s12 Vet
s12 Vet
Revision History
This manual has a revision number. This revision number changes whenever the
manual is updated due to software or technical specification change. Contents of this
manual are subject to change without prior notice.
Document No.: J/S-I-019
Revision number:V1.0
Release time: 2022.04
Copyright © 2022 Guangdong Biolight Meditech Co., Ltd. All rights reserved.
Statement
Guangdong Biolight Meditech Co., Ltd. (hereinafter called Biolight) holds the
copyright of this manual, and we are also entitled to deal with this manual as confidential
files. This manual is only used for operation, maintenance and service of product,
someone else can not publish the manual.
This manual contains exclusive information protected by copyright laws and we reserve
its copyright. Without written approval of manufacturer no parts of this manual shall be
photocopied, Xeroxed or translated into other languages.
The contents contained in this manual are subject to amendments without notification.
I
Manufacturer's Responsibility
Only under the following circumstances will manufacturer be responsible for the safety,
reliability and performance of the instrument:
All the installation, expansion, readjustment, renovation or repairs are conducted by
the personnel certified by manufacturer.
All the replaced components, assorted accessories and consumables involved in
maintenance are original or approved by Biolight.
The storage condition, operation condition and electrical status of the instrument
conform to the product specification. The electrical installation of the relevant room
complies with the applicable national and local requirements;
The instrument is used in accordance with the user’s manual.
NOTE: Provides application tips or other useful information to ensure that you get
the most from your product.
II
In order to protect your rights, please fill in the warranty card after the installation of the
equipment, and give the second warranty card ("Biolight company retention") to the
installer or mail it back to the after-service department of Biolight.
Please note that the following conditions will not be covered by the warranty:
The customer does not fill in or return the equipment warranty card within 30
days after the installation acceptance;
The equipment series number provided by the customer is not correct (our
company confirms the equipment series number is warranty or not).
Within the warranty period, all products can enjoy free after-sales service; however,
please note that, even within the warranty period, due to the following reasons, the
products need to be repaired, the company will carry out maintenance service, you need to
pay maintenance fees and accessories fees:
Man-made damage;
Improper use;
The voltage of the power network exceeds the product’s specified range;
Irresistible natural disasters;
Replace or use parts, accessories, consumables that are not approved by
Biolight or maintained by non-authorized personnel of Biolight;
Other faults not caused by the product itself.
After the expiration of the warranty, Biolight can continue to provide charged
maintenance services. If you do not pay or delay in paying the maintenance fee, Biolight
will suspend the maintenance service until you pay for it.
After service
Manufacturer: Guangdong Biolight Meditech Co., Ltd.
Address: No.2 Innovation First Road, Technical Innovation
Coast, Hi-tech Zone, Zhuhai, P.R. China
Fax: +86-756-3399919
Post code: 519085
Toll-free consultation hot line: +86-400-8818-233
III
About this manual
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a prerequisite
for proper product performance and correct operation and ensures patient and operator
safety.
This manual is based on the maximum configuration and therefore some contents may
not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the
equipment so that it can be obtained conveniently when needed.
The manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practiced and terminology as required for monitoring
patients.
All illustrations in this manual serve as examples only. They may not necessarily reflect
the setup or data displayed on your product.
Conventions:
Bold Italic text is used in this manual to quote the referenced chapter or sections.
【】is used to enclose screen texts.
→ is used to indicate operational procedures.
IV
Content
VI
Chapter 11 Temperature(TEMP) ........................................................................ 11-1
11.1. Introduction ..................................................................................................... 11-1
11.2. Displaying the TEMP Parameter Area ............................................................ 11-1
11.3. TEMP Display ................................................................................................. 11-1
11.4. Preparing for TEMP Monitoring ..................................................................... 11-2
11.5. TEMP Settings ................................................................................................ 11-2
11.6. TEMP Troubleshooting ................................................................................... 11-3
Chapter 12 NIBP ......................................................................................................... 12-1
12.1. Introduction ..................................................................................................... 12-1
12.2. Safety Information .......................................................................................... 12-1
12.3. NIBP Measurement Limitations ...................................................................... 12-3
12.4. Measurement Modes ....................................................................................... 12-3
12.5. NIBP Display .................................................................................................. 12-4
12.6. Preparing for NIBP Measurements ................................................................. 12-5
12.7. Starting and Stopping NIBP Measurements .................................................... 12-7
12.8. NIBP Settings .................................................................................................. 12-8
12.9. Assisting Venous Puncture ............................................................................ 12-10
12.10. NIBP Maintenance ...................................................................................... 12-11
12.11. NIBP Troubleshooting ................................................................................ 12-11
Chapter 13 IBP............................................................................................................ 13-1
13.1. Introduction ..................................................................................................... 13-1
13.2. Safety information ........................................................................................... 13-1
13.3. IBP measurement ............................................................................................ 13-2
13.4. IBP Display ..................................................................................................... 13-3
13.5. Setting IBP ...................................................................................................... 13-5
13.6. IBP troubleshooting......................................................................................... 13-9
Chapter 14 Cardiac output (C.O.) ............................................................................. 14-1
14.1. Introduction ..................................................................................................... 14-1
14.2. Safety Information .......................................................................................... 14-1
14.3. C.O. Measurement Limitation ......................................................................... 14-1
14.4. C.O. Display .................................................................................................... 14-2
14.5. C.O. Equipment Connecting ........................................................................... 14-3
14.6. C.O. Measurement........................................................................................... 14-3
14.7. C.O. Setting ..................................................................................................... 14-6
14.8. C.O. Troubleshooting ...................................................................................... 14-7
Chapter 15 Carbon Dioxide (CO2) ............................................................................ 15-1
15.1. Introduction ..................................................................................................... 15-1
15.2. Safety information ........................................................................................... 15-1
15.3. CO2 Measurement Limitations ....................................................................... 15-2
15.4. Monitoring Procedure ..................................................................................... 15-2
15.5. Display ............................................................................................................ 15-7
15.6. Settings CO2 .................................................................................................... 15-8
15.7. Entering Intubation Status ............................................................................. 15-11
15.8. Zeroing .......................................................................................................... 15-11
VII
15.9. Calibration ..................................................................................................... 15-12
15.10. Exhaust Emission ........................................................................................ 15-13
15.11. Announcements ........................................................................................... 15-13
15.12. CO2 Troubleshooting................................................................................... 15-15
Chapter 16 Drip Monitor (DM) ................................................................................. 16-1
16.1. Introduction ..................................................................................................... 16-1
16.2. Safety information ........................................................................................... 16-1
16.3. DM measurement ............................................................................................ 16-2
16.4. DM module indicator ...................................................................................... 16-5
16.5. DM display ...................................................................................................... 16-5
16.6. Setting DM ...................................................................................................... 16-6
16.7. DM check ........................................................................................................ 16-6
Chapter 17 Review ...................................................................................................... 17-1
17.1. Review Overview ............................................................................................ 17-1
17.2. Reviewing Page ............................................................................................... 17-1
Chapter 18 Calculations ............................................................................................. 18-1
18.1. Introduction ..................................................................................................... 18-1
18.2. Safety information ........................................................................................... 18-1
18.3. Drug Calculation ............................................................................................. 18-1
18.4. Hemodynamic Calculations ............................................................................ 18-4
18.5. Oxygenation Calculation ................................................................................. 18-6
18.6. Ventilation Calculations .................................................................................. 18-9
18.7. Nephridium Calculation ................................................................................ 18-11
Chapter 19 Recording................................................................................................. 19-1
19.1. Recorder .......................................................................................................... 19-1
19.2. Recording Type ............................................................................................... 19-1
19.3. Starting Recordings ......................................................................................... 19-2
19.4. Stopping Recordings ....................................................................................... 19-2
19.5. Recording Flags .............................................................................................. 19-2
19.6. Setting the Recorder ........................................................................................ 19-3
19.7. Installing Recording Paper .............................................................................. 19-4
19.8. Clearing Jam Paper.......................................................................................... 19-5
19.9. Cleaning Recorder ........................................................................................... 19-5
Chapter 20 Other Functions ...................................................................................... 20-1
20.1. Analog Signal Output ...................................................................................... 20-1
20.2. Network Settings ............................................................................................. 20-1
20.3. HL7 Settings ................................................................................................... 20-4
Chapter 21 Battery ..................................................................................................... 21-1
21.1. Introduction ..................................................................................................... 21-1
21.2. Installing a Battery .......................................................................................... 21-1
21.3. Battery Guidelines ........................................................................................... 21-1
21.4. Battery Maintenance ....................................................................................... 21-2
21.5. Battery Recycling ............................................................................................ 21-4
VIII
Chapter 22 Maintenance and Cleaning .................................................................... 22-1
22.1. Introduction ..................................................................................................... 22-1
22.2. Cleaning of the Monitor .................................................................................. 22-2
22.3. Disinfection ..................................................................................................... 22-2
22.4. Cleaning and Sterilizing of Accessories .......................................................... 22-3
22.5. Sterilization ..................................................................................................... 22-4
Chapter 23 Maintenance ............................................................................................ 23-1
23.1. Inspection ........................................................................................................ 23-1
23.2. Maintenance Schedule..................................................................................... 23-2
23.3. Disposing of the Monitor ................................................................................ 23-3
Chapter 24 Accessories ............................................................................................... 24-1
24.1. Recommended Accessories ............................................................................. 24-2
Appendix A Product Specifications .................................................................................... 1
A.1 Safety Specifications ............................................................................................... 1
A.2 Environmental Specifications .................................................................................. 2
A.3 Physical Specifications ............................................................................................ 2
A.4 Power Specifications ............................................................................................... 3
A.5 Hardware Specifications.......................................................................................... 4
A.6 Data Storage ............................................................................................................ 7
A.7 Wireless network ..................................................................................................... 7
A.8 Measurement Specifications .................................................................................... 8
A.9 Alarm Specification............................................................................................... 25
Appendix B EMC and Radio Regulatory Compliance ..................................................... 28
B.1 EMC ...................................................................................................................... 28
B.2 The management compliance of Radio.................................................................. 34
Appendix C Default Settings ............................................................................................ 35
C.1 ECG, Arrhythmia, ST, QT..................................................................................... 35
C.2 RESP ..................................................................................................................... 38
C.3 SpO2 ...................................................................................................................... 38
C.4 TEMP .................................................................................................................... 39
C.5 NIBP ...................................................................................................................... 40
C.6 IBP......................................................................................................................... 41
C.7 C.O. ....................................................................................................................... 43
C.8 CO2 ....................................................................................................................... 43
C.9 DM ........................................................................................................................ 44
C.10 Alarm Default Settings ........................................................................................ 44
C.11 Screen Setup ........................................................................................................ 45
C.12 Color of parameters ............................................................................................. 45
C.13 Recorder .............................................................................................................. 45
C.14 Other .................................................................................................................... 46
C.15 Maintenance Item ................................................................................................ 46
Appendix D Alarm Message ............................................................................................. 48
IX
D.1 Physiological alarm ............................................................................................... 48
D.2 Technical Alarm .................................................................................................... 50
Appendix E Cybersecurity ............................................................................................... 60
E.1 Operating environment .......................................................................................... 60
E.2 Network data interface ........................................................................................... 60
E.3 User access control mechanism ............................................................................. 60
E.4 Software Environment ........................................................................................... 61
Appendix F Terminology and Definitions ........................................................................ 62
F.1 List of units ............................................................................................................ 62
F.2 Symbol list ............................................................................................................. 63
F.3 Terminology list ..................................................................................................... 64
Appendix G Toxic and Harmful Substances or elements ................................................. 68
X
S12 Vet Monitor User’s Manual
The Veterinary monitor, hereafter called the monitor, are intended to be used for
monitoring, displaying, reviewing, storing and alarming of multiple physiological
parameters of animals, including ECG, Heart Rate(HR), Respiration Rate(RR),
Temperature(TEMP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive
Blood Pressure(NIBP), Carbon dioxide(CO2), Invasive Blood Pressure(IBP) and Cardiac
output(C.O.).
WARNING:
The monitor is intended for use only by clinical professionals or under their
guidance. It must only be used by persons who have received adequate training
in its use. Anyone unauthorized or untrained must not perform any operations
on it.
4
5
1-1
S12 Vet Monitor User’s Manual
1. Display screen
2. Light induction: For example, when the environment is dark, the display brightness
can be automatically adjusted.
3. Alarm lamp
Alarm lamp with different color and flashing frequency indicates the level of
technical alarm and physiological alarm:
High level alarm: the lamp quickly flashes red.
Medium level alarm: the lamp slowly flashes yellow.
Low level alarm: the lamp lights cyan without flashing.
4. Power indicating lamp
It is a LED that lights green and orange, the status of the LED is specified as follows:
Green: When the AC mains is connected.
Orange: When the AC mains is not connected and monitor is powered by
battery.
Off: When the AC mains is not connected.
5. Charging indicating lamp
Light up: When the battery is being charged.
Off: When the battery is fully charged or no battery in monitor.
6. Phonetic holes.
1-2
S12 Vet Monitor User’s Manual
10 3
4
9 8
1. Power button
Power on: after the monitor is connected to the AC power supply, press the button to
open the monitor.
Power off: in the startup state, short or long press the power switch button to turn off
the monitor
2. Recorder
3. DM connector
4. TEMP connector
5. SpO2 connector
6. ECG connector
1-3
S12 Vet Monitor User’s Manual
7. NIBP connector
8. CO2 connector
9. C.O. connector
10. IBP connector
3
4
5 6 7
1. VGA connector(optional)
Connect secondary display with a standard VGA connector. Auxiliary display and
monitoring are performed by connecting a secondary display. The display content of the
secondary display is consistent with the monitor display.
2. Multifunction connector(optional)
The standard RJ45 interface, enabling networking with the central monitoring system,
1-4
S12 Vet Monitor User’s Manual
When other devices and monitors are unified used, the wires should be used to
connect the equipotential grounding terminal of other devices and monitors to eliminate
the potential difference between different devices and ensure safety
6. AC power socket
7. Cable retainer
1 3 2 5
4
7
6
1-5
S12 Vet Monitor User’s Manual
Select this area to enter the【Patient Management】menu and detailed description please
to refer Chapter 4 Patient Management.
2. Current configuration
3. Technical alarm area
Display technical alarm information and prompt information. Cyclic displaying when
there are multiple messages. Select this area to open the 【Alarm Informations】menu to
view the current technical alarm.
4. Physiological alarm area
Shows the physiological alarm messages, medium-level and low-level alarm
messages display on the left, while the high-level alarm messages display on the right.
Cyclic displaying when there are multiple messages. Select this area to open the 【Alarm
Informations】menu to view the current physiological alarm.
5. System status information area
Display alarm volume, network and storage devices connection status, battery and
system time. For the battery status icon please to refer chapter 21 Battery.
6. Waveform area
Show the waveforms of physiological parameter. Label displays on the top left
corner of each waveform area. Select the waveform area of a parameter and enter the
corresponding parameter setting menu.
7. Parameter area
It consists of various parameter areas, and shows parameter value, unit, alarm limit
and alarm status, etc. Label displays on the top left corner of each parameter area. Select
the parameter area of a parameter to enter the corresponding parameter setting menu.
8. Area of touch quick keys
Shows quick keys, these quick keys are used to conduct some common operations.
1-6
S12 Vet Monitor User’s Manual
The following table shows the symbols and meanings displayed in the system
information area:
No battery is installed.
Quick keys are displayed at the bottom of the monitor's main screen. Through the
quick keys, you can easily and quickly access some functions or perform operations.
1-7
S12 Vet Monitor User’s Manual
1-8
S12 Vet Monitor User’s Manual
You can set quick keys which need to be displayed on the interface, as follows:
1. Enter the 【Quick keys】setting menu in one of the following ways:
Select 【Screen Setup】 quick key → select 【Quick keys】 submenu.
1-9
S12 Vet Monitor User’s Manual
1.3.3. Menu
The styles of the various menus are basically similar, see the picture below:
5
2
3 6
4
1. Menu title:Summary of the current menu.
2. Submenu button:Press this button to enter the corresponding submenu.
3. Main display area of menu: Display menu options.
1-10
S12 Vet Monitor User’s Manual
1-11
S12 Vet Monitor User’s Manual
Chapter 2 Safety
2.1. Safety Information
WARNING:
The device is intended for animal monitoring by trained personnel in the
specified places.
The physiological data and alarm messages displayed on the monitor are for
reference only and cannot be directly used for diagnostic interpretation.
The monitor can only be used for a single patient at the same time
This monitor can only be connected to a power outlet that has a protective
ground. Do not use a removable multi-hole socket. If the power outlet is not
connected to a grounding conductor, do not use the outlet and use a
rechargeable battery to power the monitor.
Before use, you must check the equipment, cables and accessories to ensure
they work properly and safely.
To avoid explosion hazard, do not use the monitor in the presence of
flammable anesthetics or other flammable substance in combination with air,
oxygen-enriched environments, or nitrous oxide.
Do not open the shell of the device; otherwise the electric shock hazard may
exist. All maintenance and upgrades must be carried out by the personnel
trained and authorized by manufacturer only.
Do not use the monitor in nuclear magnetic resonance (MR) environments.
The operator cannot simultaneously touch the conductive parts on the patient
and the monitor.
Do not come into contact with the patient during defibrillation. Otherwise
serious injury or death could result.
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off may result in a hazard to the
patient. Remember that alarm settings should be customized according to
different patient situations and always keeping the patient under close
2-1
S12 Vet Monitor User’s Manual
CAUTION:
To ensure patient safety, use only parts and accessories specified in this
manual.
Magnetic and electrical fields are capable of interfering with the proper
performance of the monitor. For this reason make sure that all external devices
operated in the vicinity of the monitor comply with the relevant EMC
requirements. Mobile phone, X-ray equipment or MRI devices are a possible
source of interference as they may emit higher levels of electromagnetic
radiation.
To avoid contamination or infection of personnel, environment or other
equipments, the equipment and its accessories that meet the service life must be
disposed in accordance with relevant local regulations or hospital systems.
The lifetime of the monitor is 5 years.
Before turning on the device, make sure that the voltage and frequency of the
power supply meet the label of the device or the requirements specified in this
manual.
Please install or carry the equipment properly to prevent it from falling,
colliding, being subjected to strong vibration or other mechanical external
damage.
2-2
S12 Vet Monitor User’s Manual
The monitor must be wiped dry immediately after exposure to rain or splashes.
Please do not mix different types and brands of electrodes. Mixing the
electrodes may result in a large baseline drift or a long baseline recovery time
after defibrillation. It is forbidden to use dissimilar metal electrodes, which
may cause high polarization voltages.
NOTE:
Put the monitor in a location where you can easily see the screen and access the
operating controls.
This device uses a power plug to disconnect it from an AC power. Place the
device in a location where it is easy to plug in and out.
In normal use, the operator should stand in front of the device
Keep this manual near the device so that it can be easily and timely obtained
when needed.
This manual describes the product in the most complete configuration. This
manual describes all features and options. Your monitor may not have all of
them.
2-3
S12 Vet Monitor User’s Manual
Abbreviation of “Invasive
Equipotential grounding IBP
Blood Pressure”.
Abbreviation of “Drip
Auxiliary output DM
Monitor”.
Abbreviation of “Cardiac
Network C.O.
output”
Defibrillator synchronization
USB
output connector
Symbol for the marking of electrical and electronics devices according to Waste
Electrical and Electronic Equipment Directive.
Keep dry.
2-4
S12 Vet Monitor User’s Manual
WARNING:
The equipment should be installed by personnel designated by the
manufacturer.
The copyright of the software of this device belongs to the manufacturer. No
organization or individual can tamper with, copy or exchange any
infringement by any means or form without permission.
Equipment connected to the analog or digital interfaces must comply with the
respective IEC standards (e.g. IEC 60950 for data processing equipment and
IEC 60601-1 for medical equipment). Furthermore all configurations shall
comply with the current version of the standard for SYSTEMS IEC 60601-1.
Everybody who connects additional equipment to the signal input part or
signal output part configures a medical system, and is therefore responsible
that the system complies with current version of the requirements of the system
standard IEC 60601-1. If in doubt, consult the technical service department or
your local representative.
When the equipment is connected to other electrical equipment into a
combination with specific functions, if it is impossible to determine whether the
combination is dangerous (for example, the electric shock hazard caused by the
accumulation of leakage current), please contact the expert of the company or
the hospital to ensure that the necessary safety of all equipment will not be
damaged in the combination.
1. Unpacking
Before unpacking, examine the packing case carefully for signs of damage. If any
damage is detected, contact the carrier. If the packing case is intact, open the package.
2. Remove the monitor and accessories carefully.
3-1
S12 Vet Monitor User’s Manual
3. Keep all the packaging materials for future use in transportation or storage.
4. Check the monitor and accessories
Check the monitor and its accessories one by one in accordance with the packing list.
Check to see if the parts have any mechanical damages. In case of problems, please
contact us or our agent.
WARNING:
Keep the packing materials out of children’s reach. Disposal of the packing
materials should comply with local regulations or the hospital's waste disposal
system.
The monitor might be microbial contaminated during storage and transport.
Before use, please verify whether the packages, especially the package of
disposable accessories are intact. In case of any damage, do not apply it to the
patient.
The operating environment of the equipment must meet the specifications in this
manual.
The operating environment of the equipment should also reasonably free from the
noises, vibration, dust, corrosive or flammable, explosive substances. If it is installed in
the cabinet, make sure that there is enough space in front of the cabinet for operation,
maintenance and repair; In order to maintain ventilation, the equipment should be at least
2 inches (5cm) away from around the cabinet.
When the equipment is transferred from one place to another, condensation may
occur due to differences in temperature or humidity. At this point, you must wait for the
condensation to disappear before you can use the equipment.
WARNING:
Please ensure the monitor is working under specified conditions; otherwise, the
technical specifications mentioned in this manual will not be met, thus possibly
leading to damage of equipment and other unexpected results.
3-2
S12 Vet Monitor User’s Manual
WARNING:
Always use the accompanying power cord delivered with the monitor.
Use the battery if the integrity of the protective earth conductor or the
protective earthing system in the installation is in doubt.
Using battery
The monitor can be powered by a rechargeable lithium battery. After the battery is
installed, if the external power supply is suddenly interrupted, the monitor can
automatically use the lithium battery to supply the power. For the use of the battery,
please refer to Chapter 21 Battery.
3-3
S12 Vet Monitor User’s Manual
WARNING:
If the monitor is damaged or does not work properly, do not use it for any
monitoring procedure on a patient. And then please contact the maintenance
personnel or the manufacturer immediately.
CAUTION:
Although not recommended, you can press and hold the power on/off switch
for 5 seconds to forcibly shut down the monitor when it could not be shut down
normally or under some special situations. This may cause loss of data of the
monitor.
3-4
S12 Vet Monitor User’s Manual
NOTE:
The AC power supply of the monitor is not cuff off through the power switch.
