Switching Handbook v4
Switching Handbook v4
Switching Handbook
Switching
Handbook
A Guide to Signal Switching in Automated Test Systems
No. 1716
4
th
Edition
4th
www.keithley.com
Switching
Handbook
Fourth Edition
TABLE OF CONTENTS
SECTION 3 Issues in Switch System Design
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.2 Basic Steps for Switch System Design . . . . . . . . . . . . . . . . . .3-2
3.3 Calculating Uncertainties . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.4 Switching Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.5 Cold vs. Hot Switching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
SECTION 6 Applications
6.1 Battery Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.2 Capacitor Leakage Measurements . . . . . . . . . . . . . . . . . . . . .6-3
6.3 Continuity Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
6.4 Insulation Resistance Testing . . . . . . . . . . . . . . . . . . . . . . . . .6-7
6.5 Combining Continuity and
Insulation Resistance Testing . . . . . . . . . . . . . . . . . . . . . . . .6-10
SWITCHING HANDBOOK
6.6 Insulation Resistance Testing
of Printed Circuit Boards . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
6.7 Contact Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
6.8 Temperature Scanning . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-18
6.9 Diode Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-23
6.10 Capacitance Measurements . . . . . . . . . . . . . . . . . . . . . . . . .6-28
6.11 Accelerated Lifetime Testing
of Cellular Phone Handsets . . . . . . . . . . . . . . . . . . . . . . . . .6-32
6.12 Power Supply Burn-In Testing . . . . . . . . . . . . . . . . . . . . . . .6-34
APPENDIX A Glossary
APPENDIX B Switch Card and Switch Module
Selector Guides
APPENDIX C Safety Considerations
Index
The Switching
Function
1.1 Introduction
Many electronic test systems use relay switching to connect multiple
devices to sources and measurement instruments. In some cases, mul-
tiple sources and measuring instruments are connected to a single
device. Switching allows automating the testing of multiple devices,
thereby reducing error and costs.
Designing the switching for an automated test system demands an
understanding of the signals to be switched and the tests to be per-
formed. Test requirements can change frequently, so automated test
systems should provide the flexibility needed to handle a variety of sig-
nals. Even simple test systems often have diverse and conflicting
switching requirements.
The test definition will determine the system configuration and
switching needs. Given the versatility that test systems must offer,
designing the switching function may be one of the most complex and
challenging parts of the overall system design. A basic understanding
of relay types and switching configurations is helpful when choosing
an appropriate switch system.
Section 1 describes the effects of switching on system perfor-
mance. Relay types, switching configurations, and switching hardware
options are also discussed.
- SECTION 1
For example, the offset current of the relays and the leakage resis-
tance of the boards, connectors, and cables may degrade the integrity
of high impedance applications. Contact potential and contact resis-
tance of the relays can reduce the accuracy of low voltage and low resis-
tance circuits. Switches may reduce the bandwidth of high frequency
signals. Crosstalk between channels on the card may limit the low-level
performance. The uncertainties that can occur will depend on the type
of signals being switched.
System speed can also be a critical issue in system accuracy. For
example, an erroneous reading will occur if a measurement is taken
through a switch before the relay has had sufficient time to settle.
Often, it’s necessary to strike a compromise between system speed and
accuracy. Factors that affect system speed include the triggering time
of the hardware, the relay actuation and settling times, and software
overhead.
Given the uncertainties associated with any new system design,
switch hardware specifications must be reviewed carefully to make cer-
tain they fit the application. Section 2 provides a detailed description of
switch card and mainframe specifications. The types of uncertainties
that may arise in the system often depend on the type of signal being
switched. Section 4 provides an overview of switching by signal type.
NC
b) SPDT COM 1 Form C
NO
c) DPST COM NO
2 Form A
COM NO
NC
d) DPDT COM
NO
2 Form C
NC
COM
NO
- SECTION 1
1
2
3
4
Figure 1-2. Scanner—a
one out of n
selector switch
Ch. 1 Ch. 4
Pin 1
R1-2 R1-3
Ch. 2 Ch. 5
Pin 2
Ch. 3 Ch. 6
Pin 3
HI
Voltage Ammeter
Source
LO
Using a matrix switch card allows connecting any input to any out-
put by closing the switch at the intersection (crosspoint) of a given row
and column. The most common terminology to describe the matrix
size is M rows by N columns (M×N). For example, a 4×10 matrix switch
card, such as the Keithley Model 7012, has 4 rows and 10 columns.
Matrix switch cards generally have two or three poles per crosspoint.
As shown in Figure 1-5, a 5VDC source can be connected to any
two terminals of the device under test (DUT). A function generator
supplies pulses between another two terminals. Operation of the DUT
can be verified by connecting an oscilloscope between yet another two
terminals. The DUT pin connections can easily be programmed, so this
system will serve to test a variety of parts.
When choosing a matrix card for use with mixed signals, some
compromises may be required. For example, if both high frequency
and low current signals must be switched, take extra care when review-
ing the specifications of the card. The card chosen must have wide
- SECTION 1
Columns
1 2 3 4 5 6 7 8
+ 1
5VDC
Source – 2
3
Function Rows
Generator 4
5
Oscilloscope
6
DUT
Card 1 Card 2
7012 7012
1 Columns 10 7001 11 Columns 20
Analog
Backplane
1
2
Rows
3
HI
LO
2
Rows
3
External
column
jumpers
Card 2
6
Rows
7
8
7012
- SECTION 1
Card 1 Card 2
7012 7012
1 Columns 10 7002 11 Columns 20
Analog
Backplane
1
2
Rows
3
Columns
wired
together
Card 3 externally Card 4
7002
Analog
Backplane
5
6
Rows
7
8
7012 7012
Matrix Expansion
A large system may require more rows and/or columns than a single
card can provide. A matrix can be expanded by joining the rows and/or
columns of several cards together. For example, Figure 1-6 shows how
the number of columns can be expanded by using two Model 7012
cards to make a 4×20 matrix. Three cards will make a 4×30 matrix, and
so on. Depending upon the switch card and mainframe, the rows of the
cards may be connected together through the backplane of the main-
frame or the rows may be connected externally. The rows of the 7012
cards can be connected through the analog backplane of the Model
7001 or 7002 Switch Mainframe. When using multiple matrix cards,
check the specifications to determine if the rows can be connected
through the backplane or if they must be wired externally.
- SECTION 1
Ch. 1
Ch. 2
Ch. 3
Ch. 4
Ch. 1
Device 1
Ch. 2
Device 2
Ch. 3
Device 3
Tree
The tree switch configuration shown in Figure 1-12 is an alternative to
the cascade configuration. When compared to the cascade configura-
tion, the tree technique requires more relays for the same size system,
but the isolation between a given path and any unused paths may be
somewhat better. This will reduce crosstalk and DC leakage. The tree
switch configuration is also used at frequencies greater than 10MHz.
The advantages of the tree configuration include the absence of
unterminated stubs and the fact that the channels have similar charac-
teristics. However, multiple relays in a given path mean there will be
greater losses.
- SECTION 1
1
2
3
4
5
6 A 1
7
B 2
8
1 C 3
9
10 D 4
11
12
13
14
15
16
Figure 1-12. Multiplexer (two-tier tree Figure 1-13. Single-channel blocking matrix
switching)
A B C D A B C D
1 1
4- way
power
2 divider 2
3 3
4 4
The Model 7016A 50Ω 2GHz Multiplexer Card, the Model 7017 800
MHz Multiplexer Card, and the Model 7038 75Ω 2.0GHz Multiplexer
Card all employ the tree configuration. The System 41 RF/Microwave
Signal Routing Mainframe can also be configured as a multiplexer.
- SECTION 1
This switching hardware option is the most flexible, because of the
variety of compatible cards designed for switching various signal types
(for example, high voltage, low current). These cards also make it easi-
er to design a system that combines various switching configurations,
such as matrix, cascade, tree, etc.
These systems can be expanded easily by adding more cards
and/or mainframes and are GPIB programmable.
Instruments with Integrated Switching Capability
A measurement instrument with integrated switching (sometimes
referred to as a data acquisition system) provides the convenience of
using a single instrument rather than multiple units. With only one
instrument involved, the hardware takes up less rack space, is usually
more cost-effective, and programming and triggering are less compli-
cated. However, instruments like this may not offer as many switching
card options for various signal types nor as many switch configurations
as stand-alone mainframes do. These instruments usually have a wider
measurement range with higher resolution and better accuracy than
plug-in data acquisition boards. Keithley’s Model 2700 and 2750
Multimeter/Switch Systems, the Model 2000-20 Scanning Multimeter,
and the Model 6517A Electrometer with the Model 6521 Low Current
Scanner Card are all examples of this type of instrument.
Plug-In Data Acquisition Boards
Plug-in data acquisition boards are connected to and controlled by a
computer, rather than a separate mainframe. PC plug-in cards are a
good choice if the application’s accuracy and resolution requirements
are lower (<16-bit), if the required sampling rate is high (1kHz and
above), or if a card-based form factor is preferable for the overall sys-
tem design. This type of data acquisition system is software dependent,
and the number of channels is limited by the space available in the
computer. Some systems have external expansion slots to accommo-
date more data acquisition boards.
With this approach, the engineer will be designing the entire meas-
urement system. This can be a complex process that includes choosing
appropriate signal conditioners, isolation circuitry, filtering, scaling,
formatting, etc.
Switch Card
and Mainframe
Considerations
2.1 Physical Implementation
A physical implementation of the switching configurations described
in Section 1.4 includes a circuit board (or card) with relays, connections
for inputs and outputs, and supporting circuitry. The cards are usually
available in the scan, multiplex, matrix, cascade, tree, and isolated (or
independent) switching configurations. Switch cards or modules are
designed to plug into a mainframe that supplies drive current for the
relays and various control signals.
In terms of the time required to complete a system, purchasing the
switching instrument is generally more economical than building one
out of components. Although one may be limited to certain configura-
tions with an off-the-shelf instrument, system wiring (as opposed to
switch design) can begin at once. Combining configurations makes it
possible for the final system to meet even complex switching require-
ments.
Switch card specifications are usually stated with a specific appli-
cation in mind, such as low current or high voltage switching. The
switch card specifications are based on the performance of the com-
plete card, not just that of the switching element (relay).
This section offers an overview of specifications for both the switch
cards and mainframes. Definitions for many of the terms and specifi-
cations discussed are listed in the glossary.
Rc Voff
Channel Ideal Contact
Resistance Switch Potential
HI
Offset
Channel Channel
Rin Cin Current Rout Cout
In Out
Ioff
LO
- SECTION 2
ideal switch is shown in series with the channel resistance (Rc), which
includes both the contact resistance of the switch and the conductors
of the circuit board. The contact potential (Voff) is an offset voltage
caused by temperature gradients across the signal path. This voltage
adds to the switched voltage. The offset current (Ioff) is a spurious cur-
rent generated by the relay, the connector, and the connecting circuit
board traces. This current will combine with the unknown current to be
switched.
This section not only describes the specification, but in some
cases, provides information on how to measure a given specification.
Information on how the specification may affect system performance
may also be provided, as well as compensation techniques, if applica-
ble. Ultimately, this section is a tool to help the user determine which
card is best suited for a particular application.
2.2.1 Isolation
Isolation is a measure of the leakage resistance between paths on the
switch card. The path can be between any terminal and earth ground
(common mode) or between any two terminals. For example, this can
be the resistance between any two channels (channel to channel) or
between the HI and LO inputs of a given channel (differential).
Isolation is specified in terms of resistance and capacitance. The isola-
tion should be as high as possible to avoid errors when switching high
impedance circuits.
It’s generally unnecessary to verify the isolation capacitance on a
switch card because the capacitance is a mechanical function and
should not change over time. In contrast, the isolation resistance does
change over time as it is affected by changes in humidity and by cont-
amination due to the environment or handling of the card.
Isolation resistance measurements are usually made by sourcing a
voltage, then measuring the resulting current using an electrometer or
picoammeter. The isolation resistance is calculated by R = V/I. Refer to
the instruction manual for the switch card for isolation measurement
procedures specific to that card; however, the following paragraphs
provide a general description of how to perform isolation measure-
ments.
Channel-to-Channel. This is a measure of the isolation between any
two channels on a multiplexer switch card. The measurement is made
with one channel open and one channel closed.
Example Measurement of a Two-pole Form A Card (Ch. 1 to Ch. 2)
1. Remove all connections to the card.
2. Connect the HI and LO terminals of Ch. 1 together.
3. Connect the HI and LO terminals of Ch. 2 together.
- SECTION 2
given conditions is usually determined by the spacing (gap) between
contacts.
If the gap is too small, an arc may develop when the contacts open
and current continues to flow. Arcing is common when switching high
voltage, but if the gap is adequate, the arc will quench itself. An arc due
to an AC signal usually quenches itself as soon as the voltage level pass-
es through zero.
Arcing can damage the relay or reduce its life. In addition, the RFI
(radio frequency interference) generated may disrupt high-speed logic
circuits in the system.
Maximum Current
Factors that affect the maximum current through the switch card
include the traces on the board, the connectors, and the relays. The
specification usually includes both the maximum carry current and the
maximum switched current.
Carry current is the maximum current the relay can tolerate once
the contacts have been closed. The carry current is limited by the cross-
sectional area of the path through the switch contacts. The carry cur-
rent specification applies only when “cold switching.” Cold switching is
defined as opening and closing the switch when no current is flowing.
Contact life is much longer when cold switching is used.
