English Aerospray Stat (7122) Applications Manual 57-2000-01A
English Aerospray Stat (7122) Applications Manual 57-2000-01A
English Aerospray Stat (7122) Applications Manual 57-2000-01A
AEROSPRAY®
HEMATOLOGY
STAT
SLIDE
STAINER/CYTOCENTRIFUGE
Model
7122
Applications
Manual
57-‐2000-‐01A
©2013
ELITechGroup
Inc.
All
rights
reserved.
Printed
in
the
United
States
of
America.
No
part
of
this
publication
may
be
reproduced,
transmitted,
transcribed,
stored
in
a
retrieval
system,
or
translated
into
any
language
(human
or
computer)
in
any
form,
or
by
any
means
whatsoever,
without
the
prior
express
written
permission
of
ELITechGroup
Inc.
Wescor,
Aerospray
and
Cytopro
are
trademarks
of
ELITechGroup
Inc.
Other
trade
names
used
in
this
manual
are
trademarks
of
their
respective
owners,
used
here
for
information
only.
ELITechGroup
Inc.
makes
no
express
or
implied
warranty
regarding
this
manual,
its
quality,
performance,
or
appropriate
use
regarding
any
type
of
specific
procedure.
ELITechGroup
Inc.
may
modify
this
manual
without
notice
and
without
implying
any
obligation
or
liability
on
the
part
of
the
company.
Manufactured
in
the
United
States
of
America
by:
ELITechGroup
Inc.
370
West
1700
South
Logan,
Utah
84321-‐8212
USA
2
Table
of
Contents
Section
1
Introduction
1.1
Instrument
Overview
......................................................................
7
Using
this
Manual
..................................................................
7
Functional
Description
..........................................................
8
Key
Features
..........................................................................
9
Intended
Use
.........................................................................
9
Table
1:
General
Specifications
.............................................
10
Table
2:
Performance
Specifications
.....................................
11
Table
3:
Run
Time
Sequence
..................................................
12
Table
4:
Approximate
Reagent
Consumption
........................
12
Table
5:
Carousel
and
Rotor
Information
..............................
13
Table
6:
Explanation
of
Symbols
............................................
14
1.2
Instrument
Description
...................................................................
16
Figure
1:
Front
and
Right
Side
Panels
....................................
16
Figure
2:
Front
Panel
and
Touchscreen
.................................
17
Figure
3:
Rear
Panel
..............................................................
17
Figure
4:
Stainer
Bowl
Components
......................................
18
Figure
5:
Nozzle
Diagram
......................................................
18
Table
7:
Preventive
Maintenance
Kit
.....................................
19
Barcode
Reader
.....................................................................
20
Figure
6:
Barcode
Reader
......................................................
20
1.3
Touchscreen
and
User
Interface
.....................................................
21
Table
8:
Front
Panel/Main
Screen
Function
Keys
..................
21
Table
9:
System
Setup
Keys
...................................................
22
Table
10:
Maintenance
Function
Keys
...................................
24
Section
2
Setup
and
Preparation
for
Operation
2.1
Instrument
Setup
............................................................................
25
Unpacking
and
Installing
the
Stainer
.....................................
25
Connecting
the
Drain
Tube
and
Waste
Container
.................
26
Connecting
Power
.................................................................
26
Installing
Reagent
Bottles
......................................................
27
Installing
the
Barcode
Reader
...............................................
28
2.2
Preparing
the
Stainer
for
Operation
...............................................
29
Priming
Procedures
...............................................................
29
The
Clean
Cycle
.....................................................................
31
Performing
Tests
...................................................................
31
Reagent
Level
Monitoring
.....................................................
32
Enabling/Disabling
Reagent
Level
Detect
........................
32
Zeroing
the
Reagent
Level
Sensors
..................................
33
Section
3
Controlling
and
Customizing
Stainer
Functions
3.1
System
Setup
Menu
........................................................................
35
Accessing
the
System
Setup
Menu
........................................
35
Stain
Programs
......................................................................
35
Creating
a
Stain
Program
......................................................
36
Editing,
Renaming,
or
Adjusting
Stain
Programs
...................
36
Administrator
and
User
Accounts
.........................................
37
Creating
an
Administrator
Account
.................................
37
3
Table
of
Contents
Section
3
Controlling
and
Customizing
Stainer
Functions
(continued)
Creating
User
Accounts
...................................................
38
Managing
User
Access
....................................................
39
User
Login/Logout
...........................................................
39
Using
Reagent
Information
Tracking
.....................................
41
Modifying
Level
Detect
Functions
.........................................
41
Changing
User
Language
.......................................................
42
Setting
the
Date
and
Time
.....................................................
42
System
Log
............................................................................
42
Accessing
Logs
.................................................................
42
Exporting
Logs
.................................................................
43
Controlling
Beeper
Alerts
......................................................
43
QC/Maintenance
Tracking
....................................................
44
Enable
Stain
Slide
Tracking
..............................................
44
Enable
Cyto
Slide
Tracking
...............................................
45
Enable
Manual
Entry
.......................................................
45
Enable
Preventive
Maintenance
Tracking
.......................
45
Enable
Reagent
Tracking
.................................................
45
Restoring
Software
Defaults
.................................................
46
3.2
Recording
Specimen
and
Reagent
Information
..............................
47
Scanning
Slides
with
the
Barcode
Reader
.............................
47
Scanning
Reagent
Bottles
with
the
Barcode
Reader
.............
48
Manually
Entering
Specimen
information
............................
49
Manually
Entering
Reagent
Information
...............................
50
3.3
The
Help
Menu
................................................................................
51
Using
Help
.............................................................................
51
Section
4
Operating
the
Stainer
4.1
Operating
Instructions
....................................................................
52
Suggested
Staining
Protocol
..................................................
52
Performing
a
Hub
Pattern
Test
.............................................
52
Figure
7:
Correct
Hub
Pattern
Test
Result
.....................
52
Figure
8:
Incorrect
Hub
Pattern
Test
Result
...................
52
Loading
the
Carousel
.............................................................
53
Figure
9:
Loading
the
Carousel
......................................
54
Figure
10:
Specimen
Placement
on
Slide
.......................
54
Using
Blocking
Slides
.............................................................
55
Figure
11:
Using
Blocking
Slides
....................................
55
Performing
a
Stain
Cycle
.......................................................
55
Unloading
the
Carousel
.........................................................
56
Monitoring
Reagent
and
Waste
Levels
.................................
57
Table
11:
Reagent
Level
Detect
Display
Symbols
..........
57
Replacing
a
Reagent
Bottle
...................................................
58
Emptying
the
Waste
Container
.............................................
58
Section
5
Preventive
Maintenance
and
Safety
5.1
Preventive
Maintenance
.................................................................
59
Daily
Maintenance/Quality
Control
(QC)
..............................
59
Weekly
Maintenance
............................................................
59
Monthly
Maintenance
...........................................................
60
Using
the
Preventive
Maintenance
Log
................................
60
4
Table
of
Contents
Section
5
Preventive
Maintenance
and
Safety
(continued)
5.2
Storing
the
Instrument
....................................................................
61
Preparing
for
Long-‐Term
Storage
..........................................
61
5.3
Replacing
Fuses
...............................................................................
62
5.4
Cleaning
the
Stainer
and
Carousels
.................................................
63
Cleaning
Liquid
Spills
.............................................................
63
Cleaning
Broken
Slides
..........................................................
63
5.5
Decontaminating
the
Stainer
and
Carousels
...................................
64
Solutions
for
Decontaminating
the
Instrument
....................
64
Figure
12:
Lid
Latch
and
Locking
Pin
Hole
Locations
.....
65
Decontaminating
the
Instrument
..........................................
65
Decontaminating
Reagent
Bottles
........................................
66
5.6
Shipping
or
Disposing
of
the
Stainer
or
Carousels
..........................
67
Shipping
the
Instrument
or
Carousels
...................................
67
Hazard-‐Free
Certification
..............................................
67
Disposing
of
the
Instrument
or
Carousels
.............................
67
Section
6
Nozzle
Maintenance
and
Performance
6.1
Nozzle
Disassembly
and
Cleaning
...................................................
68
Nozzle
Disassembly
...............................................................
68
Nozzle
Cleaning
.....................................................................
68
6.2
Nozzle
Reassembly
..........................................................................
69
6.3
Manual
Priming
...............................................................................
70
6.4
Performing
the
Volume
Test
...........................................................
71
Table
12:
Volume
Test
Tolerances
.........................................
71
6.5
Performing
the
Slide
Pattern
Test
...................................................
73
Figure
13:
Correct
Slide
Pattern
Test
Result
..........................
73
Figure
14:
Incorrect
Slide
Pattern
Test
Result
........................
73
Section
7
Solving
Problems
7.1
Troubleshooting
..............................................................................
74
Table
13:
General
Troubleshooting
and
Diagnosis
................
74
7.2
Abnormal
Staining
Results
..............................................................
80
Smear
Separation
or
Tearing
.................................................
80
Improper
Alcohol
Fixation
.................................................
80
Wet
Smears
.......................................................................
80
Thick
Smears
......................................................................
80
Dirty
Slides
.........................................................................
80
Red
Cell
Artifacts
...............................................................
80
Use
Aerofix®
......................................................................
81
False-‐Positive
Howell-‐Jolley
Bodies
or
Amorphous
Precipitates
........................................................................
81
7.3
Instrument
Malfunction
..................................................................
83
Air
or
Reagent
Leaks
..............................................................
83
Reagent
Level
Detect
System
Errors
.....................................
84
Calibrating
the
Reagent
Level
Detect
System
.......................
84
7.4
Calibrating
the
Touchscreen
...........................................................
86
7.5
Service
Information
.........................................................................
87
5
Table
of
Contents
Section
8
Cytopro®
Cytocentrifuge
8.1
Cytopro
Cytocentrifuge
Information
...............................................
88
Functional
Description
..........................................................
88
Key
Features
..........................................................................
88
Intended
Use
.........................................................................
88
Appendix
A:
Critical
Reagent
Components
............................................
89
Appendix
B:
Risk
and
Safety
Phrases
.....................................................
91
Appendix
C:
Accessories
and
Supplies
..........................................................
94
6
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Using
this
Manual
This
manual
provides
instructions
to
install,
operate,
and
maintain
the
Aerospray
Hematology
Stat
Stainer/Cytocentrifuge.
The
manual
is
an
important
part
of
the
product.
Read
it
carefully
and
completely
before
setup
and
first
use
of
the
instrument.
If
additional
accident
prevention
and
environmental
protection
requirements
exist
in
the
country
of
operation,
this
manual
must
be
supplemented
by
appropriate
instructions
to
ensure
compliance.
Safety
Regulations
This
instrument
has
been
built
and
tested
in
accordance
with
safety
regulations
for
the
following
device
types:
electrical
control,
regulating,
and
laboratory.
In
order
to
maintain
this
condition
and
ensure
safe
operation,
the
operator
must
observe
all
the
instructions
and
warnings
contained
in
this
manual.
For
current
information
about
applicable
standards,
please
refer
to
the
CE
Declaration
of
Conformity
included
with
the
documents
shipped
with
this
device.
NOTE:
This
equipment
complies
with
the
emission
and
immunity
requirements
described
in
the
IEC
61326
series.
Understanding
Warnings
This
manual
uses
three
warning
levels
to
alert
you
to
important
information
as
shown
in
the
following
examples.
WARNING!
A
Warning
alerts
to
the
possibility
of
personal
injury,
death,
or
other
serious
adverse
reactions
stemming
from
the
use
or
misuse
of
this
device
or
its
components.
CAUTION:
A
Caution
alerts
to
possible
problems
with
the
device
associated
with
its
use
or
misuse.
Such
problems
include
device
malfunction,
failure,
damage,
damage
to
the
sample,
or
damage
to
other
property.
Where
applicable,
a
Caution
may
include
precautions
to
be
taken
to
avoid
the
hazard.
NOTE:
A
Note
reinforces
or
supplies
additional
information
about
a
topic.
Specific
Warnings
Pay
particular
attention
to
the
following
safety
precautions.
If
these
safety
precautions
are
ignored,
injury
or
damage
to
the
instrument
may
occur.
Each
individual
precaution
is
important.
WARNING!
Install
the
stainer
in
a
well-‐ventilated
area.
If
ventilation
is
inadequate,
operate
the
instrument
under
a
safety
hood.
7
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
WARNING!
Reagents
used
in
the
stainer
contain
moderately
hazardous
chemicals
that
require
care
in
handling.
Always
use
appropriate
safety
measures
including
gloves
and
eye
protection
when
handling
reagents.
WARNING!
Always
wear
protective
clothing
and
eye
protection
when
using
Nozzle
Cleaning
Solution
(diluted
REF:
SS-‐029C).
Dispose
of
used
solution
properly.
WARNING!
If
power
is
lost
while
the
stainer
is
running,
the
lid
will
remain
locked
until
power
is
restored.
Do
not
attempt
to
open
the
lid
while
power
is
off.
WARNING!
Electrical
shock
hazard:
Do
not
open
this
instrument
or
attempt
internal
repairs.
Refer
servicing
to
qualified
service
personnel.
Contact
Elitech
Group
Biomedical
Systems
service.
CAUTION:
This
equipment
has
been
designed
and
tested
to
CISPR
11
Class
A
and
FCC
Part
15
Class
A.
In
a
domestic
environment
it
may
cause
radio
interference,
in
which
case,
you
may
need
to
take
measures
to
mitigate
the
interference.
CAUTION:
To
avoid
serious
instrument
damage,
always
use
reagents
supplied
by
Elitech
Group.
Except
for
approved
methanol,
using
reagents
not
supplied
by
Elitech
Group
may
void
the
warranty.
CAUTION:
Only
spare
parts
supplied
or
specified
by
Elitech
Group
should
be
used
in
this
device.
Using
non-‐approved
parts
may
affect
the
performance
and
safety
features
of
the
instrument.
If
the
equipment
is
used
in
a
manner
not
specified
by
Elitech
Group
the
protection
provided
by
the
equipment
may
be
impaired.
If
in
doubt,
contact
your
Elitech
Group
representative.
Functional
Description
The
Aerospray
Hematology
Stat
Slide
Stainer/Cytocentrifuge
(Model
7122)
is
a
dual-‐
purpose,
microprocessor-‐controlled
slide
staining
and
cell
preparation
system.
In
use,
atomizing
spray
nozzles
apply
fresh
reagents
onto
microscope
slides
prepared
with
blood
or
other
body
fluid
specimens.
The
slides
are
mounted
in
a
rotating
carousel
for
processing.
8
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Key
Features
• Minimized
reagent
consumption
• Rapid
staining
• Barcode
scanner
for
tracking
specimens
and
reagents
• Reagent
and
specimen
traceability
• User
traceability
• Administrator
password
• Interactive
touchscreen
display
• Multiple
languages
• High-‐volume
staining
productivity
(12
slides
per
stain
cycle)
• Separate
reservoir,
delivery
tube,
pump,
and
spray
nozzle
for
each
reagent
• Operator-‐selectable
automatic
alcohol
fixation
function
to
fix
specimens
• Reagent
and
waste
level
monitoring
• Log
files
• 10
Stain
Intensity
Settings
The
correct
accessory
must
be
used
for
each
function.
The
Cytopro
Cytocentrifuge
Rotor
is
available
as
an
option
offering
additional
features
(see
Section
8).
Intended
Use
The
Aerospray
Hematology Stat
Slide
Stainer/Cytocentrifuge
(Model
7122)
is
intended
for
use
by
medical
professionals
to
stain
specimens
that
may
include
blood
and
other
body
fluids,
as
a
step
of
standard
laboratory
practice
in
diagnosing
disease
in
humans.
Addition
of
the
Cytopro
rotor
allows
preparation
of
slides
by
cytocentrifugation
before
staining.
