Tissue-Tek Prisma Plus User Manual
Tissue-Tek Prisma Plus User Manual
Manufactured for:
Sakura Finetek U.S.A., Inc. Torrance, CA 90501, USA
Sakura Finetek Europe B.V., 2408 AV Alphen aan den Rijn, NL
Sakura Finetek Japan Co., Ltd. Tokyo, Japan 103-0023
Manufactured by:
Sakura Seiki Co., Ltd. Nagano, Japan
(a)
(b)
(c)
Chapter 4 Troubleshooting
Troubleshooting ...................................................................................................................................... 4-1
Error Codes Displayed during System Operation ............................................................................... 4-1
Instrument Conditions and Recovery Actions ..................................................................................... 4-7
(d)
* Before using the instrument, read "Safety Precautions" to ensure the correct use of the instrument.
* The cautionary instructions provided herein are intended to ensure that the instrument will be used
safely in a manner preventing operator from injury and property damage.
* In this manual, instructions pertaining to different levels of potential hazards are classified as and
indicated under , , and .
pertains to a potential hazard where failure to observe the specified instruction may result in death or
serious injury of the operator or other person.
pertains to a potential hazard where failure to observe the specified instruction may result in a serious
injury of the operator or other person.
pertains to a potential hazard where failure to observe the specified instruction may result in damage to
this instrument or other property, or affect the process results.
* The symbols used on the labels attached to the instrument are explained.
The labels bearing one of the following symbols provide particularly important information you must
know in order to ensure safety of the operator, improve work efficiency and protect the instrument from
damage. Be sure to check these labels and understand the specified instructions before commencing
your work.
A label bearing this symbol specifies an action that must be taken. Always follow the instructions.
A label bearing this symbol specifies a prohibited item. Always follow the instructions.
A label bearing this symbol specifies a precaution. Wrong handling against the instruction may
potentially expose the operator to danger or result in damage to the instrument. Always follow
the instructions.
A label bearing this symbol is provided near a high-temperature area. Exercise caution against
burns caused by high temperatures.
(e)
● Those who have experience operating a Sakura Finetek product or products other than the
instrument must also read this manual. Also, those who will be operating the instrument
should receive training first.
Sakura Finetek will assume no responsibility for any accident that may result from an accident
caused by operating the instrument without reading this manual first. If the instrument is operated in
a manner not specified in this document, the protective functions of the equipment may be lost.
● Safety mechanisms installed in the instrument only represent the minimum measures taken
to prevent accidents. The responsibility for safe operation of the instrument lies primarily
with the owner of the instrument, and persons in charge who operate the instrument.
● Should the instrument malfunction, turn off the power and disconnect the power plug from
the power outlet. If processing is still in progress, immediately protect the tissues.
Should malfunction occur, turn off the power and contact the Sakura Finetek Technical Support
representative or local representative. Disconnecting the power plug also cuts off the main power
supply. When installing the instrument, select a place where the power outlet is not concealed and
the power plug can be disconnected any time in case of emergency.
(f)
WARNING
● A wrong operation may stop or interrupt the tissue processing.
Be careful not to let unauthorized personnel manipulate or tamper with the instrument while
operating.
A wrong operation may stop or interrupt the tissue processing. (If the power is turned off by
unauthorized personnel, the stain process may be interrupted and the staining problems may occur.
Arrange the work environment so that only the operators with expert knowledge can use the
instrument. (Also consider using external alarms, etc., just in case.)
● Do not block the intake/exhaust ports and air inlet around the instrument.
Do not block the intake/exhaust ports (air inlets) with a wall or object or use the instrument in a poorly
ventilated place or with the dustproof covers still attached, because heat will be trapped in the
instrument and cause the capacity of the instrument to drop, while fire, accident, failure, etc., may also
occur due to abnormal overheating.
● Periodically check the power cord and cables and power outlet for damage and accumulation
of dust.
Foreign particle and dust accumulating on the power outlet may cause fire. Clean the power outlet
periodically.
● Do not let tissue with flammable solvent attached go inside the drying station.
Deposited flammable solvent may be heated by the drying station and vaporize, causing fire or
explosion.
● Install the instrument indoors. Do not place the instrument in a location exposed to direct
sunlight, snow or rain.
This product is specified for indoor use, so do not use it outdoors. Ultraviolet rays and high
temperature may damage the instrument. Never use the instrument while exposing it to snow or rain
because it may cause electric shock or short-circuit.
(g)
WARNING
● Unplug the power cord from the power outlet and store it while the instrument is not in use.
For safety, if the instrument will not be used for an extended period of time, clean each part
and store the instrument in a place free from humidity and from dust and foreign matter.
Using the instrument in a humid place, place exposed to oil smoke or steam, or dusty place, may
cause its parts to deteriorate, short-circuit, fire, etc. The installation humidity range is 30 to 85% (non-
condensing). If you wish to install the instrument in a special place, consult the Sakura Finetek
Technical Support representative or local representative.
● Do not operate the power plug and switches with a wet hand. When connecting/ removing the
power plug, do so by holding the plug.
Failure to observe these instructions may result in electric shock .
● Be sure to use the supplied power cord. Also, be sure to use a correct power cord that is
suitable for the current capacity of region where the instrument is used.
If an inappropriate power cord or voltage is used, instrument malfunction due to short-circuit, fire or
voltage drop, unexpected accidents or injuries due to electric shock, etc., can occur. The preset
voltage cannot be changed by the user.
● For the power cord, a standard cord specified by local law may have to be used.
If a power cord not specified by the PSE standard, UL/CSA standard or EC standard is required,
contact the Sakura Finetek Technical Support representative or local representative.
● Do not forcibly bend, pull, twist or bundle the power cord and cables or otherwise damage the
power cord and cables.
Fire or electric shock may occur.
● This instrument has been designed and tested according to CISPR11 Class A. When the
instrument is used in a home, it may cause electromagnetic interference, in which case
measures to reduce interference may have to be taken.
● Radiant waves may prevent this instrument from operating properly, so do not use the
instrument near a radiation source of strong electromagnetic waves.
● Use the instrument under the conditions (such as voltage within 10% of the rated voltage)
stated in the Specification.
(h)
CAUTION
● Do not pinch the drain hose or allow it to slacken.
Water will not be drained properly and overflow from the instrument.
● Before staining is started, confirm that the end station has been set.
If the end station is not set, the basket cannot be put down and will remain gripped. If the next basket
is processed in this condition, the baskets may collide with each other and the tissues may be
damaged.
● Be careful not to pinch your hand when opening/closing the cover or door.
Be careful not to pinch your hand between the fixed part and moving part of the cover or door or
between the cover and door.
● Do not move the instrument while operating. Move the instrument after removing all solution
bottles, paraffin containers and other components carrying liquid inside.
Moving the instrument while operating may cause the solution or paraffin to leak. Accident or failure
may also occur.
● If the straining process is suspended, the process may be negatively affected depending on
the situation. Remember to resume operation.
Prolonged suspension may negatively affect the stain process. Keep the suspension time to the
minimum.
● When the expiration date of the current solution is passed, a message appears to urge
exchanging the solution. Exchange the solution properly in order to maintain the staining
quality.
● The safety data sheet of each solution used can be obtained from the supplier/manufacturer of
the solution. You can also download it from the website of respective supplier/manufacturer of
the solution.
● Do not insert a metal bar or other metal object into the hot air outlet of any drying station or
use a metal product to reset an over temperature protector.
Metal contact may cause electric shock.
(i)
● Confirm that no paper or foreign object is present at the cover or outlet of any drying station.
It may result in insufficient drying.
● Do not touch the drying stations and heating stations while the instrument is operating.
These stations become very hot during operation, so touching them may result in burns.
● Securely connect the water supply hose and drain hose to make sure water will not leak.
If water leaks, the facility may be damaged.
● Before using the instrument, inspect the water supply hose and drain hose to ensure that they
are free from damage and connected properly.
It may cause an unexpected problem.
● Clean the water supply strainer regularly. Also, close the supply source water tap before
cleaning.
If the water supply strainer is not cleaned regularly, the water supply volume will drop and washing
problems may occur. Check the flow rate after cleaning. If the water supply strainer is cleaned with
the supply source tap open, water will overflow when the strainer is removed.
[Corrosive Solutions]
Acids [hydrochloric acid (HCl), hydrobromic acid (HBr), hydrofluoric acid (HF), etc.], staining solutions
and reagent solutions containing hydrochloric acid (Schiff’s reagent, cold Schiff, sulfurous acid
solution, resorcin‐fuchsin staining solution, etc.), oxidizing chlorides [ferric chloride (FeCl3), mercuric
chloride (HgCl2), cupric chloride (CuCl2), sodium hypochloride (NaOCl), etc.], tioculfates [sodium
tiosulfate (Na2S2O3), etc.], organic acids [formic acid (HCOOH), acetic acid (CH3COOH), bromic
acid (C204H2), lactic acid (CH3CH(OH)COOH), etc.]
(j)
● Do not stare at the laser beam from the 2D bar code reader.
For details, refer to the operating manual. The operating manual can be downloaded from the website
of the 2D bar code reader manufacturer.
● Observe the rated water pressure stated in the Specification of the instrument.
If the water pressure is high, water may leak and cause electric shock or the instrument may stop.
● Do not use any CF card other than the one that comes with the instrument. Also, run a virus
check before inserting the CF card.
● Secure the instrument with seismic anchorage (optional) to prevent it from moving/falling in
case of earthquake.
Provide fall prevention measure to prevent the instrument from moving should a disaster occur.
● Do not put solution, etc., on the instrument. Also, do not spill solution on the painted surfaces,
caution/warning labels or touch screen.
If solution is spilled over the instrument, the instrument may fail or sustain damage.
● Do not put any obstacle within the moving range of the arm.
Failure may occur.
(k)
● Replace the activated carbon filters in the fume control unit periodically.
Replace the activated carbon filters regularly in order to maintain a better environment around the
instrument. Expired activated carbon filters cannot maintain exhaust gas at an appropriate level of
concentration. When replacing the activated carbon filters, put the lids on the solution stations or
remove the solution stations by considering the possible scattering of activated carbon particles, to
prevent solution contamination.
