Fresenius MultiFiltrate - Service Manual
Fresenius MultiFiltrate - Service Manual
Service Manual
Technical Manual
Edition: 6/03.07
1 Important Information
1.1 Organization of the Technical Manual ..................................................................................... 1-1
1.2 How to Use the Technical Manual............................................................................................ 1-2
1.3 Precautions for Working on the System ................................................................................. 1-3
1.4 Addresses .................................................................................................................................. 1-4
2 Functional Description
2.1 Extracorporeal Circuit............................................................................................................... 2-1
2.1.1 Pumps ......................................................................................................................................... 2-1
2.1.2 Heaters ........................................................................................................................................ 2-2
2.1.3 Pressure Transducer ................................................................................................................... 2-2
2.1.4 Air Detector and Venous Clamp .................................................................................................. 2-2
2.1.5 Optical Detector (Non-Opaque/Opaque Fluid Detector).............................................................. 2-3
2.1.6 Blood Leak Detector .................................................................................................................... 2-3
2.1.7 Heparin Pump.............................................................................................................................. 2-3
2.2 Weighing Units .......................................................................................................................... 2-3
2.3 Ci-Ca Module (Option)............................................................................................................... 2-4
2.4 Functional Test (T1 Test) and Error Messages....................................................................... 2-4
2.4.1 Battery Test, Part 1...................................................................................................................... 2-5
2.4.2 Scales Test.................................................................................................................................. 2-5
2.4.3 Pump Test ................................................................................................................................... 2-6
2.4.4 Pressure Transducer ................................................................................................................... 2-8
2.4.5 Optical Detector........................................................................................................................... 2-9
2.4.6 Air Detector................................................................................................................................ 2-10
2.4.7 Blood Leak Detector .................................................................................................................. 2-11
2.4.8 Heater ........................................................................................................................................ 2-12
2.4.9 Battery Test, Part 2.................................................................................................................... 2-15
2.4.10 Audible Alarm ............................................................................................................................ 2-16
2.4.11 Heparin Pump............................................................................................................................ 2-17
2.4.12 multiDataLink ............................................................................................................................. 2-18
2.4.13 Ci-Ca Module (Option)............................................................................................................... 2-18
2.5 Error Messages........................................................................................................................ 2-20
2.5.1 Alarm Messages ........................................................................................................................ 2-20
2.5.2 Warning Messages.................................................................................................................... 2-22
2.5.3 Fatal Errors................................................................................................................................ 2-30
2.5.4 Error Codes of Scales Lo-Level Routines ............................................................................... 2-33
2.5.5 Error Codes of Pressures Lo-Level Routines.......................................................................... 2-35
2.5.6 Error Codes of Heparin Pump Lo-Level Routines................................................................... 2-36
2.5.7 Error Codes of Pumps Lo-Level Routines............................................................................... 2-37
2.6 Overview of Display Identification Numbers ........................................................................ 2-37
1 Important Information
Specifications For the specifications of the respective system, refer to the current
Operating Instructions.
Circuit diagrams and The identification on the PCB permits the operator/technician to verify if
component layouts the circuit diagram/component layout matches the PCB actually
installed in the system.
Note
Informs the operator that in case of a failure to follow the steps as
described, a specific function will be executed incorrectly or will not be
executed at all, or will not produce the desired effect.
Caution
Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
operators and patients.
Test equipment and The activities described in the Technical Manual require the availability
accessories of the necessary technical test equipment and accessories.
Precautions When working on the open system, the following precautions must be
respected:
Protect the components against ingress of fluids.
Do not touch live parts.
All plugs, connections and components may only be disconnected or
connected if de-energized.
ESD precautions When repairing and when replacing spare parts, observe the applicable
ESD precautions (e.g. EN 100 015-1).
1.4 Addresses
Local service
2 Functional Description
2.1.1 Pumps
All pumps are supplied with 24 V. The nominal voltage of the blood
pump motor is 20 V. The nominal voltage of the other pumps, however,
is 24 V.
A Hall sensor in the pump housing and a permanent magnet in each
pump door monitor the state of the pump door.
The line inserting position of the pump rotors for inserting and removing
the pump segments is detected by a combination of a reed switch (in
the pump housing) and a permanent magnet (in the rotor).
The functional test covers a check of all pumps.
2.1.2 Heaters
The heater foil is applied to the outside of the heater rod and supplied
with approx. 28 V. The temperatures are controlled and monitored by
altogether four NTC sensors (two for the operating processor and two
for the safety processor). This is implemented on P.C.B. LP 122.
The air detector (P.C.B. LP 450-3) serves for the detection of air in the
extracorporeal blood circuit and operates on the ultrasound principle .
Both the transmitter and the receiver are integrated in the drip chamber
holder. Once the level in the venous drip chamber has fallen below a
certain threshold, the venous clamp is closed. This function is executed
independently of the operating or safety processor. The additional
board AD 28 increases the transmitter voltage during Preparation to
ensure that the level of saline solution is reliably detected.
The air detector is subjected to the functional test.
The blood leak detector (P.C.B. LP 125) is provided for the detection of
a potential blood loss through the membrane. It is operated applying a
two-color measuring section. In the course of this, red and green light is
alternately transmitted to a reference receiver or, through the filtrate
line, to a measuring receiver.
Both the transmitter and the receiver are integrated in the line holder.
The blood leak detector is subjected to the functional test.
One Hall sensor each signals when the piston has reached its upper
and lower end of travel. The safety system of the pump comprises a
speed monitoring device (slotted disc with optical sensor) and a motor
current monitoring function. The syringe types are set via a coding
switch (HEX switch).
The weighing units are used for managing the fluid balance during
treatment.
The weighing cells are operated according to the strain gauge principle.
Signal conditioning, including analog-to-digital conversion, is achieved
per weighing cell on the P.C.B. LP 127. The actual weights are
produced by the operating processor.
The scales are subjected to the functional test. To test the scales, a test
weight (ball) of defined value must be taken off each scale. Proper
functioning of the scales can be concluded from the correct difference
between the weights before and after lifting.
Turning power on The Ci-Ca module requires a supply voltage of 24VDC. This voltage is
provided by the multiFiltrate system's power supply unit via the
connector in the lower IV pole support of the IV pole located on the right
of the system.
After the multiFiltrate system was turned on by pressing the power
switch on its rear, the supply voltage of the module is connected. The
operating processor of the Ci-Ca module switches into the standby
mode.
If the multiFiltrate systems is then turned on via the I/O key on the front
of the system, the operating processor of the Ci-Ca module will perform
an internal processor test. This test is performed simultaneously to the
processor test of the multiFiltrate system.
The Ci-Ca module communicates with the multiFiltrate basic system via
a serial interface.
Processor test If a test is not passed successfully, the module will not establish
communications with the multiFiltrate system. It is not possible to
perform a treatment with citrate anticoagulation. The multiFiltrate
system recognizes this problem and displays a message which
proposes to use an alternative anticoagulation equipment (e.g. heparin
pump) and which has to be confirmed by the operator.
T1 test The Ci-Ca module performs its own T1 test, independent of the T1 test
of the multiFiltrate system. This test will be started automatically and
simultaneously to the multiFiltrate T1 test after the prompt whether the
starting conditions are met was confirmed with [OK].
This test cannot be skipped or deselected.
System errors If it is still possible, system errors in the Ci-Ca module are shown on the
multiFiltrate display with the indication that citrate anticoagulation is not
available.