9
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
1:
General
Specifications
Category
Characteristics
Slide
Carousel
Capacity
1
to
12
Carousel
Rotation
Speed
Reagent
application:
20
rpm
(±
5%)
Slide
Drying:
950
rpm
(±
5%)
Cytocentrifuge
Rotor
Speed
100
to
2000
rpm
(±
5%),
user
programmable
Reagent
Consumption
Refer
to
Approximate
Reagent
Consumption,
Table
4
Operating
Time
Refer
to
Run
Time
Sequence,
Table
3
Display
7
in.
WVGA
(800
x
480
pixels)
TFT
Touchscreen
Controls
Menu-‐driven
icons
Drain
Connection
Connector
on
rear
panel
accepts
male
connector
attached
to
vinyl
drain
tube.
1.8
meters
(6
ft.)
length
supplied
Ventilation
Air
is
exhausted
from
the
stainer
via
a
female
½
inch
SAE
pipe
thread
fitting
Dimensions
Width
57
cm
(22
in.)
Height
(lid
closed)
25
cm
(10
in.)
Depth
54
cm
(21
in.)
Height
(lid
open)
58
cm
(23
in.)
Weight
15.2
kg
(33.5
lb.)
Electrical
Requirements
100
to
240
VAC
(±
10%)
@
50
to
60
Hz
Power
Consumption
200
Watts
maximum
Fuses
2.0
Amp
time-‐lag
(Type
T)
(5
x
20
mm)
(2
required)
Manufactured
by
Littlefuse
(Part
No.
218002)
or
Bussman
(Part
No.
GDC-‐2A)
Ambient
Temperature
Operating
15
to
30
°C
(59
to
86
°F)
Storage
-‐10
to
50
°C
(14
to
122
°F)
Relative
Humidity
≤
80%
non-‐condensing
Altitude
≤
2000
m
(≤
6562
ft.)
10
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
1:
General
Specifications
(continued)
Category
Characteristics
Pollution
Degree
2
Heat
Dissipation
Maximum
512
Btu/hour
max
Average
During
Staining
34
Btu/hour
Average
While
Idle
10
Btu/hour
Maximum
Sound
Emission
72
dB
Table
2:
Performance
Specifications
Category
Characteristics
Reagent
Spray
Nozzles
Four
reagent
nozzles:
A,
B,
C,
D
Reagents:
A
-‐
Rinse
NOTE:
Use
only
Elitech
Group
reagents,
with
diluents
B
–
Thiazin
Stain
(Blue)
as
specified
for
Elitech
Group
concentrated
reagents.
REF
numbers
for
this
stainer
start
with
one
of
the
C
–
Eosin
Stain
(Red)
following:
SS-‐035,
SS-‐049,
SS-‐135,
SS-‐149,
SS-‐048,
or
SS-‐148.
D
–
Methanol
or
Aerofix®
Note:
Methanol
must
be
99.5%
pure
no
more
than
0.5%
water.
Stain
Settings
Number
of
slides
to
stain.
Intensity
settings:
1-‐10.
Fixation
level:
Off,
Normal,
High
11
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
3:
Run
Time
Sequence
NOTE:
Table
3
represents
a
typical
timing
sequence
for
this
instrument
using
the
12-‐slide
carousel,
presented
for
general
reference
only.
Timing
is
in
seconds.
Actual
cycle
times
may
vary.
Staining
Mode
Medium
Dark
Setting
(5)
Fixation
Pump
D
(0ptional)
(102)
Eosin
Stain
(Pump
C)
21
Thiazin
Stain
(Pump
B)
33
Rinse
(Pump
A)
27
High
Speed
Dry
33
Total
without
fixation
(in
minutes)
1.9
Total
with
fixation
(in
minutes)
3.6
Table
4:
Approximate
Reagent
Consumption
(mL
)
Staining
Mode
Medium
Dark
Clean
Cycle
Setting
(5)
Reagent
A:
Rinse
6.0
7.5
Reagent
B:
Thiazin
6.0
7.5
Reagent
C:
Eosin
6.0
7.5
Reagent
D:
Fixative
1.4
38.0
12
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
5:
Carousel
and
Rotor
Information
Only
the
following
slide
staining
carousel
or
cytocentrifuge
rotor
can
be
used
in
this
instrument.
Each
should
be
used
following
the
instructions
in
this
manual
or
the
Cytopro
Applications
Manual
(RP-‐517).
Rotor/Carousel
Maximum
Maximum
Capacity
Maximum
rpm
Sample
Volume
12-‐Slide
Carousel
(AC-‐028)
1000
rpm
12
each,
26
mm
x
76
mm
N/A
(1
x
3
inch)
microscope
slides
Cytopro
Cytocentrifuge
Rotor
(AC-‐160)
2000
rpm
8
each,
standard
Up
to
600
µL*
chambers,
plus
slides
8
each,
Cytopro
Magnum
Up
to
6
mL*
chambers,
plus
slides
*Do
not
overfill
cytocentrifuge
chambers.
See
Cytopro
Applications
Manual
or
Methods
Manual
for
detailed
instructions
and
warnings.
13
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
6:
Explanation
of
Symbols
Alternating
Current
(AC)
Authorized
Representative
in
the
European
Community
Batch
Code
Biological
Hazards
(Biological
Risks)
Catalog
Number
(Model
Number)
Caution,
Consult
Accompanying
Documents
(Attention,
see
instructions
for
use)
CE
Mark,
product
meets
the
essential
requirements
designated
in
Annex
I
of
the
In
Vitro
Device
Directive
(IVDD)
98/79/EC
Consult
Instructions
For
Use
Do
Not
Reuse
Do
not
use
if
package
is
damaged
Fragile,
Handle
with
Care
Fuse
General
Symbol
for
Recovery,
Recyclable
In
vitro
Diagnostic
Device
(In
vitro
Diagnostic
Medical
Device)
Keep
away
from
sunlight
(Keep
away
from
heat)
Manufacturer
Waste
Electrical
and
Electronic
Equipment
Reagent
stable
after
opening
for
number
of
months
specified
14
SECTION
1
INTRODUCTION
1.1
Instrument
Overview
Table
6:
Explanation
of
Symbols
(continued)
I
“On”
(Power)
O
“Off”
(Power)
Environment
Friendly
Use
Period
Serial
Number
Temperature
Limitation
–
indicates
high
and
low
limits.
(Normal
room
temperatures
are
specified
for
all
Elitech
Group
Reagents.)
Use
By
Volume
Control
General
Warning,
Caution,
Risk
of
Danger
Warning,
Biological
Hazard
Corrosive
Environment
Hazard
Flammable
Harmful/Irritant
Oxidant
Toxic
Product
to
be
used
for
manual
cleaning
only.
Do
not
pump
product
through
instrument.
15
SECTION
1
INTRODUCTION
1.2
Instrument
Description
Figure
1:
Front
and
Right
Side
Panels
4
3
2
5
1
D
C
B
A
1
–
Front
Panel
with
Touch
Screen
Display
6
2
–
Bowl
3
–
Carousel
4
–
Lid
with
Safety
Lock
5
–
Right
Side
Panel
with
Label
Indicating
Reagent
Positions:
A
–
Rinse
B
–
Thiazin
C
–
Eosin
D
–
Methanol
or
Aerofix
6
–
Reagent
Tray
16
SECTION
1
INTRODUCTION
1.2
Instrument
Description
Figure
2:
Front
Panel
and
Touchscreen
1
2
1
–
Standby/Ready
Button
2
–
Touchscreen
The
front
panel
features
an
interactive
touchscreen
display.
Refer
to
Front
Panel/Main
Screen
Function
Keys
(Section
1.3,
Table
8)
for
more
information.
Figure
3:
Rear
Panel
1
1
2
3
4
5
11
10
9
8
7
6
1
–
USB
Ports
2
–
Network
Ethernet
Connection
3
–
Exhaust
Vent
4
–
Rear
Panel
Label
5
–
Biohazard
Warning
Label
6
–Waste
Tube
Connection
7
–
Level
Detection
Connection
for
Waste
Container
8
–
Power
Switch
9
–
Fuse
Door
10
–
Power
Cord
Connection
11
–
Model/Serial
Number
Label
17
SECTION
1
INTRODUCTION
1.2
Instrument
Description
Figure
4:
Stainer
Bowl
Components
1
2
3
4
5
1
–
Nozzle
D
(Methanol
or
Aerofix)
2
–
Nozzle
C
(Eosin
stain)
3
–
Nozzle
B
(Thiazin
stain)
4
–
Nozzle
A
(Rinse)
5
–
Drive
Hub
Figure
5:
Nozzle
Diagram
1
2
3
4
1
–
Compression
Screw
2
–
Swirl
Cone
3
–
O-‐Ring
4
–
Nozzle
Housing
18
SECTION
1
INTRODUCTION
1.2
Instrument
Description
Table
7:
Preventive
Maintenance
Kit
Component
Name
Description
Manual
Priming
Tool
Primes
air-‐locked
pumps.
Silicon
Grease
Lubricates
the
nozzle
threads
for
easy
assembly.
Nozzle
Wire
Cleans
nozzle
housing
orifices.
Nozzle
Cleaning
Strainer
Strains
the
nozzle
parts
to
prevent
them
from
going
down
a
drain.
19
SECTION
1
INTRODUCTION
1.2
Instrument
Description
Table
7:
Preventive
Maintenance
Kit
(continued)
Component
Name
Description
Volume
Test
Collection
Collects
reagents
while
performing
Tubes
(small
tube)
the
Volume
Test.
Nozzle
Maintenance
Tube
Holds
Nozzle
Cleaning
Tubes
(large
Stand
tube)
and
Volume
Test
Tubes
(small
tube).
Nozzle
Cleaning
Tubes
For
soaking
nozzles
in
the
Nozzle
(large
tube)
Cleaning
Solution.
Barcode
Reader
An
optional
barcode
reader
is
available
for
the
Hematology
Stat
Stainer/Cytocentrifuge
(Model
7122).
Figure
6:
Barcode
Reader
20
SECTION
1
INTRODUCTION
1.3
Touchscreen
and
User
Interface
Users
control
all
instrument
functions
from
the
interactive
touchscreen
display.
Table
8:
Front
Panel/Main
Screen
Function
Keys
Button
Name
Description
Standby/Ready
With
instrument
power
ON:
Blue
=
Ready
Amber
=
Standby
Pressing
Standby
runs
a
System
Clean
cycle
and
places
instrument
into
standby
mode.
The
Standby/Ready
button
also
accesses
the
touchscreen
calibration
function.
Refer
to
System
Setup
Menu
(Section
3.1).
Maintenance
Accesses
features
for
verifying
proper
nozzle
performance
and
places
pumps
in
a
testing
sequence.
Accesses
the
line
priming,
Pattern
Test,
Volume
Test,
and
B-‐Line
Flush
functions.
Clean
Performs
the
Clean
cycle.
Cyto
Enters
the
Cytocentrifuge
mode.
System
Information
Shows
the
system
information,
including
serial
number
and
software
version.
Allows
access
to
the
System
Setup
features.
Refer
to
System
Setup
Menu
(Section
3.1).
Help
Opens
the
software
Help
file.
Programs
Allows
users
to
select
or
edit
programs.
Start/Load
Slides
Starts
a
Stain
or
Cytocentrifuge
cycle.
Button
is
inactive
until
a
program
is
created.
Refer
to
Creating
a
Stain
Program
(Section
3.1).
With
Slide
Tracking
enabled,
opens
the
Scan
and
Load
Slides
menu
(Section
3.2).
Number
of
Specimen
Selects
the
number
of
specimen
slides
in
the
Slides
carousel.
Users
staining
an
odd
number
of
specimen
slides
should
press
the
next
higher
specimen
slide
number
icon.
21
SECTION
1
INTRODUCTION
1.3
Touchscreen
and
User
Interface
Table
8:
Front
Panel/Main
Screen
Function
Keys
(continued)
Button
Name
Description
Back
Returns
to
the
previous
menu.
Stop
Aborts
any
operation.
22
SECTION
1
INTRODUCTION
1.3
Touchscreen
and
User
Interface
Table
9:
System
Setup
Keys
(continued)
Button
Name
Description
Set
Date/Time
Allows
users
to
set
the
date
and
time.
Restore
Defaults
Restores
programming
to
default
settings.
Login
Enters
Login
sequence
for
authorized
users.
Logout
Logs
authorized
users
out.
Users
must
log
in
again
to
use
the
stainer.
Save
Saves
the
entered
or
selected
information.
Add
Enters
programming
mode
for
creating
staining
and
cytocentrifuge
programs.
Also
allows
the
system
administrator
to
authorize
new
users.
Allows
manual
entry
of
slide
or
specimen
information.
Delete/Erase/Remove
Deletes
or
erases
the
selected
item.
Edit/Change
User
Allows
editing
of
an
existing
stain
or
cytocentrifuge
program.
Allows
manual
entry
of
slide
or
specimen
information
(stain
or
cytocentrifuge
mode).
Also
allows
system
administrator
to
edit
user
information.
Zero
Zeros
the
Level
Detect
sensors.
Calibrate
Calibrates
the
Level
Detect
system.
Unselected
Shows
an
unselected
option.
Selected
Shows
a
selected
or
enabled
option.
23
SECTION
1
INTRODUCTION
1.3
Touchscreen
and
User
Interface
Table
10:
Maintenance
Function
Keys
Button
Name
Description
Individual
Prime
Primes
the
selected
line.
Buttons
(A,
B,
C,
D)
Pattern
Test
Performs
Pattern
Test
to
ensure
nozzles
are
clear
of
debris
and
spraying
properly.
Volume
Test
Performs
Volume
Test
to
verify
the
selected
nozzle
volume
is
within
the
correct
range.
60-‐Sec
Prime
Runs
the
pumps
for
1
minute
and
primes
the
lines.
QC/PM
Shows
the
Preventive
Maintenance
and
Quality
Control
logs,
when
enabled
from
the
System
Setup
menu
(Section
3.1).
Line
Flush
Cleans
the
B
and
C
reagent
lines.
ABCD
Prime
button
Primes
all
(A,
B,
C,
D)
reagent
lines
simultaneously.
24
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.1
Instrument
Setup
Unpacking
and
Installing
the
Stainer
Follow
this
sequence
if
you
are
using
this
instrument
for
the
first
time.
Details
about
these
operations
are
given
in
the
next
three
sections.
CAUTION:
Contact
Elitech
Group
before
installing
the
instrument
if
you
observe
any
damage
to
the
packaging
or
equipment.
1
Unpack
and
inspect
the
instrument.
2
Check
that
the
contents
of
the
boxes
match
the
packing
lists
for
the
instrument
and
accessories.
3
Open
the
instrument
lid
and
remove
the
cardboard
tube
that
protects
the
hub.
NOTE:
Keep
the
box
and
packaging
material
to
repack
the
instrument
if
you
intend
to
ship
it
to
the
manufacturer
for
service.
4
Place
the
instrument
on
a
flat
surface,
free
from
dust
and
vibration
and
away
from
direct
sunlight.
NOTE:
Position
the
instrument
with
the
rear
panel
at
least
30
cm
(12
in.)
from
obstructions
or
hazardous
materials.
CAUTION:
Keep
the
drain
tube
straight
and
as
short
as
possible.
The
maximum
length
is
1.8
m
(72
in.).
The
waste
container
must
be
positioned
lower
than
the
stainer.
25
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.1
Instrument
Setup
Connecting
the
Drain
Tube
and
Waste
Container
1
Insert
the
waste
tube
connector
into
the
rear
panel
receptacle
until
you
hear
a
click.
2
Adjust
the
tube
length
to
less
than
1.8
m
(72
in.).
NOTE:
Ensure
the
waste
tube
has
no
loops
or
kinks,
and
is
as
straight
and
as
short
as
possible.
Cut
off
excess
tubing
as
needed.
3
Connect
the
drain
tube
to
the
waste
container.
If
using
a
waste
bottle
with
level
detect
(REF:
AC-‐182):
4
Connect
the
waste
monitoring
cable
to
the
rear
panel
receptacle.