* If the instrument is connected to an external exhaust device, the activated carbon filters need not be
replaced.
● Use dedicated activated carbon filters for the fume control unit.
Use of inappropriate activated carbon filters may cause explosion of powder dust generated from
activated carbon powder.
● If the instrument must be moved after installation, contact the Sakura Finetek Technical
Support representative or local representative.
Unexpected accident may occur by relocating the instrument, so contact the Sakura Finetek Technical
Support representative or local representative.
● Close the lids of solution reservoirs whenever the instrument is not operating.
If the solution reservoir lids remain open for a prolonged period of time, organic gas may be released
to the room.
(l)
● Do not use the instrument in a place subject to too low or high temperatures.
Use it in a place where the temperature is stable in a range of 10C to 40C.
● Be sure to connect the power plug to a power outlet with grounding bar.
Connect the power plug alone to a power outlet with grounding bar conforming to at least the Class D
grounding specification. If the instrument is not grounded properly, fire or electric shock may result
due to leak current.
● Lightning and power outage may cause the instrument to fail and stain process may not be
able to continue and adversely affect staining. If the power supply is unstable, use an
uninterruptible power supply.
Lightning may cause the instrument to fail and damage the tissues. If the instrument will be used in a
region where lightning damage is a possibility, use an uninterruptible power supply. If lightning has
caused an irreparable damage to the instrument, take out the basket from the instrument and take an
appropriate measure.
● Do not perform stain process if an extended period of power outage is scheduled. Staining
quality may drop. Staining problems may occur depending on when the power outage occurs.
In this case, use a uninterruptible power supply. If you find that the stain process has stopped due to
an unexpected power outage lasting for an extended period of time, remove the baskets from the
instrument and take appropriate action.
● Install the power cord and cables carefully by making sure the cord and cables will not
interfere with the work.
Failure to do so may result in the operator tripping over the power cord and cables and sustaining
injury.
● Securely connect the power cord and power outlet. If the power cord came off or the power
switch was turned off inadvertently, connect the power cord or turn on the power switch
without delay.
If they are loosely connected, heat may be generated at the connected part, voltage drop can occur
and the instrument may not operate normally. Furthermore, unexpected accidents or injuries can
occur. Also, the tissues may be damaged when the power is cut off during processing.
● Handle the power cord with care. Also, be aware of damage caused by mice and other small
animals.
To unplug the power cord from the power outlet, do not pull the cord, but hold the plug and then pull it
out. Inspect the power cord periodically. If any damage is found in the exterior of the power cord, stop
using it and contact the Sakura Finetek Technical Support representative or local representative.
(m)
● Check the staining program before the operation. If a wrong staining program is used, staining
problems may occur.
A wrong operation may interrupt the stain process or edit the program to a wrong one. If staining is
started with a wrong program, staining problems may occur. Read this manual carefully and use the
instrument correctly.
● Operating the instrument with an inappropriate temperature setting or time setting (long/short)
may cause staining problems. Conduct a verification test beforehand to confirm that staining
can be performed properly. If you have any question regarding staining programs, contact the
Sakura Finetek Technical Support representative or local representative.
● The allowable concentration limit of each substance in the work environment varies
depending on the air temperature, size of the room, additional load, if any, ventilation efficient
y, etc.
If danger is suspected, the Instrument Control Manager should measure the concentration of the
applicable substance in the area and confirm that the allowable limit in the work environment is not
exceeded. Follow the standard concentration limits specified by the environmental regulation of the
country/region in which the instrument is used.
● Solution may scatter and attach to your body while working. Be sure to wear protective gears
while working.
In all operations including stain process, solution exchange, opening/closing of doors, cleaning and
daily maintenance, the solutions used may scatter and attach to your body (eye, mucous membrane,
skin, mouth, nose, etc.) or you may inhale their gases. Be sure to wear protective gears (gloves,
organic solvent mask, goggles) while working.
● If any solution harmful to human health attached to your body or entered your mouth, take an
appropriate action according to the SDS of the solution.
Wear protective gears (mask, gloves, goggles) during work in case anything goes wrong. Use
protective gears that are resistant to all solutions on the solution list.
● You may accidentally touch an infectious source and operate the instrument with the
contaminated hands. Be sure to wear protective gears when operating the instrument.
If you have operated the instrument after handling a pathogen, disinfect the instrument properly. The
disinfection method varies depending on the type of pathogen, so choose a method appropriate for
the pathogen. Also, disinfect/sterilize or otherwise properly treat the protective gears used and then
dispose of them as infectious waste. For the disposal method, follow the applicable regulation and
guideline specified by each institution, country or region.
(n)
● Sterilize any potentially infectious waste material or infectious medical waste and then
dispose of it according to the applicable regulation or guideline in the country or region.
● When the instrument is used to handle any substance presenting biological hazard, ensure
safety according to the applicable regulation or guideline in the country or region. When
performing operations that involve any such substance, also wear gloves, mask, safety
goggles and/or other protective gears and perform the operations inside a safety cabinet by
taking appropriate measures to protect yourself as the operator.
● Handling of solutions requires expert knowledge on associated hazards due to their toxicity
and volatility, etc.
Appoint a person responsible for solutions and use solutions under the guidance/supervision of the
responsible person.
(o)
● Enhanced Wash
To repeat the process of immersing a basket into a solution and then removing it completely out of the
solution inside a station.
● Slide
A slide made of glass on which tissues are pasted.
● Basket
A special staining basket in which slides are set. Two types of baskets, 10-slide type and 20-slide type,
are available.
● Batch mode
An operation mode in which the next staining run cannot be started unless the current staining run is
completed.
● Continuous mode
An operation mode in which staining runs of different tissues can be started during the current staining
run.
● Solution station
A general term referring to both solution reservoirs and wash reservoirs.
● Start station
A solution station in which baskets are set at the start of staining.
● End station
A solution station in which stained baskets are stored.
(p)
This is a precision instrument and must be handled with care. If the instrument is handled in a rough
manner or dropped, its internal components may malfunction or get damaged. Always exercise due
care when handling the instrument.
Flow of Installation
Checking the Installation Location and Transporting the Instrument Details: P. 1-2
• Checking the installation conditions
• Checking the work area
• Transporting the instrument
Required Tools
Installation Environment
As is the case with all electronic devices, the instrument should not be installed in places subject to
excessive temperature or humidity.
Temperature and humidity should be held at constant levels to achieve the highest degree of operating
stability. The ambient operating temperature of the instrument is in a range of +10 to +40C. Its ambient
operating humidity is in a range of 30 to 85% in relative humidity (non-condensing).
Also, locate the instrument, with at least 10 cm away from walls, in a well-ventilated place not exposed to
corrosive vapor or extremely high/low temperatures or where the instrument will not come in direct contact
with air flow. Install the instrument away from direct sunlight and do not install it proximity to open
windows, ovens, open flames, hot plates, radiators, and dry ice baths. The instrument should be installed
away from any equipment carrying high voltage or current, such as a large refrigerator or oven. Since the
instrument is very heavy, the experiment bench should have sufficient bearing strength, and the
instrument should be laid flat.
The instrument should be installed near a power supply meeting the power facility requirements (voltage,
current) specified in this Operating Manual. Be sure to ground the power outlet and wire the power
properly using a dedicated cable.
If the instrument is used in a non-ventilated place, serious health risks may occur to the human body.
1-1
The instrument can support the power supplies listed below using the selector switches.
Switching of the voltage setting should only be performed by qualified service staff working
exclusively with the instrument.
Facility requirements
• Single-phase 100 VAC±10% 50/60 Hz: 15 A or more Power outlet with grounding bar
• Single-phase 115 VAC±10% 50/60 Hz: 15 A or more Power outlet with grounding bar
• Single-phase 230 VAC±10% 50/60 Hz: 7 A or more Power outlet with grounding bar
If the above facility requirements are not satisfied (such as when the power supply used is 200 VAC),
the instrument must not be used.
If the instrument generates abnormality, the breaker on the facility side may actuate and other
equipment may be affected; accordingly, connect each instrument to a dedicated power outlet
(including a dedicated breaker on the facility side).
If the instrument is used in a region where the applicable standard above is not met, provide a different
power cord conforming to the standard in the location of use.
• About grounding
Confirm with the person responsible for the facility that the power outlet to be used is Class-D grounded.
Confirm that a ventilation system, local exhaust system or exhaust port leading to the outside
of the building is available on the facility side.
Although the volumes of volatile solution gases released from the instrument vary depending on the
solutions and processing programs used and the environmental temperature, volumes of gases
released per unit time are provided below for reference. Based on these values and the volumetric size
of the installation location, check if the ventilation rate is sufficient to meet the requirements at the
installation location pertaining to the concentrations of substances in the work environment. If no
ventilation system is available, use a 38 or 75 exhaust duct adapter (optional) to connect the duct
hose to the exhaust port of the instrument, and discharge the exhaust fume to the outside of the
building.
Maximum volumes of gases released per unit time (reference)
Ethanol: 3.9×10-3 (m3/hr)
Xylene: 4.8×10-4 (m3/hr)
1-2
Transportation
When transporting the instrument on a pallet truck, etc., one person should not work alone in order to
prevent instrument damage or accident caused by rear-end collision, dropping of the instrument, etc.
When transporting the instrument, place thin iron sheets, etc., to eliminate height gaps on the floor so
that the instrument will not receive impact.
Pay due attention to safety. Particularly when transporting the instrument on a slope, remember that,
because the instrument is heavy, stopping the instrument will not be easy once it began accelerating.
When using a pallet truck, etc., for transport, confirm that the instrument and packing pallet are
secured together with L brackets, because otherwise the instrument may shift and drop from the
packing pallet. When the instrument must be pushed while being transported on a pallet truck, etc., pay
attention to where on the instrument to be pushed, because the exterior panels may be deformed or the
acrylic cover may be damaged. (Do this by holding the transport handles at the base of the system.)
Pay due attention to safety, and particularly when transporting the instrument on a slope, remember
that, because the instrument is heavy, stopping the instrument will not be easy once it began
accelerating. When lifting the instrument to move, be sure to secure a line of movement or footing.
