Chapter I. Introduction
RayScan handheld laser 3D scanner is handheld design, using multiple laser line
grid projected on the surface of the object, through the high-speed, high-
resolution industrial camera for data acquisition, through the 3D scanning software
to calculate the object's three-dimensional point cloud, to obtain the object's three-
dimensional data. Scanning is free and flexible, the whole process is completed
handheld, without the need for a tripod or other support devices. Adopt motion
viewpoint, convenient for users to observe the real-time scanning situation. The
with a beautiful and generous appearance, and the overall equipment is easy to
carry.
PAGE \* MERGEFORMAT 2
The product mainly has the following advantages:
1) Scanning is fast;
2) Non-contact measurement;
of frame points;
4) Equipped with a global error control module, which can globally control the
PAGE \* MERGEFORMAT 2
display the results of the camera calibration, can be intuitively access to the
PAGE \* MERGEFORMAT 2
1.2 System components
components:
acquisition;
PAGE \* MERGEFORMAT 2
1.3 Description of equipment
camcorde
r
lasers
camcorde
r
keypad
USB 3.0
Power interface
connector
Button Description:
4. Confirmation key "O": Click to confirm the command or start the scanning.
PAGE \* MERGEFORMAT 2
Connection sequence description:
1、 Plug the USB3.0 cable and power cord into the corresponding jacks on the
probe.
2、 Then plug the USB3.0 cable into the USB3.0 port of your computer and
PAGE \* MERGEFORMAT 2
1.4 Warnings
For your convenience, please read the instruction manual carefully and follow
the steps in the manual to prevent dangerous situations and injury to the operator
and equipment. Please keep this manual in a safe place in case you want to refer to
it again. Our company provides upgrading service for you from the beginning to
the end, no matter there is any problem when you use it, please contact us
according to the service telephone number and address on the last page of the
The scanner specifies that the power adapter operates on 220V, 50HZ, so
make sure that the power supply meets the requirements of the machine
before use.
Do not place the scanner in direct sunlight, near heating devices or heat
radiation devices.
Keep the scanner in a dry area away from dust and moisture.
Do not hit the scanner measuring head section with hard objects.
PAGE \* MERGEFORMAT 2
Do not touch the surface of the calibration block with hard objects or
chemicals.
Do not look directly at the laser emitter while the machine is in use to avoid
eye damage.
Use the power cord carefully and avoid excessive bending. A broken power
To ensure proper operation of the program, use only the software that came
with it.
PAGE \* MERGEFORMAT 2
Chapter II. System installation
This system runs on Microsoft Win7 or Microsoft Win10 operating system. The
hardware configuration requirements of the system for the computer are: CPU main
frequency I7, discrete graphics card memory 4 G or more, memory 16G or more;
Open the unit's box and gently remove the probe. The system is supplied with a
set of cables, connect one of the USB cables with the blue port to the end of the
probe and the other to the USB3.0 port on your computer. Finally connect the
power cable and press the switch at the handle, if the red light glows then the head
Never plug or unplug the power cord or turn off the power while the
PAGE \* MERGEFORMAT 2
machine is in use.
Use caution when moving the scanner and do not roll or damage the
power cord.
Remove the device dongle from the box and plug it into the USB port of your
Do not plug or unplug the dongle while in use as this may cause
Do not leave the dongle unattended after use as it may be lost and
Do not bend or damage the dongle as this may render the device
inoperable.
PAGE \* MERGEFORMAT 2
2.3 Software installation
Remove the random CD-ROM or USB flash drive and insert it into your
1. Run the Setup program on the CD-ROM or USB flash drive as shown below:
Click on "Next":
PAGE \* MERGEFORMAT 2
To install the scanning system, select either Anyone or Individual, usually select
Click on "Next":
PAGE \* MERGEFORMAT 2
Click "Next", the system automatically installs the program.
PAGE \* MERGEFORMAT 2
Installation is complete, click "Close".
PAGE \* MERGEFORMAT 2
Chapter 3 Software Operation
Before starting the software, you need to insert the dongle into the computer
and power up the hardware device and connect it to the computer. Double-click the
shortcut icon on the desktop of the computer, if it is the first time you start the
scanner program after installation, the dialog box shown in the figure will pop up.
Select the appropriate option according to the device model, if the computer
supports GPU acceleration, select "Start GPU hardware acceleration", click "OK" to
PAGE \* MERGEFORMAT 2
enter the startup interface interface, as shown in the following figure.
