100% found this document useful (1 vote)
4K views258 pages

CH200 Operation Manual

Operation manual shimadzu radspeed pro

Uploaded by

jahoyosc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
4K views258 pages

CH200 Operation Manual

Operation manual shimadzu radspeed pro

Uploaded by

jahoyosc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 258

M514-E070L

Sep. 2022

Ceiling Type X-ray Tube Support

CH-200
CHU-200

R
Operation Manual
Read this manual thoroughly before you use the product.
Keep this manual for future reference.
About the Symbols Appearing in this Operation
Manual
Throughout the text in this manual, warnings and other information essential when using
this unit, such as cautionary or prohibited items, appear classified as per the following:

Mark Description
Indicates an imminently hazardous situation which, if
not avoided, will result in serious injury or death.

Indicates a potentially hazardous situation which, if not


avoided, could result in serious injury or possibly death.

Indicates a potentially hazardous situation which, if not


avoided, may result in minor to moderate injury or
equipment damage.

Emphasizes additional information that is provided to


ensure the proper use of this product.

Indicates information provided to improve product


performance.

Indicates the location of related reference information.

Revision History
Revision Date Changes

First edition Mar. 2014

A Feb. 2015 Added List of Cable in section 7.1.

B Jul. 2016 Compliant with Standards.

C Jul. 2017 Changed software revision.

D Dec. 2017 Compliant with standard of EN ISO 15223-1:2016.

E Jun. 2018 Compliant with Standards.

F Nov. 2018 Compliant with Standards.

G Feb. 2019 Changed software revision.

H Mar. 2019 Changed the Bucky description.

• Added CHU-200.
• Changed software revision.
J Jul. 2019
• Added "4.11 Function Associated with X-ray
Diagnostic Table YSF-500(Option)".

• Compliant with Standards.


K Nov. 13 2020 • Added the power assist function as an optional
feature.

L Sep. 1 2022 Compliant with Standards.

CH-200/CHU-200 Operation Manual


0 Preface

Thank you for purchasing the "Ceiling Type X-ray Tube Support CH-200
/CHU-200" (hereafter referred to as "equipment").

This operation manual contains information for ensuring proper use of the
equipment. Read this manual thoroughly before using the equipment and
operate the equipment in accordance with the instructions in this manual. If the
precautions in this manual are disregarded, there is risk of damage to equipment
or injury to operators and patients. However, it is not possible to foresee all risks
and provide precautions for each of them. Therefore, when using the equipment
in ways not described in this manual or when any question arises in the manual
description, please contact the place of purchase or a sales representative listed
at the end of this manual.

This manual should be kept available for future reference. If the user or usage
location changes, ensure that this operation manual is always kept together with
the equipment. Periodically check to be sure that the operation manual and the
warning labels are not missing or damaged. If they are, contact your Shimadzu
service representative for replacement.

Original version is approved in English.

Notice

• All copyrights regarding this manual are property of Shimadzu Corporation.


Neither all nor part of these contents shall be reproduced or duplicated
without the express permission of Shimadzu Corporation.
• Content of this manual may be changed for improvement without notice.
Although every possible effort has been made to avoid errors while creating
this manual, immediate revision may not be possible in the event that errors
or missing information are detected.
• Screen images and illustrations contained in this manual may differ from
those in actual use, and are intended for example purposes only. Illustrations
may also use partial images.
• Other company names and product names mentioned in this manual are
trademarks or registered trademarks of their respective companies.
• The TM and  symbols are omitted in this manual.
© 2014-2022 Shimadzu Corporation. All rights reserved.
Preface

Operating Precautions

"Operating Precautions (for Both Safety and the Prevention of Danger)


in the Use of Electric Medical Equipment"

1. Nobody without the following experience and knowledge should use the
equipment.

(1) Medical (radiographic) training (if particular qualifications are required in the
country concerned, those qualifications must be held).
(2) The capacity to read and understand the operation manual.

2. When installing the equipment, pay attention to the following items:

(1) Do NOT install it near water faucet or similar equipment.


(2) Install it away from potential sources of problems such as abnormal pressure,
temperature or humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.
(3) During transportation and operation of the equipment, avoid tilting, vibration
and sharp impact against it.
(4) Keep the equipment away from areas where chemicals or gases are stored.
(5) Use only the correct electrical power source with matching frequency, voltage
and current (or wattage).
(6) Check the conditions of the battery power source (power and polarity) before
operating the equipment.
(7) Properly ground the equipment.

3. Before operating the equipment, pay attention to the following items:

(1) Check the conditions of switch contacts, polarity, dial settings, and meters, and
make sure the equipment performs correctly.
(2) Confirm that the ground is connected properly.
(3) Check all wiring for proper and correct connections.
(4) Pay attention when using more than one unit at a time, because it may lead to
an incorrect diagnosis and cause danger.
(5) Check the condition of the external electric circuit, which will be directly
connected to a patient.
(6) Check the condition of the battery power source.

ii CH-200/CHU-200 Operation Manual


Preface

4. While operating the equipment, pay attention to the following items:

(1) Do NOT exceed time or the amount of equipment use needed for diagnosis or
therapy.
(2) Observe the equipment and patient continuously for early detection of
problems.
(3) When a problem is detected with the equipment or patient, take proper action to
stop the equipment without harming the patient.
(4) Do NOT let the equipment touch the patient.

5. After operating the equipment, pay attention to the following items:

(1) Turn off the switches and return the dial to their original postions before use in
the prescribed order. Then, turn off the main power switch.
(2) Do NOT pull the power cable forcibly from the outlet.
(3) When storing the equipment, pay attention to the following items:
(i) Keep it away from water.
(ii) Store it away from potential causes of problems such as abnormal
pressure, temperature and humidity, draft, direct sunlight, dust, chlorine or
sulphur gas.
(iii)During transportation and storage of the equipment, avoid tilting, vibration
and sharp impact against it.
(iv)Store the equipment away from areas where chemicals and gases are
stored.
(4) Clean all attachments, cables and contacts, and store them in one place.
(5) Keep the equipment clean to avoid problems during the next use.

6. When the equipment is found to be out of order, do not try to repair it.
Display an appropriate sign to indicate that the equipment is out of order, and
contact your Shimadzu service representative for repair.

7. Do NOT modify any part of the equipment.

8. Preventive maintenance

(1) The equipment and its parts should be periodically checked.


(2) If the equipment has not been in operation for an extended period of time, test it
prior to actual operation to make sure it works correctly and safely.

9. Concerning other items, operate properly according to the operation manual.

CH-200/CHU-200 Operation Manual iii


Preface

Precautions in Use
When using this equipment, please observe the following precautions for the safety of
the operator and patient.

The responsibility for management of use and maintenance of medical


equipment lies with the user.
This equipment is restricted to use by, or under supervision of, a diagnostic
radiology technician or a person with a certificate indicating equal proficiency.
Repair and inspection of the inside of the equipment is dangerous. Be sure to
contact your Shimadzu service representative for repair and inspection.

Never modify the equipment.


In general, modifications are strictly prohibited by the Regulatory requirements
of the law of the country where the device is installed. Please contact your
Shimadzu service representative if it is necessary to modify the equipment.

Perform periodic inspection.


Preventive maintenance is required to maintain long-term safety and
performance of the equipment.
The "6 Maintenance" chapter in this manual gives detailed descriptions of daily
and periodic maintenance and inspection that a user should perform. As for the
maintenance and inspection that only specially trained experts can perform,
utilize the maintenance agreement program offered by your Shimadzu service
representative.

Repair and maintenance of the inside of the equipment can only be


performed by your Shimadzu service representative.
Maintenance must be assigned to specially trained experts. Contact your
Shimadzu service representative for repair and maintenance.

iv CH-200/CHU-200 Operation Manual


Preface

In order to ensure safety when using the system, read the operation
manual provided with each system component for details on usage and
relevant precautions.

Connect this equipment only to Shimadzu certificated devices or devices


that are proven to be safe and show no performance degradation in any
combination including connecting system.

If the operator has no experience in operating the equipment, be sure


that he or she receives instruction on how to operate it from Shimadzu
service representative or someone who has adequate experience in
using the equipment.
In order to operate the equipment safely, an explanation of the operation
needs to be given. When installing the equipment, your Shimadzu service
representative explains the operating procedure using this operation manual.
Follow their directions and operate the equipment correctly.
"1.7 Operator Profile"

Secure the means for the operator and the patient to communicate with
each other.
If equipment usage is deemed to put the patient at risk due to his or her
condition, refrain from conducting the study or treatment.

CH-200/CHU-200 Operation Manual v


Preface

Federal law restricts this equipment to sale by or on the order of


physician. (This caution is the prescription language required by Federal
Regulations in U.S.A.)

Be Sure to Read the Following to Prevent Explosion, Electric Shock, or


Injury

Do NOT use a power supply other than the single phase AC100 V,
0.2 kVA, 50/60 Hz and the single phase AC200/220/230/240 V, 1.0 kVA, 50/
60 Hz.
Using a power supply other than the one specified above may cause serious
accidents such as fires, smoke emissions or explosions.

Do NOT use any potentially flammable or explosive gas, such as


disinfectant sprays, near the equipment.
Use of such gas may cause an explosion.

Check the condition of the patient before conducting a study.


If equipment usage is deemed to put the patient at risk due to the his or her
condition, refrain from conducting the study or treatment.

vi CH-200/CHU-200 Operation Manual


Preface

Do NOT use the equipment in places where liquid may enter.


The equipment is not designed to be waterproof. Invasion of any liquid should
cause electric shock, system failure or malfunction.

Do NOT spill any liquid, such as contrast medium, saline, or disinfectant,


onto the equipment.
Should such liquid drip on equipment surfaces, wipe it off immediately. Any
such liquid entering into system electronics may cause failure or malfunction.
Should liquid drip on the equipment or enter the covers, immediately turn off
the power and contact your Shimadzu service representative.

When there is any abnormality in operation, or unusual smell or smoke


emission during operation, stop operation immediately and contact your
Shimadzu service representative.
Continued use may damage the equipment and cause injury.

Do NOT open the covers of the equipment.


Otherwise, electric shock may result. When opening the covers for
maintenance, contact your Shimadzu service representative.

CH-200/CHU-200 Operation Manual vii


Preface

Always be very careful when moving the equipment to avoid contact with
the patient or operator and to ensure that the patient or operator does not
become caught between the equipment and any neighboring devices.
Otherwise, it may cause injury.

Always be very careful when moving the X-ray radiography stand and X-
ray radiography table to ensure that the patient does not become caught
between these devices and this equipment.
Otherwise, it may cause injury.

Do NOT use in a location where metal fragments may enter the


equipment.
This can result in electric shocks.

Do NOT perform any maintenance work of the equipment during study.


The patient may be injured.

viii CH-200/CHU-200 Operation Manual


Preface

If the message "Alert: Wire. Stop unit and call service representative" is
displayed on the operation touch panel, stop using the equipment and
contact your Shimadzu service representative.
• The equipment is suspended by two wire ropes. If either of these ropes break,
radiography is disabled.
• A buzzer sounds continuously when the warning message is displayed on the
operation touch panel. Any continued use of the equipment while this message
is displayed may result in serious injury to the patient or operator.

Alert: Wire. Stop unit and


call service representative.

• When combined with the X-ray high-voltage generator UD150B-40/V-40/L-40,


press the hand switch to display the warning message on the touch panel of the
X-ray high-voltage generator UD150B-40/V-40/L-40.

CH-200/CHU-200 Operation Manual ix


Preface

When the equipment is combined with the X-ray Diagnostic Table


YSF-300/YSF-500, observe the following cautions when moving the
tabletop by operating the X-ray tube support:

Avoid collisions with other Avoid contact with


equipment. personnel.

Avoid trapping the


patient's fingers or hands.

Accuracy of displayed values is not guaranteed.


Displayed values measured by the measurement functions of this equipment
are not absolute values but relative values based on the capability of the
equipment used.
"7.4.3 International Standards"

x CH-200/CHU-200 Operation Manual


Preface

Cautions on Environmental Conditions

Do NOT use the equipment in an oxygen-rich environment.


The use in an oxygen-rich environment may cause fatal or serious injuries or
damage to the equipment due to easy ignition.

Be sure to use the equipment under the following environmental


conditions:
The installation of a dedicated air-conditioner in the examination room is
recommended if the building air-conditioner cannot meet the necessary
environmental conditions.
"1.5 Environmental Conditions" P.8
Note also that there must be no sudden changes in temperature or humidity.
This causes condensation, which can lead to trouble.

Cautions on Radiography

Restrict all persons other than the patient from accessing the equipment
in accordance with local regulations.
To avoid unnecessary exposure, acceptable distances (maximum access
values) to the equipment by any person other than the patient are defined for
each region.

The equipment can be operated only by qualified personnel, such as


radiology technicians or those with equivalent qualifications.

CH-200/CHU-200 Operation Manual xi


Preface

No person but the patient is allowed to stay in the examination room


during X-ray irradiation.
If the equipment is not used correctly, the operator, the patient, and other
persons may receive a greater dose of radiation than necessary. If for some
reason another person has to be in the examination room, that person must
take adequate measures to protect themselves against radiation (protective
apron, screen, etc.).
* During radiography, the radiography indicator illuminates and the buzzer
sounds an audible warning.

Perform X-ray irradiation carefully and according to the doctor's


directions when using the equipment with expectant mothers, women
who suspect they are pregnant, lactating women, or children.
Particular ways of using the equipment may increase the scatter dose
absorbed into the patient, which may cause a radiation hazard.

Always check the X-ray exposure region using the collimator lamp.
Irradiating a patient with X-rays outside the required region risks exposure of
the patient to unnecessary radiation.

During X-ray irradiation, ensure that the X-rays irradiate the necessary
region only.
To avoid unnecessary exposure, narrow down the collimator and take
protection measures, such as wearing a protective apron.

xii CH-200/CHU-200 Operation Manual


Preface

Do NOT place any unnecessary object in the location within the X-ray
exposure region.
Doing so may result in unnecessary radiation exposure to the patient.

Be sure to carry out a warm-up (running-in operation of the Xray tube


unit) before taking an X-ray radiograph.
Follow the warm-up procedure described in the X-ray high voltage generator
operation manual.

Perform the warm-up if an arc occurs.


Suddenly using the X-ray tube unit near the nominal X-ray tube voltage (above
100 kV) after using the unit at a relatively low tube voltage (80 kV max.) for a
prolonged period may result in arc. This arc occurs due to loss of the warm-up
effect at high tube voltage after the X-ray tube unit is used at a relatively low
tube voltage for a prolonged period.
In this case, warm-up the X-ray tube unit by referring to the procedure
described in the X-ray high voltage generator operation manual.

Do NOT perform unnecessary standby operations.


If standby status continues after the radiography preparation button is pressed,
wire disconnection or withstand voltage failure may result owing to the
evaporation of the X-ray tube filament.

CH-200/CHU-200 Operation Manual xiii


Preface

In order to minimize the radiation dose on the patient, make the distance
between the focus and the patient's body surface as long as possible
(Minimum 45 cm).
The shorter the distance becomes, the greater the amount of scatter dose
absorbed into the patient, which may cause a radiation hazard.

Pay extra attention when irradiating X-rays for a long time or repeatedly.
It may cause a radiation hazard.

Cautions on Cleaning and Disinfection

Be sure to turn the equipment power OFF before cleaning and


disinfecting the equipment.
Otherwise, a malfunction may occur in the equipment, or the equipment may
operate in an unintended way.
Also, thoroughly ventilate the room before turning ON the power after
disinfection work is complete.

Be sure to clean and disinfect the equipment periodically.


Cleaning and disinfection is very important to ensure that the equipment can
be used hygienically and safely. Strictly follow the methods prescribed.

xiv CH-200/CHU-200 Operation Manual


Preface

Be sure to clean the equipment frequently and after each patient use.
While doing so, do NOT directly apply or spray any disinfectant, cleaner, or
water onto the equipment. Wipe down all contact surfaces using a cloth
moistened with an appropriate disinfectant or cleaner. Make sure the cloth is
NOT too wet. If it is, liquid may enter into system electronics, causing failure or
malfunction.

Observe the following precautions when cleaning and disinfecting the


touch panel:
• Turn the equipment power OFF before cleaning and disinfecting.
• Clean and disinfect the touch panel periodically.
• Wipe the surface of the equipment with a cloth moistened with disinfectants or
cleaner. If liquid gets inside the equipment, it can cause failure or accidents.
• Do NOT rub or hit the panel surface because it tends to scar easily.
• Wipe gently with moistened soft close to remove the dust from the panel
surface.

Use the following disinfectants and cleaner:


• Chlorine disinfectants
Sodium dichloroisocyanurate solution (1 % maximum)
Sodium hypochlorite solution (1 % maximum)
• Alcohol disinfectants
Commercially available isopropyl alcohol solution (Up to 99 wt% can be used)
Rubbing alcohol (76.9 - 81.4 vol% Ethanol, Isopropyl alcohol as an additive)
• Glass cleaner (Only onto the touch panel)

CH-200/CHU-200 Operation Manual xv


Preface

Do NOT use the following disinfectants:


• Disinfectants that corrode metals, plastics, rubber, or paint
• Disinfectants unsuitable for metals, plastics, rubber, or paint
• Spray-gas type disinfectants
• Volatile disinfectants
• Disinfectants that may enter the equipment

Use disinfectants at a minimum.


Repeated disinfection over a long time may lead to discoloring and cracking on
the equipment surface, and deterioration of rubber and plastic. If any
abnormality is found on the equipment after disinfection, stop using the
equipment immediately. Contact your Shimadzu service representative for
repair.

Do NOT use an organic solvent.


Organic solvents may change the surface color. If an organic solvent adheres
to the surface, wipe it off immediately.

When disinfecting unpainted metals, do NOT use chlorine-based


disinfectants.
Chlorine-based disinfectants may corrode the surface of the equipment. If
chlorine-based disinfectants adhere to the surface, wipe them off immediately.

xvi CH-200/CHU-200 Operation Manual


Preface

When disinfecting resin parts such as reticule of the collimator, do NOT


use rubbing alcohol.
Rubbing alcohol may lead to deformation or crack of resin parts such as
reticule of the collimator. Wipe it off immediately if it adheres to the reticule of
the collimator.

When cleaning resin parts such as reticule of the collimator, use cloth
lightly moistened, not soaked, with cold or warm water mixed with
neutral detergent that does not include organic solvent.
Rubbing alcohol, organic solvents or non-neutral detergents may lead to
deformation or crack of resin parts such as reticule of the collimator. Wipe
them off immediately if they adhere to the reticule of the collimator.

On completing the work, check the following points before switching the power ON
again.

• There must be no water or disinfectant adhering to the equipment.


• The tools used in cleaning and disinfecting work must be tidied away.
"3.1 Power ON/OFF"

When turning the power ON after cleaning, make sure the examination
room is properly ventilated.
Turning the power ON while any flammable gas remains in the examination
room could lead to fire, smoke, explosion or electrocution.

CH-200/CHU-200 Operation Manual xvii


Preface

Cautions Relating to Cellular Telephones

Do NOT bring any cellular telephones or related devices into the


examination room with their power ON.
Such devices can exceed the EMC standard limitations, and under some
conditions this can impair the proper functioning of the equipment. In the worst
case, this can cause serious injuries or clinical errors.

Cautions on Electromagnetic Compatibility (EMC)

This equipment needs special precautions regarding EMC.


Install and use the equipment according to the EMC information provided in
this operation manual.
"7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)"
P.186

Make sure that electromagnetic compatibility is obtained.


All peripheral devices must satisfy EMC standards regarding emission of
electromagnetic energy and susceptibility to electromagnetic environment.
Devices that do not satisfy these standards may disturb the correct functioning
of the equipment. In the worst case, this can cause serious injuries or clinical
errors.
"7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)"
P.186

xviii CH-200/CHU-200 Operation Manual


Preface

Do NOT use this equipment adjacent to, or stacked with, other


equipment.
If adjacent or stacked use is necessary, check to be sure that this equipment
works properly in the environment.
"7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)"
P.186

Do not use accessories, transducers and cables other than those


specified or provided by Shimadzu.
The use of accessories, transducers and cables other than those specified or
provided by Shimadzu could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper
operation.

Portable RF communications equipment (including peripherals such as


antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the CH-200, including cables specified by
the manufacturer.
Otherwise, degradation of the performance of this equipment could result.

Portable RF communications equipment (including peripherals such as


antenna cables and external antennas) using TETRA (TErrestrial Trunked
RAdio) should not be used no closer than 50 cm to any part of the CH-
200, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.

CH-200/CHU-200 Operation Manual xix


Preface

Portable RF communications equipment (including peripherals such as


antenna cables and external antennas) using Bluetooth, Wireless LAN
(2.4 GHz), RFID, LTE (Band7) should not be used no closer than 45 cm to
any part of the CH-200, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.

Portable RF communications equipment (including peripherals such as


antenna cables and external antennas) using GSM 180, CDMA 1900, GSM
1900, DECT, LTE Band 1,3,4,25, UMTS should not be used no closer than
45 cm to any part of the CH-200, including cables specified by the
manufacturer.
Otherwise, degradation of the performance of this equipment could result.

Cautions When Irradiating Consecutive Pulse X-rays

Observe the following precautions when irradiating


consecutive pulse X-rays:
• Conducting studies involving irradiating consecutive pulse X-rays onto the
region where an implantable pacemaker or defibrillator is implanted may cause
these devices to malfunction.
• Refer to the "Important General Cautions", "Interactions", or other relevant
sections in the accompanying documentation of the implantable pacemaker or
defibrillator and take the prescribed measures before irradiating the implanted
region of these devices with consecutive pulse X-rays.
Fluoroscopy or radiography performed by irradiating consecutive pulse Xrays
(such as serial radiography with a few second intervals, pulsed fluoroscopy,
digital angiography, DSA, or cineradiography) can adversely affect the CMOS
circuit in implantable pacemakers and defibrillators. Such affects may cause
oversensing in these devices that can temporarily inhibit pacing pulse output
and result in an inappropriate heart rate.

xx CH-200/CHU-200 Operation Manual


Preface

Warranty
The system is warranted to be free from defects in material and workmanship for one
year from the date of delivery. If found to be defective, the system must be offered to
your Shimadzu service representative for inspection and examination. Upon
examination, your Shimadzu service representative, at its sole option, will repair or
replace at no charge, the system or any part found to be defective. Components which
wear are not warranted.

This warranty extends to original purchaser or the lessee of the new system only.
If the system is to be resold or delivered to a third party, such third party must be
provided with a copy of this manual, the installation manual and the technical manual
supplied with the system.

This warranty does not apply to the following:

1. Failure or damage due to any installation, relocation, or


service not provided by your SHIMADZU service
representative or a SHIMADZU designated contractor.

2. Failure or damage caused by the product of other companies


(except those purchased from SHIMADZU).

3. Failure or damage due to repairs using non-SHIMADZU


certified service parts.

4. Failure or damage due to non-compliance with the notices


and procedures set forth in this manual.

5. Failure or damage due to any operating environment


deviating from the requirements set forth in this manual.

6. Failure or damage due to natural disasters such as power


surge, rain, fire, earthquake, flood, and thunder.

7. Failure or damage due to installation on non-SHIMADZU


approved vehicle, ship, aircraft, or others.

8. Failure or damage due to use in non-SHIMADZU certified


countries.

9. Failure or damage in case of purchase from entity other than


SHIMADZU or Shimadzu service representative.

10. Failure or damage due to impact, drop, or shock.


Service after the expiration of the warranty is available at a reasonable cost and should
be performed by your SHIMADZU service representative.

CH-200/CHU-200 Operation Manual xxi


Preface

IN NO EVENT SHALL SHIMADZU AND ITS AFFILIATED ENTITIES BE LIABLE TO


ANY PERSON OR ENTITY FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL
OR CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY
DAMAGES RESULTING FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS
OF PROFITS, LOSS OF SAVINGS, THE COST OF PROCUREMENT OF
SUBSTITUTED GOODS, SERVICES OR TECHNOLOGIES OR FOR ANY MATTER
ARISING OUT OF OR IN CONNECTION WITH THE USE OR INABILITY TO USE
THE SYSTEM. In some jurisdictions, some of the foregoing warranty disclaimers or
damage limitations may not apply.

Shimadzu will be indemnified for any claim, liability, or damage arising out of the
misuse or non-compliance with this manual by the purchaser or lessee of the system.

Software
The information in this Operation Manual is based on the following software.

• Version: CH200.V01
• Revision: 4.3*
* : Even if the last digit of the revision is different from that above, the descriptions in
the manual are valid.

Service Life
The equipment lifetime is 10 years (based on Shimadzu's criteria) assuming the
specified maintenance checks are performed.

Disposal Precautions

When disposing of the equipment, contact your Shimadzu service


representative.
An improper disposal of this equipment may pollute the environment by
substances contained in parts.

xxii CH-200/CHU-200 Operation Manual


Preface

Action for Environment (WEEE)


To All Users of Shimadzu Equipment
in the European Union:
Equipment marked with this symbol indicates that it was sold on or
after 13th August 2005, which means it should not be disposed of
with general household waste. Note that our equipment is for
industrial/professional use only.

Contact your Shimadzu service representative when the


equipment has reached the end of its life. They will advise
WEEE Mark you regarding the equipment take-back.

With your co-operation we are aiming to reduce contamination from waste electronic
and electrical equipment and preserve natural resource through re-use and recycling.

Do not hesitate to ask your Shimadzu service representative, if you require further
information.

For California, USA Only

This product contains a battery that contains perchlorate material.


Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate

CH-200/CHU-200 Operation Manual xxiii


Preface

The Equipment to be Combined with

The "Shimadzu X-ray radiography stand " in this operation manual denotes the
following models:
• X-ray radiography stand BR-120 (without tilting)

• X-ray radiography stand BR-120T (with tilting)

The "Shimadzu X-ray radiography table" in this operation manual denotes the
following models:
• X-ray radiography table BK-120

• X-ray radiography table BK-200

For the grid indication or the operations in this operation manual, those are
described as the grid removal type of X-ray radiography table or X-ray
radiography stand.

The "Shimadzu autocollimator" in this operation manual denotes the following


model:
• R-300

xxiv CH-200/CHU-200 Operation Manual


Preface

Depending the combination with the X-ray radiography table or the X-ray
radiography stand, the operating procedure and touch panel screens of
the X-ray tube support CH-200 may differ from those indicated in this manual.
This operation manual describes the operating procedure that applies when
the equipment is combined the Shimadzu X-ray radiography table or Shimadzu
X-ray radiography stand.

Depending the combination with the collimator, the operating procedure and
touch panel screens of the X-ray tube support may differ from those indicated
in this manual.
This operation manual describes the operating procedure that applies when
the equipment is combined the Shimadzu collimator.

CH-200/CHU-200 Operation Manual xxv


Preface

Warning and Caution Labels


The following safety labels, which describe handling precautions, are attached to the
equipment. With adequate understanding of the contents on these labels and the
warning/ caution items in this manual, operate the equipment safely. Inspect the safety
labels periodically (once a year).

If any label is peeled or unreadable by stain or scratch, replace it with a new one.

For new labels, contact your Shimadzu service representative.

xxvi CH-200/CHU-200 Operation Manual


Preface

Related Documents
This manual describes the basic operating procedures for the equipment.
For more detailed information, refer to the operation manuals of the respective
components.

Document
Component Document Name
No.

