CORNET® ED-88T Electrosmog Meter User's Manual: Applications

The document provides instructions for using the CORNET ED-88T Electrosmog meter. It describes the meter's capabilities for measuring radio frequency, magnetic fields, and electric fields. It also explains the display, controls, modes, and how to interpret measurements.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

CORNET® ED-88T Electrosmog meter user’s manual 7/22/2015

The Audio sound will output from bottom hole of the enclosure, (do not plug the bottom hole)
(12) SysSetup Menu: can be enabled by push and hold the “HOLD” button then push the
st
“Mode” button to get into the SysSetup menu, use the 1 button to move the cursor in the
nd
Menu and use the 2 button to enable/disable the function. In the SysSetup menu:
CORNET Microsystem Inc., ED-88 Electrosmog meter is a Tri-mode device for quick measurement of (a) EXIT: exit the SysSetup menu, return to normal mode
(b) RF level units select: mw/m2, v/m, or dBm
both High frequency (RF) Electromagnetic wave field strength/power density level, Low frequency (LF)
(c) MAX_CLEAR bit: If the MAX_CLEAR is “ON” the MAX value can be cleared by
Magnetic field level(Gauss, Tesla), and Low frequency Electric Field(V/m) for living environments. It toggling the “HOLD” button. If it is “OFF” the MAX value can be cleared only
is an excellent device for individual or company with Electromagnetic wave safety concerns. It has by power off the meter.
RF bandwidth of 100MHz to 8GHz with high sensitivity (0.5uw/m²to 1.8w/m²) and LF bandwidth of (d) Alarm function: ON/OFF and one of the 8 trigger levels(0, -5,-10,-15,-20,-25,-30,-35dBm) can
50Hz to 10KHz (or 50Hz to 1KHz) with sensitivity of 0.1uT to 60uT(1mG-600mG) or 0.01uT to Be selected to trigger the audio Alarm. *(Alarm function is used in RF mode only).
2
1uT(0.1mG to 10mG). and E-field bandwidth of 50Hz-50KHz with sensitivity of 10v/m to 1000v/m, with (e) RESET: reset to factory default (mw/m ,MAX_CLEAR= “on”, Alarm= “off”)
nd
very fast response time (f) SAVE: push the 2 button to save the changes of setup to EEPROM memory. Do not
power off meter, Wait until save is done! (EXIT without SAVE to EEPROM the
Applications: changes will still functional, but it will lost if the meter is power off ).
(13) The LF30 mode: has high sensitivity (0.1mG-10mG), but with lower frequency range
- High frequency RF Electromagnetic wave field strength and power density measurement (50Hz –1KHz) to reduce the high frequency noise. (The Histogram and LED
segment display can go up to 30mG)
- Low frequency LF Magnetic field measurement (Gauss meter function)
The LF600 mode: with sensitivity (1mG-600mG), has higher frequency coverage (50Hz -10KHz).
- Low frequency ELF Electric field measurement (E-field meter function)
- Mobile phone base station antenna radiation power density measurement When measuring the high frequency digital/pulse type of signals,(such as switching power supply)
- Wireless communications, Analog & Digital RF (AM/FM, TDMA, GSM, DECT, CDMA,3G,4G) the LF30 mode might have lower reading than the LF600 mode, this is due to the lower frequency
- RF power measurement for transmitters coverage range of the LF30 mode.
- Wireless LAN (Wi-Fi), Bluetooth, Ultra-wide-band detection, installation, optimization (14) While in LF (magnetic/electric field measurement) or E-field mode, please hold the ED88 steady to
- Spy camera, wireless bug finder get the good stable reading, avoid fast moving of the ED88 to prevent the sudden change of the
- Cellular/Cordless phone radiation safety level, Electrical Utilities SMART METER radiation level reading caused by the Earth magnetic field or the induced electric field of nearby objects.
- AC power line, High voltage tower, power Transformer, motors and small appliance EMF (15) AC Smart Meter radiates RF signal in short burst every few minutes, use MAX function and
- Microwave oven leakage detection Histogram function of ED88, and wait a few minutes to capture the RF signals from smart meter.
- Personal living environment EMF safety evaluation

Usage guide:

(1) Put the 9V battery in the ED88, Handle the unit with right hand in vertical direction, and push the power
switch button to turn on the power, it will come up with RF meter mode.
(2) The RF sensor is located in the left hand side of the ED88; the LF sensor is located in the right
hand top side of the ED88, the E-field sensor is located in the middle top side of the ED88, please do
not cover the sensor area with hand or other objects.
(3) There are two push button below the power switch: “Mode” button and “Hold” button. “Mode”
button is used to switch in between RF mode, two LF modes, and E-field mode. The “Hold” button is
used to freeze the data reading of the ED88
(4) RF mode: measured RF field strength/power density is shown on the digital LCD display
(with dBm, v/m, or mw/m²). 8 LED lights with Red, Yellow, and Green color on the right hand
side of LCD window are used for quick RF signal level indications. 3 Red LEDs are used to
indicate the 3 safety ranges. The power level of each LED can be found in the table on the
ED88 front panel.
(5) LF mode: measured LF magnetic field strength is shown on the digital LCD display (with uT
and mGauss). Two LF modes can be selected by “Mode” button: (a) LF30mode (0.1mG-10mG)
and (b)LF600 mode (1mG-600mG). 8 color LED lights are also available to show the relative
Field strength/power density readout:
strength of the magnetic field.
(6) E-field mode: measured ELF electric field strength is shown on the digital LCD display with V/m. ED88 use 8 high brightness LED to indicate the measured power density with 3 safety
(7) Histogram: previous 30 signal level readings are recorded and shown as moving graph indications.
on the LCD display for RF, LF, and E-field modes. It can be used for finding direction of signal source
and recording bursts from digital RF signals such as signals from AC smart meter. RF LF600/LF30 E-
LED RF Power
(8) Hold: HOLD button can be used to halt the data measurement of the ED88, a “HOLD” mark Power level field Indication Action
color density
will shown on the LCD screen to indicate the “Hold” condition. Push the “Hold” button again the level level
ED88 will exit from the “Hold” condition. RED3
-5 dBm
0.18 w/m
2 30uT/3uT 500 Italy RF safety
Caution!
2
(9) MAX: Maximum measured data value since the last power-on is shown on the LCD display. up up v/m standard (0.1w/m )
(10) Average: average signal value is displayed on the LCD with “A” mark. It can be used to RED2 -10 dBm 0.058 w/m
2
20uT/2uT
200 Swiss RF safety
Caution!
v/m standard (0.04w/m2)
estimate the duty cycle of the digital RF burst signals.
100 Russian RF safety
(11) Sound function & LCD backlight: go into the “Hold mode” first, then toggle the “Mode” button RED1 -15 dBm 0.018 w/m2 10uT/1uT 2 Caution!
v/m standard (0.02w/m )
in sequence to turn-on/off the LCD backlight and the Sound signature function, a “S” mark on 75
the LCD indicates the sound mode is on. The LCD,backlight is forced off when the Sound is ON YELLOW3 -20 dBm 5.8 mw/m2 5uT/0.5uT safe
v/m
to reduce the battery current consumption. (Remember to turn-off the LCD backlight or the YELLOW2 -25 dBm 1.8 mw/m2 2uT/0.2uT
50
safe
sound function when it is not needed). Audio Sound can be used to detect very low level RF v/m
signals (down to 0.05uw/m2) especially the modern digital RF burst signals. YELLOW1 -30 dBm 0.58 mw/m
2
0.5uT/0.05uT
30
safe
v/m
©2015 CORNET Microsystems Inc., 1400 Coleman Ave #C28 Santa Clara, CA 95050 USA Tel: (408)9690205 20 WiFi Wireless LAN
www.cornetmicro.com ED88TV1.0 07/22/2015 GREEN3 -35 dBm 0.18 mw/m2 0.2uT/0.02uT safe
v/m typically in this range
-40 dBm 0.2uT/0.02uT 10vm Some signal source
GREEN2 0.06 mw/m2 safe
down down around