The software interface is mainly divided into main menu, system menu panel,
PAGE \* MERGEFORMAT 2
3.1.1 Main Menu
Click on the round icon in the upper right corner to enter the main menu
Import Frame Points (R): imports frame points from an external file to be used
Export Frame (C): saves the current frame point as an external file;
PAGE \* MERGEFORMAT 2
Remove Frame Point (Y): removes the current frame point;
Export Point Cloud (D): Save the point cloud data or mesh data acquired by 3D
Export Circular Hole (E): Saves the circular hole data acquired by 3D scanning
as an external file;
dialog box.
PAGE \* MERGEFORMAT 2
System Calibration Options: You can select "Enter [Scan Calibration] mode first
frame before scanning and real-time positioning based on the global frame during
scanning;
the device;
The next time you start the software, the Hardware Device Configuration
Prompt checkbox will pop up, and when you restart the software, the Laser
Scanning System Hardware Device Configuration dialog box will pop up.
PAGE \* MERGEFORMAT 2
System calibration: Enable "scanning calibration" mode and "laser calibration"
mode, in this mode, the user can accurately calibrate the system
measurement parameters.
Shooting frame: Users can capture the frame marking points on the surface of the
object;
Brightness Detection: Detects the brightness of the image, marking the bright
Marker Point Detection: Detects the marker point and identifies the center of the
Laser Line Detection: Detects the laser line and identifies the laser line position
gain factor, laser brightness, fill-in light brightness, etc.) to enable the
Connecting the device: connect the scanner's internal circuit controller, click the
"Connect Device" button and hear a beep, indicating that the device is
PAGE \* MERGEFORMAT 2
connected properly, and the device indicator will be purple after the
connection is normal.
2、 Accessibility
Background Plane: creates or removes a background plane and does not capture
area, the point cloud data acquired by users from multiple scans are fused
into a single point cloud with a unified coordinate system through global
registration;
Mesh processing: mesh the point cloud obtained from 3D scanning. With "mesh
streamlining".
PAGE \* MERGEFORMAT 2
3、 3D operation
Reverse Selection: Unselects the currently selected portion and selects the
unselected portion.
Clear Selection: deselects the currently selected portion of the selected state;
Undo Delete: Undo the last delete operation and restore the deleted part.
4、 3D view
Window reset: makes the contents of the view fit the window display;
PAGE \* MERGEFORMAT 2
dimensional model, including "front view, back view, left view, right view, top view,
top view".
Penetration Selection: When set to Penetration Selection Mode, it will have the
ability to select the occluded portion, otherwise only the visible portion
can be selected
5、 window pane
The Workspace Panel check box shows or hides the workspace panel;
The Left Camera Panel checkbox shows or hides the left camera panel;
The Right Camera Panel checkbox shows or hides the Right Camera Panel.
PAGE \* MERGEFORMAT 2
3.1.3 Workspace panel
1、 Display Settings
2、 model color
3、 grid element
PAGE \* MERGEFORMAT 2
Vertex: displays the triangle mesh vertices
4、 Other settings
Show world coordinate system: Show or hide the world coordinate system;
PAGE \* MERGEFORMAT 2
Show Model Wrap: Shows or hides the model wrap;
world coordinate system, global frame points, background plane and other
information.
PAGE \* MERGEFORMAT 2
3.1.5 Camera Panel
Displays information about the left camera panel and the right camera panel,
displays the camera frame rate, the camera's normal acquisition frame rate is 60 HZ,
the exposure time of the left and right cameras is generally set to "10", and the
PAGE \* MERGEFORMAT 2
The position of the camera panel can be adjusted by dragging the drag bar of
the camera panel, and the size of the camera panel can be adjusted by dragging
any of the four corners of the camera panel. Dragging the drag bar of the Camera
Panel without releasing it brings up the direction indicator button, which allows you
PAGE \* MERGEFORMAT 2
to place the Camera Panel in the corresponding position, as shown in the following
figure.
camera menu
Setting up the camera : Open or close the Camera Settings tool, where you can
PAGE \* MERGEFORMAT 2
Crosshairs : Shows or hides the image center crosshairs;
1. Open the equipment box and gently remove the probe. The system is
equipped with a set of data lines, one of the blue port USB line connected to the
end of the probe, the other end of the computer connected to the USB3.0 port.