X-ray High Voltage Generator UD150B-40/L-40/


X-ray High
V-40, D150BC-40/45 LC-40/45 VC-40/45,
Voltage M501-E052
D150BC-41 LC-41 VC-41, GSC-2002L
Generator
OPERATION MANUAL

Collimator Collimator type R-300 Operation Manual M526-E024

Digital Radiography System RADspeed Pro


System M517-E176
Operation Manual

CH-200/CHU-200 Operation Manual xxvii


Preface

Abbreviations
The following are the abbreviations used in this manual.

Abbreviations Descriptions

A/P Auto Positioning

APR Anatomical Programmed Radiography

deg Degree (Unit of angle)

Obliq Oblique Projection Radiography

Positive Beam Limitation (states that the irradiation field is


PBL
focused on size corresponding to X-ray detector)

POS Position (Location registered by Auto Positioning Function)

Source Image Distance (Radiography distance between X-ray


SID tube focus and X-ray detector)
" 8 Radiography distance indicator" P.24

Source Stand Distance (Radiography distance between X-ray


SSD tube focus and surface on Bucky of X-ray radiography stand)
" 8 Radiography distance indicator" P.24

Source Table distance (Radiography distance between X-ray tube


STD focus and X-ray radiography table)
" 8 Radiography distance indicator" P.24

FPD Flat Panel Detector

PA Posterior-Anterior (projection)

AP Anterior-Posterior (projection)

Options
The "Options" described in this manual may be provided as standard, or may be
unavailable, depending on the sales format.

xxviii CH-200/CHU-200 Operation Manual


Contents

Contents

Preface
Precautions in Use.................................................................... iv
Be Sure to Read the Following to Prevent Explosion,
Electric Shock, or Injury .......................................................................vi
Cautions on Environmental Conditions................................................xi
Cautions on Radiography ....................................................................xi
Cautions on Cleaning and Disinfection .............................................. xiv
Cautions Relating to Cellular Telephones........................................ xviii
Cautions on Electromagnetic Compatibility (EMC) .......................... xviii
Cautions When Irradiating Consecutive Pulse X-rays ........................xx
Warranty ..................................................................................xxi
Software..................................................................................xxii
Service Life .............................................................................xxii
Disposal Precautions ..............................................................xxii
Action for Environment (WEEE)
To All Users of Shimadzu Equipment
in the European Union: .......................................................... xxiii
For California, USA Only........................................................ xxiii
The Equipment to be Combined with.....................................xxiv
Warning and Caution Labels..................................................xxvi
Related Documents .............................................................. xxvii
Abbreviations ....................................................................... xxviii
Options................................................................................. xxviii

Chapter 1 Outline
1.1 Applications................................................................................ 2
1.2 Features..................................................................................... 3
1.3 Principle ..................................................................................... 5
1.4 Configuration.............................................................................. 6
1.5 Environmental Conditions .......................................................... 8
1.5.1 Operating Environment ...................................................................... 8
1.5.2 Transportation and Storage Environment .......................................... 9
1.5.3 Power Supply..................................................................................... 9
1.5.4 Grounding .......................................................................................... 9
1.5.5 Installation Requirements ................................................................ 10

CH-200/CHU-200 Operation Manual Contents-1


Contents

1.6 Classification of Equipment ......................................................11


1.7 Operator Profile ........................................................................12
1.8 Symbols....................................................................................13

Chapter 2 Configuration
2.1 Configuration ............................................................................16
2.2 Operation Panel .......................................................................19
2.3 Touch Panel .............................................................................22
2.4 Rear Mount Type, Low-ceiling Type,
and Front Mount L Type ...........................................................30
2.4.1 Rear Mount Type ............................................................................. 30
2.4.2 Low-ceiling Type .............................................................................. 32
2.4.3 Front Mount L Type.......................................................................... 33
2.5 Layout When Using Optional Functions ...................................34

Chapter 3 Operation
3.1 Power ON/OFF.........................................................................44
3.1.1 Power ON ........................................................................................ 44
3.1.2 Power OFF....................................................................................... 44
3.2 Moving X-ray Tube Support......................................................45
3.2.1 Vertical Movement ........................................................................... 45
3.2.2 Longitudinal Movement .................................................................... 48
3.2.3 Transversal Movement .................................................................... 49
3.2.4 3-Directional Movement ................................................................... 50
3.2.5 2-Directional Movement ................................................................... 51
3.2.6 Power Assist Function (Option) ....................................................... 53
3.3 Rotating X-ray Tube Support....................................................59
3.3.1 Rotation Around Vertical Axis .......................................................... 59
3.3.2 Rotation Around Horizontal Axis ...................................................... 60
3.4 Detent Mode .............................................................................63
3.4.1 Moving to Longitudinal/Transversal Detent Positions ...................... 63
3.4.2 Rotating to Preset Position Around Horizontal Axis ......................... 65
3.5 Upward retraction function........................................................66
3.5.1 Retracting upward ............................................................................ 66
3.6 Auto Collimation Function.........................................................67
3.6.1 Status Indication for System ............................................................ 67
3.6.2 Irradiation Field Size Selection Function (Option)............................ 70
3.6.3 Tracking with Radiography Conditions ............................................ 71
3.7 Lighting Collimator Lamp..........................................................74
3.7.1 Collimator Lamp Lights Up............................................................... 74
3.7.2 Adjusting Size of Irradiation Field [Reference] ................................. 75

Contents-2 CH-200/CHU-200 Operation Manual


Contents

Chapter 4 Radiography
4.1 General Radiography............................................................... 78
4.2 Vertical General Radiography.................................................. 80
4.3 Horizontal General Radiography.............................................. 82
4.4 Vertical Bucky Radiography..................................................... 84
4.4.1 Rotating X-ray Tube Unit ................................................................. 87
4.4.2 Dual X-ray Radiography Stand System ........................................... 88
4.5 Horizontal Bucky Radiography................................................. 90
4.5.1 When the Direction of Transversal Center
at X-ray Radiography Table and the Center
of X-ray Radiography Stand Have Shifted ....................................... 94
4.6 Radiography Using Tracking Device (Option).......................... 95
4.6.1 Vertical Tracking Radiography (Not using tilt function) .................... 96
4.6.2 Vertical Tracking Radiography (Using tilt function) ........................ 104
4.6.3 Horizontal Tracking Device ............................................................ 108
4.6.4 Horizontal Tracking Distance Selection Function .......................... 111
4.7 Tomography (Option)............................................................. 113
4.8 Bucky Tracking Radiography (Option) ................................... 117
4.9 Auto Positioning Function (Option) ........................................ 121
4.9.1 Configuration ................................................................................. 122
4.9.2 Movement Function ....................................................................... 125
4.9.3 Registration of A/P Tracking Position ............................................ 128
4.9.4 POS Position Registration ............................................................. 134
4.9.5 Radiography .................................................................................. 138
4.9.6 Status Display ................................................................................ 144
4.10 Function Associated with X-ray Diagnostic Table YSF-300
(Option) .................................................................................. 146
4.10.1 Features and Functions ................................................................. 146
4.10.2 Operation ....................................................................................... 149
4.11 Function Associated with X-ray Diagnostic Table YSF-500
(Option) .................................................................................. 153
4.11.1 Features and Functions ................................................................. 153
4.12 Operation of Dual X-ray Tube System ................................... 156
4.13 Program Data Synchronization .............................................. 157

Chapter 5 Troubleshooting
5.1 Information (Message) List .................................................... 160
5.2 Emergency Stop .................................................................... 168
5.3 Operation During Power Outage............................................ 169
5.4 The Area Where Cannot Be Registered
with Auto Positioning Function............................................... 170

CH-200/CHU-200 Operation Manual Contents-3


Contents

Chapter 6 Maintenance
6.1 Maintenance Check................................................................174
6.2 Daily Inspection (Inspected by User)......................................175
6.2.1 Cleaning and Disinfection .............................................................. 176
6.2.2 Inspection for Warning and Caution Labels ................................... 179
6.2.3 Checklist for Daily Inspection ......................................................... 180
6.3 Periodical Inspection ..............................................................181
6.4 Maintenance Parts List ...........................................................183

Chapter 7 Specifications
7.1 Environmental Conditions of EMC
(Electromagnetic Compatibility)..............................................186
7.2 Specifications .........................................................................191
7.2.1 Specifications................................................................................. 191
7.2.2 Distance from Spot to Reception Area........................................... 195
7.3 Labeling ..................................................................................196
7.4 Statement of Compliance [For Europe] ..................................200
7.4.1 Regulatory Information................................................................... 200
7.4.2 Company's Quality System ............................................................ 201
7.4.3 International Standards .................................................................. 201
7.4.4 Protection Against Radiation.......................................................... 201
7.5 Statement of Compliance with Standards ..............................202
7.6 Manufacturer Information .......................................................203
7.7 Information on Radiation ........................................................204
7.7.1 Deterministic Effect ........................................................................ 204
7.8 Product Safety ........................................................................205

Appendix
Equipment Malfunction Checklist ........................................................................ 208
Safety Explanation Record .................................................................................. 209

Index

Dimension

Contents-4 CH-200/CHU-200 Operation Manual


Outline
This chapter deals with the applications, features
and usage environment of the X-ray tube support
CH-200/CHU-200.
1 Outline

1.1 Applications
 Intended Use

The ceiling type x-ray tube support CH-200/CHU-200 is used to hold the X-ray tube
and collimator to position the X-ray tube unit and determine the radiography area after
locating the radiography position of patient when combined with an X-ray high voltage
generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray
radiography stand, an X-ray radiography table, and digital radiography system.

 Indications for Use

This unit is used for radiographic examinations of whole body except mammography
when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray
collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table,
and digital radiography system.

This unit must only be operated by qualified personnel, such as radiography


technicians or those with equivalent qualifications.

This system is used for total patient population including pediatric examination.

• Do NOT use the system for any other purpose than indicated above.
• Do NOT connect the system to any other equipment, either by
electrical or mechanical means, or modify the system.

Tube Support CH-200 and CHU-200 are hereafter referred to as "Tube


Support" or "equipment".

2 CH-200/CHU-200 Operation Manual


1.2 Features

1.2 Features
1
 Operation Panel

• The X-ray tube unit can be positioned by operation from the operation panel.
1
• Position information such as the exposure distance and angle of rotation are
displayed on the operation panel's screen.
1
• Operation panel carries the color liquid crystal touch panel.

• APR (anatomical program) can be displayed and changed on the operation


panel when the X-ray high-voltage generator is combined. 1
For details on the combination with the digital radiography unit or the X-ray
high-voltage generator, refer to "1.4 Configuration".

• The button arrangement on the operation panel can be modified when installing
the equipment.
1
• The X-ray tube support can be moved and lighted irradiation field lamp using the
rear operation panel. 6
(3-directional moving, rotation around the vertical axis lock release functions and
irradiation field lamps)

CH-200/CHU-200 Operation Manual 3


1 Outline

 Function

• The rotation around the vertical axis of the X-ray tube support can be unlocked
with the operation panel enabling even radiography at difficult angles demanded
in orthopedic operations. (Radiography with general radiography technique)

• The X-ray tube support can be moved vertically within a 1600 mm range.
Without using a footstool, it can perform radiography at a low position, such as
the knees, with the patient in a standing position.

• The vertical position of the X-ray tube support is selectable at the time of
installation making it suitable even for examination rooms with high ceilings.

• Aluminum rails are standard equipment enabling the quiet and smooth
movement of the X-ray tube support. They also enable accurate positioning for
conducting radiography.

• The magnetic brakes reduce noises upon locking.

• When combined with the Shimadzu collimator, the equipment can automatically
adjust the irradiation field using the auto collimation function.

* When the equipment is combined with a collimator other than the Shimadzu
collimator, the auto collimation function is not available. In this case, adjust
the irradiation field size using the knob provided on the collimator.

 Optional Function

• The vertical tracking function (option) enables automatic positioning of the X-ray
tube support when the X-ray radiography stand is combined.

• The horizontal tracking function (option) maintains a constant radiography


distance (SID) automatically when the X-ray radiography table is combined.

• The Bucky tracking function (option) enables automatic positioning of the X-ray
tube support and the Bucky when the X-ray radiography table is combined.

• The tomography function (option) enables tomography when the X-ray


radiography table is combined.

• The power assist function (option) enables the X-ray tube support to be moved
easily.

• The automatic positioning function (option) enables the X-ray tube to move to
the registered position just by pressing the button on the remote control after
selecting the APR.

4 CH-200/CHU-200 Operation Manual


1.3 Principle

• The tracking switch kit (option) enables the use of the auto collimation function,
the vertical tracking function and the horizontal tracking function in combination
with the X-ray radiography stands and tables of other manufacturers. 1
• By means of the irradiation field size selection function (option), the irradiation
field size can be selected the size below the cassette of X-ray radiography table
or stand.
1
• Patient information display function (option) displays the patient information on
the touch panel when combined with certain types of digital radiography unit. 1
• Associated with Shimadzu X-ray Diagnostic Table YSF-300 using YSF-300 link
option, the radiography is available. 1
"4.10 Function Associated with X-ray Diagnostic Table YSF-300 (Option)"

• The tomosynthesis function (option) enables tomosynthesis when combined with 1


certain types of digital radiography units.

• The energy subtraction radiography function (option) enables energy subtraction


radiography when combined with certain digital radiography units.
6
• The long view radiography function (option) enables long view radiography when
combined with certain digital radiography units. 1
• The serial radiography function (option) enables serial radiography when
combined with certain digital radiography units. 1
1.3 Principle
1
By releasing the locks on all system motions and manually operating the X-ray tube in
longitudinal motion, transverse motion, vertical motion, horizontal rotation and vertical
rotation, you can set the radiography position.

Note also that position information such as the exposure distance and angle of rotation
are displayed using a position detection mechanism.

CH-200/CHU-200 Operation Manual 5


1 Outline

1.4 Configuration
Depending on the combination with the X-ray tube support, X-ray high-voltage
generator and digital radiography unit, three modes "CH-200 tracking APR mode", and
"CH-200 non-tracking mode" are available. The difference between two modes is listed
below.

Depending the combination with the digital radiography unit or the X-ray high-
voltage generator, the operating procedure and touch panel screens of the
X-ray tube support may differ from those indicated in this manual. This
operation manual describes the operating procedure that applies when the
equipment is in "CH-200 tracking APR mode".

1 CH-200 Tracking APR Mode


When the equipment is in CH-200 tracking APR mode, selecting radiography technique
and changing radiography conditions can be done from the operation panel of the X-
ray high-voltage generator or the X-ray tube support, because the X-ray tube support
and the X-ray high-voltage generator are APR tracking.

X-ray high-voltage generator X-ray tube support

APR tracking

Digital radiography unit

: APR tracking

: Tracking

Component Model

• UD150B-40/L-40/V-40
X-ray high-voltage
• D150BC-40/LC-40/VC-40/BC-41/LC-41/VC-41 ꞏ GSC-
generator
2002L

• None
Digital radiography unit • Digital radiography units from manufacturers other than
Shimadzuthat can be combined

X-ray tube support CH-200/CHU-200

6 CH-200/CHU-200 Operation Manual


1.4 Configuration

2 CH-200 Non-tracking Mode


When the equipment is in CH-200 non-tracking mode, the X-ray tube support and the
X-ray high-voltage generator must be operated individually, because the X-ray tube
1
support, the digital radiography unit and the X-ray high-voltage generator are not
tracking.

Only the radiography technique on the equipment can be displayed and changed from
the operation panel of the X-ray tube support.
1
*• The radiography technique and the radiography condition on the X-ray
high-voltage generator cannot be changed from the X-ray tube support.
The radiography technique on the X-ray tube support cannot be
1
changed from the X-ray high-voltage generator.
• The optional tomography function cannot be installed.
1
X-ray high-voltage generator X-ray tube support
(other than 1)
1

6
Digital radiography unit

Component Model
1
X-ray high-voltage
X-ray high-voltage generator (other than 1)
generator

• None
Digital radiography unit • Digital radiography units from manufacturers other than
Shimadzu that can be combined

X-ray tube support CH-200/CHU-200

CH-200/CHU-200 Operation Manual 7


1 Outline

1.5 Environmental Conditions


To obtain proper performance, be sure to use the equipment under the specified
environmental conditions.

1.5.1 Operating Environment

Use the equipment under the environmental conditions listed below:

The installation of a dedicated air-conditioner in the examination room is recommended


if the building air-conditioning cannot meet the necessary environmental conditions.

Do NOT use the equipment in an oxygen-rich environment.


The use of the equipment in an oxygen-rich environment may cause fatal or
serious injuries or damage to the equipment due to easy ignition.

Even under the prescribed conditions, avoid rapid changes of


temperature or humidity.
Condensation may occur and cause failure. Also, rust or corrosion may occur
inside the equipment.

Item Specifications

Atmosphere No explosive or corrosive gases

Ambient Temperature 10 C to 40 C

Relative Humidity 30 % to 85 % (with no condensation)

Atmospheric Pressure 800 hPa to 1060 hPa

Environment Luminosity 150 lx to 500 lx

Ambient Noise Level Under 70 dB

8 CH-200/CHU-200 Operation Manual


1.5 Environmental Conditions

1.5.2 Transportation and Storage Environment


1
Item Specifications

Ambient Temperature -10 C to 60 C 1


Relative Humidity 10 % to 95 % (with no condensation)

1
Atmospheric Pressure 700 hPa to 1060 hPa

1.5.3 Power Supply


1

Item Specifications
1
Phase Single phase AC

Frequency 50/60 Hz

Standard Voltages
100, 200, 220, 230, 240 V
(Set to one of the voltages when installing)
6
Voltage Fluctuation
10 % of nominal voltage
Range

Supply Capacity
Single phase 100 V: 0.2 kVA
1
Single phase 200 / 220 / 230 / 240 V: 1.0 kVA

Single phase 100 V: Max. 5 


Power Supply Impedance
Single phase 200 / 220 / 230 / 240 V: Max. 4 
1

Be sure to use the power supply specified in the operation manual. 1


Using a power supply other than the one specified may cause equipment
malfunction or serious accidents such as fire, smoke emission, or explosions.

1.5.4 Grounding

Grounding resistance: less than 100 

Be sure to connect the equipment only to a (commercial) power outlet


with a ground terminal.
If the outlet does not have a ground terminal, electric shock may occur.

CH-200/CHU-200 Operation Manual 9


1 Outline

1.5.5 Installation Requirements

 Ceiling Structure

The ceiling structure must permit anchor bolts to be installed for anchoring the
equipment to the examination room ceiling.

Preferably, the ceiling should be at least 50 mm thick.

Make sure the ceiling at the installation site as follows.

• Ceiling levelness: 5/1000 mm max. (measured with a laser marker etc.)

 Anchoring

The ceiling of the room in which the equipment is installed must have anchor bolts that
can support the weight of the equipment. It is necessary to secure 500 kg pull out
strength per anchor bolt. Refer to the table below and select anchor bolts. Also, hook
anchor bolts onto a foundation rebar in order to attain greater load capacity.

Collar Dimensions Maximum


Pullout
Maximum
Boring Capacity Drill Anchor
Nominal Shear
Outer Screw Depth for Diameter Pitch
Dimensions Capacity
Diameter Depth (mm) Concrete (mm) (cm)
(kg)
Strength of
200 kg/m2

W1/4 11 8 30 1,100 800 11.3 9

W5/16 12 10 40 1,400 1,200 12.3 11

W3/8 14.3 11 45 2,200 1,800 14.6 13

W1/2 17.5 16 55 3,200 3,200 18.0 19

W5/8 21.5 20 65 4,500 5,200 22.0 23

W3/4 22.5 26 90 6,800 7,700 26.0 28

W7/8 29.5 30 100 8,100 9,600 29.0 32

The maximum allowable capacity is 40 % of corresponding maximum capacity.

10 CH-200/CHU-200 Operation Manual


1.6 Classification of Equipment

1.6 Classification of Equipment


1
This equipment is classified as follows, based on safety standards for electrical medical
equipment.
1
 Protection Method Against Electric Shock

Class I equipment

 Classification of Applied Parts


1
No Applied Parts

 Operation Mode
1
Continuous operation

 Degree of Protection Against Liquid Ingress


1
Ordinary equipment

 For Use in an Oxygen-rich Environment


6

1
Do NOT use the equipment in an oxygen-rich environment.
The use in an oxygen-rich environment may cause fatal or serious injuries or
damage to the equipment due to easy ignition.
1
 For Use in Flammable Atmosphere

Not to be used in presence of flammable anaesthetic gas


1

Do NOT use the equipment or system in the presence of flammable


anesthetics gas.
It may cause an explosion.

 Classification of Installation Type

Permanently installed equipment

CH-200/CHU-200 Operation Manual 11


1 Outline

1.7 Operator Profile

Item Details
Age that person can obtain the license of
Age
Radiological Technologist or a license equal to it.
Sex No limitation
Nationality No limitation
Radiological Technologist or person who has a
license equal to it.
Education
The capacity to read and understand the
operation manual.
Radiological Technologist or person who has a
Knowledge
license equal to it.
Language Can read and understand English.
Necessary.
Every operator needs to take training for
Experience
operating the equipment before using the
equipment.
Corrected visibility is over 0.7 in the decimal
number.
Permissible impairments
Impaired by 40 % resulting in 60 % of normal
hearing at 500 Hz to 2 kHz.

12 CH-200/CHU-200 Operation Manual


1.8 Symbols

1.8 Symbols
1
Symbols used on this system are shown and described as follows.

Symbol Location Meaning


1

On name plate Alternating current 1

Inside the equipment, where


protective earth conductor in Protective earth ground
1
power cord is connected

On warning and caution labels


Observe described items, or 1
refer to operation manual.

On name plate Refer to operation manual.


6
On name plate Refer to operation manual.
1

On name plate
Year and month of
manufacture 1

On name plate Manufacturer


1

On name plate Serial number

Next to name plate Refer to operation manual.

CH-200/CHU-200 Operation Manual 13


1 Outline

This page is intentionally left blank.

14 CH-200/CHU-200 Operation Manual


Configuration
This chapter describes the configuration and
operating panel of the X-ray tube support.
2 Configuration

2.1 Configuration
This X-ray tube support consists of the following components.

2 Moving Rail

Transversal

1 Fixed Rail

Longitudinal

Vertical
3 X-ray Tube Support Pull-up Section

5 Operation Panel
4 X-ray Tube Support Section

6 Rear Operation Panel

7 Control unit

*: Equipment suitable for use in the patient environment (except 7).

Fig. 2.1 Configuration

1 Fixed Rail
This rail is fixed onto the ceiling. The moving rail moves along this rail.

2 Moving Rail
Rail that moves along the fixed rail. The X-ray tube support section moves along this
rail.

3 X-ray Tube Support Pull-up Section


Pull-up the X-ray tube support and moves it vertically.

4 X-ray Tube Support Section


Holds the X-ray tube support and moves it vertically. And rotates it around the vertical
axis as well as the horizontal axis.

16 CH-200/CHU-200 Operation Manual


2.1 Configuration

5 Operation Panel
Used to control the movement of the equipment or to radiography setting.
Radiography conditions can be set with touch panel type color LCD, and the set screen
2
of the APR can be displayed on color LCD by selecting screen. It is also indicated a
horizontal axis rotational angle indicator for the X-ray tube support.

2
* When the equipment is in CH-200 non-tracking mode, the set screen of the APR
is not available.

Equipped with lock release buttons for the X-ray tube support's vertical direction, its 2
longitudinal and transversal movement as well as its horizontal and vertical axis
rotation, and a collimator lamp button.
The arrangement for each button controlling movement and rotation can be set at will
when installing the equipment. 2
6 Rear Operation Panel
Equipped with a vertical axis rotation release button, a lamp button and a 3-directional
release button. When the power assist function (option) is selected, the stop switch,
2
stop release switch, and single-axis move button will be added.

"3.2.6 Power Assist Function (Option)"


6
7 Control unit
Controls the X-ray tube support entirely.

 Other Components
2
• Servo tracking driver (Option): Moves the X-ray tube support upward/downward


at horizontal/vertical tracking.
Bucky tracking driver (Option): Drives the Bucky of the X-ray radiography table.
2
• Tomography driver (Option): Drives the X-ray tube support at tomography.

* When the equipment is in CH-200 non-tracking mode, the tomography 2


function is not available.

• Power assist driver (Option): Equipped with a sensor to detect a force applied to
the operation panel, this unit drives the X-ray tube support when the power
assist function is enabled.
• Auto positioning driver (Option): Drives the X-ray tube support when the Auto
Positioning Function is selected.
• Long view drive section (Option): Drives the system for long view radiography.
• Tracking switch kit (Option): The switch box is equipped with the tracking switch
and the stop switch. Also, it is equipped with the interface for the X-ray
radiography stand or the X-ray radiography table.
• Dongle Unit (Functional Extension Option): Enables the following function.
• Irradiation field size selection function
• Patient information display function

CH-200/CHU-200 Operation Manual 17


2 Configuration

• Function for performing long view radiography by holding down the


radiography button once
• Tomosynthesis function
• Energy subtraction radiography function
• Serial radiography function

 Explanation of Terminology:

• Longitudinal: It moves along with the fixed rail.


• Transversal: It moves along with the moving rail.
• Vertical: X-ray tube support moves up and down.

 Attachment Position of X-ray Tube Support

The attachment position of X-ray tube support can be chosen from the four following
places at the time of installation of X-ray tube support.

Upper left Upper right (standard)

Lower left Lower right

Fig. 2.2 Installation of X-ray Tube Support

18 CH-200/CHU-200 Operation Manual


2.2 Operation Panel

2.2 Operation Panel


2
The function of each button found on the operation panel is explained. The buttons on
the rear operation panel located on the back surface of the equipment perform the
identical functions.
Their functions are explained as per the standard arrangement. If the arrangement is
2
changed, read the explanation with respect to that new arrangement.

1 2 2

2
3 6

4 7
6
5 8

2
9 0 a b c

Fig. 2.3 Operation Panel 2


1 (Vertical release) button
This button releases the lock when moving the X-ray tube support vertically. 2
2 (3-directional release) button
This button simultaneously releases the locks for the vertical, longitudinal and
transversal directions.

3 (H-axis rotation release) button


This button is to release the lock when rotating the X-ray tube support around the
horizontal axis.

4 (V-axis rotation release) button


This button is to release the lock when rotating the X-ray tube support around the
vertical axis.

CH-200/CHU-200 Operation Manual 19


2 Configuration

5 (Vertical release) button


This button performs the same function as 1.

6 (Longitudinal release) button


This button is to release the lock when moving the X-ray tube support longitudinally.

7 (Transversal release) button


This button is to release the lock when moving the X-ray tube support transversely.

8 (3-directional release) button


This button performs the same function as 2.

9 (Tracking) button
This button is to track the X-ray radiography table or X-ray radiography stand.

0 (Collimator lamp) button


This button is to turn on the collimator lamp. The same button is also equipped on the
operation panel of collimator.

* When the X-ray tube support is combined with a collimator other than the
Shimadzu autocollimator, this function is available as an option. If you require this
function, please contact your Shimadzu service representative.

Collimator Lamp Button

Fig. 2.4 Collimator Lamp Button

a (Detent mode) button


This button is to turn ON/OFF the preset position stop mode.
The detent mode is used to stop the X-ray tube support at a specific longitudinal,
transversal and vertical setting position. It can also stop the equipment at 0 and 90
rotational angles around the horizontal axis.

*• Using the detent mode key on the touch panel, perform the above procedures.
"2.3 Touch Panel".
• Replacement with the (Upward retraction) button is possible at installation.
"3.5 Upward retraction function".