1-3
Confirm before unpacking the instrument, that the packing materials have not been damaged due to
external forces applied during transport. Two or more persons should always work together to unpack
the instrument. Be careful not to damage/scratch the instrument when unpacking. When taking the
accessory box out of the instrument, assume a natural posture.
Sleeve
Ester band
2. Remove the cushion materials put around the instrument, and take out the sleeves and polyethylene
bags.
3. Remove the cushion materials inserted into the front cover from both sides, lift the front cover, and take
out the cushion materials sandwiched between the front cover and the front door.
1-4
Tape
Cushion material
Accessory box
PP band
6. Remove the cushion materials inserted underneath the display arm, and the cushion foams wrapping
the instrument.
1-5
1. Remove the cushion materials sandwiched between the robot arm and the rail.
Cushion material
2. Remove the tape securing the holder, and remove the cushion materials inserted underneath the
holder.
Tape Cushion material
3. Remove the reinforcing cushion materials on the tray, and take out the cushion foam sheet and PP
band.
Cushion material Cushion material Cushion material PP band Cushion foam sheet
4. Remove all 3-position and 4-position reservoir trays stored in the system, and remove the tape
securing the cushion foam sheet and the water supply nozzle.
1-6
1-7
1-8
● Attach the water supply hose connect it to a water supply facility (tap water supply) meeting the
following conditions:
Pressure: Dynamic pressure = 0.098 to 0.441 MPa
Maximum static pressure = 0.74 MPa
Temperature: 30C max. (non-freezing)
Faucet: A faucet of general shape having a tip of 12 mm to 17 mm in size. (The
water supply hose cannot be connected to a faucet of any other size or to a
special shaped faucet such as chemical faucet or aerator faucet.)
Facility requirements: Use a facility meeting the following items in order to prevent health damage
caused by flow-back:
[1] Water supply facility not directly connected to public water pipes
[2] Water supply facility not directly connected to a water supply system
used for drinking water, etc.
[3] Water supply facility designed to prevent flow-back caused by reverse-
siphoning
• Water supply facility equipped with a flow-back protection mechanism
meeting the law of each country/region
1. Connect a nipple o the water supply port on the instrument via a bushing, and then screw the threaded
part of the water supply hose into the bushing.
2. Pull the flange on the other end of the water supply hose toward the hose and separate the water tap
joint.
3. Loosen the four screws on the water tap joint.
4. Tighten the four screws uniformly while pressing the black rubber in the water tap joint against the tip
of the water faucet.
5. Turn and tighten the nipple.
6. Insert the water supply hose into the water tap joint with the flange pulled back.
7. Gently pull the water supply hose by hand to confirm that it does not come off from the water tap joint.
]
[1]
[2]
How to separate the water tap joint Water tap joint Nipple Flange on the water
supply hose
1-9
Drain hose
Drain hose
Hose clamp
Drain elbow
4. Connect the other end of the drain hose to a sink having a minimum drain capacity of 15 L/min.
(Connect the hose securely so that it will not separate from the sink.)
5. Ensure that the drain port on the instrument is positioned higher than the sink, that the drain hose has
no slack in the middle, and that the outlet of the drain hose is away from the water level of the sink by
at least 100 mm.
Do not extend the drain hose using a different hose. The drain elbow can be oriented to either right or
left.
The drain hose contains vinyl chloride, so be careful not to spill organic solvents onto the hose.
After connecting the water supply hose and drain hose, confirm by referring to P. 1-42 that water is not
leaking.
How to Connect
1. Attach the 38 or 75 exhaust duct adapter to the exhaust port located at the back of the system on
the left side.
2. Insert a duct hose into the adapter and tighten it with a wire clamp, etc.
3. Bring the open end of the duct hose to near a ventilation system or local exhaust system or to the
outside of the building.
The duct hose should not be longer than 10 m, and confirm that the hose is not bent or crushed in the
middle. If the ambient temperature is low, drain water may collect inside the duct hose; accordingly,
make sure drain water will not collect in any part of the duct hose and also install a drain pan at the tip
of the duct hose or take other appropriate measure.
1-10
1. Enough maintenance space of 200 mm or more is available on the right side face of the instrument.
(1) Screw the connection nipple and bushing into the attached strainer.
Bushing
Nipple
Strainer Strainer type
(2) Screw the attached nipple, and the strainer set assembled in (1), into the water supply port on the
right side face of the instrument.
Nipple
Strainer type
(3) Screw in the water supply hose, and connect the water supply hose to the tap water supply
according to P. 1-9.
1-11
(2) Screw the attached nipple and street elbow into the water supply port on the right side face of the
instrument.
Nipple
Street elbow
Strainer type
(4) Screw in the water supply hose, and connect the water supply hose to the tap water supply
according to P. 1-9.
1-12
1. Remove the 4-position reservoir tray at the installed position of the wash reservoir.
2. Remove the cap or wash station nozzle currently set.
The cap can be removed by turning it counterclockwise. To remove the nozzle, turn the wash cap
counterclockwise to remove it, and then remove the wash station nozzle by pulling it straight up.
3. Set the cap if you want to seal the water supply part. At this time, confirm that a waterproof rubber
gasket is placed inside the hole, and then turn the cap clockwise to secure it.
4. To install the wash station nozzle, first align the small hole in the waterproof rubber gasket with the
small hole in the water supply outlet. Next, put the small projection on the bottom face of the wash
station nozzle into the small hole in the water supply outlet. Lastly, put the wash cap on and turn it
clockwise to secure the cap.
Wash port
Rubber
gasket
Cap
Wash
station
nozzle
Wash cap
1-13
[1]
Lock
[1]
[2]
Arm holder
Arm grooves
1-14
Near side
Near side
Near side
1-15
• Type
A 3-position or 4-position reservoir tray contains standard solution reservoirs and small solution
reservoirs. A special stain reservoir tray contains only special stain solution reservoirs.
• Direction
Orient each solution reservoir so that the "solution name tab" on top of the reservoir faces the front side
of the instrument, and set the reservoir by aligning the "grooves" on both sides of the reservoir with the
grooves in the tray. Pay attention to the widths of the grooves provided on both sides of the solution
reservoir. When the solution reservoir is set in the tray, the wider grooves should come to the left, while
the narrower grooves should come to the right.
Back
Left tray grooves Right tray grooves
Wider than those * Narrower than
on the right those on the left
Near side
3-position 4-position Special stain
reservoir tray reservoir tray reservoir tray
Left grooves
Wider than those
on the right Right grooves
* Narrower than
those on the left
Use a power cord appropriate for the power-supply voltage. Be sure to connect the power cord to a
power outlet with grounding bar.
1. Confirm that input voltage has been set according to the installation environment.
2. Connect the power cord to the power-supply inlet on the right side face of the instrument at the bottom.
3. Confirm that the power switch at the bottom right on the front face of the instrument is turned off, and
then connect the plug on the power cord to the power outlet on the facility side.
4. Turn on the power switch to start the instrument.
1-16
1-17
1-18
1-19
[10]
[6]
[7] [11]
[8] [9]
[1] Cover
The cover protects the operator against exposure to dangerous chemicals without
obstructing his/her view.
[2] Door
Open this door when setting or removing tissues.
[3] Water supply strainer door
Access this area when maintaining the water supply circuit.
[4] Power switch
[5] Control panel display
A touch-panel type display for operating the instrument.
[6] External interface
It can be connected to an UPS (uninterruptible power supply), 2D bar code reader,
etc.
[7] Power-supply inlet
[8] Water supply port
Tap water is supplied to this port.
[9] Drain port
[10] Exhaust outlet
This outlet can be connected to an external exhaust duct via a special attachment.
[11] Automated coverslipper link port
Remove the port cover to connect an automated coverslipper for automated operation
encompassing staining through coverslipping.
1-20
1-21
[2] [7]
[6] [8] [9]
[3]
[4]
[10]
[11]
[12]
[19]
[23] [20]
[24]
[25]
[26] [22]
[27]
[29]
[28] [30]
[31]
[33]
[32]
[42]
[34]
[35]
[36] [38]
[39]
[102]
[101]
[37]
[40]
[104]
[41] [103]
1-22
1-23
Instrument Overview
● This instrument automatically stains tissues and cell smear samples on glass slides for microscopy, as
part of histological studies and tests conducted in the fields of pathology, anatomy, clinical pathology,
etc.
The instrument supports various types of staining such as H&E, Papanicolaou etc.
● This instrument is operated from the touch-panel type display located on the right side. This display
allows the operator to easily operate the instrument according to visual cues by touching the necessary
keys displayed on the screen. Staining data can also be output from a PC by using a special compact
flash card (CF card).
● Two types of baskets can be used for storing slides: the 20-slide type for general staining and the 10-
slide type for special staining
● Solution reservoirs are largely classified into three types: standard solution reservoir, small solution
reservoir and special stain solution reservoir. The basket capacity and required solution volume of each
solution reservoir are summarized below.
Standard solution
Small solution reservoir Special stain reservoir
reservoir
20-slide basket 3 baskets 1 basket -
10-slide basket 3 baskets 1 basket 1 basket
Required solution volume 680 ~ 820 mL 255 ~ 285 mL 160 ~ 180 mL
• Continuous mode
A different staining run can be started during the current staining run, with up to 11 staining runs
performed simultaneously.
• Batch mode
In this mode, each stain process must be completed before the next sample can be stained. The start
step and end step can be specified to start a staining run.
1-24
1-25
• Set a CF card into the CF card slot from the side with an arrow, with the other side on which a label is
attached facing toward you.
• Be sure to wait for the confirmation message to appear on the screen before removing the CF card.
• Sakura does not warrant the integrity of data stored in CF cards regardless of the failure exhibited by the
instrument or monetary damage suffered in connection with the data loss/damage. Sakura does not
provide any service associated with the recovery or restoration of data in CF cards.
• CF cards other than those specified for the instrument may not be supported. CF cards must be FTP-
formatted using Microsoft Windows.
