English Aerospray Stat (7122) Applications Manual 57-2000-01A

Download as pdf or txt
Download as pdf or txt
You are on page 1of 94

 

 
 
 
AEROSPRAY®  HEMATOLOGY  STAT  
 
SLIDE  
STAINER/CYTOCENTRIFUGE  
 
 
Model  7122  
 
 

Applications  Manual  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
57-­‐2000-­‐01A  
 
 
 
 
 
 
 
 
   
 
 
©2013  ELITechGroup  Inc.  All  rights  reserved.  Printed  in  the  United  States  of  America.  
No  part  of  this  publication  may  be  reproduced,  transmitted,  transcribed,  stored  in  a  
retrieval  system,  or  translated  into  any  language  (human  or  computer)  in  any  form,  or  
by  any  means  whatsoever,  without  the  prior  express  written  permission  of  
ELITechGroup  Inc.  
 
Wescor,  Aerospray  and  Cytopro  are  trademarks  of  ELITechGroup  Inc.  
 
Other  trade  names  used  in  this  manual  are  trademarks  of  their  respective  owners,  used  
here  for  information  only.  
 
ELITechGroup  Inc.  makes  no  express  or  implied  warranty  regarding  this  manual,  its  
quality,  performance,  or  appropriate  use  regarding  any  type  of  specific  procedure.  
ELITechGroup  Inc.  may  modify  this  manual  without  notice  and  without  implying  any  
obligation  or  liability  on  the  part  of  the  company.  
 
Manufactured  in  the  United  States  of  America  by:  
 
ELITechGroup  Inc.  
370  West  1700  South  
Logan,  Utah  84321-­‐8212  USA  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2  
Table  of  Contents  
 
Section  1  
Introduction    
1.1  Instrument  Overview  ......................................................................  7  
Using  this  Manual  ..................................................................  7  
Functional  Description  ..........................................................  8  
Key  Features  ..........................................................................  9  
Intended  Use  .........................................................................  9  
Table  1:  General  Specifications  .............................................  10  
Table  2:  Performance  Specifications  .....................................  11  
Table  3:  Run  Time  Sequence  ..................................................  12  
Table  4:  Approximate  Reagent  Consumption  ........................  12  
Table  5:  Carousel  and  Rotor  Information  ..............................  13  
Table  6:  Explanation  of  Symbols  ............................................  14  
1.2  Instrument  Description  ...................................................................  16  
Figure  1:  Front  and  Right  Side  Panels  ....................................  16  
Figure  2:  Front  Panel  and  Touchscreen  .................................  17  
Figure  3:  Rear  Panel  ..............................................................  17  
Figure  4:  Stainer  Bowl  Components  ......................................  18  
Figure  5:  Nozzle  Diagram  ......................................................  18  
Table  7:  Preventive  Maintenance  Kit  .....................................  19  
Barcode  Reader  .....................................................................  20  
Figure  6:  Barcode  Reader  ......................................................  20  
1.3  Touchscreen  and  User  Interface  .....................................................  21  
Table  8:  Front  Panel/Main  Screen  Function  Keys  ..................  21  
Table  9:  System  Setup  Keys  ...................................................  22  
Table  10:  Maintenance  Function  Keys  ...................................  24  
 
Section  2  
Setup  and  Preparation  for  Operation  
2.1  Instrument  Setup  ............................................................................  25  
Unpacking  and  Installing  the  Stainer  .....................................  25  
Connecting  the  Drain  Tube  and  Waste  Container  .................  26  
Connecting  Power  .................................................................  26  
Installing  Reagent  Bottles  ......................................................  27  
Installing  the  Barcode  Reader  ...............................................  28  
2.2  Preparing  the  Stainer  for  Operation  ...............................................  29  
Priming  Procedures  ...............................................................  29  
The  Clean  Cycle  .....................................................................  31  
Performing  Tests  ...................................................................  31  
Reagent  Level  Monitoring  .....................................................  32  
Enabling/Disabling  Reagent  Level  Detect  ........................  32  
Zeroing  the  Reagent  Level  Sensors  ..................................  33  
 
Section  3    
Controlling  and  Customizing  Stainer  Functions  
3.1  System  Setup  Menu  ........................................................................  35  
Accessing  the  System  Setup  Menu  ........................................  35  
Stain  Programs  ......................................................................  35  
Creating  a  Stain  Program  ......................................................  36  
Editing,  Renaming,  or  Adjusting  Stain  Programs  ...................  36  
Administrator  and  User  Accounts  .........................................  37  
Creating  an  Administrator  Account  .................................  37  

3  
Table  of  Contents  
 
Section  3    
Controlling  and  Customizing  Stainer  Functions  (continued)  
Creating  User  Accounts  ...................................................  38  
Managing  User  Access  ....................................................  39  
User  Login/Logout  ...........................................................  39  
Using  Reagent  Information  Tracking  .....................................  41  
Modifying  Level  Detect  Functions  .........................................  41  
Changing  User  Language  .......................................................  42  
Setting  the  Date  and  Time  .....................................................  42  
System  Log  ............................................................................  42  
Accessing  Logs  .................................................................  42  
Exporting  Logs  .................................................................  43  
Controlling  Beeper  Alerts  ......................................................  43  
QC/Maintenance  Tracking  ....................................................  44  
Enable  Stain  Slide  Tracking  ..............................................  44  
Enable  Cyto  Slide  Tracking  ...............................................  45  
Enable  Manual  Entry  .......................................................  45  
Enable  Preventive  Maintenance  Tracking  .......................  45  
Enable  Reagent  Tracking  .................................................  45  
Restoring  Software  Defaults  .................................................  46  
3.2  Recording  Specimen  and  Reagent  Information  ..............................  47  
Scanning  Slides  with  the  Barcode  Reader  .............................  47  
Scanning  Reagent  Bottles  with  the  Barcode  Reader  .............  48  
Manually  Entering  Specimen  information  ............................  49  
Manually  Entering  Reagent  Information  ...............................  50  
3.3  The  Help  Menu  ................................................................................  51  
Using  Help  .............................................................................  51  
 
Section  4    
Operating  the  Stainer    
4.1  Operating  Instructions  ....................................................................  52  
Suggested  Staining  Protocol  ..................................................  52  
Performing  a  Hub  Pattern  Test  .............................................  52  
Figure  7:  Correct  Hub  Pattern  Test  Result  .....................  52  
Figure  8:  Incorrect  Hub  Pattern  Test  Result  ...................  52  
Loading  the  Carousel  .............................................................  53  
Figure  9:  Loading  the  Carousel  ......................................  54  
Figure  10:  Specimen  Placement  on  Slide  .......................  54  
Using  Blocking  Slides  .............................................................  55  
Figure  11:  Using  Blocking  Slides  ....................................  55  
Performing  a  Stain  Cycle  .......................................................  55  
Unloading  the  Carousel  .........................................................  56  
Monitoring  Reagent  and  Waste  Levels  .................................  57  
Table  11:  Reagent  Level  Detect  Display  Symbols  ..........  57  
Replacing  a  Reagent  Bottle  ...................................................  58  
Emptying  the  Waste  Container  .............................................  58  
Section  5    
Preventive  Maintenance  and  Safety  
5.1  Preventive  Maintenance  .................................................................  59  
Daily  Maintenance/Quality  Control  (QC)  ..............................  59  
Weekly  Maintenance  ............................................................  59  
Monthly  Maintenance  ...........................................................  60  
Using  the  Preventive  Maintenance  Log  ................................  60  
 

4  
Table  of  Contents  
 
Section  5    
Preventive  Maintenance  and  Safety  (continued)  
5.2  Storing  the  Instrument  ....................................................................  61  
Preparing  for  Long-­‐Term  Storage  ..........................................  61  
5.3  Replacing  Fuses  ...............................................................................  62  
5.4  Cleaning  the  Stainer  and  Carousels  .................................................  63  
Cleaning  Liquid  Spills  .............................................................  63  
Cleaning  Broken  Slides  ..........................................................  63  
5.5  Decontaminating  the  Stainer  and  Carousels  ...................................  64  
Solutions  for  Decontaminating  the  Instrument  ....................  64  
Figure  12:  Lid  Latch  and  Locking  Pin  Hole  Locations  .....  65  
Decontaminating  the  Instrument  ..........................................  65  
Decontaminating  Reagent  Bottles  ........................................  66  
5.6  Shipping  or  Disposing  of  the  Stainer  or  Carousels  ..........................  67  
Shipping  the  Instrument  or  Carousels  ...................................  67  
  Hazard-­‐Free  Certification  ..............................................  67  
Disposing  of  the  Instrument  or  Carousels  .............................  67  
     
Section  6    
Nozzle  Maintenance  and  Performance  
6.1  Nozzle  Disassembly  and  Cleaning  ...................................................  68  
Nozzle  Disassembly  ...............................................................  68  
Nozzle  Cleaning  .....................................................................  68  
6.2  Nozzle  Reassembly  ..........................................................................  69  
6.3  Manual  Priming  ...............................................................................  70  
6.4  Performing  the  Volume  Test  ...........................................................  71  
Table  12:  Volume  Test  Tolerances  .........................................  71  
6.5  Performing  the  Slide  Pattern  Test  ...................................................  73  
Figure  13:  Correct  Slide  Pattern  Test  Result  ..........................  73  
Figure  14:  Incorrect  Slide  Pattern  Test  Result  ........................  73  
 
Section  7  
Solving  Problems  
7.1  Troubleshooting  ..............................................................................  74  
Table  13:  General  Troubleshooting  and  Diagnosis  ................  74  
7.2  Abnormal  Staining  Results  ..............................................................  80  
Smear  Separation  or  Tearing  .................................................  80  
  Improper  Alcohol  Fixation  .................................................  80  
Wet  Smears  .......................................................................  80  
Thick  Smears  ......................................................................  80  
Dirty  Slides  .........................................................................  80  
Red  Cell  Artifacts  ...............................................................  80  
Use  Aerofix®  ......................................................................  81  
False-­‐Positive  Howell-­‐Jolley  Bodies  or  Amorphous    
Precipitates  ........................................................................  81  
7.3  Instrument  Malfunction  ..................................................................  83  
Air  or  Reagent  Leaks  ..............................................................  83  
Reagent  Level  Detect  System  Errors  .....................................  84  
Calibrating  the  Reagent  Level  Detect  System  .......................  84  
7.4  Calibrating  the  Touchscreen  ...........................................................  86  
7.5  Service  Information  .........................................................................  87  
 

5  
Table  of  Contents  
 
Section  8  
Cytopro®  Cytocentrifuge  
8.1  Cytopro  Cytocentrifuge  Information  ...............................................  88  
Functional  Description  ..........................................................  88  
Key  Features  ..........................................................................  88  
Intended  Use  .........................................................................  88  
 
Appendix  A:  Critical  Reagent  Components  ............................................  89  
 
Appendix  B:  Risk  and  Safety  Phrases  .....................................................  91  
 
Appendix  C:  Accessories  and  Supplies  ..........................................................  94  
 

6  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
Using  this  Manual  
This  manual  provides  instructions  to  install,  operate,  and  maintain  the  Aerospray  
Hematology  Stat  Stainer/Cytocentrifuge.  The  manual  is  an  important  part  of  the  
product.  Read  it  carefully  and  completely  before  setup  and  first  use  of  the  instrument.    
 
If  additional  accident  prevention  and  environmental  protection  requirements  exist  in  
the  country  of  operation,  this  manual  must  be  supplemented  by  appropriate  
instructions  to  ensure  compliance.  
 
Safety  Regulations  
This  instrument  has  been  built  and  tested  in  accordance  with  safety  regulations  for  the  
following  device  types:  electrical  control,  regulating,  and  laboratory.  In  order  to  
maintain  this  condition  and  ensure  safe  operation,  the  operator  must  observe  all  the  
instructions  and  warnings  contained  in  this  manual.  For  current  information  about  
applicable  standards,  please  refer  to  the  CE  Declaration  of  Conformity  included  with  the  
documents  shipped  with  this  device.  
NOTE:  This  equipment  complies  with  the  emission  and  immunity  requirements  described  
in  the  IEC  61326  series.  
Understanding  Warnings  
This  manual  uses  three  warning  levels  to  alert  you  to  important  information  as  shown  in  
the  following  examples.  
 
WARNING!  
A  Warning  alerts  to  the  possibility  of  personal  injury,  death,  or  other  serious  adverse  
reactions  stemming  from  the  use  or  misuse  of  this  device  or  its  components.  
 
CAUTION:    
A  Caution  alerts  to  possible  problems  with  the  device  associated  with  its  use  or  misuse.  
Such  problems  include  device  malfunction,  failure,  damage,  damage  to  the  sample,  or  
damage  to  other  property.  Where  applicable,  a  Caution  may  include  precautions  to  be  
taken  to  avoid  the  hazard.  
 
NOTE:  A  Note  reinforces  or  supplies  additional  information  about  a  topic.  
 
Specific  Warnings  
Pay  particular  attention  to  the  following  safety  precautions.  If  these  safety  precautions  
are  ignored,  injury  or  damage  to  the  instrument  may  occur.  Each  individual  precaution  is  
important.  
 
WARNING!  
Install  the  stainer  in  a  well-­‐ventilated  area.  If  ventilation  is  inadequate,  operate  the  
instrument  under  a  safety  hood.    

7  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
WARNING!  
Reagents  used  in  the  stainer  contain  moderately  hazardous  chemicals  that  require  
care  in  handling.  Always  use  appropriate  safety  measures  including  gloves  and  eye  
protection  when  handling  reagents.  
 
WARNING!  
Always  wear  protective  clothing  and  eye  protection  when  using  Nozzle  Cleaning  
Solution  (diluted  REF:  SS-­‐029C).  Dispose  of  used  solution  properly.  
 
WARNING!    
If  power  is  lost  while  the  stainer  is  running,  the  lid  will  remain  locked  until  power  is  
restored.  Do  not  attempt  to  open  the  lid  while  power  is  off.    
 
WARNING!  
Electrical  shock  hazard:  Do  not  open  this  instrument  or  attempt  internal  repairs.  Refer  
servicing  to  qualified  service  personnel.  Contact  Elitech  Group  Biomedical  Systems  
service.  
 
CAUTION:  
This  equipment  has  been  designed  and  tested  to  CISPR  11  Class  A  and  FCC  Part  15  Class  
A.  In  a  domestic  environment  it  may  cause  radio  interference,  in  which  case,  you  may  
need  to  take  measures  to  mitigate  the  interference.  
CAUTION:    
To  avoid  serious  instrument  damage,  always  use  reagents  supplied  by  Elitech  Group.  
Except  for  approved  methanol,  using  reagents  not  supplied  by  Elitech  Group  may  void  
the  warranty.  
 
CAUTION:  
Only  spare  parts  supplied  or  specified  by  Elitech  Group  should  be  used  in  this  device.  
Using  non-­‐approved  parts  may  affect  the  performance  and  safety  features  of  the  
instrument.  If  the  equipment  is  used  in  a  manner  not  specified  by  Elitech  Group  the  
protection  provided  by  the  equipment  may  be  impaired.  If  in  doubt,  contact  your  Elitech  
Group  representative.  
 
Functional  Description  
The  Aerospray  Hematology  Stat  Slide  Stainer/Cytocentrifuge  (Model  7122)  is  a  dual-­‐
purpose,  microprocessor-­‐controlled  slide  staining  and  cell  preparation  system.  In  use,  
atomizing  spray  nozzles  apply  fresh  reagents  onto  microscope  slides  prepared  with  
blood  or  other  body  fluid  specimens.  The  slides  are  mounted  in  a  rotating  carousel  for  
processing.    
 

8  
SECTION  1  
INTRODUCTION  
 
 1.1  Instrument  Overview    
Key  Features  
• Minimized  reagent  consumption  
• Rapid  staining  
• Barcode  scanner  for  tracking  specimens  and  reagents  
• Reagent  and  specimen  traceability  
• User  traceability  
• Administrator  password  
• Interactive  touchscreen  display  
• Multiple  languages  
• High-­‐volume  staining  productivity  (12  slides  per  stain  cycle)  
• Separate  reservoir,  delivery  tube,  pump,  and  spray  nozzle  for  each  reagent  
• Operator-­‐selectable  automatic  alcohol  fixation  function  to  fix  specimens  
• Reagent  and  waste  level  monitoring  
• Log  files  
• 10  Stain  Intensity  Settings  

The  correct  accessory  must  be  used  for  each  function.  The  Cytopro  Cytocentrifuge  Rotor  is  
available  as  an  option  offering  additional  features  (see  Section  8).  
 
Intended  Use  
 
The  Aerospray  Hematology Stat  Slide  Stainer/Cytocentrifuge  (Model  7122)  is  intended  for  
use  by  medical  professionals  to  stain  specimens  that  may  include  blood  and  other  body  
fluids,  as  a  step  of  standard  laboratory  practice  in  diagnosing  disease  in  humans.  Addition  of  
the  Cytopro  rotor  allows  preparation  of  slides  by  cytocentrifugation  before  staining.  
 

9  
SECTION  1  
INTRODUCTION  
 
 1.1  Instrument  Overview    
Table  1:  General  Specifications  
Category   Characteristics  
Slide  Carousel  Capacity   1  to  12  
Carousel  Rotation  Speed   Reagent  application:  20  rpm  (±  5%)    
Slide  Drying:  950  rpm  (±  5%)  
 
Cytocentrifuge  Rotor  Speed   100  to  2000  rpm  (±  5%),  user  programmable  
 
Reagent  Consumption   Refer  to  Approximate  Reagent  Consumption,  Table  4  
 
Operating  Time   Refer  to  Run  Time  Sequence,  Table  3  
 
Display   7  in.  WVGA  (800  x  480  pixels)  TFT  
 
Touchscreen  Controls   Menu-­‐driven  icons  
 
Drain  Connection   Connector  on  rear  panel  accepts  male  connector  
attached  to  vinyl  drain  tube.  1.8  meters  (6  ft.)  length  
supplied  
 
Ventilation   Air  is  exhausted  from  the  stainer  via  a  female  ½  inch  SAE  
pipe  thread  fitting  
Dimensions      
Width   57  cm  (22  in.)  
Height  (lid  closed)   25  cm  (10  in.)  
Depth     54  cm  (21  in.)  
Height  (lid  open)   58  cm  (23  in.)  
 
Weight   15.2  kg  (33.5  lb.)  
 