Finally connect the power cable and press the switch at the handle, if the red light
glows, it means that the probe is energized normally, the device encryption lock is
system hardware device configuration" dialog box, please set the appropriate
PAGE \* MERGEFORMAT 2
parameters, determine to enter the main program.
parts, the first time to start the program preferred to enter the "camera calibration"
mode, and then start the program, select the "system calibration" will enter the
"laser calibration" mode. The first time you start the program, you will enter the
"camera calibration" mode, then start the program again and select "system
calibration", you will enter the "laser calibration" mode. When you start the
program again for calibration, you can click the round icon in the upper right
corner to enter the main menu, and select "System Calibration Options" in "System
You can select "Camera Calibration" and "Laser Calibration" modes by selecting
Click the "System Calibration" button to enter the "Camera Calibration" mode.
PAGE \* MERGEFORMAT 2
Click on the "Shoot" button to enter the system calibration.
holding the scanner, the two cameras on the scanner are facing the calibration
board, so that the scanner's two camera lines and the marking points on the middle
vertical column line up, the rear camera is raised, the front camera is lowered, and
the two camera lines and the calibration board are at a certain angle, so that the
center of the guide dots become green. Then according to the left and right sides
and above the guide bar prompts to adjust the camera front and rear position, near
and far position and left and right position, for the first position calibration, and
PAGE \* MERGEFORMAT 2
then adjust the scanner's near and far position, to complete the angle of the five
After completing the above calibration, the scanner's two camera lines and the
marking point lines on the center vertical column are aligned, the rear camera is
lowered, the front camera is raised, and the lines of the two cameras and the
calibration board are at an angle so that the center guide dot turns green. Then
according to the left and right sides and above the guide bar prompts to adjust the
camera front and rear position, near and far position and left and right position, for
PAGE \* MERGEFORMAT 2
the 6th position calibration, and then adjust the scanner near and far position, to
After completing the above calibration, the scanner's two camera lines are
aligned with the marking point line on the center vertical column, and the scanner
is rotated around the marking point line on the center vertical column, with the
camera facing to the right, so that the center guiding dot turns green. Then
according to the left and right sides and above the guide bar prompts to adjust the
camera front and rear position, near and far position and left and right position, for
PAGE \* MERGEFORMAT 2
the 11th position calibration, and then adjust the scanner's near and far position, to
After completing the above calibration, the two camera lines of the scanner are
aligned with the marking point line on the center vertical column, and the scanner
is rotated around the marking point line on the center vertical column, with the
camera facing to the left, so that the center guiding dot turns green. Then
according to the left and right sides and above the guide bar prompts to adjust the
camera front and rear position, near and far position and left and right position, for
PAGE \* MERGEFORMAT 2
the 16th position calibration, and then adjust the scanner's near and far position, to
After completing the above calibration, the scanner's two camera links are
aligned with the marking point links on the center vertical column, and the
scanner's two cameras are pointed squarely at the calibration plate so that the
center guide dot turns green. Then according to the left and right sides and above
the guide bar prompts to adjust the camera front and rear position, near and far
position and left and right position, for the 21st position calibration, and then
PAGE \* MERGEFORMAT 2
adjust the scanner's near and far position, to complete this angle of the five
calibration.
After finishing the above position shooting, click the "Calibration" button to
PAGE \* MERGEFORMAT 2
Click the "Next" button to enter the "Laser Calibration" mode. Follow the prompts
PAGE \* MERGEFORMAT 2
The two cameras of the scanner are aligned with the line of the marking point
on the middle vertical column, and the two cameras of the scanner are facing the
calibration plate, so that the center guiding dot turns green. Then adjust the front
and back position, far and near position, and left and right position of the cameras
according to the guide bar prompts on the left and right sides and above to
perform the laser calibration, and then adjust the far and near position of the
PAGE \* MERGEFORMAT 2
After finishing the above position shooting, click the "Calibration" button to
After the laser calibration is finished, the calibration result is obtained. The figure
PAGE \* MERGEFORMAT 2
below:
PAGE \* MERGEFORMAT 2
4.3 3D Scanning Operation
marking points, reflective marking points are generally affixed to the plane or the
curvature of the position of the change is not large, the spacing of the marking
points is about 10cm, to ensure that each time the measuring head to identify the
2、 scanning operation
Scanning is initiated by clicking the button circle button on the handle for one-
touch operation. Scanning requires the operator to scan the object back and forth
until the scan is complete. After scanning is complete, click the circle button on the
Scanning process need to keep the measuring head and the object has the
PAGE \* MERGEFORMAT 2
scanning head of the indicator light shows green, the measuring head is far from
the measured object shows yellow and red, the measuring head is close to the
3、 Preservation Engineering
During or after scanning, select "Save Project (S)" or "Save Project (A)" in the main
menu to save.