After it has been turned on the rear and the [I/O] key has been pressed,
the system automatically starts the processor test. After completion of
the processor test, the display test will be performed. In this test, the
numerical characters are represented for 2 seconds in all of the three
fonts used. After this test, the functional test (T1 test) is started
automatically. Depending on the configuration in the SETUP (SETUP
automatic), the test is running in the background or (SETUP detailed)
the test steps are represented separately on the display.
At the beginning of the T1 test, the system assumes the terminal voltage
and loads the battery with a defined resistance over the entire test.
Scale I Lithium battery on P.C.B. LP 244- Replace the battery and execute
OP discharged. the calibration step in the Service
Cal. factor missing, error.
program.
Acknowledge with
[START/RESET] key.
Scale I Defect of the EPROM on P.C.B. Replace the P.C.B. and the
LP 127 of the scales specified. weighing bar and execute the
Temp. coeff. factors missing, error.
calibration step in the Service
Acknowledge with
program.
[START/RESET] key.
The +12 V and/or 12 V voltage(s) Check the fuses on the PSU board
is (are) missing. LP 128 and replace the P.C.B.
LP 128, if necessary.
Caution: All pressures must be
checked and calibrated, if
necessary.
Test weight fails to drop back. Ball Check for smoothness and adjust,
mechanics not smooth. if necessary. Weight check in the
Service program (ESC Service).
In the first step, the stop-by-OP function is tested, i.e. the pumps must
stop running. In this step, the safety processor releases the pumps, and
the operating processor does not activate the pumps. In the second
step, all of the four pumps are activated with a rate of 100 ml/min. The
safety processor checks whether the pumps are running at the correct
rate and whether the reed contact (line inserting position) is actuated. In
the third step, the operating processor activates the pumps and the
safety processor disables the pump activation, i.e. the pumps may not
be running. In the event of an error, messages are emitted. Here, the
error messages section only relates to the blood pump. The messages
are identical for the other pumps, i.e. the filtrate, substituate and
dialysate pumps.
Blood pump Blood pump door open. Close the blood pump door.
Blood pump door open. Magnet in blood pump door Replace the blood pump door.
Acknowledge with missing.
[START/RESET] key.
Hall sensor in pump housing not Plug the connector onto the
connected or defective. appropriate P.C.B. LP 123 or
replace the pump housing.
Blood pump Clock pulse generator of motor on Connect the clock pulse generator
P.C.B. LP 123 not connected. socket.
Wrong speed, rate.
Acknowledge with Control electronics (P.C.B. LP 123) Replace P.C.B. LP 123.
[START/RESET] key. of the specified pump defective.
Blood pump +24-V supply voltage missing. Check the fuse on the PSU board
LP 128. If necessary, replace
Stop.
P.C.B. LP 128.
Acknowledge with
[START/RESET] key. Caution: All pressures must be
checked and calibrated, if
necessary.
In the first step, the zero point, amplification and detuning factors are
tested. In the second step, the zero points of the pressure transducers
are checked with a tolerance of ±20 mmHg. In the third step, Part and
Pven are detuned to 300 mmHg. The values are checked with a
tolerance of ±20 mmHg. In the fourth step, PPHF and PFil are detuned
to 300 mmHg. The values are checked with a tolerance of ±20 mmHg.
In the event of an error, messages are emitted. The following table only
shows the error messages for "arterial pressure". The messages are
identical for the venous, pre-hemofilter and filtrate pressures.
Arterial pressure Lithium battery on P.C.B. LP 244- Replace the battery and execute
OP discharged. the calibration step in the Service
Factors / Offset lost.
program.
Acknowledge with
[START/RESET] key.
Arterial pressure Zero point drifted off. Check the setting and calibrate the
pressure in the Service program.
Zero outside tolerance.
Acknowledge with P.C.B. LP 343-1 (for venous Replace P.C.B. LP 343-1 (LP 450-
[START/RESET] key. pressure: P.C.B. LP 450-3) 3) and calibrate the pressure in the
defective. Service program.
Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.
Arterial pressure Amplification drifted off. Check the setting and calibrate the
pressure in the Service program.
Detuning outside tolerance
Acknowledge with P.C.B. LP 343-1 (for venous Replace P.C.B. LP 343-1 (LP 450-
[START/RESET] key. pressure: P.C.B. LP 450-3) 3) and calibrate the pressure in the
defective. Service program.
Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.
In the first step, the non-opaque state is checked. In the second step,
the optical detector is detuned and the opaque state is checked. In the
event of an error, messages are emitted.
Opt. detector Blood present in the system or Remove the blood line or the
objects inserted in the OD. objects from the OD.
Senses opaque fluid.
Acknowledge with OD adjusted improperly. Check the setting of the OD.
[START/RESET] key.
P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.
Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.
Air detector LD adjusted improperly (LDA2). Check the setting of the LD.
Clamp does not close. P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.
Acknowledge with
[START/RESET] key. Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.
In the first step, the blood leak detector is checked for being in an
acceptable state. In the second step, the blood leak detector is detuned
and the alarm state checked. In the event of an error, messages are
emitted.
Blood leak detector Lithium battery on P.C.B. LP 244- Replace the battery and execute
SP discharged. the calibration step in the Service
Calibration values incorrect
program.
or missing.
Acknowledge with
[START/RESET] key.
Blood leak detector Empty filtrate line or objects Remove the filtrate line or the
inserted in the BLD sensor head. objects.
Outside acceptable range.
Remove filtrate line. P.C.B. LP 125 and/or BLD sensor Replace P.C.B. LP 125 and the
Acknowledge with head defective. BLD sensor head, and calibrate the
[START/RESET] key. BLD.
Blood leak detector P.C.B. LP 125 and/or BLD sensor Replace P.C.B. LP 125 and the
head defective. BLD sensor head, and calibrate the
Alarm-free after signal attenuation.
BLD.
Acknowledge with
[START/RESET] key. P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.
Caution: All scales and pressures
must be calibrated.
2.4.8 Heater
In the first step, the safety relay is checked for being in the open state.
In addition, the sensors are checked for interruption and short-circuit. In
the second step, the safety relay is activated and checked for being in
the closed state. In the third step, the bag and foil sensors of OP and
SP are checked for being synchronized. Tolerances are ±1 °C. In the
fourth step, the foil sensor is detuned to >120 °C and, thus, the safety
shutoff mechanism (fire protection) is checked. In the fifth step, the bag
and foil sensors are detuned to >41 °C to check whether the heater is
switched off in case of an overtemperature of the solutions. In the sixth
step, it is checked whether the heater can be activated. In the event of
an error, messages are emitted. Here, the error messages section only
relates to the lower heater. The messages are identical for the upper
heater.
Lower heater Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.
Bag sensor broken. Sensor interrupted. Replace the heater.
Acknowledge with
[START/RESET] key. P.C.B. LP 122 defective Replace P.C.B. LP 122.
Lower heater Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.
Foil sensor broken. Sensor interrupted. Replace the heater.
Acknowledge with
[START/RESET] key. P.C.B. LP 122 defective Replace P.C.B. LP 122.
Lower heater Supply voltage for heater not Fit the socket on P.C.B. LP 122.
connected.
Sub. relay not closed,
foil defective. Fuse on P.C.B. LP 122 defective. Replace the fuse.
Acknowledge with
[START/RESET] key. Heater foil not connected. Connect the heater foil to P.C.B.