Electrical  Requirements   100  to  240  VAC  (±  10%)  @  50  to  60  Hz  
 
Power  Consumption   200  Watts  maximum  
 
Fuses     2.0  Amp  time-­‐lag  (Type  T)  (5  x  20  mm)  (2  required)  
Manufactured  by  Littlefuse  (Part  No.  218002)  or  
Bussman  (Part  No.  GDC-­‐2A)  
 
Ambient  Temperature    
Operating   15  to  30  °C  (59  to  86  °F)  
Storage   -­‐10  to  50  °C    (14  to  122  °F)  
 
Relative  Humidity   ≤  80%  non-­‐condensing  
 
Altitude   ≤  2000  m  (≤  6562  ft.)  

 
 

10  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
Table  1:  General  Specifications  (continued)  
 
Category   Characteristics  
Pollution  Degree     2  
   
Heat  Dissipation    
Maximum   512  Btu/hour  max  
Average  During  Staining   34  Btu/hour  
Average  While  Idle   10  Btu/hour  
   
Maximum  Sound  Emission   72  dB  
 
 
Table  2:  Performance  Specifications  
Category   Characteristics  
Reagent  Spray  Nozzles   Four  reagent  nozzles:  A,  B,  C,  D    
Reagents:   A  -­‐  Rinse  
   
NOTE:  Use  only  Elitech  Group  reagents,  with  diluents   B  –  Thiazin  Stain  (Blue)  
as  specified  for  Elitech  Group  concentrated  reagents.    
REF  numbers  for  this  stainer  start  with  one  of  the   C  –  Eosin  Stain  (Red)  
following:  SS-­‐035,  SS-­‐049,  SS-­‐135,  SS-­‐149,  SS-­‐048,  or    
SS-­‐148.   D  –  Methanol  or  Aerofix®  
 
Note:  Methanol  must  be  99.5%  pure  no  more  than  0.5%  
water.  
 
 
Stain  Settings   Number  of  slides  to  stain.  Intensity  settings:  1-­‐10.  
Fixation  level:  Off,  Normal,  High  
 

11  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
Table  3:  Run  Time  Sequence    
NOTE:  Table  3  represents  a  typical  timing  sequence  for  this  instrument  using  the  12-­‐slide  
carousel,  presented  for  general  reference  only.  Timing  is  in  seconds.  Actual  cycle  times  may  vary.  
 
Staining  Mode   Medium  Dark  Setting  (5)  
Fixation  Pump  D  (0ptional)   (102)  
Eosin  Stain  (Pump  C)   21  
Thiazin  Stain  (Pump  B)   33  
Rinse  (Pump  A)   27  
High  Speed  Dry   33  
   
Total  without  fixation  (in  minutes)   1.9  
Total  with  fixation  (in  minutes)   3.6  
 
Table  4:  Approximate  Reagent  Consumption  (mL  )    
 
Staining  Mode   Medium  Dark   Clean  Cycle  
Setting  (5)  
Reagent  A:  Rinse   6.0   7.5  
Reagent  B:  Thiazin   6.0   7.5  
Reagent  C:  Eosin   6.0   7.5  
Reagent  D:  Fixative   1.4   38.0  

12  
SECTION  1  
INTRODUCTION  
1.1  Instrument  Overview    
Table  5:  Carousel  and  Rotor  Information  
Only  the  following  slide  staining  carousel  or  cytocentrifuge  rotor  can  be  used  in  this  instrument.  
Each  should  be  used  following  the  instructions  in  this  manual  or  the  Cytopro  Applications  Manual  
(RP-­‐517).  
Rotor/Carousel   Maximum   Maximum  Capacity   Maximum  
rpm   Sample  
Volume  
12-­‐Slide  Carousel  (AC-­‐028)   1000  rpm   12  each,  26  mm  x  76  mm   N/A  
(1  x  3  inch)  microscope  
slides  
Cytopro  Cytocentrifuge  Rotor  (AC-­‐160)   2000  rpm   8  each,  standard   Up  to  600  µL*  
chambers,  plus  slides    
   
8  each,  Cytopro  Magnum   Up  to  6  mL*  
chambers,  plus  slides  
 
*Do  not  overfill  cytocentrifuge  chambers.  See  Cytopro  Applications  Manual  or  Methods  Manual  
for  detailed  instructions  and  warnings.  
 

13  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
Table  6:  Explanation  of  Symbols  
 
Alternating  Current  (AC)  

 
Authorized  Representative  in  the  European  Community  
 
 
Batch  Code  
 
 
 
Biological  Hazards  (Biological  Risks)  
 
 
Catalog  Number  (Model  Number)  
 
 
   
  Caution,  Consult  Accompanying  Documents  (Attention,  see  instructions  for  use)  
 
 
  CE  Mark,  product  meets  the  essential  requirements  designated  in  Annex  I  of  the  
  In  Vitro  Device  Directive  (IVDD)  98/79/EC  
 
   
  Consult  Instructions  For  Use  
 
 
Do  Not  Reuse  
 
  Do  not  use  if  package  is  damaged  
   
 
Fragile,  Handle  with  Care  
 
 
  Fuse  
   
 
General  Symbol  for  Recovery,  Recyclable  
 
   
 
In  vitro  Diagnostic  Device  (In  vitro  Diagnostic  Medical  Device)  
 
 
Keep  away  from  sunlight  (Keep  away  from  heat)  
 
  Manufacturer  
 
 
Waste  Electrical  and  Electronic  Equipment  
 
 
Reagent  stable  after  opening  for  number  of  months  specified  
 

14  
SECTION  1  
INTRODUCTION  
 
1.1  Instrument  Overview    
Table  6:  Explanation  of  Symbols  (continued)  

I    “On”  (Power)    
 
   
O   “Off”  (Power)  
 
   
Environment  Friendly  Use  Period      
 
   
  Serial  Number  
 
   
Temperature  Limitation  –  indicates  high  and  low  limits.  (Normal  room  
temperatures  are  specified  for  all  Elitech  Group  Reagents.)  
   
  Use  By  
 
   
  Volume  Control  
 
   
    General  Warning,  Caution,  Risk  of  Danger  
 
   
  Warning,  Biological  Hazard    
 
   
Corrosive  
 
   
  Environment  Hazard  
 
   
  Flammable  
 
   
Harmful/Irritant  
 
   
Oxidant  
 
   
    Toxic  
 
 
Product  to  be  used  for  manual  cleaning  only.  Do  not  pump  product  through  
 
instrument.  
 

15  
SECTION  1  
INTRODUCTION  
 
1.2  Instrument  Description    
Figure  1:  Front  and  Right  Side  Panels  
 
 
 
 
 
 
  4  
 
 
  3  
 
  2  
  5  
  1  
  D  
  C  
  B  
  A  
 
 
 
 
 
1  –  Front  Panel  with  Touch  Screen  Display   6  
2  –  Bowl  
3  –  Carousel  
4  –  Lid  with  Safety  Lock  
5  –  Right  Side  Panel  with  Label  Indicating  Reagent  Positions:  
A  –  Rinse  
B  –  Thiazin  
C  –  Eosin  
D  –  Methanol  or  Aerofix  
6  –  Reagent  Tray  

16  
SECTION  1  
INTRODUCTION  
 
1.2  Instrument  Description  
Figure  2:  Front  Panel  and  Touchscreen  
 
 
    1  

 
 
  2  
 
 
 
 
1  –  Standby/Ready  Button    
2  –  Touchscreen  
 
The  front  panel  features  an  interactive  touchscreen  display.  Refer  to  Front  Panel/Main  
Screen  Function  Keys  (Section  1.3,  Table  8)  for  more  information.  
 
Figure  3:  Rear  Panel      1    1   2              3  
 
 
 
 
 
 
 
 
 
 
 
  4  
 
  5  
 
 
  11   10    9  8   7   6  
1  –  USB  Ports  
2  –  Network  Ethernet  Connection  
3  –  Exhaust  Vent  
4  –  Rear  Panel  Label  
5  –  Biohazard  Warning  Label  
6  –Waste  Tube  Connection  
7  –  Level  Detection  Connection  for  Waste  Container  
8  –  Power  Switch  
9  –  Fuse  Door    
10  –  Power  Cord  Connection  
11  –  Model/Serial  Number  Label  

17  
SECTION  1  
INTRODUCTION  
 
1.2  Instrument  Description    
Figure  4:  Stainer  Bowl  Components  
 
                 1          2        3        4  
 
 
 
 
5    
 
 
 
 
1  –  Nozzle  D  (Methanol  or  Aerofix)  
2  –  Nozzle  C  (Eosin  stain)  
3  –  Nozzle  B  (Thiazin  stain)  
4  –  Nozzle  A  (Rinse)  
5  –  Drive  Hub  
 
Figure  5:  Nozzle  Diagram  
 
 
 
 
 
 
  1   2   3   4  
 
1  –  Compression  Screw  
2  –  Swirl  Cone  
3  –  O-­‐Ring  
4  –  Nozzle  Housing  

18  
SECTION  1  
INTRODUCTION  
 
1.2  Instrument  Description    
Table  7:  Preventive  Maintenance  Kit  
 
Component   Name   Description  
  Manual  Priming  Tool   Primes  air-­‐locked  pumps.  
 
 
 
 
 
 
  Silicon  Grease   Lubricates  the  nozzle  threads  for  
easy  assembly.  
 
 
 
 
  Nozzle  Wire   Cleans  nozzle  housing  orifices.  
 
 
 
 
  Nozzle  Cleaning  Strainer   Strains  the  nozzle  parts  to  prevent  
them  from  going  down  a  drain.  
 

  Nozzle  Tool   Unscrews  nozzles  from  the  stainer  


bowl.  
 
 
 
 
  Nozzle  Wrench   Disassembles  the  nozzle.    
 

  Nozzle  Brush   Cleans  nozzles  without  removing  


them  from  the  stainer.  

 
 
 

19  
SECTION  1  
INTRODUCTION  
 
1.2  Instrument  Description  
Table  7:  Preventive  Maintenance  Kit  (continued)  
 
Component   Name   Description  
  Volume  Test  Collection   Collects  reagents  while  performing  
Tubes  (small  tube)   the  Volume  Test.    
 
 

 
  Nozzle  Maintenance  Tube   Holds  Nozzle  Cleaning  Tubes  (large  
Stand   tube)  and  Volume  Test  Tubes  (small  
tube).  
 
 
 
 
  Nozzle  Cleaning  Tubes   For  soaking  nozzles  in  the  Nozzle  
(large  tube)   Cleaning  Solution.  
 
 
 
 
 
   
Barcode  Reader  
An  optional  barcode  reader  is  available  for  the  Hematology  Stat  Stainer/Cytocentrifuge  
(Model  7122).  
 
Figure  6:  Barcode  Reader  
 
 
 
 
 
 
 

20  
SECTION  1  
INTRODUCTION  
 
1.3  Touchscreen  and  User  Interface  
Users  control  all  instrument  functions  from  the  interactive  touchscreen  display.  
 
Table  8:  Front  Panel/Main  Screen  Function  Keys  
 
Button   Name   Description  
  Standby/Ready   With  instrument  power  ON:    
Blue  =  Ready  
  Amber  =    Standby    
   
Pressing  Standby  runs  a  System  Clean  cycle  and  
places  instrument  into  standby  mode.    
 
The  Standby/Ready  button  also  accesses  the  
touchscreen  calibration  function.  Refer  to  
System  Setup  Menu  (Section  3.1).  
 
  Maintenance   Accesses  features  for  verifying  proper  nozzle  
performance  and  places  pumps  in  a  testing  
sequence.  Accesses  the  line  priming,  Pattern  
Test,  Volume  Test,  and  B-­‐Line  Flush  functions.  
 
  Clean   Performs  the  Clean  cycle.    
 
 
  Cyto   Enters  the  Cytocentrifuge  mode.  
 
 
  System  Information   Shows  the  system  information,  including  serial  
number  and  software  version.  Allows  access  to  
the  System  Setup  features.  Refer  to  System  
Setup  Menu  (Section  3.1).  
   
  Help   Opens  the  software  Help  file.  
 
 
  Programs   Allows  users  to  select  or  edit  programs.  
 
 
 
  Start/Load  Slides   Starts  a  Stain  or  Cytocentrifuge  cycle.  Button  is  
  inactive  until  a  program  is  created.  Refer  to  
  Creating  a  Stain  Program  (Section  3.1).    
 
With  Slide  Tracking  enabled,  opens  the  Scan  
and  Load  Slides  menu  (Section  3.2).  
 
  Number  of  Specimen   Selects  the  number  of  specimen  slides  in  the  
Slides   carousel.  Users  staining  an  odd  number  of  
specimen  slides  should  press  the  next  higher  
specimen  slide  number  icon.  
 

21  
SECTION  1  
INTRODUCTION  
 
1.3  Touchscreen  and  User  Interface  
Table  8:  Front  Panel/Main  Screen  Function  Keys  (continued)  
 
Button   Name   Description  
  Back   Returns  to  the  previous  menu.  
 
  Stop   Aborts  any  operation.  
   

  OK   Indicates  completion  of  current  task.  

  System  Setup   Allows  users  to  modify  the  software  settings.  


See  System  Setup  Menu  (Section  3.1).  
 
 
Table  9:  System  Setup  Keys  
 
Button     Name   Description  
  Stain  Programs   Allows  users  to  create,  edit,  and  delete  stain  
programs.  
 
  Cyto  Programs   Allows  users  to  create,  edit,  and  delete  
cytocentrifuge  programs.  
 
  Reagents   Allows  users  to  edit  reagent  information.  
 
 
  Users   Allows  users  to  create  and  change  user  
accounts.  
 
  QC/Maintenance   Enables  slide  tracking,  preventive  maintenance  
Tracking   tracking,  and  reagent  tracking.  
 
  Level  Detect   Allows  users  to  manage  the  automatic  reagent  
level  detection  system.  
 
  Language   Allows  users  to  change  the  display  language.  
 
 
  System  Log   Allows  users  to  control  logging  functions.  
 
 
  Network  Settings   Allows  users  to  change  network  settings.  
 
 
  Beeper   Allows  users  to  change  audible  alerts.  
 
 

22  
SECTION  1  
INTRODUCTION  
 
1.3  Touchscreen  and  User  Interface  
Table  9:  System  Setup  Keys  (continued)  
 
Button     Name   Description  
  Set  Date/Time   Allows  users  to  set  the  date  and  time.  
 
  Restore  Defaults   Restores  programming  to  default  settings.  
 
 
  Login   Enters  Login  sequence  for  authorized  users.  
   
 
  Logout   Logs  authorized  users  out.  Users  must  log  in  
  again  to  use  the  stainer.  
 
 
  Save   Saves  the  entered  or  selected  information.    
   
 
  Add   Enters  programming  mode  for  creating  staining  
  and  cytocentrifuge  programs.  Also  allows  the  
system  administrator  to  authorize  new  users.    
Allows  manual  entry  of  slide  or  specimen  
information.  
 
  Delete/Erase/Remove   Deletes  or  erases  the  selected  item.  
   
 
  Edit/Change  User   Allows  editing  of  an  existing  stain  or  
  cytocentrifuge  program.  Allows  manual  entry  of  
slide  or  specimen  information  (stain  or  
cytocentrifuge  mode).  Also  allows  system  
administrator  to  edit  user  information.  
 
  Zero   Zeros  the  Level  Detect  sensors.  
 
 
  Calibrate   Calibrates  the  Level  Detect  system.  
 
 
 
  Unselected   Shows  an  unselected  option.  
   
   
Selected   Shows  a  selected  or  enabled  option.  
 
 
 
 
 

23  
SECTION  1  
INTRODUCTION  
 
1.3  Touchscreen  and  User  Interface  
Table  10:  Maintenance  Function  Keys  
 
Button   Name     Description  
  Individual  Prime   Primes  the  selected  line.  
Buttons  (A,  B,  C,  D)    
 
  Pattern  Test   Performs  Pattern  Test  to  ensure  nozzles  are  
clear  of  debris  and  spraying  properly.  
 
  Volume  Test   Performs  Volume  Test  to  verify  the  selected  
nozzle  volume  is  within  the  correct  range.  
 
  60-­‐Sec  Prime   Runs  the  pumps  for  1  minute  and  primes  the  
lines.  
 
  QC/PM   Shows  the  Preventive  Maintenance  and  Quality  
Control  logs,  when  enabled  from  the  System  
Setup  menu  (Section  3.1).  
  Line  Flush   Cleans  the  B  and  C  reagent  lines.  
 

  ABCD  Prime  button   Primes  all  (A,  B,  C,  D)  reagent  lines  
simultaneously.  

24  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.1  Instrument  Setup  
Unpacking  and  Installing  the  Stainer  
Follow  this  sequence  if  you  are  using  this  instrument  for  the  first  time.  Details  about  
these  operations  are  given  in  the  next  three  sections.  
CAUTION:  
Contact  Elitech  Group  before  installing  the  instrument  if  you  observe  any  damage  to  the  
packaging  or  equipment.  
 
1   Unpack  and  inspect  the  instrument.  
 
2   Check  that  the  contents  of  the  boxes  match  the  packing  lists  for  the  
instrument  and  accessories.    
 
3   Open  the  instrument  lid  and  remove  the  cardboard  tube  that  protects  the  
hub.  
 
NOTE:  Keep  the  box  and  packaging  material  to  repack  the  instrument  if  you  
intend  to  ship  it  to  the  manufacturer  for  service.  
 
4   Place  the  instrument  on  a  flat  surface,  free  from  dust  and  vibration  and  
away  from  direct  sunlight.  
 
NOTE:  Position  the  instrument  with  the  rear  panel  at  least  30  cm  (12  in.)  from  
obstructions  or  hazardous  materials.    
 
CAUTION:    
Keep  the  drain  tube  straight  and  as  short  as  possible.  The  maximum  length  is  
1.8  m  (72  in.).  The  waste  container  must  be  positioned  lower  than  the  stainer.    
 

25  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.1  Instrument  Setup  
Connecting  the  Drain  Tube  and  Waste  Container  
1   Insert  the  waste  tube  connector  into  the  rear  panel  receptacle  until  you  
hear  a  click.  
 
 
2   Adjust  the  tube  length  to  less  than  1.8  m  (72  in.).  
NOTE:  Ensure  the  waste  tube  has  no  loops  or  kinks,  and  is  as  straight  and  as  
short  as  possible.  Cut  off  excess  tubing  as  needed.  
 
3   Connect  the  drain  tube  to  the  waste  container.  
 
 
If  using  a  waste  bottle  with  level  detect  (REF:  AC-­‐182):  
4   Connect  the  waste  monitoring  cable  to  the  rear  panel  receptacle.  
 
 
5   Connect  the  waste  monitoring  cable  to  the  waste  container  lid.  
 
 
Connecting  Power  
1   Make  sure  the  power  switch  is  OFF  (O).  
 