5
Connect
the
waste
monitoring
cable
to
the
waste
container
lid.
Connecting
Power
1
Make
sure
the
power
switch
is
OFF
(O).
2
Plug
the
power
cord
into
the
power
connector
on
the
rear
panel
of
the
instrument.
NOTE:
Use
a
surge
protector
to
isolate
the
instrument
from
electrical
spikes
and
surges.
3
Plug
the
power
cord
into
a
properly
rated
AC
electrical
outlet.
4
Turn
the
power
switch
ON
(I).
After
a
brief
delay
the
Main
menu
will
appear.
26
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.1
Instrument
Setup
Installing
Reagent
Bottles
WARNING!
Reagents
used
in
the
instrument
contain
moderately
hazardous
chemicals
that
require
care
in
handling.
Always
handle
reagents
using
appropriate
safety
measures,
including
gloves
and
eye
protection.
NOTE:
Reagents
should
be
stored
according
to
the
conditions
specified
on
their
label.
After
opening,
reagents
are
stable
until
the
expiration
date
specified
on
the
label
unless
otherwise
indicated
by
the
symbol
shown
at
left.
1 Place
each
500
mL
reagent
bottle
in
the
correct
position.
(A)
Rinse
(B)
Thiazin
Stain
(C)
Eosin
Stain
(D)
Methanol
NOTE:
See
Appendix
A
for
complete
identification
of
all
reagents
used
in
this
A
B
C
D
stainer.
CAUTION:
To
avoid
severe
damage,
never
use
reagents
containing
organic
solvents
(such
as
ketones)
in
this
instrument,
unless
supplied
by
Elitech
Group,
or
specified
in
official
Elitech
Group
formulation
instructions.
NOTE:
Immediately
remove
spills
in
the
reagent
tray
to
preserve
the
accuracy
of
the
reagent
level
detecting
system.
2
For
all
reagents:
• Open
a
new
bottle
of
reagent.
• Record
the
reagent
letter
on
each
cap.
• Insert
the
corresponding
dip
tube
into
the
reagent
bottle
and
install
the
ring
cap.
27
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.1
Instrument
Setup
Installing
the
Barcode
Reader
A
barcode
reader
can
be
connected
to
the
stainer
for
scanning
reagent
bottles
and
specimen
slides
that
contain
barcodes.
This
allows
easy
reagent
and
specimen
information
tracking.
If
a
barcode
reader
is
not
installed,
reagent
and
specimen
information
can
be
entered
manually
(Section
3.2).
Installing
the
Barcode
Reader
1
Place
the
barcode
reader
and
stand
on
a
level
surface
near
the
stainer.
2 Plug
the
barcode
reader
into
the
left
USB
port
on
the
rear
panel
of
the
stainer.
See
Section
3.2
for
instructions
on
using
the
barcode
reader.
28
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
Priming
Procedures
NOTE:
The
instrument
is
shipped
with
alcohol
in
the
reagent
lines.
For
proper
performance,
this
alcohol
must
be
replaced
with
the
correct
reagent
for
each
reagent
line
prior
to
use.
Thoroughly
purge
and
prime
each
reagent
delivery
line
using
the
following
instructions.
1
Remove
each
spray
nozzle
with
the
provided
nozzle
tool
by
turning
counterclockwise.
D
C
B
A
2
Note
the
location
of
each
nozzle
so
you
can
return
it
to
the
original
position
during
reassembly.
3
Place
a
carousel
on
the
stainer
hub
to
prevent
stain
from
entering
the
motor
shaft.
CAUTION:
Fluid
from
priming
can
flood
and
damage
the
motor
if
the
drain
tube
is
not
properly
installed.
4
Press
Maintenance
from
the
Main
menu.
5
Press
the
A
prime
button.
Stain
should
appear
within
10
seconds.
When
properly
primed,
a
steady
stream
of
reagent
(no
sputtering
or
breaks)
flows
from
the
nozzle
receptacle.
• If
stain
appears,
proceed
to
the
next
step.
• If
stain
does
not
appear
within
10
seconds,
perform
the
manual
priming
procedure
(Section
6.3).
CAUTION:
Never
operate
a
dry
pump
for
more
than
10
seconds.
Operating
a
dry
pump
may
cause
damage
to
the
instrument.
29
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
Priming
Procedures
(continued)
6
Repeat
the
previous
steps
for
each
reagent
line
(B,
C,
and
D).
7
Press
60-‐Sec
Prime
to
prime
each
reagent
line
with
200
mL
of
reagent
to
remove
all
of
the
alcohol
from
the
reagent
lines
and
pumps.
8
To
prime
all
lines
at
the
same
time,
press
ABCD.
The
pumps
will
run
for
1
minute
and
prime
all
the
lines.
Follow
instructions
on
the
screen.
NOTE:
To
prime
individual
lines,
press
the
appropriate
individual
prime
button
(A,
B,
C,
D).
D
C
B
A
9 Return
the
nozzles
to
their
original
positions
and
tighten
clockwise
with
the
nozzle
tool.
10 With
the
nozzles
installed,
repeat
Steps
5
and
6.
A
fine
cone
of
spray
should
come
from
each
nozzle.
11 After
verifying
nozzle
performance,
a
clean
cycle
must
be
performed
before
a
stain
cycle
can
be
run.
30
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
The
Clean
Cycle
The
Clean
cycle
is
a
two-‐step
process
that
cleans
the
reagent
nozzles
and
the
carousel.
The
first
step
purges
each
reagent
line
and
cleans
the
empty
carousel
with
methanol
or
Aerofix
while
cycle
progress
is
displayed
on
the
screen.
The
second
step,
which
can
be
delayed
indefinitely,
reprimes
the
reagent
lines
with
staining
reagents.
NOTE:
Pressing
Standby/Ready
performs
a
System
Clean
cycle
before
the
instrument
goes
on
standby.
1
Place
an
empty
carousel
in
the
instrument
and
close
the
lid.
CAUTION:
Never
place
any
carousel
loaded
with
specimens
in
the
instrument
for
a
clean
cycle
(including
placing
the
instrument
in
standby
mode).
Specimens
will
be
damaged
if
they
contact
reagents
sprayed
from
the
nozzles
when
you
press
Clean
or
Standby.
2
Press
Clean.
NOTE:
Pressing
Stop
during
the
Clean
cycle
causes
the
Clean
Cycle
Cancelled
message
to
be
displayed.
Press
Clean
to
complete
the
interrupted
cycle.
3
Press
Reprime
to
reprime
the
lines.
e
4
Open
the
lid
and
remove
the
carousel
when
the
Clean
cycle
is
complete.
5
Spray
the
interior
of
the
bowl
with
70
to
100%
methanol
or
ethanol.
Wipe
the
stainer
bowl
dry
with
paper
towels.
NOTE:
Perform
the
instrument
storing
procedure
(Section
5.2),
if
the
instrument
will
remain
idle
for
more
than
1
week.
Performing
Tests
We
recommend
performing
the
Pattern
Test
and
Volume
Test
before
using
the
stainer
(see
Section
6).
31
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
Reagent
Level
Monitoring
Reagent
Level
Detect
monitors
reagent
levels
and
alerts
you
when
the
reagent
is
running
low,
or
when
the
waste
container
is
full
(when
using
the
waste
container
with
level
detect
(AC-‐182)).
You
can
turn
reagent
and
waste
container
monitoring
ON
or
OFF
from
the
Level
Detect
menu.
The
system
default
is
ON
for
reagent
monitoring
and
OFF
for
waste
container
monitoring.
NOTE:
The
instrument
must
be
installed
on
a
flat,
level
surface
for
accurate
reagent
monitoring.
CAUTION:
This
system
is
designed
to
warn
you
when
the
reagent
level
is
getting
low.
The
instrument
will
continue
running
through
these
warnings.
Always
monitor
and
replenish
(if
needed)
the
reagent
before
running
a
stain
cycle.
Enabling/Disabling
Reagent
Level
Detect
1
Press
Information
from
the
Main
menu.
2
Press
System
Setup.
3
Press
Level
Detect.
The
display
shows:
4
Press
Tray
to
enable,
or
OFF
to
disable
a
reagent
line.
Functions
are
grey
when
unselected,
blue
when
selected.
Press
External
to
enable
level
detect
for
the
waste
bottle.
5
When
finished,
press
Back
to
exit
to
System
Setup
menu.
32
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
Zeroing
the
Reagent
Level
Sensors
The
Level
Detect
function
must
be
zeroed
at
initial
setup,
when
the
stainer
is
moved,
or
if
the
level
detect
is
not
reporting
correctly.
If
zeroing
does
not
correct
the
problem,
recalibrate
the
Level
Detect
function
(Section
7.3).
1
Press
Information.
2
Press
System
Setup.
3 Press
Level
Detect
to
enter
the
Reagent
Level
Detect
Setup
menu.
NOTE:
The
stainer
should
be
turned
ON
for
at
least
30
minutes
before
zeroing
to
stabilize
level
sensors.
The
instrument
can
be
used
during
this
time.
4
Press
Zero.
The
display
shows:
33
SECTION
2
SETUP
AND
PREPARATION
FOR
OPERATION
2.2
Preparing
the
Stainer
for
Operation
Zeroing
the
Reagent
Level
Sensors
(continued)
5
Remove
all
reagent
bottles
and
press
Start.
NOTE:
Vibrations
or
bumps
to
the
instrument
or
lab
bench
can
cause
inaccuracies
in
zeroing
or
calibration.
6
After
zeroing,
press
OK.
Press
the
Back
button
to
exit
to
the
System
Setup
menu.
7
Return
the
reagent
bottles
to
their
correct
positions
in
the
tray.
NOTE:
For
accurate
reagent
level
detection
and
calibration,
dip
tubes
must
follow
their
pre-‐formed
coiled
shapes.
34
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Many
software
settings
can
be
controlled
from
the
System
Setup
menu,
including:
• Creating,
editing,
and
deleting
stain
programs
• Creating,
editing,
and
deleting
cytocentrifuge
programs
• Tracking
reagent
information
• Managing
user
accounts
• Enabling
tracking
features
for
slides,
maintenance,
and
reagents
• Managing
reagent
level
sensing
• Changing
the
display
language
• Viewing
and
exporting
the
system
log
• Changing
beeper
settings
• Setting
the
date
and
time
• Restoring
default
settings
• Fixation
Fixation
options
are
Off,
Normal,
and
High.
35
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Creating
a
Stain
Program
1 From
System
Setup,
press
Stain
Programs.
2
Press
Add.
3
Select
Enter
Program
Name,
and
enter
the
desired
name
on
the
keypad.
4
Press
Enter
on
the
keypad.
The
display
returns
to
the
Stain
Programming
Menu.
5
Select
Intensity
From
the
Stain
Programming
Menu.
Select
the
desired
stain
intensity
(1-‐10).
7 Press Save.
Editing,
Renaming,
or
Adjusting
Stain
Programs
1
From
Stain
Programs
menu
select
the
program
to
be
modified.
2
Press
Edit.
36
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Editing,
Renaming,
or
Adjusting
Stain
Programs
(continued)
3
Adjust
the
settings
as
needed.
4
Press
Save.
Administrator
and
User
Accounts
You
can
create
one
Administrator
account
and
multiple
(up
to
50)
user
accounts.
The
Administrator
controls
access
to
the
system
by
adding
and
editing
user
accounts.
Users
cannot
edit
System
Settings
unless
permitted
by
the
Administrator.
Creating
an
Administrator
Account
1
From
System
Setup,
press
Users.
2
Select
Lock
System
Setup
Access.
3
Enter
a
password
for
the
Administrator
account
(at
least
4
characters)
and
press
Enter.
4
Re-‐enter
the
password
and
press
Enter
to
confirm.
37
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Creating
User
Accounts
Note:
This
function
is
available
only
if
an
administrator
account
has
been
created.
1 Press
System
Setup.
2 Enter
the
Administrator
password.
3
Press
Enter.
4
Press
Users
to
reveal
the
Manage
Users
menu.
5
Select
Enable
Global
Login.
6
Select
Add
User.
7
Enter
a
new
user
name.
8 Press Enter.
38
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Creating
User
Accounts
(continued)
9
Enter
a
numeric
passcode
(at
least
4
numbers)
for
the
user
account.
10
Press
Enter.
11
Re-‐enter
the
passcode
to
confirm.
12
Press
Enter.
Managing
User
Access
From
the
Manage
Users
screen,
the
Administrator
has
several
options
to
manage
user
access
to
the
instrument.
• Enable
Global
Login
allows
users
to
log
in
to
the
instrument.
Users
will
log
out
manually
or
automatically
(with
user-‐selectable
time
options).
See
User
Login/Logout
below.
• Enable
Run
Login
requires
the
current
user
to
enter
a
password
to
run
a
Stain
or
Cytocentrifuge
cycle.
Global
Login
must
be
enabled
to
use
this
option.
• User
System
Access
enables
complete
control
of
the
instrument,
including
changing
the
System
Setup
options.
This
option
can
be
controlled
on
an
individual
user
basis,
if
Global
Login
is
enabled.
User
Login/Logout
With
System
Access
locked
and
Global
Login
enabled,
users
must
log
in
to
use
the
stainer:
1 Select
User
ID
and
Select
Logout
time
after
idle
from
the
drop-‐down
menu.
NOTE:
Users
can
select
how
long
the
stainer
can
be
idle
before
automatically
logging
the
user
out.
39
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
User
LogIn/Logout
(continued)
2 Press
Login.
3 Enter
the
correct
passcode
for
the
selected
user
and
press
Enter.
4 Once
Login
is
complete,
the
display
advances
to
the
Main
screen
and
the
instrument
is
ready
for
programming
and
staining.
A
Logout
button
and
the
user
name
appear
at
the
top
right
of
the
Main
screen.
NOTE:
Users
can
log
out
manually
by
pressing
the
logout
button.
40
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Using
Reagent
Information
Tracking
You
can
enter
reagent
information
to
help
track
reagent
usage
and
expiration.
Reagent
information
includes
reference
number,
expiration
date,
lot
number,
date
and
time
the
reagent
was
last
installed.
1
From
System
Setup,
press
QC/Maintenance
Tracking.
2
Select
Enable
Reagent
Tracking
by
choosing
reagent
A,
B,
C,
or
D.
This
enables
reagent
lot
number
and
expiration
date
tracking.
3
Press
Back
to
return
to
System
Setup.
4
Press
Reagents.
5
Press
Change
next
to
the
appropriate
reagent.
6
Scan
the
reagent
bottle
barcodes
(Section
3.2)
or
manually
enter
the
reagent
information
in
the
correct
fields.
7
Press
Save
for
each
reagent.
Modifying
Level
Detect
Functions
The
Level
Detect
function
alerts
you
when
the
reagent
is
running
low,
or
when
the
waste
container
is
almost
full.
You
can
turn
reagent
and
waste
container
monitoring
ON
and
OFF
from
the
System
Setup
screen.
The
system
defaults
to
ON
for
reagent
monitoring
and
to
OFF
for
waste
container
monitoring.
See
Section
2.2
for
complete
instructions.
1
From
System
Setup,
press
Level
Detect.
41
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Changing
User
Language
1
From
System
Setup,
press
Language.
2 Select the software language from the list on the left.
3
Press
OK.
Setting
the
Date
and
Time
1
From
System
Setup,
press
Set
Date/Time.
2 Select 12 for a 12-‐hour clock or 24 for a 24-‐hour clock.
3 Use the up and down arrows to modify the time and date.
4
Press
Save.
System
Log
The
instrument
records
all
login,
logout,
stain
or
cytocentrifuge
cycles,
setting
changes,
maintenance
functions
and
specimen
identification
(if
enabled).
Accessing
Logs
1
From
System
Setup,
press
System
Log.
2
Use
navigation
arrows
to
scroll
through
the
log.
42
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
System
Log
(continued)
Exporting
Logs
1 From
System
Setup,
press
System
Log.
2
Plug
a
Flash
Drive
into
the
right
USB
port.