20 CH-200/CHU-200 Operation Manual


2.2 Operation Panel

bTouch panel
Liquid crystal display operated by touching.
2

• Only single operation should be performed using a touch panel after your
2
confirmation of the display. If multiple operations are tried simultaneously,
this system does not work properly.

• Operate the touch panel by pressing down gently with a fingertip or a hard, 2
blunt object (such as a stylus pen or the back of a finger nail). Otherwise,
unintended locations may respond.

• If a sharp object is used to touch the touch panel (such as a fingernail tip), it
can cause a scratch or, if repeated, can break the panel.
2
c Grip
Hold this part to move or rotate the X-ray tube support. 2
 Rear Operation Panel

Equipped with a vertical axis rotation release button, a lamp button and a 3-directional 6
release button. When the power assist function (option) is selected, the stop switch,
stop release switch, and single-axis move button will be added.

"3.2.6 Power Assist Function (Option)" 2

2
1

2
2
3

Fig. 2.5 Rear Operation Panel

1 (V-axis rotation release) button


This button is to release the lock when rotating the X-ray tube support around the
vertical axis.

2 (Collimator lamp) button


This button is to turn on the collimator lamp.

* When combined with a collimator other than the Shimadzu autocollimator, this
function is available as an option.

3 (3-directional release) button


This button simultaneously releases the locks for the vertical, longitudinal and
transversal directions.

CH-200/CHU-200 Operation Manual 21


2 Configuration

2.3 Touch Panel


 Positioned Screen (Typical Screen)

CH-200 Tracking APR Mode Screen

1
6
7
2

3 8
9
4
0
a
5
b
c
e

Fig. 2.6 CH-200 Tracking APR Mode

CH-200 Non-tracking Mode Screen

No switching tab

Fig. 2.7 CH-200 Non-tracking Mode

1 Technique key
Shows the technique and No. of X-ray tube currently selected. They can be changed
with this key.

2 Longitudinal detent position indicator


Lights up when the X-ray tube support has reached the longitudinal setting position.

3 Transversal detent position indicator


Lights up when the X-ray tube support has reached the transversal setting position.

4 Vertical position indicator


Lights up when the X-ray tube support has reached the vertical setting position.

22 CH-200/CHU-200 Operation Manual


2.3 Touch Panel

5 Detent mode key


Changes the detent mode ON/OFF.
2
Detent mode ON:

Detent mode OFF:

This key functions in the same manner as the detent mode button on 2
* operation panel.
"2.2 Operation Panel"

6 Quick ready key 2


This is convenient for example if the patient is a child cannot be left alone.
• If the Quick ready key is ON, the radiography preparations are completed in
approximately half the normal time.
Quick ready key ON:
2
• The quick ready function is automatically cancelled when the preset interval has
elapsed. To cancel the function manually, press the Quick ready key to turn it
OFF. 2
If the X-ray tube is changed after Quick ready is turned ON, the function will turn
OFF.
Quick ready key OFF:
6
*• Selection of whether the Quick ready key is used/unused is as per the
installation settings.
• Quick ready function is not available when the equipment is in CH-200 non-
tracking mode.
2

2
The X-ray tube unit may be damaged if the Quick ready function is used
more than necessary.
"X-ray High Voltage Generator Operation Manual" 2
When combined with the X-ray high voltage generator except Shimadzu X-ray
high voltage generator (Touch panel type), the quick ready function is not
available.

7 Control cabinet basic information indicator


Shows the basic information of control cabinet. If a communication error occurs with
the X-ray high-voltage generator, the basic information of the control cabinet and the
Quick ready key are not available.

* When the equipment is in CH-200 non-tracking mode, this basic information


indicator is not available.

CH-200/CHU-200 Operation Manual 23


2 Configuration

8 Radiography distance indicator


This indicator will show the Source Image Distance (SID) between the focus of X-ray
tube unit and the X-ray detector of the X-ray radiography stand or table when the
equipment is facing the X-ray radiography table or stand.
For the general radiography technique, the radiography distance displays as SSD or
STD.

SID: Radiography distance between the SID: Radiography distance between the
X-ray tube focus and the X-ray detector of X-ray tube focus and the X-ray detector of
X-ray radiography stand X-ray radiography table

SID

SID

X-ray detector
X-ray detector

SID: Radiography distance between the SID: Radiography distance between the
X-ray tube focus and the X-ray detector of X-ray tube focus and the X-ray detector of
X-ray radiography stand for oblique X-ray radiography stand when tilting the
projection radiography Bucky

X-ray detector X-ray detector

SID SID

SSD: Radiography distance between the STD: Radiography distance between the
X-ray tube focus and the front surface on X-ray tube focus and the tabletop of the X-ray
Bucky cover of X-ray radiography stand radiography table

SSD

Tabletop STD

Front surface on the


Bucky

* The radiography distance indication can be set either cm unit or inch unit at
installation work.

24 CH-200/CHU-200 Operation Manual


2.3 Touch Panel

9 Rotation angle indicator


Shows the angle at which the X-ray tube support was rotated around the horizontal
axis. Pressing the symbol of angle indication toggles between horizontal and vertical
2
direction.

Technique key
2

2
* Pressing the technique key toggles between horizontal and vertical direction. The
function can be set on installation.
Contact your Shimadzu service representative for setting. 2
0 Collimator fully open button
Opens the collimator leaves to the cassette size on the displayed screen or to the 2
irradiation field size. This button is displayed when, for example, the leaves have been
closed down manually.

a Cassette size indicator 6


Shows the cassette size and the reference position of the irradiation field currently
inserted in the X-ray radiography table or X-ray radiography stand.

*
• The type (inch or cm) of size for cassette can be set at installation. To
make this setting, please contact your Shimadzu service representative.
2
• When the irradiation field size selection function is enabled, the
irradiation field size, the reference position of the irradiation field, the
portrait/landscape orientation and the size type (inch/cm) can be 2
changed. The cassette size display switches to the irradiation field size
display. (Refer to "3.6.2 Irradiation Field Size Selection Function
(Option)".)

• When the power assist function (option) is selected, the right half of the
2
display area shows the cassette size, and the left half the assist level.
(Refer to "3.2.6 Power Assist Function (Option)".)

b Type of grid/status indicator


Indicates the grid type or the status of insertion.

Display Meaning

Type of grid Name for type of grid registered

No grid No grid inserted

Exchanging Exchanging grid

CH-200/CHU-200 Operation Manual 25


2 Configuration

* If the equipment other than the Shimadzu X-ray radiography stand or Shimadzu
X-ray radiography table is provided with the removable grid, the indication for grid
differs, or the specified type of grid displays or the type is hidden.

c Radiography condition indicator


• In CH-200 Tracking APR Mode
Shows the radiography conditions set with X-ray high-voltage generator. Select the tab
to display the screen as required.

Fig. 2.8 Radiography Conditions Screen (CH-200 Tracking APR Mode)


If you touch any button (1 - 8 shown in Fig 2.3) on the operation panel when
radiography technique and the patient information are displayed on the touch panel,
the display will be automatically switched to the positioned screen, and you can
operate the X-ray tube support while checking its position data.

* This function can be set on installation. Contact your Shimadzu service


representative for setting.

Fig. 2.9 Panel Display

26 CH-200/CHU-200 Operation Manual


2.3 Touch Panel

However, performing with each button (9 - a shown in Fig 2.3) operations when
radiography technique and the patient information are displayed will not switch the
display to the positioned screen. 2
* To manually switch the display to the positioned screen, press the switching tab.
2

2
Fig. 2.10 Switching Tab

* When the equipment is in CH-200 non-tracking mode, the switching tab to switch
display is not available.
2
d Irradiation field size display button
By pressing the display button you can change the irradiation field size, reference
6
position of the irradiation field, portrait/landscape orientation and size type.

For details on the irradiation field size display button, refer to "3.6.2 Irradiation
Field Size Selection Function (Option)". 2
* The irradiation field size selection function is available as an option. If you require
this setting, please contact your Shimadzu service representative.
2
e Patient information tab
Press the patient information tab to display the patient information selected by the
digital radiography unit. 2

Patient information

Fig. 2.11 Patient information tab

CH-200/CHU-200 Operation Manual 27


2 Configuration

If the information on the patient information tab is inconsistent with that on the
digital radiography unit, follow the information on the digital radiography unit.

*• When the equipment is in CH-200 non-tracking mode, patient information


cannot be displayed.
• The Patient information display function is available as an option. If you require
this setting, please contact your Shimadzu service representative.
• A function is available for automatically displaying the patient information tab
upon receiving patient information from the digital radiography unit.
Contact your Shimadzu service representative to set up this function.

 Status Indication for System

Various system statuses are displayed on the upper right of a touch panel by
radiography conditions, a position, etc. of X-ray tube unit.

Fig. 2.12 Information Screen


For details on each state, refer to "3.6 Auto Collimation Function" and "5.1
Information (Message) List".

When combined with a collimator other than the Shimadzu autocollimator, no


indication other than "Not Ready" is displayed.

28 CH-200/CHU-200 Operation Manual


2.3 Touch Panel

 Information Screen

The button is displayed when X-ray tube support is not aligned or detects error. 2
1

2
Fig. 2.13 Information Screen

1 Information button
2
By pressing the button, the pop-up window will be open and displays messages.
According to the message displayed, operate the X-ray tube support.

2
Fig. 2.14 Sample Message on Information Screen
• Confirming the message displayed, press button to close the pop-up
2
window.
• Operating the X-ray tube support according to the message displayed, the
button is distinguished and the X-ray tube support is aligned.
For information, refer to "5.1 Information (Message) List".

CH-200/CHU-200 Operation Manual 29


2 Configuration

2.4 Rear Mount Type, Low-ceiling Type,


and Front Mount L Type
There are three other types of the X-ray tube support in addition to the standard type:
Rear mount type, Low-ceiling type, and Front mount L type, which hold the X-ray tube
assembly in different ways.

2.4.1 Rear Mount Type

Rear mount type is available for the operation of the X-ray tube support installed by
walls in examination rooms.

30 CH-200/CHU-200 Operation Manual


2.4 Rear Mount Type, Low-ceiling Type, and Front Mount L Type

 Attachment Position of X-ray Tube Assembly

The X-ray tube assembly with Rear mount type can be attached to one of four positions 2
shown below when installing the assembly.

2
Upper left (standard) Upper right
2

2
Lower left Lower right

Fig. 2.15 Installation of X-ray Tube Assembly 2

CH-200/CHU-200 Operation Manual 31


2 Configuration

2.4.2 Low-ceiling Type

Low-ceiling type holds the X-ray tube assembly at a position higher than the standard
type. This is to keep the distance between a radiography table and the X-ray tube focus
(1 m as standard) in examination rooms with low ceilings.

 Difference of the Specification from That of Standard Type

The X-ray tube assembly position:15 cm higher


The rotation angle around the vertical axis:- 90° to + 30°

 Attachment Position of X-ray Tube Assembly

The X-ray tube assembly with Low-ceiling type can be attached to one of two positions
shown below when installing the assembly.

Fig. 2.16 Installation of X-ray Tube Assembly

32 CH-200/CHU-200 Operation Manual


2.4 Rear Mount Type, Low-ceiling Type, and Front Mount L Type

2.4.3 Front Mount L Type


2
Front mount type holds the X-ray tube assembly at a position 80 mm lower than the
standard type. Eases positioning by preventing interference between the lower edge of
the tube support and the tabletop during lateral radiography using the radiography
table. 2

 Difference of the Specification from That of Standard Type 2


The X-ray tube assembly position:8 cm lower

 Attachment Position of X-ray Tube Assembly 2


The X-ray tube assembly with front mount L type can be attached to one of two
positions shown below when installing the assembly.

Upper left Upper right (standard)

Fig. 2.17 Installation of X-ray Tube Assembly

CH-200/CHU-200 Operation Manual 33


2 Configuration

2.5 Layout When Using Optional


Functions
When using the following optional functions, there are layout limitations according to the
combination with the X-ray radiography stand or table.

• Bucky tracking radiography


• Tomography and tomosynthesis radiography
• Auto positioning function
• Long view radiography

 Limitation for X-ray radiography stand

The X-ray tube support cannot be located at the place where its operation position
faces the back side of the X-ray detection surface of the radiography stand.

Correct Incorrect

Bucky back side

Bucky

Operation surface Operation surface

34 CH-200/CHU-200 Operation Manual


2.5 Layout When Using Optional Functions

 Limitation for X-ray radiography table

The X-ray tube support cannot be installed in the direction where its operation position 2
comes opposite to that of the radiography table. In this case, the system will not work
properly even when rotating the X-ray tube support around vertical axis 180 degrees.

2
Correct Incorrect

2
Bucky

Operation
surface Operation
surface 2
Bucky

Operating the X-ray tube support


2
by rotating it around vertical axis
180 degrees against the X-ray
radiography table is not allowed.
6
 Limitation when an optional function is enabled with both X-ray
radiography stand and table

When using an optional function by combining the X-ray radiography stand and table, 2
they cannot be arranged in such a way that their operation directions become 180
degrees opposite one another.

Operating with X-ray radiography table Operating with X-ray radiography stand
2
Rotating the X-ray tube support

Incorrect
180 degrees around vertical
axis and operating it towards
2
the X-ray radiography stand is
not allowed.

Operation
surface

Bucky

Operation directions 180 degrees opposite

CH-200/CHU-200 Operation Manual 35


2 Configuration

 Limitations when combining X-ray radiography stand and table

The following shows the layout examples considering the above-mentioned limitations:
Limitation for X-ray radiography stand, Limitation for X-ray radiography table, and
Limitation when an optional function is enabled with both X-ray radiography stand and
table.

The examples in this section assume that the X-ray tube support is mounted in the
upper right position (standard).

Attachment position of
X-ray tube assembly: X-ray tube support
Upper right (Standard) pull-up section

X-ray tube support


section

When the position is otherwise, the same concept should be applied although the
illustrations may not reflect the actual cases.

In the layout examples (Examples 1 and 2), the statuses of the X-ray tube
support are differentiated as below:

Facing the X-ray radiography stand Facing the X-ray radiography table

36 CH-200/CHU-200 Operation Manual


2.5 Layout When Using Optional Functions

Example 1 X-ray radiography stand: Longitudinal direction

Correct Correct 2
X-ray radiography
table X-ray radiography

X-ray
X-ray
radiography
table
2
radiography stand
stand

2
Correct Incorrect
X-ray
X-ray radiography radiography X-ray radiography
table stand table
2

Not allowed because the operation 2


X-ray radiography stand directions of the X-ray radiography stand
and table are 180 degrees opposite one
another. Not allowed also because the
attachment position of the X-ray tube
support is the upper right (Standard) and
2
operation position of the X-ray tube
support comes to 180 degrees opposite
around vertical axis towards the X-ray
radiography table. 2

CH-200/CHU-200 Operation Manual 37


2 Configuration

Example 2 X-ray radiography stand: Transversal direction

Incorrect Correct
X-ray radiography stand X-ray radiography stand

X-ray radiography
table
X-ray radiography
Not allowed because the attachment table
position of the X-ray tube support is the
upper right (Standard) and operation
position of the X-ray tube support comes
to 180 degrees opposite around vertical
axis towards the X-ray radiography
table.

Correct Incorrect
X-ray radiography stand X-ray radiography X-ray
stand radiography
table

Not allowed because the operation


X-ray radiography
table directions of the X-ray radiography stand
and table are 180 degrees opposite one
another.

38 CH-200/CHU-200 Operation Manual


2.5 Layout When Using Optional Functions

 Limitation when tomography function or tomosynthesis


radiography function is enabled
2
To use the tomography function or tomosynthesis radiography function, the following
limitation is added to those from "Limitation for X-ray radiography stand" to "Limitation
when combining X-ray radiography stand and table".
2
Layouts of X-ray radiography table

Allowed when the X-ray tube support fixed rail (longitudinal direction) and the X-ray
radiography table longitudinal direction are parallel.
2
Longitudinal direction Fixed rail (longitudinal direction)

Correct X-ray radiography 2


table

Longitudinal direction 2
Correct X-ray radiography
table 6

Not allowed when the X-ray tube support fixed rail (longitudinal direction) and the X-ray
2
radiography table longitudinal direction cross at right angles.

Longitudinal direction Fixed rail (longitudinal direction)


2
Incorrect

X-ray radiography
2
Longitudinal direction table

Incorrect X-ray radiography


table

CH-200/CHU-200 Operation Manual 39


2 Configuration

 Limitation when auto positioning function is enabled

To use the auto positioning function, the following limitation is added to those from
"Limitation for X-ray radiography stand" to "Limitation when combining X-ray
radiography stand and table". Note that this function is not compatible with the 2 X-ray
tube system.

The tube support unit is attached in the upper right position (standard).

Tube support unit

The tube support unit is attached in the lower right position

Tube support unit

40 CH-200/CHU-200 Operation Manual


2.5 Layout When Using Optional Functions

Layouts of X-ray radiography stand

Longitudinal direction layout


2
X-ray radiography stand: Left side X-ray radiography stand: Right side

Longitudinal direction Longitudinal direction X-ray radiography


stand 2

2
+ 90 - 90
- 90 + 90
0
0 2
Transversal direction layout

X-ray radiography stand: Left (far side) X-ray radiography stand: Right (far side)
2
+ 90
6
0
+ 90

- 90
2
0

Transversal
Transversal
direction
- 90 direction
2
X-ray radiography stand: Left (front side) X-ray radiography stand:
Right (front side)

- 90
2
- 90
0

0
+ 90
Transversal Transversal
direction + 90 direction

CH-200/CHU-200 Operation Manual 41


2 Configuration

The angles in the figures indicate angles and directions around vertical axis.
There are areas which cannot be registered with the auto positioning function.
"5.4 The Area Where Cannot Be Registered with Auto Positioning Function"

42 CH-200/CHU-200 Operation Manual


Operation
This chapter describes how to move or rotate the
device as well as how to operate the collimator
shutters automatically and the collimator lamp on/off.
3 Operation

3.1 Power ON/OFF

Before turning the power on, confirm that the following power switch is
in ON position.
• Power switch in the control cabinet of the X-ray tube support

Refer also to the manual of each component.


The location of the power button may differ depending on the combination.

3.1.1 Power ON

When turning the power supply of the X-ray tube support ON, the power supply of the
X-ray high-voltage generator, connected to the equipment, is also turned ON. (press to
light up or )

3.1.2 Power OFF

When turning the power supply of the X-ray tube support OFF, the power supply of the
X-ray high-voltage generator, connected to the equipment, is also turned OFF. (press
or )

Wait at least 10 seconds or more between turning off the power and
turning it on again. Wait 10 seconds or more before turning the power on
after pressing the stop button and releasing it. Otherwise, the equipment
may malfunction.

44 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3.2 Moving X-ray Tube Support


3

3
When moving the X-ray tube support, take ample precautions to prevent
the equipment from contacting with other devices in the surrounding
area.
3

The distance indication plates of the longitudinal, transversal, and


vertical movement are attached on the fixed rail, moving rail, and column.
3
When moving the X-ray tube support, align the X-ray tube support with
reference to those plates.
6

3.2.1 Vertical Movement


3

3
Always be very careful when moving the equipment to avoid contact with
the patient or operator and to ensure that the patient or operator does not
become caught between the equipment and any neighboring devices.
3
Otherwise, it may cause injury.

CH-200/CHU-200 Operation Manual 45


3 Operation

If the message "Alert: Wire. Stop unit and call service representative" is
displayed on the operation touch panel, stop using the equipment and
contact your Shimadzu service representative.
• The equipment is suspended by two wire ropes. If either of these ropes break,
radiography is disabled.
• A buzzer sounds continuously when the warning message is displayed on the
operation touch panel. Any continued use of the equipment while this message
is displayed may result in serious injury to the patient or operator.

Alert: Wire. Stop unit and


call service representative.

• When combined with the X-ray high-voltage generator UD150B-40/V-40/L-40,


press the hand switch to display the warning message on the touch panel of the
X-ray high-voltage generator UD150B-40/V-40/L-40.

Do NOT raise the second and third sections of the column.


Raising these sections may cause a problem to occur within the column. In the
event that a problem occurs when operating the column, stop using the
equipment and contact your Shimadzu service representative.

Second and third


column sections

46 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

Moves the X-ray tube support vertically.


The lock is released only while the (Vertical release) button is pressed. Lifting your
finger off the button will enable the lock. 3
1 While pressing the button on the operation panel, hold the grip
and move the X-ray tube support vertically.
Pushing button and release immediately, the lock is released for 5 seconds.
3

3
Fig. 3.1 Vertical Movement

The brake of an X-ray tube support is device for carrying out stillness
maintenance of the X-ray tube unit. When it is used in order to move an X-ray
3
tube unit up and down and to brake it, sound may occur.

If button is pushed once again while the lock is canceled for 5 seconds, it 3
is locked again.

2 When reaching the intended vertical position, lift your finger from the
button.
This will lock the X-ray tube support.

* The 5 seconds lock release function can be invalidated at installation work.


Contact your Shimadzu service representative.

CH-200/CHU-200 Operation Manual 47


3 Operation

3.2.2 Longitudinal Movement

Moves the X-ray tube support longitudinally.

The lock is released only when (Longitudinal release) button is pressed. Lifting your
finger off the button will enable the lock.

To stop the equipment at a longitudinal setting position, refer to "3.4.1 Moving to


Longitudinal/Transversal Detent Positions".

1 While pressing the button on the operation panel, hold the grip
and move the X-ray tube support longitudinally.

Fig. 3.2 Longitudinal Movement

2 When reaching the intended longitudinal position, lift your finger


from the button.
This will lock the X-ray tube support.

48 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3.2.3 Transversal Movement


3
Moves the X-ray tube support transversally.

The lock is released only while the (Transversal release) button is pressed. Lifting
your finger off the button will enable lock. 3
To stop the equipment at a transversal preset position, refer to "3.4.1 Moving to
Longitudinal/Transversal Detent Positions".

3
1 While pressing the button on the operation panel, hold the grip
and move the X-ray tube support transversally.
3

6
Fig. 3.3 Transversal Movement
3
2 When reaching the intended transversal position, lift your finger from
the button.
This will lock the X-ray tube support.
3

CH-200/CHU-200 Operation Manual 49


3 Operation

3.2.4 3-Directional Movement

This function simultaneously enables the vertical, longitudinal, and transversal


movement of the X-ray tube support.

The lock is released only while pressing the (3-directional release) button. Lifting
your finger off the button will enable lock.

1 While pressing the button on the operation panel, hold the grip
and move the X-ray tube support vertically, longitudinally and
transversally.

Operation Panel Rear Operation Panel

Fig. 3.4 3-Directional Movement

2 When reaching the intended position, release your finger from the
button.
This will lock the X-ray tube support.

50 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3.2.5 2-Directional Movement


3
X-ray tube support can move 2-direction simultaneously. This function can be set by
installing and two or more functions are combined with one button. Please contact your
Shimadzu service representative for setting.
3
 Vertical/Longitudinal Movement

The lock is released only while pressing the (vertical/longitudinal release)


button. Lift your finger to lock the X-ray tube support. 3
1 While pressing the button on the operation panel, hold the
grip and move the X-ray tube support vertically and longitudinally. 3

Fig. 3.5 Vertical/Longitudinal Movement


3
2 When reaching the intended position, release your finger from the
button. 3
This will lock the X-ray tube support.

CH-200/CHU-200 Operation Manual 51


3 Operation

 Vertical/Transversal Movement

The lock is released only while pressing the (vertical/transversal release)


button. Lift your finger to lock the X-ray tube support.

1 While pressing the button on the operation panel, hold the


grip and move the X-ray tube support vertically and transversally.

Fig. 3.6 Vertical/Transversal Movement

2 When reaching the intended position, release your finger from the
button.
This will lock the X-ray tube support.

 Longitudinal/Transversal Movement

The lock is released only while pressing the (longitudinal/transversal release)


button. Lift your finger to lock the X-ray tube support.

1 While pressing the button on the operation panel, hold the


grip and move the X-ray tube support longitudinally and
transversally.

Fig. 3.7 Longitudinal/Transversal Movement

2 When reaching the intended position, release your finger from the
button.
This will lock the X-ray tube support.

52 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3.2.6 Power Assist Function (Option)


3
Pressing the vertical, longitudinal, and transversal release buttons with the power
assist function enabled will allow you to move the X-ray tube support easily. The assist
level can be set or switched in three steps.
3

3
When an emergency has occurred due to failure in switches, etc.,
perform an emergency stop.
For details on emergency stop, refer to "5.2 Emergency Stop".
3

3
Confirm that no patient is close to the X-ray tube support when the
equipment is in motion.
Injury may result if the X-ray tube support makes contact with a person.
6

3
Confirm that the area around the X-ray tube support is free from
obstructions.
Damage may result if the equipment makes contact with an object. 3

CH-200/CHU-200 Operation Manual 53


3 Operation

Do NOT lean on or apply strong force to the operation panel.


If a strong force is applied repeatedly, the sensor detecting the force on the
operation panel may fail, causing the equipment to move in an unexpected
manner. Please note that Shimadzu is not responsible for any damage caused
if strong force is recorded on the log.
Approximate number of repetitions for equipment lifetime

Force [Nm]

• When a force exceeding 40 Nm is applied, a warning tone sounds, “beep, beep,


beep, beep”.
• When a force exceeding 50 Nm is applied, power assist is temporarily
suspended until the unlock button is pressed again.

Do NOT hold the operation panel grip when moving the equipment if the
power is OFF, such as during a power failure.
If the sensor detecting force applied to the operation panel malfunctions, the
equipment may move in an unintended manner. Hold the lower part of the
column when moving the equipment.
"5.3 Operation During Power Outage"

54 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3
When using the power assist function, touch only the grip on the
operation panel. 3
Otherwise, the force applied to the operation panel may not be properly
detected and the equipment may move in an unintended manner.

3
Self-diagnosis is performed to check if the sensor detecting force applied to the
operation panel is malfunctioning during startup.Do NOT touch the equipment
when the operation touch panel displays "PA is in preparation. Wait without
operating.". If the equipment detects any possibility of being touched, the
message "Preparing. Press [Close] and wait." is displayed. Close the message 3
and wait for a while.

Note that the power assist function is not available on the rear mount type. 3

CH-200/CHU-200 Operation Manual 55


3 Operation

 Operation Panel

When the power assist function is used, the screen items and the rear operation panel
will be as shown below.

2 3

1 Assist level indicator


Shows the assist level currently selected. Pressing the indicator switches the assist
level.
Holding down the indicator switches the display to the assist level setting screen.

...

*• At the time of installation, the assist level can be set to be automatically


lowered when the collimator lamp is tuned on. If you require this function,
please contact your Shimadzu service representative.
• If the power assist is unavailable due to an error of the motor or during self-
diagnosis, the indicator will be shown as below.

2 Stop switch
Forcefully disables the power assist function.

3 Stop release switch


Recovers the equipment from the status brought with the stop switch.

4 Single-axis move button


Moves the equipment in the transversal direction at a certain speed, while the button is
held down.

56 CH-200/CHU-200 Operation Manual


3.2 Moving X-ray Tube Support

3
Note that the rear operation panel on the low-ceiling type does not have the
2 stop switch, 3 stop release switch, or 4 single-axis move button.
3
 Assist Level Settings

The acceleration (Acc), deceleration (Dec), and maximum speed (Speed) can be set
for each assist level.
3

3
If the maximum speed is exceeded, the assist is temporarily interrupted or
braking is applied, but this is not a malfunction.
Although continuously applying a strong force allows for movement beyond the
maximum speed, if this state continues for a certain time, an error will occur to
protect the motor. 3

6
1 Hold down (Assist level indicator).