 
2     Plug  the  power  cord  into  the  power  connector  on  the  rear  panel  of  the  
instrument.  
NOTE:  Use  a  surge  protector  to  isolate  the  instrument  from  electrical  spikes  and  
surges.    
 
3     Plug  the  power  cord  into  a  properly  rated  AC  electrical  outlet.  
 
 
4     Turn  the  power  switch  ON  (I).  After  a  brief  delay  the  Main  menu  will  
appear.  
 
   
 
 
 
 
 
 

26  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.1  Instrument  Setup  
Installing  Reagent  Bottles  
WARNING!  
Reagents  used  in  the  instrument  contain  moderately  hazardous  chemicals  that  require  
care  in  handling.  Always  handle  reagents  using  appropriate  safety  measures,  including  
gloves  and  eye  protection.  
NOTE:  Reagents  should  be  stored  according  to  the  conditions  specified  on  their  label.  
After  opening,  reagents  are  stable  until  the  expiration  date  specified  on  the  label  unless  
otherwise  indicated  by  the  symbol  shown  at  left.  
1 Place  each  500  mL  reagent  bottle  in  the  correct  position.    
(A)  Rinse  
(B)  Thiazin  Stain  
  (C)  Eosin  Stain    
(D)  Methanol  

NOTE:  See  Appendix  A  for  complete  identification  of  all  reagents  used  in  this  
   A          B                        C        D   stainer.    
 
CAUTION:  
To  avoid  severe  damage,  never  use  reagents  containing  organic  solvents  (such  
as  ketones)  in  this  instrument,  unless  supplied  by  Elitech  Group,  or  specified  in  
official  Elitech  Group  formulation  instructions.  
 
NOTE:  Immediately  remove  spills  in  the  reagent  tray  to  preserve  the  accuracy  of  
the  reagent  level  detecting  system.    
 
2   For  all  reagents:  
• Open  a  new  bottle  of  reagent.  
• Record  the  reagent  letter  on  each  cap.  
• Insert  the  corresponding  dip  tube  into  the  reagent  bottle  and  install  
the  ring  cap.  
 

27  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.1  Instrument  Setup  
Installing  the  Barcode  Reader  
A  barcode  reader  can  be  connected  to  the  stainer  for  scanning  reagent  bottles  and  
specimen  slides  that  contain  barcodes.  This  allows  easy  reagent  and  specimen  
information  tracking.  If  a  barcode  reader  is  not  installed,  reagent  and  specimen  
information  can  be  entered  manually  (Section  3.2).    
 
Installing  the  Barcode  Reader  
1   Place  the  barcode  reader  and  stand  on  a  level  surface  near  the  stainer.  
 
2 Plug  the  barcode  reader  into  the  left  USB  port  on  the  rear  panel  of  the  
stainer.  See  Section  3.2  for  instructions  on  using  the  barcode  reader.  
 

28  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
Priming  Procedures  
NOTE:  The  instrument  is  shipped  with  alcohol  in  the  reagent  lines.  For  proper  
performance,  this  alcohol  must  be  replaced  with  the  correct  reagent  for  each  reagent  
line  prior  to  use.  
Thoroughly  purge  and  prime  each  reagent  delivery  line  using  the  following  instructions.    
1   Remove  each  spray  nozzle  with  the  provided  nozzle  tool  by  turning  
counterclockwise.    
 
 
D                        C                        B                      A   2   Note  the  location  of  each  nozzle  so  you  can  return  it  to  the  original  
position  during  reassembly.  
 
 
 
3     Place  a  carousel  on  the  stainer  hub  to  prevent  stain  from  entering  the  
motor  shaft.    
 
CAUTION:  
Fluid  from  priming  can  flood  and  damage  the  motor  if  the  drain  tube  is  not  
properly  installed.      
 
 
4   Press  Maintenance  from  the  Main  menu.  
 
 
5   Press  the  A  prime  button.  Stain  should  appear  within  10  seconds.  When  
properly  primed,  a  steady  stream  of  reagent  (no  sputtering  or  breaks)  
flows  from  the  nozzle  receptacle.  
 
• If  stain  appears,  proceed  to  the  next  step.  
 
• If  stain  does  not  appear  within  10  seconds,  perform  the  manual  
priming  procedure  (Section  6.3).  

CAUTION:    
Never  operate  a  dry  pump  for  more  than  10  seconds.  Operating  a  dry  pump  
may  cause  damage  to  the  instrument.  
 
 

29  
SECTION  2  
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
Priming  Procedures  (continued)  
6     Repeat  the  previous  steps  for  each  reagent  line  (B,  C,  and  D).    
 
 
7     Press  60-­‐Sec  Prime  to  prime  each  reagent  line  with  200  mL  of  reagent  to  
remove  all  of  the  alcohol  from  the  reagent  lines  and  pumps.  
 
 
8   To  prime  all  lines  at  the  same  time,  press  ABCD.  The  pumps  will  run  for  1  
minute  and  prime  all  the  lines.  Follow  instructions  on  the  screen.  
 
 
NOTE:  To  prime  individual  lines,  press  the  appropriate  individual  prime  button  
(A,  B,  C,  D).    
   
     D    C      B      A    
9 Return  the  nozzles  to  their  original  positions  and  tighten  clockwise  with  
the  nozzle  tool.    
 
 
10 With  the  nozzles  installed,  repeat  Steps  5  and  6.  A  fine  cone  of  spray  
should  come  from  each  nozzle.    
 
 
11 After  verifying  nozzle  performance,  a  clean  cycle  must  be  performed  
before  a  stain  cycle  can  be  run.    

30  
SECTION  2  
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
The  Clean  Cycle  
The  Clean  cycle  is  a  two-­‐step  process  that  cleans  the  reagent  nozzles  and  the  carousel.  
The  first  step  purges  each  reagent  line  and  cleans  the  empty  carousel  with  methanol  or  
Aerofix  while  cycle  progress  is  displayed  on  the  screen.  The  second  step,  which  can  be  
delayed  indefinitely,  reprimes  the  reagent  lines  with  staining  reagents.  
NOTE:  Pressing  Standby/Ready  performs  a  System  Clean  cycle  before  the  instrument  
goes  on  standby.  
1   Place  an  empty  carousel  in  the  instrument  and  close  the  lid.  
 
CAUTION:  
Never  place  any  carousel  loaded  with  specimens  in  the  instrument  for  a  clean  
cycle  (including  placing  the  instrument  in  standby  mode).  Specimens  will  be  
damaged  if  they  contact  reagents  sprayed  from  the  nozzles  when  you  press  
Clean  or  Standby.  
 
2   Press  Clean.  
 
 
NOTE:  Pressing  Stop  during  the  Clean  cycle  causes  the  Clean  Cycle  Cancelled  
message  to  be  displayed.  Press  Clean  to  complete  the  interrupted  cycle.    
 
 
 
 
 
 
 
3   Press  Reprime  to  reprime  the  lines.  
 
 
 
e  
 
 
 
4   Open  the  lid  and  remove  the  carousel  when  the  Clean  cycle  is  complete.  
 
 
5   Spray  the  interior  of  the  bowl  with  70  to  100%  methanol  or  ethanol.  Wipe  
the  stainer  bowl  dry  with  paper  towels.  
 
NOTE:  Perform  the  instrument  storing  procedure  (Section  5.2),  if  the  instrument  
will  remain  idle  for  more  than  1  week.    
 
Performing  Tests  
We  recommend  performing  the  Pattern  Test  and  Volume  Test  before  using  the  stainer  
(see  Section  6).  

31  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
Reagent  Level  Monitoring  
Reagent  Level  Detect  monitors  reagent  levels  and  alerts  you  when  the  reagent  is  
running  low,  or  when  the  waste  container  is  full  (when  using  the  waste  container  with  
level  detect  (AC-­‐182)).  You  can  turn  reagent  and  waste  container  monitoring  ON  or  OFF  
from  the  Level  Detect  menu.  The  system  default  is  ON  for  reagent  monitoring  and  OFF  
for  waste  container  monitoring.    
 
NOTE:  The  instrument  must  be  installed  on  a  flat,  level  surface  for  accurate  reagent  
monitoring.  
 
CAUTION:  
This  system  is  designed  to  warn  you  when  the  reagent  level  is  getting  low.  The  
instrument  will  continue  running  through  these  warnings.  Always  monitor  and  replenish  
(if  needed)  the  reagent  before  running  a  stain  cycle.    
 
Enabling/Disabling  Reagent  Level  Detect  
 1   Press  Information  from  the  Main  menu.  
 
 
 
2   Press  System  Setup.    
 
 
 
3     Press  Level  Detect.  The  display  shows:  
 
 
 
 
 
 
 
 
4     Press  Tray  to  enable,  or  OFF  to  disable  a  reagent  line.  Functions  are  grey  
when  unselected,  blue  when  selected.  Press  External  to  enable  level  
detect  for  the  waste  bottle.  
 
 
 
5   When  finished,  press  Back  to  exit  to  System  Setup  menu.    
 

32  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
Zeroing  the  Reagent  Level  Sensors  
The  Level  Detect  function  must  be  zeroed  at  initial  setup,  when  the  stainer  is  
moved,  or  if  the  level  detect  is  not  reporting  correctly.  If  zeroing  does  not  
correct  the  problem,  recalibrate  the  Level  Detect  function  (Section  7.3).  
 
1     Press  Information.  
 
 
 
2   Press  System  Setup.  
 
 
 
3 Press  Level  Detect  to  enter  the  Reagent  Level  Detect  Setup  menu.  

 
 

 
 
 
 
 
 
NOTE:  The  stainer  should  be  turned  ON  for  at  least  30  minutes  before  zeroing  to  
stabilize  level  sensors.  The  instrument  can  be  used  during  this  time.  
 
4   Press  Zero.  The  display  shows:    

 
 
 
 
 

33  
SECTION  2    
SETUP  AND  PREPARATION  FOR  OPERATION  
 
2.2  Preparing  the  Stainer  for  Operation  
Zeroing  the  Reagent  Level  Sensors  (continued)  
5     Remove  all  reagent  bottles  and  press  Start.  
 
 
 
 
 
 
 
 
 
NOTE:  Vibrations  or  bumps  to  the  instrument  or  lab  bench  can  cause  
inaccuracies  in  zeroing  or  calibration.    
 
6   After  zeroing,  press  OK.  Press  the  Back  button  to  exit  to  the  System  Setup  
menu.  
 
 
 
 
7   Return  the  reagent  bottles  to  their  correct  positions  in  the  tray.  
 
NOTE:  For  accurate  reagent  level  detection  and  calibration,  dip  tubes  must  
follow  their  pre-­‐formed  coiled  shapes.  
 

34  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Many  software  settings  can  be  controlled  from  the  System  Setup  menu,  including:  
• Creating,  editing,  and  deleting  stain  programs  
• Creating,  editing,  and  deleting  cytocentrifuge  programs  
• Tracking  reagent  information  
• Managing  user  accounts  
• Enabling  tracking  features  for  slides,  maintenance,  and  reagents  
• Managing  reagent  level  sensing  
• Changing  the  display  language  
• Viewing  and  exporting  the  system  log  
• Changing  beeper  settings  
• Setting  the  date  and  time  
• Restoring  default  settings  

Accessing  the  System  Setup  Menu  


1     Press  System  Information  from  the  Main  menu.  
 
 
 
2    Press  System  Setup.  
 
 
 
 
 
 
 
 
 
 
Stain  Programs  
Stain  Programs  allows  the  user  to  create,  edit,  or  erase  staining  programs  according  to  the  
user’s  specific  staining  requirements.  Up  to  12  staining  programs  can  be  programmed  and  
stored  in  memory.    
 
Programmable  staining  parameters  include  stain  intensity  and  fixation  options.  
 
• Stain  Intensity  
Selections  range  from  1  to  10.  1  is  the  lightest  and  10  the  darkest  setting.  
Settings  8,  9,  and  10  are  particularly  appropriate  for  bone  marrow  specimens.  

• Fixation  
Fixation  options  are  Off,  Normal,  and  High.  

35  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Creating  a  Stain  Program  
1 From  System  Setup,  press  Stain  Programs.    
 

 
 
 
2   Press  Add.  
 
 
3   Select  Enter  Program  Name,  and  enter  the  desired  name  on  the  keypad.  
 
 
 
 
4   Press  Enter  on  the  keypad.  The  display  returns  to  the  Stain  Programming  
Menu.  
 

5   Select  Intensity  From  the  Stain  Programming  Menu.  Select  the  desired  
stain  intensity  (1-­‐10).  

6   Select  the  desired  Fixation  (Off,  Normal,  High).  

7   Press  Save.  

 
   Editing,  Renaming,  or  Adjusting  Stain  Programs  
1   From  Stain  Programs  menu  select  the  program  to  be  modified.  
 
 
 
 
2    Press  Edit.  

36  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Editing,  Renaming,  or  Adjusting  Stain  Programs  (continued)  
3   Adjust  the  settings  as  needed.  
 
 
 
 
4   Press  Save.  
 
 
 
Administrator  and  User  Accounts  
You  can  create  one  Administrator  account  and  multiple  (up  to  50)  user  accounts.  The  
Administrator  controls  access  to  the  system  by  adding  and  editing  user  accounts.  Users  
cannot  edit  System  Settings  unless  permitted  by  the  Administrator.  
 
Creating  an  Administrator  Account  
1   From  System  Setup,  press  Users.  
 
 
2   Select  Lock  System  Setup  Access.  
 
 
 
 
 
 
 
 
 
3   Enter  a  password  for  the  Administrator  account  (at  least  4  characters)  and  
press  Enter.  
 
 
 
 
 
 
 
 
 
4   Re-­‐enter  the  password  and  press  Enter  to  confirm.  
 
 
 

37  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Creating  User  Accounts  
Note:  This  function  is  available  only  if  an  administrator  account  has  been  
created.  
1 Press  System  Setup.  
 
 
2 Enter  the  Administrator  password.  
 
 
 
 
 
 
 
 
3   Press  Enter.  
 
 
 
4   Press  Users  to  reveal  the  Manage  Users  menu.  
 
 
 
 
5   Select  Enable  Global  Login.  
 
 
 
 
 
 
 
 
6   Select  Add  User.  
 
 
 
 
7   Enter  a  new  user  name.  
 
 

8   Press  Enter.  

38  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Creating  User  Accounts  (continued)    
9   Enter  a  numeric  passcode  (at  least  4  numbers)  for  the  user  account.  
 
 
 
 
 
 
10   Press  Enter.  
 
 
 
11   Re-­‐enter  the  passcode  to  confirm.  
 
 
12   Press  Enter.  
 
 
 
 
 
Managing  User  Access  
From  the  Manage  Users  screen,  the  Administrator  has  several  options  to  
manage  user  access  to  the  instrument.  
• Enable  Global  Login  allows  users  to  log  in  to  the  instrument.  Users  will  log  
out  manually  or  automatically  (with  user-­‐selectable  time  options).  See  
User  Login/Logout  below.  
• Enable  Run  Login  requires  the  current  user  to  enter  a  password  to  run  a  
Stain  or  Cytocentrifuge  cycle.  Global  Login  must  be  enabled  to  use  this  
option.  
• User  System  Access  enables  complete  control  of  the  instrument,  including  
changing  the  System  Setup  options.  This  option  can  be  controlled  on  an  
individual  user  basis,  if  Global  Login  is  enabled.  
 
User  Login/Logout  
With  System  Access  locked  and  Global  Login  enabled,  users  must  log  in  to  use  
the  stainer:  
 
1 Select  User  ID  and  Select  Logout  time  after  idle  from  the  drop-­‐down  
menu.  
 
NOTE:  Users  can  select  how  long  the  stainer  can  be  idle  before  automatically  
logging  the  user  out.    
 

39  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
User  LogIn/Logout  (continued)  
2 Press  Login.  
 
 
 
 
3 Enter  the  correct  passcode  for  the  selected  user  and  press  Enter.  
 
 
 
 
4 Once  Login  is  complete,  the  display  advances  to  the  Main  screen  and  the  
instrument  is  ready  for  programming  and  staining.  A  Logout  button  and  
the  user  name  appear  at  the  top  right  of  the  Main  screen.  
 
NOTE:  Users  can  log  out  manually  by  pressing  the  logout  button.  

40  
SECTION  3  
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Using  Reagent  Information  Tracking  
You  can  enter  reagent  information  to  help  track  reagent  usage  and  expiration.  Reagent  
information  includes  reference  number,  expiration  date,  lot  number,  date  and  time  the  
reagent  was  last  installed.  
1   From  System  Setup,  press  QC/Maintenance  Tracking.  
 
2   Select  Enable  Reagent  Tracking  by  choosing  reagent  A,  B,  C,  or  D.  This  
enables  reagent  lot  number  and  expiration  date  tracking.  
 
3   Press  Back  to  return  to  System  Setup.  
 
4   Press  Reagents.  
 
5   Press  Change  next  to  the  appropriate  reagent.  
 
6   Scan  the  reagent  bottle  barcodes  (Section  3.2)  or  manually  enter  the  
reagent  information  in  the  correct  fields.  
 
7   Press  Save  for  each  reagent.  
 
Modifying  Level  Detect  Functions  
The  Level  Detect  function  alerts  you  when  the  reagent  is  running  low,  or  when  the  
waste  container  is  almost  full.  You  can  turn  reagent  and  waste  container  monitoring  ON  
and  OFF  from  the  System  Setup  screen.  The  system  defaults  to  ON  for  reagent  
monitoring  and  to  OFF  for  waste  container  monitoring.  See  Section  2.2  for  complete  
instructions.  
 
1   From  System  Setup,  press  Level  Detect.  
 

2   Select  the  reagent  monitoring  options  to  be  modified.  


• To  disable  monitoring,  press  OFF  next  to  the  appropriate  reagents.  
• To  enable  monitoring,  press  Tray  next  to  the  appropriate  reagents.  
• To  monitor  the  waste  container,  press  External.  
 

41  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Changing  User  Language  
1   From  System  Setup,  press  Language.  

2   Select  the  software  language  from  the  list  on  the  left.  

3   Press  OK.  
 
 
Setting  the  Date  and  Time  
1   From  System  Setup,  press  Set  Date/Time.  

2   Select  12  for  a  12-­‐hour  clock  or  24  for  a  24-­‐hour  clock.  

3   Use  the  up  and  down  arrows  to  modify  the  time  and  date.  

4   Press  Save.  
 
 
 
System  Log  
The  instrument  records  all  login,  logout,  stain  or  cytocentrifuge  cycles,  setting  changes,  
maintenance  functions  and  specimen  identification  (if  enabled).  
 