3
Press
Export.
NOTE:
The
log
files
are
exported
to
the
Flash
Drive
as
a
CSV
file
that
can
be
used
in
spreadsheet
software
programs.
2
Use
the
sliders
to
modify
the
beeper
volume
for
Cycle
Complete,
Warnings,
Errors,
or
Key
Clicks.
3
Press
Beep
On
Startup
to
turn
the
audible
startup
alert
ON
or
OFF.
43
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
QC/Maintenance
Tracking
Under
system
default
settings,
the
following
QC/Maintenance
Tracking
options
are
disabled:
• Stain
Slide
Tracking
• Cyto
Slide
Tracking
• Manual
Entry
• Preventive
Maintenance
Tracking
• Reagent
Tracking
Enable
Stain
Slide
Tracking
To
activate
Stain
Slide
Tracking:
1
From
System
Setup,
Press
QC/Maintenance
Tracking.
2
Press
Enable
Stain
Slide
Tracking.
3
Press
Back
twice
to
return
to
the
main
screen.
Verify
that
the
Start
Button
on
the
main
screen
reads
“Load
Slides.”
NOTE:
Selecting
Enable
Stain
Slide
Tracking
changes
the
Start
button
on
the
Main
menu
to
“Load
Slides.”
4
Press
Load
Slides.
The
Scan
and
Load
Slides
menu
appears.
5
Enter
slide
information.
a.
If
using
the
barcode
reader,
scan
the
specimen
slides
that
contain
barcodes.
See
Scanning
Slides
with
the
Barcode
Reader
(Section
3.2).
b.
If
entering
specimen
information
manually,
see
Manually
Entering
Specimen
Information
(Section
3.2).
6
See
Section
4
for
remaining
steps
for
running
a
stain
cycle.
44
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Enable
Cyto
Slide
Tracking
Allows
slide
tracking
in
cytocentrifuge
mode.
See
the
Cytopro
Rotor
Applications
Manual
(Aerospray
Models
7xx2)
(REF:
RP-‐517)
for
complete
information.
Enable
Manual
Entry
If
selected,
allows
manual
entry
of
slide
information
using
the
keypad
(limited
to
24
characters).
Enable
Preventive
Maintenance
Tracking
To
activate
the
tracking
prompts
for
Preventive
Maintenance
Tracking,
use
the
following
steps:
1
From
System
Setup,
select
QC/Maintenance
Tracking.
2
Select
Enable
Preventive
Maintenance
Tracking.
3
Enter
the
information
for
the
Daily,
Weekly,
and
QC
Slide
prompts
in
corresponding
fields.
See
Using
the
Preventive
Maintenance
Log
(Section
5.1).
Enable
Reagent
Tracking
To
activate
Reagent
Tracking:
1
From
the
QC/Maintenance
Tracking
menu,
select
the
reagent
(A,
B,
C,
D)
to
be
tracked.
45
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.1
System
Setup
Menu
Restoring
Software
Defaults
1
From
System
Setup,
select
Restore
Defaults.
CAUTION:
Restoring
the
system
defaults
will
remove
all
personal
settings.
• Restoring
System
Settings
will
delete
all
user
names
and
passwords.
• Restoring
Stain
Settings
will
delete
all
stain
programs
and
restore
the
default
programs.
• Restoring
Cytocentrifuge
Settings
will
delete
all
cytocentrifuge
programs
and
restore
the
default
programs.
2
Select
the
settings
you
would
like
to
restore
to
factory
defaults:
System
Settings,
Stain
Settings,
or
Cytocentrifuge
Settings.
3 Press
Restore.
4
The
display
returns
to
the
Main
menu.
46
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.2
Recording
Specimen
and
Reagent
Information
Scanning
Slides
with
the
Barcode
Reader
1 From
System
Setup
select
QC/Maintenance
Tracking.
2 Select
Enable
Stain
Slide
Tracking.
NOTE:
Selecting
Enable
Stain
Slide
Tracking
changes
the
Start
button
on
the
Main
menu
to
Load
Slides.”
3 Press
Back
twice
to
return
to
the
Main
menu.
4 Press
Load
Slides
on
the
Main
menu.
The
Scan
and
Load
Slides
menu
will
appear.
5 Scan
the
barcode
of
each
slide
in
the
batch
and
load
into
the
carousel
according
to
instructions
in
Section
4.1.
6 Verify
that
each
barcode
appears
on
the
Scan
and
Load
Slides
menu.
7
When
you
have
completed
preparations
to
stain,
(Section
4.1)
press
Start.
47
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.2
Recording
Specimen
and
Reagent
Information
Scanning
Reagent
Bottles
with
the
Barcode
Reader
1 From
System
Setup
select
QC/Maintenance
Tracking.
2 Select
Enable
Reagent
Tracking
for
each
desired
reagent
(A,
B,
C,
D).
3 Press
Back
to
return
to
the
System
Setup
menu.
4 Press
Reagents
to
reveal
the
Reagent
Information
screen.
5 Select
the
desired
Reagent
(A,
B,
C,
D)
and
press
Change.
6 Scan
the
barcode
of
each
enabled
reagent
bottle.
7
Verify
that
the
barcode
appears
on
the
Scan
and
Load
Slides
menu.
48
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.2
Recording
Specimen
and
Reagent
Information
Scanning
Reagent
Bottles
with
the
Barcode
Reader
(continued)
8
Press
Save.
9
Repeat
steps
3-‐8
for
each
reagent
bottle
that
is
enabled
in
QC
Maintenance
Tracking.
NOTE:
You
can
access
Reagent
Information
by
pressing
the
bottle
icons
on
the
right
side
of
the
Main
menu.
This
takes
you
directly
to
Reagent
Information
menu,
where
you
can
scan
or
manually
enter
reagent
information
by
pressing
Change.
Manually
Entering
Specimen
Information
With
Stain
Slide
Tracking
and
Manual
Entry
enabled
in
the
QC
Maintenance
menu:
1 Press
Load
Slides
on
the
Main
menu.
2 Press
Add
to
reveal
the
keypad.
3 Enter
slide
information
(maximum
of
24
characters)
and
press
Enter.
4 To
change
or
delete
the
entry,
select
the
entry
on
the
display
and
press
Edit
or
Remove.
5 Load
slides
and
run
stain
cycle
as
shown
in
Section
4.1.
NOTE:
Reagent
REF
number
must
be
a
valid
Elitech
Group
REF
number
for
the
selected
reagent.
Incorrect
entries
will
generate
an
error
message.
49
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.2
Recording
Specimen
and
Reagent
Information
Manually
Entering
Reagent
Information
1 Press
Reagents
from
the
System
Setup
menu,
or
press
the
reagent
status
icon
on
the
Main
menu
to
reveal
the
Reagent
Information
menu.
2 Select
the
desired
reagent
and
press
Change.
3 Select
the
desired
field
(Reagent
REF,
Expiration
Date,
Lot
Number,
or
Service
Date/Time).
4 Enter
the
information
on
the
keypad
and
Press
Enter.
5
Press
Save.
50
SECTION
3
CONTROLLING
AND
CUSTOMIZING
STAINER
FUNCTIONS
3.3
The
Help
Menu
The
Help
menu
is
a
comprehensive
onscreen
help
function
that
provides
detailed
information
on
the
following
subjects:
Help
Screens
Basic
Operation
• Loading
the
Carousel
• Programming
Number
of
Slides
• Select
Intensity
Setting
• Correct
Reagents
and
Locations
• Selecting
a
Staining
Program
Touchscreen
Calibrating
the
Touchscreen
Cleaning
Clean
Cycle
Using
Help
1
Press
Help
to
access
the
help
function.
2
Select
the
desired
topic.
3
Use
the
direction
arrows
to
navigate.
4
Press
Exit
to
return
to
the
Main
menu.
51
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
WARNING!
Treat
slides
in
accordance
with
good
laboratory
practices
and
local
regulations.
Suggested
Staining
Protocol
• Hub
Pattern
Test
(once
per
day).
• Select
or
verify
desired
stain
program.
• If
slide
tracking
is
enabled,
scan
or
enter
slide
information.
• Load
slides
into
the
carousel.
Use
blocking
slides
if
needed.
• Place
loaded
carousel
into
the
stainer
and
close
the
lid.
• Check
reagent
and
waste
levels.
• If
slide
tracking
is
not
enabled,
enter
the
number
of
slides
on
the
Main
menu.
• Perform
a
stain
cycle.
• Unload
the
carousel.
2 Hold
a
sheet
of
white
paper
towel
near
the
drive
hub,
squarely
facing
the
target
nozzle.
3 Press
the
corresponding
prime
button.
Check
the
pattern.
If
incorrect,
clean
the
nozzle
orifice
with
the
nozzle
brush
provided
in
the
Nozzle
Maintenance
Kit.
If
this
fails
to
correct
the
problem
refer
to
Nozzle
Maintenance
and
Performance
(Section
6).
Figure
7:
Correct
Hub
Pattern
Test
Result
Figure
8:
Incorrect
Hub
Pattern
Test
Result
52
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
Loading
the
Carousel
CAUTION:
Never
load
chipped
or
cracked
slides
into
the
instrument.
Slides
in
poor
condition
may
break
during
the
staining
cycle.
If
a
slide
breaks
in
the
bowl,
see
Cleaning
Broken
Slides
(Section
5.4).
CAUTION:
Keep
small
ferrous
metal
objects
away
from
the
lab
bench.
These
objects
can
be
attracted
to
the
magnets
on
the
bottom
of
the
carousel
and
cause
damage
if
spun
free
during
instrument
operation.
CAUTION:
Load
slides
in
balanced
pairs.
If
staining
an
odd
number
of
slides,
use
a
blank
slide
to
balance
the
carousel.
NOTE:
Load
the
carousel
with
similar
specimens
for
a
similar
level
of
staining.
There
is
no
guarantee
of
staining
performance
when
dissimilar
specimens
are
used.
1
Remove
the
carousel
from
the
bowl
and
place
it
on
a
solid,
level
surface.
2
Remove
the
carousel
lid
by
pressing
the
button
and
lifting
the
lid.
3
If
Slide
Tracking
is
enabled,
press
Load
Slides.
• If
using
the
barcode
reader,
scan
each
specimen
slide
barcode
before
loading
it
into
the
carousel.
Slide
Tracking
must
be
enabled
from
the
System
Setup
menu.
See
Enable
Stain
Slide
Tracking
(Section
3.1).
• If
entering
slide
information
manually,
follow
the
instructions
in
Section
3.2.
4
Insert
the
slides
into
the
carousel
with
the
first
slide
in
position
1.
• Load
slides
in
balanced
pairs
(directly
opposite
one
another)
to
balance
the
carousel.
If
staining
an
odd
number
of
slides,
use
a
blank
slide
to
balance
the
carousel.
• If
there
are
empty
slots
in
the
carousel,
use
blocking
slides
to
prevent
overspray
(see
below).
53
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
Loading
the
Carousel
(continued)
• Load
slides
with
the
labels
toward
the
outside
edge
of
the
carousel.
• Always
load
slides
with
the
specimen
facing
clockwise.
• Always
place
the
first
slide
in
position
1,
the
second
in
position
2,
and
so
on.
NOTE:
A
warning
will
sound
during
the
staining
cycle
if
the
carousel
is
unbalanced.
Figure
9:
Loading
the
Carousel
Labeled
End
Away
From
Center
Hub
Specimen
Side
Figure
10:
Specimen
Placement
on
Slide
5
Replace
the
carousel
lid
by
pressing
the
button
and
lowering
the
lid
over
the
indexing
posts.
6
Release
the
button
and
press
the
lid
handle
until
it
is
firmly
closed
and
locked.
54
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
Using
Blocking
Slides
If
the
carousel
is
not
full,
blank
slides
should
be
used
as
blocking
slides.
Blocking
slides
prevent
overspray
of
reagents
onto
the
specimen
slides.
Overspray
can
cause
slides
to
become
over-‐stained.
• Place
a
blocking
slide
in
front
of
position
1
and
2.
55
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
Performing
a
Stain
Cycle
(continued)
3
If
you
have
created
a
stain
program,
and
it
appears
on
the
display,
proceed
to
Step
4.
If
the
desired
program
does
not
appear
on
the
display,
press
Programs.
Then
select
the
desired
program
and
proceed
to
Step
4.
NOTE:
No
special
collection,
pre-‐treatment,
or
storage
conditions
are
required
for
specimen
types.
Specimens
that
are
normally
manually
hematology
stained
can
be
stained
with
the
Hematology
Stat
Stainer.
Adjust
stain
settings
to
maximize
staining
quality
for
each
specimen
type
(see
Section
3.1).
4
Press
Start.
The
display
shows
the
progress
of
the
program,
and
a
signal
tone
(if
enabled)
indicates
the
end
of
the
cycle.
NOTE:
Use
the
emergency
Stop
button
when
required,
for
example,
if
there
is
abnormal
vibration
or
noise.
This
will
abort
the
staining
cycle.
Unloading
the
Carousel
1 Remove
the
carousel
from
the
bowl
and
place
it
on
a
solid,
level
surface.
2 Remove
the
carousel
lid
by
pressing
the
button
and
lifting
the
lid.
3
Carefully
remove
each
slide
and
observe
the
specimen
using
a
microscope.
56
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
Monitoring
Reagent
and
Waste
Levels
If
enabled,
the
stainer
displays
the
approximate
reagent
and
waste
container
levels
and
other
information.
CAUTION:
You
must
monitor
the
reagent
and
waste
container
levels
on
the
display
(if
enabled)
and
by
direct
inspection
of
the
bottles.
The
monitor
will
show
the
approximate
level
of
each
reagent.
This
can
be
compared
to
the
actual
level
in
the
bottles.
• Never
allow
a
reagent
to
run
dry.
When
the
reagent
level
is
near
empty,
replace
the
reagent
bottle
with
a
new
one
(see
below).
• Never
allow
the
waste
container
level
to
go
above
the
maximum
safety
level.
57
SECTION
4
OPERATING
THE
STAINER
4.1
Operating
Instructions
NOTE:
Access
the
Reagent
Information
menu
by
pressing
the
bottle
icons
on
the
right
side
of
the
Main
menu.
Press
Change
to
scan
or
manually
enter
reagent
information.
NOTE:
Do
not
put
residual
reagent
from
a
used
bottle
into
a
new
bottle.
This
can
lead
to
an
accumulation
of
residue
on
the
slides
and
may
be
a
source
of
contamination.
WARNING!
Reagents
used
in
this
instrument
contain
moderately
hazardous
chemicals
that
require
care
in
handling.
Always
use
appropriate
safety
measures,
including
gloves
and
eye
protection,
when
handling
reagents.
Replacing
a
Reagent
Bottle
1
Remove
the
empty
reagent
bottle
from
the
tray
but
do
not
disconnect
the
dip
tube.
2
Open
the
new
bottle
and
record
the
letter
on
the
cap
for
future
use,
such
as
long-‐term
storage.
3
Open
the
Reagent
Information
menu
by
pressing
the
reagent
bottle
icon
on
the
right
side
of
the
Main
menu.
4
Select
the
desired
reagent
and
press
Change.
5
If
you
are
using
reagent
tracking,
scan
the
barcode
or
manually
enter
the
reagent
REF,
expiration
date
and
lot
number
in
the
Reagent
Information
menu
(Section
3.2).
6
Unscrew
the
cap
and
remove
the
dip
tube
from
the
empty
bottle.
7
Insert
the
dip
tube
into
the
new
reagent
bottle
and
screw
on
the
cap.
8
Place
the
new
bottle
in
the
tray.
58
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.1
Preventive
Maintenance
The
system
provides
a
Preventive
Maintenance
Log
for
tracking
the
most
recent
maintenance
activities.
See
Enable
Preventive
Maintenance
Tracking
(Section
3.1)
and
Using
the
Preventive
Maintenance
Log
in
this
section.
Daily
Maintenance/Quality
Control
(QC)
1
Check
reagent
levels
and
expiration
dates.