3
Assist level indicator
3
The assist level setting screen is displayed.

2 Press (Assist level selection) and select the assist level to set.

Assist level selection

CH-200/CHU-200 Operation Manual 57


3 Operation

3 Select the cell to set, and change the value by pressing or .


The selected cell is displayed with a bold frame.The changed value is displayed in
red.

Longitudinal

Transversal

Vertical

Item Description
The higher the value, the easier it is to accelerate.
Acc
Set the value from -2 to 2. The factory default setting is 0.

The higher the value, the easier it is to decelerate.


Dec
Set the value from -2 to 2. The factory default setting is 0.

The maximum speed can be set in three steps.The factory default


settings are 50 (longitudinal and transversal) and 60 (vertical).
Speed 40: Approx. 40 cm/sec
50: Approx. 50 cm/sec
60: Approx. 60 cm/sec (vertical movement only)

4 Press to save the setting, or to cancel it.

The setting screen can be protected by a password, prompted to be provided


before displaying the screen.
If you require this function or forgot the password, please contact your
Shimadzu service representative.

58 CH-200/CHU-200 Operation Manual


3.3 Rotating X-ray Tube Support

3.3 Rotating X-ray Tube Support


3
3.3.1 Rotation Around Vertical Axis
3
Rotates the X-ray tube support around the vertical axis.

The lock is released only while the (V-axis rotation release) button is pressed.
Lifting your finger will lock the X-ray tube support.
3
1 While pressing the button on the operation panel, hold the grip
and rotate the X-ray tube support around the vertical axis. 3
The equipment will temporally stop at the 0, 90 and 180 positions. To move
the equipment again, lift your finger from the button once, press it again and
rotate the equipment. If the equipment is difficult to move while pressing the
button, press the button and quickly remove your finger. The lock will be 3
released for only 5 seconds.

6
If button is pushed once again while the lock is canceled for 5 seconds, it
is locked again.

3
Operation Panel Rear Operation Panel

Fig. 3.8

Fig. 3.9 Rotation around Vertical Axis

CH-200/CHU-200 Operation Manual 59


3 Operation

2 When reaching the intended position, lift your finger from the
button.
This will lock the X-ray tube support.

3.3.2 Rotation Around Horizontal Axis

Rotates the X-ray tube support around the horizontal axis.


The lock is released only while the (H-axis rotation release) button is pressed.
Lifting your finger will lock the equipment.
The equipment will stop at the 0 and 90 positions when rotating around the
horizontal axis.

For details on how to rotate around the horizontal axis, refer to "3.4.2 Rotating to
Preset Position Around Horizontal Axis".

1 While pressing the button on the operation panel, hold the grip
and rotate the X-ray tube support around the horizontal axis.
The rotation angle indicator will show the angle at which the equipment is
positioned.

In case that the X-ray tube support exercises is rotated before pushing
button of an operation panel, it may be in a lock state.
Be sure to push button on the operation panel before performing rotation
operation.

60 CH-200/CHU-200 Operation Manual


3.3 Rotating X-ray Tube Support

- 180 + 180 3
+ 120 - 120

3
+ 90 - 90 + 90 - 90

3
0 0
6

Upper left Upper right


(Standard)
3

Lower left Lower right


3
Fig. 3.10 Rotation around Horizontal Axis
The (+) or (-) direction of rotation angle varies in the installation position of X-ray tube
support as shown figures.

CH-200/CHU-200 Operation Manual 61


3 Operation

2 When reaching the intended position, lift your finger from the
button.
This will lock the X-ray tube support.

• The display of rotation angle tracks automatically the radiography


* technique change. This function is set at installation work.
Contact your Shimadzu service representative for setting.
+ 90

0 + 90 - 90

- 90 0

• The X-ray tube support combined with the X-ray radiography stand or
table has the function to which it does not indicated minus (negative)
angle sign on the touch panel. The function is set at installation work.
Contact your Shimadzu service representative for setting.

62 CH-200/CHU-200 Operation Manual


3.4 Detent Mode

3.4 Detent Mode


3
This function is to stop the X-ray tube support at the preset position.

The preset positions are one is set by installing in the longitudinal/transversal direction
and the other is 0 or 90 angle around the horizontal axis. 3
*• A function is available that always stops the X-ray tube support at preset
positions or always prevents the X-ray tube support from stopping at present
positions, regardless of whether the detent mode is ON or OFF. 3
This function can be configured to buttons on the operation panel at installation.
Contact your Shimadzu service representative to set up this function.
• A function is available for preventing the X-ray tube support from stopping at the
preset position when it is moving quickly, even if the detent mode is ON. This 3
function can be set at installation.
Contact your Shimadzu service representative to set up this function.
• A function is available for automatically turning OFF the detent mode when the
general radiography technique is selected. This function can be set at
installation.
3
Contact your Shimadzu service representative to set up this function.

6
3.4.1 Moving to Longitudinal/Transversal Detent Positions

Press the (Detent mode) button on operation panel or (Detent mode) key on 3
touch panel to operate.

* For the detent mode change, press the detent mode button on operation panel, so
the detent mode key on touch panel changes at a time. 3
Also, pressing the detent mode key on touch panel, the detent mode button on
operation panel changes.
If described as "Press the (detent mode) button on the operation panel" after
this, that is equally "press the (detent mode) key on the touch panel". 3

Detent mode key

Detent mode button

CH-200/CHU-200 Operation Manual 63


3 Operation

1 Press the button on the operation panel.


will appear, and the detent mode will be valid (ON).

Fig. 3.11 Detent Mode Button

2 While pressing the , or button on the operation panel, hold


the grip and move the X-ray tube support.
lights up when reaching the longitudinal detent position and the X-ray tube
stops.
lights up when reaching the transversal detent position and the X-ray tube
stops.

• Move the equipment slowly when in the vicinity of the preset position.
• To move the equipment again from the preset position, lift your finger from the
, or button once and press again to release the lock.
• By pressing the button again, the detent mode is turned OFF, then the
detent mode key on the touch panel turns OFF ( ).

64 CH-200/CHU-200 Operation Manual


3.4 Detent Mode

3.4.2 Rotating to Preset Position Around Horizontal Axis


3
The position is set at 0 or 90 angle around the horizontal axis.

1 Press the button on the operation panel.


3
will appear, and the detent mode will be valid (ON).

Fig. 3.12 Detent Mode Button 3


2 While pressing the button, hold the grip and move the X-ray tube
support. 6
The X-ray tube support will be locked at the 0 and 90 angles.

3
• To move the equipment again from the preset position, lift your finger from
the button once and press again to release the lock.

• By pressing the button again, the detent mode is turned OFF, then the 3
detent mode key on the touch panel turns OFF ( ).

CH-200/CHU-200 Operation Manual 65


3 Operation

3.5 Upward retraction function

Always be very careful when using the upward retraction function to


avoid contact with the patient or operator and to ensure that the patient
or operator does not become caught between the equipment and any
neighboring devices.
Otherwise, it may cause injury.

3.5.1 Retracting upward

This function retracts the unit upward.

* Servo tracking driver (option) is necessary to set the upward retraction function at
installation. If you require this function, please contact your Shimadzu service
representative.

If the upward retraction function is set, the (Detent mode) buttonon the operation
panel is replaced by the (Upward retraction) button.

1 Press and hold the (Upward retraction) button for more than 1 second.
The unit is retracted upward.

Upward retraction
button

* The retraction height can be set at installation.

If an emergency situation occurs with the movement, immediately stop


the equipment.
For details, refer to "5.2 Emergency Stop".

* Touch the touch panel or press a button on the operation panel to stop the
equipment movement.

66 CH-200/CHU-200 Operation Manual


3.6 Auto Collimation Function

3.6 Auto Collimation Function


3
By using this equipment in combination with the Shimadzu collimator, the irradiation field
can be adjusted automatically in accordance with the cassette size of the X-ray
radiography table or X-ray radiography stand, and the radiography distance. 3
This function is called the auto collimation function, and it improves positioning and
radiography efficiency.

As the irradiation field size achieved in adjustment through auto collimation you can 3
select the cassette size of the Bucky, but you can also select sizes smaller than the
cassette size.

For details on how to switch the irradiation field size, "3.6.2 Irradiation Field Size
Selection Function (Option)".
3
* Automatic adjustment of the irradiation field is possible according to the
radiography distance preset in the general radiography technique.
Contact your Shimadzu service representative to set up this function.
3

3.6.1 Status Indication for System


6

3
The system status indication on the operation panel's touch panel changes in tandem
with operation of the auto collimation function.

 PBL

This is displayed when all of the stipulated conditions (see the table below), such as the
alignment of the positions of the X-ray tube and the radiography device (Bucky), are
satisfied.

This condition is called the PBL (Positive Beam Limitation) status.

CH-200/CHU-200 Operation Manual 67


3 Operation

Unit to be combined Requirements


Is X-ray tube faced toward tabletop?

Is horizontal SID in range of 90 - 150 cm?

Does the center of the X-ray tube coincide with the center of
X-ray radiography table the detection surface in the Bucky?
(Horizontal front) Is cassette equipped? (Correct size?)

Is the exposed cassette of Bucky used?


(Prohibit double exposure)*

Is the grid exchanging? (No grid or the grid is inserted.)

Is X-ray tube faced toward stand?

Is vertical SID in range of 90 - 205 cm?

Does the center of the X-ray tube coincide with the center of
the detection surface in the Bucky (longitudinal, transversal,
X-ray radiography stand or vertical)?
(Vertical front)
Is cassette equipped? (Correct size?)

Is the exposed cassette of Bucky used?


(Prohibit double exposure)*

Is the grid exchanging? (No grid or the grid is inserted.)

*: When a film cassette including CR is normally used

On changing the radiography distance (SID) within the stipulated radiography distance
range when the conditions stated above are satisfied, the collimator automatically
adjusts the area of the irradiation field to match the cassette size.

Further, by operating the collimator while the cassette size is the maximum, the
irradiation field can be narrowed down to sizes smaller than that cassette size (in this
case too, [PBL] is displayed). Once the irradiation field has been narrowed down, if the
cassette size is then changed, for example by changing the cassette, the irradiation
field automatically adjusts to the newly selected cassette size.

* If combined with X-ray radiography table or X-ray radiography stand, which


cannot recognize cassette size, the [PBL] is not displayed.

 Automatic

This is displayed when the position of the radiography equipment (Bucky) is not aligned
with the X-ray axis. The collimator automatically adjusts the irradiation field to the size
of the X-ray detector in use.

When [Automatic]* is displayed, the operator must take responsibility for


adjusting the irradiation field before starting radiography.
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

68 CH-200/CHU-200 Operation Manual


3.6 Auto Collimation Function

 Manual

This is displayed when general radiography technique is selected using the 3


radiography technique keys.The irradiation field can be adjusted manually without
pressing the (manual operation) button at the collimator.

3
When [Manual] is displayed, the operator must take responsibility for
adjusting the irradiation field before starting radiography.
3
 PBL Override

This is displayed when the (manual operation) is pressed at the collimator and the
LED lights. It is to be used if the system has a failure.
3

When [PBL Override] is displayed, the operator must take responsibility


3
for adjusting the irradiation field before starting radiography.

 Obliq 6
This is displayed when the X-ray tube support is tilted to track with the Bucky and, at
the same time, the irradiation field is coincided with the position of the Bucky, the
Bucky. 3
 Tomo

This is displayed when an initial position setup for a tomography and loading of a 3
cassette are completed. The tomography is started by push button operation of X-ray
high-voltage generator.

 Not Ready 3

• When the radiography system status condition displays "Not Ready", X-ray
exposure is not possible.

• When combined with a collimator other than the Shimadzu autocollimator,


no indication other than "Not Ready" is displayed.

This is displayed under the following conditions:

• When the wrong X-ray tube has been selected


• When the initial setting of the position for tomography has not been completed
• When the cassette has not been loaded
• When the cassette after performed Bucky radiography is intended to use1
• When the grid is exchanging

CH-200/CHU-200 Operation Manual 69


3 Operation

• When the Bucky for the selected technique and the X-ray tube are not in direct
opposition (60 degrees) 2
1: When a film cassette including CR is normally used
2: Radiography may be possible depending on the installation settings.

3.6.2 Irradiation Field Size Selection Function (Option)

The irradiation field size (including portrait and landscape orientations) can be switched
to a size equal to or smaller than the cassette size of the X-ray radiography stand or X-
ray radiography table.

*• The irradiation field size selection function is available as an option. If you


require this function, please contact your Shimadzu service representative.

1 Press the irradiation field size display area on the touch panel.

The irradiation field size selection window will open.

2 Select the irradiation field size.

Select reference position.


Select exposure field size.
Select the direction of
field (vertical/horizontal)

Irradiation field sizes (including portrait and landscape orientations) up to the size
of the cassette inserted in the Bucky of the X-ray radiography stand or X-ray
radiography table can be selected. It is also possible to switch the size type
(inches/cm).

70 CH-200/CHU-200 Operation Manual


3.6 Auto Collimation Function

Only in cases where cassettes with a dimension of 17 inches (43 cm) in the
patient's height direction are used in an X-ray radiography stand, you can select a
top, center, or bottom reference position for the irradiation field. With X-ray 3
radiography tables only the center reference is available and it cannot be
changed.

3 Press
3

3
This the irradiation field size selection window and confirms the irradiation

3
field size.

3.6.3 Tracking with Radiography Conditions


3
When the tracking with radiography conditions function is used, the irradiation field size
can be changed in accordance with the radiography program.

6
* The tracking with radiography conditions function can be used when the
irradiation field size selection function is enabled. If you require this function,
please contact your Shimadzu service representative. 3

3
If the irradiation field size tracking with the radiography conditions is bigger
than the size of the cassette loaded in the X-ray radiography table or X-ray
radiography stand, the indication on the irradiation field size display button 3
becomes "---". When "---" is indicated, radiography is not possible. Either
change the cassette size of the X-ray radiography table or X-ray radiography
stand (exchange the cassette for another), or press the irradiation field size
display button to change the irradiation field size.

CH-200/CHU-200 Operation Manual 71


3 Operation

 In CH-200 Tracking APR Mode

The irradiation field size and the reference point track the APR.

Registering the irradiation field size information in a radiography


program
Make a setting for each radiography step on the radiography program editing screen of
the X-ray high-voltage generator.

For details on radiography programs, refer to the operation manual for X-ray
high-voltage generators for diagnosis.

1 Press the radiography step name frame on the radiography program


editing screen.

2 Input the irradiation field size information at the beginning of the


radiography step comment field.

Input the irradiation field size information in the format shown below.

Format Item Details


H Size in the horizontal direction (Numeric characters)

V Size in the horizontal direction (Numeric characters)

Reference position
• T : Top reference
• C : Center reference
• B : Bottom reference
HxV
Only in cases where cassettes with a dimension of 17 inches (43
 cm) in the patient's height direction are used in an X-ray
radiography stand, you can select a top, center, or bottom
reference position for the irradiation field.
If the reference position is omitted, the reference position of the
Bucky section of the X-ray radiography stand will apply.
With X-ray radiography tables only the center reference is
available and it cannot be changed.

72 CH-200/CHU-200 Operation Manual


3.6 Auto Collimation Function

Examples
• "14x17"
The reference position is the reference position of the Bucky section of the X-
3
ray radiography stand and the size is 14 inches in the horizontal direction by 17
inches in the vertical direction.

• "14x14T" 3
The reference position is the top reference and the size is 14 inches in the
horizontal direction by 14 inches in the vertical direction.

• "10x12B" 3
The reference position is the bottom reference and the size is 10 inches in the
horizontal direction by 12 inches in the vertical direction.

Note that it is also possible to combine this format with other comments. 3
(Example: 14x17 SID120cm)

If no irradiation field size information has been input in the radiography step
comment field, or if the corresponding irradiation field size has not been input 3
correctly, the size of the cassette loaded in the Bucky of the X-ray radiography
table or X-ray radiography stand is set. The available irradiation field sizes are
shown in the table below (the vertical and horizontal values can be exchanged).
6
Size Type Irradiation Field Size
inch FULL,17x14,16x12,17x7,14x14,14x11,12x10,10x8,9.5x9.5

cm FULL,43x35,40x30,43x18,35x35,35x30,30x24,24x18,24x24 3
This is the maximum size of the detector of the X-ray radiography table or X-ray radiography stand,

3
such as 17 x 17 inches or 43 x 43 cm.

Calling the Irradiation Field Size from a Radiography Program

1 Select a radiography step for which irradiation field size information


has been entered in the comment field. 3

The auto collimation function operates in accordance with the irradiation field size
indicated in the comment for the radiography step.

You can select the irradiation field size by pressing the irradiation field size
display.

CH-200/CHU-200 Operation Manual 73


3 Operation

3.7 Lighting Collimator Lamp

3.7.1 Collimator Lamp Lights Up

Pressing the (collimator lamp) button on the operation panel or rear operation
panel will light up the collimator lamp for roughly 30 seconds.

The collimator lamp turns off after pressing the collimator lamp button while it is ON.
Another lamp button is located on the collimator. (Fig. 3.13)

Operation Panel Rear Operation Panel

Fig. 3.13 Collimator Lamp Button


The collimator lamp automatically lights up when the operating the equipment as
follows:

• Pressing the brake release button of Shimadzu X-ray radiography stand.


• Releasing the fixing selection switch for tabletop locking of Shimadzu X-ray
radiography table combined with the equipment.
• Tracking with the X-ray radiography table or the X-ray radiography stand.
(Valid for the combination with the X-ray radiography table or the X-ray radiography
stand of other manufacturer.)
• Operating the equipment by auto positioning function and moving to the registered
position.
(Valid for the combination with the X-ray radiography table or the X-ray radiography
stand of other manufacturer.)

• The installation work makes the collimator lamp preventing the


* automatic lighting.
If you require this function, contact your Shimadzu service
representative.

• When the equipment is combined with a collimator other than the


Shimadzu autocollimator, the auto lighting function is available as an
option.
If you require this function, please contact your Shimadzu service
representative.

74 CH-200/CHU-200 Operation Manual


3.7 Lighting Collimator Lamp

3.7.2 Adjusting Size of Irradiation Field [Reference]


3

For details on operating the collimator, refer to the collimator's operation


3
manual attached separately.

Adjust the irradiation field size by turning the right and left dials on the collimator. The
collimator is equipped with an aperture gauge that shows the irradiation field size
3
according to the radiography distance (SID).

3
Collimator Lamp Button Dial
3
Fig. 3.14 Adjusting the Size of the irradiation field

CH-200/CHU-200 Operation Manual 75


3 Operation

This page is intentionally left blank.

76 CH-200/CHU-200 Operation Manual


Radiography
This chapter describes how to perform radiography
by the X-ray tube support combined with an optional
component such as X-ray radiography table and
X-ray radiography stand.
4 Radiography

4.1 General Radiography


1 Confirm that the general radiography technique is selected.

• When the equipment is in CH-200 non-tracking mode, be sure to select the


general radiography technique at the X-ray tube support too by following
the procedure below.

(1)Press the technique key on


the touch panel.
The pop up window to select
the radiography technique
will appear.

(2)Select the general


radiography technique key.

• When the general radiography technique is selected, the irradiation field


can be adjusted without closing the collimator leaves in the X-ray tube
support.

2 Adjust the region of interest of patient.

3 Face the X-ray tube of the X-ray tube support toward the patient's
region of interest.

4 Move the X-ray tube support and check the radiography distance
using a tape measure provided at the bottom face of the collimator.

Tape measure

78 CH-200/CHU-200 Operation Manual


4.1 General Radiography

5 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp. 4

The operator must have responsibility in carrying out irradiation field


4
adjustment and performing X-ray radiography.
If [Manual] is displayed on a touch panel, irradiation field adjustment and X-ray
radiation will become possible.
4
6 Operate the X-ray high-voltage generator and perform radiography.
4

CH-200/CHU-200 Operation Manual 79


4 Radiography

4.2 Vertical General Radiography


1 Confirm that the general radiography technique is selected.

• When the equipment is in CH-200 non-tracking mode, be sure to select the


general radiography technique at the X-ray tube support too by following
the procedure below.

(1)Press the technique key on


the touch panel.
The pop up window to select
the radiography technique
will appear.

(2)Select the general


radiography technique key.

• When the general radiography technique is selected, the irradiation field


can be adjusted without closing the collimator leaves in the X-ray tube
support.

2 Face the X-ray tube support


toward the X-ray radiography
stand.

The indicator [Manual] is displayed on


the touch panel.
The SSD indicates the distance between
the cassette of X-ray radiography stand
and the focus of X-ray tube when the X-
ray tube is faced to the X-ray radiography
stand.

80 CH-200/CHU-200 Operation Manual


4.2 Vertical General Radiography

3 Move the X-ray tube support.


Adjust the radiography distance. 4
4 Fix the cassette holder (Option) to
the Bucky. Cassette 4
Holder

4
5 Attach the cassette to the cassette
4
holder.
Cassette

Cassette
Holder
4
6 Position standing patient at X-ray
radiography stand. 6
7 Move the X-ray tube support vertically.
Adjust the X-ray irradiation position.
4
8 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.
4

The operator must have responsibility in carrying out irradiation field


adjustment and performing X-ray radiography.
4
If [Manual] is displayed on a touch panel, irradiation field adjustment and X-ray
radiation will become possible.

9 Operate the X-ray high-voltage generator and perform radiography.

CH-200/CHU-200 Operation Manual 81


4 Radiography

4.3 Horizontal General Radiography


1 Confirm that the general radiography technique is selected.

• When the equipment is in CH-200 non-tracking mode, be sure to select the


general radiography technique at the X-ray tube support too by following
the procedure below.

(1)Press the technique key on the


touch panel.
The pop up window to select the
radiography technique will
appear.

(2)Select the general radiography


technique key.

• When the general radiography technique is selected, the irradiation field


can be adjusted without closing the collimator leaves in the X-ray tube
support.

2 Face the X-ray tube support


toward the X-ray radiography
table.

The indicator [Manual] is displayed on


the touch panel.
The STD indicates the distance between
the tabletop of X-ray radiography table
and the focus of X-ray tube when the X-
ray tube is faced to the X-ray radiography
table.

3 Move the X-ray tube support.


Adjust the radiography distance.

82 CH-200/CHU-200 Operation Manual


4.3 Horizontal General Radiography

4 Take the patient on the X-ray


4
radiography table and place the
cassette between the patient
region and tabletop.
4
5 Move the X-ray tube support
Cassette

vertically.
Adjust the X-ray irradiation position.
4
6 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.
4

The operator must have responsibility in carrying out irradiation field 4


adjustment and performing X-ray radiography.
If [Manual] is displayed on a touch panel, irradiation field adjustment and X-ray
radiation will become possible.
6
7 Operate the X-ray high-voltage generator and perform radiography.

CH-200/CHU-200 Operation Manual 83


4 Radiography

4.4 Vertical Bucky Radiography


1 Confirm that the vertical Bucky radiography technique is selected.

When the equipment is in CH-200 non-tracking mode, be sure to select the


vertical bucky radiography technique at the X-ray tube support too by following
the procedure below.
(1)Press the technique key on
the touch panel.
The pop up window to select
the radiography technique
will appear.

(2)Select the vertical bucky


radiograhy technique key.

2 Press the (detent mode)


button on the operation panel of
the X-ray tube support to turn
the detent mode ON.

3 Face the X-ray tube support


toward the X-ray radiography
stand.

84 CH-200/CHU-200 Operation Manual


4.4 Vertical Bucky Radiography

4 Move the X-ray tube support.


Adjust the radiography distance. X-ray Radiography 4
Also align the X-ray tube support with the Stand
center of the X-ray radiography stand's
horizontal direction.
4
X-ray Tube

5 Install the cassette into the Bucky. 4


Insert the grid or remove it, if required. Cassette Tray

For inserting the cassette, refer to the


operation manual of X-ray radiography 4
stand.

Cassette
4

6
When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired. 4

In case of that the cassette and the grid are not inserted correctly, the [Not
Ready] indicates on the touch panel of X-ray tube support. Be sure to lock the 4
cassette and the grid correctly.

6 Position the standing patient.

7 Move the X-ray tube support vertically.


Adjust the X-ray irradiation position.

8 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.

CH-200/CHU-200 Operation Manual 85


4 Radiography

If the vertical tracking function (option) is provided, [PBL] is displayed on the


touch panel when the focal height of the X-ray tube is aligned with the
radiography position.

Even when rotating the X-ray tube around vertical axis for each 90, the
[PBL] will be indicated on the touch panel and you can perform
radiography. Verify that the X-ray tube is facing to the proper direction,
and perform radiography.

X-ray Tube

[Automatic] *is displayed on the touch panel if the vertical tracking


function (option) is not installed or the focal height of the X-ray tube is
not aligned with the radiography position.
When [Automatic] is displayed, the operator must have responsibility in
carrying out irradiation field adjustment and performing X-ray
radiography.
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

9 Operate the X-ray high-voltage generator and perform radiography.

86 CH-200/CHU-200 Operation Manual


4.4 Vertical Bucky Radiography

4.4.1 Rotating X-ray Tube Unit


4

4
When rotating the X-ray tube unit 180 degrees around vertical axis and

4
facing to the X-ray radiography stand to take radiography, select the
general radiography technique for radiography.

Usual Position
After rotating the X-ray tube unit 180
degrees around vertical axis, face it to
the X-ray radiography stand.
4

CH-200/CHU-200 Operation Manual 87


4 Radiography

4.4.2 Dual X-ray Radiography Stand System

When the X-ray radiography stand is being used in addition to the automatic X-ray
radiography system which consists of X-ray tube support, X-ray radiography table and
X-ray radiography stand, only manual X-ray radiography is available for the 2nd set of
X-ray radiography stands.
(Auto collimation function and vertical tracking function not available)

X-ray Radiography Stand

X-ray Tube Support

2nd X-ray Radiography Stand


(Changer, Manual Stand,
Lieder Radiography Stand)

X-ray Radiography
Table

Even when it has by carrying out the automatic Bucky stand and automatic
changer unit as the 2nd set of equipment, only manual radiography
corresponds.

88 CH-200/CHU-200 Operation Manual


4.4 Vertical Bucky Radiography

When there is 2nd set of X-ray radiography stand, operate it as follows.

1 Select the vertical Bucky radiography technique as 2nd set of X-ray


4
radiography stand with touch panel of the X-ray tube support.

2 Face the X-ray tube of the X-ray tube support toward the patient's 4
region of interest.
Check the radiography distance by using the collimator's tape measure.

3 Load a cassette in the second stand and decide the patient's region
4
of interest.

4 Move the X-ray tube support vertically. 4


Adjust the X-ray irradiation position.

5 Adjust the irradiation field with the collimator while pressing the
4
collimator lamp button to light the collimator lamp.

6 Set the radiography conditions and perform radiography.


6

CH-200/CHU-200 Operation Manual 89


4 Radiography

4.5 Horizontal Bucky Radiography


1 Confirm that the horizontal Bucky radiography technique is selected.

When the equipment is in CH-200 non-tracking mode, be sure to select the


horizontal bucky radiography technique at the X-ray tube support too by
following the procedure below.
(1)Press the technique key on the
touch panel.
The pop up window to select the
radiography technique will
appear.

(2)Select the horizontal bucky


rediography technique key.