Accessing  Logs  
1   From  System  Setup,  press  System  Log.  
 
2   Use  navigation  arrows  to  scroll  through  the  log.  
 
 

42  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
System  Log  (continued)  
Exporting  Logs  
1 From  System  Setup,  press  System  Log.  
 
 
2   Plug  a  Flash  Drive  into  the  right  USB  port.    
 
 
3   Press  Export.    
NOTE:  The  log  files  are  exported  to  the  Flash  Drive  as  a  CSV  file  that  can  be  used  
in  spreadsheet  software  programs.    
 
 
 
 

Controlling  Beeper  Alerts  


1   From  System  Setup,  press  Beeper.  
 
 

2   Use  the  sliders  to  modify  the  beeper  volume  for  Cycle  Complete,  Warnings,  
Errors,  or  Key  Clicks.  
 
 
3   Press  Beep  On  Startup  to  turn  the  audible  startup  alert  ON  or  OFF.  
 

43  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
QC/Maintenance  Tracking  
Under  system  default  settings,  the  following  QC/Maintenance  Tracking  options  are  disabled:  
• Stain  Slide  Tracking  
• Cyto  Slide  Tracking  
• Manual  Entry  
• Preventive  Maintenance  Tracking  
• Reagent  Tracking  
Enable  Stain  Slide  Tracking  
To  activate  Stain  Slide  Tracking:  
1   From  System  Setup,  Press  QC/Maintenance  Tracking.  
 
2   Press  Enable  Stain  Slide  Tracking.  

 
 
 
 
3   Press  Back  twice  to  return  to  the  main  screen.  Verify  that  the  Start  Button  
on  the  main  screen  reads  “Load  Slides.”  
NOTE:  Selecting  Enable  Stain  Slide  Tracking  changes  the  Start  button  on  the  
Main  menu  to  “Load  Slides.”  
4   Press  Load  Slides.  The  Scan  and  Load  Slides  menu  appears.  
 
 
 
 
 
 
5   Enter  slide  information.  
a.   If  using  the  barcode  reader,  scan  the  specimen  slides  that  contain  
barcodes.  See  Scanning  Slides  with  the  Barcode  Reader  (Section  3.2).  
b.     If  entering  specimen  information  manually,  see  Manually  Entering  
Specimen  Information  (Section  3.2).    
6   See  Section  4  for  remaining  steps  for  running  a  stain  cycle.  

44  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Enable  Cyto  Slide  Tracking  
Allows  slide  tracking  in  cytocentrifuge  mode.  See  the  Cytopro  Rotor  
Applications  Manual  (Aerospray  Models  7xx2)  (REF:  RP-­‐517)  for  complete  
information.  
 
Enable  Manual  Entry  
If  selected,  allows  manual  entry  of  slide  information  using  the  keypad  (limited  
to  24  characters).  
 
Enable  Preventive  Maintenance  Tracking  
To  activate  the  tracking  prompts  for  Preventive  Maintenance  Tracking,  use  the    
following  steps:  
1   From  System  Setup,  select  QC/Maintenance  Tracking.  
 
 
2   Select  Enable  Preventive  Maintenance  Tracking.  
 
 
 
 
 
3   Enter  the  information  for  the  Daily,  Weekly,  and  QC  Slide  prompts  in  
corresponding  fields.  See  Using  the  Preventive  Maintenance  Log  (Section  
5.1).  
 
Enable  Reagent  Tracking  
To  activate  Reagent  Tracking:  
1   From  the  QC/Maintenance  Tracking  menu,  select  the  reagent  (A,  B,  C,  D)  
to  be  tracked.  
 

45  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.1  System  Setup  Menu  
Restoring  Software  Defaults  
1   From  System  Setup,  select  Restore  Defaults.  
 
CAUTION:    
Restoring  the  system  defaults  will  remove  all  personal  settings.
• Restoring  System  Settings  will  delete  all  user  names  and  passwords.  
• Restoring  Stain  Settings  will  delete  all  stain  programs  and  restore  the  
default  programs.  
• Restoring  Cytocentrifuge  Settings  will  delete  all  cytocentrifuge  
programs  and  restore  the  default  programs.  
 
2   Select  the  settings  you  would  like  to  restore  to  factory  defaults:  System  
Settings,  Stain  Settings,  or  Cytocentrifuge  Settings.  
 
3 Press  Restore.  
 
 
 
 
 
 
 
 
 
 
4   The  display  returns  to  the  Main  menu.  
 

46  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.2  Recording  Specimen  and  Reagent  Information  
Scanning  Slides  with  the  Barcode  Reader  
1 From  System  Setup  select  QC/Maintenance  Tracking.  

 
2 Select  Enable  Stain  Slide  Tracking.  

NOTE:  Selecting  Enable  Stain  Slide  Tracking  changes  the  Start  button  on  the  
Main  menu  to   Load  Slides.”  
 
 
 
3 Press  Back  twice  to  return  to  the  Main  menu.  
 
 
4 Press  Load  Slides  on  the  Main  menu.  The  Scan  and  Load  Slides  menu  will  
appear.  
 
 
 
 
 
 
 
5 Scan  the  barcode  of  each  slide  in  the  batch  and  load  into  the  carousel  
according  to  instructions  in  Section  4.1.  

 
 
 
6 Verify  that  each  barcode  appears  on  the  Scan  and  Load  Slides  menu.  
 

 
 
 
 
7   When  you  have  completed  preparations  to  stain,  (Section  4.1)  press  Start.  

47  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
3.2  Recording  Specimen  and  Reagent  Information  
Scanning  Reagent  Bottles  with  the  Barcode  Reader  
1 From  System  Setup  select  QC/Maintenance  Tracking.  
 
 
2 Select  Enable  Reagent  Tracking  for  each  desired  reagent  (A,  B,  C,  D).  

 
 
 
 
 
 
 
3 Press  Back  to  return  to  the  System  Setup  menu.  
 

 
4 Press  Reagents  to  reveal  the  Reagent  Information  screen.  

 
 
 
 
 
 
 
 
5 Select  the  desired  Reagent  (A,  B,  C,  D)  and  press  Change.  

 
 
 
6 Scan  the  barcode  of  each  enabled  reagent  bottle.    

 
 
 
 
7   Verify  that  the  barcode  appears  on  the  Scan  and  Load  Slides  menu.  
 
 

48  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.2  Recording  Specimen  and  Reagent  Information  
Scanning  Reagent  Bottles  with  the  Barcode  Reader  (continued)  
8   Press  Save.  
 
9   Repeat  steps  3-­‐8  for  each  reagent  bottle  that  is  enabled  in  QC  
Maintenance  Tracking.  
 
NOTE:  You  can  access  Reagent  Information  by  pressing  the  bottle  icons  on  the  
right  side  of  the  Main  menu.  This  takes  you  directly  to  Reagent  Information  
menu,  where  you  can  scan  or  manually  enter  reagent  information  by  pressing  
Change.  
 
Manually  Entering  Specimen  Information  
With  Stain  Slide  Tracking  and  Manual  Entry  enabled  in  the  QC  Maintenance  menu:  
 
1 Press  Load  Slides  on  the  Main  menu.  
 
 
2 Press  Add  to  reveal  the  keypad.  
 
 
3 Enter  slide  information  (maximum  of  24  characters)  and  press  Enter.  
 
 
4 To  change  or  delete  the  entry,  select  the  entry  on  the  display  and  press  
Edit  or  Remove.  
 
 
 
 
5 Load  slides  and  run  stain  cycle  as  shown  in  Section  4.1.  
 
NOTE:  Reagent  REF  number  must  be  a  valid  Elitech  Group  REF  number  for  
the  selected  reagent.  Incorrect  entries  will  generate  an  error  message.  
 

49  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.2  Recording  Specimen  and  Reagent  Information  
Manually  Entering  Reagent  Information  
1 Press  Reagents  from  the  System  Setup  menu,  or  press  the  reagent  status  
icon  on  the  Main  menu  to  reveal  the  Reagent  Information  menu.  
 
 
 
 
 
 
 
 
 
 
 
2 Select  the  desired  reagent  and  press  Change.  
 
 
 
 
 
 
 
3 Select  the  desired  field  (Reagent  REF,  Expiration  Date,  Lot  Number,  or  
Service  Date/Time).  
 
 
 
 
 
 
4 Enter  the  information  on  the  keypad  and  Press  Enter.  
 
 
5   Press  Save.  

50  
SECTION  3    
CONTROLLING  AND  CUSTOMIZING  STAINER  FUNCTIONS  
 
3.3  The  Help  Menu  
The  Help  menu  is  a  comprehensive  onscreen  help  function  that  provides  detailed  information  on  
the  following  subjects:  
 
 
 
Help  Screens  
Basic  Operation  
• Loading  the  Carousel  
• Programming  Number  of  Slides  
• Select  Intensity  Setting  
• Correct  Reagents  and  Locations  
• Selecting  a  Staining  Program  
Touchscreen  
Calibrating  the  Touchscreen  
Cleaning    
Clean  Cycle  

 
Using  Help    
1   Press  Help  to  access  the  help  function.    
 
2   Select  the  desired  topic.  
 
3   Use  the  direction  arrows  to  navigate.    
 
4   Press  Exit  to  return  to  the  Main  menu.    

51  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
WARNING!  
Treat  slides  in  accordance  with  good  laboratory  practices  and  local  regulations.  
 
Suggested  Staining  Protocol  
• Hub  Pattern  Test  (once  per  day).  
• Select  or  verify  desired  stain  program.  
• If  slide  tracking  is  enabled,  scan  or  enter  slide  information.  
• Load  slides  into  the  carousel.  Use  blocking  slides  if  needed.    
• Place  loaded  carousel  into  the  stainer  and  close  the  lid.  
• Check  reagent  and  waste  levels.  
• If  slide  tracking  is  not  enabled,  enter  the  number  of  slides  on  the  Main  menu.  
• Perform  a  stain  cycle.  
• Unload  the  carousel.  

Performing  a  Hub  Pattern  Test  


Use  the  Hub  Pattern  Test  to  ensure  the  nozzles  are  clear  of  debris  and  spraying  properly.  
1 From  the  Maintenance  menu,  Press  Pattern  Test.  
 

2 Hold  a  sheet  of  white  paper  towel  near  the  drive  hub,  squarely  facing  the  
target  nozzle.  
 
3 Press  the  corresponding  prime  button.  
 

Check  the  pattern.  If  incorrect,  clean  the  nozzle  orifice  with  the  nozzle  brush  provided  in  
the  Nozzle  Maintenance  Kit.  If  this  fails  to  correct  the  problem  refer  to  Nozzle  
Maintenance  and  Performance  (Section  6).    
 
Figure  7:  Correct  Hub  Pattern  Test  Result  
 
 
 
 
 
 
 
Figure  8:  Incorrect  Hub  Pattern  Test  Result  
 
 
 
 
 

52  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
Loading  the  Carousel  
CAUTION:    
Never  load  chipped  or  cracked  slides  into  the  instrument.  Slides  in  poor  condition  may  
break  during  the  staining  cycle.  If  a  slide  breaks  in  the  bowl,  see  Cleaning  Broken  Slides  
(Section  5.4).  
 
CAUTION:    
Keep  small  ferrous  metal  objects  away  from  the  lab  bench.  These  objects  can  be  
attracted  to  the  magnets  on  the  bottom  of  the  carousel  and  cause  damage  if  spun  free  
during  instrument  operation.  
 
CAUTION:    
Load  slides  in  balanced  pairs.  If  staining  an  odd  number  of  slides,  use  a  blank  slide  to  
balance  the  carousel.  
 
NOTE:  Load  the  carousel  with  similar  specimens  for  a  similar  level  of  staining.  There  is  no  
guarantee  of  staining  performance  when  dissimilar  specimens  are  used.  
 
1   Remove  the  carousel  from  the  bowl  and  place  it  on  a  solid,  level  surface.  
 
 
 
 
2   Remove  the  carousel  lid  by  pressing  the  button  and  lifting  the  lid.  
 
 
 
3   If  Slide  Tracking  is  enabled,  press  Load  Slides.  
• If  using  the  barcode  reader,  scan  each  specimen  slide  barcode  before  
loading  it  into  the  carousel.  Slide  Tracking  must  be  enabled  from  the  
System  Setup  menu.  See  Enable  Stain  Slide  Tracking  (Section  3.1).  
 
• If  entering  slide  information  manually,  follow  the  instructions  in  
Section  3.2.  

 
4   Insert  the  slides  into  the  carousel  with  the  first  slide  in  position  1.  
• Load  slides  in  balanced  pairs  (directly  opposite  one  another)  to  
balance  the  carousel.  If  staining  an  odd  number  of  slides,  use  a  blank  
slide  to  balance  the  carousel.  
 
• If  there  are  empty  slots  in  the  carousel,  use  blocking  slides  to  prevent  
overspray  (see  below).  

53  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
Loading  the  Carousel  (continued)  
• Load  slides  with  the  labels  toward  the  outside  edge  of  the  carousel.  
• Always  load  slides  with  the  specimen  facing  clockwise.  
• Always  place  the  first  slide  in  position  1,  the  second  in  position  2,  and  
so  on.  
NOTE:  A  warning  will  sound  during  the  staining  cycle  if  the  carousel  is  
unbalanced.  
 
 Figure  9:  Loading  the  Carousel    
 
  Labeled  End  Away  
  From  Center  Hub  
 
  Specimen  Side    
 
Figure  10:  Specimen  Placement  on  Slide  
 
 
 
 
 
5     Replace  the  carousel  lid  by  pressing  the  button  and  lowering  the  lid  over  
the  indexing  posts.  
 

 
 
 
6   Release  the  button  and  press  the  lid  handle  until  it  is  firmly  closed  and  
locked.  

54  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
Using  Blocking  Slides  
If  the  carousel  is  not  full,  blank  slides  should  be  used  as  blocking  slides.  Blocking  slides  
prevent  overspray  of  reagents  onto  the  specimen  slides.  Overspray  can  cause  slides  to  
become  over-­‐stained.  
 
• Place  a  blocking  slide  in  front  of  position  1  and  2.  

Figure  11:  Using  Blocking  Slides  


 
B  
 
 
3  
 
  2   1  
 
  4  
 
 
B  –  Blocking  Slide    
1  –  Slide  Position  1   B  
2  –  Slide  Position  2  
3  –  Slide  Position  3  
4  –  Slide  Position  4  
 
Performing  a  Stain  Cycle  
1   Insert  a  carousel  loaded  with  specimen  slides  and  close  the  instrument  lid.  
 
 
2   If  you  have  not  enabled  Slide  Tracking,  select  the  number  of  slides  to  be  
stained.  Slide  selection  defaults  to  full  carousel  at  the  end  of  the  run,  after  
pressing  Stop,  or  selecting  a  number  greater  than  the  full  carousel  default.  
 
NOTE:  To  stain  an  odd  number  of  specimen  slides,  select  the  next  higher  
number  listed  on  the  display.  For  example:  to  stain  3  slides,  select  4,  etc.  
 
 
 
 
 
NOTE:  If  entering  slide  information  by  barcode  reader  or  keypad,  the  number  of  
slides  is  programmed  automatically.  Adjust  the  total  number  of  slides  if  adding  
other  specimen  slides  that  have  not  been  entered  by  barcode  reader  or  keypad.  
 
NOTE:  Do  not  include  blocking  slides  in  the  total  number  of  slides.  
 

55  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
Performing  a  Stain  Cycle  (continued)  
3   If  you  have  created  a  stain  program,  and  it  appears  on  the  display,  proceed  
to  Step  4.  If  the  desired  program  does  not  appear  on  the  display,  press  
Programs.  Then  select  the  desired  program  and  proceed  to  Step  4.  
   
NOTE:  No  special  collection,  pre-­‐treatment,  or  storage  conditions  are  required  
for  specimen  types.  Specimens  that  are  normally  manually  hematology  stained  
can  be  stained  with  the  Hematology  Stat  Stainer.  Adjust  stain  settings  to  
maximize  staining  quality  for  each  specimen  type  (see  Section  3.1).  
4   Press  Start.  The  display  shows  the  progress  of  the  program,  and  a  signal  
tone  (if  enabled)  indicates  the  end  of  the  cycle.  
 
NOTE:  Use  the  emergency  Stop  button  when  required,  for  example,  if  there  is  
abnormal  vibration  or  noise.  This  will  abort  the  staining  cycle.  
 
 
 
Unloading  the  Carousel  
 
1 Remove  the  carousel  from  the  bowl  and  place  it  on  a  solid,  level  surface.  
 

 
2 Remove  the  carousel  lid  by  pressing  the  button  and  lifting  the  lid.  
 

 
3   Carefully  remove  each  slide  and  observe  the  specimen  using  a  microscope.  

56  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
Monitoring  Reagent  and  Waste  Levels  
If  enabled,  the  stainer  displays  the  approximate  reagent  and  waste  container  levels  and  
other  information.    
 
CAUTION:    
You  must  monitor  the  reagent  and  waste  container  levels  on  the  display  (if  enabled)  and  
by  direct  inspection  of  the  bottles.  The  monitor  will  show  the  approximate  level  of  each  
reagent.  This  can  be  compared  to  the  actual  level  in  the  bottles.    
     
• Never  allow  a  reagent  to  run  dry.  When  the  reagent  level  is  near  empty,  replace  the  
reagent  bottle  with  a  new  one  (see  below).  
 
• Never  allow  the  waste  container  level  to  go  above  the  maximum  safety  level.    

Table  11:  Reagent  Level  Detect  Display  Symbols  


  Reagent  unselected  in  Level  Detect  
 
 
  Reagent  Bottle  Full  
 
 
  Reagent  Bottle  2/3  Full  
 
 
  Reagent  Bottle  1/3  Full  
 
 
  Reagent  Bottle  Empty  
 
 
  Measurement  Error  (internal  level  
detect  problem)  
 
  Reagent  has  exceeded  expiration  
date  (enabled  from  QC  Maintenance  
menu)  

  Waste  bottle  empty  


 
 
  Waste  bottle  error  (such  as  external  
bottle  unplugged)  
 
  Waste  bottle  full    
 
 
 

57  
SECTION  4    
OPERATING  THE  STAINER  
 
4.1  Operating  Instructions  
NOTE:  Access  the  Reagent  Information  menu  by  pressing  the  bottle  icons  on  the  right  
side  of  the  Main  menu.  Press  Change  to  scan  or  manually  enter  reagent  information.  
 
NOTE:  Do  not  put  residual  reagent  from  a  used  bottle  into  a  new  bottle.  This  can  lead  to  
an  accumulation  of  residue  on  the  slides  and  may  be  a  source  of  contamination.  
 