Maximum Power
- SECTION 2
HI
Four-Wire Sense Source
Ohmmeter
LO
HI HI
Ch. 1 Output
LO LO
HI
Ch. 2
LO .
.
.
HI
Ch. 10
LO
11. Open Ch. 1 and then close Ch. 2. Verify that Ch. 2 is within specifi-
cation.
12. Repeat this procedure to measure the LO channel resistance for
each channel.
- SECTION 2
4. Select the lowest range of the voltmeter.
5. Close Ch. 1.
6. Allow the reading to settle and verify that the contact potential of
the channel is within the specification.
7. Open Ch. 1 and repeat the procedure for all other channels.
2.2.6 Crosstalk
Crosstalk is a measure of the high frequency signal leakage from one
channel to another. It is the result of stray capacitance, mutual induc-
tance, and leakage resistance between channels and is generally given
in decibels at a specific frequency.
Figure 2-3 shows an example of crosstalk. In this example, a 10VAC
signal source (V1) is connected to a load resistor (R) through Ch. 1 of
the switch card. An AC voltmeter (V2) is connected through Ch. 2 to a
second signal source.
The crosstalk caused by impedance (Z) between Ch. 1 and Ch. 2 is
specified in decibels as:
V2
crosstalk (dB) = 20 log ____
V1
This equation can also be expressed in terms of RF power:
P2
RF power (dB) = 10 log ____
P1
To find the maximum signal on Ch. 2 due to the signal on Ch. 1, the
equation is solved for V2:
V2 = V1 [10(crosstalk (dB)/20)]
V1
R
10V
Z
Ch. 2
Signal AC
Source V2
Voltmeter
- SECTION 2
ZS ZL VL AC
VS
50Ω 50Ω Voltmeter
VS
Insertion Loss (dB) = 20 log ____
VL
VS
VL = ____________
10[Loss (dB)/20]
If VS is 10V and the insertion loss is <1.0dB, then:
10V
VL > ________ , or 8.9V
10(1.0/20)
2.2.8 VSWR
The Voltage Standing Wave Ratio (VSWR) of a switch card specifies how
well the connectors and switching signal path are matched to the char-
acteristic impedance of the transmission line. More specifically, VSWR
is the ratio between the voltage at the maximum point of the standing
wave and the voltage at an adjacent minimum of the standing wave.
With a VSWR equal to one, the transmission line has no reflected wave
present. With a VSWR greater than one, part of the switched signal is
reflected back to the source and less than the maximum power will be
transferred to the load. Low VSWR is crucial for switching systems that
are designed for signals involving multiple components in series.
The VSWR is also related to the reflection coefficient by:
ρ+1
|
VSWR = ______
ρ–1 |
where: ρ = reflection coefficient
The reflection coefficient is the ratio of the reflected wave voltage
to the incident wave voltage, and is calculated by:
ZL – ZS
ρ = ________
ZL + ZS
where: ZL = the impedance of the load
ZS = the impedance of the switch card
- SECTION 2
For some applications, it is necessary to disconnect a card from the
analog backplane. Most switch cards have jumpers that can be
removed to disconnect the card output from the backplane.
Some mainframes also have jumpers that can be removed to iso-
late certain groups of slots. For example, the Model 7002 Mainframe
has a jumper that can be removed to isolate slots 1–5 from slots 6–12.
Similarly, the Model 707A Mainframe has removable jumpers between
slots 3 and 4.
2.3.2 Triggers
Most scanner mainframes can be triggered externally via the trigger
input and can send an output trigger pulse when the relay contacts
have settled. Precise triggering is important in test system develop-
ment to ensure synchronization between the scanner and other instru-
ments in the system, such as sources and measuring devices.
An input trigger can be used to close an individual channel or to
initiate a scan of several channels. Trigger sources include manual
(front panel button), IEEE-488 bus, Trigger Link, internal timer, and
external trigger.
The output trigger may be connected to a measuring device, such
as a voltmeter. This will ensure the measurement is not made until the
switch contacts have fully settled.
Specific details on triggering can be found in the operating manu-
als of the various mainframes.
Issues in
Switch System
Design
3.1 Introduction
Section 1 covers switching configurations and hardware options,
Section 2 addresses switch card and mainframe considerations, and
Section 3 outlines the fundamentals of switch system design. In addi-
tion, Section 3 discusses issues such as calculating uncertainties, tech-
niques for minimizing switching time, and the use of cold vs. hot
switching.
When designing and selecting switching instrumentation, there
are several important factors to keep in mind:
• Switching configurations available
• Number of device pins and number of instruments
• Electrical specifications
• Speed
• Control bus compatibility
• Physical size
• Flexibility (to adapt to changing system requirements)
• Base cost
• Expansion cost
- SECTION 3
5. What’s the required operating speed? (For example, does it require
scanning and measuring ten batteries within one second?)
6. Are the switch mainframe and cards compatible with other equip-
ment in the system with respect to triggering, controls and con-
nections?
Once these questions are answered, create a schematic of the test
system, including as much detail as possible. After reviewing the con-
figuration, it may be possible to streamline the system. For example,
testing 4-wire resistors may require four switch poles per resistor.
However, if a common connection can be made to one side of all the
resistors, then only two switch poles would be needed per resistor.
Once this information has been gathered, determining the type
and number of switch mainframes and cards required is straightfor-
ward. Consider several alternatives to find the best fit. In some cases,
mixed signal types will limit the choice of switch cards. For example, it
may be necessary for a system to route high voltage signals at some
times and high current signals at others. In this case, there are a limit-
ed number of switch cards that can handle both the high voltage and
the high current signals. Two separate groups of cards may be needed
in these circumstances.
Once the hardware has been selected, evaluate the design again to
ensure its adaptability to new applications for future system
expansion.
- SECTION 3
Trigger Execution Time
Actuation Time
Settle Time
R
10kΩ
C
185pF V
example, the voltmeter trigger time and measurement time would also
be included in the total system time.
To increase overall system speed, synchronize all the instruments
in the system using hardware triggers. Hardware triggers are usually
much faster than software triggers because software triggers must first
be parsed/processed before being sent to the instrument.
- SECTION 3
SECTION 4
Switch Considerations
by Signal Type
4.1 Introduction
The quality of a switch system depends in large part on its ability to
preserve the characteristics of the test signals routed through it. For
example, when the test signal is a low voltage, the switching system
must minimize errors such as offset voltage and IR drops. Leakage cur-
rent may be a problem for high resistance and low current switching
applications. Depending on the type of test signal involved, specific
switching techniques must be used to maintain signal integrity
through the switch system. This section describes switching tech-
niques for a variety of test signals.
- SECTION 4
V Ch. 1 VS1
Ch. 2 VS2
Ch. 30 VS30
R Ch. 1
Ch. 2
Ch. 3
Ch. n
Switch Resistance
When switching a voltage source to multiple devices, it may become
necessary to compensate for voltage drops due to switch resistance. In
particular, if the devices have low resistance, the current flowing
through the switches may cause a significant voltage drop. To prevent
this problem, many voltage sources have remote sensing to correct for
500mA
HI
Contact resistance
5V Source is 1Ω for each switch IC <4V
LO
500mA 1Ω
HI
HI Sense 1Ω
5V Source
with
Remote 6V IC 5V
LO Sense 1Ω
Sense
LO 1Ω
Figure 4-3b. Using Remote Sense through a switch ensures the proper voltage
delivered to the IC
- SECTION 4
These unwanted voltage errors may be due to thermoelectric offset
voltage in the switch card and connecting cabling, switch film contam-
ination, magnetic interference, and ground loops. These sources of
interference and ways to eliminate or reduce them are discussed in this
section. For more information on low voltage measurements, refer to
Keithley’s handbook, Low Level Measurements.
Thermoelectric Offset Voltage
The contact potential or thermoelectric offset voltage is the key speci-
fication of a switch card designed for low voltage switching. Thermo-
electric voltage is the voltage generated by thermal differences at the
junction of dissimilar metals, such as between the nickel-iron reed
relays and the copper conductor to which they are connected. The
temperature gradient is typically caused by the power dissipated by the
energized coil. The offset voltage adds directly to the signal voltage and
can be modeled as an unwanted voltage source in series with the
intended signal. The offset voltage will cause an error in the applied
stimulus to a device under test or the value measured by the voltmeter.
As shown in Figure 4-4a, the offset voltage (E) of a single pole relay
is added into the circuit. As a result, the measured voltage (VM) will be
the sum of the source voltage (Vs) and the offset voltage (E). To mini-
mize this offset voltage, a low voltage switch card uses a two-pole relay,
+ +
VS VM VS ≠ VM
– –
ET1
+ –
+ +
If ET1 = ET2
VS VM
then VS = VM
– ET2 –
+ –
- SECTION 4
Switch Film Contamination
Over time, a contaminating film can form on the surface of a relay con-
tact. This film can increase the relay contact resistance, which can
make the switched voltages erratic when measuring or sourcing low
voltage. Voltages greater than 100mV are usually sufficient to clear this
contamination.
Using scanner cards with solid state switches is one way to avoid
this problem.
Magnetic Interference
Magnetic interference can be a problem in low voltage circuits. A high
rate of change in magnetic flux, such as that produced by a switching
power supply or by switching a high current signal on and off, can
induce a pulse of many microvolts in an adjacent circuit. This can
easily cause significant error in a low voltage circuit. This type of
interference can be minimized by separating the noise source and the
sensitive circuit as much as possible, by magnetic shielding, and by
reducing the enclosed area of the noise source and signal conductors.
Twist the HI and LO wires of each channel together to minimize the
enclosed area.
Ground Loops
Ground loops can easily occur in a complex test system. If a small
potential difference exists between two ground points, some ground
currents may flow through a sensitive part of the system. This may
occur only when certain switches are closed, so it can be very difficult
to diagnose. When possible, try to maintain a single system ground
point. When this is not possible, isolation techniques using optical
coupling or balanced transformers may help by increasing the effective
resistance between the two points, thereby reducing the common
ground current to a negligible level.
Capacitive
IN OUT Load
500V
R = ______ = 25kΩ
20mA
Inductive reaction voltage (L(di/dt)) must be less than the scanner
card’s maximum voltage rating. Figure 4-6 shows two typical clamping
circuits, one using a diode for clamping DC voltages and the other
using back-to-back zener diodes for clamping AC voltages.
- SECTION 4
High Voltage
Switch
+
Inductive
IN OUT
Load
Diode Clamped
(DC voltages)
High Voltage
Switch
+
Inductive
IN OUT
Load
Zener Clamped
(AC voltages)
HI
Electrometer Volts
Voltmeter
LO
Time
High Impedance
Voltage Source
HI
Guard Electrometer Volts
Voltmeter
LO
Time
- SECTION 4
NC
Model 7058
COM or 7158
HI
From Scan Card
NO
Current
Source #1
LO
NC
COM
HI
From NO
Current
Source #2
LO
A
LO HI
Ch. 3
COM
HI
From
Current
Source #2 Ch. 4
LO
A
LO HI
Figure 4-9. Using the Model 7013 Relay Card to switch current
Model 7053
High Current Card
Jumper
HI
From
Current Ch. 1
Source #1
LO
Jumper
HI
From
Current Ch. 2
Source #2
LO
A
LO HI
Figure 4-10. Using the Model 7053 High Current Card to switch current
- SECTION 4
HI
From
Current RS Ch. 1
Source #1
LO
HI
From
Current RS Ch. 2
Source #2
LO
V
LO HI
Ch. 2 Load #2
chosen, verify that the minimum voltage drop will not be so small that
it will be difficult to measure with the voltmeter.
- SECTION 4
4A Ch. 1 1A Load #1
Common
Path Ch. 2 1A Load #2
- SECTION 4
between channels). When switching a current to be measured, high
input isolation (the isolation between input HI and LO), is not as criti-
cal as channel isolation because there is normally very little voltage
present, only the voltage burden of the ammeter. However, when
switching a current source, very high voltage can be present across the
input depending on the load resistance; therefore, high input isolation
becomes critical.
To determine the leakage current at a particular voltage, apply a
step voltage of this magnitude to the circuit. This generates a transient
current that will gradually decay to a steady value, which is the leakage
current of the system for that particular path. Once the leakage current
is determined, it can be subtracted from subsequent readings on a par-
ticular channel. However, the leakage current is dependent on the
applied voltage, so this technique cannot be used for voltage sweeps.
To prevent leakage current on the switch card, take special care
when handling to prevent degradation of performance. Handle the
switch card by the edges to avoid contaminating it with dirt, body oil,
etc. If cleaning the board becomes necessary, follow the instructions in
the manual provided with the switch card. Humidity can also increase
leakage current. Placing the switch system in a low humidity room will
minimize this effect. Switch cards are typically specified to operate in
relative humidity of 50% or less.
One way to reduce leakage currents in the test fixturing is to use
good quality insulators such as Teflon® and polyethylene. Avoid materi-
als such as nylon and phenolics.
Another way to reduce leakage current due to cabling and test fix-
turing is to use guarding. By definition, a guard is a low impedance
point in the circuit that is nearly at the same potential as the high
impedance input terminal. Figure 4-14 shows an example of guarding
as applied to switching an ion chamber to an ammeter (pA) to measure
the ion chamber current (IX). An unguarded ion chamber is shown in
Figure 4-14a. The circuit shows that the full bias voltage appears across
the insulator leakage resistance (RL), so a leakage current (IL) will be
added to the measured ion current (IM = IX + IL). The leakage resistance
(RL) is due to the insulator of the ionization chamber and the coax
cable insulation.