2 Press the (detent mode)


button on the operation panel of
the X-ray tube support to turn
the detent mode ON.

3 Face the X-ray tube support


toward the X-ray radiography table.

4 Move the X-ray tube support.


Align the X-ray tube support with the
center of the X-ray radiography table in
the longitudinal and transversal
directions.

90 CH-200/CHU-200 Operation Manual


4.5 Horizontal Bucky Radiography

5 Confirm that the Bucky of the X-


4
ray radiography table and the X-
ray tube are directly opposite each
other.
Check this as follows, and if they are not 4
in direct opposition, operate the X-ray Bucky

tube support to align them.

6 Press the collimator lamp


4
button. Beam

The collimator's beam for center


alignment will be emitted.
Move backward. 4
7 Check that the beam for center
alignment is aligned with the 4
center marking of the Bucky's
cassette tray handle.

6
Notch
If the beam is blocked by the X-ray
radiography table's tabletop, allow it to
reach the handle by one of the
following methods. Center Marking
4
• Shift the X-ray radiography table's
tabletop backward. Draw the cassette

4
• Draw out the Bucky's cassette tray. tray.

If the beam for center alignment and


center marking on the Bucky's
cassette tray handle are not aligned,
move the Bucky or the X-ray tube
support.
4
For details on how to operate the Bucky, refer to the operation manual for
the X-ray radiography table.

CH-200/CHU-200 Operation Manual 91


4 Radiography

8 Load a cassette in the X-ray


radiography table's Bucky.
Insert the grid or remove it, if required.

Cassette

For details on how to insert cassettes, refer to the operation manual for the
X-ray radiography table.

When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired.

In case of that the cassette and the grid are not inserted correctly, [Not Ready]
on the touch panel of X-ray tube support is displayed. Be sure to lock the
cassette and the grid correctly.

9 Decide the patient's horizontal position.


10Move the X-ray tube support.
Adjust the radiography distance.

11Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.

When the X-ray tube unit is directly opposing the Bucky, [PBL] will be displayed
on touch panel, and X-ray radiography is possible.

92 CH-200/CHU-200 Operation Manual


4.5 Horizontal Bucky Radiography

4
Even when rotating the X-ray tube around vertical axis for each 90, the
[PBL] will be indicated on the touch panel and you can perform
radiography. Verify that the X-ray tube is facing to the proper direction, 4
and perform radiography.

4
When the X-ray tube unit and Bucky are not in direct opposition,
[Automatic]* is displayed on the touch panel.
When [Automatic] is displayed, the operator must have responsibility in
4
carrying out irradiation field adjustment and performing X-ray
radiography.
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting. 4

12Operate the X-ray high-voltage generator and perform radiography. 6

CH-200/CHU-200 Operation Manual 93


4 Radiography

4.5.1 When the Direction of Transversal Center at X-ray Radiography


Table and the Center of X-ray Radiography Stand Have Shifted

If the direction of transversal center at X-ray radiography table and the


center of X-ray photography stand have shifted, the operator should
perform the correct positioning and do X-ray radiography.

In arrangement of the X-ray radiography table and X-ray radiography stand as shown
in a figure, the system is not discriminating the direction center position of transversal
(A) of X-ray radiography table, and the center position (B) of X-ray radiography stand.
Therefore, X-ray radiography is possible for the X-ray tube unit to the X-ray
radiography stand also in the position of (A). Moreover, perform X-ray radiography after
setting an X-ray tube unit by the correct position in performing X-ray radiography, since
X-ray radiography can do the X-ray tube unit to the X-ray radiography table also in the
position of (B).

X-ray Radiography Table Longitudinal Rail

X-ray Tube Unit Facing to (A)


X-ray Radiography Stand

Transversal Rail

(B)

X-ray Tube Unit facing to


X-ray Radiography Table
X-ray Radiography Stand

94 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

4.6 Radiography Using Tracking Device


4
(Option)
The radiography using tracking device (option) is the vertical tracking radiography or the
horizontal tracking radiography.
4

4
When an emergency has occurred due to failure in switches, etc.,
perform an emergency stop.
For details on emergency stop, refer to "5.2 Emergency Stop". 4

 Vertical Tracking Radiography 4


The vertical tracking radiography is only possible when the vertical Bucky radiography

6
technique is selected.

• In the vertical Bucky radiography technique, the X-ray tube's radiography height
tracks in accordance with the height of the cassette loaded in the X-ray radiography
stand's Bucky.
For details on radiography using the vertical Bucky radiography technique, refer
to "4.4 Vertical Bucky Radiography".
4
To operate the tracking device, use the following two procedures:

• Operating the X-ray radiography stand without tilting.


For details on radiography not using tilt function, refer to "4.6.1 Vertical
4
Tracking Radiography (Not using tilt function)".

• Operating the X-ray radiography stand by tilting it.


For details on radiography using tilt function, refer to "4.6.2 Vertical Tracking 4
Radiography (Using tilt function)".

 Horizontal Tracking Radiography

For horizontal tracking radiography, tracking is possible when the general radiography
technique or horizontal Bucky technique is selected.

• When the general radiography technique is selected, tracking is performed so as to


maintain a constant distance between the X-ray tube and the tabletop of the X-ray
radiography table.
• When the horizontal Bucky technique is selected, tracking is performed so as to
maintain a constant distance between the X-ray tube and the cassette loaded in the
X-ray radiography table's Bucky.

CH-200/CHU-200 Operation Manual 95


4 Radiography

4.6.1 Vertical Tracking Radiography (Not using tilt function)

Use the X-ray radiography stand without tilting the Bucky.

1 Operate the X-ray tube support.


For details on the operation of the X-ray tube support, refer to "4.4 Vertical
Bucky Radiography".

2 Load a cassette into the X-ray radiography stand’s Bucky.


Insert the grid or remove it, if required.

When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired.

In case of that the cassette and the grid are not inserted correctly, [Not Ready]
on the touch panel of X-ray tube support is displayed. Be sure to lock the
cassette and the grid correctly.

3 Decide the patient's standing position.

4 Press the (tracking) switch on the X-ray radiography stand or


the (tracking start) button on the operation panel of the X-ray
tube support, and move the Bucky of the X-ray radiography device
vertically.

Tracking start switch

Tracking switch

96 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

* If the tracking switch kit is incorporated, the (tracking) switch is on the switch
4
box.

Switch box

4
Tracking switch

4
This starts tracking operation based on the operating method set at installation.
During tracking operation, (the vertical motion set position indicator) flashes. 4
When the X-ray tube support is aligned with the stipulated position, (the

4
vertical motion set position indicator) becomes continuously lit.

Ensure that there is no patient at the position where the X-ray tube
4
support is to be lowered before starting tracking operations.
Contact with the X-ray tube support could result in injuries.

CH-200/CHU-200 Operation Manual 97


4 Radiography

To prevent damage to other equipment, confirm that nothing is blocking


the X-ray tube support when moving it.

Verify that the X-ray tube is facing toward the proper direction, and
perform radiography.

X-ray Tube

Even when rotating the X-ray tube around vertical axis and stopping at each
90, the [PBL] will be indicated on the touch panel and you can perform
radiography.

98 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

4
• If there is a possibility that the X-ray tube support contacts the X-ray
radiography stand, it once stops and continuous beep sounds (Approx. 2
seconds).
4
If the tracking switch is pressed again, discontinuous beep sounds and the
X-ray tube support descends at a low speed. The X-ray tube support
continues to descend until the tracking operation is cancelled by pressing
(releasing when the hold function is turned on) the tracking switch. So,
4
position the X-ray tube unit in contact with X-ray radiography stand with
care.
The tabletop of the X-ray radiography table cannot move up even if it is a
Shimadzu X-ray radiography table. 4
• It is not a malfunction that the X-ray tube support once goes beyond too far
and returns to the predetermined position for alignment.
4

6
Press a button on the operation panel of the X-ray tube support during tracking
to stop movement.
4

CH-200/CHU-200 Operation Manual 99


4 Radiography

When the rotation angle is shifted over 4 degrees from axis, or the height of X-ray
tube unit is shifted the center of Bucky, the [Automatic] is displayed.

When not aligned the axis of X-ray tube unit with the center of Bucky,
[Automatic]* is displayed on touch panel. When [Automatic] is displayed,
the operator must have responsibility in carrying out irradiation field
adjustment and performing X-ray radiography.
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

5 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.

6 Operate the X-ray high-voltage generator and perform radiography.

100 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

 About Vertical Tracking

There are three types of tracking operation: full auto, semi auto, and hold. 4

At installation you select one of the functions, full auto, semi auto, or hold. 4
For details on this setting please contact your Shimadzu service
representative.

Full Auto
4
Press the (tracking) switch on the X-ray radiography stand or the (tracking
start) button on the operation panel of the X-ray tube support, or move the X-ray
radiography stand vertically. 4
The tracking switch will flash, and the X-ray tube support will track by moving vertically.

• Even if you don't press the tracking start button, if you move the X-ray
* radiography stand vertically the X-ray tube support will track it by moving 4
vertically.

• There is no need to keep the tracking start button (tracking switch) held
down during tracking.
6
• Pressing the tracking start button (tracking switch) while tracking
operation is in progress stops the tracking operation, but it can be
restarted by pressing the tracking start button (tracking switch) again, or
by moving the X-ray radiography stand's Bucky vertically. 4
Semi Auto
Press the X-ray radiography stand's (tracking) switch or the
button on the operation panel of the X-ray tube support.
(tracking start)
4
The tracking switch will flash, and the X-ray tube support will track by moving vertically.

*
• If you move the X-ray radiography stand vertically after tracking
operation has started the X-ray tube support moves vertically to track
4
this motion.

• There is no need to keep the tracking start button (tracking switch) held
down during tracking.

• Pressing the tracking start button (tracking switch) while tracking


operation is in progress stops the tracking operation, but it can be
restarted by pressing the tracking start button (tracking switch) again.

CH-200/CHU-200 Operation Manual 101


4 Radiography

Hold
Hold down the X-ray radiography stand's (tracking) switch or the (tracking
start) button on the operation panel of the X-ray tube support.

The tracking switch will flash, and the X-ray tube support will track by moving vertically.

* Releasing the tracking start button (tracking switch) while tracking operation is in
progress stops the tracking operation.

 Necessary Conditions for Vertical Tracking

The X-ray radiography stand and the X-ray tube support can work together under the
following condition:

Condition Description

Radiography
(vertical bucky radiography)
Technique

At the center of the X-ray radiography stand and at an exposure


distance of 50 cm or more

SID: 50 cm or more
X-ray Radiography X-ray Tube Support
Stand

Center of X-ray Radiography Stand

Position of X-ray
Tube Support Horizontal toward X-ray radiography stand
Incorrect

Otherwise tracking function is not available.


Horizontal Direction (Tracking function is available if it is set at
the installation.)

102 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

 Oblique Projection Radiography

After tracking the X-ray tube support and X-ray radiography stand, operate the 4
following procedures for oblique projection radiography.

1 Adjusting the height of X-ray tube unit, tilt the equipment with
4
rotating a little for upper or lower direction.

6
The X-ray tube support can be set to move vertically even if it is tilted.
The angles to which the X-ray tube support can be tilted include 20, 40and 4
60.
When the X-ray tube support is set to move vertically while tilted, and also
under that setting, the exposure filed is coincided with the position of the
Bucky, the [Obliq] is displayed on the upper right of the touch panel.
4

4
In vertical oblique projection, Bucky tracking operates with an error of 3 cm
(Radiography distance = 100 cm).
In case the distance is over 100 cm, the error increases by 3 cm or more.

CH-200/CHU-200 Operation Manual 103


4 Radiography

4.6.2 Vertical Tracking Radiography (Using tilt function)

The tracking radiography can be performed while the Bucky of the X-ray radiography
stand is tilted.
The operation procedures other than the tilting operation are same as those operation
not in use tilting.

For details on operation not using tilting, refer to "4.6.1 Vertical Tracking
Radiography (Not using tilt function)".

1 Tilt the Bucky forward or back.

According to the predetermined angle of Bucky, the status indication on touch


panel turns from to or etc.

Except the determined angle (-20, 15, 30, 45, 60, 75, 90), is
displayed and also, the ( ) button is displayed for information.

When the Bucky is tilted, the operator must have responsibility in


carrying out exposure field adjustment and performing X-ray
radiography.

104 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

4
When the LED lights by pressing the (manual operation) button at the
collimator, the present display such as [TILT15] disappears and [PBL Override]
appears.
4

2 Aim the X-ray tube support at Bucky of X-ray radiography stand, and
4
adjust the SID.

6
The SID indication on touch panel changes and the exposure field tracks that
change.
4

The Source-to-Image Distance (SID) with the Bucky tilted shows the distance
4
to the point of intersection of the extended line of X-ray detector surface and
the X-ray radiation line.

Bucky

Bucky
X-ray detector surface

X-ray detector surface

CH-200/CHU-200 Operation Manual 105


4 Radiography

Even where the Bucky is tilted at 90 and is turned upward, select the vertical
radiography technique.

106 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

4
With a tilt type X-ray radiography stand, vertical tracking is available even with
the Bucky inclined by the setting at installation. The angle range of the X-ray
tube for this vertical tracking is 20, 40 and 60 against the Bucky.
Also, Vertical tracking operation is not available on the following states: 4
• Rotating the X-ray tube in a direction that is around horizontal axis, it does
not face to the X-ray radiography stand.

• The Bucky is not fixed ([TILT] is displayed on touch panel). 4


Incorrect

With the tilting angle of Bucky 90 (with [TILT90] displayed in the right upper 6
part of touch panel), X-ray tube moves relatively so that SID becomes the
predetermined value.
The SID can be set between 90 cm and 150 cm at installation (the same
distance at the horizontal tracking radiography). 4

4
When operating vertical tracking with tilting Bucky and X-ray detector surface
is at right angle to X-ray tube, the allowance will be 2 % of radiography
distance as shown below.

CH-200/CHU-200 Operation Manual 107


4 Radiography

4.6.3 Horizontal Tracking Device

The explanation given here assumes that the horizontal Bucky technique has been
selected.

For details on radiography using the horizontal Bucky technique, refer to


"4.5 Horizontal Bucky Radiography".

For details on radiography using the general radiography technique, refer to


"4.3 Horizontal General Radiography".

1 Operate the X-ray tube support.


For details on the operation of the X-ray tube support, refer to "4.5
Horizontal Bucky Radiography".

2 Check whether the X-ray radiography table's Bucky and the X-ray
tube focus are directly opposing each other.

3 Load a cassette into the Bucky.


Insert the grid or remove it, if required.
When inserting the grid, be sure to confirm the grid type or the status of insertion.

When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired.

In case of that the cassette and the grid are not inserted correctly, [Not Ready]
on the touch panel of X-ray tube support is displayed. Be sure to lock the
cassette and the grid correctly.

4 Decide the patient's horizontal position.

108 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

5 Adjust the tabletop to the appropriate height by depressing the ( )


or ( ) footswitch. 4

If the radiography distance (SID) of X-ray tube unit turns to preset distance 4
(1m) during the tabletop rises, then it will track maintaining the preset distance.
At the time of tabletop descent, the X-ray tube unit descends without
descending.
The radiography distance can be preset within the range between 90 cm and
150 cm at the time of installation.
4

4
6 Hold down the X-ray radiography table's (tracking start) switch
or the (tracking start) button on the operation panel of the X-ray
tube support. 4

Tracking start button


6
Tracking start switch
4

4
* If the tracking switch kit is incorporated, the (tracking) switch is on the switch
box.

Switch box
4

Tracking switch

Motion continues until the X-ray tube support is at the X-ray radiography table's
radiography height (radiography distance). Keep holding down the X-ray radiography
table's (tracking start) switch or (tracking start) button on the CH-200/CHU-
200's operation panel until motion stops and a "pip pip pip" sound is heard.

CH-200/CHU-200 Operation Manual 109


4 Radiography

The radiography distance can be set in 5 cm increments within the range


90 cm to 150 cm at installation.

During tracking operation the vertical motion set position indicator ( ) on the X-ray
tube support's touch panel flashes. When the X-ray tube support is aligned with the
stipulated position, the vertical motion set position indicator ( ) becomes
continuously lit.

Press a button on the operation panel of the X-ray tube support during tracking
to stop movement.

7 Operate the X-ray tube support.


The X-ray irradiation position is automatically adjusted.

8 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.

9 Operate the X-ray high-voltage generator and perform radiography.

110 CH-200/CHU-200 Operation Manual


4.6 Radiography Using Tracking Device (Option)

4.6.4 Horizontal Tracking Distance Selection Function


4
With the horizontal tracking function of the X-ray tube support, the radiography
distance of the X-ray tube unit with respect to a horizontal table can be changed in
accordance with the radiography program of the Shimadzu X-ray high-voltage
generator (touch panel type). 4
When the radiography technique in use is the general radiography technique, the
tracking distance between the X-ray tube unit and the tabletop surface is changed, and
when it is the horizontal Bucky tracking technique the tracking distance between the X-
ray tube unit and the film in the Bucky is changed. 4
 In CH-200 Tracking APR Mode

Registering a Radiography Distance in a Radiography Program


4
Make a setting for each radiography step on the X-ray high-voltage generator's
radiography program editing screen.

For details on radiography programs, refer to the operation manual for the X-ray 4
high-voltage generator.

* This function can be set at installation. If you require this function, please contact
your Shimadzu service representative. 6
1 Press the radiography step
name frame on the radiography 4
program editing screen.

2 Enter the radiography distance


Radiography step name frame
4
information in the radiography
step comment field.

Input the radiography distance


information in the following format.

Format Item Details


D Upper case
D***cm *** Radiography distance
cm Lower case

CH-200/CHU-200 Operation Manual 111


4 Radiography

• Example: D120 cm
The X-ray tube unit tracks at the position where the radiography distance is
120 cm.
• Combination with other comments is also possible.
(Examples: "SID120 cm", "STD120 cm")
• If the SID indication on the touch panel is set to display in inches, input "in"
instead of "cm".
• The radiography distance can be set within the range 90 cm to 170 cm (36
inches to 68 inches).
If no radiography distance information has been input in the radiography step
comment field, or it has not been input correctly, the radiography distance set at
installation is set.

Depending on the ceiling height or the height of the radiography table the X-ray
tube support may reach its upper limit in the vertical direction, meaning that
motion to the position at the input radiography distance is not possible.
Set a radiography distance that is within the possible range of motion.

Calling the Radiography Distance from a Radiography Program

1 Select a radiography step for


which radiography distance
information has been input in
the comment field.

2 Operate the X-ray tube support/X-ray radiography table to perform


horizontal tracking radiography.
For details on horizontal tracking radiography, refer to
"4.6.3 Horizontal Tracking Device".

The X-ray tube unit tracks at the radiography distance displayed as a comment for
the radiography step.

When the X-ray tube unit is used in combination with a radiography table that
can be raised and lowered, when the tabletop is raised, once the radiography
distance of the X-ray tube unit has reached the radiography distance displayed
as a comment for the radiography step, this radiography distance is maintained
in tracking beyond that point.
When the tabletop is lowered, the X-ray tube unit is not lowered and only the
tabletop moves.

112 CH-200/CHU-200 Operation Manual


4.7 Tomography (Option)

4.7 Tomography (Option)


4
* When the equipment is in CH-200 non-tracking mode, the tomography is not
available.
4

During operation of tomography, pay attention to avoid the patient from


4
rising up.
Rising up during operation of tomography may cause injury.
4

4
When an emergency has occurred due to failure in switches, etc.,
perform an emergency stop.
For details on emergency stop, refer to "5.2 Emergency Stop".
6
1 Confirm that the tomography is selected.
4
2 Press the (detent mode)
button on the X-ray tube 4
support's operation panel to turn
the detent mode ON.

3 Move the X-ray tube support.


4
Align the X-ray tube support with the
center of the X-ray radiography table in
the longitudinal and transversal
directions.

4 Turn the X-ray tube downward


so it faces the X-ray radiography
table.

CH-200/CHU-200 Operation Manual 113


4 Radiography

5 Turn the X-ray tube support so it faces the front. (The moving
direction of the X-ray radiography table and the operation panel of
the X-ray tube support are parallel.)
For details on the layout, refer to "2.5 Layout When Using Optional
Functions".

Be sure to turn the X-ray tube support so it faces the front.


Although radiography can be performed with the X-ray tube support facing
other direction, tomography images cannot be acquired.

6 Set the parameter for


tomography Pitch Speed

• Angle sets a perpendicular axis as


a center a maximum of
40 degrees.
• Speed sets up two kinds of speed,
Low or First.
• The layer height is height from
tabletop to a layer.
• The auto layer feed amount
increases and decreases the layer
height by the setting value for each
Layer Height Exposure
exposure.
• By the number of times of X-ray Auto Layer Feed
radiation, the number of times of
radiography can be set up
arbitrarily, and becomes fewer by a
unit of 1 time for every radiography.
• Pushing a key to set a parameter,
[up] and [down] keys are displayed
on the touch panel. (For pitch and
speed, adjust with cyclic operation
key.)
• Pushing of the right portions
of [up] and [down] keys, the [up]
and [down] keys disappear and the
setting value of a parameter is
saved.

114 CH-200/CHU-200 Operation Manual


4.7 Tomography (Option)

7 Load a cassette into the X-ray radiography stand’s Bucky.


Insert the grid or remove it, if required. 4

4
When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or

4
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired.

4
In case of that the cassette and the grid are not inserted correctly, [Not Ready]
on the touch panel of X-ray tube support is displayed. Be sure to lock the
cassette and the grid correctly.
4
8 Decide the patient's horizontal position.

9 Adjust the tabletop to the appropriate height by depressing the ( )


6
or ( ) footswitch.

10Move the X-ray tube support to adjust the distance (100 cm/ 4
40 inch) between X-ray tube and tabletop.
Adjust the X-ray tube support by moving vertically or operating horizontal tracking
function.
4
11Adjust the irradiation field with
the collimator while pressing the
collimator lamp button to light 4
the collimator lamp.

CH-200/CHU-200 Operation Manual 115


4 Radiography

Before performing tomography, ensure thoroughly that the X-ray tube


support is not positioned to contact the patient, and there are no
utensils, in the vicinity.
The X-ray tube support moves up to 70 cm to the left and right of the center
position during tomography. Contact with the X-ray tube support could injure a
patient or damage utensils.
In the case of the system with two X-ray tube supports, move another X-ray
tube support aside not to make contact with the X-ray tube support.

* Tomography operation (1)


The sequence of tomography shows as
illustrated below.
It operates so that the center of layer
exposed may be used as a fulcrum and Center of
the focus of the X-ray tube unit and the Pitch
Layer
position relation based on films may Layer
always come on a straight line.
While the X-ray tube unit moves for every
one X-ray radiography, the Bucky also
moves.
Layer Height
Although the X-ray tube unit and the
exposed film (Bucky) move during Film Bucky
radiography as shown in the figure of (1), (2)
(2) and (3), the position relation always
exists on a straight line across the center of layer.

(2) (3)

12Operate the X-ray high-voltage generator and perform radiography.

116 CH-200/CHU-200 Operation Manual


4.8 Bucky Tracking Radiography (Option)

4.8 Bucky Tracking Radiography (Option)


4
When the horizontal Bucky radiography technique is selected, moving the X-ray tube
support in longitudinal direction of X-ray radiography table, the Bucky of X-ray
radiography table starts to move toward the center of the X-ray tube. 4
The Bucky tracking radiography can be executed under the following conditions:

• The X-ray tube support is at the center position in the X-ray radiography table's
transversal direction. 4
• The tray is secured correctly.(Improper insertion is not permitted.)
• The grid is removed, or inserted inside for certain if the grid is an removal type.

4
(Incomplete inserting is invalid.)

For details on radiography using the horizontal Bucky technique, refer to


"4.5 Horizontal Bucky Radiography".

When an emergency has occurred due to failure in switches, etc.,


perform an emergency stop.
6
For details on emergency stop, refer to "5.2 Emergency Stop".

4
1 Press the (detent mode)
button on the X-ray tube
support's operation panel to turn 4
the detent mode ON.

2 Move the X-ray tube support to


4
bring it to the center of the X-ray
radiography table in the
transversal direction.

3 Move the X-ray tube support in the X-ray radiography table's


longitudinal direction.
The X-ray radiography table's Bucky will automatically move to and align with the
center of the X-ray tube.

For details on oblique radiography with Bucky tracking, refer to "When


Performing Oblique Radiography" P.118.

CH-200/CHU-200 Operation Manual 117


4 Radiography

4 Operate the X-ray radiography table.


For details on the operation of the X-ray radiography table, refer to "4.5
Horizontal Bucky Radiography".

5 Operate the X-ray tube support.


For details on the operation of the X-ray tube support, refer to "4.5
Horizontal Bucky Radiography".

6 Operate the X-ray high-voltage generator and perform radiography.

 When Performing Oblique Radiography

When performing oblique radiography with Bucky tracking, operate the X-ray tube
support as follows.

1 Operate the buttons on the


operation panel to position the
X-ray tube support in the
longitudinal direction.
The X-ray radiography table's Bucky
will automatically move to and align
with the center of the irradiation field.

118 CH-200/CHU-200 Operation Manual


4.8 Bucky Tracking Radiography (Option)

2 Operate the buttons on the


operation panel to turn the X- 4
ray tube to accomplish oblique
positioning.
The X-ray radiography table's Bucky 4
will automatically move to and align
with the center of the irradiation field.

If the center of the irradiation field 4


goes beyond the range of motion of
the Bucky, the Bucky remains stopped
at the outer limit of the range of
motion. If this happens move the X-ray 4
tube support again to bring the center
of the irradiation field within the
Bucky’s range of motion, whereupon
the Bucky will resume tracking and
4
move into alignment.

When the positions of the irradiation


field and Bucky are aligned, [Obliq] is
6
displayed on the touch panel.

CH-200/CHU-200 Operation Manual 119


4 Radiography

In oblique tracking, the error in the Bucky's tracking operation is 3 cm


(assuming SID = 100 cm). If the SID is greater than 100 cm, an error of greater
than 3 cm is generated.

When the Bucky Goes Out of the Tracking Range


If the Bucky goes out of the tracking range, [Automatic] is displayed at the top right of
the touch panel.

When [Automatic]* is displayed on the touch panel, the operator


must take responsibility to adjust the irradiation field before
performing radiography.
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

120 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4.9 Auto Positioning Function (Option)


4
If the desired position of the X-ray tube support is registered to the radiography program,
the X-ray tube support will move automatically to the registered position when the
operator selects the step and presses the appropriate switch on the remote control. The
operator can register the position of his/her choice for the X-ray support. After
4
registration, the X-ray tube support will move to the registered position at the touch of the
appropriate switch on the remote control.

This feature is called the "Auto Positioning Function". It reduces the amount of stress
4
operators exert when moving X-ray tubes during examinations.

4
In general, position data is registered during equipment installation. Position data can
also be registered after installation.

Note that when the long view radiography function (option) is used, for example, the auto
positioning function can be used regardless of whether or not a hand grip or patient stand
is fitted. 4
There are registration- and movement-related restrictions on the area depending on the
position of the X-ray radiography stand and whether or not a hand grip or patient stand is
fitted.
6
For details on the limitation of the area, refer to "5.4 The Area Where Cannot
Be Registered with Auto Positioning Function".

When an emergency has occurred due to failure in switches, etc.,


perform an emergency stop. 4
For details on emergency stop, refer to "5.2 Emergency Stop".