WARNING!  
Reagents  used  in  this  instrument  contain  moderately  hazardous  chemicals  that  
require  care  in  handling.  Always  use  appropriate  safety  measures,  including  gloves  
and  eye  protection,  when  handling  reagents.  
 
Replacing  a  Reagent  Bottle  
1   Remove  the  empty  reagent  bottle  from  the  tray  but  do  not  disconnect  the  
dip  tube.  
 
2   Open  the  new  bottle  and  record  the  letter  on  the  cap  for  future  use,  such  
as  long-­‐term  storage.  
3   Open  the  Reagent  Information  menu  by  pressing  the  reagent  bottle  icon  
on  the  right  side  of  the  Main  menu.  
 
4     Select  the  desired  reagent  and  press  Change.  
 
5   If  you  are  using  reagent  tracking,  scan  the  barcode  or  manually  enter  the  
reagent  REF,  expiration  date  and  lot  number  in  the  Reagent  Information  
menu  (Section  3.2).  
 
6   Unscrew  the  cap  and  remove  the  dip  tube  from  the  empty  bottle.  
 
7   Insert  the  dip  tube  into  the  new  reagent  bottle  and  screw  on  the  cap.  
 
8   Place  the  new  bottle  in  the  tray.  

Emptying  the  Waste  Container  


The  Level  Detect  function  automatically  monitors  the  waste  level  and  indicates  when  the  
waste  container  should  be  emptied.  It  is  still  necessary  to  check  waste  levels  visually  to  
ensure  the  waste  container  does  not  overfill.  
 
CAUTION:  
Dispose  of  collected  waste  according  to  local  statutes  and  safety  requirements.  
 
1   Unscrew  the  cap  from  the  full  waste  container.  
 
2   Discard  the  waste  according  to  local  regulations.  
 
3   Reinstall  the  cap  on  the  empty  waste  container.  

58  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.1  Preventive  Maintenance  
The  system  provides  a  Preventive  Maintenance  Log  for  tracking  the  most  recent  maintenance  
activities.  See  Enable  Preventive  Maintenance  Tracking  (Section  3.1)  and  Using  the  Preventive  
Maintenance  Log  in  this  section.  
 
Daily  Maintenance/Quality  Control  (QC)    
1   Check  reagent  levels  and  expiration  dates.  

2   Empty  the  waste  container  if  necessary.  

3   At  the  beginning  of  the  day:  


• Perform  a  Hub  Pattern  test.  
• Run  a  QC  slide  if  required  by  your  laboratory.  

4     If  necessary,  use  the  nozzle  brush  from  the  Maintenance  Kit  to  clean  the  
nozzle  orifices.  Press  the  individual  bristles  into  the  nozzle  openings.    
 
5   At  the  end  of  the  day:  
• Place  an  empty  carousel  in  the  bowl  and  close  the  lid.  Press  
Standby/Ready  on  the  front  panel  and  wait  until  the  end  of  the  
automatic  cleaning  process.  
 
• Spray  and  wipe  the  bowl,  interior  lid,  and  nozzles  with  70  to  100%  
alcohol.  
 
• Wipe  down  the  exterior  of  the  instrument  with  70  to  100%  alcohol.  
 
6   Ensure  the  maintenance  procedures  listed  on  the  Maintenance  Log  have  
been  performed,  and  entered  into  the  chart  or  log.  
 
Weekly  Maintenance  
1   Wipe  the  carousel  tray  and  lid  using  70  to  100%  alcohol.  
 
2   Flush  the  waste  tube  with  200  to  300  mL  of  alcohol  to  prevent  clogging.  
 
3   Perform  a  Volume  Test  (Section  6.4).  
 
4   Perform  a  Hub  Pattern  Test  (Section  4.1).  
 
5   Manually  clean  the  nozzles  if  necessary.  
 
6   Ensure  the  maintenance  procedures  listed  on  the  Maintenance  Log  have  
been  performed,  and  entered  into  the  chart  or  log.  

59  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.1  Preventive  Maintenance  
Monthly  Maintenance  
1   Disassemble  and  manually  clean  all  nozzles.  Refer  to  Nozzle  Disassembly  
and  Cleaning  (Section  6.1).  
 
2   Disinfect  any  reagent  bottle  that  is  being  reused  (Section  5.5).  
 
3   Ensure  the  maintenance  procedures  in  the  Preventive  Maintenance  (PM)  
Log  have  been  performed,  and  entered  into  the  PM  chart  or  log.  
 
Using  the  Preventive  Maintenance  Log  
With  Preventive  Maintenance  Tracking  enabled,  the  PM  Log  provides  a  convenient  and  
structured  means  of  recording  important  maintenance  and  QC  functions.  The  system  
allows  you  to  set  up  timely  prompts  that  require  response  by  the  user.  See  Enable  
Preventive  Maintenance  Tracking  (Section  3.1).  
 
1 From  the  Maintenance  menu,  press  QC/PM  to  open  the  PM  Log.  
 
 
 
 
 
 
 
 
2   Press  Record  Maintenance.  
PM  Task  entry  options:  
QC  Slide  Staining  (Drop  Down  Menu)  
  Not  Completed  
  Acceptable  
  Unacceptable  
  Inconclusive  
Disinfect  Reusable  Bottles    
Completed  (Select/Deselect)  
Drain  Check    
  Completed  (Select/Deselect)    
Manual  Nozzle  Cleaning  
  Completed  (Select/Deselect)  
 
 
 
 
 
 
3   Press  Save  to  record  entries.  

60  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.2  Storing  the  Instrument  
If  the  instrument  is  inactive  for  more  than  one  week,  you  may  want  to  perform  the  long-­‐term  
storage  procedure.  This  will  prevent  nozzles  from  clogging  when  the  machine  is  reactivated.  
 
Preparing  for  Long-­‐Term  Storage  
1   With  the  carousel  removed,  remove  and  clean  the  nozzles.  Store  parts  in  
tubes  that  are  labeled  to  indicate  their  correct  position.  
 
2   Unscrew  the  cap  and  remove  the  dip  tube  from  the  bottles.  
 
3   Place  the  end  of  the  dip  tube  in  a  bottle  of  methanol.  
 
4   Flush  at  least  250  mL  of  methanol  through  each  reagent  line  by  priming  all  
lines  simultaneously.  Leave  the  alcohol  in  the  line.  
 
CAUTION:    
Leave  methanol  in  the  reagent  lines  during  storage.  Allowing  reagent  lines  to  
run  dry  can  damage  the  instrument.  
 
CAUTION:    
Do  not  subject  the  instrument  to  freezing  temperatures.  Freezing  of  aqueous  
fluids  in  the  lines  may  cause  damage  to  the  instrument.  
 
5   Flush  the  bowl  with  water.  
 
6   Return  nozzles  to  their  original  positions.  

Preparing  for  Operation  after  Storage  


Follow  the  Setup  and  Preparation  for  Operation  instructions  (Section  2).  

61  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.3  Replacing  Fuses  
WARNING!  
To  prevent  the  risk  of  fire,  the  main  fuses  should  only  be  replaced  with  fuses  of  the  same  type  
and  rating.  Recurring  fuse  failure  indicates  serious  internal  problems,  if  this  occurs,  contact  
Elitech  Group.  
 
1   Power  OFF  the  instrument.  
 
2   Disconnect  the  power  cord  from  the  power  outlet  and  the  rear  panel  of  
the  instrument.  
 
3   Open  the  fuse  cover  by  inserting  a  screwdriver  in  the  slot  on  the  right  side  
of  the  cover  and  gently  prying  the  cover  open.  
 
4   Remove  the  fuse  holders  to  inspect  the  fuses.  
 
    Replace  the  fuses  if  necessary.  
 
6   Push  the  fuse  holders  in.  
 
7   Close  the  fuse  cover.  
 
8   Reconnect  the  main  power  cord  to  the  rear  panel  of  the  instrument  and  to  
the  power  outlet.  
 
9   Power  ON  the  instrument.  
 
 

62  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.4  Cleaning  the  Stainer  and  Carousels  
WARNING!  
All  cleaning  procedures  should  be  performed  in  a  well-­‐ventilated  room  by  authorized  and  
trained  personnel  wearing  appropriate  protection  equipment.  
 
1   Clean  the  outside  of  the  instrument  with  70  to  100%  ethanol  or  methanol.  
 
2   Clean  the  carousel  and  lid  with  70  to  100%  ethanol  or  methanol.  
 
NOTE:  Freshly  prepared  (<  24  hours  old)  10%  bleach  solution  can  be  used  as  
well.  The  10%  bleach  solution  helps  clean  the  stained  areas.  
 
Cleaning  Liquid  Spills  
Remove  any  liquid  spilled  on  the  instrument  immediately  to  avoid  damage  to  the  
equipment.  
 
WARNING!  
If  potentially  infectious  liquid  is  spilled  on  the  instrument,  the  instrument  must  be  
disinfected  in  accordance  with  all  applicable  local  regulations.  Refer  to  
Decontaminating  the  Stainer  and  Carousels  (Section  5.5)  for  instructions.  
 
Cleaning  Broken  Slides  
You  must  take  stringent  precautions  if  a  slide  breaks  inside  the  instrument  during  a  
staining  cycle,  especially  if  the  instrument  has  been  processing  dangerous  pathogens.  
Always  use  protective  gloves,  safety  glasses,  and  forceps  when  removing  broken  glass  
from  inside  the  instrument.    
 
• Glass  shards  embedded  in  the  walls  of  the  bowl  can  cause  serious  cuts  and  pose  a  
risk  of  infection.    
 
• Always  remove  embedded  shards  with  a  scraper  before  attempting  to  remove  loose  
glass.  
 
• Use  a  vacuum  or  adhesive  tape  to  pick  up  loose  glass  inside  the  stainer  bowl.  

63  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.5  Decontaminating  the  Stainer  and  Carousels  
All  parts  of  the  instrument  that  come  into  contact  with  biological  specimens,  patient  specimens,  
positive  control  specimens,  or  hazardous  material  must  be  treated  as  potentially  infectious.  
 
Before  the  instrument  is  returned  for  service,  all  outer  surfaces  must  be  decontaminated.  The  
operating  authority  must  complete  a  disinfection  declaration,  otherwise  the  instrument  may  be  
rejected  by  the  distributor  or  service  center  or  quarantined  by  customs  authorities.  
 
WARNING!  
Reagents  used  with  the  instrument  contain  moderately  hazardous  chemicals  that  require  care  
in  handling.  Always  use  appropriate  safety  measures  including  gloves  and  eye  protection,  
when  handling  reagents.  
 
WARNING!  
Authorized  and  trained  personnel  wearing  appropriate  protection  equipment  should  perform  
the  decontamination  procedure  in  a  well-­‐ventilated  room.  It  is  very  important  to  thoroughly  
decontaminate  the  instrument  before  removing  it  from  the  laboratory  or  before  performing  
any  technical  service.  This  procedure  may  not  be  effective  against  prions.  
 
WARNING!  
Prior  to  decontaminating,  disconnect  the  instrument  from  the  main  power  supply  to  avoid  any  
risk  of  fire  or  explosion.  
 
WARNING!  
The  decontamination  procedure  and  the  disinfectants  must  comply  with  the  local  applicable  
regulations.  
 
Solutions  for  Decontaminating  the  Instrument  
The  outer  surfaces  of  the  instrument  should  be  decontaminated  using  a  
decontaminating  solution  such  as:  
• 70%  ethanol  or  methanol  
• Mild  detergent  
• 10%  bleach  solution  (<  24  hours  old)  
• Decontamination  Solution  (REF:  SS-­‐133)  

64  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.5  Decontaminating  the  Stainer  and  Carousels    
Figure  12:  Lid  Latch  and  Locking  Pin  Hole  Locations  
 
 
 
 
 
 
 
 
 
 
 
  1        2  
1  –  Lid  Latch  Hole  
2  –  Locking  Pin  Hole  

Decontaminating  the  Instrument  


1   Prepare  a  suitable  container  for  all  disposables.  
 
2   Cover  the  lid  latch  and  locking-­‐pin  holes  with  waterproof  tape  to  protect  
the  interior  (Figure  12).  
 
 
3   Place  the  instrument  in  a  biological  safety  hood  or  well-­‐ventilated  area.  
 
 
4   Spray  the  inner  bowl  and  inner  lid  with  a  decontaminating  solution  such  as  
REF:  SS-­‐133.  
 
 
5   Repeat  the  spray  treatment  every  2  or  3  minutes  for  a  total  of  20  minutes.  
Do  not  allow  cleaning  solutions  to  dry  on  the  instrument  surfaces.  
   
 
6   Rinse  the  inner  bowl  and  lid  thoroughly  with  water.  
 
 
7   Spray  and  wipe  the  exterior  surfaces  with  decontamination  solution  such  
as  REF:  SS-­‐133.  
 
CAUTION:    
Do  not  flood  the  display  panel  with  excessive  moisture.  Any  moisture  that  
seeps  through  could  damage  the  internal  electronics.  
 
 

65  
SECTION  5    
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.5  Decontaminating  the  Stainer  and  Carousels  
Decontaminating  the  Instrument  (continued)  
8   Repeat  the  spray  treatment  of  exterior  surfaces  every  2  or  3  minutes  for  a  
total  of  20  minutes.  Do  not  allow  cleaning/decontamination  solutions  to  
dry  on  the  instrument  surfaces.  
 
 
9   Wipe  surfaces  thoroughly  with  a  cloth  soaked  in  water  until  you  have  
removed  all  decontamination  solution.  
 
 
10   Immerse  or  generously  spray  the  carousel  and  lid  with  decontaminating  
solution.  Allow  the  solution  to  react  for  20  minutes.  
 
 
11   Thoroughly  rinse  the  carousel  and  lid  with  deionized  or  distilled  water.  
 

Decontaminating  Reagent  Bottles  


Any  reused  reagent  bottles  should  be  disinfected  at  each  reagent  lot  change,  or  
monthly:  
1   Fill  the  reagent  bottle  with  a  fresh  (<  24  hours  old)  10%  bleach  solution.  
 
 
2   Allow  the  bleach  solution  to  react  in  the  bottle  for  10  minutes.  
 
 
3   Rinse  the  bottle  thoroughly  with  tap  water.  
 
 
4   Rinse  the  bottle  thoroughly  with  deionized  or  distilled  water.  

66  
SECTION  5  
PREVENTIVE  MAINTENANCE  AND  SAFETY  
 
5.6  Shipping  or  Disposing  of  the  Stainer  or  Carousels  
Shipping  the  Instrument  or  Carousels  
WARNING!  
You  must  disinfect  the  instrument  or  carousels  before  returning  it  to  Elitech  Group.  
The  operating  authority  must  complete  a  Hazard-­‐Free  Certification  form,  otherwise  
the  distributor  or  service  center  may  not  accept  the  instrument;  or  customs  
authorities  may  hold  it.  
 
CAUTION:    
Shipping  the  instrument  or  carousels  without  decontaminating  according  to  these  
instructions  is  dangerous  to  service  personnel.  You  will  be  charged  additional  fees  for  
decontamination  performed  by  Elitech  Group.  
 
CAUTION:    
Ship  the  instrument  or  carousels  in  containers  comparable  to  the  original  packaging.  
Hazard-­‐Free  Certification  
The  operating  authority  must  print  and  complete  the  Hazard-­‐Free  Certification  
(obtained  from  Elitech  Group  Customer  Service).  
 
Attach  the  certification  form  to  the  top  of  the  instrument  package  before  
sending  the  package  to  Elitech  Group.  
 
Disposing  of  the  Instrument  or  Carousels  
The  instrument  and  carousels  should  be  completely  decontaminated  and  disposed  of  as  
follows:  
 
Under  WEEE  Directive  2002/96/EC,  this  equipment  cannot  be  disposed  of  in  a  normal  
landfill.  Instead,  the  equipment  must  be  disposed  of  either  by:  
 
1   Routing  to  an  authorized  local  facility  approved  for  handling  hazardous  
materials.    
 
OR  
 
2   Returning  the  equipment  to  Elitech  Group.  
 

67  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.1  Nozzle  Disassembly  and  Cleaning  
Nozzle  maintenance  requires  the  Nozzle  Maintenance  Kit  and  prepared  Nozzle  Cleaning  Solution  
(diluted  REF:  SS-­‐029C).  
 
WARNING!  
Always  wear  protective  clothing  and  eye  protection  when  using  Nozzle  Cleaning  Solution  
(diluted  REF:  SS-­‐029C).  Dispose  of  used  solution  properly.  
 
NOTE:  If  the  compression  screw  cannot  be  easily  loosened,  use  light  penetrating  oil  and  a  5/8-­‐in.  
wrench  to  loosen  the  nozzle.  
 
Nozzle  Disassembly  
1   Remove  the  nozzle  using  the  nozzle  tool  from  the  Nozzle  Maintenance  Kit.  
 
 
 
2   Disassemble  the  nozzle.  See  Figure  5:  Nozzle  Diagram  in  Section  1.  
 
 
 
3   If  you  will  be  storing  the  nozzle  parts  in  Nozzle  Cleaning  solution  longer  
than  one  hour,  remove  the  o-­‐ring.  
 
 
4   Place  the  nozzle  parts  (minus  the  o-­‐ring  if  required)  in  a  50  mL  conical  tube  
that  has  been  clearly  marked  with  the  correct  nozzle  position.  
 
 
 
 
 
5   Repeat  Steps  1  through  4  for  each  nozzle.  
 
Nozzle  Cleaning  
1   Fill  each  50  mL  tube  with  25  mL  of  prepared  Nozzle  Cleaning  Solution  (REF:  
SS-­‐029C)  and  cap  the  tube.  
 
 
2   Gently  invert  the  tube  at  least  ten  times  to  ensure  all  parts  come  in  contact  
with  the  cleaning  solution.  
 
 
3   Place  the  tube  in  the  correctly  marked  position  in  the  provided  tube  stand.  
Soak  the  parts  as  long  as  possible.  
 
NOTE:  Soak  nozzle  parts  for  at  least  15  minutes.  Parts  can  be  soaked  in  Nozzle  
Cleaning  Solution  overnight.    
 
4   Repeat  steps  1  through  3  for  each  nozzle.  

68  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.2  Nozzle  Reassembly  
1   Hold  your  thumb  or  a  nozzle  strainer  over  the  end  of  the  tube  to  keep  the  
nozzle  parts  in  the  tube.  Discard  the  cleaning  solution  according  to  
applicable  statutes.  
 
