Quick Guide FDX Console

This document provides instructions for stitching multiple x-ray images together on an FDX console. It describes registering the patient, acquiring images, automatic and manual stitching processes, and additional details.

Uploaded by

Wajahat Sheikh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
339 views5 pages

Quick Guide FDX Console

This document provides instructions for stitching multiple x-ray images together on an FDX console. It describes registering the patient, acquiring images, automatic and manual stitching processes, and additional details.

Uploaded by

Wajahat Sheikh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

DOC-0038775-A

Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 1 of 5

FDX Console Image Stitching Workflow

Registration of patient
 Select the correct patient from the RIS tab or manually register
the patient.

 Select or verify that the correct stitching exposure menu from


the menu display tabs.

 Image combining or stitching exposure menus will have a WS


designation.

Acquire the first image


 After the first image is acquired two additional exposure menus
will populate.

 Expose the next one or two exposure menus as required to


include the required anatomy. A maximum of 3 exposures can
be stitched. O

 If only two exposures are required, do not delete the third


empty exposure menu. Leave it in place and continue to the
console stitching process/Image combining icon.

Console stitching process


 Once all of the required exposures are performed.
 Select the image combining icon.
 A warning message will populate prior to the system
attempting automatic stitching. If the order is not correct
select cancel and click and hold the menu and move it to
the correct position.

 The system will automatically link the images or require manual stitching
to be performed. If the images are automatically linked evaluate the stitched image
for image quality and perform the standard QA functions as required and complete the
study.
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 2 of 5

 The system will determine if automatic linking is possible. If automatic


linking fails a warning message will populate.

 If manual stitching is required, select the QA home icon.

 Then activate the manual stitching icon.

Manual Stitching workflow


 After selecting the manual stitching icon. The technologist will have a choice to select either
S value adjustment or 2 choices image movement management.
The methods of image movement are:
1. Fine adjustment
2. 2 point adjustment

 The S value adjustment icon will allow the technologist to adjust the brightness of each
image individually. Click on each image area and adjust the S value individually. Click on
the + and - radial buttons to adjust the S value or hover the mouse into the image area and
press and hold the left mouse button and move the mouse to adjust the S value.

S# adjustment icon

Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 3 of 5

 Fine adjustment selection will allow you to move each image independently to achieve
proper alignment of the stitching markers or the overlapped anatomy.

 The technologist can select the 100% zoom or Pixel size display to zoom in to the images.
Adjust the alignment markers or anatomy to correct position.

Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 4 of 5

 An alternative method to fine adjustment is to select the two point adjustment.

Select the point on one image and then select the point on the second image, the
Images will be moved to those two points.

 The cancel button, will undo either the fine adjustment or the two point
adjustment action or S# adjustment. You will exit the manual stitching
function and have to reselect QA and the manual stitching icon.

 If manual stitching has been performed and satisfactory alignment has been achieved
select the “OK” button. If you choose the initialize button, the stitched images will returned
to their misaligned positions and fine adjustment will have to be performed again.

 The stitched image will be presented for QA. Shadow masking and other QA functions can
be applied as needed. Then press the study completion icon to send the images to the
desired output location.
 If the images have been stitched incorrectly or fine adjustment must
be repeated. Select the unstitch icon on the display to repeat the process.

Additional information
 Process to follow if a third image has been acquired accidently and is blank or
unnecessarily exposed.

1. If a third image or blank image needs to be removed prior to stitching. The technologist
must add any regular (RT) exposure menu to the exposure list.

2. The technologist should then press and hold the image thumbnail until an aqua blue
box is seen around the thumbnail. Let go of the mouse button, and move the mouse
cursor to the newly added (RT) exposure menu’s blank thumbnail and press the left
mouse button to move the image to this new menu.

3. Then use the standard rejection workflow to reject that image.

4. Select the image combining icon and follow the instructions above for image combining.

Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 5 of 5

 The system is designed to adjust automatic brightness for the stitched images. If individual
image S value adjustment is desired, this function must be setup during installation. The
S/L adjustments can also be made to the entire stitched image in the QA function.

 The motion detection warning can be turned on at the time of installation and setup. The
motion detection software can identify up to 10 mm of vertical movement and 40 mm of
horizontal movement can be identified. If the movement is greater than these
measurements, the warning message will not populate. Automatic adjustment for motion
can only occur if the movement is less than 10 mm in size.

The maximum movement will identified.

 If an attempt is made to close the study prior to stitching the image, a warning message will
appear. The images must be stitched prior to closing the study. Click “OK” and stitch the
images then suspend the study or close the study.

 For more information, refer to FDX Console Options Reference Guide and locate the
appropriate section for FCR, FDR D-EVO image combining section for further detail.