CF card slot
Arrow Label
CF card
1-26
Some UPSs are not supported by the instrument. For details, contact the Sakura Finetek Technical
Support representative or local representative. The instrument provides an UPS input signal to detect a
power outage mode of the UPS connected to the power supply. Use an UPS equipped with a contact-
output interface or open-collector output that remains open in normal state and shorts upon power
outage. Always turn off the instrument power before connecting an UPS.
* When connecting the power-outage detection signal to an external circuit, loop the signal cable
around the attached ferrite core at a point as close as possible to the instrument, as shown below. If
the ferrite core is not installed, the instrument may malfunction due to external noise.
To the instrument’s power-outage detection signal input To the UPS’s power-outage detection signal output
SEIWA
1-27
● EMC
This instrument conforms to the emission and immunity requirements specified in EN61326-2-6 or
JISC1806-2-6.
Specification: 10BASE-T
* Check the local Sakura Finetek website for availability of the Sakura staining solutions and staining
kits.
1-28
Fixed station numbers are assigned to the start stations (S1 to S3), end stations (linked) (E1-1 to E1-5),
end stations (not linked) (E2-1 to E2-5), wash stations (W1 to W4), drying stations (D1A, D1B, D2A, D2B),
heating stations (H1, H2) and link station (LINK). The tables below specify the identification number of
each station attribute and the corresponding station number.
Start station S1 S2 S3
Without special solution stations
Station 52 Station 51 Station 50
With special solution stations
* S2 and S3 can also be set as solution stations.
Special solution station SP1 SP2 SP3 SP4 SP5 SP6 SP7 SP8
Without special solution stations − - - - - - - -
With special solution stations Station 21 Station 22 Station 23 Station 24 Station 25 Station 26 Station 27 Station 28
1-29
Fixed station numbers are assigned to the start stations (S1 to S3), end stations (linked), (E1-1 to E1-5),
end stations (not linked) (E2-1 to E2-5), wash stations (W1 to W4), drying stations (D1, D2) and link
station (LINK). The tables below specify the identification number of each station attribute and the
corresponding station number. The rightmost end station is selected as the PE station.
With special solution stations Station 34 Station 33 Station 32 With special solution stations Station 29 Station 30 Station 31 Station 32 Station 33
* S2 and S3 can also be set as solution stations. * E1-2 to E1-5 can also be set as solution stations.
With special solution stations Station 35 With special solution stations Station 7 Station 6 Station 5 Station 4
* The link station can be set only when the instrument is * W1 to W4 can also be set as solution stations.
connected to an automated coverslipper.
Drying station D1 D2
Without special solution stations Station 8 Station 16
Special solution station SP1 SP2 SP3 SP4 SP5 SP6 SP7 SP8
Without special solution stations - - - - - - - -
With special solution stations Station 12 Station 13 Station 14 Station 15 Station 16 Station 17 Station 18 Station 19
1-30
Fixed station numbers are assigned to the start stations (S1 to S3), end stations (linked) (E1-1 to E1-5),
end stations (not linked) (E2-1 to E2-5), wash stations (W1 to W4), drying stations (D1, D2), heating
stations (H1, H2) and link station (LINK). The tables below specify the identification number of each station
attribute and the corresponding station number. The rightmost end station is selected as the PE station.
Start station S1 S2 S3 End station (linked) E1-1 E1-2 E1-3 E1-4 E1-5
Without special solution stations Station 31 Station 30 Station 29 Without special solution stations Station 26 Station 27 Station 28 Station 29 Station 30
With special solution stations Station 35 Station 34 Station 33 With special solution stations Station 30 Station 31 Station 32 Station 33 Station 34
* S2 and S3 can also be set as solution stations. * E1-2 to E1-5 can also be set as solution stations.
With special solution stations Station 36 With special solution stations Station 7 Station 6 Station 5 Station 4
* The link station can be set only when the instrument is * W1 to W4 can also be set as solution stations.
connected to an automated coverslipper.
With special solution stations Station 8 Station 20 With special solution stations Station 28 Station 29
Special solution station SP1 SP2 SP3 SP4 SP5 SP6 SP7 SP8
Without special solution stations - - - - - - - -
With special solution stations Station 12 Station 13 Station 14 Station 15 Station 16 Station 17 Station 18 Station 19
1-31
[4] [5]
This menu screen appears while the instrument is standing by after logon.
[4] [2]
1-32
When the Closing the door is selected, be careful not to let stain process start inadvertently.
• Start Confirmation
Set whether or not to show a confirmation message at the start of staining.
• Process Start Notice
Set whether or not to notify the start of stain process by showing the start screen once staining
process has started.
• Heating Station Always Heated
When heating stations are used, set whether or not to always heat the heating stations.
• Solution Name Display
Set whether or not to display the applicable solution name (abbreviation) on each station in the
monitor screen showing a solution configuration.
• Link with Coverslipper
Set whet her to connect or disconnect the communication link with the connected coverslipper when
the link is set in the System Setup screen (P. 1-36). Even if the instrument is linked to a coverslipper,
slides are not transferred to the coverslipper if "Not in Use" is shown here. This item is displayed only
when the instrument is linked to an automated coverslipper.
* Check the local Sakura Finetek website for availability of the Sakura staining solutions and staining
kits.
1-33
1-34
1-35
[9]
[1]
[2]
[10]
[3]
[4] [11]
[5]
[6] [12]
[7]
[8]
1-36
[8] Coverslipper
If the Coverslipper link is installed, Coverslipper can be selected.
1-37
Entering Year
Touching the Modify button on the right of "Year" displays the numeric keypad. Enter the current year
using the numeric keypad. You can enter a year in the range of 2000 to 2099.
Entering Month
Every time the ▲ or ▼ key is touched on the right of "Month," the selection changes to the next month
(▲) or previous month (▼).
Entering Day
Select a desired day by touching the corresponding number in the calendar.
am/pm
You can select am or pm only when the 12 hour format is currently selected.
After all items have been entered, touching the Save key to apply the settings.
1-38
Setting an ID
You can change the existing ID of a user. Touching the User’s ID key displays the simple keyboard.
Enter a new user ID, and then touch the Save key to save the ID. A user ID can be entered using up to 20
characters. Touch the Caps Lock key to enter characters in uppercase. Touch the key again to enter
characters in lowercase.
Setting a Password
You can change a password only when you have logged on as an Administrator. Select the user whose
password you want to set, touch the Password Setup key, and then enter a new password consisting of
one to six digits on the numeric keypad. The entered password is shown in asterisks (*) in the password
field. After the password has been entered, touch the Enter key. When the screen prompts you to enter
the password again for confirmation, enter the same password and touch the Enter key again to save the
password.
Up to four "Administrator" passwords and 20 "User" passwords can be saved. The factory settings are
shown below.
1-39
Stain Process
Staining Mode Selection
Touching this function changes permissions for
"Staining Mode Selection" between Allowed and
Not Allowed.
Abort Stain Process
Touching this function changes permissions for
"Abort Stain Process" between Allowed and Not
Allowed.
Reset Solution Usage Information
Touching this function changes permissions for
"Reset Solution Usage Information" between
Allowed and Not Allowed.
Solutions Configuration Selection
Touching this function changes permissions for
"Solutions Configuration Selection" between
Allowed and Not Allowed.
Utility
System Setup
Touching this function changes permissions for
"System Setup" between Allowed and Not
Allowed.
Export Data
Touching this function changes permissions for
"Export Data" between Allowed and Not Allowed.
Language Selection
Touching this function changes permissions for
"Language Selection" between Allowed to Not
Allowed.
Water Flow
Touching this function changes permission for
"Water Flow" between Allowed to Not Allowed.
Edit
Edit Program
Touching this function changes permissions for
"Edit Program" between Allowed to Not Allowed.
Program Check
Touching this function changes permissions for
"Program Check" between Allowed to Not
Allowed.
Edit Solution Configuration and Solution
Name
Touching this function changes permissions for
"Edit Solution Configuration and Solution Name"
between Allowed to Not Allowed.
Delete Data
Touching this function changes permissions for
"Delete Data" between Allowed to Not Allowed.
1-40
When all items have been set, touch the Exit key to return to the System Setup screen.
1-41
• Water supply run-up time --- The time at which water supply to the wash stations is started during stain
process, prior to the transfer of tissues into the wash stations.
Change the setting whenever necessary before the start of staining, after changing the solution
configuration or staining program, and so on. The water supply run-up time must be changed after the
wash solution configuration has been changed. The water flow can be adjusted with the cover open.
1. Confirm that water has been drained from all wash stations.
2. Open the water supply handle.
3. Touch the Start key and wait for water to overflow from the wash stations.
4. Once water starts to overflow from the wash stations, touch the Stop key. Adjust the water supply
handle so that water will overflow within 20 to 30 seconds.
5. The water supply run-up time you have just set is displayed and the save screen appears. Touch Save
if you want to use this run-up time. Touching Exit returns to the Water Flow screen.
1-42
An example of data export is shown on P. 2-40. The heading in the parenthesis is the title when
exported..
1-43
1-44
2-1
Be sure to remove all lids. If any lid remains closed, tissues may be damaged when the baskets are
transferred to the applicable station.
Touch the Start button under "Confirm Water Supply," and you can perform a water supply test. Touch
the Stop button to stop the water supply.
2. When the Stain Process Monitor screen appears, touch the Log On key at the bottom left. When the
Enter Password sub-window appears, enter a pre-assigned password using the numeric keypad
(number keys 0 to 9) shown on the display. If you have made a mistake, touching the Backspace key
deletes the last character entered. Alternatively, you can touch the Cancel key to cancel the password
entry and return to the Stain Process Monitor screen. Refer to P. 1-39 for passwords.
3. After the password has been entered, touch the Enter key to complete the logon.
Logging off
To log off from the instrument, touch the Log Off key at the bottom left of the screen. You can log off
from the instrument even after staining has started. Once the user logs off, the instrument only accepts
the operation inputs to check the staining status, which prevents inadvertent operations by different
users. To perform other operations, the user must log on to the instrument again.
2-2
1. Touch the Menu key shown at the bottom right of the Stain Process Monitor screen.
2. Select Stain Process Menu from the Stain Process Menu, Utility Menu or Edit Menu, and then touch
the Solution Configuration Selection key.