In Figure 4-14b, a guard ring is added to the ionization chamber.
This guard circuit splits the leakage resistance into two parts. The coax
connections of Figure 4-14a have been replaced with triax connec-
tions. The voltage across RL1 is the picoammeter voltage burden, nor-
mally less than 1mV, so the resulting leakage current will be quite small.
The full bias voltage appears across RL2 (the leakage resistance
between the inside and outer shields of a triax cable). However, the
resulting leakage current does not flow through the meter, so it is not
added to the measurement.
HI
Note: RL
Cross-sectional represents pA
view the leakage
RL resistance of the IL LO
ion chamber
insulator as
well as the
cable leakage.
IM = IX + IL
b. Triax
Ion Chamber IX Connection
IM
HI
Cross-sectional Guard RL1 pA
view Ring
LO
Guard
RL2 IL
IM = IX
Electrostatic Interference
- SECTION 4
Triboelectric Currents
Triboelectric currents are generated by charges created by friction
between a conductor and an insulator, such as between the conduc-
tor and the insulation of a coax cable. This noise source can be
reduced by using special low noise cable that has a conductive coat-
ing (such as graphite) and securing the interconnect cabling to mini-
mize movement.
Electrochemical Currents
Electrochemical currents are generated by galvanic battery action
caused by contamination and humidity. Thorough cleansing of joints
and surfaces to remove electrolytic residue, including PC etchants,
body salts, and processing chemicals, will minimize the effect of these
parasitic batteries.
Settling Time
When a relay opens or closes, there is a charge transfer (on the order of
picocoulombs), which causes a current pulse in the circuit. This charge
transfer is due to the mechanical release or closure of the contacts, the
contact-to-coil capacitance, and the stray capacitance between signal
and relay drive lines. After a relay is closed, it’s important to allow suf-
ficient settling time before taking a measurement. This time can be as
long as several seconds.
If a step voltage is applied to the circuit, a transient current is gen-
erated. This current will gradually decay to a steady value. The time
needed to reach the steady value, or the settling time, can be used to
determine the proper delay time for the measurement.
See Keithley’s Low Level Measurements for more detailed discus-
sions of generated currents and guarding.
HI
Force
SMU2 Guard Row 2
Guard
HI
Force
SMU3 Guard Row 3
Guard
HI
Force
GNDU Guard Row 4
HI
Output HI
Guard Row 1
Guard
SMU1
HI
Output LO
Guard Row 2
HI
Output HI
Guard Row 3
Guard
SMU2
HI
Output LO
Guard Row 4
Drain
Gate
Source
- SECTION 4
of the switch and the Guard of the SMU is connected to the Guard ter-
minal of the switch. In this example, guarding becomes important to
avoid leakage currents because the FORCE terminal of the SMU is at
the test potential. With the guard terminal at the same potential, the
leakage current through the switch is minimized. Guarding will also
speed up the response time. The ground terminal (GNDU) of the Model
4200 is connected to a single row of the matrix.
Figure 4-16 shows two independent SMUs with the Output LO ter-
minals connected to separate rows. Note that each pin of the device
under test, the FET, is connected to a single column. The guard termi-
nals of the matrix card are not connected to the device. To connect the
SMUs to the transistor, close crosspoint Row 1/Column 2 to connect
SMU1 to the drain, Row 3/Column 1 to connect SMU2 to the gate, and
Row 2/Column 3 and Row 4/Column 3 to connect both SMU Output
LO terminals to the source terminal of the FET.
Remote sensing may be necessary if the current through a given
path is high enough to cause a significant voltage drop. Remote sensing
compensates for test lead and switch voltage drops and ensures that the
programmed output voltage of the SMU is delivered to the load. Remote
sensing allows making accurate load voltage measurements. Figure
4-17 shows SMU1 connected to four rows to enable remote sensing,
thereby allowing accurate measurement of the collector-emitter voltage
of the BJT. SMU2 is used to supply the relatively small base current and
requires only two rows. To connect SMU1 between the emitter and the
collector, close crosspoints Row 1/Column 4, Row 2/Column 5, Row
3/Column 2, and Row 4/Column 1. Note that remote sensing requires
the use of four rows and four columns. To connect SMU2 between the
base and the emitter, close crosspoints Row 5/Column 3 and Row
6/Column 1. Remote sense is not needed to source the base current,
because the lead resistance does not affect the current.
When connecting two SMUs in a matrix, the Output LO terminals
of both SMUs can be tied together and connected to the device using
one row. However, the SMU in remote sense may be outputting high
current (>1mA) and the resulting voltage drop may interfere with the
second SMU if it is used to measure small voltages (<1mV).
Once a given combination of matrix cards, sourcing, and measuring
devices is put together, it may be desirable to measure the offset current
for various pathways to characterize the system. To measure the offset
current, close a specific crosspoint and use an electrometer or SMU to
measure the current with everything in place except the device under
test. Periodically perform this system check on only the crosspoints for
low current switching. If the offset current is relatively constant, this
value can then be subtracted from subsequent measurements.
Leakage current is another source of measurement interference
that must be characterized. It is dependent on the applied voltage, con-
HI
Output HI
Guard Row 1
Guard
SMU1
HI
Sense HI
Guard Row 2
Guard
HI
Sense LO
Guard Row 3
HI
Output LO
Guard Row 4
HI
Output HI
Guard Row 5
Guard
SMU2
HI
Output LO
Guard Row 6
Collector
Base
Emitter
Figure 4-17. Using two SMUs to test a bipolar junction transistor (BJT)
- SECTION 4
4.4 Resistance Switching
Switching an ohmmeter to a device under test is common to a variety
of applications, including measuring the insulation resistance of mate-
rials, continuity testing of cables and connectors, contact resistance
measurements, and measuring the resistance of components such as
resistors, thermistors, and potentiometers.
Resistance measurements may range from less than 1Ω to greater
than 1012Ω, so the switching techniques required may vary significantly
based on the magnitude of the resistance. Low resistance measure-
ments are generally those less than 100Ω, mid-range are those from
100Ω to 107Ω, and high resistance measurements are those greater than
107Ω. As with any measurement, the exact point at which low resistance
techniques should be used depends upon the desired measurement
accuracy. For example, a 1ppm measurement of a 1kΩ resistor implies
an uncertainty of less than 1mΩ, which can only be achieved by using a
four-wire technique. Separate sections describe low resistance (Section
4.4.1) and high resistance (Section 4.4.2) switching techniques. This sec-
tion describes mid-range resistance techniques.
For scanning resistances in the mid-range (100Ω to 10MΩ), either
single-ended (one-pole) or differential (two-pole) methods can be
used. The single-ended method is shown in Figure 4-18a. Note that all
the resistors under test have a common terminal that is connected to
meter low.
Figure 4-18b illustrates the differential method. A two-pole relay is
used to connect both terminals of the unknown resistor to the meter.
The application usually dictates whether the single-ended or dif-
ferential method should be used for switching. The advantage of the
single-ended method is that it requires only half as many switches as
the differential method. The only advantage of the differential method
is that the offset voltages of the two switches tend to cancel each other.
However, given that these are microvolt-level voltages, the differential
method is not usually an issue for mid-range resistance switching.
HI
Ohmmeter
LO
- SECTION 4
timing is critical, offset compensation may not be practical, because it
takes twice as long to make a measurement.
Some DMMs provide a built-in offset compensation capability.
Two measurements are made, one at a positive current and another at
nearly zero current. The DMM then calculates the resistance
automatically.
Four-Wire Switching
Accurate measurements of low resistance in a system require a four-
wire connection to eliminate both lead wire and switch contact resis-
tance. Figure 4-19 is an example of four-wire switching. In this dia-
gram, two of the leads (HI and LO Source) supply current to the resis-
tors, while the other two leads (HI and LO Sense) measure the voltage
developed across the resistors. The source leads are connected to one
bank of switches and the sense leads are connected to the other bank
of switches.
To measure the resistance of the first resistor (R1), channels 1 (Ch.
1) and 4 (Ch. 4) are both closed. The actual resistance measurement is
made with either a DMM or a micro-ohmmeter with four-wire ohms
capability or with a separate current source and voltmeter.
As shown in Figure 4-19, two sets of two-pole relays are used for
low resistance switching. The poles in an individual relay will have
R1
Ch. 1 Ch. 4
R2
Ch. 2 Ch. 5
R3
Ch. 3 Ch. 6
HI Sense HI Source
Ohmmeter
LO Sense LO Source
HI Sense HI Source
Ohmmeter
LO Sense LO Source
Figure 4-20a. Offset voltage of Ch. 1 and Ch. 2 will be added to sense voltage
Ch. 1 Ch. 2
HI Sense HI Source
Ohmmeter
LO Sense LO Source
similar offset voltages. By using these two poles to connect the low volt-
age signal to the sense terminals of the meter, the offset voltages or the
contact potential (as it is specified), will tend to cancel each other.
When the sense leads are connected through two different relays to the
unknown resistor (R), as shown in Figure 4-20a, the offsets are not like-
ly to cancel and may actually combine to cause an even greater error.
Even a properly connected two-pole switch, like that shown in Figure
4-20b will have some differential offset voltage that will not cancel out.
The error caused by such offset voltage can be virtually eliminated by
using offset compensation.
In some applications, the resistances to be tested may have a com-
mon terminal. In this case, only one two-pole switch per resistor is nec-
essary, as shown in Figure 4-21. This approach has the advantage that
only half as many switches are required, so it is more economical. The
disadvantage is that the offset voltage of the single pole in the sense
- SECTION 4
R1
Ch. 1
R2
Ch. 2
R3
Ch. 3
HI Source LO Source
Ohmmeter
HI Sense LO Sense
Ch. 3 R3 R Ch. 6
HI
Voltage R = Current Limiting Resistor
Source pA
LO
- SECTION 4
R1 R Ch. 1
R2 R Ch. 2 Current
Switches
R3 R Ch. 3
short circuit current will not exceed the maximum current specifica-
tion of the relays on either set of switch cards. The voltage rating of
these resistors must be at least equal to the test voltage.
In some cases, the device under test may have a common terminal,
as shown in Figure 4-23. In this case, the voltage source will be applied
to all the devices simultaneously. As a result, the excitation voltage time
is different for each resistance to be measured. This will cause errors for
some time-dependent applications, such as measuring insulation
resistance or capacitor leakage. In these cases, the measured resistance
is a function of the excitation time.
R2 Ch. 2
R3 Ch. 3
HI
Voltage
Source pA
LO
Figure 4-24. Switching high resistances with common terminal and Form A
switches
Test 2
4 Row 2
Sample
3
Row 3
Row 4
+ –
Differential
Voltmeter
Figure 4-25. Van der Pauw resistivity measurements through a switch matrix
- SECTION 4
interferences and ways to avoid or minimize them are discussed in
detail in Section 4.3.1.
V
a. i C
S R
V
b. i C
Without series R
c. i With series R
Time
close to the relay as possible to limit the current when charging the
cable capacitance.
When an inductive load is connected to a voltage source, the cur-
rent will increase relatively slowly. However, when the switch is
opened, a large inductive reaction voltage will appear across the switch
contacts and may damage the contacts. The contact bounce that
occurs on closure can also produce an inductive reaction voltage
because the current is interrupted repeatedly. A voltage-clamping
device across the inductive load is usually required. Figure 4-27 illus-
trates four possible circuits for voltage clamping. For best results, the
voltage-clamping device should be located near the load.
Applications that involve switching inductive loads include testing
motors, solenoids, and transformers.
- SECTION 4
a. Varistor Clamped (AC or DC Voltages)
H H
L L
H H
L L
H H
H H C
In Switching Out Load
Card
R
L L
- SECTION 4
Impedance Matching: Given that the switch is positioned between the
measurement instruments and the DUT, matching the impedance lev-
els of all elements in the system is critical. For optimal signal transfer,
the output impedance of the source should be equal to the character-
istic impedance of the switch, the cables, and the DUT. In RF testing,
the most commonly used impedance levels are 50Ω and 75Ω.
Whichever impedance level is required, proper matching will ensure
the overall system integrity.
Termination: At high frequencies, all signals must be properly termi-
nated or the electromagnetic wave will be reflected from the terminat-
ing point. This, in turn, will cause an increase in VSWR. It may even
damage the source if the reflected power is large enough. All paths
through a system must be terminated with their characteristic
impedance.
Power Transmission: Another important consideration is the system’s
ability to transfer the RF power from instrument to DUT. Due to inser-
tion loss, the signal may require amplification. In other applications, it
may be necessary to reduce the signal power to the DUT. An amplifier
or attenuator may be needed to ensure that the required level of power
is transmitted through the switch.
Signal Filter: Signal filters can be useful in a number of circumstances,
such as removing spurious noise or limiting the bandwidth of the sig-
nal passing through the system.
Phase Distortion: As the size of a test system expands, signals from the
same source may travel to the DUT via paths of different lengths,
resulting in phase distortion. This specification is often referred to as
propagation delay. For a given conducting medium, the delay is pro-
portional to the length of the signal path. Different signal path lengths
will cause the signal phase to shift. This phase shift may cause erro-
neous measurement results. To minimize phase distortion, keep the
path lengths the same.
Cables and Connectors
Many different types of cables and connectors can be used in RF/
microwave switch systems. The signal frequency, the system impe-
dance, power rating, and test fixture/handler compatibility, etc. should
all be taken into consideration when choosing connectors and cables.