4
If
necessary,
use
the
nozzle
brush
from
the
Maintenance
Kit
to
clean
the
nozzle
orifices.
Press
the
individual
bristles
into
the
nozzle
openings.
5
At
the
end
of
the
day:
• Place
an
empty
carousel
in
the
bowl
and
close
the
lid.
Press
Standby/Ready
on
the
front
panel
and
wait
until
the
end
of
the
automatic
cleaning
process.
• Spray
and
wipe
the
bowl,
interior
lid,
and
nozzles
with
70
to
100%
alcohol.
• Wipe
down
the
exterior
of
the
instrument
with
70
to
100%
alcohol.
6
Ensure
the
maintenance
procedures
listed
on
the
Maintenance
Log
have
been
performed,
and
entered
into
the
chart
or
log.
Weekly
Maintenance
1
Wipe
the
carousel
tray
and
lid
using
70
to
100%
alcohol.
2
Flush
the
waste
tube
with
200
to
300
mL
of
alcohol
to
prevent
clogging.
3
Perform
a
Volume
Test
(Section
6.4).
4
Perform
a
Hub
Pattern
Test
(Section
4.1).
5
Manually
clean
the
nozzles
if
necessary.
6
Ensure
the
maintenance
procedures
listed
on
the
Maintenance
Log
have
been
performed,
and
entered
into
the
chart
or
log.
59
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.1
Preventive
Maintenance
Monthly
Maintenance
1
Disassemble
and
manually
clean
all
nozzles.
Refer
to
Nozzle
Disassembly
and
Cleaning
(Section
6.1).
2
Disinfect
any
reagent
bottle
that
is
being
reused
(Section
5.5).
3
Ensure
the
maintenance
procedures
in
the
Preventive
Maintenance
(PM)
Log
have
been
performed,
and
entered
into
the
PM
chart
or
log.
Using
the
Preventive
Maintenance
Log
With
Preventive
Maintenance
Tracking
enabled,
the
PM
Log
provides
a
convenient
and
structured
means
of
recording
important
maintenance
and
QC
functions.
The
system
allows
you
to
set
up
timely
prompts
that
require
response
by
the
user.
See
Enable
Preventive
Maintenance
Tracking
(Section
3.1).
1 From
the
Maintenance
menu,
press
QC/PM
to
open
the
PM
Log.
2
Press
Record
Maintenance.
PM
Task
entry
options:
QC
Slide
Staining
(Drop
Down
Menu)
Not
Completed
Acceptable
Unacceptable
Inconclusive
Disinfect
Reusable
Bottles
Completed
(Select/Deselect)
Drain
Check
Completed
(Select/Deselect)
Manual
Nozzle
Cleaning
Completed
(Select/Deselect)
3
Press
Save
to
record
entries.
60
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.2
Storing
the
Instrument
If
the
instrument
is
inactive
for
more
than
one
week,
you
may
want
to
perform
the
long-‐term
storage
procedure.
This
will
prevent
nozzles
from
clogging
when
the
machine
is
reactivated.
Preparing
for
Long-‐Term
Storage
1
With
the
carousel
removed,
remove
and
clean
the
nozzles.
Store
parts
in
tubes
that
are
labeled
to
indicate
their
correct
position.
2
Unscrew
the
cap
and
remove
the
dip
tube
from
the
bottles.
3
Place
the
end
of
the
dip
tube
in
a
bottle
of
methanol.
4
Flush
at
least
250
mL
of
methanol
through
each
reagent
line
by
priming
all
lines
simultaneously.
Leave
the
alcohol
in
the
line.
CAUTION:
Leave
methanol
in
the
reagent
lines
during
storage.
Allowing
reagent
lines
to
run
dry
can
damage
the
instrument.
CAUTION:
Do
not
subject
the
instrument
to
freezing
temperatures.
Freezing
of
aqueous
fluids
in
the
lines
may
cause
damage
to
the
instrument.
5
Flush
the
bowl
with
water.
6
Return
nozzles
to
their
original
positions.
61
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.3
Replacing
Fuses
WARNING!
To
prevent
the
risk
of
fire,
the
main
fuses
should
only
be
replaced
with
fuses
of
the
same
type
and
rating.
Recurring
fuse
failure
indicates
serious
internal
problems,
if
this
occurs,
contact
Elitech
Group.
1
Power
OFF
the
instrument.
2
Disconnect
the
power
cord
from
the
power
outlet
and
the
rear
panel
of
the
instrument.
3
Open
the
fuse
cover
by
inserting
a
screwdriver
in
the
slot
on
the
right
side
of
the
cover
and
gently
prying
the
cover
open.
4
Remove
the
fuse
holders
to
inspect
the
fuses.
Replace
the
fuses
if
necessary.
6
Push
the
fuse
holders
in.
7
Close
the
fuse
cover.
8
Reconnect
the
main
power
cord
to
the
rear
panel
of
the
instrument
and
to
the
power
outlet.
9
Power
ON
the
instrument.
62
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.4
Cleaning
the
Stainer
and
Carousels
WARNING!
All
cleaning
procedures
should
be
performed
in
a
well-‐ventilated
room
by
authorized
and
trained
personnel
wearing
appropriate
protection
equipment.
1
Clean
the
outside
of
the
instrument
with
70
to
100%
ethanol
or
methanol.
2
Clean
the
carousel
and
lid
with
70
to
100%
ethanol
or
methanol.
NOTE:
Freshly
prepared
(<
24
hours
old)
10%
bleach
solution
can
be
used
as
well.
The
10%
bleach
solution
helps
clean
the
stained
areas.
Cleaning
Liquid
Spills
Remove
any
liquid
spilled
on
the
instrument
immediately
to
avoid
damage
to
the
equipment.
WARNING!
If
potentially
infectious
liquid
is
spilled
on
the
instrument,
the
instrument
must
be
disinfected
in
accordance
with
all
applicable
local
regulations.
Refer
to
Decontaminating
the
Stainer
and
Carousels
(Section
5.5)
for
instructions.
Cleaning
Broken
Slides
You
must
take
stringent
precautions
if
a
slide
breaks
inside
the
instrument
during
a
staining
cycle,
especially
if
the
instrument
has
been
processing
dangerous
pathogens.
Always
use
protective
gloves,
safety
glasses,
and
forceps
when
removing
broken
glass
from
inside
the
instrument.
• Glass
shards
embedded
in
the
walls
of
the
bowl
can
cause
serious
cuts
and
pose
a
risk
of
infection.
• Always
remove
embedded
shards
with
a
scraper
before
attempting
to
remove
loose
glass.
• Use
a
vacuum
or
adhesive
tape
to
pick
up
loose
glass
inside
the
stainer
bowl.
63
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.5
Decontaminating
the
Stainer
and
Carousels
All
parts
of
the
instrument
that
come
into
contact
with
biological
specimens,
patient
specimens,
positive
control
specimens,
or
hazardous
material
must
be
treated
as
potentially
infectious.
Before
the
instrument
is
returned
for
service,
all
outer
surfaces
must
be
decontaminated.
The
operating
authority
must
complete
a
disinfection
declaration,
otherwise
the
instrument
may
be
rejected
by
the
distributor
or
service
center
or
quarantined
by
customs
authorities.
WARNING!
Reagents
used
with
the
instrument
contain
moderately
hazardous
chemicals
that
require
care
in
handling.
Always
use
appropriate
safety
measures
including
gloves
and
eye
protection,
when
handling
reagents.
WARNING!
Authorized
and
trained
personnel
wearing
appropriate
protection
equipment
should
perform
the
decontamination
procedure
in
a
well-‐ventilated
room.
It
is
very
important
to
thoroughly
decontaminate
the
instrument
before
removing
it
from
the
laboratory
or
before
performing
any
technical
service.
This
procedure
may
not
be
effective
against
prions.
WARNING!
Prior
to
decontaminating,
disconnect
the
instrument
from
the
main
power
supply
to
avoid
any
risk
of
fire
or
explosion.
WARNING!
The
decontamination
procedure
and
the
disinfectants
must
comply
with
the
local
applicable
regulations.
Solutions
for
Decontaminating
the
Instrument
The
outer
surfaces
of
the
instrument
should
be
decontaminated
using
a
decontaminating
solution
such
as:
• 70%
ethanol
or
methanol
• Mild
detergent
• 10%
bleach
solution
(<
24
hours
old)
• Decontamination
Solution
(REF:
SS-‐133)
64
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.5
Decontaminating
the
Stainer
and
Carousels
Figure
12:
Lid
Latch
and
Locking
Pin
Hole
Locations
1
2
1
–
Lid
Latch
Hole
2
–
Locking
Pin
Hole
65
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.5
Decontaminating
the
Stainer
and
Carousels
Decontaminating
the
Instrument
(continued)
8
Repeat
the
spray
treatment
of
exterior
surfaces
every
2
or
3
minutes
for
a
total
of
20
minutes.
Do
not
allow
cleaning/decontamination
solutions
to
dry
on
the
instrument
surfaces.
9
Wipe
surfaces
thoroughly
with
a
cloth
soaked
in
water
until
you
have
removed
all
decontamination
solution.
10
Immerse
or
generously
spray
the
carousel
and
lid
with
decontaminating
solution.
Allow
the
solution
to
react
for
20
minutes.
11
Thoroughly
rinse
the
carousel
and
lid
with
deionized
or
distilled
water.
66
SECTION
5
PREVENTIVE
MAINTENANCE
AND
SAFETY
5.6
Shipping
or
Disposing
of
the
Stainer
or
Carousels
Shipping
the
Instrument
or
Carousels
WARNING!
You
must
disinfect
the
instrument
or
carousels
before
returning
it
to
Elitech
Group.
The
operating
authority
must
complete
a
Hazard-‐Free
Certification
form,
otherwise
the
distributor
or
service
center
may
not
accept
the
instrument;
or
customs
authorities
may
hold
it.
CAUTION:
Shipping
the
instrument
or
carousels
without
decontaminating
according
to
these
instructions
is
dangerous
to
service
personnel.
You
will
be
charged
additional
fees
for
decontamination
performed
by
Elitech
Group.
CAUTION:
Ship
the
instrument
or
carousels
in
containers
comparable
to
the
original
packaging.
Hazard-‐Free
Certification
The
operating
authority
must
print
and
complete
the
Hazard-‐Free
Certification
(obtained
from
Elitech
Group
Customer
Service).
Attach
the
certification
form
to
the
top
of
the
instrument
package
before
sending
the
package
to
Elitech
Group.
Disposing
of
the
Instrument
or
Carousels
The
instrument
and
carousels
should
be
completely
decontaminated
and
disposed
of
as
follows:
Under
WEEE
Directive
2002/96/EC,
this
equipment
cannot
be
disposed
of
in
a
normal
landfill.
Instead,
the
equipment
must
be
disposed
of
either
by:
1
Routing
to
an
authorized
local
facility
approved
for
handling
hazardous
materials.
OR
2
Returning
the
equipment
to
Elitech
Group.
67
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.1
Nozzle
Disassembly
and
Cleaning
Nozzle
maintenance
requires
the
Nozzle
Maintenance
Kit
and
prepared
Nozzle
Cleaning
Solution
(diluted
REF:
SS-‐029C).
WARNING!
Always
wear
protective
clothing
and
eye
protection
when
using
Nozzle
Cleaning
Solution
(diluted
REF:
SS-‐029C).
Dispose
of
used
solution
properly.
NOTE:
If
the
compression
screw
cannot
be
easily
loosened,
use
light
penetrating
oil
and
a
5/8-‐in.
wrench
to
loosen
the
nozzle.
Nozzle
Disassembly
1
Remove
the
nozzle
using
the
nozzle
tool
from
the
Nozzle
Maintenance
Kit.
2
Disassemble
the
nozzle.
See
Figure
5:
Nozzle
Diagram
in
Section
1.
3
If
you
will
be
storing
the
nozzle
parts
in
Nozzle
Cleaning
solution
longer
than
one
hour,
remove
the
o-‐ring.
4
Place
the
nozzle
parts
(minus
the
o-‐ring
if
required)
in
a
50
mL
conical
tube
that
has
been
clearly
marked
with
the
correct
nozzle
position.
5
Repeat
Steps
1
through
4
for
each
nozzle.
Nozzle
Cleaning
1
Fill
each
50
mL
tube
with
25
mL
of
prepared
Nozzle
Cleaning
Solution
(REF:
SS-‐029C)
and
cap
the
tube.
2
Gently
invert
the
tube
at
least
ten
times
to
ensure
all
parts
come
in
contact
with
the
cleaning
solution.
3
Place
the
tube
in
the
correctly
marked
position
in
the
provided
tube
stand.
Soak
the
parts
as
long
as
possible.
NOTE:
Soak
nozzle
parts
for
at
least
15
minutes.
Parts
can
be
soaked
in
Nozzle
Cleaning
Solution
overnight.
4
Repeat
steps
1
through
3
for
each
nozzle.
68
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.2
Nozzle
Reassembly
1
Hold
your
thumb
or
a
nozzle
strainer
over
the
end
of
the
tube
to
keep
the
nozzle
parts
in
the
tube.
Discard
the
cleaning
solution
according
to
applicable
statutes.
Swirl
Cone
Grooves
(4
2
Inspect
the
nozzle
parts.
Remove
any
material
in
the
swirl
cone
grooves
by
total,
2
shown)
sliding
a
piece
of
paper
edgewise
along
each
of
the
4
grooves.
3
Thread
the
nozzle
orifice
cleaning
wire
(REF:
AC-‐059)
through
the
back
of
the
disassembled
nozzle
housing.
4
Place
the
nozzle
parts
back
into
the
tube
and
rinse
them
with
water.
5
Rinse
the
parts
again
with
alcohol.
6
Apply
a
small
amount
of
silicone
lubricant
(REF:
AC-‐103)
to
the
compression
screw
threads.
Apply
Silicone
7
Reassemble
the
nozzle
by
placing
the
compression
screw
on
the
nozzle
Lubricant
Here
disassembly
tool,
then
inserting
the
swirl
cone
into
the
compression
screw.
Nozzle
NOTE:
Hold
all
the
parts
in
a
vertical
position
during
reassembly.
Disassembly
Tool
8
Reinstall
the
nozzle
housing
over
the
swirl
cone
and
compression
screw.
9
Replace
the
o-‐ring
if
it
was
removed
earlier.
10
Return
the
assembled
nozzle
to
its
original
position
in
the
instrument.
D
C
B
A
11
Repeat
Steps
1
through
10
for
each
nozzle.
12
Perform
a
Hub
Pattern
Test
(Section
4.1).
13
Perform
a
Volume
Test
(Section
6.4).
NOTE:
You
must
perform
the
Hub
Pattern
Test
and
Volume
Test
before
operating
the
instrument.
If
the
results
are
incorrect,
manually
prime
the
instrument.
69
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.3
Manual
Priming
1
Remove
the
carousel
from
the
bowl.
2
Remove
the
nozzle
connected
to
the
line
to
be
manually
primed.
3
Insert
the
priming
tool
nozzle
adapter
(included
in
the
Nozzle
Maintenance
Kit)
into
the
nozzle
holder
and
turn
it
clockwise
to
install
the
adapter
into
the
holder.
4
Withdraw
the
priming
tool
plunger
halfway
to
create
a
vacuum.
Hold
the
plunger
in
that
position.
5
Press
Maintenance
from
the
Main
menu.
6
Press
Volume
Test.
7
Press
the
desired
prime
button
to
start
the
reagent
pump.
8
Run
the
reagent
into
the
tube
until
the
fluid
is
free
of
bubbles,
then
press
Stop.
WARNING!
Do
not
pull
the
plunger
completely
out
of
the
priming
tool.
Pulling
the
plunger
out
of
the
tool
may
result
in
splashing
or
spraying
of
reagents.
Do
not
push
the
plunger
in
while
it
is
connected
to
the
nozzle
holder.
9
Turn
the
nozzle
adapter
counterclockwise
to
remove
it
from
the
nozzle
holder.