LP 122.
Lower heater Heater still warm after last Allow the heater to cool down.
treatment or Service program.
Bag sensors,
OP-SP not synchronous Sensor board (P.C.B. LP 1220) Replace the heater.
Acknowledge with defective.
[START/RESET] key.
P.C.B. LP 122 defective Replace P.C.B. LP 122.
Lower heater Heater still warm after last Allow the heater to cool down.
treatment or Service program.
Foil sensors,
OP-SP not synchronous Sensor board (P.C.B. LP 1220) Replace the heater.
Acknowledge with defective.
[START/RESET] key.
P.C.B. LP 122 defective Replace P.C.B. LP 122.
The first step was carried out at the beginning of the functional test, with
acceptance of the starting value and loading of the battery. In the
second step, the voltage is checked after loading and is compared with
the starting value. In the event of an error, messages are emitted.
Battery test Resistor or load relay on the PSU, Replace P.C.B. LP 128.
P.C.B. LP 128, defective.
Load test failed. Caution: All pressures must be
Acknowledge with checked and calibrated, if
[START/RESET] key. necessary.
Battery test System not used for a prolonged Load the battery from the mains for
period or frequent power failure a minimum of 10 hours.
Insufficient capacity.
(preparation).
Acknowledge with Replace the battery.
[START/RESET] key. Battery defective.
Audible alarm Constant ambient noise. Repeat the test after the noise has
stopped.
Not silenced.
Acknowledge with Loud humming noise on the Check the route of the loudspeaker
[START/RESET] key. loudspeaker. cable. Replace the loudspeaker-
microphone unit or the
motherboard P.C.B. LP 124.
1 = 30 ml heparin syringe
2 = 50 ml Injectomat syringe
Heparin pump Not used in the multiFiltrate. Not used in the multiFiltrate.
CAMUS transmission error
2.4.12 multiDataLink
The test for the Ci-Ca module can not be enabled if the processor test
of the module and the multiFiltrate was not passed successfully. If these
processor tests are performed successfully, the T1 test will be started
automatically as soon as the operator has confirmed that the starting
conditions are met.
The following functions of the Ci-Ca module are tested:
Stop of the citrate and the calcium pump by the module's operating
processor
Stop of the citrate and the calcium pump by the system's watchdog.
Function of the citrate pump and the insertion switch
Function of the calcium pump and the insertion switch
If a test could not be passed successfully, a warning indicating an error
number appears on the multiFiltrate display. The test can be repeated
any number of times, however, it cannot be skipped.
The T1 test of the multiFiltrate system is performed simultaneously to
the T1 test of the Ci-Ca module.
281 Lacking Ci Hall impulse after test was started The rotor of the citrate pump is loose, jammed or
(>5 s). blocked.
282 Ci Hall impulse too early during pump stop test Insertion switch pressed --> citrate line already
(<7 s). inserted.
283 Ci Hall impulse too late during pump stop test The rotor of the citrate pump is loose, jammed or
(>9 s). blocked.
285 Ci Hall impulse too early during pump stop test Insertion switch pressed --> citrate line already
(<7 s). inserted during the test.
286 Ci Hall impulse too late during pump stop test The rotor of the citrate pump is loose, jammed or
(>9 s). blocked.
288 Ci Hall impulse too early during insertion switch Insertion switch pressed --> citrate line already
test (<7 s). inserted during the test.
289 Ci Hall impulse too late during insertion switch The rotor of the citrate pump is loose, jammed or
test (>9 s). blocked.
291 Lacking Ca Hall impulse after test was started The rotor of the Ca pump is loose, jammed or
(>5 s). blocked.
292 Ca Hall impulse too early during pump stop test Insertion switch pressed --> calcium line already
(<7 s). inserted.
293 Ca Hall impulse too late during pump stop test The rotor of the Ca pump is loose, jammed or
(>9 s). blocked.
295 Ca Hall impulse too early during pump stop test Insertion switch pressed --> calcium line already
(<7 s). inserted during the test.
296 Ca Hall impulse too late during pump stop test The rotor of the Ca pump is loose, jammed or
(>9 s). blocked.
298 Ca Hall impulse too early during insertion switch Insertion switch pressed --> calcium line already
test (<7 s). inserted during the test.
299 Ca Hall impulse too late during insertion switch The rotor of the Ca pump is loose, jammed or
test (>9 s). blocked.
Code Description
E28 End of emergency operation, the unit will turn itself off
automatically after 5 minutes.
Converted by OP from SP code 189.
Code Description
E50 Ci-Ca module: The citrate bag change takes more than
2 min; finish bag change!
Code Description
These messages cause the balancing to stop. The blood circuit will be
maintained.
For error messages no. 132 through 144 the scales must be
recalibrated.
Code Description
Code Description
Code Description
Code Description
Code Description
W177 Filtrate scale: Test weight fails to drop back (T0 test)
Code Description
W189 End of emergency operation: The unit will turn itself off
automatically after 5 minutes
Code Description
Code Description
Code Description
W331 Ci-Ca module: Filtrate rate too low for set Ca dose!,
Increase dialysate or substituate or Ca dose if
necessary!
W332 Ci-Ca module: Filtrate rate too high for set Ca dose!,
Reduce dialysate or substituate or Ca dose if
necessary!
W333 Ci-Ca module: Blood flow too low for set citrate dose!,
Increase blood flow or citrate dose if necessary!
W334 Ci-Ca module: Blood flow too high for set citrate dose!,
Reduce blood flow or citrate dose if necessary!
Code Description
E519 Self-Test
E520 Balancing
Code Description
E542 Ci-Ca module: Missing setup values from the MFT (at
end of line installation)
E550 Ci-Ca module: Too many extra Hall pulses from the Ca
pump
E560 Ci-Ca module: Too many extra Hall pulses from the
citrate pump
Scale no. 1
Code Description
Scale no. 2
Code Description
Scale no. 3
Code Description
Scale no. 4
Code Description
Code Description
PART
Code Description
PPHF
Code Description
PVEN
Code Description
PFIL
Code Description
Code Description
Code Description
No. Description
M1 Start-up screen
No. Description
No. Description
No. Description
3 Installation
3.1 Preface
For initial start-up only This technical document is intended for initial start-up only. It is not
intended for restarting systems that have been shut down or have been
put out of service temporarily.
Tester's qualification The initial start-up must be performed by the Technical Service of
Fresenius Medical Care or a person authorized by them.
The initial start-up procedure may only be performed by persons who
are qualified to properly perform the specified checks owing to their
educational background and training, their knowledge and experience
gained in practice. Furthermore, the persons performing the tests must
not be bound by any directives when performing this activity.
Test equipment and The activities described in this technical document require the
accessories availability of the necessary technical test equipment and accessories.
Any information on the specifications must be observed.
TSC/TMC/MA intervals The TSC/TMC/MA procedures for this system are to be performed after
12 months.
1 Unpacking
1.2 Check that all parts of the system have been accounted for.
2.3 Connect the power cable and secure it to prevent improper replacement.
3 Visual inspections
3.1 Fuses accessible from the outside comply with the indicated values.
4 Functional check
Blood pump
Filtrate pump
Substituate pump
Dialysate pump
4.2 The venous occlusion clamp closes after an air detector alarm.
4.3 The pressure of 2 bar applied in the venous bubble catcher may not drop by more
than 0.1 bar within 3 minutes.
