DPI260
DPI260
Safety
WARNING
CAUTION:
Identify the operating Voltage before connecting
the instrument.
AC powered instruments have black 3-core cables
that must be connected:
Brown - Live
Blue - Neutral
Green/Yellow - Earth
WARNING
Electrical safety
To comply with electrical safety standards, the label
showing the operating Voltage MUST be changed
when the Voltage setting is changed.
An internal 500 mA 20 mm fuse is fitted in the live
circuit.
Fuse replacement
Only replace the internal fuse (a.c. or d.c.) with a
ceramic HBC, F type fuse.
(OPTIONAL)
SCREEN
+ SIGNAL
- SIGNAL
+ SUPPLY
- SUPPLY
SENSE
0V
1 2 3 4 5 6
CONNECTOR
GREENPAR
RM12 BPG-6P
Transducer
TRANSMITTER
100Ω INTERFACE
PCB
0va
⊗ ⊗ ⊗
1 2 3 4 5 6
SCREEN
CONNECTOR
GREENPAR
BLUE
RED
RM12 BPG-6P
Transmitter
5 Calibration
Routine calibration adjustments should be made at
periods of 6 months, this period may be changed
depending on the frequency of use. Use an accurate
pressure standard to apply pressure and, if necessary,
adjust the instrument. Allow a warm-up time of
approximately 1 hour.
5.1 Zero
The zero adjustment is made using the 20-turn
potentiometer accessed through the front panel. For
gauge instruments make sure no pressure is applied.
For absolute instruments, apply a vacuum to the
transducer before adjustment. To prevent
unauthorised adjustments fit a sealing label over the
potentiometer access on the front panel.
5.2 Span
The span adjustment for the full-scale pressure value
is made using the 20-turn span potentiometer accessed
through the front panel. Connect the pressure standard
to the instrument and adjust to the full-scale value. To
prevent unauthorised adjustments fit a sealing label
over the potentiometer access on the front panel.
PRESS TO
PRESS TO
CLOSE
CLOSE
12V DC - DC
OPEN
OPEN
5 6 7 8
1 2 3 4 5 6 7 8
N
E
EARTH/
LK2 GROUND
8
1 2 3 4
DECIMAL
POINT
SELECT
-
SWD NO OPTION IC1
LINK LK1 +
R2
SET SET REPLACED
SPAN ZERO BY R2
ZERO SWS SWZ FOR DPI 262
RV6 R1
R4
RV5
RV4
LK7 LK6 LK3
6.4 Linearity
RV4 controls feedback between the amplified
transducer signal and the bridge excitation supply and
affects the correction for square law non-linearity.
Removing LK5 stops linearization feedback.
The following procedure only applies to transducers
where zero pressure equates to zero nominal output
from the transducer.
Note: For other applications refer to Druck.
300nS MIN
X
X
DATA
20mS CLOCK
EXPANDED
4µS
approx
1.0mS
approx
OUTPUT 1 2 3 4 5
HUNDREDS
1
THOUSANDS
BCD8 1
UNITS
TENS
BCD4 1 1
0=+ 0=+
BCD2 0=- 0=- ETC
BCD1 MSD
DIGIT 1 0 1 0 0 0 0
DIGIT 2 0 0 1 0 0 0
DIGIT 3 0 0 0 1 0 0
DIGIT 4 0 0 0 0 1 0
1
0 VOLT LOGIC 14 HOLD
2
POLARITY 15 DATA VALID
3
BCD1 x 100 16 BCD2 x 100
4
BCD4 x 100 17 BCD8 x 100
5
BCD1 x 10
1 18 BCD2 x 101
6
BCD4 x 101 19
BCD8 x 101
7
BCD1 x 10
2 20 2
BCD2 x 10
8
21
BCD4 x 102 BCD8 x 102
9
BCD1 x 103 22 BCD2 x 103
10
BCD4 x 10
3 23 BCD8 x 10
3
11
MSD1 x 104 24 MSD2 x 104
12
OVER-RANGE 25 0V ANALOGUE/
13 4 - 20 mA (POWER)
ANALOGUE
SIGNAL 4 - 20 mA
(SINK)
17
DATA 1 DATA 2 DATA 2 DATA 3 DATA 4
DATA
>300nS
HOLD
FREE RUNNING HOLD TRIGGER/HOLD
EACH 400mS Dv = AS HOLD CHANGED TO ‘0’ HOLD = ‘1’ ALLOWS UPDATE
DATA UPDATED DATA UPDATED AT END OF 400mS AT NEXT END OF
Dv PULSES TO‘0’ NO FURTHER UPDATE UNTIL CONVERSION
HOLD = ‘1’
8 FAULT FINDING
In the event of a malfunction instruments can be
returned to the Druck Service Department or Druck
agent for rectification.
The following fault finding procedures in Table 1 must
be carried out by qualified* personnel using good
engineering practise. Make sure power supply is
disconnected from the instrument before accessing
internal components.
* see the safety page (i)
FAULT ACTION/CAUSE
No display 1. CHECK power supply switched on.
2. CHECK power supply fuse.
3. CHECK power cable.
4. CHECK internal 500mA fuse.
5. REPLACE power supply module.
6. REPLACE display PCB failure.