3. Select and touch a desired solution configuration name under Solution Configuration List, and then
touch the Select key. For creating a new station configuration or editing an existing station
configuration, refer to "Setting a Solution Configuration" (P. 2-15).
"Selecting a Solution Configuration" If any of the following conditions applies to a solution station for
which a solution management option has been set, the solution management information will not be
transferred.
• The station position where solution management is performed changes when the solution
configuration changes.
• The solution at the station position where solution management is performed changes to a different
solution when the solution configuration changes.
• The management method for the station where solution management is performed changes when
the solution configuration changes.
If the solution used, station position (station number) and management method do not change after
the solution configuration is changed, the solution management information will be transferred.
Touch the Check Program key to check the staining programs that can be operated with the solution
configuration you have selected.
To check a given solution configuration, select the applicable solution configuration and touch the
View key to display the Solution Configuration View screen showing the applicable station
arrangement. You can also check the reservoir arrangement of the next configuration by touching the
Change View key at the bottom right. If the current solution configuration is appropriate, touch the
Select key. To check other solution configuration, touch the Exit key and select a desired solution
configuration again.
2-3
To perform other operation, you can return to the Stain Process Menu screen by touching the Cancel key.
1. Prepare slides to be stained. Check the solution configuration and prepare baskets (20-slide or 10-
slide baskets) together with appropriate basket adapters.
2. Attach a basket adapter to each basket by aligning the "UP SIDE" mark on top of the adapter with the
"UP SIDE" mark on one side of the basket. After confirming that both "UP SIDE" marks are facing the
same direction, tilt the basket adapter along the groove.
3. Next, store the slide into the basket one by one. Set the slides in the correct orientation so that their
tissue side aligns with the "UP SIDE" marks. When all slides have been set, return the tilted basket
adapter to the original position.
Do not use a deteriorated, worn, damaged or deformed basket/basket adapter, as it may cause
problems. Replace any such basket/basket adapter with a new one. Store the slides evenly in the
basket.
If the slides are stored unevenly, basket movements are affected during the staining operation.
Particularly when there are fewer slides, be careful not to let the basket tilt when storing the slides.
2-4
Use the 1-position arm holder and If the Start button is touched without entering a
attach a 10-slide basket value, a warning window appears. If the
load/unload adapter to the start program does not cover any station for which
station. Stain process cannot be "Management by slides" has been set, the field
started if a 10-slide basket in which to enter the number of slides is not
load/unload adapter is not set. displayed. If "Management by slides" is
selected, processing can be started at only one
station at a time.
2-5
When setting a small solution reservoir or special stain solution reservoir, attach an applicable
adapter to the start station. If not, stain process cannot be performed correctly. Also confirm that the
arm holder has only one tab. If a wrong arm holder is set, stain process cannot be performed
correctly.
6. Touching the Start key at the bottom right of the screen displays a confirmation message. Select Start
to start staining.
7. Once staining starts, the Stain Process Monitor screen appears where you can check the current
processing. Refer to P. 2-7 for the operations permitted during stain process.
The starting procedure explained above is only one example. If the instrument has been set to start
staining upon closing of the door, you need not touch the Start key because closing the door
automatically starts stain process. Refer to P. 1-33 for the settings relating to the starting procedure.
2-6
Adding a Basket
This section explains the procedure to add a basket during stain process. This procedure is basically the
same as the normal staining start operation, but a basket can be added only in the continuous mode. Also
note that no basket can be added in the following conditions:
• The instrument is performing stain process in the batch mode.
• 11 staining runs are being processed.
• The start station is full.
• An error (including a station full error) is present.
2-7
If you are using a 20-slide basket in a small solution station, attach a 20-slide basket load/unload
adapter to the start station. If not, staining process may not be performed correctly. If you are using a
special stain solution station, attach a 10-slide basket load/unload adapter to the start station. If not,
staining process will not be started.
6. Touching the Start key at the bottom right of the screen displays a confirmation message. Select Start
to start staining. If the instrument has been set to start stain process upon closing of the door, you
need not touch the Start key because closing the door automatically starts stain process. Refer to P. 2-
19 for the applicable settings.
7. Once staining starts, the Stain Process Monitor screen appears.
1. Confirm that the target basket is set in the start station. Then, touch the Start key and select a desired
staining program from Program List in the Stain Process Monitor screen, just like when you are adding
a basket.
2. Touching the Priority key activates the priority start function and displays the Stain Process Monitor
screen.
2-8
Overview Monitor
Detailed Monitor
The staining run numbers and abbreviated staining program names shown in the "No." field are also
shown in the Overview Monitor, so the current step can be visually identified at a glance. Also, you can
touch the button corresponding to a desired staining run number (No.) in Overview Monitor to check
the details of the applicable staining program. Touch the Exit key to return to the Overview Monitor.
2-9
● Error display
Up to 99 errors that have occurred since the start of staining process are displayed, from the latest
error shown on top.
You can touch the keys on the side of the screen to scroll the list to show the hidden columns.
Touching Error Description displays a screen explaining error numbers and corresponding error
descriptions. You can also touch Clear to clear the saved error data.
2-10
● Pause operation
To pause staining process, touch the Pause key on the left of the Stain Process Monitor screen.
Touching the Pause key displays the pausing window. The arm will continue to move until a position
where it can stop safety, after which all staining runs will stop.
To resume stain touch the Resume key. Note that the pause operation may make it no longer possible
to complete the process by the scheduled end time. In this case, the scheduled end time must be
extended at the time of resumption by up to the period during which stain process has been paused.
● Abort operation
To cancel all or part of the current staining runs, touch the Pause key in the Stain Process Monitor
screen and then touch the Abort key. When the abort window appears, touch the All Runs key to
cancel all staining runs. To cancel certain staining runs, touch the applicable staining run numbers to
add a check mark to the "Selected" field of each staining run.
When the Abort key is touched, a confirmation screen appears and asks if you want to abort the
selected staining runs. Selecting Yes displays the abort completion window. Open the cover and
remove the aborted baskets. Once the baskets have been removed and the cover closed, the Exit key
is displayed. Touch this key to complete the abort operation.
2-11
1. Confirm that the arm is not on a solution station in the frontmost row, and then open the door and
remove the basket.
2. Once the basket has been removed and the door closed, the basket removal confirmation window
appears. Touch the Yes key in this window. If you have not yet removed the basket, touch the No key.
After the door is opened, remove all stained baskets from the end station and PE station.
End of Operation
This section explains the procedure to complete the stain process and end the operation.
1. After all operations have completed, turn off the power switch located at the bottom right of the
instrument, and then stop the water supply.
2. Close the lids on all solution reservoirs, and then close the cover.
2-12
How to Start Stain Process by Specifying the Start and End Steps
You can start stain process by specifying the start and end steps only in the batch mode. If a desired
step is specified as the start step, the staining program will start from the specified step. If a desired step
is specified as the end step, staining process will end at the specified step and the baskets will be
transferred to the end station specified in the applicable staining program.
The start and end steps only represent temporary changes in the start and end stations. The settings
will become ineffective once the series of staining runs are completed or aborted. The start and end
steps cannot be set on the stations shown in the table below.
If the start step is set on a drying station, the drying station heater is turned on at the start of staining.
If the start step is set on a heating station, the baskets are held in the start station until the heating
station reaches the specified temperature.
Stations that cannot be set as the start step Stations that cannot be set as the end step
End stations, PE station Start stations, End stations, PE station
2-13
2. Touch the Start key and select a desired staining program under Program List.
3. Next, set a basket and select a start station under Available Start stations. Refer to P. 2-4 for details on
how to set a basket and select a start station. Once the basket has been set and start station selected,
touch the View Program key at the bottom right.
4. The View Program screen appears and the details of the currently selected staining program are
displayed. Touch the Start/End Step key at the bottom right.
5. The Select Start/End Steps screen appears.
6. To set a start step, select a step number you want to specify as the start step, and then touch the Start
Step key to the right. When the key is touched, "S" appears in front of the step number to indicate that
it has been set as the start step.
7. To set an end step, select a step number you want to specify as the end step, and then touch the End
Step key to the right. When the key is touched, "E" appears in front of the step number to indicate that
it has been set as the end step.
8. Touching the Save & Start key starts stain process. Touching the Cancel key cancels the start/end
step settings and returns to the View Program screen.
2-14
2-15
Set no more than six stations including the start station, PE station, linked end stations (E1) and
unlinked end stations (E2).
2-16
● Heating Stations
Determine whether or not to use heating stations. If Not In Use (No Heating) is selected, the heating
stations do not perform heating control and are treated as normal special stain solution stations.
Touching this function toggles the selection in the sequence of In Use, Not In Use and Not In Use (No
Heating).
Touching the Save key displays the next setup screen, or Edit Solution Configuration screen, where
you can set each solution station.
2-17
1. In the Edit Solution Configuration screen, touch the station you want to set a solution for. Select a
desired solution name from "Standard Solutions," "Sakura Solutions * " and "Kit Solutions * " in the
solution setting field.
* Check the local Sakura Finetek website for availability of the Sakura staining solutions and staining
kits.
Selecting "Standard Solutions" displays solution names of some standard solutions and solution
names registered by user. Refer to P.2-23 for how to create or edit solution names.
Selecting "Sakura solutions" displays solution names of Sakura staining solutions managed by bar
code.
Selecting "Kit solutions" displays the solutions that constitute the Sakura staining kits managed by
bar code.
2. After selecting the solution name, set the solution management method.
Touching the Management Method toggles in the sequence of "Runs: Management by runs
done/Days: Management by days/Slides: Management by slides/NONE: Not in Use."
3. Next, determine a specific usage limit. Touching the Set button in the entry field displays the virtual
numeric keypad, so enter a setting value using the numeric keypad.
Runs: Manages the number of runs. You can enter a value between 0 and 99.
Days: Manages the number of days. You can enter a value between 0 and 99.
Slides: Manages the number of slides. You can enter a value between 1 and 9999.
NONE: Not in Use. The entry field is grayed out.
4. Touching the Save key saves the settings and returns to the Edit Solution Configuration screen.
2-18
2-19
2-20
Temperature Settings
• Drying Station Temperature
You can set the temperature of drying stations.