Jumper J1-P.C.B. LP 450 set to the operation position. Place the checking block into
the air detector. LEDs DI 5 and DI 10 are light.
Jumper J1-P.C.B. LP 450 set to the operation position. Place the adjusting block
into the air detector. LEDs DI 5 and DI 10 are dark.
Scale I
_______g
Calibration: Test load on scales: display 5000 g ±1 g
Scales II
_______g
Calibration: Test load on scales: display 5000 g ±1 g
Scales III
_______g
Calibration: Test load on scales: display 5000 g ±1 g
Scales IV
_______g
Calibration: Test load on scales: display 5000 g ±1 g
5.2 Adhesive labels of citrate and calcium pumps are present and legible.
7.2 Protective earth resistance max. 0.3 (with power cable) ______
scaled to nominal voltage (maximum 480 A, see Additional conditions): _____ A
8 Functional test
9 Final check
9.2 Operating Instructions and accessories package complete and match the system.
Comments:
The system has been released for its intended use. Yes
(Attach inspection sticker.)
No
Next
inspection date:
Comments:
Identification
Technician's name:
Technician's first name and surname.
Customer/customer no.:
Number of the final customer.
Service report number:
Number of the service call.
System type including option(s):
System name with possible options and extras.
System no.:
Serial number indicated on the type label.
Inventory no.:
Inventory number assigned to the system.
Operating hours:
Operating hours, if a time meter is installed.
Equipment code:
Equipment code indicated on the system.
(e.g. EC xxx, E-code xxx)
Re: 1.1
Remove the system from its transport packaging and save the
packaging for possible future return shipments.
Re: 1.2
Check that all parts of the system have been accounted for. Inform your
contact at Fresenius immediately of any missing parts.
5
6
Legend
1 Luer locks
2 Venous clamp
3 Filtrate bag hook
4 IV pole and second IV pole (Ci-Ca module option)
5 Potential equalization
6 RS232 screwed connection
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
The same measurement kit (M28 060 1) as for the patient leakage
current measurement is used for the measurement. Place the
measuring plates on both bag trays and insert the heater bags filled
with NaCl. Measurement setup, see appendix.
When performing a direct measurement, the following precautions
also must be observed:
The system must be insulated when installed. All external
connections and the potential equalization must have been removed
from the system.
Additional conditions:
If the device leakage current is higher than 90 % of the admissible alarm
limit (450 µA), the last measured value or the first measured value must
additionally be considered for the rating.
If the device leakage current considerably increased since the last
measurement or continuously increased since the first measurement
(creeping deterioration of the insulation), or if the sum composed of the
current value plus the difference since the last measurement is
> 500 µA, the measurement has not been passed.
Example 1:
Leakage current: 470 µA
Last measured value: 450 A
470 + (470 450) = 470 + 20 = 490
OK
Example 2:
Leakage current: 470 µA
Last measured value: 390 µA
470 + (470 390) = 470 + 80 = 550
Not passed
Re: 5.3 and 5.4 Patient and device leakage current measurement
Fig.: Setup of the measurement kit for measuring the patient leakage
current and the device leakage current
3
4
Legend
Re: 5.4.1 Patient leakage current measurement for degree of protection (BF)
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
Neither the system nor the test equipment must be touched during
the measurement nor must a cable be connected to the serial
interface.
The patient leakage current measurement kit (M28 060 1) is used for
the measurement. Place the measuring plates on both bag trays and
insert the heater bags filled with NaCl. Measurement setup, see
appendix.
Documentation covers the nominal voltage during the measurement
and the patient leakage current at line voltage, scaled to the nominal
line voltage.
Example:
Measurement value: 48 µA, measured at 225 V
Nominal voltage: 230 V
Scaled to nominal voltage: 49 µA
(48 µA: 225 V x 230 V = 49 µA)
Re: 5.4.2 Patient leakage current measurement for degree of protection (CF)
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
Neither the system nor the test equipment must be touched during
the measurement nor must a cable be connected to the serial
interface.
The patient leakage current measurement kit (M28 060 1) is used for
the measurement. Place the measuring plates on both bag trays and
insert the heater bags filled with NaCl. Measurement setup, see
appendix.
Documentation covers the nominal voltage during the measurement
and the patient leakage current at line voltage, scaled to the nominal
line voltage.
Example:
Measurement value: 28 µA, measured at 244 V
Nominal voltage: 230 V
Scaled to nominal voltage: 26 µA
(28 µA: 244 V x 230 V = 26 µA)
Comments:
Irregularities encountered during the test will be recorded in this section.
Date, signature, stamp
Performance of the test has to be confirmed by indicating date, tester's
signature and stamp.
Caution
Before the Ci-Ca module is installed, the power plug of the system must
be disconnected.
The Ci-Ca module is provided with spring pressure pieces, which make
the module click into place at the IV pole on the left side of the
multiFiltrate system. If the securing lever on the rear of the module has
not been unlocked yet, it has to be done before installing the module by
removing the locking screw. The securing lever has to point up when
the module is connected. Place the module onto the lower IV pole
bearing and slide it along the multiFiltrate housing panel, over the action
point of the pressure pieces and onto the IV pole, until the pressure
pieces click into place. Make sure that the electrical connectors easily
slide into each other.
When the module is locked in place at the IV pole, the securing lever
with pressure roll has to be thrown and then secured with the locking
screw so that it cannot be opened. In this case, another action point
must be overcome.
Tester's qualification The initial start-up must be performed by the Technical Service of
Fresenius Medical Care or a person authorized by them!
The tests may only be performed by persons who are qualified to
properly perform the specified checks owing to their educational
background and training, their knowledge and experience gained in
practice. Furthermore, the persons performing the tests must not be
bound by any directives when performing this activity.
Test equipment and The activities described in this technical document require the
accessories availability of the necessary technical test equipment and accessories.
Any information on the specifications must be observed.
TSC/TMC/MA intervals The TSC/TMC/MA procedures for this system are to be performed after
12 months.
1 Visual inspections
TSC 1.1 Fuses accessible from the outside comply with the indicated values.
MA 1.6 Replace the lithium battery of operating and safety processors (P.C.B. LP
244) every 4 years. ________
2 Extracorporeal components
Blood pump
Filtrate pump
Substituate pump
Dialysate pump
TSC 2.2 Pump rotors checked for damage and rolls for smooth running.
Blood pump
Filtrate pump
Substituate pump
Dialysate pump
TSC 2.3 The venous occlusion clamp closes after an air detector alarm.
TSC 2.4 The pressure of 2 bar applied in the venous bubble catcher may not drop
by more than 0.1 bar within 3 minutes.
Jumper J1-P.C.B. LP 450 set to the operation position. Place the checking
block into the air detector. LEDs DI 5 and DI 10 are light.
Jumper J1-P.C.B. LP 450 set to the operation position. Place the adjusting
block into the air detector. LEDs DI 5 and DI 10 are dark.
MA 2.7 Blood leak detector values checked for red coloration and dimness.
3 Mechanical components
MA 3.1 Mechanical parts of scales checked for tight seat, parallelism and
smoothness.