To completely disconnect the power supply, unplug the power cord.
In case of a temporary power failure, the monitor retains patient data before
shutdown, including patient monitoring data and configuration data.
Everything you need to operate the monitor is on its screen. Almost every element on
the screen is interactive. Screen elements include parameter values, waveforms, quick
keys, information area, alarms area, menus, etc. Often you can access the same element in
different ways. For example, you can access a parameter setup menu by selecting
corresponding parameter area or waveform area.
CAUTION:
Check the touch screen whether it is damaged or breakage before use. If it is
found to be damaged or broken, please stop using the monitor immediately and
3-5
S12 Vet Monitor User’s Manual
The monitor supports the mouse with USB connector, the mouse can be plugged and
unplugged. You can use the mouse to select a screen element by moving the cursor on the
element and then click on it.
The monitor supports the barcode scanner to input patient’s medical record number
or registration number, and connects to the monitor through USB interface
Use the key to confirm what you have entered and close the onscreen keyboard.
This chapter only introduces the general settings. For the setting of parameters and
other functions, please refer to the corresponding chapter.
3-6
S12 Vet Monitor User’s Manual
Select【Brightness】quick key.
NOTE:
When the monitor enters standby mode, the screen brightness will be
automatically adjusted to the lowest.
When the AC power is interrupted and the battery is powered, the screen
brightness is automatically set to the corresponding brightness when the
battery is powered. You can still manually adjust the brightness as needed.
3-7
S12 Vet Monitor User’s Manual
If the monitor is connected to the central monitoring system, the system time of the
monitor will be automatically adjusted according to the time of the central monitoring
system.
CAUTION:
When starting to use the monitor, please modify the date and time of the device
according to the local time. Incorrect date and time setting may lead to
misjudgment of patient trend data.
You can manually turn the parameter switch on or off. The steps to set the parameter
switch are as follows:
3-8
S12 Vet Monitor User’s Manual
You can set the parameter waveform and its order displayed in the normal interface
as required. The steps are as follows:
1. Enter【Screen Layout】interface in one of the following ways:
Select【Screen Setup】quick key→select【Screen Layout】submenu.
Select【Main Menu】quick key→from【Display】column to select
【Screen Layout】.
2. Select a parameter area or waveform. From the pop-up parameter list to select
the elements needed to display in the area. The selected parameters and
waveforms are displayed according to the set position. The parameters and
waveforms that are not selected will not be displayed on the interface.
NOTE:The first line of the parameter area and the waveform area always
WARNING:
The parameters of the 【Screen Layout】 are not assigned to the display area,
which will not be displayed on the monitor interface and relevant alarms for
this parameter will still be provided
Each parameter has an independent setting menu, through which alarm and
parameter setting can be modified. You can enter the Parameter Setting Menu in the
following ways:
Select the parameter area or waveform area of a parameter
Select 【Main Menu】 quick key, from 【Parameter】 column to select
【Setup】,and then select corresponding parameter.
3-9
S12 Vet Monitor User’s Manual
Monitor mode is the most common clinical working mode used to monitor patient.
When the monitor is turned on, it automatically enters the monitor mode.
You can temperately stops patient monitoring without switching off the monitor by
entering the standby mode. You can enter the Standby Mode in the following ways:
Select either way showed in the following to enter the Standby Mode:
Press【Standby】quick key, or
Select【Main Menu】quick key→from the【Patient】column to select
【Standby】button, or
Select【Patient Management】quick key→discharging patient and the
enter Standby Mode.
WARNING:
Pay attention to the potential risk of placing the monitor to standby. In the
standby mode, the monitor stops all parameters’ measurements and disables
all the alarm indications, except for the battery low alarm.
3-10
S12 Vet Monitor User’s Manual
If the monitor automatically enters the standby mode after a patient is discharged,
choose any of the following ways to exit the standby mode:
Select【Patient Management】 to exit the standby mode and admit a new
patient.
Select【Monitor】to enter the patient information for preparing to admit a new
patient.
In Demo mode, the monitor can demonstrate its major functions when patient or
patient simulator is not connected. The Demo mode is password protected
Choose the following way to enter the Demo Mode:
1. Select【Main Menu】→from【System】column to select【Demo】.
2. Input the password→select【OK】.
Shut down and restart to exit the Demo Mode.
WARNING:
The demonstration function is mainly used to display machine performance
and to train users. In the actual clinical use, it is forbidden to use the
demonstration function, so as to prevent the medical staff from mistakenly
thinking that the monitor displays the waveform and parameters of the
monitored patient, thus affecting the patient’s monitoring and delaying the
diagnosis and treatment.
3-11
S12 Vet Monitor User’s Manual
Night mode is a special clinical monitoring mode. Under the night mode, the alarm
volume, QRS volume and screen brightness turn to be lowest automatically. To avoid
disturbing the patient, night mode may be used.
WARNING:
Verify the settings of brightness, alarm volume, QRS volume and key tone
before entering the night mode. Pay attention to the potential risk if the setting
value is low.
NOTE:The monitor resumes the previous settings after exiting the night mode.
3-12
S12 Vet Monitor User’s Manual
The privacy mode is a special monitoring mode. In the privacy mode, the monitor
does not display patient information and monitoring data to protect patient information
from non-clinicians such as visitors.
The privacy mode is only available when the patient admitted by the monitor is also
monitored by the Central monitoring system. The monitor continues monitoring the
patient, but patient data is only visible at the Central monitoring system.
The monitor has the following features after entering the privacy mode:
The screen turns blank, and prompt 【Being monitoring】at same time.
All parameters and waveforms display are shield.
Except for the low battery alarm, the monitor inactivates alarm tone and alarm
light of all other alarms.
The monitor suppresses all system prompt tone, including heart beat tone,
pulse tone, etc.
WARNING:
In Privacy mode, all audible alarms are suppressed and the alarm light is
deactivated at the monitor. Alarms are presented only at the Central
monitoring system. Pay attention to potential risk.
CAUTION:
Cannot enter the privacy mode if a low battery alarm occurs.
3-13
S12 Vet Monitor User’s Manual
The monitor automatically exits the privacy mode in any of the following situations:
The monitor disconnects from the Central Monitoring System.
The low battery alarm occurs.
You can also press 【Exit Privacy Mode】on the screen to manually exit the privacy
mode.
The monitor provides timer function, which can display up to four timers at the same
time. You can set each timer separately, which prompts when the set time arrives.
WARNING:
Do not use the timers to schedule critical patient-related tasks.
3-14
S12 Vet Monitor User’s Manual
You can set each timer independently. To set the timer, follow this procedure:
2. Set【Timer Type】:
: The timer cycles, that is, the timer counts according to the
【Cycled】
preset 【Run Time】, and starts timing again after reaching the run time. The
timer area displays the number of timer cycles.
: The timer displays the time elapsed since the timer was
【Unlimited】
started.
3. Set【Direction】
.
4. Set【Run Time】
.
5. Set【Reminder Volume】
. When the remaining time is 10 seconds, the monitor
issues a reminder tone and the timer flashes in red, prompting you that the run
time is to expire.
CAUTION:
Cannot change timer settings when a timer is running.
Set【Direction】, 【Run Time】and【Reminder Volume】only for【Normal】
or【Cycled】.
The voice assistant can be used as an auxiliary input interface. You can issue specific
control commands to the monitor via voice. After the monitor recognizes the command, it
performs the corresponding operation function.
The method of operating the voice assistant as following:
1. Select【Voice Assistant】quick key, open the voice assistant function, the icon
3-15
S12 Vet Monitor User’s Manual
2. Say the wake-up words. At this time, the top status bar will display the "
" icon, indicating that the monitor is in the state of voice recognition control
command.
3. When the control command is spoken, the monitor will recognize the
command and perform the corresponding operation, at this time, " " icon
CAUTION:
Every time before you say a control command, you must say the wake-up word
to wake up the voice assistant at first. After waking up(the top status bar
displays the " " icon), if the control command is not spoken within 7
seconds, the voice assistant will enter the sleep state again (top The status bar
displays the " " icon), you need to re-execute the “Wake-up” →
NOTE:
The specific wake-up words and supported voice commands will show in the
voice assistant screen.
3-16
S12 Vet Monitor User’s Manual
Before monitoring a new patient, discharge the previous patient. After the patient is
discharged, all patient data, including patient information, trend data, and physiological alarm
information is be deleted from the monitor, the technical alarms are reset, and monitor
settings returns to their defaults (current configuration or user-specified configuration). For
more information, see 5.2 Setting Default Configuration.
After discharging a patient, the monitor automatically admit a new patient.
WARNING:
Always discharge the previous patient before starting monitoring a new patient.
Failure to do so can lead to data being attributed to the wrong patient.
NOTE:
Discharging a patient deletes all history data of current patient in the monitor.
Select the patient information area at the top left of the screen→Select
【Discharge Patient】
.
Select【Main Menu】
, from 【Patient】column to select【Discharge Patient】
.
4-1
S12 Vet Monitor User’s Manual
Always inputs patient information as soon as the patient is admitted. For more
information please refer to 4.3.2 Editing Patient Information for details.
WARNING:
【Patient Type】 and 【Paced】 will always contain a value, regardless of
whether the patient is fully admitted or not. If you do not specify settings for
these fields, the monitor uses the default settings from the current
configuration, which might not be correct for your patient.
For paced patients, you must set 【Paced】 to 【Yes】. If it is incorrectly set to
【No】, the monitor could mistake a pace pulse for a QRS and fail to alarm
when the ECG signal is too weak.
For non-paced patients, you must set 【Paced】 to 【No】.
4-2
S12 Vet Monitor User’s Manual
Edit patient information after a patient has been admitted, or when patient
information is incomplete, or when you want to change patient information.
To edit patient information, follow this procedure:
1. Enter the 【Patient Management】 menu. For more information, please refer
to 4.3.1 Patient Management menu.
2. Select patient type according to the actual situation: 【Big Animal】, 【Small
Animal】.
3. Edit patient information as required.
If your monitor is connected with the barcode scanner, you can enter the medical
record number by scanning the barcode.
NOTE:
The setting of patient type determines the algorithm used by the monitor to
process and calculate certain measurements, as well as the safety limit and
alarm limit range applicable to certain measurements.
The monitor reloads the configuration when the patient type is changed.
You can set whether to display and edit patient room number, name, age, and so on
in the 【Patient Management】 menu by following these steps:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input password→Enter.
2. Select【Patient Management】submenu→【Field】submenu.
3. Select which the patient information needs to be displayed and edited in the
【Patient Management】 menu.
4. Select customizes the patient information section, if necessary, and enters the
name of the section.
4-3
S12 Vet Monitor User’s Manual
4-4
S12 Vet Monitor User’s Manual
WARNING:
The configuration management function is password protected. The
configuration management tasks must be performed by clinical professionals.
5-1
S12 Vet Monitor User’s Manual
3. Select【OK】
.
You can delete the saved user configurations by following these steps:
1. Select【Main Menu】quick key→from【Configuration】column to select
【Delete User Config】→input maintenance password→Enter.
2. Select the configuration you want to remove:
In the【Delete User Config】menu, the currently existing user configuration on
the monitor is displayed.
3. Select【Delete】button.
4. In the popup dialog box to select【OK】
.
CAUTION:
The current configuration in use cannot be deleted.
The monitor provides configuration transfer function. You can use a USB drive to
transfer the configuration from one monitor to another monitor that needs the same
settings without having to reset them item by item. The monitor supports transferring the
monitor configuration with USB disk.
You can export the monitor's current user configuration to a USB drive by following
these steps:
1. Connect the USB drive to the monitor’s USB port.
2. Select【Main Menu】quick key→from【Configuration】column to select
【Export User Config】→input maintenance password→Enter.
3. Select configuration that needs to be exported.
4. Select【OK】.
5-2
S12 Vet Monitor User’s Manual
You can import the user configuration to the monitor through a USB drive by
following these steps
1. Connect the USB drive to the monitor’s USB port.
2. Select【Main Menu】quick key→from【Configuration】column to select
【Import User Config】→input maintenance password→Enter.
3. Select configuration that needs to be imported.
4. Select【OK】.
You may make changes to some settings during operation. However, these changes
or the pre-selected configuration may not be appropriate for the newly admitted patient.
Therefore, the monitor allows you to load a desired configuration so as to ensure that all
the settings are appropriate for your patient.
When receiving patients, you can select loading the nearest configuration or the
specified configuration. You can set the default configuration by following these steps:
1. Select【Main Menu】quick key→from【Configuration】column to select
【New Patient Config】→input maintenance password→Enter.
2. Select【Default】or【Current】
.
【Default】: The monitor loads the default configuration specified by
the user when it receives the patient, please refer to 5.2 Selecting
Default Configuration.
【Current】: The monitor loads the nearest configuration when it
receives the patient.
5-3
S12 Vet Monitor User’s Manual
When the monitor starts, you can select whether the monitor loads the nearest
configuration or the specified configuration. You can set the default configuration by
following these steps:
1. Select【Main Menu】quick key→from【Configuration】column to select
【Boot Config】→input maintenance password→Enter.
2. Select【Default】or【Current】.
【Default】: The monitor loads the default configuration specified by
the user the user when it starts. Please refer to 5.2 Selecting Default
Configuration.
【Current】: The monitor loads the nearest configuration when it starts.
If you access the configuration management menu and use the password to access
alarm-related settings, you can set a valid time for the password, beyond which you will
need to re-enter the password.
Please follow the steps below:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Authorization】submenu.
3. Set【Retention Time】.
5-4
S12 Vet Monitor User’s Manual
You can select the parameters and waveforms you want to view in the 【Screen
Layout】 window. For details, please refer to 3.6.2 Setting Display Screen.
The conventional screen is the most commonly used clinical monitoring screen for
the monitor, and the monitor enters the normal screen after being turned on. You can also
select the screen type as needed, the steps are as follows:
1. Enter【Select Screen】interface in one of the following ways:
Select【Screen Setup】quick key→Select【Screen Select】submenu.
Select 【Main Menu】quick key→from【Display】column to select
【Screen Select】
.
2. Select screen types as needed.
The steps for setting colors of parameter values and waveforms are as follows:
6-1
S12 Vet Monitor User’s Manual
The Dynamic Trend window is located to the left of the waveform area, showing the
trend of a series of parameters in a recent period of time. You can enter the Dynamic Trend
screen in any of the following ways:
Select【Screen Setup】quick key→Select【Screen Select】submenu→Select
【Dynamic Trend】.
Select【Main Menu】quick key→from 【Display】column to select【Screen
Select】→Select【Dynamic Trend】.
In the Dynamic Trend window, the parameter label name of the trend is displayed
above each trend curve, and the trend scale is displayed on the left. Trend times are
displayed at the bottom of the window.
6-2
S12 Vet Monitor User’s Manual
You can enter the OxyCRG screen by any of the following methods:
The OxyCRG screen shows two trend curves and a compression waveform.
Follow the steps below to set the parameters of the OxyCRG screen:
1. Enter OxyCRG screen.
2. Select【Setup】.
3. Set【Trend 1】, 【Trend 2】, 【Compressed Wave】separately.
4. Select 【Scale】 submenu, set the scales of each parameter. If you want to use
the default scaleplate of the system, select the 【Default Scale】on the
OxyCRG screen.
You can review the 48-hour trend curve and compression waveform on the OxyCRG
Review Screen. Follow these steps to go to the OxyCRG Review Screen:
1. Enter OxyCRG screen.
6-3
S12 Vet Monitor User’s Manual
2. Select【Review】.
You can check the alarm status and real-time physiological data of patients on other
remote monitors in the LAN on the monitor. A remote monitor (such as a bedside monitor)
is also called other bed monitor. You can monitor the alarms of up to 16 other beds at the
same time, and you can also view the waveform of 1 other bed from the current monitor.
You can monitor the alarm of other bed through the alarm monitoring area of【Other
Bed View】 interface.
CAUTION:
Can view the monitor through the remote monitor. You can check the alarm
and waveform of this monitor from 5 remote monitors at the same time.
Through the 【Other Bed View】screen, you can check the real-time parameters and
waveforms of a remote monitor and monitor the alarm of other monitors.
6-4
S12 Vet Monitor User’s Manual
1. Other bed observation area: Display the patient information, alarm status,
information, waveform and parameters of the selected bed. You can move the
interface down to browse the contents of the interface.
2. Other bed monitoring area
Displays all remote monitors being monitored.
Display the equipment number of other bed monitor, and indicate other bed
monitor's alarm status with different background colors:
• Red: Indicates that other bed monitor is giving high priority
physiological or technical alarm, and the high priority alarm is the
highest level alarm in the current alarm of the bed.
• Yellow: Indicates that other bed monitor is giving medium priority
physiological or technical alarm. The medium priority alarm is the
highest level alarm in the current alarm of the bed.
• Cyan: Indicates that the monitor is giving low priority physiological or
technical alarm, and the low priority alarm is the highest level alarm in
the current alarm of the bed.
• Green: Indicates that the monitor is connected successfully and no
alarms have occurred.
• Gray: Indicates that the monitor is not connected successfully.
6-5
S12 Vet Monitor User’s Manual
Only add other bed monitor, the device can monitor alarm of other bed. If you have
added other bed monitor, you can up to a choice of 16 beds. Add other bed as follows:
1. Select other bed observation interface area, in the pop-up window 【Bed View
Settings (Bed number)】to select 【Bed】submenu.
2. Select the device number of the monitor to be observed in the list.
The setup interface mainly displays the device number, IP, and
patient information of the networked monitor.
Select【Show Offline Bed】to display the device numbers of all
monitors.
If you no longer need to monitor the remote monitor, you can remove it as follows:
1. Select other bed observation interface area, in the pop-up window【Bed View
Settings (Bed number)】 to select【Bed】submenu.
2. Cancel the device number of the monitor in the list. If you want to delete all
beds, you can select【Delete All】.
In the other bed monitoring area of its bed observation window, select a bed and
then other bed observation window will display the real-time monitoring interface of the
monitor. This selected bed is called the main bed.
You can follow these steps to view the current real-time alarm information of the
main bed:
1. Enter【Alarm】interface by one of the following methods:
Click on the alarm information display area of the bed observation area, and
the alarm interface will pop up.
Select other bed observation interface area, in the pop-up window【Bed View
6-6
S12 Vet Monitor User’s Manual
Other bed observation area can display up to 4 waveforms. Following these steps to
select the label name of the waveform you want to observe:
1. Select other bed observation interface area, in the pop-up window【Bed View
Settings (Bed number)】to select【Wave】submenu.
2. In proper order to select 【First Wave】, 【Second Wave】, 【Third Wave】
and 【Fourth Wave】, then select the label name of the waveform in the
pop-up list. If you select is 【Close】, then the display of one waveform will
be turned off.
Other bed observation can display all the online parameters. Select the parameter
label names you want to observe as follows:
1. Select other bed observation interface area, in the pop-up window【Bed View
Settings (Bed number)】 to select【Param】submenu.
2. Open the parameter labels you want to observe in the displayed list of online
parameters.
WARNING:
The data displayed will delay in other bed observation window. Don't rely on
other bed observation window for real-time data.
6-7
S12 Vet Monitor User’s Manual
You can enter big numerics screen in either of the following ways:
Select【Screen Setup】quick key→Select【Screen Select】submenu→Select
【Big Font Screen】.
Select【Main Menu】quick key→from【Display】column to select【Screen
Select】→Select【Big Font Screen】.
In the setting window of 【Big Font Screen】, you can select 6 parameters to observe
according to your needs. For parameters with waveform, one waveform is displayed at the
same time.
During monitoring the patient, you can freeze the waveform on the screen and then
review it to carefully observe the patient during this time. You can also export the frozen
waveform through the recorder.
2. All waveforms are frozen, that is, the waveforms are not refreshed. The data
in the parameter area is refreshed normally.
In the frozen status, you can select the control icon to browse the frozen
waveforms: the frozen waveform will move to the left or right correspondingly. At the
same time, each waveform is marked with a time scale, and the freezing time is recorded
as【0s】. As the waveform moves to the right, the time scale will be gradually changed to
6-8
S12 Vet Monitor User’s Manual
Under freezing condition, you can select button in the upper right corner of the
freezing menu to release the freezing condition.
Select the 【Record】 button in 【Freeze】, the recorder will output the waveform
selected and the parameter value at the Freezing time. The recorder can output up to 3
waveforms at one time. For the setting of frozen waveforms, please refer t to 19.6.1
Selecting the record waveforms.
6-9
S12 Vet Monitor User’s Manual
Chapter 7 Alarm
7.1. Introduction
This chapter introduces the alarm function and the settings of the monitor.
WARNING:
A potential hazard can exist if different alarm presets are used for the same or
similar equipment in any single area, e.g. an intensive care unit or cardiac
operating room.
Alarm Settings for different monitors in the same area may vary to suit the
condition of the patient being monitored. Before starting the patient
monitoring, check whether the alarm setting is suitable for the patient, and
always open certain necessary alarm limits, and ensure that the alarm limit
setting is suitable for the patient.
If your monitor is connected to the central monitoring system (CMS) or other
monitors, alarms can be displayed and controlled remotely. Remote suspension,
inhibition, or silence of monitor alarms via the CMS or other monitors may
cause a potential hazard. For details, refer to the operator’s manual of the
CMS and the other monitors.
Setting the alarm limit to the highest may invalidate the alarm system.
When the alarm sound is turned off, even if a new alarm is triggered, the
monitor will not emit an alarm sound. Therefore, the user must carefully select
whether to turn off the alarm sound. Check patient status frequently after
turning off alarm or alarm sound.
For patients who cannot be continuously treated by medical staff, the alarm
settings must be made according to the patient's condition.
Do not rely solely on an audible alarm system to monitor a patient. There may
be risks in adjusting the alarm sound to a lower volume, which may impede
operator recognition of alarm. The alarm volume should be large enough in the
7-1
S12 Vet Monitor User’s Manual
current monitoring environment and the actual clinical condition of the patient
should be paid close attention.
NOTE:
When the alarm system is powered off, the monitor will save the alarm
information before power interruption, and the stored alarm information will
not change.
The monitor has two different types of alarms: physiological alarms and technical
alarm.
Physiological alarms: Physiological alarms are triggered by a monitored
parameter value that violates set alarm limits or an abnormal patient
condition.
Technical alarms: Technical alarms are triggered by a device malfunction due
to improper operation or system problems. The problems may result in
system abnormal operation or irresponsible monitoring parameters.
Apart from the physiological and technical alarm messages, the monitor will also
display some information related to system status or patient status.
7-2
S12 Vet Monitor User’s Manual
current situation.
Prompt: Prompt patient and system status information.
When an alarm occurs, the monitor will indicate it to the user through visual or
audible alarm indications:
High priority Medium priority Low priority
Alarm signal Prompt Note
alarm alarm alarm
DO-DO-DO--D
Alarm Tone O-DO -----
ISO DO-DO-DO - DO- / /
Mode DO-DO-DO--
DO-DO -----
Displayed in the
top information
Black words, Black words, area, click on
White words, White
Alarm Information Yellow Cyan the alarm
Red background words
background background information to
view the alarm
information list.