Set a value using the numeric keypad displayed by
touching the Modify key. The adjustable range is "30
to 65" (C). The default is "65" (C). Note that the
drying stations may not be controlled at low
temperatures depending on the outside air
temperature.
2-21
1. Touch Menu at the bottom right of the Stain Process Monitor screen.
2. Select Edit Solution Configuration in the Edit Menu.
3. Under Solution Configuration List in the Solution Configuration Selection screen, touch and select the
name of the solution configuration you want to edit.
4. Next, touching the Edit key displays the Create Solution Configuration Name screen showing the
simple keyboard. To change the solution configuration name currently saved, enter a new name from
the onscreen keyboard (using of up to 20 characters). After the new solution configuration name has
been entered, touch the Save key to save the name.
5. Touching the Save key displays the Solution Configuration Setup screen explained on P. 2-16.
If a solution configuration affecting any saved staining program has been edited, the changes cannot
be saved.
1. Touch Menu at the bottom right of the Stain Process Monitor screen.
2. Select Edit Solution Configuration in the Edit Menu.
3. Under Solution Configuration List in the Solution Configuration Selection screen, touch and select the
name of the solution configuration you want to copy and then touch the Copy key. The Copy Solution
Configuration screen appears.
4. Next, select the copying destination. To create a new solution configuration to copy to, touch the Add
Configuration key. To copy to an existing solution configuration, select the desired solution
configuration from the Copy to window and then touch the Copy key.
5. Touching the Add Configuration or Copy key displays the Create Solution Configuration Name
screen showing the simple keyboard. To change the solution configuration name, enter a new name
from the onscreen keyboard (using up to 20 characters). After the new solution configuration name has
been entered, touch the Save key to save the name.
6. Touching the Save key displays the Solution Configuration Setup screen explained on P. 2-16.
2-22
2-23
2-24
2-25
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
• Copy
Touching this button copies the line corresponding to the selected "Step" field. The data is not yet
pasted.
• Paste
Touching this button pastes the data in the copied line to the line corresponding to the "Step" field
currently selected. This operation inserts the source line immediately above the "Step" line currently
selected by the cursor. Basically, only the "Time," "Delay" and "Mix" data are copied.
• Replace All Steps
Touching this button replaces the "Time," "Delay" and "Mix" data of all lines excluding the start
station, end station and PE station lines, with the corresponding data in the copied line. The solution
station names are not replaced.
• Insert Step
Touching this button inserts a new blank "Step" field directly below the line corresponding to the
selected "Step" field.
• Delete Step
Touching this button deletes the line corresponding to the selected "Step" field.
Touching the Save key saves the applicable changes and closes this screen.
2-26
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
• Select Station
You can select the station to be assigned to the selected "Station" field.
Touching the Select Station key displays the solution configuration layout. When a desired station is
touched, the data of that station is displayed at the bottom of the screen to let you check the attribute
and solution name for the station. Touching the Modify key reflects the selection in the "Station" field
in the Edit Program screen, with the selected station number shown in the "Station" field and
applicable solution name shown in the "Solution Name" field. If a non-solution station such as start
station, end station, wash station or drying station is selected, the text displayed in the "Station" field
conforms to the table below.
Touching the Save key saves the applicable changes and closes this screen.
Text displayed in
Text displayed in
Station attribute "Station" field
"Solution Name"
field
Start stations S* (Solution name)
End stations (linked) E1- * (Solution name)
End stations (not
E2- * (Solution name)
linked)
Wash stations W* Wash Station
Drying stations D* Drying Station
PE station PE (Solution name)
2-27
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
• Modify
Touch the Modify key to determine a processing time. Touching this key displays the numeric
keypad where you can set a desired time in the "Hours: Minutes: Seconds" format. The input range
of each segment is shown below:
Hours: 0 to 99
Minutes: 0 to 59
Seconds: 0 to 59
Touching Clear clears the selected segment to "0." Touching the Enter key applies the entered
processing time. You can enter "0:00:00" in some fields. But if "0:00:00" is set in all fields, you cannot
save.
Touching the Save key saves the applicable changes and closes this screen.
2-28
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
• Exact
No delay is allowed and each staining run is performed exactly according to the specified time.
• Limited (20%)
A delay of up to 20% of the specified time is allowed.
• Limited (50%)
A delay of up to 50% of the specified time is allowed.
• Unlimited
An unlimited delay is allowed.
Touching the Save key saves the applicable changes and closes this screen.
2-29
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
• Mix
Touching the Mix key displays the screen where you can set mixing.
• Enhanced Wash
Touching the Enhanced Wash key displays the screen where you can set enhanced wash.
• OFF
Touching the OFF key disables both mixing and enhanced wash.
Touching the Save key saves the applicable changes and closes this screen.
2-30
Downloaded from www.Manualslib.com manuals search engine
Setting a Staining Program
Editing a Staining Program
This section explains the procedure to edit an existing staining program.
1. Touch Menu at the bottom right of the Stain Process Monitor screen.
2. Select Edit Program from the Edit Menu.
3. Under Program List in the Select Program screen, touch and select the name of the staining program
you want to edit. The solution configuration used for the selected staining program is displayed on the
right. To change this solution configuration, touch the Change Solution Configuration key at the
bottom and select a new solution configuration under Solution Configuration List. To use the displayed
solution configuration, proceed to the next step.
4. Next, touch the Edit key. The Create Solution Configuration Name screen with the simple keyboard
appears. To change the staining program name currently saved, enter a new name from the onscreen
keyboard (using up to 20 characters). Once the staining program name has been entered, touch the
Save key. When the screen prompts you to enter an abbreviated staining program name, enter an
abbreviation from the onscreen keyboard and then touch the Save key to save the abbreviation. (Enter
the abbreviation using up to five characters.)
5. Touching the Save key proceeds to the operation explained under "Creating a Staining Program" on P.
2-26.
2-31
1. Touch Menu at the bottom right of the Stain Process Monitor screen.
2. Select Edit Program from the Edit Menu.
3. Under Program List in the Select Program screen, touch and select the name of the staining program
you want to copy, and then touch the Copy key. The Copy Program screen appears.
4. Next, select the copying destination. To create a new solution configuration to copy to, touch the Add
Configuration key. To copy to an existing program, select the desired program from the Copy to
window and then touch the Copy key.
5. Touching the Add Program or Copy key displays the Create Solution Configuration Name screen
showing the simple keyboard. To change the staining program name, enter a new name from the
onscreen keyboard (using up to 20 characters). Once the staining program name has been entered,
touch the Save key. When the screen prompts you to enter an abbreviated staining program name,
enter an abbreviation from the onscreen keyboard and then touch the Save key to save the
abbreviation. (Enter the abbreviation using up to five characters.)
6. Touching the Save key proceeds to the operation explained under "Creating a Staining Program" on P.
2-26.
2-32
1. Touch Menu at the bottom right of the Stain Process Monitor screen.
2. Select Program Check from the Edit Menu.
3. When the Program Check screen appears, select the solution configuration you want to check under
Solution Configuration List. Next, select the staining program you want to check under Program List.
4 After both the program and solution configuration have been selected, touch the Check key at the
bottom right. The "Program Check Result" screen appears.
5. This screen shows problems with the selected combination, if any. If necessary, change the staining
program and/or solution configuration by referring to the displayed check result to correct the
problems.
2-33
Applicable Solutions
- Sakura staining solutions*
- Staining kits*
* Check the local Sakura Finetek website for availability of the Sakura staining solutions and staining kits.
[1]
[2]
3. Select the station where the special solution will be used, and select Sakura Staining Solutions [3].
When the Select Solution window [4] appears, select the special solution to be used. When the
solution is selected, the station is recognized as a Sakura special solution station.
(The details of each category are shown in the table below.)
Category Description
Standard solutions Solutions other than Sakura staining solutions (including user-defined solutions)
Sakura solutions Sakura staining solutions managed by bar code
* "*" is attached at the beginning of the solution name.
Solutions that constitute the Sakura staining kits managed by bar code
Kit solutions
* "/" is attached at the beginning of the solution name.
[4]
[3]
2-34
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
How to Enter Sakura Solution Bar Code Data
1. Prepare the solution to be set to the instrument.
2. Touch the Read Barcode button on the Stain Process Menu.
3. Move to the Read Barcode screen, and confirm the station for which the special solution is set.
Only the stations for which bar code can be entered are displayed.
The bar code reading procedure and solution state vary depending on the station/text color, so use the
table below as reference.
Station color
Sakura staining
Yellow
solution station
Kit staining solution
Green
station
7. Once the solution has been set, touch the Enter button.
8. Touch the Save button and then touch the Yes button in
the figure below, and the entered bar code information
will be saved.
2-35
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
How to Use Sakura Staining Kit
[1]
[2]
Edit Menu Screen Solution Configuration Setup Screen
3. For the specified station, select Kit Solutions [4] from the categories shown in [3].
(The details of each category are shown in the table below.)
Category Description
Standard solutions Solutions other than Sakura staining solutions (including user-defined solutions)
Sakura staining solutions managed by bar code
Sakura solutions
* "*" is attached at the beginning of the solution name.
Solutions that constitute the Sakura staining kits managed by bar code
Kit solutions
* "/" is attached at the beginning of the solution name.
Touching the Kit Components button [4] displays the Select Solution Kit Components window [5],
so select the solution you want to assign to the station. When the solution is selected, the station is
recognized as a kit component solution station.
[5]
[3]
[4]
2-36
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
How to Enter Kit Component Solution Bar Code Data
1. Prepare the solution to be set to the instrument.
2. Touch the Read Barcode button on the Stain Process Menu.
3. Move to the Read Barcode screen, and confirm the station for which the special solution is set.
Only the stations for which bar code can be entered are displayed.
The bar code reading procedure and solution state vary depending on the station/text color, so use the
table below as reference.
Station color
Sakura staining
Yellow
solution station
Kit staining solution
Green
station
4. Select the station for which the kit is set, and read the bar
code of the kit.
5. Once the kit has been read, the information of the kit is
displayed on the screen.
If the information displayed on the screen is correct,
touch the Enter button.