Scale I
Scales II
Scales III
Scales IV
Scale I
Zero load: display within a range from 60 g to 4500 g _______g
Calibration: Test load on scales: display 5000 g ±1 g
_______g
Ball weight: display within a range from 43.8 g to 44.8 g
_______g
Scales II
Zero load: display within a range from 60 g to 4500 g _______g
Calibration: Test load on scales: display 5000 g ±1 g
_______g
Ball weight: display within a range from 43.8 g to 44.8 g
_______g
Scales III
Zero load: display within a range from 60 g to 4500 g _______g
Calibration: Test load on scales: display 5000 g ±1 g
_______g
Ball weight: display within a range from 43.8 g to 44.8 g
_______g
Scales IV
Zero load: display within a range from 60 g to 4500 g _______g
Calibration: Test load on scales: display 5000 g ±1 g
_______g
Ball weight: display within a range from 43.8 g to 44.8 g
_______g
MA 3.3 Rotary selector checked for easy movement and tight fit
TSC 4.2 Adhesive labels of citrate and calcium pumps are present and legible.
TSC 6.2 Protective earth resistance max. 0.3 (with power cable) ______
TSC 6.4.1 For degree of protection type "BF" according to fig. C.8
scaled to nominal voltage (maximum 480 A, see Additional conditions): _____ A
TSC 6.4.2 For degree of protection type "CF" according to fig. C.8
7 Functional test
TSC 7.1 Functional test (T1 test) checked for proper performance.
8 Final check
Comments:
The system has been released for its intended use. Yes
(Attach inspection sticker.)
No
Next
inspection date:
Comments:
1 Visual inspections
TSC 1.1 Fuses accessible from the outside comply with the indicated values.
2 Extracorporeal components
Blood pump
Filtrate pump
Substituate pump
Dialysate pump
TSC 2.2 Pump rotors checked for damage and rolls for smooth running.
Blood pump
Filtrate pump
Substituate pump
Dialysate pump
TSC 2.3 The venous occlusion clamp closes after an air detector alarm.
TSC 2.4 The pressure of 2 bar applied in the venous bubble catcher may not drop
by more than 0.1 bar within 3 minutes.
TSC 3.2 Adhesive labels of citrate and calcium pumps are present and legible.
TSC 5.2 Protective earth resistance max. 0.3 (with power cable) ______
TSC 5.4.1 For degree of protection type "BF" according to fig. C.8
scaled to nominal voltage (maximum 480 A, see Additional conditions): _____ A
TSC 5.4.2 For degree of protection type "CF" according to fig. C.8
6 Functional test
TSC 6.1 Functional test (T1 test) checked for proper performance.
7 Final check
Comments:
The system has been released for its intended use. Yes
(Attach inspection sticker.)
No
Next
inspection date:
Comments:
Identification
Technician's name:
Technician's first name and surname.
Customer/customer no.:
Number of the final customer.
Service report number:
Number of the service call.
System type including option(s):
System name with possible options and extras.
System no.:
Serial number indicated on the type label.
Inventory no.:
Inventory number assigned to the system.
Operating hours:
Operating hours, if a time meter is installed.
Equipment code:
Equipment code indicated on the system.
(e.g. EC xxx, E-code xxx)
For more detailed information on item 3.2, please refer to the Quick
Guide on the PC Service Software MFT.
5
6
Legend
1 Luer locks
2 Venous clamp
3 Filtrate bag hook
4 IV pole and second IV pole (Ci-Ca module option)
5 Potential equalization
6 RS232 screwed connection
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
The same measurement kit (M28 060 1) as for the patient leakage
current measurement is used for the measurement. Place the
measuring plates on both bag trays and insert the heater bags filled
with NaCl. Measurement setup, see appendix.
When performing a direct measurement, the following precautions
also must be observed:
The system must be insulated when installed. All external
connections and the potential equalization must have been removed
from the system.
Additional conditions:
If the device leakage current is higher than 90 % of the admissible alarm
limit (450 µA), the last measured value or the first measured value must
additionally be considered for the rating.
If the device leakage current considerably increased since the last
measurement or continuously increased since the first measurement
(creeping deterioration of the insulation), or if the sum composed of the
current value plus the difference since the last measurement is
> 500 µA, the measurement has not been passed.
Example 1:
Leakage current: 470 µA
Last measured value: 450 A
470 + (470 450) = 470 + 20 = 490
OK
Example 2:
Leakage current: 470 µA
Last measured value: 390 µA
470 + (470 390) = 470 + 80 = 550
Not passed
Re: 4.3 and 4.4 Patient and device leakage current measurement
Fig.: Setup of the measurement kit for measuring the patient leakage
current and the device leakage current
3
4
Legend
Re: 4.4.1 - Patient leakage current measurement for degree of protection (BF)
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
Neither the system nor the test equipment must be touched during
the measurement nor must a cable be connected to the serial
interface.
The patient leakage current measurement kit (M28 060 1) is used for
the measurement. Place the measuring plates on both bag trays and
insert the heater bags filled with NaCl. Measurement setup, see
appendix.
Documentation covers the nominal voltage during the measurement
and the patient leakage current at line voltage, scaled to the nominal
line voltage.
Example:
Measurement value: 48 µA, measured at 225 V
Nominal voltage: 230 V
Scaled to nominal voltage: 49 µA
(48 µA: 225 V x 230 V = 49 µA)
Re: 4.4.2 Patient leakage current measurement for degree of protection (CF)
Basic requirements:
Measurement of the protective earth resistance performed.
The system is in standby operating mode (system connected to the
mains).
Neither the system nor the test equipment must be touched during
the measurement nor must a cable be connected to the serial
interface.
The patient leakage current measurement kit (M28 060 1) is used for
the measurement. Place the measuring plates on both bag trays and
insert the heater bags filled with NaCl. Measurement setup, see
appendix.
Documentation covers the nominal voltage during the measurement
and the patient leakage current at line voltage, scaled to the nominal
line voltage.
Example:
Measurement value: 28 µA, measured at 244 V
Nominal voltage: 230 V
Scaled to nominal voltage: 26 µA
(28 µA: 244 V x 230 V = 26 µA)
Comments:
Irregularities encountered during the test will be recorded in this section.
Date, signature, stamp
Performance of the test has to be confirmed by indicating date, tester's
signature and stamp.
510 130 1 ESC tester with measuring adapter for patient leakage
current
5.2.1 Start
The system must be turned off via I/O. The power switch on the rear of
the system must be turned on.
Press and hold the Start/Reset key. Simultaneously press the I/O
until the yellow status indicator lights up.
Using the rotary selector, select the numerical code and confirm
each number by pressing the OK key.
After the password has been checked positively, all revisions are
displayed to the left of the screen. The menu fields of the Service
program are displayed to the right of the screen.
Use the rotary selector to select the desired menu field and press
OK.
Use the rotary selector to select the desired language and press OK.
Turn off the system via the I/O key immediately thereafter.
Once again turn on the system via the I/O key and restart the Service
program.
Only the error messages (E XX) and warnings (W XX) of the previous
treatments are filtered out of the events memory (see below) and
displayed on the screen.
Note
If the system does not contain any installed heparin syringe, the
machine status should be set to Heparin syringe currently not active.
If the setting fails to reflect the proper machine status, a warning is
displayed during the T1 test.
old Tubing system with heater bag downstream of the pump segment
"Fin" Tubing system with heater bag upstream of the pump segment
Activating and This soft key permits activation and deactivation of the optional mDL.
deactivating mDL
Screen display Machine status
Note
If the system does not contain any installed mDL, the machine status
should be set to mDL currently not active. If the setting fails to reflect
the proper machine status, a warning is displayed during the T1 test.