Displayed in
Alarm level symbol *** ** * / front of alarm
information.
Red Yellow Cyan
Parameter alarm background, background, background, / /
flashes flashes flashes
WARNING:
When multiple alarms of different priorities occur simultaneously, the alarm
lamp and alarm tone are prompted according to the highest level of all current
7-3
S12 Vet Monitor User’s Manual
alarms.
When there are multiple alarms in the same area at the same time, the alarm
messages are displayed circularly on the screen.
You can set the alarm properties centrally. Select【Alarm Setup】quick key or select
the corresponding button from the【Alarm】column, in the main menu to set alarm.
The steps to centrally set the properties of the parameter alarm are as follows:
1. Enter【Limit】interface in any of the following ways:
Select 【Alarm Setup】 quick key.
7-4
S12 Vet Monitor User’s Manual
You can change the password protection mode of the alarm settings and arrhythmia
settings as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Authorization】submenu.
3. Change the password protection mode of the alarm settings.
【No Password】: Change alarm setups to not be password protected.
【Password】: Change alarm switch, alarm limit and alarm level to be
protected by password.
If you use password to access alarm and arrhythmia alarm related settings, you can
set the valid time of the password, beyond which you need to re-enter the password. For
details, please refer to 5.9 Setting Password Valid Time.
NOTE:
7-5
S12 Vet Monitor User’s Manual
When the alarm volume is set to 0, the alarm tone will be turned off, and an
alarm audio off icon will appear on the screen.
When the alarm volume is set to 0, the setting of high level alarm volume is
invalid.
The minimum alarm volume determines the minimum alarm volume setting. The
steps are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Sound】submenu.
3. Select【Minimum Alarm Volume】.
NOTE:
You can set the minimum alarm volume to 0 only when you are connected to
the CMS. If the monitor is not connected to the CMS, the minimum alarm
volume can only be set to 1.
When the CMS is connected, if the minimum alarm volume is set to 0, the
minimum alarm volume will be automatically changed to 2 when the CMS is
disconnected.
You can set the alarm tone interval. The steps are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
7-6
S12 Vet Monitor User’s Manual
When the alarm volume is 0, alarm reset or the alarm is off, the monitor can provide
periodic alarm prompt tone to remind you that there is still an activated alarm in the
current system. This function is turned on by default.
You can set the alarm tone as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】button→【Pause /Reset】submenu.
3. Set 【Alarm Pause duration】
. You can set 【Alarm Pause duration】to
【1min】, 【2min】, 【3min】or【Permanent】, the default is【2min】
.
4. Set【Alarm Off Reminder】switch.
【On】: The monitor provides an alarm tone according to the set interval.
【Off】: The monitor does not provide an alarm tone.
5. . You can set 【Reminder Interval】to 【1min】,
Set 【Reminder Interval】
【2min】, 【3min】, 【5min】or【10min】, the default is【5min】
.
The monitor provides an alarm tone enhancement function. If the alarm exceeds the
set time and is not confirmed, the alarm volume can be automatically enhanced.
The steps to set the alarm tone enhancement are as follows
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
7-7
S12 Vet Monitor User’s Manual
2. Select【Alarm】→【Sound】
.
3. Set【Auto Increase Volume】to【3 Steps】, 【2 Steps】, 【1 Steps】or
.
【Off】
【 3 Steps 】 : After the alarm occurs, the alarm volume will be
automatically increased to level 3, if the set time is not confirmed.
【 2 Steps 】: After the alarm occurs, the alarm volume will be
automatically increased to level 2, if the set time is not confirmed.
【 1 Steps 】: After the alarm occurs, the alarm volume will be
automatically increased to level 1, if the set time is not confirmed.
: After the alarm occurs, the set time is not confirmed, and the
【Off】
alarm volume remains unchanged.
4. Set【Increase Volume Delay】
, select sound enhanced delay time.
For the over-limit alarm of continuous measurement parameters, the alarm delay
time can be set. If the condition of triggering alarm disappears within the delay time, the
monitor will not alarm.
Set the delay time for the alarm by following these steps:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Other】submenu.
3. Set【Alarm Delay】
.
The delay time of the apnea alarm is not affected by the alarm delay time setting.
You can set the delay time of the apnea alarm separately.
7-8
S12 Vet Monitor User’s Manual
.
【Setup】
2. Select【Apnea Delay】to set apnea alarm delay time.
You can set the length of the waveform needs to be output when an alarm occurs, the
setting steps are as follows:
1. Enter【Setup】interface in either of the following ways:
You can set whether to alarm when the monitor and the CMS are disconnected. This
function is enabled by default. The setting method is as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Other】submenu.
3. Open or Close 【CMS Disconnected】.
When the 【CMS Disconnected】 switch is turned on, the technical alarm
will be generated when the monitor and the CMS are disconnected after
successful connection.
7-9
S12 Vet Monitor User’s Manual
WARNING:
Pausing or turning off the alarm may cause the patient to be in danger, please
handle it carefully.
7-10
S12 Vet Monitor User’s Manual
Click on【Alarm Reset】quick key to reset the alarm system, and the alarm reset icon
will appear in the system status information area.
NOTE:
In the alarm reset state, if a new alarm is generated, the alarm reset icon
disappears and the alarm system is reactivated.
After the physiological alarm is reset, the sound of the currently existing
physiological alarm is shielded, and the other alarm states remain unchanged.
7-11
S12 Vet Monitor User’s Manual
information area.
You can choose to individually lock the visual signal or simultaneously lock
the visual and audible signals.
For visual latching, after the alarm condition disappears, the visual signal of
the alarm, including the alarm light, the alarm information and the
background color remain unchanged, and the alarm information text is
followed by the time of last triggering the alarm.
For audible latching, the system still emits an alarm tone after the alarm
condition disappears.
CAUTION:
Changing of alarm priority may affect the latching status of corresponding
alarm. Please determine if you need to reset the latching status for the specific
alarm when you have changed its alarm priority.
When the alarm system is reset, the latched physiological alarms are cleared.
Do not set all alarm status to latching alarm signals when used in the intensive
care unit.
The nurse call function means that when the alarm set by the user occurs, the monitor
7-12
S12 Vet Monitor User’s Manual
can output a signal to the nurse call system, call the nurse. The monitor provides a nurse
call connector, and the monitor is connected to the nurse call system of the hospital
through the randomly provided nurse call cable. After the system is connected, the
connector can implement the nurse call function.
The nurse call function must be valid only if the following conditions are met:
The nurse call function is turned on.
A user-defined alarm occurs.
The monitor is not alarm paused or off.
To set the type and priority of alarms that are sent to the nurse call system, follow this
procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Nurse Call】submenu.
3. Select【Signal Type】to set the type of nurse call signal.
【Pulse】: The nurse call signal is a pulse signal and each pulse lasts one
second. When multiple alarms simultaneously occur, only one pulse
signal is outputted. If an alarm occurs but the previous one is not cleared,
a new pulse signal will also be outputted.
【Continuous】: The nurse call signal lasts until the alarm ends. That is
to say the duration of a nurse call signal is equal to that of the alarm
condition.
4. Select 【Trigger Type】to set the work mode of the nurse call relay.
5. Select 【Alarm Priority】to set the priority of alarms sent to the nurse call
system.
6. Select 【Alarm Type】to set the type of alarms sent to the nurse call system.
WARNING:
Do not rely exclusively on the nurse call system for alarm notification.
7-13
S12 Vet Monitor User’s Manual
Remember that the most reliable alarm notification combines audible and
visual alarm indications with the patient’s clinical condition.
The monitor provides the intubation status function during RESP and CO2
monitoring. In this state, the physiological alarms related to RESP and CO2 are shielded,
and the alarm off icon is displayed in the parameter area. During the intubation process of
general anesthesia surgery, the intubation status can be selected to shield unnecessary
alarms.
The default intubation time is 2 minutes. To change the time, following this
procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Other】submenu.
3. Set【Intubation Duration】.
7-14
S12 Vet Monitor User’s Manual
Enter other bed observation interface, and when the monitored bed monitor has an
alarm triggered, the alarm light and alarm sound are prompted according to the highest
level of all alarms of the current monitor and other bed monitor. You can view and
manage other bed alarm.
The alarm delay time from the device to other bed is≤2s.
You can reset other bed alarm on the monitor. The steps to enable it to reset the bed
alarm are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Remote View】submenu.
3. Open【Reset Remote Bed’s Alarms】.
And then 【Bed View Settings (bed number)】 window→ 【Reset Remote Alarm】
button in the 【Alarm】 submenu will be activated. Click on 【Reset Remote Alarm】
button, other bed alarm will be reset.
CAUTION:
Only when the “Alarm Reset By Other Bed” function of the remote monitor is
enabled, can you reset other bed alarm on this monitor
Alarms on your monitor can be reset by remote devices if you enable this function.
To do so, follow this procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
7-15
S12 Vet Monitor User’s Manual
The monitor can provide an alarm if remote devices are disconnected. By default, the
function is enabled. To disable the alarm, follow this procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Remote View】submenu.
3. Switch off【Remote Disconnected Alarm】.
The monitor automatically performs a self-test at startup. Check that the alarm lamp
illuminates, one after the other, in red, yellow, and cyan, and that an alarm tone is heard.
This indicates that the visible and audible alarm indicators functions correctly.
To further test individual measurement alarms, perform measurements on yourself or
using a simulator. Adjust alarm limits and check that appropriate alarm behavior is
observed
When an alarm occurs, please refer to the following steps to take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Check if the alarm is eliminated.
For more information, please refer to D Alarm Information.
7-16
S12 Vet Monitor User’s Manual
Chapter 8 ECG
8.1. Introduction
The electrocardiogram (ECG) measures the electrical activity of the heart and
displays it on the monitor as waveforms and parameters. The monitor provides 3-lead,
5-lead, 6-lead, and 12-lead ECG monitoring, arrhythmia analysis, ST segment analysis
and QT/QTc measurements.
WARNING:
This equipment is not intended for direct cardiac application.
Make sure the conductive parts of electrodes and associated connectors for
applied parts, including the neutral electrode, should not contact any other
conductive parts including earth.
Use defibrillation-proof ECG cables during defibrillation.
Do not touch the patient or metal devices connected to the patient during
defibrillation.
To reduce the hazard of burns during high-frequency surgical procedure,
ensure that the monitor’s cables and transducers never come into contact with
the electrosurgery unit (ESU).
To reduce the hazard of burns during use of high-frequency surgical unit
(ESU), the ECG electrodes should not be located between the surgical site and
the ESU return electrode.
CAUTION:
Only use parts and accessories specified in this manual. Follow the instructions
for use and adhere to all warnings and cautions.
Regularly inspect the electrode application site to ensure skin quality. If there
are signals of allergies, replace the electrodes or change the application site.
8-1
S12 Vet Monitor User’s Manual
The following figures show the ECG waveform and parameter areas. Your display
may be configured to look slightly different.
Waveform Display
1 2 3 4 5 6 7
8
(1). ECG lead label
(2). ECG waveform gain
(3). ECG filter mode
(4). ECG waveform speed
(5). Paced status: If【Paced】is set to 【Yes】, is displayed; If【Paced】is set
to【No】, is displayed.
(6). Notch frequency
(7). Alarm message: Display only the highest level of alarm information.
(8). Pace pulse mark: If 【Paced】 is set to 【Yes】, the pace pulse markers are
shown on each ECG waveform when the patient has a paced signal.
8-2
S12 Vet Monitor User’s Manual
Parameter Display
4
1
2
NOTE:
The ECG parameter area and waveform area are configured to be different for
different lead type and ECG settings.
Proper skin preparation is necessary for good signal quality at the electrode sites, as the
skin is a poor conductor of electricity. Sites where leads are attached to the body must be
properly prepared to optimize contact.
Dogs and cats have enough electrolyte material on their skin and hair so that merely
moistening lead sites with 70% isopropyl alcohol is appropriate. This will usually be
sufficient for ECG monitoring for a short time, 30 to 60 minutes, depending upon the
relative humidity.
For monitoring during longer periods, an electrode paste should be used. It is best to
first wet the hair at the lead attachment site with alcohol; then place paste on the moistened
8-3
S12 Vet Monitor User’s Manual
hair and skin. It is important that the paste be in direct contact with skin. For animals with
dense undercoat, rub paste with fingers to assure that it has made contact with skin.
The following table lists the 5-lead labels and colors for AHA and IEC standards:
IEC AHA
Lead
Label Color Label Color
Right arm R Red RA White
Left arm L Yellow LA Black
Right leg (neutral) N/RF Black RL Green
Left leg F Green LL Red
Chest C White V Brown
The following table lists the 6-lead labels and colors for AHA and IEC standards:
IEC AHA
Lead
Label Color Label Color
Right arm R Red RA White
Left arm L Yellow LA Black
Right leg (neutral) N/RF Black RL Green
Left leg F Green LL Red
Chest 1 Ca White Va Brown
Chest 2 Cb White Vb Brown
8-4
S12 Vet Monitor User’s Manual
The following table lists the 12-lead labels and colors for AHA and IEC standards:
IEC AHA
Lead
Label Color Label Color
Right arm R Red RA White
Left arm L Yellow LA Black
Right leg (neutral) N/RF Black RL Green
Left leg F Green LL Red
Chest1 C1 White/ Red V1 Brown/ Red
Chest 2 C2 White/Yellow V2 Brown/Yellow
Chest 3 C3 White/Green V3 Brown/Green
Chest 4 C4 White/Brown V4 Brown/ Blue
Chest 5 C5 White/Black V5 Brown/ Orange
Chest 6 C6 White/ Purple V6 Brown/ Purple
Taking the AHA standard as an example, the 3-lead electrode placement position is
as shown:
8-5
S12 Vet Monitor User’s Manual
Taking the AHA standard as an example, the 5-lead electrode placement position is
as shown:
NOTE:
The exploring lead is used for diagnostic purposes as needed. Otherwise, it may
be left unplugged.
For 6-lead placement, you can use the position for the 5-lead placement, but with two
chest leads. The two chest leads (Va and Vb) can be placed according to the physician’s
preference.
8-6
S12 Vet Monitor User’s Manual
12-lead ECG uses 10 electrodes, which are placed on the patient’s four limbs and
chest. The limb electrodes should be placed on the soft skin and the chest electrodes
placed according to the physician’s preference.
While placing electrodes for a surgical patient, the type of surgery should be
considered, for instance, as to a chest surgery, the chest lead electrodes can be placed at
sides or backside of chest. Moreover, while using a surgical electrotome, in order to
reduce the influence of artifacts to ECG waveform, the electrodes can be placed at left and
right shoulders, close to left and right sides of abdomen; the chest lead electrodes can be
placed at left side of chest midst. Do not place the electrodes on the upper arm. Otherwise,
the ECG waveform will be very small.
WARNING:
To reduce the hazard of burns during use of electrosurgical units (ESU), the
ECG electrodes should not be located between the surgical site and the ESU
negative electrode plate.
Never entangle the ESU cable and the ECG cable together.
When using ESU, never place ECG electrodes near to the negative electrode
plate of the ESU, as this can cause a lot of interference on the ECG signal.
8-7
S12 Vet Monitor User’s Manual
It is important to correctly set the paced status before you start monitoring ECG. The
paced symbol is displayed when【Paced】is set to【Yes】. The pace pulse markers
"│" are shown on each ECG waveform when the patient has a paced signal. If【Paced】is
set to【No】or the patient’s paced status is not selected, the symbol will be shown in
the ECG waveform area.
To change the paced status, follow this procedure:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Paced】submenu.
3. Set 【Paced】to be【Yes】or【No】. If you did not set the paced status, the
monitor issues a prompt tone when pace pulse is detected. At the same time,
the paced symbol flashes and the message prompt "Suspected Pacing
Signal". Check and set the patient’s paced status.
WARNING:
For paced patients, you must set【Paced】 to【Yes】. If it is incorrectly set to
【No】, the monitor could mistake a pace pulse for a QRS complex and fail to
alarm when the ECG signal is too weak.
False low heart rate or false asystole alarms may result with certain
pacemakers because of pacemaker artifacts, such as electrical overshoot of the
pacemaker overlapping the true QRS complexes.
Do not rely entirely on rate meter alarms when monitoring patients with
pacemakers. Always keep these patients under close surveillance.
For non-paced patients, you must set 【Paced】to 【No】.
The pace pulse rejection function is disabled by default. To enable this function,
follow this procedure:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
8-8
S12 Vet Monitor User’s Manual
2. Select【Paced】submenu.
3. Switch on【Pacer Reject】.
NOTE:
When pace pulses are detected, the pace pulse marks “|” are shown on the
ECG waveforms. 【Pacer Reject】 setting has no impact on the display of pace
pulse marks “|”.
You can switch on 【Pacer Reject】 only when 【Paced】 is set to 【Yes】.
8-9
S12 Vet Monitor User’s Manual
You can set the label name of the ECG calculation lead as follows:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Setup】submenu.
3. Select【ECG 1】or【ECG 2】 to set label name of ECG calculating Lead.
WARNING:
Only when you switch on 【MultiLead Ana】can you set 【ECG 2】.
CAUTION:
【ECG 1】is the key calculation lead; 【ECG 2】is the auxiliary calculation lead.
Only when the ECG 【Lead Type】 is 5/6/12 lead can you set【MultiLead Ana】.
8-10
S12 Vet Monitor User’s Manual
If the ECG waveform is too small or clipped, you can change its amplitude by
selecting an appropriate gain setting. To do so, follow this procedure:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Gain】submenu.
3. Set the size of each ECG waveform. If you select【Auto】, the monitor
automatically adjusts the gain of the ECG waveforms.
8-11
S12 Vet Monitor User’s Manual
NOTE:
The notch filter can eliminate power frequency interference. Follow the steps below
to turn the notch switch on or off:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Setup】submenu.
3. Switch on or off【Notch Filter】.
NOTE:
Only the 【Filter Mode】is set to 【Diagnose Mode】 or 【ST】 can you switch
8-12
S12 Vet Monitor User’s Manual
According to the mains frequency of your country, you can set the frequency of the
notch to【50Hz】 or【60Hz】. If you need to change the 【Notch Frequency】, please
contact the manufacturer maintenance personnel.
This monitor provides the function of switching main lead automatically. When
switch on 【Smart Lead】(Smart lead auto switchover), the current smart leads are
automatically identified by the algorithm, and the host automatically switches the smart
leads according to the identification of the algorithm.
Steps of switching off smart lead function are as follows:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Setup】submenu.
3. Switch off【Smart Lead】.
The steps to set the alarm level for ECG lead off alarms are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Other】submenu.
3. Set【ECG Lead Off Alarm Level】.
The QRS volume is determined by【Alarm Source】in the ECG or PR alarm setting
menu. Which parameter (HR or PR) is set to【Alarm Source】and the QRS volume is
sounded according to which parameter's rhythm.
The volume of QRS sound can be set, the steps are as follows:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
8-13
S12 Vet Monitor User’s Manual
2. Select【Setup】submenu.
3. Set【QRS Volume】
When valid SpO2 measurements are available, the monitor adjusts the pitch tone of
QRS volume based on the SpO2 value. For detail, please refer to 10.5.5 Setting Pitch Tone.
The signal quality of the ECG waveform provides two display modes. The monitor
displays the signal quality of the main calculated lead waveform by default. You can set
the signal quality of the multi-lead waveform as required. The setting steps are as follows:
1. Select ECG parameter area or waveform area to enter the【ECG】menu.
2. Select【Setup】submenu.
3. Switch on【MultiLead Signal】.
Multi-lead Signal Quality: The color of the ECG signal of all leads is
indicated by the waveform color respectively. The five colors of white, red,
orange, yellow and green respectively correspond to the five signal quality
levels of extreme bad, bad, general, good and excellent.
When switch off【MultiLead Signal】,
Main-lead Signal Quality: The signal quality of the main calculation lead is
indicated by a triangular diagram of 5 grids, and 1 to 5 grids respectively
correspond to five signal quality levels of extreme bad, bad, general, good and
excellent. The signal quality is displayed above the icon (SQI) value, which
unit is "%".
Select the ECG standard according to the leads you are using. To select the ECG
standard, follow this procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Module】submenu→【ECG】submenu.
3. Set【ECG Standard】to【AHA】or【IEC】.
8-14
S12 Vet Monitor User’s Manual
WARNING:
Arrhythmia may affect heart rate. When monitoring arrhythmia patients, do
not rely entirely on the alarm information calculated by heart rate, but always
place the patients under close surveillance.
Arrhythmia function is applicable for detecting certain ventricular and atrial
arrhythmias, not all atrial or supraventricular arrhythmias. Sometimes, it may
detect wrong arrhythmia. Therefore, doctors must combine more clinical
manifestations to analyze arrhythmia information.
CAUTION:
Since the arrhythmia detection algorithm sensitivity and specificity is less than
100%, sometimes there may be some false arrhythmias detected and also some
true arrhythmia events may not be detected. This is especially true when the
signal is noisy.
The amplitude of ECG waveform will affect the arrhythmia detection and
heart rate calculation sensitivity.
If the QRS amplitude is too low, the monitor may not be able to calculate the
heart rate and false asystole may occur.
Arrhythmia detection may not be available during ECG relearning. Therefore,
the patient’s state should be closely observed during ECG relearning and
within a few minutes after completion.
8-15
S12 Vet Monitor User’s Manual
Arrhythmia Events Description
Asystole There is no fluctuation or very small and slow waveform for 6
seconds.
Vent Fib/Tach Ventricular fibrillation waveform for 4 seconds.
V-Tach More than 5 (including 5) ventricular waveforms were detected
continuously, and the heart rate was greater than the ventricular
tachycardia heart rate limit.
Vent Brady More than 3 (including 3) ventricular waveforms were detected
continuously, and the heart rate was less than the ventricular
bradycardia limit.
Extreme Tachy Non-ventricular rhythm and the heart rate are greater than the extreme
tachycardia limit.
Extreme Brady Non-ventricular rhythm and the heart rate are less than the extreme
bradycardia limit.
R on T Ventricular premature beats appear on the T wave of the previous
cardiac cycle.
Tachy Non-ventricular rhythm and the heart rate are greater than the
tachycardia limit.
Brady Non-ventricular rhythm and heart rate less than bradycardia limit.
Nonsustained V-Tach Three or four consecutive ventricular waveforms and the heart rate are
greater than the ventricular tachycardia heart rate limit.
Vent Rhythm More than 5 (including 5) ventricular waveforms were detected
continuously, and the heart rate was less than the ventricular
tachycardia heart rate limit and greater than the ventricular bradycardia
heart rate limit.
PNC One cardiac leak and one pacing pulse were detected.
PNP One cardiac leak was detected, but no pacing pulse was detected.