With the use of electromechanical switches, it becomes necessary
to use RF coaxial cable in order to route the signal path throughout the
test system with minimal losses. Coaxial cable should provide low
insertion loss, low VSWR and proper RF shielding with the appropriate
connectors on the ends of the cable assembly. Insertion loss is perhaps
the parameter of greatest concern when using coaxial cable, and given
that this specification increases proportionally with length, it is desir-
able to use the shortest length cables available for the application.
- SECTION 4
SECTION 5
Hardware
Implementation
5.1 Introduction
Section 5 describes ways to implement and test the switching hard-
ware. This section addresses connections and wiring, shielding and
grounding, hardware verification, and troubleshooting. Refer to the
user’s manual for the switching hardware for further information on
these subjects.
- SECTION 5
2700 and 2750 include isolation relays that allow the outputs to be con-
nected or disconnected from the backplane. Some matrix cards of the
Models 707A and 708A make connection to the backplane through
jumpers on the cards, while other cards that are optimized for low cur-
rent switching use external low noise jumpers.
In some applications, it is undesirable to connect a card output to
the backplane, so always check the schematics of the individual cards
before applying power to ensure backplane connections are not made
by default. When the output of a card must be isolated from the back-
plane, make connections to the output directly to the card.
Resistance Measurements
The first step to verify the system requires that resistors of known value
be placed in the test fixture. An example might be a system for testing
- SECTION 5
resistor networks. The individual resistor values should be similar but
not identical to make certain that the proper connections are made to
each element. For instance, if the nominal resistances are all 10kΩ,
then a suitable test device could have 10.1kΩ on Channel 1, 10.2kΩ for
Channel 2, etc. In this way, if one or more channels are incorrectly
wired, it will be obvious immediately.
If the range of resistance values to be tested is very wide, then tests
must be made with known resistors near both the maximum and min-
imum limits. If the system makes two-terminal connections to the
devices, a measurement near the minimum value may have excessive
error, indicating the need for four-terminal connections. This will
require twice as many switches but it will eliminate the switch and lead
wire resistances from the measurement.
At the maximum limit, a lower than expected measurement would
indicate excessive leakage resistance in the test fixture, the wiring, or
the switch cards, and appropriate steps must be taken to reduce this
error to an acceptable level. These steps include cleaning the test fix-
ture and adding electrical guarding. See Section 4.3.2 for details on
guarding.
Particularly for high resistance measurements, the system settling
time should be verified as sufficient to provide the desired accuracy.
Voltage Measurements
For voltage measurements, first verify that each voltage source is con-
nected to the correct channel. The straightforward way to check to
make sure the channels are connected properly is to connect known
voltages on each channel. For example, apply 1V to Channel 1, 2V to
Channel 2, etc.
To make certain that voltage offsets are not contributing excessive
error to low voltage (< 1mV) measurements, first take a series of read-
ings with a known stable source connected directly to the voltmeter.
Then, take another series of measurements of the same source through
a given channel of the switch system. Compare the standard deviations
from the two tests. Excessive voltage offsets can be compensated for on
each channel by replacing the voltage signal with a short circuit using
a clean copper wire. For each short-circuited channel, take at least ten
voltage readings and average them. Save the average value and use this
to correct future measurements made on that channel. For this to be
effective, the ambient temperature of the switch system must remain
relatively constant (plus or minus a few degrees Centigrade).
If this procedure is too cumbersome, then use the following
method: Connect a short-circuit to only one channel, then use the off-
set of that channel to correct measurements made on all other chan-
nels of the card.
- SECTION 5
SECTION 6
Applications
6.1 Battery Testing
Overview
Batteries are used in a rapidly expanding variety of consumer and
industrial applications. Their testing requirements typically depend on
their chemistry, size, specific use, and whether the cells are primary or
secondary.
Secondary (rechargeable) batteries are commonly tested using dis-
charge and charge cycling. The discharge characteristics of a secondary
battery provide important information about the battery’s capacity
and life. Charging/discharging a battery often takes several hours, so
it’s usually desirable to connect several batteries in series in order to
charge and discharge them simultaneously. The voltage of each battery
can be monitored during charging and discharging using a two-pole
scanner.
In some applications, it may be desirable to measure each cell volt-
age, both with and without a load. Monitoring the voltage decay over
time will give an indication of cell quality. While this is particularly true
for primary cells, this technique can also be used for secondary cells.
Using a scanner makes it possible to monitor the voltage decay of sev-
eral cells at one time.
The following paragraphs describe the switch configurations for
both the charge/discharge cycling test and the voltage monitoring test.
Switching Configurations
Figure 6-1 illustrates the switch configuration for charging/discharg-
ing a string of cells connected in series. In this configuration, a series
string of 40 cells is connected to a current source (Model 2400
SourceMeter® instrument), which charges/discharges the cells. A sepa-
rate voltage measurement instrument monitors the individual battery
voltages through a switch. (Note: In this application, the Model 2400
cannot be used for both current sourcing and voltage measurement,
because it is limited to a 5V difference between the input/output HI
and sense HI terminals.)
In this particular example, the current source is supplying ±10mA
to all 40 cells simultaneously, so all the cells can be either charged or
discharged at the same time. The Model 2700 Multimeter/Data
Acquisition System, equipped with a Model 7702 40-Channel Differen-
tial Multiplexer Module, is used to switch and measure the voltage of
each cell. Each cell has a voltage of 1.2V, so the total voltage across the
string of cells is 48V. It is important to ensure that the total cell voltage
across the string does not exceed the common mode voltage rating and
the maximum voltage level of the switch module.
To avoid damage to relays, synchronizing the switch/measure
sequence properly is critical. It’s particularly important to make certain
- SECTION 6
Model 2700/7702
Multimeter/Data Acquisition System
Charging
Ch. 1 1.2V
10mA
Discharging
Ch. 2 1.2V
HI
2400
Current
Source
Ch. 40 1.2V
LO
HI LO
Voltmeter
all other channels are open before closing a particular channel. Inad-
vertently closing multiple channels will short two or more cells and
may damage the relays. Adding a current limiting resistor or fuse in
series with each switch will help prevent this type of damage.
To monitor the voltage decay of a series string of cells, load resis-
tors must be switched across individual cells. Figure 6-2 illustrates a
switching system for testing 80 primary cells. In this example, there are
two sets of switches: one set of two-pole switches (two Model 7702
modules) is used to connect the cells to the voltmeter, and one set of
isolated switches (two Model 7705 40-Channel Control Modules) is
used to connect the load resistors (RL).
If the load resistor was connected directly across the voltmeter
input terminals, the current through the relay contacts would cause a
voltage drop and the measured cell voltage would be in error. This error
is eliminated if isolated relays are used to switch the load resistors to
each cell. The Model 7705 module has 40 isolated relays. In addition,
these extra isolated relays will allow the load resistors to be connected
or disconnected from each cell at any time during the test cycle.
APPLICATIONS -
Model 2750 Multimeter/Switch System
with two Model 7702 and two Model 7705 Modules
Ch. 81
Ch. 1 B1
RL
Ch. 82
Ch. 2 B2
RL
Ch. 160
Ch. 80 B80
RL
HI LO
Voltmeter
- SECTION 6
Switch Configuration
Figure 6-3 illustrates a capacitor leakage test system that uses the
Model 6517A Electrometer/Source, Model 7158 Low Current Scanner
Cards, and Form C Switch Cards, such as the Model 7111-S or 7169A.
The cards are installed in a Model 7002 Switch Mainframe.
In this test system, one set of switches (on the Model 7111-S or
7169A) is used to apply the test voltage to each capacitor. In the nor-
mally closed position, one end of the capacitor is connected to circuit
LO. When the switch is actuated, the capacitor is connected to the volt-
age source. Switch actuation is usually staggered (for example, two sec-
onds apart) so that each capacitor may be charged for the same period
of time before its leakage is measured. If the maximum test voltage is
110V or less, the Model 7111-S card may be used; otherwise, the Model
7169A card can be used for test voltages up to 500V. If voltages greater
than 500V must be applied, use an appropriately rated switch.
A second set of switches (on the Model 7158) connects each capac-
itor to the picoammeter after a suitable soak period. Note that before
the capacitor is switched to the picoammeter, the capacitor is connect-
ed to circuit LO. This allows the leakage current to flow continuously
while it is being charged up.
For this application, a single instrument provides both the voltage
sourcing and the low current measurement functions. The Model
7111-S or 7169A
Form C Switch Card 7158 Low Current Card
R C
R C
R C
Output Output
Diode
HI
6517A Electrometer/Source pA
LO
APPLICATIONS -
6517A is particularly useful for this application because it can display
either resistance or leakage current and will source up to 1000VDC.
After the capacitors have been tested, the voltage source should be
set to zero; some time must be allowed for the capacitors to discharge
before they are removed from the test fixture. Note that in Figure 6-3
the capacitors (C) have a discharge path through the normally closed
contact of the relays.
The resistor (R), which is in series with each capacitor, is an impor-
tant component in this test system. It limits the charging current of
each capacitor and also protects the relays in case a capacitor becomes
short-circuited. Also, the resistor limits the AC gain of the feedback
ammeter. In general, as the source capacitance increases, the noise
gain also increases. The resistor limits this gain to a finite value. A rea-
sonable value is one that results in an RC time constant from 0.5 to two
seconds. The forward-biased diode in series with the HI electrometer
(pA) terminal also serves to limit the AC gain.
A triax-to-BNC adapter (Model 7078-TRX-BNC) is used to connect
the Model 6517A to the Model 7158 card. The capacitors are connected
to the Model 7158 cards using low noise coax cables. Insulated wires
can be used to connect the Model 7111-S card to the capacitors. The
Model 7169A’s connections are made through mass terminated
connectors.
- SECTION 6
Switching Configuration
Figure 6-4 illustrates a typical continuity test circuit. Two banks of two-
pole switches are used to make four-wire resistance measurements on
20 conductors. To measure the resistance of Conductor 1 using the
Model 2700 Multimeter/Data Acquisition System in the four-wire
ohms mode, close Ch. 1. In the four-wire ohms mode, this will auto-
matically close Ch. 21 as well. This is repeated for each conductor.
To measure 20 conductors, a Model 2700 with one Model 7702 40-
Channel Differential Multiplexer Module is required. Use a Model 2750
Multimeter/Switch System with multiple Model 7702 modules if more
than 40 conductors must be tested at one time.
Conductor 1
Ch. 1 Ch. 21
Conductor 2
Ch. 2 Ch. 22
Conductor 20
Ch. 20 Ch. 40
HI Sense HI Source
Ohmmeter
LO Sense LO Source
APPLICATIONS -
7111-S 40-Channel Form C Switch Card
1
Ch. 1
Terminals 2
of Ch. 2
Multi-pin
Connector
40
Ch. 40
HI Output LO
HI
6517A
pA Voltage
Electrometer/Source or Source
487 LO
Picoammeter/Source
Switching Configurations
Figure 6-5 illustrates a test system for measuring the IR of any one ter-
minal of a multi-pin connector to all other terminals using the Model
- SECTION 6
7054 High Voltage Card #1 7054 High Voltage Card #2
Ch. 1 R R Ch. 11
1
Ch. 2 R R Ch. 12
2
.
.
.
Ch. 10 R R Ch. 20
10
HI
1000V 6517A Electrometer/Source pA
LO
APPLICATIONS -
The resistors (R) limit the charging current through the relays.
These resistors should be located near the relay to minimize cable
capacitance charge and discharge current. A typical R value is 100kΩ.
Switching Configurations
Figure 6-7 illustrates a combined continuity and IR system for testing
multiple conductors using a four-wire DMM or SourceMeter instru-
ment. The resistances R1 through R20 represent the conductor resist-
R1
Ch. 1 1 1´ Ch. 21
Ra
R2
Ch. 2 2 2´ Ch. 22
Rb
R20
Ch. 20 20 20´ Ch. 40
Output Output
HI LO LO HI
LO Source HI Source
Four-Wire
LO Sense DMM or HI Sense
SourceMeter
- SECTION 6
ances. To measure conductor resistance R1, close Chs. 1 and 21. The
resistances Ra and Rb represent leakage resistance between conduc-
tors. The leakage resistance can be measured between any two or more
conductors. To measure leakage resistance Ra, close Channels 1 and
22. This is essentially the leakage resistance between conductors 1 and
2, given that Ra is much greater than R1.
A single Model 2700 Multimeter/Data Acquisition System with a
Model 7702 40-Channel Differential Multiplexer can be used to test up
to 20 conductors. When measuring the leakage resistance with a DMM,
the maximum applied voltage is usually less than 15V. Also, the maxi-
mum measured resistance is usually not greater than 100MΩ. To test
the IR at a specified test voltage, use a test configuration such as the
Model 2400 SourceMeter instrument with Model 7011 Quad 1×10
Multiplexer Cards in a Model 7001 or 7002 Switch Mainframe.
If a higher test voltage is required or if a higher leakage resistance
must be measured, the circuit in Figure 6-8 may be used. In this dia-
gram, two Model 7154 High Voltage Scanner Cards are used to switch a
Model 2410 SourceMeter instrument and Model 2010 Digital Multi-
meter to eight conductors. This system can measure conductor resist-
ances as low as 0.1mΩ and leakage resistances as high as 300GΩ with a
test voltage as high as 1000V. Note that the Model 2410 and Model 2010
are not connected to the card outputs, but are connected to specific
channels of the scanner cards. The outputs of the cards are only used
to expand the system to measure a larger number of conductors. To
measure the resistance of R1, close Chs. 1, 10, 11 and 20. This will con-
nect the Model 2010 across R1. To measure Ra, the leakage resistance
between R1 and R2, close Chs. 1, 9, 12 and 19. This will connect the
Model 2410 across the leakage resistance (Ra).