2-37
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
7. Once the bar code of the solution has been read, the
information of the solution is displayed on the screen, as
with the kit.
10. When the Yes button is touched, the station to which the
bar code information has been set turns blue. In the case
of a kit, read all solutions constituting the kit and then
touch the Register button in the bottom right of the
screen, to complete the solution saving process.
2-38
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
Resetting the Solution Usage Information
You can select the Reset Solution Usage Information screen from the Stain Process Menu and check for
expired solutions and reset their usage information.
You can also move to the Read Bar Code screen from the Reset Solution Usage Information screen.
2-39
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
Outputting Process Report
Each report showing bar code information can be output
from the Export screen.
2-40
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
Process Report
You can output this report by touching Process Report on the Export screen. The report shows, among
the bar code data of the currently used solutions and kits, only the lot numbers and the expiration dates of
unopened and opened solutions/kits.
The Exp. (Use) field of the report shows the practical expiration date (unopened or opened, whichever
comes first). Also, the date the bar code was read is shown as the Date Opened. The report shows the
expiration dates of individual kit solutions separately. The solution usage status at the end of processing is
shown. In addition to the bar code of each station, the bar code of each kit is also shown.
The Usage field shows the solution usage status. In the case of a group, the stations actually used, and
the usage status of each station used, are shown.
[Example]
---------------------------Process Report --------------------------
Exported Date: 2017-03-08 1:07:06 PM
Start:2017-03-08 11:27:34
End :2017-03-08 12:06:33
2-41
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
Solution Configuration Report
You can output this report by touching Solution Configuration List on the Export Data screen.
The solution configuration report also shows the staining kit name.
[Example]
--------------------- Solution Configuration List -------------------
Exported Date: 2017-03-08 13:06:46
Created Date : 2017-01-27 10:40:34
Revised Date : 2017-03-08 10:37:23
Configuration
Name:HE_kit Kit
Name:H&Ekit250
Station Solution Set Management
1 Xylene -- --
2 Xylene 10 Runs
3 Xylene 10 Runs
4 HCl Alcoholic -- --
5 /Hematoxylin -- --
6 (W2) Wash Station -- --
7 (W1) Wash Station -- --
8 (D1) Drying Station -- --
9 Alcohol 95% -- -
10 Alcohol 100% -- --
11 Alcohol 100% -- --
12 Distilled Water -- --
13 /Eosin -- --
14 Distilled Water -- --
15 Alcohol 95% -- --
16(D2) Drying Station -- --
17 Xylene -- --
18 Xylene -- --
19 Xylene -- --
20 Alcohol 100% -- --
21 Alcohol 100% -- --
22 Alcohol 100% -- --
23 Alcohol 95% -- --
24(H1) -- -- --
25(H2) -- -- --
26(E1-1) End Station 1 -- --
27(E1-2) End Station 1 -- --
28(E1-3) End Station 1 -- --
29(S3) Start Station -- --
30(S2) Start Station -- --
31(S1) Start Station -- --
------------------------ Color Information --------------------------
G1 Green
G2 Cream
G3 Yellow
G4 Pink
G5 Magenta
------------------------- Station Parameters ------------------------
Mix Amplitude 30 mm
Mix Frequency 5
Mix Speed 3
Enhanced Wash Frequency 5
Enhanced Wash Speed 3
Basket Lifting Speed 1
Water Wash Cycle Time 60 seconds
Drying Station Temperature 65 C.
Heating Station(1) Temperature 60 C.
Heating Station(2) Temperature 60 C.
Heat up Heating Stations Yes
2-42
Downloaded from www.Manualslib.com manuals search engine
How to Use the Bar Code Reader, and It's Function (Options)
Barcode History Report
You can output this report by touching Barcode History on the Export Data screen.
The barcode history shows the GTIN of each barcode that has been read, lot number, and the date the
barcode was read, for up to 1,000 most recently read barcodes.
[Example]
Barcode History report
--------------------------- Barcode History --------------------------
Exported Date: 2015-06-27 16:47:23
2-43
2-44
2-45
1. Once the power is restored, an error buzzer sounds and the Power Outage Alarm screen or Arm Error
window appears on the control panel display to inform the operator that a power outage has occurred.
2. The operator must evacuate the arm to the initial position according to the onscreen instruction. If any
basket is currently set on the arm, remove the basket from the arm according to the onscreen
instruction and then place it in the station shown in brown.
3. To resume stain process, touch the Resume key. Because of the power outage, the staining may not
end by the initially scheduled end time. In this case, the specified time may be delayed by up to the
period after the power outage occurred until the Resume key was touched.
4. To abort stain process, touch the Abort key. Refer to P. 2-11 for the method to abort staining.
A prolonged power outage may cause the tissues to dry up or result in staining problems. If you
know that a power outage will occur, do not use the instrument. If a prolonged power outage has
occurred due to an unexpected problem, take an appropriate action to prevent the tissues from
drying up, and then manually perform the remainder of the staining process.
2-46
Other Errors
If the error does not accompany a number, follow the onscreen instruction shown in the error window.
In the event of an unexpected problem for which you cannot find any solution,
immediately turn off the power. Next, open the cover, remove the tissues, and protect
them so as not to let them dry up. Then, contact the Sakura Finetek Technical Support
representative or local representative.
2-47
3-1
Basket adapter
metal part
3-2
In addition to the interior of the instrument, also clean the cover, door and around the exhaust port
regularly. At this time, be sure to turn off the power and disconnect the power plug from the power
outlet. Failure to do so may result in electric shock.
<Corrosion Protection>
The Tissue-Tek Prisma uses corrosion-resistant stainless steel, but the stainless steel may rust depending
on a combination of factors such as the environment (water, air), staining method (type of solution,
concentration), use method (handling of solvent, maintenance), etc. Take note of the following points to
protect the instrument from corrosion.
Keep the solution reservoir lids closed except during the staining process. Remove those solution
reservoirs containing highly corrosive solutions, from the trays, and keep the solutions outside the
instrument. If a highly corrosive solution spilled in the instrument, wipe it off immediately.
Remove all solution trays (including the solution reservoirs) from the instrument once a week or so, wipe
down the trays using a cloth moistened with water, clean them with neutral detergent, and wipe down with
a dry cloth.
Corrosive solutions
Acids [hydrochloric acid (HCL), hydrobromic acid (Hbr), hydrofluoric acid (HF), etc.] and staining
solutions containing an acid
Reagent solutions [Schiff’s reagent, cold Schiff, sulfurous acid solution, resorcin‐fuchsin staining
solution, etc.]
Oxidizing chlorides [ferric chloride (FeCl3), sodium tiosulfate (Na22O3), organic acid (HCOOH), acetic
acid (CH3COOH), bromic acid (C204H2), lactic acid (CH3CH(OH)COOH), etc.)]
If this is not performed, the metal parts may corrode (rust, etc.) quickly. Contact the Sakura Finetek
Technical Support representative or local representative for use of metal cleaners.
3-3
If you have any question regarding cleaning, contact the Sakura Finetek Technical Support representative
or local representative.
3-4
* In the event of an unexpected problem for which you cannot find any solution, immediately
turn off the power. Then, contact the Sakura Finetek Technical Support representative or
local representative.
Error
number Name Cause Action
1 System Error The staining program does not end 30 seconds after the Contact the Sakura Finetek
scheduled time. Technical Support representative
or local representative.
2 Power Outage A power outage has occurred due to abnormal power Contact the Sakura Finetek
supply from the facility. Technical Support representative
or local representative.
The power switch was turned off during the staining Turn on the power switch.
process.
The power cord came off from the power outlet during the Connect the power cord.
staining process.
A power outage signal was sent to the instrument from the This is normal when the facility
UPS. power was cut off. Wait for the
power to be restored.
3 Power Outage Faulty memory on the CPU board of the control panel. Contact the Sakura Finetek
Recovery Failure The voltage of the battery unit installed in the memory of Technical Support representative
the CPU board of the control panel dropped. or local representative.
10 Cover Open The operator opened the cover during staining. Contact the Sakura Finetek
The cover deformed and created a gap between the Technical Support representative
detection sensor and the detection plate during the staining or local representative.
process.
The instrument malfunctioned during the staining process
due to a cover open/close sensor error.
The instrument malfunctioned during the staining process
due to a controller board input error.
The control panel connection harness connecting the CPU
board of the control panel and the controller board was
disconnected during the staining process.
The instrument malfunctioned during the staining process
due to an input error of the CPU board of the control panel.
The instrument malfunctioned during the staining process
due to a CPU error of the CPU board of the control panel.
11 Cover Open (when The operator opened the cover while the arm was Close the cover and resume the
not staining) operating other than during the staining process. process.
The cover deformed and created a gap between the Contact the Sakura Finetek
detection sensor and the detection plate while the arm was Technical Support representative
operating other than during the staining process. or local representative.
The instrument malfunctioned due to a cover open/close
sensor error while the arm was operating other than during
the staining process.
The instrument malfunctioned due to a controller board
input error while the arm was operating other than during
the staining process.
The control panel connection harness connecting the CPU
board of the control panel and the controller board shorted
while the arm was operating other than during the staining
process
The instrument malfunctioned due to an input error of the
CPU board of the control panel while the arm was
operating other than during the staining process.
The instrument malfunctioned due to a CPU error of the
CPU board of the control panel while the arm was
operating other than during the staining process.
20 Processing Halt The operator paused the instrument during the staining Resume the process.
process.
4-1
4-2
Error
number Name Cause Action
214 Deviation Error Movement of the Y-axis is inhibited by an obstacle or faulty Contact the Sakura Finetek
(Y axis) wire drive. Technical Support representative
The Y-axis home sensor is faulty. or local representative.
The Y-axis encoder is faulty.
The Y-axis motor is faulty.
The controller board is faulty.
215 Wire Cut Off (Y axis) The Y-axis wire broke or came off from the pulley, etc. Contact the Sakura Finetek
The Y-axis wire breakage sensor is faulty. Technical Support representative
The controller board is faulty. or local representative.
311 Home Position Sensor The Z-axis home sensor is faulty. Contact the Sakura Finetek
Error The controller board is faulty. Technical Support representative
(Z axis) or local representative.