Pause No heartbeat is detected within 1.75× of the average R-R interval
(when the heart rate is less than 100), or no heartbeat is detected within
1 second (when the heart rate is more than 100) and the current RR
interval is greater than 4 seconds and less than 6 seconds.
Pauses/min High The number of Pause per minute is greater than the decision limit.
Run PVCs For 3 or 4 consecutive ventricular waveforms, the heart rate is less
than the ventricular tachycardia heart rate limit and greater than the
ventricular bradycardia heart rate limit.
Couplet Two consecutive ventricular waveforms.
8-16
S12 Vet Monitor User’s Manual
Arrhythmia Events Description
Bigeminy Dominant rhythm of N, V, N, V.
Trigeminy Dominant rhythm of N, N, V, N, N, V.
Frequent PVCs The number of PVC per minute is greater than the decision limit.
PVC Occasional ventricular premature beat.
Missed Beat No heartbeat is detected within 1.75× of the average R-R interval
(when the heart rate is less than 100bpm), or no heartbeat is detected
within 1 second (when the heart rate is more than 100bpm) and the
current RR interval is less than 4 seconds.
A-Fib RR interval of normal cardiac beats is irregular and there is no P wave.
A-Fib End No atrial fibrillation was detected within the delay time after the end of
atrial fibrillation.
ECG Noise There is too much noise to analyze the waveform.
Irregular Rhythm Always an irregular rhythm.
Irregular Rhythm End No irregular rhythm was detected within the delay time after the end of
the irregular rhythm.
NOTE: The alarm level for lethal arrhythmia is always high and cannot be
8.7. ST Monitoring
ST segment of ECG waveform refers to the phase from the end of ventricular
depolarization to the beginning of ventricular repolarization, or from the end of QRS
complex (point J) to the beginning of T wave. ST segment analysis is mostly used to
8-17
S12 Vet Monitor User’s Manual
WARNING:
Factors such drugs, metabolism or conduction disorders may affect ST values.
Since ST is calculated by a fixed delay after point J, it may be affected by
changes in heart rate.
The data accuracy of ST algorithm has been tested, and its clinical significance
should be decided by doctors.
The monitor provides ST segment change information, and the clinical opinion
of this information should be decided by the doctor.
8-18
S12 Vet Monitor User’s Manual
1
2
4 5
8-19
S12 Vet Monitor User’s Manual
6 5 4
(1) ST lead
(2) The current ST value
(3) ST baseline value
(4) The current ST segment (green) and baseline ST segment (white)
(5) ST segment measurement position line
(6) Scale
ST View displays a complete QRS segment of each ST lead. You can enter 【ST
View】 to view these ST segments. The color of the current ST segment and ST value is
8-20
S12 Vet Monitor User’s Manual
the same as that of ECG waveform, usually green. ST baseline segment and baseline
value are white.
You can select the ST waveform area to enter the 【ST View】 page or enter the 【ST
View】 page through the following steps:
1. Select ECG parameter area, waveform area or ST parameter area to enter
【ECG】 menu.
2. Select 【ST】 submenu.
3. Select 【ST View】 from the bottom of the menu.
ST analysis requires valid samples. Set an ST baseline when ST values become stable.
If you do not set a baseline, the monitor will automatically save a set of baselines about 5
minutes after a valid ST measurement appears. You can also manually update the baseline
by selecting 【Set Baseline】 in the lower left corner of the 【ST View】 interface.
You can also make the following settings under the ST interface:
Select 【Show Baseline】 or 【Hide Baseline】 to show or hide ST baseline
segments and parameter values.
Select 【Show Mark】 or 【Hide Mark】 to show or hide ST reference point,
J point and ST point positions.
CAUTION:
Changing the ST baseline will affect the ST alarm.
8-21
S12 Vet Monitor User’s Manual
The following figure shows ST Graphic. The height of the bar represents the ST
value of the corresponding ST lead. The color of the bar indicates the ST alarm status:
green indicates that the ST value is within the normal range; Cyan, yellow and red
indicate that the ST value exceeds the alarm limit. The alarm color corresponds to the
level of ST alarm.
8.7.8. ST Setup
The ISO point, J point and ST point position marks are not displayed by default on
the ST segment in the waveform area. To display these marks, the steps are as follows:
1. Select ECG parameter area, waveform area, ST parameter area or ST
waveform area to enter 【ECG】 menu.
2. Select 【ST】 submenu→【Setting】submenu.
3. Switch on 【ST Mark】.
8-22
S12 Vet Monitor User’s Manual
The ST value for each beat complex is the vertical difference between the isoelectric
(ISO) point and the ST point. The ISO point provides the baseline. The ST point is at the
midpoint of the ST segment. The J point is where the QRS complex changes its slope. As
the J point is a fixed distance away from the ST point, it can be useful to help you
correctly position the ST point.
(1) ISO Baseline Point (2) J Point (3)ST Measurement Point (4) ST Value
CAUTION:
When you start monitoring or the patient's heart rate or ECG waveform has
obvious changes, it may affect the length of QT interval, thus affecting the
position of ST points, so the positions of ISO and ST points need to be adjusted.
Incorrect setting of ISO point or ST point may lead to false ST segment
depression or elevation. Please always ensure that the location of ST
measurement point is suitable for the patients under monitoring.
The steps for setting ST, J and ISO points are as follows:
1. Select ECG parameter area, waveform area, ST parameter area or ST
waveform area to enter 【ECG】 menu.
2. Select 【ST】 submenu→【Adjust】submenu.
8-23
S12 Vet Monitor User’s Manual
The setting of 【Auto Adjust】 defines the method of adjusting the ISO point and J
point. When the 【Auto Adjust】 switch is turned on, the module automatically adjusts
the positions of ISO and J points according to the current waveform. When the 【Auto
Adjust】 switch is off, you can manually adjust the positions of 【ISO】 and 【J】
through “+” and “-”.
The ISO point (isoelectric) position is given relative to the R-wave peak.
Position the ISO point in the middle of the flattest part of the baseline
(between the P and Q waves).
The J point position is given relative to the R-wave peak and helps locating
the ST point. Position the J point at the end of the QRS complex and the
beginning of the ST segment.
The ST point is located at a fixed distance relative to the J point, and the J
point is moved so that the ST point is located in the middle of the ST segment.
The ST point can be located at the positions of J+0, J+20, J+40, J+60, and
J+80.
QT interval is the time from the beginning of QRS complex to the end of T wave, that
is, the whole period of ventricular action potential depolarization (QRS interval) and
repolarization phase (ST-T). QT test can help you to judge long QT interval syndrome.
QT interval is negatively correlated with heart rate. As heart rate increases, the QT
interval shortens, while at lower heart rates QT interval gets longer. We can use several
formulas to correct QT interval according to heart rate. The QT interval corrected by heart
rate is called QTc.
8-24
S12 Vet Monitor User’s Manual
QT/QTc monitoring function is off by default, and you need to turn it on before
performing QT/QTc monitoring. Enable QT/QTc monitoring as follows:
1. Select ECG parameter area or waveform area to enter 【ECG】 menu.
2. Select 【QT】 submenu→【Setting】submenu.
3. Switch on 【QT Analysis】.
8-25
S12 Vet Monitor User’s Manual
leads. You can enter QT View to view QT-HR. For details, please refer to
8.8.4 Entering QT View.
The QT parameter area is displayed as follows. Depending on the settings, the display
of your monitor may be different.
4
3
2 5
1
6
(1) QTc alarm limit (if QTc alarm is off, the alarm off icon is displayed here)
(2) Parameter Unit
(3) Parameter Label
(4) QTc value
(5) ΔQTc value (the difference between the current value of QTC and the baseline
value; if ΔQTc alarm is off, the alarm off icon is displayed on the right side of
the value)
(6) QT value
8-26
S12 Vet Monitor User’s Manual
The current waveform is displayed at the top of the view, and the color is the
same as the ECG waveform, usually green.
The baseline segment is displayed below in white.
The starting point of QRS complex and the ending point of T wave are marked
with vertical lines.
In some cases, the algorithm may not be able to give QT measurement results
because the waveform does not meet the requirements. At this time, the reason
that cannot be analyzed will be displayed below the QT parameter area in QT
View. In addition, a prompt message "QT cannot be analyzed" will be
displayed in the technical alarm information area of the main interface.
Select the lead label at the lower left of QT View, switch leads, and highlight
the waveforms of the corresponding leads.
Setting QT baseline is helpful to quantify QTc changes. After QT valid values appear,
if you do not set QT baseline within 5 minutes, the monitor will automatically set QT
baseline.
The steps for manually setting QT baseline are as follows:
1. Select the 【Set Baseline】 button below QT View.
2. Select 【OK】 in the pop-up dialog box to set the current QT parameter value
as the baseline. The baseline value will be used to calculate the ΔQTc value.
After the new QT baseline is set, the original baseline will be discarded. The
8-27
S12 Vet Monitor User’s Manual
CAUTION:
Changing QT baseline will affect ΔQTc value and ΔQT alarm.
8.8.6. QT Setting
The monitor uses Hodges formula by default to correct QT interval according to heart
rate. If you need to select other QTc formulas, the steps are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Module】submenu→【ECG】submenu.
3. Select【QTc Formula】.
Hodges:
Bazett:
Fridericia:
Framingham:
8-28
S12 Vet Monitor User’s Manual
Between the global onset and offset of the QRS-complex, signal parts with a duration
of more than 6 ms and amplitudes not exceeding 20μV for at least three samples should
be defined as isoelectric segments – I-wave before the global QRS-ONSET and K-wave
after the global QRS-OFFSET.
Isoelectric parts (I-wave) after global QRS-ONSET or before global QRS-OFFSET
(K-wave) are excluded in the duration measurement of the respective adjacent waveform.
The ECG signal may be inaccurate due to hardware or software problems. As a result,
8-29
S12 Vet Monitor User’s Manual
the ECG waveform amplitude becomes greater or smaller. In that case, you need to
calibrate the ECG module. To do so, follow this procedure:
1. Select ECG parameter area or waveform area, set 【 Filter Mode 】 to
【Diagnose】.
2. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
3. Select【Module】submenu→【ECG】submenu.
4. Select【Calibrate】, the square wave signal will appear on the screen to
compare the amplitude of the square wave with the scale. The error range
should be within 5%. The ECG calibration must be completed by the
maintenance personnel.
2. Select【Module】submenu→【Auxiliary Output】submenu.
CAUTION:
Improper use of a defibrillator may cause injury to the patient. The operator
should determine whether to perform defibrillation or not according to the
patient’s condition.
According to AAMI specifications the peak of the synchronized defibrillator
discharge should be delivered within 60 ms of the peak of the R-wave. The
signal at the ECG output on the monitors is delayed by maximum of 25ms.
8-30
S12 Vet Monitor User’s Manual
This section lists the problems that might occur. If you encounter problems when
using the monitor or accessories, check the table below before requesting for services. If
the problem persists after you have taken corrective actions, contact your service
personnel.
8-31
S12 Vet Monitor User’s Manual
3. Check for ECG filter setting. Set ECG Filter mode to 【Monitor】.
Information.
8-32
S12 Vet Monitor User’s Manual
Impedance respiration is measured across the thorax. When the patient is breathing,
the volume of air changes in the lungs, resulting in impedance changes between the
electrodes. Respiration rate (RR) is calculated from these impedance changes, and a
respiration waveform appears on the monitor screen.
WARNING:
If you do not set the detection level for the respiration correctly in manual
detection mode, it may not be possible for the monitor to detect apnea. If you
set the detection level too low, the monitor is more likely to detect cardiac
activity, and to falsely interpret cardiac activity as respiratory activity in the
case of apnea.
The respiration measurement does not recognize the cause of apneas. It only
indicates an alarm if no breath is detected when a pre-adjusted time has
elapsed since the last detected breath. Therefore, it cannot be used for
diagnostic purpose.
If operating under conditions according to the EMC Standard IEC 60601-1-2
(Radiated Immunity 3V/m), field strengths above 3V/m may cause erroneous
measurements at various frequencies. Therefore, it is recommended to avoid
the use of electrically radiating equipment in close proximity to the respiration
measurement unit.
The impedance respiration measurement may cause rate changes in Minute
Ventilation Rate Responsive Pacemakers. Set the pacemaker rate responsive
mode off or disable the impedance respiration measurement on the monitor.
When using the electrosurgery unit, ensure proper contact of the ESU return
electrode to the patient to avoid burns at monitor measurement sites. Also
ensure that the ESU return electrode is near the operating area.
9-1
S12 Vet Monitor User’s Manual
CAUTION:
Only use parts and accessories specified in this manual, and obey all warnings
and cautions.
Respiration monitoring is not for use on the patients who are very active, as
this will cause false alarms.
8
5
As the Respiration measurement adopts the standard ECG electrode placement, you
can use different ECG cables. Since the respiration signal is measured between two ECG
electrodes, if a standard ECG electrode placement is applied, the two electrodes should be
RA and LA, or RA and LL.
For more information, see 8.4.2 Applying Electrodes.
9-2
S12 Vet Monitor User’s Manual
RA-LA RA-LL
CAUTION:
Correct electrodes placement can help to reduce cardiac overlay: avoid the
liver area and the ventricles of the heart in the line between the respiratory
electrodes.
Some patients with restricted movements breathe mainly abdominally. In these
cases, you may need to place the left leg electrode on the left abdomen at the
point of maximum abdominal expansion to optimize the respiratory wave.
In clinical applications, some patients expand their chests laterally, causing a
negative intrathoracic pressure. In these cases, it is better to place the two
respiration electrodes in the right midaxillary and the left lateral chest areas at
the patient’s maximum point of the breathing movement to optimize the
respiratory waveform.
Regularly inspect the electrode application site to ensure skin quality. If there
are signals of allergies, change the clip application site.
If you want to measure Resp and you are already measuring ECG, you may need to
optimize the placement of the two electrodes between which Resp will be measured.
Repositioning ECG electrodes from standard positions results in changes in the ECG
waveform and may influence ST and arrhythmia interpretation.
9-3
S12 Vet Monitor User’s Manual
Cardiac activity that affects the Resp waveform is called cardiac overlay. It happens
when the Resp electrodes pick up impedance changes caused by the rhythmic blood flow.
Correct electrodes placement can help to reduce cardiac overlay: avoid the liver area and
the ventricles of the heart in the line between the respiratory electrodes.
Some patients with restricted movement breathe mainly abdominally. In these cases,
you may need to place the left leg electrode on the left abdomen at the point of maximum
abdominal expansion to optimise the respiratory wave.
In clinical applications, some patients expand their chests laterally, causing a negative
intrathoracic pressure. In these cases, it is better to place the two respiration electrodes in
the right midaxillary and the left lateral chest areas at the patient’s maximum point of the
breathing movement to optimize the respiratory waveform.
9-4
S12 Vet Monitor User’s Manual
You can set up respiration lead to get the best respiratory waveform. The steps to set
up breathing leads are as follows:
1. Select the RESP parameter area or waveform area to enter the 【RESP】menu.
2. Select【Setup】submenu.
3. Set 【RESP Lead】. If the respiratory waveform is still poor after adjusting the
respiration lead or the respiration rate measurement is suspected to be
inaccurate, you can adjust the electrode position.
You can adjust the RESP waveform gain to better view the waveform amplitude. The
steps to set the RESP waveform gain are as follows:
1. Select the RESP parameter area or waveform area to enter the 【RESP】menu.
2. Select【Setup】submenu.
3. Set【Gain】.
9-5
S12 Vet Monitor User’s Manual
2. Select【Setup】submenu.
3. Switch on or off【Auto Threshold Detection】.
If 【 Auto Threshold Detection 】 is switched on, the monitor
automatically adjusts the RESP waveform detection level, or threshold.
【Auto Threshold Detection】is switched off, you have to manually
adjusts the RESP waveform threshold. For more information, see
9.5.7.Manually Adjust the RESP Waveform Detection Threshold.
9-6
S12 Vet Monitor User’s Manual
Chapter 10 SpO2
10.1. Introduction
Parameter Display 2 3
6
4
7
5 8
(1) Pleth waveform (Pleth): The amplitude of the waveform can directly reflect the
strength of the patient’s pulse signal. The waveform is not normalized.
(2) SpO2 value: Percentage of oxygenated hemoglobin in relation to the sum of
oxyhemoglobin and deoxyhemoglobin.
(3) Pleth bar: Proportional to the intensity of the pulse.
(4) SpO2 unit
(5) SpO2 alarm limits: If the SpO2 alarm is turned off, the alarm off icon is displayed
here.
(6) Perfusion index (PI): Applicable for SpO2 module. Gives the numerical value for
the pulsatile portion of the measured signal caused by arterial pulsation. PI indicates
10-1
S12 Vet Monitor User’s Manual
the signal strength of SpO2 and also partially indicates the signal quality.
Above 1 is optimal;
Between 0.3 and 1 is acceptable;
Below 0.3 indicates low perfusion. If the PI is less than 0.3, the low perfusion
status is indicated (alternating with a question mark for SpO2 measurements),
indicating that the SpO2 measurements may be inaccurate. Reposition the
SpO2 sensor or find a better site. If low perfusion persists, choose another
method to measure oxygen saturation if possible.
(7) Pulse rate (PR) alarm off icon: pulse rate alarm is turned off.
(8) Pulse rate: the number of pulses detected per minute (from the pleth waveform)
WARNING:
Use only SpO2 sensors specified in this manual. Follow the SpO2 sensor’s
instructions for use and adhere to all warnings and cautions.
Before use, the operator needs to verify the compatibility between the monitor,
probe and cable. Otherwise, it may cause injury to the patient.
When a trend toward patient deoxygenation is indicated, analyze the blood
samples with a laboratory co-oximeter to completely understand the patient’s
condition.
Do not use SpO2 sensors during magnetic resonance imaging (MRI). Induced
current could potentially causes burns. The sensor may affect the MRI image,
and the MRI unit may affect the accuracy of the oximetry measurements.
Prolonged continuous monitoring may increase the risk of undesirable changes
in skin characteristics, such as irritation, reddening, blistering or burns.
Inspect the sensor site every 2 hours and move the sensor if the skin quality
changes. Change the application site every 4 hours. For patients with poor
peripheral blood circulation or sensitive skin, inspect the sensor site more
frequently.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry
probes and pulse oximetry monitors.
10-2
S12 Vet Monitor User’s Manual
CAUTION:
Use only specified accessories in this manual. Follow the instructions for use
and adhere to all warnings and cautions.
Depending on the animal category, weight and application site, you can select the
SpO2 sensor as required.
2. Connecting SpO2 Sensor
Plug the SpO2 sensor cable into the SpO2 connector on the measurement module.
3. Applying SpO2 Sensor
Place the sensor on the animal’s tongue or ear. For dogs, cats and equines, place the
sensor on their tongue. When placing the sensor, place the optical part of the sensor in the
center of the tongue. You can also place the sensor on the animal's lips, toes, ears, prepuce
and vulva.
CAUTION:
Do not apply sensor too tightly as this results in venous pulsation which may
severely obstruct circulation and lead to inaccurate measurements.
At elevated ambient temperatures be careful with measurement sites that are
not well perfused, because this can cause burns after prolonged application.
Avoid placing the sensor on extremities with an arterial catheter, an NIBP cuff
or an intravascular venous infusion line.
10-3
S12 Vet Monitor User’s Manual
The SpO2 value displayed on the monitor screen is the average of data collected
within a specific time. The shorter the averaging time is, the quicker the monitor responds
to changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging
time is, the slower the monitor responds to changes in the patient’s oxygen saturation
level, but the SpO2 measurement is more stable. For critically ill patients, selecting
shorter averaging time will help understanding the patient’s state.
To set the averaging time, follow this procedure:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【SpO2 Setup】submenu.
3. Select【Sensitivity】and then toggle between【High】,【Med】or【Low】.
When monitoring SpO2 and NIBP on the same limb simultaneously, you can switch
on 【NIBP Simul】 to lock the SpO2 alarm status until the NIBP measurement ends. If
you switch off 【NIBP Simul】, low perfusion caused by NIBP measurement may lead to
inaccurate SpO2 readings and therefore cause false physiological alarms.
To set the 【NIBP Simul】, follow this procedure:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【Alarm】submenu.
3. Set【NIBP Simul】as to【On】or【Off】.
10-4
S12 Vet Monitor User’s Manual
The pitch tone function is on by default. The steps to turn off the pitch tone function
are as follows:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【SpO2 Setup】submenu.
3. Switch off【Pitch Tone】.
10.5. Setting PR
If the alarm source is set to PR, the QRS tone is derived from PR measurements. To
set the QRS volume, follow this procedure:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【PR Setup】submenu.
3. Set【QRS Volume】to the appropriate value.
If the SpO2 value is effective, the monitor also adjusts the QRS tone (Pitch tone)
10-5
S12 Vet Monitor User’s Manual
according to the SpO2 value. For information, see 10.5.5 Setting Pitch Tone.
Current pulse source is displayed in the PR parameter area. The PR from current pulse
source has the following characteristics:
PR is monitored as system pulse and generates alarms when you select PR as
the active alarm source;
PR is stored in the monitor’s database and reviewed in the graphic/tabular
trends; in trend graphs, as the PR curve is in the same color with that of the
PR source, it is unlikely to distinguish the PR source.
To set which pulse rate as PR source, follow this procedure:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【PR Setup】submenu.
3. Select【PR Source】, and select a suitable PR source in the drop-down list.
The drop-down list of 【PR Source】displays the currently valid PR source from top
to bottom according to the priority level. When you select【Auto】, the system will
automatically select the first option in the list as the PR source. If the PR source you set
does not exist, the system will automatically switch【PR Source】to【Auto】. When you
select 【IBP】, the system will automatically use the first pressure label in the list as the
PR source.
If you doubt the SpO2 measurements, check the patient’s vital signs first, then check
the monitor and SpO2 sensor. The following factors may influence the accuracy of
measurements:
There is excessive illumination from light sources such as a surgical lamp, a
brilirubin lamp, or sunlight;
Excessive patient movement;
Diagnostic test;
Low perfusion;
Electromagnetic interference, such as MRI device;
Electrosurgical equipment;
10-6
S12 Vet Monitor User’s Manual
10-7
S12 Vet Monitor User’s Manual
Chapter 11 Temperature(TEMP)
11.1. Introduction
The following figure shows the TEMP parameter area for temperature monitoring.
Your display may be configured to look different.
1
2
4 5
(1) Parameter label
(2) TEMP unit
(3) TEMP alarm limits: If TEMP alarm is turned off, the alarm closing icon is displayed
11-1
S12 Vet Monitor User’s Manual
here.
(4) TEMP value
(5) TEMP Difference (ΔT): TEMP Difference between two temperature sites. It displays
only when ΔT is switched on.