APPLICATIONS -
7154 High Voltage Card #1 7154 High Voltage Card #2
R1
Ch. 1 1 1´ Ch. 11
Ra
R2
Ch. 2 2 2´ Ch. 12
Rb
R3
Ch. 3 3 2´ Ch. 13
Rc
R9
Ch. 8 8 8´ Ch. 18
LO HI
HI HI
LO LO
Output Output
(not used except (not used except
for expansion) for expansion)
Switching Configurations
Figure 6-9 outlines a system to test a single five-terminal (“W”) test
pattern. In this circuit, the Model 7011 Quad 1×10 Multiplexer Card is
used to connect the voltage source to terminals 1 and 5 together or to
terminal 3. The Model 7158 Low Current Switch Card is used to connect
the electrometer to terminal 2 or 4.
- SECTION 6
7011 Quad 1×10 7158 Low Current
Multiplexer Card Scanner Card
100kΩ 1
Ch. 1
Ch. 41
1MΩ 2
100kΩ 3
Ch. 2
Ch. 42
Output 1MΩ 4
5
Output
Equivalent circuit of
“W” pattern test coupon
HI
Voltage
Source 6517A Electrometer/Source pA
LO
Figure 6-9. System to test the Insulation resistance of one test coupon
APPLICATIONS -
7169A 20-Channel Test 7158
Form C Switch Card Coupon Low Current Card
R R1
Ch. 1 Ch. 21
R R2
Ch. 2 Ch. 22
.
.
.
R R10
Ch. 10 Ch. 30
Output Output
HI
Voltage
Source 6517A Electrometer/Source pA
LO
Five “W” patterns can be tested as described using one Model 7011
card and one Model 7158 card. Model 7111 Form C Switch Cards can be
used in place of the Model 7011 card. If this card is used, de-energizing
a channel will connect the terminal to circuit LO rather than the volt-
age source. This will eliminate the need for the 1MΩ resistors. When
using either the Model 7011 or 7111 Cards, the test voltage cannot
exceed 110V.
The 100kΩ resistors are used to limit the available current in the
event of a short-circuited test coupon. Choose current limiting resis-
tors that are no more than 10% of the lowest measured resistance.
Figure 6-10 shows a circuit to test a coupon with ten two-terminal
leakage paths. Each coupon can be thought of as a bank of isolated
resistors (R1, R2, R3, R4, etc.). Each resistor requires one voltage chan-
nel and one current channel. The Model 7158 card connects the elec-
trometer or picoammeter to one end of each resistor, while the Model
7169A 20-Channel Form C Switch Card connects the voltage source to
the other end of each resistor.
To test resistor R1, first energize Ch. 1 to bias the test coupon. After
a specified “soak” period, energize Ch. 21 to measure the resulting cur-
rent. Continue this process for all of the remaining resistances.
Note that when the channels are de-energized, the corresponding
resistor terminals are connected to circuit LO, allowing any charge
- SECTION 6
across the resistances to be dissipated when the resistance is not being
measured.
The current limiting resistors (R) protect the scanner cards from
damage if the test coupon develops a short circuit.
Low noise coax cables should be used between the test coupons
and the Model 7158 cards to avoid noise currents.
For detailed information on measuring the insulation resistance of
PCBs, refer to Keithley Application Note 302, “Insulation Resistance
Testing of Printed Circuit Boards.”
Switch Configurations
As Figure 6-11 shows, all four terminals of the ohmmeter are switched
in order to prevent the test lead, path, and switch contact resistances
from being added into the measurement. Note the HI and LO Source
terminals are connected to one set of two-pole switches, while the HI
and LO Sense terminals are connected to another set of two-pole
switches. To measure the resistance of Contact 1, close both Chs. 1 and
21, then measure the resistance with the ohmmeter. To measure
Contact 2, open Chs. 1 and 21, close Chs. 2 and 22, then measure the
resistance again.
A Model 2750 Multimeter/Switch System can be used to measure
the contact resistance because it can measure resistances less than 1mΩ
and offers offset compensation and dry circuit testing capabilities.
Contacts used in low level applications are often tested under dry
circuit conditions, in which current and voltages are limited to levels
that can’t cause changes in the physical and electrical condition of the
contact junction. In general, a dry circuit is one in which the open cir-
cuit voltage is low, usually 20mV or less, and the short circuit current is
limited to 100mA or less.
In some cases, the number of switches required may be reduced by
half by connecting the samples in a unique series manner, as shown in
APPLICATIONS -
Contact #1
Ch. 1 Ch. 21
Contact #2
Ch. 2 Ch. 22
.
.
.
Contact #20
Ch. 20 Ch. 40
HI HI
Source Ohmmeter Sense
LO LO
Figure 6-12. In this example, all the contact pins are connected in
series and each of the mating sleeves is connected to a two-pole switch.
In this case, a Model 2750 Multimeter/Switch System is configured with
a Model 7701 Low-Voltage 32-Channel Differential Multiplexer Module
and can test 32 contacts. Up to 160 contacts can be tested using five
Model 7701 modules in one Model 2750 mainframe.
- SECTION 6
Ch. 1
Ch. 2
Contact Pin
and
Mating Sleeve
Ch. 32
HI LO
Source Source
Ohmmeter
HI LO
Sense Sense
2750/7701 Multimeter/Switch System
7011 7011
Bank A&B Bank C&D
Ch. 1 Ch. 21
Ch. 2 Ch. 22
.
.
.
Ch. 20 Ch. 40
Output Output
2400
SourceMeter 2182
Instrument Nanovoltmeter
APPLICATIONS -
Twenty contacts can be switched using one Model 7011 card and
one Model 7001 Switch Mainframe. The Model 7011 card is configured
into two 20-channel banks. Two relays can be simultaneously opened
or closed automatically by using the mainframe’s four-pole switch
mode.
Switching Configurations
Thermocouples
Thermocouples are the most widely used temperature sensors. They
can cover a wide temperature range and provide a robust, reliable sen-
sor. The measured thermocouple voltage is related to the difference
between the unknown temperature and a reference temperature. This
reference temperature is determined from the cold junction reference
(CJR). When thermocouples are to be switched, the reference tempera-
ture may be obtained from an internal reference junction located on
the switch card, a simulated reference temperature, or an external ref-
erence junction. The choice of reference junction is limited by the
choice of switch card and by the method of cold junction compensa-
tion implemented by the measuring instrument.
Built-In (Internal) Reference Junction. This is the most convenient
method to use if the switch card includes the temperature reference.
This reference junction is built into the switch module and uses a
thermistor, a solid-state sensor, or other type of temperature sensor.
Switch modules with built-in CJRs include the Model 7700, 7706, and
7708 modules for the Model 2700 Multimeter/Data Acquisition System
and Model 2750 Multimeter/Switch System. The Model 2001-TCSCAN
9-Channel Thermocouple Scanner Card for the Model 2000, 2001,
2002, and 2010 DMMs also has a built-in CJR.
Figure 6-14 shows multiple thermocouples connected to the
Model 2700 with a Model 7700 20-Channel Differential Multiplexer
Module installed. The Model 7700 has several strategically placed cold
- SECTION 6
2700/7700 Multimeter/Data Acquisition System
T1 Ch. 1 Voltmeter
T2 Ch. 2
. .
. .
. .
T20 Ch. 20
Cold
Junction Ref. Ch.
Reference
X6
Figure 6-14. Using the internal reference junction of the Model 2700/7700
APPLICATIONS -
2700/7702 Multimeter/Data Acquisition System
T1 Ch. 1 DMM
T2 Ch. 2
. .
. .
. .
T40 Ch. 40
Figure 6-15. Using simulated reference temperature with the Model 2700/7702
- SECTION 6
2700/7700 Multimeter/Data Acquisition System
A A
T1 Ch. 1 DMM
B A
A A
T2 Ch. 2
B A
. .
. .
. .
A A
T40 Ch. 40
B A
B A B A B A
0°C
Platinum RTDs
APPLICATIONS -
2700/7702 Multimeter/Data Acquisition System
Ch. 1 DMM
T1 Ch. 2
. RT
.
.
.
.
.
T39 Ch. 40
- SECTION 6
2700/7702 Multimeter/Data Acquisition System
Source HI
Sense HI
T1 RTD #1 Ch. 1 DMM
Sense LO
Source LO
T2 RTD #2 Ch. 2
.
.
.
APPLICATIONS -
2700/7702 Multimeter/Data Acquisition System
T1 Ch. 1 DMM
T2 Ch. 2
.
.
.
T40 Ch. 40
Switching Configurations
Figure 6-20 shows an example of testing multiple diodes in a switching
system. This configuration can be used for measuring the forward volt-
age, the reverse breakdown voltage, and the leakage current. This type
of system is often used for production tests in which performing single
point pass/fail DC tests on packaged diodes ensures compliance with
specifications and sorts out defective parts.
In this system, each diode is switched to the Model 2400
SourceMeter instrument through two two-pole switches. The Model
2400 can source and measure both current and voltage.
To connect the SourceMeter instrument to diode 1, close Chs. 1
and 21. To measure the forward voltage, apply the specified current,
then measure the resulting voltage drop. Measure the reverse break-
down voltage by sourcing a specified reverse current bias, then meas-
uring the voltage drop across the diode. To perform the leakage current
test, source a specified reverse voltage, then measure the resulting
leakage current. Repeat this procedure for all the diodes.
- SECTION 6
4-Element
7011 Banks A&B Diode Array 7011 Banks C&D
#1
Ch. 1 Ch. 21
#2
Ch. 2 Ch. 22
#3
Ch. 3 Ch. 23
#4
Ch. 4 Ch. 24
Output Output
HI Source LO Source
Series 2400
SourceMeter
HI Sense LO Sense
Note that two 2-pole relays are used to connect each diode to the
Model 2400. This is done to eliminate error from the voltage drop in the
switch and lead resistance. This is particularly important when meas-
uring the forward voltage, because the measured voltages are relative-
ly small (hundreds of millivolts) and the source current may be rela-
tively high (100mA).
For the example in Figure 6-20, two-pole general-purpose cards,
such as the Model 7011 Quad 1×10 Multiplexer Card can be used. One
Model 7011 card will support testing up to 20 diodes.
For measuring smaller leakage currents (<10nA), use the configu-
ration shown in Figure 6-21. In this example, the Model 7158 Low
Current Scanner Card is used to switch ten diodes with a common
cathode to the Model 6517A Electrometer/Source.
When measuring the leakage current with this setup, the voltage
bias is applied to all the diodes simultaneously through the normally
closed contact of each relay. When a particular channel is energized,
APPLICATIONS -
7158 Low Current Card
R
Ch. 1
R
Ch. 2
.
.
.
R
Ch. 10
HI
Voltage 6517A
Source pA
Electrometer/Source
LO
the leakage current for that diode will be measured. The current limit-
ing resistor (R) is included to prevent test failure in the event a diode is
short-circuited.
- SECTION 6
7111-S
R
Ch. 1
R
Ch. 2
.
.
.
R
Ch. 40
Guard Output HI
A
SMU or
SourceMeter
Output LO
APPLICATIONS -
6.10 Capacitance Measurements
Overview
Capacitance measurements are made to verify that capacitors meet
their manufacturer’s specifications. For quality control purposes, a
group of capacitors may be placed in an environmental chamber and
the capacitance of each capacitor tested. Multiple switches connect the
capacitors to the capacitance meter.
Test frequency and stray capacitance may impose limitations when
switching capacitance. Capacitance measurements are often made at a
high frequency (kHz –MHz ranges), so it’s important to choose switch-
es carefully to avoid errors due to bandwidth limitations. The capaci-
tance introduced to the system by both the switches and the cable
must be corrected for by the LCR meter by using open and short circuit
calibration; therefore, it’s important to determine if a particular LCR
meter is capable of performing the necessary corrections.
7011-S
Bank A Bank B Bank C Bank D
Ch. 1 C1
Ch. 2 C2
.
.
.
Ch. 8 C8
Ch. 9 Open
Circuit
Short
Ch. 10 Circuit
HI HI LO LO
CURR POT POT CURR
LCR Meter
- SECTION 6
Switching Configurations
In Figure 6-23, the Model 7011-S Quad 1×10 Multiplexer Card connects
an LCR meter to a number of unknown capacitors (C1 through C8). For
any given capacitor measurement, four relays, one in each bank, must
be closed. The outputs of the four banks are connected to the LCR
meter’s four input terminals.
The Model 7011-S card allows making four-wire connections to as
many as ten capacitors. However, two sets of channels should be used
to perform open and short calibration, so the system can only test up
to eight capacitors. When performing the open calibration, the HI
CURR and HI POT leads must be connected together using Ch. 9 of
each bank; the LO CURR and LO POT leads must also be connected
together. When performing the short calibration, all four leads are con-
nected together using Ch. 10 of each bank.
The shields of the four coaxial cables connected to the unknown
capacitors are connected to each other. This connection should be
made near the capacitor for best accuracy. For simplicity, the shields
are not shown in Figure 6-23. Space limitations on the Model 7011-S
card make it necessary to use sub-miniature coax cables. The length of
the coaxial cables for all channels should be approximately the same.
