Quick Guide FDX Console
Quick Guide FDX Console
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 1 of 5
Registration of patient
Select the correct patient from the RIS tab or manually register
the patient.
The system will automatically link the images or require manual stitching
to be performed. If the images are automatically linked evaluate the stitched image
for image quality and perform the standard QA functions as required and complete the
study.
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 2 of 5
The S value adjustment icon will allow the technologist to adjust the brightness of each
image individually. Click on each image area and adjust the S value individually. Click on
the + and - radial buttons to adjust the S value or hover the mouse into the image area and
press and hold the left mouse button and move the mouse to adjust the S value.
S# adjustment icon
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 3 of 5
Fine adjustment selection will allow you to move each image independently to achieve
proper alignment of the stitching markers or the overlapped anatomy.
The technologist can select the 100% zoom or Pixel size display to zoom in to the images.
Adjust the alignment markers or anatomy to correct position.
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 4 of 5
Select the point on one image and then select the point on the second image, the
Images will be moved to those two points.
The cancel button, will undo either the fine adjustment or the two point
adjustment action or S# adjustment. You will exit the manual stitching
function and have to reselect QA and the manual stitching icon.
If manual stitching has been performed and satisfactory alignment has been achieved
select the “OK” button. If you choose the initialize button, the stitched images will returned
to their misaligned positions and fine adjustment will have to be performed again.
The stitched image will be presented for QA. Shadow masking and other QA functions can
be applied as needed. Then press the study completion icon to send the images to the
desired output location.
If the images have been stitched incorrectly or fine adjustment must
be repeated. Select the unstitch icon on the display to repeat the process.
Additional information
Process to follow if a third image has been acquired accidently and is blank or
unnecessarily exposed.
1. If a third image or blank image needs to be removed prior to stitching. The technologist
must add any regular (RT) exposure menu to the exposure list.
2. The technologist should then press and hold the image thumbnail until an aqua blue
box is seen around the thumbnail. Let go of the mouse button, and move the mouse
cursor to the newly added (RT) exposure menu’s blank thumbnail and press the left
mouse button to move the image to this new menu.
4. Select the image combining icon and follow the instructions above for image combining.
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713
DOC-0038775-A
Quick Guide on
Effective Date: 6/8/2018
FDX Console Stitching Page 5 of 5
The system is designed to adjust automatic brightness for the stitched images. If individual
image S value adjustment is desired, this function must be setup during installation. The
S/L adjustments can also be made to the entire stitched image in the QA function.
The motion detection warning can be turned on at the time of installation and setup. The
motion detection software can identify up to 10 mm of vertical movement and 40 mm of
horizontal movement can be identified. If the movement is greater than these
measurements, the warning message will not populate. Automatic adjustment for motion
can only occur if the movement is less than 10 mm in size.
If an attempt is made to close the study prior to stitching the image, a warning message will
appear. The images must be stitched prior to closing the study. Click “OK” and stitch the
images then suspend the study or close the study.
For more information, refer to FDX Console Options Reference Guide and locate the
appropriate section for FCR, FDR D-EVO image combining section for further detail.
Confidential and Proprietary Information -- Use, Duplication FUJIFILM Medical Systems U.S.A., Inc.
or Distribution of this document without permission from 419 West Avenue • Stamford, CT 06902
FUJIFILM Medical Systems U.S.A., Inc., is strictly prohibited. (800) 431-1850 • Fax (203) 351-4713