The document describes specifications for the THC8000, a continuous analyzer that measures total hydrocarbon levels in parts per million. It has a resolution of 0.1-1 ppm, accuracy of 0.5% of full scale, and drift of less than 1% over 24 hours. It uses flame ionization detection and has outputs for alarms and calibration.

Uploaded by

kanaf11943
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
122 views2 pages

THC8000

The document describes specifications for the THC8000, a continuous analyzer that measures total hydrocarbon levels in parts per million. It has a resolution of 0.1-1 ppm, accuracy of 0.5% of full scale, and drift of less than 1% over 24 hours. It uses flame ionization detection and has outputs for alarms and calibration.

Uploaded by

kanaf11943
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

SPECIFICATIONS

www.orthodyne.be
Three ranges

THC8000 series
0 -10 ppm Display resolution 0.1 ppm
0 -100 ppm Display resolution 0.1 ppm
0 -1000 pmm Display resolution 1 ppm
Other range possible on request from 1 to 10000 ppm
Accuracy 0.5 % at full scale
Detection limit
Drift
< 0.05 ppm for the range 0-1 ppm
1% over 24 hours
Continuous THC analyser
Sample flow
Max. input pressure
Around 3 l/h
Working pressure from 0.3 to 15 Barg
Analysis of THC in PPM level
Sampling gas N2, Ar, He, Air, H2 , O2 or CO2
Sample gas connection 1/8’’ Swagelok SS
Sample flow rate Approximately 3 l/h
Sample pressure from 0.3 to 15 Barg
Combustive gas Synthetic air
Combustive gas connection 1/8’’ Swagelok SS
Combustive gas pressure 2 Barg
Combustive gas flow rate 20 l/h
Recommended quality 5.0
Fuel gas Hydrogen (other mixtures available on request)
Fuel gas connection 1/8’’ Swagelok SS
Fuel gas pressure 1 Barg
Fuel gas flow rate 2 l/h
Recommended quality 6.0
Operating temperature Around 20°C without wide temperature variations
Power supply 220 Vac, 50-60 Hz or 120 Vac, 60 Hz on request
Power consumption 500 VA
4-20 mA output 1 output for THC
RJ-45 connection Computerised system maintenance
Output relays (dry contacts, SPST 2A / 250 VAC)
1 STATUS contact relay (security alarm)
3 RANGE contact relays (active range)
1 process alarm contact relay (level 1)
1 process alarm contact relay (level 2)
2 contacts for automatic calibration
1 flow alarm contact relay
1 calibration in use contact relay

Standard rack mount 4U


Dimensions

Height > 178 mm | Depth > 450 mm | Width > 483 mm


4U 178 mm

102 mm

16132075 © www.ultrapetita.com • 0495 807 112

19.0“ (483 mm)

450 mm 448 mm

ORTHODYNE S.A. Phone : +32-4-263 90 90


Rue des Technologies, 23 Fax : +32-4-263 09 79
B-4432 ALLEUR - BELGIUM E-Mail : [email protected]
THC8000
FEATURES
• < 50 ppb resolution guaranteed on the 0-1 ppm range, if you request this type of range.
• E lectronic flame-out guard circuit.
•A  utomatic fuel shut off system.
• E asy to use.
Continuous THC analyser • L ow sample flow.
• L ower power consumption.
•A  lphanumeric screen 64 x 256 pixels.
The THC8000 is a continuous analytical system that
•A  djustable alarm and oven settings.
measures the total hydrocarbon (THC) in ppm level in
• F ast response.
various gases, such as Oxygen, Air, Carbon Dioxide,
•M  ultifunctional and real-time software.
Nitrogen, Argon, Helium and Hydrogen.
•C  E marked.

PRINCIPLE

The THC8000 module is composed of a flame ionization


detector placed in a temperature regulated chamber and
coupled with a continuous analysing technology.

Burning of the components containing hydrocarbons by the


hydrogen flame generates an ion current proportional to the
impurities introduced into the gas to be analysed.

This analyser is a standard 19 inch rack module, 4U in


height and 450 mm in depth.

One analog 4-20 mA output provides the total FID DETECTOR EXPLANATION
concentration of hydrocarbons impurities.
The above diagram shows the general
construction of a FID.
Organic compounds from the separation
column are injected into the detector housing Zero gas
where they are mixed with Hydrogen and Zero & Span GAS Span gas
Type of configuration Synthetic Air before entering the detector Calibration
Zero/Span Calib.
Automatic Calib.
Auto Calib. Sequ.*
Factors Time Purge *
nozzle where the mixture is burned.
•T
 HC 8000 : Only one Zero, Span, Measure
Calibartion Param
configuration for the measurement During this process, organic compounds alarm Levels

System overview
of THC in Oxygen, Air, Carbon are broken down into carbon fragments Data Range Alarm Level 1
Detector Alarm Level 2
Dioxide, Nitrogen, Argon, Helium and acquire a positive charge (i.e., become Save Select Range
and Hydrogen. ionized) at the surface of the anode. Analog Output Range
Range
Mode
Carbon fragments are detected by the Diagnosis
Analyser Config
collector. Settings Mass Flow Config Security alarms
4-20 mA range
Set points
The signal is then amplified and sent to the Software version
data processing system.
Applications

Software date
Analog Input
• Air separation plants A > Sample inlet Analog Output Test analog input
Maintenance
• Gas purity certification B >M  ixture between the sample
Relays
Remote access Test analog output
• Specialty gas laboratories and the Hydrogen
• Process control C > Hydrogen inlet
Test relays
• Steel industry D > Synthetic Air inlet * Optional
E > Nozzle
F > Flame tip
G > Collector
H > Anode & Ignitor THC 8000 series - Continuous THC analyser
I > Flame guard