During the scanning process the 3D scanning system will prompt to save the
4、 Data export
After scanning the data, the results can be exported to other 3D design
PAGE \* MERGEFORMAT 2
4.3.2 Frame 3D scanning operations
marking points, reflective marking points are generally affixed to the plane or the
curvature of the position of the change is not large, the spacing of the marking
points is about 10cm, to ensure that each time the measuring head to identify the
Select the "Shoot Frame" command from the system menu and shoot the light
release marking point that has been attached to the object to be measured through
PAGE \* MERGEFORMAT 2
Take all frame points, stop frame point acquisition, perform frame optimization,
3、 scanning operation
Scanning is initiated by clicking the button circle button on the handle for one-
touch operation. Scanning requires the operator to scan the object back and forth
until the scan is complete. After scanning is complete, click the circle button on the
PAGE \* MERGEFORMAT 2
handle again to stop scanning.
Scanning process need to keep the measuring head and the object has the
scanning head of the indicator light shows green, the measuring head is far from
the measured object shows yellow and red, the measuring head is close to the
4、 Preservation Engineering
During or after scanning, select "Save Project (S)" or "Save Project (A)" in the
During the scanning process the 3D scanning system will prompt to save the
5、 Data export
PAGE \* MERGEFORMAT 2
After scanning the data, the results can be exported to other 3D design
In the point cloud display window, hold down the left mouse button to select
the point cloud, hold down the middle mouse button to rotate the point cloud, and
slide the mouse wheel or keyboard up and down keys to zoom in and out of the
point cloud.
Provides several ways such as rectangle selection, folding line selection, lasso
Tool", drag the mouse to draw a rectangular box to select the point
cloud.
PAGE \* MERGEFORMAT 2
2) Folding line tool: In the 3D view menu, click "Select Tools" → "Folding
Line Tool", click the screen to select a closed polygonal area point cloud.
3) Lasso Tool: In the 3D view menu, click "Selection Tools" → "Lasso Tool",
click the mouse to select a closed area to wrap the point cloud.
"¡ú" Brush Selection "¡ú", hold down the mouse and drag, cover the
point to be deleted and then release, you can select the point cloud.
2. Point cloud operation: in the three-dimensional operation menu, you can choose
to "show the selected", "show all", "reverse selection", "delete the selected "Undo
PAGE \* MERGEFORMAT 2
Chapter V. Maintenance
Here are some of the daily maintenance and maintenance of the equipment,
performance of the equipment to use, while extending the service life of the
equipment.
When using the computer, please do not connect to the Internet, so as not to
you must access the Internet, please update the virus database of the antivirus
computer;
The use of industrial cameras should be strictly adhered to the contents of their
When the scanner is not in use, put it away and store it in a dry place;
When the calibration block is not in use, it should be held gently to avoid
PAGE \* MERGEFORMAT 2
bumping and cutting.
PAGE \* MERGEFORMAT 2
Chapter 6 Troubleshooting
1、 Problem: Opening the program prompts that the dongle cannot be found.
Solution: 1. Check whether the dongle is inserted in the correct USB interface
position and whether it is loose; 2. Contact the manufacturer to check whether the
Solution: The splicing accuracy exceeds the error allowable value (if the accuracy
requirement is not high, the allowable error of the splicing marking points can be
changed in the system setting), and there are fewer than 4 markings on a single
side or fewer than 4 markings on a common part. Post more logo points, and make
surface of the object. 2. The gain is too small or the luminosity of the camera is too
PAGE \* MERGEFORMAT 2
small when calibrating. The gain should be increased or the brightness should be
Solution: The appearance of laser line identification marking the number of points
is not enough, it may be the calibration board or equipment lens with dust, this
Solution: 1. Change the brightness of the laser; 2. Change the brightness of the fill-
in lamp; 3. Check whether the scanner or calibration plate has a clean surface
without impurities.
PAGE \* MERGEFORMAT 2