.
[2]: If two of the channels are enabled, it is called half channels mode.
[3]: If any three channels or all four channels are enabled, it is called all channels mode.
[4]: 200 μV/div and 500 μV/div are a magnification of 1 mV/div setting. For vertical accuracy
calculations, use full scale of 8 mV.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
5
Document Overview
3 Document Overview
This manual is designed to guide you to properly test the performance specifications
of RIGOL DHO900 series digital oscilloscope. For the operation methods mentioned
in the test procedures, refer to User Guide of this product.
TIP
For the latest version of this manual, download it from the official website of RIGOL (http://
www.rigol.com).
Publication Number
PVA37100-1110
Software Version
Software upgrade might change or add product features. Please acquire the latest
version of the manual from RIGOL website or contact RIGOL to upgrade the software.
1. Key
The front panel key is denoted by the menu key icon. For example,
indicates the "Default" key.
2. Menu
The menu item is denoted by the format of "Menu Name (Bold) + Character
Shading" in the manual. For example, Setup indicates the "Setup" sub-menu under
the "Utility" function menu. You can click or tap Setup to access the "Setup" menu.
3. Operation Procedures
The next step of the operation is denoted by ">" in the manual. For example,
> Storage indicates first clicking or tapping and then clicking or tapping
Storage.
4. Connector
DHO900 series digital oscilloscope includes the following models. Unless otherwise
specified, this manual takes DHO924S as an example to illustrate the methods for the
performance verification of DHO900 series.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
6
Document Overview
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
7
Overview
4 Overview
1. Self-test
2. Warm-up (make sure that the instrument has been running for at least 30 minutes)
3. Self-calibration
Self-test
After the instrument is connected to the power source, press the power key at the
lower-left corner of the front panel to power on the instrument. You can also click or
tap > Utility > Setup. Then select "Switch On" for the Power status menu. After
the instrument is connected to power source, it will start directly.
During the start-up process, the instrument performs a series of self-tests. After the
self-test, the splash screen is displayed.
Self-calibration
The self-calibration program can quickly make the oscilloscope work in an optimal
state to get the precise measurement results. You can perform self-calibration at any
time, especially when the changes of the ambient temperature reach or above 5℃.
Make sure that the oscilloscope has been warmed up or operating for more than 30
minutes before the self-calibration.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
8
Overview
- Click or tap Start, and then the oscilloscope will start to execute the self-
calibration program.
- After starting the self-calibration program, you can click or tap Exit to cancel
self-calibration operation at any time.
4. Set the vertical scale of each channel to 2 mV/div and view the offset of the
waveform of each channel. If the offset is greater than 0.5 div, check whether there
are interference signals around you and whether the power source is well
grounded. After that, perform self-calibration again.
TIP
It is recommended that users photocopy the test record form before each test and record the
test results in the copy so that the form can be used repeatedly.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
9
Overview
4.3 Specifications
The specification of each test item is provided in this manual. For other technical
parameters, refer to DHO900 Data Sheet (download it from RIGOL website:
www.rigol.com).
TIP
All the specifications are valid only when the oscilloscope has been warmed up for more than
30 minutes within the specified operating temperature range (18℃ to 28℃).
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
10
Performance Verification Test
Recommended
Device Specification Model
TIP
1. Make sure that the oscilloscope passes the self-test and the self-calibration is performed
before executing the performance verification tests.
2. Make sure that the oscilloscope has been warmed up for at least 30 minutes before
executing any of the following tests.
3. Please reset the instrument to the factory setting before or after executing any of the
following tests.
Input Impedance
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
11
Performance Verification Test
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.1 .
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the resistance
measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
12
Performance Verification Test
5.2.1 Specification
Input Capacitance
Specification 15 pF ± 3 pF
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.2 .
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
13
Performance Verification Test
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the capacitance
measurement function. Read and record the capacitance measured.
4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the capacitance
measured.
5. Turn off CH1. Measure the capacitances of CH2, CH3, and CH4 respectively using
the method above and record the measurement results.
DC Gain Accuracy
NOTE
[1]: Full Scale = 8 x Current Vertical Scale. 200 μV/div and 500 μV/div are a magnification of 1
mV/div setting. For vertical accuracy calculations, use full scale of 8 mV for 200 μV/div and 500
μV/div.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
14
Performance Verification Test
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.3 .
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
b. Click or tap the channel status label at the bottom of the screen. Then the
Vertical menu is displayed. Click or tap Probe to enter the Probe setting menu.
Set the probe attenuation ratio to "1X".