Select the temperature label according to the measurement site. To do so, follow this
procedure:
1. Select the TEMP parameter area to enter the【TEMP】menu.
2. Select【Setup】submenu.
3. Set the TEMP label name according to the measurement site.
11-2
S12 Vet Monitor User’s Manual
You can change the unit of TEMP by following the steps below:
1. Select the TEMP parameter area to enter the【TEMP】menu.
2. Select【Setup】submenu.
3. Set TEMP【Unit】.
This section lists the problems that might occur. If you encounter the problems when
using the equipment or accessories, check the table below before requesting for services.
If the problem persists, contact your service personnel.
NOTE: For the physiological and technical alarm messages, see D Alarm
Information.
11-3
S12 Vet Monitor User’s Manual
11-4
S12 Vet Monitor User’s Manual
Chapter 12 NIBP
12.1. Introduction
The monitor uses the oscillometric method for measuring the non-invasive blood
pressure (NIBP). NIBP measurement is based on the principle that pulsatile blood flow
through an artery creates oscillations of the arterial wall. The oscillometric device uses a
blood pressure cuff to sense these oscillations that appear as tiny pulsations in cuff
pressure. The oscillometric devices measure the amplitude of pressure changes in the
occluding cuff as the cuff deflates from above systolic pressure. The amplitude suddenly
increases as the pulse breaks through the occlusion in the artery. As the cuff pressure
decreases further, the pulsations increase in amplitude, reach a maximum (which
approximates to the mean pressure), and then diminish. The oscillometric method
measures the mean pressure and determines the systolic and diastolic pressures.
NOTE:
WARNING:
Be sure to select the correct patient category setting for your patient before
NIBP measurement.
Do not measure NIBP on patients with sickle-cell disease or on the limb where
skin damage has occurred or is expected.
Use clinical judgment to determine whether to perform frequent automatic
12-1
S12 Vet Monitor User’s Manual
CAUTION:
Only use parts and accessories specified in this manual. Follow the instructions
for use and adhere to all warnings and cautions.
Accuracy of NIBP measurement depends on using a cuff of proper size. It is
essential to measure limb circumference and choose a cuff with proper size.
12-2
S12 Vet Monitor User’s Manual
NIBP automatically calibrated every time when the monitor is turned on. If it
is not turned off for a long time or if the pressure is not accurate during use,
you can use the “Reset” function in the NIBP menu to calibrate. You need to
remove the cuff and windpipe before calibration, which in order to connect the
NIBP pressure sensor to the atmosphere.
Measurements are impossible with heart rate extremes of less than 40bpm or greater
than 240bpm, or if the patient is on a heart-lung machine. The measurement may be
inaccurate or impossible in the following situations:
Regular arterial pressure pulses are hard to detect;
With excessive and continuous patient movement such as shivering or
convulsions;
With cardiac arrhythmias;
With rapid blood pressure changes;
With severe shock or hypothermia that reduces blood flow to the peripheries;
On an edematous extremity;
12-3
S12 Vet Monitor User’s Manual
3 4 5
1
2
12 6
11 10 9 8 7
NOTE:
12-4
S12 Vet Monitor User’s Manual
CAUTION:
A wrong cuff size and a folded or twisted bladder can cause inaccurate
measurements.
Do not touch or apply external pressure against the cuff and air tubing during
NIBP measurement. This may cause inaccurate blood pressure values.
Use care when placing the cuff on an extremity used for monitoring other
patient parameters. Pressurization of the cuff may temporarily cause loss of
12-5
S12 Vet Monitor User’s Manual
12-6
S12 Vet Monitor User’s Manual
the cuff should be wrapped around the metatarsus just proximal to the tarsal pad or around
the hind leg just distal to the hock. The tail site should not be used for cuff placement
during anesthesia.
It is not necessary to center the cuff over the artery because of the fully encircling
bladder design. If the hair over the artery site is too thick or matted for good contact, it
should be clipped.
Start and stop NIBP measurement by selecting the NIBP quick key keys or from the
NIBP menu:
Task Via quick key Via NIBP menu
Start a manual
Select【NIBP Start/Stop】quick key Select【Start】
measurement
Select【Setup】
NIBP Auto Select【NIBP Measure】quick
submenu→set【Interval
measurement
key →Select interval time
Time】
Select【Sequence】
Select【NIBP Measure】quick
submenu→set NIBP
NIBP Sequence
key →Select【Sequence】→Select Sequence
measurement
measurement→Select
【NIBP Start/Stop】quick key
【Start】
12-7
S12 Vet Monitor User’s Manual
Task Via quick key Via NIBP menu
Start STAT Select【STAT】
Select【NIBP STAT】quick key
measurement
You can manually set the initial inflation pressure of the cuff, follow this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu.
2. Select【Setup】submenu.
3. Set【Initial Pressure】: Select the appropriate cuff pressure value as needed.
For Auto NIBP measurement, you need to set the interval between two NIBP
measurements. To set the NIBP interval, follow this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu.
2. Select【Setup】submenu.
12-8
S12 Vet Monitor User’s Manual
3. Set【Interval Time】.
Start mode defines how NIBP auto mode works. To set the 【Start Mode】, follow
this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu
2. Select【Setup】submenu.
3. Set【Start Mode】.
【Clock】: After the first measurement, the monitor automatically
synchronizes NIBP automatic measurements with the real time clock.
For example, if 【Interval】 is set to 【30min】, and you start NIBP auto
measurement at 10:03, the next measurement will be taken at 10:30, and
then at 11:00, 11:30, and so on.
【Interval】: After the first measurement, the monitor automatically
repeats measurements at set interval. For example, if 【Interval】 is set
to【30min】, and you start NIBP auto measurement at 10:03, the next
measurement will be taken at 10:33, and then at 11:03, 11:33, and so on.
The monitor can issue a reminder tone at the completion of NIBP measurement. The
NIBP End Tone is off by default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu
2. Select【Setup】submenu.
3. Switch on【NIBP End Tone】.
12-9
S12 Vet Monitor User’s Manual
NIBP measurements become outline fonts after a preset time. This avoids older NIBP
values being misinterpreted as current measurements. To set the timeout period, follow
this procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Module】submenu→【Other】submenu.
3. Set【NIBP Invalid Time】.
To display multiple sets of the latest NIBP measurements, follow this procedure:
1. Enter【Screen Layout】submenu by either of the following ways:
Select【Screen Setup】quick key→Select【Screen Layout】submenu.
Select【Main Menu】quick key→from 【Display】 column to select
【Screen Layout】.
2. In the desired parameter area, select【NIBP】→【NIBP List】.
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood
vessel and therefore help venous puncture. To assist venous puncture, follow this
12-10
S12 Vet Monitor User’s Manual
procedure:
1. Select the NIBP parameter area.
2. Select【Setup】submenu;
3. Set【Assisted Venipuncture Pressure】.
4. Select【Assisted Venipuncture】at the bottom of the menu.
5. Puncture vein and draw blood sample.
6. Select【NIBP Start/Stop】quick key or【Assisted Venipuncture】button to
manually deflate the cuff. When performing a venipuncture, observe the inflation pressure
and the remaining time of the venipuncture in the NIBP parameter area.
The NIBP leakage test checks the integrity of the system and of the valve. The NIBP
leakage test should be performed once every two years or when you doubt the NIBP
measurements. The NIBP leakage test should be performed by qualified service personnel
only.
The NIBP accuracy test should be performed once every two years or when you
doubt the NIBP measurements. The NIBP accuracy test should be performed by qualified
service personnel only.
12-11
S12 Vet Monitor User’s Manual
Chapter 13 IBP
13.1. Introduction
The method of IBP measurement is direct measuring the BP of artery or veins on the
pressure sensor mainly through liquid coupling so as to obtain the pressure curve of the
continuous BP.
The monitor can provide up to 2 channels of IBP measurement results.
WARNING:
Use only IBP transducers specified in this manual. Never reuse disposable
pressure transducers.
The operator should avoid contact with conductive parts of the accessories
when being connected or applied.
When the monitor is used with HF surgical equipment, the transducer and the
cables must be avoided conductive connection to the HF equipment to protect
against burns to the patient.
All invasive measurement involves risks to the patient. Use aseptic technique
and perform according to manufacturer’s instructions during measurement.
Mechanical shock to the IBP sensor may cause severe shifts in zero balance and
calibration, and cause erroneous readings.
13-1
S12 Vet Monitor User’s Manual
CAUTION:
Before IBP measurements, it should make sure all IBP sensors are zeroed
properly.
Before IBP measurement, it makes sure no air bubbles in the IBP sensor which
result in erroneous pressure readings.
When intracranial pressure (ICP) measurements put on a sitting patient, the
13-2
S12 Vet Monitor User’s Manual
sensor should be in line with the top of the patient's ear. Incorrect position can
result in erroneous pressure readings.
To obtain accurate pressure readings, the monitor requires a valid zero point. Zeroing
the sensor at the hospital's specified frequency, and zeroing must be performed in the
following cases:
Every time reconnecting IBP sensor and IBP sensor cable.
The monitor needs restart.
Suspecting the monitor's pressure reading is inaccurate.
When the monitor displays the message 【Zero Required】, Please refer to the
following steps to calibrating zero:
1. Connecting IBP sensor, sensor cable and module.
2. Closing the 3-way stopcock (nearby the sensor end) to the patient's valve,
and let the sensor pass through the 3-way stopcock to the atmosphere.
3. Zeroing the sensor using one way from the following methods:
Select the parameter area of the pressure (e.g. ART) and select the
【Zero】button.
Click 【Zero】 quick key→select 【IBP zero】 submenu→select
the pressure to zeroing.
4. After successful zero, close the valve to the atmosphere and open the
valve to the patient. During zero process, when pressure fluctuation or
the pressure exceeds the zero pressure range, it may fail. If fails, the
processing method as follows:
Check valve position of the 3-way stopcock near the sensor end
for ensure access to the atmosphere.
Perform zero against. Do not shake the IBP sensor and tubing
during zero calibration.
IBP measurements display the waveforms of pressure and pressure values on the
13-3
S12 Vet Monitor User’s Manual
screen. For arterial pressure, IBP parameter area displays systolic pressure, diastolic
pressure and mean pressure. For venous pressure, the IBP numeric area displays only the
mean pressure. The figure below shows the waveform and numerics for the Art pressure.
Waveform display
1 2 3
Parameter Display
5 6
7
11
8
10 9
(1) IBP label
(2) Waveform scale
(3) Waveform
(4) Pressure unit: mmHg, kPa or cmH2O
(5) Systolic pressure
(6) Diastolic pressure
(7) Diastolic pressure alarm limit
(8) Mean pressure alarm limit
(9) Mean pressure
(10) PPV measurement value
(11) Systolic pressure alarm limit
13-4
S12 Vet Monitor User’s Manual
The pressure label is identifier for each type only, so the pressure label must be set up
when making pressure measurements. You can choose a pressure label following these
steps:
1. Selecting IBP parameter area or waveform area where you need to change
labels for enter corresponding IBP menu.
2. Select【Setup】submenu.
3. Select 【Label】 where the appropriate tag name in the list.
Label name Description Label name Description
NOTE:
The same label name cannot be used for different channels about IBP
13-5
S12 Vet Monitor User’s Manual
If current pressure label is set to the extended pressure (P1 or P2), you could select
the type which displays in the parameter area, following these steps:
1. Select the parameter area or waveform area about the extended pressure for
entering the corresponding pressure menu.
2. Select【Setup】submenu.
3. Set【Measurement】:
All: Corresponding pressure in the parameter area shows all the pressure:
systolic pressure, diastolic pressure and mean pressure.
Mean only: Corresponding pressure in the parameter area only shows
the average pressure.
Auto: The system will automatically shows the pressure is displayed in
the parameter area or only the average pressure according to the
measured value of the extended pressure.
The blood pressure value displayed on the monitor is average calculation about the
collected data over a period time. The higher the sensitivity, the faster the monitor
responds when the patient's blood pressure value changes, but the measurement accuracy
is lower. Inversely, the lower the sensitivity, the slower the response of the monitor when
the patient's blood pressure value changes, but the measurement accuracy is higher. When
monitoring critically ill patients, setting up a higher sensitivity is useful for timely
analysis.
You can set up the sensitivity of the current pressure, following these steps:
1. Select the IBP parameter area or waveform area for enter the corresponding
pressure menu.
2. Select【Setup】submenu.
3. Set【Sensitivity】.
13-6
S12 Vet Monitor User’s Manual
1. Select the IBP parameter area or waveform area to enter the corresponding
pressure menu.
2. Select【Setup】submenu.
3. Make the following settings for the IBP waveform:
Set【Speed】.
Set 【Scale Type】: If 【Auto】 is selected, the upper and lower scales
of the IBP waveform will be automatically adjusted as the waveform
amplitude changes.
PPV is the pulse pressure variation. When measuring arterial pressure (excluding PA),
you can turn on PPV measurements, following these steps:
1. Select the IBP parameter area or waveform area to enter the corresponding
pressure menu.
2. Select【PPV】submenu.
3. Set【PPV Measurement】 to 【ON】.
When 【PPV Measurement】 is setting to 【ON】, the source of the PPV can
be selected.
WARNING:
The monitor will calculate the PPV based on any arterial pressure value
between heartbeats. The conditions of PPV measurement, and whether the
PPV numerical calculation has clinical significance or not, it is applicable or
not. It must be judged by a doctor.
Only a doctor can determine the clinical value of PPV information. According
to recent scientific literature, the clinical relevance of PPV information is
limited to controlled mechanical ventilation and to sedated patients without
arrhythmias.
The calculated PPV value may not be accurate under the following conditions:
a) Respiration rate is less than 8 rpm
b) During venting, the tidal volume is less than 8ml/kg
13-7
S12 Vet Monitor User’s Manual
c) The patient has acute right ventricular dysfunction (i.e., pulmonary heart
disease)
Please following steps below, set up the IBP waveform overlay display:
1. Enter the 【Screen Layout】page, following method:
Select 【Screen Layout】 quick key → select 【Screen Layout】
submenu.
Select 【Main Menu】 quick key → select 【Screen Layout】 from
the 【Display】 column.
2. Select 【IBP Overlap】 in the waveform parameter area and select the IBP
waveform to be overlapped on the same line on the left side.
3. Repeat the operation of step 2 at other locations with waveform parameter
areas if needed.
4. Select to exit Setup page. The overlapped IBP waveform can be displayed
on the main interface.
You can open the 【IBP Overlap】 menu by selecting the IBP waveform area to be
overlapped on the main screen. In the 【IBP Overlap】 menu, you can make the
following settings:
Scale
13-8
S12 Vet Monitor User’s Manual
This section describes problems you may encounter during using. You can refer to
the following table for troubleshooting. If the problem persists, please contact
maintenance staff.
alarm information.
Problem Solution
IBP parameter area and waveform 1. Check whether the display of IBP
area cannot be found on the parameters is set in the 【Screen Layout】
interface menu or not. For details, see 3.6.1 Setting
Parameters to be protected.
2. Check whether the IBP parameter switch is
turned on or not. For details, see 3.6.1 Setting
Parameters
3. Check the IBP cable, IBP sensor and
module are connected or not.
4. Check the valve position of the 3-way
stopcock is correct or not.
5. Confirm that the sensor has been zeroed.
For details, see 13.3.2 IBP Sensor zero.
P1/P2 does not display systolic and Set the displayed pressure to【All】. For details,
diastolic pressure measurements see 13.5.3 Setting up display types about
Extended Pressure.
13-9
S12 Vet Monitor User’s Manual
Problem Solution
IBP reading is unstable 1. Verify no air bubbles in the IBP sensor
system.
2. Check if the sensor is fixed.
3. Perform zero against.
4. Replace the sensor.
Zero failure 1. Check if the pipeline of the IBP sensor is
open to the atmosphere.
2. Perform zero against. Do not shake the IBP
sensor and tubing during zeroing. For details,
see 13.3.2 IBP Sensor zero.
3. If zero still fails, replace the sensor.
13-10
S12 Vet Monitor User’s Manual
C.O. (cardiac output) measurements were performed using right atrial thermodilution
to measure cardiac output and other hemodynamic parameters. The method is injecting a
cold solution into the blood circulation system and measure the blood temperature drop
caused by the cold solution at a location downstream. In the C.O. measurement window,
the temperature change is shown as a curve. The area under the curve is inversely
proportional to the C.O. value, and the monitor calculates the C.O. value based on the
curve. Since cardiac output is a continuously varying amount, multiple measurements
must be taken to obtain a reliable C.O. average. The monitor can retain 5 results, and user
can select the desired measurement to perform the average calculation.
WARNING:
C.O. measurements may be inaccurate when electrosurgical is performed.
All invasive measurements bring about risks for the patient. Sterile techniques
should be used in the measurements and follow manufacturer's instructions.
Please use the accessories specified in this manual. When using, avoid contact
with the conductive metal body.
14-1
S12 Vet Monitor User’s Manual
C.O. Measurement shows no waveform on the main interface, only the values of C.O.,
C.I. (cardiac displacement index) and TB (blood temperature) are displayed in the
parameter area.
4
1
2 5
3 6
14-2
S12 Vet Monitor User’s Manual
1. Connecting the C.O. cable to the C.O. module or the thermistor interface
sensor port, and confirming that the C.O. measurement area is displayed on the
monitor.
2. Prepare the C.O. measurement for patients according to regulations and
procedures of the hospital department.
3. Connecting the floating catheter and other C.O. measuring accessories in
accordance with the manufacturer's manual.
4. Check that the attached accessories are functioning properly.
NOTE: For an out-line probe setup, make sure that the out-line sensor is
Before performing C.O. measurement, please following steps for set it up:
14-3
S12 Vet Monitor User’s Manual
14-4
S12 Vet Monitor User’s Manual
2
5
3
14-5
S12 Vet Monitor User’s Manual
will calculate and display the average values of C.O. and C.I., and the average
will be stored and displayed in parameter area.
At the time of injection, the 3-way stopcock opens to the floating catheter end and
closes to the injection end. At the end of the measurement, close the end to the floating
catheter, open to the end of the injection, and then inhale the injection into the syringe.
In addition, in the C.O. measurement window, you can also choose:
【Stop】: Press this button during measurement to cancel this measurement.
【Setup】: Enter into the 【Setup】 page of the 【C.O.】 menu. This option
is displayed when the 【 C.O. Measuring 】 hotkey enters the C.O.
measurement window.
【 Hemodynamics 】 : Enter the 【 Calculate 】 menu. See Chapter 19
Calculation.
NOTE:
14-6
S12 Vet Monitor User’s Manual
If the measured value of C.O. is not updated once more within setting time, its
measured value is displayed as a hollow number, which indicates the current measured
value has expired. You could set up the time when the C.O. measurement value expires,
following steps:
1. Select 【Main Menu】 quick key→select 【Maintenance】 from 【System】
column→input password→Enter.
2. Select 【Module】 submenu.
3. Select 【Other】 submenu.
4. Set 【C.O. Invalid Time】.
NOTE:
This section describes the problems you may encounter during using. You can refer
to the following table for troubleshooting. If the problem still persists, please contact
maintenance staff.
14-7
S12 Vet Monitor User’s Manual
NOTE: See the Appendix D Alarm Information for physiological alarms and
technical alarm information.
Problem Solution
Cannot find out 1. Check if the display of C.O. parameter is set in the
C.O. parameter area 【 Screen Layout 】 menu. For details, see 3.6.2 Setting
on the interface Display Screen.
2. Check whether the C.O. parameter switch is on or not. For
details, see 3.6.1 Setting Parameters.
3. Check whether the C.O. cable, floating catheter and liquid
temperature sensor are connected or not.
Suspected that the 1. Check whether the floating catheter position is correct or
C.O. measurement not.
is inaccurate 2. Check out whether the catheter coefficient matches the
injection temperature, volume, and temperature measurement
methods.
3. Inject the solution quickly and smoothly.
4. Complete each injection in 4 to 5 seconds.
5. Increase the volume of the injection or lower the
temperature of the injection.
6. Check the 【Setup】 page, the patient's 【Height】 and
【Weight】 settings are correct.
14-8
S12 Vet Monitor User’s Manual
Problem Solution
2. Complete the injection smoothly within 4 to 5 seconds.
3. Check whether the C.O. cable, floating catheter and liquid
temperature sensor are connected or not.
14-9
S12 Vet Monitor User’s Manual
The monitor adopts infrared absorption technology to measure the carbon dioxide
(CO2) concentration in the breathing airway of patient. Because CO2 molecule can absorb
infrared light of special wavelength, and the amount of absorbed infrared light directly
relates to the concentration of CO2, therefore while the infrared light radiated from the
infrared light source passing through the gas sample containing CO2, part of energy will
be absorbed by CO2 in the gas. At another side of infrared light source, a photodetector is
used to measure the remaining infrared energy and convert it to electric signal, which will
be compared with the energy of infrared light source and adjusted so as to correctly reflect
the CO2 concentration in the gas sample.
There are two methods for measuring carbon dioxide in the patient’s airway:
1. Mainstream: Uses a CO2 sensor attached to an airway adapter directly inserted into
the patient’s breathing system.
2. Sidestream/Microflow: Takes a sample of the respiratory gas with a constant sample
flow from the patient’s airway and analyzes it with the CO2 sensor.
WARNING:
When placing pipes such as sampling tubes, prevent the pipes from suffocating
the patient’s throat.
CAUTION:
When the patient is being treated with aerosolized drugs, the measured EtCO2
value may be inaccurate and is not recommended for use in this situation.
The EtCO2 value measured by the CO2 module may differ from the CO2
partial pressure value measured by blood gas analysis.
The CO2 module has an automatic alarm suppression function, and the CO2
15-1
S12 Vet Monitor User’s Manual
15-2
S12 Vet Monitor User’s Manual
Before attaching the airway adapter to the CO2 sensor, verify that the airway
adapter windows are clean and dry. Clean or replace the adapter if necessary.
Follow these steps:
1) Align the arrow on the bottom of the airway adapter with the arrow on
the bottom of the sensor.
2) Press the sensor and airway adapter together until they click.
3) Wait for the airway adapter and sensor to warm up.
The monitor will display the “Sensor Warm Up…” message for approximately
1 minute while the sensor and adapter warm to operating temperature. The message
disappears when the sensor is ready for use.
Since the Masimo IRMA CO2 mainstream module has a faster heating rate,
there is no “Sensor Warm Up…” prompt.
CAUTION:
The atmospheric pressure must be set to the correct value before using the
mainstream CO2 module. Incorrect atmospheric pressure settings can result in
erroneous CO2 readings (Masimo IRMA CO2 mainstream modules do not
function because they already have automatic atmospheric pressure
compensation).
Install the sensor above the adapter to avoid liquid build-up on the adapter
window. Gas concentration at high concentrations at this location can hinder
gas analysis.
To avoid dead space, place the sensor and adapter as close as possible to the
patient.
15-3
S12 Vet Monitor User’s Manual
CO2
sensor
15-4
S12 Vet Monitor User’s Manual
Sampling
CO2 sensor tube
connector
NOTE:
Inserting the sampling tube into the sampling tube connector will automatically
start the sampling pump. After the sampling tube is pulled out, the sampling
pump stops.