Downloaded from www.Manualslib.com manuals search engine


NOTE: The European Community provided general guidelines in its Council
Recommendation of July 1999.1 ICNIRP published similar guidelines in
April 1998.2 Table I gives a sampling of the international and national field-
 Most high frequency RF antenna such as Mobile phone base station is vertical polarized strength limit values for the general public and continuous exposure (for
(in vertical direction), therefore while in RF mode, the ED88 is normally used in vertical Reference only !)
direction. For LF mode, the LF sensor is located in the top right hand of the ED88 and the
meter is normally used in Horizontal position in LF mode. For E-field mode, the E-field sensor 950Mhz 1850Mhz
is located in the top middle of the meter, point the sensor to the ELF source. Please rotate the
meter to find the maximum reading direction in either case. The maximum reading will also
42 V/m 59 V/m
increase as you approach the source. ED88 can be used to find the location of signal source. International Council Recommendation 1999/519/EC
(4.75W/m2) (9.25W/m2)
 Most of modern communication devices (Mobile phone, Wireless LAN, Wi-Fi, etc.,) use
42 V/m 59 V/m
digital RF burst signals. When measuring this type of signals, several LED lights will blinking International ICNIRP Guidelines, April 1998
at the same time. this is normal and it can be used as an indication of burst type of RF signals. (4.75W/m2) (9.25W/m2)
For continues waves (AM/FM) signals, the LED light will be stable. ED88 measures the peak 49 V/m 61 V/m
Austria Ö NORM S1120
power density of the signal with very fast sampling time. It is more accurate than the needle style (6.33W/m2) (10W/m2)
of readout which only shown the average value of signal power most of the time. 21 V/m 30 V/m
Belgium Belgisch Staatsblad F.2001-1365
 Electromagnetic wave field strength/power density reduces very fast with distance (distance (1.18W/m2) (2.31W/m2)
square), keep a good distance from the high frequency RF signal source can reduce the high 42 V/m 59 V/m
Germany 26. Deutsche Verordnung
frequency radiation effect. Alumina foil or window sun reflector film (silver color) can be used (4.75W/m2) (9.25W/m2)
as an effective and low cost shielding material for most of RF radiations. 6 V/m 6 V/m
 ED88 is designed for quick living environment RF radiation evaluation and is for reference use (0.1W/m2) (0.1W/m2)
Italy Decreto n. 381, 1998
only. Official RF safety radiation measurement procedure is complicate and should be handled 20 V/m 20 V/m
by trained technical person with lab instruments. Safety range standard listed below is for (1W/m2) (1W/m2)
reference only. ED88 is not a medical instrument, Please do not use it in medical, legal or other 51 V/m 83 V/m
The Netherlands Health Council
related applications. (6.92W/m2) (18W/m2)
 The “SENSOR” mark on the bottom of the back enclosure cover is not used in ED88 4 V/m 6 V/m
 ED88 is designed for home and personal use only, It is not for commercial rental purpose. Switzerland Verordnung 1999
(0.04W/m2) (0.1W/m2)
 The audio sound output of the sound function is the demodulated RF signal, it is good for AM 49 V/m 68 V/m
and modern digital RF signals (pulse/burst ) detection, it is not for FM or constant amplitude RF United States IEEE C95.1
(6.33W/m2) (12W/m2)
signals or LF/ELF modes. it is an excellent tool for RF signal type indication. (different RF Draft: National Quality Technology 49 V/m 61 V/m
signal such as Wifi, GSM, DECT, ….etc., all has different sound frequency signature of the China
Monitoring Bureau (6.33W/m2) (10W/m2)
demodulated RF signal) and for very low level signal detection.
 When in E-field mode, the electric field induced by human body or large objects nearby can Radio-Radiation Protection Guidelines, 49 V/m 61 V/m
affect the measurement results, hold the ED88 by hand on the lower right side of the meter, do Japan
1990 (6.33W/m2) (10W/m2)
not cover the E-field sensor area (top of the meter) by hand or other objects, keep away from
large metal door or objects. Point the top of the ED88 to the high voltage AC power line (with
ED88 at least 1 meter above the ground) when measuring the VLF/ELF E-field radiation from AC
power lines or towers. 1W/m2 = 1000mW/m2, 1mW/m2 = 1000uW/m2 = 0.1uW/cm2
1uW/m2 =0.001mW/m2 , 1uT = 10mG
Specification
Sensor type: Electric field sensor and Magnetic field sensor
Frequency range & Sensitivity: RF: 100MHz to 8GHz
2
(-60dBm to +5dBm),(0.5uw/m2 to 1.8w/m ),
(14mv/m to 26.2v/m)
LF1: 50Hz to 10KHz (0.1uT to 60uT)/(1mG to600mG)
LF2: 50Hz to 1KHz (0.01uT to 1uT)/(0.1mG to 10mG)
ELF: 50Hz to 50KHz (10v/m to 1000v/m)
RF Peak power measurement: 0.5uw/m²to 1.8w/m²
Display type: digital LCD graphic display
Unit of measurements: dBm, mw/m², v/m, uT, mG
LCD back light: 15 seconds auto-off and manual on/off control
Display of data: LCD 4 and 5 digit, 8 LED color segment,
Moving Histogram(level/time) of previous 30 recorded data,
Analog segment bar
Data update rate: Sampling rate: 10000/sec. Display update rate: 2/sec.
Error rate: RF: +/- 3.5dBm, LF: 20%, E-field: 25%
Functions: Hold, Max, Average, Sound signature, Alarm,
Sound & Alarm: Audio Sound ON/OFF control, programmable Alarm triggering level
Safety standard indication: 3 safety range indication by 3 Red LED
Battery used: 9V alkaline battery, or 9v Li-ion rechargeable battery (not included)
Battery life: >20 hours

©2015 CORNET Microsystems Inc., 1400 Coleman Ave #C28 Santa Clara, CA 95050 USA Tel: (408)9690205
www.cornetmicro.com ED88TV1.0 07/22/2015