Swirl  Cone    
Grooves  (4   2   Inspect  the  nozzle  parts.  Remove  any  material  in  the  swirl  cone  grooves  by  
total,  2  shown)   sliding  a  piece  of  paper  edgewise  along  each  of  the  4  grooves.  
 
 
3   Thread  the  nozzle  orifice  cleaning  wire  (REF:  AC-­‐059)  through  the  back  of  
the  disassembled  nozzle  housing.  
 
   
4   Place  the  nozzle  parts  back  into  the  tube  and  rinse  them  with  water.  
 
 
5   Rinse  the  parts  again  with  alcohol.  
 
 
6   Apply  a  small  amount  of  silicone  lubricant  (REF:  AC-­‐103)  to  the  
compression  screw  threads.  
 
 
Apply  Silicone   7   Reassemble  the  nozzle  by  placing  the  compression  screw  on  the  nozzle  
Lubricant  Here   disassembly  tool,  then  inserting  the  swirl  cone  into  the  compression  screw.  
 
Nozzle  
NOTE:  Hold  all  the  parts  in  a  vertical  position  during  reassembly.  
Disassembly  
 
Tool  
8   Reinstall  the  nozzle  housing  over  the  swirl  cone  and  compression  screw.  
 
 
9     Replace  the  o-­‐ring  if  it  was  removed  earlier.  
 
 
10   Return  the  assembled  nozzle  to  its  original  position  in  the  instrument.  
D                        C                        B                    A    
 
11   Repeat  Steps  1  through  10  for  each  nozzle.  
 
 
12   Perform  a  Hub  Pattern  Test  (Section  4.1).  
 
 
13   Perform  a  Volume  Test  (Section  6.4).  
 
NOTE:  You  must  perform  the  Hub  Pattern  Test  and  Volume  Test  before  
operating  the  instrument.  If  the  results  are  incorrect,  manually  prime  the  
instrument.  

69  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.3  Manual  Priming  
1   Remove  the  carousel  from  the  bowl.  
 
2   Remove  the  nozzle  connected  to  the  line  to  be  manually  primed.  
 
3   Insert  the  priming  tool  nozzle  adapter  (included  in  the  Nozzle  Maintenance  
Kit)  into  the  nozzle  holder  and  turn  it  clockwise  to  install  the  adapter  into  
the  holder.  
 
4   Withdraw  the  priming  tool  plunger  halfway  to  create  a  vacuum.  Hold  the  
plunger  in  that  position.  
 
5   Press  Maintenance  from  the  Main  menu.  
 
6   Press  Volume  Test.  
 
7   Press  the  desired  prime  button  to  start  the  reagent  pump.  
 
8   Run  the  reagent  into  the  tube  until  the  fluid  is  free  of  bubbles,  then  press  
Stop.  
 
WARNING!  
Do  not  pull  the  plunger  completely  out  of  the  priming  tool.  Pulling  the  
plunger  out  of  the  tool  may  result  in  splashing  or  spraying  of  reagents.  Do  not  
push  the  plunger  in  while  it  is  connected  to  the  nozzle  holder.  
 
9   Turn  the  nozzle  adapter  counterclockwise  to  remove  it  from  the  nozzle  
holder.  
 
10   Discard  the  collected  fluid  into  the  stainer  bowl.  
 
11   Reinstall  the  nozzle.  
 
12   Perform  a  Hub  Pattern  Test.  
 
13   Perform  a  Volume  Test.  
 

70  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.4  Performing  the  Volume  Test  
The  Volume  Test  requires  the  Nozzle  Maintenance  Kit.  
 
NOTE:  The  Volume  Test  must  be  performed  weekly.  
 
1   From  the  Maintenance  menu,  select  Volume  Test.  
 
 
2   Hold  a  Volume  Test  tube  (small  tube)  to  cover  the  selected  nozzle.  
 
 
3   Press  the  corresponding  reagent  prime  button  to  collect  the  reagent.    
 
NOTE:  With  QC/Maintenance  tracking  enabled,  enter  the  measured  volume  on  
the  keypad  and  press  ENTER.  With  QC/Maintenance  Tracking  disabled,  the  
menu  returns  to  the  Maintenance  menu.  
 
4   Remove  and  cap  the  tube.  
 
5   Record  the  nozzle  position  on  the  tube  and  place  the  tube  in  the  
appropriate  position  in  the  tube  stand.  
 
6   Repeat  Steps  2  through  5  for  each  nozzle.  
 
7   Compare  collected  nozzle  volumes  with  the  following  table.  
 
Table  12:  Volume  Test  Tolerances  
 
Nozzle/Reagent  Line   Minimum   Maximum  
A,  B,  C   9.0  mL   11.0  mL  
D   8.5  mL     10.5  mL  

NOTE:  The  stainer  normally  functions  correctly  if  nozzle  volumes  are  slightly  
higher  or  lower  than  the  specified  range.  It  is  important  that  the  B  and  C  nozzle  
volumes  are  similar,  (typically  within  1  mL).  Spray  volumes  <  7.5  mL  or  >  13.0  
mL  indicate  serious  problems  with  the  nozzles  or  reagent  delivery  lines.    
• If  the  volume  is  within  the  tolerance  range,  go  to  Step  9.  
• If  the  volume  is  outside  the  tolerance  range:  
a.   Clear  the  nozzle  orifice  with  the  nozzle  brush  found  in  the  
maintenance  kit.  
 
b.   If  necessary,  remove  the  nozzle  and  perform  the  Nozzle  
Cleaning  procedure  (Section  6.1).  
 
c.   If  the  problem  persists,  replace  the  nozzle.  
 
NOTE:  If  the  problem  persists  after  you  have  replaced  the  nozzle,  contact  Elitech  
Group.  

71  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.4  Performing  the  Volume  Test  
8   Prepare  the  Maintenance  Kit  for  future  use:  
• Empty  the  contents  of  the  tubes  into  the  stainer  bowl.  
• Rinse  the  tubes  with  water.  
• Put  the  tubes  back  into  their  original  place  in  the  Maintenance  Kit  or  
tube  stand.  

 
9   Press  Back  twice  to  return  to  the  Main  menu.  

72  
SECTION  6  
NOZZLE  MAINTENANCE  AND  PERFORMANCE  
 
6.5  Performing  the  Slide  Pattern  Test  
This  test  can  differentiate  poor  staining  results  from  sample  preparation  problems,  or  nozzle  
obstructions.  Perform  the  Slide  Pattern  test  when  a  Hub  Pattern  test  produces  a  normal  result,  
but  staining  is  still  inadequate.  
 
1   Place  a  1  x  3  inch  (2.5  x  7.6  cm)  piece  of  paper  (REF:  RP-­‐500)  in  positions  1  
and  2  of  the  carousel,  with  a  blocking  slide  in  front  of  positions  1  and  2.  
 
2   Load  the  carousel  into  the  stainer  and  close  the  lid.  
 
3   From  the  Main  menu,  press  Maintenance.  
 
4   Press  Pattern  Test.  
 
5   Press  the  corresponding  prime  button  for  the  reagent  line  to  be  tested.  
 
6   Remove  the  paper  slides.  
 
7   Repeat  Steps  1  through  6  for  each  reagent  line.  
 
8   Examine  the  paper  slides  for  each  reagent.  The  pattern  on  the  slide  should  
be  uniform,  without  any  continuous  lines  or  streaks.  
 

Figure  13:  Correct  Slide  Pattern  Test  Result  


 
 
 
 
 
 
 
 
Figure  14:  Incorrect  Slide  Pattern  Test  Result  
   
 
 
 
 
 
 
9   If  the  result  is  incorrect,  clear  the  nozzle  blockage  using  the  nozzle  brush,  
or  disassemble  and  clean  the  nozzle  (Section  6.1).  

73  
SECTION  7    
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
The  following  table  is  to  help  identify  and  solve  routine  problems  with  the  stainer.  More  difficult  
problems  may  require  technical  service.  Contact  your  Elitech  Group  representative  for  
assistance.  
 
WARNING!  
Due  to  the  electrical  shock  hazard,  do  not  open  this  instrument  or  attempt  internal  repairs.  
Refer  servicing  to  qualified  service  personnel.  Contact  your  dealer  or  Elitech  Group  Service.  
 
Table  13:  General  Troubleshooting  and  Diagnosis  
 
Problem   Solution  
There  is  no  power  to  the  stainer  when  the  power  switch   Check  the  facility  outlet  and  the  power  cord  connection.  
is  turned  ON.   Check  the  fuses.  Refer  to  the  Replacing  Fuses  procedure.  
  CAUTION:    
  Fuse  failure  may  indicate  a  serious  internal  problem.    
 
Strange  information  shows  on  the  display,  and/or  erratic   Switch  the  power  OFF,  wait  10  to  20  seconds,  then  
stainer  operation.   switch  power  ON  again.  If  problem  recurs,  install  a  
  computer-­‐type  surge  suppressor  to  protect  the  
instrument  from  power  line  transients.  If  possible,  
connect  the  stainer  to  a  power  circuit  that  is  not  shared  
by  centrifuges,  refrigerators,  air  conditioners,  or  other  
motorized  equipment.  
 
If  the  above  steps  do  not  solve  the  problem,  consult  the  
Aerospray  Service  Manual,  or  contact  your  dealer  or  
Elitech  Group  for  assistance.  
 
A  reagent  line  will  not  prime  when  power  is  ON  and  you   Follow  the  procedures  in  Section  6.3  for  priming  reagent  
press  the  prime  button.   pumps.  
   
A  reagent  line  will  not  prime,  even  with  the  priming  tool   Press  the  priming  button  and  listen  carefully  for  the  
(Section  6.3).   sound  of  the  pump.  If  you  can  hear  the  pump,  try  the  
  priming  tool  again.  If  the  problem  is  not  solved  or  if  you  
cannot  hear  the  pump  there  may  be  an  internal  problem.  
Contact  your  dealer  or  Elitech  Group  for  assistance.  
 
Stainer  bowl  fills  with  reagent  after  use.   A  small  puddle  of  stain  around  the  drain  inlet  or  the  
  bottom  of  the  bowl  is  normal.  If  the  bowl  is  filling  with  a  
large  quantity  of  stain,  check  the  external  drain  tube  for  
blockage.  Make  sure  the  drain  tube  is  properly  
connected  and  running  continuously  down  toward  the  
lab  drain  or  vented  waste  container,  with  no  loops,  rises,  
or  obstructions.  Make  sure  the  end  of  the  tube  is  not  
submerged.  This  can  prevent  proper  drainage.  The  
internal  drain  may  need  to  be  cleaned  or  replaced.  See  
the  Aerospray  Service  Manual,  or  contact  your  dealer  or  
Elitech  Group  Service.  
 
 

74  
SECTION  7  
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
Table  13:  General  Troubleshooting  and  Diagnosis  (continued)  
Problem   Solution  
Stain  is  leaking  onto  the  counter.     Check  all  external  reagent  lines  for  visible  signs  of  cracks  
  or  loose  fittings.  
   
Make  sure  the  drain  outlet  is  not  blocked.  
 
Make  sure  the  drain  tube  is  securely  attached  to  the  
drain  port  and  that  the  tubing  is  not  cracked  or  
deformed.  
 
Reagent  leaks  may  indicate  an  internal  problem  (see  
Section  7.3).  See  the  Aerospray  Service  Manual,  or  
contact  your  dealer  or  Elitech  Group  for  further  
assistance.  
 
Error  messages  on  the  screen.   If  the  display  shows  Lid  Not  Shut:  Verify  that  the  lid  is  
  fully  closed  and  latched.  If  the  Lid  Not  Shut  indication  
  remains,  contact  Elitech  Group  for  assistance.  
 
If  the  display  shows  Wrong  Rotor  after  pressing  Start:  
Make  sure  the  slide  carousel  is  properly  loaded  on  the  
drive  hub.  In  staining  mode,  the  instrument  detects  
whether  the  staining  carousel  is  present  before  
proceeding.  In  cytocentrifuge  mode,  the  instrument  will  
stop  if  it  senses  the  staining  carousel.  After  verifying  the  
carousel  is  correctly  loaded,  press  Start.  If  the  display  still  
shows  Wrong  Rotor,  there  may  be  an  internal  problem.  
Check  for  missing  carousel  magnets.  
The  microprocessor  monitors  carousel  rotation  during  a  
staining  cycle.  The  display  shows  an  error  message  if  the  
rotation  is  not  within  the  specified  range.  
 
If  the  display  shows  Motor  Drive  Error:  Check  the  stainer  
bowl  for  interference:  Turn  the  hub  or  carousel  by  hand;  
it  should  turn  freely.    
Drive  motor  or  electronic  component  malfunctions  
require  servicing  of  internal  components.  Contact  your  
dealer  or  Elitech  Group  for  assistance.    
 
If  the  display  shows  Rotor  Imbalance,  make  certain  the  
Cytopro  rotor  is  balanced,  or  the  staining  carousel  is  
seated  correctly  on  the  hub.  
See  Electronic  Failure  later  in  this  table.  
 
 

75  
SECTION  7  
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
Table  13:  General  Troubleshooting  and  Diagnosis  (continued)  
Problem     Solution  
The  stainer  fails  to  spray  reagent  during  a  staining  cycle   To  allow  programmed  staining  of  partial  loads,  the  
and/or  continues  to  run  after  the  cycle  should  be   stainer  monitors  the  position  of  the  carousel  as  it  rotates  
complete.   in  the  bowl.  In  normal  operation,  stain  is  sprayed  only  in  
the  correct  position.  This  causes  the  actual  cycle  time  to  
vary,  depending  on  the  position  of  the  carousel  at  the  
beginning  of  the  cycle.  However,  if  the  cycle  continues  
for  an  abnormally  long  period,  or  if  the  bar  graph  and  
percentage  complete  icon  do  not  change  after  1  minute,  
it  may  indicate  an  electronic  problem  or  an  internal  
problem.  To  determine  this,  press  Stop.  
 
If  the  cycle  stops:  this  indicates  a  problem  with  the  
carousel  position  sensor.  Consult  the  Aerospray  Service  
Manual,  or  contact  your  dealer  or  Elitech  Group  for  
assistance.  
 
If  the  cycle  continues:  this  indicates  an  electronic  
problem  (see  below).    
Abnormal  staining  on  entire  surface  of  all  slides.   Check  the  reagent  level  on  the  display  and/or  in  the  
  reagent  bottles.  
   
Make  sure  the  external  reagent  dip  tubes  are  securely  
attached  to  each  bottle  (Section  2.1).  
 
Open  the  lid  and  verify  that  each  reagent  pump  is  
primed,  by  pressing  the  corresponding  prime  button.  The  
nozzle  should  immediately  spray  a  fine  mist  of  reagent.  
There  should  be  no  sputtering  or  hissing  sounds  
indicating  air  in  the  reagent  lines.  
 
Watch  the  external  tubes  for  air  bubbles.  Air  bubbles  
indicate  inadequate  priming  or  possibly  an  air  or  reagent  
leak  in  the  system.  Air  in  any  reagent  line  will  cause  poor  
staining.  
Check  nozzle  performance  using  the  Slide  Pattern  
(Section  6.5)  and  Volume  Tests  (Section  6.4).  If  necessary,  
clean  nozzle(s)  using  the  procedures  in  Section  6.1.  
 
Verify  that  each  reagent  dip  tube  vent  hole  is  clear.  
 
 

76  
SECTION  7  
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
Table  13:  General  Troubleshooting  and  Diagnosis  (continued)    
Problem   Solution  
Abnormal  staining  on  entire  surface  of  all  slides   When  staining  a  full  carousel  (7  or  more  slides)  make  
(continued).   certain  you  have  not  programmed  the  stainer  for  fewer  
  slides.  
   
If  staining  a  partial  load,  load  the  slides  in  the  correct  
positions  as  indicated  by  the  markings  on  the  carousel  
(see  Section  4.1).  
 
Abnormal  staining  on  entire  surface  of  some  slides,  while   Make  certain  that  all  position  magnets  are  still  attached  
other  slides  from  the  same  carousel  appear  normal.     to  the  bottom  of  the  carousel.  Make  certain  you  have  not  
  programmed  the  stainer  for  fewer  slides  than  you  have  
  loaded.  
 
If  you  program  the  stainer  for  a  partial  load,  load  the  
slides  in  the  correct  positions  as  indicated  by  the  
markings  on  the  carousel  (see  Section  4.1).  
 
Verify  that  each  reagent  pump  is  primed  by  opening  the  
lid  and  pressing  the  corresponding  prime  button.  The  
nozzle  should  immediately  spray  a  fine  mist  of  reagent.  
There  should  be  no  sputtering  or  hissing  sounds  to  
indicate  the  presence  of  air  in  the  reagent  lines  (see  
Section  7.3).  
 
Streaks  or  bands  of  discoloration  on  one  or  more  slides.   Check  the  level  of  the  Rinse  (Reagent  A)  bottle.  Check  
  Rinse  (Reagent  A)  spray  volume  according  to  Section  6.4.  
 
Check  nozzle  spray  pattern  according  to  the  procedures  
in  Sections  4.1  and  6.5.  This  type  of  discoloration  is  
usually  caused  by  debris  or  reagent  precipitate  clogging  
the  spray  nozzle  orifice.  
 
Clean  any  nozzle  that  exhibits  a  poor  spray  pattern.  
 
Cells  are  washing  off  slides.   Check  methanol  or  Aerofix  (Reagent  D)  level  in  bottle.  
 
Check  Reagent  D  nozzle  spray  volume  and  slide  pattern  
(see  Section  6.4  and  Section  6.5).  
 

77  
SECTION  7  
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
Table  13:  General  Troubleshooting  and  Diagnosis  (continued)  
Problem   Solution  
Cells  are  washing  off  slides  (continued).   Check  the  methanol  or  Aerofix  (Reagent  D)  hub  pattern  
  and  spray  volume  (see  Sections  4.1  and  6.5).  
Make  certain  the  blood  smears  are  completely  dry  and  
not  too  thick  (see  Section  7.2).  
Increase  the  fixation  setting  to  High.  
 
Call  Elitech  Group  for  information  on  slide  quality.  
 
Refractile  artifacts  are  observed  in  erythrocytes.   Use  Aerofix  (SS-­‐048  or  SS-­‐148)  fixative.  Refer  to  Section  
7.2.  
 
High  levels  of  precipitates  observed.   Discard  leftover  thiazin  stain  (Reagent  B)  in  the  used  
  bottle  when  changing  to  a  fresh  bottle  of  stain.  While  
transferring  remaining  stain  saves  stain,  it  eventually  
concentrates  the  precipitates  to  the  point  they  appear  on  
the  slide.  
 
Staining  intensity  is  too  weak.  Granulation  is  weak.   Increase  the  intensity  setting.    