3. If the sampling pump fails to turn on, or runs intermittently, perform a “Zero”
procedure. (Refer to the 15.8 Zeroing)
4. Ensure that the CO2 sensor exhaust tube vents gases away from the sensor
environment.
5. Wait for the CO2 sensor to warm up. The monitor will display the “Sensor Warm
Up...” message for approximately 1minute while the sensor warms up to operating
temperature. The message disappears when the sensor is ready for use.
6. Applying airway adapter or cannula:
1) For intubated patients requiring an airway adapter: Install the airway adapter at the
proximal end of the circuit between the elbow and the ventilator Y section. Shown
as follows:
2) For intubated patients with an integrated airway adapter in the breathing circuit:
Connect the male connector on the straight sample line to the female port on the
airway adapter. Shown as follows:
15-5
S12 Vet Monitor User’s Manual
3) For non-intubated patients: Place the nasal cannula onto the patient. Shown as
follows:
4) For nasal or oral-nasal cannulas with oxygen delivery, place the cannula on the
patient as shown then attach the oxygen supply tubing to the oxygen delivery
system and set the prescribed oxygen flow.
CAUTION:
Always connect the airway adapter to the sensor before inserting the airway
adapter into the breathing circuit. In reverse, always remove the airway adapter
from the breathing circuit before removing the sensor.
Always disconnect the cannula, airway adapter or sampling tube from the CO2
sensor when not in use.
Pull the sampling tube out of the sampling tube connector when not in use.
Do not insert the things other than sampling tube into receptacle of sampling
tube.
The sampling tubes are disposable. Please keep the sampling tube clean, and
prevent the tube from clogging by dust. Under the conditions that the sampling
gas temperature is 37℃, the indoor temperature is 23℃, and the sampling
relative humidity is 100%, replace the sampling tube at least every 12 hours (if
filter tip is used, it can be extended to 120 hours), or replace the sampling tube
when the sampling tube is found to be leaking, damaged or contaminated.
When using a microflow CO2 module, the exhaust holes on the module must be
connected to the exhaust gas treatment system.
15-6
S12 Vet Monitor User’s Manual
② Check whether the LED aperture on the module is green and always bright. (a green
light indicates that the system is normal)
③ Exhale toward the sampling tube and check whether CO2 waveform and value are
displayed on the monitor.
④ Block the sampling tube with your fingers and wait for 10s.
⑤ Check whether the blockage alarm is displayed, and whether the LED aperture on the
module flashed red at the same time.
⑥ Under appropriate circumstances, check the tightness of the patient’s breathing
circuit connected with the sampling tube.
Status indicated by LED aperture on Nomoline ISA CO2 module:
15.5. Display
Waveform Display
Parameter Display
2
15-7
S12 Vet Monitor User’s Manual
6
3
7
4
You can set the apnea alarm time by following the steps below. The monitor will
trigger the alarm when the patient’s suffocation time exceeds the set time.
1. Select the CO2 parameter area or waveform area to enter the【CO2】menu.
2. Select【Alarm】submenu.
3. Set【Apnea Delay】.
WARNING:
It is cannot judge the cause of respiratory apnea through CO2 monitoring. If the
monitor cannot detect the patient's breath again from the moment of the last
breath to the preset time, the monitor only provides the alarm of respiratory
apnea. Therefore, the alarm of respiratory apnea should not be used for the
diagnosis of the patient.
15-8
S12 Vet Monitor User’s Manual
Temperature, water vapor in the patient’s breath, barometric pressure, and the
proportions of O2, N2O and Helium in the mixture all influence CO2 absorption.
For mainstream and sidestream/microflow CO2 module, you can set the CO2
correction by following the steps below:
1. Select the CO2 parameter area or waveform area to enter the【CO2】menu.
2. Select【Setup】submenu.
3. Please set the following contents according to the actual situation before
correction:
——【Gas Temperature】: Set the temperature of gas.
——【Barometric Pressure】: Set the atmospheric pressure.
The system default barometric pressure is 760mmHg, you can select【Main Menu】
quick key→from【System】column to select【Maintenance】→in【Other】submenu to
enter the barometric pressure value of the environment.
——【Zero Gas Type】: Select the gas type of zeroing, the options are【Zero On
Room Air】or【Zero On N2】.
15-9
S12 Vet Monitor User’s Manual
WARNING:
Please set the CO2 corrections according to actual situation, otherwise, the
measured value may be inaccurate and away from actual value.
You can select the operating mode of the CO2 module according to the actual
situation of the module. The operation steps are as follows:
1. Select the CO2 parameter area or waveform area to enter the【CO2】menu.
2. Select【Setup】submenu.
3. Set【Operation Mode】.
Measure: Select measurement mode is required when measuring with the CO2
module.
15-10
S12 Vet Monitor User’s Manual
Sleep: When not using CO2 for monitoring, it is recommended to set the CO2
module to sleep mode to increase the life of the CO2 module.
When performing intubation during general anesthesia, you can enter the intubation
mode in order to reduce unnecessary alarms. To enter the intubation mode, follow this
procedure:
1. Select the CO2 parameter area or waveform area to enter the【CO2】menu.
2. Select【Intubation Status】button.
For the details of the intubation mode, see 7.11 Intubation Status.
15.8. Zeroing
While zeroing is recommended the first time a CO2 sensor is connected to the monitor,
it is only absolutely necessary when the message “Zero Required” is displayed.
1. Ensure that the catheter or airway adapter is not connected to the patient or
close to any source of CO2 (including the patient's, your own exhaled breath
and ventilator exhaust valves).
The screen prompts【CO2 Zero In Progress…】, and the message disappears after the
zeroing is completed.
For Nomoline ISA CO2, the sampling tube should be pulled out for automatic zero
when needed.
15-11
S12 Vet Monitor User’s Manual
CAUTION:
Before zeroing, the BLT sidestream/ microflow CO2 sensor must be connected
with the sampling tube.
Before zeroing, the mainstream CO2 sensor must be connected with the airway
adapter.
Zeroing should not be performed for 20 seconds after the airway adapter or
cannula is separated from the patient's airway. Wait a moment before zero
correction to dissipate the remaining CO2 in the adapter or cannula.
Zeroing should not be performed when the airway adapter or cannula is
connected to the patient's trachea.
When the temperature is not stable, please do not adjust to zero.
When CO2 remains in the airway adapter or casing, zeroing will result in
inaccurate measurement or other error conditions. The time required for zeroing
will also increase.
When zeroing, do not rely on gas readings.
Nomoline ISA CO2 module does not require user to manually zeroing, and the
user cannot successfully send the zeroing command to the module. When
necessary, the sampling tube should be pulled out for automatic zero.
15.9. Calibration
The monitor has already been calibrated before leaving factory. User can directly
apply it measuring in normal conditions (except for the following three cases). When the
following three conditions occur, please return the CO2 module to the factory for
calibration.
After the CO2 module is used for half a year and one year;
Clinicians doubt the accuracy of readings;
After the latest calibration, atmospheric pressure or height above sea level
varies evidently.
15-12
S12 Vet Monitor User’s Manual
WARNING:
When using the sidestream/microflow CO2 measurement on patients who are
receiving or have recently received anesthetics, connect the outlet to a scavenging
system, or to the anesthesia machine/ventilator, to avoid exposing medical staff to
anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the
outlet connector of sensor.
15.11. Announcements
WARNING:
Do not position the sensor cables or tubing in any manner that may cause
entanglement or strangulation.
Reuse, disassembly, cleaning, disinfecting or sterilizing the single patient use
cannula kits and on-airway adapters may compromise functionality and system
performance leading to a user or patient hazard. Performance is not guaranteed
if an item labeled as single patient use is reused.
Before using, please check whether the airway adapter is damaged. Do not use if
damage is found.
If excessive secretions are found on the airway adapter, replace them
immediately.
When monitoring CO2 waveform, if changes or abnormal phenomena are found,
please check the airway adapter or sampling tube. If necessary, please replace it
immediately.
Note whether the baseline of CO2 waveform is too high. Sensor or patient
problems will cause the baseline to be too high.
Regularly check CO2 sensor and pipeline for excessive moisture or secretion
accumulation.
Do not operate the CO2 module when it is wet or has exterior condensation.
Do not use microflow CO2 sensors for patients who cannot tolerate the
15-13
S12 Vet Monitor User’s Manual
CAUTION:
Use only accessories provided by manufacturer.
Do not sterilize or immerse the CO2 sensor in liquids.
Clean the CO2 sensor and accessories as directed in this manual.
Do not apply excessive tension to the CO2 sensor cable.
When aerosol drugs are present, please keep the airway adapter away from the
breathing circuit. The viscous substance of the aerosol drug can pollute the
window of the airway adapter, and the airway adapter needs to be cleaned or
replaced in advance.
For further information on the use of Masimo IRMA CO2 mainstream module
and Nomoline ISA CO2 module, please refer to the user’s manual included with
the module.
NOTE:
15-14
S12 Vet Monitor User’s Manual
This chapter describes the problems that may be encountered during the use of the
monitor. You can first refer to the following table to eliminate them. If the problem
persists, please contact the maintenance personnel.
CAUTION:
For the physiological and technical alarm messages, see D Alarm Message.
Problem Solution
EtCO2 measurement value 1. Judge whether the CO2 concentration in the use
too low environment is too high. If the environmental
concentration is too high, the measured value is too
low. If it is more serious, zero will fail. Please pay
attention to the ventilation of the environment at
this time.
2. Check the sampling tube and connectors for leakage.
3. Check the patient status.
Problem Solution
15-15
S12 Vet Monitor User’s Manual
CAUTION:
DM only measures the number of drops in the set infusion tube assembly and
does not participate in drop control.
WARNING:
During measurement, the liquid level in the drip chamber should be kept below
the liquid level indicator line of the infusion monitoring module.
Please make sure that the outside of the drip chamber is not sticky with water,
otherwise the dripping rate measurement may be inaccurate.
The operator should pay attention to the length of the infusion tube and use the
extension tube when necessary to avoid accidents caused by pulling infusion tube
due to the patients turning over.
The Drip monitor (DM) measurement function isn’t intended to measure the
drips rate in the infusion process of analgesic, chemotherapy medicine and
insulinum.
16-1
S12 Vet Monitor User’s Manual
CAUTION:
In order to ensure the accuracy of dripping rate measurement, the drip monitor
module should be vertically installed or naturally hung on the infusion stand
using matching bracket.
This function is an assistive technology implementation method designed for the
high-quality infusion nursing services, and cannot replace manual monitoring
and speed control operations during infusion.
This function is suitable for working under relatively static conditions.
Therefore, avoid using it in a moving state, and avoid shaking and tilting at a
large angle. When the water mist and small water drops in the dropper are
seriously hung on the wall, it may interfere with the detection. If necessary, you
can flick the wall of the dropper with your finger to shake off the small wate
drops.
The DM module uses infrared sensing to detect, so it should be avoided in strong
light environment.
16.3. DM measurement
Insert the DM module cable into the DM socket on the monitor, and the drip monitor
interface will be displayed on the monitor.
Close the flow clamp tightly and connect the infusion tube set with the infusion
container, then squeeze the drip chamber and pour the liquid medicine to the 1/2 position
of the drip chamber. Open the flow clamp, fill the liquid medicine to the tip needle, and
then close the flow clamp tightly.
16-2
S12 Vet Monitor User’s Manual
Push the drip chamber into the slot of the drip chamber of the DM module, and clamp
the pipeline which is connected to the lower part of the drip chamber into the clamping
groove of DM module, as shown in the figure. The DM module is fixed to a suitable
position by means of supporting brackets or hanging ropes. Then, exhaust the pipeline to
ensure that the gas in the pipe set is exhausted and close the flow clamp tightly.
5
1
3
4
2
WARNING:
Ventilating operation can only be performed when the infusion is not performed
and the infusion tube is not connected to the patient.
If the unit of mL/h needs to be adopted, can switch “Drops/min” to “mL/h” in the
menu of the monitor and set the conversion parameter between the number of drops and
mL.
CAUTION:
The drip rate corresponding to the 1 mL infusion volume must be entered
16-3
S12 Vet Monitor User’s Manual
according to the relevant statement of the infusion set used. For example, the
Double-Dove tube declares that 20 drops of distilled water are equivalent to 1
mL±0.1 mL, so enter: 20 in the Drops/mL parameter setup.
Connect the infusion tube to the patient, start drip monitor (DM) measurement
through the“Start/Stop” button on the DM module, and adjust to the desired drip rate via
the flow clamp. The DM module indicator light switches from yellow to green and blinks
synchronously with the dropping of liquid drops.
During the infusion or after the infusion is completed, press the “Start/Stop” button
of the DM module, and the indicator light of the module will switch to a yellow state. At
this time, the monitor will exit the DM function and will no longer perform drip monitor
measurement.
WARNING:
In the non-infusion drip monitor state, when infusion is completed, the monitor
will not stop the infusion and send out an infusion completion alarm. Just quit
DM function, but the infusion is still continuing. If the infusion needs to be
stopped, the liquid stop clamp of the tube set needs to be operated to stop the
infusion.
16-4
S12 Vet Monitor User’s Manual
If you need to replace the liquid medicine container, please follow the steps as
below:
1) When the liquid stop clamp of the DM module is closed, remove the liquid
medicine container from the infusion pipeline;
2) Connect the infusion pipeline with a new liquid medicine container;
3) Open the liquid stop clamp through the “Reset” button on the DM module,
and then press the “Start/Stop” button to continue drip monitoring.
Status Indicator
Drip monitoring The green light is always on and flickers with
drops of liquid.
Suspension or stop of infusion Yellow is always bright.
Infusion completed and stopped Red light flashing (2Hz)
16.5. DM display
2
1
6
4 5
(1) Parameter Label
(2) Drip rate unit (main unit)
(3) The value of drip rate
(4) The value of flow rate
(5) Flow rate unit
(6) Working state diagram
The green dot blinks during drip monitoring, and when when drip monitor is
stopped, the yellow dot light without flashing; when the infusion is completed, the whole
symbol light white and red alternately.
16-5
S12 Vet Monitor User’s Manual
16.6. Setting DM
You can set the display of the DM main unit by the following steps as below:
1. Select the DM parameter area to enter 【DM】 menu.
2. Set 【Unit】of DM. The selected unit is displayed in the DM parameter area in
the form of a main unit.
In order to ensure the accuracy of the flow rate, you need to set the conversion
parameter between the number of drops and mL. The steps are as below:
1. Select the DM parameter area to enter 【DM】 menu.
2. Set 【Drop Per Milliliter】 of DM. The default value is 20.
16.7. DM check
The DM module has already been verified before leaving factory. Generally, the user
can directly measure it. Please contact the manufacturer or maintenance personnel
authorized by the hospital to repair the DM module when the following two situations
occur:
——The maintenance test is unqualified;
——Clinicians doubt the accuracy of readings.
16-6
S12 Vet Monitor User’s Manual
Chapter 17 Review
17.1. Review Overview
You can know how the patient's condition is developing through reviewing interface
to check the trend data, events, waveforms, and so on. You can also view the trend data
through the OxyCRG screen to know the changes in the patient’s condition.
CAUTION:
Changing the date and time will affect the storage of trends and events and may
result in data loss.
The review page contains graphic trends and tabular. The review page where each
submenu is located displays patient trend data in different forms.
The review pages have similar structure. We take the graphic trends review page as
an example. These contents will not be introduced in each review page.
17-1
S12 Vet Monitor User’s Manual
1
2 5
3 7
9
(1) Event type indicator: Different color blocks match different types of events.
• Red: high priority alarm event
• Yellow: medium priority alarm event
• Cyan: low priority alarm event
(2) Current window time line: indicates the time length of the current window.
(3) Waveform area: display trend curves. The color of trend curves is consistent with
the color of parameter labels.
(4) Time line:
can be moved within this time length.
Different color blocks at the time line indicate alarm events of different
types. The color of the color block is consistent with the color of the
event identifier.
(5) Cursor time
(6) Cursor
(7) Waveform area: displays the parameter value at the cursor time.
(8) Button area
(9) Slider: indicates the position of current window time in the entire time length.
Dragging this button left or right enables you to locate the trend data at a specific
time and also refreshes trend data in current window accordingly.
17-2
S12 Vet Monitor User’s Manual
Event list: displays events in a chronological order. The most recent event is
displayed at the top. The number of asterisk symbols (*) before an event an
event matches alarm priority.
In review page, the user can browse trend data in one of the following ways:
Move the slider .
Move the cursor.
Slide page.
Select to open the event list. You can select the event you want to view
17-3
S12 Vet Monitor User’s Manual
The trend table review page displays how the patient's physiological parameter trend
is developing in a tabular manner.
Choose one of the following methods to enter the tabular trends review page:
Select【Review】quick key→Select【Trend Table】submenu.
Select【Main Menu】quick key→from【Review】column to select【Tabular
Trends】.
The trend group defines the trend data displayed on the tabular trends review page. If
you have selected a 【Trend Group】 other than 【All】 and 【Standard】, you can
edit the trend group. To do so, follow this procedure:
1. Enter the tabular trends review page by either of the following ways:
Select【Review】quick key→Select【Trend Table】submenu.
Select【Main Menu】quick key→from【Review】column to select
【Tabular Trends】.
2. Select【Trend Group】 button.
3. Select the trend group submenu to edit.
Add parameters: select desired parameters from the 【Choices】 column
17-4
S12 Vet Monitor User’s Manual
CAUTION:
When 【Trend Group】 is set to 【All】 or 【Standard】, you cannot edit the
trend group.
ECG parameter and waveform are always displayed in the first row on the
trend page. It cannot be deleted or moved.
17-5
S12 Vet Monitor User’s Manual
The graphic trends review page displays trend data in a graphic form.
Choose one of the following methods to enter the graphic trends review page:
Select【Review】quick key→Select【Trend Graph】submenu.
Select【Main Menu】quick key→from【Review】column to select【Graphic
Trends】.
To set the length of time for each screen to display data as follows:
1. Enter the graphic trends review page.
2. Select 【…】 to enter setup menu.
3. Select 【Window Time】.
【8min、30min】:Each screen displays trend data for the set time, and
you can observe the trend in the last 6 hours.
【1h、2h、4h】:Each screen displays trend data for the set time, and you
can observe the trend in the last 180 hours.
17-6
S12 Vet Monitor User’s Manual
Follow these steps to select the number of waveforms to display in the graphic
trends:
1. Enter the graphic trends review page.
2. Select 【…】 to enter setup menu.
3. Select 【Wave Number】.
The monitor stores alarm events and system events in real time. Alarm event types
include physiological alarm event. When an alarm event occurs, the monitor will store the
values of relevant parameters at the time of occurrence and the relevant waveforms for 16
seconds before and after the time of occurrence.
CAUTION:
A sudden loss of power has no impact on the events stored.
Choose one of the following methods to enter the events review page:
Select【Review】quick key→Select【Event】submenu.
Select【Main Menu】quick key→from【Review】column to select【Event】.
The event review page displays a list of events in the order in which they occurred.
The most recent event is displayed at the top. The number of asterisk symbols (*) before
an event an event matches alarm priority.
The event identifier on the left side of the alarm event displays different types of
events with different color blocks:
Red: high priority alarm event
Yellow: medium priority alarm event
Cyan: low priority alarm event
The number of currently events and the total number of filtered events are displayed
17-7
S12 Vet Monitor User’s Manual
at the top right corner of the event list. For example, 3/10 indicates that there are a total of
10 selected events, and currently there are 3 events.
You can filter events by time, alarm priority, parameter category or event type. To
configure the filter, follow this procedure:
1. Enter events review page to switch on 【Filter】.
2. Select【Filter Setup】and set the desired filter criterion. Events after filtering
will be displayed in the event list.
CAUTION:
If 【Filter】 is not switch on, the relevant setting in 【Filter Setup】 will not take
effect.
To view waveforms and parameter values at the selected event time, follow this
procedure:
1. Enter the event review page.
2. Select 【Detail】.
CAUTION:
Please ensure that the best ECG lead with largest waveform amplitude and the
highest signal-to-noise ration is selected. Choosing the best ECG lead is very
important to recognize cardiac beat, classify cardiac beat and recognize
ventricular fibrillation.
On the Full Disclosure review page, you can review waveform data up to 72 hours.
You can view compressed waveforms, full waveforms and numeric values.
17-8
S12 Vet Monitor User’s Manual
Choose one of the following methods to enter the Full Disclosure review page:
Select【Review】quick key→Select【Full Disclosure】submenu.
Select【Main Menu】quick key→form 【Review】 column select 【Full
Disclosure】.
To review compressed waveforms, you must first select which parameters to store
and display. Follow these steps:
1. Enter the Full Disclosure review page.
2. Select 【Setup】 submenu to enter 【Full Disclosure Setup】 page.
3. Select 【Storage】 submenu and select the desired waveforms to be stored.
4. Select 【Display (Maximum: 3)】 submenu and select the desired waveform
to be displayed from the stored waveforms.
CAUTION:
If more waveforms are selected in the 【Storage】 column, the storage time of
these waveforms will be shortened due to the limitation of memory size. The
waveforms may not be stored for 72 hours. Please exert caution when selecting
waveforms.
When an alarm occurs, the band on the compressed waveform at the alarm time
will use different shading to indicate different alarm levels.
Red: high priority alarm
Yellow: med priority alarm
Cyan: low alarm priority
To set the length of time each compressed waveform is displayed and the ECG
17-9
S12 Vet Monitor User’s Manual
To view the full waveforms and numeric values of compressed waveforms, follow
this procedure:
1. Enter the holographic waveform review page.
2. Select 【Details】. You can perform the following operations on this page:
Select 【…】 to set 【Waveform Speed】, 【Record】 and 【Gain】.
Select 【Overview】 to return to the compressed waveform page.
You can review up to 48 hours’ trend curves and compressed waveforms on the
OxyCRG review page.
Choose one of the following methods to enter the OxyCRG Review Page:
Select 【Review】 button on the OxyCRG Review Page.
Select【Review】quick key→Select【OxyCRG】submenu.
Select 【 Main Menu 】 quick key→enter 【 Review 】 column →Select
【OxyCRG】.
17-10
S12 Vet Monitor User’s Manual
2. Set 【Zoom】.
17-11
S12 Vet Monitor User’s Manual
Chapter 18 Calculations
18.1. Introduction
The monitor provides calculation functions. The calculated values, which are not
directly measured, are computed based on the data and measurement values you enter.
The calculation is independent of other monitoring functions and the object of
calculation may not be the patient monitored by this monitor. The calculation operation
will not affect the patients being monitored.
WARNING:
The dosage of drugs must be decided by the physician in charge.
During calculation, check that the entered values are correct and the calculated
values are appropriate. We assume no responsibility for any consequences
caused by wrong entries and improper operations.