10
Discard
the
collected
fluid
into
the
stainer
bowl.
11
Reinstall
the
nozzle.
12
Perform
a
Hub
Pattern
Test.
13
Perform
a
Volume
Test.
70
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.4
Performing
the
Volume
Test
The
Volume
Test
requires
the
Nozzle
Maintenance
Kit.
NOTE:
The
Volume
Test
must
be
performed
weekly.
1
From
the
Maintenance
menu,
select
Volume
Test.
2
Hold
a
Volume
Test
tube
(small
tube)
to
cover
the
selected
nozzle.
3
Press
the
corresponding
reagent
prime
button
to
collect
the
reagent.
NOTE:
With
QC/Maintenance
tracking
enabled,
enter
the
measured
volume
on
the
keypad
and
press
ENTER.
With
QC/Maintenance
Tracking
disabled,
the
menu
returns
to
the
Maintenance
menu.
4
Remove
and
cap
the
tube.
5
Record
the
nozzle
position
on
the
tube
and
place
the
tube
in
the
appropriate
position
in
the
tube
stand.
6
Repeat
Steps
2
through
5
for
each
nozzle.
7
Compare
collected
nozzle
volumes
with
the
following
table.
Table
12:
Volume
Test
Tolerances
Nozzle/Reagent
Line
Minimum
Maximum
A,
B,
C
9.0
mL
11.0
mL
D
8.5
mL
10.5
mL
NOTE:
The
stainer
normally
functions
correctly
if
nozzle
volumes
are
slightly
higher
or
lower
than
the
specified
range.
It
is
important
that
the
B
and
C
nozzle
volumes
are
similar,
(typically
within
1
mL).
Spray
volumes
<
7.5
mL
or
>
13.0
mL
indicate
serious
problems
with
the
nozzles
or
reagent
delivery
lines.
• If
the
volume
is
within
the
tolerance
range,
go
to
Step
9.
• If
the
volume
is
outside
the
tolerance
range:
a.
Clear
the
nozzle
orifice
with
the
nozzle
brush
found
in
the
maintenance
kit.
b.
If
necessary,
remove
the
nozzle
and
perform
the
Nozzle
Cleaning
procedure
(Section
6.1).
c.
If
the
problem
persists,
replace
the
nozzle.
NOTE:
If
the
problem
persists
after
you
have
replaced
the
nozzle,
contact
Elitech
Group.
71
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.4
Performing
the
Volume
Test
8
Prepare
the
Maintenance
Kit
for
future
use:
• Empty
the
contents
of
the
tubes
into
the
stainer
bowl.
• Rinse
the
tubes
with
water.
• Put
the
tubes
back
into
their
original
place
in
the
Maintenance
Kit
or
tube
stand.
9
Press
Back
twice
to
return
to
the
Main
menu.
72
SECTION
6
NOZZLE
MAINTENANCE
AND
PERFORMANCE
6.5
Performing
the
Slide
Pattern
Test
This
test
can
differentiate
poor
staining
results
from
sample
preparation
problems,
or
nozzle
obstructions.
Perform
the
Slide
Pattern
test
when
a
Hub
Pattern
test
produces
a
normal
result,
but
staining
is
still
inadequate.
1
Place
a
1
x
3
inch
(2.5
x
7.6
cm)
piece
of
paper
(REF:
RP-‐500)
in
positions
1
and
2
of
the
carousel,
with
a
blocking
slide
in
front
of
positions
1
and
2.
2
Load
the
carousel
into
the
stainer
and
close
the
lid.
3
From
the
Main
menu,
press
Maintenance.
4
Press
Pattern
Test.
5
Press
the
corresponding
prime
button
for
the
reagent
line
to
be
tested.
6
Remove
the
paper
slides.
7
Repeat
Steps
1
through
6
for
each
reagent
line.
8
Examine
the
paper
slides
for
each
reagent.
The
pattern
on
the
slide
should
be
uniform,
without
any
continuous
lines
or
streaks.
73
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
The
following
table
is
to
help
identify
and
solve
routine
problems
with
the
stainer.
More
difficult
problems
may
require
technical
service.
Contact
your
Elitech
Group
representative
for
assistance.
WARNING!
Due
to
the
electrical
shock
hazard,
do
not
open
this
instrument
or
attempt
internal
repairs.
Refer
servicing
to
qualified
service
personnel.
Contact
your
dealer
or
Elitech
Group
Service.
Table
13:
General
Troubleshooting
and
Diagnosis
Problem
Solution
There
is
no
power
to
the
stainer
when
the
power
switch
Check
the
facility
outlet
and
the
power
cord
connection.
is
turned
ON.
Check
the
fuses.
Refer
to
the
Replacing
Fuses
procedure.
CAUTION:
Fuse
failure
may
indicate
a
serious
internal
problem.
Strange
information
shows
on
the
display,
and/or
erratic
Switch
the
power
OFF,
wait
10
to
20
seconds,
then
stainer
operation.
switch
power
ON
again.
If
problem
recurs,
install
a
computer-‐type
surge
suppressor
to
protect
the
instrument
from
power
line
transients.
If
possible,
connect
the
stainer
to
a
power
circuit
that
is
not
shared
by
centrifuges,
refrigerators,
air
conditioners,
or
other
motorized
equipment.
If
the
above
steps
do
not
solve
the
problem,
consult
the
Aerospray
Service
Manual,
or
contact
your
dealer
or
Elitech
Group
for
assistance.
A
reagent
line
will
not
prime
when
power
is
ON
and
you
Follow
the
procedures
in
Section
6.3
for
priming
reagent
press
the
prime
button.
pumps.
A
reagent
line
will
not
prime,
even
with
the
priming
tool
Press
the
priming
button
and
listen
carefully
for
the
(Section
6.3).
sound
of
the
pump.
If
you
can
hear
the
pump,
try
the
priming
tool
again.
If
the
problem
is
not
solved
or
if
you
cannot
hear
the
pump
there
may
be
an
internal
problem.
Contact
your
dealer
or
Elitech
Group
for
assistance.
Stainer
bowl
fills
with
reagent
after
use.
A
small
puddle
of
stain
around
the
drain
inlet
or
the
bottom
of
the
bowl
is
normal.
If
the
bowl
is
filling
with
a
large
quantity
of
stain,
check
the
external
drain
tube
for
blockage.
Make
sure
the
drain
tube
is
properly
connected
and
running
continuously
down
toward
the
lab
drain
or
vented
waste
container,
with
no
loops,
rises,
or
obstructions.
Make
sure
the
end
of
the
tube
is
not
submerged.
This
can
prevent
proper
drainage.
The
internal
drain
may
need
to
be
cleaned
or
replaced.
See
the
Aerospray
Service
Manual,
or
contact
your
dealer
or
Elitech
Group
Service.
74
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
Table
13:
General
Troubleshooting
and
Diagnosis
(continued)
Problem
Solution
Stain
is
leaking
onto
the
counter.
Check
all
external
reagent
lines
for
visible
signs
of
cracks
or
loose
fittings.
Make
sure
the
drain
outlet
is
not
blocked.
Make
sure
the
drain
tube
is
securely
attached
to
the
drain
port
and
that
the
tubing
is
not
cracked
or
deformed.
Reagent
leaks
may
indicate
an
internal
problem
(see
Section
7.3).
See
the
Aerospray
Service
Manual,
or
contact
your
dealer
or
Elitech
Group
for
further
assistance.
Error
messages
on
the
screen.
If
the
display
shows
Lid
Not
Shut:
Verify
that
the
lid
is
fully
closed
and
latched.
If
the
Lid
Not
Shut
indication
remains,
contact
Elitech
Group
for
assistance.
If
the
display
shows
Wrong
Rotor
after
pressing
Start:
Make
sure
the
slide
carousel
is
properly
loaded
on
the
drive
hub.
In
staining
mode,
the
instrument
detects
whether
the
staining
carousel
is
present
before
proceeding.
In
cytocentrifuge
mode,
the
instrument
will
stop
if
it
senses
the
staining
carousel.
After
verifying
the
carousel
is
correctly
loaded,
press
Start.
If
the
display
still
shows
Wrong
Rotor,
there
may
be
an
internal
problem.
Check
for
missing
carousel
magnets.
The
microprocessor
monitors
carousel
rotation
during
a
staining
cycle.
The
display
shows
an
error
message
if
the
rotation
is
not
within
the
specified
range.
If
the
display
shows
Motor
Drive
Error:
Check
the
stainer
bowl
for
interference:
Turn
the
hub
or
carousel
by
hand;
it
should
turn
freely.
Drive
motor
or
electronic
component
malfunctions
require
servicing
of
internal
components.
Contact
your
dealer
or
Elitech
Group
for
assistance.
If
the
display
shows
Rotor
Imbalance,
make
certain
the
Cytopro
rotor
is
balanced,
or
the
staining
carousel
is
seated
correctly
on
the
hub.
See
Electronic
Failure
later
in
this
table.
75
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
Table
13:
General
Troubleshooting
and
Diagnosis
(continued)
Problem
Solution
The
stainer
fails
to
spray
reagent
during
a
staining
cycle
To
allow
programmed
staining
of
partial
loads,
the
and/or
continues
to
run
after
the
cycle
should
be
stainer
monitors
the
position
of
the
carousel
as
it
rotates
complete.
in
the
bowl.
In
normal
operation,
stain
is
sprayed
only
in
the
correct
position.
This
causes
the
actual
cycle
time
to
vary,
depending
on
the
position
of
the
carousel
at
the
beginning
of
the
cycle.
However,
if
the
cycle
continues
for
an
abnormally
long
period,
or
if
the
bar
graph
and
percentage
complete
icon
do
not
change
after
1
minute,
it
may
indicate
an
electronic
problem
or
an
internal
problem.
To
determine
this,
press
Stop.
If
the
cycle
stops:
this
indicates
a
problem
with
the
carousel
position
sensor.
Consult
the
Aerospray
Service
Manual,
or
contact
your
dealer
or
Elitech
Group
for
assistance.
If
the
cycle
continues:
this
indicates
an
electronic
problem
(see
below).
Abnormal
staining
on
entire
surface
of
all
slides.
Check
the
reagent
level
on
the
display
and/or
in
the
reagent
bottles.
Make
sure
the
external
reagent
dip
tubes
are
securely
attached
to
each
bottle
(Section
2.1).
Open
the
lid
and
verify
that
each
reagent
pump
is
primed,
by
pressing
the
corresponding
prime
button.
The
nozzle
should
immediately
spray
a
fine
mist
of
reagent.
There
should
be
no
sputtering
or
hissing
sounds
indicating
air
in
the
reagent
lines.
Watch
the
external
tubes
for
air
bubbles.
Air
bubbles
indicate
inadequate
priming
or
possibly
an
air
or
reagent
leak
in
the
system.
Air
in
any
reagent
line
will
cause
poor
staining.
Check
nozzle
performance
using
the
Slide
Pattern
(Section
6.5)
and
Volume
Tests
(Section
6.4).
If
necessary,
clean
nozzle(s)
using
the
procedures
in
Section
6.1.
Verify
that
each
reagent
dip
tube
vent
hole
is
clear.
76
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
Table
13:
General
Troubleshooting
and
Diagnosis
(continued)
Problem
Solution
Abnormal
staining
on
entire
surface
of
all
slides
When
staining
a
full
carousel
(7
or
more
slides)
make
(continued).
certain
you
have
not
programmed
the
stainer
for
fewer
slides.
If
staining
a
partial
load,
load
the
slides
in
the
correct
positions
as
indicated
by
the
markings
on
the
carousel
(see
Section
4.1).
Abnormal
staining
on
entire
surface
of
some
slides,
while
Make
certain
that
all
position
magnets
are
still
attached
other
slides
from
the
same
carousel
appear
normal.
to
the
bottom
of
the
carousel.
Make
certain
you
have
not
programmed
the
stainer
for
fewer
slides
than
you
have
loaded.
If
you
program
the
stainer
for
a
partial
load,
load
the
slides
in
the
correct
positions
as
indicated
by
the
markings
on
the
carousel
(see
Section
4.1).
Verify
that
each
reagent
pump
is
primed
by
opening
the
lid
and
pressing
the
corresponding
prime
button.
The
nozzle
should
immediately
spray
a
fine
mist
of
reagent.
There
should
be
no
sputtering
or
hissing
sounds
to
indicate
the
presence
of
air
in
the
reagent
lines
(see
Section
7.3).
Streaks
or
bands
of
discoloration
on
one
or
more
slides.
Check
the
level
of
the
Rinse
(Reagent
A)
bottle.
Check
Rinse
(Reagent
A)
spray
volume
according
to
Section
6.4.
Check
nozzle
spray
pattern
according
to
the
procedures
in
Sections
4.1
and
6.5.
This
type
of
discoloration
is
usually
caused
by
debris
or
reagent
precipitate
clogging
the
spray
nozzle
orifice.
Clean
any
nozzle
that
exhibits
a
poor
spray
pattern.
Cells
are
washing
off
slides.
Check
methanol
or
Aerofix
(Reagent
D)
level
in
bottle.
Check
Reagent
D
nozzle
spray
volume
and
slide
pattern
(see
Section
6.4
and
Section
6.5).
77
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
Table
13:
General
Troubleshooting
and
Diagnosis
(continued)
Problem
Solution
Cells
are
washing
off
slides
(continued).
Check
the
methanol
or
Aerofix
(Reagent
D)
hub
pattern
and
spray
volume
(see
Sections
4.1
and
6.5).
Make
certain
the
blood
smears
are
completely
dry
and
not
too
thick
(see
Section
7.2).
Increase
the
fixation
setting
to
High.
Call
Elitech
Group
for
information
on
slide
quality.
Refractile
artifacts
are
observed
in
erythrocytes.
Use
Aerofix
(SS-‐048
or
SS-‐148)
fixative.
Refer
to
Section
7.2.
High
levels
of
precipitates
observed.
Discard
leftover
thiazin
stain
(Reagent
B)
in
the
used
bottle
when
changing
to
a
fresh
bottle
of
stain.
While
transferring
remaining
stain
saves
stain,
it
eventually
concentrates
the
precipitates
to
the
point
they
appear
on
the
slide.
Staining
intensity
is
too
weak.
Granulation
is
weak.
Increase
the
intensity
setting.
Refractile
artifacts
are
observed
in
erythrocytes
when
Mix
1
part
Aerofix
(SS-‐048)
with
3
parts
Basofix
predip
using
Basofix
predip
(SS-‐049P).
(SS-‐049P),
or
add
5
mL
Aerofix
concentrate
(SS-‐148)
to
500
mL
Basofix
predip
(SS-‐049P).
Increased
false-‐positives
for
Howell-‐Jolly
bodies
or
Disinfect
reagent
lines
using
instructions
in
Section
7.2.
amorphous
precipitates
are
observed.
Discard
leftover
thiazin
stain
(Reagent
B)
in
the
used
bottle
when
changing
to
a
fresh
bottle
of
stain.
While
transferring
remaining
stain
saves
stain,
it
eventually
concentrates
the
precipitates
to
the
point
where
they
can
be
seen
on
the
slide.
78
SECTION
7
SOLVING
PROBLEMS
7.1
Troubleshooting
Table
13:
General
Troubleshooting
and
Diagnosis
(continued)
Problem
Solution
Electronic
Failure
• An
electronic
failure
would
appear
as
an
obvious
malfunction
such
as
a
scrambled
or
totally
inoperative
display
panel.
•
• Transient
voltages
coming
through
the
power
lines
may
cause
the
stainer
to
“lose
its
place.”
1
If
this
occurs,
switch
the
main
power
OFF
for
10-‐20
seconds
and
then
back
ON
to
reset
the
instrument.
2
If
the
problem
recurs,
install
a
computer-‐type
surge
protector
to
isolate
the
instrument.
3
If
possible,
connect
the
stainer
to
a
power
circuit
not
shared
by
centrifuges,
refrigerators,
air
conditioners,
or
other
motorized
equipment.
