has met the requirements of AC204, IAS Accreditation Criteria for Calibration Laboratories, and has demonstrated
compliance with ISO/IEC Standard 17025:2017, General requirements for the competence of testing and calibration
laboratories. This organization is accredited to provide the services specified in the scope of accreditation.
Effective Date September 15, 2023
Expiration Date February 1, 2025
President
RICI COMPANY LTD.
www.ricionline.com
Anvil 0 mm to 25 mm 6.0 µm
Digital 0 mm to 25 mm 1.1 µm
Plain Anvil / Digital 25 mm to 150 mm 2.6 µm
* If information in this CMC is presented in non-SI units, the conversion factors stated in NIST Special Publication
811 “Guide for the Use of the International System of Units (SI)” apply.
•
Major Diameter
0 mm to 600 mm 8.1 µm
Taper Thread Ring Gauge Effective Diameter ULM 600
0 mm to 8 mm 2.0 µm LABC-161
8 mm to 100 mm 2.0 µm
Jigs and Fixtures Length 0 mm to 25 mm 0.69 µm ULM 600 / Slip Gauge Blocks
Measurement 25 mm to 50 mm 0.92 µm / Long Slip Gauges
50 mm to 100 mm 1.6 µm
100 mm to 200 mm 3.9 µm LABC-171
200 mm to 300 mm 5.7 µm
300 mm to 400 mm 7.6 µm
400 mm to 500 mm 9.4 µm
500 mm to 1000 mm 20 µm
1000 mm to 2000 mm 41 µm
Mechanical
Force-Compression 0.0025 kN to 3000 kN 0.2 % Load Cells
LABC-007
Force-Tension Up to 50 kN 0.03 kN Tensile Load cell
LABC-168
Proving Rings 0 kN to 50 kN 0.05 kN Reference Load Cell with
Indicator
LABC-066
Balances 0 g to 210 g 0.33 mg Class E1, F1 Weights
201 g to 2.1 kg 6.7 mg LABC-006
2.1 kg to 8 kg 11 mg
8 kg to 30 kg 77 mg
30 kg to 100 kg 94 g
100 kg to 500 kg 0.06 kg
500 g 0.72 mg
200 g 0.29 mg
100 g 0.14 mg
50 g 0.072 mg
20 g 0.031 mg
10 g 0.017 mg
5g 0.10 mg
2g 0.048 mg
1g 0.002 mg
500 mg 2.5 μg
200 mg 1.6 μg
100 mg 1.5 μg
50 mg 1.2 μg
20 mg 1.1 μg
10 mg 0.9 μg
5 mg 0.7 μg
2 mg 0.7 μg
1 mg 0.7 μg
Hydrometers 0.600 g/mL to 1.050 g/mL 0.0029 g/mL Reference Solutions
0.600 g/mL to 1.050 g/mL 0.00031 g/mL LABC-064
Anemometers/ Air Meters 25 ft/min to 9000 ft/min 7 ft/min Reference Air Meter, Wind
(Air Velocity) Tunnel
LABC-069
Densitometers 0 D to 14 D 0.021 D Reference Density Film
LABC-068
Electrical – DC/LF
Welding Machines 0 A to 600 A 1.2 % of Setting Check Master 9000
0 V to 50 V (DC) 0.56 V LABC-016
Holiday Detector 0 kV to 40 kV 0.06 kV Fluke High Voltage Probe,
Digital Multimeter
LABC-015
DC Voltage –Generate3 0 mV to 329.9999 mV 0.4 µV Multi Product Calibrator Fluke
0.33 V to 3.299999 V 6.3 µV 5522A
3.3 V to 32.99999 V 0.12 mV LABC-018
33 V to 329.9999 V 0.97 mV
330 V to 1020.000 V 3.5 mV
3.3 V to 32.9999 V
(10 Hz to 90 kHz) 4.4 mv
33 V to 329.999 V
(45 Hz to 100 kHz) 16 mV
330 V to 1020 V
(45 Hz to 8 kHz) 39 mV
1 V to 10 V
(10 Hz to 300 kHz) 2.8 mV
10 V to 100 V
(10 Hz to 300 kHz) 0.026 V
100 V to 1000 V
(10 Hz to 300 kHz) 0.039 V
3.3 mA to 32.999 mA
(10 Hz to 30 kHz) 5 µA
33 mA to 329.99 mA
(10 Hz to 30 kHz) 47 µA
0.33 A to 1.09999 A
(10 Hz to 10 kHz) 0.38 mA
1.1 A to 10.9999 A
(10 Hz to 10 kHz) 1.8 mA
11 A to 20.5 A
(45 Hz to 5 kHz) 4.9 mA
(10 Hz to 3 kHz)
1.16 nF to 3.2999 nF 0.0051 nF
(10 Hz to 1 kHz)
3.3 nF to 10.9999 nF 0.0078 nF
11 nF to 32.999 nF 0.049 nF
110 nF to 329.999 nF 0.21 nF
(10 Hz to 80 Hz)
33 μF to 109.9999 μF 0.60 μF
(0 Hz to 50 Hz)
110 μF to 329.999 μF 0.22 μF
(0 Hz to 20 Hz)
0.33 mF to 1.09999 mF 0.16 μF
(0 Hz to 6 Hz)
1.16 mF to 3.2999 mF 0.0014 mF
(0 Hz to 2 Hz)
3.3 mF to 10.9999 mF 0.034 mF
(0 Hz to 0.6 Hz)
11 mF to 32.9999 mF 0.120 mF
(0 Hz to 0.2 Hz)
33 mF to 110.00 mF 0.034 mF
Capacitance – Measure4 0 nF to 1 nF 0.016 nF Precision Multimeter Fluke
1 nF to 10 nF 0.063 nF 8846A
10 nF to 100 nF 0.63nF LABC-018
100 nF to 1 μF 6.3 nF
1The uncertainty covered by the Calibration and Measurement Capability (CMC) is expressed as the expanded
uncertainty having a coverage probability of approximately 95 %. It is the smallest measurement uncertainty that a
laboratory can achieve within its scope of accreditation when performing calibrations of a best existing device. The
measurement uncertainty reported on a calibration certificate may be greater than that provided in the CMC due to the
behavior of the calibration item and other factors that may contribute to the uncertainty of a specific calibration.
2When uncertainty is stated in relative terms (such as percent, a multiplier expressed as a decimal fraction or in scientific
notation), it is in relation to instrument reading or instrument output, as appropriate, unless otherwise indicated.
3Capability is suitable for the calibration of measuring devices in the stated ranges.
Notes:
1. Gs = Gauss; 1 Gs corresponds to 10-4 T
2. CPS = Centipoise
3. D = Degree of Darkness (optical density)
4. ppm = parts per million