Refractile  artifacts  are  observed  in  erythrocytes  when   Mix  1  part  Aerofix  (SS-­‐048)  with  3  parts  Basofix  predip  
using  Basofix  predip  (SS-­‐049P).   (SS-­‐049P),  or  add  5  mL  Aerofix  concentrate  (SS-­‐148)  to  
  500  mL  Basofix  predip  (SS-­‐049P).  
 
Increased  false-­‐positives  for  Howell-­‐Jolly  bodies  or   Disinfect  reagent  lines  using  instructions  in  Section  7.2.  
amorphous  precipitates  are  observed.    
Discard  leftover  thiazin  stain  (Reagent  B)  in  the  used  
bottle  when  changing  to  a  fresh  bottle  of  stain.  While  
transferring  remaining  stain  saves  stain,  it  eventually  
concentrates  the  precipitates  to  the  point  where  they  
can  be  seen  on  the  slide.  
 
 
 

78  
SECTION  7  
SOLVING  PROBLEMS  
 
7.1  Troubleshooting  
Table  13:  General  Troubleshooting  and  Diagnosis  (continued)  
Problem   Solution  
Electronic  Failure   • An  electronic  failure  would  appear  as  an  obvious  
malfunction  such  as  a  scrambled  or  totally  inoperative  
display  panel.    
•  
• Transient  voltages  coming  through  the  power  lines  may  
cause  the  stainer  to  “lose  its  place.”    

1    If  this  occurs,  switch  the  main  power  OFF  for  10-­‐20  
seconds  and  then  back  ON  to  reset  the  instrument.    
 
2    If  the  problem  recurs,  install  a  computer-­‐type  surge  
protector  to  isolate  the  instrument.    
 
3    If  possible,  connect  the  stainer  to  a  power  circuit  not  
shared  by  centrifuges,  refrigerators,  air  conditioners,  
or  other  motorized  equipment.  
For  more  obscure  electronic  problems,  monitor  the  
stainer  through  a  complete  staining  cycle  to  determine  if  
the  operating  sequence  is  correct.  Do  this  by  running  the  
stainer  while  watching  the  display  and  listening  to  the  
pumps.  
 
Ensure  that  each  event  occurs  according  to  the  operating  
sequence,  shown  in  Table  3  in  Section  1.  
 
If  the  problem  recurs,  contact  your  dealer  or  Elitech  
Group  for  assistance.  
 
 

79  
SECTION  7  
SOLVING  PROBLEMS  
 
7.2  Abnormal  Staining  Results  
Smear  Separation  or  Tearing  
Losing  blood  cells  from  the  slide  surface  during  a  staining  cycle  can  happen  for  several  
reasons:  
 
Improper  Alcohol  Fixation  
If  the  alcohol-­‐fix  phase  of  the  cycle  fails,  there  will  be  cell  loss  when  stain  is  
applied.  Verify  that  the  alcohol  nozzle  is  operating  properly  with  normal  spray  
pattern  and  spray  volume  (Sections  6.4,  6.5).  To  identify  the  problem  as  a  
fixation  failure,  fix  some  specimen  slides  in  methanol  prior  to  loading  into  the  
Wet  Smears  
carousel.  Select  a  fixation  setting  that  increases  fixation  (7,  8,  or  9).  
 
Wet  Smears  
Loading  wet  blood  smears  into  the  carousel  may  contribute  to  smear  tearing.  
Allow  smears  to  dry  for  several  minutes  before  beginning  a  stain  cycle.  
 
Thick  Smears  
Smears  that  are  too  thick  can  cause  smear  separation.  If  the  smear  has  a  thick  
“wedge”  cross  section,  cells  can  be  torn  away  from  the  slide  surface  in  the  
thicker  regions,  while  the  thinner  “monolayered”  regions  remain  usable.  Severe  
tearing  that  starts  in  the  thick  region  may  also  extend  into  the  monolayer.  
 
Dirty  Slides  
Dirty  slides  are  a  major  source  of  cell  loss  during  staining.  We  strongly  
recommend  that  you  use  new,  premium  quality  slides.  
 
Even  with  premium  quality  slides,  random  loss  may  occur  due  to  inadequate  
slide  surface  quality.  When  the  stainer  is  functioning  properly,  slide  quality  
accounts  for  most,  if  not  all,  cell  loss  problems.  
 
Red  Cell  Artifacts  
Refractile  bodies  or  inclusions  (seen  as  dark  ringed  objects  on  red  cells)  are  
thought  to  be  caused  by  the  presence  of  water  during  fixation.  Often  called  
“water  spotting,”  this  effect  is  common  in  Romanowsky  staining.  Methanol  
used  for  fixation  must  be  anhydrous  (0.5%  or  less  water).  
Methanol  that  is  exposed  to  the  atmosphere  will  absorb  a  considerable  amount  
of  water,  particularly  when  the  relative  humidity  is  high.  The  first  line  of  
defense  to  avoid  water  spotting  is  to  make  certain  the  methanol  used  for  fixing  
smears  is  free  of  water.  If  in  doubt,  replace  it  with  reagent  known  to  be  
anhydrous.    
 
 
   

80  
SECTION  7  
SOLVING  PROBLEMS  
 
7.2  Abnormal  Staining  Results  
Red  Cell  Artifacts  (continued)  
When  the  relative  humidity  is  above  60  percent,  water  spotting  may  be  seen  
even  when  using  anhydrous  methanol  for  fixation.  This  occurs  because  of  
moisture  accumulation  and  condensation  inside  the  carousel  chamber.  Relative  
humidity  within  the  chamber  will  tend  to  build  up  with  repeated  use,  leading  to  
further  problems.  If  water  spotting  occurs:  
 
Use  Aerofix®  
Water  spotting  can  be  eliminated  by  using  Aerofix.  Aerofix  is  available  as  an  
additive  (REF:  SS-­‐148,  135  mL  bottle,  enough  for  nine  500  mL  bottles  of  
methanol)  or  premixed  with  methanol  (REF:  SS-­‐048,  500  mL  bottle).  
 
• Pre-­‐fix  Slides  
Water  spotting  can  usually  be  eliminated  by  pre-­‐fixing  slides  in  anhydrous  
methanol  before  loading  into  the  carousel.  Allow  the  slides  to  dry  for  2  to  3  
minutes  before  staining.  
 
• Wipe  Carousel  Chamber  Dry  
Between  cycles,  leave  the  stainer  lid  open  to  dry  out  the  bowl.  If  necessary,  
wipe  the  interior  of  the  carousel  chamber  dry  to  reduce  the  relative  
humidity.  Or,  irrigate  the  chamber  with  methanol  from  a  wash  bottle.  
 
False-­‐Positive  Howell-­‐Jolly  Bodies  or  Amorphous  Precipitates  
Artifactual  Howell-­‐Jolly  bodies  can  be  caused  by  precipitate  build  up  from  
transferring  the  residual  thiazin  stain  (Reagent  B)  to  the  fresh  bottle  when  
reloading  reagents.  Such  artifacts  may  also  indicate  microbial  contamination  of  
the  reagent  lines  and  pumps.    
 
If  such  artifacts  are  encountered,  and  if  you  suspect  microbial  contamination,  
use  the  following  solution  to  disinfect  the  reagent  lines  and  pumps.  
 
Replace  precipitated  stain  with  fresh  stains  after  flushing  the  B  line  with  the  
disinfecting  solution.  
 
Disinfectant  Solution:  
10%  freshly  prepared  dilution  of  household  bleach  (5.25%  Sodium  
Hypochlorite).  
 
Dilution:    
         100  mL  household  bleach  
 +    900  mL  deionized  water  
 
 =  1000  mL  

81  
SECTION  7  
SOLVING  PROBLEMS  
 
7.2  Abnormal  Staining  Results  
 
Procedure:  
NOTE:  Do  not  remove  the  nozzles.  
 
1   Pump  150  mL  of  prepared  solution  through  each  affected  reagent  line  (A,  B  
and  C).  Wait  20  minutes.  
   
2   Pump  another  100  mL  of  disinfectant  solution  through  each  affected  
reagent  line  and  let  sit  for  20  minutes.  
 
3   Flush  each  affected  reagent  line  with  150  mL  of  deionized  water.    
 
 
4   Reprime  the  stainer  with  fresh  reagents  (see  Section  2.2)  and  clean  nozzles  
according  to  instructions  in  Section  6.1  and  6.2.  
 
5   Perform  a  Volume  Test  according  to  instructions  in  Section  6.4  and  record  
on  PM  Chart.  
 
CAUTION!  
To  avoid  damage  to  the  instrument,  do  not  leave  disinfectant  solution  in  the  
reagent  lines  longer  than  recommended.  This  procedure  should  only  be  
performed  on  an  “as  needed”  basis.  
 
 
 

82  
SECTION  7  
SOLVING  PROBLEMS  
 
7.3  Instrument  Malfunction  
Air  or  Reagent  Leaks  
Repriming  the  instrument  is  usually  unnecessary  unless  a  reagent  bottle  runs  completely  
dry.    
 
An  air  leak  is  usually  to  blame  if  a  smooth  and  continuous  liquid  spray  fails  to  come  from  
the  nozzles.  Carefully  inspect  all  components  in  the  external  reagent  delivery  lines.  Look  
for  loose  connections,  cracks,  or  breaks  that  might  allow  air  to  be  drawn  in  when  the  
pump  operates.  Replace  any  defective  part  or  assembly.  
   
An  internal  leak  may  cause  fluid  to  leak  from  the  line  when  the  pump  is  not  running.  If  
an  abnormal  liquid  spray  still  occurs  after  all  the  external  reagent  delivery  line  
components  have  been  verified,  the  instrument  may  require  service.  Contact  your  
dealer  or  Elitech  Group  for  assistance.  
                   Reagent  Delivery  L ines      
A  reagent  line  leak  between  the  pump  outlet  and  the  nozzle  will  cause  fluid  to  leak  into  
the  interior  of  the  stainer  housing  and  ultimately  onto  the  counter.  If  you  observe  this,  
the  instrument  will  require  service.  Contact  your  dealer  or  Elitech  Group  for  assistance.  
 
WARNING!  
A  break  or  malfunction  in  the  reagent  delivery  system  can  potentially  release  up  to  
500  mL  of  highly  flammable  anhydrous  alcohol  in  and  around  the  instrument.  If  this  
occurs,  carefully  shut  off  the  power  to  the  instrument  and  consult  the  SDS  for  
information  in  handling  alcohol  spills.  Do  not  use  the  instrument  again  until  any  leaks  
are  repaired.  
 
WARNING!  
Electrical  shock  hazard⎯ do  not  open  this  instrument  or  attempt  internal  repairs.  
Refer  servicing  to  qualified  service  personnel.  Contact  your  dealer  or  Elitech  Group  
Service.  

83  
SECTION  7  
SOLVING  PROBLEMS  
 
7.3  Instrument  Malfunction  
Reagent  Level  Detect  System  Errors  
Reagent  A-­‐D  Not  Calibrated  
During  the  second  part  of  calibration,  if  no  bottles  are  detected,  the  display  shows  an  
error  message.  
 
Calibrate  again,  making  sure  that  the  reagent  bottles  are  inserted  in  those  tray  positions  
that  have  been  enabled  in  the  level  detection  system.  
 
Level  Detection  (LD)  Unstable  
If  movement  was  detected  on  bottles  while  calibrating/zeroing,  the  display  shows  an  
error  message.  
 
NOTE:  
While  zeroing  or  calibrating,  do  not  bump  the  instrument  or  lab  bench.  Ensure  that  no  
nearby  equipment  vibrations  can  be  transmitted  to  the  stainer.  
 
Calibrating  the  Reagent  Level  Detect  System  
If  the  Reagent  Level  Detect  System  is  reporting  incorrectly  and  zeroing  (Section  2.2)  
does  not  correct  the  problem,  calibrate  the  system  as  follows:  
 
1     Press  System  Information  from  the  Main  menu.  
 
 
 
2     Press  System  Setup.    
 
 
 
3     Press  Level  Detect.  
 
 
 
 
 
 
 
 
4     Press  Calibrate.  Follow  the  display  prompts.  
 
 
 
 
 

84  
SECTION  7    
SOLVING  PROBLEMS  
 
7.3  Instrument  Malfunction  
Calibrating  the  Reagent  Level  Detect  System  (continued)  
5     Remove  all  reagent  bottles  and  press  Start.    
 
 
 
 
 
 
 
 
 
 
NOTE:  Any  vibrations  or  bumps  to  the  instrument  or  lab  bench  can  cause  
inaccuracies  in  zeroing  or  calibration.    
 
NOTE:  Calibration  requires  full,  unopened  (caps  and  seals  in  place)  500  mL  
bottles  of  reagent,  placed  in  the  correct  tray  positions  (due  to  different  densities  
of  each  reagent  type).  
 
6     Place  the  correct  reagent  bottles  in  all  enabled  positions,  and  press  Start.    
 
 
 
 
 
 
 
 
 
 
NOTE:  The  calibration  function  ignores  any  disabled  reagent  line.  
 
7     Press  OK.  Press  Back  twice  to  return  to  the  Main  menu.  
 
 
8     Return  the  reagent  bottles  to  the  tray  as  indicated  in  Section  2.1  to  
prepare  for  staining.  
 
NOTE:  For  accurate  reagent  level  detection  and  calibration,  dip  tubes  must  
follow  their  pre-­‐formed  coiled  shapes.  

85  
SECTION  7  
SOLVING  PROBLEMS  
 
7.4  Calibrating  the  Touchscreen  
1   Press  and  hold  Standby/Ready  for  5  seconds.  A  calibration  screen  with  a  
target  appears.  
 
2   Press  the  center  of  the  target  with  a  finger,  stylus,  or  similar  tool.  Another  
target  will  appear  in  a  different  location.  
 
3   Continue  to  press  the  center  of  the  targets  until  you  have  pressed  all  the  
targets  (five  total).    After  the  fifth  target  is  pressed,  the  instrument  will  
save  the  touch  screen  calibration  and  return  to  the  Main  menu.  
 

86  
SECTION  7  
SOLVING  PROBLEMS  
 
7.5  Service  Information  
Elitech  Group’s  Service  Department  will  help  you  resolve  any  questions  about  the  operation  or  
performance  of  your  Aerospray  Stainer/Cytocentrifuge.  
 
Customers  in  the  United  States  should  contact  us  by  telephone.  Outside  the  U.S.,  our  authorized  
dealers  offer  full  local  service  and  support.    
 
ELITechGroup  Inc.  
370  West  1700  South  
Logan,  Utah  84321-­‐8212  USA  
 
Telephone:  
800  453  2725  (United  States  &  Canada)  
(+1)  435  752  6011  (International  calls)  
 
Fax:  
(+1)  435  752  4127    
 
Email:  
[email protected]  (Service)  
[email protected]  (Sales)  
 
Web  Page:  
www.elitechgroup.com  
 
European  Authorized  Representative:  
Medical  Technology  Promedt  Consulting  
Altenhofstr.  80  
D-­‐66386  St.  Ingbert  
Germany  
 
Telephone:  +49(0)68  94-­‐58  10  20  
Fax:  +49(0)68  94-­‐58  10  21  
Email:  info@mt-­‐procons.com  

87  
SECTION  8  
CYTOPRO®  CYTOCENTRIFUGE  
 
8.1  Cytopro  Cytocentrifuge  Information  
Functional  Description  
The  Cytopro  Cytocentrifuge  rotor  allows  rapid  sedimentation  of  specimen  cells  onto  
microscope  slides  for  staining  or  other  purposes.  Up  to  eight  disposable/reusable  
sample  chamber  assemblies  with  absorbent  pads  and  glass  microscope  slides  can  be  
loaded  into  the  Cytocentrifuge  rotor.  
Cytocentrifuge  and  staining  functions  are  independent  of  one  another.  
The  Cytopro  rotor  reduces  cell  loss  during  collection  and  prevents  accidental  damage  
to  the  collected  specimen.  The  rotor  is  sealed  to  control  aerosol  release  during  
cytocentrifugation.  See  the  Cytopro  Rotor  Applications  Manual  (Aerospray  Models  
7xx2)  (RP-­‐517)  for  complete  information.  
 
Key  Features  
Adding  the  Cytopro  Cytocentrifuge  rotor  transforms  the  stainer  into  a  standard  
cytocentrifuge  with:  
• Single,  Dual,  and  Cytopro  Magnum  chambers  
• Reusable  or  disposable  chambers  (single  and  dual)  
• Eight  slides  and  chambers    
• User-­‐programmable  memory  locations  for  settings  (speed,  acceleration  rate,  and  
time)  
• Easy  switching  between  staining  and  cytocentrifuge  modes  
• Autoclavable  rotor  

NOTE:  Pressing  Cyto  brings  up  the  Cytocentrifuge  mode.  Pressing  Back  returns  to  stain  
mode.  
 
WARNING!  
The  Cytopro  rotor  lid,  rotor  gaskets  and  related  components  are  intended  to  be  part  of  
biosafety  system  as  specified  in  international  and  national  biosafety  guidelines.  They  
cannot  be  relied  on  as  the  only  means  of  safeguarding  workers  and  the  environment  
when  handling  pathogenic  microorganisms.  
 
Intended  Use  
The  Cytopro  Cytocentrifuge  rotor  is  an  in  vitro  diagnostic  medical  device  for  
professional  use  only.  It  is  an  accessory  for  fixing  biological  cell  suspensions  on  glass  
microscope  slides  for  cytological  examination.  
 