CH-200/CHU-200 Operation Manual 121


4 Radiography

4.9.1 Configuration

The system consists of the X-ray tube support, remote control, and X-ray high-voltage
generator, which support the auto positioning function.

[Inside Examination Room] [Outside Examination Room]

No. Unit Detail Remarks

Combine X-ray tube support,


X-ray Tube standard type.
1
Support Rear mount or low-ceiling type
cannot be combined.

Provided as standard for the use


2 Remote Control
inside the examination room.

X-ray High- Shimadzu X-ray high-voltage Refer to the


3
voltage Generator generator (touch panel type) operation manual.

Remote Control Optional for the use outside the


4
(option) examination room.

 Remote Control

The remote control provided with this system is intended for the use with the auto
positioning function.
For a standard use, install the remote control in the vicinity of the X-ray tube support to
operate inside the examination room.
For the operation outside the examination room, install it as an option for the use with
the X-ray high-voltage generator. In this case, the position data registered on each
switch are common with both remote controls.

122 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
1

3 4

4
4

6 5
4

4
No. Switch Name Function Remarks

Stops the equipment movement


1 Stop Switch
urgently.
6
Blinks during the equipment
2 Indicator movement. Lights up when position
is registered.

Moves to the position 4


Moves the X-ray tube support to the in accordance with
A/P Tracking
3 position registered on the the radiography
Switch
radiography step. stand/table height to
keep the SID. 4
Moves to the
Moves the X-ray tube support to the
registered position
position that the operator has
POS Switch regardless of the
4
4
registered voluntarily.
radiography stand/
(Movement to the POS Position)
table height.

Registers a position on A/P tracking


switch or POS switch. Press together
5 Key Switch
with the A/P tracking switch or POS
switch to register.

Restarts the equipment after it has


Stop Reset
6 been urgently stopped since the Stop
Switch
switch was pressed.

CH-200/CHU-200 Operation Manual 123


4 Radiography

* Remote Control Overlay


The remote control overlays are inserted into the POS switch.
A symbol corresponding to the position registered for each POS switch is printed
on the surface of the overlay. The symbols can be selected during system
installation.

* Magnetic piece on the rear of remote control


A magnetic piece is assembled to the rear of the remote control so that it can be
attached or detached on the surface such as steel wall

Keep the items such as magnetic card, floppy disk, or clock those influenced
by magnetic away from the remote control when handling the remote control.
If the magnetic card saved data of medical records is close to the remote
control magnetic piece, the saved data may be damaged.

124 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4.9.2 Movement Function


4
Each unit movement using auto positioning function is as follows.
Unit Movement Status
4
Vertical
4

4
Longitudinal

4
X-ray Tube
Transversal
Support
6
Horizontal axis
rotation*1
(Option) 4

4
Vertical axis
rotation*2

Irradiation field
X-ray
tracking
Collimator*3
on H/V axes

X-ray
Radiography Bucky tracking
Table*4

CH-200/CHU-200 Operation Manual 125


4 Radiography

*1: When the horizontal axis rotation is available


*2: When the vertical axis rotation is available
*3: When combined with Shimadzu auto collimator
*4: When combined with Shimadzu X-ray radiography table

* The horizontal axis rotation function and vertical axis rotation function are not
provided with the standard auto positioning due to optional.
Contact your Shimadzu service representative to set up this function.

The following two approaches can be used to move the X-ray tube support by means of
the auto positioning function.

• Movement to the A/P Tracking Position


Select this method to move the X-ray tube support to the position registered on
the radiography program. When this method is used, the equipment moves to a
position equivalent to the height of the radiography stand/table in order to
maintain the SID.

The equipment will move to the registered position regardless of the radiography
stand/table height if the general radiography technique is selected.

• Movement to the POS Position


The X-ray tube support moves to the position registered to each POS switch on
the remote control.
The equipment will move to the registered position regardless of the radiography
stand/table height.

126 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
If the auto positioning function is used, the X-ray tube support will not

4
stop if it makes contact with an object or person.
In case of emergency, let go of the A/P tracking switch or POS switch or press
the Stop switch.

Confirm that no patient is close to the X-ray tube support when the
equipment is in motion.
4
Injury may result if the X-ray tube support makes contact with a person.

6
Also confirm that the area around the X-ray tube support is free from
obstructions.

4
Damage may result if the equipment makes contact with an object.

Hand
Grip
4
Cassette
Holder

CH-200/CHU-200 Operation Manual 127


4 Radiography

4.9.3 Registration of A/P Tracking Position

 Preparing Registration

In CH-200 Tracking APR Mode


Before using the auto positioning function, configure and register the radiography steps
at the X-ray high-voltage generator.
Register A/P tracking positions at the X-ray high-voltage generator and then at the X-
ray tube support.
(For details, refer to the operation manual for the X-ray high-voltage generator.)

1 Hold down the [Setup] button on


the operation panel for 1
second or more.
The setup screen is displayed.

2 Press the [Program] key.


The radiography program registration
screen is displayed.

3 Press the [Program] key at the


radiography program
registration screen.

4 Press the [Edit] key.


The radiography step editing screen is
displayed.

128 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

5 Press .
The number table corresponding to 4
A/P tracking positions (A/P No.) is
displayed on screen.

4
The SID of the radiography technique
displayed by the radiography
4
technique button and the tube angle
are shown in the A/P No. Table. When
general radiography techniques are
used, the distance between the X-ray
4
tube and the floor is displayed instead
of the SID.

Each technique has its own A/P No.


4
table. There are four kinds, as listed
below:
Vertical Bucky Radiography Technique: S 6
Horizontal Bucky Radiography Technique:T
Tomography Technique: TM
General Radiography Technique: M
4
With the vertical Bucky, horizontal
Bucky, or general radiography
techniques, positions from 1 through 30
can be registered. On the other hand,
4
only 1 position can be registered with
the tomography technique.

6 Press the No. to register.


4
The color changes on the line of the
A/P No.
Select the No. of the SID and tube
angle you would like to register.

In the example, the following


selections have been made:
Horizontal Bucky radiography
technique, No. 2, SID: 120,
Tube angle: 0

When making a new registration, select


the No. with which no SID and tube
angle are displayed (No. 8 in the example).

CH-200/CHU-200 Operation Manual 129


4 Radiography

7 Press .
The selected A/P No. is registered to
the radiography step. Then the setting
screen for radiography step is
displayed.

Confirm that the button has


changed to .

The details of the A/P No. is as


follows.
P-T02
Registration No.
S: Stand position
T: Table position
M: Manual position
TM: Tomography position

* To restore, (not registering the


APR tracking position to the
radiography step), register No. 0
as listed on the A/P No. table.

8 Press .
The confirmation screen is displayed.
Press the [Save] key.

The A/P No. is reflected on the


radiography step, and is displayed.

130 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

 Registering the Position

X-ray Tube Support Operation 4


After the registration at the X-ray high-voltage generator, register the position with the
X-ray tube support.

(If the SID and/or tube angle is not registered to the selected A/P No. or will be
changed.)
4
1 Display the radiography condition screen.
4
2 Select the radiography step in which the A/P No. is displayed.

3 Press the switching tab. 4


The positioned screen is displayed.

Confirm that is displayed.


4

6
Press to display a pop-up screen

4
of the registered data.

Registered data is listed below:


No.: A/P No.
SID: SID
U/D: Distance between the X-ray tube
4
and the floor
TubeRot.: Tube angle
SwivelRot.: Angle around the vertical 4
axis*
*: The angle when the X-ray tube
faces up to the X-ray radiography
stand should be 0 degree.

For details on X-ray radiography stand layout, refer to "Limitation when auto
positioning function is enabled" P.40.

CH-200/CHU-200 Operation Manual 131


4 Radiography

4 Move the X-ray tube to the


desired position to register.

Remote Control Operation

1 Press the Key switch on the


remote control.

2
A/P Tracking
While pressing the Key switch, Switch

press the A/P tracking switch for


3 seconds or more.
When the registration is complete, an
electronic sound occurs from the Key Switch

remote control.

During the registration, the LED lamps


of the remote control light up gradually
in a direction from left to right. When
the registration is complete, all the
LED lamps light up.

Be sure to press the Key switch at registration before pressing the A/P
tracking switch. Otherwise, the equipment may start moving.

132 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
If the selected A/P No. is also registered on other steps, the position
settings in all those steps will be changed after you changed the setting
of the A/P No. 4

4
• If, after you select a radiography step, the radiography technique at the time
the A/P No. was registered is different from that displayed at the technique
button (e.g. when the technique has been changed by a technique button),
4
will not be displayed and you will not be able to register the position.

will not be displayed.


4

• If no A/P No. is registered for the radiography step(s), you will not be able to
register a position at the X-ray tube support. Also, the position can also not
4
be registered if no SID value is displayed except the general radiography
(i.e. if "- - -" is displayed).

CH-200/CHU-200 Operation Manual 133


4 Radiography

4.9.4 POS Position Registration

X-ray Tube Support Operation

1 Move the X-ray tube support to


the position to register.

Remote Control Operation

1 Press the Key switch on the


remote control.

2 While pressing the Key switch,


press the POS switch to register
POS Switch
for 3 seconds or more.
When the registration is complete, an
electronic sound occurs from the Key Switch

remote control.

* The position can be registered


on any of the 4 switches.

134 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
If the vertical position X-ray tube support is located at upper or lower limit, that
position cannot be registered.
The X-ray tube support positions indicated below cannot be registered.
Change the position that can be registered. 4
For details on the area where positions cannot be registered, refer to "5.4
The Area Where Cannot Be Registered with Auto Positioning Function".
If the X-ray tube support makes contact with the radiography stand when
rotated around the vertical axis, set the circumference angle of the vertical axis 4
to within 30° or over -90°.
Incorrect Correct

X-ray Tube Set an angle within


4
Support The angle setting 30° or over -90°.
exceeds 30°.
X-ray Radiography Stand

Even if the X-ray tube support and the center of the radiography stand are
positioned apart, set the circumference angle of the vertical axis to within 30°
6
or over -90°.
Incorrect
4
The position
cannot be
registered only by
locating apart. 4

CH-200/CHU-200 Operation Manual 135


4 Radiography

• If the X-ray tube support make contact with the X-ray radiography stand
even when the circumference angle of the vertical axis is set to within 30 or
over -90, set the angle to 0.
Set the position of the X-ray tube support so it is either aligned with the
center of the stand or is a safe distance from the stand's forward exposure
handles.

Incorrect Correct

X-ray Radiography Stand Center Vertical Axis: only 0

within 30
Incorrect Correct
or over -90

Forward
Exposure within 30°
Handle or over -90°

• After the registration is completed, operate with A/P tracking or POS


moving to move the X-ray tube support from the position that is not
registered, and then confirm the desired position is registered.
For details on A/P tracking operation, refer to "A/P Tracking Operation"
P.138. For detail on POS moving operation, refer to "POS Moving
Operation" P.141.

136 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
Accuracy in position
• When moving the X-ray tube support to registered position with auto
positioning function, the allowance of registered position and one after
4
actual travelling is 2 cm.
Height
Longitudinal
4
Transversal

4
2 cm
2 cm

SID=100 cm

X-ray Radiography Stand


6
Accuracy in vertical axis
• When rotating the X-ray tube support to registered position in vertical axis 4
with auto positioning optional function, the allowance of registered position
and one after actual travelling is 6 cm.
X-ray radiography table Center of vertical axis
4

4
X-ray tube
support rotated
to registered
6 cm position
SID=100 cm

Side cassette
(option)

CH-200/CHU-200 Operation Manual 137


4 Radiography

4.9.5 Radiography

There are two types of the radiography operations: A/P tracking operation (movement
to A/P tracking position) and POS moving operation (movement to POS position).

 A/P Tracking Operation

1 Select the radiography step whose A/P No. is registered.


The A/P tracking switch on the remote control will light up if a properly-registered
radiography step is selected.

2 Press the A/P tracking switch


on the remote control.
Press and hold until the X-ray tube
support stops moving. A/P Tracking
Switch

The X-ray tube support will not move from positions that cannot be registered.
Change the position of the support before attempting to move it. (Refer to P.
134 and P. 136.)

138 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

3 The X-ray tube support moves automatically.


When the X-ray tube support has come to the set position, an electronic sound 4
occurs from the remote control.

4
While the support moves to the set
position, the LED lamps of the remote
control turn on/off repeatedly from side 6
to side. When the movement is
complete, all the LED lamps light up.

4
The X-ray tube support stops moving if the A/P tracking switch is released.
Also, press the buttons on operation panel, so that the positioning operation
stops.

If the X-ray tube support does not stop even though the switch on the
remote control has been released, press the Stop switch to make it stop
suddenly.

CH-200/CHU-200 Operation Manual 139


4 Radiography

The X-ray tube support moves to registered position despite of height for
X-ray radiography stand or X-ray radiography tables as follows:
• Operation with general radiography
• Registered at position except center of X-ray radiography stand with vertical
Bucky radiography

Pay attention to the position of the stand/table during use.

4 Load a cassette into the Bucky of the X-ray radiography stand or the
X-ray radiography table.
Insert the grid or remove it, if required.

When inserting the grid into the Bucky, be sure to confirm the grid type
or the status of insertion. In case that the grid is incorrectly inserted, or
the status of insertion or grid type differ from radiography condition,
appropriate radiography images may not be acquired.

In case of that the cassette and the grid are not inserted correctly, the [Not
Ready] indicates on the touch panel of X-ray tube support. Be sure to lock the
cassette and the grid correctly.

140 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

5 Decide the patient's vertical or horizontal position.


4
The X-ray tube support may fail to move toward the A/P tracking position
because it reaches to the upper point or lower one with height for X-ray
radiography table or stand.
4
On this occasion, change the height for X-ray radiography table or stand to
operate X-ray tube support.
Upper point
4

6
6 Adjust the irradiation field with the collimator while pressing the
collimator lamp button to light the collimator lamp.

7 Operate the X-ray high-voltage generator and perform radiography.


4

 POS Moving Operation 4


1 Press a POS switch on the
remote control.
Press the POS switch that
4
corresponds to the position to which
you would like to move the X-ray tube
support.
POS Switch

CH-200/CHU-200 Operation Manual 141


4 Radiography

2 The X-ray tube support moves to the registered position


automatically.
When the movement is complete, an electronic sound occurs from the remote
control.

Release the POS switch to stop the unit's movement.


Also, press the buttons (except touch panel and collimator lamp button) on
operation panel, so that the positioning operation stops.

If the X-ray tube support does not stop even though the switch on the
remote control has been released, press the Stop switch to make it stop
suddenly.

When moving the X-ray tube support, pay attention to the position of the
radiography table/stand.

142 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

4
When the X-ray tube support moves toward the X-ray radiography stand at a
high position that will result in a collision, the X-ray tube support will stop short
of X-ray radiography stand before making contact.
4

2 Moves 4
Stops 1 Lowers

4
If the X-ray tube support stops, lower the Bucky of X-ray radiography stand.
Press the switch again to move the X-ray tube support.
When the hand grip or screen is attached to the X-ray radiography stand
4
(detection-enabled X-ray radiography stand only), the X-ray tube support will
not start moving if located at a position that may result in contact with these
components.
Remove the hand grip or screen and then press the switch on the remote 6
control.

CH-200/CHU-200 Operation Manual 143


4 Radiography

4.9.6 Status Display

The following tables show/describe the status of lamps on the remote control and
sounds emitted by the system during operation.

 Lamp Display Status

Name Display Movement

• Movement complete
ON
• Position registration complete

• When equipment is not in motion


(Switch released while equipment in motion;
Display Lamp OFF
operation canceled)
• When position registration is not operated

• During equipment movement


Blinking
• During position registration

• The radiography step that was selected is


ON
compatible with the auto positioning function.

When the equipment cannot function as described


below
• No A/P No. is registered at the X-ray high-
A/P Tracking
Switch Lamp* voltage generator for the selected radiography
OFF step.
• The radiography technique at the time the A/P
No. was registered is different from that
displayed at the technique button.
• A different X-ray tube is selected.

*It synchronizes with button on the X-ray tube support touch panel.

144 CH-200/CHU-200 Operation Manual


4.9 Auto Positioning Function (Option)

 Sound Emitted by System during Operation

Name Movement 4
Equipment has been activated by the A/P tracking switch
or POS switch.

4
[Single tone]
Stop switch has been released. (A tone will sound when
any Stop switch is released on any piece of equipment.)

When the operation is stopped as described below


• The X-ray tube support reaches to the upper point or
[A tone sounds two lower in vertical. 4
times.] • The X-ray tube support causes the collision to the X-
ray radiography stand.
• Error occurred.

[A tone sounds three Equipment has reached destination.


4
times.] Position has been registered.

Beep
(Prolonged sound)
When the equipment reaches to the height for tabletop of
X-ray radiography table. 4
Beep repeated When the equipment moves down below the height for
(Short period sound) tabletop of X-ray radiography table.

Beep repeated When the equipment moves longitudinally/transversally or 6


(Long period sound)* downward vertically.

*This system sound can be set at installation. For personnel safety, it is recommended
that the sound is set. 4

CH-200/CHU-200 Operation Manual 145


4 Radiography

4.10 Function Associated with X-ray


Diagnostic Table YSF-300 (Option)
When performing radiography by using a Bucky equipped with the YSF-300, use an X-
ray tube support that has functions corresponding to the YSF-300.

4.10.1 Features and Functions

 Features

• The tabletop of the YSF-300 can be operated with the joystick of the X-ray tube
support.
• Radiography can be performed with the Bucky of the YSF-300.

* The following radiography techniques are not available:


• Bucky Tracking Radiography
• Tomography
• Vertical Tracking Radiography

* Only the rear mount type of X-ray tube support CH200 can be combined with
the YSF-300. The low-ceiling type of X-ray tube support cannot.

 Operation Panel

The figure below shows the differences with respect to the buttons of the standard X-
ray tube support.

Refer to Fig. 2.3.

2 1

Fig. 4.1

146 CH-200/CHU-200 Operation Manual


4.10 Function Associated with X-ray Diagnostic Table YSF-300 (Option)

1 Joystick
Tilting the joystick causes the tabletop of the YSF-300 to move in either the longitudinal
or transverse direction. 4
Example: The tabletop moves to each direction, corresponding to the joystick
operations.
4

1
4
3 4

1 2
3
2
4
(Tabletop movement)
4
When used in combination with the X-ray wall stand (optional), the joystick can be set
as a tracking switch for vertical tracking movement. 4
* Vertical tracking movements can be performed by pushing the joystick.
6
X-ray Wall Stand
4

X-ray Tube Support


4
Bucky

4
* Vertical tracking movement is not possible with the Bucky of the YSF-300.

The X-ray tube support has no function for avoiding interference with the
YSF-300. Take care not to collide with the YSF-300.

2 Detent mode button


On the standard operation panel, the detent mode button is located at position1.
When the joystick is provided with an operation panel, the detent mode button is shifted
to the position2.

Refer to Fig. 3.11.

CH-200/CHU-200 Operation Manual 147


4 Radiography

 Touch Panel

Radiography Technique Indication


When the Bucky of the YSF-300 is used, select the horizontal Bucky radiography
technique.

* Select the horizontal Bucky radiography technique regardless of the tilt angle of
the YSF-300.

(Horizontal) (Vertical)

Warning Indication
If the X-ray tube support is positioned inside the operating range within which the
tabletop of YSF-300 can be tilted, the warning symbol is displayed on the touch panel.
When the warning symbol is displayed, the tabletop of the YSF-300 cannot be tilted at
all.

Warning Symbol

The limitations on tilting can be set to inactive. Take care not to collide
with when tilting the YSF-300.

148 CH-200/CHU-200 Operation Manual


4.10 Function Associated with X-ray Diagnostic Table YSF-300 (Option)

4.10.2 Operation
4
 Operation of X-ray High-voltage Generator

1 Press the radiography program


4
key on the touch panel to select
horizontal Bucky radiography.
Confirm that the technique key indicates
the horizontal Bucky radiography
4
technique.

4
 Operation of X-ray Diagnostic Table YSF-300

4
* Refer to the operation manual of the X-ray diagnostic table YSF-300 for detailed
information.

1 Instruct patient to grasp hand grips or hand rail during all table 6
adjustments.

2 Tilt the tabletop. 4


Tilt the tabletop so that it is level (angle of 0.
For radiography at the vertical position, tilt the tabletop to an angle of 88

4
When the tabletop is tilted to the horizontal or vertical the X-ray tube support
recognizes the horizontal or vertical position even if there is a tolerance of up
to 15 and controls the SID or collimator. Make sure that the horizontal 4
position is 0 or the vertical position is 88. The Source Image Distance (SID)
of the X-ray tube support is not displayed except at the angle of 0or 88.

CH-200/CHU-200 Operation Manual 149


4 Radiography

Take care not to collide with the cable of the X-ray tube support when operating
the X-ray diagnostic table YSF-300.

X-ray Diagnostic Table


YSF-300

Cable of the X-ray Tube


Support

 Operation of X-ray Tube Support

1 Move the X-ray tube support.


Move the X-ray tube support to the predefined position in detent mode operation.

 Operation of X-ray Diagnostic Table YSF-300


(Inserting Cassette)

Position the Bucky according to the following procedure to insert the cassette.

1 Push the collimator lamp switch.


The center beam of the collimator lights.

2 Align the center beam of the


collimator with the center marking
of the cassette tray handle.

Notch

3 Insert the cassette into the Bucky.


Center Marking

150 CH-200/CHU-200 Operation Manual


4.10 Function Associated with X-ray Diagnostic Table YSF-300 (Option)

4
On the status indication on the touch panel of the X-ray tube support,
[PBL] is not displayed, but [Automatic]* is displayed.
The operator must take responsibility for carrying out irradiation field
adjustment and performing X-ray radiography.
4
*For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

 Operation of X-ray Tube Support


4
(Irradiation Field Adjustment)

1 Move the tabletop to suit the Joystick 4


patient's position and adjust the
irradiation field.
Operate the joystick to move the
tabletop.
4
* Move the tabletop with the operating
joystick: the irradiation field lamp 6
automatically lights.

* When the horizontal Bucky radiography technique is selected or the X-ray tube 4
unit is oriented downward with the general radiography technique selected, the
tabletop can be moved.

Operate the tabletop with care when using the joystick. 4


• Take care not to trap the patient's fingers, hands or feet.

• Take care not to make contact with any person except the operator.

CH-200/CHU-200 Operation Manual 151


4 Radiography

 Operation of X-ray High-voltage Generator (Radiography)

1 Finely adjust the X-ray


radiography conditions of the X-
ray high-voltage generator.

2 Hold down the radiography


preparation button.

3 When the radiography button or


radiography ready indicator lights, Radiography Radiography
Preparation Button
press the radiography button on
Button
the operation panel while holding
down the radiography preparation
button.
• Radiography operation starts.
• The radiography button and the
radiography indicator on the operation
panel light during radiography.
• The radiography is completed when
the radiography indicator on the
operation panel goes off.

4 Release the radiography button.

152 CH-200/CHU-200 Operation Manual


4.11 Function Associated with X-ray Diagnostic Table YSF-500 (Option)

4.11 Function Associated with X-ray 4


Diagnostic Table YSF-500 (Option)
When performing radiography by using a Bucky equipped with the YSF-500, use an X-
ray tube support that has functions corresponding to the YSF-500.
4

4.11.1 Features and Functions


4

 Features
4
• The tabletop of the YSF-500 can be operated with the joystick of the X-ray tube
support.


Radiography can be performed with the Bucky of the YSF-500.
Bucky Tracking Radiography and Vertical Tracking Radiography are available
4
(Only when the table is set to the horizontal position).

* The following radiography techniques are not available: 6


• Tomography

 Operation Panel 4
The figure below shows the differences with respect to the buttons of the standard X-
ray tube support.

Refer to Fig. 2.3.


4

Fig. 4.2

CH-200/CHU-200 Operation Manual 153


4 Radiography

1 Joystick
Tilting the joystick causes the tabletop of the YSF-500 to move in either the longitudinal
or transverse direction.

Example: The tabletop moves to each direction, corresponding to the joystick


operations.

1
4
3
1 2
3 4
2
(Tabletop movement)

* Move the tabletop with the operating joystick: the irradiation field lamp
automatically lights.

* When the horizontal Bucky radiography technique is selected or the X-ray tube
unit is oriented downward with the general radiography technique selected, the
tabletop can be moved.

Operate the tabletop with care when using the joystick.

• Take care not to trap the patient's fingers, hands or feet.

• Take care not to make contact with any person except the operator.

The X-ray tube support has no function for avoiding interference with the
YSF-500. Take care not to collide with the YSF-500.

154 CH-200/CHU-200 Operation Manual


4.11 Function Associated with X-ray Diagnostic Table YSF-500 (Option)

 Touch Panel

Radiography Technique Indication 4


When the Bucky of the YSF-500 is used, select the horizontal Bucky radiography
technique. Make sure that the table is set to the horizontal position.

Warning Indication 4
If the X-ray tube support is positioned inside the operating range within which the
tabletop of YSF-500 can be tilted, the warning symbol is displayed on the touch panel.
When the warning symbol is displayed, the tabletop of the YSF-500 cannot be tilted at
all.
4

Warning Symbol 6

4
Take care not to collide with the cable of the X-ray tube support when operating
the X-ray diagnostic table YSF-500.
4

X-ray Diagnostic Table


4
YSF-500

Cable of the X-ray Tube


Support

CH-200/CHU-200 Operation Manual 155


4 Radiography

4.12 Operation of Dual X-ray Tube System


1 Confirm that the numeric
number "2" as 2nd tube in the
radiography technique key and
also in the tube selected
indicator are displayed on the
touch panel.
When these both numbers are in
agreement, the X-ray radiography can
be taken by selected radiography
technique with the selected X-ray tube.

If the displayed tube number does not agree with the number and radiography
technique of the selected X-ray tube, a warning message will be displayed on
the touch panel.
Indication Tube Number of
Radiography Technique Number of Selected X-ray Tube

Another tube is selected.

2 For positioning of an X-ray tube support, operate in the same


procedure as the case of single X-ray tube system.

Before performing tomography, retract the other X-ray tube support and
make sure that the two equipments do not come into contact with each
other.
The X-ray tube support moves up to 70 cm left and right of the center position
during tomography.
Contact with the X-ray tube support could injure a patient or damage to
utensils.

156 CH-200/CHU-200 Operation Manual


4.13 Program Data Synchronization

4.13 Program Data Synchronization 4


When setting about the radiography technique or the X-ray detector is added or changed,
follow the steps below to update the program data in the X-ray tube support to make it
consistent with the data in the X-ray high-voltage generator. 4
1 Press the tab on the touch panel of X-ray tube support.
Press and hold it approximately for 5 seconds.
4

A confirmation message will appear asking if you really want to synchronize


program data.
4

Synchronize program data? 6

4
2 Press [OK].
The transfer process of program data is started, and the progress status is
displayed. 4

Receiving program data, please wait.


4

When the screen notifies you of the completion of transfer ([OK] will be displayed),
the synchronization process has already been completed. Click [OK] to close the
message screen.

Synchronization is completed.

CH-200/CHU-200 Operation Manual 157


4 Radiography

This page is intentionally left blank.

158 CH-200/CHU-200 Operation Manual


Troubleshooting
This chapter describes information (Message) list
and operation during a Power Outage.
5 Troubleshooting

5.1 Information (Message) List


 Error Information

If an error occurs in the system, the and the button is displayed in the
upper right of the touch panel. (Some error messages may be displayed due to the
combination of equipment.) When the button is pressed, the error information
listed below are indicated in the pop-up window. In case that the error information is
displayed, take action in reference to the following list. If a same message is displayed
again after the corresponding action below has been taken, take note of the
identification number and error code, and contact your Shimadzu service
representative.