18-1
S12 Vet Monitor User’s Manual
Select【Calculations】quick key.
Select【Main Menu】quick key→from【Calculations】column to select
【Drug】.
2. Set 【Drug Name】 and 【Patient Type】. If the selected drug is affected by
weight, switch on 【Weight Participation】 and enter the patient’s weight.
3. Enter the drug-related information such as total amount, volume and dose of
drugs.
4. Select 【Calculate】 button to calculate. Red arrow marks are displayed
before the calculation results.
CAUTION:
If available, the patient category and weight from the patient demographics
menu are automatically entered when you first access drug calculation. You can
change the patient category and weight. This will not change the patient category
and weight stored in the patient demographic information.
The Titration Table shows informations on the currently used drugs. You can
check the dose received by the patient at different infusion rate through the Titration
Table. The procedure for viewing the titration table is as follows:
1. Access drug calculation page by either of the following ways:
Select【Calculations】quick key.
Select【Main Menu】quick key→from 【Calculations】 column to
select 【Drug】.
2. Select 【Titration】 sub menu.
3. Select 【Dose Type】 at the bottom of the interface to set the unit type of drug
dose in the titration table.
4. Select 【Step】 to set the interval between two adjacent titration table item.
You can choose the sorting method of titration table:
【Dose】: The titration table is listed in the sequence of increased drug
18-2
S12 Vet Monitor User’s Manual
dose.
【INF Rate】: The titration table is listed in the sequence of increased
infusion rate.
g series:mcg, mg, g;
Dose/hr unit series: unit, kU, MU; Dose/h=Dose/min×60
mEq series: mEq;
g series:mcg, mg, g;
Dose/min unit series: unit, kU, MU; DoseMin=DoseMin
mEq series: mEq;
18-3
S12 Vet Monitor User’s Manual
The monitor provides the hemodynamic calculation function. The monitor can
save the results of up to 20 calculations, which are displayed in groups.
2. Enter the correct value for each parameter. For a patient who is being
monitored, the currently measured values are automatically taken, and the
height and weight are derived from the patient information entered.
18-4
S12 Vet Monitor User’s Manual
Output
Unit Full Name Formula
Parameter
C.I. mL/min/m2 cardiac index
body surface BSA=HT0.725×WT0.425×
BSA m2
area 0.007184
SV mL stroke volume SV = 1000× C.O. /HR
2
SVI mL/m stroke index SVI= SV/BSA
systemic
5
SVR dyn*s/cm vascular
resistance
systemic
dyn*s*m2/c
SVRI vascular SVRI = SVI /BSA
m5
resistance index
pulmonary
PVR dyn*s /cm5
vascular
18-5
S12 Vet Monitor User’s Manual
Output
Unit Full Name Formula
Parameter
resistance
pulmonary
dyn*s*m2/c
PVRI vascular PVRI= PVR×BSA
m5
resistance index
LCW kg*m left cardiac work LCW= 0.0136 × APMAP × C.O.
left cardiac work
LCWI kg*m/m2 LCWI = RCW×BSA
index
left
LVSW g*m ventricularstrok LVSW = 0.0136 ×MAP×SV
e work
left ventricular
2
LVSWI g*m/m stroke work LVSWI = LVSW/BSA
index
right cardiac
RCW kg*m
work
right cardiac
RCWI kg*m/m2 RCWI= RCW/BSA
work index
right ventricular
RVSW g*m RVSW = 0.0136 × MPAP× SV
stroke work
right ventricular
2
RVSWI g*m/m stroke work R VSWI= RVSW /BSA
index
EF % ejection fraction EF= 100 × SV /EDV
The monitor provides the oxygenation calculation function. The monitor can save
the results of up to 20 calculations, which are displayed in groups:
18-6
S12 Vet Monitor User’s Manual
18-7
S12 Vet Monitor User’s Manual
g/L, g/dL,
Hb hemoglobin
mmol/L
mL/dL,
CaO2 arterial oxygen content
mL/L
mL/dL, venous oxygen content
CvO2
mL/L
RQ Respiratory quotient
Output
Unit Full Name Formula
Parameters
BSA=HT0.725×WT0.425×
BSA m2 body surface area
0.007184
arteriovenous
mL/L,
C(a-v)O2 oxygen content C(a-v)O2=CaO2-CvO2
mL/dL
difference
oxygen extraction
O2ER % O2ER=(CaO2-CvO2)/ CaO2
ratio
DO2 mL/min oxygen transport DO2=C.O.×CaO2
PAO2=【FiO2×(ATMP-water
partial pressure of
mmHg, pressure) 】-(PaCO2×1.25)
PAO2 oxygen in the
kPa Wherein the water pressure is
alveolar
selected to be 47mmHg (6.3kPa)
18-8
S12 Vet Monitor User’s Manual
Output
Unit Full Name Formula
Parameters
mmHg, alveolar-arterial
AaDO2 AaDO2=PAO2-PaO2
kPa oxygen difference
mL/L, capillary oxygen CcO2 = Hb ×1.34
CcO2
mL/dL content +0.031×PAO2
Qs/Qt= (CcO2-CaO2)/( CcO2
Qs/Qt % venousad mixture
-CvO2)
The monitor provides the ventilation calculation function. The monitor can save the
results of up to 20 calculations, which are displayed in groups.
18-9
S12 Vet Monitor User’s Manual
TV mL tidal volume
RQ respiratory quotient
Output
Unit Full Name Formula
Parameter
PAO2=【FiO2×(ATMP-water
partial pressure of
pressure) 】-(PaCO2×1.25)
PAO2 mmHg,kPa oxygen in the
Wherein the water pressure is
alveolar
selected to be 47mmHg (6.3kPa)
alveolar-arterial
AaDO2 mmHg,kPa AaDO2=PAO2-PaO2
oxygen difference
Pa/FiO2 mmHg,kPa oxygenation ratio Pa/FiO2= PaO2/FiO2
arterial to alveolar
a/AO2 % a/AO2= (100 ×PaO2)/PAO2
oxygen ratio
MV L/min minute volume MV=(TV/1000)×RR
18-10
S12 Vet Monitor User’s Manual
Output
Unit Full Name Formula
Parameter
volume of
Vd=【( PaCO2-PeCO2) ×TV】
Vd mL physiological dead
/PaCO2
space
physiologic dead
Vd/Vt= (PaCO2-PeCO2)/
Vd/Vt % space in percent of
PaCO2×100%
tidal volume
The monitor provides the nephridium calculation function. The monitor can save the
results of up to 20 calculations, which are displayed in groups.
Select【Calculations】quick key→【Nephridium】submenu.
2. Enter the correct value for each parameter. For a patient who is being
monitored, the currently measured values are automatically taken, and the
height and weight are derived from the patient information entered.
3. Select 【Calculate】 to calculate the value of each output parameter. The
calculated value is greater than the normal upper limit is indicated by an up
arrow “↑”; the calculated value is lower than the normal lower limit is
indicated by a down arrow “↓”.
Select 【Range】 to show the normal range of each parameter.
Select 【Unit】 to show the unit of each parameter.
18-11
S12 Vet Monitor User’s Manual
Cr umol/L creatinine
Output
Unit Full Name Formula
Parameter
Cosm=Uosm×(Urine/24/60)/
Cosm mL/min osmolar clearance
Posm
18-12
S12 Vet Monitor User’s Manual
Output
Unit Full Name Formula
Parameter
free water CH20=Urine/24×(1-
CH2O mL/h
clearance Uosm/Posm)
sodium potassium
Na/K % Na/K=URNaEx/URKEx
ratio
urine to plasma
U/Posm U/Posm =Uosm/Posm
osmolality ratio
blood ureanitrogen
BUN/Cr BUN/Cr
creatinine ratio
clearance of CNa(mL/24hrs)=URNa)×Ur
CNa mL/24h
sodium ine/SerNa
urine-serum
U/Cr UCr/Cr
creatinine ratio
fractional excretion FENa%=(URNa×Cr)/(SerNa
FENa %
of sodium ×Ucr) ×100%
18-13
S12 Vet Monitor User’s Manual
Chapter 19 Recording
19.1. Recorder
This monitor uses the thermal recorder which supports various record types. It can
output the patient information, measurement data, review data and three waveforms at
best.
1
2
3
The records can be divided into the following types according to trigger modes:
1. Real-time record of manual startup;
2. The circular record of automatic strartup of the recording meter in line with
the given time interval.
3. The alarm record triggered by out-of-limit parameter and so on.
4. Record started by manual operation and related to special function.
19-1
S12 Vet Monitor User’s Manual
You can start recording by manual way through the following means:
Press 【Real-time Record】 quick key below the monitor interface to start
real-time recording.
Select 【Record】 button in the current window or above the menu to start the
associated record of the special function.
You can stop recording by manual way through the following means:
In the process of real-time recording, click the 【Real-time Record】 quick
key.
When the printing of the record report is finished, there are the following flags:
For automatically stopped recordings: Print “***END***” at the end of the
report.
For manually or abnormally stopped recordings: There is no flag printing at
the end of the report.
19-2
S12 Vet Monitor User’s Manual
This section describes the definition of the main setting items. Users can refer to
these definitions to select other similar setting items in the device according to their
needs.
When starting a real-time recording, the length of recording depends on your setting
of recording duration.
1. Open the 【Record Setup】 menu.
2. Set 【Record Duration】 to:
【8s】: Record the waveform of 4 seconds before and after the current
time.
【Continue】: Record the waveform 5 seconds before and after the
current time until you manually stop recording.
You can set a certain time interval, and the recorder automatically starts recording
according to the set time interval.
1. Open the 【Record Setup】 menu.
2. Set 【Cycle Record Interval】.
3. After the setting is completed, the recorder starts each recording at the set
interval.
19-3
S12 Vet Monitor User’s Manual
You can set the duration of every period recording in the following ways:
【8s】: Record the waveform of 4 seconds before and after the current
time.
You can set how long the waveform needs to be recorded when an alarm occurs, as
follows:
You can set to record the output NIBP measurement results when NIBP
measurement is completed, as follows:
1. Open the 【Record Setup】 menu.
2. Set 【NIBP Trigger】 to ON or OFF.
If the record paper runs out, please install the record paper as the following step:
1. Press both sides of the recorder door with one hand and pull outwards to open
19-4
S12 Vet Monitor User’s Manual
CAUTION:
Must use the thermo-sensitive paper that meets requirements; otherwise, it will
lead to recording failure, bad-quality record or damage of thermo-sensitive
printing head.
Do not pull out the recording paper during recorder printing, otherwise the
recording meter may be damaged.
Unless for paper replacement or fault remedy, don’t keep the recorder door
open.
After long-time service, some paper scrap and impurity will accumulate on the
printing head, and affect printing quality as well as the service life of printing head and
roll shaft. The recorder can be cleaned according to the following methods:
1. Before cleaning, the measures such as wearing anti-static wrist strap shall
be adopted to avoid the damage to recording meter resulting from static;
19-5
S12 Vet Monitor User’s Manual
CAUTION:
Don’t use any article that can damage the thermo-sensitive parts of recorder
during cleaning.
Don’t heavily press the printing head of recorder.
19-6
S12 Vet Monitor User’s Manual
The monitor has an auxiliary output port that can provide “analog signal output”.
Connect the monitor to equipment such as an oscillograph, and then do some associated
setup, after that you can output the analog signal to the oscillograph through the port.
The setting ways of analog signal output are as below:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Module】submenu→【Auxiliary Output】submenu.
3. Select 【Analog Output】, set the analog output signal as required.
CAUTION:
Analog output function is seldom used in clinic. If you need t know more
detailed information, please contact the service personnel.
20-1
S12 Vet Monitor User’s Manual
The steps for setting the wireless network frequency and antenna type are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select 【Network Setup】 submenu → 【WLAN】 submenu.
3. Set the 【Frequency】 and 【Antenna】 of the wireless network according
to the usage.
【Frequency】:【5G】 or 【2.4G】.
20-2
S12 Vet Monitor User’s Manual
The monitor can be connected to the central monitoring system via wired network or
wireless network.
The device number of the networked monitor will be displayed when the central
monitoring system and other beds are monitored. The steps for setting the device
number of the monitor are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select 【Network Setup】 submenu → 【CMS】 submenu.
3. Set 【Device No.】 of the monitor.
Please refer to the Central Monitoring System User’s Manual for detailed
instructions.
NOTE: This monitor can only be connected to the central monitoring system
20-3
S12 Vet Monitor User’s Manual
The real-time data, waveforms, and alarms of the monitor can be transmitted to
the hospital’s monitoring system through the HL7 protocol. The operating steps are as
follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
20-4
S12 Vet Monitor User’s Manual
Chapter 21 Battery
21.1. Introduction
The monitor can be fitted with rechargeable battery to ensure the normal use of
the monitor in case of intra-hospital patient transfer or whenever the power supply is
interrupted. When the monitor is switched on with AC power, the battery can be
charged regardless of whether the monitor is switched on or not. Since we do not
provide external charging equipment, the battery can only be charged in the monitor.
In case of sudden power failure, the system will automatically use battery to supply
power to the monitor, thus not causing interruption of monitoring work.
On-screen battery symbols indicate the battery status as follows:
Indicate that the battery works correctly. The solid portion represents the
battery’s charge.
The power supply of battery can only function for a certain period. Excessively
low voltage of battery will trigger a high priority technical alarm 【Battery Low】. At
this moment, the monitor shall immediately connect with alternating current power
supply to charge the battery.
In case of long-term monitoring, a backup battery shall be installed and used after
the AC power is plugged in. The AC power plug must be plugged into the special
interface of the hospital.
The battery of this monitor must be installed and replaced by maintenance personnel
trained and authorized by our company.
The service life of the battery depends on the frequency and time of use. If
21-1
S12 Vet Monitor User’s Manual
lithium batteries are properly maintained and stored, their service life is about 3 years.
If batteries are used improperly, their life may be shorter. We recommend replacing
lithium batteries every 3 years.
In order to ensure the maximum capacity of the battery, please pay attention to
the following instructions:
The battery performance must be checked every two months. Before the
monitor is repaired or when you suspect that the battery is the source of the
fault, battery performance inspection is also required.
When the battery is used or stored for three months or when the running time of
the battery is significantly shortened, the battery performance is optimized
once.
If the monitor is not used for a long time, please optimize the battery
performance every three months. Because not taking out the battery will
shorten the battery life.
If the lithium battery is put on hold when its charge is 50% of its full charge,
the storage life of the lithium battery is about 6 months. After 6 months, the
lithium battery must be used up before being charged to full capacity. The
monitor is powered by the lithium battery, and the battery is taken out of the
monitor and then put on hold when the battery is 50% of the full charge.
WARNING:
Keep the battery out of the reach of children.
Use only batteries specified in the manufacturer.
If the battery shows signs of damage or signs of leakage, replace it immediately.
Do not use a faulty battery in the monitor.
A battery should be optimized before it is used for the first time. A battery
21-2
S12 Vet Monitor User’s Manual
CAUTION:
Over time and with the use of batteries, the actual storage capacity of batteries
will decrease. For old batteries, the full capacity icon does not mean that the
battery storage capacity can still meet the manufacturer’s specifications, nor
does it mean that the battery power supply time can still meet the
manufacturer’s specifications. During optimization, if the battery power
supply time is obviously shortened, please replace the battery.
The performance of a battery may deteriorate over time. To check the performance
of a battery, follow this procedure:
1. Disconnect the monitor from the patient and stop all monitoring and
measuring procedures.
2. Connect the monitor to AC power and charge the battery continuously until
the battery is full.
3. Disconnect AC mains and allow the monitor to run on the battery until it shuts
off.
4. The operating time of a battery reflects its performance directly. If the power
21-3
S12 Vet Monitor User’s Manual
supply time of the battery is obviously lower than the time stated in the
specification, please consider replacing the battery or contact maintenance
personnel.
CAUTION:
If the power supply time is too short after the battery is fully charged, the
battery may have been damaged or malfunctioned. The power supply time of
the battery depends on the equipment configuration and operation. For
example, frequent NIBP measurement will also shorten the power supply time
of the battery.
When a battery has visual signs of damage, or no longer holds a charge, it should be
replaced. Removed the old battery from the monitor and recycle it properly. To dispose of
the batteries, follow local laws for proper disposal.
WARNING:
Do not disassemble batteries, or dispose of them in fire, or cause them to short
circuit. They may ignite, explode, or leak, causing personal injury.
21-4
S12 Vet Monitor User’s Manual
22.1. Introduction
Keep your equipment and accessories free of the dust and dirt. To avoid damage to
the equipment, follow these rules:
Please dilute the detergent and disinfectant according to the manufacturer’s
instructions, or use as low a concentration as possible.
Do not immerse the monitor in liquid.
Do not pour liquid on the monitor or accessories.
Do not allow liquid to enter the cabinet.
Abrasive materials (such as steel wool or silver polishing agent) and any strong
solvent (such as acetone or detergent containing acetone) as well as liquids with
strong conductivity (such as physiological saline) shall not be used.
Please do not clean or disinfect the equipment when it is running or when it
is exposed to direct sunlight.
Ensure that all parts of the equipment are completely dry after cleaning and
disinfection.
WARNING:
Disconnect the power cord from the socket before cleaning the monitor.
CAUTION:
If you accidentally pour liquid on the monitor or accessories, please contact the
22-1
S12 Vet Monitor User’s Manual
Monitors should be cleaned regularly. If there is heavy pollution or lots of dust and
sand in your place, the monitor should be cleaned more frequently. Before cleaning the
monitor, consult the hospital’s regulations for cleaning the monitor.
Use a soft cloth that cannot bear balls, wet and clean it with an appropriate amount of
water or alcohol-based detergent (such as 70% ethanol). Do not use strong solvents such
as acetone or tichlorothylene. Be careful when cleaning the monitor’s screen, which is
more sensitive than the case. The interface and metal parts of the equipment should be
avoided. After cleaning, the equipment should be placed in a ventilated and cool
environment to dry.
CAUTION:
Interfaces and metal parts may be corroded after contacting with detergent.
22.3. Disinfection
You can disinfect the equipment according to the hospital’s disinfection procedures.
Clean the monitor before disinfection. The following table lists the recommended
disinfectants:
Name Type Manufacturer
Isopropyl alcohol, 70% Liquid -
Sodium hypochlorite, 0.5% Liquid -
Alcohol, 70% Liquid -
Hydrogen peroxide, 3% Liquid -
22-2
S12 Vet Monitor User’s Manual
CAUTION:
Do not immerse accessories in water or disinfectant.
Do not wet the pins of the accessories.
Frequent disinfection of accessories can cause damage to them. It is suggested
that according to hospital regulations, accessories should be disinfected only
when necessary.
When cleaning and disinfecting NIBP airpipe, liquid should be prevented from
entering the airpipe.
Use only the detergents and disinfectants specified in this manual.
Use a soft cloth that cannot bear balls, wet and clean the accessories with an
appropriate amount of water or alcohol-based detergent (such as 70% ethanol). After
cleaning, place the accessories in a cool and ventilated environment to dry.
You can disinfect the accessories of the monitor according to the disinfection
procedures of the hospital. Recommended disinfectants include:
22-3
S12 Vet Monitor User’s Manual
22.5. Sterilization
22-4
S12 Vet Monitor User’s Manual
Chapter 23 Maintenance
WARNING:
Failure on the part of the responsible individual hospital or institution
employing the use of this equipment to implement a satisfactory maintenance
schedule may cause undue equipment failure and possible health hazards.
The safety checks or maintenance involving any disassembly of the equipment
should be performed by professional servicing personnel. Otherwise, undue
equipment failure and possible health hazards could result.
If necessary, please contact the manufacturer for product circuit diagrams,
parts lists, calibration instructions or other equipment maintenance related
information.
If there is a problem with the monitor, please contact the maintenance
personnel or us.
23.1. Inspection
Before use, after continuous use for 6-12 months, maintenance or upgrade, qualified
maintenance personnel should conduct a comprehensive inspection to ensure the normal
operation and work of the monitor.
Items to be inspected shall include:
The environment and power supply meet the requirements.
There is no mechanical damage to the monitor and accessories.
The power cord has no abrasion and good insulation performance.
Use the specified accessories.
The alarm system functions normally.
The recorder works normally and the recording paper meets the specified
requirements.
The performance of the battery.
Various monitoring functions are in good working condition.
Grounding impedance and leakage current meet the requirements.
23-1
S12 Vet Monitor User’s Manual
If any damage or abnormal phenomenon is found, please do not use the monitor and
immediately contact the medical engineer of the hospital or the maintenance personnel of
the company.
The following tasks, except visual inspection, startup detection, touch screen
calibration, battery inspection and recorder inspection, can only be completed by
professional maintenance personnel. Please contact the maintenance personnel in time
when the following maintenance is required. Before testing or maintenance, the equipment
must be cleaned and disinfected.
NIBP Leakage detect 4. For other module, at least once every two
Pressure calibration
CO2 CO2 gas detect
Static pressure detect
IBP
Pressure calibration
DM Drip rate detect
Nurse call detect
Analog output test When you suspect that the function is not normal.
Defibrillation synchronization test
Electrical Safety Tests
Select test items based on IEC 1. After repairing or replacing the power
23-2
S12 Vet Monitor User’s Manual
60601-1 module.
2. Or after the monitor falls.
3. At least once every two years or as required.
Other Tests
Power-on test Before use.
Network print test 1. During the first installation.
2. After repairing or replacing the printer.
Recorder check 1. Before the first use.
2. After repairing or replacing the recorder.
1. During the first installation.
Functionality test
Battery 2. After replacing the battery
check Every two months or when the running time of
Performance test
the battery is significantly shortened.
After the equipment reaches its service life, please dispose of the monitor and its
accessories according to local regulations.
WARNING:
For the disposal of parts and accessories, if there is no corresponding
regulation, local regulations on disposal of hospital waste can be followed.
23-3
S12 Vet Monitor User’s Manual
Chapter 24 Accessories
WARNING:
Use only the accessories specified in this chapter. Use of other accessories may
damage the monitor or fail to meet the specifications claimed in this manual.
The accessories listed in this chapter must be used together with the
monitoring equipment of our company. The user has the responsibility to read
the operating instructions of the equipment (including accessories) or contact
us for consultation to confirm the matching between the accessories and the
equipment. Otherwise, it may cause injury to the patient.
Disposable accessories can only be used once. Repeated use may cause
performance degradation or cross infection.
Do not open the disposable or sterilized accessory package too early, so as not
to cause the accessory to fail or become contaminated.
CAUTION:
If the use or storage environment of accessories exceeds the specified
temperature or humidity range, the performance of accessories may not meet
the claimed specifications. If the performance of accessories is degraded due to
aging or environmental conditions, please contact customer service personnel.
If there are signs of damage to the package of the accessory or the accessory
itself, please do not use the accessory.
Do not use the accessory if it expires.
Disposable accessories must be handled in accordance with local regulations or
hospital systems.