With the Model 7011-S card, the test frequency may be as high as
100kHz. The minimum unknown capacitance that can be switched
with this card is on the order of 10pF. There is no maximum capaci-
tance other than determined by the card’s maximum current rating.
The system shown can be expanded easily by adding more Model
7011-S cards. The backplane of the scanner mainframe will connect the
appropriate banks of each card. Remove any bank-to-bank jumpers
from all cards. As more switch cards are added to expand the system,
the maximum allowable test frequency may decrease.
The Model 7011-C version is not recommended for this application
because the added coupling due to this card’s mass-terminated con-
nectors may cause an unacceptable amount of error.
To measure capacitors at 1MHz or higher, an RF switch card, such
as the Model 7062 RF Switch Card, is essential. The Model 7062 consists
of two banks of five RF switches. As shown in Figure 6-24, two Model
7062 cards provide four-terminal connections from a four-terminal
capacitance bridge to test up to four capacitors. One card switches the
HI and LO CURR terminals of the bridge, while the second card switch-
es the HI and LO POT terminals.
The fifth channel of each switch bank is connected to a jumper to
allow short-circuit calibration. When no channels are selected, the
input to the card is connected to the output port, which is effectively a
sixth channel. By connecting coax cables to all four output ports as
APPLICATIONS -
7062 #1 7062 #2
Open
Circuit
Ch. 10 Ch. 25
Short
Circuit
Ch. 5 Ch. 30
Ch. 7 Ch. 22
C2
Ch. 2 Ch. 27
Ch. 6 Ch. 21
C1
Ch. 1 Ch. 26
HI HI LO LO
CURR POT LCR Meter POT CURR
- SECTION 6
7011-S
Bank A Bank B Bank C Bank D
Ch. 1 C1
Ch. 2 C2
.
.
.
Ch. 8 C8
Ch. 9 R3
R1 R2
HI
pA
LO
Ch. 10
HI HI LO LO
CURR POT POT CURR
LCR Meter
NOTE: All bank-to-bank jumpers must be removed from card.
Open-circuit and short-circuit channels are omitted for clarity.
APPLICATIONS -
the leakage has been measured, the capacitor should be discharged by
first opening Ch. 9/Bank A and Ch. 9/Bank C. Then close Ch. 9 on Bank
B and Bank D, which will place resistor R3 across capacitor C1.
The diode and resistor R1, in series with the picoammeter input,
serve two functions. First, the resistor will limit the current in case a
capacitor becomes shorted. Second, the diode minimizes the picoam-
meter noise introduced when measuring capacitors of 1nF or greater.
The diode is photosensitive, so it should be enclosed with R1 in a light-
tight electrostatic shield.
The capacitance of C1 can now be measured by first opening Ch. 9,
Banks B and D, then closing Ch. 10, Banks A through D. This will con-
nect the LCR bridge to capacitor C1. Ensure that all capacitors are dis-
charged before connecting the LCR bridge.
This system is relatively low in cost; however, it has several perfor-
mance limitations:
• The maximum test voltage is 110VDC. It’s capable of measuring
a leakage current of 1nA with fair accuracy, so it can verify capac-
itor leakage resistance of 100V/1nA or 100GΩ.
• The LCR bridge test frequency should be no more than 100kHz.
• The minimum capacitance that can be measured is on the order
of 10pF.
Switching Configurations
Figure 6-26 is an overview of a cell phone lifetime test system based on
Keithley’s System 40 Microwave/RF Switch System. Given that this sys-
tem must be able to switch RF signals, audio signals, and DC signals, a
variety of switch cards are required.
In this example, high frequency switch cards are required to con-
nect the communications analyzer to the handset. Therefore, the RF
- SECTION 6
Mobile Phone
RF Test Fixtures
7016A
GSM Comm.
Analyzer
Audio
7011
Power
DMM Audio 7053
Digital
Control
Power Supplies
7020
System 40
7002 Switch Pass/Fail PASS FAIL
Switch Mainframe Solution Indication
carrier signal is routed with the Model 7016A 50Ω 2GHz Multiplexer
Card.
Isolated relays, such as those used in the Model 7013 Relay Switch
Card, can be used to control certain test system requirements, such as
the conveyor.
APPLICATIONS -
6.12 Power Supply Burn-In Testing
Overview
“Burn-in” testing is a method of evaluating the quality of power sup-
plies by monitoring the output voltage with a load connected for a
specified period of time. This test is performed to ensure the power
supply will continue to function properly over its entire service life. As
the temperature in the environmental/burn-in chamber reaches the
upper and lower limits of the temperature range, the power supply out-
put is cycled on and off repeatedly. Even though power supplies may
have several outputs, monitoring a single output is usually sufficient
for burn-in testing. It is not uncommon to age and monitor thousands
of power supplies at once. When designing a burn-in test system, one
task is to determine the best way to switch all the output voltages to be
monitored.
To monitor the output voltage, a voltmeter is connected in parallel
with the load resistance. The load is chosen to emulate the current
draw the power supply undergoes in its intended application. For
accelerated stress testing, the load resistance may be selected to draw
a level of current higher than the rated current of the power supply.
The power supply output is often cycled and measured repeatedly,
then is left on continuously for an extended period. The duration of the
test cycle can range from less than an hour to many days, depending on
the manufacturer’s requirements. Figure 6-27 illustrates a typical
burn-in test cycle.
Switching Configuration
Typically, burn-in systems use switching to monitor the output voltage
and the environmental chamber temperature, and to control the
incoming AC power. Figure 6-28 illustrates a system for testing multi-
ple power supplies. For the sake of simplicity, this figure shows only
two power supplies under test. In this system, a Model 2750/7708
Multimeter/Switch System is used to monitor the voltage across the
“ON” Voltage
“OFF” Voltage
15s 15s 40 Minute Constant Cycle
“ON” “OFF”
- SECTION 6
2750/7708 Multimeter/Switch System
Line HI
Solid State Power
Power Supply RL Ch. 1 Voltmeter
Input Relay Module
#1 LO
Ribbon
Cable
Power
Supply RL Ch. 2
#2
Digital I/O
Board
Personal
Computer
Thermo-
couple Ch. 40
APPLICATIONS -
APPENDIX A
Glossary
The following terms are defined CHANNEL — One of several signal
to clarify or expand on how they paths on a switching card. For
relate specifically to the material scanner (or multiplex) cards,
in this handbook. the channel is used as a
switched input in measuring
circuits or as a switched out-
ACTUATION TIME — The time
put in sourcing circuits. For
between application of the
switch cards, each channel’s
nominal relay coil voltage and
signal paths are independent
the final closure of the relay
of other channels. For matrix
contacts after the contact
cards, a channel is established
bounce interval.
by the actuation of a relay at a
AUTOMATIC CJC (COLD JUNCTION row and column crosspoint.
COMPENSATION) — A system for Also see PATH.
multi-channel thermocouple
CHANNEL CROSSTALK — Coupling
measurements where the
of a signal from one channel
temperature at each thermo-
to another or to the output by
couple/switch card junction
stray capacitance, inductive
is sensed and compensated
coupling, or radiation.
for automatically.
Crosstalk is expressed in deci-
BANDWIDTH — The range of fre- bels at a specified load
quencies that can be impedance and at a specific
switched, conducted, or frequency.
amplified within certain lim-
CHANNEL ISOLATION — On a
its. Under given load condi-
switching card, the isolation
tions, bandwidth is defined by
from signal HI and LO of one
the –3dB (half-power) points.
channel to signal HI and LO
BANK — A group of relays with a of any other channel (or the
common connection for scan- output on switch or scanner
ning or multiplex applica- cards). Specified as resistance
tions. and capacitance, except for
BREAK-BEFORE-MAKE — RF cards (decibels and fre-
Disconnecting the present quency range). Also see PATH
circuit before connecting a ISOLATION.
new circuit. Also known as COAXIAL CABLE — A two-
BREAK/MAKE. conductor cable that has a
CARRY CURRENT — The maximum center conductor surrounded
continuous current of closed by a shield conductor, the two
relay contacts. Most relays are being coaxial and separated
rated higher for carry current by an insulator.
than switched current. (Heat COIL RESISTANCE — A nominal
is generated by I2R losses for value of the resistance of the
carry current and I2R losses coil winding of a relay.
plus arcing for switched cur-
rent.)
A-2 APPENDIX A
COLD JUNCTION — The junction in CONTACT LIFE — The maximum
a thermocouple circuit that is number of expected closures
at a known temperature. Also before failure. Life is depen-
known as REFERENCE JUNCTION. dent on the switched voltage,
COLD SWITCHING — Closing the current, and power. Failure
relay contacts before applying usually occurs when the con-
voltage and current and tact resistance exceeds the
removing voltage and current end of life value.
before opening the contacts. CONTACT POTENTIAL — A voltage
(Contacts do not make or produced between contact
break current.) Also see DRY terminals due to the tempera-
CIRCUIT SWITCHING. ture gradient across the relay
COLUMN — As viewed on the contacts, and the reed-to-ter-
schematic of a matrix minal junctions of dissimilar
card/module, the vertical sig- metals. (The temperature gra-
nal lines that are connected dient is typically caused by
via relays to the horizontal the power dissipated by the
rows. energized relay coil.)
COMMON MODE — Between two CONTACT RATING — The voltage,
signal lines and a third line current, and power capacities
(e.g., from signal HI and LO to of relay contacts under speci-
chassis ground or guard). fied environmental condi-
tions. See CARRY CURRENT and
COMMON MODE ISOLATION — On a
SWITCHED CURRENT.
switching card/module, the
isolation from signal HI and CONTACT RESISTANCE — For a relay,
LO to guard (or shield) for a 3- the resistance in ohms across
pole circuit, or from signal HI closed contacts. For a Keithley
and LO to chassis ground for switching card/module, also
a 2-pole circuit. Specified as includes the tape resistance
resistance and capacitance. and connector terminal resis-
tance. Also see PATH
COMMON MODE REJECTION RATIO —
RESISTANCE.
The ability of an instrument
to reject interference from a CONTACTS — The surfaces of cur-
common mode voltage at its rent carrying elements where
input terminals with respect electric circuits are opened or
to ground. Usually expressed closed.
in decibels at a frequency. CROSSPOINT — The intersecting
COMMON MODE VOLTAGE — A volt- point of a column and row in
age between input low and a relay matrix. Specified as
chassis ground of an instru- (column, row) or (row, col-
ment. umn). A crosspoint can con-
sist of one or more poles of
CONTACT BOUNCE — The intermit-
Form A (normally open) relay
tent and undesired opening of
switching.
relay contacts during closure.
GLOSSARY A-3
CROSSTALK — See CHANNEL ELECTROCHEMICAL EFFECTS — The
CROSSTALK. property that generates cur-
CURRENT SURGE LIMITING — The rent between circuit board
circuitry necessary to protect conductors due to contami-
relay contacts from excessive nation of the card surface.
transient current. Minimized with proper han-
dling and cleaning.
DIFFERENTIAL INPUT ISOLATION —
On a switching card, the isola- ELECTROMECHANICAL RELAY — A
tion from signal HI to LO. relay that uses an electromag-
Specified as resistance and net to move an armature and
capacitance. make or break the relay
contacts.
DIFFERENTIAL MULTIPLEXER — A
type of switching card/mod- EMI — Abbreviation for
ule where one input is routed Electromagnetic Interference.
to one of several outputs (e.g., A term that defines unwanted
sourcing) or one of several electromagnetic radiation
inputs is routed to one output from a device that could
(e.g., measuring), and each interfere with desired signals
channel uses a 2-pole relay in test or communication
configuration (one pole for equipment, for example. RFI
signal HI and one pole for (Radio Frequency Inter-
signal LO). ference) and EMI are often
used interchangeably.
DIGITAL I/O — A TTL-level
input/output port that is pro- FLOATING — The condition where
grammable by the mainframe. a common mode voltage
exists between earth ground
DRY CIRCUIT SWITCHING —
and the instrument or circuit
Switching below specified
of interest. (Low of circuit is
levels of voltage (e.g., 20mV)
not at earth potential.) The
and current to minimize any
condition where signal LO in
physical and electrical
the system is electrically iso-
changes in the contact junc-
lated from earth ground.
tion. Also see COLD SWITCHING.
FORM A — A single-pole, normal-
DRY REED RELAY — A glass-
ly open contact configuration.
enclosed, hermetically sealed,
Also called SPST-NO. A two-
magnetically actuated con-
pole configuration is called 2
tact. No mercury or other wet-
Form A.
ting material is used.
FORM B — A single-pole, normal-
DUT — An abbreviation for
ly closed contact configura-
Device Under Test.
tion. Also called SPST-NC. A
two-pole configuration is
called 2 Form B.
A-4 APPENDIX A
FORM C —A single-pole, double- ISOLATED SWITCH — A type of card
throw contact configuration. where each channel is electri-
Also known as a transfer cally isolated from all other
switch. A two-pole configura- channels. Also called
tion is called 2 Form C or INDEPENDENT SWITCH CARD/
DPDT. MODULE.
GUARDING — A technique that ISOLATION — On a switching
reduces leakage errors and card/module, the impedance
decreases response time. between any stated terminals.
Consists of a guard conductor Specified as resistance and
driven by a low-impedance capacitance. See INPUT
source surrounding the lead ISOLATION.
of a high-impedance signal. ISOTHERMAL BLOCK — The ther-
The guard voltage is kept at or mally conductive block that
near the potential of the equalizes the temperature of
signal. all thermocouple connec-
INDEPENDENT SWITCH CARD/ tions.