312 Wire slack (Z axis) The Z-axis wire assembly is loose. Contact the Sakura Finetek
Technical Support representative
or local representative.
314 Deviation Error Movement of the Z-axis is inhibited by an obstacle or Contact the Sakura Finetek
(Z axis) faulty wire drive. Technical Support representative
The Z-axis home sensor is faulty. or local representative.
The Z-axis encoder is faulty.
The Z-axis motor is faulty.
The controller board is faulty.
315 Wire Cut Off The Z-axis wire broke or came off from the pulley, etc. Contact the Sakura Finetek
(Z axis) The Z-axis wire breakage sensor is faulty. Technical Support representative
The controller board is faulty. or local representative.
430 Sensor Opened An abnormal value was input to the signal amplification Contact the Sakura Finetek
(Drying Station 1) part of the controller board due to a disconnected Technical Support representative
temperature sensor. or local representative.
The drying station connection harness connecting the
temperature controller board and the controller board was
disconnected.
A disconnected harness was recognized by CPU1 due to
a signal amplification error of the controller board.
Normal judgment was disabled due to a CPU1 error of the
controller board.
431 Sensor Shorted An abnormal value was input to the signal amplification Contact the Sakura Finetek
(Drying Station 1) part of the controller board due to a shorted temperature Technical Support representative
sensor. or local representative.
The drying station connection harness connecting the
temperature controller board and the controller board
shorted.
A shorted harness was recognized by CPU1 due to a
signal amplification error of the controller board.
Normal judgment was disabled due to a CPU1 error of the
controller board.
432 Temperature Low Power could not be supplied to the heater due to a Contact the Sakura Finetek
(Drying Station 1) disconnection or connection failure of the heater. Technical Support representative
Power could not be supplied to the heater due to a SSR or local representative.
error of the temperature controller board.
Power could not be supplied to the drying heater due to a
disconnection of the drying station connection harness
connecting the temperature controller board and the
controller board.
Power could not be supplied to the heater due to a CPU1
error of the controller board.
The heater voltage selector switch is set incorrectly.
The power-supply voltage selector switch is set incorrectly.
433 Thermal Fuse (Drying The SSR3 on the temperature controller board Contact the Sakura Finetek
Station 1) experienced a short-circuit failure. Technical Support representative
The heater remained on due to a driver error of the or local representative.
controller board.
A false detection occurred due to an error of the SSR1 for
power-supply voltage detection.
The drying station connection harness connecting the
temperature controller board and the controller board was
disconnected.
The instrument malfunctioned due to an over-temperature
protection input error of the controller board.
The heater voltage selector switch is set incorrectly.
The power-supply voltage selector switch is set incorrectly.
4-3
4-4
4-5
4-6
* In the event of an unexpected problem for which you cannot find any solution, immediately turn
off the power. Then, contact the Sakura Finetek Technical Support representative or local
representative.
4-7
4-8
The saved solution names The control file storing the solution name information was lost. Contact the Sakura Finetek Technical Support
were initialized. The control file storing the solution name information is corrupted. representative or local representative.
The saved solution The control file storing the solution configuration information was Contact the Sakura Finetek Technical Support
configuration was lost. representative or local representative.
initialized. The control file storing the solution configuration information is
corrupted.
The saved staining The control file storing the staining program information was lost. Contact the Sakura Finetek Technical Support
program was initialized. The control file storing the staining program information is representative or local representative.
corrupted.
Sakura solutions or kit The definition file (solution, kit data base) provided by the dealer Contact the Sakura Finetek Technical Support
solutions are not was lost. representative or local representative.
supported. The definition file (solution, kit data base) provided by the dealer is
corrupted.
The arm cannot be moved Foreign matters are attached to the slide shaft of each axis. Contact the Sakura Finetek Technical Support
manually by hand when the The bearing is too tight to move the slide shaft of each axis due to representative or local representative.
power is off. rust, scratches or deformation.
Each axis does not move due to an obstacle present in its moving Remove the obstacle.
direction.
Each axis does not move in the specified direction due to an Contact the Sakura Finetek Technical Support
abnormality of the axis drive wire and its transmission mechanism. representative or local representative.
A basket cannot be placed The solution reservoir is not installed in the correct position. Install the solution reservoir in the correct
in the correct position position.
during operation. The basket adapter has deformed. Contact the Sakura Finetek Technical Support
representative or local representative.
The tray is not installed in the correct position. Install the tray in the correct position.
The home detection mechanism of each axis has not been adjusted Contact the Sakura Finetek Technical Support
correctly. representative or local representative.
A solution reservoir different from the one set in the solution Install a solution reservoir matching the
configuration was installed. solution configuration.
No solution reservoir is installed. Install a solution reservoir matching the
solution configuration.
The staining process was aborted or ended, but the basket has not Remove the basket.
been removed.
The lid of the solution reservoir remains on. Remove the lid of the solution reservoir.
The paraffin tray in the drying station is facing the wrong direction. Install it in the correct direction.
4-9
The instrument power is Contaminants entered the water supply solenoid valve seat. Contact the Sakura Finetek Technical Support
off, but water is leaking The water supply solenoid valve is damaged. representative or local representative.
from the water supply
nozzle.
When the facility was The flow rate adjustment valve is clogged. Use the flow rate adjustment valve to adjust
operating under the the amount of water supplied.
conditions within the use The strainer is clogged. Clean the strainer.
environment of the The O-ring of the water supply nozzle is damaged. Contact the Sakura Finetek Technical Support
instrument, the water level The wash reservoir is damaged. representative or local representative.
reached by supplying The wash reservoir is not installed properly.
water to the instrument is
lower than before.
When the facility was The open angle of the flow rate adjustment valve was changed. Use the flow rate adjustment valve to adjust
operating under the the amount of water supplied.
conditions within the use
environment of the The pressure relief valve is damaged. Contact the Sakura Finetek Technical Support
instrument, the water level representative or local representative.
reached by supplying The water supply capacity on the facility side has changed. Use the flow rate adjustment valve to adjust
water to the instrument is the amount of water supplied.
more than before.
Water is leaking from the A component of the water supply part is damaged. Contact the Sakura Finetek Technical Support
instrument due to a The component connection point at the water supply part is leaking representative or local representative.
problem with water supply water.
or drain. The connection of the drain hose is loose or clogged.
The sink is leaking water.
The solution odor is No fume cartridge is set. Install a fume cartridge.
leaked from the instrument The fume cartridge was set without unpacking it. Unpack the fume cartridge.
when the instrument is The fume cartridge is not installed properly. Install the fume cartridge correctly.
operating. The fume cartridge has reached its life. Replace the fume cartridge.
The exhaust fan stopped. Contact the Sakura Finetek Technical Support
representative or local representative.
The air inlet port in the system was blocked. Check the air inlet port for foreign matters.
Each gasket of the fume control unit deteriorated. Contact the Sakura Finetek Technical Support
representative or local representative.
The exhaust outlet port is blocked. Do not block the exhaust outlet port.
When connecting the exhaust duct, it is not connected correctly. Contact the Sakura Finetek Technical Support
The exhaust duct is too long. representative or local representative.
The exhaust duct was crushed.
The exhaust system on the facility side is not operating. Check the exhaust system on the facility side.
Solution odor leaks from The door is open. Close the door.
the instrument when the The cover is open Close the cover.
operation is stopped with There is no lid on the solution reservoir. Put the lid on the solution reservoir.
solution still in the
instrument.
The cover does not open The cover was pushed open at an angle. Contact the Sakura Finetek Technical Support
smoothly. representative or local representative.
The cover (top plate) is bitten. Check for foreign matters and remove them, if
found.
The rotary damper and rack are bitten or damaged. Check for foreign matters and remove them, if
found.
The roller pack is bitten or damaged. Contact the Sakura Finetek Technical Support
The spring stay is damaged. representative or local representative.
The cover does not close The cover closes at an angle or does not close fully. Contact the Sakura Finetek Technical Support
smoothly or fully. The cover is distorted. representative or local representative.
The cover (top plate) is bitten. Check for foreign matters and remove them, if
found.
The rotary damper and rack are bitten or damaged. Contact the Sakura Finetek Technical Support
The roller pack is bitten or damaged. representative or local representative.
The spring stay is damaged.
The cover stay adjustment is inappropriate.
4-10
The instrument was stopped for a long period of time Wipe off any water, etc., attached to the
with water still attached to the sink or tray. instrument.
Noises generated by Noise from the gear meshing part at the bottom of the Z- Contact the Sakura Finetek Technical
each drive axis axis motor Support representative or local
became louder than Noise from the gear meshing part at the bottom of the X- representative.
before. axis motor
Noise from the gear meshing part at the bottom of the Y-
axis motor
The solution reservoir, Dropped. Exercise caution when using the
basket or basket instrument.
adapter has An attempt was made to install the instrument under Avoid using the instrument under
deteriorated or unreasonable conditions. unreasonable conditions.
damaged. Effect of solution due to the staining process. Contact the Sakura Finetek Technical
Support representative or local
representative.
Smoke came from the The terminal block of the primary power line became Contact the Sakura Finetek Technical
instrument. loose and the terminals were overheated and smoke Support representative or local
generated from electric wire. representative.
Smoke generated due to an abnormal drying station.
Smoke generated due to an abnormal heating station.
The insulation of the power cable was lost and the leak
current continued to flow, causing abnormal heat
generation.
Solution or water was Human error Contact the Sakura Finetek Technical
spilled inside the Support representative or local
instrument. representative.
The bar code cannot The bar code was read more than 10 seconds after Read the bar code within 10 seconds
be read. touching the Read button. after touching the Read button.
A bar code reader other than the specified product was Use the specified bar code reader.
used.
The bar code reader is faulty. Contact the Sakura Finetek Technical
Power is not supplied to the bar code reader. Support representative or local
The circuit on the instrument side is faulty. representative.
The bar code is in poor state (not printed clearly, dirty,
scratched, etc.)
The bar code pertains to an expired product. Be sure to use an unexpired product.
The product code DB file provided by the dealer is Contact the Sakura Finetek Technical
damaged. Support representative or local
The bar code reader is not set correctly. representative.
4-11