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
15
Performance Verification Test
f. In the Horizontal menu, select "Average" for the Acquisition item. Then click or
tap the input field of Averages to set the number of averages to 32 with the
pop-up numeric keypad.
g. Adjust the trigger level to prevent the signals from being triggered by mistake.
5. In the Measure menu, click or tap Vertical measurement item to select "Vavg". The
Vavg result is displayed in the "Result" sidebar at the right side of the screen. Read
and record the measurement result of Vavg1.
6. Adjust Fluke 9500B to make it output a DC signal with -3 div voltage (Vout2).
8. Calculate the relative error of this vertical scale: |(Vavg1 - Vavg2) - (Vout1 - Vout2)| /
Full Scale x 100%.
b. Adjust the output voltage of Fluke 9500B to 3 x the current vertical scale and -3
x the current vertical scale respectively.
d. Calculate the relative error of each vertical scale: |(Vavg1 - Vavg2) - (Vout1 -
Vout2)|/Full Scale x 100%.
10. Turn off CH1. Test the relative error of each scale of CH2, CH3, and CH4
respectively using the method above and record the test results.
Test Result
Vertical
Channel Calculation Limit Pass/Fail
Scale Vavg1 Vavg2
Result[1]
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
CH1 50 mV/div ≤1%
100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
16
Performance Verification Test
Test Result
Vertical
Channel Calculation Limit Pass/Fail
Scale Vavg1 Vavg2
Result[1]
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
CH2 100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
CH3 100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
CH4
100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
17
Performance Verification Test
Test Result
Vertical
Channel Calculation Limit Pass/Fail
Scale Vavg1 Vavg2
Result[1]
10 V/div ≤1%
NOTE
[1]: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale x 100%; wherein,
Vout1 is 3 x the current vertical scale and Vout2 is -3 x the current vertical scale.
DC Offset Accuracy
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.4 .
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
18
Performance Verification Test
d. Set the offset to 1 V, as shown in Test Record From. Close the Vertical menu.
e. Click or tap the horizontal label at the top of the screen. Then the Horizontal
menu is displayed. In this menu, set the time base to 1 ms/div. Select "Average"
for the Acquisition item. Then click or tap the input field of the Averages item
to set the number of averages to 16 with the pop-up numeric keypad. Close the
Horizontal menu.
f. Adjust the trigger level to prevent the signals from being triggered by mistake.
5. In the Measure menu, click or tap Vertical measurement item to select "Vavg". The
Vavg result is displayed in the "Result" sidebar at the right side of the screen. Read
and record the measurement result.
7. Set the output of Fluke 9500B to 0 V. Read the value from the "Result" sidebar and
record the measurement result.
8. Repeat Step 2-7. Measure and record the average results of different vertical scales
and offsets according to Test Record From.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
19
Performance Verification Test
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
20
Performance Verification Test
Bandwidth
NOTE
[1]: Amplitude Loss (dB) = 20 x lg (Vrms2/Vrms1); wherein, Vrms1 is the measurement result of
amplitude effective value at 1MHz and Vrms2 is the measurement result of amplitude effective
value at full bandwidth.
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.5 .
a. Turn on CH1.
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
21
Performance Verification Test
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
b. Click or tap the channel status label of CH1 at the bottom of the screen. Then
the Vertical menu is displayed. Click or tap Probe to enter the Probe setting
menu. Set the probe attenuation ratio to "1X".
e. Click or tap the horizontal label at the top of the screen. Then the Horizontal
menu is displayed. In this menu, you can set the horizontal time base to 500 ns/
div.
4. Output a Sine signal with 1 MHz frequency and 6 div amplitude (e.g. 50 mV/div
vertical scale, 300 mVpp amplitude Sine signal) via the Fluke 9500B. Adjust the
horizontal time base to display five cycles of the waveforms on the screen.
5. In the Measure menu, click or tap Vertical measurement item to select "AC.RMS".
The AC.RMS measurement result is displayed in the "Result" sidebar at the right
section of the screen. Read the value and record the measurement result as Vrms1.
6. Use the Fluke 9500B to output a Sine signal with the maximum bandwidth of the
oscilloscope and 6 div amplitude. The bandwidth frequency differs for different
oscilloscope models. For details, please refer to Specification. Adjust the horizontal
time base to display five cycles of the waveforms on the screen.
7. Check the AC.RMS "Result" sidebar at the right section of the screen. Read and
record the data as Vrms2.
8. Calculate the amplitude loss with the following formula: Amplitude Loss (dB) = 20
x lg (Vrms2/Vrms1).
9. Repeat Step 2-7. Measure Vrms1 and Vrms2, and calculate the amplitude loss of
500 mV/div vertical scale.