The  Cytocentrifuge  rotor  can  be  used  with  the  following  cell  suspensions:  
• Bronchoalveolar  liquid  (BAL)  
• Cerebrospinal  fluid  (CSF)  
• Urine  
• Synovial  fluid  
• Others  

88  
APPENDIX  A  
Critical  Reagent  Components  
The  following  information  identifies  the  critical  chemicals  of  each  reagent  used  in  this  instrument.  
Reagent(s)   Critical  Components  
SS-­‐035A,  SS-­‐035A-­‐EU,  SS-­‐035AG,  or  SS-­‐135/149A  (as   <1%  Potassium  Phosphate    
diluted)  Hematology  Reagent  A  Rinse  contains:   <1%  Sodium  Phosphate  
  <1%  Eosin  
<1%  Benzyl  Alcohol  
 
SS-­‐035/049B,  SS-­‐035/049B-­‐EU,  SS-­‐035/049BG,  or  SS-­‐ <1%  Potassium  Phosphate    
135/149B  (as  diluted)  Hematology  Reagent  B  Thiazin   <1%  Sodium  Phosphate  
Stain  contains:   <1%  Azure  B  
  <1%  Methylene  Blue  
<1%  Benzyl  Alcohol  
<0.1%  Triton  X-­‐100  
 
SS-­‐035C,  SS-­‐035C-­‐EU,  SS-­‐035CG,  or  SS-­‐135/149C  (as   <1%  Potassium  Phosphate  
diluted)  Hematology  Reagent  C  Eosin  Stain  contains:   <1%  Sodium  Phosphate  
  <1%  Eosin    
<0.2%  Formaldehyde  
<0.1%  Triton  X-­‐100  
 
SS-­‐048,  SS-­‐048-­‐EU,  or  SS-­‐148  (as  diluted)  Aerofix   >95%  Methyl  Alcohol  
Fixative  High  Humidity  contains:   <4%  Ethylene  Glycol  
  <2%  Polyvinylpyrrolidone  (PVP)  
  <1%  Azure  B  
 
SS-­‐049A,  SS-­‐049A-­‐EU,  or  SS-­‐049AG  or  SS-­‐135/149A  (as   <3%  Ethyl  Alcohol  
diluted)  Basofix  Reagent  A  contains:   <3%  Methyl  Alcohol  
  <1%  Potassium  Phosphate  
<1%  Sodium  Phosphate  
<1%  Eosin  
<  0.2%  Formalin  
 
SS-­‐049C,  SS-­‐049C-­‐EU,  or  SS-­‐049CG  or  SS-­‐135/149C  (as   <10%  Diethylene  Glycol  
diluted)  Reagent  C  Basofix  Eosin  Stain  contains:   <5%  Ethyl  Alcohol  
  <5%  Methyl  Alcohol  
<1%  Potassium  Phosphate  
<1%  Sodium  Phosphate  
<1%  Eosin  Y    
 
SS-­‐049P,  SS-­‐049P-­‐EU,  or  SS-­‐149P  (as  diluted)  Basofix   <99%  Methyl  Alcohol  
Pre-­‐dip  Fixative  contains:   <1%  Polyvinylpyrrolidone  
  <1%  Azure  B  
 
SS-­‐135/149A,  Reagent  A  Rinse  Reagent  Concentrate   <1%  Potassium  Phosphate  
contains:   <1%  Sodium  Phosphate  
  <0.2%  Formalin  
<1%  Eosin  
 
SS-­‐135/149B,  Reagent  B  Thiazin  Stain  Concentrate   <10%  Potassium  Phosphate  
contains:   <10%  Sodium  Phosphate  
  <2%  Azure  B  
<1%  Methylene  Blue  
 

89  
APPENDIX  A  
Critical  Reagent  Components  
Reagent(s)   Critical  Components  
SS-­‐135/149C,  Reagent  C  Eosin  Stain  Concentrate   <10%  Potassium  Phosphate  
contains:   <10%  Sodium  Phosphate  
  <0.2%  Formaldehyde  
<3%  Eosin  
 
SS-­‐135S  Surfactant  contains:   <70%  Germall  II  (Diazolidinyl  Urea)  
  <5%  Triton  X-­‐100  
 
SS-­‐148,  Aerofix  Fixative  Concentrate  High  Humidity   55-­‐70%  Ethylene  Glycol  
contains:   30-­‐45%  Polyvinylpyrrolidone  (PVP)  
  <1%  Azure  B  
 
SS-­‐029,  SS-­‐029C  and  SS-­‐029CG  (as  diluted)  Nozzle   40-­‐50%  Methyl  Alcohol  
Cleaning  Solution  contains:   1-­‐5%  Oxalic  Acid  
 
SS-­‐029C,  SS-­‐029CG  Nozzle  Cleaning  Solution   95-­‐99%  Deionized  Water    
Concentrate  contains:   1-­‐5%  Oxalic  Acid  
   
SS-­‐133  Decontamination  Solution  when  diluted  as   <2%  Germicidal  Detergent  
directed  contains:   >98%  Deionized  Water  
   
SS-­‐MeOH  Anhydrous  Methanol  contains:   ≥99.5%  Methyl  Alcohol,  Anhydrous  
   
SS-­‐133  Decontamination  Solution  Concentrate   <30%  Germicidal  Detergent  
contains:   >70%  Deionized  Water  
   
 
 

90  
APPENDIX  B  
Risk  and  Safety  Phrases  
Reagents  SS-­‐035A,  SS-­‐035A-­‐EU,  and  SS-­‐035AG  Reagent  –  A  Rinse,  and  SS-­‐135/149A  –  Rinse  Reagent  
concentrate  after  dilution  as  directed,  are  not  associated  with  any  Risk  and  Safety  Phrases.  They  are  not  
associated  with  any  EU  Symbols  of  Danger.  
 
R  Phrases:   None  
S  Phrases:   None  
 
Reagents  SS-­‐035/049B,  SS-­‐035/049B-­‐EU,  and  SS-­‐035/049BG  –  Reagent  B  Thiazin  Stain,  and  SS-­‐135/149A  –  
Rinse  Reagent  concentrate  after  dilution  as  directed,  are  not  associated  with  any  Risk  and  Safety  Phrases.  
They  are  not  associated  with  any  EU  Symbols  of  Danger.  
 
  R  Phrases:   None  
  S  Phrases:   None  

Reagents  SS-­‐035C,  SS-­‐035C-­‐EU,  and  SS-­‐035CG  –  Reagent  C  Eosin  Stain,  and  SS-­‐135/149A  –  Rinse  Reagent  
concentrate  after  dilution  as  directed,  are  not  associated  with  any  Risk  and  Safety  Phrases.  They  are  not  
associated  with  any  EU  Symbols  of  Danger.  
R  Phrases:   None  
S  Phrases:   None  
 
SS-­‐MeOH  Aerospray  Reagent-­‐Grade  Methanol  is  associated  with  the  following  Risk  and  Safety  Phrases.    
European  Symbols  of  Danger  are:  F  and  T.  
 
R  11:   Highly  Flammable  
R  23/24/25:   Toxic  by  inhalation,  in  contact  with  skin  and  if  swallowed  
R  39/23/24/25:   Toxic:  danger  of  very  serious  irreversible  effects  through  inhalation,  in  contact  with  skin  and  
if  swallowed  
S  7:   Keep  container  tightly  closed  
S  16:   Keep  away  from  sources  of  ignition  –  No  smoking  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  45:   In  case  of  accident  or  if  you  feel  unwell,  seek  medical  advice  immediately  (show  label  where  
possible)  
 
Reagents  SS-­‐048  or  SS-­‐048-­‐EU  –  Aerofix  Fixative,  500  mL,  and  SS-­‐148  –  Aerofix  Additive  for  methanol  
reagent  after  dilution  as  directed,  are  associated  with  the  following  Risk  and  Safety  Phrases.  The  associated  
EU  Symbols  of  Danger  are  F  and  T.  
 
R  11:   Highly  Flammable  
R  23/24/25:   Toxic  by  inhalation,  in  contact  with  skin  and  if  swallowed  
R  39/23/24/25:   Toxic:  danger  of  very  serious  irreversible  effects  through  inhalation,  in  contact  with  skin  and  
if  swallowed  
S  7:   Keep  container  tightly  closed  
S  16:   Keep  away  from  sources  of  ignition  –  No  smoking  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  45:   In  case  of  accident  or  if  you  feel  unwell,  seek  medical  advice  immediately  (show  label  where  
possible)  
 
Reagents  SS-­‐049A,  SS-­‐049AG,  SS-­‐049A-­‐EU–  Basofix  Rinse  Reagent,  and  SS-­‐135/149A  –  Rinse  Reagent  
concentrate  after  dilution  as  directed,  are  not  associated  with  any  Risk  and  Safety  Phrases.  They  are  not  
associated  with  any  EU  Symbols  of  Danger.  
 
R  Phrases:   None  
S  Phrases:   None  
 

91  
APPENDIX  B  
Risk  and  Safety  Phrases  
Reagents  SS-­‐049C  and  SS-­‐049C-­‐EU  –  Basofix  Eosin  Stain,  SS-­‐049CG  or  SS-­‐135/149C  –  Eosin  Stain  concentrate  
after  dilution  as  directed,  are  associated  with  the  following  Risk  and  Safety  Phrases.  The  associated  EU  
Symbol  of  Danger  is  Xn.  
 
R  20/21/22:   Harmful  by  inhalation,  in  contact  with  skin  and  if  swallowed  
R  68/20/21/22:   Harmful:  possible  risk  of  irreversible  effects  through  inhalation,  in  contact  with  skin  and  if  
swallowed  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  45:   In  case  of  accident  or  if  you  feel  unwell  seek  medical  advice  immediately  (show  the  label  
where  possible)  
 
Reagents  SS-­‐049P  and  SS-­‐049P-­‐EU  –  Basofix  Pre-­‐dip  Fixative  and  SS-­‐149P  –  Basofix  Pre-­‐dip  Fixative  
concentrated  when  diluted  as  directed  are  associated  with  the  following  Risk  and  Safety  Phrases.  The  
associated  EU  Symbols  of  Danger  are  F  and  T.  
 
R  11:   Highly  Flammable  
R  23/24/25:   Toxic  by  inhalation,  in  contact  with  skin  and  if  swallowed  
R  39/23/24/25:   Toxic:  danger  of  very  serious  irreversible  effects  through  inhalation,  in  contact  with  skin  and  
if  swallowed  
S  7:   Keep  container  tightly  closed  
S  16:   Keep  away  from  sources  of  ignition  –  No  smoking  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  45:   In  case  of  accident  or  if  you  feel  unwell,  seek  medical  advice  immediately  (show  label  where  
possible)  
 
Reagents  SS-­‐135/149A  –  Rinse  Reagent  Concentrate  is  not  associated  with  any  Risk  and  Safety  Phrases.  It  is  
not  associated  with  any  EU  Symbols  of  Danger.  
 
R  Phrases:   None  
S  Phrases:   None  
 
Reagents  SS-­‐135/149B  –  Thiazin  Stain  Reagent  Concentrate  is  not  associated  with  any  Risk  and  Safety  
Phrases.  It  is  not  associated  with  any  EU  Symbols  of  Danger.  
 
R  Phrases:   None  
S  Phrases:   None  
 
Reagents  SS-­‐135/149C  –  Eosin  Stain  Reagent  Concentrate  is  not  associated  with  any  Risk  and  Safety  
Phrases.  It  is  not  associated  with  any  EU  Symbols  of  Danger.  
 
R  Phrases:   None  
S  Phrases:   None  
 
Reagent  SS-­‐135S  –  Hematology  Surfactant  is  associated  with  the  following  Risk  and  Safety  Phrases.  The  
associated  EU  Symbol  of  Danger  is  Xi.  
 
R  43:   May  cause  sensitization  by  skin  contact  
R  52/53:   Harmful  to  aquatic  organisms,  may  cause  long-­‐term  adverse  effects  in  the  aquatic  
environment  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  61:   Avoid  release  to  the  environment.  Refer  to  special  instructions/safety  data  sheet  
 
 
 
 

92  
APPENDIX  B  
Risk  and  Safety  Phrases  
 
Reagents  SS-­‐148  –  Aerofix  Additive  for  Methanol  is  associated  with  the  following  Risk  and  Safety  Phrases.  The  
associated  EU  Symbol  of  Danger  is  Xn.  
 
R  22:   Harmful  if  swallowed  
S  phrases:     None  
 
SS-­‐029  Aerospray  Nozzle  Cleaning  Solution  and  SS-­‐029C  when  diluted  as  instructed  are  associated  with  the  
following  Risk  and  Safety  Phrases.  European  Symbols  of  Danger  are:  F  and  T.  
 
R  11:   Highly  Flammable  
R  23/24/25   Toxic  by  inhalation,  in  contact  with  skin  and  if  swallowed  
R  39/23/24/25:   Toxic:  danger  of  very  serious  irreversible  effects  through  inhalation,  in  contact  with  skin  and  
if  swallowed  
S  7:   Keep  container  tightly  closed  
S  16:   Keep  away  from  sources  of  ignition  –  No  smoking  
S  36/37:   Wear  suitable  protective  clothing  and  gloves  
S  45:   In  case  of  accident  or  if  you  feel  unwell,  seek  medical  advice  immediately  (show  label  where  
possible)  
 
SS-­‐029C  and  SS-­‐029C-­‐EU  Aerospray  Nozzle  Cleaning  Solution  is  associated  with  the  following  Risk  and  Safety  
Phrases.  European  Symbol  of  Danger  is:  None.  
 
R:   None  
S:   None  
 
SS-­‐133  Decontamination  Solution  Concentrate  is  associated  with  the  following  Risk  and  Safety  Phrases.  
European  Symbols  of  Danger  is  C  
 
R  34   Causes  burns  
R  22   Harmful  if  swallowed  
R  67   Vapors  may  cause  drowsiness  and  dizziness  
S  24/25   Avoid  contact  with  skin  and  eyes  
S  26   In  case  of  contact  with  eyes,  rinse  immediately  with  plenty  of  water  and  seek  medical  advice  
S  36/37/39   Wear  suitable  protective  clothing,  gloves  and  eye/face  protection  
S  45   In  case  of  accident  or  if  you  feel  unwell  seek  medical  advice  immediately  (show  the  label  
where  possible)  
 
SS-­‐103  O-­‐ring/Nozzle  Thread  Grease  is  associated  with  the  following  Risk  and  Safety  Phrases.  European  
Symbols  of  Danger  are:  None.  
 
R:   None  
S:   None  
 
SS-­‐149P  Basofix  Pre-­‐dip  Fixative  Concentrate  when  diluted  as  instructed  is  associated  with  the  following  
Risk  and  Safety  Phrases.  European  Symbols  of  Danger  are:  None.  
 
R:   None  
S:   None  
 
 
 

93  
APPENDIX  C  
Accessories  and  Supplies  
Only  replacement  parts  supplied  by  Elitech  Group  should  be  used  in  this  instrument.  Use  of  non-­‐approved  
parts  may  affect  the  performance  and  safety  features  of  this  product.  
 
ACCESSORIES                       REFERENCE  NUMBER  
Slide  Carousel  (12-­‐Slide  Capacity)  ...........................................................  AC-­‐028  
Nozzle  Tool  ..............................................................................................  AC-­‐034  
Nozzle  Hex  Wrench  .................................................................................  AC-­‐035  
5  L  Space-­‐Saver  Container  w/cap  (For  Concentrate  Reagents)  ...............  AC-­‐038  
Drain  Tube,  1.8  meter  (6  foot)  Length  .....................................................  AC-­‐041  
500  mL  bottle  with  cap  (pack  of  5  bottles)  ..............................................  AC-­‐043-­‐05  
Slide  Carousel  (30-­‐Slide  Capacity)  ...........................................................  AC-­‐057  
Nozzle  Orifice  Cleaning  Wire  ...................................................................  AC-­‐059  
Reagent  Pump  Priming  Tool  ....................................................................  AC-­‐069  
Aerospray/Cytopro  Safety  Shield  ............................................................  AC-­‐110  
Cytopro  Cytocentrifuge  Rotor  .................................................................  AC-­‐160  
10  L  Waste  Container  (without  level  detect)  ..........................................  AC-­‐170  
1D  Barcode  Scanner  ................................................................................  AC-­‐181  
10  L  Waste  Container  (with  level  detect)  ................................................  AC-­‐182  
Nozzle  Maintenance  Kit  ...........................................................................  AC-­‐184  
2D  Barcode  Scanner  ................................................................................  AC-­‐185  
 
SUPPLIES  
Nozzle  Cleaning  Solution,  355  mL  ...........................................................  SS-­‐029  
Nozzle  Cleaning  Solution  Concentrate,  250  mL  (Dilutes  to  500  mL)  .......  SS-­‐029C  or  SS-­‐029C-­‐EU  
Nozzle  Cleaning  Solution  Concentrate,  1.89  L  (Dilutes  to  3.785  L)  .........  SS-­‐029CG  
Reagent  A  (Rinse),  500  mL  .......................................................................  SS-­‐035A  or  SS-­‐035A-­‐EU  
Reagent  B  (Thiazin  Stain),  500  mL  ...........................................................  SS-­‐035/049B  or  SS-­‐035/049B-­‐EU  
Reagent  C  (Eosin  Stain),  500  mL  ..............................................................  SS-­‐035C  or  SS-­‐035C-­‐EU  
Reagent  C  (Eosin  Stain),  3.785  L  (1  gal)  ...................................................  SS-­‐035CG  
Reagent  D  (Methanol),  500  mL  ...............................................................  SS-­‐MEOH  
Reagent  A  (Rinse),  3.785  L  (1  gal)  ............................................................  SS-­‐035AG  
Reagent  B  (Thiazin  Stain),  3.785  L  (1  gal)  ................................................  SS-­‐035/049BG  
Reagent  A  (Rinse)  Concentrate,  135  mL  (Dilutes  to  4.6  L)  ......................  SS-­‐135/149A  SYSTEM  
Reagent  B  (Thiazin  Stain)  Concentrate,  135  mL  (Dilutes  to  4.6  L)  ...........  SS-­‐135/149B  SYSTEM  
Reagent  C  (Eosin  Stain)  Concentrate,  135  mL  (Dilutes  to  4.6  L)    .............  SS-­‐135/149C  SYSTEM  
Surfactant  (used  with  SS-­‐135  A,  B,  C)  ......................................................  SS-­‐135S  
Reagent  D  (Aerofix®  Fixative  High  Humidity),  500  mL  ............................  SS-­‐048  or  SS-­‐048-­‐EU  
Aerofix®  Additive  for  Reagent  D  (Methanol),  135  mL  .............................  SS-­‐148  or  SS-­‐148-­‐EU  
Reagent  A  (Basofix™  Rinse),  500  mL  .......................................................  SS-­‐049A  or  SS-­‐049A-­‐EU  
Reagent  C  (Basofix™  Eosin  Stain),  500  mL  ...............................................  SS-­‐049C  or  SS-­‐049C-­‐EU  
Basofix™  Pre-­‐dip  Fixative,  500  mL  ...........................................................  SS-­‐049P  or  SS-­‐049P-­‐EU  
Basofix™  Pre-­‐dip  Concentrate  .................................................................  SS-­‐149P  
  Decontamination  Solution  Concentrate  ..................................................  SS-­‐133  
O-­‐Ring/Nozzle  Thread  Grease  (3  grams)  .................................................  SS-­‐103  
Preventive  Maintenance  Chart,  pad  of  24  sheets  ...................................  SS-­‐265  
Aerospray  Hematology  Stat  (Model  7122)  Applications  Manual  ............  RP-­‐460  
Paper  Test  Slides  .....................................................................................  RP-­‐500  
 
 

94  

You might also like