MODULE — A type of card LATCHING RELAY — A relay that
where each channel is electri- maintains its contacts in the
cally isolated from all other last position assumed without
channels. Also called ISOLATED the need to keep the coil
SWITCH CARD/MODULE. energized.
INPUT ISOLATION — On a switching LEAKAGE CURRENT — Error current
card, the isolation from signal that flows through insulation
HI to LO (or guard) for a 2- resistance when a voltage is
pole circuit. Specified as resis- applied.
tance and capacitance.
LOW NOISE CABLE — A cable that is
INSERTION LOSS — The attenuation coated with a conductive
of signals being routed layer between the braid and
through a switching card. inner insulator (coax and
Specified as a decibel value triax) and under the inner
over a frequency range. shield (triax). This reduces
INSULATION RESISTANCE — The triboelectric currents caused
ohmic resistance of insula- by friction due to vibration,
tion. It degrades quickly as movement, or temperature
humidity increases. fluctuation.
I/O — Abbreviation for input/
output, which refers to the
transmission of information
to an external device (output),
and the reception of informa-
tion from an external device
(input).
GLOSSARY A-5
MAINFRAME — A switching instru- MULTIPLEX — Connecting one
ment that operates according instrument to multiple
to user commands to connect devices under test or multiple
signals among sourcing and instruments to one device
measuring instruments and under test. Also see SCAN.
devices under test. Signal MULTIPLEX CARD — See SCANNER
switching is performed on CARD.
cards that are inserted into
NOISE — Any unwanted electron-
the mainframe rack. A main-
ic signal.
frame is also referred to as a
Scanner, Multiplexer, Matrix, NON-LATCHING RELAY — A relay
or Programmable Switch. that maintains its closed con-
tact position when the coil is
MAKE-BEFORE-BREAK —
energized and its open con-
Connecting a new circuit
tact position when the coil is
before disconnecting the pre-
not energized.
sent circuit. Also known as
Make/Break. NORMALLY CLOSED CONTACTS — A
contact pair that is closed
MATRIX — Connecting multiple
when the relay is not ener-
inputs to multiple outputs.
gized.
MATRIX CARD/MODULE — A type of
NORMALLY OPEN CONTACTS — A
card with a switching configu-
contact pair that is open
ration that has columns and
when the relay is not
rows of relay crosspoints.
energized.
With a matrix card, since any
one point can be connected NORMAL MODE — Between two
to any other, you can simulta- signal lines (e.g., from signal
neously have one input with HI to signal LO).
multiple outputs, multiple NORMAL MODE REJECTION RATIO —
inputs with one output, or The ability of an instrument
multiple inputs with multiple to reject an AC interference
outputs. (usually of line frequency)
MERCURY WETTED RELAY — A reed across its input terminals.
relay in which the contacts Usually expressed in decibels
are wetted by a film of mer- at a frequency.
cury. Usually has a required NORMAL MODE VOLTAGE — A volt-
position for proper operation; age applied between the input
some types are position high and input low terminals
insensitive. of an instrument.
MODULE — See SWITCH CARDS/
MODULES.
A-6 APPENDIX A
OFFSET CURRENT — A current that PROPAGATION DELAY — The speci-
comes from a switching card fied amount of time for a sig-
even though no signals are nal to be routed through a
applied. It comes mostly from previously closed channel of
the finite coil-to-contact a switching card. The delay
impedance. It is also generat- must be considered, for
ed by triboelectric, piezoelec- example, when the switched
tric, and electrochemical signal is used to synchronize
effects present on the card. other signals.
PATH — One of many signal paths REED RELAY — A relay using one
on a matrix switching card. A or more glass-enclosed, her-
path is established by the metically sealed, magnetically
actuation of a relay at a row actuated contact members.
and column crosspoint. Also Some types are dry to
see CHANNEL. improve isolation or make
PATH ISOLATION — On a matrix them position independent.
switching card, the isolation Some types are mercury wet-
from signal HI and LO of one ted to improve switched cur-
path to signal HI and LO of rent rating and make contact
any other path. Specified as resistance more predictable.
resistance and capacitance. Also see ELECTROMECHANICAL
Also see CHANNEL ISOLATION. RELAY and SOLID STATE RELAY.
PATH RESISTANCE — On a matrix REFERENCE CHANNEL — On a ther-
switching card, the resistance mocouple scanner card, the
per conductor of a closed channel that measures the
path, including the contact temperature of the isothermal
resistance, wiring resistance, block.
and connector terminal resis- REFERENCE OUTPUT — The output
tance. Also see CONTACT signal representing the tem-
RESISTANCE. perature of the reference
PIEZOELECTRIC CURRENTS — The channel, usually the isother-
current caused by mechanical mal block, on some thermo-
stress to certain insulating couple scanner cards/
materials. To minimize, the modules.
stress is removed from the RELAY — An electrically con-
insulators, and materials with trolled mechanical device that
low piezoelectric effect are opens and closes an electrical
used. contact. A relay provides iso-
POLE — A set of mating contacts lation of control signals from
in a relay or switch: normally switched signals.
open, normally closed, or
both.
GLOSSARY A-7
RELAY DRIVE — The total current SHIELDING — A metal enclosure
available from the power sup- for the circuit being measured
ply of the mainframe to ener- or a metal sleeving surround-
gize the relay coils on switch- ing wire conductors (coax or
ing cards. This current is der- triax cable) to lessen interfer-
ated at elevated temperatures. ence, interaction, or current
RELAY DRIVE CURRENT — The leakage. The shield is usually
amount of current necessary grounded.
to energize a relay. Usually SINGLE-ENDED MULTIPLEXER — A
specified per channel. type of switching card/mod-
RELAY SETUP — A configuration of ule where one input is routed
open and closed relays. to one of several outputs (e.g.,
sourcing) or one of several
RELEASE TIME — The time
inputs is routed to one output
between the removal of the
(e.g., measuring). Each chan-
coil voltage and the stabilized
nel uses a single-pole relay for
opening of the contacts.
signal HI and a single com-
ROW — As viewed on the mon terminal is connected to
schematic of a matrix signal LO for all channels.
card/module, the horizontal
SOLID STATE RELAY — A relay that
signal lines that are connect-
switches electric circuits by
ed via relays to the vertical
use of semiconductor ele-
columns.
ments without moving parts
SCAN — Sequential connecting or conventional contacts.
(usually break-before-make)
SMA — A type of miniature coax-
of one instrument to multiple
ial connector used in situa-
devices under test or multiple
tions requiring shielded cable
instruments to one device
for signal connections.
under test. Also see MULTIPLEX.
SMB — A type of miniature coax-
SCANNER — See MAINFRAME.
ial connector used in situa-
SCANNER CARD — A type of tions requiring shielded cable
switching card where one for signal connections.
input is routed to one of sev-
SWITCH/MEASURE SYSTEM— An
eral outputs (e.g., sourcing),
instrument that integrates a
or one of several inputs is
switch mainframe and a digi-
routed to one output (e.g.,
tal meter into a single chassis.
measuring). The actual
Signal switching is performed
switching can be nonsequen-
on cards/modules that are
tial (multiplex) or sequential
inserted into the chassis.
(scan). Also known as a
Multiplex Card. SWITCHED CURRENT — The maxi-
mum current level that can be
SETTLE TIME — The time required
reliably handled while open-
for establishing relay connec-
ing and closing contacts. Also
tions and stabilizing user
see CARRY CURRENT.
circuits.
A-8 APPENDIX A
SWITCHING CARD/MODULES — The VOLTAGE CLAMPING — The circuitry
general classification of relay necessary to protect relay
cards. May be subdivided contacts from excessive volt-
into: independent switch age caused by switching cur-
cards (1 input to 1 output), rent into inductive loads.
scanner or multiplex cards VSWR — Abbreviation for Voltage
(1:N, N:1) and matrix cards Standing Wave Ratio.
(M:N). Expressed as a ratio of the
T/C COLD JUNCTION — See COLD highest voltage to the lowest
JUNCTION. voltage found along the signal
TEMPERATURE OFFSET — On a path.
thermocouple scanner card,
the maximum temperature For further reading on switching
difference between any two terminology, see the following
channels on the card’s references:
isothermal block.
TEST FIXTURE — An enclosure,
ANSI/EIA RS-473-1981.
sometime shielded, used for
Definitions and Terminology for
mounting single or multiple
Relays for Electronic Equipment.
devices for testing purposes.
American National Standards
THERMAL OFFSET VOLTAGE — See Institute, 1981.
CONTACT POTENTIAL.
ANSI/IEEE Std. 100-1992. IEEE
TRIAXIAL CABLE — A three-conduc- Standard Dictionary of Electrical
tor cable that has a center and Electronics Terms. 5th edi-
conductor surrounded by an tion. Institute of Electrical and
inner shield conductor that is Electronics Engineers, 1992.
in turn surrounded by an
Engineers’ Relay Handbook. 5th
outer shield conductor.
edition. National Association of
TRIBOELECTRIC CURRENT — The Relay Manufacturers, 1996.
current caused by friction
between a conductor and
insulator when a cable flexes
due to vibrations, movement,
or temperature fluctuation.
Also see LOW NOISE CABLE.
TRIGGER — An external stimulus
that initiates one or more
instrument functions. Trigger
stimuli include: the front
panel, an external trigger
pulse, and IEEE-488 bus X,
Talk, and GET triggers.
GLOSSARY A-9
APPENDIX B
B-2
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comments
HIGH DENSITY
7011-C 40 Multiplexer 2 form A 110V 1A 60VA <1µV <100pA 2MHz Connector Yes Four independent 1×10
multiplexers, connec-
tion to backplane
7011-S 40 Multiplexer 2 form A 110V 1A 60VA <500nV <100pA 2MHz Screw term. Yes Four independent 1×10
multiplexers, connec-
tion to backplane
7012-C 4×10 Matrix 2 form A 110V 1A 60VA <1µV <100pA 2MHz Connector Yes Rows connect to analog
backplane
7012-S 4×10 Matrix 2 form A 110V 1A 60VA <500nV <100pA 2MHz Screw term. Yes Rows connect to analog
backplane
7013-C 20 Isolated Switch 2 form A 110V 1A 60VA <1µV <100pA 10MHz Connector Yes
7013-S 20 Isolated Switch 2 form A 110V 1A 60VA <500nV <100pA 10MHz Screw term. Yes
7015-C 40 Multiplexer 2 form A 175V 34mA 0.3VA <5µV <1nA 500kHz Connector Yes Solid state switch for
high reliability
7015-S 40 Multiplexer 2 form A 175V 34mA 0.3VA <5µV <1nA 500kHz Screw term. Yes Solid state switch for
high reliability
7018-C 28 Multiplexer 3 form A 110V 1A 60VA <5µV <100pA 2MHz Connector Yes 3 pole switching
7018-S 28 Multiplexer 3 form A 110V 1A 60VA <5µV <100pA 2MHz Screw term. Yes 3 pole switching
7035 36 Multiplexer 2 form A 60V 1A 30VA <1µV <100pA 10MHz Connector Yes 9 independent 1×4
multiplexers
7036 40 Isolated Switch 1 form A 60V 1A 30VA <4µV <100pA 10MHz Connector Yes 40 independent channels
of 1-pole switching
APPENDIX B
7111-S 40 Multiplexer 1 form C 110V 1A 60VA <500nV <100pA 2MHz Screw term. Yes Four independent 1×10
multiplexers, connection
to backplane
Switching Cards for 7001 and 7002 Mainframes continued
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comments
GENERAL PURPOSE
7052 4×5 Matrix 3 form A 200V 200mA 10VA <5µV <100pA 3MHz Screw term. Yes Switched High, Low,
Guard
7056 10/20* Multiplexer 2 form A 150V 250mA 10VA <100µV 10MHz Screw term. Yes 10 channel 2-pole,
20 channel 1-pole
7066 10 Isolated switch 2 form A 350V 2A 125VA <30µV 3MHz Screw term. Yes
7156-D 10/20* Multiplexer 2 form A 150V 500mA 10VA <100µV 3MHz Connector Yes
7156-M 10/20* Multiplexer 2 form A 150V 750mA 30VA <100µV 3MHz Connector Yes Mercury wetted contacts
7164-D 20* Multiplexer 2 form A 100V 500mA 10VA <50µV 3MHz Connector Yes
7164-M 20* Multiplexer 2 form A 100V 750mA 30VA <75µV 3MHz Connector Yes Mercury wetted contacts
7166 10 Isolated switch 2 form A 300V 750mA 30VA 75µV typ. 3MHz Connector Yes Mercury wetted contacts
LOW VOLTAGE
7059 10 Multiplexer 2 form A 200V 100mA 2VA <1µV 3MHz Screw term. Yes
7064 20* Multiplexer 2 form A 40V 100mA 2VA <1µV 10MHz Screw term. Yes
HIGH VOLTAGE
7054 10 Multiplexer 1 form A 1000V 500mA 10VA 1MHz Screw term. Yes
7154 10 Multiplexer 2 form A 1100V 500mA 10VA <35µV 1MHz Screw term. Yes
7169A 20 Isolated switch 1 form C 500V 500mA 10VA <30µV 2MHz Connector Yes Coupon testing. Use
with 7002.