For
more
obscure
electronic
problems,
monitor
the
stainer
through
a
complete
staining
cycle
to
determine
if
the
operating
sequence
is
correct.
Do
this
by
running
the
stainer
while
watching
the
display
and
listening
to
the
pumps.
Ensure
that
each
event
occurs
according
to
the
operating
sequence,
shown
in
Table
3
in
Section
1.
If
the
problem
recurs,
contact
your
dealer
or
Elitech
Group
for
assistance.
79
SECTION
7
SOLVING
PROBLEMS
7.2
Abnormal
Staining
Results
Smear
Separation
or
Tearing
Losing
blood
cells
from
the
slide
surface
during
a
staining
cycle
can
happen
for
several
reasons:
Improper
Alcohol
Fixation
If
the
alcohol-‐fix
phase
of
the
cycle
fails,
there
will
be
cell
loss
when
stain
is
applied.
Verify
that
the
alcohol
nozzle
is
operating
properly
with
normal
spray
pattern
and
spray
volume
(Sections
6.4,
6.5).
To
identify
the
problem
as
a
fixation
failure,
fix
some
specimen
slides
in
methanol
prior
to
loading
into
the
Wet
Smears
carousel.
Select
a
fixation
setting
that
increases
fixation
(7,
8,
or
9).
Wet
Smears
Loading
wet
blood
smears
into
the
carousel
may
contribute
to
smear
tearing.
Allow
smears
to
dry
for
several
minutes
before
beginning
a
stain
cycle.
Thick
Smears
Smears
that
are
too
thick
can
cause
smear
separation.
If
the
smear
has
a
thick
“wedge”
cross
section,
cells
can
be
torn
away
from
the
slide
surface
in
the
thicker
regions,
while
the
thinner
“monolayered”
regions
remain
usable.
Severe
tearing
that
starts
in
the
thick
region
may
also
extend
into
the
monolayer.
Dirty
Slides
Dirty
slides
are
a
major
source
of
cell
loss
during
staining.
We
strongly
recommend
that
you
use
new,
premium
quality
slides.
Even
with
premium
quality
slides,
random
loss
may
occur
due
to
inadequate
slide
surface
quality.
When
the
stainer
is
functioning
properly,
slide
quality
accounts
for
most,
if
not
all,
cell
loss
problems.
Red
Cell
Artifacts
Refractile
bodies
or
inclusions
(seen
as
dark
ringed
objects
on
red
cells)
are
thought
to
be
caused
by
the
presence
of
water
during
fixation.
Often
called
“water
spotting,”
this
effect
is
common
in
Romanowsky
staining.
Methanol
used
for
fixation
must
be
anhydrous
(0.5%
or
less
water).
Methanol
that
is
exposed
to
the
atmosphere
will
absorb
a
considerable
amount
of
water,
particularly
when
the
relative
humidity
is
high.
The
first
line
of
defense
to
avoid
water
spotting
is
to
make
certain
the
methanol
used
for
fixing
smears
is
free
of
water.
If
in
doubt,
replace
it
with
reagent
known
to
be
anhydrous.
80
SECTION
7
SOLVING
PROBLEMS
7.2
Abnormal
Staining
Results
Red
Cell
Artifacts
(continued)
When
the
relative
humidity
is
above
60
percent,
water
spotting
may
be
seen
even
when
using
anhydrous
methanol
for
fixation.
This
occurs
because
of
moisture
accumulation
and
condensation
inside
the
carousel
chamber.
Relative
humidity
within
the
chamber
will
tend
to
build
up
with
repeated
use,
leading
to
further
problems.
If
water
spotting
occurs:
Use
Aerofix®
Water
spotting
can
be
eliminated
by
using
Aerofix.
Aerofix
is
available
as
an
additive
(REF:
SS-‐148,
135
mL
bottle,
enough
for
nine
500
mL
bottles
of
methanol)
or
premixed
with
methanol
(REF:
SS-‐048,
500
mL
bottle).
• Pre-‐fix
Slides
Water
spotting
can
usually
be
eliminated
by
pre-‐fixing
slides
in
anhydrous
methanol
before
loading
into
the
carousel.
Allow
the
slides
to
dry
for
2
to
3
minutes
before
staining.
• Wipe
Carousel
Chamber
Dry
Between
cycles,
leave
the
stainer
lid
open
to
dry
out
the
bowl.
If
necessary,
wipe
the
interior
of
the
carousel
chamber
dry
to
reduce
the
relative
humidity.
Or,
irrigate
the
chamber
with
methanol
from
a
wash
bottle.
False-‐Positive
Howell-‐Jolly
Bodies
or
Amorphous
Precipitates
Artifactual
Howell-‐Jolly
bodies
can
be
caused
by
precipitate
build
up
from
transferring
the
residual
thiazin
stain
(Reagent
B)
to
the
fresh
bottle
when
reloading
reagents.
Such
artifacts
may
also
indicate
microbial
contamination
of
the
reagent
lines
and
pumps.
If
such
artifacts
are
encountered,
and
if
you
suspect
microbial
contamination,
use
the
following
solution
to
disinfect
the
reagent
lines
and
pumps.
Replace
precipitated
stain
with
fresh
stains
after
flushing
the
B
line
with
the
disinfecting
solution.
Disinfectant
Solution:
10%
freshly
prepared
dilution
of
household
bleach
(5.25%
Sodium
Hypochlorite).
Dilution:
100
mL
household
bleach
+
900
mL
deionized
water
=
1000
mL
81
SECTION
7
SOLVING
PROBLEMS
7.2
Abnormal
Staining
Results
Procedure:
NOTE:
Do
not
remove
the
nozzles.
1
Pump
150
mL
of
prepared
solution
through
each
affected
reagent
line
(A,
B
and
C).
Wait
20
minutes.
2
Pump
another
100
mL
of
disinfectant
solution
through
each
affected
reagent
line
and
let
sit
for
20
minutes.
3
Flush
each
affected
reagent
line
with
150
mL
of
deionized
water.
4
Reprime
the
stainer
with
fresh
reagents
(see
Section
2.2)
and
clean
nozzles
according
to
instructions
in
Section
6.1
and
6.2.
5
Perform
a
Volume
Test
according
to
instructions
in
Section
6.4
and
record
on
PM
Chart.
CAUTION!
To
avoid
damage
to
the
instrument,
do
not
leave
disinfectant
solution
in
the
reagent
lines
longer
than
recommended.
This
procedure
should
only
be
performed
on
an
“as
needed”
basis.
82
SECTION
7
SOLVING
PROBLEMS
7.3
Instrument
Malfunction
Air
or
Reagent
Leaks
Repriming
the
instrument
is
usually
unnecessary
unless
a
reagent
bottle
runs
completely
dry.
An
air
leak
is
usually
to
blame
if
a
smooth
and
continuous
liquid
spray
fails
to
come
from
the
nozzles.
Carefully
inspect
all
components
in
the
external
reagent
delivery
lines.
Look
for
loose
connections,
cracks,
or
breaks
that
might
allow
air
to
be
drawn
in
when
the
pump
operates.
Replace
any
defective
part
or
assembly.
An
internal
leak
may
cause
fluid
to
leak
from
the
line
when
the
pump
is
not
running.
If
an
abnormal
liquid
spray
still
occurs
after
all
the
external
reagent
delivery
line
components
have
been
verified,
the
instrument
may
require
service.
Contact
your
dealer
or
Elitech
Group
for
assistance.
Reagent
Delivery
L ines
A
reagent
line
leak
between
the
pump
outlet
and
the
nozzle
will
cause
fluid
to
leak
into
the
interior
of
the
stainer
housing
and
ultimately
onto
the
counter.
If
you
observe
this,
the
instrument
will
require
service.
Contact
your
dealer
or
Elitech
Group
for
assistance.
WARNING!
A
break
or
malfunction
in
the
reagent
delivery
system
can
potentially
release
up
to
500
mL
of
highly
flammable
anhydrous
alcohol
in
and
around
the
instrument.
If
this
occurs,
carefully
shut
off
the
power
to
the
instrument
and
consult
the
SDS
for
information
in
handling
alcohol
spills.
Do
not
use
the
instrument
again
until
any
leaks
are
repaired.
WARNING!
Electrical
shock
hazard⎯ do
not
open
this
instrument
or
attempt
internal
repairs.
Refer
servicing
to
qualified
service
personnel.
Contact
your
dealer
or
Elitech
Group
Service.
83
SECTION
7
SOLVING
PROBLEMS
7.3
Instrument
Malfunction
Reagent
Level
Detect
System
Errors
Reagent
A-‐D
Not
Calibrated
During
the
second
part
of
calibration,
if
no
bottles
are
detected,
the
display
shows
an
error
message.
Calibrate
again,
making
sure
that
the
reagent
bottles
are
inserted
in
those
tray
positions
that
have
been
enabled
in
the
level
detection
system.
Level
Detection
(LD)
Unstable
If
movement
was
detected
on
bottles
while
calibrating/zeroing,
the
display
shows
an
error
message.
NOTE:
While
zeroing
or
calibrating,
do
not
bump
the
instrument
or
lab
bench.
Ensure
that
no
nearby
equipment
vibrations
can
be
transmitted
to
the
stainer.
Calibrating
the
Reagent
Level
Detect
System
If
the
Reagent
Level
Detect
System
is
reporting
incorrectly
and
zeroing
(Section
2.2)
does
not
correct
the
problem,
calibrate
the
system
as
follows:
1
Press
System
Information
from
the
Main
menu.
2
Press
System
Setup.
3
Press
Level
Detect.
4
Press
Calibrate.
Follow
the
display
prompts.
84
SECTION
7
SOLVING
PROBLEMS
7.3
Instrument
Malfunction
Calibrating
the
Reagent
Level
Detect
System
(continued)
5
Remove
all
reagent
bottles
and
press
Start.
NOTE:
Any
vibrations
or
bumps
to
the
instrument
or
lab
bench
can
cause
inaccuracies
in
zeroing
or
calibration.
NOTE:
Calibration
requires
full,
unopened
(caps
and
seals
in
place)
500
mL
bottles
of
reagent,
placed
in
the
correct
tray
positions
(due
to
different
densities
of
each
reagent
type).
6
Place
the
correct
reagent
bottles
in
all
enabled
positions,
and
press
Start.
NOTE:
The
calibration
function
ignores
any
disabled
reagent
line.
7
Press
OK.
Press
Back
twice
to
return
to
the
Main
menu.
8
Return
the
reagent
bottles
to
the
tray
as
indicated
in
Section
2.1
to
prepare
for
staining.
NOTE:
For
accurate
reagent
level
detection
and
calibration,
dip
tubes
must
follow
their
pre-‐formed
coiled
shapes.
85
SECTION
7
SOLVING
PROBLEMS
7.4
Calibrating
the
Touchscreen
1
Press
and
hold
Standby/Ready
for
5
seconds.
A
calibration
screen
with
a
target
appears.
2
Press
the
center
of
the
target
with
a
finger,
stylus,
or
similar
tool.
Another
target
will
appear
in
a
different
location.
3
Continue
to
press
the
center
of
the
targets
until
you
have
pressed
all
the
targets
(five
total).
After
the
fifth
target
is
pressed,
the
instrument
will
save
the
touch
screen
calibration
and
return
to
the
Main
menu.
86
SECTION
7
SOLVING
PROBLEMS
7.5
Service
Information
Elitech
Group’s
Service
Department
will
help
you
resolve
any
questions
about
the
operation
or
performance
of
your
Aerospray
Stainer/Cytocentrifuge.
Customers
in
the
United
States
should
contact
us
by
telephone.
Outside
the
U.S.,
our
authorized
dealers
offer
full
local
service
and
support.
ELITechGroup
Inc.
370
West
1700
South
Logan,
Utah
84321-‐8212
USA
Telephone:
800
453
2725
(United
States
&
Canada)
(+1)
435
752
6011
(International
calls)
Fax:
(+1)
435
752
4127
Email:
[email protected]
(Service)
[email protected]
(Sales)
Web
Page:
www.elitechgroup.com
European
Authorized
Representative:
Medical
Technology
Promedt
Consulting
Altenhofstr.
80
D-‐66386
St.
Ingbert
Germany
Telephone:
+49(0)68
94-‐58
10
20
Fax:
+49(0)68
94-‐58
10
21
Email:
info@mt-‐procons.com
87
SECTION
8
CYTOPRO®
CYTOCENTRIFUGE
8.1
Cytopro
Cytocentrifuge
Information
Functional
Description
The
Cytopro
Cytocentrifuge
rotor
allows
rapid
sedimentation
of
specimen
cells
onto
microscope
slides
for
staining
or
other
purposes.
Up
to
eight
disposable/reusable
sample
chamber
assemblies
with
absorbent
pads
and
glass
microscope
slides
can
be
loaded
into
the
Cytocentrifuge
rotor.
Cytocentrifuge
and
staining
functions
are
independent
of
one
another.
The
Cytopro
rotor
reduces
cell
loss
during
collection
and
prevents
accidental
damage
to
the
collected
specimen.
The
rotor
is
sealed
to
control
aerosol
release
during
cytocentrifugation.
See
the
Cytopro
Rotor
Applications
Manual
(Aerospray
Models
7xx2)
(RP-‐517)
for
complete
information.
Key
Features
Adding
the
Cytopro
Cytocentrifuge
rotor
transforms
the
stainer
into
a
standard
cytocentrifuge
with:
• Single,
Dual,
and
Cytopro
Magnum
chambers
• Reusable
or
disposable
chambers
(single
and
dual)
• Eight
slides
and
chambers
• User-‐programmable
memory
locations
for
settings
(speed,
acceleration
rate,
and
time)
• Easy
switching
between
staining
and
cytocentrifuge
modes
• Autoclavable
rotor
NOTE:
Pressing
Cyto
brings
up
the
Cytocentrifuge
mode.
Pressing
Back
returns
to
stain
mode.
WARNING!
The
Cytopro
rotor
lid,
rotor
gaskets
and
related
components
are
intended
to
be
part
of
biosafety
system
as
specified
in
international
and
national
biosafety
guidelines.
They
cannot
be
relied
on
as
the
only
means
of
safeguarding
workers
and
the
environment
when
handling
pathogenic
microorganisms.
Intended
Use
The
Cytopro
Cytocentrifuge
rotor
is
an
in
vitro
diagnostic
medical
device
for
professional
use
only.
It
is
an
accessory
for
fixing
biological
cell
suspensions
on
glass
microscope
slides
for
cytological
examination.
The
Cytocentrifuge
rotor
can
be
used
with
the
following
cell
suspensions:
• Bronchoalveolar
liquid
(BAL)
• Cerebrospinal
fluid
(CSF)
• Urine
• Synovial
fluid
• Others
88
APPENDIX
A
Critical
Reagent
Components
The
following
information
identifies
the
critical
chemicals
of
each
reagent
used
in
this
instrument.