NOTE:
For accessories with safe service life, see the package of accessories for service
24-1
S12 Vet Monitor User’s Manual
life.
Please refer to the package of accessories for sterilization accessories. If the
package of the accessories of the sterilization package is damaged, please do
not use it.
ECG
Accessories Specification Model / PN
3-lead,IEC,Snap (12PIN) 15-031-0013
5-lead,IEC,Snap (12PIN) 15-031-0002
5-lead,IEC,Clip (12PIN) 15-100-0370
6-lead,IEC,Snap (12PIN) 15-031-0051
12-lead,IEC,Snap (12PIN) 15-031-0001
5-lead,IEC,Clip (12PIN)(plug-in) 15-031-0022
ECG cable 3-lead,AHA,Snap (12PIN) 15-031-0014
5-lead,AHA,Snap (12PIN) 15-031-0004
6-lead,AHA,Snap (12PIN) 15-031-0050
12-lead,AHA,Snap (12PIN) 15-031-0003
3-lead,AHA,Snap (12PIN)(plug-in) 15-100-0196
5-lead,AHA,Clip (12PIN)(plug-in) 15-031-0021
5-lead,AHA,Snap (12PIN)(plug-in) 15-100-0197
Electrode with snap
Reusable 15-100-0077
clips
SpO2
Accessories Specification Model / PN
SpO2 sensor Ear clip SpO2 sensor 15-100-0325
Reusable ,Tongue clip (small) 15-100-0079
Tongue clip
Reusable ,Tongue clip (large) 15-100-0189
SpO2 Extension
Reusable 15-100-0357
cable
24-2
S12 Vet Monitor User’s Manual
TEMP
Accessories Specification Model / PN
Reusable, Surface 15-031-0005
TEMP Probe
Reusable, Coelom 15-031-0012
NIBP
Accessories Specification Model / PN
Disposable cuff,3-5.5CM 15-100-0164
Disposable cuff,4-8CM 15-100-0165
NIBP cuff Disposable cuff,6-11CM 15-100-0166
Disposable cuff,7-13CM 15-100-0167
Disposable cuff,9-14.5CM 15-100-0168
NIBP air pipe Reusable 15-031-0008
CO2
BLT Mainstream CO2
Accessories Model / PN
CO2 sensor 15-100-0199
Airway adapter 15-100-0212
Airway adapter 15-100-0213
24-3
S12 Vet Monitor User’s Manual
IBP
Accessories Model / PN
IBP sensor PT-1/15-100-0053
24-4
S12 Vet Monitor User’s Manual
Accessories Model / PN
IBP cable 15-100-0029
IBP extension cable (4PIN to 6PIN) 15-031-0023
IBP cable 15-100-0098
IBP sensor PT-01/15-100-0097
IBP cable 15-100-0136
IBP cable 15-100-0137
IBP cable 15-100-0143
DM module
Accessories Model / PN
DM module 16-100-0113
C.O.
Accessories Model / PN
C.O. interface cable (6PIN) 15-100-0148
Sensor cable SP4042
Sensor cap SP5045
Tricyclic syringe 15-100-0169/RFF:620161
24-5
S12 Vet Monitor User’s Manual
Mainframe I No mark
Fixed
parameter See Chapter
(ECG, 22
TEMP, Maintenanc
CF Continuou
RESP, NIBP, IP21 Not suitable e and
SpO2) NA s
Cleaning of
CO2
BF
DM
Note:
I: Class I, internally and externally powered equipment.
When you doubt about the protecting earth integrality or protecting earth lead of the
equipment, you’d better change the equipment to internally powered equipment.
CF: Type CF applied part.
BF: Type BF applied part.
NA: Not applicable
Not suitable: Equipment is not suitable for use in the presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
1
S12 Vet Monitor User’s Manual
Mainframe 0℃~40℃
Operating
temperature 5℃~40℃
CO2 module (external)
CAUTION:
The equipment must be used under the specified environmental specifications,
otherwise it will not meet the technical specifications claimed in this manual
and may lead to unexpected consequences such as equipment damage. If the
performance of the equipment changes due to aging or environmental
conditions, please contact the maintenance personnel.
2
S12 Vet Monitor User’s Manual
Frequency 50Hz/60Hz
Input power 100VA
Standard requirements According to IEC 60601-1 and IEC 60601-1-2
A.4.2 Battery
3
S12 Vet Monitor User’s Manual
A.5.1 Display
Host display
Type Color TFT LCD
S12 Vet Size(diagonal) 12.1 inch
A.5.2 Recorder
4
S12 Vet Monitor User’s Manual
Keys
Physical keys 1 power switch key
Touch screen Support
Others
Mouse input Support (optional)
Keyboard input Support (optional)
Barcode scanner Support (optional)
Voice assistant Support (optional)
A.5.6 Connectors
5
S12 Vet Monitor User’s Manual
6
S12 Vet Monitor User’s Manual
Conforming IEEE802.11a/b/g/n
standards
Operating 2.4GHz ~ 2.495 GHz, 5.15GHz ~ 5.35GHz, 5.47GHz ~
frequency
7
S12 Vet Monitor User’s Manual
A.8.1 ECG
A.8.1.1 Standard
8
S12 Vet Monitor User’s Manual
9
S12 Vet Monitor User’s Manual
10
S12 Vet Monitor User’s Manual
Measurement
HR, PVCs, ST, QT, and Arrhythmia analysis
parameter
Multi-lead Supports synchronous analysis of at least 2 leads, one of which
synchronous is the key monitoring lead and the other is the auxiliary lead. It
analysis function is on except 3-lead mode.
Automatic, Manual; Default manual
(3-lead mode is fixed as manual)
Smart Lead Switch Automatic mode: the algorithm automatically identifies the
current smart leads, and the host automatically switches the key
monitoring leads according to the identification of the algorithm.
Measurement
10~400 bpm
range
HR measurement Resolution 1 bpm
range and accuracy Accuracy ±1% or ±1 bpm, whichever is greater
Detection
0.20mVp-p
sensitivity
Display 12-lead ST segment values at the same time and support
ST Display
ST graphic display.
Measurement
-2.0mV~2.0mV
range
ST measurement Resolution 0.01mV
range and accuracy -0.8mV~0.8mV: ±0.02mV or ±10%,
Accuracy whichever is greater
Other: Unspecified
ST Update period 10s
11
S12 Vet Monitor User’s Manual
12
S12 Vet Monitor User’s Manual
A.8.2 RESP
Measurement
RR and respiration waveform
parameter
Source RA-LA, RA-LL (default)
64 kHz
Excitation waveform
Error: ≤±10%
Excitation current ≤0.3mA RMS
Respiration Apnea Fixed high priority alarm
Alarm Adjustable delay time: 10~60s, error ±3s or ±10%
Cardiac interference
Fixed high priority alarm
alert
13
S12 Vet Monitor User’s Manual
A.8.3 NIBP
A.8.3.1 Standard
Measurement
SYS, DIA, MAP,PR
parameters
Mode of operation Manual, Auto, STAT, Sequence
Intervals for periodic 1min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min,
measurement time 30 min, 1h, 1.5h, 2h, 3h, 4h, 8h
At least 5 groups are supported, and each group individually
Sequence mode
sets the interval and number of periodic measurement.
STAT mode cycle
5 min
time
Measurement range
0~300 mmHg
of cuff pressure
14
S12 Vet Monitor User’s Manual
Sensor calibration
One year (recommend)
time
Unit mmHg, kPa
Systolic 30~270 mmHg
Dynamic pressure
Diastolic 10~220 mmHg
measurement range
Mean 20~235 mmHg
Dynamic Pressure Systolic 35~255mmHg: ±8 mmHg
Measurement Error Diastolic 15~195mmHg: ±8 mmHg
of Simulator Mean 22~215mmHg: ±8 mmHg
Static pressure
±3 mmHg(±0.4kPa)
accuracy
Pressure resolution 1 mmHg or 0.1kPa
Measurement
PR measurement 40 ~ 240 bpm
range
range and accuracy
Accuracy ±3bpm or ±3%, whichever is greater
Maximum
<120s
measurement time
First overpressure
297±3 mmHg
protection point
Second overpressure
315±10 mmHg
protection point
Note: The accuracy of NIBP cannot be determined by using a simulator, but under many
conditions, it is still necessary to use a simulator to test its performance (for example, a
simulator is required for quality control in the production process), and the simulator of
the model specified by the manufacturer shall be used for this performance test.
A.8.4 SpO2
A.8.4.1 Standard
15
S12 Vet Monitor User’s Manual
A.8.4.2 Specification
A.8.5 TEMP
A.8.5.1 Standard
16
S12 Vet Monitor User’s Manual
Parameter T1,T2,TD
Probe YSI400 series probe (2252Ω@25℃, accuracy ±0.1ºC)
Measurement site Surface and coelom
0.0ºC ~ 50.0ºC ( 32.0℉ ~
Measurement range
Measurement range and 122.0℉)
accuracy Resolution 0.1ºC
Accuracy of circuit ±0.1℃ (±0.2℉) (without sensor)
Power supplied to probe <20μW
Updated time Every about 1~2s
Minimum measurement Surface: ≤100s
time Coelom: ≤80s
A.8.6 IBP
A.8.6.1 Standard
17
S12 Vet Monitor User’s Manual
greater
Resolution 1%
Measurement
-50 mmHg ~+360mmHg
Static pressure range
measurement range Resolution 1 mmHg
and accuracy ±2mmHg or ±2% (whichever is greater,
Accuracy
without sensor)
Measurement
Dynamic pressure -50 mmHg ~+360mmHg
range
measurement range
±2mmHg or ±2% (whichever is greater,
and accuracy Accuracy
without sensor)
Including sensor At least d.c.~10Hz
Frequency response
Only host At least d.c.~12Hz
IBP zero range -200mmHg~+200mmHg
Measurement
30bpm ~300bpm
range
PR
Resolution 1bpm
Accuracy ±1% or ±1bpm whichever is greater
Nominal
5uV/V/ mmHg
sensitivity
Output
300Ω~3000Ω
Pressure sensor impedance
Volumetric
<0.04 mm3 /100 mmHg
displacement
Error ±2%
IBP analog output
≤35ms
delay
18
S12 Vet Monitor User’s Manual
A.8.7 CO2
A.8.7.1 Standard
Measurement
EtCO2, FiCO2, a CO2 waveform and awRR
parameter
Measurement
Mainstream, Sidestream
method
19
S12 Vet Monitor User’s Manual
Respironics
<20s
LOFLO C5
Masimo ISA <10s (Report concentration and achieve
Capno highest accuracy)
97% of the design error can be reached
BLT Capno_M within 8s, and the design error can be
reached within 20s.
Respironics
<15s
CAPNOSTA 5
Masimo IRMA <10s
EtCO2/FiCO2 display
0.1% or 1mmHg
resolution
BLT Capno_S 3~150 bpm
Respironics
2~150 bpm
LOFLO C5
Masimo ISA
awRR measurement 0~150 bpm
Capno
range
BLT Capno_M 3~150 bpm
Respironics
0~150 bpm
CAPNOSTA 5
Masimo IRMA 0~150 bpm
awRR measurement
±1 bpm
accuracy
BLT Capno_S 50±10mL/min
Sampling frequency Respironics
50±10mL/min
and accuracy of gas LOFLO C5
(only sidestream) Masimo ISA
50±10mL/min
Capno
BLT Capno_S <3s (including delay time and rise time)
Response time Respironics
<3s (including delay time and rise time)
LOFLO C5
20
S12 Vet Monitor User’s Manual
Masimo ISA
<3s (including delay time and rise time)
Capno
BLT Capno_M About 70ms (rising time)
Respironics
<60ms (rising time)
CAPNOSTA 5
Masimo IRMA <90ms (rising time)
A.8.7.4 The effects on CO2 measuring values caused by the interfering gases
Respironics LOFLO C5
Gas or Vapor Gas Level Quantitative Effects
Nitrous oxide 60% No Additional Effect
Halothane 4% No Additional Effect
Enflurane 5% No Additional Effect
Isoflurane 5% No Additional Effect
Sevoflurane 5% No Additional Effect
Xenon 80% Negatively bias Carbon Dioxide
values by up to an additional 5
mmHg at 38 mmHg
Helium 50% No Additional Effect
Metered dose inhaler Unspecified Unspecified
propellants
Desflurane 15% Concentrations greater than 5% will
21
S12 Vet Monitor User’s Manual
CO (Carbon 1 vol% – 1) – 1) – 1) – 1)
22
S12 Vet Monitor User’s Manual
5)
monoxide)
NO (Nitrogen 0.02 vol% – 1) – 1) – 1) – 1)
monoxide) 5)
O2 5) 100 vol% – 2) – 2) – 1)
Note 1: Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2: Negligible interference with N2O / O2 concentrations correctly set, effect
included in the specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO2 readings by 6%. This means that if measuring on a mixture containing
5.0 vol% CO2 and 50 vol% Helium, the actual measured CO2 concentration will typically
be (1-0.06) * 5.0 vol% = 4.7 vol% CO2.
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
Respironics CAPNOSTA 5
Gas or Vapor Gas Level
Nitrous oxide 60%
Halothane 4%
Enflurane 5%
Isoflurane 5%
Sevoflurane 5%
Xenon 80%
Helium 50%
Metered dose inhaler propellants Unspecified
Desflurane 15%
Ethanol Unspecified
Isopropanol Unspecified
Acetone Unspecified
Methane Unspecified
Additional notes regarding cross-sensitivity compensation errors:
Xenon: The presence of Xenon in the exhaled breath will negatively bias Carbon Dioxide
23
S12 Vet Monitor User’s Manual
Ethanol, Isopropanol, Acetone, Methane: CO2 accuracy will not be affected by the
presence of 0.1% ethanol, 0.1% isopropanol, 0.1% acetone or 1% methane.
Quantitative effects of humidity and condensation: Full accuracy specifications will not
be maintained for all non-condensing humidity levels.
A.8.8 C.O.
A.8.9 DM
24
S12 Vet Monitor User’s Manual
Drip rate
5~200 Drops/min(1mL of conventional tube =20 drops)
measurement range
Drip accuracy ±2 and ±2% (whichever is greater)
A.9.1 ECG
A.9.2 RESP
A.9.3 NIBP
25
S12 Vet Monitor User’s Manual
A.9.4 SpO2
A9.5 TEMP
A.9.6 IBP
26
S12 Vet Monitor User’s Manual
A.9.7 CO2
A.9.8 C.O.
27
S12 Vet Monitor User’s Manual
The monitor complies with IEC 60601-1-2. All accessories listed in the accessories
listed in the accessories of this manual meet the requirements of IEC 60601-1-2 when
used with this equipment.
CAUTION:
The monitor conforms to the electromagnetic compatibility requirements in
IEC 60601-1-2, ISO 80601-2-55, IEC 80601-2-30, IEC 80601-2-49, ISO
80601-2-61, IEC 60601-2-34 standards.
The user shall install and use according to the electromagnetic compatibility
information provided by the accompanying documents.
Portable and mobile RF communication equipment may affect the
performance of this monitor, and strong electromagnetic interference should
be avoided during use, such as close to mobile phones, microwave ovens, etc.
The guidelines and the manufacturer’s statement are detailed in the appendix.
The device needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
Portable and mobile communication equipment may affect the performance of
this monitor.
Other devices that have RF transmitter or source may affect this device (e.g.
cell phones, PADs, PCs with wireless function).
WARNING:
The monitor should not be used close to or stacked on top of other equipment.
If it must be used close to or stacked on top of other equipment, it should be
observed and verified that it can operate normally under its used
28
S12 Vet Monitor User’s Manual
configuration.
Class A equipment is intended to be used in industrial environment. Due to
conduction disturbance and radiation disturbance of this monitor, there may
be potential difficulties in ensuring electromagnetic compatibility in other
environments.
In addition to cables sold by the manufacturer of this monitor as spare parts
for internal components, the use of accessories and cables other than those
specified may result in increased emission or reduced immunity of this
monitor.
Even if other equipment meets the emission requirements of corresponding
national standards, this monitor may still be interfered by other equipment.
A warning that operation of the EQUIPMENT or SYSTEM below the
minimum amplitude or value may cause inaccurate results. The minimum
amplitude or value of patient physiological signal: the minimum amplitude of
ECG signal is 0.5mV, the minimum value of PR is 30bpm and the minimum
value of SpO2 is 70%.
Table 1
Guidance and manufacture’s declaration – electromagnetic emission
The monitor is intended for use in the environment specified below. The customer or the user of
the monitor should assure that it is used in such environment.
29
S12 Vet Monitor User’s Manual
IEC 61000-3-3
If the system is operated within the electromagnetic environment listed in Table
Guidance and Declaration – Electromagnetic Immunity, the system will remain safe and
provide the following essential performance:
Operating mode
Accuracy
Function
Accessories identification
Date stored
Alarm
Detect for connection
Parameter
Table 2
Guidance and manufacture’s declaration – electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of monitor should assure that it is used in such an environment.
Electrical fast 2 kV for power 2kV for power Mains power quality should
transient/burst supply lines supply lines be that of a typical
IEC 61000-4-4 1 kV for input/output 1 kV for commercial or hospital
±1 kV line(s) to ±1 kV line(s) to
Surge
line(s) line(s)
IEC 61000-4-5
±2 kV line(s) to earth ±2 kV line(s) to
earth
30
S12 Vet Monitor User’s Manual
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table 3
Guidance and manufacture’s declaration – electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of monitor should assure that it is used in such an environment.
31
S12 Vet Monitor User’s Manual
3.5
d P 150 KHz to 80 MHz
V 1
3.5
d P 80 MHz~800 MHz
E1
7 80MHz~2.7GHz
d P
E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the compliance
level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band
(2400-2483.5MHz) is exempt from the ESSENTIAL PERFORMANCE requirements, but remains
safe.
32
S12 Vet Monitor User’s Manual
a
Field strengths from fixed RF transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the monitor is used exceeds the
applicable RF compliance level above, the monitor should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the monitor.
b
Over the frequency range 0.15 MHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4
Recommended separation distances between portable and mobile RF communications
equipment and the monitor
The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the monitor can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the monitor.
Rated Separation distance according to frequency of transmitter
maximum
150kHz~80MHz 80MHz~800MHz 80 MHz~2.7 GHz
output power of
transmitter (w) d = 1.2 P d = 1.2 P d = 2.3 P
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
33
S12 Vet Monitor User’s Manual
RF Parameter
Radio frequency Operating frequency Modulation Transmission power
transmitter
WiFi 2.4GHz~2.495GHz DSSS and <20dBm (average value)
IEEE802.11a/b/g/n 5.15 GHz~5.35GHz OFDM <30dBm (Peak)
5.47 GHz~5.725GHz
5.725 GHz~5.82GHz
The radio module used in the device is complied with the essential requirements and
other relevant provisions of The Radio Equipment Directive.
WARNING:
Keep a distance of at least 20 cm away from the monitor when WIFI function is
in use.
34
S12 Vet Monitor User’s Manual
C.1.1 ECG
C.1.2 Arrhythmia
35
S12 Vet Monitor User’s Manual
C.1.3 ST
36
S12 Vet Monitor User’s Manual
C.1.4 QT
37
S12 Vet Monitor User’s Manual
C.2 RESP
C.3 SpO2
38
S12 Vet Monitor User’s Manual
C.4 TEMP
39
S12 Vet Monitor User’s Manual
C.5 NIBP
40
S12 Vet Monitor User’s Manual
C.6 IBP
41
S12 Vet Monitor User’s Manual
42
S12 Vet Monitor User’s Manual
C.7 C.O.
C.8 CO2
43
S12 Vet Monitor User’s Manual
C.9 DM
44
S12 Vet Monitor User’s Manual
C.13 Recorder
45
S12 Vet Monitor User’s Manual
C.14 Other
46
S12 Vet Monitor User’s Manual
47
S12 Vet Monitor User’s Manual
This chapter lists some of the most important physiological and technical alarm
information, and some alarm information may not be listed.
48
S12 Vet Monitor User’s Manual
49
S12 Vet Monitor User’s Manual
This chapter lists the main technical alarms, the level of technical alarms, the cleared
status of alarm reset alarm prompts, and the measures to be taken after the alarm occurs.
Some alarm messages may not be listed.
After different technical alarm is reset, the alarm prompt will be cleared to different
degrees. The following three types of technical alarms are given in this section according
to the status of alarm being cleared.
Completely clear: the technical alarm is completely clear. The monitor has no
alarm indication.
Sound and light can be cleared: the technical alarm displays as prompt
information.
Not clearable: the sound of technical alarm is shielded.
50
S12 Vet Monitor User’s Manual
51
S12 Vet Monitor User’s Manual
52
S12 Vet Monitor User’s Manual
53
S12 Vet Monitor User’s Manual
54
S12 Vet Monitor User’s Manual
55
S12 Vet Monitor User’s Manual
56
S12 Vet Monitor User’s Manual
57
S12 Vet Monitor User’s Manual
58
S12 Vet Monitor User’s Manual
59
S12 Vet Monitor User’s Manual
Appendix E Cybersecurity
This chapter mainly describes the information related to cybersecurity of the
monitor.
Hardware environment
Monitor software is only applicable to S12 Vet monitor hardware
platform.
Screen: 12.1″ LCD screens with 4:3 aspect ratio and 800*600 resolution.
Peripherals: nurse call module, recorder.
Software environment
Main board: S12MB
Operating system: Linux-3.2.0 kernel + Busybox filesystem.
Database: sqlite-3.16.2
Network environment
Apply to LAN
The communication interface between the monitor and the CMS is wired or wireless
Ethernet, using the standard TCP/IP protocol family, and the application layer data format
follows the Central Monitoring System Network Communication Protocol during
transmission.
60
S12 Vet Monitor User’s Manual
61
S12 Vet Monitor User’s Manual
62
S12 Vet Monitor User’s Manual
Symbol Explanation
- Minus
– Negative
% Percent
/ Per; Divide;Or
~ To
+ Plus
= Equal to
< Less than
63
S12 Vet Monitor User’s Manual
> Greater than
≤ Less than or equal to
≥ Greater than or equal to
± Plus or minus
× Multiply
© Copyright
64
S12 Vet Monitor User’s Manual
65
S12 Vet Monitor User’s Manual
66
S12 Vet Monitor User’s Manual
67
S12 Vet Monitor User’s Manual
68
S12 Vet Monitor User’s Manual
y
CO2 ○ ○ ○ ○ ○ ○
accessor
y
Stand Carts ○ ○ ○ ○ ○ ○
stand
Wall ○ ○ ○ ○ ○ ○
stand
○: It means that the content of the toxic and harmful substances in all homogeneous materials of the
component is below the limit specified in SJ/T11363-2006.
×: Indicates that the content of the toxic and harmful substances in at least one homogeneous material
of the component exceeds the limit requirements specified in SJ/T11363-2006.
69
Product name: Veterinary Monitor
Fax: 0756-3399919
PN:22-075-0022