DHCP active / inactive This soft key informs the mDL of whether a DHCP server is active in the
connected network. If this is the case, the mDL receives its IP
addresses from this server. If no, the IP addresses must be set
manually.
Note
Before setting the mDL parameters, please clarify with the responsible
hospital network administrator whether a DHCP server is available for
automatically assigning the IP addresses. If no, the administrator has to
specify defined addresses which will then have to be entered manually.
The port numbers depend on the application which further processes
the multiFiltrate data. The administrator of this software will have to
specify the appropriate values if these are different from default values.
Activating/deactivating the This soft key permits activation and deactivation of the optional
Patient/Case ID patient/case ID.
The screen displays the zero load weights of the scales 1 to 4. Upon
start of the Service program, the scales are not tared. Hence, a display
value within a range from 2500 g to 3500 g is expected for scales 1 and
2. A display value within a range from 250 g to 350 g is expected for
scales 3 and 4. Each of the scales (1 to 4) is provided with its own menu
field for taring and calibrating purposes. The tare and calibrate
procedures are described by the example of scale 1. The same
procedures apply to scales 2 to 4.
Requirement The system must be turned on via I/O for at least 5 minutes to bring the
scales to operating temperature.
Using the rotary selector, select the Tare field under Scale I and
press OK.
To calibrate the scales 1 and 2, place the test weight centrally on the
rear part of the scale pan.
To calibrate the scales 3 and 4, suspend the test weight from a hook.
Use the rotary selector to select the Calibrate field of the loaded
scale and press OK.
While calibration is in progress, the field selected is represented light-
green. The time out is approx. 10 seconds.
Use the rotary selector to select the Pressure sensor open? OK
to confirm! field and press OK.
Test equipment Pressure gauge with a measurement range from 0350 mmHg
(0500 mbar), disposable syringe, artery forceps
Measuring instrument
Use the rotary selector to select the Pressure sensor open? OK
to confirm! field and press OK.
While the self-test is in progress, the field selected is represented light-
green. The time out is approx. 10 seconds.
Note
If an error occurs during calibration, check and, if necessary, calibrate
the corresponding evaluation board (P.C.B. LP 450-3 or P.C.B. LP 343-
1).
Use the rotary selector to select Verify system values and press
OK.
Using this service step, all relevant parameters as well as the function
of the pumps, pressure transducers and scales can be verified.
Press the Start/Reset key to activate the heater.
Caution
If heater bags are not inserted and if there is no fluid in the system, the
heater will become very hot.
Ci-Ca module (option) Citrate and calcium pump can be limitedly be activated in this menu.
Press the STOP key.
The pumps stop
Press the START/RESET key
The pumps start running at a delivery rate of 500ml/h if the
insertion switch is pressed (tubing inserted)
Screen Temperature
UFVxxxx UF volume in ml
Legend
x Numerical character
y Lower-case letter
Y Upper-case character
Using this menu, the treatment types can be programmed in the order
as they are used most frequently, while those treatment types that are
not desired can be hidden.
Use the rotary selector to select the desired treatment modes and
press [OK].
Turn off the system via the I/O key immediately thereafter.
Once again turn on the system via the I/O key and restart the Service
program.
Note
The selected treatment modes remain active only if you turn the system
off and back on again immediately after terminating the selection.
All settings relevant for the Ci-Ca treatment are performed in this menu.
Use the rotary selector to select the desired field and press [OK].
Use the rotary selector to enter the desired value and press [OK].
Note
To reduce the risk of using the wrong citrate or calcium containers, only
one type of container (container size and concentration) should be used
throughout the hospital or a comparable organization institution. Save
the same settings for citrate and calcium containers in the setup of all
multiFiltrate systems of this organization institution.
Note
Carry out all steps described below in the Service program.
Measurement setup
Test equipment Pressure gauge, disposable syringe, venous tubing system with
occlusion clamp or artery forceps or test tubing, stop watch
With the open line end in front, insert the measuring instrument into
the venous occlusion clamp.
Using the syringe, build up a pressure of 2 bar. Then, using the
artery forceps, occlude to the direction of the syringe so that the
pressure gauge is still under load.
Start the stop watch. After 3 minutes have elapsed, the pressure
applied may not drop by more than 0.1 bar.
Zero point Check the pressure transducers in the Service program in the Check
system values menu item. With open pressure transducers, the
pressures displayed must be within a range from 5 mmHg to
+5 mmHg.
Measurement setup
P3 P1 P2 P4 P5 DI 5 DI 10
Jumper J1
MP3
MP2
P1 P2 P3 P4 P5 MP3 MP2
DI 5
DI 10
Jumper J1
Setting the zero point Setting the zero point with open pressure transducers.
Connect the negative pole of the multimeter to MP3 (0 V).
Connect the positive pole of the multimeter to MP2.
Use the potentiometer P3 to set the following voltage.
Venous 3 V 0.2 V
Setting the amplification For verification and setting purposes, apply a pressure of 300 mmHg
20 mmHg to the pressure transducers.
Use the potentiometer P4 to set the following voltage.
Note
Whenever the P.C.B. LP 450-3 is replaced, the air detector must be
recalibrated.
Test equipment Adjusting block and checking (reference) block M36 067 1
Check using the checking Set the jumper J1 / P.C.B. LP 450-3 to the operation position
block (red) (NORM OP).
Fill the spherical recesses of the checking block with grease and
remove any excess grease so that only the recesses themselves will
be struck with grease.
Place the greased checking block into the drip chamber holder. It
must be ensured that the ultrasonic sensors must correctly click into
place into the spherical recesses of the adjusting block. The block
must not touch the holder wall, but must hang freely between the
sensors.
Desired function: The LEDs DI 5 and DI 10 on the P.C.B. LP 450-3 are
light.
If the desired function fails to be fulfilled, the ultrasonic detector must be
set.
Check using the adjusting Set the jumper J1 / P.C.B. LP 450-3 to the operation position
block (grey) (NORM OP).
Fill the spherical recesses of the adjusting block with grease and
remove any excess grease so that only the recesses themselves will
be struck with grease.
Place the greased adjusting block into the drip chamber holder. It
must be ensured that the ultrasonic sensors must correctly click into
place into the spherical recesses of the adjusting block. The block
must not touch the holder wall, but must hang freely between the
sensors.
Calibration Start the Service program and select "Verify system values".
Hold the neutral filter (640 560 1) (to be used in 2 layers) into the
optical detector. Close the flap to minimize external light.
Slowly turn the potentiometer P5 on P.C.B. LP 450-3 until the
audible signal turns silent.
Slowly turn potentiometer P5 until the audible signal is sounded.
Then turn it for another half-revolution.
Setting the zero point Setting the zero point with open pressure transducers.
Connect the negative pole of the multimeter to the shield connection
of a jack connector.
Connect the positive pole of the multimeter to MP31.
Use the potentiometer R24 to set the following voltage.
PHF 2 V 0.2 V
Setting the amplification For verification and setting purposes, apply a pressure of 300 mmHg
20 mmHg to the pressure transducers (see Measuring instrument,
chapter 5.1.9, Calibrating the Pressures).
Use the potentiometer R25 to set the following voltage.
Calibrate A filtrate tubing set filled with water must be inserted in the BLD
measuring head.