ID Error Information

Error code

Fig. 5.1 Error Information

ID Information Action
Stop using the X-ray radiography stand
E01 Wire error (stand up/down) and immediately contact your
Shimadzu service representative.

Stop using the X-ray tube support and


Alert: Wire cut. Turn off the unit. Contact
E02 immediately contact your Shimadzu
your service representative.
service representative.

160 CH-200/CHU-200 Operation Manual


5.1 Information (Message) List

ID Information Action
E03 Position error (longitudinal)
5
E04 Position error (tube angle)

E05 Position error (bucky)

E06

E07
Position error (transverse)

Position error (up/down)


5
E08 Position error (table up/down)

E09

E10
Position error (stand up/down)

Position error (stand tilt angle)


Turn OFF the power of the X-ray high 5
voltage generator, and turn it ON again.
E11 Position error (swivel angle)

5
E12 Communication error (stand)

E13 Alert for AC motor (longitudinal)

E14 Alert for AC motor (tube rotation)

E15 Alert for AC motor (bucky)

E16 Alert for DC motor (transverse) 5


E17 Alert for AC motor (up/down)

E18 Alert for AC motor (stand up/down)

E19 Position detection error (longitudinal) 6


E20 Position detection error (tube angle)

E21 Position detection error (bucky)

E22 Position detection error (transverse) Turn OFF the power of the X-ray high
voltage generator, and turn it ON again.
5
E23 Position detection error (up/down)
Position detection error (stand up/
E24

5
down)

E25 Position detection error (swivel rot.)

E26 Encoder error (stand up/down)

Confirm that nothing is touching the


E27 Detected button input at startup.
buttons and/or touch panel of the X-ray
tube support when turning the power
5
ON.

E28 Grid error of table

E29 Grid error of stand

E30 Position is not registered (U/D limit).

E31 Communication error (table)


Turn OFF the power of the X-ray high
E32 Communication error (X-ray Generator)
voltage generator, and turn it ON again.
E33 Communication error (collimator)

E34 Motor error (H-leaf)

E35 Motor error (V-leaf)

E36 Motor error (filter)

CH-200/CHU-200 Operation Manual 161


5 Troubleshooting

ID Information Action
E37 Sensor position detection error (H-leaf)

E38 Sensor position detection error (V-leaf)

E39 Sensor position detection error (filter)


Turn OFF the power of the X-ray high
E40 Force sensor error
voltage generator, and turn it ON again.
E41 Power Assist is not available.

E42 Motor drive is not available.

E43 Wrong power assist setting

Contact your Shimadzu service


E60 Usage Information is not registered
representative.

 System Status Information

In case that the set position or angle of the X-ray tube support is not aligned,
or etc. and the button is displayed in the upper right of the touch panel. When the
button is pressed, the system status information listed below are indicated in the
pop-up window. In case that the system status information is displayed, take action in
reference to the following list.

ID System Status Information

Fig. 5.2 System Status Information

ID Information Action
N01 Insert the cassette. Insert the cassette correctly.

N02 Exchange the cassette. Replace the cassette with a new one.

N03 Stand tilt is not locked. Lock the stand tilt.

Select the radiography technique that


N04 Another tube is selected. the No. of the X-ray tube to be operated
is registered.

Set the tilt angle of the X-ray tube


N05 Set tube angle within 3 degrees.
support within 3 degrees.

N06 Swivel center is displaced. Align the swivel position.

N07 Transverse center is displaced. Align the transversal center.

N08 Longitudinal center is displaced. Align the longitudinal center.

Set the angle of the X-ray tube support


N09 Set tube angle to 0 degrees.
to 0 degrees.

162 CH-200/CHU-200 Operation Manual


5.1 Information (Message) List

ID Information Action

N10 Unknown cassette size


Confirm the cassette size.
Refer to the operation manual for the X-
5
ray radiography table or stand.

Set the X-ray radiography stand angle

5
N11 Set stand tilt angle to 0 degrees.
to 0 degrees.

Turn OFF the power of the X-ray high


voltage generator, and turn it ON again.
N12 Low battery! If the same message is displayed

5
again, contact your Shimadzu service
representative.

Tilt the X-ray radiography table to a


N13 Table is not vertical or horizontal.
horizontal or vertical position.

N14 Table frame is in the exposure field.


Move the tabletop frame so it will not
interfere the irradiation field. 5
N15 Grid is being exchanged. Remove the grid, or insert it inside.

N16 Grid is installed. Remove the grid.

N17 Set the TOMO exposure number.


Set the tomography exposure number 5
to 1 or more.

Remove the cassette from the X-ray


N18 Cassette is loaded in the stand.

6
radiography stand.

Remove the cassette from the X-ray


N19 Cassette is loaded in the table.
radiography table.

Adjust the X-ray irradiation field position


N20 Longitudinal beam field is displaced.
in the longitudinal direction.

Adjust the X-ray irradiation field position


5
N21 Vertical beam field is displaced.
in the vertical direction.

Move the X-ray tube unit in the range of


5
N22 SID is out of range.
SID.

Manually rotate the X-ray tube around


the horizontal axis before slowly
N23 Tube rotation pin lock is jammed.
rotating it again toward the position
fixed by the pin.

Manually move the X-ray tube


5
longitudinally before slowly moving it
N24 Longitudinal pin lock is jammed.
again toward the position fixed by the
pin.

Press the (manual operation) button to


N25 Set the collimator mode to Auto. turn the LED off and place the
collimator in the automatic mode.

Move the patient stand and set the pin


N26 Set the patient stand.
to the setting base.

Press the SET button and wait until the


N27 Set the tube and bucky positions. motion completes and "Long View"
appears at the top right on the screen.

Remove the hand grip from the X-ray


N28 Hand grip is mounted in the stand.
radiography stand.

Point the top or bottom position with the


N29 Set the top or bottom position. line marker and press the button to set
the position.

CH-200/CHU-200 Operation Manual 163


5 Troubleshooting

ID Information Action
Lower the tabletop of the X-ray
N30 Lower the table height.
radiography table.

Increase the number of exposures or


N31 Expand the exposure region.
the irradiation field size.

Press the irradiation field size display


button and fix the irradiation field size,
N32 Set the irradiation field size.
or replace the cassette with one of a
larger size.

Adjust the irradiation field size to a


N33 Adjust to a proper view field.
proper setting.

Set the tilt angle of the X-ray tube


N34 Set tube angle within 60 degrees.
support within 60 degrees.

N35 Remove the screen from the stand. Remove the screen from the stand.

Change the tomography conditions (i.e.


N36 Set the appropriate TOMO setting. exposure distance, tube swing angle
and total exposure time).

Move the Bucky towards the center of


N37 Adjust the stand height. the movable range on the stand in
vertical direction.

Move the X-ray tube support toward the


N38 Adjust the longitudinal position.
center of the bucky of the table.

Move the X-ray tube support toward the


N39 Adjust the transverse position.
center of the bucky device of the table.

Release the foot switch and move the


Pinching was detected in the stand part.
N40 bucky 5 cm or more to a counter
Move the bucky to a counter direction.
direction.

Release the lock and operate the grip. If


the same message is displayed
N41 Operate after unlocking.
repeatedly, contact your Shimadzu
service representative.

Operate without applying excessive


force to the grip. If the same message is
N42 Excessive force was detected.
displayed repeatedly, contact your
Shimadzu service representative.

Release the lock again and operate the


grip. If the same message is displayed
N43 Release lock again for operation.
repeatedly, contact your Shimadzu
service representative.

Close the message and wait for a while


N44 Preparing. Press [Close] and wait.
without touching the equipment.

Time for periodic inspection. Contact your Shimadzu service


N60
Contact your service representative. representative.

Time for replacement : Wire rope. Contact your Shimadzu service


N61
Contact your service representative. representative.

Contact your Shimadzu service


N62 Memory write failure.
representative.

164 CH-200/CHU-200 Operation Manual


5.1 Information (Message) List

 Auto Positioning Information

If you attempt to use the Auto Positioning Function, the auto positioning information 5
listed below will be displayed on a pop-up window on the touch panel, such as when
operating, the X-ray tube support in a position that cannot be registered. Should this
occur, refer to the list below for appropriate actions.
5
ID Auto Positioning
Information 5

5
Fig. 5.3 Auto Positioning Information

5
ID Information Action
Register A/P No. for the radiography
A01 Current APR has no A/P number. step by the X-ray high-voltage
generator. 6
Select the technique registered to the
A02 Different from registered technique. A/P No., or change the registered
content of the A/P No.

A03 A/P position is not registered.


Register the position by the remote
5
control.

Manually move the X-ray tube support


A04 Pin lock is jammed. to the position other than that fixed by a
pin. 5
Manually move the equipment to the
A05 Out of range of movement. area where the equipment activates
with auto positioning function.

A06 This position cannot be registered.


Move the X-ray tube support to the 5
position where it can be registered.

Turn OFF the power of the X-ray high


voltage generator, and turn it ON again.
A07 Position is not registered (stand pin). If the same message is displayed
again, contact your Shimadzu service
representative.

Set the swivel angle between 30 and - Set the angle around the swivel axis
A08
90 degrees. within 30 degrees or over -90 degrees.

Set the swivel angle between 0 and -90 Set the angle around the swivel axis to
A09
degrees. within 0 degrees or over -90 degrees.

Moving toward stand. Lower the position of the Bucky on the


A10
Retract bucky on the stand. X-ray radiography stand.

Turn OFF the power of the X-ray high


voltage generator, and turn it ON again.
A11 Position is not registered (U/D limit). If the same message is displayed
again, contact your Shimadzu service
representative.

CH-200/CHU-200 Operation Manual 165


5 Troubleshooting

ID Information Action
Vertical tracking pos. is out of range.
Change the vertical position of the
A12 Change the height of bucky on the
Bucky for the X-ray radiography stand.
stand.

Lower the height position of tabletop for


Vertical tracking pos. is out of range.
A13 the
Change the table height.
X-ray radiography table.

Set the angle around the swivel axis to


A14 Set the swivel angle to 0 degrees.
0 degrees.

Move the patient stand in a safe


Moving toward stand.
A15 position to keep it clear of the X-ray
Move the patient stand to a safe place.
tube support.

Moving toward stand. Remove the hand grip from the X-ray
A16
Remove the hand grip. radiography stand.

A17 Stand tilt is not locked. Lock the stand tilt.

 Other Information

If wire disconnection is detected or if the STOP switch is pressed due to an emergency,


a message will appear with a red message bar in the upper section of the touch panel.
If this message is displayed, refer to the list below and take appropriate action.

Fig. 5.4 Other Information

ID Information Action
Stop using the X-ray tube support and
Alert: Wire cut. Turn off the unit. Contact
L1 immediately contact your Shimadzu
your service representative.
service representative.

Wait until the equipment completes the


Preparing for operation.
L2 preparation. Or press the Stop Reset
Please wait a while.
Switch on the remote control to restart.

L3 Stop button is pressed. Release OK! Turn the stop switch to restore.

Select the radiography technique in


L4 Another tube is selected. which the number of the X-ray tube to
be operated is registered.

To stop the movement, press a button


on the operation panel of the X-ray tube
Push button on operation panel to stop
L5 support.
the movement.
In the event of an emergency, such as a
switch failure, press the STOP switch.

166 CH-200/CHU-200 Operation Manual


5.1 Information (Message) List

ID Information Action

L6 Program data is being set!!


Finish defining a program on the X-ray
high-voltage generator.
5
After confirming safety in the area, raise
Caution: Press again to move up the
L7 the X-ray radiography table by pressing

5
tabletop.
the up switch again within 5 seconds.

Turn OFF the power of the X-ray high


voltage generator, and turn it ON again.
L8 Rear operation panel is not available. If the same message is displayed

5
again, contact your Shimadzu service
representative.

PA is in preparation. Wait without Wait for a while without touching the


L9
operating. equipment.

CH-200/CHU-200 Operation Manual 167


5 Troubleshooting

5.2 Emergency Stop


When an emergency has occurred in the movement of the unit due to failure in
switches, etc., press the stops switch provided on Shimadzu X-ray radiography
table or Shimadzu X-ray radiography stand or press the power OFF button provided on
the X-ray high-voltage generator. The motor will be stopped.

*• When the tracking switch kit is incorporated, the stop switch is on the
switch box.
• When the power assist function is used, the stop switch is also on the
rear operation panel (except for the low-ceiling type).
• When the auto positioning function is used, the stop switch is also on the
remote control.

If an emergency stop is performed, the message shown below appears.

Stop button is pressed.


Release OK!

To recover from the emergency stop, follow the steps below.

• Turn the stop switch clockwise.


• Press the stop reset switch on the rear operation panel (when the power assist
function is used, except for the low-ceiling type).
• Press the (release) switch on the remote control (only when the auto
positioning function is used).

Make sure that the equipment does not move abnormally when the stop
switch is turned clockwise to recover from the emergency stop.
If the equipment moves abnormally again, stop the equipment immediately
using the emergency stop procedure.

168 CH-200/CHU-200 Operation Manual


5.3 Operation During Power Outage

5.3 Operation During Power Outage


5
During power outage, pressing the lock release button will not disengage the lock.

For the safety of the patient as well as other individuals, apply a 15 to 20 kg force on
the device, for example, vertically to move the X-ray tube support vertically. This 5
enables you to move the X-ray tube support contrary to the lock.

6
Fig. 5.5 Operating during Power Outage

CH-200/CHU-200 Operation Manual 169


5 Troubleshooting

5.4 The Area Where Cannot Be


Registered with Auto Positioning
Function
In the case of position registration with auto positioning function, it may not to be
registered according to the equipment position or angle of vertical axis for X-ray tube.

The equipment cannot be moved to a position that cannot be registered.

The table below shows the areas where position registration is possible (relative
positioning between the X-ray tube and X-ray radiography stand) and the rotation angle
of the X-ray tube around the vertical axis for each area.

Area Angle of X-ray tube around the vertical axis


(1) 180

(2), (3) 30 to 90

(4) 0 to 90

(5) 0*

*Only combined with tilt type of X-ray radiography stand can be registered for auto
positioning. In this case, the X-ray tube must be aligned with the center of horizontal
direction of X-ray radiography stand.

For details on layout limitation, refer to "Limitation when auto positioning


function is enabled" P.40.

Refer to the figure below for the relationship between the X-ray tube and X-ray
radiography stand for each area.

170 CH-200/CHU-200 Operation Manual


5.4 The Area Where Cannot Be Registered with Auto Positioning Function

X-ray radiography stand: Left side


Area (4) Area (2) Area (1)
5

C
Film surface 5

D Center axis A
B 5
Area (5) Area (3)

5
Area (4) Area (2)

A, B, C and D in the figure are the distances from the X-ray radiography stand's 5
detection surface.

When the equipment is combined with a tilt type X-ray radiography stand, the distance
changes according to the tilt angle of the X-ray radiography stand's Bucky. Refer to the
figure below for each distance.
6
Tilt angle A B C D
0* 76 cm 112 cm
5
90 96 cm 132 cm 92 cm 148 cm

Except 0 and 90 97 cm 133 cm

*Including X-ray radiography stand without tilting.


5
When the hand grip (P type) or patient stand is fitted to the X-ray radiography stand
(only in cases where the X-ray tube support is combined with an X-ray radiography
stand that is detectable), area (3) and area (5) in the figure is designated as an area in 5
which the X-ray tube support cannot be moved (the area in which position registration
is possible is not changed). Refer to the table below for each distance.

condition A B C D
Hand grip
126 cm 162 cm 143 cm 199 cm
fitted

Patient stand
121 cm 157 cm 108 cm 164 cm
fitted

CH-200/CHU-200 Operation Manual 171


5 Troubleshooting

X-ray radiography stand: Right side

Area (1) Area (2) Area (4))

Film surface
C

A Center axis D
B
Area (3) Area (5)

Area (2) Area (4)

A, B, C and D in the figure are the distances from the X-ray radiography stand's
detection surface.

When the equipment is combined with a tilt type X-ray radiography stand, the distance
changes according to the tilt angle of the X-ray radiography stand's Bucky. Refer to the
figure below for each distance.

Tilt angle A B C D
0* 58 cm 58 cm

90 78 cm 78 cm 92 cm 148 cm

Except 0 and 90 78 cm 79 cm

*Including X-ray radiography stand without tilting.

When the hand grip (P type) or patient stand is fitted to the X-ray radiography stand
(only in cases where the X-ray tube support is combined with an X-ray radiography
stand that is detectable), area (3) and area (5) in the figure is designated as an area in
which the X-ray tube support cannot be moved (the area in which position registration
is possible is not changed). Refer to the table below for each distance.

condition A B C D
Hand grip
108 cm 108 cm 143 cm 199 cm
fitted

Patient stand
103 cm 103 cm 108 cm 164 cm
fitted

172 CH-200/CHU-200 Operation Manual


Maintenance
This chapter describes points for daily and periodic
inspections as well as maintenance parts.
6 Maintenance

6.1 Maintenance Check


• The Ceiling Type X-ray Tube Support CH-200/CHU-200, a precision medical
device, requires periodical maintenance checks as well as daily checks before and
after the use of the device in order to keep its life as long as possible.
• Daily and periodical checks would prevent the device from various possible
troubles, including freeze of the application, and support the proper and safe
operation of the device.
• Daily and periodical checks including wire ropes check are necessary to maintain
safety and high-performance of the device. If no check is performed, the equipment
life may decrease, or the equipment may be severely damaged or pose a serious
health risk to human body.
• The maintenance check not only includes daily and periodical checks, but also
covers replacement of maintenance parts and periodic replacement of parts.
• A part of the periodical checks and the replacement of parts require knowledge and
skills sufficient to use special tools or deal with dangerous situations in the process.
• For more information on periodical checks and periodic replacement of parts,
contact your Shimadzu service representative.

If any abnormality is found during the maintenance check, stop using the
equipment and contact your Shimadzu service representative.
When a user performs the maintenance check for himself or herself, please
keep in mind his or her own safety.
Shimadzu shall not be liable for any damage resulting from the checks other
than those by Shimadzu or your Shimadzu service representative.

174 CH-200/CHU-200 Operation Manual


6.2 Daily Inspection (Inspected by User)

6.2 Daily Inspection (Inspected by User)


6
Inspect the following items before operating the equipment, and if any abnormal points
are observed, please contact your Shimadzu service representative:

6
1. Check for cuts or twists in the cables.

1. Check for exterior damage.


6
(dents, fractures, etc.)
2. Check if the equipment wobbles or shakes. 3. For moving or rotating the
equipment:

2. Check if the grip wobbles or shakes.


• Check the lock release
function.
6
• Check if it locks properly.
• Check if it moves smoothly.

3. Check the function of each button.


3. Check the abnormal sound.
3. Check the abnormal odor.
1. Removing unnecessary items.
6
3. Check the balance of the vertical motion.
4. Check the warning message. 2. Is the collimator fixed?
4. Check the Buzzer.
6
Fig. 6.1

CH-200/CHU-200 Operation Manual 175


6 Maintenance

6.2.1 Cleaning and Disinfection

Be sure to turn the equipment power OFF before cleaning and


disinfecting the equipment.
Otherwise, a malfunction may occur in the equipment, or the equipment may
operate in an unintended way.
Also, thoroughly ventilate the room before turning ON the power after
disinfection work is complete.

Be sure to clean and disinfect the equipment periodically.


Cleaning and disinfection is very important to ensure that the equipment can
be used hygienically and safely. Strictly follow the methods prescribed.

Be sure to clean the equipment frequently and after each patient use.
While doing so, do NOT directly apply or spray any disinfectant, cleaner, or
water onto the equipment. Wipe down all contact surfaces using a cloth
moistened with an appropriate disinfectant or cleaner. Make sure the cloth is
NOT too wet. If it is, liquid may enter into system electronics, causing failure or
malfunction.

176 CH-200/CHU-200 Operation Manual


6.2 Daily Inspection (Inspected by User)

6
Observe the following precautions when cleaning and disinfecting the
touch panel:
• Turn the equipment power OFF before cleaning and disinfecting.
6
• Clean and disinfect the touch panel periodically.
• Wipe the surface of the equipment with a cloth moistened with disinfectants or
cleaner. If liquid gets inside the equipment, it can cause failure or accidents.
• Do NOT rub or hit the panel surface because it tends to scar easily.
6
• Wipe gently with moistened soft close to remove the dust from the panel
surface.
6

6
Use the following disinfectants and cleaner:

6
• Chlorine disinfectants
Sodium dichloroisocyanurate solution (1 % maximum)
Sodium hypochlorite solution (1 % maximum)
• Alcohol disinfectants
Commercially available isopropyl alcohol solution (Up to 99 wt% can be used)
Rubbing alcohol (76.9 - 81.4 vol% Ethanol, Isopropyl alcohol as an additive) 6
• Glass cleaner (Only onto the touch panel)

Do NOT use the following disinfectants:


6
• Disinfectants that corrode metals, plastics, rubber, or paint
• Disinfectants unsuitable for metals, plastics, rubber, or paint
• Spray-gas type disinfectants
• Volatile disinfectants
• Disinfectants that may enter the equipment

CH-200/CHU-200 Operation Manual 177


6 Maintenance

Use disinfectants at a minimum.


Repeated disinfection over a long time may lead to discoloring and cracking on
the equipment surface, and deterioration of rubber and plastic. If any
abnormality is found on the equipment after disinfection, stop using the
equipment immediately. Contact your Shimadzu service representative for
repair.

Do NOT use an organic solvent.


Organic solvents may change the surface color. If an organic solvent adheres
to the surface, wipe it off immediately.

When disinfecting unpainted metals, do NOT use chlorine-based


disinfectants.
Chlorine-based disinfectants may corrode the surface of the equipment. If
chlorine-based disinfectants adhere to the surface, wipe them off immediately.

When disinfecting resin parts such as reticule of the collimator, do


NOTuse rubbing alcohol.
Rubbing alcohol may lead to deformation or crack of resin parts such as
reticule of the collimator. Wipe it off immediately if it adheres to the reticule of
the collimator.

178 CH-200/CHU-200 Operation Manual


6.2 Daily Inspection (Inspected by User)

6
When cleaning resin parts such as reticule of the collimator, use cloth
lightly moistened, not soaked, with cold or warm water mixed with
neutral detergent that does not include organic solvent.
6
Rubbing alcohol, organic solvents or non-neutral detergents may lead to
deformation or crack of resin parts such as reticule of the collimator. Wipe
them off immediately if they adhere to the reticule of the collimator. 6
On completing the work, check the following points before switching the power ON
again.

• There must be no water or disinfectant adhering to the equipment.


6
• The tools used in cleaning and disinfecting work must be tidied away.

"3.1 Power ON/OFF" 6

6
When turning the power ON after cleaning, make sure the examination
room is properly ventilated.
Turning the power ON while any flammable gas remains in the examination
room could lead to fire, smoke, explosion or electrocution.
6

6
6.2.2 Inspection for Warning and Caution Labels

Periodically (once a year) inspect the labels attached on the equipment. 6


If any label is peeled or unreadable by stain or scratch, contact your Shimadzu service
representative for replacement of a new one.

"Warning and Caution Labels" P.xxvi

CH-200/CHU-200 Operation Manual 179


6 Maintenance

6.2.3 Checklist for Daily Inspection

Date of inspection:
Inspected by:

1 Visually inspect the items/areas listed below. Take the


recommended corrective action if you find any abnormalities.
Item/Area Corrective Action
 Equipment is free of unnecessary items. Remove unnecessary items.
 No cables are jammed, twisted, or stripped. Untangle or replace cables.

 No breakage (dents, fractures, etc.) is on the cover. Contact your Shimadzu service
representative

2 Manually inspect the following. Take the recommended corrective


action if you find any abnormalities.
Item/Area Corrective Action
 Collimator is locked in place. Lock completely in place.

 Grip does not wobble or shake. Contact your Shimadzu service


representative

3 Turn the equipment ON and inspect the following. Take the


recommended corrective action if you find any abnormalities.
Corrective
Item/Area
Action
 Equipment does not make an abnormal sound when moved/rotated.
 Equipment emits unusual smell.
 Lock is released to move/rotate the equipment.
Contact
 Lock is engaged for moving/rotating the equipment. your
 Equipment moves smoothly. Shimadzu
service rep-
 Buttons on the operation panel are in good working order. resentative
 The vertical motion of the equipment is in balance.
 The instrument does not move by itself simply by pressing the stop release
switch on the operation panel (when the power assist function is used).

4 Inspect warning message and buzzer.


• Warning Message
Warning Situation Corrective Action
Terminate the operation immediately and
Wire rope cut warning Wire rope is
promptly contact your Shimadzu service
message is displayed. disconnected.
representative

• Buzzer
Warning Situation Corrective Action
Terminate the operation immediately and
The buzzer sounds Wire rope is
promptly contact your Shimadzu service
continuously. disconnected.
representative

180 CH-200/CHU-200 Operation Manual


6.3 Periodical Inspection

6.3 Periodical Inspection


6
Even without any problem in daily inspection, the following items should be inspected
periodically.

Be sure to perform periodic inspection (every 6 months).


Failure to do this may cause serious accidents or significantly shorten the 6
lifetime of the equipment.
A pop-up window will appear six months after performing periodical inspection
and display a message. The message will be repeatedly displayed until
periodical inspection is performed. 6

Periodic inspections mainly check the equipment performance and the internal
mechanisms. The inspections require good knowledge of the internal mechanisms and
6
can also be dangerous. Contact your Shimadzu service representative to request a
periodic inspection. It is recommended to conduct periodic inspections every 6 months.
A fee is charged for periodic inspections after expiry of the warranty periods. 6

CH-200/CHU-200 Operation Manual 181


6 Maintenance

The maintenance work involves dangerous tasks. Be sure to request your Shimadzu
service representative to perform the periodic inspection.

• Periodical Inspection

Item Inspection Points


Bolts fixing, ceiling rail, X-ray tube
support, X-ray tube, collimator, and Is there loosening in bolts?
roller of each part

• Is the wire rope severed?


• Is the wire rope scratched or snagged?
• Is the wire rope corroded?
• Is the wire pulley scratched?
• Are there any scratches or obstructions
around the rail or in the wheels?
Wire rope • Check whether the wire cut detection
switch and the operation panel's wire rope
cut warning indicator are activated
properly.
Remove
the cover • Apply an aerosol oil (RO-300 or
and check equivalent) to the wire ropes on periodical
inspection.
Shafts and stoppers
(pins, screws, etc.) for Is it worn, deformed, bent, loose or dropped?
pulleys and sprockets

Bearings Is it worn, rusted or needs oiling?

Power supply Is input/output voltage appropriate?

Ceiling rails
(Fixed rail and moving Are there any damage or dirt on the rail?
rail)

Remove Are there any damage or dirt on the magnet


Magnet
the cover surface?
and check Lock pin mechanism Check the jolt or misaligned mechanism.