B-3
Switching Cards for 7001 and 7002 Mainframes continued
B-4
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comments
Control
7019-C Dual Matrix 1 form A 200V 1A 10VA <25µV <100pA 2MHz Connector Yes 6-wire resistance
3×6 measurements
7020 80 Digital I/O Connector Yes 40 inputs/40 outputs
7020-D**
7021 30/20 Mux/ 2 form A 110V 1A 30VA <3µV <100pA 10MHz Connector Yes Dual multiplexers.
Digital I/O Up to 30 channels, 10
digital inputs, 10 digital
outputs.
7022 30/20 Matrix/ 2 form A 110V 1A 30VA <3µV <100pA 10MHz Connector Yes 5 row×6 column matrix,
Digital I/O 10 digital inputs,
10 digital outputs.
7037 30/20 Isolated/ 1 form A 110V 1A 30VA <4µV <100pA 10MHz Connector Yes 30 independent channels
7037-D** Digital I/O of 1-pole switching, 10
digital inputs, 10 digital
outputs
7065 Hall Effect measurement
buffer card
LOW CURRENT
7058 10 Multiplexer 1 form C 200V 100mA <250µV <1pA 3MHz 2-lug triax Yes Can also be used as a
voltage switch
7152 4×5 Matrix 2 form A 200V 500mA 10VA <20µV <1pA 60MHz Connector Yes
APPENDIX B
7153 4×5 Matrix 2 form A 1300V 500mA 10VA <50µV <1pA 60MHz Connector Yes
7158 10 Multiplexer 1 form C 30V 100mA <200µV <1pA 1MHz BNC Yes
Switching Cards for 7001 and 7002 Mainframes continued
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comment
HIGH CURRENT
7053 10 Multiplexer 2 form C 300V 5A 100VA <1mV 1MHz Screw term.
RF
7062 double 1×5 2 isolated 1 pole,
switches 5 throw 24V 50mA 0.5VA <20µV 500MHz BNC Yes 50Ω unterminated
7063 double 1×5 2 isolated 1 pole,
switches 5 throw 5V 50mA 0.5VA <20µV 500MHz BNC Yes 50Ω terminated
7016A double 1×4 2 isolated 1 pole,
switches 4 throw 30V 500mA 10VA <6µV 2GHz SMA Yes Optional 50Ω
termination
7017 double 1×4 2 isolated 1 pole, 30V 1A 10VA <25µV 800MHz SMA Yes 108 closures contact life.
switches 4 throw
7038 12 Three 1×4 1 pole, 24V 10mA 10W @ <15µV 2GHz 75Ω SMB Yes 75Ω characteristic
multiplexers 1 of 4 tree 1.2GHz receptacle impedance
B-5
Switching Cards for 7001 and 7002 Mainframes continued
B-6
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comment
* An even numbered channel and an odd numbered channel cannot be turned on at the same time (only in the 20 channel mode for 10/20 channel cards).
** Cards with a -D suffix feature D-sub connectors.
APPENDIX B
Switching Cards for 707A and 708A Mainframes
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comments
LOW CURRENT
7072 8×12 Matrix 2 form A 200V 1A 10VA <20µV <1pA 15 MHz 3-lug triax Yes Optimized for semiconductor
applications.
7072-HV 8×12 Matrix 2 form A 1300V 1A 10VA <20µV <1pA 4 MHz 3-lug triax Optimized for semiconductor
applications.
7172 8×12 Matrix 2 form A 200V 1A 10VA <30µV <500fA 30 MHz 3-lug triax Yes Optimized for semiconductor
applications.
7174A 8×12 Matrix 2 form A 200V 2A <100fA 30 MHz 3-lug triax Yes Optimized for semiconductor
applications.
GENERAL PURPOSE
7071 8×12 Matrix 3 form A 200V 500mA 10VA <5µV <100pA 3 MHz connector Yes Also provides screw terminal
connection.
7071-4 Dual 4×12 Matrix 3 form A 200V 500mA 10VA <5µV <100pA 3 MHz connector Yes Screw terminals available on
row connections.
7074-D Eight 1×12 Multiplexer 3 form A 200V 500mA 10VA <5µV <100pA 5 MHz connector Yes
B-7
B-8
Switching Cards for 707A and 708A Mainframes continued
Max.
No. of Card Contact Max. Max. Max. Contact Offset Recomm. Connection
Channels Config. Config. Voltage Current Power Potential Current Frequency Type CE Comments
HIGH FREQUENCY
7073 8×12 Matrix 1-pole form A 200V 1A 30VA <2µV <200pA 30MHz BNC Yes
7173-50 4×12 Matrix 2 form C 30V 0.5A 10VA <15µV <200pA 200MHz BNC Yes
PROTOTYPE
7070 96 Yes 7070 includes ribbon cable
to extend
7070-PCA 96 backplane outside mainframe.
APPENDIX B
Switch/Control Modules for Model 2700 and 2750 Multimeter/Switch Systems
Max. Current
# Analog Type of Max. Switched Measurement Digital Switch
Module Inputs Configuration Differential* 4-pole Connector Voltage Current Channels I/O Speed Other
7700 20 Multiplexer 1 × 20 or 1 × 10 Screw 300V 1A 2 channels N/A 3 ms Maximum power
w/CJC two 1 × 10 terminals @ 3A = 125VA.
7701 32 Multiplexer 1 × 32 or 1 × 16 D-sub 150V 1A N/A N/A 3 ms Maximum power
two 1 × 16 = 125VA.
7702 40 Multiplexer 1 × 40 or 1 × 20 Screw 300V 1A 2 channels N/A 3 ms Maximum power
two 1 × 20 terminals @ 3A = 125VA.
7703 32 Multiplexer 1 × 32 or 1 × 16 D-sub 300V 500mA N/A N/A 1 ms Reed relays.
two 1 × 16
7705 40 Independent N/A N/A D-sub 300V 2A N/A N/A 3 ms Maximum power
SPST = 125VA.
7706 20 Multiplexer 1 × 20 or 1 × 10 Screw 300V 1A N/A 16 Digital 3 ms (2) ±12V analog
w/CJC two 1 × 10 terminals Out Only output channels &
100kHz event
counter/totalizer.
Max. power = 125VA.
7707 10 Multiplexer/ 1 × 10 or 1×5 D-sub 300V 1A N/A 32 Digital 3 ms Max. power = 125VA.
Digital I/O two 1 × 5 I/O (4) eight-bit word I/O.
B-9
APPENDIX C
Safety Considerations
Many electrical test systems or instruments are capable of measuring
or sourcing hazardous voltage and power levels. It is also possible,
under single fault conditions (e.g., a programming error or an instru-
ment failure), to output hazardous levels even when the system indi-
cates no hazard is present.
These high voltage and power levels make it essential to protect oper-
ators from any of these hazards at all times.
Protection methods include:
• Design test fixtures to prevent operator contact with any hazardous
circuit.
• Make sure the device under test is fully enclosed to protect the
operator from any flying debris.
• Double insulate all electrical connections that an operator could
touch. Double insulation ensures the operator is still protected, even
if one insulation layer fails.
• Use high-reliability, fail-safe interlock switches to disconnect power
sources when a test fixture cover is opened.
• Where possible, use automated handlers so operators do not require
access to the inside of the test fixture or have a need to open guards.
• Provide proper training to all users of the system so they understand
all potential hazards and know how to protect themselves from
injury.
It is the responsibility of the test system designers, integrators, and
installers to make sure operator and maintenance personnel protec-
tion is in place and effective.
The following safety precautions should be observed before using any
Keithley product and any associated instrumentation. Although some
instruments and accessories would normally be used with non-
hazardous voltages, there are situations where hazardous conditions
may be present. Keithley products are intended for use by qualified
personnel who recognize shock hazards and are familiar with the safe-
ty precautions required to avoid possible injury. Read the operating
information provided in each product's manual carefully before using
any Keithley product.
C-2 APPENDIX C
Maintenance personnel perform routine procedures on the product to
keep it operating, for example, setting the line voltage or replacing con-
sumable materials. Maintenance procedures are described in the man-
ual. The procedures explicitly state if the operator may perform them.
Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe
installations and repairs of products. Only properly trained service per-
sonnel may perform installation and service procedures.
Exercise extreme caution when a shock hazard is present. Lethal volt-
age may be present on cable connector jacks or test fixtures. The
American National Standards Institute (ANSI) states that a shock haz-
ard exists when voltage levels greater than 30V RMS, 42.4V peak, or
60VDC are present. A good safety practice is to expect that hazardous
voltage is present in any unknown circuit before measuring.
Users of these products must be protected from electric shock at all
times. The responsible body must ensure that users are prevented
access and/or insulated from every connection point. In some cases,
connections must be exposed to potential human contact. Product
users in these circumstances must be trained to protect themselves
from the risk of electric shock. If the circuit is capable of operating at or
above 1000 volts, no conductive part of the circuit may be exposed.
As described in the International Electrotechnical Commission (IEC)
Standard IEC 664, these instruments are Installation Category I, and
signal lines must not be directly connected to AC mains.
For rack mounted equipment in which the power cord is not accessi-
ble, in the event of fire or other catastrophic failure, the user must pro-
vide a separate power disconnect switch.
Do not connect switching cards directly to unlimited power circuits.
They are intended to be used with impedance limited sources. NEVER
connect switching cards directly to AC mains. When connecting
sources to switching cards, install protective devices to limit fault cur-
rent and voltage to the card.
Before operating an instrument, make sure the line cord is connected
to a properly grounded power receptacle. Inspect the connecting
cables, test leads, and jumpers for possible wear, cracks, or breaks
before each use.
For maximum safety, do not touch the product, test cables, or any other
instruments while power is applied to the circuit under test. ALWAYS
remove power from the entire test system and discharge any capacitors
before: connecting or disconnecting cables or jumpers, installing or
removing switching cards, or making internal changes, such as
installing or removing jumpers.
Do not touch any object that could provide a current path to the com-
mon side of the circuit under test or power line (earth) ground. Always
C-4 APPENDIX C
To clean an instrument, use a damp cloth or mild, water based cleaner.
Clean the exterior of the instrument only. Do not apply cleaner direct-
ly to the instrument or allow liquids to enter or spill on the instrument.
Products that consist of a circuit board with no case or chassis (e.g.,
data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board
becomes contaminated and operation is affected, the board should be
returned to the factory for proper cleaning/servicing.
I-2 INDEX
Ground loops, 4-5, 4-7, 5-4 Microwave Monolithic Integrated
Circuits (MMICs), 4-33
H Multiplex switch, 1-5
Hardware verification and trou- Multi-position switch, 4-34
bleshooting, 5-4 to 5-6
Hot switching, 2-5, 3-6 O
Offset compensation, 2-8, 4-6,
I 4-23 to 4-25
Ice point, 6-20 Offset current, 1-3, 1-7, 2-3, 2-9,
Ideal switch, 1-2, 2-2 to 2-3 3-3, 3-4, 4-8, 4-16, 4-21, 4-29,
Impedance matching, 4-35, 4-36 5-6
In-rush current, 4-31 to 4-32 Offset voltage, 2-3, 2-7, 2-8, 4-2,
4-5, 4-6, 4-8, 4-23, 4-24, 4-27
Insertion loss, 2-10 to 2-11, 4-34,
4-35 P
Insulation resistance, 1-5 to 1-6, Path isolation, 2-4, 4-34
4-9, 6-4
Phase distortion, 4-35
testing, 6-6, 6-7 to 6-10
Phase-matched cables, 4-36
testing combined with conti-
nuity testing, 6-10 to 6-11 Platinum RTDs, 6-21 to 6-22
testing of printed circuit Plug-in data acquisition boards,
boards, 6-11 to 6-15 1-15
Integrated scanners, 1-15 Pole, 1-3
Isolated switch, 1-10, 4-11 Power transmission, 4-35
Isolation, 2-3 to 2-4 Printed circuit boards, 6-6, 6-10,
6-11 to 6-15
L
R
LCR meter, 6-28 to 6-31
Radio Frequency Integrated
Leakage resistance, 1-3, 2-3, 2-9, Circuits (RFICs), 4-32 to 4-33
4-17, 5-5, 6-3 to 6-4, 6-11
Radio Frequency Interference
M (RFI), 2-5, 3-6
Magnetic interference, 4-7 Reactive loads, 4-7 to 4-8, 4-15,
4-31 to 4-32
Mainframe specifications, 2-12 to
2-13 RC time constant, 3-4 to 3-5, 6-6
Mass-terminated cables and Reflection coefficient, 2-11
connectors, 5-2, 6-29 Relay actuation time, 3-4
Matrix switch, 1-5 to 1-10, 1-14, Relays, 1-2 to 1-15, 2-2 to 2-12
4-19 to 4-22, 4-34 latching, 2-8, 4-6
Maximum signal levels, 2-4 to mercury-filled or -wetted, 2-6
2-5
Relay welding, 4-31
Microwave switching, 4-32 to
4-36 Remote sensing, 4-3 to 4-4
INDEX I-3
Removable jumpers, 2-13, 4-13, Switching speed, 3-4 to 3-6
5-2 to 5-3, 6-29 System uncertainties, calculat-
Resistance switching, 4-23 to ing, 3-3 to 3-4
4-31
low, 4-23 to 4-27 T
high, 4-27 to 4-31 Temperature scanning, 6-18 to
6-23
Response time, 3-4, 4-9 to 4-10,
4-21 Termination, 4-35
I-4 INDEX
A GREATER MEASURE OF CONFIDENCE
Switching Handbook
Switching
Handbook
A Guide to Signal Switching in Automated Test Systems
No. 1716
4
th
Edition
4th
www.keithley.com