Via PC with Service Software MFT
Initialization
After turning the system on, the heparin pump will be initialized by
moving the slide carriage up and down once, using the two control keys.
Press and hold the corresponding key during the up and down
movement. The end points are defined after completion of the
initialization.
Directory tree Path C:\MFT\Software must be created. The subordinate paths are
created automatically.
The Word file "System_Configuration_**.doc contains information on
the modifications of the respective software version.
Light (green) paths must be created
Dark (blue) paths are created automatically
'@1*8
7SJX[EVI
:C<<<
7]WXIQC/SRJMKYVEXMSRCIRHSG
7]WXIQC/SRJMKYVEXMSRCHIHSG
&SSXC(MWOC>Q
&V ,SWX
%1&;4
FV <<<W
(V
&SSX
>Q0F
1JX\\\I\I &SSX0FF
>Q0F
1JX\\\I\I
>Q0F
1JX\\\I\I
>Q0F
1JX\\\I\I
7V
%1&;4
WV<<<W
>Q
4VGSQ
4VGSQ
8VERWJIV '31
8VERWJIV '31
>1FEX
>1FEX
>6'31I\I
Note
In case of an update of the display processor the preset system
parameters will be lost. The have to be saved via the Service Software
and then reinstalled on the display processor of the multiFiltrate.
After an update, the mDL parameters, the parameters for language,
filtrate bag monitoring and heparin pump activation and the settings for
Ci-Ca treatments must be checked and, if necessary, readjusted.
Preparation
The system is disconnected from external power supply (on the rear).
Open the rear wall of the system housing.
Connect the interface cable to the parallel port (LPT1) of the PC and
the other end to the dongle ICD cable. If necessary, use the cross
protection adapter.
Fit the 10-pin ribbon cable socket of the dongle onto the processor
board to be programmed.
Turn on the system using the power switch (on the rear).
Start the PROG16-68HC16 program on the PC.
Note
After having turned the multiFiltrate on, avoid waiting too long until
starting the program because, otherwise, the battery will be loaded for
an unnecessarily long period.
Select the file name (e.g. sr-x.xx.s2) of the new software in the BR
(OP) / SR (SP) directory, and confirm by pressing the ENTER key or
clicking the Öffnen (Open) button.
Note
Enter the "*.*" file name and display the files by clicking the Öffnen
(Open) button. Select the desired file using the cursor and apply it by
pressing the Enter key or clicking the Öffnen (Open) button.
Erasing the Flash Click the eraser symbol in the function bar, or call up EM Erase
module by double-clicking the pertinent line in the left-hand window.
The status window displays the process. Once the process is
completed, the message Module has been erased is emitted.
Programming the Flash Click the lightning symbol in the function bar, or call up PM
Program module by double-clicking the pertinent line in the left-
hand window.
This process is displayed in the status window. The addresses are
incremented $xxxxx. At the end of the program, Programmed is
displayed.
To exit the menu, click the closing symbol (X), or select FILE and
Exit (ALT-X) from the menu bar.
Completion
Turn off the system using the power switch on the rear.
Remove the 10-pin ribbon cable socket of the dongle of the printed
circuit board.
Although their software versions are having the same number (e. g.
1.51), they differ by the stored program code and by the saved data.
Note
Erase the flash as described above and then program it using the
"SR-x.xx.S2" software (SR = SP).
Check: Jumper J7 must be connected.
Turn off the system using the power switch.
Turn on the system using the power switch.
Press ESC and I/O to turn the system on (initialization).
Turn the system off after several seconds.
Turn the system on in the Service mode (I/O and Start).
Calibration of the blood leak detector.
Turn the system off after several seconds.
Erase the flash as described above and then program it using the
"BR-x.xx.S2" software (BR = OP).
Check: Jumper J7 must not be connected.
Turn off the system using the power switch.
Turn on the system using the power switch.
Press ESC and I/O to turn the system on (initialization).
Turn the system off after several seconds.
Turn the system on in the Service mode (I/O and Start).
Calibration of all pressures (TM 5.1.9)
Calibration and taring of all scales (TM 5.1.8)
The correct weight will be displayed only after having saved "5 kg".
Enter date, time, serial number and EC-code.
T1 test
selection of a new treatment mode (once).
correct identification on the printed circuit board (BR (OP) or SR
(SP)).
Note
In case of a failure of the Ci-Ca module, the entire module is replaced.
The module is not directly repaired on site.
Use the rotary selector to select Set up? [OK] to confirm! and
press [OK].
Use the rotary selector to select all messages during preparation
(status line is blue) up to the message Start connection? [OK] to
confirm! and confirm them by pressing [OK].
Establish an alarm-free condition of the venous pressure transducer
and have the optical detector sense opaque.
Use the rotary selector to select Start treatment? [OK] to confirm!
and press [OK].
The functional tests of the citrate and the calcium drip counter are
performed analogously.
The functional test of the citrate and the calcium pump is successfully
passed if the simulated treatment is completed without any alarm. If
both pumps perform several revolutions without alarm it can be
assumed that both pumps and their Hall sensors are in full working
order.
The functional tests of the insertion switches of the citrate and the
calcium pump are performed analogously.
The functional tests of the line occlusion of the citrate and the calcium
pump are performed analogously. The test is performed while the
system is turned off.
Caution
After each transmission of data from the service program to the dialysis
system, the operator of the service program must check the data at the
dialysis system for plausibility. The operator himself or herself is
responsible for the correctness of the data.
Caution
After each transmission of data from the service program to the dialysis
system, the dialysis system must be switched off and on again before a
patient treatment is started.
Caution
This service program is only intended for service purposes. Ensure that
the interface cable to the PC is not connected during patient treatment.
6.3 Preparation
Caution
When installing the software an older PC Service Software 4008
version already installed will automatically be uninstalled.
To retain an existing version of the service software, start the installation
with "setup.exe /noremove". In this case, select an installation path
differing from the path of the previous installation to avoid overwriting.
Caution
For Windows NT, 2000, or XP, you must have administrators rights to
install the PC Service Software MFT.
Connect the PC and the Multifiltrate via the supplied interface cable.
Caution
This is not a standard cable. It is therefore imperative to use the
enclosed interface cable.
Caution
Install the SmartCardReader drivers before connecting the
SmartCardReader to the PC!
These will be installed when the PC Service Software Multifiltrate is
installed.