Tracking movement
(Vertical, Horizontal, and Is the accuracy in stop operation properly?
Bucky)
Check by
Is the tomography operation normally, or the
actual Tomography
image properly?
operation
Movement
(Vertical, Longitudinal, Is the force of each operation properly?
and Transversal)

182 CH-200/CHU-200 Operation Manual


6.4 Maintenance Parts List

6.4 Maintenance Parts List


6
To maintain the performance of the system, the following parts need replacing at
regular intervals. When replacing them, use the genuine parts and contact your
Shimadzu service representative.
6
• Maintenance Parts List

Part Required Replacement Implemented


Item Name
number quantity cycle by
6
Wire rope (main) 563-72780 2 5 years

Sub Wire Pulley

6
Parts for (incl. wire rope 563-64883-01 1 10 years
periodical (sub))
inspection
Sub Wire Pulley
(incl. wire rope 563-64883-02 1 10 years
(sub))

Fuse, 313005 Shimadzu


service
6
(250 VAC, 5 A, 072-01664-33 2 2 years
SLOWBLOW) representative

Fuse, 313010
Maintenance
parts
(32 VAC, 10 A,
SLOWBLOW)
072-01664-37 1 2 years 6
Memory battery
(Control unit)
074-73307-02 1 2 years
Lithium, CR2450,
3 V, 600 mAh 6
*1: The replacement period of the main wire rope differs depending on the frequency
of use. The main wire rope must be replaced every 5 years when the frequency of
use is under 260 days per year or 50 patient studies per day. Replacement may be 6
required in less than 5 years depending on the environment and conditions of use.

*2: In case the equipment is not used for 1 year or more contact your Shimadzu
service representative and make them remove the battery.
6

CH-200/CHU-200 Operation Manual 183


6 Maintenance

If the proper maintenance has not been scheduled, the wire cable will wear
with use quickly, and be broken in some cases. In the case where the wire
cables were broken concurrently, it is concerned that severe incident may
occur. It is recommended that an engineer properly trained by the official
Shimadzu Group maintain the system.

Scheduled maintenance policy


• Inspect the condition of wire cables every 6 months after installation of the
system is highly recommended. In the case where worn of cables is found,
immediately replacement of cables is needed.

• In the case where notification on the display screen is indicated after


installation, inspection by authorized engineer for maintenance service is
needed.

• Even if the above alert is displayed, but no damage to the wire cables is
observed as a result of the inspection, continue use of the system is
possible, but 6month inspection is highly recommended.

When the wire rope reaches the replacement time, a pop-up window will
appear and display a message. The message will be repeatedly displayed until
the wire ropes are replaced.

Prepare some locking mechanism to keep a off status of Molded Case Circuit Breaker.
When replacing the part, service engineer isolates from the supply mains.

184 CH-200/CHU-200 Operation Manual


Specifications
This chapter describes specifications.
7 Specifications

7.1 Environmental Conditions of EMC


(Electromagnetic Compatibility)

This section describes the electromagnetic compatibility (EMC) specification


for X-ray tube support CH-200.

The equipment satisfies the EMC (Electromagnetic Compatibility) standard below:

IEC60601-1-2:2007/2014

The equipment is suitable for installing at a professional healthcare


facility environment except below:
• Medical treatment areas with high-powered medical electrical equipment (High
frequency surgical equipment, short-wave therapy equipment)
• Inside the radio frequency shielded room of an MRI.

 Classification of EMI in Accordance with IEC 60601-1-2: 2007/


2014

Group 1, Class A

The system uses radio-frequency energy only for its internal function and is not
intended to deliver energy to the patient. But little leakage radio-frequency energy does
harm to highsensitive equipment.

The system main power line in the clinical site should be connected to the domestic
power sources which are separated from the public main network.

For replacement parts of internal components, make sure to apply the


cables supplied by Shimadzu.
The use of non-cable devices, accessories, or cables other than those sold by
Shimadzu as replacement parts for the internal components may result in
increased emissions or decreased immunity of the equipment.

186 CH-200/CHU-200 Operation Manual


7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)

7
The emissions characteristics of the system make it suitable for use in
industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is
7
normally required) the system might not offer adequate protection to radio-
frequency communication services.
The user might need to take mitigation measures, such as relocating or re-
orienting the system. 7
 Performance to be EMC Immunity Tested (Essential
Performance) 7
Essential performances of this equipment are as followings:



Positioning of X-ray irradiation field
Holding the X-ray tube unit 7
If an essential performance of this equipment is degraded or lost due to
electromagnetic disturbances, the following may occur.

• Operations for positioning the X-ray irradiation field cannot be performed. 6


• The X-ray tube unit moves during X-ray exposure.

CH-200/CHU-200 Operation Manual 187


7 Specifications

 List of Cables

Cable
No. Cable Name Shield Manufacturer
Length
1 POWER SUPPLY CABLE 16 m S Shimadzu

2 EARTH CABLE 30 m U Shimadzu

3 CH MAIN BODY SIGNAL CABLE 30 m S Shimadzu

4 CH POWER CABLE 30 m S Shimadzu

5 REMOCON CABLE 28 m S Shimadzu

6 X-RAY I/F CABLE AC 6m S Shimadzu

7 X-RAY I/F CABLE A 7m S Shimadzu

8 X-RAY I/F CABLE B 6m S Shimadzu

9 AC SERVO UVW CABLE 30 m S Shimadzu

10 AC SERVO ENC CABLE 30 m S Shimadzu

11 SIGNAL CABLE 1 15 m U Shimadzu

12 SIGNAL CABLE 3 15 m U Shimadzu

13 RELAY CABLE 1 4m - Shimadzu

14 CABLE LEAF H/V 25 m S Shimadzu

15 AC SERVO ENC CABLE J4 30 m S Shimadzu

16 AC SERVO UVW CABLE J4 30 m S Shimadzu

17 SERVO ENC CABLE J4 03A6 30 m S Shimadzu

18 SERVO UVW CABLE J4 03A6 30 m S Shimadzu

S: Shielded / U: Unshielded

The cables listed above are the parts specified to be compliant with the
standards. These parts are not provided to the equipment.

188 CH-200/CHU-200 Operation Manual


7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)

 Guidance and Manufacturer's Declaration - Electromagnetic


Immunity
7
Immunity test IEC 60601 test level Compliance level

8 kV contact
2 kV4 kV6 kV8 kV 7
Electrostatic discharge (ESD) contact
2 kV4 kV8 kV15 kV
EN 61000-4-2/IEC 61000-4-2 2 kV4 kV8 kV15 kV
air
air

Electrical fast transient / burst


2 kV for power supply
lines
2 kV for power supply
lines
7
EN 61000-4-4/IEC 61000-4-4
1 kV for input/output lines 1 kV for input/output lines

1 kV line(s) to line(s) 1 kV line(s) to line(s)


7
Surge
EN 61000-4-5/IEC 61000-4-5 2 kV line(s) to earth 2 kV line(s) to earth

0% UT: 0.5 cycle


0% UT: 0.5 cycle
at 0°, 45°, 90°, 135°, 180°,
at 0°, 45°, 90°, 135°, 180°,
Voltage dips, short interruptions
and voltage variations on power
225°, 270°, 315°
0% UT: 1 cycle at 0°
225°, 270°, 315°
0% UT: 1 cycle at 0° 7
40% UT: 5 cycles at 0°
supply input lines
EN 61000-4-11/IEC 61000-4-11 70%UT: 25 (50Hz) / 70%UT: 25 (50Hz) /
30 (60Hz) cycles at 0°
0% UT: 250 (50Hz) /
30 (60Hz) cycles at 0°
0% UT: 250 (50Hz) / 6
300 (60Hz) cycles 300 (60Hz) cycles

Power frequency (50/60 Hz)

7
magnetic field 30 A/m 30 A/m
EN 61000-4-8/IEC 61000-4-8

150 kHz - 80 MHz 150 kHz - 80 MHz

Conducted RF
EN 61000-4-6/IEC 61000-4-6
3 Vrms outside ISM bands,
6 Vrms in ISM bands
3 Vrms outside ISM bands,
6 Vrms in ISM bands
7
(80 % AM at 1 kHz) (80 % AM at 1 kHz)

80 MHz - 2.7 GHz 80 MHz - 2.7 GHz

Radiated RF 3 V/m (80 % AM at 1 kHz) 3 V/m 7


EN 61000-4-3/IEC 61000-4-3
(Refer to IEC 60601-1- (80 % AM at 1 kHz) (See
2:2014, table 9.) Table 7.1)

NOTE
• UT is the a.c. mains voltage prior to application of the test level.
• The ISM bands between 150 kHz and 80 MHz are 6.765 - 6.795 MHz, 13.553 MHz -
13.567 MHz, 26.957 - 27.283 MHz, and 40.66 - 40.70 MHz.

CH-200/CHU-200 Operation Manual 189


7 Specifications

Table 7.1 Test specifications for ENCLOSURE PORT IMMUNITY to


RF Wireless Communications Equipment

Test Frequency [MHz] Modulation Test Level

385 Pulse modulation: 18 Hz*1 19 V/m

450 FM ±5 kHz deviation: 1 kHz sine*2 28 V/m

710

745 Pulse modulation: 217 Hz*1 9 V/m

780

810

870 Pulse modulation: 18 Hz*1 28 V/m

930

1462 Pulse modulation: 217 Hz*1 10 V/m

1720

1845 Pulse modulation: 217 Hz*1 19 V/m

1970

2450 Pulse modulation: 217 Hz*1 19 V/m

5240

5500 Pulse modulation: 217 Hz*1 9 V/m

5785

*1:The carrier shall be modulated using a 50 % duty cycle square wave signal.
*2:As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while
it does not represent actual modulation, it would be worst case.

190 CH-200/CHU-200 Operation Manual


7.2 Specifications

7.2 Specifications
7
7.2.1 Specifications
7
Table 7.2 Specifications

Item Description 7
Max, allowable weight of
47 kg
suspension

Balancing system Spring-balancing type


7
Ceiling Fixed rail of 4 m or 5.5 m length
Rail
fixture Moving rail of 2 m, 2.6 m or 3.3 m length

Rail
Fixed rail of 4 m or 5.5 m length 7
Moving rail of 2 m, 2.6 m or 3.3 m length

Cabling in the

Ceiling Cable hose


longitudinal and
transversal directions
6
fixture using cable hoses
Cabling
Cabling in the
Cable bear (Option)
longitudinal and
transversal directions
7
using cable bears

Vertical 160 cm Electro-magnetic lock


Movement
of X-ray Longitudinal 295 cm*1
Lock pin mechanism
can be skipped when
7
Tube
moving at a speed
Support Transversal 140 cm*1
higher than 25 cm/s.*2

 180 *3 7
Around vertical Stops at every 90 angle.
axis The equipment can be fixed at an intended angle
with the electro-magnetic lock.
Rotation of
X-ray tube  180 to  120
support Stop at 0 and  90 angle.
Around The equipment can be fixed at an intended angle
horizontal axis with the electro-magnetic lock.
Lock pin mechanism can be skipped when
rotating at a speed higher than 80/s.

FFD (Focus- 3 digits, Digital display, Unit: cm (1 cm


Indication Film-distance) increment) or inch (0.5 inch increments)

X-ray tube angle 3 digits, Digital display, Unit: deg

CH-200/CHU-200 Operation Manual 191


7 Specifications

Item Description

SID 50 to 999 cm

(Unit: deg)
Ceiling tube support
(: clockwise,
mount position
: counterclockwise)
Display
range Upper left 0 to 120, 0 to 180
Rotation angle
Lower left 0 to 120, 0 to 180

Upper right (Standard) 0 to 180, 0 to 120

Lower right 0 to 180, 0 to 120

X-ray tube assembly tracks the center of the X-


ray field of the Bucky of Shimadzu's stand or
Vertical tracking other manufacturer's stand.
Tracking unit Tracking reference position: Top reference,
(Option) center reference, bottom reference

X-ray tube assembly moves to keep FFD


Horizontal
constant for Shimadzu's table or other
tracking
manufacturer's table.

Angle 8, 20, 30, 40


Tomography
Speed Maximum 40/sec
(Option)
Layer height 0 - 25 cm over tabletop

Size types: Inches or cm, vertical and horizontal


dimensions interchangeable
• When the size types are inches
1717, 1714, 1414, 1411, 1210, 108,
Irradiation field
177*4
• When the size types are cm
Tracking
4343, 4335, 3535, 4030, 3024, 2418,
Switch kit
(Option) 4318*4

Radiography
Top reference, center reference, bottom
reference
reference
position*4

Tracking
In conformity with the tracking function
operation

Bucky
Bucky tracking Bucky of BK-120, or BK-200 tracks X-ray tube
tracking
function assembly.
(Option)

Operation of the
Option for Joystick switch for operating the table top of
table top of YSF-
YSF-300 YSF-300.
300

Power assist Power assist Enables the X-ray tube support to be moved
(Option) function easily.

Auto X-ray tube assembly moves to the registered


Auto positioning
positioning position automatically by using the remote
function
(Option) controller.

192 CH-200/CHU-200 Operation Manual


7.2 Specifications

Item Description

Long view
Long view
radiography
Performs the operations required for long view
radiography in combination with the BK-120,
7
unit (Option)
function BK-200, BR-120, or BR-120T.

The irradiation field size can be selected the size


below the cassette of X-ray radiography table or 7
stand.
Size types: Inches or cm, vertical and horizontal
Irradiation field dimensions interchangeable
size selection
function
• When the size types are inches
1717, 1714, 1414, 1411, 1210, 108,
7
177, 1612, 9.59.5
• When the size types are cm
4343, 4335, 3535, 4030, 3024, 2418,
4318, 3530, 2424
7
Patient The patient information can be displayed on the
information touch panel in combination with some digital

Dongle Unit
display function radiography equipment. 7
(Functional Function for
Extension performing long All necessary radiography for long view
view radiography radiography can be performed by holding down
Option)
by holding down the radiography button once when combined 6
the radiography with certain radiography equipment.
button once

Tomosynthesis
function
Operations required for tomosynthesis can be
performed when combined with certain 7
radiography equipment.

Energy
Operations required for energy subtraction
subtraction
radiography
radiography can be performed when combined
with certain radiography equipment.
7
function

Serial Operations required for serial radiography can


radiography
function
be performed when combined with certain
radiography equipment.
7
Tube support of standard type with control panel
Front mount
attached to the front of the X-ray tube unit.

Tube support of type with control panel attached


to the opposite side of the X-ray tube unit. Eases
Rear mount*6
operations when X-ray tube support is installed
against the wall of the examination room.
Tube Tube support of front mount type with X-ray tube
support unit attached 150 mm higher. Maintains SID in
(option)*5 Low ceiling type
supine position radiography in examination
rooms with low ceilings.

Tube support of front mount type with X-ray tube


unit attached 80 mm lower. Eases positioning by
Front mount L preventing interference between the lower edge
of the tube support and the tabletop during
lateral radiography using the radiography table.

CH-200/CHU-200 Operation Manual 193


7 Specifications

Item Description

500 mm (width)  506 mm (height)  400 mm


Control Dimension
(depth)
cabinet*9
Weight 50 kg

Standard height
286 cm*7
of ceiling

A square of 250 cm  450 cm or more, with door


Installation area
opening of 60 cm width.
Installation Gross mass
requirement 250 kg*8
(Weight)

Single phase AC100 V, 0.2 kVA, Max. 5 ,


50/60 Hz
Power supply
Single phase AC200/220/230/240 V, 1.0 kVA,
Max. 4 , 50/60 Hz

*1: In the case of a 4 m fixed rail and a 2 m moving rail. For rails of other dimensions,
refer to "Table 7.3". In optional tomography radiography and optional auto
positioning, the longitudinal travel distance is 10 cm shorter.
*2: This is available in the longitudinal and transversal directions only when equipped
with a position detecting mechanism.

*3: - 90 to 30 for Low-ceiling type


*4: Can be set when the equipment is combined with certain X-ray radiography stands
only.
*5: Additions and changes to the tube support option offer ways to maintain operability
when the layout of the examination room has changed.
*6: Power assist, auto positioning and long view radiography are not available on a
rear mount support type.
*7: The ceiling height of 286 cm satisfies the specification of the X-ray tube support;
but the height can be less than this dimension but the vertical travel range may be
restricted.
*8: In the case of a 4 m fixed rail and a 2 m moving rail.
*9: CH-200 only.

Table 7.3 Length of Rail and Moving Range of Device (Unit: m)


Rail Moving Range of Device
Fixed Rail Moving Rail Longitudinal Transversal
4 2 2.95 1.4
4 2.6 2.95 2.0
4 3.3 2.95 2.7
5.5 2 4.45 1.4
5.5 2.6 4.45 2.0
5.5 3.3 4.45 2.7

194 CH-200/CHU-200 Operation Manual


7.2 Specifications

7.2.2 Distance from Spot to Reception Area


7
Distance from spot to reception area varies according to the unit combination.
Unit Combination Distance from

X-Ray Tube Unit Collimator


X-Ray
Radiography table
Spot to Reception
Area (mm)
7
BK-120 390-1725
CH-200/CHU-200
BK-200 390-1545
(Ceiling height:
2860 mm)
R-300
BK-120MK 384-1539 7
BK-12HK 384-1374

CH-200/CHU-200 Operation Manual 195


7 Specifications

7.3 Labeling
Location where the labels are attached on:

1, 4 - 6

2, 3

196 CH-200/CHU-200 Operation Manual


7.3 Labeling

 Name Plate

1 Name plate 7
CH-200

7
CHU-200

 Warning Label

2 Caution label

CH-200/CHU-200 Operation Manual 197


7 Specifications

3 Caution label

 Other Labels

4 Label, CE (For CE)

5 Label, cTÜVus mark (For USA)

6 Refer to operation manual label

198 CH-200/CHU-200 Operation Manual


7.3 Labeling

 Package for Transport

The label describing the storage environment and transport environment is shown 7
below.

7
Symbol Location Meaning

Package
Keep away from rain in case of
transport. 6

CH-200/CHU-200 Operation Manual 199


7 Specifications

7.4 Statement of Compliance


[For Europe]

This section describes the statement of compliance [for Europe] for X-ray tube
support CH-200.

7.4.1 Regulatory Information

For Europe:

This product complies with the requirement of the Medical Device Directive 93/42/EEC
and RoHS Directive 2011/65/EU.

Product Name: CEILING TYPE X-RAY TUBE SUPPORT

Model Name: CEILING TYPE X-RAY TUBE SUPPORT CH-200

Manufacturer: SHIMADZU CORPORATION Medical Systems Division

1, NISHINOKYO-KUWABARACHO,
NAKAGYO-KU, KYOTO, 604-8511, JAPAN

Authorized Representative Shimadzu Europa GmbH


/ Importer:
Albert-Hahn-Strasse 6-10,
47269 Duisburg, Germany

When any serious incident occurred in relation to the system, report that to
your Shimadzu service representative and to the competent authority of the
Member State in you are established.
“Serious incident” means any incident that directly or indirectly led, might have
led or might lead to any of the following:
• The death of a patient, user or other person
• The temporary or permanent serious deterioration of a patient's, user's or other
person's state of health
• A serious public health threat

200 CH-200/CHU-200 Operation Manual


7.4 Statement of Compliance [For Europe]

7.4.2 Company's Quality System


7
The company's quality management system complies with the requirements of Annex
II, excluding Section 4 of the MDD 93/42/EEC, which is certified by TUV Rheinland
LGA Products GmbH (Notified under No.0197)
7

7.4.3 International Standards 7


This equipment conforms the following international standards.

• IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+A1:2013


7
• IEC 60601-1-2:2014 / EN 606061-1-2:2015
• IEC 60601-1-2:2007 / EN 60601-1-2:2007


IEC 60601-1-3:2008+A1:2013 / EN 60601-1-3:2008+A11:2016
IEC 60601-1-6:2010+A1:2013 / EN 60601-1-6:2010+A1:2015
7
• IEC 60601-2-54:2009+A1:2015+A2:2018 / EN 60601-2-54:2009+A1:2015+A2:2019
• EN ISO 10993-1:2009+AC:2010
• ISO 14971:2007,Corrected version 2007-10-01 / EN ISO 14971:2012 6
• IEC 62304:2006 / EN 62304:2006+AC:2008
• IEC 62366:2007+A1:2014 / EN 62366:2008+A1:2015
• ISO 15223-1:2016 Corrected version 2017-03 / EN ISO 15223-1:2016
• EN 1041:2008 7

7.4.4 Protection Against Radiation 7


Refer to the standard about the acceptance and performance testing.
(ex:DIN 6868-150:2013-06)
7

CH-200/CHU-200 Operation Manual 201


7 Specifications

7.5 Statement of Compliance with


Standards

This section describes the Statement of Compliance with Standards for X-ray
tube support CH-200.

• X-RAY EQUIPMENT for RADIOGRAPHY ... CEILING TYPE X-RAY TUBE


SUPPORT CH-200 IEC 60601-2-54:2009+A1:2015+A2:2018
• X-RAY EQUIPMENT for RADIOGRAPHY ... CEILING TYPE X-RAY TUBE
SUPPORT CH-200 EN 60601-2-54:2009+A1:2015+A2:2019

202 CH-200/CHU-200 Operation Manual


7.6 Manufacturer Information

7.6 Manufacturer Information


7
Manufacturer: SHIMADZU CORPORATION

Address:
Medical Systems Division

1, NISHINOKYO-KUWABARACHO,
7
NAKAGYO-KU, KYOTO, 604-8511, JAPAN

CH-200/CHU-200 Operation Manual 203


7 Specifications

7.7 Information on Radiation

7.7.1 Deterministic Effect

Deterministic effects may occur when radiation dose on certain organs or body tissues
exceeds the threshold. The areas of interest of diagnostic radiology are the skin and
the crystalline lens. The threshold at which a deterministic effect occurs on these body
tissues is 1 Gy to 3 Gy.

The operator must control exposure dose properly in consideration of the above, when
using the equipment.

204 CH-200/CHU-200 Operation Manual


7.8 Product Safety

7.8 Product Safety


7

7
Do NOT operate this equipment if there is any uncertainty as to the
proper functioning of the system. refer all servicing to qualified service
personnel. 7

This equipment must be grounded. To minimize the shock hazard, make


sure of performing the ground work according to Installation Manual
7
(M514-E350).

7
The operator must set the focal spot to skin distance as large as possible
in order to keep the absorbed does to the patient as low as reasonably
achievable.
7

CH-200/CHU-200 Operation Manual 205


7 Specifications

This page is intentionally left blank.

206 CH-200/CHU-200 Operation Manual


Appendix
Appendix

Equipment Malfunction Checklist

Fill following blanks and contact your Shimadzu service representative.

Hospital:

Phone:

FAX:

Serial Number:

Date of Installation:

Question Answer
 Did the malfunction injure anyone?

 When did the malfunction occur?

 Did the system malfunction suddenly? Were there


any indications of a problem beforehand?
 Was there a power failure or thunderbolt when the
malfunction occurred?
 Has the system come in contact with moisture?

 Has the equipment been subjected to a strong


impact?
 How many patients per day are diagnosed using
the malfunctioning equipment?
 When was the last periodic inspection?

208 CH-200/CHU-200 Operation Manual


Appendix

Safety Explanation Record


8

8
 Name

 Phone

 Extension
Hospital
 FAX

 Address
8
 Room number

 Unit Name
8
Equipment  Serial number

 Installation Date

Shimadzu
 Dept./Section 8
 Phone
Rep.
 FAX
6
Administrator/
Direction Shimadzu Rep.
Operator
Date
Direction Subject
Signature Signature
8

• At the time of installation or when changing the designated operator, your


Shimadzu service representative will explain this manual's safety description of
this device's maintenance to operation staff and record the new information.

• This manual is important. We request all maintenance personnel to keep it in a


safe place close to the equipment.

CH-200/CHU-200 Operation Manual 209


Appendix

This page is intentionally left blank.

210 CH-200/CHU-200 Operation Manual


Index
Index

Numerals I
2-directional movement ......................................... 51 Information button ..................................................29
3-directional movement ......................................... 50 Installation area ....................................................194
3-directional release button ......................... 19,20,21 Irradiation field size display button .........................27

A L
Adjusting size of irradiation field ............................ 75 Length of rail and moving range of device ...........194
Applications ............................................................. 2 Lighting collimator lamp ..........................................74
Assist level indicator .............................................. 56 Longitudinal ............................................................18
Attachment position of X-ray tube support ........ 18,33 Longitudinal detent position indicator .....................22
Longitudinal movement ..........................................48
B Longitudinal release button ....................................20
Longitudinal/transversal movement ........................52
Balancing system ................................................. 191

M
C
Maintenance parts list ..........................................183
Cassette size indicator ........................................... 25
Max, allowable weight of suspension ...................191
CH-200 Non-tracking Mode ..................................... 7
Moving rail ..............................................................16
CH-200 Tracking APR Mode ................................... 6
Moving to longitudinal/transversal detent positions 63
Check sheet for daily inspection .......................... 180
Moving X-ray tube support .....................................45
Collimator fully open button ................................... 25
Collimator lamp button ................................. 20,21,74
N
Configuration ......................................................... 16
Control cabinet basic information indicator ............ 23 Name plate ...........................................................197
Control unit ....................................................... 16,17
O
D
Operating environment .............................................8
Daily inspection (Inspected by user) .................... 175 Operation during power outage ............................169
Detent mode .......................................................... 63 Operation panel ............................................16,17,19
Detent mode key .................................................... 23
P
E
Patient information tab ...........................................27
Equipment malfunction checklist ......................... 208 Periodical inspection ............................................181
Power OFF .............................................................44
F Power ON ...............................................................44
Power supply ............................................................9
Features ................................................................... 3
Fixed rail ................................................................ 16
Q
Front mount L type ................................................. 33
Quick ready key ......................................................23
G
R
Grip ........................................................................ 21
Grounding ................................................................ 9 Radiography condition indicator .............................26
Radiography distance indicator ..............................24
H Rear operation panel ....................................16,17,21
Rotating the X-ray tube support .............................59
H-axis rotation release button ................................ 19
Rotating to the preset position
around horizontal axis .........................................65
Rotation angle indicator ..........................................25

212 CH-200/CHU-200 Operation Manual


Index

Rotation around horizontal axis ..............................60


Rotation around vertical axis ..................................59

S
Safety explanation record .....................................209
Single-axis move button .........................................56
Specifications .......................................................191
Standard height of ceiling .....................................194
Stop release switch ................................................56
Stop switch .............................................................56

T
Technique key ........................................................22
Touch panel ............................................................21
Tracking button .......................................................20
Transversal .............................................................18
Transversal detent position indicator ......................22
Transversal movement ...........................................49
Transversal release button .....................................20
Type of grid/status indicator ...................................25

V
V-axis rotation release button ............................19,21
Vertical ...................................................................18
Vertical detent position indicator ............................22
Vertical movement ..................................................45
Vertical release button ............................................19
Vertical/longitudinal movement ..............................51
Vertical/transversal movement ...............................52

W
Weight ..................................................................194

X
X-ray tube support pull-up section ..........................16
X-ray tube support section .....................................16

CH-200/CHU-200 Operation Manual 213


Index

This page is intentionally left blank.

214 CH-200/CHU-200 Operation Manual


Dimension
Dimension

216 CH-200/CHU-200 Operation Manual


Dimension

CH-200/CHU-200 Operation Manual 217


Dimension

218 CH-200/CHU-200 Operation Manual


Dimension

CH-200/CHU-200 Operation Manual 219


Dimension

220 CH-200/CHU-200 Operation Manual


Dimension

CH-200/CHU-200 Operation Manual 221


Dimension

222 CH-200/CHU-200 Operation Manual


Dimension

CH-200/CHU-200 Operation Manual 223


Dimension

This page is intentionally left blank.

224 CH-200/CHU-200 Operation Manual

You might also like