Reagent(s)
Critical
Components
SS-‐035A,
SS-‐035A-‐EU,
SS-‐035AG,
or
SS-‐135/149A
(as
<1%
Potassium
Phosphate
diluted)
Hematology
Reagent
A
Rinse
contains:
<1%
Sodium
Phosphate
<1%
Eosin
<1%
Benzyl
Alcohol
SS-‐035/049B,
SS-‐035/049B-‐EU,
SS-‐035/049BG,
or
SS-‐ <1%
Potassium
Phosphate
135/149B
(as
diluted)
Hematology
Reagent
B
Thiazin
<1%
Sodium
Phosphate
Stain
contains:
<1%
Azure
B
<1%
Methylene
Blue
<1%
Benzyl
Alcohol
<0.1%
Triton
X-‐100
SS-‐035C,
SS-‐035C-‐EU,
SS-‐035CG,
or
SS-‐135/149C
(as
<1%
Potassium
Phosphate
diluted)
Hematology
Reagent
C
Eosin
Stain
contains:
<1%
Sodium
Phosphate
<1%
Eosin
<0.2%
Formaldehyde
<0.1%
Triton
X-‐100
SS-‐048,
SS-‐048-‐EU,
or
SS-‐148
(as
diluted)
Aerofix
>95%
Methyl
Alcohol
Fixative
High
Humidity
contains:
<4%
Ethylene
Glycol
<2%
Polyvinylpyrrolidone
(PVP)
<1%
Azure
B
SS-‐049A,
SS-‐049A-‐EU,
or
SS-‐049AG
or
SS-‐135/149A
(as
<3%
Ethyl
Alcohol
diluted)
Basofix
Reagent
A
contains:
<3%
Methyl
Alcohol
<1%
Potassium
Phosphate
<1%
Sodium
Phosphate
<1%
Eosin
<
0.2%
Formalin
SS-‐049C,
SS-‐049C-‐EU,
or
SS-‐049CG
or
SS-‐135/149C
(as
<10%
Diethylene
Glycol
diluted)
Reagent
C
Basofix
Eosin
Stain
contains:
<5%
Ethyl
Alcohol
<5%
Methyl
Alcohol
<1%
Potassium
Phosphate
<1%
Sodium
Phosphate
<1%
Eosin
Y
SS-‐049P,
SS-‐049P-‐EU,
or
SS-‐149P
(as
diluted)
Basofix
<99%
Methyl
Alcohol
Pre-‐dip
Fixative
contains:
<1%
Polyvinylpyrrolidone
<1%
Azure
B
SS-‐135/149A,
Reagent
A
Rinse
Reagent
Concentrate
<1%
Potassium
Phosphate
contains:
<1%
Sodium
Phosphate
<0.2%
Formalin
<1%
Eosin
SS-‐135/149B,
Reagent
B
Thiazin
Stain
Concentrate
<10%
Potassium
Phosphate
contains:
<10%
Sodium
Phosphate
<2%
Azure
B
<1%
Methylene
Blue
89
APPENDIX
A
Critical
Reagent
Components
Reagent(s)
Critical
Components
SS-‐135/149C,
Reagent
C
Eosin
Stain
Concentrate
<10%
Potassium
Phosphate
contains:
<10%
Sodium
Phosphate
<0.2%
Formaldehyde
<3%
Eosin
SS-‐135S
Surfactant
contains:
<70%
Germall
II
(Diazolidinyl
Urea)
<5%
Triton
X-‐100
SS-‐148,
Aerofix
Fixative
Concentrate
High
Humidity
55-‐70%
Ethylene
Glycol
contains:
30-‐45%
Polyvinylpyrrolidone
(PVP)
<1%
Azure
B
SS-‐029,
SS-‐029C
and
SS-‐029CG
(as
diluted)
Nozzle
40-‐50%
Methyl
Alcohol
Cleaning
Solution
contains:
1-‐5%
Oxalic
Acid
SS-‐029C,
SS-‐029CG
Nozzle
Cleaning
Solution
95-‐99%
Deionized
Water
Concentrate
contains:
1-‐5%
Oxalic
Acid
SS-‐133
Decontamination
Solution
when
diluted
as
<2%
Germicidal
Detergent
directed
contains:
>98%
Deionized
Water
SS-‐MeOH
Anhydrous
Methanol
contains:
≥99.5%
Methyl
Alcohol,
Anhydrous
SS-‐133
Decontamination
Solution
Concentrate
<30%
Germicidal
Detergent
contains:
>70%
Deionized
Water
90
APPENDIX
B
Risk
and
Safety
Phrases
Reagents
SS-‐035A,
SS-‐035A-‐EU,
and
SS-‐035AG
Reagent
–
A
Rinse,
and
SS-‐135/149A
–
Rinse
Reagent
concentrate
after
dilution
as
directed,
are
not
associated
with
any
Risk
and
Safety
Phrases.
They
are
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
Reagents
SS-‐035/049B,
SS-‐035/049B-‐EU,
and
SS-‐035/049BG
–
Reagent
B
Thiazin
Stain,
and
SS-‐135/149A
–
Rinse
Reagent
concentrate
after
dilution
as
directed,
are
not
associated
with
any
Risk
and
Safety
Phrases.
They
are
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
Reagents
SS-‐035C,
SS-‐035C-‐EU,
and
SS-‐035CG
–
Reagent
C
Eosin
Stain,
and
SS-‐135/149A
–
Rinse
Reagent
concentrate
after
dilution
as
directed,
are
not
associated
with
any
Risk
and
Safety
Phrases.
They
are
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
SS-‐MeOH
Aerospray
Reagent-‐Grade
Methanol
is
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbols
of
Danger
are:
F
and
T.
R
11:
Highly
Flammable
R
23/24/25:
Toxic
by
inhalation,
in
contact
with
skin
and
if
swallowed
R
39/23/24/25:
Toxic:
danger
of
very
serious
irreversible
effects
through
inhalation,
in
contact
with
skin
and
if
swallowed
S
7:
Keep
container
tightly
closed
S
16:
Keep
away
from
sources
of
ignition
–
No
smoking
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
45:
In
case
of
accident
or
if
you
feel
unwell,
seek
medical
advice
immediately
(show
label
where
possible)
Reagents
SS-‐048
or
SS-‐048-‐EU
–
Aerofix
Fixative,
500
mL,
and
SS-‐148
–
Aerofix
Additive
for
methanol
reagent
after
dilution
as
directed,
are
associated
with
the
following
Risk
and
Safety
Phrases.
The
associated
EU
Symbols
of
Danger
are
F
and
T.
R
11:
Highly
Flammable
R
23/24/25:
Toxic
by
inhalation,
in
contact
with
skin
and
if
swallowed
R
39/23/24/25:
Toxic:
danger
of
very
serious
irreversible
effects
through
inhalation,
in
contact
with
skin
and
if
swallowed
S
7:
Keep
container
tightly
closed
S
16:
Keep
away
from
sources
of
ignition
–
No
smoking
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
45:
In
case
of
accident
or
if
you
feel
unwell,
seek
medical
advice
immediately
(show
label
where
possible)
Reagents
SS-‐049A,
SS-‐049AG,
SS-‐049A-‐EU–
Basofix
Rinse
Reagent,
and
SS-‐135/149A
–
Rinse
Reagent
concentrate
after
dilution
as
directed,
are
not
associated
with
any
Risk
and
Safety
Phrases.
They
are
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
91
APPENDIX
B
Risk
and
Safety
Phrases
Reagents
SS-‐049C
and
SS-‐049C-‐EU
–
Basofix
Eosin
Stain,
SS-‐049CG
or
SS-‐135/149C
–
Eosin
Stain
concentrate
after
dilution
as
directed,
are
associated
with
the
following
Risk
and
Safety
Phrases.
The
associated
EU
Symbol
of
Danger
is
Xn.
R
20/21/22:
Harmful
by
inhalation,
in
contact
with
skin
and
if
swallowed
R
68/20/21/22:
Harmful:
possible
risk
of
irreversible
effects
through
inhalation,
in
contact
with
skin
and
if
swallowed
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
45:
In
case
of
accident
or
if
you
feel
unwell
seek
medical
advice
immediately
(show
the
label
where
possible)
Reagents
SS-‐049P
and
SS-‐049P-‐EU
–
Basofix
Pre-‐dip
Fixative
and
SS-‐149P
–
Basofix
Pre-‐dip
Fixative
concentrated
when
diluted
as
directed
are
associated
with
the
following
Risk
and
Safety
Phrases.
The
associated
EU
Symbols
of
Danger
are
F
and
T.
R
11:
Highly
Flammable
R
23/24/25:
Toxic
by
inhalation,
in
contact
with
skin
and
if
swallowed
R
39/23/24/25:
Toxic:
danger
of
very
serious
irreversible
effects
through
inhalation,
in
contact
with
skin
and
if
swallowed
S
7:
Keep
container
tightly
closed
S
16:
Keep
away
from
sources
of
ignition
–
No
smoking
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
45:
In
case
of
accident
or
if
you
feel
unwell,
seek
medical
advice
immediately
(show
label
where
possible)
Reagents
SS-‐135/149A
–
Rinse
Reagent
Concentrate
is
not
associated
with
any
Risk
and
Safety
Phrases.
It
is
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
Reagents
SS-‐135/149B
–
Thiazin
Stain
Reagent
Concentrate
is
not
associated
with
any
Risk
and
Safety
Phrases.
It
is
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
Reagents
SS-‐135/149C
–
Eosin
Stain
Reagent
Concentrate
is
not
associated
with
any
Risk
and
Safety
Phrases.
It
is
not
associated
with
any
EU
Symbols
of
Danger.
R
Phrases:
None
S
Phrases:
None
Reagent
SS-‐135S
–
Hematology
Surfactant
is
associated
with
the
following
Risk
and
Safety
Phrases.
The
associated
EU
Symbol
of
Danger
is
Xi.
R
43:
May
cause
sensitization
by
skin
contact
R
52/53:
Harmful
to
aquatic
organisms,
may
cause
long-‐term
adverse
effects
in
the
aquatic
environment
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
61:
Avoid
release
to
the
environment.
Refer
to
special
instructions/safety
data
sheet
92
APPENDIX
B
Risk
and
Safety
Phrases
Reagents
SS-‐148
–
Aerofix
Additive
for
Methanol
is
associated
with
the
following
Risk
and
Safety
Phrases.
The
associated
EU
Symbol
of
Danger
is
Xn.
R
22:
Harmful
if
swallowed
S
phrases:
None
SS-‐029
Aerospray
Nozzle
Cleaning
Solution
and
SS-‐029C
when
diluted
as
instructed
are
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbols
of
Danger
are:
F
and
T.
R
11:
Highly
Flammable
R
23/24/25
Toxic
by
inhalation,
in
contact
with
skin
and
if
swallowed
R
39/23/24/25:
Toxic:
danger
of
very
serious
irreversible
effects
through
inhalation,
in
contact
with
skin
and
if
swallowed
S
7:
Keep
container
tightly
closed
S
16:
Keep
away
from
sources
of
ignition
–
No
smoking
S
36/37:
Wear
suitable
protective
clothing
and
gloves
S
45:
In
case
of
accident
or
if
you
feel
unwell,
seek
medical
advice
immediately
(show
label
where
possible)
SS-‐029C
and
SS-‐029C-‐EU
Aerospray
Nozzle
Cleaning
Solution
is
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbol
of
Danger
is:
None.
R:
None
S:
None
SS-‐133
Decontamination
Solution
Concentrate
is
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbols
of
Danger
is
C
R
34
Causes
burns
R
22
Harmful
if
swallowed
R
67
Vapors
may
cause
drowsiness
and
dizziness
S
24/25
Avoid
contact
with
skin
and
eyes
S
26
In
case
of
contact
with
eyes,
rinse
immediately
with
plenty
of
water
and
seek
medical
advice
S
36/37/39
Wear
suitable
protective
clothing,
gloves
and
eye/face
protection
S
45
In
case
of
accident
or
if
you
feel
unwell
seek
medical
advice
immediately
(show
the
label
where
possible)
SS-‐103
O-‐ring/Nozzle
Thread
Grease
is
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbols
of
Danger
are:
None.
R:
None
S:
None
SS-‐149P
Basofix
Pre-‐dip
Fixative
Concentrate
when
diluted
as
instructed
is
associated
with
the
following
Risk
and
Safety
Phrases.
European
Symbols
of
Danger
are:
None.
R:
None
S:
None
93
APPENDIX
C
Accessories
and
Supplies
Only
replacement
parts
supplied
by
Elitech
Group
should
be
used
in
this
instrument.
Use
of
non-‐approved
parts
may
affect
the
performance
and
safety
features
of
this
product.
ACCESSORIES
REFERENCE
NUMBER
Slide
Carousel
(12-‐Slide
Capacity)
...........................................................
AC-‐028
Nozzle
Tool
..............................................................................................
AC-‐034
Nozzle
Hex
Wrench
.................................................................................
AC-‐035
5
L
Space-‐Saver
Container
w/cap
(For
Concentrate
Reagents)
...............
AC-‐038
Drain
Tube,
1.8
meter
(6
foot)
Length
.....................................................
AC-‐041
500
mL
bottle
with
cap
(pack
of
5
bottles)
..............................................
AC-‐043-‐05
Slide
Carousel
(30-‐Slide
Capacity)
...........................................................
AC-‐057
Nozzle
Orifice
Cleaning
Wire
...................................................................
AC-‐059
Reagent
Pump
Priming
Tool
....................................................................
AC-‐069
Aerospray/Cytopro
Safety
Shield
............................................................
AC-‐110
Cytopro
Cytocentrifuge
Rotor
.................................................................
AC-‐160
10
L
Waste
Container
(without
level
detect)
..........................................
AC-‐170
1D
Barcode
Scanner
................................................................................
AC-‐181
10
L
Waste
Container
(with
level
detect)
................................................
AC-‐182
Nozzle
Maintenance
Kit
...........................................................................
AC-‐184
2D
Barcode
Scanner
................................................................................
AC-‐185
SUPPLIES
Nozzle
Cleaning
Solution,
355
mL
...........................................................
SS-‐029
Nozzle
Cleaning
Solution
Concentrate,
250
mL
(Dilutes
to
500
mL)
.......
SS-‐029C
or
SS-‐029C-‐EU
Nozzle
Cleaning
Solution
Concentrate,
1.89
L
(Dilutes
to
3.785
L)
.........
SS-‐029CG
Reagent
A
(Rinse),
500
mL
.......................................................................
SS-‐035A
or
SS-‐035A-‐EU
Reagent
B
(Thiazin
Stain),
500
mL
...........................................................
SS-‐035/049B
or
SS-‐035/049B-‐EU
Reagent
C
(Eosin
Stain),
500
mL
..............................................................
SS-‐035C
or
SS-‐035C-‐EU
Reagent
C
(Eosin
Stain),
3.785
L
(1
gal)
...................................................
SS-‐035CG
Reagent
D
(Methanol),
500
mL
...............................................................
SS-‐MEOH
Reagent
A
(Rinse),
3.785
L
(1
gal)
............................................................
SS-‐035AG
Reagent
B
(Thiazin
Stain),
3.785
L
(1
gal)
................................................
SS-‐035/049BG
Reagent
A
(Rinse)
Concentrate,
135
mL
(Dilutes
to
4.6
L)
......................
SS-‐135/149A
SYSTEM
Reagent
B
(Thiazin
Stain)
Concentrate,
135
mL
(Dilutes
to
4.6
L)
...........
SS-‐135/149B
SYSTEM
Reagent
C
(Eosin
Stain)
Concentrate,
135
mL
(Dilutes
to
4.6
L)
.............
SS-‐135/149C
SYSTEM
Surfactant
(used
with
SS-‐135
A,
B,
C)
......................................................
SS-‐135S
Reagent
D
(Aerofix®
Fixative
High
Humidity),
500
mL
............................
SS-‐048
or
SS-‐048-‐EU
Aerofix®
Additive
for
Reagent
D
(Methanol),
135
mL
.............................
SS-‐148
or
SS-‐148-‐EU
Reagent
A
(Basofix™
Rinse),
500
mL
.......................................................
SS-‐049A
or
SS-‐049A-‐EU
Reagent
C
(Basofix™
Eosin
Stain),
500
mL
...............................................
SS-‐049C
or
SS-‐049C-‐EU
Basofix™
Pre-‐dip
Fixative,
500
mL
...........................................................
SS-‐049P
or
SS-‐049P-‐EU
Basofix™
Pre-‐dip
Concentrate
.................................................................
SS-‐149P
Decontamination
Solution
Concentrate
..................................................
SS-‐133
O-‐Ring/Nozzle
Thread
Grease
(3
grams)
.................................................
SS-‐103
Preventive
Maintenance
Chart,
pad
of
24
sheets
...................................
SS-‐265
Aerospray
Hematology
Stat
(Model
7122)
Applications
Manual
............
RP-‐460
Paper
Test
Slides
.....................................................................................
RP-‐500
94