1 2 3 4
+24V
+12V
+5V
Ethernet (4-Wire) mDL
U+
4kV isoliert
U+ LP244
RJ45 Ethernet (4-Wire) RxD LP123
H DGND +5V H
PGND TxD +24V TS BPTS +12V
+5V RS BPRS +24V
RS232-ISO (PC)Teil der LP124 (Motherboard)
TxD CLK BPCLK
LP244
3 RxD-PC RxD STOP BPSTOP
TxD RxDC-1 RxDD-2 DGND
2 TxD-PC PGND
Blut-Pumpe
RxD TxDC-1 TxDD-2 LP123
5 GND-PC
+24V +24V TS DFTS
+5V +5V RS DFRS
Display-Rechner
+12V TxD CLK DFCLK
TxD RxD-DR RxD STOP DFSTOP
RxD TxD-DR DGND
PGND
Dialysat-Pumpe
+5V
+12V LP123
DGND
+24V TS SUBTS
G CTS +5V RS SUBRS G
DSR RxD-Pump TxD CLK SUBCLK
CD TxD-Pump RxD STOP SUBSTOP
DGND
PGND
SUB-Pumpe
Backlight-Converter Schutz-Rechner
LP123
+12V
DGND Betriebs-Rechner +24V TS FTS
Dis-Dim +5V RS FRS
Dis-Off TxD CLK FCLK
RxD STOP FSTOP
DGND
Drehencoder LP129 (Teil) PGND
Filtrat-Pumpe
1 DGND
2 LP122
3 SW CD
4 Out-B DSR +24V +5VRef-SR +5VRef-SR
5 Out-A CTS +5V
F 6 +5V +5VRef-BR +5VRef-BR S-Temp2 S-Temp2 F
Dis-Off S-Temp1 S-Temp1 S-Temp4 S-Temp4
Dis-Dim S-Temp3 S-Temp3 S-RelSens S-RelSens
I/O LED-Dim LED-Dim SCK-BR SCK-BR S-Rel S-Rel
I/O
DGND Dis-Dim Dis-Dim MOSI-BR MOSI-BR
PCS3-BR PCS3-BR
S-PWM S-PWM
Start
Start/Reset SUB-Plasma
DGND S-HT S-HT
Heizung
S-HT-120 S-HT-120
+5V
Stop +12V
Stop
DGND -12V D-Temp1 D-Temp1 D-Temp2 D-Temp2
DGND D-Temp3 D-Temp3 D-Temp4 D-Temp4
D-PWM D-PWM D-RelSens D-RelSens
IO
Ton-Aus I/O IO D-Rel D-Rel
Ton-Aus Start
DGND Start Start D-HT D-HT
Stop
Stop Stop D-HT-120 D-HT-120 LP129 (Teil)
Ton-Aus
Ton-Aus Ton-Aus
ESC
ESC ESC ESC OK ESC DGND Dialysat-Heizung
E DGND OK OK AGND LED-A Alarm E
PGND
ODSTAT ODSTAT
LP343-1
Druck arteriell LP450
+12V
Part Part +24V Luftdetektor LED- W Warnung
T-Part T-Part +12V Pegel Heben /
AGND LDA1/LDA2 LDA1 Senken
LDSA LDSA Hell-Dunkel
LP343-1 LDACLR LDACLR Druck venös
ODSA ODSA
Filtrat-Druck
+12V Pven Pven VKLZU VKLZU LED-IO IO
PFil PFil T-Pven T-Pven LDA2 LDA1/LDA2
T-PFil T-PFil LDA1 LDA1
AGND LVUPSW LVUPSW OD OD
LVDWNSW LVDWNSW VLVUP LVUP
AGND VLVDWN LVDWN
LP343-1
Druck PGND
D +12V D
Prä-Hämofilter
PPHF PPHF NOTHALT NOTHALT
T-PPHF T-PPHF LP125
AGND
+12V MISO-SR MISO-SR
MOSI-SR MOSI-SR
LP127 SCK-SR SCK-SR
1 +5VRef +12V TA1 TA1 PCS3-SR PCS3-SR
2 DMS+ -12V LED-gn LED-gn +5V
3 DMS- +5V LED-rd LED-rd
4 AGND F-Scl f-Scl1
IO
F-Syn F-SYN IO
Blutleck- Start
F-Tk f-SclTk1 AGND Start PersonalRuf
Waage-1 Detektor Stop
AGND Stop
Ton-Aus Teil der LP124
DGND Teil der LP128 Ton-Aus
ESC
ESC
LP127 Akku Lade- OK
+24V OK
und Test-
1 +5VRef +12V AkkuSens AkkuSens
Schaltung
2 DMS+ -12V Akku-Test Akku-Test
C 3 DMS- +5V Akku-Rel Akku-Rel Akku-In C
4 AGND F-Scl f-Scl2 DGND
F-Syn F-SYN PB-Akku 18V
F-Tk f-SclTk2
AGND Sense+24V
DGND Sense+12V
Sense+5V
Waage-2
+5VRef-BR Sense+5VREF-BR
1 Sense+5VREF-SR +5VRef-SR
1K
2 SD-Ball RxD-BR TxD-SR
1K
Prüfgewichte Hubmagnete
TxD-BR RxD-SR
1
Teil der LP124
2 F-Ball +5V +5VRef-SR
NOTHALT
LP127 Sense-Overvoltage
1 +5VRef +12V Heparinpumpe
B Teil der LP124 (Motherboard) B
2 DMS+ -12V
3 DMS- +5V Ton-BR Ton-BR Ton-SR Ton-SR HP-DOOR Key-up
4 AGND F-Scl f-Scl3
F-Syn F-SYN Vol-Control Vol-Control Test-Alarm Test-Alarm HP-CLK Key-down
F-Tk f-SclTk3
Waage-3
AGND
DGND Sense -12V Mikrofon SPEAKER 8Ohm/1W
LP127
1 +5VRef +12V PGND PGND
2 DMS+ -12V AGND AGND +5V
3 DMS- +5V DGND DGND +24V
4 AGND F-Scl f-Scl4
F-Syn F-SYN -12V DGND
F-Tk f-SclTk4 PGND
AGND
DGND TxDB-2 RxD
RxDB-2 TxD
Waage-4
A A
-12V Tit el Revision
PGND
multiFiltrate
Blockschaltbild I/O C
AGND
Datum 16-Apr-2004
DGND
Datei C:\PRJ\MFT\HW\BLOCK-IO\RE VI SI~1\BLK_IO_3.SCH Blatt 1 von 1
1 2 3 4
7.2 AC Wiring
7)879& 7834
'328630
79&
28'79&&%+
(
7-
79& 79&
+2(
7-
(-%0
(-%0
(
7)8(-%0
'328630
(-%0
28'(-%0&%+
7834
7.3.2 Description
6
6
5
7
4
7.4.1 Description
Pump coding
2 6
10
7.5.1 Description
10 Scale I 29 mDl
11 Scales II 30 mDl
13 Scales IV 32 n.c.
16 Display processor
6
32
27 33
26
17 7
20
21
19
9
22
D 14
15
23 14
24
SI 1
31 13
18
12
16
11
29
10
28 34
25 30
7.7.1 Description
The EEPROM used for storage is provided with a synchronous serial interface (SPI interface). By means of the
jumper J1 in position 1-2, the EEPROM can be protected from being overwritten inadvertently.
: 7-
6IKYPEXSV :
: 7-
6IKYPEXSV : HMK
7-
7- : ERE
: 7-
6IKYPEXSV : HMK
7-
: ERE
: 7-
6IKYPEXSV :
&EXX 7-
'LEVKI 9&EXX
7.8.2 Description
Fuse arrangement
3 SI 8 (5 A) 12 SI 3 (100 mA / -12 V)
11
10
12
9
8
13
7
14
6
5 15
4 16
3
17
2
18
1
4'
(MKMXEP (MKMXEP
1SXSVSPE
-RTYXW 3YXTYXW
1','>
%REPSK
%REPSK
-RTYXW
3YXTYXW
68' 7IVMEP
-RXIVJEGI
Note
Jumpers are intended for developing purposes only. Do not make any
changes!
2 GND
2 GND
Operating processor
(P.C.B. LP 1102 )
MFT Citrate
24V DC Ci-Ca motherboard drop counter
RS 232 (P.C.B. LP MS 0407)
Citrate
pump
Signal and activation Voltage supply
Calcium
Voltage supply drop counter
Signal processing and
distribution Calcium
pump