10. Turn off CH1. Measure CH2, CH3, and CH4 respectively using the method above
and record the measurement results.
Test Result
Vertical
Channel Amplitude Limit Pass/Fail
Scale Vrms1 Vrms2
Loss[1]
-3 dB to 3
CH1 50 mV/div dB
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
22
Performance Verification Test
Test Result
Vertical
Channel Amplitude Limit Pass/Fail
Scale Vrms1 Vrms2
Loss[1]
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH2
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH3
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH4
-3 dB to 3
500 mV/div dB
NOTE
[1]: Amplitude Loss (dB) = 20 x lg (Vrms2/Vrms1); wherein, Vrms1 is the measurement result of
amplitude effective value at 1MHz and Vrms2 is the measurement result of amplitude effective
value at full bandwidth.
NOTE
[2]: For the number of years that the instrument has been used, please calculate it according to
the date in the verification certificate provided when the instrument leaves factory.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
23
Performance Verification Test
CH1
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.6 .
2. Output a Sine signal with 10 MHz frequency and 1 Vpp amplitude via Fluke 9500B.
b. Turn on CH1.
- Click or tap the channel status label at the bottom of the screen to enable
CH1.
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
turn CH1 on.
c. Click or tap the channel status label of CH1 at the bottom of the screen. Then
the Vertical menu is displayed. Click or tap Probe to enter the Probe setting
menu. Set the probe attenuation ratio to "1X".
f. Click or tap the horizontal label at the top of the screen. Then the Horizontal
menu is displayed. In this menu, you can set the horizontal time base to 1 ns/
div.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
24
Performance Verification Test
h. Click or tap the trigger label at the top of the screen. Set the trigger level to 0 V.
4. Click or tap > Cursor. Then the "Result" sidebar is displayed at the right
section of the screen. Click or tap the "Result" sidebar and select Setting in the
pop-up window. In the Mode item, select Manual to enable the manual cursor
measurement function. Then measure the offset (ΔT) of the middle point of the
signal (namely the crossing point of the rising edge of the current signal and the
trigger level line) relative to the screen center and record the measurement result.
5. Calculate the time base accuracy, namely the ratio of ΔT to the horizontal position
of the oscilloscope. For example, if the offset measured is 1 ns, then the time base
accuracy is 1 ns/1 ms = 1 ppm.
6. Calculate the time base accuracy limit by using the following formula: ±(25 ppm +
5 ppm/year x Number of Years that the Instrument Has Been Used).
NOTE
[2]: For the number of years that the instrument has been used, please calculate it according to
the date in the verification certificate provided when the instrument leaves factory.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
25
Appendix: Test Record Form
Test Result
Vertical
Channel Scale Calculation Limit Pass/Fail
Vavg1 Vavg2
Result[1]
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
CH1
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
26
Appendix: Test Record Form
Test Result
Vertical
Channel Calculation Limit Pass/Fail
Scale Vavg1 Vavg2
Result[1]
100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
CH2 100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
CH3 100 mV/div ≤1%
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
CH4 10 mV/div ≤1%
20 mV/div ≤1%
50 mV/div ≤1%
100 mV/div ≤1%
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
27
Appendix: Test Record Form
Test Result
Vertical
Channel Calculation Limit Pass/Fail
Scale Vavg1 Vavg2
Result[1]
200 mV/div ≤1%
500 mV/div ≤1%
1 V/div ≤1%
2 V/div ≤1%
5 V/div ≤1%
10 V/div ≤1%
NOTE
[1]: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale x 100%; wherein,
Vout1 is 3 x the current vertical scale and Vout2 is -3 x the current vertical scale.
Channel Vertical
Setting Scale Offset Test Result Min. Max.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
28
Appendix: Test Record Form
Test Result
Channel Vertical
Amplitude Limit Pass/Fail
Scale Vrms1 Vrms2
Loss[1]
-3 dB to 3
50 mV/div dB
CH1
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH2
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH3
-3 dB to 3
500 mV/div dB
-3 dB to 3
50 mV/div dB
CH4
-3 dB to 3
500 mV/div dB
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 Performance Verification
29
Appendix: Test Record Form
NOTE
[1]: Amplitude Loss (dB) = 20 x lg (Vrms2/Vrms1); wherein, Vrms1 is the measurement result of
amplitude effective value at 1MHz and Vrms2 is the measurement result of amplitude effective
value at full bandwidth.
NOTE
[2]: For the number of years that the instrument has been used, please calculate it according to
the date in the verification certificate provided when the instrument leaves factory.
DHO900 Performance Verification Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
30